1. General view Refer to relevant figure on the fold out part of the cover 1 Rear panel 4 Powder drum 2 Front panel Display 8 ProphyPen handpiece 3 Sensitive control panel 8 Compressed air supply filter Before use the compressed air first passes through a filter ref 8 with a purge button ref 13 A rapid disconnection system with an automatic valve enables it to be connected to the mains 9 Foot control connector Connection for the foot control pedal 10 On off button O 11 Air pressure adjustment Air pressure is preset in the factory For any adjustment of the pressure contact an approved installer of dental equipment 12 Drinking water supply connector Its rapid disconnection system with an automatic valve enables it to be connected to the supply A removable filter is located on the water inlet pipe 1 Introduction 13 Purge of compressed air filter Press this purge button to eliminate air remaining in the filter 14 Power socket Connection to mains electricity is by a standard connector conforming to IEC 60 320 15 Fuse housing Contains the device s fuses Recommendations For your safety connections water air and electricity should be carried out by an approved installer of dental equipment The device requires the following connections see technical specifications chapter Carry out the connection to the drinking water supply using an automatic cutoff valve o
2. Air Max User s manual Notice d utilisation e Manual de uso Benutzerhandbuch Manuale d uso Handleiding voor de gebruiker Illustrations 1 Introduction 2 Installation 3 Starting and stopping the device sscesseeseeee 4 Using polishing mode 5 Maintenance 6 Troubleshooting 7 Technical specifications The Air Max is an air polisher It enables treatment to be carried out in optimal aseptic conditions Air Max is equipped with a ProphyPen handpiece General characteristics Front panel Display e Polishing function via the powder drum e Irrigation via the water supply with flow regulation on the handpiece connector e Foot control via a pedal e Integrated heating system for the spray for optimal patient comfort The Air Max is a prophylactic device for polishing after scaling or in periodontal and implant maintenance Parts included 6 On off button O I Lights up when the on off button ref 10 is in position Sensitive control panel 7 Purge key gt A short press purges the handpiece powder circuit for a predetermined length of time of a few seconds Rear panel One Air Max device One mains cord One single pedal foot control ProphyPen handpiece with titanium nozzle Powder Max introductory powder kit e Pack three plastic nozzles
3. 3 Replace the fuses ref 27 by fuses of the same type and value 4 Push the fuse holder back into its housing Fault Possible cause Solution Does not operate general Unplugged or defective mains cord Check and or replace the mains cord Fuses blown Replace the fuses Defective wall power outlet Contact your electrician Device switched off 0 Set the switch to the on position I Foot control disconnected and or defective Check and or replace the foot control No powder air or water No air or water coming from the pipes connected to the mains supply Contact your approved installer Handpiece and or nozzle blocked Disconnect the handpiece and check that air and powder are coming out of the cord _ if not retum the device to Satelec After Sales Service _ if they are unscrew the handpiece nozzle unblock the nozzle with a file and place the handpiece in an ultrasound tank for 10 minutes No powder or water but air present Powder drum empty or too full Fill the drum or remove the excess Water water inlet not connected or incorrect water setting Check the water inlet pipe or adjust the water flow on the handpiece cord Powder damp Empty and clean the drum and refill with new powder Handpiece and or nozzle blocked Disconnect the handpiece and check if water comes out of the cord If it doesn t return the de
4. tum the screw in the opposite direction Replacing the fuses It is imperative that the powder circuit is purged before any cleaning takes place The powder drum Carefully empty the drum of powder if the device is not to be used for several hours overnight for example Residual dampness of the air alters the powder s properties To clean the powder drum unscrew the reservoir cap and remove the residue using the dental chair s suction system The ProphyPen handpiece After each use unscrew the ProphyPen handpiece nozzle ref 20 and clean in an ultrasound tank for 10 minutes fragments the powder dry well and sterilize the nozzle and ProphyPen handpiece in an autoclave Cleaning the filters The water filter Cleaning the water filter must be done in relation to the hardness index of the water Proceed as follows 1 Shut off the water at the tap on the water supply side and purge the circuit 2 Set the on off button to the O position and unplug the mains lead 3 Unscrew the two parts of the filter ref 21 and 23 using two 10mm open ended spanners Extract and replace the cartridge ref 22 Reassemble by reversing the operation retightening the two parts of the filter ref 21 and 23 uh Proceed as follows 1 Set the on off button to the O ref 24 Position and unplug the mains lead ref 25 2 Insert the end of a screwdriver ref 26 in the slot to open the protective cover
5. Y SPAIN Z I du Phare B P 216 F 33708 MERIGNAC Cedex Tel 33 0 556 34 06 07 Fax 33 0 556 34 92 92 e mail satelec acteongroup com 130 Gaither Drive Suite 100 MOUNT LAUREL NJ 08034 Tel 1 856 222 9988 Fax 1 856 222 4726 e mail info us acteongroup com Industriestrasse 9 D 40822 METTMANN Tel 49 0 21 04 95 65 10 Fax 49 0 21 04 95 65 11 e mail info de acteongroup com Av Principal 11 H Poligono Indus Can Clapers E 08181 SENTMENAT BARCELONA Tel 34 93 715 33 66 Fax 34 93 715 32 29 e mail info es acteongroup com Unit 1B Steel Close Little End Road GB EATON SOCON PE19 8TT Tel 44 0 1 480 477 307 Fax 44 0 1 480 477 381 e mail info uk acteongroup com C ACTEON group MIDDLE EAST Numan Center 1st Floor N 111 Gardens Street CHINA PHILIPPINES KOREA INDIA PO Box 468 AMMAN 11953 JORDAN Tel 962 6 553 4401 Fax 962 6 553 7833 e mail info me acteongroup com Office 401 12 Xinyuanxili Zhong Street Chaoyang District BEIJING 100027 Tel 86 10 646 570 11 2 3 Fax 86 10 646 580 15 e mail info cn acteongroup com 3F King s Court Il Bldg Pasong Tamo Street Makati City 1200 MANILA Tel 632 811 29 14 18 Fax 632 811 24 88 e mail info ph acteongroup com 8F Hanil B D 132 4 1Ga Bongrae dong Joong gu SEOUL Tel 82 2 753 41 91 Fax 82 2 753 41 93 e mail info kr acteongroup com E 91 G I D C Electron
6. e drum 2 Unscrew the cap ref 18 3 Fill the drum with Satelec PowderMax powder up to the level marker ref 19 Testing the effectiveness of the powder jet Purging the polishing circuit 1 Point the handpiece at the test card and press the pedal 2 Set the water flow on the handpiece connector If there is not enough water powder accumulates on the card If there is too much water the polishing effect is less effective If the setting is correct the powder is immediately removed by the water Where the powder jet is not powerful enough clean the nozzle see chapter Cleaning the polishing system or have the air pressure adjusted by an approved installer of dental equipment Polishing Proceed as follows 1 Use the ProphyPen polishing handpiece 2 Press the pedal to start polishing It is important to observe a jet surface angle of from 45 to 60 and to keep to a spray tooth distance of 4 to 6mm 1 2 3 Release the pedal to stop polishing Point the ProphyPen handpiece towards the cuspidor Press the gt key ref 7 to purge the circuit Cleaning and disinfecting the device The casing of the Air Max as well as the cords should be cleaned and disinfected every day using alcohol or disinfectant dental wipes Cleaning the polishing system The air filter When water is present purge by turning the black screw ref 13 clockwise To stop the purge
7. f the Aquastop type or to the dental chair water supply Caution always turn off the water supply when you leave the dental practice Setting up the equipment Position the device in the place where it is going to be used Connect the foot control ref 16 to the connector on the device ref 9 Connect the water supply pipe with its filter to the connector located at the rear of the device ref 12 Connect the air supply pipe to the air filter located at the rear of the device ref 8 Plug in the power socket ref 14 Open the supply taps water air This device should only be used by professionals dental practitioners and hygienists Before being used for the first time it is imperative to clean disinfect using alcohol or disinfectant dental wipes and or sterilize the accessories casing of the device handpiece tips keys etc recommended that the accessories be plugged in and the spray operated for a few minutes to rinse the device s internal water circuits Important It is imperative that the water supply is turned off when leaving the dental practice Switching on the power Set the power switch to the on position I ref 10 To operate Press the foot control pedal ref 17 Switching off the device Set the on off switch to the 0 position ref 10 Filling the powder drum Proceed as follows 1 Put the device into standby mode to depressurize th
8. ic Estate Sector 26 GANDHINAGAR 382044 Tel 91 79 323 80 00 Fax 91 79 642 58 51 e mail info in acteongroup com 1 11400 N9 Version 2 17 av Gustave Eiffel e BP 216 e 33708 MERIGNAC cedex France e Tel 33 0 556 34 06 07 Fax 33 0 556 34 92 92 E mail satelec acteongroup com acteongroup com
9. vice to Satelec After Sales Service If it does unscrew the handpiece nozzle unblock the nozzle with a file and place the handpiece in an ultrasound tank for 10 minutes Electrical Operational Symbols and markings Class 1 type BF device Intermittent duty 10 min 5 min Mains power supply factory set 110 to 150 V AC 50 60 Hz 230 to 260 V AC 50 60 Hz Fuse protection dimensions 5x20 mm 0 5 A 230 V or 1 A 115 V Power consumption 70 VA Insulation 4kV 1 ms Manufactured in accordance with current legislation Conforms to the general IEC 601 1 EN 60601 1 standard Retention of the memorized configuration 10 years at 25 C Mechanical Air flow rate at inlet 10 l min max Water flow rate at exit 0 to 120 ml min Rate of flow of powder 1 to 3g min Specifications of the powder white to slightly yellow powder impalpable slightly salty taste and mint or lemon flavor Powder drum capacity 50g Dimensions and use Water pressure at inlet 1 to 5 bar 14 to 72 PSI Air pressure at inlet 4 5 to 6 bar 65 to 87 PSI Overall dimensions Width 250mm Height 110mm Depth 300mm Weight 2 7kg Fixing mount on a table Operating temperature 10 to 40 C 30 to 75 RH Storage transport temperature 40 to 70 C 10 to 100 RH PNA Alternating current Type BF device A Warning please refer to the accompanying documentation Marking C FRANCE GERMAN
Download Pdf Manuals
Related Search
Related Contents
Pike SL / Race / Team Coil Action de formation : ISNTALLATION ET MAINTENANCE DES CESI W1943SB W1943TB W2043SE W2043TE W2243S - Migros Service Manual XC Series Descargar PROPOSAL: POTENTIAL SONAR SYSTEMS FOR R/V scie circulaire 1200w modele 91104.1 mode d`emploi Foglio Istruzioni.indd Sony XM SD22X User's Manual Copyright © All rights reserved.