Veeco Dektak 6M Surface Profiler
Contents
1. Y Adjustment Veeco Dektak 6M Surface Profiler POC Stephan Koev SYSTEM CHECK AND START UP 1 Turn on the monitor and start the Dektak32 application on the Desktop 2 The system will initialize itself and will let you know when it is ready in the bottom left corner of the screen LOADING UNLOADING A SAMPLE 1 Load your sample on the positioning stage 2 3 4 5 NOTE The profiler is set up such that the scan is performed from front to back on the stage left to right on the monitor therefore it is necessary to place your sample on the stage at a rotated angle of 90 degrees so that the bottom of your sample is facing the monitor Monitor Scan Press the Sample Positioning Window Icon l at the top Once your sample is positioned click on the Tower Down Icon ED located at the top of the screen The tower and stylus assembly will move down to your sample touch the surface and then the needle will rise up A video image of the sample will appear on the screen as the tower moves down Adjust the zoom and focus knobs on the sample viewing microscope for best results 6 Adjust the sample illumination intensity by pressing the Light Bulb Icon Y l in the upper left corner and moving the slider that appears or by using the Up and Down Arrows 7 To unload press the Tower Up Icon Tt and wait for the tower to completely rise Y our sample can then be removed from the stage 8 Return
2. the profiler to its standby mode by closing the Dektak32 program and turning off the monitor MAKING MEASUREMENTS 1 Position the cross hairs where you would like to start measuring e Inthe Sample Positioning Window navigate by turning the manual X Y Dials and rotating the stage to the structure you would like to measure e Use the zoom feature on the camera if needed by turning the manual Zoom Dial e To change the intensity of the video image use the Up and Down Arrows to achieve the desired image 2 Click on the Scan Routines Icon located at the top of the screen 3 Under scan parameters click on one of the options to bring up the scan parameters dialog box Scan Parameters for Routine 1 of 1 C Nominal Parameters Stylus Type IL Length Duration Sample 1000 um 10 secs S000 points Resolution Scan Type 0 333 umisample Standard Scan ka Stylus Force Meas Range Frofile Additional Parameters Hills valleys ali Hillstvalleys J gt Input the variables based on your sample and preferences Stylus type Leave at 12 5um unless installing a new tip Length scan length desired Duration Enter the duration of the scan depending on the accuracy you need The number of sampling points based on this time will be shown in the next box Resolution resolution desired Scan types Leave on standard scan Stylus force Set between 1 15 mg If measuring resist it should be betwee
3. e average step height hit F72 and it will appear on the left hand side of the screen under Analytic Results The individual results can be deleted by selecting them and hitting Delete 3 Analytical Functions When in the dataplot screen press the Calculator Icon or click on the analysis menu and select analytical functions In the dialog box that appears it is possible to select multiple parameters to calculate including roughness height waviness and geometry Hit Compute to finish the calculation More details on this function can be found in the user s manual EXPORTING DATA 1 In the dataplot view click on the file menu and select Export Save the data as Scan Data unless you are creating an auto program which can be found in the user s manual Select if it should be saved as either Tab Delimitated or Comma Delimitated Select if you would like to Overwrite the existing file or Append the data to an existing text file Any further questions regarding the use of this tool that are not covered in this SOP or the users manual should be brought to the attention of the POC
4. n 1 3 mg Measurement range Can be set from between 6 5um and 1mm Profile select hills valleys or hills and valleys 4 Once all variables are set select OK and then click on the Sample Positioning Icon k on the top of the screen 5 Adjust the measurement location 1f necessary and then hit the Run Current Scan e JC Routine Icon 6 If there is a problem during the scan you can either hit Escape or press the amp icon at the top of the screen TRACE ANALYSIS 1 Trace Leveling e After the measurement is complete the program will bring you to the Dataplot Screen e The cursors are labeled as R and M corresponding to Reference and Measurement respectively e Place the cursors in the positions you would like the plot to be leveled by dragging them with the mouse e Ifyou would like the leveling to be done over a range of data points near each cursor then click on the bottom of the cursor and expand the bandwidth e Hit F8 or the Leveling Icon it to level the plot 2 Average Step Height Measurement Place the Measurement Cursor over the section of the plot to measure and the average step height will be displayed on the bottom right of the screen under ASH A summary of the data will also be displayed in this area including Reference cursor position and width Measurement cursor position and width average step height and the distance between the cursors If you would like to store th
Download Pdf Manuals
Related Search
Related Contents
Registre santé sécurité au travail : mode d`emploi 456405_a Titelblatt - AL-KO Intel IIX999LSGE01 development software Teac SR60iP Operating Instructions Manual Kenmore 253.25002 Dehumidifier User Manual Copyright © All rights reserved.
Failed to retrieve file