F1001-Series Coriolis Mass Flowmeter
Contents
1. E Calibration Press E to enter gt E Inner Para Press E to enter System Para Press E to enter 86 Password 000000 E 8 Flow Config Press E to enter Output Config Press E to enter a 10 Unit Config Press E to enter ol 11 PressureComp Press E to enter L 12 Other Press E to enter D 13 Addons Functions Press E to enter E 14 E Change Password Press E to enter 60 g s us e FlowCal XXX X so Freq x xxx Hz Time x xxx us 76 Temperature Cal El X XXXX Mass Unit t Output AUTO Sim Press E to enter OW Analyse Switch OFF EnterOldPassword 000000 Temp Correction X XX Drv xxx L xx Rxx Temperature Zero X XX 20 C OilDen g mL X XXX EnterNewPassword 000000 Pred xxxxxx us Ind Density Zero X XX Volume Unit m3 InputPress MPa XX XX Flow Sim Start Press E to enter 20 C WaterDen g mL X XXX NewPasswordA gain 000000 H Period K2 us XXXXXX X PUL CUR XxXxx x Hz XX XX mA Menu Language ENG
2. external pressure sensor connected to the flowmeter s corresponding terminals FlowPress xX XX 141 11 Coefficient of pressure compensation for flow as percentage of calibration coefficient correction for i E edit CompCoe pressure deviation of 1MPa Default value is 0 01 MPa InputPress MPa XX XX AR Actual pressure value for manual compensation mode Other 113 4 Press E to enter EAA Miscellaneous functions such as flow simulation etc Press E to start automatic flow simulation procedure The sequence of frequency and current test Output AUTO Sim signals appears on the flowmeter s outputs Signal on the current output changes from 4 to 20 mA in Press E to enter steps of 0 5mA frequency on the pulse output changes from O to 12000 Hz in steps of 375 Hz Values change every 5 seconds 146 12 Input mass flow rate used in flow simulation Signals on pulse and current outputs will correspond to 45 FlowSimulate t h X XX 42 Flow Sim Start 147 12 Flow simulation mode starts by pressing E Simulated flow rate is specified in the Menu item 45 Press E to enter E edit Simulation stops after pressing unu G Hold Select the menu item which will be displayed if no button is pressed within 2 minutes Select Mass 8 First Menu Mass er to move to Menu item 1 Volm to move to Menu item 2 If Hold option is selected moving to other 3
3. CAUTION The flowmeter is not intended for use at nuclear facilities A CORIOLIS MASS FLOWMETER USER MANUAL 1 2 Principle of operation The flowmeter consists of the following units as shown in Figure 1 1 e Flow sensor 1 e Transmitter 2 Figure 1 1 Appearance of the flowmeter Transmitter can be mounted right on the flow sensor integral type or separate from it Separate type The sensor is a measuring chamber with inlet and outlet flanges for mounting on a pipeline Inside the measuring chamber there are two parallel U shaped flow tubes which vibrate by means of an electromagnetic coil and a magnet The principle of operation is based on the Coriolis effect Figure 1 2 shows the forces affecting the flow tube through which the measured liquid is flowing during the half cycle fluctuations when the tube moves up At this time the liquid flowing into the tube creates resistance to its upward movement and downward pressure on the tube Absorbing vertical momentum by driving around the tube s bend the liquid flowing out of the pipe pushes the tube up This makes the tube twist When the tube is moving down in the second half of the oscillation cycle it twists in the opposite direction This twisting is called Coriolis effect Figure 1 2 Forces affecting the tube on the upward moving www qpeus com CORIOLIS MASS FLOWMETER USER MANUAL Twist angle of the sensor tube is directly pr
4. Temperature measurement is performed using the built in platinum temperature sensor Pt100 The measured temperature is used for automatic correction of the flow and density when the medium temperature changes Temperature correction factors of flow and density are entered in the flowmeter memory at the factory and can be changed through the menu display or Modbus Flowmeters have the manual correction of flow when the medium pressure changes When manual correction mode is enabled in the menu the user should specify the actual pressure value taken from the external pressure sensor see Menu items 39 41 The sensor generates the primary electrical signal containing information about the time delay between the signals of the detectors The primary signal is transmitted to the electronic transmitter placed directly on the sensor or separate from it The transmitter with digital signal processor processes the primary signal calculates mass and volume flow values temperature correction generates output signals and displays the information on the flowmeter s display Features of this measurement principle e direct measurement of mass flow rate in the pipeline without additional errors e maintaining high accuracy and stability over long time e ability to measure the flow of high viscosity liquids non Newtonian fluids fluids containing solid or gas inclusions e reliable operation under conditions of vibration and misalignment of t
5. 500 000 50 200 20 000 1 000 000 70 000 1 000 000 50 000 1 000 000 100 wWww gpeus com 1 CORIOLIS MASS FLOWMETER USER MANUAL The flow ranges are presented for water at temperature of 20 25 C pressure of 0 1 0 2 MPa and density of 1 000 kg m under standard conditions For liquids of different density the volumetric flow range should be calculated by dividing these flow range limits under standard conditions by actual density value If the measured flow rate is less than low flow cutoff value the flowmeter will indicate zero flow and accumulation of mass and volume will pause Low flow cutoff value is set to 1 of the maximum flow rate Cutoff value can be changed through the menu display or through Modbus Flowmeter can measure flow over 1 of the upper limit of the full flow range but measurement error in the range of 1 to the lower limit of the full flow range 2 is not standardized However this error can be estimated by the formula 1 1 Medium density measurement range is 200 3000 kg m 1 3 3 Accuracy of Relative basic error of measurement of mass flow mass on pulse and digital output measurement signals 5m calculated as Om t do Z Qu 100 1 1 where accuracy class Z zero stability according to Table 1 2 kg h Qu measured mass flow rate kg h Note For the accuracy flow range corresponding to a given accuracy class according to Table 1 2 the value of
6. Current Output Mass VolumeFlow Unit m3 h First Menu Mass GasContainCorre X XXX PWD Succeed NewPWD 000000 Water x xxx g mL Oil x xxx g mL MultiPointCorrect OFF Figure 2 14 Display menu structure Reverse Flow Tol DEC MODBUS Address XX Density Unit g mL W ContainCorre X XXX L Period K1 us XXXXXX X Restore Para Press E to enter Mass Inventory t 2 Baud Rate X XXX 29 9600 37 Temperature Unit X C J Parity Bit No X TimeValue XX XX Volminventory m3 X XXX 30 None 31 J Stop Bits One LowerDen h mL X XXXX No X CorrectValue XX XXX SN Table 2 7 Displayed parameters MMM wo3 sSn Indication format panton Description actions Measured parameters display 1 Version and data of electronics Showed on flowmeter s power up 2 0 Total mass and instant mass flow rate GPE Ver 2 28 2012 03 Flow XXX XXX Mass XXX XXX Flow XXX XXX Volm XXX XXX Oil XXX XXX A 34 Water XXX V4 if
7. Baud rate 1200 2400 4800 9600 bit s Data format 1 start bit 8 data bits No parity bit and 1 or 2 stop bits or 1 Odd Parity or Even Parity and 1 stop bit Can be set in the Menu items 30 31 Default data format is 9600 8 N 1 Maximum length of data frame 1 byte address 253 byte data 2byte CRC 256 Bytes Frame structure Function 01 Read Coil Status Query Slave address Function Coils quantity CRC 01H OOH 08H OOH 08H Starting address Response Function Coil status CRC 01H Slave address Byte count 01H Error Error code CRC 01H Function 81H Slave address Exception code 01H Function is not supported 02H Incorrect Starting address or Coils quantity 03H Coils quantity is out of range of 1 200 04H Failed to read coil Function 04 Read Input Registers Query Slave address Function Starting address Registers quantity CRC 04H Response Slave address Function Data value CRC 04H Byte count Error Function Error code CRC 84H 01H Slave address Exception code 01H Function is not supported 02H Incorrect Starting address or Registers quantity WWW eus com i CORIOLIS MASS FLOWMETER USER MANUAL 03H Registers quantity is out of range of 0 125 04H Failed to read registers Function 05 Write Single Coil Query Slave address Function Coil address Data value Response Slave address Function Coil address Data value Error Sla
8. That pressure must be not less than critical pressure Pcr calculated as Pcr 2 9 AP 1 3 py kPa 1 6 where AP pressure drop on the flowmeter kPa Py saturated steam pressure at working conditions background information kPa If the pressure calculated by this formula exceeds the actual pressure in the pipeline a safety valve should be installed to increase the pressure www gpeus com CORIOLIS MASS FLOWMETER USER MANUAL 1 5 Explosion protection The flowmeters of explosion proof modification GPE F1001 Ex have a combined type of protection flameproof enclosure complied with GOST R 51330 1 and the input and output intrinsically safe level ib complied with GOST R 51330 10 The sensor explosion proof grade is showed in Table 1 7 Table 1 7 Sensor explosion proof grade Temperature code Explosion proof grade 100 1Ex ibllCT4X 200 1Ex iblICT3X 350 1Ex iblICT1X Transmitter explosion proof grade is 1Exdfib lICT6X Explosion proof grade is written on the name plates attached to the body of the sensor of explosion proof modification and to the transmitter The name plates appearance is showed one flowmeter ordered The X letter in the explosion proof grade means the special requirements the measured medium temperature must not exceed the maximum temperature according to the explosion proof grade temperature group explosion protection is provided under pressure not
9. e excess pressure in the pipeline e high temperature of the medium Operation of flowmeters of Ex modification must be performed in accordance with the requirements of Chapter 7 3 of the Electrical Installations Code and other regulations of using of electrical equipment in explosive environment CAUTION Installation and operation of the flowmeter in conditions of pressure or temperature exceeding their maximum allowable values is prohibited Do not use the flowmeter with the cover opened and also without the chassis grounding Heh y To prevent leakage of the gas filling the sensor s outer shell do not open the cover 1 WWw gpeus com A CORIOLIS MASS FLOWMETER USER MANUAL 2 3 Installation on pipeline 2 3 1 Installation When selecting the installation place you should follow these rules location e There should be no strong vibration high temperatures or strong magnetic fields at the installation place Therefore do not install the flowmeter close to transformers power units and other mechanisms that create vibration and electromagnetic interference e The flowmeter should not be installed in the strained section of the pipeline and should not be used as a pillar for the pipeline e It is recommended to provide moisture protection for the flowmeter e Flowmeter should be installed in easily accessible places Free space around the flowmeter should be provided for easy instal
10. 350 290 390 185 188 150 A 1 2 A 1 4 1100 1050 1600 380 280 440 320 323 200 A 1 2 A 1 4 1380 1160 1740 420 335 535 625 628 Overall width of the body excluding transmitter Table A 2 Flowmeter flange dimensions DN PN d1 D6 BYA h h2 mm MPa mm mm mm mm ae 1 6 2 5 4 10 40 60 90 12 2 4 14 6 4 8 35 41 70 100 16 4 3 4 14 1 6 2 5 4 15 46 65 95 12 2 4 14 ot 6 4 11 6 40 46 75 105 16 4 3 4 14 1 6 2 5 4 27 3 65 85 115 13 3 4 14 5 6 4 24 8 58 65 100 140 20 4 3 4 18 1 6 2 5 4 41 1 85 110 150 13 3 4 18 a 6 4 37 76 84 125 170 22 4 3 4 22 1 6 2 5 4 52 3 99 125 165 18 2 4 18 os 6 4 47 88 99 135 180 22 4 3 4 22 1 6 2 5 4 79 5 132 160 200 20 2 8 18 a 6 4 17 121 132 170 215 24 4 3 8 22 1 6 2 5 4 101 7 156 190 235 21 3 8 22 es 6 4 94 150 156 200 250 25 5 4 5 3 5 8 26 1 6 2 5 4 154 211 250 300 26 2 8 26 6 4 142 204 211 280 345 31 5 4 5 3 5 8 33 1 6 2 5 4 200 285 320 375 35 3 12 30 ov 6 4 198 260 284 345 415 37 5 4 5 3 5 12 36 www eus com CORIOLIS MASS FLOWMETER USER MANUAL Figure A 2 Outline dimensions of connection kit flanges Table A 3 Connection kit flange dimensions and weight DN PN d1 DYA D1 D mm MPa mm mm mm mm ang 1 6 2 5 4 10 40 60 90 12 2 4 14 0 7 6 4 8 34 70 100 16 4 4 14 1 0 1 6 2 5 4 15 46 65 95 12 2 4 14 0 8 je 6 4 11 6 39 75 105 16 4 4 14 1 1 1 6 2 5 4 27 3 65 85 115 13 3 4 14 1 2 ae 6 4 24 8 57 100 140 20 4 4 18 2 3 1 6 2 554 41 1 85 110 150 13 3 4 18 2 1 6 4 37 75 12
11. CORIOLIS MASS FLOWMETER USER MANUAL Table 2 4 Notes for Figure 2 7 1 Transmitter base 2 Bolts 3 Clamp 4 Nuts CAUTION Do not install the transmitter with the cable entry directed vertically upwards 2 3 5 Thermal If thermal insulation of the pipeline and the flowmeter is necessary see the insulation recommendations in Figure 2 8 Figure 2 8 Recommendation for thermal insulation of the flowmeter www qpeus com 1 CORIOLIS MASS FLOWMETER USER MANUAL 2 3 6 Cooling For high temperature modification of the flowmeter with medium temperature above 200 C possibility of external cooling is provided see Figure 2 9 Figure 2 9 External cooling supply 2 3 7 Transmitter If in the process of installation the front side of the transmitter appeared to be not in rotation front of the user and display is not visible it is possible to rotate the transmitter at an angle of 90 or 180 so that the display is facing the user To do this unscrew the four bolts 1 see Figure 2 10 Then turn the transmitter 2 by 90 or 180 in the desired direction and tighten those 4 bolts to ensure sealing Figure 2 10 Rotation of the transmitter CAUTION To avoid excessive twisting of wires inside the flowmeter do not turn the transmitter at an angle over 180 regarding its initial position www gpeus com EER CORIOLIS MASS FLOWMETER USER MANUAL 2 4 Wiring 2 4 1 Basic operations El
12. DN50 DN200 0 27 300 8 0 5 Maximum length of the connection cable for the separate type flowmeter is 300 m Drive coil power circuits are electrically isolated from other circuits by means of the undamaged transformer according to GOST R 51330 10 Insulation between primary and secondary windings can withstand voltage of at least 1 5 kV Ambient temperature for the flowmeters of Ex modification must be between 20 and 40 C www qpeus com A CORIOLIS MASS FLOWMETER USER MANUAL 1 6 Delivery set The base delivery set and the accessories for the flowmeter are shown in Figures 1 6 1 7 and in Tables 1 12 1 13 Figure 1 6 Base delivery set Table 1 12 Base delivery set Item Den Base delivery Special No P set order Coriolis Mass Flowmeter Cable F1001 series length 2 User manual 3 Passport of the flowmeter 4 Packing box 5 Verification methods document www qpeus com CORIOLIS MASS FLOWMETER USER MANUAL Figure 1 7 Accessories Table 1 13 Accessories 1 Connection kit flanges gaskets stud bolts nuts washers 2 Flowmeter s replacement mounting part CAUTION Upon receipt of the flowmeter it is necessary to e Check the packaging box for damage e Make sure that delivery set is complete e Make sure the flowmeter model matches the order data If the package is damaged delivery set or flowmeter model doesn t match the order you should draw up a statement www
13. Menu item 52 OW Analyze Switch if set to On Den XXX XXX Temp XX XX 14 1 Density and temperature of the measured medium Configuration 15 1 p f fiouri f E86 ress E button to enter the submenus for configuring the flowmeter Press E to enter Submenus of the Configuration menu Total mass and instant volumetric flow rate This menu item is not displayed when Menu item 52 13 1 OW Analyze Switch if set to On Mass flow rate of pure oil and percentage of water in the oil water This menu item is displayed only The following submenus can be accessed only after entering the 6 digit password If the entered password is invalid a message Wrong password will be displayed To return to the Menu item 4 Configuration press To enter the password again press Password 000000 Flow Config Press E to enter ee Allows user to specify low flow cutoff flow direction and the way of its counting Zero point adjustment The value displayed is the time delay between the signals coils taken as Zero Calibration X XXX us Lae zero point Press E to start zeroing procedure described in paragraph 2 5 4 118 8 If the current mass flow rate is less than specified Low Flow Cutoff value then the flow rate is E edit assumed as zero and totalizers will pause 119 8 Press E to reset mass and volume totalizers resettable totalizer Low Flow Cuto
14. Z is assumed to be 0 Absolute basic error of measurement of medium density Ap is 1 kg m Absolute basic error of measurement of medium temperature is 1 C Additional error of measurement of density caused by a change of medium temperature is 0 03 kg m for every 10 C of deviation from the density calibration temperature Additional error of measurement of density caused by a change of pressure is 0 015 kg m for every 100 kPa of deviation from the density calibration pressure www gpeus com CORIOLIS MASS FLOWMETER USER MANUAL Relative basic error of measurement of volumetric flow volume on pulse and digital output signals dy calculated as Sy 50 Ap p 100 Z Qy 100 1 2 where 6 accuracy class Ap absolute basic error of measurement of medium density kg m p measured medium density kg m Z zero stability according to Table 1 2 L h Qy measured volumetric flow rate L h Note For the accuracy flow range corresponding to a given accuracy class according to Table 1 2 the value of Z is assumed to be O Relative basic error of measurement of mass flow mass on current output signal Sim calculated as m Om 0 2 lmax 4 16 Qm Qumax 1 3 where dy Relative basic error of measurement of mass flow mass Imax 20 MA maximum value of current output signal Qu measured mass flow rate kg h Qumax upper lim
15. calibration parameters of the flowmeter cannot be changed entering the menu section 5 Calibration and section 7 System Para is not available Such a state of the switches is necessary to change of some Modbus registers through digital interface such as the Modbus register map version or display language see Appendices D and E If any of switches 1 8 is set to ON the entrance to the Calibration and System www gpeus com si CORIOLIS MASS FLOWMETER USER MANUAL Para sections will also be impossible To enter the Calibration and System Para sections and change or just see the calibration coefficients it is necessary to set all the switches to OFF positions see Figure 2 16 Figure 2 16 Protection switch block in OFF positions www eus com i CORIOLIS MASS FLOWMETER 2 5 6 Maintenance USER MANUAL The flowmeter put into operation does not require special maintenance other than periodic inspection to verify the following e Compliance with the working conditions e Whether the voltage and the other parameters of the power supply conform to the requirements of parameter 1 3 4 Power supply e Visibility of the information on the nameplates and other labels e Cleanness of the exterior surfaces of the flowmeter e Tightness of connections of the flowmeter to the pipeline e No damage to the flowmeters parts is visible Examination period for the flowmeter depends on operation conditions an
16. gpeus com a CORIOLIS MASS FLOWMETER USER MANUAL 1 7 Model codes F1001 series coriolis mass flowmeter s model codes are presented in Table 1 14 The example of the flowmeter s modification for the order is showed below ee ist dst st ic tir i tet oft R ER G 1 16 1 24 A A F1001 U 050 1 0 1 Table 1 14 Flowmeter model codes Sensor version T Triangle shaped size from 1 2 to 1 M Micro bent shaped size from 1 2 to 1 U U shaped size from 1 1 2 to to 8 Nominal diameter DN 010 DN 10 mm 015 DN 15 mm 025 DN 25 mm 040 DN 40 mm 050 DN 50 mm 080 DN 80 mm 100 DN 100 mm 150 DN 150 mm 200 DN 200 mm X Special order 2 Connection options A ANSI D DIN J JIS X Special order Max Process Pressure 1 6 PN 1 6MPa 2 5 PN 2 5MPa 4 0 PN 4 0MPa 6 4 PN 6 4MPa 4 Max Process temperature 1 Process temperature 50 125 C 2 Process temperature 50 200 C Separate type only 3 Process temperature 50 350 C separate type only NX Non Explosion EX Explosion proof wWww gpeus com CORIOLIS MASS FLOWMETER USER MANUAL 1 DC 24V 2 AC220V P Pulse 4 to 20mA R RS485 4 to 20mA Pulse H Hart Pulse 4 to 20mA DE LE 0 1 Accuracy 0 1 0 2 Accuracy 0 2 0 5 Accuracy 0 5 wWww gpeus com A CORIOLIS MASS FLOWMETER USER MANUAL 2 OPERATI ON AND MAI NTENANCE 2 1 Model selection recommendations One of the most important conditions for
17. human health and the environment during and after the life of the flowmeter and after utilization Utilization of the flowmeter is performed separately for groups of materials plastic parts metal parts and fastening parts 4 VERI FI CATI ON Verification of the flowmeters is performed in accordance with the document Instructions GSI F1001 Series Coriolis Mass Flowmeter Verification procedure WWW eus com i CORIOLIS MASS FLOWMETER USER MANUAL APPENDIX A Outline dimensions and weight A 1 1 1 L B E S A x 1 B L1 A 1 3 A 1 4 Figure A 1 Outline dimensions www gpeus com 1 CORIOLIS MASS FLOWMETER USER MANUAL Integral type Figure A 1 1 A 1 2 Separate type Figure A 1 3 A 1 4 Table A 1 Outline dimensions and weight oer L mm Bi C max Weight kg MPa MPa A 1 2 A 1 4 010 A 1 1 A 1 3 150 170 350 290 260 130 90 11 14 015 A 1 1 A 1 3 180 192 350 290 260 108 95 11 14 025 A 1 1 A 1 3 200 248 450 400 290 188 115 15 18 040 A 1 2 A 1 4 520 547 470 660 280 182 150 30 33 050 A 1 2 A 1 4 558 588 550 750 290 198 165 39 38 080 A 1 2 A 1 4 780 808 710 1040 320 227 205 80 83 100 A 1 2 A 1 4 920 946 860 1290
18. menu items will not occur CheckSystem OFF 149 12 8 State ON E edit Enable disable system check function ae 50 12 A9 Flow Filtering s X XXX E edit Flow rate averaging time from 0 5 to 10 s in steps of 0 5 s TRE 144 12 Den Filtering s X X E edit Density averaging time from 0 5 to 30 s in steps of 0 5 s Addons Function eee Entering parameters for measurement of two component media Press E to enter E52 g P p OW Analyse si 153 13 52 Switch E edit Enable disable oil water analyze function 154 13 53 20 C OilDen X XXX g mL E edit Density of pure oil under standard conditions r 155 13 ar 54 20 C WaterDen X XXX g mL E edit Density of pure water under standard conditions 156 13 Flow compensation by percentage of gas in the oil water It works only if oil water analyze function is oe ee EUR CS CSS E edit enabled in the Menu item 52 157 gt 13 Flow compensation by percentage of pure water in the oil water not realized in current versions of 56 WContainCorre X XXX E L edit flowmeters Densitv Rande OFF 158 13 Enable disable density range check When enabled and If density of the medium is out of range of y J ON E edit specified in menu items 58 and 59 then the volumetric flow will not be calculated UpperDen X XXXX E The upper limit of density range see Menu item 57 152 13 E edit Change Password 18 4 Password
19. or less than the nominal diameter of the pipeline When mismatch pipeline diameter and the nominal diameter of the flowmeter tapered transitions can be used They can be made independently at that to ensure minimum loss of pressure the central cone angle must not exceed 30 C WWw gpeus com CORIOLIS MASS FLOWMETER USER MANUAL 2 2 Safety precautions Installation operation maintenance of flowmeters should be performed by persons studied this manual and safety instructions for working with electrical devices All operations on calibration and usage of flowmeters must comply with the requirements for protection against static electricity Installation of flowmeter in the pipeline and its removal from the pipeline should be performed without pressure in the pipeline and with the power supply switched off Electrical connection should also be performed only when the power supply is switched off During the installation commissioning and maintenance shall be prohibited e replacement of electronic components when the flowmeter is powered on e connecting the flowmeter to the power supply with output voltage other than specified in this manual e using electrical devices and tools without protective grounding and also in case of their malfunctions During installation the hazardous factors are e power supply voltage of 220V AC or higher in case of the external power supply located in close vicinity of the installation place
20. 0 Current primary default 1 Pulse 2 Digital 3 Current secondary 4 Discrete output 5 Discrete input For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7 To write in registers the switches 2 4 6 8 must be in ON position 32 bit registers single precision floating point values complied with IEEE 754 format 4 bytes high bytes first Functions Read 04 Write 16 For example value 1 5 corresponds to OxBF OxCO 0x00 0x00 in the order from low to high address in memory ENS EE TE 0149 0150 RW Low density cutoff g cm 0155 0156 ORW Low density for density calibration giem a 0157 0158 High density for density calibration gicm 0159 0160 Period for low density us 0161 0162 Period for high density us 0163 0164 Temperature correction coefficient for density 100 C gg 0189 0190 Flow rate averaging time from 0 5 to 10 s in steps of 0 05 s RW 0191 0192 RW Temperature averaging time s 0193 0194 Density averaging time from 0 5 to 30 s in steps of 0 05 s 0195 0196 RW Low mass flow cutoff kg h 0197 0198 Low volume flow cutoff L h 0199 0200 Upper limit of density range g cm 0201 0203 RW Lower limit of density range g cm 0231 0232 Ro Zero point standard deviation us a 0233 0234 RO Zero point us www gpeus com A CORIOLIS MASS FLOWMETER USER MANUAL 0247 0248 RO Mass
21. 187 242 V 15 VA www qopeus com EER CORIOLIS MASS FLOWMETER USER MANUAL 1 3 5 Output signals The flowmeter has the following output signals e pulse output e current output e digital output RS 485 interface To display the values of the mass flow volume flow and other measured parameters the flowmeter s transmitter has built in display 1 3 5 1 Pulse output Pulse output signal is a periodical pulse signal with the frequency which is proportional to the measured value of the mass flow rate considering the damping time specified in the menu 49 The pulse output can be configured to denote mass flow rate volumetric flow rate or density of the medium Pulse output is active The total number of pulses generated on the pulse output corresponds to the mass or volume of the fluid passing through the flowmeter since the measurement start Maximum frequency fmax of the pulse output signal calculated as fmax Qmax 3 6 m Hz 1 5 where Qmax upper limit of the full mass flow range kg h m pulse weight g pulse Pulse output signal frequency range is 0 10000 Hz Maximum frequency can be increased up to 12000 Hz The amplitude of the pulse output signal is 13 V Default pulse weight is presented in Table 1 4 Table 1 4 Default pulse weight am so 1s 2 wo 0 oo 10 v0 2 20 40 vse oo abuse CARE ERES ES EE 1 3 5 2 Current output The value of the current in the current output circuit rang
22. 5 170 22 4 4 22 3 7 1 652 554 52 3 99 125 165 18 2 4 18 2 8 oe 6 4 47 87 135 180 22 4 4 22 4 6 1 6 2 5 4 79 5 132 160 200 20 2 8 18 4 8 a 6 4 77 120 170 215 24 4 8 22 7 2 1 6 2 5 4 101 7 156 190 235 21 3 8 22 7 0 Le 6 4 94 149 200 250 25 5 4 5 8 26 10 7 1 6 2 5 4 154 211 250 300 26 2 8 26 13 2 ne 6 4 142 203 280 345 31 5 4 5 8 33 25 4 1 6 2 5 4 200 285 320 375 35 3 12 30 24 0 a 6 4 198 259 345 415 ETS 4 5 12 36 38 5 www gpeus com 1 CORIOLIS MASS FLOWMETER USER MANUAL APPENDI X B Wiring diagrams Transmitter Sensor gt Left signal coil Left signal coil Right signal coil Right signal coil Yellow Drive coil Green Drive coil Blue Temperature White Temperature compensation Black Figure C 1 Separate type transmitter connection Power supply gi lt Output signal J SISS N L Figure C 2 Power supply connection www gpeus com i CORIOLIS MASS FLOWMETER USER MANUAL Output signal Power supply F F l I B A Power supply F F l I B A Figure C 4 Current output connection Output signal Power supply anasoj SSN F F l l B A Figure C 5 Digital output connection www gpeus com 1 CORIOLIS MASS FLOWMETER USER MANUAL APPENDIX C Modbus register map version 2 xx Modbus register map version V2 17 Interface mode RS 485 asynchronous half duplex
23. 7 Time delay for point 4 D4 us 0148 0149 Time delay for point 5 D5 us 0150 0151 Time delay for point 6 D6 us 0152 0153 Time delay for point 7 D7 us 0154 0155 Correction coefficient for point 0 KO 0156 0157 Correction coefficient for point 1 K1 0158 0159 Correction coefficient for point 2 K2 0160 0161 Correction coefficient for point 3 K3 www gpeus com DI CORIOLIS mass FLOWMETER USER MANUAL 0162 0163 RW Correction coefficient for point 4 K4 0164 0165 RW Correction coefficient for point 5 K5 0166 0167 RW Correction coefficient for point 6 K6 Address Mode Description Menu item 0168 0169 RO Mass flow rate p 0172 0173 RO Temperature C P 0174 0175 RO Volumetric flow rate L s po 0176 0177 RO Mass total kg 0178 0179 Volume total L 0180 0181 RO Current value on the current output mA 0182 0183 RO Frequency of the signal on the pulse output Hz 0184 0185 RO Measuring tubes oscillation frequency Hz 0186 0187 Left coil voltage mV 0188 0189 RO Right coil voltage mV 0190 0191 RO Drive coil load 0202 0203 RO Internal temperature 7m For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7 Reserved for future modifications www eus com i CORIOLIS MASS FLOWMETER USER MANUAL APPENDIX D Modbus register map version 3 xx ProLink Modbus register map version V3 02 Genera
24. EE F1001 Series Coriolis Mass Flowmeter Ha USER MANUAL Direct mass measuring High accuracy Digital signal processing Straight upsteam downstream piping not required High viscosity liquids measuring O O v Q Q a z gt gt GPE Inc EER CORIOLIS MASS FLOWMETER USER MANUAL General information WWw gpeus com This user manual provides basic technical specifications instructions for usage storage and transportation and other information necessary for the proper operation of the F1001 Series Coriolis Mass Flowmeter hereinafter the flowmeter or F1001 Modbus is the abbreviation for Modicon Modbus Protocol and it is a registered trademark of Modicon Inc ProLink is a registered trademark of Micro Motion Inc Golden Promise Equipment Inc hereinafter called GPE reserves the right to make changes in the design of the flowmeters not deteriorated of their consumer characteristics without prior notice If you need additional information on the GPE equipment please contact your local dealer or the head office Any use of the material of this publication in whole or in part without written permission of the copyright holder is prohibited CAUTION Before you start operating the flowmeter you should carefully read this manual Before starting the installation use or maintenance of the flowmeters make sure that you fully read and understood the contents of the
25. N position Reserved for future modifications WWW eus com CORIOLIS MASS FLOWMETER USER MANUAL 32 bit registers single precision floating point values complied with IEEE 754 format 4 bytes high bytes first Functions Read 04 Write 16 For example value 1 5 corresponds to OxBF OxCO 0x00 0x00 in the order from low to high address in memory Address Mode Deserpion menuitem 0100 0101 Calibration coefficient g s us 60 0102 0103 Temperature correction coefficient for flow 100 C et 0104 0105 High density for density calibration gicm ns DS 0106 0107 Period for high density Ropes 0108 0109 Low density for density calibration g cm 0110 0111 Period for low density us e 0112 0113 Temperature correction coefficient for density 100 C 0114 0115 RW Pulse weight puise weight a 0116 0117 Value relating to 20 mA 0118 0119 Low flow cutoff a foe gh 0120 0121 Zero point wept 4 0122 0123 Density of pure oil under standard conditions g cm CE 0124 0125 RW Density of pure water under standard conditions gem 4 D nt at nette ee 0134 0135 RW ___ Correction coefficient of the density for simulation mode 78 0136 0137 RW Zero density for simulation mode 7e o Time delay for point 0 DO us 0140 0141 Time delay for point 1 D1 us 0142 0143 RW Time delay for point 2 D2 us Us Us 0144 0145 Time delay for point 3 D3 us 84 0146 014
26. OWMETER USER MANUAL 2 4 5 Grounding Transient processes induced by lightning welding big power consuming equipment or switches may lead to distortion of the flowmeter s measuring or damage it For protection from transient processes one should provide a connection of the ground terminals located on the transmitter body see Figure 2 13 with the ground through a wire designed for operation under high currents For grounding it is recommended to use copper wire of at least 2 5 mm AWG13 section Ground wires should be as short as possible and have a resistance of less than 1 Ohm Transmitter can be grounded through the pipeline if the pipeline provides the ground Figure 2 13 Grounding CAUTION There should be no potential applied or induced on the ground wire Do not use the same wire for grounding of two or more devices WWw gpeus com 1 CORIOLIS MASS FLOWMETER USER MANUAL 2 5 Operation and maintenance 2 5 1 Basic recommendations 2 5 2 Power Up 2 5 3 Display operating To ensure reliable operation of the flowmeter and maintain accuracy it is necessary to satisfy the following requirements e To prevent damage to the measuring tubes of the flowmeter because of hydroblow opening closing valves on the inlet pipe should be done gradually e Operation at flow rates close to the upper limit of the full flow should last no more than 2 hours per day Right after power up the flowmeter performs a
27. UTION When using the flowmeter in hazardous areas apply the requirements for explosion protection provided in paragraph 2 4 2 Installation with explosion protection www gpeus com EER CORIOLIS MASS FLOWMETER USER MANUAL 2 4 2 Installation with Installation of flowmeters in explosive environment must be performed in explosion protection accordance with requirements of This manual Section 3 4 of Operational Code for Electrical Installations Section 7 3 of Electrical Installations Code GOST R 51330 0 GOST R 51330 1 GOST R 51330 10 Instruction BCH332 74 MMCC Instructions for installation of electrical equipment power and lighting lines in hazardous areas Other regulations in force within the enterprise During installation you should pay attention to the specific operating conditions mentioned in Section 1 5 Explosion protection Before installation the flowmeter should be inspected You should pay special attention to explosion proof grade labels warning labels make sure that there is no damage to the flameproof shell and flow sensor Also make sure that grounding terminal and the seals for cables and covers are available and in good condition also check the connecting cables condition Upon completion of the electrical installation the electrical resistance of the ground line should be checked It must not exceed 1 Ohm A copper wire with section of at lea
28. ay 75 kg h default 78 t h Address Mode Description Menu item 0040 RW oO O Density unit 91 gicm default 96 kg L 92 kg m RW Temperature unit 32 C default 33 F Volumetric flow rate unit 0 mL s 24 L s 17 L min 29 m day 138 L h default 19 m h Pressure unit 0 MPa 7 Bar default 12 kPa Mass total unit 60 g 61 kg default 62 t Volume total unit 0 mL 41 L default 43 m Device type code 40 Core Processor 21 41 42 RFT9739 1 700A 2700A eek Q gt N 0041 RW Q3 O1 CD 0042 0044 0045 R 0046 0120 0125 0126 0136 0313 0419 0420 0421 RW 28 0423 0424 RO Alarmscode d o 0521 RW Bytes sequence in floating point 32 bit registers 0 0 1 2 3 default 1 2 3 0 1 2 1 0 3 2 3 3 2 1 0 1138 Output signals 0 None 1 Current Pulse RS485 default 2 Fieldbus H1 or Profibus PA 1166 Output channel A type 0 Current primary default 1 Pulse www gpeus com KR NO NO A A N O1 i CORIOLIS MASS FLOWMETER USER MANUAL 2 Digital 3 Current Secondary 4 Discrete output 5 Discrete input Address Mode Description Menu item Output channel B type 0 Current primary default 1 Pulse 2 Digital 3 Current Secondary 4 Discrete output 5 Discrete input Output channel C type
29. changing is used for protection against unauthorized access to the settings of the Press E to enter E8 flowmeter LowerDen The lower limit of density range see Menu item 57 EX Coriolis mass flowmeter MMM woo sna 87 88 89 90 EnterOld PassWord EnterNew PassWord NewPassWord Again PWD Succeed Calibration Press E to enter FlowCal Temp Correction H Density D2 H Period K2 L Density D1 L Period K1 Density TempComp Zero Correction Inner Para Press E to enter Freq Time 000000 000000 000000 NewPWD 000000 X XXX g s us X XX X XXX g mL XXXXXX X US X XXX g mL XXXXXX X US X XX 00 XX XX HZ XX XX US edit E88 edit E89 edit E90 114 14 E14 lt nm S 161 5 E edit 162 5 E edit 163 5 E edit 164 5 E edit 165 5 E edit 166 5 E edit 167 5 E edit If the entered password is invalid a message Wrong password will be displayed To return to the Menu item 14 Change Password press To enter the password again press Input new password Confirm new password New password is displayed Setting of calibration parameters provided for measuring flow density temperature Calibration coefficient K Temperature correction coefficient for flow rate as percentage of calibration coeffici
30. conditions 72 Oil X XXX g mL J Density of oil in oil water under working conditions X X XXXX XXXX XXXX XXXX XXXX XX XXXX 174 6 Internal parameters for pressure compensation mode GPE EN kg CX XX 175 36 Menu language EN version of the display menu X XX 9600 8 N 1 ty YY Modbus data den register map version Y YY XXXXXX 1 68 6 Display low volume flow cutoff L h totalizer mode 0 disabled 1 enabled XXXXXX 0 J low density cutoff g mL bytes order for float point values default value is 0 System Para 11 1 System parameters of the flowmeter Press E to enter E76 Intended to use for the representatives of the manufacturer only WIT gt f Correction coefficient of the temperature Tk for simulation mode in the range of 0 9 1 1 76 LAMPE CON an T Tk xTdev T0 Tdev is temperature deviation 77 Temperature Zero X XX B Zero temperature TO for simulation mode in the range of 20C to 20C 179 7 Correction coefficient of the density ok for simulation mode in the range of 0 9 1 1 78 DEDEGE DIED p pkx pdev p0 pdev is density deviation 79 DensityZero X XX Eu E Zero density p0 for simulation mode in the range of 0 1 g cm to 0 1 g cm a 80 Menu Language ia fae E Select display menu language Switching from English language to Russia or Chinese Enable disable multipoint correction function used for correction of the calibration coefficient in 182 7 E edit several poi
31. d defined by the maintenance organization in agreement with the operating organization In case of the flowmeter s failure it is necessary to follow the instructions of the paragraph 2 5 7 Troubleshooting CAUTION Working in inappropriate operating conditions may lead to failure of the flowmeter or increase of the measuring error 2 5 7 Troubleshooting 2 8 Table 2 8 Troubleshooting Probable reason Repairing way www gpeus com 1 When the flowmeter IS powered there IS no indication on display nor any signal on the flowmeter s outputs 2 When the flowmeter IS powered the measured parameters are displayed correctly put there are no signals on the flowmeter s outputs 3 When there is a certain actual flow the flowmeter shows zero flow rate on the Wrong connection of the power wires to the flowmeter Power wires break Power supply voltage does not meet the requirements of this manual Wrong connection of the output wires to the flowmeter or secondary devices The actual flow rate is less than the lower limit of the measuring range for this size of the flowmeter The actual flow rate is less than the low flow cutoff Possible faults their probable reasons and repairing ways are presented in Table Check the power cable connection according to the wiring diagram in Appendix Check the power wires and replace them if needed Check the power supply voltage and
32. densing at 25 C Up to 40 A m 50 Hz class V1 according to GOST R 52931 IP65 4 years Pt100 not less than 12 years See Appendix A Sensor stainless steel Transmitter aluminum alloy Flowmeter does not contain precious metals 1 It is possible to produce flowmeters with special characteristics according to the order 2 Temperature range for OLED display is 40 70 C Flow ranges of the flowmeters with accuracy 0 1 0 2 and 0 5 are shown in Table 1 2 Normal operation of the flowmeter guaranteed within the full flow range according to the Table 1 2 Operation of the flowmeter with the flow rate exceeding the upper limit of the full flow range is not allowed The lower limit of the full measuring flow range depends on the parameters of the medium and should be specified in the order www gpeus com EER CORIOLIS MASS FLOWMETER USER MANUAL Table 1 2 Measuring mass and volumetric flow range for liquid PAE Full flow range Accuracy flow range kg h L h Zero stability 10 20 1 000 70 1000 50 1 000 0 1 15 40 2 000 150 2 000 100 2 000 0 2 25 120 6 000 400 6 000 300 6 000 0 6 40 600 30 000 2 000 30 000 1 500 30 000 3 50 1 000 50 000 3 500 50 000 2 500 50 000 5 80 2 400 120 000 8 000 120 000 6 000 120 000 12 100 4 000 200 000 15 000 200 000 10 000 200 000 20 150 10 000 500 000 35 000 500 000 25 000
33. e modification code 200 T1 for the flowmeters with temperature modification code 350 Explosion protection type of input and output intrinsically safe circuit level ib is ensured by the following means external power supply and connection of the external devices to the pulse current and digital outputs of the flowmeters of Ex modification must be implemented in accordance with GOST R 51330 1 electric load of anti spark circuit elements of the flowmeter does not exceed 2 3 of their certified values www gpeus com 1 CORIOLIS MASS FLOWMETER USER MANUAL the values of parameters of the left right signal coils drive coil and the temperature sensor circuits do not exceed the limits in accordance with GOST R 51330 10 spark safety barrier with Zener diodes is applied electrical clearances and creepage distances comply with GOST R 51330 10 Insulation resistance between the sensor outer shell and electrical circuits elements can withstand the testing voltage of 500 V AC rms internal capacity and inductance of the circuit do not accumulate energy dangerous for spark ignition gas mixtures of category IIC current conducting elements and electronic components of the flowmeter circuit are protected from the environment influence with the shell which provides the protection degree IP 65 according to GOST 14254 Block diagram of the flowmeter is shown in Figure 1 3 Explosion proof ty
34. ectrical connections should be performed in the following sequence see Figure 2 11 e Remove the back cover 1 of the transmitter e Lead signal cables 2 and power cable 7 through the cable entries 3 e Loose the terminal block screws 4 e Connect wires according to the wiring diagram shown in Appendix e Tighten the terminal block screws e Tighten the clamp of the cable entry e f necessary install the blind plug 5 instead of the unused cable entry e Connect the ground cable to the ground terminal 6 e Tighten the cover of the transmitter Figure 2 11 Electrical connection of the flowmeter Table 2 5 Notes for Figure 2 11 1 Back cover of the transmitter 2 3 4 5 6 7 Maximum length of the power cable is 300 m with minimum wire section 0 8 Signal cable Cable entries Terminal block Blind plug Ground terminal Power cable mm AWG 18 www gqpeus com 1 CORIOLIS MASS FLOWMETER USER MANUAL Connecting to the current and pulse outputs should be performed with a twisted pair wire with the maximum length of 150 m and minimal wire section 0 5 mm AWG20 Sensor and transmitter of the flowmeter of separate type are connected through a special 9 core shielded cable with maximum length of 300 m Connection diagram is shown in Figure C 1 of Appendix After mounting and electrical connection the zero point adjustment should be performed see paragraph 2 5 4 Zero point adjustment CA
35. ed with fluid Close the valve located before the flowmeter e Make sure that the fluid in the flowmeter is completely still e Staying in the root menu see Figure 2 14 press button several times to move to the Menu item Configure Press E button and enter the password default password is 000000 To enter the password press to increment the currently selected digit to move to the next digit press button After successful password entering a message Zero calibration will appear Press E to start zero calibration procedure A message Zeroing will appear Press E to confirm that Zero calibration procedure lasts for 30 seconds the remaining time is displayed Zero calibration should also be performed if the flowmeter indicates some flow when there is no actual flow at the moment 2 5 5 Protection switch The flowmeter has a protection switch block see Figure 2 15 which protects the flowmeters calibration coefficients from unauthorized access The protection switch block is located at the front side of display board to the right from the display The protection switch block contains 8 switches To access the protection switch block it is necessary to remove the sealing wire or the sticker from the front cover of the transmitter then remove the front cover Figure 2 15 Protection switch block in default positions At factory switches 2 4 6 8 are set to ON see Figure 2 15 In this position the
36. eline e Using the flowmeter s replacement mounting part fix and center flanges and weld them to the pipeline www gpeus com si CORIOLIS MASS FLOWMETER USER MANUAL CAUTION In the process of installation the flowmeter may be used instead of the replacement mounting part only in the following cases installation is carried out using a gas welding when installing using an electric arc welding the power source is connected in a way that prevents the welding current to run through the flowmeter see Figure 2 3 Right Wrong Figure 2 3 Connecting the power source for arc welding using the flowmeter As a result the installation place must appear as shown in Figure 2 4 where the length L is the sum of the length of the flowmeter and thickness of two gaskets me mn PP gagne pus a mnt Figure 2 4 Pipeline preparation for installation of the flowmeter CAUTION When using filters or gas separators the length L should be enough for their installation www gpeus com EER CORIOLIS MASS FLOWMETER USER MANUAL 2 3 4 Installation Prior to installation it is necessary e Thoroughly clean the pipeline of cinder sand and other solid particles e Inspect the flanges and the measuring tubes of the flowmeter and remove solid mechanical and other particles from them e Remove the preservative grease from the flowmeter by passing through it kerosene gasoline or diesel fuel To install the flowmeter in the p
37. ent correction for temperature deviation of 100 C High density value corresponding to the medium of high density water Used for calibration for density measuring Measuring tubes oscillation frequency for the medium of high density water Used for calibration for density measuring Low density value corresponding to the medium of low density air Used for calibration for density measuring Input current password Measuring tubes oscillation frequency for the medium of low density air Used for calibration for density measuring Temperature correction coefficient for density as percentage of measured density correction for temperature deviation of 100 C Correction of the zero flow point in the range of 19 to 19 7 1 3 E68 169 6 Inner parameters of the flowmeter used for diagnostic purposes Measuring tubes oscillation frequency Current time delay between left and right signal coils Drv XXX YY 70 36 Drive coil load as ogee XXX and its standard deviation within 8 seconds as percentage YY 69 L XX mV R XX mV tro Voltage on the left L and right R signal coils Pred XXXXX XX US 71 36 Measuring tubes oscillation period 70 Ind XX X C Internal temperature of the flowmeter PUL XXXX X HZ 172 6 Frequency of the signal on the pulse output 11 CUR XX XX MA Current value on the current output MMM woo sna Water X XXX g mL 73 6 Density of water in oil water under working
38. enuitem 1 English default 2 Russian 1 Gpe default 2 ProLink RW 27 0012 Current output 0 Mass flow rate default 1 Volumetric flow rate 2 Density Modbus Baud rate 0 9600 bit s default 1 4800 bit s 2 2400 bit s 3 1200 bit s www gpeus com CORIOLIS MASS FLOWMETER USER MANUAL Pulse output 0 Mass flow rate default 1 Volumetric flow rate 2 Density Menu item O NO N O1 0017 RW Flow direction 19 0 Single direction default 1 Bidirectional Mass flow rate unit 0 g s 1 kg s 2 kg min 3 t day 4 kg h default 5 t h Density unit 0 giem default 1 kg L 2 kg m 0020 RW Temperature unit 0 C default 0 F Volumetric flow rate unit 0 mL s 1 L s 2 L min 3 m day 4 L h 5 m h default Mass total unit 1 kg 2 t default 0023 Volume total unit O mL 1 L 2 m default 0257 RO__ Calculated pressure from external pressure sensor Bar 0273 0275 0451 Input external pressure Bar 0521 RW Bytes sequence in floating point 32 bit registers 0 0 1 2 3 default 1 2 3 0 1 2 1 0 3 2 3 3 2 1 0 For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7 NIS gt N oO Q3 CO OO oO OO O1 O0 gt N O1 N O Q gt To write in registers the switches 2 4 6 8 must be in O
39. es from 4 to 20 mA and is proportional to the measured flow rate The current value of 4 mA corresponds to zero flow rate The current value of 20 mA corresponds to the upper limit of the full flow range of the flowmeter Qmax The current output can be configured to denote mass flow rate volumetric flow or density of the medium Current output is active The current signal parameters are presented in Table 1 5 Table 1 5 Current output signal parameters Current output signal Current value 4 20 mA Load resistance 250 600 Ohm 1 3 5 3 Digital output Digital interface complies with the requirements EIA TIA 422 B and recommendations RTU V 11 and provides the opportunity of networking and transferring of all measured parameters The digital interface specifications are presented in Table 1 6 Table 1 6 Digital interface specifications www gpeus com CORIOLIS MASS FLOWMETER USER MANUAL Digital interface Standard EIA RS 485 Data transfer protocol Modbus RTU Data transfer baud rate 1200 2400 4800 9600 bit s Maximum distance 300 m Data format 8 data bits 1 start bit no parity bit 2 stop bits default Data format can be changed in the menus 30 31 The following measured parameters can be transmitted through the digital interface mass volume flow rate mass volume density and temperature of the medium Digital interface can also be used to calibrate and to configure the flowmeter Flowmete
40. exceeding the maximum allowable pressure for the given modification connection of external circuits to the flowmeter must be implemented through the cable entries complied with GOST R 51330 1 unused cable entry must be closed with the end cap supplied by the flowmeter s manufacturer or other end cap complied with GOST R 51330 1 connection of the external devices to the pulse current and digital outputs of the flowmeters of Ex modification must be implemented in accordance with GOST R 51330 1 Explosion protection type of flameproof enclosure is implemented by putting the electrical parts of the flowmeter into the flameproof enclosure in accordance with GOST R 51330 1 which prevents the explosion from coming out of the flowmeter into the explosive environment Explosion protection of the enclosure is ensured by the following means the shell withstands the explosion test at the test pressure of 4 times the pressure of the explosion axial thread length and number of full turns of thread engagement comply with GOST R 51330 1 requirements the gaps and lengths of flat and cylindrical flameproof joints comply with GOST R 51330 1 requirements maximum flowmeter s surface temperature in working conditions must not exceed the temperature range in accordance to GOST R 51330 0 for the temperature groups T4 for the flowmeters with temperature modification code 100 T3 for the flowmeters with temperatur
41. ff X XXX kg h Reset Totalizer O0 SIG Option of direct SIG or reverse BID direction of flow If SIG is chosen then only direct flow BID MMM Flow Direction will be accumulated direction corresponding to the arrow on the sensor body If BID is chosen then flow will be counted on both direct and reverse directions wioo sna DEC 121 8 The way of counting reverse flow whether it will be added INC or subtracted DEC from the Reverse Flow Tol INC E edit accumulated total mass and volume Massinventory X XXX 22 8 Accumulated total mass This value is also displayed in Menu item 1 Flow Mass Volminventory XXXX XX 116 8 Accumulated total volume This value is also displayed in Menu item 2 Flow Volm Output Config 110 4 D oo Configuration of the outputs of the flowmeter Pulse Weight XXX XX g N a E Input mass flow rate volume flow rate or density corresponding to one pulse on the pulse output Pulse Output e The pulse output can be configured to denote mass flow rate Mass volumetric flow rate Volm j Dens or density Dens of the medium 20mA Value XXX X 2 f B Input mass flow rate volume flow rate or density corresponding to 20mA on the current output Current Output T The current output can be configured to denote mass flow rate Mass volumetric flow rate j Dens Volm or density Dens of the med
42. flow rate 0249 0250 RO Density 3 0251 0252 RO Temperature 3 0253 0254 RO Volumetric flow rate 2 0257 0258 Calculated pressure from external pressure sensor kgf cm wee astres mode senao Mena tem 0259 0260 RO Mass totalizer resettable et 0261 0262 RO Volume totalizer resettable Pp 0263 0264 RO Mass totalizer inventory 0265 0266 RO Volume totalizer inventory Coefficient of pressure compensation for flow PSI Coefficient of pressure compensation for density PSI 0271 0272 Flow calibration pressure kgf cm 0273 0274 Pressure relating to 4 mA kgf cm 0275 0276 Pressure relating to 20 mA kgf cm d 43 0285 0286 RO Measuring tubes oscillation frequency Hz 68 0287 0288 RO Left coil voltage mV 69 0289 0290 RO Right coil voltage mV 69 0291 0292 RO Drive coil load 69 0293 0294 RO Live mass flow rate zero point not counted kg h 0451 0452 Input external pressure kgf cm AS registers _ 0072 RW Calibration coefficient g s us 0073 0074 Format XXXXXX Example 23 350 K 23 35 g s us 0075 0076 RW Temperature correction coefficient for flow Kt 100 C 61 Format XXXX Example 5 00 Kt 5 0 100 C For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7 Reserved for future modifications www eus com
43. gnostics Lighting period Lights continuously right after power up Self diagnostics test failed Lights continuously some time later after Wrong zero calibration power up Flashing The color is red Malfunction of the flowmeter Flow rate is less than the lower limit of the flow range for this size of the flowmeter Flashing Lights within 3 4 of the period dark within 1 4 of the period To determine a fault in the sensor part of the flowmeter it is necessary to check the resistance of the coils first Resistance value should be in the ranges specified in Table 2 10 Table 2 10 Coils diagnostics Circuit Contact Resistance range Left coil Brown red 60 75 Right coil Orange yellow 3 4 60 75 Drive coil Blue green 5 6 6 30 lee IE Gray white 7 8 75 175 sensor Lane Gray black 7 9 75 175 sensor Reclamation act appearance and procedure of returning the faulty flowmeter and its warranty repair are descripted in the passport of the flowmeter www eus com i CORIOLIS MASS FLOWMETER USER MANUAL 3 TRANSPORTATION AND STORAGE 3 1 Transportation While transportation of the flowmeter it is recommended to follow these recommendations e The flowmeter should be transported in a transport container which should protect the flowmeter from mechanical damage e It is recommended to lay a waterproof paper inside the transport container e Transportation should be performed at ambient temperatures f
44. he pipeline of the medium temperature and pressure changing e no moving parts inside the flowmeter oscillation amplitude of the tubes is very small and they can be considered fixed and parts subject to wear which extends the life time of the flowmeter e no obstructions to the flow inside the flowmeter ability to measure temperature and density of the liquid e no need for straight pipe sections before and after the flowmeter as well as flow conditioning plates etc www gpeus com CORIOLIS MASS FLOWMETER USER MANUAL 1 3 Technical specifications 1 3 1 Technical parameters overview Brief description of the technical specifications of the flowmeter is presented in Table 1 1 Table 1 1 Technical specifications Parameter Value Note 1 3 2 Range of measurement Nominal diameter mm Accuracy Process pressure Medium temperature Explosion proof grade modification F1001 Ex Atmospheric pressure Environment temperature Relative humidity Resistance to the external magnetic field Vibration resistance Enclosure protection Verification period Temperature sensor Service life Outline dimensions Materials used 10 15 25 40 50 80 100 150 200 0 1 0 2 0 5 Up to 6 4 MPa Or up to 25 MPa special order 50 350 C 1Exd ib IICT6X transmitter 1ExibliC T1 T4 X sensor 84 0 106 7 kPa 40 55 C or 50 70 C special order 90 3 non con
45. hould be provided only on one side of the cable near the power supply e lt is not recommended to lay the signal cables in the same conduit with power wires and also near sources of strong electromagnetic fields If necessary grounding of signal wiring can be done at any point in the signal circuit For example it is possible to ground the negative terminal of the power supply The electronics housing is grounded to the sensor housing 2 4 4 Waterproofing The flowmeter meets all the requirements for moisture protection grade specified in the Technical Specifications section In order to ensure the required protection grade after the installation or maintenance of the flowmeter the following requirements should be fulfilled see Figure 2 12 e Seals in the transmitter should not have dirt or damage If necessary clean or replace the seals It is recommended to use the original sealing elements from the manufacturer e Electrical cables should fit the cable entry and must not be damaged e Covers and the cable entries must be tightened firmly e Unused cable entries must be plugged e Just before the cable entry the cable should have U shaped loop to prevent any liquid from going into transmitter when the liquid flows down the cable Do not install the transmitter with the cable entry directed vertically upwards Figure 2 12 Recommendations for orientation of cables and cable entries www gpeus com EER CORIOLIS MASS FL
46. ipeline do the following steps see Figure 2 5 e Arrange the flowmeter so that the arrow on the sensor s body corresponds to the normal direction of flow e Slide the bolts through the holes of one of the pipeline flanges and flowmeter s flange put washers and nuts Nuts should not be finally tightened yet e Put the gasket between the pipeline flange and flowmeter s flange and align it It is recommended to avoid protrusion of the gasket into the pipeline opening e Install the gasket between the other pair of flanges slide the bolts through the holes of the flanges put washers and nuts Nuts should not be finally tightened yet e Tighten the nuts in the sequence shown in Figure 2 6 When installing the flowmeter bending and torsional load on the joints and also mated flanges misalignment should be minimized Figure 2 5 Installation of the flowmeter in the pipeline www qpeus com 1 CORIOLIS MASS FLOWMETER USER MANUAL Table 2 3 Notes for Figure 2 5 Mounting part 1 Flowmeter flanges Connection kit flanges 2 3 Gaskets 4 Nuts 5 Washers 6 Bolts or stud bolts Figure 2 6 The sequence of tightening the flange bolts Transmitter of the separate type of the flowmeter can be mounted as shown in Figure 2 7 Transmitter can be mounted with brackets or clamps to the rack pipe or wall Figure 2 7 Installation of transmitter of separate type modification www qpeus com
47. it of the full mass flow range kg h Relative basic error of measurement of volumetric flow volume on current output signal v calculated as dv dy 0 2 Imax 44 16 Qy Qvmax 1 4 where Relative basic error of measurement of volumetric flow volume Imax 20 MA maximum value of current output signal Qy measured volumetric flow rate L h Qymax upper limit of the full volumetric flow range L h Additional error of measurement of mass volumetric flow rate caused by a change of medium temperature is 0 05 of the maximum flow rate for every 10 C of deviation from the zero calibration temperature Additional error of measurement of mass volumetric flow rate caused by a change of pressure is 0 02 of the maximum flow rate for every 100 kPa of deviation from the zero calibration pressure The effect of changes in temperature and pressure can be adjusted by zero calibration under the actual pressure and temperature see paragraph 2 5 4 Zero point adjustment 1 3 4 Power supply Electrical power of the flowmeters depending on its voltage modification is provided by an external DC 24V power supply or AC 220 V with frequency 50 1 Hz Parameters of the power supply are presented in Table 1 3 Table 1 3 Power supply parameters Nominal voltage Voltage range Maximum power consumption www gpeus com RA CORIOLIS MASS FLOWMETER USER MANUAL DC 24 V 18 30 V 15 VA AC 220 V
48. ium MODBUS XXX 129 9 Address of the flowmeter in Modbus network Address E edit 4800 2400 Data transfer baud rate in Modbus network Baud Rate Parity control in data transfer protocol Select NO for no parity Even for 1 even parity bit Odd for 1 odd parity bit Stop Bits ene 124 gt 9 Two E edit Select one or two stop bits in data transfer protocol in Modbus network Parity Bit MMM wioo sna Unit Config 111 4 Press E to enter Select units for parameters displayed in the menu items 1 2 3 it Mass Unit g kg t ant Select units for mass totalizer MassFlow Unit g s kg s kg min yee Select units for mass flow rate t min t h E edit Volume Unit mL L m nes Select units for volume totalizer mL s L s L min 136 10 VolumeFlow Unit min L h m h Select units for volumetric flow rate Density Unit g mL kg L ae Select units for density Temperature Unit C F jaa Select units for temperature PressureComp 112 4 er Press to enter Switching On Off and changing parameters of pressure compensation for flow Pressure compensation modes Select None to disable pressure compensation select Manu None l for manual compensation when the actual pressure value is entered manually in the Menu item 41 PressComp Mode Manu i Auto select Auto for automatic pressure compensation when actual pressure is measured by an
49. l information The addresses in the tables below are decimal the starting address is 1 For example address 127 corresponds to hexadecimal address 0x00 7EH 126 decimal Read Write mode WO write only RO read only RW read and write Coils Functions Read 01 Write 05 Read 0 Off 1 On Write Ox0000 Off OxFFOO On L nd ress Mode Desemptn Menuiem 0002 RW Start Stop totalizers 0 Stop totalizers 1 Start totalizers 0003 RW Reset totalizers resettable 0 Abort 1 Reset totalizers 0004 RW Reset totalizers inventory 0 Abort 1 Reset totalizers 0005 Zero calibration 16 0 Abort 1 Start zero calibration 0056 Reset mass totalizer resettable 0 Abort 1 Reset mass totalizer 0057 RW Reset volume totalizer resettable 0 Abort 1 Reset volume totalizer ape Pressure compensation 0 Off default 1 On For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7 To write in registers the switches 2 4 6 8 must be in ON position 16 bit registers integer values 2 bytes high byte first AERE TA 04 Write 06 ee La map version 1 GPE default 2 ProLink 0016 RO Transmitter software version Format XXX X S www gpeus com 1 CORIOLIS MASS FLOWMETER USER MANUAL 0039 RW Mass flow rate unit 33 70 g s 13 kg s 74 kg min 77 t d
50. lation and maintenance e Display of the flowmeter should be installed at a place which is convenient for the operator to read displayed data e The installation place of the flowmeter should be selected to ensure a minimum temperature of the transmitter s surface In direct sunlight the surface temperature may rise by up to 30 degrees compared to the ambient temperature so if installation in the shade is not possible it is recommended to use a sunshade Figure 2 1 Basic requirements for the installation place CAUTION If there is a strong vibration or pipeline strained at the installation place it is necessary to provide external support for the pipeline before and after the flowmeter The support base must be firm In this case installation of the flowmeter in such places including mobile units is permitted www gpeus com CORIOLIS MASS FLOWMETER USER MANUAL 2 3 2 Orientation Flowmeter can be installed in horizontal vertical or inclined sections of the pipeline It is optimal to install the flowmeter in horizontal section The flowmeter should be installed so that its measuring tubes are always filled with fluid and the arrow on the sensor body coincided with the flow direction In these conditions the flowmeter will operate properly in any orientation Flowmeter does not require collateral straight pipeline sections before and after the flowmeter nor installation of additional equipment to level the flow
51. lue CRC address quantity count Response Slave address Starting address Registers quantity 41H 00H 00H 02H BFH 16H Error WWW eus com i CORIOLIS MASS FLOWMETER USER MANUAL General information The addresses in the tables below are decimal the starting address is 1 For example address 127 corresponds to hexadecimal address 0x007EH 126 decimal Read Write mode WO write only RO read only RW read and write Coils Functions Read 01 Write 05 Read 0 Off 1 On Write 0x0000 Off OxFFOO On address Mode Desorpton Menuitem 0001 WO Output emulation 12 1 Start output emulation 0003 WO Reset totalizer resettable 18 1 Reset totalizer Reset totalizer inventory a 1 Reset totalizer Zero calibration 16 1 Start zero calibration 0009 RW Reverse flow total count mode 20 0 Subtract from the totalizer default 1 Add to the totalizer Oil water analyze pure oil counting 0 Off default 1 On RW 81 0013 Multipoint correction 0 Off 1 On 0082 RW Pressure compensation 39 0 Off default 1 On For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7 To write in registers the switches 2 4 6 8 must be in ON position 16 bit registers integer values 2 bytes high byte first Functions Read 04 Write 06 address Mode Description m
52. manual This condition is necessary to ensure safe operation and proper functioning of flowmeters For advice contact your local agent of GPE or the support service Tel Fax 1 646 619 1289 CAUTION This manual applies only to F1001 Series Coriolis Mass Flowmeters This document does not cover other products of GPE and the products of other companies CORIOLIS MASS FLOWMETER USER MANUAL Table of contents 1 PRODUCT DESCRIPTION AND TECHNICAL DATA 2 OPERATION AND MAINTENANCE 3 TRANSPORTATION AND STORAGE 4 VERIFICATION 1 Applications 2 Principle of operation 1 3 Technical specifications 1 3 1 Technical parameters overview 1 3 2 Range of measurement 1 3 3 Accuracy of measurement 1 3 4 Power supply 1 3 5 Output signals 1 3 5 1 Pulse output 1 3 5 2 Current output 1 3 5 3 Digital output 1 3 6 Display 1 4 Pressure drop 1 5 Explosion protection 1 6 Delivery set 1 7 Model codes ok 2 1 Model selection recommendations 2 2 Safety precautions 2 3 Installation on pipeline 2 3 1 Installation location 2 3 2 Orientation 2 3 3 Pipeline preparation 2 3 4 Installation 2 3 5 Thermal insulation 2 3 6 Cooling 2 3 Transmitter rotation 2 4 Wiring 2 4 1 Basic operations 2 4 2 Installation with explosion protection 2 4 3 Wiring recommendations 2 4 4 Waterproofing 2 4 5 Grounding 2 5 Operation and maintenance 2 5 1 Basic recommendations 2 5 2 Power Up 2 5 3 Display
53. nts of flow rate time delay Restore Para 183 7 Restore of all coefficients and settings of the flowmeter except for calibration parameters in the Press E to enter E edit menu items 60 67 to factory defaults 81 MultipointCorrect 83 Correction Point X a a Number of correction coefficient 0 7 which values are specified in the menu items 84 and 85 84 No X TimeValue yy yyy 185 7 The value of time delay between left and right signal coils YY YYY us for the correction coefficient E edit number X 85 No XCorrectValue XX XXX 176 7 Correction coefficient in the range of 0 9 to 1 1 applied to the calibration coefficient K for time delay E edit point specified in the menu item 84 MMM woo snad i CORIOLIS MASS FLOWMETER USER MANUAL After installation and electrical connection a zero point adjustment zero calibration procedure should be performed This procedure sets the base point corresponding to zero flow Without actual flow there is a certain time delay between signal coils anyway that delay assumed as zero point To perform this procedure follow these steps 2 5 4 Zero point adjustment e Apply power to the flowmeter and let it warm up for at least 30 minutes e Let the measuring medium flow through the flowmeter as long as the thermal balance is established Close the valve located after the flowmeter e Make sure that the flowmeter s tubes are completely fill
54. operating 2 5 4 Zero point adjustment 2 5 5 Protection switch 2 5 6 Maintenance 2 5 7 Troubleshooting 3 1 Transportation 3 2 Storage 3 3 Utilization Appendix A Outline dimensions and weight Appendix B Wiring diagrams Appendix C Modbus register map version 2 xx Appendix D Modbus register map version 3 xx ProLink www gpeus com EER CORIOLIS MASS FLOWMETER USER MANUAL 1 PRODUCT DESCRI PTI ON AND TECHNI CAL DATA 1 1 Applications WWw gpeus com The flowmeter is designed to measure the mass and volume flow density mass and volume of liquids and use the gathered information for technological purposes or commercial accounting The flowmeter is used as the counter of gasoline liquefied petroleum gas kerosene diesel fuel oil oil water and other liquids in the chemical petrochemical oil food pharmaceutical and other industries and public municipal facilities The flowmeter is used in technological processes automatic monitoring and control systems in various industries for stationary technological plants land mobile refueling and pumping equipment and in commercial accounting systems The flowmeter is designed for use in explosive safe and explosive environment The flowmeter of explosion proof modification GPE F1001 Ex has a combined type of protection flameproof enclosure complied with GOST R 51330 1 and the input and output intrinsically safe level ib complied with GOST R 51330 10
55. oportional to the amount of fluid passing through the tube within the certain time Electromagnetic speed detectors situated on both inlet left and outlet right side of the tube measure the tubes oscillations frequency Mass flow rate is determined by measuring the time delay between the signals of those detectors When there is no flow the tubes dont twist and there is no time delay between left and right detectors signals Mass flow rate Q is proportional to the time delay At between the detectors signals Q K At 3 6 kg h 1 0 where K calibration coefficient g s us At time delay between the detectors signals us The density of the medium is determined by measuring the period of oscillation of the flow tubes which is proportional to the density of the medium The density measuring channel is calibrated for two different kind of medium with exactly known measured by standard density meter density water and air For each medium the flowmeters shows the flow tubes oscillation period corresponding to the medium density The measured value of the oscillation period is shown in Menu item 70 see the display menu The medium density and the corresponding oscillation period for both water and air are entered in the menu display Menu items 62 65 or through Modbus Due to the linear dependency of the oscillation period of the density the actual medium density can be determined by the oscillation period measured by the flowmeter
56. pe Transmitter Connection a Inside connection for Integral type Special cable for separate type Intrinsically safe type Sensor Fuse Power transformer 3 e Signal output circuit Optical Coupler Reliable Power transformer Fuse HVAC 24VDC m Detect circuit Figure 1 3 Block diagram of the flowmeter Input and output parameters of intrinsically safe circuits are shown in Tables 1 8 and 1 9 Table 1 8 Input parameters of intrinsically safe circuits Parameter value Parameter name Left right Temperature Drive coil signal coils sensor Maximum input voltage Ui V 5 4 10 5 5 4 Maximum input current li mA 72 70 72 Maximum input capacity Ci pF 50 50 50 Maximum input inductance Li mH 2 2 3 5 0 010 Maximum input power Pi W 0 097 0 184 0 097 www qpeus com EER CORIOLIS MASS FLOWMETER USER MANUAL Table 1 9 Output parameters of intrinsically safe circuits Parameter value for the circuit of Parameter name Left right coil Drive coil Temperature power power sensor power Maximum output voltage Uo V 5 4 10 5 5 4 Maximum output current lo mA 72 70 72 Maximum output capacity Co UF 10 1 10 Maximum output inductance Lo mH 5 4 5 5 Maximum output power Po W 0 097 0 184 0 097 Parameters of the sensor coil windings are shown in Table 1 10 Table 1 10 Coil windings parameters Left right signal coils 0 13 20 0 5 Drive coil DN10 DN40 0 13 300 11 0 5 Drive coil
57. profile flow conditioning plates etc But if two or more flowmeter are installed in the same section of the pipeline the distance between them should be at least 2 m Installation recommendations are presented in Figure 2 2 b u Hi 1 Figure 2 2 Recommendations for installation Table 2 2 Notes for Figure 2 2 Recommendations It is recommended to install the flowmeter tubes downward to improve their filling with fluid and to prevent accumulation of gas in them For vertical or inclined pipeline orientation it is recommended to install the 2 flowmeter in a pipeline section with the upward flow direction to improve filling of the pipe with fluid In the event of bending pipe it is recommended to install the flowmeter in the lower section of the pipeline Do not install the flowmeter on a horizontal pipe before the drainage section with free flow because in this case filling the flowmeter s tubes wit fluid is not guaranteed 2 3 3 Pipeline To prepare for the installation of the flowmeter it is necessary preparation e Check the completeness of the connection kit and conformity of all mounting parts to the flowmeter s modification e Cut the pipeline section of length Linst Linst Lm 2 Lg 2 Lf 2 1 where Lm flowmeter length see Appendix Lg gasket thickness Lf connection kit flange thickness minus depth of landing on the pipeline e Install connection kit flanges on the pip
58. r supports two versions of the Modbus register map e GPE register map version 2 xx is supported by the GPE Integrator software and set as default at factory e ProLink register map version 3 xx compatible with ProLink Il software Register map and description for GPE version 2 xx are presented in the Appendix for ProLink version 3 xx in the Appendix Register map version switching can be performed by the GPE Integrator or changing the value of the corresponding Modbus register see Appendix using third party serial port tools 1 3 6 Display The flowmeter s OLED graphic display contains 4 lines with 16 symbols per line The following measured parameters can be displayed Mass flow rate Volumetric flow rate Medium density Medium temperature Total mass Total volume Display operation is provided by the means of three optical sensor buttons below the display Display operation described in the paragraph 2 5 3 Display operating 1 4 Pressure drop The pressure drop on the flowmeter AP at maximum flow rate pressure and temperature is not more than 0 13 MPa for water medium In the process of measurement flow of liquids it is necessary to consider the cavitation effect liquid boiling which may occur in certain conditions of the flow Cavitation may cause the flowmeter work abnormally To prevent this it is necessary to keep the certain pressure at the distance of 5 DN after the flowmeter
59. reliable operation of the flowmeter and to obtain reliable measurement results is correspondence of the flowmeters modification to the process parameters The list of the process parameters required for optimal flowmeter modification selection is presented in Table 2 1 Table 2 1 Process parameters for modification selection Item No Process parameter Measured medium name Composition and percentage of liquid components Composition and percentage of solid impurities in fluid Composition and percentage of gas inclusions in fluid Measured medium density Measured medium viscosity Flow range CON O o A OO N gt Required accuracy 9 Measured medium temperature 10 Pipeline pressure 11 Allowable pressure drop 12 Presence of regulation and control components in the system 13 Pipeline diameter 14 Pipeline orientation at the installation place 15 Ambient temperature 16 Explosion protection grade and requirements CAUTION To avoid erroneous self dependent selection of the modification of the flowmeter please send the completed questionnaire to the nearest GPE company representative Flowmeter size should be selected according to the actual flow rates in the pipeline which may differ from the calculated design values Flowmeter size should be chosen so that the actual flow rate of the medium was in the second third of the flow range Therefore nominal diameter DN of the flowmeter can be either equal
60. rom 40 to 70 C with relative humidity up to 100 non condensing at 35 C e Protection against atmospheric precipitation must be provided e Flowmeters can be transported by all kinds of roofed transport including air transport in heated hermetic compartments in accordance with the rules in force for this type of transport e The requirements in the package handling marks should be fulfilled e Method of stacking the boxes on the transport vehicle must prevent their movement e The boxes must not be the object to sharp blows while loading unloading e Period of stay in the transportation conditions must be not more than 3 months e After transportation at a temperature below 0 C container with the flowmeter should be unpacked at least 12 hours after its storage in a warm room When the flowmeter is transported out of package it is necessary to follow recommendations given in Figure 3 1 Figure 3 1 Out of package transportation rules www gqpeus com A CORIOLIS MASS FLOWMETER USER MANUAL 3 2 Storage Flowmeters can be stored in unheated rooms with air temperature of 20 to 70 C and relative humidity up to 95 non condensing at 25 C Flowmeters can be stored in a transport boxes stacking of up to 3 boxes in height and also without package For the long term storage it is recommended to use the manufacturer s package 3 3 Utilization Flowmeters are free of harmful substances and components that are hazardous to
61. self test and if it succeeds the flowmeter begins to measure flow to generate output signals and display the measured values Controlling of the flowmeter s display is performed via three buttons of optical type below the display Thus display operating doesn t require opening the front cover of the transmitter which is important in explosive environment in conditions of high humidity or precipitations and other conditions with a high probability of contamination of the internal structural elements of transmitter or ingress of moisture liquids foreign objects etc To push the optical button one should briefly put a finger or other opaque object close to the display glass in the area of the button At that time the LED next to the display lights for a moment Flowmeter can display the measured parameters shown in Table 2 6 To change the display page next pair of parameters press the button Display sequence corresponds to Table 2 6 Table 2 6 Displayed parameters sequence Displayed parameter Display format Mass flow rate Flow XXX XXX Mass total Mass XXX XXX Volume flow rate Flow XXX XXX Volume total Volm XXX XXX Density Den XXX XXX Temperature Temp XX X If the displayed value becomes more than 999 999 the units are automatically switched to bigger ones to be able to display that value If no button is pressed within two minutes flowmeter is automatically switched to display mass or volume flow rate Menu i
62. set its value in accordance with the requirements of this manual Check the output wires connection according to the wiring diagrams in Appendix Check the computer s port used to connect the flowmeter to the computer through digital interface Make sure that the same port number is set in the Gpe Integrator settings Fully open the valves to set the flow rate within the measuring flow range Decrease the low flow cutoff 1 CORIOLIS MASS FLOWMETER USER MANUAL outputs value set in the menu value to be less than the actual flow rate Probable reason Repairing way Deviation of the zero point because of temperature and pressure deviation from temperature and pressure values on previous zero Calibration 4 When there is no actual flow the flowmeter indicates a certain flow Perform zero calibration in accordance with the paragraph 2 5 4 Zero point adjustment 5 In the menu it is not possible Atleast one of the switches Set all the switches to OFF to enter the of the protection switch positions see Figure 2 15 for Calibration block is set to the ON the time needed to perform and System position changes in those sections Para sections special LED indicator is provided for indication in case of some faults of the flowmeter LED is located over the display and its color and lighting period depends on the kind of the flowmeter s fault according to Table 2 9 Table 2 9 LED dia
63. st 2 5 mm2 AWG13 should be used for grounding Unused cable entry must be closed with the end cap supplied by the flowmeter s manufacturer or other end cap complied with GOST R 51330 1 During installation check the explosion proof surfaces of mounting parts involved in providing explosion protection Scratches dents chips on the explosion proof surfaces of those parts which are in the explosion path are not permitted After the completion of the electrical connections it is necessary to close the transmitter covers firmly and put the latch lock on the covers www gpeus com A CORIOLIS MASS FLOWMETER USER MANUAL 2 4 3 Wiring During electrical connection of the flowmeter you should follow these recommendations recommendations e Wire cores must be cleaned and fixed on terminals to prevent a short circuit between each other and the body of the flowmeter ground e t is recommended to use separate power supplies or a multichannel power supply with isolated channels to power the flowmeter and each of its outputs e f it is necessary to calculate the load resistor value the total load impedance should be calculated as the sum of the cable resistance external load resistance resistance of safety barriers load resistance of the secondary equipment e To minimize interference when transmitting of analog output signal 4 20 mA and digital signal it is recommended to use a shielded twisted pair cable Grounding of the cable s
64. ted 04H Failed to fetch Slave ID Minimal query period is 32 us Recommended period is 125 us www eus com i CORIOLIS MASS FLOWMETER USER MANUAL Example of queries Read Reverse flow total count mode Query Slave address Starting address Coils quantity paw om Lo om eon con an cE Response Slave address Byte count Coil status OOH 44H Error Result Reverse flow is subtracted from the totalizer values Reset totalizer resettable Query pan em D om un mn Perform zero calibration Query pate ose oo ow ree oon om oe Response 05H OOH 04H FFH OOH C3H Error www gpeus com CORIOLIS MASS FLOWMETER USER MANUAL Read mass flow rate Query Slave address 41H 04H 00H A7H OOH 02H CEH E8H Function Starting address Registers quantity CRC Response Slave address Function Data value CRC 41H 04H 04H 43H B4H 74H DOH Byte count Error Slave address Function Error code CRC 41H 84H 01H 83H 14H Result mass flow rate is 360 9126 kg s Write mass unit of kg Query Slave address Function Register address Data value CRC 41H 06H OOH 15H OOH 01H 57H OEH Response Slave address Function Register address Data value CRC 41H 06H OOH 15H OOH 01H 57H OEH Error Slave address Function Error code CRC 41H 86H 01H 74H Write calibration coefficient K 1100 g s us Query Slave address Function Starting megisiels eye Data va
65. tems 1 or 2 depending on the state of the menu item 47 First Menu The flowmeter can be configured via display menu The structure of the menu is shown in Figure 2 14 description of the menu items presented in Table 2 7 In the editable menu items press E button to switch to edit mode e lf the menu item has several predefined options press or to move between those options The currently selected option is flashing To accept selection press E The message Saved Y N about saving changes will appear Press or to move between Y yes and N no options Select Y to accept changes or N to cancel then press E to quit edit mode e lf the menu item has numeric value its editing is performed by changing the digits one by one and moving to the next digit on the right The digit in currently selected position is flashing Press to increment the digit in current position from 0 to 9 To move to the next digit press Press E to finish editing then select Yes to accept changes or No to cancel then press E to quit edit mode www gpeus com MMM S A O O 3 Root menu 1 Mass xxx xxx t Flow xxx xxx t h EL Volm xxx xxx m3 Flow xxx xxx m3 h 2a C 1 Water XX X IOil XXX XXXt 4 Den xxx xxx g mL Temp xx xx C El 4 E Configuration Press E to enter
66. ve address Function Error code CRC po Exception code 01H Function is not supported 02H Incorrect Coil address 03H Data value is out of range of Ox0000 OxFFOO 04H Failed to write single coil Function 06 Write Single Register Query D ee a a Exception code 01H Function is not supported 02H Incorrect Register address 03H Data value is out of range of 0 65535 04H Failed to write single register Function 08 Diagnostics supports only subfunction code 00 Query Slave address Subfunction CRC Response Error www gpeus com a CORIOLIS MASS FLOWMETER USER MANUAL Slave address Function Error code CRC o CS PS ES Exception code 01H Function or subfunction is not supported 03H Incorrect Data value 04H Failed to diagnose Function 16 Write Multiple Registers Query Slave address Function Starting Registers Byte Data value CRC address quantity count ook oh o l an Slave address Starting address Registers quantity o EC o o Error pam fom a Exception code 01H Function is not supported 02H Incorrect Starting address or Registers quantity 03H Registers quantity N is out of range of 1 123 or Byte count is not equal to N 2 04H Failed to write multiple registers Function 17 Report Slave ID Query Response Error CRC AE A ES Exception code 01H Function or subfunction is not suppor