ProTime Microcoagulation System - Accriva Diagnostics

image

Contents

1. 16 memory clear 30 month setting 27 OID entering and changing 24 setting 25 password changing 30 entering and changing 24 performance characteristics accuracy 33 comparisons 33 patient self testing 33 precision 32 sensitivity 33 PID entering and changing 24 setting 25 preparation contents 9 unpacking and inspection 9 professional guidelines reporting results 31 programming mode 27 quality control 33 reporting 31 expected values 32 reportable range 29 safety featu
2. 14 34 35 test methodology 2 test procedures 10 therapeutic range setting 28 TR limits setting 28 troubleshooting 14 year setting 27 IR5246 1 08
3. 35 SAFETY STANDARDS 36 INDEX 39 This manual is published by International Technidyne Corporation ITC for use with the ProTime Microcoagulation System and Tenderlett Plus fingerstick blood collection device Questions or comments regarding the contents of this manual can be directed to the address at the back of this manual or to your ITC representative Please read the instructions before use 2000 2001 2002 2003 2004 2005 This document is the copyright of ITC and must not be copied or reproduced in any form without prior consent ITC reserves the right to make technical improvements to equipment and documentation without prior notice as part of a continuous program of product development 0123 2 INTENDED USE The ProTime Microcoagulation System is a portable battery operated instrument with a disposable cuvette for quantitative determination of prothrombin time from fingerstick whole blood or anticoagulant free venous whole blood The product is intended for use in the management of patients treated with oral anticoagulants by a healthcare professional and for patient self testing Additional information for medical professionals to monitor patients is located at the end of this instruction manual Materials are available through ITC for professional
4. CHARGE ERROR TURN OFF TRY AGAIN Power supply error Check the AC DC power module repeat the test If problem persists call ITC Technical Support TEMP ERROR TURN OFF TRY AGAIN Temperature not in range Check for proper operating temperature Repeat the test If problem persists call ITC Technical Support PHOTO ERROR TURN OFF TRY AGAIN LED blocked or other photo system error Repeat the test If problem persists call ITC Technical Support MAIN MENU OPTIONS The options presented in the MAIN MENU are Each of these options will lead to sub menus within the selection The following buttons are used to navigate the menu The button is used to move the highlight bar to select the option The button is used to select the option that is highlighted 17 Run Test To run the test select the RUN TEST menu item The instrument does a SELF CHECK procedure which may take up to 60 seconds The following screen is displayed for this period of time The test sequence continues as described in the Test Procedure section When the test completes the result will appear as one of the following screens depending on PID OID selections see SET UP below Note X is used for illustrative purposes only in the following examples If PID is ON and OID is ON If PID is OFF and OID is ON If PID is ON and OID is OFF If PID is OFF and OID is OFF 18 RUN LQC When RUN LQC is selected from the MAIN MENU the f
5. It takes only a few seconds for ProTime to draw the blood into the cuvette Watch the screen for the next prompt 7 Remove Tenderlett Plus Remove Tenderlett Plus immediately when prompted to do so CAUTION Failure to do so will result in an error message ProTime allows you six seconds Do not press the button after the Tenderlett Plus is removed from the ProTime instrument while you are testing your INR This will interrupt the test procedure and you will have to start over with a new blood sample The instrument then progresses to the test and displays the TESTING screen 14 8 Read the Result In a few minutes the result is ready Press the button to turn off Press the button to go to the MAIN MENU if you want to run another test review the data in memory print results transfer results to a computer or perform set up functions Notes The result remains displayed for 5 minutes or until the button or the button is pressed If the cuvette has not been removed and the instrument is left unattended for 5 minutes the instrument will display the following messages before shutting down and powering OFF What Does the Result Mean The result indicates the clotting activity of blood When ProTime is used as a self testing instrument the healthcare provider may program ProTime with the upper and lower limits that are right for the patient In this case ProTime will display OTR if results are out
6. connecting cables or other components have been removed The materials that are provided are listed below Note Inspect each component for damage when unpacking If damage is observed contact your ProTime representative Materials Provided Article Quantity ProTime Microcoagulation Instrument 1 ProTime Microcoagulation System Information and Training CD PROTIMEPRO only 1 AC DC power module ITC Part No IR5226 1 ProTime Microcoagulation Operator s Manual 1 Note An AC power cord is provided in the United States and Canada only For use outside the US and Canada the customer must obtain a power supply cord that is internationally harmonized and marked lt HAR gt 2 conductor 0 75 mm2 minimum wire rated 300 V with a PVC insulated jacket The cord and plug cap must be suitable for medical use The cord must have a molded on plug cap rated 250 V 2 5 A Materials Needed But Not Supplied ProTime cuvettes Tenderlett Plus incision device Optional Materials Article Personal Computer Interface Cable ITC No PROCABLE ITC Part No IR5313X Printer Label Maker ITC No LBLKIT Replacement Printer Label Maker Interface Cable ITC No LBLCABLE ITC Part No IR5314X Replacement Battery ITC No PROBATTERY 10 Charging the Battery The battery in the instrument needs to be charged before the first use 1 Connect the AC DC power module to an electrical service outlet using the AC power c
7. AC power cord and the ProTime instrument The screen will show CHARGE COMPLETE when the battery is fully charged The AC DC power module should be disconnected after the CHARGE COMPLETE message is seen The AC DC power module that has been supplied by ITC has been selected specifically for use with your ProTime Microcoagulation System Do not use any other AC DC power module When the battery indicator on the screen shows less than 25 charge remaining please refer to the Charging the Battery section for instructions To maximize battery life allow your ProTime instrument to discharge completely before re charging Avoid charging the ProTime instrument for frequent short periods of time such as charging for a few minutes removing from the AC DC power module and then recharging again Avoid storing and charging the ProTime instrument at extreme temperatures Built in ProTime Battery Features The ProTime instrument will shut off after 5 minutes if left unattended in order to preserve the battery charge If the instrument shuts off automatically press the button to re start the instrument Before the start of each test the ProTime instrument checks the amount of battery charge If there is not enough charge in the battery to run a test the ProTime screen will display PLEASE CHARGE IT If this occurs follow the charging instructions located in the Charging the Battery section Whether or not the batteries are charged you may continue to run tests
8. Then ProTime displays lt 0 6 INR LOW 0 6 0 79 INR lt 0 8 0 8 7 0 Result 7 1 9 9 Result followed by 10 0 12 0 INR gt 9 9 gt 12 0 INR HIGH 30 CHANGE PASSWORD If CHANGE PASSWORD option is selected from the PROGRAM MODE menu the user must enter the password up to 6 digits as a left justified number without any blanks The rules for entering a numerical PID or OID are also applicable for entering the 6 digit numerical password see page 24 After password has been entered the CONFIRMATION OF PASSWORD screen appears If the button is pressed the newly entered password is accepted by the instrument If the button is pressed the user will repeat the change password procedure by returning to the previous CHANGE PASSWORD screen Note Upon manufacturing the default password is set to 000000 CLEAR MEMORY Upon this selection from the PROGRAM MODE menu all results will be erased from the instrument memory To prevent accidental loss of data the following confirmation screen will precede the data removal from the instrument memory Selecting NO will return the instrument to the MAIN MENU and no results will be erased from the instrument memory 31 PERFORMANCE CHARACTERISTICS Reporting ProTime Results The ProTime Microcoagulation System reports a result as International Normalized Ratio INR and in PT seconds The ProTime system calculates the INR directly from whole blood clotting time
9. Note In the case of a 24 HOUR selection AM PM acronyms will not appear on the screens 23 Changing the Hour The time is pre set to Eastern Standard Time EST For example the time is 1 25 PM Change the hour by pressing the button until the correct hour appears in the highlight bar In this example the correct hour is 10 AM To change the hour press the button until 10 AM appears in the highlight bar The PM changes to AM at midnight Once the correct hour appears press the button to set the hour The procedure will then advance to the MINUTES screen Changing the Minutes The hour on the top line has changed to the time just set and PM has changed to AM Change the minute by pressing the button until the correct minute appears Once the correct minute appears press the button to set the minutes The procedure will then automatically advance to the DAY screen Changing the Day Only the day can be changed The month and year can be changed only in PROGRAM MODE Change the day by pressing the button until the correct day appears Once the correct day appears press the button to set the day 24 SET PID OID The PID OID selection is indicated only by the location of the highlight text on the screen is unchanged Use the button to select PID OID options Selecting PID OID ON enables both a patient ID and operator ID to be entered Selecting PID ON enables only a patient ID to be entered Selecting OID ON e
10. Upon selecting THIS RESULT the last recorded test is sent from the instrument memory to a printer or to a computer If ALL RESULTS is selected all results up to 50 are printed sent from the instrument memory to a printer or to a computer If MAIN MENU selection is made the user is returned to the main menu Note If the AUTO SEND feature is on results will be automatically sent if the PROCABLE is connected prior to the start of a test To use a laser printer the data must first be transferred to an IBM compatible personal computer and the results can be printed from that computer The following screen displays until the PRINT DATA or SEND DATA process is complete 22 SET UP Go to the MAIN MENU screen and choose the SET UP option by using the button to scroll and the button to select SET UP The following screen appears SET LANGUAGE Press the button to select SET LANGUAGE The following options are displayed when SET LANGUAGE is selected Highlight a language selection and press the button to set it Note English will be highlighted as the default language in a new instrument After changing the language the new language becomes the default The instrument will automatically turn off after changing the language selection SET TIME DAY Select SET TIME DAY from the SET UP menu The following screen is displayed 12 HOUR MM DD will be highlighted as a default selection Highlight your TIME DAY preference then press the button
11. from human plasma or serum that has been tested using US FDA recognized methods and found to be non reactive for HIV antibody and for hepatitis B surface antigen Handle with appropriate care and dispose of cuvettes and blood collection materials in accordance with standard methods of biohazard control The use of accessory equipment e g printers etc not identified in this manual either in the patient vicinity or that does not comply with either the equivalent safety requirements of this equipment or UL IEC 60601 1 or IEC 60601 1 2 may lead to a reduced level of safety with the resulting system Limitations The ProTime instrument uses only fresh capillary or venous whole blood Plasma or serum cannot be used Glass tubes or syringes must not be used to collect venous samples Use only plastic syringes without anticoagulants to collect venous samples Poor fingerstick blood collection technique may affect results In clinical trials no significant difference was observed between fingerstick and venous specimens run on ProTime During those trials ProTime measured patients with an INR range of 0 8 to 7 0 ProTime is programmed to calculate and report INR results as follows If the calculated INR is Then ProTime displays lt 0 6 INR LOW 0 6 0 79 INR lt 0 8 0 8 7 0 Result 7 1 9 9 Result followed by 10 0 12 0 INR gt 9 9 gt 12 0 INR HIGH Results may be affected in patients receiving heparin or wh
12. Cuvettes Tenderlett Plus ITC Catalogue Number of Device Do Not Reuse Single Use Only Upper and Lower Temperature Limitations For Storage or Use For in vitro Diagnostic Use Attention Read Accompanying Documentation or Instructions Consult Instructions for Use Class II Protection Against Electrical Shock Input Port for DC Power Cord from AC DC Power Module Polarity VDC and A Input Output Port for Data Transfer Name and Address of Manufacturer 5 PRINCIPLES OF OPERATION The ProTime Microcoagulation System measures the PT using fibrin clot formation and detection The ProTime cuvette is a self contained micro volume reaction cell constructed of precision molded plastic There are two user options within the ProTime Microcoagulation System the standard ProTime cuvette and the ProTime3 cuvette These cuvettes differ from each other in the amount of blood that needs to be collected and tested The standard ProTime cuvette has five micro channels which contain the dried reagents required to perform triplicate testing of the PT assay and two levels of controls The ProTime3 cuvette has three functional micro channels Two micro channels perform the controls and one micro channel performs the PT test The standard ProTime uses the Tenderlett Plus device for performing the fingerstick and it is designed to hold 65 L of blood approximately 3 drops needed to fill all five micro channels The ProTime3 uses the Tenderlett P
13. D is ON If PID is OFF and OID is ON 20 LQC RECORDS The result and ID information will be stored in the database as a QC record QC result will be displayed as follows depending upon the OID ON OFF selections and the LQC NOR ABN selections For LQC Normal with OID on For LQC Normal with OID off For LQC Abnormal with OID on For LQC Abnormal with OID off PRINT DATA OR SEND DATA Press the button to move the highlight bar to the SHOW RESULTS line on the MAIN MENU Then press the button to view the SHOW RESULTS menu and observe the following screen 21 The instrument auto send feature allows transmitting of test results directly to a serial printer or to a computer by using the PROCABLE Contact ITC Customer Service to order the PROCABLE To use the AUTO SEND feature press the button on the PRINT DATA or SEND DATA display until AUTO SEND is highlighted Press the button and depending on the current settings in the instrument one of the following displays will appear Example Printing Data Press the button to set the feature to ON or OFF and then press the button to save the setting Highlight the SHOW RESULTS line in the MAIN MENU Press the button to view the SHOW RESULTS menu Press the button to move the highlight bar to the PRINT DATA or SEND DATA line Press the button to access the PRINT DATA or SEND DATA option and select THIS RESULT or ALL RESULTS option in the PRINT DATA or SEND DATA screen
14. ProTime Microcoagulation System For in vitro Diagnostic Use Operator s Manual TABLE OF CONTENTS INTENDED USE 2 SUMMARY AND EXPLANATION 2 SYSTEM FEATURES 3 IMPORTANT LABELS AND SYMBOLS 4 PRINCIPLES OF OPERATION 5 SAFETY FEATURES AND QUALITY CONTROL 5 INSTRUMENT SPECIFICATIONS 7 SERVICE AND MAINTENANCE 7 PREPARING THE INSTRUMENT 9 TEST PROCEDURE 10 TROUBLESHOOTING 14 MAIN MENU OPTIONS 16 SET UP 22 FOR PROFESSIONALS ONLY PROGRAM MODE 27 PERFORMANCE CHARACTERISTICS 31 TECHNICAL ASSISTANCE 34 SUGGESTED READING
15. al regulations for human blood contaminated waste INSTRUMENT SPECIFICATIONS Size 2 77 h x 4 5 w x 8 75 l inches Weight 1 6 pounds Operating Temperature Room temperature 15 C to 30 C 59 F to 86 F PT Test Temperature 37 1 0 C Battery Type Nickel Metal Hydride NiMH Operating Time On Battery Approximately 2 hours constant run or 10 complete test cycles per charge Tests may also be run while ProTime is plugged into AC DC power module Anticipated Battery Life 500 charges 100 240 VAC Power Supply Charger Input 100 240 volts AC 47 63 Hz Output 15 volts DC 2 0 A SERVICE AND MAINTENANCE Routine Maintenance and Cleaning DO NOT immerse the ProTime instrument or allow fluid to enter the cuvette housing Inspect and clean the outside of the cuvette slot as required Remove residual dried blood or other foreign matter on the outside of the instrument using gauze dampened with a 10 dilution of household bleach in water or with gauze dampened with isopropyl alcohol or other disinfectant DO NOT use other solvents or strong cleaning solutions as they may damage the plastic components of the instrument Servicing The ProTime Instrument Other then replacement of the rechargeable battery as described in this manual the ProTime instrument is not user serviceable Should service be required please contact Technical Support at 1 732 548 5700 1 800 631 5945 or e mail us at techsupport itcmed com If you are a
16. amples had a correlation of 0 92 The home test results compared to a reference laboratory results 368 samples had a correlation of 0 87 These correlation values mean that home testing is comparable to lab testing Precision Precision testing was conducted with two levels of standard control plasma substrate preparations Level I and Level II A Standard ProTime Cuvette Control Precision n Mean SD Within day 17 0 9 0 06 Level I Day to day 5 days 4 day 1 0 0 08 Within day 19 3 2 0 19 Level II Day to day 5 days 4 day 3 2 0 12 B ProTime3 Cuvette Control Precision n mean SD Within day 18 0 9 0 07 Level I Day to day 5 days 4 day 0 9 0 12 Within day 20 4 0 0 19 Level II Day to day 5 days 4 day 4 2 0 22 33 Accuracy INR results generated by the ProTime and ProTime3 cuvettes using venous and fingerstick whole blood samples were compared to INR values obtained using standard laboratory plasma PT methods with samples collected in 3 2 sodium citrate tubes The following accuracy data was obtained A Standard ProTime Cuvette vs Lab Plasma Sample Type Regression Equation r n Fingerstick Y 0 94x 0 38 0 95 229 Venous Y 0 91x 0 44 0 94 232 B ProTime3 Cuvette vs Lab Plasma Sample Type Regression Equation r n Fingerstick y 1 05x 0 07 0 95 229 Venous y 0 97x 0 19 0 95 219 Clinical Performance Comparison A ProTime3 vs ProTime Line
17. ar regression containing clinical fingerstick results from three sites yielded a regression equation as below y 1 02x 0 14 r 0 94 n 229 B Patient Self Testing In a trial of patient self testing PST in the home vs professional testing in the clinical and reference lab equivalent ProTime results were obtained ProTime Professional vs Patient Self Testing y 94x 0 13 r 0 92 n 421 ProTime Patient Self Testing vs Reference Lab y 77x 0 38 r 0 87 n 368 Sensitivity The ProTime instrument is sensitive to deficiencies in vitamin K dependent coagulation factors known to influence the PT test i e Factors II VII and X Hematocrit levels between 20 and 60 do not significantly affect test results Quality Control The ProTime instrument has been designed with multiple systems to ensure proper instrument function The instrument self check at startup checks temperature and timing functions battery level and optical electrical and mechanical functions The instrument does not require further calibration Each ProTime cuvette has two integral reagent controls that ensure assay reliability and performance Both levels of control produce quantifiable clotting endpoints that are compared to pre set acceptance limits programmed in the instrument Other in process instrument QC features and the integral reagent controls function together to ensure correct sample size and collection technique correct test procedure instrumen
18. at these steps until the desired field length is entered To save the entered PID value press and hold the button for approximately 2 seconds until the second audio beep is heard Note These audio beeps will be heard regardless of BEEP setting ON OFF After the cuvette is inserted the user has the option to set an operator ID OID and or a patient ID PID if the OID and or PID have been enabled in the SET UP section of the MAIN MENU The PID may contain up to twelve digits the OID can be a value with up to six digits Note In the different scenarios related to SET PID OID ON OFF settings some of the following four screens are not presented to the user For example none of these four screens will appear if both PID and OID are set to OFF value The default numerical PID and OID values are zero If OID is set to the ON position the following screen will appear Use the and buttons to move the cursor and to enter the numeric ID The CONFIRM OID confirmation screen will follow the ENTER OID screen If button is pressed the user will return to the previous screen otherwise if the button is pressed and PID is set to the ON position the program will proceed to the following ENTER PID screen Use the and buttons to move the cursor and to enter the numeric ID 26 The CONFIRM PID confirmation screen will follow the ENTER PID screen After setting of both requested PID and or OID the WARMING screen will appear in the center of the disp
19. based on a conversion equation that was established in clinical trials The result in plasma equivalent seconds is then calculated from the INR result Historically laboratories have reported prothrombin time results in seconds specifically the time it takes for the plasma sample to clot following the addition of the reagents Reporting the results in seconds is problematic since the clotting time depends on the sensitivity of the reagent used Results cannot be compared from one laboratory to another The INR system was introduced by the World Health Organization WHO to standardize PT reporting such that patient results could be uniform across different laboratories INR is calculated using the formula INR PT PTn ISI where PT is the time for the patient plasma sample to clot PTn is the mean clotting time of a non anticoagulated population and ISI is the international sensitivity index that is assigned by the manufacturer of the reagent according to a standard protocol developed by the WHO The ISI value ranges from 1 0 to approximately 3 0 A low ISI indicates that the reagent has the highest sensitivity to vitamin K dependent clotting factors Typically the variable PTn is generated by the laboratory with each new reagent lot by running 20 samples and calculating the mean The ProTime system enhances clinical utility for the physician and patient by providing results in both formats so that anticoagulant dosing can be managed using a familiar f
20. cedure Repeat test SAMPLE NOT SEEN TURN OFF TRY AGAIN Can be caused by small samples clots air bubbles or a seal problem in one or more channels Check for proper collection technique Verify cup is completely filled Use another cuvette SAMPLE TOO LARGE TURN OFF TRY AGAIN Sample oversize Review directions and repeat the test SAMPLE TOO SMALL TURN OFF TRY AGAIN Sample undersize Repeat test Exceed the fill line on the cup to ensure sample size SAMPLE ERROR TURN OFF TRY AGAIN Incorrect fluid movement Air bubble detected in one or more of the channels Turn off and on again and repeat test with new cuvette and fresh fingerstick 16 Screen Display Source of Error Method of Control CUVETTE EXPIRED Expired cuvette Check to make sure cuvettes have not expired The expiration date is located on the packaging and alongside the barcode on the cuvette BARCODE ERROR REMOVE CUVETTE TRY AGAIN Instrument cannot read barcode accurately Visually inspect barcode If scratched discard cuvette If dirty wipe clean If barcode is good review correct procedure and repeat test BATTERY ERROR TURN OFF TRY AGAIN Power supply battery error Repeat the test If it occurs again replace the battery If problem persists call ITC Technical Support INSTRUMENT ERROR TURN OFF TRY AGAIN Instrument set up data log or communication error Repeat the test If problem persists call ITC Technical Support
21. dvised to return the instrument to ITC for service or repair prior to shipping please clean the instrument as described above Instrument Disposal If instrument disposal is required follow local regulations for the disposal of electronic devices For battery disposal see the Battery Replacement section Battery Information The ProTime Microcoagulation System is designed to run either on AC power supplied by the AC DC power module or on the rechargeable battery supplied within the unit 8 Rechargeable Battery Facts Batteries discharge naturally over time approximately 5 per month Battery capacity the amount of charge the battery will hold is lower at low temperatures Batteries that are charged to their maximum capacity will generate heat if they continue to be charged The ProTime instrument uses a rechargeable NiMH Nickel Metal Hydride type battery The maximum capacity of any rechargeable battery will gradually decrease over time To ensure maximum life of the rechargeable battery read and follow the information in Care of the ProTime Battery for Maximum Life section Care of the ProTime Battery for Maximum Life A new instrument an instrument that is used infrequently or an instrument with a new replacement battery should be plugged in for at least 8 hours before use to ensure that the battery is completely charged The instrument screen will show CHARGING BATTERY when the AC DC power module is connected to the
22. e Microcoagulation System be powered from an uninterruptible power supply or battery Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical domestic commercial or hospital environment NOTE UT is the AC mains voltage prior to application of the test level 39 INDEX attention label 4 battery care 8 battery information 8 charging 10 battery replacement 8 blood coagulation systems 2 blood collection 11 data history 19 LQC records 20 patient records 19 print send 21 finger incision preparation 11 insert a cuvette 11 intended use 2 introduction 2 ISI setting 29 main menu
23. e blood samples The clotting time is a measure of the functionality of the patient s hemostatic system Specific coagulation activating reagents are employed in various clotting tests to measure specific components of the hemostasis system Clotting times are prolonged in the case of either decreased procoagulant activity or increased anticoagulant activity Summary of the Blood Coagulation System The events leading to the formation of a fibrin clot are simplified in coagulation theory into two coagulation pathways the intrinsic and the extrinsic pathways There are twelve clotting factors or proteins involved in this cascade scheme numbered I through XIII excluding VI The prothrombin time test measures the extrinsic coagulation pathway and is sensitive to coagulation Factors VII X V II and Fibrinogen I With the exception of Factor V vitamin K is a required co factor for biosynthesis of these factors in the liver The prothrombin time PT test uses thromboplastin as the active reagent to initiate the extrinsic pathway The prothrombin time test will be prolonged in patients with liver disease or vitamin K deficiency The test is widely used to monitor oral anticoagulant therapy that suppresses the synthesis of vitamin K dependent clotting factors 3 SYSTEM FEATURES 1 Tenderlett Plus Tenderlett Plus LV The finger incision and blood collection device used with ProTime and ProTime cuvettes 2 ProTime ProTime3 Cuvette The cuvette
24. e screen for optimal operation Do not try to hasten the test process by performing the fingerstick while the WARMING screen is displayed as blood may clot before it is drawn into the cuvette and an error may occur FOLLOW THE INSTRUCTIONS ON THE SCREEN Troubleshooting Guide Screen Display Source of Error Method of Control INR HIGH REPEAT TEST IF SAME CALL MD Patient s PT result is too high INR gt 12 0 Repeat the test If it occurs again patient should be referred to their physician IMMEDIATELY and re tested at the lab INR LOW REPEAT TEST IF SAME CALL MD Patient s PT result is too low INR lt 0 6 Repeat the test If it occurs again patient should be referred to their physician IMMEDIATELY and re tested at the lab ON BOARD QC OUT OF RANGE TRY AGAIN Level I or II control is too high or too low Repeat the test Check for adequate storage of cuvettes or sampling technique Confirm results with the lab NO CLOT DETECTED REPEAT TEST IF SAME CALL MD At least one channel did not clot Repeat the test If it occurs again patient should be referred to their physician IMMEDIATELY and re tested at the lab TIME OUT TRY AGAIN Time ran out before starting test Test again with new Tenderlett Plus and a fresh fingerstick You may reuse the cuvette if no blood was in contact with the cuvette TEST INTERRUPTED TRY AGAIN Cuvette removed or operator interrupted test Review correct pro
25. e user will be given two more chances to enter the correct password in the ENTER PASSWORD screen After three unsuccessful attempts to enter the correct password the program will return to the MAIN MENU and the user should contact ITC Technical Support After performing all programming functions the user will return to the MAIN MENU After the correct password has been entered the PROGRAM MODE menu screen appears SET MM YYYY Selecting SET MM YYYY set month and year will guide you through the following sequence of screens Change the month by pressing the button until the correct month appears 28 Once the correct month appears press the button to set the month The procedure will then advance to changing the year screen Change the year by pressing the button until the correct year appears Once the correct year appears press the button to set the year The procedure will then advance to MAIN MENU SET TR LIMITS To set therapeutic range TR limits select the SET TR LIMITS from PROGRAM MODE menu screen The following screens appear for selection of the upper and lower TR limits Set upper limit by pressing the button until the desired upper limit is displayed Select the limit using the button The upper limit can be set to one of eight values ranging from 2 0 to 5 5 with an incremental step of 0 5 or to the value 9 9 Once the upper limit is set the lower limit will be available for setting Set lower limit by pressing the button unt
26. il the desired lower limit is displayed Select the limit using the button The lower limit can be set to 0 0 1 2 or to one of six values ranging from 1 5 to 4 0 with an incremental step of 0 5 If limits are not changed the instrument will use its default limits 9 9 and 0 0 Once the limits are set the instrument will return to the MAIN MENU 29 SET ISI If SET ISI is selected from the PROGRAM MODE the following screen is displayed To change the ISI used in the calculation access the PROGRAM MODE from the SET UP on MAIN MENU and enter the ISI of the local laboratory reagent Displayed plasma equivalent values are indicative of the results obtained using a laboratory reagent with an ISI of 1 0 If the ISI of the reagent used in the local facility is significantly different from 1 0 the operator may wish to select an ISI more closely aligned with their lab reagent Change the pre programmed ISI by pressing the button until the desired ISI is displayed Press the button to set The values used for the normal control population change as the theoretical ISI is changed The normal control population values are determined from clinical data REPORTABLE RANGE In clinical trials no significant difference was observed between fingerstick and venous specimens run on ProTime During those trials ProTime measured patients with an INR range of 0 8 to 7 0 ProTime is programmed to calculate and report INR results as follows If the calculated INR is
27. lay The test sequence then continues as described in TEST PROCEDURE SET BEEP If SET BEEP is selected the following screen appears To turn the BEEP SOUND ON press the CHANGE button The SET BEEP ON screen appears Then press the button to turn the beep sound ON Note Regardless of BEEP setting the beep will sound under the following conditions power on power off and following a key hold 27 FOR PROFESSIONALS ONLY PROGRAM MODE Note The Program Mode feature is only available to professional users The following information is not contained in the Operator s Manual for patients performing self testing For PROTIMEPRO to access the Professionals Only section of the training CD enter the following code when prompted by the software 54321 the training CD is not provided with PROTIMEINT If the PROGRAM MODE option is selected from the SET UP Menu the following screen appears Note The PROGRAM MODE option is password protected The default password in a newly manufactured instrument is 000000 which the user can then change to a password of their own choosing see the Change Password section If you forget or lose your password an emergency password is available from ITC Technical Support by calling 1 732 548 5700 1 800 631 5945 or e mail us at techsupport itcmed com If the entered password matches the password stored in the instrument the instrument will enter the PROGRAM MODE If the entered password is incorrect th
28. lowing safety standard requirements and directives CSA C22 2 601 1 Medical Electrical Equipment General Requirements for Safety EN 60601 1 UL IEC 60601 1 Medical Electrical Equipment General Requirements for Safety EN 60601 1 2 IEC 60601 1 2 Medical Electrical Equipment Part 1 2 General Requirements for Safety Collateral Standard Electromagnetic Compatibility Requirements and Tests Directives 89 336 EEC and as amended by 91 263 EEC 92 31 EEC 93 68 EEC and 98 13 EC 98 79 EC Equipment Classifications As Defined Per UL 60601 1 2003 IEC60601 1 2nd Edition Protection against electrical shock Class II Internally Powered Equipment with no applied parts Protection against ingress of liquids Ordinary no protection as defined by IEC 529 Product cleaning and disinfection Only according to recommendations of the manufacturer s accompanying documentation Mode of operation of equipment Continuous Degree of safety of application in the presence of flammable anesthetic mixture with air oxygen or nitrous oxide Not Suitable NOTE As defined in the above standards the classification of Not Suitable DOES NOT MEAN that the instrument is not suitable for use in an Operating Room OR environment Rather it is intended to indicate that the instrument is not suitable for use in the direct presence of a flammable anesthetic mixture with air oxygen or nitrous oxide All relevant docume
29. lus LV low volume device for performing the fingerstick and it collects 27 L of blood approximately 1 large drop needed to fill the three micro channels of the ProTime3 cuvette The instrument draws the precise volume of blood into the micro channels of each cuvette which contain thromboplastin and other reagents An array of LEDs detects the motion of sample reagent mixtures as they move through a precision restriction in each channel The blood is pumped back and forth until a clot forms obstructing the channel and slowing the flow of blood The instrument detects the clot when the blood movement decreases below a predetermined rate SAFETY FEATURES AND QUALITY CONTROL Each ProTime and ProTime3 cuvette has two channels for performing the two levels of control which work simultaneous with the channel s used for testing the PT assay The built in safety features of the instrument and integral reagent controls work together to ensure that the instrument and reagent systems are working properly and that the test procedure was performed correctly Two levels of quality control are performed with each and every test Calibration The ProTime instrument and cuvettes are pre calibrated No additional calibration is required Reagents ProTime cuvettes are pre loaded with dried thromboplastin stabilizers and buffers The thromboplastin has a high sensitivity measured as an ISI near 1 0 Each cuvette performs the prothrombin time test and in additio
30. n has one channel for a Level I control and one channel for a Level II control The controls consist of purified plasma extracted coagulation factors and anticoagulants 6 Operating Precautions For in vitro Diagnostic Use The ProTime instrument is designed for use only with ProTime cuvettes The ProTime instrument will not perform a test if the cuvette is past its expiration date or has been improperly stored The ProTime instrument is designed to be used for testing in a stationary position DO NOT perform testing while carrying or holding the instrument In order to charge the ProTime instrument the AC power cord should be plugged into an electrical service outlet and the AC DC power module while the DC power cord from the AC DC power module is plugged into the DC port in the back of the ProTime instrument DO NOT expose the ProTime instrument to extreme temperature above 35 C 95 F Such exposure could affect the performance of any type of electronic instrumentation DO NOT drop the ProTime instrument and do not use the results if the instrument is dropped during a test DO NOT attempt to open the ProTime instrument other than for battery replacement as there are no user serviceable parts DO NOT remove the AC DC power module from the ProTime instrument by pulling on the cord Patient specimens and used cuvettes are potentially infectious The cuvettes include materials that have been prepared
31. nables only an operator ID to be entered Selecting OFF disables both a patient ID and operator ID Note The selection is highlighted and remains the default until changed again by the user Entering and Changing Numerical Information PID OID PASSWORD By using a combination of the button and button and the triangular cursor on the display screen numeric information can be entered into the instrument for operator identification number OID patient identification number PID and PASSWORD fields When entering numeric information into a field pressing a button has the following actions CHANGE Pressing the button once will increment the digit in the current cursor position by one starting at 0 and going through 9 and back to 0 again OK Pressing the button briefly enters the digit displayed in the current cursor position and then moves the cursor right to the next field position Pressing the button for 2 seconds or longer saves the complete numeric field EXAMPLE Entering a PID number With the PID feature enabled the following screen appears after a cuvette is inserted into the instrument 25 To enter a PID the first position in the PID field starting from the left is indicated by a triangular cursor Press the button until the desired digit is displayed Press the button briefly to enter the digit in that position of the field The cursor will automatically move to the next field position Repe
32. ntation is kept on file at ITC 37 Guidance and Manufacturer s Declaration Electromagnetic Emissions The ProTime Microcoagulation System is intended for use in the electromagnetic environment specified below The customer or the user of the ProTime Microcoagulation System should assure that it is used in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF Emissions CISPR 11 Group 1 The ProTime Microcoagulation System uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF Emissions CISPR 11 Class B Harmonic Emissions IEC 61000 3 2 Class A Voltage Fluctuations Flicker Emissions IEC 61000 3 3 Complies The ProTime Microcoagulation System is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes 38 Guidance and Manufacturer s Declaration Electromagnetic Immunity The ProTime Microcoagulation System is intended for use in the electromagnetic environment specified below The customer or the user of the ProTime Microcoagulation System should assure that it is used in such an environment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Elec
33. o have an abnormal response to heparin Correlation of results reported by the ProTime instrument to laboratory results depends on the precision of the laboratory method and on the ISI value of the laboratory reagent and instrument system Do not disturb instrument while test is in progress As with all diagnostic tests ProTime Microcoagulation System test results should be scrutinized in light of a specific patient s condition and anticoagulant therapy Any results exhibiting inconsistency with the patient s clinical status should be repeated or supplemented with additional test data Follow doctor s instructions if you have difficulty performing the test or receive a result outside of the prescribed therapeutic range 7 Reagent Preparation and Storage Bring the foil pouched cuvette to room temperature prior to use ProTime cuvettes are ready to use No additional preparation is required Store the foil pouched cuvettes in a refrigerator 2 C to 8 C 36 F to 46 F An unopened cuvette pouch is stable when stored at 2 C to 8 C until the date printed on the pouch Unopened cuvette pouches may be stored at room temperature for 60 days Once the pouch has been opened the cuvette must be used within 16 hours Cuvette or Tenderlett Disposal The ProTime cuvette and Tenderlett Plus are for single use only and are not to be reused After use they contain human blood and should be disposed of in accordance with loc
34. ollowing screen is displayed and the user can select NOR ABN NORMAL ABNORMAL The instrument performs SELF CHECK procedure which may take up to 60 seconds The testing proceeds as described above except that no prompt for PID will be shown for a QC test whether PID is ON or OFF When the test completes the result will be displayed as follows depending on OID ON OFF and LQC NOR ABN selection NOTE The result and ID information will be stored in the database as a QC record For LQC Normal with OID on For LQC Normal with OID off For LQC Abnormal with OID on For LQC Abnormal with OID off 19 SHOW RESULTS Press the button to move the highlight bar to the SHOW RESULTS line on the MAIN MENU Press the button to view the SHOW RESULTS menu and observe the following screen DATA HISTORY Press the button to view the most recent result in the DATA HISTORY memory The instrument memory holds 50 results Pressing the button will scroll through individual results The results are stored in memory from the most recent to the oldest Press the button to return to the MAIN MENU PATIENT RECORDS The result and ID information will be stored in the database as one of the following screens depending upon the PID OID ON OFF selections Note MM DD YYYY and HH MM are used for illustrative purposes only in the following examples of date and time screens If PID is ON and OID is OFF If PID is OFF and OID is OFF If PID is ON and OI
35. once the ProTime instrument is plugged into its AC DC power module Battery Replacement Refer to the instructions provided with the replacement battery The used battery should be disposed of in accordance with local regulations for NiMH batteries 9 SPECIMEN COLLECTION Fingerstick whole blood is the recommended specimen The Tenderlett Plus device is to be used with the ProTime cuvette and the Tenderlett Plus LV low volume device is to be used with the ProTime3 cuvette The Tenderlett Plus device will collect approximately 65 L of blood approximately 3 drops while the Tenderlett Plus LV device will collect approximately 27 L of blood approximately 1 large drop Samples should be analyzed immediately after collection No additional sample preparation is required For venous samples collect venous whole blood into an anticoagulant free plastic syringe in place of fingerstick sampling steps 3 and 4 of the Test Procedure section Immediately dispense venous sample into the Tenderlett Plus collection cup filling the Tenderlett Plus collection cup Follow instructions from step 5 of the Test Procedure section Note Serum plasma or whole blood collected with any anticoagulants is NOT suitable samples PREPARING THE INSTRUMENT Unpacking and Inspection 1 Remove any protective packaging that may be present around the instrument 2 Examine the packaging material to be sure that the AC DC power module AC power cord see note below
36. ord 2 Connect the DC power cord from the AC DC power module into the DC port on the back of the instrument The instrument screen will show CHARGING BATTERY when the AC DC power module is connected to the AC power cord and the ProTime instrument The screen will show CHARGE COMPLETE when the battery is fully charged 3 Allow the battery to charge for at least eight hours TEST PROCEDURE 1 Turn on the ProTime Instrument Press the button to start The ProTime instrument performs a self check procedure that may take up to 60 seconds ProTime will prompt you through the test Watch the screen and follow the prompts 2 Insert a Cuvette Make sure the ProTime cuvette is brought to room temperature before use Wait for the prompt Insert the cuvette into the slot with the printing face up and the bar code down The WARMING screen will appear as follows 11 3 Prepare for Finger Incision While the cuvette is warming prepare the finger Wait for the prompt before incising the finger and collecting blood It is easier to collect blood if the hands are warm Follow these steps to ensure a good sample Wash the hands in warm water or rub hands together to stimulate blood flow Apply firm pressure to the palm and finger Massage the hand and push blood into the fingertips Cleanse middle or ring finger and dry To prevent contamination do not touch the site after cleansing 4 Blood Collection When this screen ap
37. ormat Since results reported in PT seconds depend on the sensitivity ISI of the reagent employed the physician has the option of changing the ISI value in ProTime so that the ProTime results reported in PT seconds closely match the results reported by the hospital laboratory To change the ISI used in the calculation the physician simply accesses the programming screens and enters the ISI of the historical laboratory reagent Please see the For Professionals Only Program Mode section The default parameters used to calculate PT seconds are ISI 1 0 which is the sensitivity of the thromboplastin reagent used in the ProTime cuvette and PTn 13 1 seconds which was established in clinical trials The PTn value used in the calculation is automatically adjusted when the reporting ISI is changed Table 1 Options for programming ISI and associated PTn ISI 1 0 1 2 1 4 1 6 1 8 2 0 2 2 2 4 2 6 PTn 13 1 12 9 12 8 12 6 12 5 12 3 12 2 12 0 11 9 Table 2 Relationship of INR to PT seconds with varied ISI PT seconds vs ISI INR ISI 1 0 ISI 1 6 ISI 2 0 ISI 2 4 1 0 13 1 12 6 12 3 12 0 1 5 19 6 16 2 15 1 14 2 2 0 26 1 19 4 17 4 16 0 2 5 32 6 22 3 19 5 17 6 3 0 39 2 25 0 21 3 19 0 3 5 45 7 27 6 23 0 20 2 4 0 52 2 30 0 24 6 21 4 4 5 58 7 32 3 26 1 22 5 32 Table 2 shows examples of how the calculation of PT seconds is affected by changing the ISI As the ISI increases the plasma e
38. pears it is time for the finger incision CAUTION Blood collection must be finished within 2 10 minutes to prevent early clotting of the sample ProTime will keep time If the 2 10 minute interval has expired and the button has not been pressed during this interval a TIME OUT error message is displayed To run another test repeat procedure from Step 2 Place the Tenderlett Plus device firmly against the side of the finger Place thumb on top of the device as shown and press the red trigger firmly using the other thumb Wipe away the first trace of blood Gently massage from the base of the finger to force blood to the tip so that a large drop of blood forms 12 Touch the large drop of blood to the collection cup Keep adding blood until the blood level fills the cup above the line For Tenderlett Plus LV ensure the cup is filled completely Ensure the blood extends all the way to the bottom of the cup Add another drop if you are not sure you have enough 5 Snap Tenderlett Plus to ProTime Hold the device at an angle and place the front end of the device into the slot in the instrument Press down to click the Tenderlett Plus in place You should hear a soft click Note Proper engagement of the Tenderlett Plus to the cuvette is critical to prevent a sample error 6 Start the Test Press the button to start the test This signals ProTime to draw the sample into the cuvette 13
39. performs the PT test 3 Display Panel The display panel prompts you through the test procedure displays results and error messages 4 MENU SCROLL Button The button is used to move through menu screens and to scroll through results 5 START SELECT Button The button is pressed to turn ProTime on and off to start the test and to select menu items 6 Cuvette Entry Port The cuvette entry port accepts only ProTime and ProTime3 cuvettes 7 AC DC Power Module The AC DC power module is used to connect the ProTime instrument to the AC power cord to recharge the battery 8 Input DC Port Plug the DC power cord from the AC DC power module into this port when connecting the ProTime instrument to the AC DC power module 9 Output Data Port Use this port to download records in the ProTime memory 1 Tenderlett 2 ProTime Cuvette 3 Display Panel 4 MENU SCROLL Button 5 START SELECT Button 6 Cuvette Entry Port 8 Input DC Port 9 Output Data Port 7 AC DC Power Module 4 IMPORTANT LABELS AND SYMBOLS Before using the ProTime Microcoagulation System it is essential that the contents of this Operator s Manual and any instructions accompanying the ProTime cuvettes and Tenderlett Plus incision devices are read and understood by the operator These materials make reference to various symbols that are explained below Start Select Menu Scroll Expiration Date of Cuvettes Serial Number of Device Lot Number of
40. quivalent PT decreases at any given INR For example a patient with INR 3 00 would have PT seconds 39 2 if the ISI is the default ISI 1 0 or PT seconds 21 3 if the ISI is reset to ISI 2 0 Expected Values ProTime measures both normal and therapeutic prothrombin times in fresh whole blood Results are displayed in plasma equivalent seconds and INR Expected values for patients taking oral anticoagulants depend on the patient s disease state and the target values established by the physician Sample Condition INR PT Seconds ISI 1 0 Normal 0 8 1 2 10 4 15 7 sec Low anticoagulation 1 5 2 0 19 6 26 1 sec Moderate anticoagulation 2 0 3 0 26 1 39 2 sec High anticoagulation 2 5 4 0 32 6 52 2 sec Note The ProTime instrument converts whole blood clotting time to INR then calculates PT seconds The default ISI used in this calculation is 1 0 Please see the For Professionals Only Program Mode section if you wish to select an ISI that is more closely aligned with the local lab reagent The ProTime instrument has been tested extensively by doctors nurses and patients Eighty four people ages 7 to 81 participated in the first home use trial The trial was conducted to see how well ProTime results compare to the lab Comparisons of this type are described by correlation values and a correlation value near 1 0 means the comparison is good The home tests compared to tests run at the clinic using ProTime with venous samples 421 s
41. res calibration 5 cuvette disposal 7 limitations 6 precautions 6 reagent preparation 7 reagents 5 storage 7 service and maintenance battery facts 7 instrument disposal 7 routine maintenance 7 servicing 7 set up options 22 beep 26 day 23 hour 23 language 22 minutes 23 OID 24 password 24 PID 24 specifications 7 specimen collection 9 technical assistance
42. rothrombin Time Ratio on the Results of Oral Anticoagulant Therapy N Engl J Med 1993 329 696 702 Fairweather RB Ansell J van den Besselaar AMHP Brandt JT Bussey HI Poller L et al College of American Pathologists Conference XXXI on Laboratory Monitoring of Anticoagulant Therapy Arch Pathol Lab Med 1998 122 768 81 Gosselin R Owings JT White RH Hutchinson R Branch J Mahackian K et al A Comparison of Point of Care Instruments Designed for Monitoring Oral Anticoagulation with Standard Laboratory Methods Thromb Haemost 2000 83 698 703 Hirsh J Antithrombotic Therapy in Deep Vein Thrombosis and Pulmonary Embolism Am Heart J 1992 123 1115 22 Hirsh J Poller L The International Normalized Ratio A Guide to Understanding and Correcting its Problems Arch Intern Med 1994 Feb 14 154 3 282 8 Hubbard AR Margetts SM Weller LJ Macnab J Barrowcliffe TW An International Collaborative Study on the INR Calibration of Freeze Dried Reference Plasmas Br J Haematol 1999 Mar 104 3 455 60 Levine M HJ Landefeld Raskob G Hemorrhagic Complications of Anticoagulant Treatment CHEST 1992 102 352s 63s Technical Assistance For further information and technical assistance call ITC Technical Support at 1 732 548 5700 1 800 631 5945 or e mail us at techsupport itcmed com Ordering Information For further information on ordering and supplies contact your ProTime distributor 36 SAFETY STANDARDS The ProTime instrument complies with the fol
43. s to train patient self testers using the ProTime Microcoagulation System ProTime instruments intended for patient self testing are available in the U S by prescription only These instruments include patient specific product instructions For in vitro Diagnostic Use SUMMARY AND EXPLANATION What Does The ProTime Microcoagulation System Do The ProTime Microcoagulation System is designed for testing prothrombin time PT and International Normalized Ratio INR The reagents for a prothrombin time test are in the cuvette Whole blood clotting time is converted to INR than the result is calculated for plasma equivalent PT seconds This test is done to check the status of patients receiving oral anticoagulation therapy What is INR The International Normalized Ratio INR was developed to help the doctor compare an individual s prothrombin time results from one lab to another An advantage of reporting an INR is to allow for normalization of comparisons from one lab or instrument to another The precision of the INR is improved when a reagent with a lower ISI is used Note ISI stands for International Sensitivity Index ProTime uses this number to calculate PT seconds from INR Blood Coagulation Test Methodology Traditional coagulation tests measure the time required for the formation of a fibrin clot following the addition of a coagulation activating reagent Laboratory assays typically use plasma recovered from anticoagulated citrated whol
44. side of the therapeutic range The OTR will not display if results are within the limits If no limits are set ProTime will display only the result Note If OTR Out of Therapeutic Range appears after the INR result the result is out of the therapeutic range TR limits that have been preset by the professional see the For Professionals Only Program Mode section TROUBLESHOOTING Most often an error message indicates a problem with blood collection or a mistake in the test procedure If an error message appears read the instructions again and repeat the test with a new cuvette The Troubleshooting Guide section gives the messages for the most common errors with possible causes and solutions As with all diagnostic tests the ProTime Microcoagulation System test results should be scrutinized in light of a specific patient s condition and anticoagulant therapy Any results exhibiting inconsistency with the patient s clinical status should be repeated or supplemented with additional test data If you get a persistent error write down the error message and call ITC Technical Support at 1 732 548 5700 1 800 631 5945 or e mail us at techsupport itcmed com Please have the ProTime serial number and cuvette lot number ready when you contact ITC Technical Support 15 Troubleshooting Guidelines Become skilled at obtaining blood with the fingerstick technique This will help avoid sample errors Follow all of the directions on th
45. t functionality and reagent integrity A fault message is displayed instead of PT results when any instrument or reagent quality criterion is not met When a fault message is displayed the user should review the product instructions and repeat the test 34 Additional external control materials may be used to check instrument function reagent integrity and user technique Each institution should establish a quality assurance program for prothrombin time testing that complies with existing local state and federal regulations as applicable As with all diagnostic tests the ProTime Microcoagulation System test results should be scrutinized in light of a specific patient s condition and anticoagulant therapy Any results exhibiting inconsistency with the patient s clinical status should be repeated or supplemented with additional test data TECHNICAL ASSISTANCE For further information and technical assistance contact ITC Technical Support at 1 732 548 5700 1 800 631 5945 or e mail us at techsupport itcmed com 35 SUGGESTED READING Adcock DM Kressin DC Marlar RA Effect of 3 2 vs 3 8 Sodium Citrate Concentration on Routine Coagulation Testing Am J Clin Pathol 1997 107 105 10 Brien WF Baskerville JC Taberner DA Crawford L Calculation vs Calibration Curve for INR Determination Results of an Interlaboratory Proficiency Scheme Am J Clin Pathol 1999 111 193 201 Eckman MH Levine HJ Pauker SG Effect of Laboratory Variation in the P
46. trostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient Burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply lines 1 kV for input output lines Mains power quality should be that of a typical domestic commercial or hospital environment Surge IEC 6100 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical domestic commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 lt 5 UT gt 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles lt 5 UT gt 95 dip in UT for 5 sec lt 5 UT gt 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles lt 5 UT gt 95 dip in UT for 5 sec Mains power quality should be that of a typical domestic commercial or hospital environment If the user of the ProTime Microcoagulation System requires continued operations during power mains interruptions it is recommended that the ProTim