Home
CT Analyzer Brochure
Contents
1. Your Benefits gt Field verification of CTs up to the 0 1 accuracy class due to extremely high accuracy 0 02 typical gt Compact and lightweight lt 8 kg 17 4 Ibs gt Automatic assessment according to IEEE and IEC standards gt Reduced testing time typically lt 1 min gt High noise immunity for on site testing Highly Accurate CT Verification Made Mobile The ideal way of testing a current transformer As energy is supplied by many different sources power system grids for generation transmission and distribution are expanded continuously This makes the use of additional metering and protection CTs necessary To test all of these CTs in a cost effective and reliable way the ideal CT test device fulfills the following requirements Mobility Test engineers often have to maintain several CTs within one utility The ideal CT test device would therefore be an all in one solution light enough to be carried by one person It should be able to measure all parameters without the need for any further equipment such as a burden box Accuracy Correct billing is only possible if metering CTs work within their specifications for all secondary burdens and levels of primary current that are defined in the standards To test and calibrate these metering CTs measurement equipment delivering reliable results up to class 0 1 CTs is needed Primary Nominal Current Injection Primary Current
2. x ae a Remote Control SS x gt Full access to all functions of the CT Analyzer via a PC using oer crel mem the remote interface Object erudite coef O o gt Optimizes the integration into automated testing procedures o in production lines ken 600 A tare 5 sla p pes g oO z gt Data export into Excel and Word pan Jr gt Customizable testing and reports Rated burden 7 va w Op budeni 7 mwa op come 7 F Network Simulation gt NetSim is a software tool for network simulation part of the Test Universe software suite for relay testing gt Easy transfer of CT Analyzer measurement data to NetSim gt Accurate modeling of power systems for network studies and fault simulation testing of protection relays gt Behavior analysis of protective relays in case of CT saturation Data Handling and Reporting gt Test reports can be saved on the Compact Flash Card and transferred to a PC gt Data and protocols can be shown on a PC via the Excel file loader program gt Customizable report templates are available for example Different standards classes and applications gt Single multi core and multi tap CTs gt Three phase testing gt Core testing V OMICRON Verification for different bu rdens and currents POWER Current ratio error in at of rated curr
3. is the most complete testing system for protection and metering CTs according to IEEE and IEC standards It allows all types of single and multi ratio current transformers to be tested on site in power system grids Manufacturer of CTs transformers or GIS use the CT Analyzer in their production facilities and test development labs The CT Analyzer offers a wide range of measurements such as CT ratio and phase angle accuracy with consideration of nominal and operational burden for various currents CT winding resistance CT excitation saturation unsaturated and saturated ALF and FS direct and indirect Burden impedance CT residual magnetism OMICRON Automated testing procedure The CT Analyzer is designed to accurately measure all relevant CT parameters and compare them to the requirements of the defined IEEE or IEC standard Due to this automated assessment testing engineers receive the pass or fail decision within seconds Step 1 Step 3 POWER Current ratio error in at of rated current W W cos Data type 1 5 10 20 50 100 120 200 String value ues ae 0 021 0 018 0 013 o0 0 009 0 008 yam 0 8 Float value 0 008 0 010 0 010 0 008 0 006 0 004 0 003 0 002 String value 0 005 0 001 0 000 0 001 0 000 0 000 0 001 0 001 ewe 1 Float value 0 005 0 001 0 000 0 001 0 000 0 000 0 001 0 001 H
4. the standard are ideally created automatically by the test device Mobility Accuracy Safety Handling Secondary Voltage Injection CT as an Electrical Model PP V VAG gt 30 kg 66 1 Ibs Not including additional equipment e g external burden box Mobility lt 8 kg 17 4 lbs ideal for handling on site Measurement of class 0 1 metering CTs Excellent noise suppression guarantees highly accurate on site testing even if active lines are close to the test object Not sufficient for high accuracy metering CTs Sensitive to transient distortion from life signals Accuracy due to the use of 50 Hz test signals Voltages for saturation tests can be 2 000 V or more Safety Maximum output voltage of 120 V Re wiring is required for each type of test One step test determining all parameters lt 1 min e g ratio polarity saturation winding resistance Handling Automated assessment to standard and Test results must be assessed manually integrated report functionality Extraordinary Features Flux density B RemAlyzer gt Software based tool to determine the residual magnetism in current transformers gt Analysis of the remanence condition before putting into operation the CT to assure proper function gt Simplifies power grid failure analysis after unwanted operation of protective relays gt Demagnetizes the CT core after measurement
5. EUROPE S SELECTION OF TEST amp MEASUREMENT EQUIPMENT FOR HIRE Nationwide Low Call g ry OMICRON CT Analyzer Revolution in Current Transformer Testing and Calibration A OMICRON Revolutionary Way of CT Testing Current transformers are used for relaying and metering purposes in electrical power systems They connect the high power primary side to the protection and metering equipment on the secondary side Depending on the application they are used for current transformers are designed differently Protection current transformers As it is used to feed protective relays the CT must be accurate during normal and fault conditions Failures in transformation could lead to misoperation of the relay along with unwanted and costly outages To test CTs according to the requirements of modern protection systems it is compulsory to consider transient components and auto reclosure systems Metering current transformers CTs for metering purposes must provide high accuracy up to class 0 1 to guarantee correct billing It is therefore essential to test and calibrate the metering current transformer as the entire metering chain is only as accurate as the instrument transformers feeding the meter In contrast to protection CTs metering CTs must go into saturation directly above the nominal primary current level to protect the connected metering equipment CT Analyzer a new way of testing CTs The CT Analyzer
6. Injection Approx 2 tons of equipment high current source huge cables current booster burden box etc gt 30 kg 66 1 Ibs Not including additional Mobility equipment e g external burden box Mobility High accuracy but complicated wiring makes Not sufficient for high accuracy metering CTs testing error prone Accuracy _ Sensitive to transient distortion from life signals due to the use of 50 Hz test signals Accuracy Uses dangerously high voltages and currents eg Typical levels of A to 800A d primary nominal current injection Safety ypical current levels of 500 A to are use Safety Re wiring is required for each type of test Handling e g ratio polarity saturation winding resistance Test results must be assessed manually Requires several people to set up and Handlin J conduct the test Safety OMICRON Equipment for testing CTs on site must comply to applicable safety standards and regulations However the ideal test device avoids the use of high test currents and voltages and conducts tests with as low test voltages as possible to reduce the operator s health and safety risks Handling Short measurement times and an automated assessment to the respective IEC and IEEE standards characterize modern test equipment All relevant parameters should be measured in one test cycle without the need for rewiring Printable test reports including all measured data and the assessment to
7. a Europe South and East Asia Australia and the Middle East together with a worldwide network of distributors and representatives make the company a market leader in its sector The following publications provide further information on the solutions described in this brochure Datasheet CT SB2 Switch Box For a complete list of available literature please visit our website Americas Asia Pacific Europe Middle East Africa OMICRON electronics Corp USA OMICRON electronics Asia Limited OMICRON electronics GmbH 12 Greenway Plaza Suite 1510 Suite 2006 20 F Tower 2 Oberes Ried 1 Houston TX 77046 USA The Gateway Harbour City 6833 Klaus Austria Phone 1 713 830 4660 Kowloon Hong Kong S A R Phone 43 5523 507 0 1 800 OMICRON Phone 852 3767 5500 Fax 43 5523 507 999 Fax 1 713 830 4661 Fax 852 3767 5400 info omicron at info omicronusa com info asia omicron at OMICRON L205 January 2012 Subject to change without notice www omicron at www omicronusa com a world class service every day In every sector You can take advantage of expert advice to ensure you get the best most appropriate and cost effective equipment for the job We supply a wide variety of industries so if there is another way to do the job or save you time and money we ll pass on the benefit of our experience for free If we don t have a particular item rather than hiring you something that won t do the job we would rath
8. ent Ea cos Phi Data type 1 5 10 20 50 100 120 200 pe Existing measurement data can be loaded to the CT Analyzer re String value 0 023 0 023 0 021 0 018 0 013 0 010 0 009 0 008 ata ny time Float value 0 023 0 023 0 021 0 018 0 013 0 010 0 009 0 008 aa String value 0 008 0 010 0 010 0 008 0 006 0 004 0 003 0 002 gt Recalculation of the CT para meters for different burdens Float value 0 008 0 010 0 010 0 008 0 006 0 004 0 003 0 002 String value 0 005 0 001 0 000 0 001 0 000 0 000 0 001 0 001 a nd Pomiary EH rrents i Float value 0 005 0 001 0 000 0 001 0 000 0 000 0 001 0 001 No fu rther on site measurements are necessa ry to verify i String value 0 007 0 005 0 004 0 003 0 003 0 003 0 004 0 004 whether change in the burden will influence the accuracy Float value 0 007 0 005 0 004 0 003 0 003 0 003 0 004 0 004 ofa CT Manual Testing QuickTest gt Use of the CT Analyzer as a multimeter with an integrated current and voltage source gt Perform manual tests L Z R ratio polarity burden etc for Alling oa wv trouble shooting and quick verification on site Frequency S000 is gt VT ratio check S 673me 89 91 1 002 O 000 5 44 85 157 0 0015 Ta12Zkn 3 7 2OpF CT SB2 Switchbox gt Automated testing of multi tap CTs without the need for rewiring Includes termina
9. er direct you to an alternative supplier You will always be provided with full instructions and if you still need help call our technical team on Nationwide Low Call 0333 6000 600 Our aim is to save you time frustration and money For full details check our price match guarantee online With Inlec you ll get the most accurate reliable and well maintained equipment available Prices are regularly reviewed to ensure you always enjoy the best value for money We have made a significant investment in test equipment so we ensure that it s well packed to minimise damage and delay You won t waste your time contacting Inlec Every request for equipment is logged and carefully considered Listening to our customers helps keep our product range up to date and O relevant If you are unhappy about any aspect of our service please let us know so we can put it right subject to availability and conditions Nati ide L Call r Fa T UKAS Mas Online sas MES mmen U m 008 Accredited Contractor www chas gov uk CERT NO GB93 1773 ENG Inlec UK Ellerbeck Way Stokesley Business Park Stokesley N Yorkshire TS9 5JZ United Kingdom
10. hout accessories Ordering information Name Order No Description Standard Package VE000656 CT Analyzer Standard Package Advanced Package VE000654 CT Analyzer Advanced Package Upgrade Standard Advanced VESM0653 Software upgrade from Standard to Advanced CT SB2 Kit VEHZ0696 Hardware extension to connect and test multi ratio CTs Primary Resistance Kit VEHZ0684 4 pol cable and Kelvin clamps RemAlyzer VESM0657 Software License Accessories Accessories part of Standard and Advanced Package Coax cables VEHK0651 with banana plugs 2x3 mMm 2 X98 Tt 1x10 m 1x32 8ft Battery clamps VEHZ0652 with 4 mm 0 2 in banana sockets primary side connection VEHZ0656 with 4 mm 0 2 in banana sockets secondary side connection 20 mm 0 8 in opening width 2 x red 2 x black Flexible terminal adapters a l VEHS0009 with 12 x4 mm 0 2 in banana socket Compact Flash Card VEHZ0654 128 MB Memory space for at least 416 test reports Compact Flash card reader VEHZ0655 USB 2 0 Compact Flash card reader User manual VESD0605 User manual Carry bag VEHPO018 CT Analyzer carry bag Grounding PE cable VEHK0615 1x6m 1 x 19 7 ft 6 mm 0 01 sq in protective earth connection USB RS232 converter cable p E EAS VEHZ0014 with Nullmodem cable Training CT VEHZ0643 300 5 class 0 5 FS 5 CT Analyzer PC soft
11. ls for burden and primary resistance tests CTs with up to six taps can be tested Automatic wiring check before measuring Use attached to the CT Analyzer or as a standalone unit VV VV SEAS Fesistan Excitati Ratio Wy Guessing Nameplates Location s Determination of unknown CT data Mt tilt ceeds tsiviecaanardiuedansd area mned aanetleuediveae gt Older CTs can be classified and put into service without on 4 Lome a contacting the manufacturer RSAstandard ANSI45 PM Pou gt Determinable parameters include fergie i Lee a i a Sean Resistan Excitati Ratio S RAG a gt Knee point 2000 08 ans 3 08 gt Power Factor ANGAS PMi Mn gt Nominal and operating burden VE FF T gt Winding resistance primary and secondary a oe Si Technical Data Technical features Standard Package gt Excellent noise immunity to disturbances from energized power lines close to the measurement gt Automatic assessment according to IEC 60044 1 or IEEE C57 13 up to accuracy class gt 0 3 gt Determination of ALF ALFi and FS FSi Ts and composite error for nominal and connected burden gt CT ratio and phase measurement with consideration of nominal and connected secondary burden gt Currents from 1 up to 400 of the rated value gt Different burdens full 2 4 burden gt CT winding resistance measurement primary and secondary gt CT excitation curve unsaturated and satu
12. pical 0 1 guaranteed 0 05 typical 0 2 guaranteed Environment Conditions Operating Temperature Storage Temperature 10 C up to 50 C 14 F up to 122 F 25 GUD TO 70 C 13 F upto Ibo Humidity Relative humidity 5 up to 95 not condensing Phase Displacement peeontilen aie EMC The product adheres to the Accuracy 1 min typical 3 min guaranteed electromagnetic compatibility EMC Directive 2004 108 EC CE conform Winding Resistance Resolution 1mQ EMC Emission Accuracy 0 05 typical International IEC 61326 1 Class A 0 1 1 MQ guaranteed Europe EN 61326 1 Class A USA FCC Subpart B of Part 15 Class A Power Supply Input Voltage 100 Vac to 240 Vac EMC Immunity Permissible Input Voltage 85 Vac to 264 Vac International IEC G1276 4 Frequency 50 60 Hz Europe EN 61326 1 Permissible Frequency 45 Hz to 65 Hz Input Power 500 VA Safety The product adheres to the low voltage Connection Standard AC socket 60320 Directive 2006 95 EC CE conform International IEC 61010 1 Output Europe EN 61010 1 Output Voltage O Vac to 120 Vac USA UL 61010 1 Output Current GA OSA 15 A ea Canada CSA C22 2 No 1010 1 92 Output Power O VA s to 400 VA wes 1 500 VA oak Certificates from Independent Test Institutes KEMA Test Report PTB Test Report Wuhan HV Research Test Report Physical Dimensions Size W x H x D 360 x 285 x 145 mm 9 2x7 2 x 3 7 in Weight 8 kg 17 4 lbs wit
13. rated gt Saturation characteristic recording gt Direct comparison of excitation curve to a reference curve gt CT phase and polarity check gt Secondary burden measurement gt Automatic demagnetization of the CT after the test gt Small and lightweight lt 8 kg 17 4 Ibs gt Short testing time due to fully automatic testing gt High level of safety using patented variable frequency method max 120 V gt Nameplate guesser function for CTs with unknown data gt Remote control interface gt QuickTest Manual testing interface gt Display readable in bright sunlight gt Simulation of measured data with different burdens and currents gt Easily adaptable reports customizable gt Knee point voltage from 1 V up to 4 kV can be l annal T measured a N CT anaves Additional features Advanced Package gt Automatic assessment for accuracy class gt 0 1 inclusive classes defined in the IEEE C57 13 6 standard gt Measurement of transient behavior of TPS TPX TPY and TPZ type CTs gt Automatic assessment according to IEC 60044 6 gt Determination of the transient dimensioning factor Ktd gt Knee point voltage from 1 V up to 30 kV can be measured Considering Duty C O C O C O e g auto reclosure system Technical data CT Analyzer OMICRON Current Ratio Accuracy Ratio 1 2000 Ratio 2000 5000 Ratio 5000 10000 0 02 typical 0 05 guaranteed 0 03 ty
14. ware toolset VESMO800 remote control software QuickTest Excel File Loader etc Power cord depends country dependent Additional accessories OMICRON Calibration CT VEHZ0649 2000 1 2000 5 Coax cable VEHK0657 with Kelvin clamps class 0 02 3m O01 Ty Pluggable winding VEHK0658 Pluggable 23 turns Coax cables VEHK0654 3 m 9 8 ft L winding VEHK0652 6 m 19 7 tt VEHK0653 10 m 32 8 ft VEHK0655 15 m 49 2 ft VEHK0656 100 m 328 1 ft with banana plugs CT SB2 upgrade kit ws i be i pr T ae a lt a VEHZ0696 CT SB2 inclusive accessories Transport case VEHPO068 Transport case with wheels Primary resistance Kit 0 VEHZ0684 4 pole cable 15 m 49 2 ft CT SB2 to CTprim 2x Kelvin clamps 11 OMICRON is an international company serving the electrical power industry with innovative testing and diagnostic solutions The application of OMICRON products allows users to assess the condition of the primary and secondary equipment on their systems with complete confidence Services offered in the area of consulting commissioning testing diagnosis and training make the product range complete Customers in more than 140 countries rely on the company s ability to supply leading edge technology of excellent quality Broad application knowledge and extraordinary customer support provided by offices in North Americ
Download Pdf Manuals
Related Search
Related Contents
Vignette - Papyrus - Université de Montréal Page 1 Page 2 Page 3 ティーマックス 一 045~アングルヘッド ` チタン製 Fujitsu STYLISTIC Q704 256GB 3G 4G Black SERVICE MANUAL HEATED LANDING ZONE (HLZ 18 Bedienungsanleitung Samsung C100 دليل المستخدم Rosco STSK7665 Manual Getting started with the digital MEMS microphones expansion board Guide des formations 2014 / 2015 - Paroisse Notre-Dame Copyright © All rights reserved.