Contents
1. mdi Stericheck SVP2 UGL SVZZP2X0935E H Incubate the Canisters VI Clamp off both tubing lines approximately 6 cm from the canister inlets Remove the tubing from the pump Caution To preserve the sterility of the closed system do not remove the needle with the tubing from the last media bottle This could allow contaminated air to enter the tubing and canisters Now cut the tubing approximately 2 cm above the clamps with sterile scissors so that the canisters remain closed to the environment Fold over and insert the tubing onto the air vents on top of the canisters Remove the canisters from the drain tray and incubate for time and temperature as recommended by the pharmacopoeia being followed Observe for the presence or absence of turbidity to indicate the presence or absence of microorganisms Page 18 of 18
2. the canisters UGL SVZZP2X0935E Page 5 of 18 HU User Guide EE A Set up the Steripump Set up the peristaltic pump Please refer to Steripump User Manual for more details Pic 2 mdi Steripump Unit UGL SVZZP2X0935E Page 6 of 18 User Guide mdi Stericheck SVP2 B Position the Canisters and Tubing on the Pump Open the sterilized plastic tray that contains the Stericheck canisters and tubing needle set by peeling back the lid in an aseptic environment Remove the canisters and tubing needle set from the plastic tray Place the canisters upright into the canister openings of the drain tray Pic 3 Stericheck canisters placed on the drain tray Load the canister tubing into the pump head Refer installation gt ipump head bAa Pic 4 Placement of PVC tubing in the Ster Open the pouch that contains the caps and plugs and place them inside the blister package or on a surface free of contamination UGL SVZZP2X0935E Page 7 of 18 User Guide mdi Stericheck SVP2 Steripump Ur Vented Needle Canisters Pic 5 Schematic Set up of Stericheck SVP2 C Prewet the Membranes Reference Document Please refer to the following United States Pharmacopoeia USP European Pharmacopoeia EP any other international pharmacopeia for information on the rinse solution you should use for your application mdi Steripump User M
3. A User Guide mdi Stericheck SVP2 for Ampoules or Collapsible Bags Advanced Microdevices Private Ltd 20 21 Industrial Area Ambala Cantt 133006 India Tel 0171 2699290 2699274 Fax 0171 2699221 2699008 Mail info mdimembrane com Web www mdimembrane com 1 Introduction UGL SVZZP2X0935E User Guide mdi Stericheck SVP2 Sterility testing is critical to quality control of sterile pharmaceutical product manufacture and has become all the more important because of the regulatory and commercial considerations International pharmacopoeias recommend use of procedures that ensure elimination of secondary contamination for accurate and reproducible results As per USP Pharmacopeial articles are to be tested by the Membrane Filtration method where the nature of the product permits The membrane filtration method involves filtration of samples drawn from an aseptically manufactured product lot through the membrane filter followed by washing away of any growth inhibiting substances cutting the membrane filter into two equal halves and incubating these in nutrient media suitable for growth of aerobic and anaerobic microorganisms The sterility of the lot under test is confirmed only if no growth is observed after the specified number of days of incubation The USP also states that the presence or absence of microbial growth observed as turbidity should not be due to any external contamination false posi
4. Turn on the pump Alternatively if using the foot control accessory press the pedal down with foot to keep the pump running VII Invert the rinse container and place it in the bottle support system UGL SVZZP2X0935E Page 9 of 18 User Guide mdi Stericheck SVP2 Bottle Holder Bar Stabilizing Bar Bottle Insertion Pic 7 Bottle Support System UGL SVZZP2X0935E Page 10 of 18 User Guide A O mdi Stericheck SVP2 UGL SVZZP2X0935E Pic 8 Indicative position of rinse container VIII Transfer the sterile rinse solution approximately 25 ml to pass on into each Stericheck canister IX Aseptically place a red cap onto the top of each canister s air vent to allow the rinse solution to start filtering through the membranes X Take off rinse solution container from the bottle support assembly and keep it upright on your work surface XI Take off the red caps from the top of each canister when approximately 1 cm 0 39 in of liquid remains in each canister Important mdi recommends to leave a small amount of fluid on the membrane at the base of canister to ensure that the test product is suspended in rinse solution before filtration XII Turn off the pump If using the foot control accessory lift your foot off the pedal to turn off the pump Prepare the Product Samples to be Tested l Decontaminate the surfaces of each product container carefully ll Remove Stericheck needle from the
5. anual for operating the peristaltic pump Important O Remove the red caps from the air vents on top of the canisters to fill the canisters O Replace the red caps on the top air vents while emptying the canisters Procedure to prewet the membranes l Decontaminate the surfaces of the rinse container carefully particularly the top septum area that is to be pierced later in this process lI Remove the protective cap from the Stericheck product testing needle Il Insert aseptically the Stericheck needle into the rinse container while holding the container upright IV Insert the vented needle into the rinse container so as to make an X with the SteriCheck needle UGL SVZZP2X0935E Page 8 of 18 User Guide mdi Stericheck SVP2 Pic 6 Indicative position of rinse container Important If using a flame swiftly flame the needle allow it to cool and then insert into the rinse container Insert the needle up to its base to ensure the needle opening is inside the container V Switch on the Steripump and adjust pump at the appropriate speed Caution The pump speed should be kept at moderate level as higher speed may allow the fluid to splash up and wet the canister vent This may cause water logging of the canister vent membrane and prevent the canister from filling Although special flow directors inside the mdi Stericheck canisters minimize the chances of this happening VI
6. e vent needle into the rinse container so as to make an X with the Stericheck needle Important VI UGL SVZZP2X0935E If using a flame swiftly flame the needle and cool it and then insert into the rinse solution bottle Set the pump speed at 55 or lower and turn on the pump If using the foot control accessory press the pedal with foot to keep the pump running Invert the rinse bottle and place it into the bottle support system Pic 11 Indicative position of rinse container Pump 100 ml of rinse solution into each canister Now turn off the pump Page 14 of 18 User Guide mdi Stericheck SVP2 UGL SVZZP2X0935E VII Remove the rinse solution bottle from the bottle support system and place it upright on your work surface Pic 12 Indicative position of rinse container VIII Caution Aseptically replace the red caps onto the top air vents of the canisters Keep the needles in the bottle Turn on the pump and allow the 100 ml of rinse fluid to filter slowly through the canisters until the canisters are empty Then turn off the pump Repeat the above steps continue rinsing 100 ml of rinse solution per canister at a time according to your validated standard operating procedure until the appropriate volume of rinse solution filters through the canisters tubing and membranes The needle from the rinse solution bottle should not be removed in orderto preserve closed
7. ge 3 of 18 User Guide mdi Stericheck SVP2 Specifications Material of Construction of Stericheck SVP2 Membrane Cellulose Nitrate Nylon 66 PVDF Membrane Pore Size 0 45um Canisters SAN Styrene Acrylo Nitrile plain transparent Tubing PVC plain transparent Filter on Vented Needle 0 2 um PTFE Membrane Filter on Canister Vent 0 2 um PTFE Membrane Canister Dimension 51 mm Diameter x 120 mm Height Water Flow Rate gt 0 3 lpm 10 psi at 25 C Sterilization ETO Sterilized Maximum Operating Temperature 45 C continuous Pic 1 Tubing with Needles and Vent Filter for SVP2 UGL SVZZP2X0935E Page 4 of 18 O User Guide EEE 3 3 1 Using Stericheck SVP2 The set up requirements and steps to be sequentially followed for using SVP2 Canisters are as below Procedure of U se Requirements a mdiSteripump b Stericheck SVP2 Canister pack presterilized c Product to be tested d Incubators e Sterile Fluid Thioglycollate FTG Medium 100 ml f Sterile Soybean Casein Digest SCD Medium 100 ml g Indian US European pharmacopoeia h Aseptic laboratory environment i Rinse solution as recommended in referred pharmacopoeia USP EP IP Other Steps A Set up the Steripump B Position the canisters and the tubing on the pump C Prewet the membranes D Prepare the product samples to be tested E Test the product F Rinse the tubing canisters and membranes G Add media to the canisters H Incubate
8. r Important If you are using a flame quickly flame the needle and allow it to cool before inserting it into the next container VII Turn on the pump If you are using the foot control accessory press your foot down on the pedal to turn on the pump The pump will return to its preset speed IX Repeat the earlier steps until the last sample in the lot is transferred and filtered X Now insert the SteriCheck needle into a sterile vented container empty rinse container and turn on the pump XI Continue pumping until all residual product clears the tubing and filters through the canisters Then turn off the pump Important In order to preserve the sterility of the closed system keep the needle in the product container until you are ready to transfer it to the rinse fluid container UGL SVZZP2X0935E Page 13 of 18 User Guide O mdi Stericheck SVP2 F Rinse the Tubing Canisters and Membranes Reference Document Please refer to the following United States Pharmacopoeia USP European Pharmacopoeia EP any other international pharmacopeia for information on the rinse solution to be used for your application The volume of rinse solution is predetermined during the validation procedure Procedure Remove the red caps from the air vents on top of the canisters Take out the Stericheck needle from the product container and insert it up to its base into the rinse solution bottle Insert th
9. rinse container Important Ifyou are using a flame quickly flame the needle and allow it to cool Page 11 of 18 User Guide mdi Stericheck SVP2 E Test the Product l Break open the ampoule with an ampoule cutter tilt it at 45 and insert the Stericheck needle Zs LL VA Pic 9 Indicative position of product container Il Turn on the pump Alternatively if using the foot control accessory press the pedal down with foot to keep the pump running Caution The pump speed should be kept at moderate level as higher speed may allow the fluid to splash up and wet the canister vent This may cause water logging of the canister vent membrane and prevent the canister from filling Il Aseptically place a red cap onto the top of each canister s air vent IV Pump slowly to allow the liquid in the ampoule to transfer to the canisters until only the tip of the Stericheck needle is submerged in the liquid at the bottom of the ampoule Turn off the pump before the ampoule is empty so that air is not drawn into the system V Using the foot control accessory is recommended to switch on off the pump UGL SVZZP2X0935E Page 12 of 18 User Guide mdi Stericheck SVP2 VI Decontaminate the second ampoule before testing as earlier Pic 10 Indicative position of product container VII Remove the Stericheck needle from the first container and aseptically insert it into the second containe
10. system sterility If needle is taken out contaminated air may enter the tubing and canisters Page 15 of 18 User Guide mdi Stericheck SVP2 G Add Media to the Canisters Sterile Fluid Thioglycollate FTG media and sterile Soybean Casein Digest SCD media in separate 100ml bottles with rubber septum caps are to be prepared and kept ready for transfer to the canisters Special pre installed color coded clamps on the canister tubing will help avoid mixing of the two different media types in the canisters during transfer l Remove the red caps from the air vents of the canisters Il Lift one of the canisters from the drain tray Aseptically place a red plug into the bottom outlet port of the canister Place the canister again in the drain tray and repeat the procedure for the other canister Important Secure the red plug firmly by twisting a half turn while pushing it into place on the bottom outlet port of the canister Il Clamp off one tubing line by closing the blue clamps pre installed Place this clamp as close as possible to the Y connector of the needle spiking device IV Clean and decontaminate the FTG media bottle septum Insert the Stericheck needle into the FTG media bottle V Also insert the vented needle into the media bottle so as to make an X with the Stericheck needle Important If you are using a flame quickly flame the needle and allow it to cool before inserting it in
11. tives or presence of any inhibitory substances false negatives Such situations that compromise the reliability of the test is a serious issue as retesting has very limited scope and frequent investigations and revalidations is a highly undesirable activity Therefore it is very important for the microbiologist to ensure that no external contaminants are introduced during the sterility test procedure and that it meets all the other pharmacopoeia needs This guide describes the procedure to use of mdi Stericheck SVP2 the closed sterility testing device for liquid ampoules Page 2 of 18 2 Closed Sterility Testing Device UGL SVZZP2X0935E User Guide O mdi Stericheck SVP2 Stericheck Closed Sterility Test System mdi Stericheck Closed Sterility Test System offers the complete sterility testing solution Right from sampling filtration media exposure until incubation the entire activity takes place in a closed loop to do away with the possibility of any extraneous contamination and therefore false positives The absence of anti microbial bacteriostatic components and complete washing away such substances in the drug product ensures no false negatives Advantages Fast Pre sterilized and ready to use No membrane handling No false positives or negatives The Stericheck system incorporates disposable Stericheck devices and a specially designed easy to use Steripump for aseptic transfer of fluids Pa
12. to the media bottle VI Turn on the pump Then invert the media bottle and place it in the bottle support system Transfer the media into one canister Caution Keep the pump speed at 35 40 when pumping Fluid Thioglycollate Medium to reduce aeration VII Turn off the pump when the tubing is clear of media VIII Remove the FTG media bottle from the bottle support and set it upright on your work surface IX Open the blue clamp from the tubing Clamp red clamp the other tubing line where media has just been transferred through as before UGL SVZZP2X0935E Page 16 of 18 UGL SVZZP2X0935E User Guide mdi Stericheck SVP2 X Remove the Stericheck needle and the vented needle from the FTG media bottle and insert these into the SCD media bottle Important If you are using a flame quickly flame the needle and allow it to cool before inserting it into the second media bottle Xl Turn pump on invert the media bottle and place it in the bottle support system XII Transfer the media from second bottle into the second canister Then turn off the pump when the line is clear of media Remove the media container from the bottle support and place it upright on your work surface Caution To preserve the sterility of the closed system do not remove the needle from the last media bottle as this can potentially allow contaminated air to enter the tubing and canisters Page 17 of 18 User Guide O