1. Pa n AMERICAN DIAGNOSTIC CORPORATION ADC Inspected in the U S A sl in cu e tel 631 273 9600 1 B00 232 5670 H NY 11788 el o31 2 5 1 OUU zaeni USA fax 631 273 9659 www adctoday com i One TH email info adctoday com United Kingdom IB p n 9359 00 rev 8 english
2. Adult Size 11 Arm Cuff Air Release Valve amp Air Tube and Con with size Inflation Bulb nector identifier Front Splash Resistant Housing Extra Large LCD Te Display Rear Systolic Assist On Off Zeroing Button Button Display Explanations 1 Systolic Assist Memory Indicator 2 Heartbeat Mark 3 Diastolic Pressure 4 Weak Battery Mark 5 Deflation Rate During measurement Installing and Replacing Batteries 1 Rotate the clip 90 2 Loosen the screw on the battery cover using a coin or screwdriver 3 Install the batteries with the polarities facing in the correct direc tion 4 Replace the battery cover and tighten the screw 5 Rotate the clip back Note You should replace the batteries if e The weak battery mark appears in the display e Nothing appears in the display when the power is switched on The supplied batteries may be discharged earlier than store bought batteries Always replace both batteries together Warning Remove batteries when the unit is not in use for extended periods of time Batter ies are hazardous waste please dispose of them properly Assembling your Sphygmomanometer Your inflation system is pre assembled and ready for use You will need to simply connect your digital gauge to the complete inflation system by inserting the twist lock connector pre assem bled on a bladder tube into the air port at the bottom of the digital gauge twist to lock Measur
3. infant or child the cuff can inflate to over 300mmHg with just two full compressions of the bulb To prevent discomfort or injury to the patient and damage to the instrument bulb should only be partially squeezed so that each stroke inflates the cuff in 40mmHg to 60mmHg increments until inflated to the desired level Warning If luer lock connectors are used in the construction of tubing there is a possibility they might be inadvertently connected to intravascular fluid systems allowing air to be pumped into a blood vessel Immediately consult a physician if this occurs Warning Do not allow a blood pressure cuff to remain on patient for more than 10 minutes n inflated above 10 mmHg This may cause patient distress disturb blood circulation and contribute to the injury of peripheral nerves A Warning Safety and effectiveness with neonate cuff sizes 1 through 5 is not established Assembly Your ADC DIAGNOSTIX e sphyg digital aneroid sphygmomanometer consists of a digital manometer gauge complete inflation system ADCUFF calibrated nylon cuff latex free infla tion bladder squeeze bulb and the ADFLOW valve a zippered carrying case and operating in structions Parts and Features Extra large LCD is easier to read Virtually indestructible Backlight display for viewing in any light condition Splash resistant housing for use in harsh environments Manometer works with virtually any cuff Gauge Hang Tab
4. or practice which if not corrected or discontinued immediately could lead to patient injury illness or death NOTE Systolic Assist and pulse measurement functions are NOT motion tolerant Pressure readings ARE In a dynamic environment systolic assist and pulse rate display may not operate correctly Caution Federal law restricts this device to sale by or on the order of a physician or li dsed healthcare practitioner A Caution Remove the batteries when the unit will not be used for an extended period ime Caution Your digital aneroid is a medical device designed for use by trained medical profes Boat Keep out of the reach of children Caution Leaking batteries can damage the unit A Caution Your digital aneroid is not field serviceable Caution You should not use any tool to open the device nor should you attempt to si anything inside the device Caution To stop the operation of your digital aneroid at any time open the air release valve ou would on any mechanical blood pressure instrument and the air in the cuff will be rapidly exhausted Caution Your digital aneroid is designed for normal blood pressure measurement at the Ber arm or thigh on healthy skin Proper use of this device is essential for accurate measure ment Caution Do not inflate your digital aneroid unit to more than 300 mmHg Warning When using with an infant or child cuff extra care must be taken to prevent over tion With smaller cuffs
5. ation of Digital Gauge 10 Error Displays and Troubleshooting 11 Care and Maintenance 12 Disinfection and Sterilization 12 Specifications 13 Stana Sia i ra ad Aa at at d n ea a eae 13 Limited Warranty 14 Product Registration 15 This manual is intended to assist the user in the safe and efficient operation of the e sphyg Sphygmomanometer The product must be used in accordance with the procedures contained in this manual and must not be used for purposes other than those described herein It is essential to read and understand the entire manual It is also important that your patients be acknowledged on blood pressure measurement described in Measurement Procedure Introduction Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the auscultatory method within the limits prescribed by the American Na tional Standards Institute ANSI AAMI ISO 81060 1 2007 Manual Electronic or Automated Sphygmomanometers Your new ADC e sphyg gauge provides you with all of the functionality of a traditional aneroid gauge with none of the problems associated with these mechanical devices Using state of the art technology your e sphyg gauge will
6. aused by abuse misuse accident or negligence Incidental special or consequential damages Some states do not allow the exclusion or limitation of incidental special or consequential damages so this limitation may not apply to you To Obtain Warranty Service Send item s postage paid to ADC Attn Service Dept 55 Commerce Dr Hauppauge NY 11788 Please include your name and address daytime phone no proof of pur chase a brief note explaining the problem and 2 00 to cover the cost of return shipping and han dling Implied Warranty Any implied warranty shall be limited in duration to the terms of this war ranty and in no case beyond the original selling price except where prohibited by law This warranty gives you specific legal rights and you may have other rights which vary from state to state To Register Your Product visit us at www adctoday com and follow the links FOR QUESTIONS COMMENTS OR SUGGESTIONS CALL TOLL FREE 1 800 ADC 2670 2013 ADC All rights reserved No one is permitted to reproduce or duplicate in any form this manual or any part thereof without the expressed written per mission of ADC ADC assumes no responsibility for any injury to anyone or for any illegal or improper use of the product that may result from failure to use the product in accordance with the instructions cautions warnings or statement of intended use published in this manual The 7002 8002 e sphyg is a trademark of ADC
7. e sphyg Digital Aneroid User s Manual d IP PP ad AMERICAN DIAGNOSTIC CORPORATION A Special Thank You Thank you for choosing an ADC Diagnostix e sphyg blood pressure instru ment We re proud of the care and quality that goes into the manufacture of each and every item that bears our name Only the finest materials are used to assure you of a timeless instrument de signed for optimum performance You ll quickly appreciate the results for you now own one of the finest sphygmo manometers that money can buy With proper care and maintenance your ADC blood pressure instrument is sure to provide you with many years of dependable service Please read the following instructions and general information which will prove helpful in allowing you to enjoy your ADC product Thank you for your patronage It is indeed our pleasure to serve you Sincerely American Diagnostic Corp ADC e sphyg Models 7002 8002 Table of Contents INTOUCTIONS 252 ola t a aa n a al a tees 4 5 e Intended use e Warnings and Cautions Assembly pie ai ones beer dees amant 6 Parts and Features sisi das a ae at a seu 6 Display Explanations T Installing and Replacing Batteries 7 Assembling your Sphygmomanometer 8 Measurement Procedure 8 9 Backlight Control mc ce ae died teste 9 Calibr
8. ement Procedure Auscultatory Method 1 Patient Position The patient should sit or lie comfortably The arm should be fully sup ported on a flat surface at heart level If the arm s position varies or is not level with the heart measurement values obtained will not be consistent with the patient s true blood pressure 2 Apply the cuff To attach the digital gauge plug the connector into the gauge and twist clockwise to ensure a secure connection The e sphyg gauge can be either clipped on the gauge hang tab or held in your hand during use ADCUFF nylon cuffs are specially designed to promote the accurate determination of blood pressure Index and range markings ensure use of the correct cuff size The artery mark indicates proper cuff positioning Place the cuff over the bare upper arm with the artery mark positioned directly over the brachial artery The bottom edge of the cuff should be positioned approximately one inch above the antecubital fold Wrap the end of the cuff not containing the bladder around the arm snugly and smoothly and engage adhesive strips To verify a correct fit check that the Index Line falls between the two Range Lines 3 Activate the digital aneroid and inflate the cuff Press the Oy On Off button to activate your digital gauge After zeroing your digital gauge is ready to measure Close the ADFLOW valve by turning thumbscrew clockwise Palpate the radial artery while inflating the cuff Be sure to i
9. f the e sphyg will not be used for an extended period of time remove batteries Disinfection and Sterilization Disinfecting Glutaraldehyde type liquid disinfectant may only be used on the cuff Follow instruc tions for use provided with the glutaraldehyde product Prolonged use of this disinfectant may cause discoloration DO NOT use glutaraldehyde type liquid disinfectant on the digital gauge bulb valves or bladder Sterilizing The blood pressure cuff may be gas sterilized DO NOT use steam or heat to sterilize the cuff bulb valve or bladder DO NOT attempt to sterilize the digital gauge Specifications Power Source MEASUREMENT RANGE Pressure Pulse Pressure Pulse Pressure Sensor Inflation Deflation Auto shut off Backlight auto shut off OPERATION ENVIRONMENT Temperature Humidity STORAGE ENVIRONMENT Temperature Humidity Dimensions Weight Battery Life Two 1 5V DC LRO3 AAA Alkaline Batteries 0 299 mmHg 30 240 beats min Accuracy 3 mmHg Within 5 of reading Semi conductor Manual Manual Air Release Valve 1 min after last key operation 15 sec after last key operation 32 115 F 0 46 C 85 Relative Humidity max 4 131 F 20 55 C 90 Relative Humidity max non condensing 3 L X 238 W X 118 H 3 14 oz G W w o Batteries Type BF 1500 readings with back light on 3000 readings with back light off Device and cuff are designed to provide spec
10. ial protection against electrical shocks Specifications are subject to change without notice Standards ADC s digital aneroid sphygmomanometer is accurate to 3 mmHg and conforms to applicable sections of the following international standards EN ISO 81060 1 2012 Non invasive sphygmomanometers Part 1 General Requirements EN ISO 81060 1 2012 Non invasive sphygmomanometers Part 3 supplementary requirements for electro ANSI AAMI ISO 81060 1 2007 Manual Definition mechanical blood pressure measuring systems Electronic or Automated Sphygmomanometers Definition Important Warning Caution Latex Free Authorized European Represenative s Information Manufacturer s Information Circumference Size Conforms to EU Standards Temperature Limit Humidity Limitation Limited Warranty American Diagnostic Corporation ADC warrants its products against defects in materials and workmanship under normal use and service as follows 1 Warranty service extends to the original retail purchaser only and commences with the date of delivery 2 The entire sphygmomanometer is warranted for one year 3 The manometer is warranted to function and remain accurate to 3mmHg over its full range when compared to a reference standard for life What is Covered Repair or replacement of parts and labor What is not covered Transportation charges to and from ADC Damages c
11. nflate cuff quickly by squeezing bulb rapidly Inflate cuff 20 30 mmHg above the point at which the radial pulse disappears 4 Position the Stethoscope Position the chestpiece in the antecubital space below the cuff distal to the brachium Do not place the chestpiece underneath the cuff as this impedes accurate measurement Use the bell side of a combination stethoscope for clearest detection of the low pitched Korotkoff pulse sounds 5 Systolic Assist Function and Pulse Rate Display Your digital e sphyg gauge is equipped with a systolic assist function to help you more easily determine the onset of systolic blood pressure As you deflate the cuff a heart symbol will flash when the pressure is within about 20 mmHg of systolic pressure Once you see the flashing symbol pay close attention for the start of phase 1 Korotkoff sounds Press the Systolic Assist button to record the systolic pressure in memory The digital gauge will also measure the patient s pulse while you are meas uring blood pressure and display this information on the LCD screen once the cuff has been de flated Note Systolic assist and pulse display features are NOT motion tolerant Excessive move ment may interfere with accurate display of these features Excessive movement will NOT inter fere with accurate determination of blood pressure using the e sphyg 6 Deflate the cuff Open the valve to deflate the cuff gradually at a rate of 2 5 mmHg per sec
12. ond 2 3 is preferred The display will show the average deflation rate in the lower right corner of the LCD screen Caution Excessive deflation rates in excess of 7mmHg will lead to inaccurate measurement and will make it difficult to pinpoint the readings on the display 7 Measurement Record the onset of Korotkoff sounds as the systolic pressure and the dis appearance of these sounds as diastolic pressure Some healthcare professionals rec ommend recording diastolic 1 and diastolic 2 Diastolic 1 occurs at phase 4 After measurement is completed open valve fully to release any remaining air in the cuff The digital gauge will now display the patient s pulse Remove cuff Calibration of Digital Gauge Gauge accuracy can be checked visually whenever it is switched on If the unit is used more than 15 000 times a calibration reminder message CAL will appear on the digital display Any digital gauge that displays this message should be returned for re calibration to ADC s Service Depart ment Only ADC or an authorized service technician is qualified to perform this calibration serv ice Do not attempt to recalibrate the gauge yourself In addition to the visual inspection indicated above we also recommend you perform the following calibration check on an annual basis e Connect your gauge to a high quality known pressure standard using ADC s Y Test Kit part number 991 e Pressurize the gauge to 300 mmHg e Reduce
13. provide you with performance and reliability exceeding the most popular aneroid models While traditional mechanical gauges can be easily knocked out of calibration when dropped your digital gauge will remain accurate even when subjected to rough treatment With its helpful on screen calibration reminder you will be able to use your new digital gauge with confidence and with the knowledge that it will remain calibrated beyond the lifespan of a more traditional mechanical gauge Your digital e sphyg aneroid also has several features not available in a more traditional me chanical gauge The systolic assist function will give you a visual alert when you are approaching the patient s estimated systolic pressure allowing you to focus your attention to detect the begin nings of phase 1 Korotkoff sounds The digital gauge will also monitor the patient s pulse during the measurement process and display this information onscreen for you to record We hope you enjoy the convenience and accuracy of your new digital e sphyg gauge Using state of the art technology your e sphyg will provide digital accuracy with tank like reliability Intended Use ADC aneroid digital aneroid and mercurial sphygmomanometers are used by professional healthcare providers and individuals trained in auscultatory blood pressure technique to deter mine systolic and diastolic pressure in humans WARNINGS A warning statement in this manual identifies a condition
14. s Remove and reinsert the batteries If this Data Error error keeps occurring return the device to your local distributor or service center Open the valve im mediately to rap idly exhaust the air in the cuff The cuff was inflated to a pressure over 300 mmHg Care and Maintenance Storage After measurement wrap cuff around gauge and bulb and store in zippered carrying case Cleaning Digital Manometer Your ADC DIAGNOSTIX brand e sphyg gauge requires minimal care and maintenance The manometer may be cleaned with a slightly dampened soft cloth but should not be dismantled under any circumstances Cuff Sponge with a damp cloth or alcohol pad If necessary it can be washed in cool water with mild soap or detergent disinfectant Remember to remove bladder from cuff prior to washing After washing it should be rinsed and allowed to air dry NEVER IRON Be sure to remove the bladder and all components before cleaning the cuff Bladder and Bulb Remove bladder from cuff and disconnect bulb from tubing and valves Wipe with a damp cloth To help preserve it from cracking or discoloring wipe with a soft cloth moistened with ethanol Valves Clean the surface with a dry fabric such as soft cotton material Remove dust from filter screen Carry Case Wipe with a damp cloth and dry thoroughly with a dry cloth If necessary it can be washed with cool water and mild soap Do not allow to air dry Batteries I
15. the pressure at a rate no greater than 10 mmHg per second and stop to check at the following test points 300mmHg 240 mmHg 160 mmHg 120 mmHg and 80 mmHg Your gauge should be within 3 mmHg of your reference standard ADC recommends using the most sensitive pressure standard possible when performing this calibration check as your ability to measure the accuracy of a gauge depends on the sensitivity of the pressure standard If you are using a reference pressure standard that is only calibrated to 3 mmHg such as a mercury column or aneroid gauge then you will only be able to deter mine the accuracy of the test gauge to within 6 mmHg If you use a reference pressure stan dard that is 01 mmHg such as a digital pressure standard you will be able to determine the accuracy of the gauge being tested to within 3 1 mmHg In the event that the gauge is ever in need of calibration simply return to ADC Damaged or bro ken parts will be replaced as needed at a minimum charge Refer to the warranty for specific de tails of warranty coverage ERROR DISPLAYS and TROUBLESHOOTING Error Message Description Solution The cuff was deflated Heartbeat rate too rapidly detection error Keep the deflation rate at about 2 3 mmHg per second Press the LO button A Pressure of over to switch the unit off or 15mmHg remaining the unit will for more than automatically switch 3 minutes is detected off after 60 second