1. Inspired CD technologies Bringing breath to life 1061 Main Street 24 North Huntingdon PA 15642 office 724 861 5510 fax 724 861 5530 www inspiredtechnologiesinc com 2007 Inspired Technologies Inc ViAspire Liquefier SmartDose technology patent pending ViAspire and SmartDose are trademarks of Inspired Technologies Inc 300P LTO7 REV 0 VIAspire Oxygen Portables 300 and 600 Product Manual CAUTION Federal USA law restricts this device to sale by or on the order of a physician Inspired technologies Bringing breath to life Welcome Welcome to the VIAspire Personal Oxygen System from Inspired Technologies The VIAspire Liquefier creates a steady supply of liquid oxygen in your home making deliveries of oxygen to your home no longer necessary By docking the VIAspire Oxygen Portables you can fill and transport your liquid oxygen quickly and easily letting you use the portable for your daily oxygen needs You no longer need to run oxygen tubing through your house and you are free to go wherever your day may take you This manual is for the VIAspire Oxygen Portables 300 and 600 Before You Begin Please read this entire manual before you use your VIAspire Liquid Oxygen Portables 300 and or 600 If you are unable to understand the warnings cautions and instructions contact your medical equipment provider before using this equipment Otherwise injury or damage to the e
2. 0 to 95 RH non condensing Operating 15 to 95 RH 18 VlAspire Oxygen Portable Manual 300P LTO7 REVO VIAspire Oxygen Portables 300 and 600 Product Manual Insert This insert applies to the use of the 300 and 600 Liquid Oxygen Portable with PB coupler The VIAspire 300 and 600 Liquid Oxygen Portable with SmartDose technology is also available with a PB coupler suitable for use with liquid oxygen LOX systems which have a male PB Coupler 100 LOX These portables with a PB coupler cannot however be used with the VIAspire Liquefier This inserts applies to the following Inspired Technologies part numbers 300P 0601 and 600P 0601 All the pages in the VIAspire Oxygen Portable 300 and 600 Product Manual still apply except for the following Page 1 The portables with the PB Coupler are not compatible with the VIAspire Liquefier Another source of LOX is required Pages 7 10 Filling a VIAspire Portable 300 or 600 These directions are not applicable because the Portable will be filled from another liquid oxygen system not the VIAspire Liquefier Follow the instructions from the Liquid Oxygen System on filling a portable All terms abbreviations and symbols safeguards cautions warnings notes and Indications for Use in this manual still apply 300P LTO2 Rev 0 page 1 of 1
3. Portable Failure to do so may result in damage to the unit or personal injury Use in accordance with your prescription For your safety the VIAspire Liquid Oxygen Portable must be used in accordance with the prescription determined by your physician Seek medical advice before using oxygen Under certain circumstances oxygen therapy can be hazardous Seeking medical advice before using oxygen is advisable 300P LTO7 REV 0 Warnings About Handling ENNE Use extreme care when filling Use extreme care A when filling your VIAspire Liquid Oxygen Portable ER Keep unit is in an upright position Always ensure that the portable is in an upright position A Turning the portable on its side will cause venting of oxygen from the relief valves Cautions and Notes Do not disassemble The ViAspire Liquid Oxygen Portable contains no user serviceable parts If service is required contact your home medical equipment provider or authorized service center gt Use only AA Alkaline type batteries Only use AA Alkaline type batteries Inspired Technologies does not recommend use of rechargeable batteries gt Federal law restricts this device to sale by or on order of a physician Not for use in emergency transport vehicles The ViAspire Liquid Oxygen Portable is not intended to be used in emergency transport vehicles ambu lances or helicopters or other similar areas where high levels of Electromagnetic Interferenc
4. cloth to clean and dry the couplers on both units Clean and dry On the VIAspire Portable turn the Patient Flow Indicator to the 0 position and assure that the CF SD switch is in the SD position VlAspire Oxygen Portable Manual 300P LT07 REV 0 Align the coupler on the bottom of the VIAspire Portable with the coupler in the dock area on top of the Liquefier Press firmly on the top of the VIAspire Portable keeping it in an upright position The VIAspire Liquefier and Portable are now connected 300P LTO7 REVO Open the Vent Valve on the VIAspire Portable to begin filling the portable Hold down the VIAspire Portable in place throughout the filling process The hissing sound you hear is the sound of some oxygen escaping this is normal You may see vapor around the connection This is normal and due to the great difference in tempera tures between the liquid oxygen and the warm air surrounding the ViAspire Portable ViAspire Oxygen Portable Manual When the portable is full close the valve You will know that it is full when the hissing sound changes tone and some liquid comes out of the vent line Lift the portable off of the VIAspire Liquefier If the portable does not separate easily do not use force The VIAspire Liquefier and Portable may be frozen together which sometimes occurs A sli
5. the red LED battery indicator will light up with each breath The red LED battery indicator will be lit solid not blinking with each breath if the batteries are totally depleted When the red LED indicator lights up replace the batteries as soon as possible or switch to continuous flow by changing the CF SD Continuous Flow SmartDose switch to the CF setting Continuous Flow mode does not require batteries 2 Adjust the cannula breathing tube to the proper position so that you will be able to breathe the oxygen comfortably The green LED battery indicator will light up with each breath regardless of the liquid oxygen con tents even if the portable has no liquid oxygen left 3 Turn the flow rate selection knob clockwise until the prescribed flow rate from 0 to 5 appears in the window Assure that the CF SD switch is in the SD position 5 To stop the oxygen delivery system turn the flow rate selection knob counter clockwise back to the off 0 position 12 VlAspire Oxygen Portable Manual 300P LTO7 REV 0 300P LTO7 REVO VliAspire Oxygen Portable Manual 13 Monitoring Battery Power The VIAspire Oxygen Portables 300 and 600 are each powered by four AA alkaline batteries A blinking red light on the LED means it is time to replace the batteries If this happens replace the batteries or switch to continuous flow by chang S ing the CF SD Continuous Flow SmartDose Delivery switch to th
6. both the VIAspire Liquefier reservoir and the VIAspire Portable 300 or 600 with a clean dry oil free lint free cloth before each portable fill Accessories Inspired Technologies recommends using only cannula spe cifically intended to be used with Oxygen delivery systems Cannula should be capable of delivering up to 8 LPM of con tinuous flow maximum length that can be used is 10 meters 35 ft Contact your homecare dealer for recommendations and details Refer to the cannula manufacturer s instruc tions for proper cleaning and maintenance procedures 300P LTO7 REVO VlAspire Oxygen Portable Manual 17 Transporting amp Storing VIAspire Portables Preparing for Shipping Prior to shipping the VIAspire Portable 300 or 600 remove the batteries and empty any residual oxygen from the unit Allow the portable to sit for at least 8 hours to allow it to empty Storage Store the VIAspire Portable 300 or 600 in a cool clean dry area when not in use Avoid storing in direct sunlight Battery disposal Dispose of all alkaline batteries in accordance with local state and federal regulations Specifications Batteries 4 AA Cell Required alkaline only Oxygen 22 25 psig Pressure LOX Content Model 300 0 33 liters approx Volume Model 600 0 6 liters approx Weight Model 300 3 5 Ib empty Model 600 4 5 Ib Temperature Storage Transport 20 to 50 C Operating 5 to 40 C Humidity Storage
7. e CF setting Continuous Flow mode does not require batter ies and the VIAspire Portable will now deliver a continuous flow of oxygen If the indicator displays a non blinking solid red light this means the batteries are dead and no oxygen is being delivered in SmartDose mode Using the VIAspire Portable 300 or 600 in continu ous flow CF mode will use up oxygen at a faster rate Monitor the contents indicator as described under Checking the Level of Liquid Oxygen to be sure that you do not run out of oxygen When using the ViAspire Portable in continuous flow CF mode the LED indicator will NOT light up green or red with each breath Turning the flow selector knob to zero will not affect the flow Turn the CF SD switch to SD mode to stop flow 14 VlAspire Oxygen Portable Manual 300P LTO7 REV 0 Checking the Level of Liquid Oxygen The VIAspire Portable has a built in tool to check the amount of liquid oxygen Follow these steps for using the Content Indicator Hold the short strap directly above the black Content Indi cator spring scale chamber on the portable as shown below Let the portable stabilize do not bounce or shake it then read the exposed color scale for the approxi mate amount of liquid oxygen If the portable is empty only the red colored band of the scale will be exposed In this case you need to fill the portable before you can use it ViAspire Oxygen Portable Manua
8. e EMI may be expected Have a full cylinder of oxygen on hand Inspired Technologies recommends the medical equipment pro vider leaving a full cylinder of oxygen with the patient after setting up the ViAspire Portable and instructing the patient to always keep a full cylinder on hand PE DED Allow portable to rest for at least hour after filling Inspired Technologies recommends that the A VIAspire Portable be allowed to rest for at least hour after filling prior to being used This allow the LOX to vaporize in the inner container and you will be able to better maintain your prescription flow setting 300P LTO7 REVO ViAspire Oxygen Portable Manual 5 Indications for Use The Model 300P and or 600P Liquid Oxygen Portable will provide a source of supplemental oxygen for ambulatory home healthcare patients by vaporizing medical grade liquid oxygen and then dispensing it to the patient via an integral electronic conserving device The Model 300P and or 600P Liquid Oxygen Portable is neither a life sustaining nor a life supporting device Provider Checklist Before leaving the VIAspire Liquid Oxygen Portable with a patient the following checklist MUST be completed O Check all parts for shipping damage In case of damage do not use Contact Inspired Technologies Inc for further instructions O Make sure the VIAspire Liquid Oxygen Portable is com patible with the liquid oxygen source Use a liquid oxygen reserv
9. ght twisting motion will release them Wait until the units warm up for 3 to 5 minutes and then they will separate with slight effort Do not touch any frosted parts If any liquid is leaking from the liquefier reservoir or the VIAspire Portable after separation reattach the portable immediately then separate the units again If liquid continues to leak leave the units connected and call your medical equipment provider Close the Coupler Access Door After a short period of time the VIAspire Liquefier will begin liquefying to replenish the liquid oxygen level Should any excessive leakage occur to the point that a stream of liquid is present leave the area and call your medical equipment provider immediately Liquid or cold gaseous oxygen can cause frostbite on the skin Oxygen will result in increased fire hazard For additional information and instructions on proper use of your ViAspire Liquefier refer to the user guide that came with it 10 VlAspire Oxygen Portable Manual 300P LTO7 REV 0 Selecting the Oxygen Delivery Mode The VIAspire Portables 300 and 600 automatically dose in two modes of pulsed oxygen delivery as well as continuous flow modes When you are resting and your breath rate is slower the VIAspire Portable 300 and or 600 are designed to deliver oxygen at the lowest flow rate possible This minimizes nasal dryness and also results in quieter therapy If you notice a change in the velocity of
10. l 15 Troubleshooting The following troubleshooting chart will help you analyze and correct minor system malfunctions If the suggested procedures do not help call your medical equipment pro vider Do not attempt any other maintenance EZANA To avoid electric shock hazard do not remove the portable case The case should only be removed by a qualified technician No indicator lights e Check that there are new batteries in the unit four AA Alkaline Batteries e Check that the cannula tubing is properly connected and that the tube is not blocked e Check that the CF SD switch is in the SD SmartDose position 16 VlAspire Oxygen Portable Manual 300P LTO7 REV 0 Cleaning and Maintenance Removing Frost Under certain environmental conditions frost may form on the vaporizer coil and on the case You can remove it by tapping the portable or wiping the case After allowing the portable to warm to room temperature unzip the bottom and remove the condensation pad Wring out and allow it to dry thoroughly before replacing Preventative Maintenance Contact your medical equipment provider annually to sched ule preventative maintenance for the VIAspire Portable 300 or 600 There are no user serviceable components inside the portable with the exception of the condensation pad Keep the exterior of the VIAspire Portable 300 or 600 clean using a clean damp oil free lint free cloth Clean and dry the couplers on
11. oir with a compatible coupler O Instruct the patient on the safe operation of the VIAspire Liquid Oxygen Portable and all associated accessories and review ALL warnings O Leave a copy of this instruction guide with the patient 6 ViAspire Oxygen Portable Manual 300P LTO7 REV 0 VlAspire Portable Overview The illustration shows the parts of the VIAspire Portable Flow Rate Oxygen Outlet Selector Knob Connector Continuous Flow LED Battery SmartDose Switch Indicator Contents Indicator Batteries 4 AA cells Fill Valve ALKALINE ONLY LOX Fill Coupler Filling a ViIAspire Portable 300 or 600 After the VIAspire Liquefier has liquefied enough oxygen you can fill your portable by following these instructions EZS f the Liquid Oxygen Coupler Access Door on top of the VlAspire Liquefier is opened and the unit continues to run do not fill the portable Contact your medical equipment provider to service the equipment The VIAspire Portable is designed to accept 93 pure liquid oxygen from the VIAspire Liquefier Most oxygen portables are not compatible Attempting to use an incompatible portable device may damage the VIAspire Portable or Liquefier or could lead to injury 300P LTO7 REV 0 VlAspire Oxygen Portable Manual 7 Instructions Open the Liquid Oxygen Coupler Access Door on the top of the V1IAspire Liquefier This will stop the liquefication process Use a clean dry oil free lint free
12. on for hazards that might cause serious injury Important safeguards are indicated throughout this guide pay special attention to all safety information Read all instructions before using Save these instructions 300P LT07 REV 0 General Dangers and Warnings In order to ensure the safe installation assembly and opera tion of the VIAspire Liquid Oxygen Portable these instruc tions MUST be followed Dangers i si Bl AA Fire hazard Do not lubricate Do not allow grease or oil from your hands or other source to come into contact with the internal tubing of the VIAspire Liquid Oxygen Portable These solutions may be flammable and cause injury Oxygen causes rapid burning Do not smoke in the vicinity of your VIAspire Liquid Oxygen Portable or when you are near a person utilizing oxygen therapy Keep the ViAspire Liquid Oxygen Portable at least 5 feet 1 6M from hot sparking objects or open sources of flame Do not lay the cannula down Do not lay the cannula down while the oxygen is flowing High concentrations of oxygen can cause rapid burning Warnings al 300P LTO7 REVO Read instruction manual first Do not install assemble or operate this equipment without first reading and understanding this instruction manual and the warning labels on the product If you are unable to understand the warnings cautions and instructions contact your medical equipment provider or technical personnel if a
13. oxygen on each breath or a quieter delivery of oxygen you have probably slowed down your breathing pattern and the SmartDose Technology has adjusted the flow of oxygen to the appropriate level Sport Delivery Mode with SmartDose technology The VIAspire Portables 300 and 600 have SmartDose tech nology which will track your breath rate to determine if you are exerting more energy and therefore may need additional oxygen When you are at rest the VIAspire Portables 300 and 600 will deliver a resting volume of oxygen As you begin to exert your breath rate will climb and the portable will deliver additional oxygen While many oxygen portables require the patient to turn the oxygen flow rate selector knob up and down as they move and rest the VIAspire Portables 300 and 600 will adjust oxygen volumes to your needs automatically 300P LTO7 REV 0 VlAspire Oxygen Portable Manual 11 Starting Oxygen Delivery fromthe 4 each breath you should now be receiving a set dosage VIAspire Portable of gaseous oxygen To receive oxygen from your VIAspire Portable follow these steps The green LED battery indicator will light up with each breath signaling that the VIAspire Portable is being trig gered to deliver oxygen The LED monitors only battery power and does not indicate level of liquid oxygen available 1 Push the cannula breathing tube in reservoir firmly onto the oxygen outlet port When it is time to replace the batteries
14. pplicable before attempting to install or use this equipment otherwise injury or damage may occur Liquid Oxygen used ONLY by patient The use of this device is limited to an oxygen patient Liquid oxygen MUST be used ONLY by an oxygen patient and is not intended to be distributed to any other individual for any purpose ViAspire Oxygen Portable Manual 3 warnine A A 4 ViAspire Oxygen Portable Manual Use only authorized parts Do not use parts accessories or adapters other than those authorized by Inspired Technologies Inc Do not use if Do not use the ViAspire Portable 300 or 600 if it Is not working properly Has been dropped or damaged Has been submersed in water Call a qualified technician for examination and repair Unintended oxygen leakage If any unintended leakage of oxygen is detected do not attempt to use the product Turn the product off If leakage persists place the product outdoors and notify your Home Health Care Provider or Service Representative of this condition High pressure relief valves The ViAspire Liquid Oxygen Portable is equipped with several High Pressure Relief Valves to ensure the user s safety When activated this safety feature will make a loud noise If this noise occurs and persists turn the unit off and contact your Home Health Care Provider or Service Representative Always supervise children Children should always be supervised around the VIAspire Liquid Oxygen
15. quipment may occur When using electrical and oxygen products especially when children are present you must follow some important safety precautions Please read the following before you start using your VIAspire Portable Product Classification The VIAspire Liquid Oxygen Portable is classified as oO IEC Class II equipment tT Type B Applied Part Not suitable for use in the presence of a flammable anesthetic mixture with air or with nitrous oxide 300P LTO7 REVO VlAspire Oxygen Portable Manual 1 Terms Abbreviations and Symbols This guide and the VIAspire Portable product labeling uses the following terms abbreviations and symbols IEC Symbols A m t A Smoking is not allowed in the area REF Indicates a Model Number General SN Indicates a warning sign Serial Number Class I Indicates Equipment LOX Liquid Oxygen Type B Follow Operating Equipment Instructions Warning A Fire dangerous voltage Hazard Important Safeguards 2 ViAspire Oxygen Portable Manual When using electrical products especially when children are present basic safety precautions should always be followed Read all instructions before using Important information is highlighted by these terms CGT information for preventing damage to the product Urgent safety information for hazards that will cause serious injury or death Information to which you should pay special attention Important safety informati