TWCS Series User Manual
Contents
1. The Torque Wrench Calibration System is a high precision machine designed to measure and verify the calibration of hand torque tools Considered one of the most flexible and user friendly calibration systems available LSI s TWCS is offered in three models a Standard model TWCS Super model STWCS and an Ultra model UTWCS The TWCS model operates with one load cell on a single unit Select from five available torque capacities The STWCS features interchangeable smart torque cartridges Switch load cells easily when a different torque capacity is needed and enjoy the convenience For flexibility without changing out cartridges the UTWCS allows up to three stationary load cells to test hand torque tools of different capacities all on the same unit Instantly change torque capacities with the push of a button Available torque cartridge capacities range from 3 in lbs to 260 ft lbs depending on model In all models the load cells are protected during use by mechanical and software stops to safeguard from overload damage Each model includes three test modes Manual for needle indicating tools Peak for dial indicating or digital tools and First Peak for click type torque tools Unique to the industry the STWCS and UTWCS both feature an on screen Calibration Assistant program to guide the user through the calibration steps and print a certificate compliant with ISO 6789 standards Connect to any parallel printer with the RS232 cable and adapter
2. 2 4 1 Determine Model Being USed oconconnccnnncinncnoncnnncnnnconecnnnconenonenonnnrnnn ron EA E A 12 2A VIABREA IEEE re O A E E A aa 12 ZAS SS IWESTUTWESUSES E E A E A giao eat 12 3 TWCS USER GUIDE iaa 13 m ea a D eS Oe alie M nes Inle Aeee E a da 13 3 2 Test Modes 22 ana ia dada 14 32 Malal ii A A A 14 IL A A A A A aaa a oe OTe cat dp N E Re otter 14 ZONA ENSURE Sao ot ees Bele te i E o Re ts 14 32 4 gt Shunt Calibracion od 14 3 3 TWGS Calibration Test Procedure iivc st ccsccccecteden catteeestececacdecseeisntaceedereseaeeetdcsvutedeaneesctiesvadesctenevenvcestacsussceuacte sa 15 IT PREFEQUISILES ieee cloned E E E avs Sec state dao 15 3 3 2 gt ConmnectiS PrMt o a 15 3 33 Pre Galibration W Or S E A a A e Ea ias 15 334 O T p ici E eee sah A A EE E ES 15 390 0 Select Test Mode EE EE E E E T A E 15 3 3 6 Exercise Torque Bo e M ENE E A E AAEE E E E A 15 IM LOTO QUE EEES L 1 cc A A T ce adhe es See EEE E aa at ee 15 33 97 Mount Torque Eolo ANA EE E ai E A A E E 15 3 33 9 Test Ko Ke ULE Toldo A aad nine eet 16 33 10 Print T st Point DI i E ii T A T E eed eee 17 3 3 11 Calibration Documentation ooooocnnoccccnoooccncnooonnnnononnnnnnonnnnnononnnconnnnnnnnnnnnnnnnnnnnnnnnnnnnnononnnnnnnnnnnnnnnnnnnnannnnnns 17 4 STWCS UTWCS USER GUIDE coo a nnmnnn 18 At Tihe Gontrol Panel e a EE ca 18 AZ Test Mo des Sect E S TATEA Ee A TASE PTEE EEA E TAE TEE T TATE 19 A MaN b a E A e e 19 A Pt
3. screen will look like the first example below The screen indicates reading 1 of 5 is ready to be taken at the 20 in Ib test point mark and the unit is in Peak mode The reading will be viewable until stored following the directions above then the second screen below appears stating which number of test is being stored Screen display when in Peak mode from initial screen and beginning testing after machine is turned on This screen does not show at what point the series of test point measurements is but the display when storing a reading works the same as above Readings stored may be printed but will not be in a Calibration Report Certificate format Store up to 1 800 test point readings with the built in data logger 4 2 3 First Peak First Peak is used with click type torque tools Much like Peak mode First Peak dynamically adjusts the display to reflect the torque applied When beginning at the Cal Certificate menu and Start is selected as force is applied to the tool being tested and increases the display moves through the first series of screens below If the torque decreases for an instant the display stops at the highest read torque and indicates the test is finished The reading can no longer be adjusted Press STORE to save To reset the First Peak mode display press ON CLEAR then ZERO TORQUE to begin testing again After the specified number of test points has been stored a calibration report will automatically print if
4. Number 200 1 Calibrated By Phone Number Date Calibrated Recalibration Date Your Company Name 888 555 1212 05 09 07 05 09 08 Testing Procedure Wrench Capacity 200 00 in 1b Number of Measuring Points 3 Number of Tests per Point ae Wrench Tolerance 45 Testing Performed in First Peak Mode This is to certify that the torque assembly tool accompanied by this document has been calibrated to Standards traceable to the National Institute of Standards N I S T Calibration performed in accordance to ISO 6789 2003 E Calibration Data Measuring 40 00 120 00 200 00 Points in 1b in 1b in 1b Test Deo RI ZA A a a re a a a e a 1 36 46 122 48 198 34 2 36 96 120 36 197 34 3 36 08 118 00 198 46 4 37 10 117 28 201 48 5 37 24 118 60 200 98 Se E a a r POAN SE E A D E AR A A AA A EEE es es 4 41 60 124 80 208 00 High 37 24 122 48 201 48 Low 36 08 117 28 197 34 4 38 40 15 20 192 00 In Tol FAIL OK OK TWCS Series User Manual Page 35 of 44 060 1000 0045 00E 4 8 Calibration Data Template REPORT OF CALIBRATION Name and Location Wrench Type Serial Number Performed By Date Time Temperature Humidity Recalibration Date Calibration Interval Months Wrench Capacity Number of Measuring Points Number of Test Per Point Wrench Tolerance Testing Mode Manual Peak First Peak Circle One Counter Clockwise Torque CCW Cloc
5. for a test point adjust the tool to the next test point setting or mark Easy Loader Attachment installed Adjust the tool to the appropriate test point mark and place onto the STWCS UTWCS with the arm of the wrench extending parallel to the Easy Loader Attachment Loosen the quick release lever and adjust the slide so the wrench handle is supported in the middle Then adjust the actuator wheel to give the maximum amount of travel in the direction of the test Insert the stop pins so the wrench is contained on both sides and tighten the quick release lever to hold the slide in place The stop pins should be located on either side of the wrench handle such that when the tool reaches its click point the stop pins meet as close as possible in the middle of the handle of the torque tool This center point on the tool handle is identified by a line or marking The length of a wrench may change when changing test point marks and will require readjusting the slide to the center mark of the handle See diagram of Easy Loader in section 3 3 8 4 7 14 Test Torque Tool Refer to the Force Application diagrams in section 5 Technique for Good Measurement for the correct ways to test single arm torque tools Using the Easy Loader attachment turn the actuator wheel so the wrench is moved in desired direction for testing To start test over press ON CLEAR If satisfied that the test was performed correctly press STORE Repeat steps 4 7 11 4 7 1
6. needs to be selected Choose CW clockwise or CCW counterclockwise under the User Setup menu See section 4 6 5 6 Rule 4 7 8 Setup Calibration Procedure Press OPTION then F1 for the setup function See section 4 6 1 Use the Pre Calibration Worksheet completed to enter the settings that follow 4 7 8 1 Torque Tool Capacity Enter the full capacity of the torque tool under Wrench Cap using F1 and F2 When finished press OPTION for the next setting The following commands are used to navigate F1 Go to next space F2 Increase digit UNITS Backspace ZERO ANGLE Decrease digit OPTION Next setting MODE Previous setting 4 7 8 2 Number of Measuring Points Select the number of measuring points using F1 or F2 choices range from 1 10 The number should be a minimum of 3 The point of measurement will then be automatically calculated for you This is the mark the tool should be adjusted to for testing For example if the tool has a capacity of 200 in lbs and 4 measuring points is selected the marks to measure will be divided evenly at 50 in Ibs 100 in lbs 150 in lbs and 200 in lbs Press OPTION to continue 4 7 8 3 Verify Measuring Points Verify that the measuring points are at the desired values The First Point Second Point and Third Point settings etc will display by pressing OPTION for each and may be adjusted by choosing a different number of measuring points When finished
7. specified or press SEND to manually print Screens displayed when in First Peak mode from initial screen and beginning testing after machine is turned on The screens do not show at what point the series of test point measurements is but the display when storing a reading works the same as above Readings stored may be printed but will not be in a Calibration Report Certificate format Store up to 1 800 test point readings with the built in data logger TWCS Series User Manual Page 20 of 44 060 1000 0045 00E 4 3 STWCS Interchangeable Torque Cartridges 4 3 1 Lock Test Position An interchangeable cartridge with a collar around the top of the cylinder marked Lock in one direction and Test in the other should be turned to the locked position when not in use A collar with the lock feature is standard on the lowest capacity load cell cartridges 3 in Ib 25 in lb as they are extremely sensitive and may be easily damaged 4 3 2 Changing Cartridges Smart Torque Cartridges may be changed with the unit turned off or while in Option mode There are four screws around the top of the cylinder of the cartridge that must be removed After removing the screws lift the entire cylinder that houses the load cell out unplugging it from the STWCS Take the new cartridge and set it inside the STWCS with the connector lined up from the bottom of cartridge to the inside of unit Using gentle pressure plug the cartridge in The mounting
8. the STWCS UTWCS and is a procedure that should be done regularly whether daily or before each use according to your company policy Press OPTION to find this menu and follow the instructions in section 4 6 4 4 5 4 Calibration Assistant Go to section 4 7 Calibration Assistant and follow the instructions This leads through the entire calibration procedure including any setup and preparation the actual testing of the torque tool and printing out a customized calibration certificate Easy Cal Report 4 5 5 Going Forward Please read through entire User Manual There is detailed information regarding other functions care and maintenance techniques for achieving reliable readings etc Please contact LSI with any questions 4 6 The Option Menus Press OPTION on the control panel to display the first of several Option menus Scroll through each menu by continuing to press OPTION General information about each of these menus is described in order of appearance in this section Navigate the Option menus using the following commands OPTION starts the main menu and is used to scroll through the menu items MODE scrolls in reverse through the menu items F1 and F2 are used to make selections and also to navigate and enter data SEND backs out of menu levels to the main menu UNITS does the reverse of F1 when entering numbers or letters in a field ZERO ANGLE does the reverse of F2 when entering numbers or lette
9. to the displayed value and displaying that as the new value The percentage that the error is multiplied by is Kd Kd changes over time While the force is stable Kd decreases from the Kd entered in the User Setup menu to the Dampen Min M value at a rate defined by the Decay Rate C The Recovery R value is used to increase Kd when the force starts to change based on duration and amount of change The Kd value will be shown as it changes over time when Display Filter is chosen 4 6 5 6 User Setup Rule Rule Choose Rule using F1 Go to Torque Rule using OPTION Angle Rule not active on this model Select CW Positive or CCW Positive using F1 or F2 CW Clockwise CCW Counterclockwise This corresponds to the direction the tool is being tested on the unit and displays the readings taken in the direction selected as positive numbers This setting may need to be changed frequently When performing tests to result in a Calibration Certificate for signature it is important that the readings are positive numbers If performing the test in Clockwise direction choose CW Positive Counterclockwise CCW positive If negative readings are recorded the Calibration Certificate will always produce a FAIL test for the torque tool When done press SEND to return to the previous menu Press SEND again to return to the User Setup main menu 4 6 5 7 User Setup I O Polarity 1 0 Polarity The 1 O Polarity menu acc
10. 3 the number times required for the test points that was specified in the setup procedure Remember to adjust the tool to the correct mark for each test point See section 4 6 1 Cal Certificate for additional test procedure information Note To achieve reliable results consistently there are controlling factors to be considered and adjusted if needed The actual performance of each test is key and requires consistent movement of the torque wrench by the user For this reason the Easy Loader Attachment is recommended so testing is consistent no matter who the operator is There are also software settings regarding filtering and dampening that may contribute to the quality of the test results See section 5 Technique for Good Measurement 4 7 15 Print Calibration Certificate Upon completion of the last test point when all readings specified have been taken the STWCS UTWCS can automatically generate a printed calibration certificate All readings must be within the specified tolerances to conform to ISO 6789 1992 E Your Company Name Anytown USA Report of Calibration Serial Humidity Wrench Type Temperature Performed By Final Approval Date 08 06 07 Time 04 01 23 pm Recalibration Date 08 05 08 Calibration Interval 12 months Certified Calibration Equipment Used Device Larson Systems Inc Super Torque Wrench Calibration System Device Serial Number 1234567890 Torque Cartridge 200 00 in lb Cart Serial
11. A A LR A E E Seer cece 19 AS A a 20 4 3 STWCS Interchangeable Torque Cartridges ccccsccsceecsseseeeeseeeeeeeeeeeeeeeeeesseesseesseeseeeeeeseeeseeeeeeeeseeseenteaes 21 4 3 KOCK TeStPOSITION 4 28 A ee do eat hd a alee leh Ad od bank de leh ee ed 21 4 3 2 gt Changing Cartridge a 21 Asi BS STONING 5 2511 a Sekt oekaki ar eel Mu aah santa oa iad EAE el My al outs liad iia At Sy ae 21 AA UTWCES Load Cell Operati saia a A ia 21 4At Tum onthe UTWCS eese a a A A ceed pee cee aaa eari 21 4 42 Switch Load CellS ocio denies tae ele eae ele ite ce ce ee 21 4 5 Quick Start to Torque Tool TeSting cccscscsccseeeceeeeeeeeeeeseeeeeeeeeneeeeeeseseeesseeeseesenesseesseeesneeseesseesseeesneeeneeeneeeaes 22 4 5 1 T men STWESTUTW OS e r ra r Eaa edna ees a a anA a A tee detec eS 22 EZ USES AU A E E 22 46 3 Cali CHECK E E E E A ES 22 45 4 Calibration ASSIStM Ei aaa 22 CN GORG FONN O NN 22 46 The Option Menus ici iia 23 4 6 1 Cal Certificate Setup Start siii a aa eaa a a ae a nene ieai r ee aiaee 23 4 6 2 Option Store LENAPE ei a aida 24 4 6 3 Option Offset PoWer ooiiii taa 24 AGA Option A A 24 4 6 5 Option UsSer Setup et etsa bes A toot eat das ca ta a eS he Sak bad Lo Ms e ok Maas a EER at 25 4 6 6 Option Diagnostic isccct2e5 scl ei A lee hed dod Ses soe ed Sas Moe dy hed cod Sed toed aU aag lead loeb 29 AG Option CAD MENUS isot Pot ct Soak ahh Daa cat Oud tak ets glad ad Dedos Sal te to Nat ods Thad bab eta E
12. Larson Systems Inc LSZ Torque Wrench Calibration Systems User Manual for STWCS UTWCS TWCS Series User Manual Page 2 of 44 060 1000 0045 00E About This Manual This manual could contain technical inaccuracies or typographical errors Changes are periodically made to the information contained herein These changes will be incorporated in new editions of the manual Copyright 2007 Larson Systems Inc All rights reserved No part of this manual may be reproduced by any means without written permission of the author except portions necessary for internal use only by the purchaser of the LSI system T INTRODUCTION Susurros a a a aa 7 1 1 Standard TWCS Features a NaS 8 141 417 TWES Torque Specifications e e r ect entities a e a E E tests 8 1 2 Super TWCS STWCS Features rririk ana 9 1 2 1 STWCS Torque Specifications esi a id ee ia lado Dota a 9 1 3 Ultra TWCS UTWCS Featur8S comicios 9 1 3 1 STWCS UTWCS Torque Specifications cccccceccececceceeeeeeeeeeeeeeeeaeeceaeeceaeeeeaeeeseaeessaeeseaeeseneeeseeeeseneeeeaes 9 T4 ETA a raa Ea T Eaa aa Aaa E Aa A Aaa aaa aaa a aaa a aaar PaA AAAA AAAA aa deaa Eaa adadan saani iias 10 1 5 Tester DiM NS ONS A a a cab ees 10 2 GETTING STARTED ocoiutaisiddliii 11 VAa gt UNPACKING AEA O E E A E A T E A A E E E E TT 11 PAP R E Eea Toe aac E T E E A A A A A S 11 23 E Ee ii TE A A T 11 24 Usingthe Manual ieee aE debe cies ded estw eves AEEA INE Aaa E AAE NA ENEE EEN 12
13. User Setup menu See section 4 6 5 User Setup and 4 6 5 4 Units Performs reverse direction of F1 when using Option menus OPTION Initiates display menu system for entering settings and selecting functions See Section 4 6 for more information ZERO ANGLE Performs reverse direction of F2 when using Option menus Angle function is not active ZERO TORQUE Zero out current torque reading before starting a new measurement STORE Store up to 2 000 test points with built in data log The data can be later sent to a printer or computer via RS232 cable interface SEND RS232 output can be enabled to send test point data to a printer Also used to back out of menu levels in the Option menus OFF Turns the STWCS or UTWCS off ON CLEAR Turns the STWCS or UTWCS on and acts as an enter or save settings button in the Option menus Also resets the Peak or First Peak mode Note Lights at top of control panel are not functional on STWCS or UTWCS equipment 4 2 Test Modes Determine the test mode appropriate for the tool being tested Select the test mode by pressing MODE until desired mode is displayed Each of the test modes and how they work are described below with display screens referenced Mode can be selected and used at the initial screen after the machine is turned on or in the Cal Certificate menu when Start is selected See section 4 6 1 If unsure what mode should be used for testing a particu
14. a timesaving plus for the management of calibration verification and ensuring torque tools are operating within the specified guidelines The STWCS and UTWCS also include a built in computer that can access calibration linearization and a 1 800 test point data logger To allow for portability and ease of use a rechargeable battery with an eight hour capacity is standard on the STWCS and UTWCS and is optional on the TWCS An A C adapter provides auxiliary power and recharges the battery LSI s TWCS model series is equipped for measuring in English and Metric units including the Newton 25 pin RS232 Interface 120V AC Adapter Battery Recharger TWCS Series User Manual Page 7 of 44 060 1000 0045 00E 1 1 Standard TWCS Features e Three Test Modes Accommodates calibration for all types and classes of torque tools needle indicating dial digital indicating or click type e Fourth Mode for Shunt Calibration Verification Easily validate calibration accuracy of tester e Accuracy Better Than 0 4 10 100 of full scale e Push Button Zeroing Zero out torque reading before each new test e Backlit LED Display Easily and comfortably view test results e Digital Calibration No tools required to calibrate from control panel e NIST Traceable Calibration In compliance with ISO 6789 2003 E clear _OFF e Clockwise Counterclockwise Operation Test in either direction e Vertical Horizontal Mounting Horizontal mounting is standard
15. ained and there is not comparable packaging available at your location Larson Systems Inc 763 780 2131 10073 Baltimore Street NE 1 877 780 2131 E Minneapolis MN 55449 4425 FAX 763 780 2182 www larsonsystems com info larsonsystems com TWCS Series User Manual Page 44 of 44 060 1000 0045 00E
16. cially if different people are performing the calibration tests The Easy Loader ensures a slow steady movement is carried out for every test Keep in mind that the location of the stop pins in relation to the handle of the tool should be as close as possible to the center mark on the handle at the point reading is taken or the point of the click 5 7 Force Application Diagrams Figure 1 Figure 1 Illustrates proper movement when testing a wrench in a horizontal mount position Figure 2 Illustrates proper movement when testing a Figure 2 screwdriver Figure 3 Illustrates proper movement when testing a wrench when device is vertically mounted Figure 3 TWCS Series User Manual Page 40 of 44 060 1000 0045 00E 5 8 Dampening and Filtering Settings If unable to use the STWCS UTWCS in an environment conducive to consistent readings dampening and filtering settings may be used to compensate Not applicable to the TWCS 5 9 Verify Calibration Make sure the machine is not out of calibration The TWCS calibration can be verified in the Shunt Test Mode See Section 3 2 4 Calibration verification of a an STWCS UTWCS can be done in the Option Menu under Cal Check See section 4 6 4 for instructions 6 TROUBLESHOOTING AND ERROR CODES Error 4 Past Full Scale Load cell is being stressed beyond capacity Release from testing Damage may occur Error 75 Device Not Enabled SEND pressed without a printer or PC i
17. cter field by pressing F1 and backup using UNITS Press F2 to advance through and make selections from the alphabet in lower case characters numerals spaces and the alphabet in upper case To navigate through selections in reverse press ZERO ANGLE Once finished with a single screen advance to the next screen by pressing OPTION To go back to a screen press MODE The first four screens are used to enter the name of the calibrating company The next four represent the calibrating company s phone number The last two screens are for the date calibrated and the date recalibration is due This information will appear on the Report of Calibration certificate generated for a torque tool being tested 4 6 6 Option Diagnostic Diagnostic This function allows user to troubleshoot the tester and print out reports that may be helpful in diagnosing a problem Press F1 to choose Move through each diagnostic category by pressing OPTION 4 6 6 1 View Batt Volts Choose to view the battery volts using F1 or F2 The best operating voltage is 6 2 to 7 0 volts Press OPTION to continue TWCS Series User Manual Page 29 of 44 060 1000 0045 00E 4 6 6 2 Keypad Check Press F1 to check the keypad The next screen above will flash and stop at Press any key Check the keys by pressing each one and leave ON CLEAR till last to exit As each key is pressed the display should show its name if not the keypad is n
18. eeeeseeeeeeeeeneeesaeeeesaeeeeaeeeeeeessaeesenaessaeenseeeeaaes 37 ALTO Tool Types and CIMAS OS A iii 38 5 TECHNIQUE FOR GOOD MEASUREMENT ccccccceseeeeeseeseeeeeeeeeeeeeeseeeeeeesseaeeeeesees 39 51 The ENVIO NAME A aint talca 39 5 2 Proper sa ia 39 DiS A TT 39 5 4 Using Worksheets Provided ci tt tecatezeceencerscecescesnageneVuteceusdecueaceescedench snag ane ated aseecusanchscedanceaned crias 39 539 Torque Tool Caliber ato rns in aci ii 39 5 6 A A canned antceahectsaacedsuvecsuectcoar ct antcesicazsaces seneeccpeetvececsanieee 39 57 Force Application DIAGTaMS ic 40 5 8 Dampening and Filtering Settings oonmmniiniin ninos 41 5 9 Verify Calibration iaa 41 6 TROUBLESHOOTING AND ERROR CODES ooooocccccccccnnooccccccccccnnnnnnnananannnnnnnnnnnnnnnannnnnnnnnns 42 T CARE AND MAINTENANCE scscsctecesccctscetccctetessnezecetccetecescnscedstasecatetcsececstavecsdetcnececusaaeceseies 43 AAA eo noo Aza 43 7 2 Calibration Verification and Calibration Options oomnconnicnninnnnncnnnnnnrecnrcrcncecnnerc nanena 43 7 3 Packaging for Shipment to LSI 0 cccceecceeeseenseenseeeeeeneeeeeeeeneeseeeseasseasseeeseeeseeeseeesaeeseesseaseeasseeseneeseeeseeeseneeneeeaes 43 13 1 Andividual torque cartridge S xico alla la lp da 43 13 25 Conta Panels Circuit Boards scort ias 43 33 Packaging Entire Testing Unit Loi ia 43 1 INTRODUCTION STWCS shown with optional Easy Loader Attachment ZERO option ZERO A
19. eflect the torque applied as it increases But when the torque decreases for an instant the display stops at the highest reading and will no longer adjust Press SEND to print the reading displayed if RS232 capable or record manually using the worksheet provided in section 4 8 or 4 9 Press ON CLEAR to reset First Peak mode and begin a new measurement As in Peak mode ZERO TORQUE will zero out the last display reading Example of display in First Peak tok _49 18 4n 20 3 2 4 Shunt Calibration This mode does not apply to the calibration of a torque tool but to the calibration of the TWCS unit The number displayed in this mode must equal or be relatively close to the number under Shunt Value shown as left on the Calibration Report received from LSI with your unit Call LSI if numbers do not match Example of display when checking the calibration of TWCS in Shunt mode Page 14 of 44 P 060 1000 0045 00E 3 3 TWCS Calibration Test Procedure 3 3 1 Prerequisites A TWCS properly installed for testing torque instruments must be used The torque cartridge capacity must be greater than the highest capacity torque tool that will be tested on it If it is not there is danger of overloading the torque cartridge 3 3 2 Connect the Printer If the TWCS is RS232 capable optional with a 25 pin connector on the back of unit printing may be done on a parallel printer Use an LSI RS232 interface cable to parallel adapt
20. en done at the factory The machine will always know which load cell capacity is in use and will adjust the calibration each time the load cell is switched 4 5 Quick Start to Torque Tool Testing This section of the User Manual is designed to fast track the user to performing torque tool calibration testing resulting in a customized Calibration Report or Certificate To start torque tool testing right away without generating a specific report just turn the machine on and follow the TWCS instructions in section 3 3 If preferred read through the manual and use the Calibration Assistant procedure for torque tool testing starting at section 4 7 4 5 1 Turn on STWCS UTWCS Press ON CLEAR to start The LED display cycles through various screens listing manufacturer serial number load cell data or torque cartridge information current testing mode etc until it stops at a blank test screen However it is not yet ready for torque tool calibration testing Refer to the beginning of section 4 6 for helpful guidance in navigating the Option menus referenced in this section 4 5 2 User Setup Complete the User Setup process described in section 4 6 5 Press OPTION until the User Setup menu shown is displayed Enter printing parameters time and date company information backlight timer preference direction of test and other settings Enter each item as instructed up to section 4 6 6 4 5 3 Cal Check This verifies the shunt calibration of
21. en out of service Contact LSI to schedule recalibration service The tester may be sent back to LSI in its original packaging with a usual turnaround time of less than one week Larson Systems also offers onsite calibration Onsite service can be scheduled for multiple locations in an area for economy when possible Or if preferred calibration may be scheduled with an outside calibration lab 7 3 Packaging for Shipment to LSI 7 3 1 Individual torque cartridge s Fasten securely to the wood frame with the collar placed in the Lock position The collar with lock feature is standard on the lower capacity load cells as these units are extremely sensitive and could be easily damaged Call LSI to obtain a wood frame if needed 7 3 2 Control Panels Circuit Boards While the control panel and circuit board assembly may be removed it is recommended that the entire core unit be shipped If necessary to disconnect the panel unscrew the control panel from the unit and slowly pull forward Behind the panel is the circuit board assembly Carefully unplug the wiring from the circuit board When the circuit board is re installed it is important that the correct plugs from the unit be attached back to their corresponding plugs on the circuit board If reconnected incorrectly the entire circuit board will be shorted out 7 3 3 Packaging Entire Testing Unit Please call LSI to obtain packaging and instructions if the original crate carton was not ret
22. end data while not in Manual Mode Even though first test was done in Manual Mode only allows tests sent in same mode Choose OPTION gt EndPrint then proceed Err86 Must use in Ib units Units were changed during test cycle and SEND pressed Test cycle must be completed using the same unit of measure Er66 Peak Only Optn F2 end form Tried to send data while not in Peak Mode Even though first test was done in Peak Mode only allows tests sent in same mode Choose OPTION gt EndPrint then proceed Er87 Computer asked for stored result data but none has been stored Er67 Cert Info Must Clear User pressed STORE after changing mode Clear stored data to continue Er88 Computer asked DHT to send present contact test when none is available Error 70 key n a User pressed key not available in mode being used Error 89 Must be at least 1 Min A display sleep time entered of less than one minute Err 71 Must be in Manual Mode ZERO ANGLE pressed while in Peak Mode Must be in Manual Mode to zero length Err 92 Call LSI Initialize No Power up check indicates that E2 prom memory has a fault Press F2 and call LSI Error 72 Not available this model User pressed the ZERO ANGLE when model does not include a length scale Problem not listed Call LSI toll free at 1 877 780 2131 or locally at 763 780 2131 The Diagnostics Option menu may be helpful if printing or display problems occu
23. entical control panels and the same underlying software The two models are covered together starting at section 4 with any differences in procedure noted 2 4 2 General Startup Turn on the testing unit by pressing ON CLEAR Commanas referring to the keypad will be shown in brackets throughout the manual It is recommended to begin by getting to know the TWCS starting with its control panel and go to the section that pertains to the correct model Familiarize yourself with how the buttons work Then follow the procedure set forth in the section of the manual that is applicable 2 43 STWCS UTWCS Users Immediate torque tool testing may be done right from the initial screen that displays after the machine is turned on without any further setup using the TWCS Calibration Test Procedure in section 3 3 Torque readings will display and can be stored for manual recording or printing Press STORE to save and SEND to print When performing torque tool tests in this way the display does not guide the user through the test point adjustments and the results are printed as a list of readings that are not ina report or certificate format Use the worksheet found in section 4 8 or 4 9 to record the data in an organized report format Make copies of blank worksheets to keep on hand To perform torque tool testing to result in a concise calibration report summary or certificate follow the Quick start procedure explained below or follow the instruction
24. er to enable this function After connecting the cables turn on the TWCS and printer 3 3 3 Pre Calibration Worksheet Use the Calibration Data Template and or Pre Calibration Worksheet found in section 4 8 and 4 9 respectively to determine classification options for torque tool testing record test point data and provide certification documentation Keep copies of these forms on hand for future use Refer to the chart in section 4 10 for information on Tool Types and Classes to complete the worksheets and select the correct test mode and acceptable tolerance Use the information collected on the forms to complete the torque tool testing process and perform the correct number of test points and record the measurements of each When determining tool type and class look for how the tool adjusts For example does it operate with a graduated measurement scale or is it adjusted using a digital display 3 3 4 Unit Selection Press UNITS until the desired unit of measurement is displayed Choose from in oz in Ib ft lb N m dN m N mm kg m kg cm and kg mm The unit convention must match that of the torque tool 3 3 5 Select Test Mode Press MODE to select Manual Peak or First Peak mode Refer to the Pre Calibration worksheet and Tool Types and Classes chart in this manual for assistance See section 3 2 for information on each test mode and its respective display data 3 3 6 Exercise Torque Tool Before actual testing type 2 too
25. esses a setup feature for users who need to use the external inputs and outputs on the tester with certain auxiliary equipment To select press F1 and the second display example shown above appears Press F1 to advance through the H fields underlined Press F2 to change any H to L These 4 fields represent polarities of 2 3 4 and 5 respectively from left to right and define the active voltage level for those polarities as high or low Press OPTION to advance to the third display example shown above This screen operates the same Press F1 to advance through the H fields underlined and press F2 to change any H to L These eight fields represent polarities of 6 7 8 9 10 11 12 and 13 from left to right This menu feature is not commonly used Users wanting to use this feature should contact LSI for further instruction 4 6 5 8 User Setup Cal Info Cal Info Use this menu to enter calibration information This will store the name of the company that performed the last calibration of the tester the calibrating company s phone number when calibration was done and when the tester is due for its next calibration A new tester will show Larson Systems Inc information The display screens below show the 10 data entry screens available for tester calibration information There are eight fields or characters that may be entered on each screen in the underlined areas as shown Navigate to each chara
26. for the unit a dry non humid non condensing environment kept at room temperature 20 C 5 C 68 F 9 F is suggested This should be a location that also provides protection from debris or falling objects Keep in mind accessibility to a printer and power outlet depending on your model s capabilities When a relatively permanent location has been chosen fasten securely to a flat vibration free surface using the mounting holes on the core unit If vertical mounting is desired the front panel will need to be unscrewed and carefully turned for upright reading of the display There are wires attached to the circuit board on the back of the panel that need to stay in place Screw the panel in place once turned Then mount vertically to a flat vibration free surface making sure the unit is level and perpendicularly positioned While mounting is not required with the Easy Loader assembly installed securing the core unit is recommended to achieve consistent results The core unit must be mounted horizontally in this case When using the TWCS at different locations for operation on its battery if applicable continue to fasten it securely to a stable surface 2 4 Using the Manual 2 4 1 Determine Model Being Used TWCS The software and instructions for this model are very different from the STWCS or UTWCS and are covered separately starting at section 3 STWCS or UTWCS While these models each possess different features they do have id
27. formed in the same mode TWC S Series User Manual 3 2 1 Manual Manual mode is used with needle indicating torque tools When the force applied increases the display dynamically updates to reflect the torque measurement When the force decreases the reading on the display will recede Manually record the reading taken at the appropriate point during the test on the worksheet provided in section 4 8 or 4 9 If RS232 capable with a printer connected SEND will forward the reading to be printed In this way a series of readings may be printed as each measurement is taken Example of display when in Manual mode Peak mode is used with dial indicating or digital torque tools As the force applied to the tool increases the display dynamically adjusts to reflect the torque measurement If the torque subsides the display locks at the highest reading captured If a greater force is subsequently applied the display will continue to adjust to a torque measurement that exceeds the previous reading but will not adjust to a lesser reading Press SEND to print the reading displayed if RS232 capable or record manually using the worksheet provided in section 4 8 or 4 9 Press ON CLEAR to reset Peak mode and begin a new measurement ZERO TORQUE will zero out the last reading Example of display when in Peak mode Cee 3 2 3 First Peak First Peak mode is used with click type torque tools Again the displayed reading dynamically adjusts to r
28. g used the display will stream as shown below From the last screen choose F1 to clear all test points stored in memory or F2 to keep all data in storage When done with data stored clear memory to start a new series of test points P TWCS Series User Manual Page 23 of 44 060 1000 0045 00E 4 6 2 Option Store EndPrint Store From this selection view stored test point data clear a single test point from memory or clear all test points from memory Scroll through test points using F1 or F2 To select a test point as in the first example below press OPTION The display will read like the second example below Press F1 to clear that test point or press OPTION again and the third example below will be displayed Here press F1 to clear all test points from memory or OPTION to return to the test point reading Unless F1 is used to clear any test points they will be saved EndPrint This selection ends the printout in progress and prints out the statistical summary 4 6 3 Option Offset Power Offset A torque offset may be entered using this selection F1 This means when ZERO TORQUE is pressed the torque displayed will adjust to the offset specified as opposed to zero To do this press OPTION until display reads Tor Off Advance the cursor in the field underlined above using F1 or UNITS to backspace Change the values using F2 or ZERO ANGLE When finished press ON CLEAR to accept and exit o
29. gin on the torque tool At the Cal Certificate menu select Start F2 It is recommended that a few preliminary tests be done to warm up the tester and promote consistent results Be sure to clear out any readings captured before starting the calibration procedure Stored memory may need to be cleared to begin testing The LED screen will prompt clearing memory if needed Follow the instructions displayed and make the selection to Clear F1 or Keep F2 data Clear F1 must be selected to start a new set of tests 4 7 10 Select Test Mode Use MODE to select from the three available test modes Choose between Manual Mode Peak Mode and First Peak Mode depending on the type and class of the torque tool Refer to the Pre Calibration Worksheet for this information The test modes and how they work are described in section 4 2 4 7 11 Exercise Torque Tool Before testing type 2 tools must be exercised at full capacity at least five times The torque tool must turn in the direction being tested and torque indication must be triggered i e click type wrenches must click five times 4 7 12 Zero Torque Always press ZERO TORQUE to clear the last torque reading before each new test 4 7 13 Mount Torque Tool Without Easy Loader attachment installed Adjust the torque tool to the test point being measured and place the torque tool onto the STWCS or UTWCS Then go to section 4 7 14 When the required number of measurements is completed
30. handle is identified by a line or marking The length of some torque tools will change after adjusting to the next test point mark and will require adjustment of the slide on the Easy Loader for proper positioning Slide Quick Release Lever Actuator Wheel UTWCS shown with Easy Loader Attachment for illustration purposes 3 3 9 Test Torque Tool Refer to the Force Application diagrams in section 5 Technique for Good Measurement for recommended position and movement to use when testing Record the test point reading press ON CLEAR and then ZERO TORQUE Repeat the steps from 3 3 7 to 3 3 9 for the required number of times determined necessary for each test point Do not record test point measurements that were not performed in accordance with recommended procedures simply redo the test Note As stated earlier if the tool being tested is used both clockwise and counter clockwise it should be tested both ways The calibration can be different in each direction depending on usage Make sure documentation of calibration testing shows CW or CCW for the testing direction TWCS Series User Manual Page 16 of 44 060 1000 0045 00E 3 3 10 Print Test Point Data At any time when a test point measurement is displayed the data can be sent to an attached printer if RS232 capable by pressing SEND The reading displayed will be printed on one line along with the unit measurement A single reading or series of individ
31. he factory and has been found to be the most reliable for achieving accurate results Ifthe tester is capturing the force too late the first peak drop may need to be set at a smaller value Filter Filtering is the first submenu under Filter F2 Using F1 or F2 choose between None Low Medium High User 1 User 2 User 3 and User 4 Once the desired selection is displayed press OPTION Toggle between Yes and No using F1 or F2 to display filter Press OPTION to select The next four submenus are accessed using the OPTION key after each The directions follow on the next page along with filtering information TWCS Series User Manual Page 27 0f 44 060 1000 0045 00E TWCS Series User Manual Page 28 of 44 060 1000 0045 00E Dampening Kd Enter value using F1 and F2 Dampen Min M Enter value using F1 and F2 Decay Rate C Enter value using F1 and F2 Recovery R Enter value using F1 and F2 To review entries made under Filtering press OPTION Press SEND to save and return to the previous menu Press OPTION to continue to next User Setup menu item During normal operation the display shows some number of discrete force values every second These values can be thought of as the value being displayed and the next value to be displayed When the two values are subtracted from each other the result is an error Filtering works by multiplying this error by a percentage and then adding that partial error
32. hen the display shows SMC X it is showing the connected cell s location in memory TWCS Series User Manual Page 30 of 44 060 1000 0045 00E 4 6 6 10 Print Cell Info On STWCS up to 16 load cells can be assigned to a tester in locations numbered from 1 to 16 Data about each load cell can be printed individually F1 or collectively F2 4 6 6 11 Print All E2 Mem Choosing F1 will print out specifications currently saved in memory from User Setup and Calibration Setup menus 4 6 7 Option Calib Menus Calib Menus Set by LSI or authorized calibration house or person Accessed by pass code 4 6 8 Option Factory Setup Factory Setup Used only by LSI 4 7 Calibration Assistant 4 7 1 User Setup Menu If preferences and information have already been entered following the directions in section 4 6 5 continue to the next step in section 4 7 2 Otherwise press OPTION to scroll to the User Setup menu and follow the directions under Option Menus in section 4 6 5 Press F1 to begin Enter each item as instructed up to section 4 6 6 Then return to the test procedure that follows at section 4 7 2 4 7 2 Test Procedure The steps that follow represent the procedure recommended for testing torque wrenches per ISO standard 6789 2003 E Section 4 of this standard highlights accuracy testing Obtaining a copy for reference is suggested 4 7 3 Prerequisites An STWCS or UTWCS for testing torque instruments must be used The tor
33. holes on the outside of the cylinder should match up to the unit and screws used to secure the cartridge Exit the Option mode or turn the STWCS on to begin testing 4 3 3 Storing Store any cartridge not in use in a safe place and left in the locked position if applicable The wood frame received with the extra cartridge s can serve as a holder to store the cartridge s Do not store a cartridge lying on its top or side If shipping a cartridge back to LSI screw it to the wood frame securely and turn to the locked position if applicable before packaging 4 4 UTWCS Load Cell Operation 4 4 1 Turn on the UTWCS When the UTWCS is turned on the LED will cycle through screens regarding the manufacturer serial number etc and finally will show the load cell information with its capacity for the first location only The corresponding button by that load cell will light up This is the cell currently in use Each time the machine is turned on it will default to the cell in the first position 4 4 2 Switch Load Cells To test with one of the other load cells just push the corresponding button for that cell The LED will indicate a switch is taking place and will note when the changeover is complete In addition the button pushed will light up next to the cell chosen This process takes a few seconds Switching between the cells is that simple There is no information to enter into the UTWCS software regarding the load cells That has be
34. hree modes available are Manual for needle indicating instruments Peak for dial indicating or digital and First Peak for click type tools There is also a fourth mode Shunt for checking the calibration of the TWCS ZERO TORQUE Zero out current torque reading before starting a new test ZERO ANGLE Not an active button on this model OFF Turns the TWCS off ON CLEAR Turns the TWCS on and resets Peak and First Peak modes SEND Use when a printout of current reading displayed is needed and TWCS is equipped with RS232 capability If not connected to a printer or not RS232 capable RS232 NOT READY appears on the LED screen When connected for printing and SEND is pressed the printout shows the single reading on the display with a minus sign if performed counterclockwise and the unit measurement used A series of test point measurements can then be printed one at a time on the same page if desired for documentation 3 2 Test Modes Determine the test mode appropriate for the tool being tested Choose between Manual Peak or First Peak modes Select by pressing MODE on the control panel after turning on the machine If unsure which test mode to use refer to the chart in section 4 10 and the worksheets provided in section 4 8 and 4 9 Make extra copies of these worksheets ahead of time and keep on hand for future use Keep in mind all readings for a single tool and series of test point measurements must be per
35. kwise Torque CW Measuring Points Units Test GIVI Or RAJUIN 4 or 6 High Low 4 or 6 OK FAIL This is to certify that the torque assembly tool accompanied by this document has been calibrated to Standards traceable to the National Institute of Standards N I S T Calibration performed in accordance to ISO 6789 2003 E Final Approval 4 9 Torque Tool Pre Calibration Worksheet 5 Tool Classification 1 Units Point Tool 2 Rule C Peak 5 CCW CW AS D Peak 5 Peak Type I 4 Measuring Points Number of Measuring Points 6 Tolerance NON 0 10D U4 gu DDN TATE 1 2 3 4 5 6 7 8 9 1 NN 0 10D Ub UN OMAN U4 BWN KF 7 Calibration Interval months Notes 4 10 Tool Types and Classes Class A Wrench torsion or flexion bar Tolerance 6 Class B Wrench rigid housing with scale or dial or display Tolerance 4 Class C Wrench rigid housing and electronic measurement Tolerance 4 Class D Screwdriver with scale or dial or display Tolerance 6 Class E Screwdriver with electronic measurement Tolerance 4 Class A Wrench adjustable graduated or with Indicating Torque Tools Type 1 display Tolerance 4 Class B Nominal value Wrench fixed adju
36. lar torque tool a tool classification chart is located in section 4 10 This chart is used in conjunction with the worksheets before it Keep in mind that all readings for a single tool and series of test point measurements must be performed in the same mode 4 2 1 Manual Manual mode is used with needle indicating torque tools As the force applied to the tool increases the display dynamically adjusts to reflect the amount of torque measured When the force applied recedes the reading will adjust downward Because of the continual reading adjustment the actual reading must be stored during the application of torque force Press ZERO TORQUE between readings and remember to change the test point mark on the tool being tested when the test point changes After all readings are stored for each test point a report will be generated as output to a printer if specified to do so Otherwise press SEND to forward to the printer When in Manual mode from Cal Certificate menu and Start is selected the LED screen will look like the first example below with the top line flashing between 1st Pt 20 00 and Test 1 05 The T in the bottom left corner means the unit is in Manual mode A reading is now ready to be taken for the first test point at the 20 in lb mark of tool Capacity and will be test number 1 of 5 to be taken The second screen shows a test reading pressing STORE as described above results in the third LED screen below sta
37. le above except the backlight will be dark Enter 99 99 to turn off Backlight Time out feature Units Choose Units F2 to remove any unit measurements that will not be needed from the UNITS selection menu Scroll through unit measurements using OPTION Units listed will show as used if included as a menu selection See first display example above When removed F1 units will be listed as not available as in the next display example If a unit has been removed it may be added back at any time through this User Setup menu by pressing F2 when unit needed is displayed When done press SEND to save and return to the previous menu Press OPTION to continue to next User Setup menu item 4 6 5 5 User Setup 1st Peak Filter 1st Peak Choose F1 to enter First Peak Drop value This value is used in the calibration testing of click type torque tools Enter the desired value in the fields shown underlined above using F1 and F2 Press SEND to save and return to the previous menu Press OPTION to continue to next menu item When testing a click type torque tool the force applied increases over time however just before the click the force applied decreases and a reading is measured This point between the highest force applied and the measurement read is the First Peak Drop The value shown in the second LED example above compensates for the reduced force just before the reading The value of 50 is the default fixed at t
38. ls must be exercised at full capacity at least five times The torque tool must turn in the direction being tested and torque indication must be triggered i e click type wrenches must click five times If the tool being tested is used both clockwise and counterclockwise also test it both ways and be sure to note 3 3 7 Zero Out Torque Press ZERO TORQUE to clear the last torque reading before each new test 3 3 8 Mount Torque Tool Without Easy Loader Attachment installed Adjust the torque tool to the test point mark to be measured and place the torque tool onto the TWCS Its position should be level to the tester and centered Make sure the display reads zero before testing Vianual age 15 of 44 160 1000 0045 00 Easy Loader Attachment installed Adjust the tool to the test point mark to be measured and load the torque tool onto the TWCS with the arm of the wrench extending parallel to the Easy Loader Attachment Loosen the quick release lever and adjust the slide so the wrench handle is supported in the middle Adjust the actuator wheel to give the maximum amount of travel in the direction of the test Insert the stop pins so the wrench is contained on both sides and tighten the quick release lever to hold the slide in place The stop pins should be on either side of the wrench handle such that when the tool reaches its click point the stop pins meet as close as possible in the middle of the handle This center point on the tool
39. n operation as the output device Use RS232 Error 39 Printer Error External device time out Data sent to printer port when time out condition occurred Error 80 Keybrd Not Working A power up check indicates that the keyboard is stuck Call LSI Error 40 Send PC Error External device time out Data sent to PC port when time out condition occurred Error 81 Manual A D Not Working A power up check indicates the Manual Mode A D is not working Call LSI Error 60 Storage Maxed STORE pressed when maximum number of tests allowed has already been stored Error 83 Specs used with stored data Spec changes not allowed when specs are currently being used by stored test data Choose OPTION gt Store Then press F1 to clear data stored to proceed Er62 Manual Only Print or Clear Tried to store test while not in Manual Mode Only allowed to store tests done in the same mode Error 84 Specs used with print out Spec changes not allowed when specs are currently assigned to data being printed Choose OPTION gt EndPrint to stop printing then proceed Er63 Peak Only Send to end set Tried to store test while not in Peak Mode Even though first test done in Peak Mode only stores multiple tests done in one mode Err85 Must use in Ib units Units were changed during test cycle and STORE pressed Test cycle must be completed using same unit of measure Er65 Manual Only Optn F2 end form Tried to s
40. ot functioning properly After exiting press OPTION to go to next diagnostic menu 4 6 6 3 Screen Check Press F1 to check the LED screen The next screen above will flash and the LED cursor will start to cycle through each of the fields When satisfied that each field is operating correctly press ON CLEAR to exit and then OPTION to go to next diagnostic menu 4 6 6 4 Output Diag Test Using F1 scroll through the numbers 1 2 3 or 4 and make selection F2 toggles between on and off Press OPTION to go to next diagnostic menu The standard tester does not include the outputs affected by these selections 4 6 6 5 RS232 Protocols No selection needed Information only listing the fixed protocols set at the factory 4 6 6 6 RS232 Report On No selection needed Information only 4 6 6 7 Check RS232 Port Press F1 to check the RS232 Port connection If printer is not connected a loop back device may be attached to perform the check When the check is performed Greetings User will appear on the display If connected to a printer and the RS232 port checks out a printout is generated Greetings user The RS232 port is working OK Have a nice day 4 6 6 8 e Processor Data No selection needed Information only 4 6 6 9 Display Load Cell Data Using F1 and F2 scroll through the 16 possible locations for any Load Cell assigned to the tester and also show the attached cell W
41. press OPTION to continue nua Page 32 of 44 60 1000 0045 001 4 7 8 4 Tests Per Point Select the number of tests repetitions per measuring point using F1 or F2 Choices range from 1 to 10 and should normally be entered as 5 or 10 depending on the type and class of the torque tool as determined in the tool classification section on the Pre Calibration Worksheet used When finished press OPTION to continue 4 7 8 5 Tolerance Enter the tolerance percentage of the tests to be performed using F1 and F2 This number should be 4 or 6 depending on the type and class of your torque tool A chart of tool types and classes with tolerances is shown under section 4 10 of this manual This is the tolerance for deviation from an exact reading When finished press OPTION to continue 4 7 8 6 Calibration Interval Enter the calibration interval in months using F1 and F2 This is the length of time between calibrations that the torque tool is deemed within its specifications and good The suggested interval is 12 months but may vary with frequency of use and quality of the torque tool To scroll through and review the setup parameters entered press OPTION When finished press SEND to save setup and return to the Cal Certificate menu to start torque tool testing or press ON CLEAR to save and exit out of the menu program 4 7 9 Start Calibration Procedure Once the Cal Certificate Setup menu is completed testing can be
42. que cartridge capacity must be greater than the highest capacity torque tool that will be tested on it If it is not there is danger of overloading the torque cartridge To run the calibration procedure and generate a calibration printout a printer must be connected TWCS Series User Manual Page 31 of 44 060 1000 0045 00E 4 7 4 Connect the Printer Printing can be done on any printer with a parallel port However an LSI RS232 interface cable to Parallel Adapter 025 0000 0458 00 must be used Turn on printer and STWCS UTWCS after connecting the cables 4 7 5 Pre Calibration Worksheet Make copies of the worksheets found in sections 4 8 or 4 9 of this manual to keep on hand for future use Fill out either worksheet to prepare for each series of test point measurements by selecting parameters for the particular tool being tested and to customize the resulting Calibration Report Doing so ahead of time will make entering the data into the machine much faster and easier When determining tool type and class look for how the tool adjusts For example does it operate with a graduated measurement scale or is it adjusted using a digital display 4 7 6 Select Unit Type Press UNITS and choose from in lb ft lb in oz kg cm kg M kg mm g mm N mm or N M The unit convention must match that of the torque tool 4 7 7 Direction for Testing Rule In order to generate positive results for Certificate purposes the direction of the test
43. r SEND to return to the previous menu The angle offset is not active for TWCS models and should be ignored Power This option has only two choices They are Stay On F1 or Off F2 4 6 4 Option Cal check Cal Check Use this function to check the shunt calibration number This can be done as often as daily or before each use It is recommended to follow your company policy for determining frequency of calibration verification When Cal Check is selected using F1 the options displayed are Load Cell or Panel Press F1 to choose Load Cell and two numbers will display following instruction cues The numbers should be negative and match or be relatively close to each other as shown in the third LED example above The top number is the shunt calibration reading established at the factory during calibration and recorded in memory To exit when finished comparing the numbers press ON CLEAR If the numbers do not closely match call LSI TWCS Series User Manual Page 24 of 44 060 1000 0045 00E 4 6 5 Option User Setup User Setup To enter the main User Setup menu press F1 The display will ask for a pass code or press OPTION to continue to the next Option menu To enter pass code press UNITS SEND and F1 in sequence There are several submenus for entering printing parameters time and date company information backlight timer preference direction of testing etc Press OPTION to move to each menu 4 6 5 1 Use
44. r See section 4 6 6 7 CARE AND MAINTENANCE 7 1 General The TWCS is a low maintenance tester with very few moving parts A little oil should be applied from time to time to the screw on the Easy Loader actuator arm This will ensure that the movements are smooth and will prolong the life of the ACME screw Keep oils and other liquids away from and off of the electronics portion of the tester If liquid or oil should contact the device wipe it off with a soft cloth or paper towel Never apply any cleaner or solvent directly to the electronics faceplate a mild plastic cleaner applied to a wipe may be used Take measures to protect the TWCS from debris or falling objects the load cell contains sensitive electronics and can be severely damaged by such forces 7 2 Calibration Verification and Calibration Options Follow your company policy for frequency of calibration verification or use the following guidelines to assure the accuracy of the tester LSI has made the verification of calibration accuracy on its testers easily checked on a daily or more frequent basis if necessary It is recommended that the calibration be verified at least daily when used daily or before each use To complete the verification on an STWCS or UTWCS follow the instructions in section 4 6 4 of this manual under Option Cal Check Directions for verifying the calibration of a TWCS are found in section 3 2 4 If the unit does not pass verification it must be tak
45. r Setup RS232 Power RS232 By selecting RS232 F1 the communication can be set up to send raw data or a report to equipment such as a data logger or printer The following is a list of data format options to be defined within the RS232 menu before data can be sent and printed Next to each item in the first column are the choices available Use F1 or F2 to select or scroll to the choice desired Press OPTION to move to the next item once selection or input has been made Output Format Report Data Print Report Yes No Print Out Format Standard Full Stats Printer Format Print Heading No Heading Printer Format Print Summary No Summary Printer ch width 80 Single 80 Double Printer Protocol Hardware None Delay 01 sec Delay 025 sec Delay 05 sec Printer Baud Rate 9600 19200 1200 2400 4800 Paper Type Plain Paper Roll Feed Lines per Page F1 advance to next digit F2 increment digit Dbl Wdth On Off Oki 1F 1E Epson OE 14 Other TWOS Serie WUS series Define Dbl Wdth On Str Off Str F1 advance to next digit F2 increment digit Define in Decimal Hex F1 advance to next digit F2 increment digit Factory setup for Printer Protocol is Hardware for Printer Baud Rate 9600 When finished press ON CLEAR to accept Iser Manual Power Select Power F2 to set or turn off the timer for auto shut off feature The timer is di
46. rque tool calibration steps with LED screen instructions and print a calibration certificate ready for signature Easy Cal Report e English Metric Units Same as TWCS except STWCS offers g mm not dN m e Printer Compatible Connect to any parallel printer using the RS232 cable and adapter e Rechargeable NiCad Battery Go cordless for up to eight hours of continuous operation OFF SEND Mi STORE 1 2 1 STWCS Torque Specifications See Section 1 3 1 below 1 3 Ultra TWCS UTWCS Features e All Standard STWCS Features except interchangeable cartridges Poo Me a e Three Stationary Load Cells on Single Tester No cartridges to change E ar e Switch Between Load Cells in Seconds Push a button and switch sii between 200 in lb 108 or 260 ft lb capacities instantly ote e Custom Configure Load Cells Special order one two or three load cells 25 100 200 8 500 in Ib 108 amp 260 ft lb torque capacities available as a factory option 1 3 1 STWCS UTWCS Torque Specifications Accuracy is better UEA ee a di Toad Ea A A id 1 4 Optional Equipment e Easy Loader Attachment Easily mount any single arm torque wrench and test it the same way every time regardless of who the operator is factory option e Power Tool Adapter Kit Test powered torque tools such as power screwdrivers with complete confidence and safety includes adapters for 1 4 1 2 and 3 8 drives e Drive Adapters Test a variety of torque
47. rs in a field 4 6 1 Cal Certificate Setup Start Setup Press F1 to begin entering test point calibration specifications Enter each setting desired using F1 and F2 Press OPTION to advance to the next setting Enter new settings each time another type or class of tool is being tested The settings are described in detail starting at section 4 7 8 Setup Calibration Procedure When all calibration settings have been entered press SEND to save setup and continue or press OPTION to scroll through the items again Start Select Start F2 to begin the tool testing procedure Press ZERO TORQUE to clear out any prior reading Then test the torque tool See Section 5 Technique for Good Measurement If the test was not performed correctly press ZERO TORQUE or ON CLEAR to start over When satisfied with the reading press STORE to record Follow the steps outlined in sections 4 7 12 4 7 14 and repeat this testing procedure the number of times specified for each test point If in the middle of the testing procedure a button is pushed to take the machine out of testing mode before all tests are completed press ON CLEAR to return to and continue testing If attempting to scroll back to continue the testing procedure by pressing OPTION the display will read Abort in place of Start at the Cal Certificate menu Press ON CLEAR to continue testing or Abort F2 to quit Note If Start is selected and memory is bein
48. s in section 4 7 Both describe data input procedures and torque tool testing that begins in the User Setup and Cal Certificate menus and allows creation of a customized report or certificate Easy Cal Report The STWCS and UTWCS user will find a Quick Start to Torque Tool Testing in section 4 5 for their unit that may be followed to get to the actual torque tool testing and print a calibration report or certificate without reading the entire manual Although the Option Menus section 4 6 contains valuable information about menu navigation and functions it is primarily designed to be informative rather than directional Either choose to take the time to absorb all features and functions or start immediately with the Quick start instructions that refer to other sections of the manual as you go for direction Because the instructions for using the TWCS are less involved a Quick Start section is not included for that model 3 TWCS USER GUIDE 3 1 The Control Panel Current Software PN 050 1000 0010 00M Version 6 0 6 8 TWCS LARSON SYSTEMS INC Minneapolis MN USA SEND ZERO Not an active button ANGLE on this model ON CLEAR OFF Features Blue Splash Faceplate UNITS Use to choose unit of measurement between in oz in lb ft lb kg mm kg cm kg M dN m N mm and N M Must correspond to the unit convention of the tool being tested MODE Use to select the test mode that applies to the tool being tested The t
49. splayed in HH MM format and starts each time a button is pressed When the machine is idle for the time set it will turn off automatically To set the timer move the cursor F1 to the field to be modified underlined in LED example above and use F2 to change each value Enter 99 99 to turn off the Power Time out feature When finished press SEND to save and return to the previous menu Press OPTION to move to the next User Setup menu 4 6 5 2 User Setup Dampening Date and Time Dampening Press F1 to choose Dampening This feature is used to provide varied levels of noise reduction or dampening that may contribute to false readings when testing Dampening averages the force measurements to eliminate the noise in the display caused by vibration Because the dampening feature slows the measurement process it is recommended the dampening be set at the minimum level necessary to achieve a stable readable display Use F1 or F2 to select between None Low Medium and High Press SEND to save and return to the menu where Date may be selected Date Choose Date F2 from the first menu above and the next menu shown will appear Scroll through the three format choices using F1 or F2 Press OPTION to select when the desired format is displayed Enter the correct date using F1 to advance cursor to the next digit field and F2 to increase digit value When finished press OPTION to advance to the Time setting menu View the t
50. stment Tolerance 4 Class C Wrench adjustable non graduated Tolerance 4 Class D Screwdriver adjustable graduated Tolerance 6 Class E Nominal value Screwdriver fixed adjustment Tolerance 6 Class F Screwdriver adjustable non graduated Tolerance 6 Class G Wrench flexion bar adjustable graduated Tolerance 6 Setting Torque Tools Type Il gt Series User Manual Page 38 of 44 Q060 1000 0045 00I 5 TECHNIQUE FOR GOOD MEASUREMENT Following a few simple guidelines will go a long way to achieve the best possible readings Continuously practicing these guidelines will result in time saved and take the stress out of torque tool calibration Although these practices are mentioned previously in the manual this section is designed to further explain their importance and serve as a reminder 5 1 The Environment Choose a location that is free of vibration and kept at room temperature Humidity can affect readings too a dry non condensing environment is best If intending to generate a report instantly proximity to a power outlet and printer is recommended for convenience 5 2 Proper Installation Securely fastening the unit to a flat vibration free surface is highly recommended using the mounting holes on the base Keep the machine itself as motionless as possible to help produce reliable results without continually retesting to validate If mounting the uni
51. t vertically a perpendicularly level and vibration free surface is best See section 2 3 for mounting information 5 3 Preparation Warm up the machine by doing a few preliminary tests to get started Also exercise the tool being tested if required See section 4 7 11 5 4 Using Worksheets Provided Fill out one or both of the worksheets found in Section 4 8 and 4 9 depending on your needs For TWCS users this will keep testing done in an organized manner with tool information and data all in one place and also provide a report format for certification For STWCS or UTWCS users the worksheet makes it easy to enter the correct parameters for the tool being tested into the Cal Certificate Setup menu prior to testing and assists in creating a customized report 5 5 Torque Tool Calibration To achieve optimum results proper tool placement on the torque tool tester and using consistent motion is very important Pressure that is exerted from motion outside of the Force Application Diagram on the following page will likely contribute to inaccurate readings It may be helpful to practice technique using a torque tool already deemed in compliance with ISO standards The movement must be smooth steady and centered It should not be jerky leveraged or done too quickly 5 6 Easy Loader Attachment Available as a factory option the Easy Loader Attachment is recommended to help achieve the best readings possible for single arm torque tools espe
52. tak ads Deed 31 46 3 Option Factory Setup cite et og sek i Baten tattoo dele dad tos Stink dato eB Tied os gata e ad baton else akin fia 31 Ay Calibration ASsista nt scccceciiscaes sates a ea e a A aa Aaaa aa a E aaa Aa aE seams araa aaa aa aaaea a Tenan dana 31 IESO A 31 A MESTRE UI es E 31 AS O NN 31 4 7 4 Connect the Printer ccecccceeccesecceeeececeeeeeeeeeeseneeseeeesaneeceaeecaeeceaneesaeesneeseaeeesaneesneeseneeseaeessaeeseeesseneeted 32 4 7 5 O ATI A 32 4 1 6 Select Unit Type cstv cited enti velvet iia 32 4 7 7 Direction for Testing RUG eecseeseseceecceseceecceeeeeseneeceesecseeeseensecesaneeaeseaeseanesaeseeseeseeeeaneeanseanteanteees 32 4 7 8 Setup Calibration Procedure ccccccseccceecceeeeceeeeceeeeneeeeneeceneeseneeceaaeeseaeeseaeessaceesaceeseceeseneessaeesseeeeeeeeeteaeess 32 4 7 9 Start Calibration Procedure ii nidii eeaeee eaa ea eaa EnA a eaaa En ea a ioaea taiteet foana ir eioan 33 4 7 10 Select Test Modes hn eine at lia 33 4 11 Exercise Torque Tooli srei eei nate e ae ae bet lie 33 E AA A e Fe NETE E A E EREE Ea De EEEE EO E eee 33 4 7213 Mount Torgue Tool asaini a E ren 34 4 14 Test Torque Tool s a a eei eeta ae za 34 4 7 15 Print Calibration Certificate s ni eei eaae ea aaea aE E a Ea a aoaea iiaa Sa pedana Er enia seen 34 4 8 Calibration Data Template cooncicicniccii aa area aeaa aaraa ran aa ada 36 4 9 Torque Tool Pre Calibration Worksheet cccscccssceeseeeeeeeeeese
53. ting the test number being stored Screen displayed when in Manual mode from initial screen and beginning testing after machine is turned on This screen does not show at what point the series of test point measurements is but the display when storing the reading works the same Readings stored here may be printed using SEND but will not be in a Calibration Report Certificate format Stored readings will not need to be cleared for each tool tested The data logger will accept up to 1 800 test point readings when stored this way 4 2 2 Peak Peak mode is used with dial indicating or digital torque tools As the force applied to the tool increases the display dynamically adjusts to reflect the torque measurement If the force subsides the display locks at the highest torque reading captured If a greater force is subsequently applied the measurement displayed will adjust to a torque measurement that exceeds the previous reading but will not adjust to a lesser reading Press STORE to save reading To reset the Peak mode display press ON CLEAR then ZERO TORQUE to begin a new test point measurement TWCS Series User Manual Page 19 of 44 060 1000 0045 00E After the specified number of test points has been stored a calibration report will automatically print if specified to do so in the User Setup menu Otherwise press SEND to manually print When in Peak mode starting at the Cal Certificate menu and Start is selected the LED
54. tools with 1 4 3 8 1 2 or 3 4 adapters e Inkjet Printer Produce calibration certificates ready for signature on STWCS or UTWCS includes RS232 interface cable and adapter e RS232 to Parallel Printer Cable Adaptor Send data to any 8 5 x 11 or 4 parallel printer and generate a customized report depending on series model e Smart Torque Cartridges Available for STWCS only in 3 25 100 200 amp 500 in lbs 108 amp 260 ft lbs 1 5 Tester Dimensions 4 1 2 Minimum 25 1 2 Maximum 11 1 2 Deep 5 1 2 TWCS 6 STWCS 5 1 8 UTWCS 0 265 Y UTWCS shown with Easy Loader Attachment for illustration purposes Installment of the Easy Loader Attachment will vary for each model TWCS Series User Manual Page 10 of 44 060 1000 0045 00E 2 GETTING STARTED 2 1 Unpacking Your new TWCS model from LSI is shipped completely assembled Once the shipping container is opened this User Manual is easily found with instructions to READ FIRST The TWCS testing unit and Easy Loader if ordered is bolted to the bottom of the shipping container Use care when removing from the packaging and protective poly bag to avoid any damage Other items ordered may be in their own carton s also bolted to the bottom or inside of the shipping container If possible leave any cartons intact inside the container and store for use later on Occasionally an item may be secured to the arm of the Eas
55. ual readings may be printed to kept as a record for that tool 3 3 11 Calibration Documentation Using the Calibration Data Template in section 4 7 pass or fail for the torque tool tested can be documented At the 4 or 6 line calculate what that would be for the test point in each column On the next two lines for each column fill in the high and low measurements taken On the next line calculate 4 or 6 for the test point in each column The high and low measurements should fall between the and numbers If so Pass or OK can be written in the appropriate columns If not write in Fail The Calibration Data Template for the tool should be filled out with a copy of the Calibration Report for the unit used for testing attached or other supporting information and test point measurements taken 4 STWCS UTWCS USER GUIDE 4 1 The Control Panel Current Software PN 050 1000 0005 00J Version 4 22 STWCS UTWCS OPTION ZERO ALS ANGLE J ON MODE ZERO CLEAR TORQUE OFF SEND A STORE LARSON SYSTEMS INC Minneapolis MN USA Features Yellow Splash Faceplate F1 F2 Make selections from LED display enter data or navigate screen fields UNITS Use to choose unit of measurement between in oz in lb ft lb g mm kg mm kg cm kg M N mm and N M Must correspond to the unit convention of the torque tool being tested Unit measurements not being used may be edited out of this menu through the
56. vertical mounting possible without the Easy Loader Assembly installed e Single Torque Capacity Available in 100 200 500 in lbs 108 amp 260 ft lbs with a 1 4 3 8 or 1 2 drive an adapter kit is available e Operates in Both English and Metric Unit Measurements I n oz in lb ft lb kg mm kg cm kg m dN m N mm and N M units e Rugged Cast Aluminum Housing Heavy duty and durable for use in any workspace e 12 Month Warranty Guaranteed satisfaction e Optional Features RS232 printing capability and portable operation with NiCad rechargeable battery are both available as a factory option ZERO py UNITS TORQUE ey MODE ZERO SEND ANGLE 1 1 1 TWCS Torque Specifications Accuracy is better than 0 4 of reading from 10 100 full scale 1 2 Super TWCS STWCS Features e All TWCS Standard Features plus Electronics Upgrade F1 F2 e Accuracy Better Than 0 2 10 100 of full scale Were CD zRo NGLE JA e Separate Function for Shunt Calibration Verification Validate a A a calibration accuracy with confidence in seconds a AS e Built in Data Logger Store up to 1 800 test points internally e Interchangeable Torque Cartridge Capability Order additional Smart Torque Cartridges that can be swapped out easily 3 25 100 200 amp 500 in Ib 108 amp 260 ft lb capacities available STWCS only e User Programmable Force Filter Allows for increased accuracy of readings e Calibration Assistant Follow to
57. wo format choices above using F1 or F2 Press OPTION to select the format displayed Enter the current time in the corresponding screen shown below in the underlined fields Use F1 to advance to next digit field and F2 to increase digit value Press OPTION to review the Date and Time entries or SEND to save and return to the previous menu Press OPTION to continue to next menu item 4 6 5 3 User Setup Company City Company City Choose F1 to enter your Company name F2 to enter your City In the corresponding screens above where underlined enter the first letter or digit in field using F2 or ZERO ANGLE Advance to next character using F1 or go back a field using UNITS to edit There are eight characters on each of four possible screen lines Advance to the next line if needed using OPTION Review the entries using OPTION or when done press SEND to save and return to the previous menu Press OPTION to continue to next item TWCS Series User Manual Page 26 of 44 060 1000 0045 00E 4 6 5 4 User Setup Backlight Units Backlight Choose Backlight F1 to enter Backlight Time out setting in HH MM format Advance between the digit fields underlined above using F1 and UNITS Adjust digit values using F2 and ZERO ANGLE When finished press SEND to save and return to the previous menu If tester is not used for the amount of time entered it will go to sleep The display will read like the third LED samp
58. y Loader if ordered 2 2 Items Shipped Locate the following items inside the shipping container e TWCS testing unit e Power adapter Battery charger e Three 8 Y X 11 white envelopes 1 User Manual 2 Packing List 3 Signed Calibration Report e Optional equipment ordered If the Easy Loader Attachment was ordered it will be assembled and already connected to the testing unit the pins that align the torque tool handle on the Easy Loader and the small actuator wheel are in a small box with the power adapter e Interchangeable torque cartridges STWCS only Extra cartridges if ordered will be inside the shipping container in separate packaging and fastened to the bottom or inside of the container Within its packaging will be the cartridge s attached by screws to a wood frame The wood frame makes it easy to safely package and ship back to LSI for recalibration and also provides a convenient holder for the cartridge when not in use e Printer or larger equipment shipped separately Check over the packing list and make sure all items listed were received Keep items not needed right away in a safe place The TWCS unit may now be plugged into an 110V power outlet using the adapter Allow the battery to charge for eight hours before operating on its own Note The STWCS and UTWCS both have the battery installed as a standard feature for portable use an option on the TWCS 2 3 Installation In choosing a location