1. APPENDIX A TROUBLESHOOTING Problem Rotor Fault Solution The EFM 100 field mill continuously monitors the speed of the rotor and ensures it is spinning at the correct speed If the rotor is found to be spinning either too fast or too slow the supervisory circuitry will change the fault status in the data stream from a 0 no fault to a 1 rotor fault The EFM 100 display software will indicate a Rotor Fault The field mill rotor may have become obstructed If the rotor becomes coated with ice during a freezing rain ice storm you can wait for the ice to melt naturally it wouldn t be safe to try and clear the ice manually when the ground and roof are ice covered The field mill will not be damaged by a jammed rotor and will begin to work normally once the ice melts away Problem Data often appears clipped at 20 kV m Solution If your data is routinely clipped at 20 kV m or 20 kV m your field mill is probably too sensitive Reduce the field mill sensitivity by changing the Sensitivity Plug to a lower value See Chapter 2 32 APPENDIX B EFM 100 DATA FORMAT EFM 100 Data Format The EFM 100 Electric Field Mill transmits data sentences down the fiber optic cable at the rate of ten readings per second The EFA 10 converts the optical signal to standard RS232 at 9600 baud 8 data bits no parity 1 stop bit Electric Field Sentence lt p gt lt cee ee gt lt f gt lt cs gt lt cr gt lt lf gt lt p gt polarity of el
2. INTRODUCTION ies ese ee es ee eo ee oe ee ee oe ee ee ee oe ee ee ee ee ee eg 1 THEORY OF OPERATION ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 2 INSTAL ATION SS SS Se Ge Se ed de Ged ee Ge de Ge Ge ee ee de Ge ee ee ede ee eed de ed Gee vee de eg 5 EFM 100 DISPLAY SOFTWARE INSTALLATION ee ee ee ee ee ee ee ee ee ee ee ee ee ee 6 HARDWARE INSTALLATION ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 6 CHOOSING AN OUTDOOR MOUNTING LOCATION ee ee ee ee ee ee ee ee 7 EPM 1 POWER MODULE ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee dee 9 ANALOG OUTPUT EER EE GE SSA EE NOME EEE EE Ee ie 10 EFA 10 FIBER OPTIC ADAPTER ee ee ee ee ce ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 11 ST FIBER OPTIC CONNECTORS ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 11 CALCULATING A PRECISE SENSITIVITY VALUE ee ee ee ee ee ee ee ee ee ee 14 GROUNDING DETAIL ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee nnne 15 CONFIGURING THE SOFT WAREL cccsssccscccssssccccscscccccssscccccssscccccssscccceescccceees 17 EFM OPTIONS GENERAL o oo occ ee ee cece cece ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 17 EFM OPTIONS SOUND ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 18 EFM OPTIONS ALARMS ee ee ee
3. 3 28 feet The electric fields which accompany thunderstorms normally measure in the thousands of Volts per meter usually abbreviated to kV m Lightning can be detected as a sudden change in the electric field The electric charge contained in a thundercloud also generates an electric field This field can be measured on the ground CHAPTER 1 INTRODUCTION For an accurate electric field reading the field mill needs to be mounted flush with the ground Mounting the field mill flush with the ground is not practical however as water dirt insects etc will collect around the sense electrodes and contaminate the electrode insulators A AA AA A A A A A A AOA ee C A Figure 1 Electric Field Mill at Ground Level Mounting the electric field mill above the ground surface will enhance the electric field resulting than an incorrect high reading Sensitivity Plugs are provided to reduce the sensitivity of your field mill and compensate for the higher field mill readings 4 4 4 4 4 WY Figure 2 Intensification of Electric Field CHAPTER 1 INTRODUCTION The EFM 100 s electric field mill senses electric field by repeated exposing and shielding a series of sense electrodes An electric field mill uses a mechanical chopper to alternately shield and expose several sense plates to an electric field When the sense plates are exposed to the electric field an electric char
4. Grounding Detail The EFM 100 field mill needs to be grounded for proper operation and for lightning and electrical safety Connect the connector end of the green ground wite to the field mill ground connector Connect the other end of the green ground wire and the field mill mounting hardware to your building s central ground through a 6 bare ground wire or other suitable wire according to your local electrical code Grounding wire and clamps are available at your local hardware store or electric supplier IMPORTANT The EFM 100 field mill needs to be grounded for proper operation and for lightning and electrical safety Connect the field mill green ground wire and the field mill mounting hardware to your building central ground through a suitable sized grounding wire according to your local electrical code 15 CHAPTER 2 INSTALLATION EFM 100 Field Mill Connect to EFM 100 Ground Connector See Figure 13 Green Field Mill Ground Wire Provided Copper Ground Wire Not Provided Ground connection Not Provided Figure 16 Field Mill Grounding Detail 16 CHAPTER 3 CONFIGURING THE SOFTWARE Configuring the Software EFM Options General EFM 100 Options Field Mill Selection Number of Data Sources E Max 4 Shop Roof z 6 Field Mill Details d oun Data Source Name Shop Roof V Enabled Data Source Point Serial US Com Port COM1 Y Network localhost Port ET D
5. ee Re Re ee 8 Figure 7 EPM 1 Power Module Connectof esse sees sees ee ee ee ee ee ee ee eke ee ee ee ee ee 9 Figure 8 EPM 1 Voltmeter Connections Differential Preferred Connection 10 Figure 9 EPM 1 Voltmeter Connections Single Ended sesse esse ee ee ee ee see ee tense ee ee ee 10 Figure 10 EFA 10 Fiber Optic Adapter iese see esse ee see ee ee ee ee ee ee ee ee ee ee ee ee ee ee 11 Figure 11 Female Fiber Optic Connector iee see ee ee Re ee ee ee ee ee ee ee ee ee ee 11 Figure 12 Male Fiber Optic Connectof esse ee ee in EEEE E 12 Figure 13 EFM 100 Field Mill ConnectHons esse esse ese ee ese see ee ee ee ee ee ee ee ee ee ee ee ee ee 12 Figure 147 Sensitivity PITS su ed de ee EE dalte oe ee ae ee ee es 13 Figure 15 Correcting a Field Mill using a Reference Field Mill ee 14 Figure 16 Field Mill Grounding Detail ee ee Re ee ee ee ee ee ee ee ee ee ee 16 Figure 17 EFM Options General ees ees ee se ee see ee ee ee ee ee ee ee ee ee ee ee ee ee ee 17 Figure 18 EFM Options Sound sees ee see ese ee see ee eens ee ee ee ee ee ee ee ee ee ee ee ee ee 18 Figure 19 EFM Options Alarms esse ese ee se ee ee ee ee ee ee O ee ee ee ee N E RTAS 19 Figure 20 EFM Options Data Logging esse esse ees ee ees see ee see ee ee ee ee ee ee ee ee ee ee ee ee ee 21 Figure 21 EFM Options Data Sharing i e esse esse ee ese ee E ee ee ee See A E ee
6. ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 19 EFM OPTIONS DATA LOGGING ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 21 EFM OPTIONS DATA SHARING ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 22 91 Did d U Le R ose seso O eE E E EEEE EEEE 23 INTERPRETING THE DATA DISPLAY ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 25 MAINTENANCE ees es ss ee es ve ee ee ee se ee ie eise ede ede ie ee dei dee ed gie ee eed si ee 29 TROUBLESHOOTING se dd Se de Sa De see ee Ge ee Gee Gee Gee Gee Gee dee de ee dee dese deed eg ds 31 EFM 100 DATA FORMAT sesse sesse dees ee ee sie esse deeg ee ede ie sed eo ees Ge gebede ee 33 ELECTRIC FIELD SENTENCE terdee es EGO R SE Re OES DR EE OES D SEE DEERE EE E Re GE EDE ER ALTERA ED Ee Eie De 33 SPECTFICATIONG ssccssssssssssesssseesccsscsscesesseseseesseseesscsessessessesesseesessessesseseeseeseones 35 Table of Figures Figure 1 Electric Field Mill at Ground Level ee ee ee ee ee ee eke ee ee ee ee 3 Figure 2 Intensification of Electric Field ese ee ee ee ee ee ee ee eke ee ee ee ee 3 Figure 3 Field Mill Block Diagram esse ee ee ee ee Re ee ee ee ee ee ee ee ee ee ee ee 4 Figure 4 EFM 100 Connection Diagram esse sees ees ees cece ee ee ee ee ee ee ee ee ee ee ee ee ee ee 5 Figure 5 A poor choice of mounting locatOnS ee ee Re ee ee ee ee 8 Figure 6 An excellent choice of mounting location
7. scale of the vertical amplitude axis 24 CHAPTER 4 OPERATION Interpreting the Data Display Often the first indication of an approaching thundercloud is a positive field reading followed by a field reversal to a negative field as the cloud moves overhead AR Figure 23 Approaching Thundercloud fair weather electric field Figure 24 Increasing Field due to Approaching Thundercloud CHAPTER 4 OPERATION Cloud Direction gt Figure 25 Thundercloud Directly Overhead Figure 26 Electric Field Polarity Reversal With the thundercloud directly overhead the polarity of the field has reversed to a negative electric field CHAPTER 4 OPERATION Figure 27 Departing Thundercloud cloud overhead Figure 28 End of Storm Oscillation Once the cloud has passed over the field will often reverse back to positive before decaying to a normal fair weather electric field reading of about 0 1 kV m This field reversal at the end of the storm has been referred to as the End of Storm Oscillation ESO 27 CHAPTER 4 OPERATION Nearby lightning t Distant lightning Figure 29 Step Changes in Field Magnitude Indicate Lightning Step changes in the magnitude of the electric field indicate lightning Closer lightning produces larger field changes than distant lightning The EFM 100 can detect lightning out to about 30 mile
8. C BOLTEK BOLTEK CORPORATION Lightning Detection EFM 100 Atmospheric Electric Field Monitor Installation Operators Guide EFM100 1000120 050205 BOLTEK LIGHTNING DETECTION EFM 100 Atmospheric Electric Field Monitor Disclaimer EFM 100 lightning data is only approximate and should not be used for safety applications Strike and storm locations indicated and alarm statuses may be erroneous and should not be used to safeguard personnel equipment or data Neither Boltek Corporation nor its affiliates shall be liable to the purchaser of this product or third parties for damages losses costs or expenses incurred by purchaser or third parties as a result of use misuse accident or abuse Notwithstanding the above Boltek Corp s liability shall not exceed the purchase price of the equipment THIS EQUIPMENT IS NOT TO BE USED FOR SAFETY PURPOSES USE OF THIS PRODUCT SHALL BE DEEMED TO BE ACCEPTANCE OF BOLTER S STANDARD TERMS AND CONDITIONS ENCLOSED AND AVAILABLE AT WWW BOLTEK COM IF YOU DO NOT AGREE TO BE BOUND BY THESE TERMS AND CONDITIONS PLEASE RETURN THIS PRODUCT TO THE PLACE OF PURCHASE FOR A FULL REFUND 2004 2005 Boltek Corporation Boltek Corporation 2316 Delaware Ave PMB 254 Buffalo NY 14216 USA 210 Glendale Ave Suite 168 St Catharines ON L2T3Y6 Canada Email info boltek com Web www boltek com Phone 905 734 8045 Fax 905 734 9049 FCC Compliance Statement For United States Users Thi
9. G THE SOFTWARE High Field Alarm Delay The High Field Alarm Delay is the length of time the field magnitude must remain above the setpoint before the alarm is activated If the electric field magnitude drops below the setpoint before the delay expires the delay timer will be reset For example with a delay of 30 seconds and setpoint of 1 kV m the electric field must remain above 1 kV m or below 1 kV m for 30 seconds before the alarm activates High Field Alarm Duration High Field Alarm Duration is the length of time the High Field Alarm will remain active once the field drops below the alarm setpoint EFM Options Data Logging EFM 100 Options es Data Logging General dy IV Enable Data Logging Sound Log Location C Program Files Boltek EFM D ata B Field Mil Shop Roof Alarms Data Sharing Figure 20 EFM Options Data Logging Field Mill Selector Select the field mill for which you would like to configure data logging Enable Data Logging Checkbox Check the Enable Data Logging checkbox if you would like field mill data to be logged to the hard drive for the field mill selected in the selection box above Logged data is viewable using the EFMView Electric Field Mill Data Viewer Select File Open on the main menu to view the data file Logging data will consume approximately 1 3 megabytes of hard disk space per field mill per day 21 CHAPTER 3 CONFIGURING THE SOFTWARE Data files have a fil
10. S 232 Connector i f to RS 232 Cable i Field Mill or USB Cable Fiber Optic Connector Figure 10 EFA 10 Fiber Optic Adapter The EFA 10 converts this optical data stream to an electrical signal compatible with your computer s RS232 COM port A USB adapter cable is provided to convert this RS232 signal to USB if a USB connection is preferred ST Fiber Optic Connectors The EFM 100 transmits its data over a fiber optic cable The fiber optic cable uses St style connectors on each end Keyway Locking Tabs Figure 11 Female Fiber Optic Connector If the field mill installer is not familiar with ST fiber optic connectors he should familiarize himself with the connectors on the ground before mounting the field mill on the roof 11 CHAPTER 2 INSTALLATION Align the key in the male connector with the keyway in the female connector before inserting the male connector into the female connector Insert the male connector into the female connector and rotate the locking collar until you feel the tabs stop in the detents Align with keyway and locking tabs before inserting Fiber Core Figure 12 Male Fiber Optic Connector Sensitivity Connector Ground Connection Power Connector Fiber Optic Data Connector Figure 13 EFM 100 Field Mill Connections 12 CHAPTER 2 INSTALLATION Power Connector Connect the black cable from the field mill s power connector to the EPM 1 Power Module s field mill power connector The cabl
11. V Enable High Field Alarm Setpoint 5 kV m General dy Sound Delay 5 seconds Duration 1 minutes High Field Alarm Setpoint 1 kVm Data Logging Delay 5 seconds E Duration 1 minutes Data Sharing onea Figure 19 EFM Options Alarms Field Mill Selector Select the field mill for which you would like to enable or disable alarms 19 CHAPTER 3 CONFIGURING THE SOFTWARE Enable Lightning Alarm Checkbox Check this checkbox to enable the Lightning Alarm for the currently selected field mill Enable Very High Field Alarm Checkbox Check this checkbox to enable the Very High Field Alarm for the currently selected field mill Enable High Field Alarm Checkbox Check this checkbox to enable the High Field Alarm for the currently selected field mill Lightning Alarm Sensitivity Lightning is detected by the software as a step change in electric field value Electric field normally changes slowly as clouds slowly pass overhead Electric charge is redistributed during a lightning strike causing an almost instantaneous change in electric field Lightning Alarm Sensitivity is the size of the step change that is required to activate the lightning alarm Lower values make the lightning alarm more sensitive and extend the range Higher values reduce the sensitivity and reduce the detection range The minimum recommended value for the Lightning Alarm Sensitivity is 0 1 kV m Increasing the value to 0 2 or 0 5 is recommen
12. are in order to receive the remote data When Data Sharing is enabled the display software will accept remote connections and deliver its field mill data over those connections for others to view To start the EFM 100 display software select Start All Programs Boltek EFM 100 Electric Field Monitor Zoom Handles Alarm Annunciator Location Descriptor Boltek Electric Field Monitor DER File Tools Help Time Range Selector Amplitude Range Selector Zoom Handles 15 40 19 41 Current Field Reading Figure 22 EFM 100 Software Display 23 CHAPTER 4 OPERATION Location Descriptor describes the location of the field mill When the software is configured to display data from multiple field mills the descriptor is important in identifying the source of the data The History Graph shows the most recent data in graphical form The history graph is useful in identifying trends in electric field Zoom Handles are used for zooming in on an area of interest in the trend graph There are zoom handles for magnifying the display horizontally the time axis or vertically the amplitude axis Click and drag a zoom handle to adjust The Current Field Reading is the amplitude of the electric field at the present time The current field reading is updated approximately once per second The Time Range Selector is used to change the scale of the horizontal scale axis The Amplitude Range Selector is used to change the
13. ata Up to four field mills can be monitored using the included software Field mills can be directly connected to the display computer USB or COM port or connected to a remote computer with data transmitted over a network Your EFM 100 package should contain 1 EFM 100 field mill 1 fiber optic cable Standard length 100 feet 30 meters Optionally 150 feet 45 meters or 200 feet 60 meters 1 UTP power cable Standard length 100 feet 30 meters Optionally 150 feet 45 meters or 200 feet 60 meters 1 EFA 10 fiber optic adapter 1 EPM 1 power module 1 AC wall adapter power supply 120VAC to 12VDC for North America 220VAC to 12VDC for Europe 1 DB 9 male to DB 9 female RS232 cable 1 CHAPTER 1 INTRODUCTION 1 field mill grounding wire 5 sensitivity plugs 0 75X 0 5X 0 33X 0 25X 0 2X 1 mast 7 4 NPT X 6 1 mounting flange 1 CDROM containing Windows software and USB drivers 1 user manual this is it Unpack your EFM 100 and make sure all the parts are included Theory of Operation Electric fields develop wherever there is a difference in electric potential If the electric field gets high enough you can feel your hair stand on end if this happens outdoors during a thunderstorm crouch down with your feet together as you are about to be struck by lightning An electric field is what attracts your hair to a charged comb or a charged balloon Electric field is measured in Volts per meter 1 meter
14. ata Logging Data Sharing oe Figure 17 EFM Options General Number of Data Sources Enter the number of field mills for which you would like to display data The minimum number is 1 field mill The maximum number is four field mills If you have more than four field mills you will require a second computer for each additional four field mills creating a multi monitor display Data Source Name Enter a descriptive name for each field mill If you are displaying the data from more than one field mill the name will be used to identify the source of the data Select individual field mills in the Field Mill Selection box above 17 CHAPTER 3 CONFIGURING THE SOFTWARE Enabled Checkbox Check this checkbox to enable the display of data for the currently selected field mill Data Source Point Select the type of connection to the field mill Serial USB or Network Select Serial USB for field mills connected to the computer directly Select Network to display the data from a remote computer Com Port For field mills connected directly to the computer select the COM port allocated to the connection For field mills connected using a RS232 COM port this is usually COM1 or COM2 For USB connected field mills this is usually COM3 COM9 Network For remote field mills enter the host and domain name or the IP address of the computer which is sharing a field mill Port For remote field mills enter the port number of the
15. cm Warranty 1 Year EFM 100 Software Specifications Operating System Microsoft Windows XP 2000 NT 98 95 Alarms High Field Alarm Very High Field Alarm Lightning Alarm High Field Alarm Range 0 to 20 kV m Very High Field Alarm Range 0 to 20 kV m History Graph 5 min 15 min 30 min 1 hour 2 hour 4 hour 8 hour 24 hour Handles on the graph for zooming in Vertical Scales 0 5 kV m 1 kV m 2 kV m 5kV m 7 5 kV m 10 kV m 15 kV m 20kV m Handles on the graph for zooming in 36
16. connected to unused interfaces Changes or modifications not expressly approved by the manufacturer could void the user s authority to operate the equipment For Canadian Users This Class B digital apparatus meets all requirements of the Canadian Interference Causing Equipment Regulations Cet appareil num rique de la class B respecte toutes les exigences du R glement sur le materiel brouilleur du Canada CAUTIONS EFM 100 lightning data is only approximate and should not be used for safety applications Strike and storm locations indicated and alarm statuses may be erroneous and should not be used to safeguard personnel equipment or data Install the EFM 100 Field Mill on a calm clear day when no thunderstorms are expected If you are not experienced in safe antenna installation using appropriate safety equipment you should refer installation to an experienced antenna installer See Television Antennas in the Yellow Pages LIGHTNING HAZARD EPM 1 connections are not optically isolated from the roof mounted field mill Dangerous voltages may occur should lightning strike at or near the field mill IMPORTANT The EFM 100 field mill needs to be grounded for proper operation and for lightning and electrical safety Connect the field mill green ground wire and the field mill mounting hardware to your building central ground through a suitable sized grounding wire according to your local electrical code Table of Contents
17. dO 1aa14 Computer Not Provided _ EFA 10 Ground Fiber Optic Adapter connection Provided Not Provided Provided Figure 4 EFM 100 Connection Diagram CHAPTER 2 INSTALLATION EFM 100 Display Software Installation Insert the provided EFM 100 Software CD into your CD ROM drive The setup program should start automatically If the setup program does not start automatically you can start it manually by clicking on Start Run type d setup exe then click OK Once the software is installed you can run the EFM 100 display software by clicking on Start All Programs Boltek EFM 100 Electric Field Monitor Instructions for installing USB Serial Adapter Driver Software 1 Plug the USB adapter cable into the computer s USB port 2 The Welcome to the Found New Hardware Wizard should appear If it doesn t go to step 8 3 Ensure that the provided USB driver CD has been inserted into your CD ROM drive 4 Select the option Install the software automatically Recommended 5 Press Next 6 Windows should find the Prolific USB to Serial Com Port driver 7 Press Finish when prompted by Windows to complete the installation If the Found New Hardware Wizard did not appear in step 2 Open Device Manager by doing the following a Click on Start and then Control Panel b Switch to Classic View if in Category View c Double click on System d Select the Hardwar
18. ded if precipitation is causing nuisance alarms Lightning Alarm Duration Lightning Alarm Duration is the length of time the lightning alarm will remain active after a lightning event is detected For the Lightning Alarm to clear there must be no lightning events detected for the period of the lightning alarm duration Very High Field Alarm Setpoint Very High Field Alarm Setpoint is the magnitude of electric field required to trigger a Very High Field Alarm For example with a setpoint of 5 kV m the alarm will activate if the field goes above 5 kV m or below 5 kV m Very High Field Alarm Delay The Very High Field Alarm Delay is the length of time the field magnitude must remain above the setpoint before the alarm is activated If the electric field magnitude drops below the setpoint before the delay expires the delay timer will be reset For example with a delay of 30 seconds and setpoint of 5 kV m the electric field must remain above 5 kV m or below 5 kV m for 30 seconds before the alarm activates Very High Field Alarm Duration Very High Field Alarm Duration is the length of time the Very High Field Alarm will remain active once the field drops below the alarm setpoint High Field Alarm Setpoint High Field Alarm Setpoint is the magnitude of electric field required to trigger a High Field Alarm For example with a setpoint of 1 kV m the alarm will activate if the field goes above 1 kV m or below 1 kV m 20 CHAPTER 3 CONFIGURIN
19. e Chapter 1 Intensification of Electric Field for a description of why this is necessary Figure 14 Sensitivity Plugs With no plug installed the field mill has a relative sensitivity of 1 0 Sensitivity plugs reduce the sensitivity to compensate for the enhanced electric field of a field mill mounted above the surface of the ground Lower numbers represent lower sensitivities For example the 0 25X plug will reduce the sensitivity to Va of the field mills normal sensitivity Fortunately most installations won t require a precisely calculated correction factor Ona clear day simply choose the sensitivity jumper which produces field values closest to the normal fait weather electric field of 0 1 kV m If electric field readings during 13 CHAPTER 2 INSTALLATION thunderstorms routinely exceed 20 kV m the limit of the field mill you should change to a lower value Sensitivity Plug after the storm and when it is safe to do so Calculating a Precise Sensitivity Value If you require a precisely calibrated field mill for scientific or other precision applications you will need to calculate a precise correction factor then order or assemble a custom Sensitivity Plug The most accurate method for calculating an exact geometric correction factor is to temporarily run two field mills stmultaneously a temporary reference field mill and the permanent field mill The temporary reference field mill is mounted flush with the surface of t
20. e supplies 12VDC to the field mill and carries the analog output signals back to the EPM 1 See the Lightning Hazard caution in the EPM 1 Analog Output section of this manual Fiber Optic Connector Connect the orange fiber optic cable from the field mill s fiber optic connector to the EFA 10 s fiber optic connector The fiber optic cable carries electric field data from the field mill to the EFA 10 and ultimately to your computer See the detailed drawings of the fiber optic connector and note the keyway and locking tabs on the connectors Orient the key and locking slots properly on the male connector before attempting to insert the male connector into the female connector Ground Connector Connect the connector end of the green ground wire to the field mill ground connector The other end of the green ground wire connects to the field mill mount s ground connection When mounting the field mill it is important to properly ground the field mill and the field mill mount The field mill is grounded through the green ground wire The field mill mounting hardware mast and or tripod needs to be grounded though suitable sized copper ground wire to your building central ground point See Grounding Detail for more information Sensitivity Connector The sensitivity connector accepts a sensitivity plug to reduce the sensitivity of the field mill Several different values are provided so you can choose an appropriate value for your installation Se
21. e tab e Press the Device Manager button Under Port COM amp LPT look for a USB Device entry with a yellow exclamation point by it Right click on the USB Device and select Update Driver This should bring up the Welcome to the Hardware Update Wizard Ensure the product driver CD provided is inserted in your CD ROM drive Select the option Install the software automatically Recommended Press Next Now begin at step 5 above and continue Hardware installation The EFM 100 field mill mounts on a 3 4 NPT National Pipe Thread threaded pipe Note that 2 NPT pipe measures 1 outside diameter A short length of 4 NPT threaded pipe and a 4 NPT mounting flange are provided for mounting the field mill on a horizontal surface Longer lengths of threaded pipe are available at your local hardware store Antenna mounting tripods are available at your local 6 CHAPTER 2 INSTALLATION Radio Shack store or your local television antenna installer see Television Antennas in your Yellow Pages Install the EFM 100 Field Mill on a calm clear day when no thunderstorms are expected Your local antenna installer can do a safe and professional field mill installation and can also ensure that the field mill is properly grounded If you are not experienced in safe antenna installation using appropriate safety equipment you should refer installation to an experienced anten
22. ectric field or lt ee ee gt electric field level 00 00 to 20 00 lt f gt fault 0 Normal 1 Rotor Fault lt cs gt checksum in hex 00 to FF lt cr gt carriage return lt lf gt line feed Example 00 33 0 C9 Represents 0 33kV m with no faults Checksum can be ignored for most applications 33 APPENDIX C SPECIFICATIONS Specifications EFM 100 Hardware Specifications Electric Field Range 20 kV m to 20 kV m Typical Range of Interest 5 kV m to 5 kV m Response Time Resolution Digital Data Cable Digital Data Format RS232 Analog Data Output DC Power Supply AC Power Supply Motor Type Mount Temperature Range Dimensions Weight Shipping Weight 0 1 seconds Digital Output 0 01 kV m Analog Output Infinite 50 125 Multi Mode Fiber Optic Cable with ST Connectors lt S gt lt EE EE gt lt C gt lt CR gt lt LF gt S sign or C checksum E electric field CR carriage return LF line feed Example 02 34 3A 9600 baud 8 bits 1 stop bit no parity Differential 1V per kV m Single Ended 0 5V per kV m 12 15 VDC 0 5 Amps 120 VAC 60Hz 220VAC 50 HZ in Europe Ball Bearing Brushless DC Motor 3 4 NPT Threaded Pipe Note 3 4 NPT measures 1 O D 40 F to 140 F 40 C to 60 C 6 7 dia x 5 17 cm dia x 13 cm 2 2 Ibs 1 kg 13 Ibs 5 75 kg 35 APPENDIX C SPECIFICATIONS Shipping Dimensions 18 5 x 13 x 8 8 47 x 33 x 22
23. ee ek ee ee ee 22 Figure 22 EFM 100 Software Display iese esse se ee ese ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee ee 23 Figure 23 Approaching Thundercloud ie ee ee ee ee 25 Figure 24 Increasing Field due to Approaching Thundercloud ee ee 25 Figure 25 Thundercloud Directly Overhead ee ee ee ee ee ee ee 26 Figure 26 Electric Field Polarity Reversal iese esse ee see se ee see see ee ee ee ee ee ee ee 26 Figure 27 Departing Thundercloud ee ee ee ee ee ee 27 Figure 28 End of Storm Oscillaton ee ee ee ee ee ee ee ee R ee ee ee 27 Figure 29 Step Changes in Field Magnitude Indicate Lightning eie ee 28 Figure 30 Precipitation NOISE nasran reiro ee E eE E EE E E EEE TE ee ee ee ee 28 Figure 31 Field Mill Cleaning ese ee ee e E E ee ee ee E e 29 CHAPTER 1 INTRODUCTION Introduction Congratulations on your purchase of a Boltek EFM 100 Electric Atmospheric Electric Field Monitor The EFM 100 is a low cost high quality atmospheric electric field monitor which uses your personal computer to display and record data The EFM 100 not only detects nearby lightning but can also detect the high electric field conditions which preceded lightning EFM 100 lightning data is only approximate and should not be used for safety applications Strike and storm locations indicated and alarm statuses may be erroneous and should not be used to safeguard personnel equipment or d
24. ename consisting of the field mill name the file date and a efm suffix For example Shop Roof 03252004 efm Data files are comma delimited ASCII text files allowing the files to be loaded into spreadsheets or other third party software Logfile Location This is the directory where the data files are stored Click on the Open icon B to browse for the directory EFM Options Data Sharing EFM 100 Options Data Sharing Field Mil Shop Roof Enable Sharing Data Logging vi Figure 21 EFM Options Data Sharing Field Mill Selector Select the field mill for which you would like to configure data sharing Enable Sharing Checkbox Check the Enable Sharing checkbox if you wish to share your field mill data with others If enabled your EFM 100 software will list on the Listen Port for incoming connections Once a connection has been made the software will begin transmitting field mill data to the remote computer There is no limit to the number of concurrent connections Listen Port This is the port on which your computer will listen for incoming connections Each field mill being shared on one computer must have a unique port number 22 CHAPTER 4 OPERATION Operation The EFM 100 display software interprets and displays the field mill readings and status For remote field mills the software also attempts to establish and maintain a network connection with the remote computer s EFM 100 display softw
25. field mill from the computer which is sharing the field mill See EFM Options Data Sharing for more information EFM Options Sound EFM 100 Options Lighbring Alarm ss V WAV Fie l LigrtrmaDetected wav ri B General d WAV Repeat Intervat 60 seconds SS Very High Field Alam 7 WAY Fie WeiHighElectncField wav E Ld WAV Repeat Irtervat 60 seconds Data Logging r High Field Aleem FE IV WAY File HighElectncField wav B P Data Sharing WAV Repeat Irtervat 60 seconds ore Figure 18 EFM Options Sound 18 CHAPTER 3 CONFIGURING THE SOFTWARE Configure the sound to play out the computer speakers whenever an alarm activates You may select a different WAV file for each of the three alarms Lightning Alarm Very High Field Alarm and High Field Alarm WAV File Checkbox Check the checkbox to enable the playing of WAV file sound when the corresponding alarm activates Clear the checkbox to disable the playing of the WAV sound WAV Filename The WAV Filename is the file which will play out the computer speakers when an alarm is activated Click on the Browse icon B to select the WAV file from a list WAV Repeat Interval Enter the number of seconds between repeats of the WAV sound EFM Options Alarms EFM 100 Options Enable Alarms Lightning Alarm Field Mil Shop Roof Sensitviy 01 kv m Duration 2 minutes MV Enable Lightning Alarm IV Enable Very High Field Alarm Very High Field Alarm I
26. ge is drawn from ground to the plates through a sense resistor When the sense plates are shielded from the field the charge flows back to ground again through the sense resistor This moving charge is an electric current which is measured as an AC voltage across the sense resistor The size of the voltage is proportional to the size of the electric field applied to the plates Sense Electrodes Rotor Rotor Sensor ffs Sensitivity Resistor A Te Tt EA D a pas 100kQ wal Sense Resistor Conditioning H and Supervision Low Pass Filter D Motor Drive Fiber Optic Output Analog Output Figure 3 Field Mill Block Diagram Charge flowing onto and off of the sense electrodes will develop a voltage across the sense resistor This voltage is amplified and fed into an analog switch along with an out of phase version of the signal CHAPTER 2 INSTALLATION Installation Connect and test your EFM 100 on the ground before permanently mounting the sensor unit Install the software on your computer and connect the field mill as shown below The field mill should respond to charged objects brought near it Most plastic objects can be easily charged by rubbing them on your clothes EFM 100 Field Mill Provided 12VDC Power Supply EPM 1 Provided Power Module Provided Cat3 or Cat5 UTP Cable Provided PeplAosd JON IM Punoi9 seddog PEplAoid aqe Sn
27. he ground to prevent the electric field enhancement which occurs when objects are mounted above the ground The permanent field mill is mounted in its permanent location The correction factor needed is equal to the reference field mill reading divided by the permanent field mill reading If two field mills are not available an approximate correction factor can be determined by moving a single field mill from one location to another during a period of stable electric field a Permanent field mill requiring correction Temporary reference field mill Ka mounted flush with the ground l Figure 15 Correcting a Field Mill using a Reference Field Mill 14 CHAPTER 2 INSTALLATION reference correction permanent Once you have calculated the required sensitivity reduction correction factor you can calculate the resistor value needed for the Sensitivity Plug resistor as OOS 100 kQ correction For most installations it is not necessary to correct using a reference field mill It is usually sufficient to choose a sensitivity plug which produces a fair weather electric field of approximately 0 1 kV m and does not saturate above 20 kV m or 20kV m during most thunderstorms Electric field trends and sudden changes lightning are of much more interest than an exact absolute value Alarm setpoints can be lowered to increase alarm sensitivity or increased to reduce the frequency of nuisance alarms
28. na installer See Television Antennas in the Yellow Pages Choosing an Outdoor Mounting Location If possible choose a mounting location for your field mill which will provide easy access for maintenance While your field mill should run for many years with little or no maintenance cleaning or de icing may occasionally become necessary A location which is easy to access can make this a much easier and safer task The choice of a mounting location can affect the operation of the field mill In general higher is better CHAPTER 2 INSTALLATION Mounting the field mill close to tall objects will result low readings as the tall object will shield the field mill sensor 4 4 TERE TOT m Figure 5 A poor choice of mounting locations Avoid mounting the field mill in a location where it will be shielded by buildings tree or other tall objects Mount your field mill with similar considerations given to an anemometer Figure 6 An excellent choice of mounting locations CHAPTER 2 INSTALLATION EPM 1 Power Module The EPM 1 power module is a connection point for sending DC power up to the field mill and connecting to analog signals from the field mill Field Mill Power Cable Connector Bower Supply Ee AE Figure 7 EPM 1 Power Module Connectors Field Mill Power Cable Connector Connect the black cable from the EPM 1 Power Module s field mill power connector to the field mill
29. power connector The cable supplies 12VDC to the field mill and carries the analog output signals back to the EPM 1 See the Lightning Hazard caution in the EPM 1 Analog Output section of this manual Power Supply Connector Connect the provided 12VDC power supply to the EPM 1 s power supply connector CHAPTER 2 INSTALLATION Analog Output Analog output is available on the EPM 1 Power Module Output voltage is 1V per kV m differential ranging from 20V corresponding to 20kV m to 20V corresponding to 20kV m Note that EPM 1 connections are directly connected to the field mill and a lightning hazard exists at those connections LIGHTNING HAZARD EPM 1 connections are not optically isolated from the roof mounted field mill Dangerous voltages may occur should lightning strike at or near the field mill DC Voltmeter 1 Volt per kV m Figure 8 EPM 1 Voltmeter Connections Differential Preferred Connection If a differential connection is not possible a single ended connection may be made DC Voltmeter 0 5 Volt per kV m Figure 9 EPM 1 Voltmeter Connections Single Ended 10 CHAPTER 2 INSTALLATION EFA 10 Fiber Optic Adapter The EFA 10 Fiber Optic Adapter converts the optical data from the field mill to an electrical signal compatible with your personal computer Data is transmitted optically from the field mill at 9600 baud 8 data bits 1 stop bit no parity R
30. s Figure 30 Precipitation Noise Rain snow and dust can carry an electric charge If charged particles contact the field mill sense plates the electric charge will transfer from the particle to the sense plate and will be detected as noise on the field mill reading During periods of heavy rain the precipitation noise can get quite severe Once the storm passes the noise will disappear Precipitation noise can reduce the ability of the software to detect distant lightning but nearby lightning can still be detected 28 CHAPTER 5 MAINTENANCE Ci f Maintenance Your EFM 100 should run for years with very little maintenance An occasional cleaning may be all that is necessary In most installations even that will not be necessary as rain will wash away any accumulated dirt Periodically inspect your field mill to ensure the Teflon insulators are free from dirt or contamination Watch for spider webs or other debris which could short an electrode to ground the case Teflon insulators must be free of dirt or contamination Clean away any spider webs or debris shorting insulators to ground Figure 31 Field Mill Cleaning 29 APPENDIX A TROUBLESHOOTING Troubleshooting Problem Signal Lost Fault Solution The EFM 100 software will report a signal lost fault if it does not receive data from the field mill for more than a few seconds You will need to determine why the data is not reaching the software Possible causes
31. s equipment is tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio or television reception However there is no guarantee that interference will not occur in a particular installation If this equipment does cause interference to radio and television reception which can be determined by turning the equipment on and off the user is encouraged to try to correct the interference by one or more of the following measures e Reorient or relocate the receiving antenna e Increase the separation between the equipment and the receiver e Connect the equipment into an outlet on a circuit different from that to which the receiver is connected e Consult the dealer or an experienced radio TV technician for help WARNING The connection of a non shielded equipment interface cable to this equipment will invalidate the FCC Certification of this device and may cause interference levels which exceed the limits established by the FCC for this equipment It is the responsibility of the user to use a shielded interface cable with this device If this equipment has more than one interface connector do not leave cables
32. will depend on the type of field mill connection USB RS 232 or network Remote Field Mill If the field mill is a remote network connected field mill the problem will need to be solved at the remote computer Troubleshoot at the remote computer as a USB RS232 connected field mill USB RS232 Field Mill If you are experiencing a Signal Lost Fault on field mill directly connected to your computer s USB or RS232 port you will first need to determine if there is data coming from the field mill down the fiber optic cable Check the power supply to the field mill It may be possible to see the optical signal coming down the fiber At the computer unplug the fiber optic ST connector from the EFA 10 Fiber Optic Adapter If the field mill is transmitting data the signal should be visible as a faint red dot in the center of the connector s white tip As the dot is very small pin prick in size it may be necessary to dim the room light to see it If the red dot is visible the field mill is transmitting properly Check your computer connections and configuration It is possible to view the field mill data using HyperTerminal or another terminal program First exit the EFM 100 display software then start HyperTerminal Data is transmitted from the field mill at 9600 baud 8 data bits no parity 1 stop bit RS232 connected field mills usually show up on COM1 or COM2 while USB connected field mills usually show up on COM3 COM4 or even higher com ports 31
Download Pdf Manuals
Related Search
Related Contents
Toastmaster 6104MEX User's Manual Belkin Neoprene Sleeve Untitled - Canadian Appliance Source Guía de usuario del Nokia 6300 - File Delivery Service Thank - Ozone und Bedienungsanleitung Assembly instructions and フラットLEDツイン シリーズ Un exemple d`utilisation du Code_Aster : calcul d`[...] CARELYS Copyright © All rights reserved.