Operating instructions BD balances

image

Contents

1. BD balances Startup procedure 1 Unpacking standard equipment The standard equipment contains e BD balance e Weighing pan permanently mounted with BD6000 e AC adapter e Operating instructions Mounting the weighing pan iiaa m gt 5 4 Location F i iis al With BD6000 Remove packaging material underneath the weighing pan Weighing pan is already mounted With BD202 remove transport arrestment Unscrew arrestment screw 1 Pull out transport arrestment 2 Models with circular weighing pan Insert the weighing pan with the stem in the middle of the pan support 3 Caution Do not alter position of stop screw 5 Notes e Keep packaging for transport or storage in a safe place e For transport move weighing pan to position 4 The best results are obtained when the balance is correctly sited e Firm horizontal support as free from vibrations as possible e Avoid direct sunlight e No excessive temperature fluctuations e No drafts Startup procedure 2 Attaching the AC adapter Plug AC adapter into rear of the balance AZO Connect AC adapter to power supply N a Installing the batteries 1 Lift off circular weighing pan rectangular pan is permanently fixed Turn balance over Slide off cover in the direction of the arrow Take out battery holder Insert 8 AA batteries so that the plus and minus terminals coincid
2. 2 Place empty container on the balance Press L_ gt 0T lt 3 Press and hold AG until SEt is displayed 10 is the preset reference quantity e Pressing S allows a reference quantity of 10 20 30 50 100 or 5 to be selected The selected value remains stored even if the balance is switched off 4 Load reference quantity and confirm with L The current piece number and PC pieces will be displayed If the reference weight is too small Add and the current piece number are displayed e To assure counting accuracy increase reference quantity Confirm with e Ignore Add prompt Confirm with L ri v O Counting Add weighing sample to the container and read off piece number The current reference weight remains stored until the balance is switched off or the entry is changed Switching between piece number weight units Press 5 75 2 Note If no entry has been made for 30 seconds the timeout function switches to normal weighing operation Operation 8 Plus minus and percent weighing Requirement F 100 must be selected in the menu Setting target weight 1 Set display to zero 2 Press and hold A until SEt 100 appears in the display 3 Load target weight 4 Confirm with LE The target weight is now de fined The current target weight remains stored until the balance is switched off or the entry is changed Note If no entry has been made
3. 1 5 V batteries Maintenance Cleaning Clean housing and weighing pan with a cloth and if necessary a mild cleaning agent Protect balance and weighing pan against contamina tion Batteries You can operate BD balances with environmentally friendly batteries Remove discharged or old batteries from balance and dispose of as hazardous waste Accessories by ordering Accu Pac rechargeable NiCd battery Carrying case Antitheft device RS 232 interface RS cable male 25 pin female 25 pin female 9 pin Extension cable DATA I O 5m MiniMettler connector solderable Calibration weights 200 g 500 g 1000 g 2 x 2000 g AC adapter 100 120 V 220 240 V 240 V 220 240 V US connector EURO connector UK connector AUSTRALIA connector OE Le A Battery holder without batteries Battery compartment cover Weighing pans 102 mm for BD202 102 mm for BD601 120 mm for BD1201 Pouch 14 Order No 590103 590116 590101 590102 210492 210491 210493 216152 33930 216522 216524 216526 216532 590108 590109 590110 590111 590112 590115 590118 590113 590114 590117 To protect your METTLER TOLEDO product s future METTLER TOLEDO service assures the quality measuring accuracy and preservation of value of all METTLER TOLEDO products for years to come Please send for full details about our attractive terms of service Thank you P709106 Subject to technical
4. and hold Enter reference weight or reference quantity a Operation 4 press briefly press and hold On off switching Weighing Taring 3 Menu key Transmit data to a printer An RS 232 interface must be fitted for this see Accessories page 14 In the menu Select menu options Calibrate entry into menu store settings When balance is unloaded press _ All segments of the display panel light up The balance model is displayed 0 0 is displayed The balance is in the weighing mode Switching off the balance Press and hold until Off is displayed Set display to zero Press L_ gt 0T lt Load weighing sample Read display when stability detector o disap pears Set display to zero Load empty container on balance The weight is displayed Taring Press L_ gt 97 lt Add weighing sample to container The net weight of the contents is displayed Remove container from balance The weight of the container tare is displayed as a negative value The tare weight remains stored until the is pressed again or the balance is switched off Operation Calibration 1 Before calibrating the balance must be switched on for at least 5 minutes 2 Have required weight See Accessories page 14 on hand BD202 200 g BD601 500 g BD1201 1000 g BD6000 4000g 3 Unload weighing pan completely 4 Press and hold until CAL appear
5. changes and to the availability of the accessories supplied with the instruments Mettler Toledo GmbH 2000 709106A Printed in Switzerland 0001 2 12 Mettler Toledo GmbH CH 8606 Greifensee Switzerland Phone 41 1 944 22 11 Fax 41 1 944 30 60 Internet http Avwww mt com
6. e with the specifications on the holder 6 Insert battery holder and check plug in connection 7 Replace cover ok ODN Notes e When the AC adapter is attached the batteries are bypassed e The balance can also be operated with a storage battery see Accessories Accu Pac page 14 e To prolong battery life activate automatic shut down in the menu see page 11 GO Operation Instrument overview 4 1 Circular or rectangular weighing pan 2 Display panel 2 ANN 3 Keypad S amp S Green function Press briefly 3 Blue function Press and hold 4 5 6 7 4 AC adapter socket 5 Model plate pmm cI F 6 Antitheft device 7 Cover of RS 232 socket Operation BD balances have two operator control levels the 2 operator control levels weighing mode and the menu General operation is in the weighing mode e Weighing mode In addition to the basic functions weighing and taring page 4 onwards other functions are available such as unit switching and counting or percent weighing Selection of the func e Menu page 9 onwards tions and setting the balance parameters takes place in the menu page 9 onwards bor wi de hw 2 3 J 1 Control key press briefly Switch on press briefly Zero setting taring return to weighing mode Of press and hold Switch off 2 Function key press briefly Switching between unit 1 unit 2 or piece number In the menu Change the settings press
7. for 30 seconds the timeout function switches to normal weighing operation Percent weighing 1 Set target weight see above 2 Load weighing sample The weight of the sample is displayed in percentage of the target weight Switching between percent weighing weight units A Plus minus weighing a 1 Set target weight see above ae 2 Tare Press L_ gt 07 lt 3 Load weighing sample The percentage by which the weighing sample differs from the target weight is displayed Switching between plus minus weighing weight units Press S Menu Menu overview E gt gt You can use the menu to configure the balance for your special requirements The selection line with halftone backing represents the current settings of the balance You can access it with the L key The choices shown in halftone appear only when the RS 232 interface is built in see Accessories page 14 Fro Bunk rogentez ofie Erorr fisse e 200 iP E i ESE 4 Feoot oz oz int 3 Roes sa e 4808 e 0 v EE a e re ra eo Menu operation Entry into the menu Selecting menu options Changing the settings Storing changed settings Return to the weighing mode without storing changes The possible settings are listed in the description of the menu options Press and hold until MENU is displayed Select next menu option with Change the settings of a menu option using amp Press and hold key unti
8. isplay has occurred within the specified time Inactive Ee 0 5 minutes 1 1 minute 2 2 minutes Reset Reset resets the menu to the factory settings To reset settings Press and hold the until r donE is displayed The balance returns to the weighing mode Error messages 12 Overload Underload Note If an overload or underload error appears when the balance is switched on recalibrate balance No stability e in taring e in calibration e when reference weight is loaded for counting or percent weighing No or wrong calibration weight loaded gt Load requested calibration weight Reference quantity too small gt Increase piece number Reference weight too small gt Increase reference weight Internal error gt off on switching or calibration If the error can not be rectified return balance to customer service You will find further details on the warranty certificate bill Technical data 13 BD202 BD601 BD1201 Readability 0 01 g 0 19 0 19 Maximum load 200 g 600 g 1200 g Taring range by subtr 0 200g 0O 600g 0O 1200g Reproducibility 0 01 g 0 07 g 0 19 Linearity 0 01g 0 19 0 19 Stabilization time typical 3 0 s 3 08s 355 Balance housing W x D x H 185 x 220 x 60 mm Weighing pan 102mm 102mm 120 mm Weight ca 800 g Power supply 6 12 V DC 70 mA BD6000 1g 6000 g 0 6000 g 0 79 1g 3 08S 120x175 mm with AC adapter Accu Pac or 8 AA
9. l StorEd is displayed Quit the menu with L_ gt 7 lt The balance operates with the settings currently stored Note If no entry has been made in the menu for 30 seconds the timeout function switches to normal weighing op eration The balance operates with the current stored settings Menu 10 Menu factory settings rn in rr o or ri Rx ml Cr E x Menu description T J O F no Unit1g Unit 2 g Int 2 A off S Stb b 2400 P E Function no count 100 Unit 1 OZ oz t dwt Ht St tt no function Unit 1 ing Unit 2 in g inactive median integration time automatic shutdown inactive Transmission mode stable single value Baud rate 2400 Parity even no function counting Plus minus and percent weighing Gram Ounce Troy ounce Pennyweight Carat Grain Pound Momme Hong Kong tael Singapore tael Taiwan tael During operation only t is displayed 11 INT RESET CAL Menu Unit 2 All units listed under unit 1 except taels can also be selected for unit 2 Integration time The integration time influences the speed of the balance and the reproducibility of the values 2 Median integration time 3 Maximum integration time 1 Minimum integration time Weighing speed Reproducibility 1 fast good 2 t l 3 slow very good Automatic shutdown The automatic shutdown switches off the balance when no change in the d
10. s Re lease key 5 After attainment of stability the value of the re quired calibration weight flashes 6 Place calibration weight on the balance The balance calibrates itself 7 When zero flashes remove weight The operation is complete when C donE is dis played The balance returns to the weighing mode Disabling the calibration Switch 1 to On position Calibration disabled Switch 1 to Off position Calibration enabled Changing the switch position 1 Remove circular weighing pan rectangular pan is permanently mounted 2 Turn balance over 3 Unscrew screws of baseplate and remove base plate 4 Move switch to desired position 5 Replace baseplate Note When the balance is open do not touch the leads of the sensor Operation Switching weight unit ig PC Printing Requirement Different weight units for unit 1 and unit 2 must be selected in the menu Switching Switching between unit 1 and unit 2 Press S If a function has been selected in the menu switching can be effected between unit 1 unit 2 and or piece number If an RS 232 interface is installed weighing results can be transmitted to a printer and printed out Parameters of the data transmission are defined in the menu see interface description Start printout Press Operation 7 Counting Requirement F count must be selected in the menu Setting reference 1 Set display to zero

Download Pdf Manuals

image

Related Search

Related Contents

none TGA0P2557S00 Instructions / Assembly  NO APARECIMENTO DE LESÕES POR ESFORÇOS REPETITIVOS  Gebrauchsanweisung Leica VT1000S V. 2.0 RevD  KIT PUERTA CORREDERA Manual de usuario  Istruzioni per l`uso  Mio 100 – Manuale Uso Manutenzione e Garanzia 4 anni  ISAS Trieste La sicurezza nei laboratori  KitchenAid KAWE977B User's Manual    Lightolier IS:1964 User's Manual  