User manual CuroCell S.A.M.®

image

Contents

1. Zi Recommended user weight Counteract shear force Machine wash at 95 C Class Il equipment Double Insulated i XXXX XX XX Year Month Day CE CE marked acc to MDD 93 42 EEC Category Patient information category Heel function gt lt 1 Chlorine X Do not iron lt 3 gt gt gJ T The mattress should only be used lying in the lengthwise direction on the mattress 12 1 Foot placement y Y e Recycling Directly on bed base TX Do not wash Q Tumble dry RS Do not dry clean Pan IP class Ingress Protection Rating 9 Spare parts Cover Curocell S A M and CuroCell S A M CF10 Cover Olivia blue Art No 294 080200 294 080210 294 085200 294 085210 294 090200 294 090210 294 105200 294 105210 294 120200 294 120210 Air cells incl quick connector and check valve Art No CC 250570 CC 250571 CC 250573 CC 250574 CC 250576 CC 250577 CC 250578 CC 250580 CC 250579 CC 250581 Pump Art No CC 250101 CC 250102 Art No CC 908032 If other spare parts are needed please contact Care of Sweden or your local dealer for further assistance Size 80x200 cm 80x210 cm 85x200 cm 85x210 cm 90x200 cm 90x210 cm 105x200 cm 105x200 cm 120x200 cm 120x200 cm Size 80x200x10 cm 80x210x10 cm 85x200x10 cm 85x210x10 cm 90x200x10 cm 90x210x10 cm 105x200x10 cm 105x210x10 cm 120x200x10 cm 120x210x
2. If your questions are not answered by the above information please contact Care of Sweden or your local dealer for further assistance 10 8 Technical description 8 1 Technical specification SPECIFICATION Model e Vonage o ovos O e Dimensions towa sss OO e T T Oo Oooo eseo ining 26708 Fruity erations 10 20 non condensing fT storage 10 909 non condensing C itimaspherie pressure 700hPa 1003 25hPa O O Cession ass Type BF PQ MATTRESS SPECIFICATION Model CuroCell S A M CuroCell S A M CF10 CuroCell A M CF16 Dimensions WxLxH pF 80 85 90 105 120x200 210x10 16 cm NOTE The manufacturer can at any time change the specifications for the product WARNING Follow the operations and maintenance instructions carefully in order to prevent bodily injuries as well as damages to the product or other equipment 8 2 Labelling The mattress are labelled with product name production date article number size care and wash instructions EAN bar code as well as space for own labelling The product also displays which side is up down and head foot 11 8 3 Symbol Definitions Article No Read the user manual pa Not in domestic refuse follow recycling instructions I Place on top of existing mattress Machine wash at 70 C me Do not tumble dry Wipe off Temperature range Manufacturer GROY anna RL Ss nuns Group of function gt x S lt S lt
3. Care of sweden Supporting Life User manual 2014 04 17 CuroCell S A M Static Antidecubitus Mattress User manual item number 95 001298 7 331345 040226 CE Contents Important safety instructions Introduction Product description Installation and start up Cleaning and reconditioning Storage Maintenance Troubleshooting Technical description Spare parts Options and accessories Warranty Other information o Oo UN A W 10 10 10 11 13 14 15 15 IMPORTANT SAFETY INSTRUCTIONS Read all instructions prior to use ae ee a a ae ae a a a a ee ee ee ee ce ee ee ee ee ee eee WARNING To minimise the risk of fire electric shocks or bodily injury 1 The product must only be used for its intended purpose as described in this manual The product may not be combined assembled or repaired with other parts e g pump and mattress accessories or spare parts other than those described in this manual or other documentation from Care of Sweden 2 Donot use the pump in wet rooms or where there is a risk of the pump coming into contact with water or another liquid Never handle a pump that has come into contact with water liquid Pull the plug out of the electrical socket immediately and send the product to an authorised service technician for servicing 3 DO NOT USE in the vicinity of or in contact with fire sources hot surfaces such as fire burning cigarettes or hot lamps Although the product has undergone fir
4. 1 for instructions MATTRESS 1 Disconnect the tubes and let the air out also open the CPR valve 2 Remove the cover See instructions in chapter 4 1 for cleaning covers If the product is equipped with a foam core this should be removed and cleaned separately 3 Move the mattress to a clean area which is suitable for cleaning 4 Wipe off the cells all tubes and the CPR valve with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 5 Wipe off and disinfect all the work surfaces with an alcohol based cleaning agent intended for this purpose 6 Put the mattress together If cells for any reason have loosened from the tubes these must be replaced according to drawing in chapter 2 3 FOAM CORE CuroCell S A M CF16 Clean the affected area with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 Press out the water carefully Let the foam core dry in a warm ventilated area without direct exposure of sunlight The foam core needs to be completely dry before it is used again NOTE Check the hygiene cover each time the product is cleaned If damaged it must be replaced or repaired Also check the pump tube connectors and power cord during cleaning Damaged parts must be replaced or repaired 9 5 Storage and packing It is advisable to s
5. 10 cm Description CuroCell S A M CF16 Cover top part grey Art No 431 080200 431 080210 431 085200 431 085210 431 090200 431 090210 431 105200 431 105210 431 120200 431 120210 Air cells incl connection Art No CC 250630 CC 250631 CC 250633 CC 250634 CC 250636 CC 250637 CC 250638 CC 250639 CC 250640 CC 250641 CuroCell S A M pump without quick connection CuroCell S A M pump incl quick connection Description Tubing set incl CPR 13 Size 80x200 cm 80x210 cm 85x200 cm 85x210 cm 90x200 cm 90x210 cm 105x200 cm 105x210 cm 120x200 cm 120x210 cm tubes with CPR and quick Size 80x200x10 cm 80x210x10 cm 85x200x10 cm 85x210x10 cm 90x200x10 cm 90x210x10 cm 105x200x10 cm 105x210x10 cm 120x200x10 cm 120x210x10 cm 10 Options and accessories 10 1 Options Inner cover Soft Enhances the comfort further can also be used for very sensitive users This is included for CuroCell S A M CF16 Place the air cells inside the cover with the soft foam side up Sturdy transportation bag Transportation and storage bag Cover With welded seams 10 2 Accessories Cover Curocell S A M and CuroCell S A M CF10 CuroCell S A M CF16 Inner cover Soft Inner cover Soft with foam pocket Art No Size Art No Size CC 928040 80x200 cm CC 928060 80x200 cm CC 928041 80x210 cm CC 928061 80x210 cm CC 928043 85x200 cm CC 928063 85x200 cm CC 928044 85x210 cm CC 928064 85x210 cm CC 928046 90x200
6. E 514 23 Tranemo SWEDEN Visit address Fabriksgatan 5A SE 514 33 Tranemo SWEDEN Cargo address Byns vag 4A SE 514 33 Tranemo SWEDEN Supplied by Care of Sweden and CuroCell S A M are trademarks of Care of Sweden AB Care of Sweden AB 2014 16
7. be used in all kinds of caring environments and it is very functional and cost effective Always read the user manual prior to use 1 1 General information This CuroCell mattress is a medical device in accordance to MDD 93 42 EEC According to statutory regulations made by the authorities regarding medical devices the manufacturer is required to report all accidents or incidents involving there products We would be very grateful for all information involving accidents or incidents relating to our products by reporting them immediately to us here at Care of Sweden This system has been tested and approved according to the following standards EN 60601 1 SS 876 00 04 SS EN ISO 12182 SS EN ISO 14971 SS EN ISO 10993 EN 597 1 EN 597 2 EN 14126 1 2 Intended use CuroCell S A M can be used in all types of medical environments and is intended to assist in the prevention and treatment of pressure ulcers up to and including category 2 The recommended user s weight is up to 200 kg It is available in two models As an overlay mattress CuroCell S A M to be placed on top of an existing mattress or as a mattress replacement CuroCell S A M CF16 We recommend that repositioning of the user is performed regularly The time interval must be assessed by the responsible personnel depending on user s status diagnostic and general condition The mattress must always be cleaned between patients NOTE e When used together
8. cm CC 928066 90x200 cm CC 928047 90x210 cm CC 928067 90x210 cm CC 928048 105x200 cm CC 928068 105x200 cm CC 928029 105x210 cm CC 928070 105x210 cm CC 928049 120x200 cm CC 928069 120x200 cm CC 928051 120x210 cm CC 928071 120x210 cm Art No Description CC 9091 Transportation and Storage Bag CC 9093 Transportation and Storage Bag to CuroCell S A M with Inner cover Soft 58 900909 Transportation and Storage Bag to CuroCell S A M CF16 CC 9098 Pump bag black 10 3 Other accessories To ensure the best level of comfort we recommend use of accessories from Care of Sweden since our products are designed to work well together Please contact Care of Sweden or your local dealer for accessories or more information on our products 14 11 Warranty 11 1 Scope This CuroCell system is covered by a 2 year warranty for manufacturing defects The warranty does not apply to normal wear and tear or product impairment or damage stemming from negligence or improper handling care 11 2 Service and repairs This CuroCell mattress is a medical device in accordance with MDD 93 42 EEC that is used as an aid in preventing and treating pressure ulcers For this reason we recommend that the pump be regularly serviced and inspected to maintain functionality and performance Like other technical devices a pump that is properly cared for works better and lasts longer Service and maintenance must always be performed by Care of Sweden or one of its auth
9. e testing it could become damaged if it comes into contact with fire sources 4 Position and use the product in a manner that prevents it from becoming trapped or damaged NOTE Be particularly aware of trapping damage when using bed rails 5 Never use the product if the pump power cord or plug is defective the pump casing is damaged or it is not functioning properly Contact a service technician for examination and repair 6 Regularly check product functionality through a hand check see section 3 4 Also check that the comfort setting is correct as indicated in the table on the pump 7 Continual monitoring is required when the product is used for individuals needing special supervision such as children 8 The mattress are protected with a hygiene cover Please try to avoid the simultaneous use of multiple hygiene cover with this mattress Because of the risk of vapour permeability can be destroyed 9 The hygiene cover does not allow liquid or air to penetrate but is vapour permeable Make sure that the patient is positioned correctly to avoid the risk of suffocation 10 Prevent damage to the hygiene cover from sharp objects 11 If the cover are supplied with handles on the long sides theses handles are only to be used for lifting and moving the mattress itself All other use will be at your own responsibility and is not covered by the product warranty 1 Introduction CuroCell S A M is a static air mattress that can
10. el 2 Tube connections with quick connection and without quick connection Back side 1 Bed hooks 2 Main power switch side For switching on off the pump Green lamp indicates when the pump is running 3 Power cord 2 2 Front panel Comfort settings The comfort setting adjusts the air pressure in the mattress When the knob is turned clockwise the air supply and pressure in the mattress increase When the knob is turned counter clockwise the air supply and pressure in the mattress decrease Appropriate air pressure is set between 1 and 8 The table on the pump gives guidance 2 3 Mattress 1 Cells air bladder 2 Tubes 3 Tubeholder 4 CPR Rapid deflation 2 4 Other parts Hygiene cover Inner cover Soft accessory Transport and Storage Bag accessory Pump bag accessory Integrated foam core only CuroCell S A M CF16 3 Installation and Start up 3 1 Installation When unpacking the mattress check that none of the parts are damaged In the event of damage contact Care of Sweden or your local dealer before using the product 1 Place the mattress on top of an existing mattress If possible secure it by pulling the elastic bands on the bottom over the corners of the underlying mattress Note If the mattress have an integrated foam core CuroCell S A M CF16 place the mattress directly on to the bed base Use the straps on the cover to fasten the mattress to the bed base 2 Hang t
11. he pump by the foot end of the bed or place it on a flat sturdy surface Place the power cord so there is no risk of stumbling over it run it over with the bed wheels or getting it jammed when raising or lowering the bed NOTE A pump bag is available as an option 3 Connect the air tubes coming from the mattress to the side of the pump 4 Check to make sure that the power is set to 0 off Plug in the power cord into an approved outlet 220 V NOTE e Make up the bed with sheets to increase user comfort e Check that the CPR valve is closed 3 2 Start up and settings NOTE Always read the user manual prior to use 1 Start the pump by pressing 1 on on the side of the pump The green light is lit 2 The pump starts filling the mattress with air It takes approximately 15 20 minutes 3 If necessary the patient can lie on the mattress even when not inflated in this case the underlay will carry the load When using a mattress with an air underlay this must be inflated 3 3 Comfort settings kg 0 20 1 2 The air pressure in the mattress can be adjusted according to the 30 50 2 3 patient s desire for comfort Always consult the prescriber or other a gt authorised personnel for the appropriate setting TA E Adjust the pressure with the comfort setting knob according to the Pee f table on the pump 190 200 8 8 NOTE e When the head end of the bed is raised the pressure s
12. hould be increased at least one 1 step e For amputated users when only part of the mattress is under load the pressure should be increased by 1 2 steps e Recommended users weight is up to 200 kg 3 4 Hand check Hand check is used to ensure that the mattress is properly inflated This should be done regularly once per shift after changing the position in the bed or changing the comfort settings Open the cover Place a flat hand under the air cells directly underneath the user s pelvic region and make sure that the user does not lay directly on the underlying mattress There should be about 1 2 fingers margin If you can feel the user s body resting on your hand you need to increase the comfort settings to next highest step Wait 10 minutes and perform a new hand check 3 5 CPR Ll In case of an emergency where CPR cardiopulmonary resuscitation is necessary Rapidly pull the CPR band on the tubes between the pump and the mattress Remove the tube connections on the side of the pump for even faster deflation Note Make sure that the CPR connector is properly connected after use 3 6 Low pressure By an abnormally low air pressure see chapter 7 for trouble shooting WARNING If the air pressure remains low search for damages on the mattress and the tubes Repair when needed or contact the supplier 4 Cleaning and reconditioning It is important to follow the instructions below before using the product f
13. or another user Consult the hygiene responsible or the supplier for help and instructions when unsure 4 1 Cleaning PUMP VARNING Do not open the pump housing risk of electric shock Service and maintenance must be performed by authorised technicians Do not get the pump wet or submerge it in any liquid Wipe off the pump with a damp cloth and a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 If using a different agent choose one that does not affect the plastic of the pump housing MATTRESS COVER Covers can be wiped off with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 Excrements and blood spots must be removed immediately with cold water Carefully follow local instructions and usage instructions for the cleaner 7 O u Bat BS Multi layer covers should be divided before washing Wash with like colours Chlorine and phenol based cleaners could have a negative impact on the PU surface lt jy and should be avoided If chlorine is used we recommend a mixture of max 1 INNER COVER SOFT accessories w Q ul BX Rx Clean the affected area with a mild cleaning solution for this purpose such as alcohol with or without tensides Isopropanol 70 oxidating solutions or chlorine max 1 4 2 Reconditioning PUMP See section 4
14. orised technicians Only use spare parts that have been approved by Care of Sweden Use of non approved spare parts invalidates the warranty A warranty claim must be submitted to Care of Sweden before the product is returned Please contact Care of Sweden or your local dealer if you have questions regarding warranty and service 11 3 Life expectance This product s expected life is estimated to be three years 12 Other information 12 1 Disassembly and recycling Except for certain parts of the pumps energy recovery is possible for almost all material in CuroCell products through incineration in waste incineration facilities Pump A used CuroCell pump must not be disassembled It must instead be taken to a recycling centre and sorted as electronic waste Mattress A used CuroCell mattress must be taken to a recycling centre and sorted as combustible waste For more information contact Care of Sweden or your local dealer 12 2 Returns and complaints Complaints are to be reported to the supplier Return of a product shall follow the supplier s instructions and can only be done after the complaint has been reported Damages due to the transportation should be reported to the transportation agent 15 bead Manufactured by Care of sweden Supporting Life Contact Phone 46 771 106600 Fax 46 325 12840 Email info careofsweden se Internet www careofsweden com Address Care of Sweden AB P O Box 146 S
15. tore the mattress and pump in the product bag accessory Handle the packed product very carefully Do not place any heavy objects on top of it 6 Maintenance 1 Check the power cord and plug for damages and scratches 2 Check the mattress cover to make sure it is undamaged and that cover and cells are correctly assembled 3 Check that air is coming from the outlets on the side of the pump 4 Check for damaged or pinched tubes and connectors Contact the manufacturer or your local dealer for spare parts 7 Trouble shooting Problem Solution The pump does not start Check that the power cord is connected to an electrical outlet Check that all fuses are OK The user lies directly Pressure might be incorrectly set Adjust it according to the on the bed weight table and wait a few minutes Check again using the hand check see chapter 3 4 Check that the CPR valve is closed The mattress moves around Check that the mattress is secured with the elastic bands to the underlying mattress Or if the mattress have an integrated foam core CuroCell S A M CF16 secured by using the straps on the cover to fasten the mattress to the bed base The pump makes a noise Check how the pump is hung on the bed Remove the pump or vibrates and listen whether this makes a difference Can be resolved by putting the pump on a flat steady surface or by placing a towel between the pump and bed Pump bag accessory can also be used
16. with positioning pillows read the pillow user manual carefully for correct positioning of the patient e f you are using this product with an evacuation sheet it is the user s responsibility to ensure that it is safe to use for evacuation e Be careful when using bed rails or other protection on the bed so that the mattress does not become crushed or damaged e The mattress must not be lifted with a person lying on it or used for any other type of transport other than a possible evacuation using an evacuation sheet e This mattress is unsuitable for x ray examinations because of the risk of blurred images or artefacts that may lead to diagnostics errors 1 European Pressure Ulcer Advisory Panel Quick reference guide 2009 1 3 Other information Action plan Pressure ulcers To achieve the best possible result for this product it is recommended that a well planned and well structured working environment be used In our guideline Action plan you will find examples which can be applied and put into practice when using our products Mattress guide This practical guide gives you as a care giver help in choosing the correct mattress from the Care of Swedens range CuroCell S A M is a mattress in function group B This information can be found on our website www careofsweden com or you can order a copy from our Customer services by sending your details to info careofsweden se 2 Product description 2 1 Pump 1 Front pan