1. 0 1 L min Accuracy 7 of read value OXYGEN e Pressure included between 280 kPa and 600 kPa 2 8 6 bar e Max required flow 90 L min NITROUS OXIDE e Pressure included between 280 kPa and 600 kPa 2 8 6 bar e Max required flow 15 L min MEDICAL COMPRESSED AIR e Pressure included between 280 kPa and 600 kPa 2 8 6 bar e Max required flow 90 L min No 3 on front panel O2 N20 AIR scale 0 6 bar 100 240Vac 45 60Hz 12Vdc 3 0 Ah Pb battery it grants 120 minutes operation e Lack or low O pressure with consequent cut off of N20 delivery e Lack or low N20 pressure e Lack or low Xe pressure e Lack or low AIR pressure e Fresh gas occlusion 3 8 Safety devices AGAINST THE ADMINISTRATION OF HYPOXIC MIXTURES MIX LIFE It always guarantees a minimum concentration of 25 oxygen on mixtures which includes nitrous oxide IN CASE OF LACK OR LOW OXYGEN PRESSURE CUT OFF Visible and audible alarm with immediate cut off of nitrous oxide delivery AGAINST OVERPRESSURE IN FLOWMETER BOX Pressure sensor for the protection of flowmeter box components IN CASE OF LACK OR COMPRESSED AIR LOW PRESSURE All the devices gas feeding supplied by compressed air are automatically supplied by oxygen AGAINST THE SIMULTANEOUS DELIVERY OF AIR AND N20 Selection by membrane key on the flowmeter front panel Control for activation of Setting of MANUAL modality on ventilator MAN with automatic de
2. consumption are reduced to the minimum The heated valves group reduces the condensation and heats the fresh gas The transition from one ventilation modality to another is completely controlled by the ventilator without any user s action on valves group LUNG VENTILATOR VC VAC O 20 40 60 80 15 1 2 99 490 02 09 11 P 09 25 01 010429 User s interface 12 TFT high resolution colour display with membrane keyboard and encoder Control modality Electronic by microprocessor Dead space compensation system Automatic Flow generation Electronic system Gas feeding Medical compressed air or Oxygen with pressure included between 280 kPa and 600 kPa 2 8 6 bar Option turbine for gas feeding independent from gas pipelines system Morpheus_E_te_rev2 5 8 Autotest Primary test at anaesthesia unit s start up a control test of medical gas supply circuit losses INSP and EXP flow sensors operation pressure sensor patient circuit losses back up battery state oxygen cell integrity of audible alarm is automatically performed This test takes around 15 seconds Subtest the unit for anaesthesia has a subtest which is activated by the user in the ventilator menu This subtest permits to verify the dead space and losses or to perform the oxygen cell calibration Ventilation modalities APCV PCV APCV TV PSV APNOEA BACK UP VC VAC SIMV PS volumetric SPONT MAN
3. MORPHEUS E Anaesthesia Unit code OM3 SE rev 2 dated 09 01 2012 INTENDED USE The MORPHEUS E is an anaesthesia unit and it can be used on adult children and newborn patients The MORPHEUS E is suitable for administration of Oxygen Air Nitrous Oxide Halothane Enflurane Isoflurane Sevoflurane Desflurane mixtures GENERAL DESCRIPTION The MORPHEUS E anaesthesia unit is completed with Morpheus_E_te_rev2 electronic gas mixing system electronic lung ventilator with 12 TFT colour display valves group open semi closed closed heated with soda lime absorber of 1 Kg capacity SIARETEX rapid connection device Selectatec compatible for 2 vaporizers gas supply group TECHNICAL DATA Structure Wheels Drawer Cylinder support Support for 2 vaporizers Auxiliary power supply outlets Work shelf lighting Dimensions Weight Environmental conditions Light aluminium alloy and plastic moulds Pivoting antistatic wheels diameter 100 mm 2 with brakes No 3 full extension drawers No 2 vertical cylinders supports on the back side for cylinders up to 10 litres capacity and round rubber pads On horizontal guide SIARETEX rapid connection device Selectatec compatible for 2 vaporizers No 1 SCHUKO 220 Vac outlet max 6 A 12Vdc by led 71 x 77 x 138 L x P x H cm without monitor 72 kg without accessories e Temperature from 10 to 40 C e Relative humidity fro
4. UAL Integrated NEONATAL ventilation mode VC VAC BABY Ventilation modalities optional CPAP SIMV PS by pressure BILEVEL S ST Breathing rate From 5 to 120 bpm step 1 bpm I E Ratio 1 1 1 1 5 1 2 1 3 2 1 3 1 Inspiratory time From 0 2 to 5 sec Inspiratory pause From 0 to 60 of inspiratory time SIMV rate From NO 1 119 bpm Tidal volume From 5 to 1500 ml lt 50ml step 1ml 50 100ml step 5ml gt 100ml step 10ml Minute volume From 1 to 30 liters PEEP OFF 3 30 cmH20 step 1 cmH20 Inspiratory flow From 1 to 80 L min Minimum resolution 1 Oximeter Automatic calibration procedure Bronchomanometer 20 80 cmH20 Flow trigger From OFF 1 15 L min step 1 L min Pressure trigger From 1 to 9 cmH20 under the PEEP level Safety Electronic and mechanical limit of airways pressure Self diagnosis system Morpheus_E_te_rev2 6 8 Alarms e Low High airways Pressure Low High Breathing Rate Low High O2 Concentration Low High Tidal Volume Electric Power Supply e Apnoea Low Battery Low Gas Supply Disconnected Patient Circuit Can Bus Failure Flow sensor Internal to the valves group by magnetic perturbance reusable Measured parameters e PAW PEEP Rate I E FiO2 Vte ExoMV e MAP Pplateau Tpause Ti Te Fi Fe Cs Ri Ventilation Curves CURVES Pressure Flow Volume LOOPs Vo
5. lume Pressure and Flow Volume Measurement RANGE automatic Trend Scale and 72 hours trend period setting Foreseen Trends PAW PEEP VTe ExpMV Rate Events Memory storage up to 100 events per machine including the alarms Configuration and gas type selection optional Software for analysis of CO2 O2 N20 AG automatic identification MAC based on weight e Mainstream external device e Sidestream built in device Note Consult the relevant technical data sheets for mainstream and sidestream modules ELECTRIC POWER SUPPLY Electric power supply 100 240Vac 45 60Hz Maximum power 120 Watt Back up battery 12Vdc 3 Ah pb battery which guarantees an autonomy of around 120 minutes Charging time Around 10 hours Morpheus_E_te_rev2 7 8 CONFORMITY TO DIRECTIVES Class and type according with IEC 601 1 Class Type B Class according with 93 42 EEC Dir ve Class IIb IEC 601 1 IEC 601 1 1 IEC 601 1 2 IEC 601 1 4 EN 1281 1 UNI EN 740 UNI EN ISO 9703 3 EN 4135 93 42 EEC Dir ve ACCESSORIES Standard accessories User s Manual O2 supply hose N20 supply hose Air supply hose O2 cylinder supply hose N20 cylinder supply hose e Top Special CO2 absorber canister of 1 kg O cell Adult silicone patient circuit Adult Mapleson C adult patient circuit Manual ventilation KIT SHUKO VDE electric power supply cable Other op
6. m 10 to 90 non condensing ELECTRONIC GAS MIXING SYSTEM Me phe 4 5 a Morpheus_E_te_rev2 It has the function to regulate the capacity and the concentration of gas mixture Air O2 N20 by displaying them on the TFT colour monitor and deliver it to the anaesthetic gas vaporizer It allows to select the mixture to be delivered Air O2 or N20 O2 and the O2 enrichment for delivered mixture in case of emergency The anaesthesia module includes a device which guarantees a minimum concentration of 25 oxygen in all gas erogating conditions MIX LIFE device Through the three pressure gauges on the front panel it allows the continuous control of medical gas feeding pressure coming from the gas pipelines system The electronically controlled flowmeter box is equipped with a safety traditional calibrated flowmeter indicating the total delivered flow The electronic flowmeter box provides the option on demand to use an alternative anaesthetic gas inspite of N20 the Xenon Xe 2 8 Auto test Oxygen rotameter Nitrous oxide rotameter Air rotameter Medical gas supply Gauges Electric power supply Back up battery Alarms Morpheus_E_te_rev2 At electronic flowmeter start up a control test is automatically performed Scale 0 1 15 L min Resolution 0 1 L min Accuracy 7 of read value Scale 0 1 9 L min Resolution 0 1 L min Accuracy 7 of read value Scale 0 1 15 L min Resolution
7. tional accessories See current export price list SIARE applies the UNI EN ISO 13485 2004 Quality System and 93 42 EEC Dir ve SIARE ENGINEERING INTERNATIONAL GROUP s r l Via Pastore 18 40056 Crespellano Bologna ITALY Tel 39 051 969802 Fax 39 051 969366 E mail mail siare it http www siare it Morpheus_E_te_rev2 8 8
8. viation of exit of fresh gas for fresh gas to the manual system of anaesthesia unit valves group or to a TO manual ventilations AND FRO circuit with visual indicator Automatic deactivation of manual ventilation systems directly by ventilator control Oz emergency by pass By apposite membrane key on the front shelf max flow 35 L min IN gas sockets on gas e No 3 sockets for distribution system O2 N20 AIR supply group e No 2 sockets for cylinder O2 N20 OUT gas sockets on gas No 1 sockets for O2 supply group e No 1 sockets O2 AIR for active scavenger feeding e No 1 fresh gas connector for external use for ex TO AND FRO selectable by apposite membrane key on the front shelf AUX Other e Socket for recycle of exhaust monitor gas e Connection for anaesthetic gas scavenging optional device active type or passive type Morpheus_E_te_rev2 4 8 BREATHING SYSTEM Compact system with automatic connections easy dismountable and autoclavable It allows the ventilation in modality real open circuit semi closed circuit closed circuit at low flows The system also allows the spontaneous and manual ventilation in case of anaesthesia unit breakdown or machine off Top special CO2 absorber canister of 1 Kg with rapid connection this allows canister replacement also during interventions the canister is autoclavable and reusable The recycling system is a selective type hence the soda lime and fresh gas

Download Pdf Manuals

image

Related Search

PDF pdf pdf editor pdf to word pdffiller pdf to jpg pdf combiner pdf merger pdf converter pdf reader pdf to excel pdf editor free pdf to png pdfescape pdf24 pdf compressor pdf to word converter pdf history pdf viewer pdf drive pdf-xchange editor pdf to jpeg pdf files pdf converter free pdf to excel converter pdf to jpg converter