        Model 5600 User`s Manual
         Contents
1.    Model LMU 209 Load Monitoring Unit       1 0    1 1    INTRODUCTION       Figure 1 LMU 209 Load Monitoring Unit    DESCRIPTION    The LMU 209 Load Monitoring Unit is a high precision  ultra stable DC measuring amplifier that  is capable of various low level mV signals  The LMU 209 offers two calibrated outputs  a voltage  output and a current output  both according to the industrial standard  Coarse gain and offset range  are selected with easily accessible DIP Switches  Fine tuning can be accomplished by adjusting  the 10 turn potentiometers  An additional low pass filter reduces bandwidth and helps to improve  signal noise ratio  The bridge power supply can be selected between 5 V or 10 VDC  The current    1 2    output is configured from 0 20 mA or 4   20 mA without any calibration     TECHNICAL SPECIFICATIONS       Supply    18 28 V   70 mA       Ripple Voltage    max  1 Vpp   50 Hz       Voltage Output    O to  10 V   Rload 3 kQ       Current Output    0 4 20 mA   Rload 0 800 Q       Operating Temperature Range     20 to  60   C       Housing Dimensions    44 x 74 x 51 4 mm       Housing Material    PA  UI94 V 0 polycarbonate   aluminum  option        Protection Class  Weight    IP 52  polycarbonate  or IP 65  aluminum   100 g       EMC Tests    EN 61000 4       Frequency Response    0 500 Hz   3 dB        F R  with selectable low pass filter    0 3 kHz    3 dB        Sensitivity  default     1 mvV V       Sensitivity Ranges          Range 1   0 5 to 1 5 mV V  Range 2
2.   1 5 to 4 0 mV V             User   s Manual    Magtrol Model LMU 209 Load Monitoring Unit          Adjustment Sensitivity    10 turn potentiometer       Zero Adjustment Ranges     75  to  75   in 5 ranges       Zero Fine Adjustment    10 turn potentiometer       Bridge power supply    5 10 VDC  selectable        Input resistance Sensor 5V  120 Q to 10 kQ  10V  330 Q to 10 kQ    100   10 V or 20 mA     0 8        Calibration Signal                         Noise max  20 mVpp  Linearity Error 0 05    TC Zero  lt  0 01     C  2 0 SAFETY  USE CAUTION AND READ THIS OPERATING MANUAL  THOROUGHLY BEFORE BEGINNING YOUR INSTALLATION  PROCESS  SETUP  CONFIGURATION AND OPERATION OF THE LMU       209 SHOULD ONLY BE CONDUCTED BY QUALIFIED PERSONNEL        2 1  Proper Use oF EQUIPMENT  e The LMU 209 Load Monitoring Unit must not be used in life supporting applications   e The LMU 209 Load Monitoring Unit must not be used in safety critical applications     2 2 SAFETY PRECAUTIONS    Non compliance with the safety regulations could result in serious damage to the unit     e Always separate the signal lines from power lines  drives  converters  and never interrupt  the signal lines unless using designated junction boxes     e Attach the cable shield  braid  at both ends to a proper ground  Do not allow a ground  current to flow over the cable shield     e Attach all machine parts to one well defined potential by using large copper wires  Allowing  different potentials will create currents tha
3.  output will also go to 100      Definition  Usur 0to 10V   lir 0 20 mA or 4 20 mA       Magtrol Model LMU 209 Load Monitoring Unit User   s Manual       4 5 Grounpine  HFE     The LMU 209 provides outstanding EMC capabilities but still needs to be connected to an earth   ground  This connection should be made with a short wire with at least AWG22  If an earth ground  is not possible  attach this connection to 0 V or a mounting plate or case  Make sure that the HFE  is connected to a proper earth ground to sink the energy which is absorbed by the EMC circuit  integrated in the LMU 209    4 6 Power    Bridge power can be selected between 5 V DC or 10 V DC  As a general rule  low impedance sensors  should be driven with 5V whereas high impedance sensors can be operated with 10V  If in doubt   select the lower supply voltage  This helps to reduce sensor overheating     4 7 6 WIRE TECHNIQUE    Contact Magtrol SA for additional information    4 8 Low Pass FILTER    The onboard low pass filter is switched on and off by toggling the appropriate DIP Switch  It is  recommended to reduce the bandwidth  because it eliminates high frequency noise and gives more  stability to the readings  If an application needs more speed  switch off the low pass filter in order  to achieve a bandwidth up to 3000 Hz     5 0 MAINTENANCE    Any repairs should be performed exclusively by Magtrol     Ist Edition     January 2009       Testing  Measurement and Control of Torque Speed Power    Load Force Weig
4. MU 209     CALIBRATION    The LMU 209 contains a calibration button for quick and easy calibration  By pressing the  CALIBRATE button  the LMU 209 creates an internal signal of 1mV V to the input     Calibration procedure  1  Calculate the sensor signal in mV V for a 10V output span   Connect the sensor to the LMU 209   Adjust the output to 0 00 V with the ZERO potentiometer  see Figure 6   Press the CALIBRATE button  see Figure 6   Adjust the voltage output with the 10V potentiometer  see Figure 6      ad en sell Sad    according to the following formula     10V x 1  mV V   Sensor signal  mV V        U Volt       OUT      Example     A 5 kN force sensor should measure a force of 3 5 kN  whereas this sensor is specified with  2 00 mV V at 5KN  Therefore  the sensor   s output will be 1 40 mV V at 3 5 KN  According to the  formula above  the output needs to be adjusted to 7 143 volts when the calibration button is pressed   After this  the LMU 209 will deliver 10 00 V  or 20 mA  at the outputs when the sensor is loaded  with 3 5 KN     The voltage output can handle bipolar signals and will therefore also deliver  10 V  In the example  above  negative forces can also be measured as long as the sensor itself is designed to do so   CURRENT OUTPUT    The LMU 209 offers two simultaneous calibrated outputs  a voltage output and a current output   The relationship between these two outputs is strictly proportional  In other words  if the voltage  output is forced to 100   the current
5. ht    Tension    Displacement        MAGTROL INC MAGTROL SA    Subsidiaries in     70 Gardenville Parkway Route de Montena 77   Buffalo  New York 14224 USA 1728 Rossens   Fribourg  Switzerland Germany   France   Phone   1 716 668 5555 Phone   41  0 26 407 3000 China    India   Fax   1 716 668 8705 Fax   41  0 26 407 3001 Worldwide Network eaman   E mail  magtrol  magtrol com E mail  magtrol   magtrol ch of Sales Agents Certification          7    
6. nse     Figure 4 Connection of Resistive 6 wire Sensor  contact Magtrol SA for information     Auxiliary  Load       o  O Excitation   ZS sense   P Signal   w Signal   B   Sense   a Excitation     a    15V    Figure 5 Use of LMU 209 Auxiliary Power    3 2 DIP SwitcHes       OFF Position    DIP Switches    ON Position       Excitation 10V  N A    N A   N A    N A    N A    N A    N A    Cut Off Frequency 3 kHz  Zero Current Out 0 mA                                     d  m  SUA       CIE   Miom 2  aim       ifo m    100 m  Ainm  gt       CUI  gt   Gio p 7       Y   Cre Pa       Excitation 5 V       Sensitivity Range 1 ON       Sensitivity Range 2 ON       Offset Range  50    20         Offset Range  25    20         Offset Range  0    20         Offset Range  25    20         Offset Range  50    20         Cut off Frequency 500 Hz       Zero Current Out 4 mA                Magtrol Model LMU 209 Load Monitoring Unit User   s Manual       3 2 1    4 0    4 1    Default Position                                           SW 1 ON 5V Excitation   SW 2 ON Range 1 selected   SW 3 OFF   SW 4 OFF   SW5 OFF   SW 6 ON Offset adjusting range 0   20   SW7 OFF   SW 8 OFF   SW 9 ON Bandwidth is limited DC   500 Hz  SW 10 ON Current Out 4   20 mA       OPERATION    The LMU 209 is designed to amplify low level signals from force  pressure or strain sensors based  on strain gauge principles     After supplying the LMU 209 with power  it is immediately ready for measuring and will deliver  a meas
7. t cannot be controlled     3 0 INSTALLATION    For quick and easy installation  select the operating mode as shown in    default position    section 3 2 1       CAUTION     3 1 ConneEcTION DIAGRAMS    MAKE SURE THE SWITCHES ARE IN THEIR CORRECT END POSITION  IF  IN DOUBT  TOGGLE THE SWITCH AGAIN WITH THE APPROPRIATE TOOL   CONFIGURATION CAN BE DONE IN THE POWERED OPERATE MODE        Magtrol Model LMU 209 Load Monitoring Unit User   s Manual                                                             3 1 1 Supply Connection and Signal Output Connection   Ub 1 Power supply  18 28 V  OV 2 Power supply 0 V  Ground   Uout 3 Calibrated voltage output  lout 4 Calibrated current output  OV 5   Signal 0 V for Usur and lour   24VDC Q   OVPWR    U Out Ou  1Out O gt   OV Signal O wn  Figure 2 Connection to Power and Control Unit  3 1 2 Sensor Connection   15V 16   Auxiliary supply output for external sensors  draw max  20 mA   E  15   Positive bridge excitation  5 V 10 V   Sen  14   Sense input    6 wire technique   S  13   Positive signal input  S  12   Negative signal input  Sen  11   Sense input      6 wire technique   E  10   Negative bridge excitation 0 V  Earth 9 Connection for grounding                   o HFE  oQ Excitation   2O    P Signal     D Signal   rO  m Excitation     a  15V       Figure 3 Connection of Resistive 4 wire Sensor       User   s Manual    Magtrol Model LMU 209 Load Monitoring Unit          oo HFE    oQ Excitation     ZO Sense   P Signal     D Signal     EO Se
8. uring signal at the signal outputs which is in perfect proportion to the input signal  coming  from the connected sensor  Depending on the sensor   s offset  it may be necessary to adjust the  potentiometer ZERO     OFFSET CALIBRATION    The LMU 209 is optimized for static measurement  as in force measuring and weighing applications   If an exact proportional signal without any offset or zero errors is desired  the zero of the whole  measuring chain has to be adjusted after installation  If the zero error is within a few percent  this  error can be adjusted by using the zero potentiometer  However  if the zero offset is larger  the DIP   switches can be used to adjust the range  The DIP switches can also be used to adjust for preload  or packaging weights     ZERO potentiometer 10V potentiometer    Calibrate    Calibrate button  gt  Pop Owy x sexsors    7  01 2006  A ma       Figure 6 DIP switches and Zero Potentiometer       User   s Manual    Magtrol Model LMU 209 Load Monitoring Unit       4 2    ADJUSTING INPUT SENSITIVITY    The default input sensitivity of the LMU 209 is adjusted to ImV V in conjunction with a bridge  supply power of 5 V  To adapt the LMU 209 to other signals  other ranges are easily selectable with  the DIP switch or by adjusting the calibration with the potentiometer in large ranges     CAUTION  ADJUSTING THE CALIBRATION WILL RESULT IN A LOSS OF FACTORY  CALIBRATION  THEREFORE ONLY QUALIFIED PERSONNEL SHOULD    4 3    4 3 1    4 4    RECALIBRATE THE L
    