Home
        User Manual revision A6
         Contents
1.    Run Recovery   Power cycling   Recover last experiement   Accessing Log Files   Identifying Serial and Version Numbers    Appendix A Concepts and Glossary    Concepts  Glossary    Technical Assistance    Declaration of conformity       The copyright of this instruction book is the property of Bibby Scientific  This instruction book is supplied by Bibby Scientific on the express  understanding that it is to be used solely for the purpose for which it is supplied  It may not be copied  used or disclosed to others in whole  or part for any purpose except as authorised in writing by Bibby Scientific  Bibby Scientific reserves the right to alter  change or modify this    document without prior notification     Bibby Scientific Limited  Beacon Road    Stone    Staffordshire   ST15 OSA    United Kingdom   Tel   44 0 1785 812121   Fax   44 0 1785 810405  e mail  enquiries techne com  www techne com    PrimePro rev A 6    Page    28  29    29    30  30  30  30  Si  31  31    32    32  33    36    37       Polymerase Chain Reaction  PCR  denotes the amplification of DNA templates catalyzed by DNA polymerase in the  presence of primers  dNTPs  divalent cations  like Mg 2   and a buffer solution  The ability to visualize and quantify  the amplification of DNA as it occurs during PCR Is called Real Time PCR or Quantitative PCR  qPCR   This is made  possible by the use of fluorescent chemistries  an optical system that can capture the emitted fluorescence at every  PCR cycle  and softwa
2.   Assay 1  Channel 2   Assay 2  Channel 3   Assay 3  Channel 4   Assay 4       Log of starting quantity target    Figure 7 Standard Curves for Four Multiplexed Assays    The Prime Pro 48 includes two excitation LED arrays  452 486 nm and 542 582 nm  and four filter channels   Table 1   which enable detection of up to four separate targets in a single reaction  Figure 7      Prime Pro 48 is factory calibrated for certain dyes within each channel  marked in Table 1   but also supports  additional dyes that are excited and detected within the instrument specifications    IA PrimePro rev A 6       2 1  UNPACK THE PRIME PRO 48    1 Remove the computer and accessories from the box    2 Lift the Prime Pro 48 instrument out of the crate  Place it on a flat surface and remove the foam packaging   NOTE Keep the box and packaging in case of a return    3 Check to ensure that all components are present and intact     Your system comes with     Accompanying  literature                               2 2  PLACE PRIME PRO 48 ON THE BENCH    Prime Pro 48 requires 5 cm  2 inches  of unimpeded space at the  front and back for ventilation and 7 5 cm  3 inches  above the  instrument so that the lid can be opened safely  Make sure you  have easy access to the power switch on the lower right back  corner of the Prime Pro 48 instrument and that there are two wall  outlets  100 240 VAC  50 60 Hz  5A  within 2 m  6 feet  of the  instrument              Figure 8 Prime Pro 48 Space requirements    2 3  
3.   For Genotyping experiments  Select a reporter dye for Alleles 1 and 2      Select a Quencher  Quencher molecules absorb fluorescent emissions of reporter dyes when in  close proximity  By default  the quencher is set to None for DNA binding dye chemistry and Non   fluorescent for Hydrolysis probes     NOTE  BHO and MGB are considered non fluorescent quenchers   NOTE Fluorescent quenchers such as Tamra are not recommended for use in the Prime Pro 48   For Genotyping experiments  Select a quencher for Alleles 1 and 2     e For Relative Quantification experiments  Specify the PCR Efficiency         4 Click go close the Assays dialog box and return to Plate Layout       Airay Mama   PCR Efficiency  9      Actay      Atay          Figure 15 Assay Dialog Box  Relative Quantification Experiments    5 For Relative Quantification experiments  Select at least one Reference assay   6 For all experiment types  If you want to use controls  select a control type for each assay from the Role  drop down list  Options for roles in the drop down list change according to the type of experiment     NOTE The Role you assign has no affect on the analysis calculations of your experiment  The Role is just a label  for your convenience  For example  if you want to define an assay as a no reverse transcription control  or  you want to define an assay as a control sample that you know to be negative for the target you are  amplifying  you can select the Role    Negative    or    NTC     Data from the
4.  Prime Pro 48 system is set up and ready to use  register your Prime Pro by going to  www techne com warranty asp and completing a short questionnaire  Registering your Prime Pro ensures that    you will receive software updates in the future  While you are visiting the web site  take advantage of the  following online resources to support your research     Prime Pro Customer Support  knowledge database  warranty information  webinars  and seminar  series www techne com  Online Ordering www techne com    Tradeshows  workshops  and meeting presence www techne com or contact your local dealer    ow PrimePro rev A 6    3 1  PRIME PRO 48 WORKFLOW    1 Prepare the sample plate  load it into the Prime Pro 48  and close the lid  See Load the Plate on page 12   Z Launch the Prime Pro 48 software on the PC   3 Define and name the experiment by selecting the application  detection chemistry  starting material  and    specific method for your application  See Define a New Experiment on page 13     TIP To use a pre defined thermal profile and plate layout for your experiment  click  Templates and select one of the template experiments saved on your computer     4 Review the thermal profile and adapt it if needed  See Set Up the Thermal Profile on page 14     5 Set up the plate layout by defining assays  samples  and standards and assigning them to wells  See  Define the Plate Layout on page 15    6 Start the run  The Monitor Run tab opens  See Monitor Run on page 21    WARNING Do not ope
5.  negative control or NTC is not  used in calculations to normalize the data     Z For Genotyping and High Resolution Melt experiments  If you want controls with unique names that  are not included with the software  set them up from the Options menu  After controls are created  they  are available for use in the Role drop down list in the Assays section     a Click the Options menu   b Click Control Types   c Use the arrows to select the Number of Control Types   d Select the Color  To change the color  select the round color icon next to the assay name  Double   click a new color in the color palette   e An Abbreviation automatically populates   f Click OK  8 Proceed to set up samples    PrimePro rev A 6          3 10  SET UP SAMPLES    Samples  Nurnber of Samples    3    yr    Sample Name  Standard    Unknown    NTC     El Apply Sample Color       Figure 17 Sample Dialog Box    On the Plate Layout tab  click ps    or Set Up Samples to open the Samples dialog box   Use the arrow keys to select the number of samples    For each sample  define a name and color    Click CA o close the Samples dialog box and return to Plate Layout    For Relative Quantification and HRM experiments  Select at least one Reference sample   Proceed to assign assays and samples to wells     O 01 A W N S    ot PrimePro rev A 6       3 11  ASSIGN ASSAYS AND SAMPLES TO WELLS       Figure 18 Plate Layout Tab  Assigning Assays and Samples    1 Left click and drag the mouse to highlight one or more wells o
6. 36     5 2  RUN RECOVERY    If the Prime Pro 48 instrument loses connection with the computer  the Prime Pro 48 often continues to run   After you reconnect your computer with the Prime Pro 48  you can usually retrieve the run data because the  data file of your last experiment is stored in the instrument  To Recover the run data  follow the Power Cycling  directions below to turn the computer and the Prime Pro 48 off and on  After Power Cycling  select Options    Recover Last Experiment in the main menu     5 3  POWER CYCLING    1 Turn off the power on the Prime Pro 48  The power switch is on the back of the instrument   2 Shut down the computer so that the power is off   3 Wait three minutes     NOTE Make sure that the Prime Pro 48 and the computer are powered off for three minutes before you proceed    to step 4   4 Turn on the Prime Pro 48 instrument   5 Turn on the computer   6 Double click the Prime Pro 48 icon on the computer desktop to start the Prime Pro 48 software     Communication between the computer and the Prime Pro 48 instrument is established within five  minutes     NOTE If you are power cycling Prime Pro 48 because the Prime Pro 48 instrument and the computer lost  connection  a warning message might appear during the connection time that says the experiment file  may not have completed  The message asks you if you want the software to attempt to Recover the run   Click Yes     7 When Status  Instrument Connected displays at the bottom of the screen  the co
7. AM  542 582 604 644 KORC    452 486 562 5960 HEX  VIC   542 582 665 705 Cy5  0670       a  If you use ROX as a passive reference for normalization  your plate layout cannot include an assay whose reporter dye is measured in channel 2     PrimePro rev A 6 ER       3 8  ASSAYS AND REPORTER DYES FOR GENOTYPING    When defining a plate layout  genotyping experiments need special setup of assays and reporter dyes  Assign  at least one well for each homozygous Allele 1 Wild Type  homozygous Allele 2 Mutant  and tor heterozygous  controls  Make sure to select a different reporter dye for each Allele  See Genotyping and High Resolution Melt  on page       HER       Figure 15 Example Plate Layout for a Genotyping Experiment    3 9  SET UP ASSAYS    1 On the Plate Layout tab  click Q or Set Up Assays to open the Assays dialog box     2 Use the arrow keys to select the number of assays   3 For each assay      Define a name and color      For Genotyping experiments  Click the yellow box    C    which appears after setting role as Wild    Type  Heterozygous or Mutant  this will let you further define the control type name and will  automatically define an abbreviation     meme                    en  smn        e  tex   texa          Select a Reporter dye  thereby setting the channel   If your dye is not listed  select a reporter with the most similar excitation and emission range to  your dye  refer to the Channel table on Multiplexing Real Time PCR on page 8      ot PrimePro rev A 6     
8. CONNECT PRIME PRO 48    1 Connect one end of the Ethernet cable to the  Ethernet port on the computer  Connect the other  end to the Ethernet port on the rear panel of the Eco  48  A   Equipment must be connected to reliable and  suitable protective earth connection   2 Connect the Eco 48 power cord to the AC power inlet  on the rear panel  and then to the wall outlet  B     A suitably approved mains power cord set may be  used  It must be ensured that the cord set meets the  host country requirements    3 Connect the computer power cord to the wall outlet Figure 9 Prime Pro 48 Connection     C    PrimePro rev A 6 s          2 4  TURN ON THE PRIME PRO 48    1 Turn on the computer  wait up to five minutes for Microsoft Windows to boot fully  then turn on the  Prime Pro 48 instrument  A   The instrument runs a series of self tests that take up to 20 minutes     2 At any time after turning on the instrument  double click the Prime Pro 48 icon on the computer desktop  to start the Prime Pro 48 software  B   Communication between the computer and the Prime Pro 48  instrument is established within five minutes  When the READY indicator lights on the front panel stop  flashing and remain solid  the instrument is ready     3 Open the Prime Pro 48 by pressing the round silver button on the front to raise its handle  while lifting the  handle from the bottom until the Prime Pro 48 pops open  C         Figure 10 Prime Pro 48 Startup Sequence    2 5  REGISTER YOUR PRIME PRO 48  Once your
9. EE    READY        STATUS    2O OD     ERROR    Sees    READY    CT   STATUS    OOOOO  ERROR    Er z    PrimePro rev A 6       Description    Power on    Initializing   Conducting self tests and  heating the thermal block      Ready Idle    Run In Progress    Do not switch off or open the  lid while a run is in progress     Fatal Error  Run Terminated  Error light flashing  Instrument  may have overheated or  encountered a hardware failure     Camera Initialization Error   2 long flashes of the Error light   Try power cycling the instrument  to resolve     Temperature Sensor Error   4 long flashes of the Error light   Hardware problem  Call  customer support        Lights    READY        STATUS    60000  ERROR          READY    SS   STATUS    OOO 010  ERROR    SSS        READY    Ce  STATUS    22 OS TS eS   ERROR    CE     READY    ooo o o  STATUS    90000  ERROR    E     READY    CT   STATUS    OOOOO  ERROR        ETT       Description    Run Complete    Communication to PC error  Try  power cycling the instrument to  resolve     Software Updating    Non Fatal Error    Decide whether you want to  terminate the run     Hardware Failure Solid Error  light  Hardware problem  Call  customer support     Camera Trigger Error   3 long flashes of the Error  light  Try power cycling the  instrument to resolve     Temperature Response Error   5 long flashes of the Error light   Hardware problem  Call  customer support        4 2  MINIMUM SYSTEM REQUIREMENTS    Operating The following 
10. Operation Manual    PrimePro rev A 6       1    TT  LLL  11111  ALTT  AT TT  LLL  TTT  bobby                   INTENDED USE  The Prime Pro 48 Real Time PCR System is intended to support the Real Time polymerase chain  reaction  PCR  application needs of life science researchers  This includes gene expression quantification and  analysis as well as genotyping by allelic discrimination or high  resolution melting  The system is able to support  other applications and protocols as well  Prime Pro 48 features high quality optical and thermal modules to provide  optimal performance and data quality  The system includes data analysis software that is preloaded on a computer  and provided on a separate USB drive for installation on additional computers as needed  Additional accessories  and consumables are provided or available tor purchase to ensure the best user experience     Use of the Prime Pro for specific intended uses  such as polymerase chain reaction  PCR   Real Time qPCR  or high   resolution melting  HRM  may require the user to obtain rights trom third parties  It is solely the user s responsibility  to obtain all rights necessary for the intended use of Prime Pro 48     This document and its contents are proprietary to Prime and its affiliates  and are intended solely for the contractual  use of Its customer in connection with the use of the product s  described herein and for no other purpose  This  document and its contents shall not be used or distributed for any ot
11. afety requirements for electrical equipment for measurement  control  and laboratory use   General requirements   EN6101 2 010 2003 Safety requirements for electrical equipment for measurement  control  and laboratory use    Part 2 010  Particular requirements for laboratory equipment for the heating of materials    CE mark affixed 2014     Other Standards   CFR47 Part 15  2013 Radio Frequency devices  Subpart B  Class A   UL 61010 1 2012 Electrical Equipment For Measurement  Control  and Laboratory Use  Part 1  General  Requirements   UL 61010A 2 010 2002 Electrical Equipment for Laboratory Use  Part 2  Particular Requirements for Laboratory  Equipment for the Heating of Materials   CAN CSA C22 2 No  61010 1 2012 Safety requirements for electrical equipment for measurement  control  and  laboratory use   Part 1  General requirements   CAN CSA C22 2 No  61010 2 010 2004 Safety Requirements for Electrical Equipment for Measurement  Control   and Laboratory Use   Part 2 010  Particular Requirements for Laboratory Equipment for the Heating of Materials    o  gt   z de  Mr 5  Marriott     Signed         sli iv     Technical Director  Bibby Scientific Ltd    Bibby Scientific Ltd   Stone   Staffs   ST15 OSA   UK    Bibby Scientific    Tel   44  0  1785 812121   Fax  44  0  1785 813748       PrimePro rev A 6       Bibby Scientific    Bibby Scientific Ltd   Beacon Road Stone  Staffordshire ST15 OSA   United Kingdom   Tel   44  0 1785 812121   Fax   44  0 1785 813748   e mail  info bi
12. allelic discrimination  assays using hydrolysis probes provide a rapid and sensitive method to geno   type samples  These assays can refer to a single nucleotide polymorphism  SNP  or insertion deletion assays  Two  variants alleles are interrogated at the same time  multiplex qPCR   Most frequently  one probe is labeled with a  FAM dye and the other with a VIC dye  Samples with FAM signal and no VIC signal are homozygous for allele 1   samples with VIC signal and no FAM signal are homozygous for allele 2  and samples with both FAM and VIC  Signal are heterozygous  Figure 5         Figure 5 Allelic Distribution Scatter Plot Figure 6 HRM Difference Plot    High Resolution Melt  HRM  enables the detection of almost any genetic variation  SNPs  mutations   Because  HRM assays only require primers and a dye  no probes or DNA sequencing   the method is simpler and cheaper  than traditional genotyping approaches  After the amplification phase  the amplicon is slowly heated until it  melts  The melting temperature and profile are directly linked to the amplicon sequence     HRM s power comes from the resolution of the sample s melt profile  It requires a high  quality optical system  and precise thermal uniformity  HRM PCR amplicons below 300 bp provide the best resolution  The shape of the  resulting melting curves  which is sensitive to almost any genetic change  determines sample identity  To easily  cluster equivalent samples  a reference curve  e g  Wild Type  is subtracted fro
13. are pending  US patents  13 417121  12 753806  14 620998  European patents 7783905 8  12178744 4  10759513 4  10007637 1 and  India patent 9450DELNPO8  Further patents are pending     PrimePro rev A 6       Revision History  Table of Contents    Chapter 1 Overview   Real Time PCR   Absolute and Relative Quantification  Genotyping and High Resolution Melt  Multiplexing Real Time PCR    Chapter 2 Setup   Unpack the Prime Pro 48   Place Prime Pro 48 on the Bench  Connect Prime Pro 48   Turn on the Prime Pro 48  Register your Prime Pro 48    Chapter 3 Workflow   Prime Pro 48 Workflow   Load the Plate   Define a New Experiment   New session tab   Set Up the Thermal Profile   Define the Plate Layout   Assays and reporter dyes   Assays and reporter dyes for genotyping  Set up assays   Set up samples   Assign assays and samples to wells  Define standards   Auto calculate serial dilutions  Manually enter dilutions   Assign dilutions automatically  Assign dilutions manually   Monitor Run   Export Results and Data   Templates and Sample Sheets  Making a Sample Sheet for Import    Chapter 4 System Information   Components   System Requirements   Specifications and Environmental Requirements    Page    3    CO N DD UW UWI    O O LO MO WO WO    zi        11  12  is  14  14  15  15  16  16  18  19  20  20  20  20  21  21  22  23  24    25  25  27    i    PrimePro rev A 6 ER          Symbols  Electromagnetic Compatibility    Cleaning And Maintenance    Chapter 5 Troubleshooting  Introduction
14. are the minimum requirements for running the Pro 48 software    e Operating system must be Windows 7  not starter  and in UK or USA English  Either 32 or 64 bit variant are    compatible     e Memory  RAM   4GB DDR3    e Communication  Gigabit Ethernet and USB 2 0    e Processor  Core 13  3rd generation  with at least 1 7 Ghz clock speed  ideally 2 4GHz clock speed      Example PC  Dell Vostro 2420 or direct equivalent    The system should have all firewalls disabled as a precaution  Some firewalls disrupt communication between the  PC and the Prime Pro 48  The PC running the Prime Pro hardware should be a designated unit with no 3rd party  software on the system  Network ports being used by the Prime Pro 48 should be set to not allow the operating  system to shut down the port to save energy  Instructions for PC set up can be found on the Prime Pro 48 quick    Start guide that shipped with the unit     4 3  SPECIFICATIONS AND ENVIRONMENTAL REQUIREMENTS    Optical    Thermal    Operational    Physical    Environmental       Light Source    Detector    Thermal Cycling  Thermal Uniformity  Sample Format  Reaction Volume  Warmup Time   Typical PCR Run Time  Sensitivity of Detection  High Resolution Melt  Multiplexing   Passive Reference    Dimensions   Weight   Electrical  Temperature Range    Humidity Range    Protection    Two sets of 48 LEDs  452 486 nm and 542 582 nm   CCD camera  4 filters    505 545 nm  562  596 nm  604 644  nm  and 665 705 nm    Proprietary hollow silver 
15. ation of PCR  Xn  is described by the following equation   Xn   Xo    1 E  n  where X   number of target molecules at cycle n  X     initial number of target molecules  E     efficiency of target amplification  and n   number of cycles    Amplification efficiency  E   is described by the following equation  E   10      re      1   The acceptable range of assay efficiency   90  to 110   or a slope between  3 1 and  3 6   A slope of  3 32 indicates 100  efficiency  meaning that the number of template molecules doubled in each  PCR cycle     Common reference genes    o High expression  18S ribosomal RNA  18S   Beta actin  ACTB   Beta 2 microglobulin  B2M    glyceraldehyde 3 phosphate dehydrogenase  GAPDH   and phosphoglycerokinase  PGK       Medium expression  Transferrin receptor  TfR    o Low expression  Transcription factor IID TATA binding protein  TBP  and glucuronidase  GUS   Always validate your reference genes to ensure that the genes you select are stable in your experiments     El PrimePro rev A 6       6 2  GLOSSARY    Term    Absolute Quantification    Allelic Discrimination  Amplicon   Assay   Baseline   used to    Channel    Ci    Cq    Dark Quencher    Delta Rn  ARn     Derivate Melt Curve    DNA Binding Dye    dsDNA    Dual Labeled Hydrolysis Probe    Dynamic Range    Efficiency    Endogenous Control    End Point Analysis    Exogenous Control    FAM  6 carboxy fluorescein   Filter    Definition   An assay that quantifies unknown samples by interpolating their qua
16. ature control and thermal  uniformity across the sample plate  Standard Fast protocol performs 40 PCR  cycles in approximately 40 minutes    Motors   Stirring Paddles   Silver hollow Thermal Block  containing  the conductive fluid    4 1 2 OPTICAL SYSTEM    Two LED arrays provide individual sample  well excitation   Four detection filters support almost all  PCR chemistries and multiplex detection   ROX is optional    CCD camera acquires high quality data in  all wells and filters at each PCR cycle    Factory calibrated optics support SYBR Green   FAM  HEX  VIC  ROX  Cy5 and Q670 dyes  You  can also use other dyes that are compatible with  the excitation and emission wavelengths     U O U  gt     CCD Camera  Filter Slide  Green LED Array  Blue LED Array    Figure 28 Prime Pro 48 Thermal System                          Mi   gt  ib       SE  ni P mi       E       Figure 29 Prime Pro 48 Thermal System          PrimePro rev A 6 ER       4  System Information       4 1 3  LIGHTS    The Pro 48 has three indicator lights on the front  Ready  Status  and Error  The following tables show the  meaning of each combination of off  on  and flashing lights  If Power Cycling is advised for an error light  see  Power Cycling on page 30     Lights    READY    C  o  STATUS    OOOO     ERROR    SVT EEE Zz     STATUS    OOOO O  ERROR    CD     READY    eas   STATUS    OJOJOO  ERROR    CD       READY  a        STATUS    QT HU        Z S  OT NN    CD     READY    CT   STATUS    CHOC CC  ERROR    C
17. ays and samples to wells  See Assign Assays and Samples to Wells on page 19    Define standards  Standard Curve Quantification experiments only   See Define Standards on page 20   Select the Rox Normalization checkbox if you are using Rox passive reference dye to normalize across your  plate     Um B WN    You can lay out the plate any time between defining the experiment and analyzing the data  However  you will  only be able to see deconvoluted data while monitoring the run    3 7  ASSAYS AND REPORTER DYES    An assay is the set of primers or primers probe used to quantify a nucleic acid target sequence  Assays can have  different roles  such as Unknown  Standard  Negative  Positive  or NTC  Non Template Controls  Each assay is  associated with a reporter dye which identifies the assay during analysis  Reporter dyes can belong to one of  four    channels     each of which is defined by a specific excitation and emission range     You can assign up to four assays per well  Within each well  assays cannot use reporter dyes from the same  channel  see following table   If they did  data trom assays using the same channel would be indistinguishable  during analysis  A red outline around a well indicates that it contains more than one reporter dye from the same  channel and requires correction before you can analyze your data     Table 2 Channels and Reporter Dyes    Excitation  nm  Emission  nm  Fluorophores Calibrated on the Pro 48  Reporter     452 486 505 545 SYBR Green    F
18. bby scientific com  www  bibby scientific com    Bibby Scientific Middle East Ltd  BPO Box 27842   Engomi 2433   Nicosia Cyprus   Tel   357 22 660 423   Fax   357 22 660 424   e mail  sales bibby scientificme com  www bibby scientific com    Bibby Scientific France   Batiment le Deltaparc Silic Paris Nord 2   7 rue du Canal  BP 55437 Villepinte  95944 ROISSY Charles de Gaulle  France   Tel   33  0  1 48 63 78 00   Fax   33  0  1 48 63 78 01   Email  ventes bibby scientific fr  www bibby scientific com    Bibby Scientific US Ltd Bibby Scientific Singapore   3 Terri Lane Suite 10 Prudential Tower  Level 26  Burlington NJ 08016 30 Cecil Street  Singapore 049712  USA Tel    65 6631 2976   Tel  800 225 9243 Fax  44  0  1785 810405   Fax  609 589 2571 Email  info bibby scientific com  www  bibby scientific com www  bibby scientific com    
19. block with Peltier  based system    Q0 1  C   48 well plate   5 20 ul     20 minutes   Less than 40 minutes for 40 cycles   1 copy   Supported resolution to 0 1  C   Detection of up to four targets simultaneously  four plex   Optional  ROX    34 5 cm Wx 31cmDx32cmH    13 6 in  W x 12 2 in  D x 12 6 in  H    13 6 kg  30 Ib  including power supply   100 240 VAC  50 60 Hz  5A   Operating  15  C to 40  C  59   F to 104   F    Storage  10  C to 100  C  50   F to 212   F    Operating  15 90  Relative Humidity   Storage  5 95  Relative Humidity   IP20    PrimePro rev A 6    4  System Information       4 4  SYMBOLS    The Prime Pro 48 has three indicator lights on the front  Ready  Status  and Error  The following tables show the  meaning of each combination of off  on  and flashing lights  If Power Cycling is advised for an error light  see  Power Cycling on page 30     Symbol Description Symbol Description    Do Not Throw in Trash  At end  of useful life  recycle the system  or device    EC REP European Representative Fuse  replacement fuses must  meet the stated rating    CAUTION  Hot Surface          Humidity Range  on packaging  Manufactured By  indicates acceptable shipping    and storage limits        Mark of European Conformity  Model Number  device complies with the  EMC Directive  2004 108 EC     and the Low Voltage Directive   2006 95 EC     i   UH               Serial Number Temperature Range  on  packaging  indicates  acceptable shipping and  storage limits        a Prim
20. but  unassigned to a  layout    re    ProStudy   Before and after a  but not   during a run    See directions for    Making a Sample    Sheet for Import     Import Plate  Template    A plate template  is a plate layout   and does not  include a thermal  profile    de   ProStudy   Before and after a  run  but not  during a run   The Plate tab    must be active        Export Plate  Template    A plate template plate   is a plate layout  thermal  and does not   include a thermal plate  profile      CSV   ProStudy   Before and after a run  run  but not during    a run    PrimePro rev A 6 ER       3 18  MAKING A SAMPLE SHEET FOR IMPORT    To create your own sample sheet for import into Prime Pro 48   use a program like Excel to create a   csv file  The sample sheet m   Calibri    can contain up to 99 assay names and 48 sample names     F Format Painter B Z    1 In the same column  enter the heading    Assay Name    k Clipboard  and    Sample Name        2 Make sure there is a space between the assay list and 4    the sample list  Assay Name  3 Give each assay and sample a unigue name  Assay 1  4 Save the file as a   csv file  Assay 2    Assay 3 Make sure there          is a space  Sample Name between the  Sample 1 assay list and  Sample 2 sample list   Sample 3       Figure 27 Example of a Sample Sheet    a PrimePro rev A 6    4 1 1  COMPONENTS    THERMAL SYSTEM    w  gt     Proprietary hollow silver thermal block filled  with circulating conductive fluid provides  Superior temper
21. ds used to quantity nucleic acids by qPCR are the absolute and relative quantification  methods     The absolute quantification method is based on a standard curve generated from serial dilution of a nucleic acid  template of known concentration  Figure 3   Quantification of unknown samples is determined by interpolat   ing the sample Ca from the standard curve   Throughout the rest of this document  absolute quantification is  referred to as a standard curve experiment      Slope  Efficiency  R     Log of starting quantity target    Figure 3 Five Point  10 Fold  Standard Curve       The slope of the standard curve measures the efficiency of the assay  E   10 Ysepel    1   A slope outside the ac   ceptable range  slope  3 1 to  3 6 and E value between 90 and 110   typically indicates a problem with the  template or assay design  The R2 value  a measure of reaction performance  should be  gt  0 99 for the assay to  accurately quantify unknown samples     The relative quantification method measures the level of gene expression in a sample relative to the level of  expression of the same gene in a reference sample  In addition  the level of expression of every gene in the assay  is normalized to the expression of a reference gene     The results  RQ value  obtained are expressed as relative levels  or fold change  in gene expression compared to  the reference or control sample  Figure 4         Figure 4 Relative Quantification Experiment    KA PrimePro rev A 6       Genotyping  
22. e  format you chose to output  If a type of data is not available   it appears grayed out in the Export Options area  In the  Export Options dialog box  select the desired file format and  components to export  then click       2 PrimePro rev A 6    P    H    efiefiefiefielief    ie  elielielieli  ded dd dl    dddddd  ld ef ef ef  fed ef ef ed  s  Adelei  A Alee       Figure 25 Plate Layout View in the Monitor Tab    Export Options   mmm G    Exceriners Fie Name  Evaluation Plate  Exot Location  C  Program Piles  amp 6  PCRimax Eco  Browse    a CV _ Sv Qo PDF Report     Excel  Export Options    O Component Data E Plate Layout  O Component Meli Dala E Thermal Prolia       n       on    Figure 26 Export Options dialog box    3  Workflow       3 17  TEMPLATES AND SAMPLE SHEETS    The table below explains the differences between a template  a plate template  and a sample sheet  It also  describes where and when templates and sample sheets can be saved  imported and exported  The commands  below are available under the File menu in Prime Pro 48  Only Import Plate Template and Export Plate Template    are available in Prime Pro 48 Study     Save a  Template    Command    Description A template is a  layout and    profile       ecot    ProStudy    File Extension  Where    When Before a run    After saving  the  template is  available for   use  It Is listed in  the Templates tab  on startup in    Pro 48     Import a Sample  Sheet    A sample sheet is  a list of assays  and samples  
23. ePro rev A 6       4 5  ELECTROMAGNETIC COMPATIBILITY    This equipment complies with the emission and immunity requirements described in IEC 61326 1 2005 and IEC  61326 2 6 2005  To confirm proper operation        The electromagnetic environment should be evaluated prior to operation of the system      Do not use this system in close proximity to sources of strong electromagnetic radiation   e g  unshielded intentional RF sources   as these may interfere with proper operation      If you notice any interference  discontinue using the system until all issues are resolved  Resolution may    include moving cords from other equipment away from the system  plugging the system into an outlet on  a different circuit from other equipment  or moving the system away from the other equipment  If you  continue to have difficulties  contact Prime     4 6  CLEANING AND MAINTENANCE    Clean the block and housing as needed  following these directions     CAUTION If hazardous or biohazardous material is spilled onto or into the equipment  clean it immediately   1 Turn the system off and allow the block to cool completely   2 Using a lint free cloth slightly dampened with clean water  gently wipe the surfaces of the equipment  If a    stronger cleaning agent is needed  use a lint free cloth slightly dampened with 95  isopropyl alcohol     Follow these practices for proper maintenance of your Prime Pro 48       Every time you use the system  visually check it to confirm there is no obvious phy
24. elative Quantification    Reporter Dye    Rn  Normalized Reporter Signal     ROX  carboxy X rhodamine   Slope    Definition   dsDNA binding agents typically attached to the 3    end of hydrolysis probes   MGBS increase the Tm value of probes  thus leading to smaller probes    Hairpin probes containing a sequence specific loop region flanked by two  inverted repeats  A quencher dye at one end of the molecule quenches the  reported dye at the other end  Sequence specific binding leads to hairpin  unraveling and fluorescent signal generation    Simultaneous analysis of more than one template in the same reaction    An assay with all necessary components except the template    The use of control genes with a constant expression level to normalize the  expression of other genes in templates of variable concentration and quality    A fluorescence dye such as ROX that the software uses as an internal reference  to normalize the reporter signal during data analysis    Element used for heating and cooling in a qPCR machine    Molecule that absorbs fluorescence emission of a reporter dye when in close  proximity  BHQ is a quencher    The coefficient of correlation between measured Cq values and the DNA  concentrations  It is a measure of how closely the plotted data points fit the  standard curve  The closer to 1 the value  the better the fit  R2 is ideally  gt  0 99   A passive dye or active signal used to normalize experimental results    Genes with a wide and constant level of expre
25. face to ensure the seal is secure   8 Place the plate adapter with your loaded and sealed plate into a compatible centrifuge rotor along with    the seconds supplied plate adapter for balance  Centrifuge the plate at  250 g for 30 seconds  Do not spin more than 500 g  Verify that there are no air bubbles at the  bottom of the wells    9 Open the Prime Pro 48 lid and place the plate on the block  aligning the notch against the top  left  corner     WARNING Forcing the plate into any other orientation could damage the instrument     WARNING Be careful not to touch the heated lid above the plate  It heats to 105  C  221  F  when the  instrument is turned on and could result in burns     10 Close the Prime Pro 48 lid  The heated lid automatically creates a seal around and on top of the plate to  prevent evaporation   11 Proceed to Define a New Experiment     Ni PrimePro rev A 6       3 3  DEFINE A NEW EXPERIMENT    1 Double click the Prime Pro 48 icon on the desktop to open the software  The New Experiment tab opens                    Figure 11 Main Real Time PCR Chemistries    2 Click an Application Option that is the specific method or protocol you want to use for your experiment   When you select the application  the software automatically configures options for downstream setup  and analysis  For example  High Resolution Melt  HRM  is associated with DNA Binding dyes and so the  other three detection chemistries are grayed out for High Resolution Melt experiments     Experime
26. ght a Standard Assay well and click   enoti  the Assign button beside the appropriate a       dilution quantity  Figure 22   Unit   Dalina Staniarda     Assign Quantity    ig   i pe ml  h E  L ROR NOT m   ae  ML    3 18  MONITOR RUN    WARNING   Do not open the lid while a run is in progress   This allows extraneous light into the system and  will corrupt the data     NOTE   If you have not yet defined the plate layout   Define the Plate Layout on page 15   you  will only be able to view progress against the  Thermal Profile on this tab     Select assays to view in Amplification Plot  B Shows current stage of the Thermal  Profile highlighted in orange Figure 23 Monitor Tab  C Amplification View shows deconvoluted  data in real time for each well  D Amplification Plot shows deconvoluted  data in real time for selected wells       Amplification Plot   Highlight wells in the  Amplification View to  display a subset of  data in the main graph       Figure 24 Amplification Plot    PrimePro rev A 6          A Plate layout view shows sample type  sample identity   dilution and assays   B Select by assay  sample or call using the drop down arrow   Select the identity of the assay using the drop down arrow   D Turn Show Plate Layout on and off by selecting the box    OO    3 18  EXPORT RESULTS AND DATA    Results  as well as other data  can be exported by selecting  File   Export in the main menu  The options available here vary  depending on the type of experiment you ran and the fil
27. her purpose and or otherwise communicated   disclosed  or reproduced in any way whatsoever without the prior written consent of Prime  Prime does not convey  any license under its patent  trademark  copyright  or common law rights nor similar rights of any third parties by  this document     The instructions in this document must be strictly and explicitly followed by qualified and properly trained personnel  in order to ensure the proper and safe use of the product s  described herein  All of the contents of this document  must be fully read and understood prior to using such product s      FAILURE TO COMPLETELY READ AND EXPLICITLY FOLLOW ALL OF THE INSTRUCTIONS CONTAINED  HEREIN MAY RESULT IN DAMAGE TO THE PRODUCT S   INJURY TO PERSONS  INCLUDING TO USERS OR  OTHERS  AND DAMAGE TO OTHER PROPERTY     Prime DOES NOT ASSUME ANY LIABILITY ARISING OUT OF THE IMPROPER USE OF THE PRODUCT S   DESCRIBED HEREIN  INCLUDING PARTS THEREOF OR SOFTWARE  OR ANY USE OF SUCH PRODUCT S   OUTSIDE THE SCOPE OF THE EXPRESS WRITTEN LICENSES OR PERMISSIONS GRANTED BY Prime IN  CONNECTION WITH CUSTOMER S ACQUISITION OF SUCH PRODUCT S      FOR RESEARCH USE ONLY  PrimePro rev A 6     2010 2014 Bibby Scientific  All rights reserved        This instrument is covered by US patents US8003370  US8008046  US8232091  US9068947  US8987685   US8035811  Germany patents DE202010010523 4  DE2120100000393  China patents ZL201020267691 9   ZL201090000931 7 and US registered design USD653764  The following patents 
28. increase in  the electronic state of a photon  containing molecule   Carboxy 2  4 4  5  7 7  hexachlorofluorescein    An enhancement of the traditional melt curve analysis which increases the  detail and information captured    A probe that is not hydrolyzed by Tag polymerase    Hybridization probes can be used for melt curve analysis    Examples include Roche FRET and Molecular Beacons    A probe that is hydrolyzed by the 5  endonuclease activity of Tag polymerase   An exogenous control added to a multiplex gPCR assay to monitor the  presence of inhibitors in the template    Carboxy 4  5  dichloro 2  7   dimethoxyfluorescein    Light Emitting Diode  A light that is illuminated by the movement of electrons  in a semiconductor material  LED lights do not have filaments that burn out  and do not get very hot    A view of an amplification plot using linear dRn values  y  axis  versus PCR  cycles  x axis     A view of an amplification plot using log dRn values  y axis  versus PCR cycles   x axis     A self quenched fluorogenic primer and a corresponding unlabeled primer   When the primer is incorporated into DNA during PCR the fluorophore is de   quenched  leading to an increase in fluorescent signal     See Derivative Melt Curve        6 2  GLOSSARY    Term    Minor Groove Binders  MGBs     Molecular Beacons    Multiplexing  No Template Control  NTC   Normalization    Passive Reference    Peltier    Quencher    R2  Coefficient of Correlation     Reference    Reference Genes    R
29. m the other curves to generate a  difference plot  Figure 6   Captions added to illustrate the different genotypes only     Resources   Livak KJ  1999  Allelic discrimination using fluorogenic probes and the 5    nuclease assay  Genet Anal Biomol Eng  14  143 149   POLAND server  http  Awww biophys uni duesseldort de local POLAND poland html  Wojdacz TK  Dobrovic A   Hansen LL  2008  Methylation sensitive high resolution melting    Nature Protocols 3 12   1903 1908    PrimePro rev A 6       The simultaneous detection of multiple targets in a single reaction is called multiplexing  An advantage  of multiplexing is that it conserves sample  allowing more data to be obtained from the same amount of    material  Another advantage is that multiplexing permits the inclusion of an internal control reference assay for    normalization purposes  significantly increasing data precision     Channel Dye  Channel 1  A   505 545 nm    Channel 2  A   604 644 nm      SYBR Green   FAM              ROX   Texas Red  HEX   JOE  TET  VIC   Gy 5250676     Channel 3       562 596 nm             Channel 4       665 705 nm        Table 1 Examples of Prime Pro 48 Compatible Dyes    Validating a multiplex qPCR assay can be challenging  The advent of more advanced qPCR master mixes has  significantly reduced the amount of optimization typically required  making multiplex qPCR a much more  attractive alternative  Validation of assays using a standard curve Is a must to ensure data accuracy     Channel 1 
30. modify it based on your reagent s recommended protocol  You can set up cycle parameters in the  Thermal Profile at any time after defining the experiment  but only before starting the run     Click Q to add a new stage  such as a reverse transcription incubation at the beginning or additional PCR  Cycling stages  The stage will appear at the end of the cycle and you can drag it to the desired location   Alternatively  you can drag the Q icon to the location within the profile where you would like the new stage  to be added       The camera icon 9 indicates when the Pro 48 collects image data  In multi step PCR  you can select  whether to collect data at the annealing or extension step  Extension is the default  To move it to annealing   mouse over the annealing step and click the dim camera icon that appears  Only one step can be designated to  collect image data       To remove a stage  drag it to the 4 trash can or highlight it and press Delete    a PrimePro rev A 6       3 6  DEFINE THE PLATE LAYOUT    Annealing Annealing    intercalating dyes Hydrolysis probes   SYBR   Green   TaqMan          Figure 14 Plate Layout Tab    The Plate Layout tab lets you define how your samples  assays  and standards are laid out on the plate loaded in  the Prime Pro 48  The analysis software uses the plate layout to calculate data values  Plate layout involves the  following steps    1 Set up assays  See Set Up Assays on page 16    Set up samples  See Set Up Samples on page 18    Assign ass
31. mputer and the    instrument are connected   8 Wait until the Ready light on the Prime Pro 48 instrument is solid blue  This takes up to twenty minutes     PrimePro rev A 6       5 4  RECOVER LAST EXPERIMENT    After you follow the directions for Power Cycling  you can Recover the experiment that was running when the  connection was lost between the Prime Pro 48 Instrument and the computer     Select Options   Recover Last Experiment in the main menu    The name of your latest experiment with the word    Recovered    appears in the New Sessions tab at the  bottom of the screen    To confirm that your data is Recovered  go to the Monitor Run tab  If your data is Recovered  you will  see the data in the Amplification Plot and the Amplification View    Select File   Save As to save your Recovered data     5 5  ACCESSING LOG FILES    Prime Pro 48 stores a log file that helps with many troubleshooting issues  Prime Technical Support may ask for  this file  To access the log file  follow the procedure here     Pro    LO CON O UT    Turn on the Prime Pro 48 and the computer    Double click the Prime Pro 48 icon on the computer desktop to start the Prime Pro 48 software   Communication between the computer and the Prime Pro 48 instrument is established within five minutes   When Status  Instrument Connected displays at the bottom of the screen  the computer and the  instrument are connected    Close the Prime Pro 48 software on the computer    Navigate to C ProgramFiles Prime Prime P
32. n the lid while a run is in progress  This allows extraneous light into the system    and will corrupt the data     7 When the run is complete  open the Prime Pro 48 lid  Press the plate release lever and remove the plate  from the block  Dispose of any hazardous materials in biohazard  caustic material  or other appropriate  containers  according to your local safety regulations     Cer ba forra p p Dey er ae g Ma    nein jad ee ee a Ma maiia    Men zar pou Current eco Ma rar f Da pira  ip VA bas rol bam Mai    PrimePro rev A 6 ER       3 2  LOAD THE PLATE    1 Thaw all necessary reagents  templates  primers  probes  and master mix     2 Turn on the PC  then the Prime Pro 48  and wait until the Pro 48 Ready light is solid blue    o Confirm that the block and optical path are clear of visible contaminants and there is no physical  damage to the system  such as dents  frayed cords  or damaged levers    4 Place a 48 well plate into the Prime Pro 48 sample loading dock  aligning the notch with the matching  indentation on the adapter    5 Turn on the dock light and incline the dock to a comfortable angle for pipetting    6 Pipette samples and gPCR reagents into the plate according to your protocol    WARNING Wear protective gloves and eyewear when handling any material that might be considered    caustic or hazardous     7 Seal the plate with an Prime Pro 48 optical seal  Holding the plate in place on the Prime Pro 48 sample  loading dock  drag the squeegee firmly across the sur
33. n the plate layout diagram  Wells turn  yellow when they are highlighted  as shown in columns 1 and 2 of Figure 18     2 Click the Assign button for up to four assays and one sample in the left pane of the window to assign the  assays and sample to the highlighted wells     3 To change the role of an assay in a given well  highlight the well and then select the desired Assay Role  from the drop down list     NOTE For quantification experiments that will be combined using the Prime Pro 48 Study software  for at least  one plate in the study  you must specify     e Standard Curve studies  At least two wells with the role    Standard     but with different quantities  e Relative Quantification studies  At least one well with the role    Unknown    or    Positive    and a sample  assigned    Any plate meeting these specifications can be used as the mother plate in your study   The mother plate Is  the plate against which the other experiments in the study will be compared      4 For Standard Curve experiments  Proceed to define standards  For other experiments  Click   to start the  run now     ia nr  Sample Name and Colour                    gt     Assay Role and Colour             O   a de  20000    Quantity      gt     Multiplexed Well with 3 dyes           gt     To clear setting highlight and press     Delete          Figure 19 Well in the Plate Layout    PrimePro rev A 6 E       3 12  DEFINE STANDARDS    When you set an Assay Role to Standard  a small orange  Standards butt
34. nt Type Options   Quantification Relative Quantification or Standard Curve  Genotyping Genotyping PCR or Genotyping Single Read  High Resolution Melt PCR with HRM Curve or HRM Curve Only       Table 1 Examples of Prime Pro 48 Compatible Dyes    3 Select a Detection Chemistry     Detection Chemistry Type  NUPCR Probe based detection chemistry  DNA Binding Dye SYBR green assays    Hydrolysis Probe 5    nuclease assays    Other Non hydrolitic assays       4 Select a Starting Material   5 Enter an experiment name of up to 20 characters   6 Click z  The Setup window opens  with the Thermal Profile tab visible     PrimePro rev A 6          3 4  NEW SESSION TAB    Use the New Session tab to create  open  and select multiple experiments  The tab is on the bottom left of the  screen  To add a new experiment  select the page with a star icon  You can also right click on the New Session  tab to rename an experiment  make a new experiment  open an experiment  or close an experiment        Figure 12 New Session Tab    3 5  SET UP THE THERMAL PROFILE       A Drag to move the stage   B  Double click the temperature plateau to adjust temperature and duration  C Type new temperature   D Drag the bar up and down to adjust the temperature   E Data collection point   F Type new cycle time   G Toggle two  three  four  or five step PCR   H Click or type to add or remove cycles    When you define the experiment a corresponding default thermal profile is selected automatically  You can  use this or 
35. ntities from  a standard curve based on a serial dilution of a sample containing known  concentration    An assay that discriminates between two alleles  gene variants     A fragment of DNA synthesized by a pair of primers during PCR    The set of primers or primers probe used to quantify an amplicon    The initial PCR cycles when little fluorescence signal is generated  This will be  subtract the background    The combination of excitation and emission spectra used to monitor  amplification for a given assay    Threshold Cycle  See Ca    Quantification Cycle  The cycle number at which the fluorescent signal crosses  the threshold  It is inversely correlated to the logarithm of the initial copy  number    A quencher without any native fluorescence  Black Hole Quencher  BHQ  dyes  are an example    The normalized Fluorescence of an amplification plot with background and ROX  normalization dye correction    A plot of temperature  x axis  versus the derivate of fluorescence with respect to  temperature   dF dT   y axis   Used to analyze the Tm of an amplicon    A dye that increases its fluorescence in the presence of double stranded DNA   Double stranded DNA    See hydrolysis probe    The range of template concentration over which accurate Cq values can be  determined  Extrapolation is not recommended    See Slope    An RNA or DNA template that is spiked into each sample at a known  concentration    Qualitative analysis of PCR data at the end of PCR  Allelic discrimination assay
36. on appears to the right of the assay role        1 Click St open the Set Up Standards pane in the lower left of the  window    2 Select the units that are used in your standards  and then enter the  quantity     3 13  AUTO CALCULATE SERIAL DILUTIONS    1 To auto calculate serial dilutions  click Define Standards     Figure 20 Well in the Plate Layout    The Dilutions dialog box opens        2 Enter the number of points in the standard curve  the Nabea Porte Statina Quantic Oddion Fasie  quantity of the most concentrated standard  and the fe  600 10 Z    desired dilution factor  and then click  3 14 MANUALLY ENTER DILUTIONS    1 Enter the value of the first standard into the first Quantity field below Units   2 Press Enter to commit the value and open the next Quantity field     3 15 Assign Standard Dilutions to Wells  You can assign standard dilutions to wells manually or automatically   3 16 ASSIGN DILUTIONS AUTOMATICALLY    1 Left click and drag the mouse over a group of  Standard Assay wells       Vertical Wells     Points on Standard Curve    Horizontal Wells     Replicates    The Apply Standards button becomes active when you  have selected a suitable group of wells     2 Click WApphuStandardaj The dilutions and replicates are  automatically added in the highlighted group of wells     eee      mm er  ong           mn ser  ong            ne     ze      eee       Figure 21 Selecting Standard Assay Wells    ow PrimePro rev A 6       3 17  ASSIGN DILUTIONS MANUALLY    Highli
37. re that can quantify the amplification     The two most commonly used gPCR chemistries are DNA binding dyes and hydrolysis probes  Figure 1   DNA    binding dyes fluoresce when bound to double stranded DNA  Hydrolysis probes fluoresce when the reporter  molecule is removed trom its quencher molecule by the 5    exonuclease activity of DNA polymerase     Annealing Annealing    intercalating dyes Hydrolysis probes   SYBR   Green   TagMan          Figure 1 Main Real Time PCR Chemistries    Little fluorescence is generated during initial PCR cycles  Figure 2   Data from these early cycles define the baseline  for the assay  Initial   As fluorescence approaches the level of optical detection  the reaction reaches the exponential  phase  which is the region where the Cq is determined  Ca is the PCR cycle at which the fluorescent signal crosses  the detection threshold level and is used for quantification  Finally  as reaction components are consumed and  amplicons become abundant  the generation of additional fluorescent signal slows down and eventually reaches  a reaction plateau     4  Plateau    2  Exponential    Fluorescence    Threshold Line    ee eee eee eee    1  Initial                   Figure 2 The Three Phases of gPCR    Resources  Saiki RK  Scharf S  Faloona F  Mullis KB  Horn GT  Erlich HA  Arnheim N  1985  Science 230  1350 1354  Higuchi R  Fockler G  Dollinger G  and Watson R  1993  Biotechnology  N Y   11  1026 1030    PrimePro rev A 6          The two primary metho
38. ro and find the GetLogs exe file  If you did not install the Prime    program on the C drive  your drive letter may be different    Open the GetLogs exe Tile    Select the Include Previously Uploaded Logs check box    Click Get Log Files  When the file finishes transferring  Done appears on the GetLogs dialog box   Save the zipped log file on your computer in an easy to remember place    Email the zipped log file to Prime Technical Support at technehelpObibby scientific com    5 6  IDENTIFYING SERIAL AND VERSION NUMBERS    To identity the Prime Pro 48 instrument ID  serial  number  the software version number  and the instrument  firmware version number  select Help   About Prime Pro 48 License Information on the main menu     PrimePro rev A 6 ER       6 1  CONCEPTS    The weight of one genome  g     size of genome in bp  x  618 g mol bp  x Avogadro s number  One human genome  g     3 x 10   bp  x  618 g mol bp  x  6 02 x 1023    3 08 x 101  g mol  One haploid cell  sperm egg    3 08 pg of DNA One diploid cell   6 16 pg of DNA    There is approximately one copy of every non repeated sequence per 3 08 pg of human DNA   The average cell contains 10 20 pg of total RNA   About 90 95  of total RNA is rRNA  18S  5 8S and 28S   1 3  is mRNA     RNA concentration  ug ul     A     DNA concentration  ug ul     A      40   D  1000  where D   dilution factor and A    absorbance at 260 nm   x60 20   D  1000  where D   dilution factor and A    absorbance at 260 nm  The exponential amplific
39. s   genotyping  are an example    An RNA or DNA template that is spiked into each sample at a known  concentration    The most commonly used reporter dye at the 5    end of a hydrolysis probe   Components used to limit the bandwidth or the excitation or emission energy to    the next component of the optical path     PrimePro rev A 6 ER          6 2  GLOSSARY    Term    Fluorophore    Fluorescence    HEX  High Resolution Melt  HRM     Hybridization Probe    Fluorophore    Fluorescence    HEX  High Resolution Melt  HRM     Hybridization Probe    Hydrolysis Probe    Internal Positive Control  IPC     JOE    LED    Linear View    Log view    LUX Primer Set    Melt Curve    PrimePro rev A 6    Definition   The functional group of a molecule that absorbs energy at a specific  wavelength and emits it back at a different wavelength    The immediate release of energy  a photon of light  as a result of an increase in  the electronic state of a photon  containing molecule   Carboxy 2  4 4  5  7 7  hexachlorofluorescein    An enhancement of the traditional melt curve analysis which increases the  detail and information captured    A probe that is not hydrolyzed by Tag polymerase    Hybridization probes can be used for melt curve analysis    Examples include Roche FRET and Molecular Beacons   The functional group of a molecule that absorbs energy at a specific  wavelength and emits it back at a different wavelength    The immediate release of energy  a photon of light  as a result of an 
40. sical damage such as  dents  frayed cords  or damaged levers  If you see any damage  discontinue use and contact Techne  Technical Support  technehelp bibby scientific com         Once a year  run a known test sample to confirm accurate analysis    CAUTION The Prime Pro 48 contains materials that may be hazardous to the environment if not disposed of  properly  Be sure to dispose of materials according to all local  state provincial  and national  regulations     PrimePro rev A 6 ER          5 1  INTRODUCTION    For most errors on the Prime Pro 48  an on screen message opens with instructions for correcting the error  Error  lights on the front of the instrument may also indicate a problem and how to fix it  See Lights on page 47     If connection is lost between the Prime Pro 48 instrument and the computer  use the Run Recovery section to  Recover your run data  See Run Recovery on page 30     This Troubleshooting section also has directions on how to access files and information that will help   Prime Technical Support troubleshoot your problem  For instance  if you contact a Prime Technical Support  representative  they might ask for copies of run  specific files for troubleshooting purposes  such as a log file   For technical questions  visit the Prime Pro 48 support pages on the Prime website for access to frequently asked  questions     For problems with run quality or performance  contact Prime Technical Support  For more information  see  Technical Assistance on page 
41. ssion  Typically used to normalize  the expression of other genes  Examples of commonly used reference genes   16S 18S  GAPDH  and b actin    An assay used to measure the expression of a target gene in one sample relative  to another sample and normalized to a reference gene    Fluorescent dye used to monitor amplicon accumulation  This can be a dsDNA  binding dye or a dye attached to a probe  Each dye is associated with a certain  channel    Reporter fluorescent signal divided by fluorescence of the passive reference dye   The most commonly used passive reference dye    The slope of a standard curve  It is a measure of assay efficiency    E   106159 1  where a slope of  3 32 is equal to 100  efficiency  E  or an  exact doubling of template molecules in each PCR cycle  Acceptable efficiencies  range from  3 6  90   to  3 1  110    Overly high efficiencies indicate qPCR  inhibition  usually due to contaminants in the sample  Overly low efficiencies    typically indicate problems with the reaction mix concentration     PrimePro rev A 6 ER       6 2  GLOSSARY    Term    Standard    Standard Curve    Standard Deviation  SD     TAMRA  Target    Template    Threshold    TET  Tm  Unknown    Y Intercept    6 3  TECHNICAL ASSISTANCE    Definition   A serial dilution of a target of known concentration used as template to  generate a standard curve    A plot of Cq values against the log of target amount  Used to determine an  assay s dynamic range  efficiency  slope   R2  and sensitivit
42. y  y intercept    The SD of replicate Cg measurements is a measure of the precision of the  assay    Tetramethyl 6 carboxyrhodamine  Commonly used as a quencher    The DNA or RNA sequence to be amplified    See Target  Template can also refer to a saved experiment that can be used as a  model for new experiments in the software    A level set above the background signal generated during the early cycles of  qPCR  When adjusted manually  it should be set in the middle of the  exponential stage of qPCR    Carboxy 2  4 7 7  tetrachlorofluorescein    The temperature at which 50  of dsDNA is single stranded  melted     A sample containing an unknown amount of template    In a standard curve  the value that crosses the y axis at x   1  single copy    target      For technical assistance  go to www techne com or contact your local dealer     MSDSs    Material safety data sheets  MSDSs  are available on the Prime website at www techne com     Product Documentation    Product documentation  you can obtain PDFs from the Prime website  www techne com    NA PrimePro rev A 6    Declaration of Conformity    Real time Thermal Cycler Models  PrimePro 48    This product complies with the requirements of the EU Directives listed below     2004 108 EC EMC Directive   2006 95 EC Low Voltage Directive  L V D     Compliance with the requirements of these Directives is claimed by meeting the following standards   EN61326 1  2013 Emissions Class A and Immunity  Basic Requirements    EN61010 1 2010 S
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Open Access - Lund University Publications  Origin Storage DELL-1000SH/5-NB62 hard disk drive  TTFSC01A  Memorex MC2862 User's Manual    Copyright © All rights reserved. 
   Failed to retrieve file