Home
        Model DM8 Vibration Type Density Meter CONTENTS
         Contents
1.      7     Sets density of calibration solution 2    Density data  g cm3  for the 2nd calibration  calibration solution 2  of two point calibration  can be input  As a rule  the input density is data at the same reference temperature as that  of    4        Note   If relation between temperature and density is well known  a density at a temperature can be input  However  the density data at  the reference temperature which is same as that of    4    is recommended to achieve accurate calibration     Function No     8     Sets temperature of calibration solution 2    IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 4  OPERATION gt  4 5    Enter the temperature of calibration solution 2 of which density at the reference temperature  is set at    7      Function No     9     Sets temperature coefficient of calibration solution 2    Enter the value which is  100000 times the temperature coefficient  g cm3   C  of the  calibration solution 2     Function No     A     Starts calibration 2  press SET key   Pressing    SET    key executes 2nd calibration computation of two point calibration      3 Maintenance mode  MAINT   Mode for adjusting and checking operating  conditions     During this mode  analog and digital outputs are held  at the value just before holding    Function No  in maintenance mode is from    1     to    D      see table 4 1    Function No     1      Cancellation of protection  enter 77     Function to avoid data changing due to carelessness  When    0    is d
2.      im  a    Analog output signal                               Ld     NE ES  Digital output signal   A Contact output for abnormal status  Figure 3 20 Cable Inlet of Density Converter    Figure 3 21 shows terminal arrangement of the density detector     Terminals Function    Output Frequency    A2  Temperature Sensor    Power Supply for Pre amp  Ap   Impressed Voltage for Pick up Electrode  Earth Terminal    Figure 3 21 Terminal Arrangement of Density Detector       IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 13       3 4 1 Wiring between Detector and Converter    This wiring is to be carried out with special cable of specified length  The cable is terminated as  illustrated in Figure 3 22  Pay attention not to stain or wet terminated         _ Density detector Density converter        gt        Figure 3 22 DM8W Cable     lt  Connection to Density Detector  gt   Remove the cover of the terminal box  using the spanner in accessories  and insert the special  cable into the cable inlet  Connect each conductor to the respective terminals     A flexible fitting is used at the inlet of the density detector  1 and the cable from the detector runs  though metal conduit to the duct  For Model VD6DF explosion proof detector  wiring should be  made complying with the Recommended Practice    The explosionproof flexible fitting for wiring should be used at the cable inlet  G3 4 female  of the detector    1  The flexible fitting is not in
3.    Flow rate shown by FM in the VD6SM sampling unit of Yokogawa varies according to  viscosity of sample  If the viscosity of sample liquid becomes higher  flow rate indication also  becomes higher than real flow rate  see Figure 4 4      Specific gravity of liquid  p 0 85                                                                                                                                                                                25 cP  20 cP   40cP 30cP 10 cP   50 cP   35 cP 15 cP   60 cP 5 cP   1cP  20     E    15  o  D  E     2 10  5  Cc  2  w  2 5  D      0 5 10 15 20  Actual flow rate  L min   Figure 4 4 Flow Rate Characteristics for Sample Liquid Viscosity     8  Confirm that the sample liquid temperature T and pressure P are in the specified range    9  Visually check that the sample liquid is not leaking from the connections     4 3 4 Turning ON the Power    Supply the converter and a receiving unit  recorder  computer  etc   with power with specified  voltage and frequency  The receiving unit should be ready to receive the data and controller  should be adjusted to perform accurate control     4 3 5 Data Input    Set the mode transfer switch to    MAINT     maintenance mode  and input data to the function No      6   7   8      9     A    and    B      For details of data to be input  refer to Section 4 2  Data is input with     SHIFT        NCR A    and    SET    keys  For details of these keys operation  refer to Figure 4 1 2     When temperature coe
4.    Table 4 2 Error Table       Display Contents Error Release and Remedy       E001 Abnormal detector density signal When cause is removed       E002   Abnormal detector temperature sinal       E003 Abnormal on converter When cause is removed       E004 Out of measuring temperature                            E005   Abnormal first calibration Check and reset the data required for   E006   Abnormal second calibration calibration    E007 lllegal data of EEPROM Check and reset the loaded constant  NOTE    E001  002  003  004 and 007 are cancelled temporally by pressing the FUNCTION No  key  in maintenance or calibration mode  however in the measuring mode  the error indication is  displayed again  These errors are detected only in the measuring mode     E005 and E006 are errors that occur only in the calibration mode  You can clear these errors  by pressing the  SET  key  After that  the density at the reference temperature by the previous  calibrating data is displayed     Check of Measured Density Value    It is recommended that measured density values are checked at intervals of several weeks to  determine the period for calibration and cleaning      1  Measure the density using a hydrometer to compare it with the density indication  at the  reference temperature  on the vibration type density meter  When sampling the liquid  check  the pressure and the temperature are not largely different from those of the normal state      2  Measure density correctly at the reference tempe
5.    vi Yv F  Strainer body  316 SS  STEAM IN X X Element  316 SS  or Ni  option   b          KN   gt o    PI  Pressure gauge  0 to 1MPa or 0 to 2MPa  316 SS  TI  Thermometer  O to 100   C or 0 to 150   C  316 SS      Bi FM  Flowmeter  tapered metal tube flowmeter  1 to 10 L min  316 SS  CONDENSATE OUT Y BV1  Ball valve  316 SS    NV1 to 4  Needle valve  316 SS  DD  Density detector   V1  V2  Valve   ST1  Steam trap    a       Figure 5 1 Cleaning the Figure 5 2 Flow Diagram    Vibrator   5  Remove the blind plugs again from the pouring ports of the standard solution and pour the    cleaning liquid about 200 mL  After a while  fix the blind plugs and purge the sample liquid  by blowing 0 05 MPa G air  After the liquid is blown out  increase the air pressure up to 0 2 to  0 3 MPa G blowing air for 2 or 3 minutes to dry the inside of the vibrator     Exchange of Desiccant in the Detector    Desiccant is used to protect the oscillation amplifier in the detector from humidity     Checks the detector terminal box one or two times a year and replace the desiccant if humidity is  detected  Two desiccants are provided as accessories  For the explosion proof detector  confirm  that there are no explosive gases using a gas detector before removing the terminal box cover     IM12T03A01 02E 4th Edition   Feb  23  2015 00    9 2    5 2 1    5 2 2     lt 5  MAINTENANCE gt  5 2    Troubleshooting    Experiences are required for repair of the vibration type density meter  after parts are replac
6.  1 MPa      2  The pipe length between the sample inlet and the SAMPLE IN should not exceed 10 m  Use  stainless steel pipe of nominal diameter 15 mm or Sch 40 to 80      3  Install stop valves and drain valves near the sampling unit     Main process pipe line              Pin  Pressure at inlet  Pout  Pressure at outlet     2 MPa  gt  Pin    Sample conduits  1  gt   Pout   0 1 MPa      10 m max      Tap valve A Stop valve    Ka  Drain valve  Sampling unit    for vibration type density meter        Drain valve    Figure 3 14 Sample Liquid Conduits    IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 10        4  Avoid sharp bending of the pipe to prevent accumulation of sludge     Sampling unit    4    Sampling ee    Figure 3 15 Example of Wrong Piping    Main pipe         Sludge is likely to collect here     3 3 2 Air Piping for Maintenance    When cleaning the vibrator or calibrating the meter with standard solution  blow away liquid from  the vibrator by air pressure  Air should be clean and dried with 0 3 to 0 5 MPa G     Mount a    stop    valve and a    pressure regulator    on the pipe from the air source  and fix a flexible  tube with a copper tube with outside diameter 10 mm should be provided for connecting with the  detector     Connect the air pipe to the sampling unit only for cleaning the vibrator or calibration with the  standard solution     Sampling unit       D Connect when cleaning the density detector    Air supp
7.  5 2 3    Terminal  Box       Checking the Oscillation Amplifier    First  visually inspect the capacitance pickup electrode  This electrode should be installed near  the vibrator with a small gap between them  Without this gap  the electrode will not function  properly  If there is moisture or dirt between the electrode and vibrator  clean the gap with a cloth     Take care not to damage the vibrator     Then  check the electrical circuitry using a DC voltmeter and other appropriate measuring    instruments  For the types of instruments and the checkpoints  see Figure 5 3     Black A5       B4 and B5 when    density detector is  i operating normally      Ur     1   I   i ZS  1mS   I   Do   I   I   I   LR Red    Piezoelectric  1 element    TB Black    Check point   Measuring instrument Measuring conditions Normal value Remedial measure  for abnormal value      A5 B5 DC voltmeter 12 V 10   Check converter INPUT       For measurement of    the meter input impedance    200 V 10   assembly        should be at least 10 MO 5 8 to 6 4 V Replace D8        O IN  E Multimeter co  X 10 KO Range   Check pickup electrode        O B4 B5  Oscilloscope Disconnect leadwires from terminals   6 to 8 V p p Check for defective circuit    IN  R  and B  on PCB  connect an 150 to 200 V p p   components and replace    oscillator  0 to 2000 Hz  output 5 V     200 mV p p them   O B5 and apply an input of 100 mV rms   4 to 6 Vp p     O B5  DC voltmeter Lower 100 mV rms input  3to0V     Note   Regarding 
8.  Carry out either or both of wirings as necessary     Shielded 2 conductor  for one signal  or 4 conductor  for two signals  cable are used for wiring   For connecting the cable to the converter  remove the sheath of conductor about 50 mm from its  end  and attach a solderless lug for M4 screw omit  The shield of the cable should be grounded at  the receiving unit     Wiring method  metal conduit and others  is same as that of Section 3 4 1     3 4 3 Wiring for Digital Output Signal    The Model DM8C density converter outputs digital signals through RS 232C interface besides  analog signal  for details of digital signal  refer to    Standard Specifications    in Chapter 1      When using this signal  use a shielded 3 conductor cable for wiring  Cable length from the  converter to a receiving unit should be 10 m or less     Cable termination is same as that of Section 3 4 2  however  remove the sheath of conductor 80  mm from the end     Wiring method is also same as that of Section 3 4 1     3 4 4 Wiring for Abnormal Status Contact Output    When an error  See Table 4 2  occurs with an abnormality in the density measuring system  a  contact signal is output     For wiring of this contact output  use 2 conductor cable  When connecting the cable to the  converter  remove the sheath of conductor 80 mm from its end and attach a solderless terminal  for M4 screw     Wiring method is same as that of Section 3 4 1     IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLAT
9.  External Dimensions  1 3 1 Detector    General Purpose and Flameproof Detector Models VD6D and VD6DF  Unit  mm       Steam Connection Rc1 4  1  Sample Inlet    Rc1 4  1            Steam Outlet Rc1 4  1        2 inch Pipe         342       200    For VD6DF O0 B FM   1  1 4NPT   2  3 ANPSM    IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 1  GENERAL    1 7       Sanitary Use Detector Model VD6DS    Unit  mm       Gland 166    Sample Inlet  1    Sample Output  1    a 2 inch Pipe  2 EE   Horizontal or Vertical   E     orizonta     IR hb     O       0       Lo E L    3  Electrical Wiring Port  G3 4  LI J  8160  s 146  ge    1  The ends of sample inlet and outlet are connected with 1 8 inch      pipe in welding  The pipe may be removed by loosing the gland     IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 1  GENERAL    1 8       1 3 2 Converter Model DM8C    Unit  mm    183    e    M    138  240 116                                           CT    a 8 alle dr is                                                                                                             14  Y  Y   T I  Wiring port holes  5 027  200 Mounting bracket  applicable with any mounting   with rubber plugs  34 34 34 34 method to panels  walls or pipes    gt  Mounting panel thickness       Max  12                   O    Ps  y  Le A Pa                               lso  SAA   e    D ir  23 31  35       Purge air inlet Rc1 8  1               Pipe mounting bracket       Purge air outlet   with seal s
10.  General Specifications  Measurement object  Liquid density  Measurement principle  Vibration density measurement  Measurement range     Density  0 5 to 2 0 g cm3  Temperature   10 to 100  C  Distance between Detector and Converter  Up to 2 km    Power supply  90 to 132 V AC or 180 to 264 V AC  50 60 Hz  Power consumption  20 VA    m Characteristics    1 1 2     overall characteristics after combing the detector and the converter   Repeatability  5 x10 4 g cm  for digital output   1   of span  for analog output   Linearity   0 5   of span  when span is 0 2 g cm or less    1   of span  when span is more than 0 2 g cm3   Temperature characteristics   0 5   of span  10  C  Compensating error for changes in the  measuring liquid temperature and detector temperature   Flow characteristics     0 1  of span in the 0 to 5 L min range  Pressure characteristics   0 0005 g cm3  98 kPa change  Viscosity error   0 1  of span in the 0 to 1500 cP range    Detector     1 General Purpose Detector Model VD6D    Detector construction  Non explosion protection  rain proof construction  Case material  Cast Aluminum alloy   Case coating  Epoxy resin  baked finish   Case color  Jade green  equivalent to Munsell 7 5BG4 1 5     IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 1  GENERAL gt     Wetted part materials    Base  316 SS   Vibrator  316 SS or Ni  Au Brazing  BAu 4   Measuring liquid temperature   10 to 100  C  Measuring liquid pressure  2 MPa G or less  Maximum pressure  4 9 MPa G  Ste
11.  PIPING AND WIRING gt  3 5        c  Release the lock screw for the vibrator turning the screw counterclockwise completely  with a flat chip screw driver  Turn it enough  if not  the vibrator may touch the screw during  measurement causing an error  In addition  pay attention not to bend or damage the vibrator  because it is made of a thin pipe     CAUTION    The vibrator tube has extremely thin wall  hence must be carefully handled     Since assembling the vibrator tube requires special skill  the exchange of the vibrator should be  made by Yokogawa      d  Pull out the rubber sheet avoiding applying force to the vibrator  refer to Figure 3 7      Lock screw  Use flat blade screw driver  to turn this screw fully counterclockwise         Pull rubber sheet  in this direction    Rubber sheet    Pull out this rubber sheet  after loosening the set screw     Figure 3 7 How to Remove the Rubber Sheet     e  Remount the cover removed in the item  b   Also remount the cover in the item  a  to the  case and lock it     Now the removing the rubber sheet finishes  Keep the rubber sheet which is necessary for  transportation of the density converter     IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 6    3 2 Installation of Density Converter    3 2 1 Installation Area    The density converter should be installed in the place described below        The vicinity of the sampling unit where maintenance or calibration is made easy  The density  
12.  cover 1 pe   O Ring 1 bag  Silica gel 2 packs  Fuse for the converter  3A  1 pe  for Converter  DM8C                    IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 1  GENERAL gt        1 2  1 2 1    Suffix Code    Option Code    Models and Suffix Codes    General Purpose Detector    Description       General Purpose Liquid Density Detector        S3   N1    Vibrator material       316 SS  Ni              Suffix Code             Style B          Option Code    Description                Flameproof Liquid Density Detector    316 SS  Ni                    Suffix Code    Style B    NEC Class   Division 1 Group C and D   explosion proof                Option Code    Description       Sanitary Use Liquid Density Detector          316 SS                Suffix Code          Style B             Option Code    Description            Vibration Type Liquid Density Converter           A1   A2       Power supply                90 to 132 VAC  50 60Hz  180 to 264 V AC  50 60Hz              Option   Air purge connector                 Style D              1 2 5 Special Cable    Suffix Code          IAP1  IAP2       Rc1 4  1 4NPT female          Option Code    Description       Special Cable for Liquid Density Meter       Cable length    LOD       Length  unit  m            Note   Enter the cable length in     LOOOO in m       Example  L0050 for 50 m  L0100 for 100 m  L2000 for 2 km             StyleA       IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 1  GENERAL    1 6    1 3
13.  iii  1  GENERAL oia e 1 1  1 1 Standard SpecificationS         esse ese Roe Rae RE AREA ER Re ER Re ER Rae AE Rae AREA ER Re ran 1 1   1 1 1 General Specifications               ccecccceecceeeeeeeeeeeeeeeeeeeeeseaeeeeeeseneeesaneesens 1 1   1 1 2 BE EE 1 1   1 1 3 Converter Model DM 1 2   1 1 4 Special Cable Model DM  1 4   1 1 5 lee 1 4   1 2 Models and Suffix Codes              cccscccceeeeeseeeeseeeeseeeesaeeeeeeeeeseeeeseeeeseeeeseneeeeeeeeses 1 5   1 2 1 General Purpose DeteCtOF         iese eke ke Re RA Re Re ee ke ee 1 5   1 2 2 Flameproof Detector sies EE EE EE ica Ee gee 1 5   1 2 3 sanitary Use Detector    ae ei se ese anie ES Re ese ES Ee ee ke se ese KS Ed ge SE 1 5   1 2 4 Converter A 1 5   1 2 5 Special Cable  rada 1 5   1 3 External Dimensions sitio SEENEN ae 1 6   1 3 1 BEL GR EE N EE EE ias 1 6   1 3 2 Converter Model DM 1 8   1 3 3 Special Cable Model DM  1 9   1 4 Cautions in Handling Flameproof  Explosion proof Instruments               1 10   1 4 1 Outline of Explosion proof Instruments                        enn 1 10   1 4 2 Labeling the Explosion proof Specifications                                     1 10   1 4 3 Installation Area and Ambient Conditions                     ee ee ke 1 11   1 4 4 External Wiring Work        esse ese see ee ee ee ee ee ee ee ee ee 1 11   1 4 5 Maintenance Procedure            iese ee RA Re ek Re RA eke ee 1 11   1 4 6 Ee EE 1 11   2  PRINCIPLES OF OPERATION                  22202444400000000000nnn0nnn ann nn ek Re Ge RR
14.  kPa G  Size  26 x 84 mm copper or stainless pipe should be used to connect air source to the purge air inlet of  the converter     The air inlet is Rc1 8 female connection  If specified  a connector for an Rc1 4 female or 1 4NPT  female screw is provided     IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 12       3 4    Wiring  The vibration type density meter requires the following wirings   1  Wiring between analog output signal  Wiring for analog output signal  Wiring for digital output signal  10 m or less length is recommended    Wiring for contact output for abnormal status  Wiring for power supplying  6  Wiring for grounding  Figure 3 19 is a diagram of these wirings     Detector Special cable Converter    Closed when a  Alarm failure is detected or  the power fails        Density signal after being  converted to the reference  RS 232C signal   temperature state     Density signal before being  converted to the reference  temperature state  e Temperature signal    0to1VDC  Density signal after being  converted to the reference  4 to 20 mA DC temperature state       Power supply  Figure 3 19 Wiring Diagram    The cable inlet of the density detector is located at the bottom of the terminal box  The density  converter has five cable inlets and their size is 27 mm  A cable can pass any inlet of them  one  cable to one inlet   Figure 3 20 shows an example of allocated inlets     Inlet of special cable    Power supply  ga SC  i
15.  nun 2 1  2 1 Measuring Principles  iisises sin Ke EER KERKE RREREGAE KORE Eed E KOEKE EReRE KOEN KORE KERK KERE anna EERS 2 1   2 1 1 Density Detector    ie SE se ee se SE en een ida ibi 2 1   2 1 2 Density Converter             nssneenneenneennennnennnennnennnennnennnnnnennnennnennnnn nenn 2 1   2 2 Main e ET 2 2   2 2 1 Density Detector EE 2 2   2 2 2 Density Converter    ese ee ke Re Re Re Re Re Re Re ee ee Re nenn 2 3    IM12T03A01 02E 4th Edition   Feb  23  2015 00    INSTALLATION  PIPING AND WIRING              ccccseeceeseseeeeeeseeeeeeeeseeeeeeees 3 1  3 1 Installation of Density DetectOF         ee see ee RE KEER KEER KEER KEER KEER KEER KEER KEER KEER ee 3 1  3 1 1 Mounting Density Detector AAA 3 1   3 1 2 Sampling Unit for Vibration Type Density Meter  3 3   3 2 Installation of Density Converter         usnensnensnennnennnennnennnennnannnennnannnannnannnannnannn 3 6  3 2 1 Installation Area  iii iia ia a 3 6   3 2 2 Installation of Density Converter  3 6   3 3 PUPUING BE 3 9  3 3 1 Sample Liquid Condunp nos 3 9   3 3 2 Air Piping for Maintenance  3 10   3 3 3 Stean PIPING soriana 3 11   3 3 4 Piping for Air PUG EE 3 11   3 4 WIN 3 12  3 4 1 Wiring between Detector and Converter  3 13   3 4 2 Wiring for Analog Output Signal                    erenneenneenneennenneennnen 3 14   3 4 3 Wiring for Digital Output Sons 3 14   3 4 4 Wiring for Abnormal Status Contact Output    3 14   3 4 5 Wiring for Power Supply          ecceeceeceeceeeeeeeeeseeeeeeeeeeeseeeeeeeeenetenee
16.  response   e Design the piping not to allow air bubbles to remain in the pipe which may cause an error   e Piping should not have sharp bending where solids may accumulate   e Maintenance such as cleaning should be made easily        Select element material most suitable for measuring liquid  Especially  for foodstuff  measurement  the selection should be done strictly      3  Installation of a Density Meter Sampling Unit  For the density detector with the sampling unit or with the sampling equipment user provided   installing place must  e be near to sample tap on the main pipe   e be free from vibration    e meet the ambient conditions specified by the hazardous area  when the detector is model  VD6DF flameproof type      Moreover  the sampling unit should be fixed to a concrete base drained well      4  Removing a Packing Sheet for Transportation from the Density Detector    The vibrator is fixed to the amplifier case to avoid being damaged during transportation     Remove the rubber sheet used for the fixing  The rubber sheet is inserted between the case  and the coupler of the vibrator assembly  Remove the sheet according the steps below  For  component names  refer to the section 4 1 1      a  Remove the lock with a hexagonal bar wrench  Release the cover by turning it  counterclockwise and pull it down not to touch the internal assembly      b  Remove the cover  with desiccant  from the vibrator     IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION 
17. 0 3 to 0 5 MPa G air is requires for cleaning the detector      Can heating steam for the density detector be stopped during operating time     4 3 2 Check for Wiring    Check that whole wiring of the measuring system including a receiving unit  for example   recorder  is completed     When the detector is explosion proof type  place the terminal box cover as completely as before  if it is once released     4 3 3 Measuring Liquid Flow    Operating procedure for introducing sampling liquid is explained here  assuming the piping in  Figure 3 5 in Section 3 1 2 is equipped  When pipe arrangement is different from that of Figure  3 5  carry out corresponding operations described here  It is important that air bubble does not  remain in the pipe     If calibrating is to be made with standard solution  conduct the measuring liquid after calibration  work to omit cleaning the vibrator  However  if pressure compensation is to be made  see Section  4 3 6   conduct the measuring liquid to obtain compensation value beforehand      1  First  set each valve in the sampling unit as follows   Ball valve    BV1     Fully closed    Needle valve    NV 1     Fully closed  Needle valve    NV2     Fully closed  Needle valve    NV3     Fully closed  Needle valve    NV4     Fully closed       Stop    valve    V1       Fully closed       Stop    valve    V2       Fully closed     Note  V1 and V2 marked by   are used only for steam tracer    2  For the sampling conduit with the steam tracing pipe  op
18. 1    display flashes during computation and the density at the reference temperature is  displayed after computation     Confirm that the density value is almost the same as that of measured at  2       5  Before carrying out the second calibration  change the density of the measuring solution in  the sampling unit  Perform the operations of  1  to  4   In the operation corresponding to  3    input the data to function No     7        8    and    9     In the operation corresponding to  4   use the  function No     A to specify calibration computation     IM12T03A01 02E 4th Edition   Feb  23  2015 00    4 4 3    4 5  4 5 1    4 5 2    4 5 3     lt 4  OPERATION gt  4 12    Calibration Error    If an error indication  E005 in the first calibration or E006 in the second calibration  is displayed  when calibration computation finishes  examine the data input to function No     3        4        5        7        8    or     9    and repeat the computation using the incorrect data     Operation    Operation Procedure    Start the density measuring system  and monitor operating conditions for a while to confirm   that there is no abnormality  Adjustment for the density measuring system during operation is  unnecessary  however  daily check of liquid flow rate and pressure is preferable  When the steam  tracing pipe is installed in the sampling unit  do not heat the detector     If an error indication is displayed  make a countermeasure immediately  Errors are listed in Table  4 2  
19. 14   Figure 3 8 Pipe Mounting    IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 7        1  Panel Mounting    The pipe mounting bracket should be removed from the converter before panel mounting   Panel cutout dimensions are shown in Figure 3 9     Before inserting the converter into front of the panel  remove the clumping brackets from both  sides of the converter case  After the case is inserted into a panel  remount the clumping brackets  on the same pace to hold the converter on the panel     Figure 3 10 illustrates that the converter is mounted on a panel     Unit  mm                                     212  1 I  275  1  aj  Y  Figure 3 9 Panel Cutout Dimensions Figure 3 10 Panel Mounting     2  Wall Mounting    Remove the pipe mounting bracket from the converter before wall mounting  Make two holes for  8 mm screws in a wall as illustrated in Figure 3 11     Unit  mm    Le  gt    2 M8 Screws     S    ote AJ     Figure 3 11 Holes for Wall Mounting Figure 3 12 Wall Mounting                   Y                   Using two mounting brackets  mount the converter on the wall  see Figure 3 12      IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 8     3  Pipe Mounting    Mounting bracket is applied to the pipe of nominal size 50A  diameter 60 5 mm   As shown in  Figure 3 13 mount the converter on the vertical pipe  inclination causes no problem         Ri       L          Figure 3 13 
20. FUNCTION SHIFT br INCRA data           Data sheet       SHIFT  gt  key    Pressed thi key  displayed  number flashes  Flashing   digit shifts right with every  pressing of the key     FUNCTION No  key                                           Mode transfer switch   DSPL  measuring mode    Mode for normal operation  In this mode  density at the reference  temperature  density at measuring temperature or measuring  temperature value is displayed by transferring function No    CAL  calibration mode    Mode for calibration   MAINT  maintenance mode      Mode for check of operating condition  setting a detector constant   a reference temperature     Cover of terminal board    Figure 4 2 Component Names of Density Converter    IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 4  OPERATION gt        4 2    4 2 1    Operation    Operation Mode    Operation of the Model DM8 vibration type density meter ruled by the mode as shown in Table  4 1  Measuring mode  DSPL   calibration mode  CAL  and maintenance mode  MAINT  are  transferred with the mode transfer switch                                                                                                                             Table 4 1 Operation Modes and Function No   MODE sw   FUNCTION Contents Display  Measuring No display   Displays density  at reference temperature  g cm3  mode 1 Displays density  measured value  g cm3   DSPL  2 Displays measuring liquid temperature   C  Calibration 3 Sets density of calibration solu
21. Feb  2015  YK  IM 12T03A01 02E 4th Edition  Feb  23  2015 00  All Rights Reserved Copyright    2006  Yokogawa Electric Corporation      Safety Precautions    m Safety  Protection  and Modification of the Product    e In order to protect the system controlled by the product and the product itself and ensure  safe operation  observe the safety precautions described in this user   s manual  We assume  no liability for safety if users fail to observe these instructions when operating the product        If this instrument is used in a manner not specified in this user s manual  the protection  provided by this instrument may be impaired        If any protection or safety circuit is required for the system controlled by the product or for  the product itself  prepare it separately        Be sure to use the spare parts approved by Yokogawa Electric Corporation  hereafter  simply referred to as YOKOGAWA  when replacing parts or consumables     e Modification of the product is strictly prohibited      The following safety symbols are used on the product as well as in this manual     ZN WARNING    This symbol indicates that an operator must follow the instructions laid out in this manual in order  to avoid the risks  for the human body  of injury  electric shock  or fatalities  The manual describes  what special care the operator must take to avoid such risks     A CAUTION    This symbol indicates that the operator must refer to the instructions in this manual in order to  prevent the 
22. ION  PIPING AND WIRING gt  3 15    3 4 5 Wiring for Power Supply  This wiring is to supply the converter with power     Use shielded 2 conductor cable for wiring  Cable termination and wiring method are the same as  those for analog output signal in Section 3 4 2     3 4 6 Wiring for Grounding  As arule  the case of the density detector and the density converter should be earthed     The grounding terminal of the detector is located on the base for the mounting bracket  and of the  converter  on the bottom of the case  Perform wiring so that ground resistance is 100Q or less   Class D  using a wire of 2 mm2 nominal cross sectional area     IM 12T03A01 02E 4th Edition   Feb  23  2015 00    4th Edition   Feb  23  2015 00     lt 4  OPERATION gt  4 1    4  OPERATION    4 1 Component Names and Function  4 1 1 Density Detector    Vibrator assembly  Case       Terminal box    Capacitance  pickup electrode    Oscillation  amplifier        Coupler  Vibrators    Piezoelectric  element    Cover    Lock screw   Keep loose during detector operation     Figure 4 1 Density Detector    IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 4  OPERATION gt  4 2       4 1 2 Density Converter    Data Display  Function No  and data are displayed     Front door    Function No  display Data display    Fastener  Nyratch           Internal assembly fixing screw   for transportation                    INCR A key   Increases the number of a  flashing digit     SET key   Used to enter a displayed       
23. Pipe Mounting    IM 12T03A01 02E 4th Edition   Feb  23  2015 00    3 3    3 3 1     lt 3  INSTALLATION  PIPING AND WIRING gt  3 9    Piping  The vibration type density meter requires the following piping    1  Sample liquid conduit   2  Air piping for maintenance   3  Steam piping   4  Piping for air purge    Sample liquid conduit is a pipe to conduct measuring liquid to the density detector  Air piping   for maintenance is to obtain air to blow away liquid from the vibrator at maintenance  vibrator  cleaning  calibration with standard solution   Steam piping which is installed according to  requirement is to obtain a steam to heat the sample liquid conduit  Air purge piping is necessary  when the converter is installed in dusty environment     The sample liquid conduit referred here indicates pipes between the main process pipe and the  sampling equipment  For piping within the sampling equipment  refer to Section 3 1 2     Sample Liquid Conduit    This piping conducts the measuring liquid to the sampling unit  The main process pipe should   have two sampling taps  sample inlet and sample return  as shown in Figure 3 14  The sample   conduit from the sample inlet is connected to    SAMPLE IN    of the sampling unit and the other   conduit from the sample return to    SAMPLE OUT    of the sampling unit    The piping procedure is as follows     1  Select the tap locations for the sample inlet and the sample return to make the pressure  difference between them become at least 0
24. User   s    Manual Model DM8  Vibration Type Density Meter    IM 12T03A01 02E    vigilantplant     YOKOGAWA 4 IM 12T03A01 02E    i f 4th Edition  Yokogawa Electric Corporation         Introduction    This manual describes the specifications  installation  operation  maintenance  and  troubleshooting for the Model DM8 Vibration Type Density Meter  To use this manual  correctly  read this manual thoroughly     This manual is applied to the following product   Detector  VD6D  Style B   VD6DF  Style B   VD6DS  Style B   Converter  DME8C  Style C   Special Cable  DM8W  Style A    For the Model VD6SM Sampling System  refer to IM 12T3S1 01E     m Notes on Handling User s Manuals    Please hand over the user s manuals to your end users so that they can keep the user s  manuals on hand for convenient reference     Please read the information thoroughly before using the product     The purpose of these user s manuals is not to warrant that the product is well suited to any  particular purpose but rather to describe the functional details of the product     No part of the user s manuals may be transferred or reproduced without prior written  consent from YOKOGAWA     YOKOGAWA reserves the right to make improvements in the user s manuals and product at  any time  without notice or obligation     If you have any questions  or you find mistakes or omissions in the user s manuals  please  contact our sales representative or your local distributor     Media No  IM 12T03A01 02E 4th Edition   
25. ained in the pipe     Assuming that the pipe is filled with a liquid as shown in Figure 2 1  the lateral free oscillation of  the pipe can be shown as the following equation       D                   Joint Joint  Figure 2 1 Pipe Filled with Liquid    D12   D22  f  av d Ce Pry DA o  1   Px Di2 D2    fx   Free transfers oscillation frequency for px  Hz   C   Constant determined by oscillation mode   I   Oscillating pipe length  m    E   Young s modulus of pipe  kg m     p     Density of pipe material  kg cm    px   Density of liquid measured  kg cm    D1   Pipe outside diameter  m    D2   Pipe inside diameter  m     In equation  1   the values other than fx and px are determined by the pipe material and  construction  Thus  the density px of the liquid can be obtained by measuring the free transfers  oscillation frequency fy     2 1 2 Density Converter    The density converter computes the liquid density using the oscillation frequency signal and  voltage of the temperature     Each value of I  E  pl  D1  D2 or px in equation  1  is a function of liquid temperature  hence the  value of f  is also a function of temperature  To obtain the correct density  the factors A t  and B t   depending temperature should be previously compensated for the temperature as follows    A t     f         Se  2   V1   px Bit     where  A t    Az   1 0060   1 9814 x 1047   9 7683 x 10 8  72    Dm   B1 1 4 5x105 T 30    A t     A   131072    100  B     B  300  T  Liquid temperature    C      Note  Bo
26. al   RT   RIV Multiplexer VIF  Ref  temp  Span 200 V  Ref  temp  Zero  Oto1V vi Gees  4 to 20 mA Zero Integrator Display    Span Controller  TD Key  Driver  Digital Output O O O O  Fail Alarm LED  7 00000  OG Badoo  o     Figure 2 3 Block Diagram of Density Converter    IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 1    3  INSTALLATION  PIPING AND WIRING    Model DM8 vibration type density meter consists of a density detector and a density converter   The density detector is generally installed in field and sampling piping is made to conduct  measuring liquid  The density converter is generally installed in the vicinity of the density detector   It must not be installed in hazardous area  It can be installed in a control room  though it is far  away from the detector  receiving digital signal     3 1 Installation of Density Detector    The density detector is shipped with a sampling unit if it is specified  For the detector without the  sampling unit  user should install sampling device suitable to measuring system     3 1 1 Mounting Density Detector    The detector  without the sampling unit  should be installed vertically  1  The pipe mounting  bracket is applicable to vertical or horizontal pipe  JIS 50A  as shown in Figure 3 1 and 3 2    1  The detector should be installed vertically to keep the vibrator pipe vertical      1  Pipe Mounting    The pipe mounting bracket is attached to the detector to mount it on a vertical pi
27. am tracing  Available  Process connection  Rc1 4  Electrical connection  G3 4    Mounting  JIS 50A  2 inch  pipe mounting  Ambient temperature   10 to 50  C  Weight  Approx  12 kg     2 Flameproof  Explosionproof  Detector Model VD6DF  Detector construction   TIS  d2G3 or  FM  Class    Division 1  Groups C and D  Flameproof construction  Process connection  Rc1 4 or 1 4NPT female  only for VD6DF O10 B FM   Electrical connection  G3 4 or 3 4NPSM female  only for VD6DF O0 B FM     Specifications are the same as for the  1  General Purpose Detector except for the above    construction      3 Sanitary Use Detector Model VD6DS    Process connection  Special joint for connection to JIS 6A pipe  within gasket   Wetted part materials  Added to the standard model   Gasket  Teflon   O Ring  Viton  Stream tracing  Not available    Specifications are the same as for the  1  General Purpose Detector except for the above two items     Temperature detector protecting tubes are detachable      Note   These detectors cannot be used with highly corrosive liquids and solutions likely to stick to sensors  If it is desired to be applied  to solutions containing slurry or sludge  consult with YOKOGAWA  For measuring NaOH solutions  use sensors with a nickel    vibrator     1 1 3 Converter Model DM8C    Display  Digital display  five digits LED  Display contents     Density  g cm3  after conversion to reference temperature  center temperature     Density  g cm  at the measuring temperature  Measuri
28. aning the vibrator     Install the sampling equipment according to the following  For details of piping from process to  the equipment  refer to the section 3 3      1  Flow Diagram    Figure 3 5 shows a flow diagram of a typical sampling equipment  Model VD6SM sampling unit of  Yokogawa is also designed according to this diagram     SAMPLE IN       SAMPLE OUT    Element specifications    a    F  Strainer body  316 SS  ER Xvi Xv2 Element  316 SS  or Ni  option   br a PI  Pressure gauge  0 to 1MPa or 0 to 2MPa  316 SS      Tl  Thermometer 0 to 100   C or 0 to 150   C  316 SS     gt  Bi FM  Flowmeter  tapered metal tube flowmeter  1 to 10 L min  316 SS    NV1 to 4  Needle valve  316 SS  DD  Density detector   V1  V2  Valve   ST1  Steam trap    D   SC    Figure 3 5 Flow Diagram    In Figure 3 5  each element of the sampling equipment is shown F  filter  is usually 80 meshes   prevents a solid from entering into the sampling line  TI  thermometer   FM  flowmeter  and PI   pressure gauge  are necessary to monitor measuring liquid  NV1  needle valve  is required to  control flow rate of a sample liquid  BV1  ball valve  and NV4  needle valve  are used for stopping  measuring liquid when the vibrator is cleaned or calibration is performed using standard solution    The needle valve is also used for adjusting flow rate   NV2  needle valve  and NV3  needle  valve  are used for draining liquid from the vibrator or for picking up sample for manual analysis     When viscosity of a measur
29. at the reference temperature  without function No  display   the  density value at measuring liquid temperature  No 1  or the measuring liquid temperature  No 2   can be displayed by selecting function No  Use FUNCTION key to select the FUNCTION No      2  Calibration mode  CAL   Mode for calibration  Sets or specifies data required  for calibration     During this mode the analog and digital outputs are held  at the values just before the holding    However  digital data output stops during calibration function No     6    or    A        When calibration computation ends holding is released and measured data are again output until  function No     6    or  A is transferred to another No     IM 12T03A01 02E    4th Edition   Feb  23  2015 00     lt 4  OPERATION    4 4    Measuring mode  DSPL     Maintenance mode  1    MAINT     Transfer to  Calibration mode    Calibration start    While in calibration  computation    Calibration end    Calibration mode  function No  transfer  or waiting for mode  transfer       3   lt   O         ke  6  E  ES  9  GE      g  2      O    Function SC  transfer  2    Maintenance mode  3 E       MAINT  aac  Measured E gt     Measuring mode  DSPL  value value       dls The density value after temperature conversion  density value at the measuring temperature and measuring liquid temperature  are transmitted in digital form   Only the density value after temperature conversion is transmitted in analog form     2  The density converter displays CAL 1 
30. ber  the set value becomes large than the standard solution density     2  Purge the inside of the vibrator with air as follows  Fully close the ball valve    BV1    and the  needle valve    NV4    in the sampling unit  Fully open the needle valve    NV2    and    NV3      connect an air pipe to the outlet port of the needle valve    NV3    and blow air for one or two  minutes  Pneumatic pressure should be 0 2 to 0 3 MPa 1       For sample liquid remains in the vibrator  blow it out by 0 05 MPa G pneumatic pressure  and increase the pressure up to 0 2 to  0 3 MPa to dry the inside of the vibrator     3  Pouring 18 to 20 mL of standard solution 1 into the vibrator  avoiding air bubble entering it is  described here     Firstly  remove the blind plugs from the pouring ports for the standard solution  two places   located on the detector base plate  Secondly  pour the standard solution into the pouring  port using the injector  accessory  so that the solution flows along the inner wall of the  vibrator tube  Using the pouring port near to the sample inlet is recommended  If the injector  is stained  clean it with alcohol      4  Read the measured density value at the reference temperature  It is displayed with the  mode selector switch 3 set to    DSPL    and function No  not displayed      5  Since calibration with standard solution 1 is made two times  blow out the used standard  solution from the vibrator  Before blowing  plug the pouring orts  two places   and blow air  acco
31. converter is designed dustproof and rainproof structure so that it can be installed outdoors     e However  when it is used with the Model VD6DF flameproof detector  it cannot be installed  the hazardous area  must be installed in the nonhazardous area  Further  when digital  output signal is used for a receiver  able length between the converter and the receiver  should be 10 m or less        Free from vibration   Vibration may cause illegal contact in electric circuits     Avoiding from direct sunshine     This instrument can be used in an ambient temperature of  10 to 55  C  However  direct  sunshine in summer may heat the instrument over the range        Free from corrosive gases and dusts     The structure of this converter is dustproof and rainproof  However  for maintenance  with  opening front door of the converter   installation in better environment is recommended  For the  same reason  the followings are conditions for selecting installation area    e Avoiding excessive humidity     e Avoiding wide temperature variation     3 2 2 Installation of Density Converter    The density converter can be installed on a panel or a wall or 2 inch pipe  The pipe should be  vertical to fix the mounting bracket on it  see Figure 3 8     Note  Mounting attitude is free from the converter performance  Mounting should be made of the following procedures     Pipe mounting  Mounting bracket hardware  gt            gt  iN Washer  Converter  x Spring washer  Ne D Bolt  M8x10   Bolt  M8x
32. display flashes seven  times      When the second calibration is finished  purge the standard solution from the vibrator and screw  the blind plug into the pouring port  fully close the needle valves    NV2    and    NV3    and remove the  connected air pipe     4 4 2 Calibration with a Sample    Calibration with sample liquid should be made according to the following procedure      1  Sample the measuring liquid which is measured by a hydrometer  Sampling should be  made when the liquid density is stable     NOTE    The density of the sample liquid should not change between sampling the liquid and finishing the  first calibration      2  Measure the density using the hydrometer and the temperature of the sampled liquid     Examine the temperature coefficient if it is not known  Using densities at high temperature   reference temperature   approx  5  C  and low temperature  reference temperature    approx  5  C   the temperature coefficient should be calculated as following     Temperature coefficient  g cm3   C      Density at high temperature     Density at low temperature    Difference between high and low temperatures     Note   The density values should be those at the reference temperature as far as possible     3  Set the mode transfer switch of the converter to    CAL    and input the data corresponding to  function No     3        4    and    5        4  For specify the first calibration computation  select the function No     6    and press    SET    key        CAL 
33. e  300 Q  e Output format  Measured data    E RR RE ER      CpLlr    space   space _t     Note  Data are output in the order of density  data converted into that at a reference temperature   density  at measuring temperature    and measuring liquid temperature  Hold data  data immediately before entering the mode  is output during the time of calibration  or maintenance mode  However  holding is released immediately after calibration or maintenance mode     Calibration status    CALIBRATION START Cele  space vc     Note   Nothing is output during calibration  However  when a parameter error occurs  output becomes as follows       ERROR      CRLE     5 or 6       CALIBRATION _ END Cpl  space _t     Note   If the function No  key is pressed during calibration  calibration is canceled  and the output becomes as follows     CALIBRATION _ CANCEL CRLE  space  Fail alarm    ERROR    _ERROR        CRLF    space     Note  When multiple errors occur  each error No  for the respective errors is output   Example    ERROR   1_ ERROR   3 CRLF    space    IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 1  GENERAL gt  1 4    Output signal span  0 05 to 0 5 g cm3 settable  Reference temperature set range  0 to 100  C  in increments or decrements of 1  C   Contact output on failure  One point  Contact closed on failure or power failure  Contact open when normal   Permissible voltage  220 V DC  250 VAC  Permissible current  2A  resistive load   Permissible contact power  60 W  Fault detect
34. e of the explosion   protection constructions  approval number  symbols for the type and working ambient  temperature range     Table 1 1 TIIS Flameproof Type  Explosion proof  Construction of Vibration Density Detector    Model and codes Type Approval No  Symbol for the type  including  explosion class and ignition group        Model VD6DF N1 No  T21726 d2G3  Model VD6DF S3 No  T21727 d2G3    FM Explosion proof Specifications    The Model VD6DF density detectors complying with FM explosion proof specifications is labeled  data plate which is marked with an approval mark  classification of hazardous location  working  ambient temperature range  and handling cautions     IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 1  GENERAL    1 11    1 4 3 Installation Area and Ambient Conditions    Model VD6DF density detector can be installed in hazardous area where specified gases are  present     However  do not install the detector at the place where explosive gases may be continuously  present  and in which the gas concentration is continuous or for long period  equal to or more  than  the lowest gas explosion limit     Environmental conditions at the installation site are very important to the detector   Before installing the detector  check the temperature  humidity and altitude of the site     The conditions shall not exceed the specified ranges  The temperature range   10 to 50  C  in  indicated on the data plate  and the altitude shall be 1000 meters or less above sea lev
35. eaes 3 15   3 4 6 Wiring for Groundrmg  rca 3 15  OPERATION WEEN 4 1  4 1 Component Names and Function        eers ee Re ER Re ER Rae AREA ER RE ER Re ER Rae ee 4 1  4 1 1 Density Detector    ss Ee aaa A 4 1   4 1 2 Density    Converter iii te cc oda da 4 2   4 2 eel dle  RE     Pon E E ET e ra ReE 4 3  4 2 1 Operation Mode  tirita ernten erahnen 4 3   4 3 Preparation for Measurement                   24422044400nnn0nnnnnnnnnnannnnnnnnnannnnnnnnnnnnnnnnnn 4 7  4 3 1 Check for Prpimg nn 4 7   4 3 2 Check TOR Wing    einen ch las 4 7   4 3 3 Measuring Liquid Flow esse sees ee Re Re Re Re Re Re Re Re Re Re ee 4 7   4 3 4 Tuming   ON the TE 4 8   4 3 5 Data Input    ae een nen einfahren 4 8   4 3 6 Pressure Compensation           ee AA AR RA Re ek nana 4 9   4 4 Cali 711 RE 4 10  4 4 1 Calibration with Standard Solution        ee ke ke 4 10   4 4 2 Calibration with a Sample           eek ek Re RA Re Re Re Re ee 4 11   4 4 3 Calibration Eitorf    4 12   4 5 Operation iii N EE EE aSa 4 12  4 5 1   TAN 4 12   4 5 2 Check of Measured Density Value nennen 4 12   4 5 3 aie EE EE ia 4 12    IM12T03A01 02E 4th Edition   Feb  23  2015 00    5  MAINTENANCE E 5 1  5 1 Daily Inspection Maintenance comicos 5 1   5 1 1 Cleaning the Vibrator in the Detechor   AA 5 1   5 1 2 Exchange of Desiccant in the Detector A 5 1   5 2 Troubleshooting      esse sesse ses eke ek ee GR ek SE KEN aan nam aan ann iia 5 2   5 2 1 Checking the Vibrators AAA 5 2   5 2 2 Checking the RIDE 5 2   5 2 3 Checking the O
36. ed   adjustment may be necessary  If trouble occurs  as a rule  repair should be made by Yokogawa   In this case  detailed information on abnormal phenomena will be effective for speedy recovery     Searching method for abnormality is explained below  For the explosion proof detector  confirm  that there are no explosive gases using a gas detector before removing the terminal box cover     Checking the Vibrators    Remove the cover of the detector  so that the vibrator can be visually inspected     Carefully check the entire surface of the two tubular vibrators  If they have dents  scratches   holes or show signs of the sample liquid leakage  the vibrator assembly should be exchanged    CAUTION     Make sure that the lock screw is not in contact with the vibrator assembly     CAUTION    The vibrator tube has extremely thin wall  hence must be carefully handled  Since assembling the  vibrator tube requires special skill  the exchange of the vibrator should be made by Yokogawa     Checking the RTD    Measure electrical resistance between the terminal A2 and A3  When the RTD is normal  the  resistance shows the value indicated in the following table     Table 5 1 Resistance at each temperature of RTD  Pt 1000Q     Temperature    C  10 15 20 25 30 35 40 45 50  Resistance  Q    1039 0   1058 5   1077 9   1097 3   1116 7   1136 1   1155 4   1174 7   1194 0                                           IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 5  MAINTENANCE gt     5 3      
37. el and the  relative humidity range 45 to 85       1 4 4 External Wiring Work    External wiring for flameproof detector shall be carried out according to the flameproof metal  conduit wiring  For TIIS explosion proof specifications  the flameproof packing type is used for  leading external cables     1 4 5 Maintenance Procedure    Do not remove the detector cover in a hazardous area unless power is OFF  However  if detector  maintenance with power on is unavoidable  a gas detector check should first be made to  determine whether a hazardous atmosphere exits in the installation area     1 4 6 Repairs    Regarding to repairs of flameproof detector  consults with service personnel   The following points must be observed during repairs of flameproof detector     The detector must be restored to their original conditions  electrically and mechanically to  maintain their flameproof properties        WARNING    The gaps  path lengths and mechanical strength of enclosures are important factors in  establishing flameproof properties  Therefore  service staff should be careful not to damage point  surfaces and not to shock enclosures        IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 2  PRINCIPLES OF OPERATION gt  2 1    2  PRINCIPLES OF OPERATION    2 1 Measuring Principle  2 1 1 Density Detector    Model VD6D Density Detector is a vibration type density meter  Measuring principle is that the  lateral free oscillation of a pipe is a function of the density of the liquid cont
38. ellow    5 K9141TV 1 Fuse  250 V  3 A    6 K9000ZE 1 Transformer Assembly     Copyright 1999  1st Edition  1987  YK  CMPL 12T3H1 01E   YOKOGAWA  gt  Subject to change without notice  7th Edition   Feb  2015  YK     Yokogawa Electric Corporation       Item       RON       CMPL 12T3H1 01E    Accessories Assembly    Part No   B1003JZ    L9811FP   K9141TV  G9612BC  G9612BE    Qty             oo Ko    2    Description    Cable Gland   Cable Gland   Fuse  250 V  3A    Connector Rc1 4  for  AP1   Connector 1 4NPT  for  AP2     7th Edition   Feb  2015  YK     Revision Information    e Title   Model DM8 Vibration Type Density Meter  O Manual No    IM 12T03A01 02E    Feb  2015 4th Edition  DM8C Style C to D  P 1 3  1 4  1 5  2 3  4 2  4 5  4 12 and CMPL 12T3H1 01E   Comment addition  P 4 11     Jun 2012 3rd Edition  P 4 9 Figure 4 5 Flow Diagram added  P 5 1 Figure 5 2 Flow Diagram added  CMPL 12T3H1 01E    revised to 6th edition     Apr  2009 2nd Edition  P 1 4 Standard Accessories  P 1 5 DM8C Style B to C  change of CMPL 12T3E1 01E P 2 Item 3 and    CMPL 12T3H1 01E P 1 Item 1     Dec  2006 1st Edition  Newly published  For Style  B     E If you want have more information about Yokogawa products  you can visit  Yokogawa   s homepage at the following web site   Homepage  http   www yokogawa com an    IM 12T03A01 02E 4th Edition   Feb  23  2015 00    
39. en the steam main valve to heat the  sampling conduit     Keep the    stop    valve    V1    closed  not heat the detector if it is not remarkably cooled      3  Prepare a sample receptacle  liquid pan  just under the    drain    valve provided in the sample  conduit  First  fully open the    drain    valve  then gradually open the    tap    valve     When the sample liquid begins to flow out of the drain valve  close the    tap    valve      4  The same procedure as  3  should be made for the sample return conduit  Prepare a  sample receptacle  liquid pan  just under the drain valve  First  fully open the    drain    valve   then gradually open the    stop    valve  When the sample liquid begins to flow out of the    drain     valve  close the    stop    valve     IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 4  OPERATION gt  4 8     5  Fully open    BV1    and    NV4    and half open    NV1    in the sampling unit  Close    drain    valve  opened in  3   Gradually open the    tap    valve located on the sample inlet conduit  As sample  begins to flow out of the    drain    valve on the sample return conduit  fully open the    stop    valve  and fully close the    drain    valve when no retaining air is found in the liquid      6  Open    NV     to exhaust retaining air in the pipe  When the sample liquid without air flows out  from the    drain    port  fully close the    NV2         7  Adjust opening of NV1 to make sample flow rate approx  5 L min for the detector  
40. er will be charged repair fee regardless of  warranty period      Failure of components which are out of scope of warranty stated in instruction manual        Failure caused by usage of software  hardware or auxiliary equipment  which Yokogawa  Electric did not supply        Failure due to improper or insufficient maintenance by user     e Failure due to modification  misuse or outside of specifications operation which Yokogawa  does not authorize        Failure due to power supply  voltage  frequency  being outside specifications or abnormal      Failure caused by any usage out of scope of recommended usage        Any damage from fire  earthquake  storms and floods  lightning  disturbances  riots  warfare   radiation and other natural changes     m Yokogawa does not warrant conformance with the specific application at the  user site  Yokogawa will not bear direct indirect responsibility for damage due  to a specific application     m Yokogawa Electric will not bear responsibility when the user configures the  product into systems or resells the product     m Maintenance service and supplying repair parts will be covered for five years    after the production ends  For repair for this product  please contact the  nearest sales office described in this instruction manual     IM12T03A01 02E 4th Edition   Feb  23  2015 00    Model DM8    Vibration Type Density Meter    IM 12T03A01 02E 4th Edition    CONTENTS            WEG DE e E i     Safety EN ii        After sales Warranty AA
41. etector      Note  Detector constant A changes with each calibration   Function No     B      Sets detector constant B    Sets the detector constant B  Since the detector constant B is also an inherent characteristic  of the detector  this is indicated in each detector     Function No     C      Displays detector calibration coefficient C  span     Indicates either the one point or two point calibration has been performed  1 is indicated for  one point calibration and calibration coefficient  Span  for two point     Function No     D      Displays detector calibration coefficient D  zero     Indicates either the one point or two point calibration has been performed  0 is displayed for  one point calibration  and calibration coefficient  zero point  than 0 is displayed for two point  calibration     IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 4  OPERATION gt  4 7    4 3 Preparation for Measurement  4 3 1 Check for Piping    Check the following points      Is the pressure difference 100 kPa or more between both sampling tap        Is the diameter of the sampling pipe appropriate  Is length of the inlet pipe too long      Note   The sample inlet pipe is usually JIS 15A or Sch 40 to 80 stainless steel pipe  The pipe length between sampling tap and the  detector should be 10 m or less        Are valves and meters  pressure gauge  thermometer  provided in the sampling pipe for  normal operation and maintenance   e Is air for cleaning the detector provided    Note   Dried 
42. evitable  The Detector not moved for usual check or maintenance does not require the flexible tube        Figure 3 23 Wiring to Density Detector Terminals Figure 3 24 Spanner for Open closing  Terminal Box Cover   Cable   Terminal box case          Be sure to screw in  at least 12 mm  5 threads        G3 4 screw  TIIS explosionproof     Lock nut 3 4NPSM  FM explosionproof     Explosionproof flaxible fitting       Figure 3 25 Explosion proof Flexible Fitting    IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 14     lt  Connection to Density Converter  gt     The density converter is installed in field or indoors  control room   The connector for passing  through cable should be sealed for air purge  The connector with a ring packing as shown in  Figure 3 26 should be used  or the inlet of metal conduit is filled with silicon sealing compound   Duct or protective pipe for the cable should be provided to prevent the cable from damage     Cable    Converter case        e  HI G3 4 screw    Lock nut    Flaxible fitting          Figure 3 26 Packing Gland Type Connector Figure 3 27 Cable Inlet for Metal Conduit  for Passing through Cable    Take the special cable into the converter from the leftmost inlet and connect the each conductor  to the respective terminals     3 4 2 Wiring for Analog Output Signal    Output signal of 4 to 20 mA DC output  load resistance 5500 or less  and 0 to 1 V DC  load  resistance 250 kQ or more  are provided 
43. fficient of measuring liquid that is to be input into function No     9     is  unknown jet  refer to    4 4 2 Calibration with Sample Liquid  2      IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 4  OPERATION    4 9       4 3 6 Pressure Compensation    Pressure compensation means to compensate the shift generated by measurement under  operating pressure  The compensation should be carried out as necessary only for calibration  under atmospheric pressure     Measured value is affected up to  0 00005 g cm by 100 kPa pressure variance     Therefore  for liquid under pressure approximating to the atmospheric pressure calibrated  value under atmospheric pressure can be used  Pressure compensation is to be carried out to  minimize effects generated by high operating pressure     Compensation data is obtained by conducting sample liquid  Measure the data according to the  following procedure when the liquid density is stable  The procedure assumes that sample liquid  is already conducted according to Section 4 3 3      1  Fully open the needle valve    NV4    and read P from the pressure gauge  At the same time   obtain density at the reference temperature indicated on the density converter      2  Fully close the ball valve    BV1     and open the needle valve    NV3    to equalize the sample  liquid pressure in the detector to the atmospheric pressure  Obtain density at the reference  temperature indicated on the density converter      3  Calculate the compensation data us
44. g the 2nd calibration two point calibration  so  the  two point calibration using standard solution and sample liquid is described as follows  Since the  following is described assuming that the density converter is installed in the vicinity of the density  detector  for the connector installed away from the detector  please edit the manual for it and  carry out the calibration according to the manual     Calibration with Standard Solution    Provide two kinds of standard solution whose density at reference temperatures is already known   1  One density  standard solution 1  should be near to the low limit of the measuring range and  the other to the high limit  standard solution 2   Temperature coefficients  g cm3   C  of these  standard solutions should be obtained beforehand    SI When measuring the density of the standard solution  use a hydrometer whose minimum scale is 0 0005 kg cms3     Calibration should be made in the following procedure  During calibration  only the density  detector and the density converter are to be driven and the unit to receive output signals from the  converter can remain stopped      1  Set the converter to    CAL    mode and input data corresponding to function No     3        4        5        7         8    and    9     For pressure compensation  subtract the compensation value obtained in Section  4 3 6 from density of standard solution  and set them to function No     3    and    7         Note  When the compensation value is a minus num
45. gulations and the cautions marked on the analyzer shall be strictly  observed when using Model VD6DF in a hazardous location        WARNING    Model VD6D and Model VD6DS Density Detector  which are not    explosion proof instrument      cannot be installed in hazardous location           Model VD6DF density detector has either TIIS or FM flameproof type of explosion proof  specifications  The TIIS type has passed tests conducted by a public authority in accordance with  the Labor Safety and Health Law  Japan  which regulates domestic explosion proof electrical  equipment  The FM type has been certified to have an explosion proof construction meeting the  requirements of NEC  National Electrical Code  by Factory Mutual Research Corporation  U S A    FM   a testing organization  It also conforms to requirements of OSHA  Occupational Safety and  Health Act  U S A       Items 1 4 2 to 1 4 6 cover general cautions in using the Model VD6DF density detector  For  further details  see the following publications        For TIIS flameproof requirements       RECOMMENDED PRACTICE    for Electrical Equipments for Use in Explosive Gas  Atmosphere published by The Research Institute of Industrial Safety  Ministry of Labor        For FM Flameproof requirements  National Electrical Code  Chapter 5  Special Occupancies     Labeling the Explosion proof Specifications  TIIS Flameproof type of explosion protected construction    The Model VD6DF density detector is labeled data plate on which typ
46. heet        Mounting pipe    4 d  1  When option code is specified  Rc1 4 or 1 4 NPT female connector is attached    2 inch pipe     Panel cutout dimensions Wall mounting hole drilling dimensions    212 1 222   lt   gt          2  9 holes or M8 screws                         E m       275 1             IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 1  GENERAL    1 9       1 3 3 Special Cable Model DM8W  Unit  mm   t    Density detector Density converter        gt     130 from A   from B 100           BROWN Ja   60 from A   from B 80 BROWN        Note  Cable length is specified by the suffix code of     LOIOOD   O is specified in meter   e g  for 50 m   L0050  for 100 m   L0100  for 2 km   L2000    IM 12T03A01 02E 4th Edition   Feb  23  2015 00    1 4    1 4 1    1 4 2     lt 1  GENERAL    1 10    Cautions in Handling Flameproof   Explosion proof Instruments    Outline of Explosion proof Instruments    Specifications of    Explosion proof Instruments    conform to the regulations of relevant public  organizations     Model DM8 Vibration Type Liquid Density Meter consists of a density detector and a density  converter  Model VD6DF density detector  explosion proof instrument  can be installed in  hazardous area where explosive gases may be generated     It should be noted  however  that the method of installation  ambient conditions  and the handling  of these instruments must conform to the regulations of relevant public organizations  The  applicable explosion proof re
47. ing contents  Detector failure and converter failure  Failure output   Analog signal  Falls down to about  10   of the output signal span  Digital signal  Error message outputs  Output signal hold  Holds in the CAL  or Maintenance mode   Settable range for temperature coefficient  0 to 0 002 g cm3   C  Calibration procedure  One touch calibration by strong calibration liquid density  one point calibration   Ambient temperature   10 to 55  C  Power supply  90 to 132 V AC or 180 to 264 V AC  50 60 Hz  Case construction  Dust and rain proof construction  Coating color   Door  Equivalent to Munsell 2 8GY6 4 0 9  Case  Equivalent to Munsell 2 0GY3 1 0 5  Coating finish  Baked finish epoxy resin  Mounting  To panel  wall or 2 inch pipe  Air purge connector  Rc1 8  Rc1 4  or 1 4NPT female is also optionally available  Electrical connection  Five holes  27 mm dia   Attached with four plastic waterproof plugs equivalent to JIS A15  and  one plastic waterproof plug equivalent to JIS A20     Weight  Approx  7 0 kg   1 1 4 Special Cable Model DM8W  Type  Six conductor double shield cable  Insulator  Polyethylene  Sheath  Polyvinyl chloride    Insulation resistance  1000 MQ km  Conductor resistance  15 31 Q km  Finished O D   15 8 mm                      Weight  Approx  0 3 kg m  1 1 5 Standard Accessories  Syringe  for injecting standard  solution or solvent  EES  Brush  for cleaning the detector  1 pe   Allen wrench for terminal box 1 pe  for Detector  VD6   Allen wrench for locking the
48. ing liquid is to be lowered  for example  the pour point of measuring  liquid is high  arrange the density detector and the liquid pipe should be warmed individually   Steam pipe can be connected to the density detector and the liquid pipe can also be traced by  steam pipe  V1     stop    valve  is to supply the steam to the density detector and V2     stop    valve   is to supply the steam to tracing pipe  ST1  steam trap  is used to drain condensed water      Note1   Kind of valve used in the sampling equipment varies with the purpose of use  This instruction manual gives a detailed name of  equipment  for example  ball valve  used for limited purpose  and shows a general name within        when equipment type is not  restricted  for example    stop    valve      Note2   Density Detectors of Model VD6D  general purpose type  and VD6DF  flameproof  have steam pipe connections  however   Model VD6DS  sanitary use  has not the connections     Note3   Model VD6SM sampling unit  a product of Yokogawa  is applied to Model VD6D  general purpose type  or Model VD6DF   flameproof type   It is not recommended for the Model VD6DS  sanitary use  used for food     IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 4       Density meter sampling unit    Anchor bolt       Concrete base    Figure 3 6 Installation Example for Sampling Unit  Model VD6SM      2  Note of Designing a Density Meter Sampling Unit  e Minimize liquid pipe length to improve
49. ing the densities obtained in  1  and  2    Compensation data  g cm3     Density under working pressure     density under atmospheric pressure    When calibration the compensation data should be subtracted from the density value of the  standard solution     SAMPLE IN       SAMPLE OUT      A i Element specifications    F  Strainer body  316 SS  are Xvi Zu Element  316 SS  or Ni  option   br oi Pl  Pressure gauge  0 to 1MPa or 0 to 2MPa  316 SS      Tl  Thermometer 0 to 100   C or 0 to 150   C  316 SS    ST FM  Flowmeter  tapered metal tube flowmeter  1 to 10 L min  316 SS  CONDENSATE OUT Y e BV1  Ball valve  316 SS    NV1 to 4  Needle valve  316 SS  DD  Density detector   V1  V2  Valve   ST1  Steam trap    Figure 4 5 Flow Diagram    IM12T03A01 02E 4th Edition   Feb  23  2015 00    4 4    4 4 1     lt 4  OPERATION    4 10    Calibration    The vibration type density meter should be calibrated as occasion demands to keep the specified  accuracy     Calibration has two methods  one point calibration and two point calibration  and each method is  carried out by either sample liquid or standard solution  The calibration with sample liquid is not  requiring stopping operation and pressure compensation resulting in accurate measurement  On  the other hand  in calibration with standard solution  the solution density is once determined at  the fixed value  density measurement of sample liquid should not be repeated at a laboratory     The one point calibration is equal to that omittin
50. instrument  hardware  or software from being damaged  or a system failure from  occurring     CAUTION    This symbol gives information essential for understanding the operations and functions     NOTE    This symbol indicates information that complements the present topic     E Warning and Disclaimer    The product is provided on an    as is    basis  YOKOGAWA shall have neither liability nor  responsibility to any person or entity with respect to any direct or indirect loss or damage arising  from using the product or any defect of the product that YOKOGAWA can not predict in advance     IM12T03A01 02E 4th Edition   Feb  23  2015 00      After sales Warranty    E Do not modify the product     E During the warranty period  for repair under warranty consult the local sales  representative or service office  Yokogawa will replace or repair any damaged  parts  Before consulting for repair under warranty  provide us with the model  name and serial number and a description of the problem  Any diagrams or  data explaining the problem would also be appreciated     e  Ifwe replace the product with a new one  we won t provide you with a repair report     e Yokogawa warrants the product for the period stated in the pre purchase quotation  Yokogawa shall conduct defined warranty service based on its standard  When the  customer site is located outside of the service area  a fee for dispatching the maintenance  engineer will be charged to the customer     E In the following cases  custom
51. isplayed  a function No   is not entered  To release the protection  display    77    and press    SET    key  To reactivate the  protection  display  78  and then press the  SET  key  To avoid the possibility of hindering  normal measurement  when releasing or reactivating the protection  do not press the  SET   key after displaying other numbers  In addition  the reactivation of protection does not take  effect until you perform another operation  e g   change the function No  to  2   after you  release the protection    Function No     2      Displays frequency  density   Checks operation of this converter  Frequency corresponding to measured density are  displayed  When the converter is normal  the frequency ranges from 600 to 700 Hz to 1000  or 1500 Hz      Note   Displayed frequency varies with measured density  liquid temperature and detector constant  If density is 0 5 to 2 0 g cm3   temperature is 25   C  detector constant  A  is 17074 and detector constant  B  is 36384  the frequency is 640 to 1400 Hz for  normal operation     Function No     3      Displays frequency  temperature   Checks operation of this converter  A frequency signal converted from liquid temperature  voltage is displayed  If temperature is 25   C  the frequency is about 25000 to 35000 Hz for  normal operation    Function No     4      Checks output signal  Checks zero point and span of an analog output signal     Note   The data at shipping is described in a data sheet in the converter    Functi
52. lectric element to maintain vibration corresponding to the sample liquid density     The capacitance pickup electrode installed between the two vibrating pipes detects the lateral  oscillation frequency of the vibrator     IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 2  PRINCIPLES OF OPERATION gt  2 3       The amplifier converts the output of the capacitance pickup electrode into an AC voltage   to amplify it  This frequency signal is  together with the temperature signal from the RTD   transmitted to the converter  A part of the frequency signal is fed back to the piezoelectric element  to maintain the vibrator oscillation     2 2 2 Density Converter  Figure 2 3 is a block diagram of the density converter     As shown in the diagram  the density converter receives the density signal  frequency  F and  temperature signal  voltage  RT from the density detector and finally outputs a 4 to 20 mA  analog signal and 0 to 1 V DC signal corresponding to the liquid density converted to that at  reference temperature  The digital signals of densities at measuring temperature and reference  temperature  and measuring temperature and measuring temperature value are also output   The converter circuit comprises three printed circuit boards  Changing displayed items or setting  constants are performed by keys or transfer switches on the front panel     Frequency of Density  F   Multiplexer                   gt   Fo                            EEPROM  eference frequency  Temp  Sign
53. ly 300 to 500 kPa G    Flexible tube Reducing Stop valve  valve    Figure 3 16 Air Piping for Maintenance    IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 11    3 3 3 Steam Piping    This is installed to heat measuring liquid whose pour point is high  and to decrease its viscosity   The sampling unit with steam trace pipes should be used for this purpose     Connect a pipe from steam source to the sampling unit    STEAM IN    inlet  The sample liquid  conduit also should have a steam tracer pipe  Steam of pressure 0 3 to 0 5 MPa and temperature  of 140 to 160  C is preferable     Installing the tracer pipe for the sample liquid conduit should be performed after pressure  retentive and leak test of the conduit     Notes for the tracer piping are as follows      1  The tracer pipe of the sample liquid conduit should be installed such that the entire conduit  can be heated  The pipe should also be covered with insulating material  see Figure 3 17       2  The steam trap discharge outlet should be open to the atmosphere  see Figure 3 18             Heat insulating material  Trace pipe    Sampling line    Figure 3 17 Steam Tracer Pipe    Steam trap    Sampling unit    Ur    Drain pit  Figure 3 18 Piping at Steam Trap Discharge Outlet    3 3 4 Piping for Air Purge  If the density converter is installed in dusty environment  air purging is recommended     Air purge should be performed continuously by clean dried air with pressure of 50
54. ng liquid temperature    C   Set density value for the calibration liquid  g cm3   displayed on call     Temperature coefficient set value for the calibration liquid  x10 5 g cm3   C   displayed on call     Output signal set value      displayed on call   Setting for output range low limit  g cm3   displayed on call   Setting for output range high limit  g cm   displayed on call     Reference temperature  center temperature  set value    C   displayed on call   Temperature coefficient set value for the measuring liquid  x10 5 g cm3   C   displayed on call   Fault contents display    IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 1  GENERAL gt  1 3       Output signal   Analog output  4 to 20 mA DC  load resistance 550 Q or less   and 0 to 1 V DC   load resistance 250 kQ or more   isolated output   Density  g cm  after conversion to the reference temperature  Digital output  RS 232C   1  Communication specifications  e Asynchronous system  Start bit  1 bit  Stop bit  2 bit  Parity  none     Baud rate  1200 bps     Data format  ASCII  Data length  8 bit  e Wiring system  Two wire system  output only    2  Output signal     Transmitting contents    Measured data  Density  g cm3  after conversion to the reference temperature  density  g   cm3  at the measured temperature  measured liquid temperature    C     Calibration state  calibration start  error No   calibration end  Failure alarm  error No      Signal level  Output voltage  ON  6   1 V  OFF   6   1 V  Output impedanc
55. on No     5      Checks LED  press SET key   Function to check LED data display  When    SET    key is pressed  all displays flash five times  if it is normal    Function No     6      Sets low limit of the output range    Sets a low limit density value  g cm  of an analog output range      Note   Set the value so that high limit   low limit is greater than equal to 0 05 g cm3  You cannot enter a value that fails to meet this  condition     Function No     7      Sets high limit of the output range    Sets a high limit density value  g cm3  of an analog output range      Note   Set the value so that high limit   low limit is greater than equal to 0 05 g cm3  You cannot enter a value that fails to meet this  condition     Function No     8      Sets reference temperature of measuring liquid  Sets a reference temperature    C  of measuring solution   Function No     9      Sets temperature coefficient of measuring liquid    Sets the value which is  1 x 105     100000  times the temperature coefficient of measuring  solution  g cm3   C  to change it to plus integer      Note   Purified water is an exception  its maximum density is at approx  4   C  Generally  temperature coefficients of any liquids are  minus number     Function No     A      Sets detector constant A    IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 4  OPERATION gt  4 6    Sets the detector constant A  Since the detector constant A is an inherent characteristic of  the detector  this is indicated in each d
56. or CAL 2     Figure 4 3 Digital and Analog Signal Output Statuses       Eight function Nos  from    3    to    A    are provided in the calibration mode  see Table 4 1   Among  them     6    and    A are to specify the calibration computation and others are to input data required  for the calibration     Function No     3     Sets density of calibration solution 1    Density value  g cm3  of a calibration solution for one point calibration or 1st calibration of  two point calibration can be input  The input density is a data at the reference temperature      Note   If relation between temperature and density is well known  a density at a temperature can be set  However  the density data at  the reference temperature is recommended to achieve accurate calibration     Function No     4     Sets temperature of calibration solution 1    Input the temperature of the calibration solution 1 of which density at the reference  temperature is set at    3        Function No     5     Sets temperature coefficient of calibration solution 1    Enter the value which is  100000 times the temperature coefficient  g cm3   C  of the liquid  for calibration 1    Note   The temperature coefficient  0 00086 g cm3   C becomes the value 086     Function No     6     Starts calibration 1  press SET key     Pressing    SET    key executes calibration computation  or also executes the first calibration  of the two point calibration    Note   For calibration procedure  refer to Section 4 4   Function No
57. pe  For mounting  on a horizontal pipe  remove the mounting bracket from the detector to turn it 90   and fix it   again to the detector  Figure 3 1 shows installation on a vertical pipe and Figure 3 2 shows on a  horizontal pipe        Figure 3 1 Installation on a Vertical Pipe Figure 3 2 Installation on a Horizontal Pipe    IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 2     2  Bracket Mounting    This bracket is not the pipe mounting bracket  Release the four bolts  with spring washers  to  remove the pipe mounting bracket from the detector  Using the four bolts and spring washers   fix the detector on the bracket as shown in Figure 3 4  Figure 3 3 shows position of holes to be  drilled on the bracket     Unit  mm    70  0 2       pe  gt  4 89 Holes  40  0 2            Figure 3 3 Position of Holes on the Bracket          Density detector       Figure 3 4 Mounting on the Bracket    IM 12T03A01 02E 4th Edition   Feb  23  2015 00     lt 3  INSTALLATION  PIPING AND WIRING gt  3 3       3 1 2 Sampling Unit for Vibration Type Density Meter  This section describes a sampling that is provided by user for the density meter     The sampling equipment is equipped with valves  a thermometer  a flowmeter  etc   and is  located near to process pipe and the density detector     The sampling equipment is used to provide good condition to measure liquid density  easy to  monitor measuring and maintenance condition such as calibration and cle
58. rature which is same as the sample liquid  temperature      3  Compare the density measured at  2  with the displayed value when sampling the liquid  If  the difference of them exceeds the allowable error range  clean the vibrator     For cleaning procedure  refer to Section 5 1 1     Shutdown    When shutdown sample following steps and the vibration type density meter also stops  drain the  sample from the detector     IM 12T03A01 02E 4th Edition   Feb  23  2015 00    5     5 1  5 1 1    5 1 2     lt 5  MAINTENANCE gt  5 1    MAINTENANCE    Daily Inspection Maintenance    Cleaning the Vibrator in the Detector    Dirt particles may be deposited on the inner wall of the vibrator pipe  depending upon the nature  of the process liquid  This can cause an error  Clean the pipe at regular intervals in the following    manner     1  Close valves BV1 and NV4 and open NV2 and NV3     2  Connect the air pipe to the outlet of the needle valve NV2  Purge the sample liquid in the  vibrator from the drain port of the needle valve NV3 by blowing 0 05 MPa G air     3  Remove the blind plugs from the pouring port of the standard solution  two places  and pour  cleaning liquid about 200 mL into the vibrator     4  After a while  fix blind plugs again on the pouring ports and purge the cleaning liquid in the       vibrator as described in  2   Wash the inside of the vibrator pipe with a brush  accessory  as  shown in Figure 5 1     SAMPLE IN       SAMPLE OUT   t   a    i Element specifications 
59. rding to  2      6  Carry out the operations of  3  and  4  to confirm that the same results as that of  4  is  achieved     Note   If the result differs from  4   repeat the operations of  3  and  4      7  For the calibration 1 computation  press    SET    key at the function No     6        CAL 1    flashes  during computation and the density at the reference temperature is displayed after the  computation finishes  NOTE  a     Check that this value is the correct density  NOTE  b    of  the standard solution 1     IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 4  OPERATION gt  4 11    NOTE     a  Time for calibration computation varies with density value difference between pre calibration  and after calibration  The larger the difference is  the longer the time is  It can take up to  about 7 minutes     b  The density value after calibration doesn   t always coincide with the density of the standard  solution 1 even if they are compared in the same temperature   Especially when the solution temperature during calibration largely different from the  reference temperature an error is apt to occur   However  the error is usually within allowable range  if it is too large  examine each set  value      8  The first calibration is finished by  7   Then  use the standard solution 2 carry out the second  calibration whose procedure is same as the first calibration  Repeat the operations of  1  to   7   For the calibration computation  set the function No  to    A     CAL 2 
60. scillation Amplifier AAA 5 3   Customer Maintenance Parts List    ese Re EE RE ER KERE EE CMPL 12T3E1 01E  Customer Maintenance Parts List    ees ee Re EE RE ER EE EE CMPL 12T3H1 01E  Revision Information  lt a i    IM 12T03A01 02E 4th Edition   Feb  23  2015 00    1     1 1    1 1 1     lt 1  GENERAL gt  1 1    GENERAL    Density is one of the fundamental physical quantities required when the property or composition  of a liquid sample is being determined in industrial processes  Especially  density measurement  of a sample liquid is indispensable in many industries such as electric manufactures  chemical  industries  oil refineries and food related applications     The Model DM8 Vibration Type Density Meter is a higher reliable meter full of operation and  maintenance functions  The meter consists of a detector and a converter  The detector has a  vibrator and an RTD  resistance temperature detector   and outputs detected density signal   frequency  and temperature signal  voltage  to the converter  The converter is equipped with a  microprocessor to convert the frequency signals from the detector into density values and display  them  lt also calculates the density at the reference temperature according to these signals and  digitally displays them  Moreover  the converter outputs digital transmission signals besides  analog transmission signals and is further provided with a variety of functions such as one touch  calibration  self diagnosis     Standard Specifications   
61. th A and B are constants of the detector which has inherent values     IM12T03A01 02E 4th Edition   Feb  23  2015 00     lt 2  PRINCIPLES OF OPERATION gt  2 2       From equation  2  and  3   the density px can be obtained     A 2  EE  Px   BE RE AE ER EE RE ET EE EK  3   The px in equation  3  represents the liquid density at measuring temperature  The density pTB  at the reference temperature can be obtained by the following equation  4    MEE TITEL MEE EE een een  4     a   Temperature coefficient of density for measuring liquid  g cm3   C   Tx   Liquid temperature at density measurement    C   Tg   Reference temperature    C     2 2 Main Components  2 2 1 Density Detector    Figure 2 2 is a schematic diagram of the density detector     As shown in the figure  the density detector consists mainly of a vibrator assembly  a capacitance  pickup electrode  and an amplifier     Sample    Vibrator D                Pick up electrode  Preamplifier e    TO Density signal  KE                                             Temperature signal          Temperature  measuring element    Piezo electric element          Vibrator cross sectional view of Density Detector    Figure 2 2 Schematic Diagram of Density Detector    The vibrator assembly consists of a sample path formed by connecting the ends of two thin  tubular vibrators whose upper ends are connected to a base  The connector incorporates  an RTD to measure the sample liquid temperature  The vibrator assembly also contains a  piezoe
62. tion 1 g cm3  mode  CAL  4 Sets temperature of calibration solution 1 a   5 Sets temperature coefficient of calibration solution 1   x10 5 g cm3   C  6 Starts calibration 1  press SET key  CAL 1  displays density after calibration   7 Sets density of calibration solution 2 i g cm3  8 Sets temperature of calibration solution 2 OC  9 Sets temperature coefficient of calibration solution 2   x10 5 g cm3   C  A Starts calibration 2  press SET key  CAL 2  displays density after calibration   Maintenance 1  Cancellation of protection  enter 77  77  mode 2  Displays frequency  density  Hz   MAINT  3  Displays frequency  temperature  Hz  4  Checks output signal    5  Checks LED  press SET key    LED    flash 5 times   6  Sets low limit of the output range   g cm3  7  Sets high limit of the output range g cm3  8  Sets reference temperature of measuring liquid OC  9  Sets temperature coefficient of measuring liguid   x10 5 g cm3 C  A   B   C   D           Sets detector constant A  Sets detector constant B  Displays detector calibration coefficient C  span   Displays detector calibration coefficient D  zero                                                                                                                        1  Measuring mode  DSPL   Mode for normal operation    An analog output in this mode is density value at the reference temperature     Moreover  density value at measuring liquid temperature and temperature of measuring liquid are  digitally output  The density value 
63. to repair parts  consults with service personnel   Figure 5 3 Oscillation Amplifier Check Points       IM 12T03A01 02E    F0502 ai    4th Edition   Feb  23  2015 00       Customer  Maintenance Model VD6D  VD6DF  VD6DS    Parts List Liquid Density Detector                Item Part No  Qty Description   1     1 Terminal Box Cover   2 K9000SK 1 Cover   3 K9000CC 2 Plug  for Models VD6D and VD6DF   4 K9000DB 2 Plug  for Model VD6DS    5 L9818JB 4 Gasket  for Model VD6DS    6 G9303AR 1 O Ring   7 G9303BA 1 O Ring   8 G9303NE 2 O Ring  for Model VD6DS    9 K9000AK 1 Chain Assembly  for Model VD6DF   10 K9324PC 1 Desiccant  2 packs    11 K9000AC 1 Clamp  12 DO117XL A 2 U Bolt Assembly  13 L9826AG 2 Bracket  14 L9826AC 1 Bracket         Copyright 1999  1st Edition  1987  YK  CMPL 12T3E1 01E  YOKOGAWA 4 Subject to change without notice  5th Edition   Apr  2009  YK     Yokogawa Electric Corporation    Accessories Assembly                Item Part No  Qty Description  1 K9000PQ 1 Accessories Assembly  2 L9827AR 1 Brush  3 K9000NB 1 Injector  4 L9827AE 1 Allen wrench  5 E9135GZ 1 Spanner  6 K9324PC 1 Desiccant  2 packs   7 K9003NG 1 O Ring Set  used for item 6 and 7 in page 1        CMPL 1 2T3E1  01 E 5th Edition   Apr  2009  YK     Customer  Maintenance Mode  DMSG  Parts List Liquid Density Converter     Style D           Item Part No  Qty Description   1 K9005CA 1 Digital Board Assembly   2 K9005AA 1 Main Board Assembly   3 K9000YR 1 Cable Assembly  red    4 K9005YR 1 Cable Assembly  y
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Owner`s Manual - Northern Tool + Equipment    Kramer PIE Compressor User Manual  simulation du comportement thermique d`une cellule d`habitat  User`s Manual  岡崎市こども発達センター等整備運営事業事業契約書(案)(PDF形式    Copyright © All rights reserved. 
   Failed to retrieve file