1. MILLIPORE Automatic Sanitization Module User Manual Model TANKASMIN i ST pa Legal Information Notice Warranty The information in this document is subject to change without notice and should not be construed as a commitment by Millipore Corporation Millipore Corporation assumes no responsibility for any errors that might appear in this document This manual is believed to be complete and accurate at the time of publication In no event shall Millipore Corporation be liable for incidental or consequential damages in connection with or arising from the use of this manual We manufacture and sell water purification systems designed to produce pure or ultrapure water with specific characteristics uS cm T TOC CFU ml Eu ml when it leaves the water purification system provided that the System is fed with water quality within specifications and properly maintained as required by the supplier We do not warrant these systems for any specific applications It 1s up to the end user to determine if the quality of the water produced by our systems matches his expectations fits with norms legal requirements and to bear responsibility resulting from the usage of the water Millipore Corporation Millipore warrants its products will meet their applicable published specifications when used in accordance with their applicable instructions for a period of one year from shipment of the products MIL
2. on a Put the ASM onto the top of the Reservoir by following the steps below reservoir Action Remove the lid coming with the storage reservoir This lid will no longer be used Take the ASM and verify that the o ring 1s well positioned inside the lid Screw the ASM with its lid onto the reservoir until it 1s hand tight Since there is an o ring inside the lid there is no need to overtighten it Electrical Make sure the Water System is not powered when connecting the connection ASM electrical cable In addition make sure the ASM UV Lamp is inside the Reservoir e The ASM is connected on the side of the water system e Push the connector into the socket located on the left side of the system Continued on next page THE INFORMATION IN THIS PAGE IS MEANT TO BE USED BY A QUALIFIED MILLIPORE SERVICE REPRESENTATIVE ASM installation Continued Software The ASM must be activated into the water system software activation e In the Service Menu go in CONFIG and ASM Activation e Choose Yes and press Y HG Inlet Valve Status ld amp Password Buzzer UW 185 nm Activation UU 25d nm Activation Water Sensor Registering e The ASM UV lamp timer has to be registered ASM UV lamp In Standby Menu go in Maintenance And follow the screen instructions Install UU 1855 Lamp Sanitise Clean Install Quantum Suitability Tests Language This procedure should be The Millipore trained ASM UN insta
3. L RIGHTS RESERVED THIS BOOK OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHERS About Millipore Telephone Internet Site Address Manufacturing Site Document See the business card s on the inside cover of the User Manual binder www millipore com bioscience Millipore SAS 67120 Molsheim FRANCE Rev 0 06 2008 Safety Information Statement The ASM and other equipment must be operated according to the instructions in this manual In particular the hydraulic and electrical specifications must be followed and met It is important to use this equipment as specified in this manual using this equipment in a different manner may impair the safety precautions of the water system Symbols The light from an UV Lamp is very damaging to your eyes and to your skin Do not remove the ASM UV Lamp at any time unless its power cable is totally removed from the adjacent Water System The UV Lamp contains metallic Mercury Please dispose of any used UV Lamp in a manner that is environmentally safe Danger UV radiation This sticker is used to refer to a position on the Water System Cabinet or inside Danger This sticker is used to refer to a position on the Water System cabinet or inside Ground connection This sticker is used to refer to a position on the Water System Cabinet or inside Electrical danger This sticker is used to refer to a position on the Water System Cabi
4. LIPORE MAKES NO OTHER WARRANTY EXPRESSED OR IMPLIED THERE IS NO WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE The warranty provided herein and the data specifications and descriptions of Millipore products appearing in Millipore s published catalogues and product Literature may not be altered except by express written agreement signed by an officer of Millipore Representations oral or written which are inconsistent with this warranty or such publications are not authorized and 1f given should not be relied upon In the event of a breach of the foregoing warranty Millipore s sole obligation shall be to repair or replace at its option the applicable product or part thereof provided the customer notifies Millipore promptly of any such breach If after exercising reasonable efforts Millipore 1s unable to repair or replace the product or part then Millipore shall refund to the customer all monies paid for such applicable product or part MILLIPORE SHALL NOT BE LIABLE FOR CONSEQUENTIAL INCIDENTAL SPECIAL OR ANY OTHER INDIRECT DAMAGES RESULTING FROM ECONOMIC LOSS OR PROPERTY DAMAGE SUSTAINED BY ANY CUSTOMER FROM THE USE OF ITS PRODUCTS Continued on next page Legal Information continued Trademarks Copyright The M mark Millipore and Elix are registered trademark of Millipore Corporation All other trademarks are trademarks of their respective manufacturers 2008 MILLIPORE CORPORATION PRINTED IN FRANCE AL
5. cycles Validate to exit the ASM UV Lamp Schedule Continued on next page ASM programming Continued Example e This table shows an example of ASM UV Lamp Schedule display ASM UL LARP SCHEDULE Duration 13 min AS min 43 min BS min 15 min 43 min e Another example is shown where only cycle N 1 is programmed the ASM UV lamp 1s powered at 7 00 for 45 min ASM UL LARP SCHEDULE H Time Duration 1 AT Ae min Z 65 36 BA min 3 12166 BE mi q 14 68 Aa min a 21 66 BA min Total 43 min ASM UV It is possible to view the amount of time remaining before the water lifetime system prompts you to replace the ASM lamp This can be viewed in Consumable Status found in the READY Menu READY MEHO CONSLMABLES STATUS Water Quality Pretreatment Hare ASM Lamp Print Menu Progard Installed 26 Oct 2666 View Operation UW 239 am Lamp Call Millipore UW 185 am Lamp Service Tracking Quantum InFormation A18 UM Lamp Maintenance Alert messages UV lamp replacement Definition Alert messages are displayed when a maintenance action is needed within a few days or is needed immediately Refer to the Water system User Manual for more information on Alert messages Summary The Alert messages are listed below Description Action CHECK ASM UV LAMP UV lamp issue Contact Millipore REPLACE ASM UV LAMP UV lamp replacement will be needed IN xx DAYS in xx days Order new UV lamp Co
6. llation is performed by a Millipore service engineer confirms reoistered Hext trained service engineer the ASM UN installation by maintenance in YSB days Press to continue or pressing e Press to Press to exit ASM programming Manager Menu access ASM Lamp Schedule Access the Manager Menu in Standby Mode go in e Standby Menu e access the Manager Menu e go to Setup e and then to ASM Lamp Schedule LOGIN amp PASSWORD MARAGER MEHU Logins Change ID and Pass POD Flow Stop Passwords Date and Time Temp Comp Mode abcdeFohijk lmnop Set Points Flow Calibration qrstuvuxyzad2243 Units UW 254 nm Activation BTS9 _ W UU 185 nm Activation User Parameters Hetwork Settings The ASM Lamp Schedule software function is used to program the illumination cycles of the ASM UV lamp Definition e There are 5 cycles which allow a daily total of 45 minutes illumination e Each cycle can be programmed to start at a certain time This time can be set in a range of 24 hours from midnight 00 00 to 23 59 e The duration of each cycle can be adjusted from 0 to 45 minutes depending on the number of cycles The value 0 means that the cycle is disabled Programming e Press A or to adjust values e Press 2 or Q to navigate A e Press Vv to validate and to exit Set the time by changing the hour and minute Set the duration of the cycle Repeat the steps above for other
7. m An electrical board that powers the UV lamp A glass tube that protects the UV lamp UV lamp A lamp that emits UV rays Reservoir lid A plug that can be screwed on all Millipore reservoir and o ring opening Ballast MAHSMIHIN Reservoir l1 d p em Connector Continued on next page Product information Continued Weight and dimensions Electrical specifications Environmental specifications ASM welght The ASM weighs 1 kg ASM dimensions Specification Power cable The maximum length of the power cable 1s 5 m The table shows the value of each specification Specification 12 VA Supply Voltage to ASM 24 VDC The ASM is intended for indoor use only under the conditions stated below Specification Ambient storage temperature between 5 40 C below 3000 m Installation category Pollution degree THE INFORMATION IN THIS PAGE IS MEANT TO BE USED BY A QUALIFIED MILLIPORE SERVICE REPRESENTATIVE ASM installation UV lamp Before proceeding make sure the ASM is not electrically connected to the Water System The following procedure shows the installation of the UV lamp Step Action _ a screws ek rl connector onto the 4 pins of the UV Lamp Insert the UV lamp into the quartz sleeve Continued on next page THE INFORMATION IN THIS PAGE IS MEANT TO BE USED BY A QUALIFIED MILLIPORE SERVICE REPRESENTATIVE ASM installation Continued ASM
8. nce on it Ordering information Catalogue numbers The consumables catalogue numbers are found below Catalogue number Automatic Sanitization Module TANKASMIN 254 nm UV lamp ZLXUVLPLI 215 mm quartz sleeve ZF3000219 10
9. net or inside Table of Contents Product informato M it cicesicial 1 FURON ae E A A l SS a E E E A O A l COMPONIST iaE l W C1 MEAN CUMS HSI aa 2 PICCIRICAl pocitos ns 2 Enyironmcatal Speci CatOmsia 21 8 ccs adenc tess eta seu havatebaded a aes 2 PALS VE IHS CAN ATO Dis iccrsstscsucscctiavsctsctenculerncussiassacsduasseupeaieadeaseiehaacsneoevedsdaseunesiaadeeebedesentebicodede 3 A OANE E T ae eae E E EE E 3 SINO AE Odds 4 ars ico AAA T E O 4 DOM WV ARG AC ULV LOM Seis ass id 5 Resisterias ASM UVY TADA it 5 AS VE DFO SEAMING iia decos 6 Manager Moni ACCESS dt Do edo 6 ASM Lamp SCRE CUIC sier e A A A 6 Example ae 7 O 7 Mamina nee iia 8 AIETE MESSA E Sea ee O 8 UN lamp teplacoment adas 8 OA A AAA O 9 OFGCTING MformatON iia ii Erain as Ee aaa 10 Catalo ue NUM tati 10 Product information Function Purpose Components The Automatic Sanitization Module ASM 1s designed to be fitted on Millipore water storage reservoirs such as the 30 L 60 L or 100 L PE Reservoirs or the SDS 200 350 The ASM is electrically powered and controlled by an adjacent water system The ASM illumination cycles are programmed via the water system keypad and software The ASM makes use of the germicidal properties of ultra violet UV light at 254 nm The ASM prevents the formation of biofilm inside the Reservoir The table describes the main components of an ASM Connector A plastic part that connects the ASM directly on the water syste
10. ntact Millipore REPLACE ASM UV LAMP UV lamp replacement is needed Contact Millipore e The ASM UV lamp 1s normally replaced every 730 days e It is recommended to have a Millipore Field Representative replace the UV lamp Contact Millipore for a copy of these instructions if you want to replace the UV Lamp by yourself e If the UV lamp 1s replaced then the ASM UV Lamp needs to be registered via the Maintenance Software This resets the timer used to prompt the replacement of the ASM UV Lamp Install UY 185 Lamp Ea Sanitise Clean Install Quantum E Suitability Tests Install ALG Lamp Language Install L POD Pak 1 Manacer Menu Install 6 POD Pak 2 Install E POD Pak 3 This procedure should be The Millipore trained ASM UW installation is performed by a Millipore service engineer confirms reoistered Hext trained service engineer the ASM UY installation by maintenance in T36 days Press to continue or pressing e Press to Press to exit Continued on next page Maintenance Continued Quartz sleeve e Periodically visually check that the quartz sleeve is clean and free of cleaning anything that would interfere with the transmission of UV light e If necessary clean with an ethanol soaked cloth in order to maintain the effectiveness of the UV lamp e After cleaning with ethanol rinse the Quartz Sleeve with purified water A Disconnect the ASM from the water system before doing any maintena