1. TIGER is a portable gas detector that uses Photo ionization technology to detect a large range of Volatile Organic Compounds VOC s which can be dangerous from both a poisoning and explosive perspective The TIGER uses a Photo lonization Detector PID to measure gas concentrations The patented fence electrode technology minimises the effects of moisture and contamination avoiding the need for compensation Survey is the default mode of operation This mode is often used in applications such as Head Space Sampling and Leak Detection where several areas or Zones are to be monitored and readings data logged All sensor readings are real time measurements and alarm levels are set manually Like its predecessor the TIGER also has an intuitive graphical interface allowing easy access to instrument settings Two soft keys A and B can be configured to suit the user s application so many functions can be selected without entering the main menu structure This improves efficiency of use particularly with repeatable tasks TIGER PC PC Software maintains the intuitive look and feel by adopting the same graphical symbols TIGER PC also helps manage logged data files and multiple instrument settings in a clear and concise way The TIGER uses a Lithium ion battery pack that not only boasts an impressive running time and short charge time but also allows the battery pack to be replaced in potentially hazardous environments Field replaceable
2. Ensure that the Sensor Seal on the inside of the Sensor Cover is not disturbed Carefully lift the MiniPID Sensor from the Instrument Body ensuring that the two Inlet Outlet Seals remain in place in the Instrument Body Using the special Removal Tool MiniPID Stack 846216 provided locate its prongs in the slots in the side of the Mini PID Sensor body Using the forefinger to restrain the sensor the internal parts are spring loaded and careless disassembly will leave you hunting for the spring squeeze the tool to release the lamp housing The lamp may now be removed To replace the lamp or install a new one reverse the procedure ensuring the seals are all in place When replacing the Sensor Cover ensure that the markers are aligned correctly and that the cover is a snug fit The instrument MUST be re calibrated after fitting a replacement or cleaned lamp CAUTION Never refit a damaged lamp Unrivalled Detection Page 35 of 41 www lonscience com TIGER MANUAL lon Science Ltd Maintenance Lamp Cleaning The TIGER PID relies on an ultraviolet light source ionising VOC gases as they pass across the lamp window This process may result in a very fine layer of contamination appearing on the detector window that must be removed on a regular basis CAUTION The TIGER is a sensitive detector Internal components must be handled with clean hands and clean tools The TIGER lamp is fragile Handle with grea
3. TIGER Instrument User Manual V1 0 For Sales amp Service Contact geotecn 2650 E 40th Ave e Denver CO 80205 Phone 303 320 4764 e Fax 303 322 7242 1 800 833 7958 www geotechenv com Register your instrument online to receive your extended warranty Unrivalled Detection Www Ionscience com Register your instrument online for extended warranty Thank you for purchasing your lon Science product The standard warranty of your instrument can be extended to up to five years on Tiger and two years on other lon Science product To receive your extended warranty you must register your instrument online within one month of purchase terms and conditions apply Visit www ionscience com instrument registration Part number 861265 TIGER MANUAL Warnings lon Science Ltd USER MANUAL STATIC HAZARDS MATERIAL EXPOSURE SERVICING BATTERY CHARGING BATTERY REPLACEMENT BATTERY CONNECTION BATTERY USE FUNCTIONAL TEST USB CONNECTION Proper Use Type of safety protection USB CONNECTION Unrivalled Detection Read and understand this user manual completely before operating the PhoCheck Tiger instrument Do not use abrasive or chemical detergents to clean the PhoCheck Tiger instrument as this may reduce the antistatic properties of the materials used clean it using a damp cloth only The PhoCheck Tiger must not be exposed to atmospheres known to have an adverse effect o
4. TRES FIFMWArS A o e o la o dd 24 Downloading Data Logged Readings sise 26 SoOnWare Discla IMSS a e es ed o e ea 28 A Ode Ou ee da D a anaes ne a ae 29 Recharging Ballesta te entente A 29 Replacement Exchange of Battery Packs 4 30 Replacement of Non rechargeable Batteries in Battery Pack 31 DIAGNOSHCS si tendent ici 32 A sente en On nice un E E a tuuenecreesceacoaammoacuerounmenes 33 CAD AMOR doo 33 MENOZ AA A 34 PEIRE Fiter DISC COO 221 etc dae 34 PID Sensor Lamp Replacement and Cleaning occooccccncocncoccconcocnnonocononcnnonnnononnnonnnnononnnnnnnnonnnannnannnnnneninanos 35 Lamp A AA nn re ee 36 Tiger parts D Seen 37 Tiger MANS MDI decano 37 Frontend titer assembly nse ee ee 37 PCCOS SONICS A A A a 38 Instrument Warranty and Service ius een nn teen nan dense eue Eaa 39 A uaneess ee mencae 39 EE RER E E NS Nine ee ee eae 39 Contact Detalla 39 Technical Speciica MOM 40 MANUAL OG ur DEN 41 Unrivalled Detection Page 6 of 41 www lonscience com TIGER MANUAL lon Science Ltd Statements Responsibility for Use TIGER instruments detect a large range of gases which are potentially dangerous from both a poisoning and or an explosive perspective TIGER T instruments have many adjustable and selectable features allowing the detector to be used in a variety of ways lon Science Ltd can accept no responsibility for the incorrect adjustment of features that cause harm or damage to persons or property TIGER c
5. 846216 Removal Tool MiniPID Stack Used to remove Stack Pellet from the MiniPID Sensor A6 861205 Protective Removable Rubber Boot 8 Supplied with every TIGER this is a replacement Filter Cap 12 Black cap that covers the PTFE filter and accepts the 880202 Inlet Nozzle 14 861230 USB Cable Angled B Connector Supplied with every TIGER this is a replacement 861266 Leather instrument holster TIGER can be carried on a belt around the waist allowing hands free operation 861412 Leather instrument harness TIGER can be carried against upper body allowing hands free operation 861523 880201 A 861474 O rings for use with Inlet Nozzle 14 pack of 5 861476 A 31057 A 31063 A 861413 A 861414 A 861415 A 861240 Lithium ion battery pack intrinsically safe Replacement spare A 861241 Alkaline battery pack intrinsically safe 3 x AA batteries included Replacement spare Accessory box as supplied with the TIGER x5 PTFE Filter Disc A 861406 Flexible Probe Assembly 300mm replaces the standard probe internal pipe material PTFE A 880204 Sample cleaning compound x2 cotton buds Removable tool miniPID stack A 861472 Pack of 10 PTFE Filter Discs A 861511 Anti vibration charger cradle Allows wall mounting of vehicle charging includes charge cradle LA4TM600 MiniPID Lamp 10 6eV ppm LA4SM700 MiniPID Lamp 11 7eV For more information please visit www ionscience com Tiger
6. Off key for 3 seconds indicated by a 3 second countdown to turn TIGER off Note If the TIGER fails to shut down press the Escape Esc key once or twice to return to the main screen then press and hold the Enter On Off key again Batteries Check that your TIGER has sufficient charge for use The battery icon see Understanding the display screen should show at least two full segments Rechargeable Batteries TIGER instruments leave the factory with the Rechargeable Battery Pack A2 fully charged However prolonged periods of storage may result in the battery pack discharging We recommend charging the instrument for seven hours before use See the Batteries section of this manual pages 33 to 35 Alkaline Batteries An alkaline AA Battery Pack A3 is also supplied with the TIGER For more information please see the Batteries section of this manual Response Factor TIGER instruments leave lon Science Ltd factory calibrated on isobutylene and with a response factor of 1 0 The reason we calibrate the instruments with a response factor of 1 0 is because this is the response factor of isobutylene Set Alarm Levels We recommend that alarm levels are set to user specifications as soon as possible before the TIGER is used for the first time Please see the Using your TIGER section of the manual for details of how to set alarm levels TIGER PC The full functionality of your TIGER can only b
7. allow connection to the TIGER Instrument while in potentially hazardous atmospheres The TIGER instruments ingress protection rating is reduced to IP 20 when its battery pack is removed so avoid changing batteries in dusty or wet environments Two battery packs are available for use with the TIGER instrument A lithium ion Rechargeable Battery Pack A2 and non rechargeable AA Battery Pack A3 that may be fitted with 3 AA Alkaline batteries The rechargeable pack is recommended for normal operation with the Non rechargeable pack available when power is not available and the use of the instrument is needed The rechargeable pack is normally fitted as standard to the instrument when shipped Recharging Batteries Ensure the TIGER is charged for at least 7 hours before using it for the first time To ensure optimum charging the TIGER should be switched off during charging If left on the TIGER will take longer to charge but should not suffer any damage The TIGER should be charged in a non hazardous environment only To charge your TIGER first connect the Charger Cradle A4 to the mains and switch on A red light will indicate that the charger is ready Place the TIGER in the Charger Cradle such that the contacts on the TIGER are aligned with those in the cradle There is no need to remove the Instrument Boot 8 from the TIGER during charging During charging the charger will display an orange light A green light indicates t
8. and select Accessories Unrivalled Detection Page 38 of 41 www lonscience com TIGER MANUAL lon Science Ltd Instrument warranty and service Warranty Standard Warranty can be extended to up to 5 years on the TIGER when registering your instrument via our website www ionscience com instrument registration To receive your Extended Warranty you need to register within one month of purchase Terms and Conditions apply To register your TIGER instrument simply fill in the online form You will need to enter your instrument serial number to hand The serial number is located on the front label of the instrument towards the bottom end of the unit Beneath the bar code there will be a number beginning with T followed by 6 digits E g T 123456 You will then receive a confirmation email that your Extended Warranty Period has been activated and processed Full details along with a copy of our Warranty Statement can be found by visiting www ionscience com instrument registration Service lon Science Ltd also offers a number of service options for your TIGER that allows you to choose the instrument cover to best suits your requirements At lon Science Ltd we recommend that all of our gas detection instruments be returned for service and factory calibration once every 12 months Contact lon Science Ltd or your local distributor for service options in your area Find your local distributor by visiting www ionscie
9. from Permanently Off Permanently On On in low ambient Light or On For Limited Time The Backlight Timeout can be set from 1 to 30 seconds Sound The three icons represent key press alarm and crescendo Crescendo increases the sound as the upper alarm is approached Sound on each of these may be enabled or disabled by ticking or un ticking the box Sound volume is adjusted with the slider Gas This allows users to set the two alarm levels a lower and higher alarm to the user s preference In this option it also allows the user to select the response factor for the instrument Lock The TIGER may be locked in any configuration to avoid unauthorised changes by the operator A tick in the box locks the instrument Soft keys in Button A column are enabled and Buttons in column B disabled Sleep during PC connect will send the instrument into sleep mode while it is connected to the PC software The instrument will awaken as soon as it is disconnected from the PC software Clock Select 24 or 12 hour format using the radial buttons Set the time and date and tick the box on the right to set the time on your TIGER Alternatively tick the lower box to synchronise your TIGER with the time on your computer Zones Use this table to define and name up to 128 separate zones The name field is limited to eight characters including spaces Calibration Calibration Zero on Carbon Canister 1 0 RF 100 0 ppm Respons
10. the transparent Filter Clamp when the Inlet Nozzle is removed The Inlet Nozzle can be removed for cleaning or replacement by unscrewing it from the transparent Filter Clamp Refit the Inlet Nozzle using fingers only avoid using tools as this may damage the filter housing To ensure the assembly is sealed place a finger over the Inlet Nozzle to block the flow while the instrument is running A flow alarm should occur if sealed correctly PTFE Filter Disc 861221 The Filter Disc should be changed after every 100 hours of use This frequency should be increased for dusty or moisture laden environments or whenever the filter appears dirty when viewed through the transparent upper surface of the Filter Clamp Filter Disc changing should be conducted in a suitably clean environment with clean hands and equipment to avoid contamination of the new Filter Disc To change the Filter Disc unscrew the Filter Housing Cap lift off the Filter Clamp and O Ring and lift the Filter Disc from the Instrument Body Carefully place a new Filter Disc into the Instrument Body Under no circumstances should a Filter Disc be used once it has been removed Replace the Filter Clamp ensuring the locating lugs are correctly positioned in the cut outs in the Instrument Body and that the O Ring is correctly seated Replace the Filter Housing Cap Do not over tighten Unrivalled Detection Page 34 of 41 www lonscience com TIGER MANUAL lon Science Ltd Main
11. to the main screen Unrivalled Detection Page 15 of 41 www lonscience com TIGER MANUAL i lon Science Ltd Using your Tiger High alarm Att 3rd Screen SPAN 1 ESPAN 1 gas concentration set in TIGER PC SPAN 2 ESPAN2 gas concentration set in TIGER PC 2nd screen A D 2 Digital number from PID Bootloader Bootloader Version 4th Screen IRN IRN Internal reference number Battery E Type and voltage Data log memory available E Unrivalled Detection Page 16 of 41 www lonscience com TIGER MANUAL lon Science Ltd gt ps a RCE Tiger PC software instillation The Tiger PC has its own specific software package used to set up the instrument and download logged data The software is available to download from the ion science website Once the software is downloaded from the website double click on the zip file and follow the installation procedure Unrivalled Detection Page 17 of 41 www lonscience com TIGER MANUAL lon Science Ltd Using Tiger PC software Connecting TIGER to a PC 1 Double click the TIGER icon on your desktop and open TIGER PC The Home page should appear Instrument s T000000461988 i T000000A9A7E0 TIGER TIGER LTPC software For technical enquiries please call 44 0 1763 208503 Unrivalled Detect www ionscience com 14 43 15 Connect Unrivalled Detection Page 18 of 41 www ionscience com TIG
12. 000000461988 Configuration L ar Session No Zone Start Date Gas Firmware f d N f 4 i 0 001 01 05 2015 09 08 37 Gas 1 0 Session 0 Service Snapshots Processing datalog done Expand the datalog folder and select the session of interest Details of the data collected during that session are presented in numerical and graphical form Use the menu function at the top right hand corner of the window to print export or delete the data CAUTION The delete function deletes all logged data from your TIGER Ensure all valuable data is exported to your PC before selecting Delete Unrivalled Detection Page 26 of 41 www lonscience com TIGER MANUAL g lon Science Ltd SAD S jet Tiger PC software The Snapshots Screen A snapshot records the settings and calibration data on your TIGER at any particular point in time The Snapshots screen displays a list of all those stored on your PC Click Delete to remove a selected snapshot To restore the stored settings to your TIGER first ensure that your instrument is running correctly and connected to your PC as detailed previously Ensure that your instrument is in normal Survey mode is not in an alarm condition and that no data logging or Health and Safety readings are being calculated Click Restore against the relevant snapshot Home Help Instrument s 1000000461988 Configuration 30 04 2015 16 01 52 Delete Restore Logo Firmw
13. ER MANUAL lon Science Ltd Tiger PC software The Help Screen This screen will show you if a new version of PC software is available You can then install the upgrade if required Home Instrument s 7000000461988 TigerPCLT fi Contact Technical Version 1 1 0 1 5 t Copyright 2015 lon Science Limited Al rights Reserved Registered to lon Science Ltd Feature Name Status Valid Until Write TigerPC to Restart TigerPC for memory stick selection to take effect 08 34 41 Connect LT 2 Click on Instrument s If your TIGER has previously been connected to your PC your instruments IRN Internal Reference Number will appear Note If while using this software your TIGER should inadvertently be turned off or disconnected from your computer click on Instrument s again and proceed as below 3 Switch TIGER on and when fully running up connect to a USB socket on your PC using the USB cable supplied If the Found New Hardware screen appears follow the prompts to install your instrument on your PC 4 Click on Read Your instrument number will appear under instrument s if it was not there already and the Instrument Summary page will appear Unrivalled Detection Page 19 of 41 www lonscience com TIGER MANUAL lon Science Ltd Tiger PC software pea Please connect an instrument to a USB po
14. NUAL lon Science Ltd Using your Tiger Instrument functionality is broken into two parts Application and Setup Application settings are initially selected via the soft keys A and B Setup functions like Backlight Sound Calibration and Alarm setting are adjusted within Options Many of the screens have a 2 second time out which returns to the main screen if no other keys are pressed 2 On To switch the TIGER on press the Enter On Off key once Startup screen 1 appears showing the TIGER logo Startup screen 2 contains variable text sent from the TIGER PC T configuration screen The lower half of the screen contains the instrument IRN Internal Reference Number and firmware version The third screen shows TIGER checking that the lamp has struck When OK appears the working screen will follow If the lamp fails to strike turn the TIGER off wait for 30 seconds and try again If the problem persists change the lamp or contact lon Science Ltd or your distributor Off To switch the TIGER off press and hold the Enter On Off key 3 second count down takes place before the instrument shuts down During this count down the instrument activates the upper alarm viz Red LED s flash and audible alarm sounds This is done to alert the user to avoid accidental switch off ZRO Zeroing HE Pressing the Zero soft key presents two zero options selected with the Up or Down keys The upper symbol represen
15. alkaline battery packs non rechargeable are also available for when power to recharge is unavailable The TIGER has true USB capability so the instrument can be connected directly to a PC via a standard USB cable It also offers fast data download A loud audio output vibration and bright LED s indicate alarm conditions Orange and Red LED s indicate High and Low conditions respectively This colour scheme is used within TIGER PCH to offer clear indication of alarm levels within logged data Unrivalled Detection Page 8 of 41 www lonscience com TIGER MANUAL lon Science Ltd Getting started Thank you for choosing TIGER from lon Science Ltd We hope that your TIGER will give you many years of active and trouble free service The Manual lon Science Ltd recommend that you familiarise yourself with this manual before using your TIGER The Statements and Introduction to TIGER sections contain important information which should be read before you turn your TIGER on for the first time Inlet Nozzles Ensure that the inlet nozzle supplied is fitted to your TIGER Details on how to fit and change inlet nozzles is detailed in the Maintenance section If the supplied inlet nozzle is unsuitable for your application details of alternative inlet nozzles can be found in the Maintenance section as well Turn On Press the Enter On Off key once to turn TIGER on Turn Off Press and hold the Enter On
16. an be used as a personal safety device It is the user s responsibility to respond appropriately to an alarm situation Inadequate performance of the gas detection equipment described in this manual may not necessarily be self evident and consequently equipment must be regularly inspected and maintained lon Science Ltd recommends that personnel responsible for equipment use a regime of regular checks to ensure it performs within calibration limits and that a record be maintained which logs calibration check data The equipment should be used in accordance with this manual and in compliance with local safety standards IMPORTANT It is essential that the TIGER is always used with a supplied 0 5 micron PTFE Filter Disc fitted to the front of the instrument Without a filter particles of debris and dust can be drawn into the detector inhibiting the function of the instrument These filters are consumable and should be changed after every 100 hours of use The frequency of replacement should be increased for dusty or moisture laden environments Filters are available from your distributor or at www ionscience com Quality Assurance TIGER has been manufactured in compliance with 1S09001 2008 which ensures that the equipment supplied to our customers has been designed and assembled reproducibly from traceable components and leaves lon Science Ltd calibrated to stated standards Disposal Dispose of TIGER its components and any used batterie
17. are Datalog Service Snapshots Processing datalog done Be aware that this process will replace all setup and calibration files On the restore screen click Yes When complete click Close and restart your TIGER Your TIGER will now be restored to the settings and calibration data stored at the time of that snapshot The snapshot menu also allows review of stored data when an instrument is not attached Expand the menu until the latest or relevant snapshot is displayed Double click on the snapshot and all the data stored in that snapshot can be accessed Unrivalled Detection Page 27 of 41 www lonscience com TIGER MANUAL lon Science Ltd J ao S le Tiger PC software Software Disclaimers Termination of Software Licence This License is effective until terminated This License will terminate automatically without notice from lon Science Ltd if you fail to comply with any provision of this License Upon termination you agree to destroy delete or purge the written materials and all copies of the SOFTWARE including modified copies if any Disclaimer of Warranty The SOFTWARE and accompanying materials including the user s manual are provided as is without warranty of any kind including the implied warranties of merchantability and fitness for a particular purpose even if lon Science Ltd have been advised of that purpose Furthermore lon Science Ltd does not warrant guarantee
18. ational safety and environmental requirements Unrivalled Detection Page 31 of 41 www lonscience com TIGER MANUAL lon Science Ltd Diagnostics Basic faults or diagnostics are presented as symbols Should a fault occur most can be corrected by pressing Enter or Esc to clear the fault message All fault conditions cause the TIGER to alarm Pump failure Pump blocked or pump failure Ry The flow of gas through the instrument has fallen below 100cc minute Check the probe and filter for signs of blockage Water or dirt in the probe a bent probe dirty filter on the inlet or blockage of the exhaust finger over hole on the back can all cause low flow If the blockage can be removed press Esc to clear the alarm If the fault persists send the instrument to your distributor for service Battery dead Battery low or battery failure The TIGER will shut down when the battery level falls below 2 Recharge the battery as instructed in the Batteries section of this manual ensuring that all connections are sound and the indicator lights on the charger are in order If the battery fails to charge fit another battery pack if available If using alkaline batteries replace them If the fault persists send the instrument and charger to your distributor for service Lamp failure The PID lamp has failed to strike illuminate this may occur at switch on or during use Switch the TIGER off and replace the lamp See the Maintenan
19. ce section Memory full Memory cannot receive more data The data log memory is full This will only happen if the Log Full box is set to alarm on the TIGER PC configuration screen Press the Esc key to continue but the TIGER will no longer continue to log data Select recycle in TIGER PCH and the TIGER will overwrite the oldest data and no alarm will be raised System error Total system failure t 68212 The instrument s firmware is corrupted In the unlikely event of this message appearing contact lon Science Ltd or your nearest authorised service centre Unrivalled Detection Page 32 of 41 www lonscience com TIGER MANUAL lon Science Ltd Maintenance Calibration lon Science Ltd recommends an annual service and calibration for users who require a traceable calibration During this service the lamp and detector are brought back to factory specifications and new Factory Calibration data is stored Due to the linear output of the lon Science PID detector a two point calibration is often adequate TIGER scales its linear output across a ZERO level clean air reference and the SPAN 1 user defined gas concentration TIGER offers the options of Factory Calibration or Custom Calibration Factory Calibration is set by lon Science Ltd during instrument manufacture or on re calibration Custom Calibration can be set by the instrument user For Factory Calibration contact lon Science Ltd or yo
20. ct Custom Calibration i and press Enter to confirm On confirming the selection the user is presented with a 15 second count down Remove both caps from the Carbon Filter Assembly A 31057 and then fit onto the probe of your TIGER Press Enter to start the zero countdown At the end of the count down a tick v will appear indicating that the Zero has been accepted Disconnect the Carbon Filter Assembly and replace the cap ends The useful life of the Carbon Filter Assembly will be shortened if the Carbon Filter Assembly is open to atmosphere for prolonged periods Unrivalled Detection Page 33 of 41 www lonscience com TIGER MANUAL i lon Science Ltd Maintenance Press Enter again and the gas and concentration for Span 1 previously set up in TIGER PCH are displayed along with a 15 second countdown ESPANA Attach the Span 1 gas using the Calibration Adaptor 861476 supplied in the Accessory Box A 861267 see Accessories and press Enter to start the Span 1 countdown At the end of the count down a tick v will appear indicating that the Span 1 has been accepted For two point calibration press Enter and the calibration is complete Inlet Nozzle Should the Inlet Nozzle become cantaminated or damaged replacement Inlet Nozzles may be obtained from your distributor or from lon Science Ltd Please note that a small O Ring at the base of the Inlet Nozzle ensures the Inlet Nozzle is sealed this can bee seen in
21. e Ne retirez pas le couvercle du capteur dans les zones dangereuses N ouvrez aucune partie du PhoCheck Tiger dans une zone dangereuse sauf pour le remplacement de la batterie L entretien du PhoCheck Tiger doit se d rouler dans un environnement non dangereux et tre confi uniquement des centres de service agr s lon Science Ne proc dez pas a l entretien de l instrument lorsqu il est sous tension Otez la batterie avant l entretien Le remplacement de composants risque de nuire la s curit intrins que Chargez le PhoCheck Tiger et sa batterie lithium ion uniquement dans un environnement non dangereux Ne remplacez jamais de piles alcalines dans un endroit dangereux ou potentiellement explosif N utilisez que des piles alcalines Duracell Procell MN1500 Les batteries lithium ion et les piles alcalines du PhoCheck Tiger ont t sp cialement concues pour permettre la connexion a l instrument PhoCheck Tiger dans des atmospheres potentiellement dangereuses Assurez vous que les connexions lectriques sont propres et intactes avant la connexion L indice de protection des instruments PhoCheck Tiger tant ramen a IP 20 lors du retrait de la batterie evitez de remplacer les batteries dans des environnements poussiereux ou humides N utilisez que les batteries fournies avec le PhoCheck Tiger Ne remplacez jamais des piles alcalines primaires dans un endroit dangereux ou potentiellement explosif N utilisez que des pile
22. e Factor SPAN 1 Soft Menu The various functions may be allocated to the soft buttons in any order of preference by dragging and dropping the icons see page 12 Zeroing This function allows either an assumed constant zero or for the zero to be tracked compensating for the shift in the zero of the PID lamp during operation The function may be enabled or disabled by ticking or un ticking the box Constant or tracked zero is selected by use of the radio buttons associated with the icons There are various options for zeroing a TIGER The symbols below are found on the configuration screen of TIGER PC Unrivalled Detection Page 22 of 41 www lonscience com TIGER MANUAL lon Science Ltd Top J IND E pei Tiger PC software Zero at switch on When selected the TIGER will automatically set its zero reading Zeroing based on ambient air When deselected the o PO A Fixed zero D When selected TIGER uses the fixed calibration Zero If used in conjunction with the Zero at switch on feature the instrument will zero at switch on and then remain at that level instrument will use Its calibration zero Zero tracking When selected the Zero level will move negative if a cleaner ambient air is detected This ensures 0 0 ppm is displayed in clean air and ensures sub ppb levels are always detected Fixed zero Log Full If you select the bell symbol your TIGER will alarm when the memory logis full Alternati
23. e realised through the TIGER PC software lon Science Ltd recommend that you load the software supplied with your instrument and set up your TIGER according to the instructions in the TIGER PCH Software section of this manual Unrivalled Detection Page 9 of 41 www lonscience com TIGER MANUAL lon Science Ltd Understanding the keypad General Description The keypad comprises two soft keys A and B Up and Down arrow keys an Escape Esc key and an Enter On Off button In general setup and application settings are selected and adjusted via the soft keys options are selected by the arrow keys and confirmed by the Enter key single press is used as a switching operation A continuous press is used to adjust numbers or change gas selections by automatically rolling Keypad Function Descriptions Soft keys A and B rely on graphical prompts on the display to indicate their functionality Note Pressing both soft keys together switches the flashlight torch on and off Up and Down keys are used to adjust settings and navigate through the menu structure Enter On Off key is used to accept adjustments and select functions also to turn the TIGER on and off Escape Esc key is used to abort an adjustment or exit from a menu Unrivalled Detection Page 10 of 41 www lonscience com TIGER MANUAL lon Science Ltd Understanding the display screen Screen Display The display is divided into fo
24. g 1 6 Ib Instrument Anti static PC ABS Polycarbonate Acrylonitrile Butadiene Styrene Rubber Boot Anti static TPE Thermoplastic Polyolefin Elastomeric www lonscience com Page 40 of 41 TIGER MANUAL lon Science Ltd p ree c 2 q EES jew Manual log Instrument PC Manual version Amendment Issue Date E Firmware Software Manual created 16 06 15 V0 5 02 V1 1 0 2 Unrivalled Detection Page 41 of 41 www lonscience com
25. ganic compounds PhoCheck TIGER Il 1G Exia IIC T4 Ga ITSOSATEX26890X ExiallC T4 Ga IECEx ITS 10 0036X Class Division 1 Groups A B C D T4 Ta 15 C to 45 C for Li ion Battery Ta 15 C to 40 C for Alkaline Battery I Required Coding Ex II 1 G Ex ia IIC T4 Ga LISTED ITSO9ATEX26890X IECEx ITS 10 0036X intertek gt k nterte Intertek 0359 Chester UK 3193491 Conforms to UL Std 913 61010 1 Certified to CAN CSA Std C22 2 No 61010 1 Patents incl US Pat 7 046 012 7 821 270 EU Pat 1474681 E lon Science Ltd Cambs UK SG8 7UJ www ionscience com 101706597MAN 001 3193491 Conforms to UL Std 913 61010 1 8 Certified to CAN CSA Std C22 2 No 61010 1 Electrical apparatus for explosive gas atmospheres General requirements 00000000 Explosive atmospheres Equipment protection by intrinsic safety I Electrical equipment for measurement control and laboratory use EMC requirements Group 1 Class B equipment emissions section only Electrical equipment for measurement control and laboratory use EMC requirements Industrial location immunity immunity section only Electromagnetic compatibility Electrical apparatus for the detection and measurement of combustible gases toxic gases or oxygen Immunity Type 2 industrial environments Class A Code of Federal Regulations 15 Subpart B Radio Frequency Devices Unintentional Radiators Quality Management Sys
26. hat charging is complete Discharged The battery icon on the TIGER will display the level of charge Fully Charged Note 1 Only use the Charger Cradle supplied with your TIGER to charge your instrument 2 lon Science Ltd recommends keeping your TIGER on charge at all times when not in use as batteries can lose power over time Unrivalled Detection Page 29 of 41 www lonscience com TIGER MANUAL lon Science Ltd Batteries Replacement Exchange of Battery Packs To replace the Rechargeable Battery Pack A2 proceed as follows a Ensure TIGER is switched Off b Remove the Instrument Boot 8 from your TIGER this is most easily achieved by starting at the front end of the instrument step 4 below C Release the clip at the rear end of the instrument and lift the Rechargeable Battery Pack A2 away from the Instrument Body A1 coincidently sliding it slightly backwards d Reverse the procedure to install the replacement battery pack ensuring that the Boot is a snug fit around the front of the instrument and the lights are not obscured e Charge the TIGER for 7 hours before use Item numbers in bold refer to the illustrations and parts list on page 41 Unrivalled Detection Page 30 of 41 www ionscience com TIGER MANUAL lon Science Ltd Batteries Replacement of Non rechargeable Batteries in Battery Pack WARNING BATTERY REPLACEMENT Never replace primary Alkaline batte
27. iger PC software The Firmware Screen This screen displays the current firmware version installed on your TIGER It also provides a facility to download updated versions to your TIGER At this point you can also check the View Release Notes information which is a description of the changes in the new version as shown here To download a firmware update connect TIGER to your computer as detailed previously Ensure that your instrument is in normal Survey mode is not in an alarm condition and that no datalogging or Health and Safety readings are being collected Check that the battery pack is sufficiently charged at least two bars IMPORTANT The Firmware upgrade process will delete all data from the instrument To avoid loss of any data during the upgrade process take a snapshot of the instrument before proceeding The snapshot can be re installed after the update process is complete Home Help Instrument s T000000461988 s Firmware Version of Instrument V0 5 01 Configuration Logo Firmware selected View Release Notes Datalog 0 5 1 Service t Processing datalog done ee A When you are ready to send the new Firmware to the instrument Select Send to Instrument You will now see a message that will warn you that all the data will be lost on your instrument If this is acceptable to continue confirm with Yes A WARNING message will appear to continue select Yes Un
28. is cotton bud to polish the PID lamp window Use a circular action applying light pressure to clean the lamp window Never touch the lamp window with fingers 3 Continue polishing until an audible squeaking is made by the cotton bud with compound moving over the window surface usually within fifteen seconds 4 Remove the residual powder with short blast of air from the can of air duster 5 The instrument MUST now be re calibrated Unrivalled Detection Page 36 of 41 www lonscience com TIGER MANUAL lon Science Ltd Tiger parts Tiger main assembly P i pan f f po CLA a3 a2 6 5 Front end filter assembly 6 DS CIC Instrument Body 861274 Rechargeable Battery LE instrument Bot 881205 1 e feensorsea eer 1 10 niet Outlet Sea 881215 2 Ci Fier Housing can e6218 1 ir LU CL E IL A EL 16 Probe Seal asoma TT Unrivalled Detection Page 37 of 41 www ionscience com TIGER MANUAL lon Science Ltd Accessories lon Science Ltd has developed an exclusive range of high quality accessories to compliment the TIGER Please see a selection of these below Part Number Accessory Description 2 1mm DC Jack Cigar Lighter Plug Car Charger Lead Allows the 1 jawu 01 Charger Cradle A4 to be powered from a standard vehicle cigar lighter 12 volt only PID Sensor Seal 9 Seal between the MiniPID Sensor A6 and the 861214 Sensor Cover A5
29. larm sound crescendo or percentage volume Press Enter to turn selection on or off For percentage volume press Enter again to confirm selection The frame will flash Use Up and Down keys to change the percentage volume and press Enter to confirm the chosen value Press Esc twice to return to the main screen E Calibration Only two options are presented Factory Calibration and Custom Calibration Select the preferred option with the Up and Down keys and confirm with Enter Whichever calibration standard is selected the TIGER will operate against that calibration Readings taken against unsatisfactory calibrations may be unreliable Factory Calibration Not for operator use Return to lon Science Ltd or your distributor for calibration see the Maintenance section of the manual page 37 Custom Calibration see the Maintenance section of the manual page 32 A Alarms Upper att and Lower at alarm settings are displayed Select the subject alarm using the Up and Down keys and press Enter The arrow s on the chosen alarm will flash Adjust the level using the Up and Down keys and press Enter Repeat if necessary for the other alarm Press Esc twice to return to the main screen Note The Lower alarm setting must never be greater than the Higher alarm i Information Pressing the Information soft key allows access to a number of other information screens Use the Down key to scroll through the screens Press Esc to return
30. n Thermoplastic polyolefin or Anti static PC ABS Do not remove the sensor cover in the hazardous area No part of the PhoCheck Tiger may be opened in a hazardous area except for replacement of the battery pack The PhoCheck Tiger must be serviced in a Non Hazardous environment and by lon Science authorised service centres only Do not service the instrument live Remove battery pack before servicing Substitution of components may impair intrinsic safety Charge PhoCheck Tiger and its Lithium ion battery packs in a Non Hazardous environment only Never replace primary Alkaline battery cells while in a potentially explosive or hazardous location Use only Duracell Procell Alkaline batteries MN1500 The PhoCheck Tigers Lithium ion and Alkaline battery packs have been specially designed to allow connection to the PhoCheck Tiger Instrument while in potentially hazardous atmospheres Ensure all electrical connections are clean and undamaged before connection The PhoCheck Tiger instruments ingress protection rating is reduced to IP20 when its battery pack is removed so avoid changing batteries in dusty or wet environments Use supplied battery packs only on the PhoCheck Tiger Never replace primary Alkaline battery cells while in a potentially explosive or hazardous location Use only Duracell Procell Alkaline batteries MN1500 The PhoCheck Tiger must be functionally checked prior to entering a hazardous area after every occasion when a co
31. nce com Contact Details UK Head Office lon Science Ltd The Way Fowlmere Cambridge SG8 7UJ UK Tel 44 0 1763 207206 Fax 44 0 1763 208814 Email info ionscience com Web www ionscience com Unrivalled Detection USA Office lon Science LLC 33 Commercial Drive Waterbury VT 05676 USA Tel 1 802 244 5153 Fax 1 802 244 8942 Email info ionscienceusa com Web www ionscience com Page 39 of 41 German Office lon Science Messtechnik GMBH Laubach 30 Metmann Neandertal 40822 GERMANY Tel 49 2104 14480 Fax 49 2104 144825 Email info ism d de Web www ism d de www lonscience com TIGER MANUAL lon Science Ltd J ao S le Technical specification Response time Detectable Range Resolution Accuracy Linearity Battery Data log Alarm visual Alarm audible Flow Rate Temperature Dimensions Weight Materials Unrivalled Detection T90 lt 2 second 0 1 ppm 0 ppm 5 000 ppm 0 1 ppm 5 displayed reading one digit 5 displayed reading one digit Lithium ton 24 hours Alkaline Duracell Procell 8 5 hours MN1500 Including date time 80 000 Flashing Red and Amber LED 95 dBA 30 cm 2 220 ml min in Ambient conditions Operating 20 to 60 C 4 to 140 F Storage 25 to 60 C 13 to 140 F Certified to 15 to 45 C 5 to 140 F Instrument 340 x 90 x 60 mm Instrument 0 720 k
32. nnection has been made to the USB port The instrument must complete its start up routine and display legible readings If the LCD display fails to show an intelligible and uncorrupted display the instrument must not enter a hazardous area The USB port can only be used in a Non Hazardous environment If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Intrinsically safe The USB port can only be used in a Non Hazardous environment www lonscience com Page 3 of 41 TIGER MANUAL Avertissement lon Science Ltd MODE D EMPLOI RISQUES LIES AUX PROPRIETES NTISTATIQUES EXPOSITION AUX MATERIAUX ENTRETIEN CHARGE DE LA BATTERIE REMPLACEMENT DES PILES CONNEXION DES PILES DE LA BATTERIE UTILISATION DES PILES DE LA BATTERIE ESSAI DE FONCTIONNEMENT CONNEXION USB Utilisation appropri e Type de protection de s ret CONNEXION USB Veuillez lire enti rement et comprendre ce mode d emploi avant d utiliser l instrument PhoCheck Tiger N utilisez pas de d tergents abrasifs ou chimiques pour nettoyer l instrument PhoCheck Tiger au risque de r duire les propri t s antistatiques des mat riaux utilis s Employez uniquement un chiffon humide N exposez pas le PhoCheck Tiger des atmosph res connues pour avoir un effet ind sirable sur la polyol fine thermoplastique ou le PC I ABS antistatiqu
33. or make any representation regarding the use or the results of the use of the SOFTWARE or written materials in terms of correctness accuracy reliability current revision or otherwise lon Science Ltd specifically does not warrant the SOFTWARE after you assume operation If the SOFTWARE or written materials are defective you not lon Science Ltd or its dealers distributors agents or employees assume the entire risk and costs of all necessary servicing repair or correction except as stated below Defective CD or Memory Stick As the only warranty under this Agreement lon Science Ltd warrants to the original Licensee only that the CD s or Memory Stick s on which the software is recorded is free from defects in materials and workmanship under normal use and service for a period of ninety 90 days from the date of delivery as evidenced by a copy of the Receipt This limitation will apply where allowed Replacement of Defective CD or Memory Stick lon Science Ltd entire liability and the original Licensee s exclusive remedy under this agreement are at the discretion of lon Science Ltd to either a return of payment as evidenced by a copy of the Receipt or b replacement of the disk or memory stick that does not meet lon Science Ltd limited warranty and which is returned to lon Science Ltd with a copy of the Receipt If failure of the disk or memory stick has resulted from accident abuse or misapplication lon Science Ltd will have no re
34. rivalled Detection Page 24 of 41 www lonscience com TIGER MANUAL lon Science Ltd Tiger PC software IMPORTANT Do not attempt to operate your TIGER during the update process The firmware on your TIGER will be updated and the new version will be shown on the Firmware screen WARNING DO NOT DISCONNECT INSTRUMENT 15 47 27 FirmwareWrite bar at the bottom of the screen will indicate the progress of the update IMPORTANT Do not disconnect the TIGER at this point Please follow the next part before disconnecting the TIGER After the PC Software has completed sending the upgrade to the TIGER the instrument will then have to finish the upgrade process The TIGER will flash the two torch LED s and the screen on the Tiger will remain blank for approximately 30 Seconds A bar will then appear and travel across the screen A message will be displayed as follows verifying file system The Firmware is now installed on to the instrument and will start up automatically Unrivalled Detection Page 25 of 41 www ionscience com TIGER MANUAL lon Science Ltd Tiger PC software Downloading Data Logged Readings Connect your TIGER to your PC as described previously Any fresh data logged readings will be downloaded as the software reads your TIGER a Go to the datalog screen and a list of the datalog sessions is presented Home Help Instrument s 1
35. rt and press Read T000000461988 bu Firmware Service 1000000461988 Delete Processing datalog done pe Home i Help Instrument s T000000461988 a Configuration Logo Firmware Datalog i Service nstrument T000000461988 3 Snapshots Summary pe Features OK Config OK Logo File not found Factory Cal OK User Cal none TigerSelect Cal none Firmware OK Datalog OK Would you like to create a snapshot of your instrument This may be used in future to recover your instrument Processing datalog done wm 2 Unrivalled Detection Page 20 of 41 www ionscience com TIGER MANUAL lon Science Ltd Tiger PC software 5 This page gives the current status of your TIGER If the snapshot screen appears click Yes You can delete the snapshot later if you wish see TIGER PC Software Tiger PC Would you ike to create a snapshot of your instrument This may be used in future to recover your instrument The Configuration Screen Use this screen to configure your TIGER more borage Sot Meru ec A 4 O a Sutton 6 hate TI000004 908 1 niorma on FE Optera Cor e eton 2 Detain me Zero Logo Sand er Response Factor D Zone Uno DAT ls o Serica vame SAO a 09 44 53 Comfigivrte Done m Unrivalled Detection Page 21 of 41 www ionscience com TIGER MANUAL lon Science Ltd Top J IND E pei Tiger PC software Backlight Choose
36. ry cells while in a potentially explosive or hazardous location Use only Duracell Procell Alkaline batteries MN1500 a Ensure TIGER is switched Off b Remove the AA Battery Pack A3 as in b and c on page 34 C Remove the screws retaining the battery cover and lift it off This reveals a set of 3 x AA cells contained in a removable battery holder d Remove the battery holder e Replace the exhausted batteries Use Duracell PROCELL Alkaline batteries MN1500 f Check all batteries have the correct polarity before replacing the battery holder g Assemble the battery holder into the AA Battery Pack A3 refit the battery cover and fix in place using the securing screws Replace the Instrument Boot 8 ensuring that it is a snug fit around the front of the Instrument Body A1 and the lights are not obscured CAUTION Fitting batteries or connecting the battery pack with the wrong polarity may result in permanent damage to the instrument CAUTION Non rechargeable batteries must not be replaced in the field Batteries should be loaded into the battery holder in a safe environment only Only the assembled AA Battery Pack A3 may be changed in the field CAUTION Do not make any connections to the USB port of this instrument whilst in a hazardous area Note When loading batteries check for correct battery polarity before connecting them Note Dispose of used batteries in accordance with all local and n
37. s alcalines Duracell Procell MN1500 Le PhoCheck Tiger doit subir un essai de fonctionnement avant de p n trer dans une zone dangereuse chaque fois qu une connexion a t effectu e au port USB L instrument doit terminer sa routine de d marrage et afficher des valeurs lisibles Si l cran LCD n affiche pas de donn es correctes et intelligibles n utilisez pas l instrument dans une zone dangereuse Le port USB peut uniquement tre utilis dans un environnement non dangereux Si l quipement est utilis d une mani re non sp cifi e par le fabricant la protection fournie pat l quipement peut tre alt r e S curit intrins que Le port USB peut uniquement tre utilis dans un environnement non dangereux Unrivalled Detection www lonscience com Page 4 of 41 TIGER MANUAL Declaration of conformity lon Science Ltd Manufacturer Product Product description Directive 94 9 EC Tamb 15 C to 45 C with Lithium ion Battery pack Tamb 15 C to 40 C with Alkaline Battery pack Certificate Number Notified body Report number Intertek Standards BS EN 60079 0 2012 BS EN 60079 11 2012 BS EN61326 1 2006 BS EN61326 1 2006 BS EN50270 2006 CFR 47 2008 Other Standards BS EN ISO 9001 2008 BS EN 80079 34 2011 lon Science Ltd The Way Fowlmere Cambridge UK SG8 7UJ TIGER Intrinsically safe photo ionisation gas detector for detecting volatile or
38. s in accordance with all local and national safety and environmental requirements This includes the European WEEE Waste Electrical and Electronic Equipment directive lon Science Ltd offers a take back service Please contact us for more information The TIGER field case material is recyclable polypropylene Calibration Facility lon Science Ltd offers a calibration service including the issue of certification confirming calibration with equipment traceable to national standards A TIGER calibration kit is available from your distributor or service centre or at www ionscience com lon Science Ltd recommends annual return of all instruments for yearly service and calibration Legal Notice Whilst every attempt is made to ensure the accuracy of the information contained in this manual lon Science Ltd accepts no liability for errors or omissions or any consequences deriving from the use of information contained herein It is provided as is and without any representation term condition or warranty of any kind either express or implied To the extent permitted by law lon Science Ltd shall not be liable to any person or entity for any loss or damage which may arise from the use of this manual We reserve the right at any time and without any notice to remove amend or vary any of the content which appears herein Unrivalled Detection Page 7 of 41 www lonscience com TIGER MANUAL lon Science Ltd J ao S le Introduction to Tiger
39. sponsibility to either replace the item or refund payment Any replacement disk or memory stick will be warranted for the remainder of the original warranty period or thirty 30 days whichever is longer This warranty gives you limited specific legal rights You may have other rights in some places No Right to Reply No oral or written information or advice given by lon Science Ltd its dealers distributors agents or employees will create a warranty or in any way increase the scope of the obligations of lon Science Ltd under this agreement and you may not rely on any such information or advice Limitation of Liability lon Science Ltd will not be liable for any direct indirect consequential or incidental damages including damages for loss of business profits business information or possibility of such damages The above limitation will apply where allowed under local laws Governing Law The laws of the United Kingdom govern this Agreement Unrivalled Detection Page 28 of 41 www lonscience com TIGER MANUAL lon Science Ltd Batteries Warnings BATTERY CHARGING Charge TIGER and its Lithium ion battery packs in a Non Hazardous environment only BATTERY REPLACEMENT Never replace primary Alkaline battery cells while in a potentially explosive or hazardous location Use only Duracell Procell Alkaline batteries MN1500 BATTERY CONNECTION The TIGER Lithium ion and Alkaline battery packs have been specially designed to
40. t 0 Three sound projection lines indicate the volume level However there are four sound levels and the quietest level has no line Alarm Bell The icon consists of 2 parts the Bell and the sound bars When the instrument reaches a Low alarm the Bell and one sound bar will flash when a High alarm is reached the Bell and both sound bars flash Flashlight Torch When the flashlight torch is illuminated the light beams appear Lock The Lock icon will appear when the configuration of the TIGER has been locked in TIGER PC Soft keys in Button A column are enabled and Buttons in column B disabled see page 13 Main Screen Area During normal operation this area displays the readings in four large numbers with the units of measurement below During setup and adjustment a function bar overlays the main screen Unrivalled Detection Page 12 of 41 www lonscience com TIGER MANUAL lon Science Ltd Understanding display screen Soft key areas The following icons will appear in the soft key areas as soft key options They are selected using the Up and Down keys Those labelled optional will only appear if your TIGER has that functionality These icons can be reorganised using the TIGER PC software ZERO Zero AH Zone Button A Y Button B 1 Information Options E Single Data Log ZERO Zeroing RF Response Factor as Zones Unrivalled Detection Page 13 of 41 Www Ionscience com TIGER MA
41. t care First ensure that the TIGER is turned off and that you are in a clean environment such that the sensor parts will not be contaminated by dust oil or grease Remove the lamp as detailed on page 34 Inspection of the lamp may reveal a layer of contamination on the detection window that presents itself as a blue hue To check for confirmation hold the lamp in front of a light source and look across the window surface Clean the window using the PID Lamp Cleaning Kit A 31063 supplied USE of PID Lamp Cleaning Kit A 31063 The container of cleaning compound contains Aluminium Oxide as a very fine power CAS Number 1344 28 1 A full material safety data sheet MSDS is available on request from lon Science Ltd The key issues are listed below Always replace the lid after using the cleaning compound Hazard identification e May cause irritation of respiratory tract and eyes Handling e Do not breathe vapour dust Avoid contact with skin eyes and clothing e Wear suitable protective clothing e Follow industrial hygiene practices Wash face and hands thoroughly with soap and water after use and before eating drinking smoking or applying cosmetics e The Compound has a TVL TWA of 10 mg m3 Storage e Keep container closed to prevent water adsorption and contamination To clean the lamp 1 Open the vial of Aluminium Oxide polishing compound With a clean cotton bud collect a small amount of compound 2 Use th
42. tem Requirements Application of Quality Systems for equipment manufacture On behalf of lon Science Ltd declare that on the date this product accompanied by this declaration is placed on the market the product conforms to all technical and regulatory requirements of the above listed directives Name Mark Stockdale Signature Unrivalled Detection Position Technical Director Date 20th January 2010 www lonscience com Page 5 of 41 TIGER MANUAL lon Science Ltd do te Contents Md A 3 Averissemen te or 4 Declaration Of conformity rinda A A A A ato 5 Statements USA a da 7 FRESDONSIDINGY TOF USC naaa RS doc nee ieee niet 7 IMPORTAN Tora As 7 Quallly ASSULAMCO ias 7 AS AA o de RS einai eed iac T calibration Facility sia a a a E E O R TE 7 Legal Noli ida T Introduction to Tiger do 8 GOtIng Stare ini A dia 9 The Manta e il a acs 9 MEENOzAlE Sa a des dde etal edo 9 PUNO a do EA 9 TOO Matt E a O 9 A A ee ni pe A A NS ne te dee ates 9 RESPONSE ri e dl cn nn 9 SETAL Leve Si tada 9 Understanding THE KEY Dad aiii a 10 General DESEO GIN andan 10 Keypad Function DESCRIDTONS a de se o dd 10 Understanding the display SOT Ori e Ca 11 Seen DIS a e e a 11 SAUS COME ta A e oe 11 MAIN SCT AIO en eae ee bee aa Dee ie ci cost 12 OL A E OS 13 Using your Tiger EE 14 Tiger PC software Instillati ni caia eee here 17 Connecting o tec thee atest hee A cea ectateeees 18 THE COMINO e e a 21 A O de A tt 23
43. tenance PID Sensor Lamp Replacement and Cleaning When using your TIGER in conditions of high ambient humidity the PID may show unexpected readings appearing to increase This occurs due to dust or other small particles within the detector becoming hydrated with humidity This causes these particles to conduct a signal between the electrodes The problem can be resolved by the user in the field using the procedure below and a can of computer duster air In normal use the lamp should be cleaned after every 100 hours of use based on 30 ppm for 100 hours Reduce this if TIGER is used in heavily gas contaminated environments Please note that some esters amines and halogenated compounds may accelerate window fouling in these cases cleaning may be required after every 20 hours of use Cleaning frequency will also depend upon alarm levels set and prevailing environmental conditions CAUTION The TIGER is a sensitive detector Internal components must be handled with clean hands and clean tools The TIGER lamp is fragile Handle with great care Never touch the window and do not drop To remove the MiniPID Sensor for cleaning or lamp replacement first ensure that the TIGER is turned off and that you are in a clean environment such that the sensor parts will not be contaminated by dust oil or grease Remove the Sensor Cover see image on right The centre screw may be turned with a small coin or a suitable flat bladed screwdriver
44. ts an absolute zero The lower symbol represents a relative zero which follows the drift of the PID detector Having made your selection press Esc Ifthe relative zero has been selected TIGER will zero itself before returning to the main screen Zones Press the Zone soft key to display the currently selected Zone The TIGER is defaulted to Zone 1 Zones are set up in TIGER PC only and the name given to the zone s will appear on the screen Use the Up and Down keys to select alternative zones El Single Data Log optional Press the Single Data Log soft key to take a single data log reading The single data log symbol appears with a tick ll Y Press Esc to return to the main screen Press the Single Data Log soft key again to take another reading The fixed LCD memory icon flashes off when a single log is taken Options Press the Options soft key to give access to various adjustable features selected with Of the Up and Down keys Press Enter to confirm selection e Backlight Four options are presented Permanently Off Permanently On On in low ambient light and Timed Set the time in TIGER PC Select the preferred option with the Up and Down keys and confirm with Enter Unrivalled Detection Page 14 of 41 www lonscience com TIGER MANUAL lon Science Ltd Using your Tiger Sound The sound options are presented as set up in TIGER PCH Use the Down key to select either keypad beep a
45. ur distributor Factory Calibration offers a safe set of three point calibration data This should be used if the custom calibration fails and will keep the unit working until a good custom calibration is completed For Custom Calibration first set up the parameters in TIGER PC see the TIGER PC Software section of the manual TIGER allows you to custom calibrate using any gas concentration from 10ppm A cylinder of the selected gas at each of the chosen concentrations is required Demand flow regulators are recommended for the Tiger If flow regulators are to be used then a 0 3L min is recommended a flow adaptor should be used to avoid over pressurising the instrument sold separately Have the cylinder s of gas regulator s and the zero carbon filter included in the Calibration Kit A 861418 for your TIGER ready before starting the procedure Alternatively a known clean air supply may be used as the zero gas Please ensure you are familiar with the entire calibration procedure before attempting to calibrate your TIGER Note The calibration of your TIGER must be carried out in a clean air environment Ensure all parts of the Calibration Kit are available and ready for use Never calibrate the zero with the span gas connected Press the Options soft key 2 on your TIGER to access the adjustable features OLE Then use the Up or Down key to select calibration A Press Enter to confirm selection Sele
46. ur sections Fixed LCD Status icons fill the top of the screen offering instrument status at a glance The Icons display only when a function is selected The main central viewing screen will display readings in large numbers only 4 digits and decimal place will display 0 001 ppm to 5 500 ppm Two soft key areas have been set aside as soft key indicators The area between the soft key indicators displays the measurement units Status icon if Memory Status This icon only appears if the instrument has the data logging functionality Four sections within the border fill as the data log memory is filled Empty border 100 memory available through to full where all segments are present USB The icon will appear when connected to a host device Battery Status This icon consists of a border and four segments The segments fill or empty to indicate 0 25 26 50 51 75 and 76 100 full When discharging the border will flash for 1 minute before the instrument shuts down When charging the segments will successively fill until 100 charge is reached Bla Unrivalled Detection Page 11 of 41 www ionscience com TIGER MANUAL lon Science Ltd Understanding the display screen N 4 _ r x H 0 0 Backlight Light beam lines appear when the backlight is on Sound The main body of the icon is present at all times The disable lines appear if all 3 alarm sounders are disabled and the volume is a
47. vely if you wish new data to overwrite the oldest stored data in the memory and continue to store new data during data logging select the recycle symbol Vibrate To set your TIGER to vibrate under alarm conditions select the vibrate symbol Calibration Use this area to define your Custom Calibration parameters First connect TIGER to your PC as detailed above under Connecting your TIGER to a PC TIGER PC offers a two point calibration Calibration zero span Send this information to your TIGER The calibration procedure is Zero on Carbon Canister detailed under the Maintenance section of this manual 1 0 RF 100 0 ppm Startup Text Response Factor SPAN 1 Enter the text you wish to appear on the startup screen of your TIGER Send to Instrument When you have configured your instrument or completed your changes send them to your TIGER by clicking the Send to Instrument box If the message There was a problem sending to the instrument appears click OK and return to the Home screen Repeat the procedure under Connecting your TIGER to a PC above If the Summary screen for your TIGER appears visit the configuration screen again If not repeat the entire procedure If your PC still fails to read or write to your TIGER seek advice from your distributor or from lon Science Ltd Unrivalled Detection Page 23 of 41 www lonscience com TIGER MANUAL lon Science Ltd T
Download Pdf Manuals
Related Search
Related Contents
Bedienungsanleitung - GMC Evénements de la semaine du 7 au 13 juillet 2008 DP-25 digital piano Honda 08V67-SJC-1000-81 User's Manual Klip Xtreme KMA-004 mobile device charger TAXI_3 MALLES AVANT.indd - Hachette Kenwood eXcelon KAC-X541 User's Manual Manual de Instrucciones Pam pandereta Pass Labs XP-30 User's Manual Copyright © All rights reserved.