UDS Troubleshooting Guide
Contents
1. b Patient passed gas during my CMG study causing the Pabd channel to rise Yes this will happen Document the event by adding a comment on rise of Pabd Patient passed gas c Pves catheter fell out this can be due to taping To avoid this in female patients tape the catheter onto the outer labia and the inner thigh of the patient Be sure that Pves and Pabd tubing are out of the way when the patient voids For men apply a thin strip of tape on the shaft of the penis and run the tape down the Pves catheter Secure well by folding the tape in half around the catheter d EMG Channel is not monitoring correctly Have the patient move a few inches forward if sitting down Sometimes when they sit on the seat the EMG patches reflect a solid red line with no recruitment detected when the patient is asked to cough or squeeze You should always have your patient cough at the beginning of each study to ensure all channels are responding appropriately Apply tape over EMG patches on female patients so when she voids the EMG patches do not get wet This is usually not an issue for males Add a note on the report if EMG patches get wet artifact or if EMG was active or quiet during voiding synergistic dyssynergic Remember UDS is like telling a story to someone who is not usually present i e physician reading tracings You want to present a clean study with documented events that tell the story of the patient s symptoms you are tryi
2. As long as theyreturn to baseline so thatPdet is not continuously negative you are fine 8 Why do Pabd pressures drift sometimes a Stool will affect pressure in Pabd You may have your patient go to the bathroom to have a BM before you place catheters if they say they need to b You may try to digitally remove the stool if the patient is a Spinal Cord patient Be aware if patient has Autonomic Dysreflexia recommend to all SCI patients to do their bowel program the night prior to their UDS study c Gas patient movement talking laughing or straining will all cause Pabd artifact d When patients begin voiding at Maximum Capacity Pabdmay drift down as the pelvic floor relaxes This is normal 9 How often does my equipment need to be calibrated Check with your manufacturer for recommendations or refer to your UDS User Manual It is probably good practice and easy to remember to CHECK the calibrations on the first day of every month Actual calibration should only be done if the Calibration Check proves some inaccuracy c o LABORIE 2012 This information printed with permission of author for use by LABORIE LABORIE www laborie com
3. LABORIE UDS Troubleshooting Guide www laborie com The following information has been provided by Myra Joseph RN who is a member of the Society of Urologic Nurses and Associates SUNA She currently serves as SUNA s Urodynamic Special Interest Group Leader and Alamo Chapter President Myra uses LABORIE UDS equipment This information reflects her views and experience and is not that of SUNA or LABORIE At recent SUNA Symposium s attendees were asked to complete a survey whose goal it was to provide solutions to common problematic issues that many Clinicians encounter during UDS testing Following are some of the most common issues that came up and the answers provided 1 What is Artifact There are several different types of artifact when it comes to UDS Here are a few examples and how they can be explained on your study a Patient kicked the Uroflow cup during the test If the Uroflow is salvageable move the Uroflow Start Stop and Peak Flow markers to measure the voiding not the artifact If it is not salvageable obtain a mechanical fill at the end of your CMG study Fill up your patient s bladder again on Fast Fill at the end of the study to the same volume at capacityas on the CMG portion of the study Confirm the patient has a strong urge to void then switch to the non instrumented Uroflow test screen Remove the Pves catheter and allow the patient to voidafter giving the patient Permission to Void
4. ee Pves and Pabd both rise at the same level and the number you mark at the peak of the cough on Pves or Pabd should measure very close or exact Sometimes have the patient perform a Valsalva maneuver patient pushes and bears down using pelvic floor this allows the numbers to rise more gradually and match much better than with coughing c o LABORIE 2012 LABORIE This information printed with permission of author for use by LABORIE o www laborie com e Check to see if the female patient has a cystocele You can count on challenges during these studies and lots of artifact they may be quite challenging f Confirm the bladder is empty at the start of the test Always measure the PVR by draining the patient s bladder after voiding and prior to starting the test If the patient is unable to void fora Uroflow the amount you catheterize will be referred to as the Residual g Start the pump on Slow Fill if you have a patient who has a diagnosis of low capacity bladder or urgency frequency Starting the filling too fast may induce instability sooner h A patient who comes in with an indwelling Foley catheter has not had their bladder stretched for awhile and usually cannot tolerate much volume i If no instability is noted by 200cc on the type of patients mentioned above you may increase the pump to medium fill Sometimes Pves and Pabd channel rises or falls in response to different things such as talking passing gas etc
5. emind them not to talk push or move during this phase recommend playing classical jazz or contemporary music during procedures my patients always comment on how enjoyable and relaxing this is 5 How do you handle nervous patients and what if they are unable to void a Iplay music during my procedures it helps to relax the patient have all kinds of music available and ask the patient what they prefer to listen to b explain to the patient exactly what am doing and why using simple terminology before insertion of catheters c Ifthe patient is unable to void run water in the sink dim the lights in the room leave the room and give them some privacy to void 3 5mins Patients have a bell by them to ring when they finish voiding so may know when to enter the room This works well most times d If patient is unable to void after giving them privacy start pulling out the catheters one at a time Pves is first if no results then Pabd if no results then EMG patches If this still doesn t work allow the patient to go into a private bathroom with urinal for men or hat for women and these patients always seem to void in the bathroom alone note that the patient was unable to void until all catheters were removed and note how much the patient voided as well as their PVR 6 What am I looking for with Urgency or Stress Incontinence a Remember when performing a study you are looking to reproduce the patient
6. n the rectum or catheter repositioning It is not recommended to use the Equalize button during the filling phase of the study as this is artificially correcting the pressure vs stopping the study and troubleshooting the problem The equalize button is intended to balance the Pabd pressure at the beginning of c o LABORIE 2012 This information printed with permission of author for use by LABORIE the study when it is off by 1 2 cm H20 after all troubleshooting efforts are exhausted During the voiding phase the pelvic floor may relax so much that Pabd goes down slightly 4 What about Early Sensation due to cool water being used to infuse into the bladder First recognize that this is what most patients will feel during the start of the filling phase of the study Tell your patient that if they feel a cool sensation of water going in to the bladder this is normal Then ask the patient to identify when they feel the First Sensation that there is actually fluid within the bladder but without a need to actually void yet then First Desire to void when they would void at the next convenient moment Next have them identify when they feel a Strong Desire to void and lastly have them identify when they are at Capacity when they cannot take any more and need to void immediately Tip Explain to the patient that when it is time to void they need to relax and be still so that the pressure generated in their bladder during voiding is measured correctly R
7. ng to recreate You want to document what happened during the filling and the voiding phase of the study you are trying your best to reproduce symptoms but sometimes it is not achieved in spite of your best efforts 2 What is the difference in zeroing to atmosphere vs zeroing to patient The International Continence Society ICS recommends zeroing to atmosphere and not to the patient This means that your baseline before filling will start at the patient s actual resting pressures So every patient will have their own unique baseline pressures When you zero to the patient which is nottecommended you are taking the patient s resting pressures Pves Pabd being measured by the catheters and zeroing that number so that their resting pressure has been manipulated What do you do if the Pves and Pabd channels drift during the filling and voiding phases of the CMG At the start of the test the Pves and Pabd channels should be very similar so that the actual bladder pressure on Pdet channel is at or near zero Pdet should be between 5 and 5 A downward drift of pressures is usually due to aconnection problem or shifting of the position of the catheter You can always ask the patient to cough and confirm that pressures are responding fully on both channels If Pves drifts upward andPabd remains stable then Pdet will rise and it may mean a bladder compliance problem Mark any artifacts including things such as peristaltic movement withi
8. s symptoms With urgency you are trying to demonstrate instability during or at the end ofthe filing phase of your study i e a rise in the Pves channel during the filling phase b With stress incontinence you are performing stress maneuvers cough and valsalva every 100cc until you see leakage demonstrated with your eyes If the patient does not leak and they are sitting have them stand up and try your maneuvers again with them standing c With men who have a diagnosis of Stress Incontinence Post TURP and do not leak with the Pves catheter in place recommend you perform a 2 fill Fast fill them up to their prior capacity right where they are able to hold it Remove the Pves catheter and try the stress maneuvers again using the Pabd catheter as your guide for leak point pressures 7 Why does the Pdet channel drift down and numbers go negative sometimes a At the start of your study you need to make sure your Pdet pressures at rest are ranging from 0 5 cmH20 before you begin the filling phase Ideally starting at zero on Pdet only is your goal b This may require adjusting the Pves or Pabd catheter by pushing them in more or pulling back some The Pves catheter may be against the bladder wall or the Pabd catheter may be caught in a fold of the rectum c Make sure there is no stool in the rectum as this will affect rectal pressures d Confirm your catheter is in the bladder by having the patient cough several times You should s