1. 1 2 3 4 Configuration Connected to server 127 0 0 1 FSE Engineering Supplier ENG2 0 0 30b 50096 927 Rev For additional information contact your Medivators Representative or Technical Support at 1 800 444 4729 Pg 13 of 18 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS sePROCESSNGSYSTEIS AND INSTALLATION INSTRUCTIONS Attach the barcode to the corresponding endoscope Prep the surface with alcohol and place the barcode on a flat surface on the scope control handle so it is visible when placed in the ADVANTAGE Plus basin See image below TJF 160F e For more detailed instructions refer to the User Manual System Parameters The system parameters located in the Management Application have a number of user options the customer needs to decide whether they want to use Adjust the system parameters as needed and be sure to save the changes Record changes in Installation Checklist e Chemical Name Enter tRapicide PA e Location Enter the facility name e Request Patient Data at beginning of cycle Enter yes or no e Request Specialist Doctor Data at beginning of cycle Enter yes or no e Request Assistant Data at beginning of cycle Enter yes or no e Use Alcohol If using alcohol set to yes select parameter set accordingly e Use Detergent If using detergent set to yes select parameter set accordingly 50096 927 Rev A For additional information contact your Medivators Pg
2. REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS 6 Dimensions e The ADVANTAGE Plus physical dimensions are 59 H x 48 W x 36 D weight 400 LBS e If the optional air compressor is installed the ADVANTAGE Plus will need an additional 5 25 at the back of the unit e lf the active vapor management system is installed the ADVANTAGE Plus will need an additional 2 at the back of the unit e Space is required on either side of the unit approximately a combined 12 for easy access to the internal components If this is not available the ADVANTAGE Plus contains a base with casters allowing the system to be moved There should be adequate room for the ADVANTAGE Plus to be moved for access to accommodate maintenance f Endoscope and User Information e To ensure a smooth and trouble free installation the customer must supply a complete list of users and endoscopes that will be reprocessed in the ADVANTAGE Plus e Fill out the ADVANTAGE Plus Site ID Form Document Number 50092 380 8 Network Remote Access e The ADVANTAGE Plus is equipped with a PC based computer running Windows XP which will allow full networking along with the capability of remote diagnostics e To allow remote diagnostics an internet connection would be required highly recommended The ADVANTAGE Plus Site Requirement Checklist is utilized to ensure the installation site meets the required specifications If all requirements are met the site
3. 14 of 18 Representative or Technical Support at 1 800 444 4729 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS 2 reprocessine systews AND INSTALLATION INSTRUCTIONS Note If the Specialist Doctor Data is going to be used a barcode needs to be assigned The barcode and the Doctor s name can be conveniently placed on the barcode reader mount General In the Windows control panel verify the correct date and time adjust if necessary The audible error alarm on the ADVANTAGE Plus uses the Window application Again in the control panel verify the sound and adjust if necessary 50096 927 Rev A For additional information contact your Medivators Pg 15 of 18 Representative or Technical Support at 1 800 444 4729 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS INSTALLATION VERIFICATION Water Temperature Calibration In the LIO application under Program Selection run the Water Sample Utility Ensure the Disinfection Blocks are in place Once the lid opens automatically measure the temperature of the incoming water with a thermometer It should measure 95 F 35 C if not adjust the external mixer valve until the desired temperature is reached Press START to finish the process Waterline Disinfect In the LIO application run the Waterline Disinfect Utility Test with scopes Place an endoscope in both basins and run a complete cycle Ensure the cycle completes wit
4. than 12 gpg 200 ppm for optimal performance e The pre filter assembly is wall mounted and has the dimensions of 26 H x 38 W x 6 D and must be located within 3 feet of the ADVANTAGE Plus The incoming water connection to the pre filter assembly is 72 NPT on the right hand side All fittings and hoses for the output of the pre filter assembly are included e Ashut off valve must be installed before the pre filtration system It is important that the valve be mounted within reach of the operator e tis highly recommended that the facility water be analyzed to determine if the supplied pre filter is adequate Contact your Medivators Representative or Technical Support to arrange for water analysis 2 Drain e Adrain system capable of draining 5 GPM less than 18 from the floor and vented will ensure the proper operation of the ADVANTAGE Plus The supplied drain line has a 1 OD 50096 927 Rev A For additional information contact your Medivators Pg 2 of 18 Representative or Technical Support at 1 800 444 4729 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS Air Supply The ADVANTAGE Plus system utilizes medical grade house air dry oil free If this is not available an optional Air Compressor kit will need to be ordered with the ADVANTAGE Plus Required air pressure is 45 to 145 PSI with a minimum flow volume of 1 5 CFM A ID flexible air hose EPDM braided 150
5. type assigned to user 4 Barcode Enter barcode number assigned to user 5 Tag Leave blank 6 Pers Code Leave blank 7 Password Enter personal password not required e Turn the edit function off to save the information e The barcode labels are located in the installation kit 50096 927 Rev A For additional information contact your Medivators Pg 11 of 18 Representative or Technical Support at 1 800 444 4729 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS e The group type determines the level of access the user will have to the ADVANTAGE Plus systems 1 Operator Hospital Day to day users lowest user level of system access scopes changing disinfectant 2 Maintenance Hospital Biomed filter changes basic maintenance 3 Administrator Hospital Nurse manager highest user lever of system access user scope entry data management Parameter Sets Install the correct Parameter Set for each endoscope type The Parameter Sets selected will depend on the scope disinfectant and cycle type Wash Detergent Alcohol The Parameter Sets do not come preinstalled on the ADVANTAGE Plus To obtain the correct Parameter Sets please contact Medivators Technical Support This must be done before Endoscope Entry Endoscope Entry As part of the pre installation the customer supplied a list of Endoscopes on the Site ID Form Confirm the list with the operator and add remove as n
6. MEDIVATORS DVAN TAG ENDOSCOPE REPROCESSOR SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS 2 MEDIVATORS A REPROCESSING SYSTEMS 50096 927 Rev A i MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS SITE REQUIREMENTS The purpose of this section is to provide the customer with requirements that need to be met to ensure the ADVANTAGE Plus will be properly installed Applicable Documents e ADVANTAGE Plus Site ID Form Document Number 50092 380 e ADVANTAGE Plus User Manual It is the responsibility of the customer to meet these requirements to ensure proper installation of the ADVANTAGE Plus The ADVANTAGE Plus will be delivered on a pallet along with multiple boxes It is important to keep all the boxes received with the ADVANTAGE Plus together It is the responsibility of the customer to provide qualified personnel to provide the connections for the ADVANTAGE Plus to the building supply Water Electrical Drain Air and Network Internet during the installation process Any deviation from these specifications may cause operational issues that would not be covered by the Warranty 1 Water Supply e The water supply must be connected to a mixer valve providing a water temperature of 95 F 35 e The incoming waterline must be a minimum of 72 ID providing a flow rate of 3 2 GPM and a water pressure of 40 to 87 PSI e Ensure water hardness is less
7. ameter Sets Parameter sets inputted and changes noted below General Date and time correct sound checked Water Temperature Calibration Water temperature set to 95 F 35 C Test with Scopes Cycles run with scopes completed without any errors Waterline Disinfect Waterline disinfect performed Data Backup Data backed up CD left with ADVANTAGE Remote Access Internet access available System Parameter Changes Parameter Set Changes Notes 50096 927 Rev A For additional information contact your Medivators Representative or Technical Support at 1 800 444 4729 Pg 17 of 18 Manufactured in the USA by Z MEDIVATORS 9 REPROCESSING SYSTEMS AMinntech Corporation Business Group Minntech Corporation 14605 28th Avenue North Minneapolis MN 55447 USA Phone 763 553 3300 Toll Free 800 328 3340 Fax 763 553 3387 website www minntech com Minntech BV Minntech Asia Pacific Sourethweg 11 438 Alexandra Road 6422PC Heerlen 03 01 Alexandra Point The Netherlands Singapore 119958 Phone 31 45 5 471 471 Phone 65 6227 9698 31 45 5 429 695 Fax 65 6225 6848 email info minntechbv com Minntech Beijing Representative Office AVIC PLAZA Room 1508B No 10 Yi Dongsanhuan Zhonglu Beijing 100022 China Phone 8610 6567 8446 Fax 8610 6567 8445
8. ards and there are no restrictions The hose can be cut if needed Having some slack in the drain hose may aid in the movement of the ADVANTAGE Plus for service 5 Air e f using house air the installation location should already be fitted with a suitable air line based on the pre installation criteria Attach the hose to the barbed fitting located on the back of the unit secure with a clamp e If the optional air compressor is to be used it should already be connected from a previous step 6 Electrical Locate the electrical power cord located in the installation kit Measure the voltage at the wall receptacle it must read 240 VAC x 1096 confirm it is a GFI protected outlet Connect the cord to the main electrical connection located on the lower back panel of the ADVANTAGE Plus Plug the other end into the wall receptacle 7 Ventilation 50096 927 Rev A If the customer is not going to connect the ADVANTAGE Plus to an external fume extraction system the black cap on the vent on the back of the ADVANTAGE Plus should remain in place and the internal fan remains turned off If the facility has supplied a 4 flexible duct connected to an external vent the black cap can be removed and the internal fan turned on The flexible duct can then be attached to the vent connection with a clamp If the facility is using the optional vapor management system it should already be installed in a previous step For additional informat
9. eeded e In the Management application select the Instrument screen Enter the Endoscope information based on the list supplied by the customer Turn the Edit function back on and enter the information below for each endoscope 50096 927 Rev A For additional information contact your Medivators Pg 12 of 18 Representative or Technical Support at 1 800 444 4729 gt MEDIVATO ADVANTAGE PLUS SITE REQUIREMENTS 2 REPROCESSING systems AND INSTALLATION INSTRUCTIONS General Instruments Ins umet 2 falF H180 nt typo Connection block Brand e code number 2 8511 Olympu j GIF H180 5 6 7 8 9 10 11 1 Enter the scope model number 2 Enter the Number of channels obtained from parameter set found in the hookup guide 1 17 501 has 5 channels Leak test number of leak test connections in most cases 1 3 Enter the hookup number from the hookup guide 4 Select the parameter set from the hookup guide add a brief program description in the comments 5 Enter the last four digits of the endoscope serial number 6 Select instrument type from pull down menu connection block will auto populate 7 Enter the endoscope brand 8 Enter the number from the barcode that will be used on the scope 9 Re enter the scope model number 10 Enter the scope serial number 11 Enter the facility s scope inventory number this field must be filled in nix ais EZ Dlx EET 2 S Tm en
10. he floor vented Air Supply Medical grade house air 45 to 125 PSI minimum flow volume of 1 5 CFM Optional air compressor available adds 5 25 to depth of unit Electrical Supply 240 VAC 10 single phase 60 Hz 10 A NEMA 6 15R outlet GFI protected accessible by user Ventilation 10 to 12 room air exchanges per hour Optional ventilation system available 110 VAC adds 2 to depth of unit Unit Dimensions 59 H x 48 W x 36 D 400 LBS Add 12 of open space to the total width to aid with service access Endoscope List ADVANTAGE Site ID Form 50092 380 completed Remote Access Internet connection optional 50096 927 Rev A For additional information contact your Medivators Pg 6 of 18 Representative or Technical Support at 1 800 444 4729 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS The purpose of this document is to provide the requirements that need to be met for the proper installation of the ADVANTAGE Plus Applicable Documents e ADVANTAGE Plus Site ID Form PN 50092 380 e ADVANTAGE Plus User Manual Packages e ADVANTAGE Plus e Computer box e Pre filter e Installation Kit e Air Compressor optional e Vapor Management System optional e Printer optional Additional Items Needed e Blank writable CD Pre installation criteria must be met to ensure proper operati
11. hout any errors DATA BACKUP After the installation is complete go to the Management application and export the system parameters user information physician assistant and endoscope information Close the LIO Server and Management applications and run the Backup utility located in the MDS folder on the desktop Open My Computer on the desktop and open the C drive In the Program Files folder locate the MDS folder Place a blank writable CD into the CD Drive and save the entire MDS folder to the CD Label the CD with ADVANTAGE Plus Backup SN Keep with the ADVANTAGE Plus and inform the customer the CD must remain with the ADVANTAGE Plus 50096 927 Rev A For additional information contact your Medivators Pg 16 of 18 Representative or Technical Support at 1 800 444 4729 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS This checklist should be used to ensure the proper installation of the ADVANTAGE Plus Automated Endoscope Reprocessor Upon completion of the installation the unit will be ready for operation ADVANTAGE PLUS INSTALLATION CHECKLIST Unpacking All packages are accounted for Assembly All installation specifications have been met User Entry All user information entered barcodes assigned Endoscope Entry All endoscope information entered barcodes assigned System Parameters System parameters checked and changes noted below Par
12. ion contact your Medivators Pg 10 of 18 Representative or Technical Support at 1 800 444 4729 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS 8 Placement e When all the connections are completed move the ADVANTAGE Plus to its final location and ensure a good fit Turn on the water air and the computer Maker sure there are no leaks and the computer powers up e Load the containers of Rapicide PA Part A and Part B ito the ADVANTAGE Plus Connect the disinfection blocks in both basins they are located in the installation kit Fill the Alcohol and Detergent reservoirs as needed COMPUTER The computer will boot up with the ADVANTAGE Plus Server application and the LIO application Open the MDS folder on the desktop and start the Management application When asked for a login name select SysAdmin from the pull down menu Call Medivators Technical Support for the password of the day NOTE The computer comes with a mouse and keyboard which can be used to ease the data entry that follows User Entry As part of the pre installation the customer supplied a list of Users on the Site ID Form Confirm the list with the operator and add remove as needed Users e Enter the User information based on the list supplied by the customer Turn the Edit function back on and enter the information below for each user 1 Name User name 2 Login Login name 3 Group Enter group
13. is suitable for the installation of the ADVANTAGE Plus 50096 927 Rev A For additional information contact your Medivators Pg 4 of 18 Representative or Technical Support at 1 800 444 4729 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS sePROCESSNGSYSTEIS AND INSTALLATION INSTRUCTIONS Dimensions and Clearances Note Add 3 to the width for the barcode stand 4g 138cem 48 122cm w barcode bracket gt Figure 1 Front Dimensions 36 2 g2cm 41 25 105 3 f 59 59 14986 cm 14986 cm E qur 424 96 5cm 106 93 V 106 93 cm Air Compressor 50096 927 Rev A For additional information contact your Medivators Pg 5 of 18 Representative or Technical Support at 1 800 444 4729 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS This checklist should be utilized to ensure the installation site meets the required specifications If all specifications are met the site is suitable for the installation and use of the Medivators Advantage Plus Automated Endoscope Reprocessor SITE REQUIREMENTS Water Supply Minimum flow rate of 3 2 GPM with a pressure of 40 to 87 PSI Water Temperature 95 F 35 C Hardness less than 12 gpg 200ppm Pre Filter wall space 26 H x 38 W x 6 D within 3 feet of the ADVANTAGE Plus Drain 5 GPM less than 18 from t
14. mpressor box Vapor Management System optional e f the optional vapor management system is used install per the instructions located in the vapor management system box Printer e f the optional printer is used install per the instructions located in the printer box Water Supply Pre Filter e The pre filter should be already mounted and plumbed e Verify that the correct filters are installed e Connect the supply line located in the pre filter box to the output of the pre filter and to the left right barbed fitting water inputs on the back of the ADVANTAGE Plus secure with the included clamps e locate the valve control cord located in the pre filter box Connect it to the pre filter valve assembly and to the location marked Sanitize on the back of the ADVANTAGE Plus e Locate the disinfectant line in the pre filter box Connect it to the barbed fitting on the pre filter valve assembly and the gray quick connect fitting on the back of the ADVANTAGE Plus For additional information contact your Medivators Pg 9 of 18 Representative or Technical Support at 1 800 444 4729 E gt 22 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS 49 REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS 4 Drain Locate the drain hose located in the installation kit and attach the fitting to the drain fitting on the back of the ADVANTAGE Plus Place the open end of the drain hose into the facility drain Ensure that the line slopes downw
15. on of the ADVANTAGE Plus It is the responsibility of the customer to provided qualified personnel to connect the ADVANTAGE Plus to the building supply during the installation process 50096 927 Rev A For additional information contact your Medivators Pg 7 of 18 Representative or Technical Support at 1 800 444 4729 E gt 22 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS 49 REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS INSTALLATION Unpacking Move the ADVANTAGE Plus and additional boxes to a suitable location for unpacking and ensure all of the boxes are accounted for Remove the packaging around the ADVANTAGE Plus unlatch the base from the pallet and using the supplied cardboard ramps carefully remove the ADVANTAGE Plus from the pallet Leave the base cover off until instructed to install Assembly Before assembly move the ADVANTAGE Plus to the installation site Once assembled the ADVANTAGE Plus will not fit through standard door ways 1 Base Cover Place the base cover on the base of the ADVANTAGE Plus Align the screw holes in the cover with the base and use the 4 screws to secure 2 Computer Computer Mount Locate the computer mount in the computer box Remover the upper panel on the back of the ADVANTAGE Plus The computer mount attaches to the back of the ADVANTAGE Plus with a single threaded stud located on the upper basin assembly and with 2 screws mounted under the base assembly on the frame Once at
16. psi minimum is required to connect the ADVANTAGE Plus to the house air This is not supplied with the ADVANTAGE Plus The hose connects to a barbed fitting on the back of the ADVANTAGE Plus Electrical Supply 240 VAC 10 single phase 60 HZ 10 A GFI protected outlet NEMA 6 15R should be located within the operator s reach NOTE It is recommended to install a surge protector to protect the ADVANTAGE Plus from power fluctuations voltage spikes etc Ventilation It is the Customer s Responsibility to ensure the room is in compliance with this requirement This is not a requirement of the ADVANTAGE Plus but of the chemicals that are used for disinfection It is a health and safety requirement to have 10 to 12 air changes per hour the room During cycle operation and the chemical heating process chemical vapor may be produced The air exchange will minimize staff exposure to chemical vapors A 4 flexible duct may be connected from the back of the ADVANTAGE Plus to the facilities exhaust system to aid in fume extraction The ADVANTAGE Plus may also be supplied with an active vapor management system to aid in chemical vapor management Please refer to the purchase order to determine the correct configuration of the ADVANTAGE Plus 50096 927 Rev A For additional information contact your Medivators Pg 3 of 18 Representative or Technical Support at 1 800 444 4729 MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS
17. tached the computer can be placed on the mount align the four holes and attach with the 4 screws Replace the back panel Computer Cables 50096 927 Rev A Locate the computer cables in the computer box There are 4 cables ASI serial cable PS2 mouse cable USB keyboard cable and computer power cable Attach the cables to the computer and to the marked locations on the back of the ADVANTAGE Plus For additional information contact your Medivators Pg 8 of 18 Representative or Technical Support at 1 800 444 4729 y 3 50096 927 Rev A MEDIVATORS ADVANTAGE PLUS SITE REQUIREMENTS REPROCESSING SYSTEMS AND INSTALLATION INSTRUCTIONS Barcode Reader e Locate the barcode reader box and the barcode reader mount located in the installation kit e Remove the right service cover and locate the 4 holes on the upper edge of the cover Carefully remove the gasket material along the edge and place it back once the barcode reader mount is attached e Align the 4 holes on the mount with the holes on the cover and attach using the Screws washers and nuts found in the barcode reader box e Attach the USB extension cable to the front of the barcode reader mount and then connect cable to the computer e Connect the barcode reader to the USB connection on the mount and place the bar code reader on the cradle Air Compressor optional e fthe optional air compressor is used install per the instructions located in the air co