1.               and Veriti   System User Guide    Running the test    Chapter 8 Temperature non uniformity test    The temperature non uniformity test uses the 0 1 mL probe assembly to test the  temperature uniformity of six different wells in each sample block     Note  If necessary  press F5  Cancel  to exit the test     To run the temperature non uniformity test                    1  Complete the procedures in    Setting up the 0 1 mL probe assembly    on  page 97   2  Press F1  Run  to start the temperature non uniformity test  The TNU  Performance screen opens with the setpoint value displayed   TNU Performance  Sample temp   xx x  C Cover temp   xxx  C  Setpoint is 55  C  Sample must be within 1 0  C of setpoint  Run  F1 F2 F3 F4 F5  The TNU Performance screen counts down the time until the setpoint is stable   TNU Performance  Sample temp   xx x C Cover temp   xxx C  Stabilizing at setpoint    x xx  F1 F2 F3 F4 F5  Note  The sample blocks must be within  1 0   C of the setpoint  In addition   the cover must be within  1   C of 55   C  It may take several minutes for the  GeneAmp   9800 to ramp up   3  When the TNU Performance screen opens to display the setpoint temperature   press F5  Cancel    TNU Performance  Sample temp   xx x C Cover temp   xxx C  Setpoint is 55  C  Sample must be 1 0  C of setpoint  Cancel  F1 F2 F3 F4 F5  4  When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer   Refer to the instructions in your Temperature Ve
2.            Fail    appears after the setpoint temperature s  for which the  test failed     TNU Performance    TNU at 95  C is xx xx   Fail    TNU at 55  C is xx xx   Fail       F1 F2 F3 F4 F5    e If the test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors  were not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See  How to obtain support  on page viii        pi    Ending the test Press F5  Cancel      P Hp    Remove the 0 2 mL probe assembly from the sample block   Power off the digital thermometer  then clean off the oil     Wait for the sample block to reach room temperature   25   C   then remove the    dual 96 well temperature verification plates from the sample block       Nre PHYSICAL INJURY HAZARD  Hot Surface  Use care    when working around the heated cover and sample block to avoid being burned    by hot components     5  Clean the oil off the sample block     46 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti   System User Guide       Chapter 4 Data Sheet  Calibration Verification Test    Data Sheet  Calibration Verification Test    When you run the calibration verification test  record the setpoint values for well R3  on this data sheet  At the end of the test  compare the values that are displayed on the  thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date T
3.          5  Press F1  Temp      The GeneAmp   9700 is automatically configured for the calibration  verification test  and the Calibration Verification screen opens     Calibration Verification  Block temp   xx x C Cover temp   xxx C             Place probe in well R3    Press Run    Run       38 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 4 Calibration verification test    Running the test In this test  you take temperature readings of the sample well at two different set  points using the 0 2 mL probe assembly     Note  Press F5  Cancel  if you want to exit the test     To run the calibration verification test     1  Press F1  Run   The Calibration Verification screen opens with the setpoint  value displayed     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 85  C  Cover must be within 1  C of 105  C    El F2 F3 F4 F5       Note  The cover must be at 105   C   1   C  It may take several minutes for the  system to ramp up     The Calibration Verification screen counts down the time until the setpoint is  reached     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Stabilizing at setpoint X xXX    F1 F2 F3 FA F5       2  When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer
4.       C4       C8       C12       E1       E4       E8       E12       H1       H4       H8       H12                Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  115  and Veriti  System User Guide       Chapter 9 Data Sheet  Temperature Non Uniformity Test    116 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Temperature Verification Tests for the  Veriti  Thermal Cycler    This chapter covers     OVVIE o Les de deo re Dee aee bad aR RE  Calibration verification tesb iiussese asse eon enden  Temperature non uniformity test            ll lessen   Data Sheet  Calibration Verification Test                      Data Sheet  Temperature Non Uniformity Test    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 10 Overview    Overview    118    Materials When you perform temperature verification tests on the Veriti   Thermal Cycler  you  use the materials provided in your kit     Assembly Figure 19 shows a properly assembled system  The frame is positioned around the  wells with the channel facing you  and the probe wire is threaded through the channel  to prevent damaging the wire when the thermal cycler heated cover is closed        Figure 19 The Temperature Verification System for Veriti  Thermal Cycler    Thermal Cycler Temperature Verification System For Ge
5.      0 2 mL probe 2  Place the 2720 temperature verification frame on the sample block   assembly   3  Use a cotton swab to coat the following wells with mineral oil   Al  A12  C4  C9  F4  F9  H1  H12   4  Place the 0 2 mL probe assembly into well A1  then as the test progresses  move  the 0 2 mL probe assembly to each of the test wells    5  Thread the probe wire through the channel in the 2720 temperature verification  frame to prevent damage to the probe and lead wires  see Figure 3 on page 8     6  Make sure the probe is connected to the digital thermometer    7  Pull the lever down to engage the heated cover and 2720 temperature  verification frame   IMPORTANT  Seat the probe properly and close the heated cover carefully  If  the probe wire is crushed when you close the heated cover  the probe can be  damaged    8  Power on the digital thermometer   Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  13    and Veriti  System User Guide    Chapter 2 Temperature non uniformity test    Configuring the 1  Power on the thermal cycler  The Main menu opens  For best results  allow the  2720 thermal thermal cycler 15 to 30 minutes to warm up to ambient temperature before  cycler performing the test     2  Press F4  Util   The Utilities screen opens   3  Press F1  Diag   The Diagnostics screen opens     4  Pre
6.     1  Decontaminate the probe assembly  see below      2  Send the equipment to the manufacturer  see below      Gently wipe the probe cone with a cotton swab that has been moistened with bleach  solution  20  bleach  80  water      IMPORTANT  Do not submerge the probe in bleach solution  Do not disassemble the  probe assembly       Norme CHEMICAL HAZARD  Sodium hypochlorite  bleach  is a    liquid disinfectant that can be corrosive to the skin and can cause skin  depigmentation  Read the MSDS  and follow the handling instructions  Wear  appropriate protective eyewear  clothing  and gloves     1  Inform Alpha Technics of the recalibration order by visiting the Alpha Technics  web site at http   www alphatechnicsonline com and following the online  instructions    Ship the Temperature Verification kits to   Alpha Technics  17151 Gillette Avenue  Irvine  CA  92614 5602    2  Ensure that the probe assembly is decontaminated     3  Pack the Model 4500 digital thermometer and probe assembly in the black case     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  129  and Veriti  System User Guide    Appendix A    4  Create a package that contains the     Black case containing digital thermometer and probe assembly    Address and contact information     Billing address    Shipping address    Name and phone number of a contact  preferably the person most  familiar with the thermometer    5  Send the package freight prepaid to Alpha Tech
7.     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    vi    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Preface    Safety information    Note  For general safety information  see this Preface and Appendix D     Safety    on  page 137  When a hazard symbol and hazard type appear by a chemical name or  instrument hazard  see the    Safety    Appendix for the complete alert on the chemical  or instrument     Safety alert Four safety alert words appear in Applied Biosystems user documentation at points  words in the document where you need to be aware of relevant hazards  Each alert  word   IMPORTANT  CAUTION  WARNING  DANGER    implies a particular  level of observation or action  as defined below        IMPORTANT    Indicates information that is necessary for proper instrument  operation  accurate chemistry kit use  or safe use of a chemical        CAUTION    Indicates a potentially hazardous situation that  if not avoided   may result in minor or moderate injury  It may also be used to alert against  unsafe practices           WARNING    Indicates a potentially hazardous situation that  if not avoided   could result in death or serious injury        DANGER    Indicates an imminently hazardous situation that  if not avoided   will result in death or serious injury  This signal word is to be limited to the  most 
8.   Damages  claims  and returns    134    Damages    Claims    Returns    If shipping damage to the product is discovered  contact the shipping carrier and  request inspection by a local agent  Secure a written report of the findings to support  any claim  Do not return damaged goods to Applied Biosystems without first  securing an inspection report and contacting Applied Biosystems Technical Support  for a Return Authorization  RA  number     After a damage inspection report is received by Applied Biosystems   Applied Biosystems will process the claim unless other instructions are provided     Do not return any material without prior notification and authorization     If for any reason it becomes necessary to return material to Applied Biosystems   contact Applied Biosystems Technical Support or your nearest Applied Biosystems  subsidiary or distributor for a return authorization  RA  number and forwarding  address  Place the RA number in a prominent location on the outside of the shipping  container  and return the material to the address designated by the   Applied Biosystems representative     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    GeneAmp   PCR 9700 Technical Note       This appendix addresses apparent discrepancies between the published instrument  specifications for accuracy and the pass fail criterion used during the routine  temperature verification     Temperature Accuracy V
9.   Inthe Tools Menu  select Run Temperature Verification         gt  INED         2007 09 04 I 04 05 PM      2  Press to begin the test  Perform the test as directed by the touchscreen  As  the test progresses  move the 0 5 mL probe assembly to the remaining test wells   B8  F1  F10      Temperature Verification Test   Step 1 of 6       You will need the temperature verification kit  TVK    Connect the TVK and turn it on now        6606660606   OOO0O0000000  000000000   D000000000   OOOO0000000   l OO00000000    Touch the right arrow when ready       3  When the test is complete  record zonal temperatures in the  Data Sheet   Calibration Verification Test  on page 127     Nx    Temperature Verification Test   Step 6 of 6    X    Left Zone                  Right Zone       Specification         0 75   C    Press the cross button when you want to exit 2     4  Press X to return to the Tools menu     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  121  and Veriti   System User Guide    Chapter 10 Calibration verification test    Ending the test     1  Ifthe heated cover is closed  lift the lever and place the heated cover in the fully  opened position       wr  PHYSICAL INJURY HAZARD  Hot Surface  Use care    when working around the heated cover and sample block to avoid being burned  by hot components     2  Wait for the sample block to reach room temperature  725   C   then remove the  the 0 5 mL probe assembly and thermal isolat
10.   Model 4500     3  Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature    field     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Enter actual block temperature    F1 F2 F3 FA F5       Note  The digital thermometer displays a four digit value  round this off to  three digits before typing it in the Calibration Verification screen     Note  Record this value on the Calibration Verification Test Data Sheet   page 47  to keep a permanent record of the test     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  39  and Veriti  System User Guide    Chapter 4 Calibration verification test    4  Press Enter  The GeneAmp   9700 automatically begins the second reading   45   C setpoint   and the Calibration Verification screen opens with the setpoint  value displayed     Calibration Verification  Block temp   xx x  C Cover temp   xxx  C       Setpoint is 45  C  Cover must be within 1  C of 105  C    F1 F2 F3 F4 F5       Note  The cover must be at 105   C   1   C  It may take several minutes for the  system to ramp up     5  Repeat steps 2 through 4 for the second temperature reading     The GeneAmp   9700 evaluates the calibration of the sample block temperature  for the setpoint values that you entered and displays the results  A summary  screen opens at the end of the test     Calibration Verification    Actual temperature at 85  C       
11.   Round this off to three  digits before typing it in the Calibration Verification screen     Record this value on the Calibration Verification Test Data Sheet  page 18  to  keep a permanent record of the test   4  Press Enter     The thermal cycler automatically begins the second temperature reading  45   C  setpoint   The Calibration Verification screen displays the setpoint value     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 45  C  Cover must be within 30  C of 45  C    F1 F2 F3 F4 F5       Note  The cover must be at 45   C   30   C  It may take several minutes for the  system to ramp up     5  Repeat steps 2 through 4 for the second temperature reading     The thermal cycler evaluates the calibration of the sample block temperature for  the setpoint values that you entered  then displays the results  A summary  section opens when the test ends     Calibration Verification    Actual temperature at 85  C       Actual temperature at 45  C    F1 F2 F3 FA F5       6  If you entered values on the Calibration Verification Test Data Sheet  page 18    compare those values with the actual test results     7  Press F1  Accept      Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  11  and Veriti  System User Guide    Chapter 2 Temperature calibration verification test    12    Evaluating the When the thermal cycler completes the calibration verification test  one of two  results screens opens  dep
12.   xxx  C    Place probe in well xx    Press Run    Run    F1 F2 F3 F4 F5       8  Slide the heated cover back  and repeat steps 4 to 7 of    Setting up the 0 02 mL  probe assembly    on page 55 and steps 2 through 7 of this procedure  Complete  these steps for all 12 wells to be tested        Left Sample Block Right Sample Block       L1  L2  L3  L4  L5  and L6 R1  R2  R3  R4  R5  and R6                9  The GeneAmp   9700 evaluates the uniformity of the sample block temperature  for the setpoint values that you entered and displays the results  A summary  screen opens at the end of the test                    F1 F2 F3 F4 F5    If you entered values on the Temperature Non Uniformity Test Data Sheet   compare those values with the actual test results     10  Press F1  Accept   To interpret the results  see    Evaluating the results    on  page 59     58 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 5 Temperature non uniformity test    11  When you complete all measurements  be sure to       Press Cancel       Remove the 0 02 mL probe assembly from the sample blocks     Power off the digital thermometer  then clean off the oil     Remove the 384 well temperature verification plates from the sample    blocks     IMPORTANT  Make sure the sample blocks are at room temperature   25   C   before removing the reaction plates     Evaluating the When the GeneAmp   9700 completes the temperatu
13.  006 ce cee cee nen eae  Data Sheet  Temperature Non Uniformity Test    llle ee eee    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 6 Calibration verification test    Calibration verification test    Required  equipment    Setting up the  20 uL probe  assembly    64    Use the calibration verification test to verify the temperature calibration of your  GeneAmp  PCR System 9700  Auto Lid Sample Block Module  The verification  test consists of the following procedures  which you must perform in sequence     Setting up the 20 uL probe assembly  csse eo eese Rec a a RN 64  Configuring the GeneAmp   9700        sss e 65  I4 nuQ  ODIUM 66  Evaluating the desull amp   ois esos oh ae rU beoe doe ete Rete PSP dd 68      Nr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the Dual 384 Well Sample Block Module Temperature Verification  Kit  PN 4308354   which includes       Light mineral oil     Two 384 well temperature verification plates     20 uL probe assembly     Digital thermometer with a 9V battery installed    Note  This kit is also used for the Auto Lid sample block module   1  Make sure the heated cover is in the rear position  If not  press and release the  red button to automatically move the heated cover to the rear position  2  Place both 384 well temperature veri
14.  5800   Toll Free  In North America    1 800 345 5224  Fax   1 650 638 5884    08 2009      Applied  E   Bibsystems    Part Number 4411933 Rev C    
15.  C is xx xx   Fail        NU at 37  C is xx xx   Fail          F1 F2 F3 F4 F5    e  fthe test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors  were not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See  How to obtain support  on page viii           30 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Ending the test 1     Rn duas GEO    Chapter 3 Temperature non uniformity test    Press F5  Cancel    Remove the 0 2 mL probe assembly from the sample block   Power off the digital thermometer  then clean off the oil     Wait for the sample block to reach room temperature   25   C   then remove the  9700 temperature verification frame from the sample block       wr  PHYSICAL INJURY HAZARD  Hot Surface  Use care    when working around the heated cover and sample block to avoid being burned  by hot components       Clean the oil off the sample block     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  31    and Veriti  System User Guide    Chapter 3 Data Sheet  Calibration Verification Test    Data Sheet  Calibration Verification Test    32    When you run the calibration verification test  record the setpoint values for well A6  on this data sheet  At the end of the test  compare the values that are displayed on the  thermal cycler with the values that are reco
16.  Continue with  Configuring the GeneAmp  9700  below   Configuring the 1  Complete the procedures in    Setting up the 0 5 mL probe assembly    above       GeneAmp  9700 2  Power on the thermal cycler  The Main menu opens   For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test   Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  83    Use this test to verify the temperature uniformity of the GeneAmp  PCR System  9700  0 5 mL Sample Block Module  The temperature non uniformity test consists  of the following procedures  which you must perform in sequence     Setting up the 0 5 mL probe assembly   02 5    4c veda etes ey E Rs 83  FORTIS HE TES ET  84  Evaluating the tesulis Lo ned phe hs he bbe ne 1 e E Ru P RE DE DEI ee 87    This test requires the 0 5 mL Sample Block Module Temperature Verification Kit   PN 4309924   which includes      Cotton swabs     Light mineral oil     GeneAmp  0 5 mL temperature verification frame     0 5 mL probe assembly     Digital thermometer Model 4500 with a 9V battery installed    1  Ifthe heated cover is in the forward position  lift the lever  then slide the heated    and Veriti  System User Guide    Chapter 7 Temperature non uniformity test    3  Press F4  Util   The Utilities screen opens   4  Press F1  Diag   The Diagnostics screen opens     5  Press F3  TmpVer   The Temperature Verification screen opens     Temperature Verifica
17.  DARE e    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 1 About this manual    About this manual    This manual describes the Temperature Verification System and the tests for which it  is used on the following thermal cyclers       2720 Thermal Cycler      GeneAmp  PCR System 9600    GeneAmp  PCR System 9700  0 5 mL Sample Block Module    GeneAmp  PCR System 9700  96 Well Sample Block Module      GeneAmp  PCR System 9700  Dual 96 Well Sample Block Module    GeneAmp  PCR System 9700  Dual 384 Well Sample Block Module      GeneAmp  PCR System 9800  e Applied Biosystems Veriti   96  Well Fast Thermal Cycler    This manual contains a separate chapter for each instrument  Each instrument  chapter provides step by step instructions for performing two different tests  which  are described in    Temperature verification tests  on page 4     About the Temperature Verification System    Description The Temperature Verification System is a kit consisting of a probe  a digital  thermometer  frames to prevent ambient air from entering the sample block  light  mineral oil  and cotton swabs  The system is used to verify temperatures of the  sample block on your thermal cycler     The table below indicates which Temperature Verification System kit you should    order for your thermal cycler     Figure 1 on page 3 shows the major components of the 0 2 mL 96 Well Temperature    Verification Kit     T
18.  Model 4500     4  Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature  field     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Enter actual block temperature    Cancel       F1 F2 F3 FA F5    Note  The digital thermometer displays a four digit value  round this off to  three digits before typing it in the Calibration Verification screen     Note  Record this value on the Calibration Verification Test Data Sheet   page 88  to keep a permanent record of the test     80 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 7 Calibration verification test    5  Press ENTER  The GeneAmp   9700 automatically begins the second reading   45   C setpoint   The Calibration Verification screen opens with the setpoint  value displayed     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 45  C  Cover must be within 30  C of 45  C    F1 F2 F3 F4 F5       Note  The cover must be within  1   C of 105   C   6  Repeat steps 3 and 4 for the second reading     7  The GeneAmp   9700 evaluates the calibration of the sample block temperature  for the setpoint values that you entered and displays the results  A summary  screen opens at the end of the test     Calibration Verification    Actual temperature at 85  C       Actual temperature at 45  C    F1 F2 F3 FA F5       If you enter
19.  Serial Communication Cable    1  Power on the Veriti   Thermal Cycler  For best results  allow the thermal cycler  15 to 30 minutes to warm to ambient temperature before performing the test   2  In the Main Menu  select Tools Menu     3  Connect the 4500 Digital Thermometer     a  Attach the free end of cable  from the probe assembly  to the input  connector port on the 4500 thermometer     b  Attach the 9 pin DIN female end of the RS232 cable to the thermometer  communication port  and the 9 pin DIN male end to the communication  port of the instrument        4  Ifthe heated cover is closed  lift the lever and place the heated cover in the fully  open position     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  123  and Veriti  System User Guide    Chapter 10 Temperature non uniformity test    5  Place the 0 5 mL Thermal Isolation Frame onto the sample block        6  Using a cotton swab  coat the following wells with mineral oil   Al  A10  B3  B8  E3  E8  F1  F10    123456789 10       OE    AAR      OQOQG  QOGQOQOQOOO       7  Place the 0 5 mL probe assembly into well Al  Thread the probe wire through  the channel in the temperature verification frame to prevent damage to the probe  assembly and lead wires  see Figure 19 on page 118      IMPORTANT  Seat the probe properly and the close the heated cover carefully   If the probe wire is crushed when you close the heated cover  the probe can be  damaged     8  Power on the digi
20.  Sheet  Calibration Verification Test    i 0020  000 cae sane ceased eben  Data Sheet  Temperature Non Uniformity Test    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti   System User Guide    91    Chapter8 Calibration verification test    Calibration verification test    Required  equipment    Setting up the  0 1 mL probe  assembly    Use the calibration verification test to verify the temperature calibration of your  Applied Biosystems GenAmp 9800 Fast Thermal Cycler with 96 Well Aluminum  Sample Block Module  The calibration verification test consists of the following  procedures  which you must perform in sequence     Setting up the 0 1 mL probe assembly     1 0 20    00 ce cece cette 92  Configuring the GeneAmp   9800          sss 93  Ronee the 1680s EMO 93  Pevalwatiie the resul  o ck anche hoe ae he ogee s ec a dee Pe a ee os 96    This test requires the following items in your Temperature Verification kit   PN 4351630        Digital thermometer Model 4500 with a 9V battery installed    Cotton swabs     Light mineral oil     0 I mL temperature verification probe assembly    1  Ifthe heated cover is in the forward position  lift the lever  then slide the heated  cover back     2  With the instrument powered off and the sample block at room temperature   place the 0 1 mL temperature verification probe assembly on the sample blocks     Make sure the undercut of the 0 1 mL temperature verification probe assemb
21.  and 2720 Thermal Cycler     Temperature calibration verification test     Checks the temperature of the  sample block against specifications for temperature accuracy    Temperature non uniformity test     Checks the uniformity of sample block  temperature from well to well    You can perform these tests as frequently as necessary  depending on the type of  work your laboratory performs  For example  a laboratory that does forensic typing  may need to validate the instrument before or after running samples to have the test  results considered valid in a court case  Another laboratory may want to validate a  protocol     Typically  the temperature calibration verification test should be performed at least  once a year  and the temperature non uniformity test should be run at the same time  and frequency     Running these tests verifies that the thermal cycler is operating to Applied  Biosystems operating specifications     The kit is supplied with a storage case  Figure 2  that includes       Model 4500 digital thermometer with a 9V battery installed    Probe assembly     Cotton swabs     Light mineral oil     Temperature verification frames       Figure 2 Temperature Verification Kit    Figures 1 and 2 show the major components of the kit     Note  If any part is damaged or missing  contact the shipping carrier and Applied  Biosystems immediately  See    Damages  claims  and returns  on page 134     Applied Biosystems recommends that your Temperature Verification System be  r
22.  applies to all Applied Biosystems electrical and electronic  symbols on products placed on the European market after August 13  2005              instruments  Symbol Description  Do not dispose of this product as unsorted municipal waste   E Follow local municipal waste ordinances for proper disposal provisions    c   to reduce the environmental impact of waste electrical and electronic  equipment  WEEE    io European Union customers   Call your local Applied Biosystems Customer Service office for  ES equipment pick up and recycling  See www appliedbiosystems com  for a list of customer service offices in the European Union   Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  139    and Veriti   System User Guide       Appendix D Safety    140 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Index    C    Calibration Verification Test  system 2720 9   12  system 9600 107   110  system 9700 23   26  system 9700 with 0 5 mL sample block module 78   81  system 9700 with Auto Lid sample block  module 64   67  system 9700 with dual 384 well sample block  module 50   53  system 9700 with dual 96 well sample block  module 37   41  system 9800 92   95  system Veriti 119   122  CAUTION  description vii  claims  processing 134  conventions  bold text viii  for describing menu commands viii  IMPORTANTS  viii  in this guide viii  italic text viii  Notes viii  u
23.  assembly             0 0 0 c cece eee eee eeeee 55  RHE Motes e haa ere Di hr Eee uetdo Suida eg Ade ST  Evaluating the results i  erorte nereis aKu E rk deen ee eee et ES 59    This test requires the Dual 384 Well Sample Block Module Temperature Verification  Kit  PN 4308354   which includes      Light mineral oil     Left and right 384 well temperature verification plates     0 02 mL probe assembly     Digital thermometer Model 4500 with a 9V battery installed    1  If the heated cover is in the forward position  lift the lever  then slide the heated  cover back   2  Place both 384 well temperature verification plates on the sample blocks     IMPORTANT  The temperature verification plates must be installed in both  sample blocks throughout all temperature verification testing     The plate labeled L1 to L6 is for the  eft sample block  Figure 12   The plate  labeled R1 to R6 is for the right sample block  Figure 13   Position the plates so  the angled corners face away from you        Figure 12 Left temperature Figure 13 Right temperature  verification plate verification plate    Note  The plates help you locate the probes within the sample blocks     3  Use the stick end of a cotton swab to coat wells L1 to L6 and R1 to R6 with  mineral oil  the wells are too small for the swab end     4  Place the 0 02 mL probe assembly into well L1     Note  As the test progresses  you move the 0 02 mL probe assembly to each of  the test wells  L1 to L6 and R1 to R6      Thermal Cycler 
24.  be within 1  C of 105  C    F1 F2 F3 FA F5       Note  The cover must be within  1   C of 105   C  It may take several minutes  for the GeneAmp  9700 to ramp up   2  During the calibration     a  The Calibration Verification screen counts down the time until the setpoint  is reached     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Stabilizing at setpoint    x xx    F1 F2 F3 FA F5       b  When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer   Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     3  Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature  field     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Enter actual block temperature    F1 F2 F3 F4 F5       Note  The digital thermometer displays a four digit value  round this off to  three digits before typing it in the Calibration Verification screen     Note  To keep a permanent record of the test  record this value on the  Calibration Verification Test Data Sheet  page 74      66 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter6 Calibration verification test    4  Press Enter  The GeneAmp   9700 automatically begins the second reading   45   C setpoint   The Calibration Verification screen opens 
25.  calibrated    Calibration Verification screen displays the following message     Calibration Verification    Calibration is good          F1 F2 F3 F4 ES       Does not pass the  calibration verification  test             Calibration Verification screen displays the following message     Calibration Verification    Instrument may require service   Contact Applied Biosystems  Technical Support        F1 F2 F3 F4 F5       e Ifthe test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors were  not made     e  fthe test fails again  contact Applied Biosystems Technical  Support  See    How to obtain support  on page viii        Note  As a result of the Temperature Tolerance Stack  between the sample block   temperature probe and thermometer  the instrument firmware is programmed to  accept a Pass Fail result of  0 75   C  Therefore  passing results may vary from  84 25   C to 85 75   C at the 85 degree setpoint and 44 25   C and 45 75   C at the 45   degree setpoint  Ranges greater than   0 75 degrees cause the test to fail     Ending the test 1  Press F5  Exit         2  Remove the 0 2 mL probe assembly from the sample block     3  Power off the digital thermometer  then clean off the oil     Wait for the sample block to reach room temperature  725   C   then remove the  9700 temperature verification frame from the sample block       rr  PHYSICAL INJURY HAZARD  Hot Surface  Use care    when working around the heated cover and s
26.  test  compare the values that are displayed on the  thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date Tested by   Probe serial no     Thermometer  serial no     Setpoint value  Well A6       85   C 45  C                                                                                              Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 2 Data Sheet  Temperature Non Uniformity Test    Data Sheet  Temperature Non Uniformity Test    When you run the temperature non uniformity test  record the setpoint values for the  wells that are listed on this data sheet  At the end of the test  compare the values that  are displayed on the thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date       Tested by       Probe serial no        Thermometer  serial no           Setpoint value 94  C 37   C       A1       A12       C4       c9       F4       F9       H1       H12                   Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 2 Data Sheet  Temperature Non Uniformity Test    20 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Temperature Verification Tests for the  GeneAmp  9700 Thermal Cycler    This chapter cove
27.  to the communication  port of the instrument    TVK serial  comm  cable  Euteghnics  4  Ifthe heated cover is closed  lift the lever and place the heated cover in the fully  open position   Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  119    and Veriti  System User Guide    Chapter 10 Calibration verification test    5  Place the 0 5 mL Thermal Isolation Frame onto the sample block        6  Using a cotton swab  coat the following wells with mineral oil   B3  B8  F1  F10    123456789 10          7  Place the 0 5 mL probe assembly into well B3  Thread the probe assembly wire  through the channel in the temperature verification frame to prevent damage to  the probe assembly and lead wires  see Figure 19 on page 118      IMPORTANT  Seat the probe properly and the close the heated cover carefully   If the probe wire is crushed when you close the heated cover  the probe can be  damaged     8  Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     9  Make sure the probe assembly is connected to the digital thermometer and that  the thermometer is powered on     10  Slide the heated cover forward and pull the lever down     120 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 10 Calibration verification test    Running the test     1
28. 00  9800   and Veriti  System User Guide    Chapter 7 Calibration verification test    Calibration verification test    Required  equipment    Setting up the  0 5 mL probe  assembly    Use the calibration verification test to verify the temperature calibration of your  GeneAmp  PCR System 9700  0 5 mL Sample Block Module  The calibration  verification test consists of the following procedures  which you must perform in  sequence     Setting up the 0 5 mL probe assembly     0 0 0 0    0c ce cece eee eens 78  Configuring the GeneAmp   9700    0 0    cece ccc ce n 79  Euonning tbe Test  iu oae eee Rawle Oh er Oh ave X deed 80  Evaliti  g the resul  o deca cess hk eek ae erae aeger sec tede me DR Re d p a 82    This test requires the 0 5 mL Sample Block Module Temperature Verification Kit   PN 4309924   which includes       Cotton swabs     Light mineral oil     GeneAmp  0 5 mL temperature verification frame     0 5 mL probe assembly     Digital thermometer Model 4500 with a 9V battery installed    1  Ifthe heated cover is in the forward position  lift the lever  then slide the heated  cover back     2  Place the GeneAmp  0 5 mL temperature verification frame on the sample  block     3  Usea cotton swab to coat well A6 with mineral oil   4  Place the 0 5 mL probe assembly into well A6     5  Thread the probe wire through the channel in the 0 5 mL temperature  verification frame to prevent damage to the probe and lead wires     6  Make sure the probe is connected to the digital th
29. 10   C of the sample block  temperature  the following display opens     Wait 3 minutes   Time MM SS Blk 40 0C  This display shows the current sample block temperature     Blk 40 0C     and a  clock that counts up from 0 in minutes and seconds   Time MM SS       When the clock reaches 3 minutes  the following display opens     Record Temperature  Time MM SS Blk 95 0C      Measure the temperature of well E1 using the digital thermometer  Record this    temperature as T 40      108 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter9 Calibration verification test    6  Press ENTER     The temperature of the sample block and heated cover go to 95  C  and the  following display opens     Cvr   xxC Blk   xx xC  This display shows the current temperature of the block cover  Cvr   xxC  and  sample block  BIk   xx xC      When the temperature of the block cover is within  10   C of the sample block  temperature  the following display opens     Wait 3 Minutes  Time MM SS Blk 95 0C    This display shows the current sample block temperature   BIk 95 0C   and a  clock that counts up from 0 in minutes and seconds     Time MM SS         When the clock reaches 3 minutes  the following display opens     Record Temperature  Time MM SS Blk 95 0C  7  Measure the temperature of well E1 using the digital thermometer  Record this  temperature as T 95      IMPORTANT  To end the test at any time  press the STOP 
30. 3 Rev  C    08 2009    Contents    Preface    Introduction    Safety information            liliis hr vii  How to usehis  guide 2 25 50  cea Ier a a ade ER DES psu a Rd E viii  How to obtain Support  raire sonia RR ras viii  About this man  al   2  oix R Eu eR ERR LEL X Pa Ba e 2  About the Temperature Verification System               llle 2  Probe assembly 5  oi fa Gated oh we diet wee eru eng ee RUE A RR E EU d dd 5  Digital thermometer             0 000 cee lh 5    Temperature Verification Tests for the 2720 Thermal Cycler    OVERVIEWS  ca pale adui Sun aat ulia d Furor eue ter a tof ena ea cbe Pessoa Stig duse oad 8  Temperature calibration verification test     0 0 0    eee eee eens 9  Temperature non uniformity test    0 0    2 ee 13  Data Sheet  Calibration Verification Test            liliis 18  Data Sheet  Temperature Non Uniformity Test          0 0 0 0  e eee eee 19    Temperature Verification Tests for the GeneAmp   9700 Thermal Cycler    OVERVIEWS  ca nett dialtone il usd ue s auum dud uv et fe ele nate eae oe 22  Temperature calibration verification test    0 0    0  eee ee eee 23  Temperature non uniformity test    0 2    0  re 27  Data Sheet  Calibration Verification Test            liliis  32  Data Sheet  Temperature Non Uniformity Test             liliis 33    Temperature Verification Tests for the GeneAmp  9700  Dual 96 Well    Module   OVerVIeW  sz La de eset Patet ue t dad SAMO P UD RE Eum ced DE 36   Calibration verification test              lille ee 3
31. 5   C setpoint   The Calibration Verification screen opens with the setpoint  value displayed     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 45  C  Cover must be within 1  C of 105  C    F1 F2 F3 F4 F5       Note  The cover must be within  1   C of 105   C   6  Repeat steps 3 and 4 for the second reading     7  The GeneAmp   9700 evaluates the calibration of the sample block temperature  for the setpoint values that you entered and displays the results  A summary  screen opens at the end of the test     Calibration Verification    Actual temperature at 85  C       Actual temperature at 45  C    F1 F2 F3 FA F5       If you entered values on the Calibration Verification Test Data Sheet  compare  those values with the actual test results     8  Press F1  Accept   To interpret the results  see    Evaluating the results  on  page 54   9  When you complete all measurements  be sure to     Press Exit     Remove the 0 02 mL probe assembly from the sample blocks     Power off the digital thermometer  then clean off the oil       Remove the 384 well temperature verification plates from the sample  blocks       NITE BURN HAZARD  Make sure the sample blocks have cooled    at least to room temperature  less than 25  C  before removing the probe  The  blocks and probes retain heat longer than expected     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  53  and Veriti   System User Guide    Chapter 5 Ca
32. 6  c6 bcc e eect ew e ne 107  Runtime thedest    secos oes RS err Rue e el oe ol 108  Calculsinp test teslis    occ ove EAS Seed chr er ORA DER PX RUE d eiu 109  Ending Ube teal  es oka es Was Ge e esce do edd A DA geo s 110      Nr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the 0 2 mL Temperature Verification Kit  PN 4317939   which  includes      Cotton swabs     Light mineral oil     9600 temperature verification frame     0 2 mL probe assembly     Model 4500 digital thermometer with a 9V battery installed           If the heated cover is in the forward position  turn the knob counterclockwise  completely  then slide the heated cover back     Place the 9600 temperature verification frame on the sample block   Using a cotton swab  coat well E1 with mineral oil     Place the 0 2 mL probe assembly into well E1     A BR WON      Thread the probe wire through the channel in the 9600 temperature verification  frame to prevent damage to the probe and lead wires  see Figure 18 on  page 106      6  Make sure the probe is connected to the digital thermometer     7  Slide the heated cover forward  then turn the cover knob clockwise until the  white mark on the knob is aligned with the white mark on the cover     IMPORTANT  Seat the probe properly and the close the heated cover carefully   If the probe wire is crushed when you close the cover  the probe can be  
33. 7   Temperature non uniformity test    0 0    ee re 42   Data Sheet  Calibration Verification Test     2 2 0 0    auauua aaa 47  Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  iii    and Veriti   System User Guide    Data Sheet  Temperature Non Uniformity Test          0 00 00 cece eee 48    Temperature Verification Tests for the GeneAmp  9700  Dual 384 Well    Sample Block Module    Calibration verification test            llle 50  Temperature non uniformity test      0 0    ee ee 55  Data Sheet  Calibration Verification Test      0    0    eee ee 60  Data Sheet  Temperature Non Uniformity Test          0 0 0 0 c eee eee 61    Temperature Verification Tests for the GeneAmp   9700  Auto Lid Sample    Block Module    Calibration verification test sieer nesrin ens e a E E EE 64  Temperature non uniformity test      0 0    et eee 69  Data Sheet  Calibration Verification Test     2 0    unasuasua eaaa 74  Data Sheet  Temperature Non Uniformity Test          0 0 0 0 0 eee eee 75    Temperature Verification Tests for the GeneAmp   9700  0 5 mL Sample    Block Module    Calibration verification test oss 0    teens 78  Temperature non uniformity test     2 0    eee 83  Data Sheet  Calibration Verification Test      0    suaus aaua eaaa 88  Data Sheet  Temperature Non Uniformity Test          0 0 000 eee eee 89    Temperature Verification Tests for the GeneAmp   9800    Calibration verification test ses osred tenes 92  Temperature non 
34. 800  101    and Veriti  System User Guide    Chapter 8 Temperature non uniformity test    12  When you complete all measurements   a  Press F5  Cancel    b  Remove the 0 1 mL probe assembly from the sample blocks     c  Power off the digital thermometer  then clean off the oil     IMPORTANT  Make sure the sample blocks are at room temperature   25   C   before removing the reaction plates     Evaluating the When the GeneAmp   9800 completes the temperature non uniformity test  the TNU  results Performance screen opens  Use the table below to evaluate the results        If the       Then          Temperature of the  sample block wells is  uniform        Pass  appears after each setpoint temperature     TNU Performance    TNU at 95  C is xx xx   Pass    TNU at 55  C is xx xx   Pass    F1 F2 F3 FA F5       Temperature variation  of the sample block  wells exceeds  performance  specifications              Fail  appears after the setpoint temperature s  for which the  test failed     TNU Performance    TNU at 95  C is xx xx   Fail    TNU at 55  C is xx xx   Fail       F1 F2 F3 FA F5    e Ifthe test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors were  not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See    How to obtain support  on page viii           102 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Ch
35. 86  system 9700 with Auto Lid sample block  module 69   73  system 9700 with dual 384 well sample block  module 55   59  system 9700 with dual 96 well sample block  module 42   46  system 9800 97   102  system Veriti 123   126  Temperature Verification System 2   4  tests  Calibration Verification Test 9   12  23   26  37   41   64   68  78   82  92   96  119   122  Temperature Non Uniformity Test 13   17  27   31   42   46  55   59  69   73  83   87  977   102   123   126  text conventions viii  training  information on viii    U    user attention words  described viii    W    WARNING  description vii  warranty  damages  claims  returns 134  exceptions 133  limitations 133  period 132    warranty claims 132  warranty exceptions 133  warranty period  effective date 132          142 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti   System User Guide    www appliedbiosystems com    Worldwide Sales and Support    Applied Biosystems vast distribution and  service network  composed of highly trained  support and applications personnel    reaches 150 countries on six continents    For sales office locations and technical support   please call our local office or refer to our   Web site at www appliedbiosystems com     Applied Biosystems is committed to providing the  world s leading technology and information  for life scientists     Headquarters   850 Lincoln Centre Drive   Foster City  CA 94404 USA   Phone   1 650 638
36. Actual temperature at 45  C    F1 F2 F3 F4 F5       6  Ifyou entered values on the Calibration Verification Test Data Sheet  page 47    compare those values with the actual test results     7  Press F1  Accept      40 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 4 Calibration verification test    Evaluating the When the GeneAmp   9700 completes the calibration verification test  one of two  results screens opens  See the table below to evaluate the results        If the sample block  module       Then the          Is properly calibrated       Calibration Verification screen displays the following message     Calibration Verification    Calibration is good       F1 F2 F3 FA F5       Does not pass the  calibration verification  test          Calibration Verification screen displays the following message     Calibration Verification       Instrument may require servic  Contact Applied Biosystems  Technical Support           F1 F2 F3 FA F5          e If the test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors  were not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See  How to obtain support  on page viii        Ending the test        Press F5  Exit      Boom    Remove the 0 2 mL probe assembly from the sample block   Power off the digital thermometer  then clean off the oil     Wait for the samp
37. C  Place Probe in well L1  Press RUN          EUR                Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  43  and Veriti  System User Guide    Chapter 4 Temperature non uniformity test    Running the test The temperature non uniformity test uses the 0 2 mL probe assembly to test the  temperature uniformity of 6 different wells in each sample block     Note  Press F5  Cancel  if you want to exit the test     To run the temperature non uniformity test     1  Press F1  Run   The temperature non uniformity test starts  and the TNU  Performance screen opens with the setpoint value displayed     TNU Performance  Sample temp   xx x C Cover temp   xxx  C    Setpoint is 95  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 FA F5          Note  The sample block must be at 95   C   1 0   C  In addition  the cover must  be at 105   C  1 0   C  It may take several minutes for the GeneAmp   9700 to  stabilize at the setpoint temperature     The TNU Performance screen counts down the time until the setpoint is stable     TNU Performance    Sample temp   xx x C Cover temp   xxx  C    Stabilizing at setpoint X XX    F1 F2 F3 F4 F5       2  When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     3  Using the numeric keys  enter the value that is displayed on the digit
38. E  DELAY IN REPAIR OR REPLACEMENT  OR FOR LOSS OF  REVENUE OR PROFITS  LOSS OF GOOD WILL  LOSS OF BUSINESS  OR  OTHER FINANCIAL LOSS OR PERSONAL INJURY OR PROPERTY  DAMAGE     NO AGENT  EMPLOYEE  OR REPRESENTATIVE OF   APPLIED BIOSYSTEMS HAS ANY AUTHORITY TO MODIFY THE  TERMS OF THIS LIMITED WARRANTY STATEMENT OR TO BIND  APPLIED BIOSYSTEMS TO ANY AFFIRMATION  REPRESENTATION   OR WARRANTY CONCERNING THE PRODUCT THAT IS NOT  CONTAINED IN THIS LIMITED WARRANTY STATEMENT  AND ANY  SUCH MODIFICATION  AFFIRMATION  REPRESENTATION  OR  WARRANTY MADE BY ANY AGENT  EMPLOYEE  OR  REPRESENTATIVE OF APPLIED BIOSYSTEMS WILL NOT BE BINDING  ON APPLIED BIOSYSTEMS  UNLESS IN A WRITING SIGNED BY AN  EXECUTIVE OFFICER OF APPLIED BIOSYSTEMS     THIS WARRANTY IS LIMITED TO THE BUYER OF THE PRODUCT  FROM APPLIED BIOSYSTEMS AND IS NOT TRANSFERABLE     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  133  and Veriti   System User Guide    Appendix B    Limited product warranty    Some countries or jurisdictions limit the scope of or preclude limitations or exclusion  of warranties  of liability  such as liability for gross negligence or wilful misconduct   or of remedies or damages  as or to the extent set forth above  In such countries and  jurisdictions  the limitation or exclusion of warranties  liability  remedies or damages  set forth above shall apply to the fullest extent permitted by law  and shall not apply  to the extent prohibited by law   
39. F3 F4 F5       e If the test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors  were not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See  How to obtain support  on page viii        Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide       Chapter 7 Temperature non uniformity test    Temperature non uniformity test    Required  equipment    Setting up the    0 5 mL probe cover back   assembly 2  Place the GeneAmp  0 5 mL thermal insulation frame on the sample block   3  Use a cotton swab to coat the following wells with mineral oil   Al  A6  A10  B3  C5  C10  D1  D7  E8  F1  F5  F10  4  Place the 0 5 mL probe assembly into well A1   Note  As the test progresses  you move the 0 5 mL probe assembly to each of  the test wells   5  Thread the probe wire through the channel in the 0 5 mL temperature  verification frame to prevent damage to the probe and lead wires   6  Make sure the probe is connected to the digital thermometer   7  Slide the heated cover forward and pull the lever down   IMPORTANT  Seat the probe properly and close the heated cover carefully  If  the probe wire is crushed when you close the heated cover  the probe may be  damaged   8  Power on the digital thermometer   Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500   9 
40. GeneAmp   9800 completes the calibration verification test  one of two  results screens opens  See the table below to evaluate the results        If the sample block  module       Then the          Is properly calibrated    Calibration Verification screen displays the following message     Calibration Verification    Calibration is good       F1 F2 F3 F4 F5       Does not pass the  calibration verification  test             Calibration Verification screen displays the following message     Calibration Verification    Instrument may require service   Contact Applied Biosystems  Technical Support        F1 F2 F3 F4 F5    e Ifthe test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors  were not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See    How to obtain support  on page viii           Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 8 Temperature non uniformity test    Temperature non uniformity test    Required  equipment    Setting up the    Use this test to verify the temperature uniformity of the Applied Biosystems  GeneAmp   9800 Fast Thermal Cycler with 96 Well Aluminum Sample Block  Module  The temperature non uniformity test consists of the following procedures   which you must perform in sequence     Setting up the 0 1 mL probe assembly              00  eee cee eens 97  Configuring th
41. L         138 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Instrumentation safety    Safety symbols The following table describes the safety symbols that may be displayed on  Applied Biosystems instruments  Each symbol may appear by itself or with text that  explains the relevant hazard  These safety symbols may also appear next to  DANGERS  WARNINGS  and CAUTIONS that occur in the text of this and other  product support documents        Description       Indicates that you should consult the manual for further information  and to proceed with appropriate caution     Indicates the presence of an electrical shock hazard and to proceed  with appropriate caution     Indicates the presence of a hot surface or other high temperature  hazard and to proceed with appropriate caution     Indicates the presence of a laser inside the instrument and to proceed  with appropriate caution     Indicates the presence of moving parts and to proceed with  appropriate caution     Indicates the presence of a biological hazard and to proceed with  appropriate caution     Indicates the presence of a radiological hazard and to proceed with  appropriate caution     Indicates the presence of a slipping hazard and to proceed with  appropriate caution     Indicates the presence of an ultraviolet light  in the instrument   and to  proceed with appropriate caution     PPPPPPPDP  I       Environmental The following symbol
42. L6 and R1 to R6 with  mineral oil  the wells are too small for the swab end     4  Place the 20 uL probe assembly into well L1     Note  As the test progresses  move the 20 uL probe assembly to each of the test  wells  L1 to L6 and R1 to R6      5  Thread the probe wire through the channels in the 384 well temperature  verification plate to prevent damage to the probe and lead wires     6  Make sure the probe is connected to the digital thermometer   7  Press and release the red button to automatically close the heated cover     IMPORTANT  Seat the probe properly  If the probe wire 1s crushed when the  heated cover closes  the probe may be damaged     8  Power on the digital thermometer   Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500   Configuring the 1  Power on the thermal cycler  The Main menu opens         GeneAmp      9700 For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test     2  Press F4  Util   The Utilities screen opens   3  Press F1  Diag   The Diagnostics screen opens     4  Press F3  TmpVer   The Temperature Verification screen opens     Temperature Verification    Temp   Calibration Verification   NU   Temperature Non Uniformity      Temp     TNU      F1 F2 F3 F4 F5                   5  Press F2  TNU  to automatically configure the GeneAmp   9700 for the  temperature non uniformity test  The TNU Performance screen op
43. NU      The thermal cycler automatically starts to configure the temperature non   uniformity test  with the cover setpoint of 35   C and the sample starting with the  setpoint of 37   C  It may take several minutes for the system to stabilize at the  cover and sample setpoint temperatures  After the system is configured  the  TNU Performance screen opens     TNU Performance  Sample temp   xx x C Cover temp   xxx C    Place Probe in well C4  Press Run    Run    F1 F2 F3 F4 F5          Running the test The temperature non uniformity test uses the 0 2 mL probe assembly to test the  temperature uniformity of 8 different wells in the sample block     Note  Press F5  Cancel  if you want to exit the test     To run the temperature non uniformity test     1  Place the probe into the indicated well  then press F1  Run   The temperature  non uniformity test for the setpoint temperature of 94   C starts  and the TNU  Performance screen opens with the setpoint value displayed     TNU Performance  Sample temp   xx x C Cover temp   xxx C    Setpoint is 94  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 F4 F5          Note  The sample block increases from 37   C to 94   C  The sample block must  be at 94   C 1 0   C  It may take several minutes for the GeneAmp   9700  thermal cycler to stabilize at the setpoint temperature        28 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 3 Temperat
44. Setpoint is 85  C  Cover must be within 10  C of 85  C    F1 F2 F3 FA F5       The Calibration Verification screen counts down the time until the setpoint is  reached     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Stabilizing at setpoint    x xx    F1 F2 F3 F4 F5       Note  The cover must be within  10   C of 85   C  It may take several minutes  for the GeneAmp   9800 to ramp up     3  When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     4  Using the numeric keys  enter the value that is displayed on the digital  thermometer into the    Enter actual block temperature    field     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Enter actual block temperature    F1 F2 F3 F4 F5       Note  Round the four digit value on the digital thermometer to three digits  before you enter it into the Calibration Verification screen     Note  You can record this value on the Calibration Verification Test Data Sheet   page 103  to keep a permanent record of the test     94 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter8 Calibration verification test    5  Press Enter  The GeneAmp   9800 begins the second reading  45   C setpoint    and the Calibration Verification screen opens with t
45. Slide the heated cover back  and repeat steps 4 through 8 of    Setting up the 0 1   mL probe assembly    on page 97 and steps 2 through 7 of this procedure at  50   C  Complete these steps for all eight wells to be tested     Al  A12  C4  C9  F4 F9  H1  and H12    The following screen opens     TNU Performance    Sample temp   xx x  C Cover temp   xxx  C    Place probe in well Al  Press Run    Run    F1 F2 F3 FA F5       The GeneAmp   9800 then sets the heated cover setpoint to 105   C     Slide the heated cover back  then repeat steps 4 through 8 of    Setting up the 0 1   mL probe assembly    on page 97 and steps 2 through 7 of this procedure at  95   C for all eight wells being tested     Al  A12  C4  C9  F4  F9  H1  and H12    When the TNU Performance screen opens to display the setpoint temperature   press F5  Cancel      TNU Performance    Sample temp   xx x C Cover temp   xxx  C    Setpoint is 105  C  Sample must be 1 0  C of setpoint    F1 F2 F3 FA F5          The GeneAmp   9800 evaluates the uniformity of the sample block temperature  for the setpoint values that you entered and displays the results  A summary  screen opens at the end of the test           If you entered values on the Temperature Non Uniformity Test Data Sheet   compare those values with the actual test results     Press F1  Accept   To interpret the results  see    Evaluating the results    on  page 102     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9
46. Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  55  and Veriti  System User Guide    Chapter 5 Temperature non uniformity test    Configuring the  GeneAmp   9700      Thread the probe wire through the channels in the 384 well temperature    verification plate to prevent damage to the probe and lead wires       Make sure the probe is connected to the digital thermometer     Slide the heated cover forward and pull the lever down     IMPORTANT  Seat the probe properly and close the heated cover carefully  If  the probe wire is crushed when you close the heated cover  the probe may be  damaged       Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     Continue with    Configuring the GeneAmp   9700    below       Complete the procedures in    Setting up the 0 02 mL probe assembly    above       Power on the thermal cycler  The Main menu opens     For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test       Press F4  Util   The Utilities screen opens     Press F1  Diag   The Diagnostics screen opens       Press F3  TmpVer   The Temperature Verification screen opens     Temperature Verification    Temp   Calibration Verification   NU   Temperature Non Uniformity            Temp J   TNU jJ    F1 F2 F3 F4 F5             6  Press F2  TNU   This automatically configures the G
47. Thermal Cycler Temperature  Verification System    For GeneAmp  PCR System 2720  9600  9700  9800  and  Veriti  Thermal Cyclers    User Guide    E   Brem       Copyright 2009  Life Technologies Corporation  All rights reserved     Information in this document is subject to change without notice  Applied Biosystems assumes no responsibility for any errors that may appear in this  document     APPLIED BIOSYSTEMS DISCLAIMS ALL WARRANTIES WITH RESPECT TO THIS DOCUMENT  EXPRESSED OR IMPLIED  INCLUDING  BUT NOT LIMITED TO THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE  IN NO EVENT SHALL APPLIED  BIOSYSTEMS BE LIABLE  WHETHER IN CONTRACT  TORT  WARRANTY  OR UNDER ANY STATUTE OR ON ANY OTHER BASIS FOR  SPECIAL  INCIDENTAL  INDIRECT  PUNITIVE  MULTIPLE OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING  FROM THIS DOCUMENT  INCLUDING BUT NOT LIMITED TO THE USE THEREOF     NOTICE TO PURCHASER     PLEASE REFER TO THE SPECIFIC GENEAMP   PCR SYSTEM USER MANUAL FOR LIMITED LABEL LICENSE OR DISCLAIMER  INFORMATION     The Applied Biosystems Thermal Cycler Temperature Verification System is covered by U S  Patent No  5 224 778     TRADEMARKS     Trademarks of Life Technologies Corporation and its affiliated companies  AB  Design   Applied Biosystems   GeneAmp   Veriti    Windows and  Excel are registered trademarks of Microsoft Corporation in the United States and other countries     All other trademarks are the sole property of their respective owners     Part Number 441193
48. TmpwVer   Temperature Verification  Upgrad   Firmware Upgrade    Hard System  TmpVer Upgrad Exit  F1 F2 RE F4 F5     Screen display missing after F1  Diag  is pressed     5  Press F3  TmpVer  to open the Temperature Verification screen     Temperature Verification    Temp   Calibration Verification   NU   Temperature Non Uniformity          TNU    F1 F2 F3 FA F5         Temp          6  Press F1  Temp   This action opens the Calibration Verification screen and  configures the GeneAmp   9800 for the calibration verification test     Calibration Verification  Block temp   xx x C Cover temp   xxx C             Place probe in well A6  Press Run    Run    F1 F2 F3 FA F5       7  Continue with  Running the test  below     Running the test Use the digital thermometer to take temperature readings of the sample well that is  connected to the 0 1 mL probe assembly  You take a reading at each of two setpoint  temperatures     Note  If necessary  press F5  Cancel  to exit the test     To run the calibration verification test     1  Complete the procedures in    Configuring the GeneAmp   9800    above     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  93  and Veriti  System User Guide    Chapter8 Calibration verification test    2  Press F1  Run  to start the calibration verification test  The Calibration  Verification screen opens with the setpoint value displayed     Calibration Verification  Block temp   xx x C Cover temp   xxx C       
49. a is returned to Applied Biosystems within the ninety  90  day warranty  period  Applied Biosystems will replace the defective media     Applied Biosystems does not warrant that the operation of the instrument or its  operating software will be uninterrupted or error free     Any applicable warranty period under these sections begins on the earlier of the date  of installation or ninety  90  days from the date of shipment for hardware and  software installed by Applied Biosystems personnel  For all hardware and software  installed by the buyer or anyone other than Applied Biosystems  and for all other  products  the applicable warranty period begins the date the product is delivered to  the buyer     Warranty claims must be made within the applicable warranty period  or  for  chemicals or other consumable products  within thirty  30  days after receipt by the  buyer     132 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Appendix B Limited product warranty    Warranty The above warranties do not apply to defects resulting from misuse  neglect  or   exceptions accident  including without limitation  operation with incompatible solvents or  samples in the system  operation outside of the environmental or use specifications  or not in conformance with the instructions for the instrument system  software  or  accessories  improper or inadequate maintenance by the user  installation of software  o
50. able1 Thermal cycles and applicable kit types                         Cycler Temperature Verification Kit       Thermal cycler Temperature Verification Kit PN   e GeneAmp  9700  96 Well 0 2 mL Single Channel 96 Well 4317939   e GeneAmp  9700  Dual 96 Well   Temperature Verification Kit   e 2720   e GeneAmp   9600   GeneAmp   9700  Dual 384 Well 0 02 mL Single Channel Dual 384 Well 4308354  Temperature Verification Kit   e GeneAmp   9700  60 Well 0 5 mL Single Channel 60 Well 4309924   e Veriti   Temperature Verification Kit   GeneAmp   9800 0 1 mL Single Channel 9800 Fast Thermal   4351630    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti   System User Guide       Chapter 1 About this manual       Digital thermometer    Frame for GeneAmp  9700  9600   and 2720 thermal cycler                                     T O0  m m OO uU _  gt                             GR1918    Probe assembly           Plate for GeneAmp  9700 with  dual 96 well sample block module                                  Figure 1 Components of the 0 2 mL 96 Well Temperature Verification Kit    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  3  and Veriti  System User Guide    Chapter 1 About this manual    Temperature  verification tests    Kit parts    Recalibration    You use the Temperature Verification System to perform the following tests on your  GeneAmp  PCR Systems  Veriti   Thermal Cycler 
51. ace  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     Required This test requires the Dual 384 Well Sample Block Module Temperature Verification  equipment Kit  PN 4308354   which includes     Light mineral oil    Two 384 well temperature verification plates    20 uL probe assembly    Digital thermometer with a 9V battery installed    Note  This kit is also used for the Auto Lid sample block module     Setting up the 1  Make sure the heated cover is in the rear position  If not  press and release the  20 uL probe red button to automatically move the heated cover to the rear position     assembly 2  Place both 384 well temperature verification plates on the sample blocks     IMPORTANT  The temperature verification plates must be installed in both  sample blocks throughout all temperature verification testing     The plate labeled L1 to L6 is for the  eft sample block  Fig 14   The plate  labeled with R1 to R6 is for the right sample block  Fig 15   Position the plates  so the angled corners face away from you        Figure 16 Left temperature Figure 17 Right temperature  verification plate verification plate    Note  The plates help you locate the probes within the sample blocks     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  69  and Veriti  System User Guide    Chapter 6 Temperature non uniformity test    3  Use the stick end of a cotton swab to coat wells L1 to 
52. al  thermometer in the    Enter actual block temperature    field     TNU Performance    Sample temp   xx x C Cover temp   xxx C       Enter actual block temperature 00 0      F1 F2 F3 FA F5       Note  The digital thermometer displays a four digit value  round this off to  three digits before typing it in the TNU Performance screen     Note  Record this value on the Temperature Non Uniformity Test Data Sheet   page 48  to keep a permanent record of the test     44 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 4 Temperature non uniformity test    4  Press Enter  The GeneAmp   9700 automatically begins the second reading   55   C setpoint   and the TNU Performance screen opens with the setpoint value  displayed     TNU Performance    Sample temp   xx x  C Cover temp   xxx  C    Setpoint is 55  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 FA F5       Note  The sample block must be at 55   C  1 0   C  In addition  the cover must  be at 105   C  1 0   C  It may take several minutes for the GeneAmp   9700 to  stabilize at the setpoint temperature     5  Repeat steps 2 and 3 for the second reading   6  Press Enter  The TNU Performance screen opens with the following prompt     TNU Performance  Sample temp   xx x  C Cover temp   xxx  C    Place probe in well xx    Press Run    Run    F1 F2 F3 F4 F5       7  For the GeneAmp   9700 with Manual Dual 96 Well Sample Block Module   Slid
53. alue  Well R3       85   C 45  C                                                                                              Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 5 Data Sheet  Temperature Non Uniformity Test    Data Sheet  Temperature Non Uniformity Test    When running the temperature non uniformity test  record the setpoint values for the  wells listed on this data sheet  At the end of the test  compare the values that are  displayed on the thermal cycler with the values that are recorded here     Note  You can photocopy this page                                                                             Date   Tested by   Probe serial no    Thermometer   serial no    Setpoint value 94  C 37   C  L1  L2  L3  L4  L5  L6  R1  R2  R3  R4  R5  R6  Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  61    and Veriti  System User Guide    Chapter 5 Data Sheet  Temperature Non Uniformity Test    62 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Temperature Verification Tests for the GeneAmp   9700  Auto Lid Sample Block Module    This chapter covers    Calibration verification 168t ciae se aes ee os d ERES ER 1 RES ROSES HEEL ODE  Temperature non uniformity testi us dda ed a ed eR UR KR ER hob aeg  Data Sheet  Calibration Verification Test            
54. ample block to avoid being burned    by hot components     4  Clean the oil off the sample block     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide       Chapter 3 Temperature non uniformity test    Temperature non uniformity test    Required  equipment    Setting up the    Use this test to verify the temperature uniformity of the GeneAmp 9700 thermal  cycler with a 0 2 mL sample block module  The test consists of the following  procedures  which you must perform in sequence     Setting up the 0 2 mL probe assembly scr ccs cece eee ee E Ree 27  Configuring the GeneAmp  9700        llis e 28  Bunning durtest  ooo eR bL ce IDA AP PP eerie EF ie bee ee 28  Exaluatidae the regulis esoo eies Ez deb eer e bod genau o eee deae enu 30  Ending ihe desk    ecco ss eec n Der qme sea c gere 31      Nr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the 0 2 mL Sample Block Module Temperature Verification Kit   PN 4317939   which includes      Cotton swabs     Light mineral oil     9700 temperature verification frame     0 2 mL probe assembly     Model 4500 digital thermometer with a 9V battery installed    1  Ifthe heated cover is in the forward position  lift the lever  then slide the heated    0 2 mL probe cover back   assembly 2  Place the 9700 temperature verification frame on the samp
55. apter 8 Data Sheet  Calibration Verification Test    Data Sheet  Calibration Verification Test    When running the calibration verification test  record the setpoint values for well A6  on this data sheet  At the end of the test  compare the values that are displayed on the  thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date Tested by   Probe serial no     Thermometer  serial no     Setpoint value  Well A6       85   C 45  C                                                                                              Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  103    and Veriti   System User Guide    Chapter 8 Data Sheet  Temperature Non Uniformity Test    Data Sheet  Temperature Non Uniformity Test    When running the temperature non uniformity test  record the setpoint values for  wells A1  A12  C4  C9  F4  F9  H1  and H12 on this data sheet  At the end of the test   compare the values that are displayed on the thermal cycler with the values that are    104    recorded here     Note  You can photocopy this page        Date       Tested by       Probe serial no        Thermometer  serial no           Setpoint value    95   C    55   C       A1       A12       C4       c9       F4       F9       H1       H12                   Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti   System User Guide    Temperature Ver
56. ature of the  sample block wells is  uniform        Pass  appears after each setpoint temperature     TNU Performance    TNU at 94  C is xx xx   Pass    TNU at 37  C is xx xx   Pass    F1 F2 F3 FA F5       Temperature variation of  the sample block wells  exceeds performance  specifications           Fail  appears after the setpoint temperature s  for which the  test failed     TNU Performance    TNU at 94  C is xx xx   Fail    TNU at 37  C is xx xx   Fail          F1 F2 F3 FA F5      If the test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors were  not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See    How to obtain support  on page viii           Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  73    and Veriti  System User Guide    Chapter 6 Data Sheet  Calibration Verification Test    Data Sheet  Calibration Verification Test    74    When running the calibration verification test  record the setpoint values for well R3  on this data sheet  At the end of the test  compare the values that are displayed on the  thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date Tested by   Probe serial no     Thermometer  serial no     Setpoint value  Well R3       85   C 45  C                                                                                              Thermal Cycler Temperature Ve
57. block module  The test consists of the following  procedures  which you must perform in sequence     Setting up the 0 2 mL probe assembly      cessus eee RR 23  Configuring the GeneAmp  9700         csse e 23  Bunning the test     025 ceci one AREE ER UD PUPS HEVATRR E EET e  24  Exaluatine the regulis uu codex Re dete dee Oeo d ge ae e S Ded Ras 26  Ending ibetesk   2 co ctis cbr RR dr SR ORI ee RE ee E EYES 26      Nr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the 0 2 mL Sample Block Module Temperature Verification Kit   PN 4317939   which includes      Cotton swabs     Light mineral oil     9700 temperature verification frame     0 2 mL probe assembly     Model 4500 digital thermometer with a 9V battery installed            If the heated cover is in the forward position  lift the lever  then slide the heated  cover back     Place the 9700 temperature verification frame on the sample block   Using a cotton swab  coat well A6 with mineral oil     Place the 0 2 mL probe assembly into well A6     A BR wWoN      Thread the probe wire through the channel in the 9700 temperature verification  frame to prevent damage to the probe and lead wires  see Figure 5 on page 22      6  Make sure the probe is connected to the digital thermometer   7  Slide the heated cover forward and pull the lever down     IMPORTANT  Seat the probe properly and close the heated cove
58. cation Tests for the  2 20 Thermal Cycler    This chapter covers     oi EP 8  Temperature calibration verification tests iis osa enar A X deg 9  Temperature non uniform  ty  Sls snos ek var sgoh bad us page up dace 13  Data Sheet  Calibration Verification Test      ic    00500  eb e E n 18  Data Sheet  Temperature Non Uniformity Test          llle  19  Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600 9700 9800  7    and Veriti  System User Guide    Chapter 2 Overview    Overview    Materials When you perform a temperature verification test on a 2720 thermal cycler  use the  materials in the kit  PN 4317939      Assembly Figure 3 shows a properly assembled system  Note that the frame is positioned  around the wells with the channel facing you  The location tab on the frame should  align with the locator tab on the thermal cycler  The probe wire is threaded through  the channel to prevent damaging the wire when you close the thermal cycler heated  cover                             GR1916    Figure 3 Proper assembly of the Temperature Verification System for 2720  thermal cycler       Figure 4 Temperature verification frame    8 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 2 Temperature calibration verification test    Temperature calibration verification test    Required  equipment    Setting up the  0 2 mL probe  assembly    Configuring the  2720 the
59. ck wells test failed   exceeds performance    specifications  per TNU Performance     NU at 94  C is xx xx   Fail        NU at 37  C is xx xx   Fail          F1 F2 F3 F4 F5    e  fthe test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors  were not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See  How to obtain support  on page viii              16 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide       Ending the test 1     Todas GRO    Chapter 2 Temperature non uniformity test    Press F5  Cancel    Remove the 0 2 mL probe assembly from the sample block   Power off the digital thermometer  then clean off the oil     Wait for the sample block to reach room temperature  725   C   then remove the  2720 temperature verification frame from the sample block       wr  PHYSICAL INJURY HAZARD  Hot Surface  Use care    when working around the heated cover and sample block to avoid being burned  by hot components       Clean the oil off the sample block     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  17    and Veriti  System User Guide    Chapter 2 Data Sheet  Calibration Verification Test    Data Sheet  Calibration Verification Test    18    When you run the calibration verification test  record the setpoint values for well A6  on this data sheet  At the end of the
60. d right dual 96 well block temperature verification plates    0 2 mL probe assembly     Model 4500 digital thermometer with a 9V battery installed    1  For the GeneAmp   9700 with Manual Dual 96 Well Sample Block Module   If the heated cover is in the forward position  lift the lever  then slide the heated  cover back     For the GeneAmp  9700 with Auto Lid Dual 96 Well Sample Block Module   If the heated cover is in the forward position  lift the lever  then press the red  button to slide the heated cover back     2  Place both dual 96 well temperature verification plates on the sample block   The plate labeled with L1 to L6 is for the left sample block  The plate labeled  with R1 to R6 is for the right sample block  Position the plates so the angled  corners face away from you     Note  The plates are used as guides to help you locate the probes within the  sample blocks     3  Use a cotton swab to coat wells L1 to L6 and R1 to R6 with mineral oil     4  Place the 0 2 mL probe assembly into well L1  As the test progresses  move the  0 2 mL probe assembly to each of the test wells  L1 to L6 and R1 to R6      5  Thread the probe wire through the channel in the dual 96 well temperature  verification plate to prevent damage to the probe and lead wires  see Figure 7 on  page 36      6  Make sure the probe is connected to the digital thermometer     42 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guid
61. damaged     8  Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  107  and Veriti   System User Guide    Chapter 9 Calibration verification test    Running the test    1  Power on the GeneAmp  9600  The Main menu opens     For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test       Press the OPTION key three times to move the pointer to UTIL  then press    ENTER  The Utilities menu opens     Select Function  DIR CONFIG DIAG DEL         Press the OPTION key twice to move the pointer to DIAG  then press ENTER     The following menu opens        Enter Diag Test  1          REVIEW History file      Press 5  then press ENTER to run the Verify Calibration Diagnostic Test     Test  5      Note  To ensure maximum accuracy  the temperatures of the heated cover and  the sample block are the same in this test  This condition prevents the heated  cover from affecting the accuracy of the probe assembly     The temperature of the sample block and heated cover increases to 40  C  and  the following display opens     Going to 409C      Cvr   xxC Blk   xx xC  This display shows the current temperature of the block cover   Cvr   xxC  and sample block  Blk   xx xC      When the temperature of the block cover is within  
62. e    Chapter 4 Temperature non uniformity test    7  For the GeneAmp   9700 with Manual Dual 96 Well Sample Block Module   Slide the heated cover forward and pull the lever down     For the GeneAmp   9700 with Auto Lid Dual 96 Well Sample Block Module   Press the red button to slide and close the heated cover     IMPORTANT  Seat the probe properly and the close the heated cover carefully   If the probe wire is crushed when you close the heated cover  the probe can be  damaged     8  Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     Configuring the 1  Power on the thermal cycler  The Main menu opens         GeneAmp      9700 For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test     2  Press F4  Util   The Utilities screen opens   3  Press F1  Diag   The Diagnostics screen opens     4  Press F3  TmpVer   The Temperature Verification screen opens     Temperature Verification    Temp   Calibration Verification   NU   Temperature Non Uniformity     Temp      TNU       F1 F2 F3 F4 F5                5  Press F2  TNU   The GeneAmp   9700 is automatically configured for the  temperature non uniformity test  starting with the setpoint of 95   C  and the  TNU Performance screen opens  The following screen opens when the cover  reaches the setpoint     TNU Performance  Sample temp   xx x C Cover temp   xxx 
63. e GeneAmp  9800    ccce eet e eee ees 98   pGgnun 40 5 EP  otk e Oh nares PA a ee eee tas 99  Evaluating the Tesulf amp  cessas escas dee bcd aden d ma e ee ne ERA 102    This test requires the following items in your Temperature Verification Kit       Digital thermometer Model 4500 with a 9V battery installed    Cotton swabs     Light mineral oil     0 I mL temperature verification probe assembly    1  Ifthe heated cover is in the forward position  lift the lever  then slide the heated    0 1 mL probe cover back   assembly 2  With the instrument powered off and the sample block at room temperature    place the 0 1 mL temperature verification probe assembly on the sample block    3  Use a cotton swab to coat wells Al  A12  C4  C9  F4  F9  H1  and H12 with  mineral oil    4  Place the 0 1 mL probe assembly into well A1   As the test progresses  move the 0 1 mL probe assembly to each of the test wells   Al  A12  C4  C9  F4  F9  H1  and H12     5  Thread the probe wire through the undercut of the 0 1 mL probe assembly to  prevent damage to the probe and lead wires    6  Make sure the probe is connected to the digital thermometer    7  Shde the heated cover forward and pull the lever down   IMPORTANT  Seat the probe properly and close the heated cover carefully  If  the probe wire is crushed when you close the heated cover  the probe may be  damaged    8  Power on the digital thermometer   Refer to the instructions in your Temperature Verification Kit for details on  operating th
64. e digital thermometer  Model 4500    9  Continue with    Configuring the GeneAmp   9800  on page 98    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  97    and Veriti  System User Guide    Chapter 8 Temperature non uniformity test    Configuring the Complete the procedures in    Setting up the 0 1 mL probe assembly    on  GeneAmp   9800 page 97   Power on the GeneAmp   9800   For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test   In the main menu  press F4  Util  to open the Utilities screen     Press F1  Diag  to open the Diagnostics screen     Press F3  TmpVer  to open the Temperature Verification screen   Temperature Verification  Temp   Calibration Verification   NU   Temperature Non Uniformity   Temp   TU   F1 F2 F3 F4 F5  6  Press F2  TNU  to open the TNU Performance screen and configure the 9800  for the temperature non uniformity test   TNU Performance  Sample temp   xx x C Cover temp   xxx C  Place Probe in well A1  Press Run  Run  F1 F2 F3 F4 F5  Note  Ifthe cover temperature is not within  1   C of the heated cover setpoint   the following screen opens   TNU Performance  Sample temp   xx x C Cover temp   xxx C  Cover setpoint is 55C  Cover must be within 1 0C of setpoint  F1 F2 F3 F4 F5  7  Continue with    Running the test    on page 99   98 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800            
65. e screen  opens with the following prompt     TNU Performance  Sample temp   xx x  C Cover temp   xxx  C    Place probe in well xx    Press Run    Run       Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  15  and Veriti   System User Guide    Chapter 2 Temperature non uniformity test    7  Open the hinged heated cover  Repeat steps 4 through 7 of the    Setting up the  0 2 mL probe assembly    on page 13 and steps 1 through 6 of this procedure   Complete these steps for all 8 wells to be tested  A1  A12  C4  C9  F4  F9  H1   HI2     8  The 2720 thermal cycler evaluates the uniformity of the sample block  temperature for the setpoint values that you entered and displays the results  A  summary screen opens at the end of the test              F1 F2 F3 FA F5    If you entered values on the Temperature Non Uniformity Test Data Sheet   page 19   compare those values with the actual test results     9  Press F1  Accept      Evaluating the When the thermal cycler completes the temperature non uniformity test  the TNU  results Performance screen opens  See the table below to evaluate the results           If the    Then     Temperature of the  Pass  appears after each setpoint temperature   sample block wells is    uniform    TNU Performance     NU at 94  C is xx xx   Pass   NU at 37  C is xx xx   Pass    F1 F2 F3 FA F5             Temperature variation of    Fail  appears after the setpoint temperature s  for which the  the sample blo
66. e the heated cover forward and pull the lever down     For the GeneAmp   9700 with Auto Lid Dual 96 Well Sample Block Module   Press the red button to slide and close the heated cover     8  Repeat steps 4 through 7 of    Setting up the 0 2 mL probe assembly    on page 42  and steps 1 through 7 of this procedure  Complete these steps for all the  remaining 11 wells to be tested  L2 to L6 and R1 to R6      The GeneAmp   9700 evaluates the uniformity of the sample block temperature  for the setpoint values that you entered and displays the results  A summary  screen opens at the end of the test                       F1 F2 F3 F4 F5    9  Ifyou entered values on the Temperature Non Uniformity Test Data Sheet   page 48   compare those values with the actual test results     10  Press F1  Accept      Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  45  and Veriti   System User Guide    Chapter 4 Temperature non uniformity test    Evaluating the After the GeneAmp   9700 completes the temperature non uniformity test  the TNU  results Performance screen opens  See the table below to evaluate the results        If the       Then          Temperature of the  sample block wells is  uniform          Pass    appears after each setpoint temperature     TNU Performance    TNU at 95  C is xx xx   Pass    TNU at 55  C is xx xx   Pass    F1 F2 F3 F4 F5       Temperature variation of  the sample block wells  exceeds performance  specifications     
67. ecalibrated once a year  For instructions on having your system recalibrated by the  manufacturer  refer to Appendix A     Recalibration        4 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Probe assembly    Chapter 1 Probe assembly    The probe assembly  shown below  consists of a cone that measures the temperature  of the sample well  The temperature is measured by a thermistor bead in the cone tip     GR1914    Digital thermometer    Overview The handheld digital thermometer  shown below  has a temperature range of 0 to       100   C and is accurate to within 4                                           GR1915           0 1   C     IMPORTANT  Do not try to recalibrate or perform any service on the digital  thermometer  The only user serviceable component in the unit 1s the battery     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  5    and Veriti  System User Guide    Chapter 1 Probe assembly    Precautions    Temperature  display  differences    Using the probe  while running  programs    When you use the digital thermometer       Make sure the probe assembly is connected to the input connector at the top of  the digital thermometer  If a probe is not connected when the thermometer is  powered on  the unit displays the error message    Error No Probe Found  Install  Probe after Shut Down         To power on the digital thermometer  pres
68. ed values on the Calibration Verification Test Data Sheet  compare  those values with the actual test results     8  Press F1  Accept   To interpret the results  see    Evaluating the results  on  page 82   9  When you complete all measurements  be sure to     Press Exit     Remove the 0 5 mL probe assembly from the sample block     Power off the digital thermometer  then clean off the oil     Remove the 0 5 mL temperature verification frame from the sample block       NITE BURN HAZARD  Make sure the sample blocks have cooled    at least to room temperature  less than 25   C  before removing the probe  The  blocks and probes retain heat longer than expected     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  81  and Veriti   System User Guide    Chapter 7 Calibration verification test    82    Evaluating the When the GeneAmp   9700 completes the calibration verification test  one of two  screens opens  See the table below to evaluate the results     results       If the sample block  module       Then the          Is properly calibrated    Calibration Verification screen displays the following message     Calibration Verification    Calibration is good       F1 F2 F3 F4 ES       Does not pass the  calibration verification  test             Calibration Verification screen displays the following message     Calibration Verification       Instrument may require servic  Contact Applied Biosystems  Technical Support           F1 F2 
69. emp 2 at  85  C 45  C  Thermometer    Date Tested by   Probe serial no  serial no        Left Right Left Right  Zone Zone Zone Zone                                                                                                    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  127  and Veriti   System User Guide    Chapter 10 Data Sheet  Temperature Non Uniformity Test    Data Sheet  Temperature Non Uniformity Test    128    When you run the temperature non uniformity test  record the setpoint values for the  wells listed on this data sheet  At the end of the test  compare the values that are  displayed on the thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date       Tested by       Probe serial no        Thermometer serial no           Measurement    Block TNU 1 at 95  C    Block TNU 2 at 60  C                                  Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Hecalibration    This appendix describes how to order a recalibration service for your Temperature  Verification System     Obtaining recalibration    Introduction    Decontaminating  the probe  assembly    Sending  equipment to  Alpha Technics    Applied Biosystems recommends that the Model 4500 digital thermometer and  Probe Assembly be recalibrated on an annual basis     To have your thermometer and probe recalibrated 
70. ending on the results  See the table below to evaluate the results        If the sample block  module       Then the          Is properly calibrated    Calibration Verification screen displays the following message     Calibration Verification    Calibration is good          F1 F2 F3 FA F5       Does not pass the  calibration verification  test             Calibration Verification screen displays the following message     Calibration Verification    Instrument may require service   Contact Applied Biosystems  Technical Support        F1 F2 F3 FA F5    e If the test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors were  not made     e If the test fails again  contact Applied Biosystems Technical  Support  See  How to obtain support  on page viii        Note  As a result of the Temperature Tolerance Stack  between the sample block   temperature probe and thermometer  the instrument firmware is programmed to  accept a Pass Fail result of  0 75   C  Therefore  passing results may vary from  84 25   C to 85 75   C at the 85 degree setpoint and 44 25   C and 45 75   C at the 45   degree setpoint  Ranges greater than   0 75 degrees cause the test to fail     Ending the test 1  Press F5  Exit         2  Remove the 0 2 mL probe assembly from the sample block     3  Power off the digital thermometer  then clean off the oil     Wait for the sample block to reach room temperature  725   C   then remove the  2720 temperature verifica
71. eneAmp   9700 for the    temperature non uniformity test  The TNU Performance screen opens     TNU Performance  Sample temp   xx x C Cover temp   xxx C       Place Probe in well A1  Press Run    Run    F1 F2 F3 FA F5       7  Continue with    Running the test  on page 57     56 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Running the test    Chapter 5 Temperature non uniformity test    The temperature non uniformity test uses the 0 02 mL probe assembly to test the  temperature uniformity of six different wells in each sample block     Note  If necessary  press F5  Cancel  to exit the test     To run the temperature non uniformity test     l     Complete the procedures in    Setting up the 0 02 mL probe assembly    on  page 55     Press F1  Run  to start the temperature non uniformity test  The TNU  Performance screen opens with the setpoint value displayed     TNU Performance  Sample temp   xx x C Cover temp   xxx  C    Setpoint is 94  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 F4 E5          Note  The sample blocks must be within  1 0   C of the setpoint  In addition   the cover must be within  1   C of 105   C  It may take several minutes for the  GeneAmp   9700 to ramp up       The TNU Performance screen counts down the time until the setpoint is stable     TNU Performance    Sample temp   xx x C Cover temp   xxx  C  Stabilizing at setpoint    x xx    F1 F2 F3 FA F5       W
72. ens     TNU Performance  Sample temp   xx x C Cover temp   xxx C    Place Probe in well L1  Press Run    Run    F1 F2 F3 F4 F5          Running the test The temperature non uniformity test uses the 20 uL probe assembly to test the  temperature uniformity of six different wells in each sample block     Note  If necessary  press F5  Cancel  to exit the test     70 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 6 Temperature non uniformity test    To run the temperature non uniformity test     1  Press F1  Run  to start the temperature non uniformity test  The TNU  Performance screen opens with the setpoint value displayed     TNU Performance  Sample temp   xx x C Cover temp   xxx  C    Setpoint is 94  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 FA F5          Note  The sample blocks must be within  1 0   C of the setpoint  In addition   the cover must be within  1   C of 105   C  It may take several minutes for the  GeneAmp   9700 to ramp up     2  Take a temperature reading     a  The TNU Performance screen counts down the time until the setpoint is  stable     TNU Performance    Sample temp   xx x C Cover temp   xxx  C    Stabilizing at setpoint    x xx       F1 F2 F3 FA F5    b  When the  Stabilizing at setpoint  value decreases to 0  read the digital  thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital therm
73. erification Test    e    In the System Specification Sheet  the thermal cycler accuracy specifications  are published as  0 25   C    The single channel Temperature Verification Kit  TVK  consists of the  hardware  thermometer and probe and an associated procedure        The TVK does not measure the original manufacturing specification of the  thermal cycler  The purpose of the TVK test is to detect any system failure that  may result in a shift in the original manufacturer s calibration  This kit should be  used for temperature verification  not system calibration        For the GeneAmp   9700  the system has an aggregate accuracy of 40 25   C     The TVK has accuracy and discrepancy in addition to the original  manufacturing specification     Temperature calibration of Applied Biosystems thermal cyclers is extremely  stable  and no measurable change in the calibration accuracy is detectable unless  a significant change occurs in the instrument characteristics after its  manufacture     Temperature Tolerance stack up between TVK and the original manufacturing  specification is  0 75 degrees Celsius        Note  Asa result of the Temperature Tolerance Stack  the instrument firmware has  been pre programmed to accept a Pass Fail result of  0 75 degrees Celsius  therefore  passing results may vary from 84 25   C up to 85 75   C        Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  135    and Veriti   System User Guide    Appendi
74. ermometer   7  Slide the heated cover forward and pull the lever down     IMPORTANT  Seat the probe properly  and close the heated cover carefully  If  the probe wire is crushed when you close the heated cover  the probe may be  damaged     8  Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     9  Continue with    Configuring the GeneAmp   9700  on page 79     78 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 7 Calibration verification test                               Configuring the 1  Complete the procedures in    Setting up the 0 5 mL probe assembly  on  GeneAmp  9700 page 78   2  Power on the thermal cycler  The Main menu opens   For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test   3  Press F4  Util   The Utilities screen opens   4  Press F1  Diag   The Diagnostics screen opens   5  Press F3  TmpVer   The Temperature Verification screen opens   Temperature Verification  Temp   Calibration Verification   NU   Temperature Non Uniformity   Temp     TNU    F1 F2 F3 F4 F5  6  Press F1  Temp  to automatically configure the GeneAmp   9700 for the  calibration verification test  The Calibration Verification screen opens   Calibration Verification  Block temp   xx x C Cover temp   xxx C  Place probe in we
75. est Data Sheet  page 32  to  keep a permanent record of the test   4  Press Enter     The thermal cycler automatically begins the second temperature reading  45   C  setpoint   The Calibration Verification screen displays the setpoint value     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 45  C  Cover must be within 30  C of 45  C    F1 F2 F3 F4 F5       Note  The cover must be at 45   C   30   C  It may take several minutes for the  system to ramp up     5  Repeat steps 2 through 4 for the second temperature reading     The thermal cycler evaluates the calibration of the sample block temperature for  the setpoint values that you entered  then displays the results  A summary  section opens when the test ends     Calibration Verification    Actual temperature at 85  C       Actual temperature at 45  C    F1 F2 F3 F4 F5       6  Ifyou entered values on the Calibration Verification Test Data Sheet  page 32    compare those values with the actual test results     7  Press F1  Accept      Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  25  and Veriti   System User Guide    Chapter 3 Temperature calibration verification test    26    Evaluating the When the thermal cycler completes the calibration verification test  one of two  results screens opens  depending on the test results  See the table below to evaluate the    results        If the sample block  module       Then the          Is properly
76. ested by   Probe serial no     Thermometer  serial no     Setpoint value  Well R3       85   C 45  C                                                                                              Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  47    and Veriti   System User Guide    Chapter 4 Data Sheet  Temperature Non Uniformity Test    Data Sheet  Temperature Non Uniformity Test    When you run the temperature non uniformity test  record the setpoint values for the  wells L1 to L6 and R1 to R6 listed on this data sheet  At the end of the test  compare  the values that are displayed on the thermal cycler with the values that are recorded  here     Note  You can photocopy this page                                                                          Date   Tested by   Probe serial no    Thermometer   serial No    Setpoint value 95   C 55   C  L1  L2  L3  L4  L5  L6  R1  R2  R3  R4  R5  R6  48 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Temperature Verification Tests for the GeneAmp    9700  Dual 384 Well Sample Block Module D    This chapter covers    Calibration ventication  e8L  iius ouo e iem hoe Roe Ro e eh ds  Temperature nop uniformity Test su sea r kr eor de RARO p ac eadera  Data Sheet  Calibration Verification Test iisususssino dee eam age der pg eas  Data Sheet  Temperature Non Uniformity Test           llle essen     Thermal C
77. extreme situations        Except for IMPORTANTS  each safety alert word in an Applied Biosystems  document appears with an open triangle figure that contains a hazard symbol  These  hazard symbols are identical to the hazard symbols that are affixed to Applied  Biosystems instruments  see    Safety symbols  on page 139      MSDSs The MSDSs for any chemicals supplied by Applied Biosystems or Ambion are  available to you free 24 hours a day  For instructions on obtaining MSDSs  see     MSDSs    on page vii        IMPORTANT  For the MSDSs of chemicals not distributed by Applied Biosystems  or Ambion contact the chemical manufacturer        Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  vii  and Veriti  System User Guide    Preface    How to use this guide    Text conventions    User attention  words    This guide uses the following conventions     Bold text indicates user action  For example   Type 0  then press Enter for each of the remaining fields       Italic text indicates new or important words and is also used for emphasis   For example     Before analyzing  always prepare fresh matrix       A right arrow symbol    gt    separates successive commands you select from a  drop down or shortcut menu  For example     Select File  gt  Open  gt  Spot Set     Two user attention words appear in Applied Biosystems user documentation  Each  word implies a particular level of observation or action as described below     Note      Prov
78. fication  Block temp   xx x C Cover temp   xxx C       Setpoint is 85  C  Cover must be within 1  C of 105  C    F1 F2 F3 FA F5       Note  The cover must be within  1   C of 105   C  It may take several minutes  for the GeneAmp  9700 to ramp up     3  The Calibration Verification screen counts down the time until the setpoint is  reached     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Stabilizing at setpoint    x xx    F1 F2 F3 F4 F5       When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer  Refer to the instructions in your Temperature Verification Kit for  details on operating the digital thermometer  Model 4500     4  Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature    field     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Enter actual block temperature    Cancel       F1 F2 F3 F4 F5    Note  Round off the four digit value on the digital thermometer to three digits  before you enter it into the Calibration Verification screen     Note  Record this value on the Calibration Verification Test Data Sheet   page 60  to keep a permanent record of the test     52 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 5 Calibration verification test    5  Press ENTER  The GeneAmp   9700 automatically begins the second reading   4
79. fication System For GeneAmp   PCR Systems 2720  9600  9700  9800  125  and Veriti  System User Guide    Chapter 10 Temperature non uniformity test    Ending the test     1  Ifthe heated cover is closed  lift the lever and place the heated cover in the fully  opened position       wr  PHYSICAL INJURY HAZARD  Hot Surface  Use care    when working around the heated cover and sample block to avoid being burned  by hot components     2  Wait for the sample block to reach room temperature  725   C   then remove the  the 0 5 mL probe assembly and thermal isolation frame from the sample block     3  If you are performing a repeat measurement  clean the probe assembly and all  wells of mineral oil  then perform the repeat measurement  Otherwise  power off  the Veriti  Thermal Cycler and Model 4500 digital thermometer     4  Clean the oil off the sample block     Note  Follow the decontamination procedures in user guide for cleaning oiled    wells   126 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 10 Data Sheet  Calibration Verification Test    Data Sheet  Calibration Verification Test    When you run the calibration verification test  record the setpoint values for well E1  on this data sheet  At the end of the test  compare the values that are displayed on the  thermal cycler with the values that are recorded here     Note  You can photocopy this page        Avg Probe Temp 1 at Avg Probe T
80. fication plates on the sample blocks     IMPORTANT  The temperature verification plates must be installed in both  sample blocks throughout all temperature verification testing     The plate labeled L1 to L6 is for the  eft sample block  Figure 14   The plate  labeled R1 to R6 is for the right sample block  Figure 15   Position the plates so  the angled corners face away from you        Figure 1 4 Left temperature Figure 15 Right temperature  verification plate verification plate    Note  The plates help you locate the probes within the sample blocks     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter6 Calibration verification test    3  Use the stick end of a cotton swab to coat well R3 with mineral oil  The wells  are too small for the swab end     4  Place the 20 uL probe assembly into well R3     5  Thread the probe wire through the channels in the 384 well temperature  verification plate to prevent damage to the probe and lead wires     6  Make sure the probe is connected to the digital thermometer   7  Press and release the red button to automatically close the heated cover     IMPORTANT  Seat the probe properly  If the probe wire is crushed when the  heated cover closes  the probe may be damaged    8  Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     Configuring t
81. g up the 20 uL probe assembly    on page 69 and  steps 1 through 6 of this procedure  Complete these steps for all 12 wells to be  tested        Left Sample Block Right Sample Block       L1  L2  L3  L4  L5  and L6 R1  R2  R3  R4  R5  and R6                8  The GeneAmp   9700 evaluates the uniformity of the sample block temperature  for the setpoint values that you entered and displays the results  A summary  screen opens at the end of the test                    F1 F2 F3 F4 F5    If you entered values on the Temperature Non Uniformity Test Data Sheet   compare those values with the actual test results     9  Press F1  Accept   To interpret the results  see    Evaluating the results    on  page 73     72 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 6 Temperature non uniformity test    10  When you complete all measurements  be sure to     Press F5  Cancel      Remove the 20 uL probe assembly from the sample blocks     Power off the digital thermometer  then clean off the oil     Remove the 384 well temperature verification plates from the sample    blocks     IMPORTANT  Make sure the sample blocks are at room temperature   25   C   before removing the reaction plates     Evaluating the When the GeneAmp   9700 completes the temperature non uniformity test  the TNU  results Performance screen opens  See the table below to evaluate the results        If the       Then          Temper
82. he     1  Power on the thermal cycler  The Main menu opens                                      GeneAmp   9700 For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test   2  Press F4  Util   The Utilities screen opens   3  Press F1  Diag   The Diagnostics screen opens   4  Press F3  TmpVer   The Temperature Verification screen opens   Temperature Verification  Temp   Calibration Verification   NU   Temperature Non Uniformity    Temp J   TNU J  F1 F2 F3 F4 F5  5  Press F1  Temp  to automatically configure the GeneAmp   9700 for the  calibration verification test  The Calibration Verification screen opens   Calibration Verification  Block temp   xx x C Cover temp   xxx C  Place probe in well R3  Press Run  Run  Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  65    and Veriti   System User Guide    Chapter 6 Calibration verification test    Running the test Use the thermometer to take temperature readings of the well that is connected to the  20 uL probe assembly  You must take readings at two different setpoint temperatures     Note  If necessary  press F5  Cancel  to exit the test     To run the calibration verification test     1  Press F1  Run  to start the calibration verification test  The Calibration  Verification screen opens with a setpoint value displayed     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 85  C  Cover must
83. he instrument     4  Place the 0 2 mL probe assembly into well A1     5  Thread the probe wire through the channel in the 9600 temperature verification  frame to prevent damage to the probe and lead wires  Make sure the probe is  connected to the digital thermometer  see Figure 18 on page 106      6  Slide the heated cover forward  then turn the cover knob clockwise until the  white mark on the knob is aligned with the white mark on the cover     IMPORTANT  Seat the probe properly and the close the heated cover carefully   If the probe wire is crushed when you close the heated cover  the probe can be  damaged     7  Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     Running the test 1  Power on the GeneAmp   9600   For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test   Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  111    and Veriti  System User Guide    Chapter 9 Temperature non uniformity test    2  Setup a two temperature CYCL program with the following parameters     Setpoint  1 Temperature   95   C    Hold Time   2 00 minutes    Ramp Time   0 00 minutes      Setpoint  2 Temperature   40   C    Hold Time   2 00 minutes     Ramp Time   0 00 minutes     Cycles   99    Note  Refer to the GeneAmp   PCR System 9600 User s Manual  PN 0093   8660  f
84. he setpoint value displayed     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 45  C  Cover must be within 30  C of 45  C    EL F2 F3 FA F5       Note  The cover must be within  30   C of 45   C     6  Repeat steps 3 and 4 for the second reading     The GeneAmp   9800 evaluates the calibration of the sample block temperature  for the setpoint values that you entered and displays the results  A summary  screen opens at the end of the test     Calibration Verification    Actual temperature at 85  C       Actual temperature at 45  C    F1 F2 F3 FA F5       If you entered values on the Calibration Verification Test Data Sheet  compare  those values with the actual test results     7  Press  F1  Accept  To interpret the results  see    Evaluating the results    on  page 96   8  When you complete all measurements  be sure to     Press Exit     Remove the 0 1 mL probe assembly from the sample blocks     Power off the digital thermometer  then clean off the oil       Remove the 0 1 mL temperature verification probe assembly from the  sample blocks       Nee BURN HAZARD  Make sure the sample blocks have cooled    at least to room temperature  less than 25   C  before removing the probe  The  blocks and probes retain heat longer than expected     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  95  and Veriti   System User Guide    Chapter8 Calibration verification test    96    Evaluating the When the 
85. hen the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer  Refer to the instructions in your Temperature Verification Kit for  details on operating the digital thermometer  Model 4500     Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature  field     TNU Performance    Sample temp   xx x C Cover temp   xxx C       Enter actual block temperature  00 0     F1 F2 F3 FA F5       Note  Round off the four digit value on the digital thermometer to three digits  before you enter it into the Calibration Verification screen     Note  Record this value on the Temperature Non Uniformity Test Data Sheet   page 61  to keep a permanent record of the test     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  57  and Veriti  System User Guide    Chapter 5 Temperature non uniformity test    5  Press ENTER  The GeneAmp   9700 automatically begins the second reading   37   C setpoint   The TNU Performance screen opens with the setpoint value  displayed     TNU Performance    Sample temp   xx x  C Cover temp   xxx  C    Setpoint is 37  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 F4 F5          Note  The sample blocks must be within  1 0   C of the setpoint   6  Repeat steps 3 and 4 for the second reading   7  Press ENTER  The TNU Performance screen opens with the following prompt     TNU Performance  Sample temp   xx x  C Cover temp 
86. ication System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti   System User Guide    Chapter 9 Temperature non uniformity test    Temperature non uniformity test    Required  equipment    Setting up the  0 2 mL probe  assembly    Use this test to verify the temperature uniformity of the GeneAmp  9600  The  temperature non uniformity test consists of the following procedures  which you  must perform in sequence     Setting up the 0 2 mL probe assembly    2    66 6     6 bccn eee e e e EH  Runnte thedest  2i oco ees ae S er d Rue e e ees 111  Calculsinp test results  ccc ove CUL Reed cher ORA G DER C  X a ee E 113  Ending he dose  oo osse Le eec e exces do edd ur abdo ga I dos 113      Nr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the 0 2 mL Temperature Verification Kit  PN 4317939   which  includes      Cotton swabs     Light mineral oil     9600 temperature verification frame     0 2 mL probe assembly     Model 4500 digital thermometer with a 9V battery installed    1  Ifthe heated cover is in the forward position  turn the knob completely  counterclockwise  then slide the heated cover back     2  Using a cotton swab  coat the following wells with mineral oil   Al  A4  A8  A12  C1  C4  C8  C12  El  E4  E8  E12  H1  H4  H8  H12    3  Place the 9600 temperature verification frame on the sample block with the  channel facing the front of t
87. ides information that may be of interest or help but is not critical to the  use of the product        IMPORTANT      Provides information that is necessary for proper instrument  operation  accurate chemistry kit use  or safe use of a chemical        How to obtain support    For the latest services and support information for all locations  go to   www appliedbiosystems com    At the Applied Biosystems web site  you can     Access worldwide telephone and fax numbers to contact Applied Biosystems  Technical Support and Sales facilities     Search through frequently asked questions  FAQs      Submit a question directly to Technical Support       Order Applied Biosystems user documents  MSDSs  certificates of analysis   and other related documents       Download PDF documents     Obtain information about customer training     Download software updates and patches   For recalibration  please contact Alpha Technics  Formerly Alpha Sensors  Inc    Eutechnics Division  at   www alphatechnics com    Calibration standards are traceable to NIST  U  S  National Institute of Standards and  Technology      viii Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Introduction       This chapter covers    Aboni Mis manal  nn ho Rnd rp  About the Temperature Verification System  2 044 40466 5404ee0 eneoeerseevaws  Probe Aer a hee s deca du thee dd eee Sie eee a     Da BOE NET  ao ior SR DU ARE Se es SOL eee ERO
88. ification Tests for the  GeneAmp   9600    This chapter covers     OVVIE o Les de deo re Dee aee bad aR RE  Calibration verification tesb iiussese asse eon enden  Temperature non uniformity test            ll lessen   Data Sheet  Calibration Verification Test                      Data Sheet  Temperature Non Uniformity Test    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    105    Chapter 9 Overview    Overview    Materials When you perform temperature verification tests on the GeneAmp   9600  you use  the materials provided in your kit     Assembly Figure 18 shows a properly assembled system  The frame is positioned around the  wells with the channel facing you  and the probe wire is threaded through the channel  to prevent damaging the wire when the thermal cycler heated cover is closed                             GR1916    Figure 18 The Temperature Verification System for GeneAmp  9600    106 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter9 Calibration verification test    Calibration verification test    Required  equipment    Setting up the  0 2 mL probe  assembly    Use this test to verify the temperature calibration of your GeneAmp   9600  The  calibration verification test consists of the following procedures  which you must  perform in sequence     Setting up the 0 2 mL probe assembly   2    64 
89. ion frame from the sample block     3  If you are performing a repeat measurement  clean the probe assembly and all  wells of mineral oil  then perform the repeat measurement  Otherwise  power off  the Veriti  Thermal Cycler and Model 4500 digital thermometer     4  Clean the oil off the sample block     Note  Follow the decontamination procedures in user guide for cleaning oiled    wells   122 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 10 Temperature non uniformity test    Temperature non uniformity test    Required  equipment    Setting up the  0 5 mL probe  assembly    Use this test to verify the temperature uniformity of the Veriti  Thermal Cycler  The  temperature non uniformity test consists of the following procedures  which you  must perform in sequence     Setting up the 0 5 mL probe assembly    coc 64 6 0   6 bccn eee e e ktina 123  Runing  he 4e8L   2   eco paw nee fae S esr d duce e ee oe ee 125  Ending The testo cid sov cade ee uu eXenER Ib S CH ODDS E Weis 126      wr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the 0 5 mL Temperature Verification Kit  PN 4309924   which  includes      Cotton swabs     Light mineral oil     Temperature verification frame     0 5 mL probe assembly     Model 4500 digital thermometer with a 9V battery installed    
90. key  This returns you  to the    Review History File  display  Press 5  then press ENTER to return to the  Verify Calibration Diagnostic test     The figure below shows temperatures during the calibration verification test     95    40    Temp  oc Time  in minutes        Calculating test Use the following information to calculate the results of the test  Refer to the  results calibration label in your user manual for the High and Low Offset values     Note  If there is more than one GeneAmp   9600 in your laboratory  verify that the  serial number on the calibration label matches the serial number on the instrument  that you are testing     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  109  and Veriti  System User Guide    Chapter 9 Calibration verification test    110    Ending the test    Average block temperature at the 95   C hold    Use the following formula to calculate the average block temperature at the 95  C  hold     Block Average at 95   C   T 95      High Offset    If the block average is more than 0 75 degrees Celsius above or below 95   C  your  GeneAmp   9600 must be recalibrated     For example     If the measured temperature of well E1 is 95 2   C  and the High Offset that is printed  on your calibration label is 0 1  you would make the following calculation     Block Average at 95   C   95 2       0 1    95 3   C          In this example  because 95 3   C does not differ by  0 75 degrees Celsius from the  p
91. le block    3  Use a cotton swab to coat the following wells with mineral oil   Al  A12  C4  C9  F4  F9  H1  H12   4  Place the 0 2 mL probe assembly into well C4  then as the test progresses  move  the 0 2 mL probe assembly to each of the test wells    5  Thread the probe wire through the channel in the 9700 temperature verification  frame to prevent damage to the probe and lead wires  see Figure 5 on page 22     6  Make sure the probe is connected to the digital thermometer    7  Slide the heated cover forward and pull the lever down   IMPORTANT  Seat the probe properly and close the heated cover carefully  If  the probe wire is crushed when you close the heated cover  the probe can be  damaged    8  Power on the digital thermometer   Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  27    and Veriti   System User Guide    Chapter 3 Temperature non uniformity test    Configuring the 1  Power on the thermal cycler  The Main menu opens  For best results  allow the  GeneAmp  9700 thermal cycler 15 to 30 minutes to warm up to ambient temperature before  performing the test     2  Press F4  Util   The Utilities screen opens   3  Press F1  Diag   The Diagnostics screen opens     4  Press F3  TmpVer   The Temperature Verification screen opens     Temperature Verification       F1 F2 F3 F4 F5    5  Press F2  T
92. le block to reach room temperature  725   C   then remove the    dual 96 well temperature verification plates from the sample block       NIIS PHYSICAL INJURY HAZARD  Hot Surface  Use care    when working around the heated cover and sample block to avoid being burned    by hot components     5  Clean the oil off the sample block     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  41    and Veriti  System User Guide       Chapter 4 Temperature non uniformity test    Temperature non uniformity test    Required  equipment    Setting up the  0 2 mL probe  assembly    Use this test to verify the temperature uniformity of the GeneAmp   9700 with Dual  96 Well Sample Block Module  The temperature non uniformity test consists of the  following procedures  which you must perform in sequence     Setting up the 0 2 mL probe assembly    260640 c ss ee cee RR ene Rn 42  Configuring the GeneAmp  9700    0 0    cece ccc cette eens 43  Rianing  he 1886    och es eek d dhe eee ewe bbe a FES eo Gee ee E E 44  Esaluateng tbe desplis  1202s eer Ret eSI o ed RA o S ER TEE 46  Podne Me dest ocolos mere bra b eee ER Ooo Se Re RM oce e 46      rr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the 0 2 mL Sample Block Module Temperature Verification Kit   PN 4317939   which includes       Cotton swabs     Light mineral oil     Left an
93. led R1 to R6 is for the right sample block  Figure 11   Position the plates so  the angled corners face away from you        Figure 10 Left temperature Figure 11 Right temperature  verification plate verification plate    Note  The plates help you locate the probes within the sample blocks     3  Use the stick end of a cotton swab to coat well R3 with mineral oil  the wells are  too small for the swab end     4  Place the 0 02 mL probe assembly into well R3     50 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 5 Calibration verification test    5  Thread the probe wire through the channels in the 384 well temperature  verification plate to prevent damage to the probe and lead wires     6  Make sure the probe is connected to the digital thermometer   7  Shde the heated cover forward and pull the lever down     IMPORTANT  Seat the probe properly  and close the heated cover carefully  If  the probe wire is crushed when you close the heated cover  the probe may be  damaged     8  Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     9  Continue with    Configuring the GeneAmp  9700    below     Configuring the 1  Complete the procedures in    Setting up the 0 02 mL probe assembly    above   GeneAmp  9700    2  Power on the thermal cycler  The Main menu opens     For best result
94. libration verification test    54    Evaluating the When the GeneAmp   9700 completes the calibration verification test  one of two  screens opens  See the table below to evaluate the results     results       If the sample block  module       Then the          Is properly calibrated       Calibration Verification screen displays the following message     Calibration Verification    Calibration is good    F1 F2 F3 FA F5       Does not pass the  calibration verification  test             Calibration Verification screen displays the following message     Calibration Verification          Instrument may require servic  Contact Applied Biosystems  Technical Support        El F2 F3 F4 F5    e lf the test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors were  not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See    How to obtain support  on page viii           Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 5 Temperature non uniformity test    Temperature non uniformity test    Required  equipment    Setting up the  0 02 mL probe  assembly    Use this test to verify the temperature uniformity of the GeneAmp  PCR System  9700  Dual 384 Well Sample Block Module  The temperature non uniformity test  consists of the following procedures  which you must perform in sequence     Setting up the 0 02 mL probe
95. ll A6  Press Run  Run  F1 F2 F3 F4 F5  7  Continue with    Running the test    on page 80   Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  79    and Veriti   System User Guide    Chapter 7 Calibration verification test    Running the test Use the digital thermometer to take temperature readings of the sample well that is  connected to the 0 5 mL probe assembly  You must take readings at two different  setpoint temperatures     Note  If necessary  press F5  Cancel  to exit the test     To run the calibration verification test   1  Complete the procedures in    Configuring the GeneAmp   9700  on page 79     2  Press F1  Run  to start the calibration verification test  The Calibration  Verification screen opens with the setpoint value displayed     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 85  C  Cover must be within 10  C of 85  C    F1 F2 F3 F4 F5       Note  The cover must be within  1   C of 105   C  It may take several minutes  for the GeneAmp   9700 to ramp up     3  The Calibration Verification screen counts down the time until the setpoint is  reached  When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Stabilizing at setpoint    x xx       F1 F2 F3 F4 F5    Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer 
96. ly  is facing outwards     3  Use a cotton swab to coat well A6 with mineral oil   4  Place the 0 1 mL temperature verification probe assembly into well A6     5  Thread the probe wire through the undercut of the 0 1 mL temperature  verification probe assembly to prevent damage to the probe and lead wires     6  Make sure the probe is connected to the digital thermometer   7  Slide the heated cover forward and pull the lever down     IMPORTANT  Seat the probe properly  and close the heated cover carefully  If  the probe wire is crushed when you close the heated cover  the probe may be  damaged     8  Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     9  Continue with    Configuring the GeneAmp   9800    on page 93     92 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti   System User Guide    Chapter8 Calibration verification test    Configuring the 1  Complete the procedures in    Setting up the 0 1 mL probe assembly  on  GeneAmp  9800 page 92     2  Power on the GeneAmp   9800     For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test     3  When the main menu opens  press Util to open the Utilities screen   4  Press Diag to open the Diagnostics screen    Diagnostics   Hard   Hardware Diagnostics   System   System Performance Tests    
97. n the GeneAmp   9600 because the heated sample block  cover affects the probe temperature  If you suspect any temperature calibration  problems  perform the calibration verification test that is described on page 107     112 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 9 Temperature non uniformity test    Calculating test Use the following information to calculate the results of the test     results    Note  If there is more than one GeneAmp   9600 in your laboratory  verify that the    serial number on the calibration label matches the serial number on the instrument  that you are testing     For the 16 Setpoint  1 measurements  95   C hold   subtract the lowest measured  temperature from the highest measured temperature     For the 16 Setpoint  2 measurements  40   C hold   subtract the lowest measured  temperature from the highest measured temperature     If either result is more than   1   C from the setpoint  your GeneAmp   9600 must be  serviced by an Applied Biosystems service representative     Ending the test 1   2     Remove the 0 2 mL probe assembly from the sample block   Power off the digital thermometer  then clean off the oil     Wait for the sample block to reach room temperature   25   C   then remove the  9600 temperature verification frame from the sample block       errem PHYSICAL INJURY HAZARD  Hot Surface  Use care    when working around the heated cover and sam
98. nd the original warranty period     Applied Biosystems warrants that all optional accessories supplied with its Thermal  Cycler Temperature Verification System for GeneAmp  PCR Systems  such as  peripherals  printers  and special monitors  will be free of defects in materials and  workmanship for a period of ninety  90  days from the date the warranty begins   Applied Biosystems will repair or replace  at its discretion  defective accessories  during this warranty period  After this warranty period  Applied Biosystems will pass  on to the buyer  to the extent that it is permitted to do so  the warranty of the original  manufacturer for such accessories     With the exception of consumable and maintenance items  replaceable products or  components used on or in the instrument are themselves warranted to be free of  defects in materials and workmanship for a period of ninety  90  days     Applied Biosystems warrants that chemicals and other consumable products will be  free of defects in materials and workmanship when received by the buyer  but not  thereafter  unless otherwise specified in documentation accompanying the product     Applied Biosystems warrants that for a period of ninety  90  days from the date the  warranty period begins  the tapes  diskettes  or other media bearing the operating  software of the product  if any  will be free of defects in materials and workmanship  under normal use  If there is a defect in the media covered by the above warranty and  the medi
99. neAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 10 Calibration verification test    Calibration verification test    Required  equipment    Setting up the  0 5 mL probe    Use this test to verify the temperature calibration of your Veriti   Thermal Cycler  The  calibration verification test consists of the following procedures  which you must  perform in sequence     Setting up the 0 5 mL probe assembly               0 00 eee eee ee eee 119  Rung 1 PET 121  Ending hetes cd oos eese eve e eeER DIL EA OPES E Wed d 122      wr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the 0 5 mL Temperature Verification Kit  PN 4309924   which  includes      Cotton swabs     Light mineral oil     Temperature verification frame     0 5 mL probe assembly     Model 4500 digital thermometer with a 9V battery installed     Serial Communication Cable    1  Power on the Veriti  Thermal Cycler     For best results  allow the thermal cycler 15 to 30 minutes to warm to ambient       assembly temperature before performing the test   2  In the Main Menu  select Tools Menu   3  Connect the 4500 Digital Thermometer    a  Attach the free end of cable  from the probe assembly  to the input  connector port on the 4500 thermometer    b  Attach the 9 pin DIN female end of the RS232 cable to the thermometer  communication port  and the 9 pin DIN male end
100. nics at the address in step 1     130 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Instrument Warranty Information B    This chapter covers     Limited product Warraniy   a Loss e otia ho rg dem eoe De Ert De SOR e dde    Damages  claus  und TONOS su sed oaa ek pacer e eX Coane Res    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    131    Appendix B Limited product warranty    Limited product warranty    Limited warranty    Warranty period  effective date    Warranty claims    Applied Biosystems warrants that all standard components of its Thermal Cycler  Temperature Verification System for GeneAmp  PCR Systems will be free of  defects in materials and workmanship for a period of ninety  90  days from the date  the warranty period begins  Applied Biosystems will repair or replace  at its  discretion  all defective components during this warranty period  After this warranty  period  repairs and replacement components may be purchased from   Applied Biosystems at its published rates  Applied Biosystems also provides service  agreements for post warranty coverage  Applied Biosystems reserves the right to use  new  repaired  or refurbished instruments or components for warranty and  post warranty service agreement replacements  Repair or replacement of products or  components that are under warranty does not exte
101. nnel in the dual 96 well temperature  verification plate to prevent damage to the probe and lead wires  see Figure 7 on  page 36      6  Make sure the probe is connected to the digital thermometer     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  37  and Veriti  System User Guide    Chapter 4 Calibration verification test    7  For GeneAmp  9700 with Manual Dual 96 Well Sample Block Module   Slide the heated cover forward then pull the lever down     For GeneAmp   9700 with Auto Lid Dual 96  Well Sample Block Module   Press the red button to slide and close the heated cover     IMPORTANT  Seat the probe properly and close the heated cover carefully  If  the probe wire is crushed when you close the heated cover  the probe can be  damaged     8  Power on the digital thermometer   Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500   Configuring the 1  Power on the thermal cycler  The Main menu opens           GeneAmp      9700 For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test     2  Press F4  Util   The Utilities screen opens   3  Press F1  Diag   The Diagnostics screen opens     4  Press F3  TmpVer   The Temperature Verification screen opens     Temperature Verification    Temp   Calibration Verification   NU   Temperature Non Uniformity      Temp     TNU      F1 F2 F3 F4 F5          
102. obe in well xx    Press Run    Run       Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  29  and Veriti   System User Guide    Chapter 3 Temperature non uniformity test    7  Slide the heated cover back  Repeat steps 4 through 7 of the  Setting up the 0 2   mL probe assembly  on page 27 and steps 1 through 6 of this procedure   Complete these steps for all 8 wells to be tested  C4  C9  F4  F9  Al  A12  H1     H12     8  The GeneAmp   9700 thermal cycler evaluates the uniformity of the sample  block temperature for the setpoint values that you entered and displays the  results  A summary screen opens at the end of the test              F1 F2    F3 FA F5    If you entered values on the Temperature Non Uniformity Test Data Sheet   page 33   compare those values with the actual test results     9  Press F1  Accept      Evaluating the When the thermal cycler completes the temperature non uniformity test  the TNU  results Performance screen opens  See the table below to evaluate the results        If the       Then          Temperature of the  sample block wells is  uniform     Pass  appears after each setpoint temperature     TNU Performance     NU at 94  C is xx xx   Pass   NU at 37  C is xx xx   Pass    F1 F2 F3 FA F5             Temperature variation of  the sample block wells  exceeds performance  specifications           Fail  appears after the setpoint temperature s  for which the  test failed     TNU Performance     NU at 94 
103. ometer  Model 4500     3  Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature  field     TNU Performance    Sample temp   xx x C Cover temp   xxx  C       Enter actual block temperature 00 0     F1 F2 F3 FA F5       Note  The digital thermometer displays a four digit value  round this off to  three digits before typing it in the TNU Performance screen     Note  To keep a permanent record of the test  record this value on the  Temperature Non Uniformity Test Data Sheet  page 75      Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  71  and Veriti  System User Guide    Chapter 6 Temperature non uniformity test    4  Press Enter  The GeneAmp   9700 automatically begins the second reading   37   C setpoint   The TNU Performance screen opens with the setpoint value  displayed     TNU Performance    Sample temp   xx x  C Cover temp   xxx  C    Setpoint is 37  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 F4 F5          Note  The sample blocks must be within  1 0   C of the setpoint   5  Repeat steps 2 and 3 and for the second reading   6  Press Enter  The TNU Performance screen opens with the following prompt     TNU Performance  Sample temp   xx x  C Cover temp   xxx  C    Place probe in well xx    Press Run    Run    F1 F2 F3 F4 F5       7  Press and release the red button to automatically open the heated cover  Repeat  steps 4 through 6 of    Settin
104. on page 8      ON      Make sure the probe is connected to the digital thermometer     7  Pull the lever down to engage the heated cover and 2720 temperature  verification frame     IMPORTANT  Seat the probe properly and close the heated cover carefully  If  the probe wire is crushed when you close the cover  the probe can be damaged     8  Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500              Power on the thermal cycler  The Main menu opens     For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test     2  Press F4  Util   The Utilities screen opens     and Veriti  System User Guide    Chapter 2 Temperature calibration verification test    3  Press F1  Diag   The Diagnostics screen opens     4  Press F3  TmpVer   The Temperature Verification screen opens     Temperature Verification    Temp   Calibration Verification   NU   Temperature Non Uniformity      Temp     TNU      F1 F2 F3 F4 F5                   5  Press F1  Temp   The thermal cycler is automatically configured for the  calibration verification test and the Calibration Verification screen opens     Calibration Verification  Block temp   xx x C Cover temp   xxx C             Place probe in well A6  Press Run    Run    F1 F2 F3 F4 F5       Running the test In this test  you take temperature readings of the sample well at two differen
105. on verification test     1  Press F1  Run  to start the calibration verification test  The Calibration  Verification screen opens with the setpoint value displayed     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 85  C  Cover must be within 10  C of 85  C    F1 F2 F3 F4 F5       Note  The cover must be at 85   C  10   C  It may take several minutes for the  system to ramp up     The Calibration Verification screen counts down the time until the setpoint is  reached     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Stabilizing at setpoint X xXX       24 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 3 Temperature calibration verification test    2  When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     3  Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature    field     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Enter actual block temperature    F1 F2 F3 F4 F5       The digital thermometer displays a four digit value  Round this off to three  digits before typing it in the Calibration Verification screen     Record this value on the Calibration Verification T
106. opens  See the table below to evaluate the results     results       If the sample block  module       Then the          Is properly calibrated       Calibration Verification screen displays the following message     Calibration Verification    Calibration is good    F1 F2 F3 FA F5       Does not pass the  calibration verification  test             Calibration Verification screen displays the following message     Calibration Verification          Instrument may require servic  Contact Applied Biosystems  Technical Support        El F2 F3 F4 F5    e lf the test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors were  not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See    How to obtain support  on page viii           Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 6 Temperature non uniformity test    Temperature non uniformity test    Use this test to verify the temperature uniformity of the GeneAmp  PCR System  9700  Auto Lid Sample Block Module  The temperature non uniformity test consists  of the following procedures  which you must perform in sequence     Setting up the 20 uL probe assembly     2 06  ics eens eee ees 69  Configuring the GeneAmp  9700    lisse e 70  puni gd PPP 70  Evaluating thi results 2i soles ee eee E eec Ot COR A dd e baie e d 73      Nr  PHYSICAL INJURY HAZARD  Hot Surf
107. or details on how to set up a CYCL program     3  On the third cycle  measure the temperature of well A1 90 seconds into Setpoint   1  95   C setpoint temperature  using the digital thermometer  The time   remaining clock on the run time display reads    0 30     30 seconds   Record this  temperature     4  Again on the third cycle  measure the temperature of well Al 90 seconds into  Setpoint  2  40   C setpoint temperature  using the digital thermometer  The  time remaining clock on the run time display reads    0 30     30 seconds   Record  this temperature     The figure below shows when to measure the temperatures     Measure well A1 Measure well A4  90 seconds into 90 seconds into  Setpoint  1 Setpoint  1    Measure well A1   Measure well A4       90 seconds into  Setpoint  2    90 seconds into  Setpoint  2              95            Move the  probe  to well A8    Move the  probe  to well A4         Cycle  3   Cycle  4    5  After you measure the second temperature of well A1  turn the cover knob  completely counterclockwise  then slide the heater cover back     6  Move the probe assembly to the next well to be measured     7  Slide the heater cover forward  then turn the cover knob clockwise until the  white mark on the knob and the white mark on the cover are aligned     8  Repeat the measurements on wells A4  A8  A12  C1  C4  C8  C12  El  E4  E8   E12  H1  H4  H8  H12     Note  The temperature display on the digital thermometer may not match the  temperature display o
108. ple block to avoid being burned  by hot components     Clean the oil off the sample block     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  113    and Veriti   System User Guide    Chapter 9 Data Sheet  Calibration Verification Test    Data Sheet  Calibration Verification Test    114    When you run the calibration verification test  record the setpoint values for well E1  on this data sheet  At the end of the test  compare the values that are displayed on the  thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date Tested by   Probe serial no     Thermometer  serial no     Setpoint value  Well E1       95   C 40   C                                                                                              Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 9 Data Sheet  Temperature Non Uniformity Test    Data Sheet  Temperature Non Uniformity Test    When you run the temperature non uniformity test  record the setpoint values for the  wells listed on this data sheet  At the end of the test  compare the values that are  displayed on the thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date       Tested by       Probe serial no        Thermometer  serial no              Setpoint value 95  C 40  C       A1       A4       A8       A12       C1 
109. r carefully  If  the wire is crushed when you close the heated cover  the probe can be damaged     8  Power on the digital thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     p lt       Power on the thermal cycler  The Main menu opens     For best results  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test     2  Press F4  Util   The Utilities screen opens     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  23  and Veriti  System User Guide    Chapter 3 Temperature calibration verification test    3  Press F1  Diag   The Diagnostics screen opens     4  Press F3  TmpVer   The Temperature Verification screen opens     Temperature Verification    Temp   Calibration Verification   NU   Temperature Non Uniformity             TNU    F1 F2 F3 F4 F5          Temp          5  Press F1  Temp   The thermal cycler is automatically configured for the  calibration verification test and the Calibration Verification screen opens     Calibration Verification  Block temp   xx x C Cover temp   xxx C             Place probe in well A6  Press Run    Run    F1 F2 F3 F4 F5       Running the test In this test  you take temperature readings of the sample well at two different setpoint  temperatures using the 0 2 mL probe assembly     Note  Press F5  Cancel  if you want to end the test     To run the calibrati
110. r interfacing  or use in combination with software or products  not supplied or  authorized by Applied Biosystems  and modification or repair of the product not  authorized by Applied Biosystems     THE FOREGOING PROVISIONS SET FORTH APPLIED BIOSYSTEMS     SOLE AND EXCLUSIVE REPRESENTATIONS  WARRANTIES  AND  OBLIGATIONS WITH RESPECT TO ITS PRODUCTS  AND   APPLIED BIOSYSTEMS MAKES NO OTHER WARRANTY OF ANY KIND  WHATSOEVER  EXPRESSED OR IMPLIED  INCLUDING WITHOUT  LIMITATION  WARRANTIES OF MERCHANTABILITY AND FITNESS  FOR A PARTICULAR PURPOSE  WHETHER ARISING FROM A STATUTE  OR OTHERWISE IN LAW OR FROM A COURSE OF DEALING OR USAGE  OF TRADE  ALL OF WHICH ARE EXPRESSLY DISCLAIMED     Warranty THE REMEDIES PROVIDED HEREIN ARE THE BUYER S SOLE AND   limitations EXCLUSIVE REMEDIES  WITHOUT LIMITING THE GENERALITY OF  THE FOREGOING  IN NO EVENT SHALL APPLIED BIOSYSTEMS BE  LIABLE  WHETHER IN CONTRACT  TORT  WARRANTY  OR UNDER  ANY STATUTE  INCLUDING WITHOUT LIMITATION  ANY TRADE  PRACTICE  UNFAIR COMPETITION  OR OTHER STATUTE OF SIMILAR  IMPORT  OR ON ANY OTHER BASIS  FOR DIRECT  INDIRECT   PUNITIVE  INCIDENTAL  MULTIPLE  CONSEQUENTIAL  OR SPECIAL  DAMAGES SUSTAINED BY THE BUYER OR ANY OTHER PERSON OR  ENTITY  WHETHER OR NOT FORESEEABLE AND WHETHER OR NOT  APPLIED BIOSYSTEMS IS ADVISED OF THE POSSIBILITY OF SUCH  DAMAGES  INCLUDING WITHOUT LIMITATION  DAMAGES ARISING  FROM OR RELATED TO LOSS OF USE  LOSS OF DATA  FAILURE OR  INTERRUPTION IN THE OPERATION OF ANY EQUIPMENT OR  SOFTWAR
111. rded here     Note  You can photocopy this page        Date Tested by   Probe serial no     Thermometer  serial no     Setpoint value  Well A6       85   C 45  C                                                                                              Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 3 Data Sheet  Temperature Non Uniformity Test    Data Sheet  Temperature Non Uniformity Test    When you run the temperature non uniformity test  record the setpoint values for the  wells that are listed on this data sheet  At the end of the test  compare the values that  are displayed on the thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date       Tested by       Probe serial no        Thermometer  serial no           Setpoint value 94  C 37   C       C4       c9       F4       F9       A1       A12       H1       H12                   Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  33  and Veriti   System User Guide    Chapter 3 Data Sheet  Temperature Non Uniformity Test    34 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Temperature Verification Tests for the  GeneAmp  9700  Dual 96 Well Module    This chapter covers   Overview  Calibration verification test  Temperature non uniformity test  Data Shee
112. re non uniformity test  the TNU  results Performance screen opens  See the table below to evaluate the results        If the       Then          Temperature of the  sample block wells is  uniform        Pass  appears after each setpoint temperature     TNU Performance    TNU at 94  C is xx xx   Pass    TNU at 37  C is xx xx   Pass    F1 F2 F3 FA F5       Temperature variation of  the sample block wells  exceeds performance  specifications           Fail  appears after the setpoint temperature s  for which the  test failed     TNU Performance    TNU at 94  C is xx xx   Fail    TNU at 37  C is xx xx   Fail          F1 F2 F3 FA F5      If the test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors were  not made     e If the test fails again  contact Applied Biosystems Technical  Support  See    How to obtain support    on page viii           Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  59    and Veriti   System User Guide    Chapter 5 Data Sheet  Calibration Verification Test    Data Sheet  Calibration Verification Test    60    When running the calibration verification test  record the setpoint values for well R3  on this data sheet  At the end of the test  compare the values that are displayed on the  thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date Tested by   Probe serial no     Thermometer  serial no     Setpoint v
113. rification Kit for details on  operating the digital thermometer  Model 4500   Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  99    and Veriti   System User Guide    Chapter 8 Temperature non uniformity test    5  Using the numeric keys  enter the value that is displayed on the digital  thermometer into the    Enter actual block temperature    field     TNU Performance    Sample temp   xx x  C Cover temp   xxx  C       Enter actual block temperature 00 0      F1 F2 F3 F4 F5       Note  Round the four digit value on the digital thermometer to three digits  before you enter it in the TNU Performance screen     Note  You can record this value on the Temperature Non Uniformity Test Data  Sheet  page 104  to keep a permanent record of the test     6  Press Enter to open the TNU Performance screen with the setpoint value  displayed     TNU Performance  Sample temp   xx x  C Cover temp   xxx  C    Setpoint is 55  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 F4 F5          Note  The sample blocks must be within  1 0   C of the setpoint     7  Press Enter  The TNU Performance screen opens with the following prompt     TNU Performance  Sample temp   xx x  C Cover temp   xxx  C    Place probe in well xx    Press Run    Run       100 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    8     10     11     Chapter 8 Temperature non uniformity test    
114. rification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 6 Data Sheet  Temperature Non Uniformity Test    Data Sheet  Temperature Non Uniformity Test    When running the temperature non uniformity test  record the setpoint values for the  wells listed on this data sheet  At the end of the test  compare the values that are  displayed on the thermal cycler with the values that are recorded here     Note  You can photocopy this page                                                                             Date   Tested by   Probe serial no    Thermometer   serial no    Setpoint value 94  C 37   C  L1  L2  L3  L4  L5  L6  R1  R2  R3  R4  R5  R6  Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  98800  75    and Veriti   System User Guide    Chapter 6 Data Sheet  Temperature Non Uniformity Test    76 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Temperature Verification Tests for the GeneAmp   9700  0 5 mL Sample Block Module    This chapter covers    Calibration ventication  e8L  655050 nas e em oe Re  Ro e on dA  Temperature nop uniformity Testo isa r kr eor e RARO Pp c ead dee d  Data Sheet  Calibration Verification Test occ cae sio hee aac ae der ras  Data Sheet  Temperature Non Uniformity Test          llle     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  97
115. rmal  cycler    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     Use this test to verify the temperature calibration of your 2720 thermal cycler that  has a 0 2 mL sample block module  The temperature calibration verification test  consists of the following procedures  which you must perform in sequence     Setting up the 0 2 mL probe assembly scos ccc icem en 9  Configuring the 2720 thermal cycler             0    cece cece eee eee 9  Bunning hete lolo tiers gies ETETE TPI EEEREN EEES DE PPS ee 10  Exaluatdape ths regulis esoo eers poder eere bod Ha eee Te a gat 12  Ending ihe desk    occorre Dex dece sare a cr a eed 12      wr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the 0 2 mL Sample Block Module Temperature Verification Kit   PN 4317939   which includes       Cotton swabs     Light mineral oil     2720 temperature verification frame     0 2 mL probe assembly     Model 4500 digital thermometer with a 9V battery installed    1  Lift the lever and open the hinged heated cover   Place the 2720 temperature verification frame on the sample block   Using a cotton swab  coat well A6 with mineral oil     Place the 0 2 mL probe assembly into well A6     TE M      Thread the probe wire through the channel in the 2720 temperature verification  frame to prevent damage to the probe and lead wires  see Figure 3 
116. rmity test    Evaluating the When the GeneAmp   9700 completes the temperature non uniformity test  the TNU  results Performance screen opens  See the table below to evaluate the results        If the       Then          Temperature of the  sample block wells is  uniform     Pass  appears after each setpoint temperature     TNU Performance    TNU at 94  C is xx xx   Pass    TNU at 37  C is xx xx   Pass    F1 F2 F3 FA F5          Temperature variation of  the sample block wells  exceeds performance  specifications           Fail  appears after the setpoint temperature s  for which the  test failed     TNU Performance    TNU at 94  C is xx xx   Fail    TNU at 37  C is xx xx   Fail          F1 F2 F3 FA F5    e lf the test fails  repeat the procedure to make sure that the  thermometer was not misread or that data entry errors were  not made     e Ifthe test fails again  contact Applied Biosystems Technical  Support  See    How to obtain support  on page viii           Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  87    and Veriti  System User Guide    Chapter 7 Data Sheet  Calibration Verification Test    Data Sheet  Calibration Verification Test    88    When running the calibration verification test  record the setpoint values for well A6  on this data sheet  At the end of the test  compare the values that are displayed on the  thermal cycler with the values that are recorded here     Note  You can photocopy this page        Da
117. rogrammed target temperature  your instrument would not need to be recalibrated     Note  The offset is the number of degrees Celsius that the temperature of well E1  differed from the average temperature of the block when the instrument was  calibrated at the factory    Average block temperature at the 40   C hold    Use the following formula to calculate the average block temperature at the 40  C  hold     Block Average at 40   C   T 40      Low Offset    If the block average is more than 0 75 degrees Celsius above or below 40  C  your  GeneAmp   9600 must be recalibrated     For example     If the measured temperature of well E1 is 39 9   C  and the Low Offset that is printed  on your calibration label is  0 1  you would make the following calculation     Block Average at 40   C   39 9       0 1    39 8   C          In this example  because 39 8   C does not differ by  0 75 degrees Celsius from the  programmed target temperature  your instrument would not need to be recalibrated     1  Remove the 0 2 mL probe assembly from the sample block   2  Power off the digital thermometer  then clean off the oil     3  Wait for the sample block to reach room temperature   25   C   then remove the  9600 temperature verification frame from the sample block       NITE PHYSICAL INJURY HAZARD  Hot Surface  Use care    when working around the heated cover and sample block to avoid being burned  by hot components     4  Clean the oil off the sample block     Thermal Cycler Temperature Verif
118. rs   Overview  Temperature calibration verification test  Temperature non uniformity test    Data Sheet  Calibration Verification Test  Data Sheet  Temperature Non Uniformity Test    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    21    Chapter 3 Overview    Overview    Materials When you perform a temperature verification test on a GeneAmp   9700 with a 96   Well Sample Block Module  use the materials in the kit  PN 4317939      Assembly Figure 5 shows a properly assembled system  Note that the frame is positioned  around the wells with the channel facing you  The location tab on the frame should  align with the locator tab on the thermal cycler  The probe wire is threaded through  the channel to prevent damaging the wire when you close the thermal cycler heated  cover                             GR1916    Figure 5 Proper assembly of the Temperature Verification System for  GeneAmp  9700 thermal cycler       Figure 6 Temperature verification frame    22 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 3 Temperature calibration verification test    Temperature calibration verification test    Required  equipment    Setting up the  0 2 mL probe  assembly    Configuring the  GeneAmp  9700    Use this test to verify the temperature calibration of your GeneAmp   9700 thermal  cycler that has a 0 2 mL sample 
119. s  allow the thermal cycler 15 to 30 minutes to warm up to  ambient temperature before performing the test     3  Press F4  Util   The Utilities screen opens   4  Press F1  Diag   The Diagnostics screen opens     5  Press F3  TmpVer   The Temperature Verification screen opens     Temperature Verification    Temp   Calibration Verification   NU   Temperature Non Uniformity            Temp J   TNU jJ    F1 F2 F3 FA F5             6  Press F1  Temp  to automatically configure the GeneAmp   9700 for the  calibration verification test  The Calibration Verification screen opens     Calibration Verification  Block temp   xx x C Cover temp   xxx C             Place probe in well R3  Press Run    Run    F1 F2 F3 FA F5       7  Continue with    Running the test    below     Running the test Use the digital thermometer to take temperature readings of the sample well that is  connected to the 0 02 mL probe assembly  You must take a reading at two different  setpoint temperatures     Note  If necessary  press F5  Cancel  to exit the test     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  51  and Veriti  System User Guide    Chapter 5 Calibration verification test    To run the calibration verification test   1  Complete the procedures in Configuring the GeneAmp  9700  on page 51     2  Press F1  Run  to start the calibration verification test  The Calibration  Verification screen opens with the setpoint value displayed     Calibration Veri
120. s the on off switch on its front      The temperature measured by the probe assembly appears on the digital  thermometer display in degrees Celsius      When you complete a test  press the on off switch to power off the digital  thermometer     The sample temperature that is displayed on the thermal cycler display screen differs  from the temperature that 1s displayed on the digital thermometer during heating or  cooling transitions  This is because the thermal cycler displays the calculated sample  temperature  while the digital thermometer measures and displays the sample block  temperature  The sample temperature that is displayed by the thermal cycler is  calculated based on the ramp rate and reaction volume     Applied Biosystems recommends that you use the probe assembly and digital  thermometer only for the temperature verification test and the temperature non   uniformity test  which are described in this manual     If you use the probe assembly and digital thermometer while running programs other  than those used in these two tests  be aware that the accuracy of the probe data  reported by the assembly decreases  The decrease is due to the variability introduced  into the test by placing the probe in alternate well locations and or the selection of  temperatures other than those recommended by Applied Biosystems     6 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Temperature Verifi
121. ser attention words viii    D  damage  reporting 134  damaged items  returning 134  DANGER  description vii  data sheets  system 2720 18   19  system 9600 114   115  system 9700 32   33  system 9700 with 0 5 mL sample block module 88   89  system 9700 with Auto Lid sample block  module 74   75  system 9700 with dual 384 well sample block  module 60   61  system 9700 with dual 96 well sample block  module 47   48  system 9800 103   104  system Veriti 127   128  digital thermometer 5   6  Calibration Verification Test 50    E    equipment overview 2    H    hazard icons  See safety symbols  on instruments 138  hazard symbols  See safety symbols  on instruments    IMPORTANT  description vii  italic text  when to use viii    M    menu commands  conventions for describing viii    MSDSs  about vii  obtaining viii    P    probe assembly 5    R    RA number 134   RA number  See return authorization number  recalibration 4  129   130   reporting  damages 134   return authorization  RA  number 134  returning damaged items 134   returns 134    S    safety symbols  on instruments 138    sample block  verifying temperature uniformity  evaluating the results 59    symbols  safety 139  systems covered 2    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  141    and Veriti  System User Guide    T    Temperature Non Uniformity Test  system 2720 13   17  system 9600 111   113  system 9700 27   31  system 9700 with 0 5 mL sample block module 83   
122. ss F3  TmpVer   The Temperature Verification screen opens     Temperature Verification       F1 F2 F3 F4 F5    5  Press F2  TNU      The thermal cycler automatically starts to configure the temperature non   uniformity test  with the cover setpoint of 35   C and the sample starting with the  setpoint of 37   C  It may take several minutes for the system to stabilize at the  cover and sample setpoint temperatures  After the system is configured  the  TNU Performance screen opens     TNU Performance  Sample temp   xx x C Cover temp   xxx C    Place Probe in well A1  Press Run    Run    F1 F2 F3 F4 F5          Running the test The temperature non uniformity test uses the 0 2 mL probe assembly to test the  temperature uniformity of 8 different wells in the sample block     Note  Press F5  Cancel  if you want to exit the test     To run the temperature non uniformity test     1  Place the probe into the indicated well  then press F1  Run   The temperature  non uniformity test for the setpoint temperature of 94   C starts  and the TNU  Performance screen opens with the setpoint value displayed     TNU Performance  Sample temp   xx x C Cover temp   xxx C    Setpoint is 94  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 F4 F5          Note  The sample block increases from 37   C to 94   C  The sample block must  be at 94   C 1 0   C  It may take several minutes for the 2720 thermal cycler to  stabilize at the setpoint temperature        14 Thermal Cycler Temperature Verifica
123. st    ho  ooced e TETEE VETE TPI EREET EEE EEE ET EEEE E 39  Exaluatidae the regulis os eies p deb eere bod genau d nega oed 41  Ending We desk    occorre Dr dee sare sa dre 41      Nr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the 0 2 mL Sample Block Module Temperature Verification Kit   PN 4317939   which includes      Cotton swabs     Light mineral oil     Left and right dual 96 well block temperature verification plates     0 2 mL probe assembly     Model 4500 digital thermometer with a 9V battery installed    1  For GeneAmp  9700 with Manual Dual 96 well Sample Block Module   If the heated cover is in the forward position  lift the lever  then slide the heated  cover back   For GeneAmp   9700 with Auto Lid dual 96  Well Sample Block Module   If the heated cover is in the forward position  lift the lever  then press the red  button to slide the heated cover back     2  Place both dual 96 well temperature verification plates on the sample block   The plate labeled with L1 to L6 is for the left sample block  The plate labeled  with R1 to R6 is for the right sample block  Position the plates so the angled  corners face away from you     Note  The plates help you locate the probes within the sample blocks   3  Using a cotton swab  coat well R3 with mineral oil   4  Place the 0 2 mL probe assembly into well R3     5  Thread the probe wire through the cha
124. t  Calibration Verification Test  Data Sheet  Temperature Non Uniformity Test    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    35    Chapter 4 Overview                               Overview  Materials When you perform the temperature verification test on a GeneAmp  9700 with a  dual 96 well sample block module  you use the materials in the kit  PN 4317939    Assembly Figure 7 show the right frame of a properly assembled temperature verification  system  Note that the probe wire 1s threaded through the channel to prevent damage  to the wire when you close the thermal cycler heated cover   e e e e e   Figure 7 The temperature verification system for GeneAmp  9700 with a dual  96 well sample block module  right frame shown   Figure8 Left frame Figure9 Right frame   36 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 4 Calibration verification test    Calibration verification test    Required  equipment    Setting up the  0 2 mL probe  assembly    Use this test to verify the temperature calibration of a GeneAmp   9700 with a dual  96 well sample block module  The calibration verification test consists of the  following procedures  which you must perform in sequence     Setting up the 0 2 mL probe assembly 1    ss cascade een ERR Rn ae  Configuring the GeneAmp   9700    2    0 0 cece e 38  Bunning the te
125. t setpoint  temperatures using the 0 2 mL probe assembly     Note  Press F5  Cancel  if you want to end the test     To run the calibration verification test     1  Press F1  Run  to start the calibration verification test  The Calibration  Verification screen opens with the setpoint value displayed     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 85  C  Cover must be within 10  C of 85  C    F1 F2 F3 F4 F5       Note  The cover must be at 85   C  10   C  It may take several minutes for the  system to ramp up     Note  The Calibration Verification screen counts down the time until the  setpoint is reached     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Stabilizing at setpoint X XX       10 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 2 Temperature calibration verification test    2  When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     3  Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature    field     Calibration Verification  Block temp   xx x  C Cover temp   xxx  C       Enter actual block temperature    F1 F2 F3 FA F5       The digital thermometer displays a four digit value
126. tal thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     9  Make sure the probe assembly is connected to the digital thermometer and that  the thermometer is powered on     10  Slide the heated cover forward and pull the lever down     124 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 10 Temperature non uniformity test    Running the test 1  In the Tools menu  select Run TNU Test      gt          2007 09 04 I 04 05 PM    2  Press to begin the test  Perform the test as directed by the touchscreen  As  the test progresses  move the 0 5 mL probe assembly to the remaining test wells   Al  A10  B3  B8  E3  E8  F1  F10    Temperature Non uniformity Test   Step 1 of 10    You will need the temperature verification kit  TVK    Connect the TVK and turn it on now      COOOOOOOOO  COOOOOO000    9000000000  5000000000  000000000  200000000    Touch the right arrow when ready       3  When the test is complete  record the TNU in the    Data Sheet  Temperature  Non Uniformity Test  on page 128     jn    Temperature Non uniformity Test   Step 10 of 10         EC               r                   TNU 95   C TNU 60   C Results  Overall   0 17  94 80   85 13    0 16  59 88   60 21   Pass  Press the cross button when you want to exit      4  Press X to return to the Tools menu     Thermal Cycler Temperature Veri
127. te Tested by   Probe serial no     Thermometer  serial no     Setpoint value  Well A6       85   C 45  C                                                                                              Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 7 Data Sheet  Temperature Non Uniformity Test    Data Sheet  Temperature Non Uniformity Test    When running the temperature non uniformity test  record the setpoint values for the  wells listed on this data sheet  At the end of the test  compare the values that are  displayed on the thermal cycler with the values that are recorded here     Note  You can photocopy this page        Date       Tested by       Probe serial no        Thermometer  serial no              Setpoint value 94  C 37   C       A1       A6       A10       B3       c5       C10       D1       D7       E8       F1       F5       F10                   Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  89  and Veriti  System User Guide    Chapter 7 Data Sheet  Temperature Non Uniformity Test    90 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Temperature Verification Tests for the  GeneAmp   9800    This chapter covers     Calibration verification test          2 00 02  cece eee eee  Temperature non uniformity test        2 2    ce eee eee eee  Data
128. tion    Temp   Calibration Verification   NU   Temperature Non Uniformity                  Temp     TNU jJ    F1 F2 F3 FA F5       6  Press F2  TNU  to automatically configure the GeneAmp   9700 for the  temperature non uniformity test  The TNU Performance screen opens     TNU Performance  Sample temp   xx x C Cover temp   xxx  C    Place Probe in well A1  Press Run    Run    F1 F2 F3 F4 F5          7  Continue with    Running the test  below   Running the test     The temperature non uniformity test uses the 0 5 mL probe assembly to test the  temperature uniformity of 12 different wells in the sample block     Note  If necessary  press F5  Cancel  to exit the test     To run the temperature non uniformity test     1  Complete the procedures in    Setting up the 0 5 mL probe assembly    above     2  Press F1  Run  to start the temperature non uniformity test  The TNU  Performance screen opens with the setpoint value displayed     TNU Performance  Sample temp   xx x C Cover temp   xxx  C    Setpoint is 94  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 FA F5          Note  The sample block must be within  1 0   C of the setpoint  In addition  the  cover must be within  1   C of 105   C  It may take several minutes for the    GeneAmp  9700 to ramp up   84 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 7 Temperature non uniformity test    3  The TNU Performance screen counts do
129. tion System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti   System User Guide    Chapter 2 Temperature non uniformity test    2  When the sample block reaches 94   C  the TNU Performance screen counts  down the time until the setpoint is stable     TNU Performance    Sample temp   xx x  C Cover temp   xxx  C    Stabilizing at setpoint X XX       F1 F2 F3 FA F5    3  When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the Model 4500 digital thermometer     4  Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature    field   TNU Performance    Sample temp   xx x C Cover temp   xxx C    Enter actual block temperature 00 0     F1 F2 F3 FA F5          The digital thermometer displays a four digit value  Round off the value to three  digits before you enter it in the TNU Performance screen     Record the value on the Temperature Non Uniformity Test Data Sheet  page 19   to keep a permanent record of the test     5  Press Enter  The following TNU Performance screen opens and the system  automatically begins decreasing its setpoint temperature to 37  C     TNU Performance  Sample temp   xx x C Cover temp   xxx  C    Setpoint is 37  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 F4 F5          6  When the setpoint temperature reaches 37   C  the TNU Performanc
130. tion frame from the sample block       Nr PHYSICAL INJURY HAZARD  Hot Surface  Use care    when working around the heated cover and sample block to avoid being burned    by hot components     4  Clean the oil off the sample block     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide       Chapter 2 Temperature non uniformity test    Temperature non uniformity test    Required  equipment    Setting up the    Use this test to verify the temperature uniformity of the 2720 thermal cycler with a  0 2 mL sample block module  The temperature non uniformity test consists of the  following procedures  which you must perform in sequence     Betting up the 0 2 mL probe assembly 1    cece cs eee ERR 13  Contigunng the 2720 thermal eyelet   5  oce eee Re meme 14  Bunning the test  PPP ce E abso bne ed 14  Exaluatite tig tess  esoo eeen Ez der ane eh kd Han ae nae a ee 16  Ending ibetesk   2  ecco sse aka A n Oe ee sore c a ders IT      wr  PHYSICAL INJURY HAZARD  Hot Surface  Use care when    working around the heated cover and sample block to avoid being burned by hot  components     This test requires the 0 2 mL Sample Block Module Temperature Verification Kit   PN 4317939   which includes      Cotton swabs     Light mineral oil     2720 temperature verification frame     0 2 mL probe assembly     Model 4500 digital thermometer with a 9V battery installed    1  Lift the lever and open the hinged heated cover
131. uniformity test             llle 97  Data Sheet  Calibration Verification Test     2 0 2    0    eee 103  Data Sheet  Temperature Non Uniformity Test           0 0 00 cece eee 104    Temperature Verification Tests for the GeneAmp   9600    BU MI ARDET EPI 106  Calibration verification test            liliis 107  Temperature non uniformity test      0 0      ee e 111  Data Sheet  Calibration Verification Test      0 0    00  eee 114  Data Sheet  Temperature Non Uniformity Test          00 0 0 0 eee eee 115    Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti   System User Guide    Temperature Verification Tests for the Veriti  Thermal Cycler    OVOLVIOW     esset rentas epa  Prts eu te eto Datis al OG s weds dole dee Tikes cas ts rade  Calibration verification test         n a aaaeeeaa  Temperature non uniformity test    0 2 0    cee ren  Data Sheet  Calibration Verification Test            llle  Data Sheet  Temperature Non Uniformity Test             lille    Appendix A Recalibration    Obtaining recalibration  os icer ra i e ODE aa e a EA RII    Appendix B Instrument Warranty Information    Limited product warranty isre eaa pA EAA A tees  Damages  claims  and returns            00 00 cee eee    Appendix C GeneAmp   PCR 9700 Technical Note    Temperature Accuracy Verification Test 2 1 2    0    cee a    Appendix D Safety  Instrumentation safety 138    Symbols on instruments      sassa aana naana nannaa tees    Index
132. ure non uniformity test    2  When the sample block reaches 94  C  the TNU Performance screen counts  down the time until the setpoint is stable     TNU Performance    Sample temp   xx x  C Cover temp   xxx  C    Stabilizing at setpoint X XX       F1 F2 F3 FA F5    3  When the    Stabilizing at setpoint    value decreases to 0  read the digital  thermometer     Refer to the instructions in your Temperature Verification Kit for details on  operating the Model 4500 digital thermometer     4  Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature    field   TNU Performance    Sample temp   xx x C Cover temp   xxx C    Enter actual block temperature 00 0     F1 F2 F3 FA F5          The digital thermometer displays a four digit value  Round off the value to three  digits before you enter it in the TNU Performance screen     Record the value on the Temperature Non Uniformity Test Data Sheet  page 33   to keep a permanent record of the test     5  Press Enter  The following TNU Performance screen opens and the system  automatically begins decreasing its setpoint temperature to 37  C     TNU Performance  Sample temp   xx x C Cover temp   xxx  C    Setpoint is 37  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 F4 F5          6  When the setpoint temperature reaches 37   C  the TNU Performance screen  opens with the following prompt     TNU Performance  Sample temp   xx x  C Cover temp   xxx  C    Place pr
133. well xx    Press Run       Run  F1 F2 F3 F4 F5  Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  85    and Veriti   System User Guide    Chapter 7 Temperature non uniformity test    86    8     10     11     Slide the heated cover back  and repeat steps 4 through 7 of    Setting up the 0 5   mL probe assembly  on page 83 and steps 2 through 7 of this procedure   Complete these steps for all 12 wells to be tested     Al  A6  A10  B3  C5  C10  D1  D7  E8  F1  F5  and F10  The GeneAmp   9700 evaluates the uniformity of the sample block temperature    for the setpoint values that you entered and displays the results  A summary  screen opens at the end of the test                 F1 F2 F3 FA F5    If you entered values on the Temperature Non Uniformity Test Data Sheet   compare those values with the actual test results     Press F1  Accept   To interpret the results  see  Evaluating the results  on  page 87   When you complete all measurements  be sure to      Press Cancel      Remove the 0 5 mL probe assembly from the sample block      Power off the digital thermometer  then clean off the oil      Remove the 0 5 mL temperature verification frame from the sample block     IMPORTANT  Make sure the sample block is at room temperature  725   C   before removing the frame     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800     and Veriti  System User Guide    Chapter 7 Temperature non unifo
134. with the setpoint  value displayed     Calibration Verification  Block temp   xx x C Cover temp   xxx C       Setpoint is 45  C  Cover must be within 1  C of 105  C    F1 F2 F3 FA F5       Note  The cover must be within  1   C of 105   C   5  Repeat steps 2 and 3 for the second reading     6  The GeneAmp   9700 evaluates the calibration of the sample block temperature  for the setpoint values that you entered and displays the results  A summary  screen opens at the end of the test     Calibration Verification    Actual temperature at 85  C       Actual temperature at 45  C    F1 F2 F3 FA F5       If you entered values on the Calibration Verification Test Data Sheet  compare  those values with the actual test results     7  Press F1  Accept   To interpret the results  see    Evaluating the results    on  page 68   8  When you complete all measurements  be sure to     Press F5  Exit      Remove the 20 uL probe assembly from the sample blocks     Power off the digital thermometer  then clean off the oil     Remove the 384 well temperature verification plates from the sample  blocks     IMPORTANT  Make sure the sample blocks are at room temperature   25   C   before removing the reaction plates     Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  67  and Veriti  System User Guide    Chapter6 Calibration verification test    68    Evaluating the When the GeneAmp   9700 completes the calibration verification test  one of two  screens 
135. wn the time until the setpoint is stable   When the    Stabilizing at setpoint    value decreases to 0  read the thermometer     TNU Performance    Sample temp   xx x  C Cover temp   xxx  C    Stabilizing at setpoint    x xx       F1 F2 F3 FA F5    Refer to the instructions in your Temperature Verification Kit for details on  operating the digital thermometer  Model 4500     4  Using the numeric keys  enter the value that is displayed on the digital  thermometer in the    Enter actual block temperature    field   TNU Performance    Sample temp   xx x C Cover temp   xxx  C    Enter actual block temperature 00 0     F1 F2 F3 FA F5          Note  The digital thermometer displays a four digit value  round this off to  three digits before you enter it in the TNU Performance screen     Note  Record this value on the Temperature Non Uniformity Test Data Sheet   page 89  to keep a permanent record of the test     5  Press ENTER  The GeneAmp  9700 automatically begins the second reading   37  C setpoint   The TNU Performance screen opens with the setpoint value  displayed     TNU Performance  Sample temp   xx x  C Cover temp   xxx  C    Setpoint is 37  C  Sample must be within 1 0  C of setpoint    F1 F2 F3 FA F5       Note  The sample block must be within  1 0   C of the setpoint   6  Repeat steps 3 and 4 for the second reading   7  Press ENTER  The TNU Performance screen opens with the following prompt     TNU Performance  Sample temp   xx x  C Cover temp   xxx  C    Place probe in 
136. x C    136 Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Safety       This appendix covers        Instrumentation safety    0    cent e ne e ene 138  Symbols on instruments        0 0    n ees 138  Thermal Cycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800  137    and Veriti   System User Guide    Appendix D Safety    Instrumentation safety    Symbols on instruments    Electrical The following table describes the electrical symbols that may be displayed on  symbols on Applied Biosystems instruments   instruments       Symbol Description         Indicates the On position of the main power switch     O Indicates the Off position of the main power switch     Indicates a standby switch by which the instrument is switched on to  the Standby condition  Hazardous voltage may be present if this  switch is on standby     Indicates the On Off position of a push push main power switch     Indicates a terminal that may be connected to the signal ground  reference of another instrument  This is not a protected ground  terminal     earth ground before any other electrical connections are made to the  instrument     Indicates a terminal that can receive or supply alternating current or  voltage     Indicates a terminal that can receive or supply alternating or direct  current or voltage     Q Indicates a protective grounding terminal that must be connected to  Cmm  Sa  C
137. ycler Temperature Verification System For GeneAmp   PCR Systems 2720  9600  9700  9800   and Veriti  System User Guide    Chapter 5 Calibration verification test    Calibration verification test    Required  equipment    Setting up the  0 02 mL probe  assembly    Use the calibration verification test to verify the temperature calibration of your  GeneAmp  PCR System 9700  Dual 384 Well Sample Block Module  The calibration  verification test consists of the following procedures  which you must perform in  sequence     Setting up the 0 02 mL probe assembly               00 cece cece teenies 50  Configuring the GeneAmp  9700        00    n 51  Euonning tbe Test  1 oco asian ode aan eo Rawle Oh er e ove X ae 51  Evaluating the Tesulis  iaces exse exce der nm ace der e wha eee as 54    This test requires the 0 02 mL Dual 384 Well Sample Block Module Temperature  Verification Kit  PN 4308354   which includes      Light mineral oil     Left and right 384 well temperature verification plates     0 02 mL probe assembly     Digital thermometer Model 4500 with a 9V battery installed    1  Ifthe heated cover is in the forward position  lift the lever  then slide the heated  cover back     2  Place both 384 well temperature verification plates on the sample blocks     IMPORTANT  The temperature verification plates must be installed in both  sample blocks throughout all temperature verification testing     The plate labeled L1 to L6 is for the  eft sample block  Figure 10   The plate  labe
    