Home
        BTA-060 - Operating Instructions PASCAL CS
         Contents
1.  The partially edited  value will not be saved     When the right most digit is selected  the middle button confirms the whole value   Stor   appears on the display to confirm that the value has been stored and the device switches  back to the menu item for the value     You can store the partially edited value at any digit position by holding the middle button  until  Stor  appears on the display  approx  two seconds         Figure 9  Button functions for entering numeric values a  limit to allowed values  b  changing one digit   c  incrementing the more significant digit when stepping over zero       BTA 060  Rev 1K1 Page 9 18    6 2 Main menu    The main menu contains the following functions                                Display Designation Description  I  IC    Basic functions Setting the unit  setting the zero point   oe   min  max  pointer  I         Display functions All settings relating to the display  Li tli   _ cl Switch point settings Configuration of the switching outputs  i  optional   L_ Lit System data Displaying system data  versions  serial  _ _I_i number   reset to factory settings     C L       Return Return to display mode       Table 3  The items in the main menu    6 3 Basic menu  bASE     The basic menu contains the following items                                Display Designation Description    I   Measuring unit Setting the measuring unit via a parameter  Lot th list  LI Set zero point Setting the applied pressure as zero point  l il  Teach lower r
2.  switch back to the corresponding menu entry  The selected value will not be  saved     The figure below depicts the button functions in a parameter list  E g  if you are in the  parameter list for the unit  you can scroll thru the available units with the arrow buttons   With the middle button you store the displayed unit   Stor  appears on the display to  confirm that the changed unit has been stored and the device switches back to the menu  item for selecting the measuring unit  unit         BTA 060  Rev 1K1 Page 8 18    previous  list entry  list entry    y    menu item of  parameter list               next  list entry    A  LPS      Stor  V    Figure 8  Button functions in a selection menu and example with parameter list for measuring units       Setting a numerical value  Numerical values are entered digit by digit  The selected digit flashes and is incremented  with the up arrow button and decremented with the down arrow button  The more  significant digit will also be incremented or decremented when stepping over zero  If a  change of the active digit would exceed the allowable value  e g  the lower or upper range  limit  the allowable value will be shown  With the opposite arrow button you can return to  the previous value     You confirm the selected digit with the middle button and proceed to the next digit     You can cancel the value entry at any time by pressing both arrow buttons simultaneously   The device will then switch back to the corresponding menu entry 
3. 1 Page 14 18    The switch point  SP  must be between the upper range limit  URL  and the reset point   The reset point  rP  must be between the lower range limit  LRL  and the switch point  The  minimum distance between switch point and reset point  minimal hysteresis  is 0 5  of the  measuring range     0 5  of MR Setting range for  Setting range for     See  the reset point  LRL ic ais p    Figure 13  Setting ranges for switch point and reset point    You can define delays for the switch point as well as the reset point  e g  to avoid that  short pressure peaks trigger the switch          switch point  SP     fF         f       4        n iMn    reset point  rP    f  4   f                                              delay delay  switch  ds  reset  dr         _ ea                0  Figure 14  Output delays for a hysteresis function normally open  Hno   When you select a frame function  the menu items for switch point and reset point are    replaced by the upper and lower frame limits  The minimum difference of the frame limits is  also 0 5  of the measuring range                                         Display Designation Description  I     Frame high Upper frame limit in the selected measuring  iol l unit    i     Frame low Lower frame limit in the selected measuring  IL l unit  T m j Delay switch Output delay when entering the frame  aTa   Delay reset Output delay when leaving the frame            Output function Configuring the output  normally open    LiLiL   normally 
4. BTA 060  Rev 1K1 Page 18 18    
5. Labom Operating Instructions for pressure transmitter  IE PASCAL CS  Type Series CS2100 and CS2110       GMT  A       Table of contents    1    2  3    4    5  6    General Information asi catenins oc seen lactate neler mien eaedaeletnet dv ensleiesene 2  1 1 Intended USG srenencierenenonenonenone nenn ee re eee eee eee eer 2  1 2 General Safety NGI SG secs cesses citecieccs cece aeesaiaisasaisneecccisendscudeneueaedseameaeniaueuuseedsegeae 2  Ue EIA RING cans tee aoe ee eee 2   Transportation and Storage cis oosccctds cucesueccwsstatecwestusaouecdude ceceduiscueaiutecnesiadasveduddedesssuvejane 2   Installation and COMmMiISSIOnINng sicececncitetecscacetecnntt tieerietseecsGebicemieebieeite ee reiteteehoiaemase 2  3 1 Electrical Connection scsi siccteiee decease aegensecus nc cies esgedeccassaweaeusaeus senses vee wesnsscones 3  3 2  Adjusting the display UMit tata ecclesia eae Ue eile dee oe da ee a ee eee 4  3 3 Devices with flush mounted diaphragm              cceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeees 4   OS ON acne ee cetera E A E E E 5  4 1 Setup   Parameterization           cccccccecceeeecccccccceeeeeeeeeeeeecaeeeeeeeeeeeeseeseaaaeeeeeeeeeeeaeaes 5  4 2 Maintenance Sasser arth itenis ieccitantalan shanti atienbhaniaenibbeaipierbleapiantdensderiemiauroucucs 5   REMOVA ca tii le eae ceed gets 5   User Man  al  seisine ea eE a EE AE EE DE 6  6 1     System operating principles celal eseis neces eve een de eed ee eines eee een 6  62 Main MENU senisesse siesau kaia 
6. NE S  This is described in a separate  manual     4 2 Maintenance    When properly installed in accordance with applicable specifications  this device is  maintenance free  However  we recommend an annual recalibration of the device     The device contains no user serviceable or user replaceable parts or components     Before attempting to remove the device  be sure to first relieve the process system  pressure  Shut off the pressure source for all feed lines and relieve the pressure in them     Attempting to remove the device from a still pressurized system may result in  severe injury     Be sure to de energize the power supply to the device before disconnecting the electrical  connections  Once this is done  the device may be mechanically removed     Be sure that residue in the process system and in the device itself do not present a danger  to humans or the environment  During dismantling and removal  the device must be easy  to remove from the process system  Do not apply any force during removal work  After the  device has been removed  seal off the measuring point and mark the open process  accordingly     Removed devices may contain hazardous deposits and residue  When removing  or transporting these devices  be sure to take appropriate safety precautions        BTA 060  Rev 1K1 Page 5 18    6 User Manual    This chapter describes the handling and parameterization of the device with the three  buttons on the display head     You find an overview of the menu tree on th
7. aieiaa aikana eTa 10  6 3 Basic menu  PASE  peepee een me eee eee ee eee eee 10  64     Display menu  ATP  isss eena a uteseats EEE AREA E 13  6 5  Switch point Menu 2 ccrcncrescaccuare ceseuats cxaueta tiene cesenaancusceteddagustancvence cueedte mundane  14  Me SU MENU ssas a aaa ees 17  6 7 Overview of the menu tree    ceeeecececceceeeeeeeeeeeeeaaeeeeeeeeeeeeeeecaaeeeeeeeeeeneesaaeees 18         Do you have questions about this device  Call our hotline at   49  0  4408 804 444 BTA 060    LABOM Mess  und Regeltechnik GmbH   Im Gewerbepark 13   27798 Hude   Germany Rev 1K1  Tel    49  0  4408 804 0   Fax   49  0  4408 804 100   e mail  info labom com   www labom com Page 1 18    1 General Information    These operating instructions contain information necessary for the proper installation and  use of this device  In addition to these instructions  be sure to observe all statutory  requirements  applicable standards  the additional technical specifications on the  accompanying data sheet  see www labom com  as well as the specifications indicated on  the type plate     1 1 Intended Use    The pressure transmitter PASCAL CS21xx is intended for measuring the relative and  absolute pressure of gases  vapors and liquids as specified in the data sheet  For a correct  function the permissible overload pressure indicated on the type plate must not be  exceeded     1 2 General Safety Notes    The installation  set up  service or removal of this device must only be done by trained   q
8. ange     0 bar   i m Min  pointer  low  Display resp  delete the min  pointer  i Max  pointer  high  Display resp  delete the max  pointer  IIl ae     pm L       Return Return to the main menu       Table 4  The items in the basic menu       BTA 060  Rev 1K1    Page 10 18          6 3 1 Setting the measuring unit  unit   The device can operate with the units shown below  The selected unit applies to data  entries  e g  for set points  and to the displaying of numerical values  e g  the min  max   pointer                              Display Unit Display Unit  bir     EPA e  no Hi   mbar eA mPa  PS   aA         7 m C7 Return             Table 5  Parameter list for the measuring unit    As an example the steps needed to change the unit from bar to PSI are shown below        Figure 10  Operator actions for changing the unit from bar to PSI    6 3 2 Correcting the Zero Point  tlr     You can correct the zero point by up to 20  of the measuring range with the  teach lower  range  menu item  t1r   When the menu item is selected the entry points to   rEt    To  trigger the function go to  YES  with one of the arrow buttons and confirm with the middle  button  This extra step prevents any unintentional zero shift while navigating the menu     With the final confirmation  the applied pressure is stored as zero point   donE  appears on  the display to confirm that the zero point has been adjusted and the device switches back  to the menu item  Teach Lower Range    t1r               Dis
9. been reset to factory settings  and the device switches back to the menu item  Reset   rES                  Display Designation Description    EL        Return Return to  rEs   Pe  L    L Confirm  yes  Resetting the device to factory settings                Table 16  Parameter list for resetting the device to factory settings       BTA 060  Rev 1K1    Page 17 18    6 7 Overview of the menu tree    Main menu Sub menu Description    bASE Menu with basic functions    Setting the measuring unit  bar  mbar  PSI     kPa   MPa  mA     Setting the lower range  4 mA  to the applied pressure    Display resp  delete the min  pointer    Hy Display resp  delete the max  pointer    All settings relating to the display  Setting the display period for the unit  Setting the display period for the measured value    Display direction  0    normal  180     turned     Setting the decimal places  zero to three fixed decimal  places or automatic        Configuring the switching outputs  only if available     Switch point or upper frame limit of  the first switching output    Reset point or lower frame limit of the  first switching output    Output delay at the switch point of the  first switching output    Output delay at the reset point of the  first switching output    Output function of the first switching  output  Hno  Hnc  Fno  Fnc     Menu items for the second switching output    Ss  System information and reset    Hard  and software versions  serial number    Reset to factory settings       
10. closed  hysteresis   frame    Menu items for second switch point      C L       Return Return to  sp        Table 12  Menu items for a switching output with frame function       BTA 060  Rev 1K1 Page 15 18    6 5 1 Configuring the output function  out 1 2     You can choose a hysteresis or frame function as the output function  Furthermore you  can define whether the output is normally open or normally closed                                Display Designation Description  tt Hysteresis  If the pressure is above the switch point the  rice normally open switch is closed  At the lower range limit the    switch is open   tt Hysteresis  If the pressure is above the switch point the  rt  normally closed switch is open  At the lower range limit the    switch is closed   L    Frame  Inside of the frame the switch is closed  At  IIIJ normally open the lower range limit the switch is open   L        Frame  Inside of the frame the switch is open  At the  a normally closed lower range limit the switch is closed      I _  Return Return to  out 1 or out 2   Pee             Table 13  Parameter list for output function                switch point  SP  or  upper frame limit  FH     reset point  RP  or  lower frame limit  FL     Hysteresis   normally open  Hno     Frame   normally open  Fno     Frame   normally closed  Fnc        Figure 15  The output functions       BTA 060  Rev 1K1    Page 16 18       Hysteresis   normally closed  Hnc     6 6 System menu    The system menu contains the follo
11. e last page of this document     6 1 System operating principles    6 1 1 System feedback to operator when buttons are pressed    When you press a button  the switching output LEDs flash acknowledging the pressed  button  The left and right arrow buttons are indicated by flashing the left or the right LED   When you press the left and right arrow buttons at the same time  both LEDs will flash   Both LEDs flash rapidly if you press the middle button                    Button Feedback   Q Left arrow button Left LED flashes   O Right arrow button Right LED flashes  O   Eo Arow buttons at the   Both LEDs flash      Middle button Both LEDs flash rapidly                   Table 2  Feedback to operator when buttons are pressed    The switching outputs are not affected by the LED flashing  When there is no button  pressed  the output states are displayed     6 1 2 Display mode   Measured value screen    When the device is switched on  it goes into display mode  The currently measured value  is displayed  or it is displayed alternately with the unit  see 6 4 1      If the measured value is greater than the maximum number that can be displayed by the  system     this can be caused by setting a fixed decimal point  see Sec  6 4 3       the  maximum number that can be displayed will flash     By pressing the middle button  the selected unit will be displayed  The unit will continue to  be displayed as long as the middle button is pressed     The arrow buttons have no function in display 
12. e ranges of  lt  2 bar  changes in the installation  position will also cause the zero point position to shift  This zero point offset can be  subsequently corrected by making an appropriate adjustment  see 6 3 2         BTA 060  Rev 1K1 Page 2 18    After the mechanical installation and the electrical connection are both complete  the  device is ready for use as soon as the voltage supply is switched on     3 1 Electrical Connection    Make all electrical connections with the voltage supply switched off     Permissible supply voltage Uy   14 30 VDC  Permissible load Rs    Uv   14 V  22 mA  4 pole connector 8 pole connector  for switching Explanations  outputs       plus side of supply  o     2  S1 S2_ gt      minus side of supply  o s foo    x  0      n c  not connected  i a o Pgm1        s 5          S1    o S1 S2 common pin of switching  6 4 outputs  see below   s1 2  o   Pgm2    7 a   S1 switching output 1  o        oe 1 ay   S2 switching output 2     O  Pgm1 2 programming pins       Figure 1  Pin assignment for M12 circular connector  device side     3 1 1 Connecting the switching outputs  optional     The switching outputs are potential free   They are electrically isolated from the           supply side  see Figure 2   S1  Therefore you can connect the load on the  high side  PNP style  or the low side   NPN style  as long as you use only one S2    switching output               Wu       Figure 2  Switching outputs isolated from supply    Due to the limited number of p
13. essing both arrow  buttons at the same time  The return is indicated by a blinking   rEt    By pressing both  buttons for more than one second  you return to the display mode  Cancelling the menu  mode is indicated in the display by a blinking   ESc       escape      If no button is pressed for five minutes in menu mode  the device automatically switches  back to display mode          measured  value    upper menu       next menu  item    Figure 6  Button functions in menu mode       BTA 060  Rev 1K1 Page 7 18    An example of button functions available in menu mode is shown below        Figure 7  Button functions in menu mode  example     For the sake of simplicity  the return to the next higher menu and directly to display mode  will not be shown anymore     6 1 5 Setting values  There are two types of values that can be altered         values that can be selected from a predefined parameter list      numerical values    Selecting a value from a parameter list    Parameter lists     for example  the units list     behave like a menu  You can scroll through  the list in both directions with the arrow buttons  Each list contains the   rEt     item  which  allows you to return to the next higher menu level     The middle button stores your selection   Stor  appears on the display to confirm that the  value has been stored  and the device returns to the higher level menu item     You can cancel the selection by pressing both arrow buttons at the same time  The device  will then
14. ins either the low side or the high side is combined internally  and routed to Pin 5  Therefore you have to connect both loads as shown below if you want  to use both switching outputs        Figure 3  Connecting options with shared low side       BTA 060  Rev 1K1 Page 3 18       Figure 4  Connecting options with shared high side    Use an appropriate free wheeling diode  if you want to switch inductive loads   The default values fort he switching units are as follows  if not specified otherwise                             switching unit 1 switching unit 2  output function hysteresis  normally open hysteresis  normally open  switch point 40  of measuring range 80  of measuring range  reset point 20  of measuring range 60  of measuring range       Table 1  Default settings for switching outputs    3 2 Adjusting the display unit    You can turn the display unit approx  300  to optim ize  the readability  To do so hold the stainless steel    housing with one hand and turn the display unit with j  the other hand into the wanted position  bs 4 300  The turning angle is limited by an internal limit stop  Do LLM __    not try to force the display unit beyond that point  It  might get destroyed     3 3 Devices with flush mounted diaphragm  For type series CS2110 with flush mounted diaphragm observe the following hints     Do not remove the protective cap or protective wrapping from the diaphragm before  immediate installation to prevent soiling or damage     Do not touch the diaphrag
15. m as this might cause deformation or damage  Any deformation  may negatively influence the zero point or other characteristics of the device     If required you can find further information about diaphragm seals in the separate  operating instructions BTA 062        BTA 060  Rev 1K1 Page 4 18    4 Operation    During device operation  take care that the device remains within its intended temperature  range  that its maximum operating pressure is not exceeded and that it is not harmed by  the medium to which it is exposed  No other monitoring is necessary     Permissible ambient temperature   20    to  85     Permissible overload pressure  See type plate  depends on pressure range     The measured value is shown on the four digit display  The LEDs above the display  indicate the switch states of the switching outputs  The LEDs are on  when the switch is  active     4 1 Setup   Parameterization    You can set or change all adjustable parameters of the device at the device itself  This is  described in detail in the user manual  see below      The three buttons on the display module are capacitive  not mechanical  therefore they do  not move when pressed  Capacitive buttons sense the presence of your finger when  pressed  Withdraw the finger at least one centimeter after pressing a button  This helps  the device to clearly detect individual keypresses     If the device comes with an 8 pole connector  you can also configure it with a  communication modem and the PC Software COMLI
16. matic The decimal point is set so that the decimal  PILI cl places are fully used  IT   No decimal place  LILILILI  UW 7   One decimal place  LILI  CU 17    Two decimal places  LILI  CU 7    Three decimal places  a        L       Return Return to  dEcpP         Table 10  Parameter list for setting the decimal point    Please note that when the decimal point is fixed the measured value may not be displayed  if there are insufficient digits left of the decimal point  In this case the maximum number  that can be shown on the display will appear flashing  e g   99 99   when two decimal  places are set for a measured value of 110 mbar     6 5 Switch point menu    The switch point menu contains the functions for setting the first and second switch point   The menu items vary  depending on whether you select a hysteresis or frame function   Independently from the output function you can define switching delays                                         Display Designation Description  m D   Switch point Switch point in the selected measuring unit  D H Reset point Reset point in the selected unit  Je      Delay switch Output delay at the switch point  Lt l l  mr    Delay reset Output delay at the reset point             Output function Configuring the output  normally open    LiLiL   normally closed  hysteresis   frame   Menu items for second switch point        L       Return Return to  sp        Table 11  Menu items for a switching output with hysteresis function       BTA 060  Rev 1K
17. mode     6 1 3 Activating the Menu mode   Key lock    A key lock prevents an unintentional misconfiguration of the device  You have to press  both arrow buttons simultaneously for at least two seconds to enter the operator menu   The first entry of the main menu  bASE  will then appear on the display  If you hold both  buttons for more than four seconds  the device switches back to display mode and shows  the currently measured value again        BTA 060  Rev 1K1 Page 6 18          measured  value    measuring  unit            operator  menu    Figure 5  Button functions in display mode  with example    6 1 4 Menu mode   Operator menu    When you enter the menu mode  you always begin with the first main menu item  bASE  In  menu mode you can navigate the menu with the arrow buttons  The middle button selects  the menu item resp  enters the submenu  If a value is just displayed  e g  the maximum  pointer  you can also return back to the menu item with the middle button     The menu item   ret    return   which allows you to go back to the next highest menu  level  is available in every menu  When you are in the main menu    rEt     returns you to  display mode     At the end of a menu  typically  the   rEt   item  you return to the first menu item by  repressing the down arrow button again  Similarly  you can jump from the first menu item  to the end of the menu or a value list with the up arrow button     You can return to the next higher menu level from every menu item by pr
18. n the device i s  PL installed upside down  I         i   Decimal places Setting the decimal places  zero to three  LIL ti fixed decimal places or automatic      pm L       Return Return to  disp        Table 8  The items in the display menu    6 4 1 Display period for measured value unit  td   tu     The unit can be displayed in two ways  either by pressing the middle button in display  mode or alternately with the measured value  When displayed alternately with the  measured value the display periods for the measured value and the unit can be selected  independently of one another     If the period for displaying the unit is set to zero  only the measured value will be  displayed   6 4 2 Rotating the display by 180   rot     You can rotate the 7 segment display by 180   so that it can be read when the device is  put in place upside down  The function for the arrow buttons are also swapped in this  case  so that the device can be operated the same way in either position                             Display Designation Description  mo Standard  09   ITE   Upside down  1809 Display rotated 180  for upside down  LILI operation  E C L       Return Return to  rot        Table 9  Parameter list for rotating the display       BTA 060  Rev 1K1 Page 13 18    6 4 3 Decimal point setting  decP     You can set a fixed decimal point or allow the system to compute the best position for the  decimal point                                      Display Designation Description   I I_       Auto
19. play Designation Description     CL _   Retum Return to  t1r   Pe  Li L Confirm  yes  Setting the applied pressure as the zero point    l  0 bar                    Table 6  Parameter list for correcting the zero point    The steps needed to adjust the zero point are shown below  starting from display mode         BTA 060  Rev 1K1 Page 11 18       Figure 11  Operator actions for correcting the zero point    6 3 3 Min  and max  pointers   Lo   Hi   The device has min  max  pointers for minimum and maximum pressure values  You can  display and reset them in this menu  Resetting a pointer is confirmed by showing            on the display                             Display Designation Description  4 Z1 tJ   Value of min  max Value of min  max  pointer in the selected  lIi l   pointer measuring unit  7 i 7 Clear Reset the stored pointer value      C i   Return Return to  Lo  or  Hi        Table 7  Parameter list for min  max  pointer    The steps needed to reset the minimum pointer are shown below        Figure 12  Operator actions to reset the minimum pointer       BTA 060  Rev 1K1 Page 12 18    6 4 Display menu  di SP     The display menu for configuring the display contains the following items                                   Display Designation Description  l l Display period for Can be set between 0 5 and 99 9 s  a a measured value  time  tgs data     Display period for unit   Can be set between 0 0 and 99 9 s  l  L  time unit   eens ae Rotate 180   Rotate screen by 180  whe
20. ualified personnel using suitable equipment and authorized to do so by the plant  operator     Improper installation or use of these devices or the use of damaged or defective  devices may result in severe injury or property damage     1 3 CE Marking    The CE marking on these devices certifies their compliance with the applicable EU  Directives for placing products on the market within the European Union     The following guideline applies to model series PASCAL CS21xx   EMC Guideline EMC 2004 108 EG    2 Transportation and Storage    Store and transport these devices only under clean  dry conditions  Avoid exposure to  shocks and excessive vibrations     Permissible storage temperature   40   85       3 Installation and Commissioning    Before installing the device  be sure that the device is suitable for the intended process  application with respect to pressure range  resistance to overpressure  compatibility with  the monitored medium  thermal stability and pressure port fitting type  The used gaskets  must be compatible with the used process connection and resistant to the medium     Complete the mechanical installation before making the electrical connections   Before placing the device in service  check it carefully for leaks under pressure     When checking the operation of the device  be sure to check the zero point signal with  respect to the installed position  In the standard design  the transmitter is preset at the  factory for vertical installation  For pressur
21. wing items                       Display Designation Description  a  ee Information Display of Hardware and software version   UU L serial number  S G E Reset Reset to factory settings  E     C L       Return Return to  sys              Table 14  The items in the system menu    6 6 1 Information  inFo   The following device information is available in the system menu                                Display Designation Display Designation  III    JU   Hardware version 1 ee l Serial number 1  I LILI I  Hwi  Itt I  Sn 1    III  I_   Hardware version 2    oI Serial number 2  PILI    HW2  JMI L  Sn 2      1         Software version 1    oF Serial number 3  _ILILI       SW  JAI J  Sn 3      I   t    Software version 2   a A   Serial number 4  JILILIL    SW2   lll l  Sn 4          L       Return Return to  inFo              Table 15  Menu items in the information menu    Due to the limited number of alphanumerical segments on the display  hardware and  software versions are split into two separate items  and serial number into four items     6 6 2 Reset to factory settings  x        ES     You can reset the device to the configuration as delivered with the menu item  Reset    rES   When the menu item is selected the entry points to   rEt       To trigger the function  go to  YES  with one of the arrow buttons and confirm with the middle button  This extra  step prevents any unintentional reset while navigating the menu      donE  appears on the display to confirm that the device has 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
e_ap03g quick guide_001-SP.p65  ThinClient ソリューションカタログ  Lenmar PowerPort Kit  Certification Report  Advanced clean-boot troubleshooting in Windows XP    eBOX730-860-FL A1 User Manual  基本 CMYK    ebode PowerMid Premium    Copyright © All rights reserved. 
   Failed to retrieve file