Home
SERVO-i Adult Ventilator Data Sheet
Contents
1. Critical Care TECHNICAL SPECIFICATIONS SERVO i ADULT SERVO i CO Analyzer optional Standard Size E COs Analyzer Module E Sensor Weight mE COs Analyzer Module E Sensor E Airway adapter Connectors and cables E COs Analyzer Module E Sensor Power source E CO Analyzer Module supply voltage E Power consumption E Sensor EN 864 ISO 9918 IEC EN 60601 1 Type BF defibrillator proof 154 x 90 x 43 mm 32 0 x 42 4 x 21 6 mm 0 45 kg 18g 10g 15 pole D sub female connector 20 pole 2 4 m cable Powered from the SERVO i lt 8 Wat 12 V during warm up lt 6 5 W at 12 V during normal operation Powered from the CO Analyzer Module SERVO i CO Analyzer Performance Measuring method Parameters Measuring range Step response time Warm up time Oxygen concentration compensation Barometric pressure compensation Digitizing rate Airway adapter dead space Mainstream dual wavelength non dispersive infrared Capnogram CO End tidal concentration etCO CO Minute elimination VCO Tidal CO elimination VTCO 0 to 100 mm Hg CO partial pressure 0 to 13 3 kPa CO partial pressure 0 to 13 2 CO volume at a barometric pressure of 1013 hPa lt 25 ms 10 to 90 step response 30 s to initial CO indication max 5 min to full specification Automatic Values supplied from the SERVO i Ventilator System Automatic Val
2. Invasive ventilation controlled modes only Mean peak pressure 10 cmH 0 or at least 35 cmH 0O Expiratory minute volume 50 Expiratory minute volume 50 Breathing frequency 40 Breathing frequency 40 Mean end expiratory pressure 5 cmH 0 Mean end expiratory pressure 3cmH O End tidal carbon dioxide concentration 25 End tidal carbon dioxide concentration 25 TECHNICAL SPECIFICATIONS SERVO i ADULT Ventilation Modes Invasive ventilation Controlled ventilation PC E VC E PRVC Supported ventilation E VS PS CPAP Combined ventilation m SIMV VC PS SIMV PC PS E SIMV PRVC PS E Bi Vent Automode Optional Can be configured with alternative flow patterns VC with flow adaptation VC without flow adaptation VC with decelerating flow Optional Optional Comes with the corresponding controlled ventilation mode Optional Comes with the corresponding controlled ventilation mode Optional Pressure controlled ventilation on two independently adjustable levels allowing unrestricted spontaneous breathing on both levels Optional Control mode VC lt gt Support mode VS Control mode PC lt gt Support mode PS Control mode PRVC lt gt Support mode VS Optional Critical Care SERVO i ADULT 7 Ventilation modes Non invasive Ventilation optional NIV PC NIV PS Ventilation modes NA
3. H 295 mm cart a table rail or pole 15 E Patient Unit W 300 x D 205 x H 415 mm 30 mm diameter ae Unit 15 kg User interface 5 Screen Unit 15 kg User Interface 5 kg Method of triggering Flow pressure and Edi Type TFT LCD module optional Size 31 cm 12 1 diagonal Max operating pressure Approximately 115 cmH O Viewing area 246 0 x 184 5 mm Bias flow 2 l min Inspiratory channel Gas supply Pressure drop Max 6 cmH20 at a flow of 1 I s Inlet gas pressure Air Oo 200 650 kPa 2 0 6 5bar Internal compressible factor Max 0 1 ml emH 0 29 94 PSI Gas delivery system Microprocessor controlled valve Connection standards available AGA DISS NIST or French Inspiratory flow range 0 to 3 3 I s standard Unavailable gas loss of gas The flow from an unavailable pressure gas air or O is automatically Expiratory channel compensated for so that the Pressure drop Max 3 cmH20 at a flow of 1 I s patient gets the preset volume Internal compressible factor Max 0 1 ml cemH 0 ave PEEP regulation Microprocessor controlled valve Rise time expiratory flow lt 12 ms for 10 90 response measurement at flow of 0 05 3 2 I s Conical fittings Male 22 mm female 15 mm Expiratory flow range 0 to 3 2 I s In accordance with ISO 5356 1 Gas exhaust port Male 30 mm cone 6 seRvo i ADULT Critical Care TECHNICAL SPECIFICATIONS SERVO i ADULT Airway pressure upper E Invasive ventilation E Non Invasi
4. AVA level cmH20 uV 0 15 Edi trigger sensitivity uV 0 1 2 0 Tpause S 0 1 5 Tpause of breath cycle time 0 30 Flow trigger sensitivity level 0 100 fraction of bias flow Press trigg sensitivity cmH O 20 0 Insp rise time of breath 0 20 cycle time Insp rise time s 0 0 4 Insp cycle off of peak flow 1 70 Insp cycle off in NIV of peak 10 70 flow Nebulizer time min 5 30 Backup settings Setting range Parameter Inspiratory tidal volume ml 100 2000 PC above PEEP cmH 0 5 120 PEEP PC above PEEP in NIV 5 32 PEEP CMV frequency breaths min 4 100 I E ratio 1 10 4 1 Tinsp s 0 1 5 10 seRvo i ADULT Critical Care TECHNICAL SPECIFICATIONS SERVO i ADULT Parameter Setting range Oxygen breaths 100 for 1 minute Start breath Initiation of 1 breath In SIMV mode initiation of 1 mandatory breath Insp or exp 0 30 seconds 2 minute silence and reset of latched alarms On Off Automode On Off Pause hold Alarm silence reset Compliance compensation Automode optional with Universal Basic Edition Nebulizer optional Backup ventilation Suction Support 5 30 min Continuous Off Backup On Off Pre oxygenation time Max 2 min Post oxygenation time Max 1 min Suction phase time No maximum level Adjustable oxygen level 21 100 Saving of data Recording of current waveform 20 seconds of data will be and
5. ET must be signed see conditions in the SERVO i MR declaration Order no 66 71 670 and the Interhospital Transport Declaration Order no 66 64 721 respectively Critical Care SERVO i ADULT 3 UBE UEE Stress Index x lt NIV NAVA NAVA Heliox Nasal CPAP X lt NIV PC NIV PS Bi Vent Y Sensor Measuring CO Analyzer Nebulizer Alarm output connector Open Lung Tool E Automode E SIMV PRVC PS B PRVC VS SIMV VC PS E E E VC E E E SIMV PC PS o E E PC fal E E PS CPAP E E E All patient category software E o KEY TO ABBREVIATIONS NAVA Neurally Adjusted Ventilatory Assist NIV Non invasive ventilation SIMV Synchronized Intermittent Mandatory Ventilation PRVC Pressure Regulated Volume Control VS Volume Support VC Volume Control PS Pressure Support PC Pressure Control CPAP Continuous Positive Airway Pressure 4 l SERVO i ADULT Critical Care TECHNICAL SPECIFICATIONS SERVO i ADULT The system General CE The device complies with Onis requirements of Medical Device Directive 93 42 EEC Classification Class equipment According to IEC EN 60 601 1 Standards IEC EN 60 601 1 Type B IEC EN 60 601 2 12 EN 794 1 Noise level lt 50 dBA measured at 0 3 m distance IP classification IP 20 Electromagnetic compatibility EMC Emission According to IEC EN 60601 1 2 Edition 2 2001 Immunity Extended test to According to IEC EN 60601 1 2 30 v m T
6. VA optional NAVA NIV NAVA Neurally Adjusted Ventilatory Assist via endotracheal tube or tracheostomy Neurally Adjusted Ventilatory Assist via non invasive patient interfaces Waveform and loop presentations Real time waveforms up to 4 waveforms can be displayed simultaneously E Pressure E Flow E Volume E CO mE Edi Loops E Volume Pressure E Flow Volume Requires SERVO i CO Analyzer option Requires SERVO i NAVA option A reference loop and three overlaying loops can be displayed Displayed simultaneously with Open Lung Tool graphical trends if requested 8 SERVO i ADULT Critical Care TECHNICAL SPECIFICATIONS SERVO i ADULT DIRENE Trended Displayed Trended Monitoring value value Monitoring value value End Expiratory Flow Ye Yi Breathing frequency Yes Yes T a at ow Mis Mii Spontaneous breaths per No Yes sae aul lk S gz minute RRsp Dynamic Compliance Yes Yes Peak Airway Pressure Yes Yes Siess iniex Yes Yes Mean Airway Pressure Yes Yes Inspiratory Resistance Yes Yes Pause Airway Pressure Yes Yes Expiratory Resistance Yes Yes End Expiratory Pressure Yes Yes Elastance Yes Yes CPAP Pressure Yes Yes Time Constant Yes No Inspired Tidal Volume Yes Yes eo Yes Yes Expired Tidal Volume Yes Yes Work of Breathing patient Yes Yes Inspired Minute Volume Yes Yes Work of Breathing ventilator Yes Yes Expired Minute Volume Yes Yes Shallow Breathing Index SBI Yes Yes Le
7. VENTILATION SERVO i ADULT MAQUET Seo 5 HIGHLIGHTS m For adult and pediatric patients m Intuitive user interface m Designed for cost efficiency m Flexible placement m Modular interchangeable plug in units m Enables Heliox delivery MAQUET GETINGE GROUP CRITICAL CARE NAVA improved synchrony and unique monitoring capabilities m For invasive as well as non invasive ventilation m The Open Lung Tool and Stress Index tools for lung protection m Available in MR and Transport editions for continuity of ventilatory care SERVO i VENTILATOR FAMILY SERVO i ADULT SERVO i Configurations The SERVO i Ventilator family consists of four configurations UBE UEE E x Infant Adult Universal Basic Edition Universal Extended Edition Standard Configuration Options Not applicable The SERVO i Ventilator family addresses the very different requirements of neonatal pediatric and adult needs from a single ventilation platform All four configurations are the same ventilator equipped with different functions They can be customized and upgraded with different options for future needs GERVO i Adult and SERVO i Infant can also be upgraded to SERVO i Universal The same ventilator can be used at the bedside during transport and in the MR room facilitating training operation and maintenance increasing efficiency and flexibility An agreement with MAQU
8. akage fraction in NIV Yes Yes Supply pressure Oz and air Yes No Ti Ttot Yes No Battery remaining time Yes No E ratio Yes No Barometric pressure Yes No Total PEEP Yes No Edi min Yes Yes Event log Alarms O2 Concentration measured Yes Yes Ventilator settings CO End tidal concentration Yes Yes Apnea periods etCO Immediate functions CO Minute elimination CO Yes Yes Service log eS ore ERT est results Tidal CO elimination VTCO gt Yes Yes Breventive MalnteHance MVe sp MVe Yes No Service history Spontaneous Exp Minute Yes Yes Configuration log Volume MV sp Stored trend values for up to 24 hours TECHNICAL SPECIFICATIONS SERVO i ADULT Parameter settings Setting range Parameter Inspiratory tidal volume ml 100 2000 Inspiratory minute volume 0 5 60 l min Apnea time to alarm s 15 45 Automode Trigger timeout s 7 12 PC PS above PEEP cmH20 0 120 PEEP PC PS above PEEP in NIV 0 32 PEEP cmH30 PEEP cmH30 0 50 PEEP in NIV cmH20 2 20 CPAP pressure cmH30 CMV frequency breaths min 4 100 SIMV frequency breaths min 1 60 Breath cycle time SIMV s alo Prigh cmH20 PEEP 1 50 Thigh S 0 2 10 TpEEP s 0 2 10 PS above Pyigh cmH20 0 120 Prigh O concentration 21 100 I E ratio 1 10 4 1 Critical Care SERVO i ADULT 9 Parameter settings Setting range Parameter Tinsp 9 0 1 5 N
9. he EMC Declaration Information to the Responsible Organization is available from MAQUET Patient range all ventilation modes Operating conditions Operating temperature 10 to 40 C Relative humidity 15 to 95 non condensing Atmospheric pressure 660 to 1060 hPa Lowest pressure in breathing 400 cmH jO system Weight 10 250 kg Non operating conditions Impact Peak acceleration 15 g Pulse duration 6 ms Number of impacts 1000 Storage temperature 25 to 60 C 13 to 140 F Storage Relative Humidity lt 95 Storage Atmospheric Pressure 470 to 1060 hPa Power supply Power supply automatic range 100 120 V AC 10 selection 50 60 Hz or 220 240 V AC 10 50 60 Hz Plug in battery module E Battery backup Two battery modules are delivered with the ventilator Up to six battery modules can be included Rechargable 12 V 3 5 Ah each Approximately 3 h battery At least 3 h when using six batteries Battery capacity E Recharge time E Battery backup time External 12 V DC Max power consumption 12 0 V 15 0 V DC 10 A At 100 120 V 2 A 190 VA 140 W At 220 240 V 1 A 190 VA 140 W Critical Care SERVO i ADULT l 5 TECHNICAL SPECIFICATIONS SERVO i ADULT The ventilator General User interface Dimensions See dimensional drawings Weight Approximately 5 kg page 15 Attachment Can be attached to the mobile E User Interface W 355 x D 53 x
10. heal tube 8 mm inner diameter Min 0 1 ml water min at 0 1 gas flow s Min 0 3 ml water min at 0 5 gas flow s Min 0 5 ml water min at 1 0 gas flow s Sterile water 55 C 131 F Max 10 ml Max 50 dBA measured at 0 3 m distance Aeroneb Nebulizer Systems optional Weight Aeroneb module Dimensions Aeroneb module Connection cable length Particle size water Flow rate Volume nebulizer unit Residual volume Approx 216 g H 105 mm x L 108 mm x W 60 mm 2 0m 1 5 mass median aerodynamic diameter MMAD gt 2 ml min m Pro 10 ml m Solo 6 ml 10 14 seRvo i ADULT Critical Care TECHNICAL SPECIFICATIONS SERVO i ADULT SERVO i Mobile Cart optional SERVO i IV Pole optional Weight 20 kg Max load total Dimensions H 1015 mm x L 640 mm x W 560 mm z see dimensional drawing Gas trolley optional below Max load 2x10 kg bottles Docking Dockable to SERVO i Mobile 5 z Cart SERVO i Drawer kit optional Dockable to a separate wall Weight 4 5 kg clamp Dimensions H 240 mm x L 210 mm x W 300 mm Compressor Mini optional See separate data sheet SERVO i Holder optional Weight 3 5 kg Dimensions H 352 mm x L 247 mm x Service W 159 mm Regular maintenance Once every 12 months or at see dimensional drawing least after 5000 operating hours below SERVO i Shelf Base optional For inaccuracies and more detailed technical specifications
11. nation Requires SERVO i CO Analyzer option Up to 21 600 breaths A cursor can be moved with the main rotary dial or via the touch screen When moving it along the graph the numeric parameter values valid for that actual moment will be shown Time resolution of the x axis can be selected in five different steps Hours and minutes when values are measured Critical Care SERVO i ADULT 11 Stress Index optional Presented values Stored values Cursor function Time marking m SI Stress Index a PEEP Positive End Expiratory pressure m VT Expiratory Tidal volume m 20 values stored in graph E Stored trend values up to 24 hours A cursor can be moved with the main rotary dial When moving it along the graph the numeric parameter values valid for that actual moment will be shown Hours and minutes when values are measured NAVA optional Standards Size E Edi module E Edi catheter cable Weight Edi module Power source Edi module supply voltage Parameters Edi catheters IEC EN 60601 1 Type CF defibrillator proof 154 x 90 x 21 mm Length 2 m 0 25 kg Powered from SERVO i lt 3 W at 12 V normal operation Edi waveform Edi leads waveforms NAVA estimated pressure waveform Pest 6 Fr length 49 cm 6 Fr length 50 cm 8 Fr length 100 cm 8 Fr length 125 cm 12 Fr length 125 cm 16 Fr length 125 cm 12 seRvo i ADULT
12. on prevention within life sciences MAQUET specializes in solutions therapies and products for surgical interventions interventional cardiology and intensive care MCV00008324 REVA
13. parameter values recorded 10 seconds before and 10 seconds after activation Communication Interface Serial port Second serial port optional RS 232C isolated For data communication via the Communication Interface Emulator CIE See information above Alarm output connector optional Connector Ratings Network connection optional Data transfer optional Screen picture transfer optional 4 pole Modular connector Max 40 V DC Max 500 mA Max 20 W MIB Medical Information Bus monitor connection Via Ventilation Record Card Via Ventilation Record Card Non invasive Ventilation optional Max leakage compensation level E Adult Leakage overrange detection Disconnect detection Disconnect flow E Low mE High E Disabled Connect detection 65 l min Automatic Automatic Configurable 7 5 l min 40 l min Deactivates disconnect detection Manual or automatic via bias flow TECHNICAL SPECIFICATIONS SERVO i ADULT Open Lung Tool OLT optional with Universal Basic Edition Three simultaneous graphical trends presented breath by breath Values stored breath by breath Cursor function Zoom function Time marking 1 EIP and PEEP End Inspiratory Pressure and Positive End Expiratory pressure 2 VT and VT Inspiratory and Expiratory Tidal volume 3 C dyn i and VTCO Dynamic inspiratory Compliance VTi EIP PEEP and Tidal CO elimi
14. please Weight 1 2kg refer to the User s manual Dimensions H 29 mm x L 205 mm x W 159 Hear ESEI ORDERING INFORMATION see dimensional drawing below SERVO i Ventilator and accessories See separate information SERVO i Version 6 1 System Flow Chart Order no 5 66 70 102 Gas cylinder restrainer optional Max load 2 x 5 liter bottles Critical Care SERVO i ADULT 15 TECHNICAL SPECIFICATIONS SERVO i ADULT Dimensional drawings SERVO i on Mobile Cart SERVO i Holder E A A A E z E ite Q amp Dv mM 10 gt Y 415 mm 438 mm SERVO i Shelf Base MAQUET GETINGE GROUP MAQUET Medical Systems USA 45 Barbour Pond Drive Wayne NJ 07470 www maquetusa com ca maquet com Maquet Critical Care AB 2011 All rights reserved e MAQUET reserves the right to modify the design and specifications contained herein without prior notice Order No 66 70 107 Printed in Sweden 111116 Rev 12 English e Data Sheet GETINGE GROUP is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences We operate under the three brands of ArjoHuntleigh GETINGE and MAQUET ArjoHuntleigh focuses on patient mobility and wound management solutions GETINGE provides solutions for infection control within healthcare and contaminati
15. ues supplied from the SERVO i Ventilator System 87 Hz lt 5 ml TECHNICAL SPECIFICATIONS SERVO i ADULT Y Sensor measuring optional Size E Y Sensor Module E Y sensor adult Weight E Y Sensor Module E Y sensor adult Sensor material Tubing Power source Y Sensor Module supply voltage Y Sensor measuring Performance Measuring method Parameters Measuring range Airway adapter dead space 154 x 90 x 43 mm Length 84 mm 0 4 kg 10 5g Makrolon polycarbonate 2 0 m Medical grade PVC Powered from the SERVO i lt 5 W at 12 V normal operation Fixed orifice differential pressure Airway pressure Airway flow Inspiratory and expiratory volumes 2 to 180 l min lt 9 0 ml Critical Care SERVO i ADULT 13 Servo Ultra Nebulizer optional Weight Dimensions Connection cable length Nebulizer T piece connections Internal volume Ultrasonic generator frequency Particle size water Output from nebulizer water minimum water flux Buffer liquid Max medication temperature Volume medication cup Noise level Approx 125 g H 105 mm x L 108 mm x W 60 mm 2 0m Inlet outlet 22 15 mm outer inner diameter and 22 mm inner diameter ISO standard Pediatric patient tubes Adapters 22 10 mm outer diameter and 15 10 mm outer diameter 60 ml 2 4 MHz Mass Median Diameter MMD approximately 4 0 um measured distally in endotrac
16. ve Ventilation Expired minute volume Upper alarm limit Expired Minute Volume Lower alarm limit 16 120 cmH 0 16 40 cmH O 0 5 60 I min 0 5 40 I min No patient effort Apnea alarm 15 45s Automatic return to support mode on patient triggering No consistent patient effort Respiratory frequency High end expiratory pressure Low end expiratory pressure High continuous pressure O concentration Gas supply Battery Yes described in User s manual 1 160 breaths min 0 55 cmH2O0 0 47 cmH 0 Note Setting the alarm to 0 zero is equal to alarm off Set PEEP level 15 cmH20 exceeded for more than 15 seconds Set value 5 vol or 18 vol Below 200 kPa 2 0 bar 29 PSI and above 650 kPa 6 5 bar 94 PSI Limited battery capacity 10 min No battery capacity less than 3 min Low battery voltage End tidal CO upper and lower 0 5 20 limit Leakage out of range in NIV Technical Autoset alarm limits specification m High airway pressure E Upper minute volume E Lower minute volume mE Upper respiratory frequency E Lower respiratory frequency mE High end expiratory pressure E Low end expiratory pressure E Upper end tidal carbon dioxide concentration etCO E Lower end tidal carbon dioxide concentration etCO 4 100 mm Hg 0 5 14 kPa Yes Described in the User s manual Yes Described in the User s manual