1. OHAUS Ranger Count 3000 Series Instruction Manual Ranger Count 3000 Series EN 1 1 INTRODUCTION This manual contains installation operation and maintenance instructions for the Ranger Count 3000 Series Please read the manual completely before using the scale 1 1 Safety Precautions Please follow these safety precautions e Verify that the AC input voltage printed on the data label matches the local AC power supply Do not drop loads on the platform Make sure that the power cord does not pose a potential obstacle or tripping hazard Use only approved accessories and peripherals Operate the scale only under ambient conditions specified in these instructions Disconnect the scale from the power supply when cleaning Do not operate the scale in hazardous or unstable environments Do not immerse the scale in water or other liquids Do not place the scale upside down on the platform e Only use weights within the scale s capacity as specified in these instructions e Service should be performed only by authorized personnel 2 INSTALLATION 2 1 Package Contents e Scale e Power Cord e Instruction Manual CD e Pan e Sub platform e Warranty Card 2 2 Installing Components Install the sub platform amp metal pan as shown below Press to lock the sub platform into place The scale can be operated without the pan if desired Figure 2 1 Installing sub platform and metal pan EN 2 Ranger Co
2. Gross Set the status OFF disabled ON the gross weight is printed Net Set the status OFF disabled ON the net weight is printed Tare Set the status OFF disabled ON the Tare weight is printed Header Set the status OFF disabled ON the Header is printed Note See section 7 3 1 for how to enter Header line Footer Set the status OFF disabled ON the Footer is printed Note See section 7 3 1 for how to enter Footer line Mode Set the status OFF disabled ON the Mode is printed Unit Set the status OFF disabled ON the Unit is printed Info Set the status OFF disabled ON the reference information is printed APW Set the status OFF disabled ON enabled Ranger Count 3000 Series EN 21 Accu Set the status OFF disabled RESULT the Accumulation result is printed ALL all the Accumulation data is printed Layout LRSOUE oet the format of the data output to a printer or computer Format Set the printing format MULTI amulti line single column style printout is generated SINGLE a single line printout is generated Feed Set the paper feed LINE move a paper up one line after printing 4LF move a paper up four lines after printing FORM a form feed is appended to the printout Data Transfer 4858 5 Output weighing results directly to a PC application OFF disabled ON enabled Windows XP Setup 1 Click Start Menu in Windows XP and click Settin
3. 1950 2275 2600 2925 3250 325 650 975 1300 1625 1950 2275 2600 2925 3250 3575 Elevation in feet O0 1060 2130 3200 4260 5330 6400 7460 8530 9600 10660 1060 2130 3200 4260 5330 6400 7460 8530 9600 10660 11730 GEO value 098 546 5 4 4 3 3 2 2 1 1 0 0 10 Heus me 3 5 s 7 8 8 T 7 5 2806 2928 11 10 10 9 9 8 8 7 7 6 6 2925 304 n 1 10 10 9 9 8 8 7 7 6 3041 3156 12 1 f 10 10 9 9 8 8 7 7 31556 3309 12 12 f 10 10 9 9 8 8 7 33909 34 13 12 12 n 1 10 10 9 9 8 8 3421 35 91 13 13 12 12 11 11 10 100 9 9 8 3591 36 2 14 13 13 12 12 11 10 10 9 9 36 1 3750 14 14 13 13 12 12 11 1 10 10 9 3750 3858 15 14 14 13 13 12 12 10 10 3858 4005 15 15 14 14 13 13 12 12 f 10 4005 4112 16 15 15 14 14 13 13 12 12 f maz 4278 16 16 15 15 14 14 13 13 12 12 4289 4326 17 16 16 15 15 14 14 13 13 12 12 4326 4432 17 17 16 16 15 15 14 14 13 13 12 4432 4538 18 17 17 16 16 15 15 14 14 13 13 4538
4. 46745 18 18 17 17 16 16 15 15 14 14 13 46 45 4751 19 18 18 17 17 16 16 15 15 14 14 47551 4858 19 19 18 18 17 17 16 16 15 15 14 4858 5006 20 19 19 18 18 17 17 16 16 15 15 50906 5174 20 20 19 19 18 18 17 17 16 16 15 5198 52722 21 20 20 19 19 18 18 17 17 16 16 52722 53 91 21 21 20 20 19 19 18 18 17 17 16 18 5704 5877 23 23 22 22 21 21 20 20 19 19 18 58 17 5932 24 23 23 22 22 21 21 20 20 19 19 5932 6048 24 24 23 23 22 22 21 21 20 20 19 60 49 62 90 25 24 24 23 23 22 22 21 21 20 2 6290 6330 25 25 24 24 23 23 22 22 21 21 20 63 30 6455 26 25 25 24 24 23 23 22 22 21 21 6455 6624 26 26 25 25 24 24 23 23 22 22 21 66724 6757 27 26 26 25 25 24 24 23 23 22 22 67 57 6935 27 27 26 26 25 25 24 24 23 23 22 69 95 7121 28 27 27 26 26 25 25 24 24 23 23 7121 7386 28 28 27 27 26 26 25 25 24 24 23 73986 7524 29 28 28 27 27 26 26 25 25 24 24 75724 7752 29 29 28 28 27 27 26 26 25 25 24 77 82 80756 30 29 29 28 28 27
5. Short Press Press less than 1 seconds Long Press Press and hold for more than 2 seconds Weight Piece Weight ZU ees ZU CID I 0 0 UO LU LUS Z U UU UID US MRS v v v Y V 5 Ea APW Lo Wt Lo Aut enu Mode 7 8 9 10 11 1213 U Clos Ooh W N k d ak k JI O10 N CO 0 QO Figure 3 2 LCD Displays TABLE 3 2 LCD Symbols Item Description st Item Description 6 Pointer symbols 15 Percent symbol not used 8 Calibration Menu Mode symbol 17 __ Pieces symbol tonne symbol not used _ Low Average Piece Weight Symbol 18 Scale symbol only shown when 2 platform is connected and turned on EN 6 Ranger Count 3000 Series The colored LED indicators on the left side of the control panel are used in Checkweigh mode section 3 7 and will light up according to the following rules 6 Red Loads Upper limit Green Loads 2 Lower limit and lt Upper limit X Yellow Loads gt Lower limit Power Input Weigh Below Access a 1 RS232 and Option Interface Port below cover I Lock Switch Figure 3 3 Below view of Ranger Count 300
6. the first stable gross weight is tared ON ACC stable gross loads within the accept limits are tared in Check weighing mode Beeper Opt bP BPt Set whether or not the beeper sounds when APW has been optimized OFF disabled ON enabled Beeper Signal bP 5 set how the beeper responds in Check Weighing mode OFF the beeper is disabled ACCEPT the beeper will sound when the weight is within the Accept range UNDER the beeper will sound when the weight is below the Under setting OVER UNDER the beeper will sound when the weight is above the Over setting OVER the beeper will sound when the weight is below the Under setting or above the Over setting Beeper Key bP Es Set whether the beeper sounds when a button is pressed OFF no sound ON sound Accumulation oet the accumulation functionality OFF accumulation disabled AUTO automatic accumulation MAN manual accumulation End Setup End Advance to the next menu or return to the top of the current menu Ranger Count 3000 Series EN 17 4 4 Readout Menu Enter this menu to set user preferences Reset no yes Stable Range 0 5 1 2 5 Filter low medium high Auto Zero Tracking off 0 5 1 3 Light off on auto Auto off off 1 5 10 End Headout Exit menu Reset ESEE Reset the Read menu to factory defaults NO not reset YES reset Stable Range SEAbLE oet the amount the reading can vary while the stability
7. 27 26 26 25 235 8056 85745 30 30 29 29 28 28 27 27 26 26 25 85 45 9000 31 30 30 29 29 28 28 27 27 26 26 EN 26 Ranger Count 3000 Series 5 LEGAL FOR TRADE When the scale is used in trade or a legally controlled application it must be set up verified and sealed in accordance with local weights and measures regulations It is the responsibility of the purchaser to ensure that all pertinent legal requirements are met The Menu Lock switch limits changes to the Cal Setup Readout Unit and Print menus The switch in type approved models may set some scale settings as required by the approval agency The switch may be secured using paper seals wire seals or plastic ties Note When LEGAL FOR TRADE is set to ON LFT Switch in locked position the menu settings are affected as follows Calibration menu is not accessible Dual Scale Capacity 2 Graduation 2 and Units are locked at their current settings Zero Range setting is locked at 296 otable Range setting is locked at 1d Auto Zero Tracking setting is locked at 0 5d Stable Only is locked On Auto Print Continuous is disabled Lb oz is locked Off 5 1 Settings Before verification and sealing perform the following steps 1 Verify that the menu settings meet the local weights and measures regulations 2 Perform a calibration as explained in Section 3 9 3 Set the switch to Locked See figure 5 1 5 2 Verifi
8. against overcharging During battery operation the battery symbol indicates the battery charge level The scale will automatically turn off when the batteries are fully discharged For maximum operating time the battery should be charged at room temperature TABLE 2 1 41 to 70 Remaining 71 to 100 Remaining Notes When battery symbol blinks fast approximately 30 minutes working time is left When Lo bRE is displayed the scale will shut off Charging the scale must be performed in a dry environment CAUTION Battery is to be replaced only by an authorized Ohaus service dealer Risk of explosion can occur if the rechargeable battery is replaced with the wrong type or if it is not properly connected Dispose of the lead acid battery according to local laws and regulations EN 4 Ranger Count 3000 Series 3 OPERATION 3 1 Displays Weight Piece Weight s C us ZU Lh e h h a onus pT LA SOCIO e v v v v v v v v v v L APW Lo Wt Lo Auto O Menu Mode A diria mj Count 7 l Pest Li Ibi U l KED NET a 9 LER La Pest Sam S RANGER COUNT3000 2 3 Tare Lo Figure 3 1 Ranger Count 3000 Control panel with LCD displays TABLE 3 1 Control Functions ON ZERO PRINT Target M Turn scale Send the Initiate
9. diritti riservati Printed in China Impreso en la China Imprim en Chine Gedruckt in China Stampato in Cina
10. even if several settings are changed The event counters can be viewed by pressing and holding the MENU button While the button is held the display will show MENU followed by Audit libration Helease the button when Audit is displayed to view the audit trail information alibration enu Mode The audit trail information is displayed in the format CFGxxx and CAL xxx The scale then returns to normal operation EN 28 Ranger Count 3000 Series 6 MAINTENANCE 6 1 Cleaning The housing may be cleaned with a cloth dampened with a mild detergent if necessary Do not use solvents chemicals alcohol ammonia or abrasives to clean the housing or control panel 6 2 Troubleshooting Table 6 1 lists common problems and possible causes and remedies If the problem persists contact OHAUS or your authorized dealer TABLE 6 1 Symptom Possible Cause No power to scale Verify connections and voltage 6 Improper calibration Perform calibration y Unstable environment Move scale to suitable location Move the scale to suitable Unstable environment Cannot calibrate location Incorrect calibration weight e Use correct calibration weight Cannot access mode Mode not enabled Enter menu and enable mode Cannot access unit Unit not enabled Enter menu and enable unit Err 5 Tare out of range Adusi tare value to be within Display overflow Weight exceeds 6 digits Ere 95 Calibration data error Calibration data not present B Disp
11. symbol remains on 0 5d 0 5 scale division 1d 1 scale division 2d 2 scale division 50 5 scale division Filter F LEE oet the amount of signal filtering LOW less stability faster stabilization time MED normal stability stabilization time HI greater stability slower stabilization time AZT Ret oet the automatic zero tracking functionality OFF disabled 0 5d the display will maintain zero until a change of 0 5 divisions per second has been exceeded 1d the display will maintain zero until a change of 1 divisions per second has been exceeded 3d the display will maintain zero until a change of 3 divisions per second has been exceeded Light L GHE sets backlight functionality OFF always off ON always on AUTO turns on when a button is pressed or the displayed weight changes Auto off R OFF oet the automatic shut off functionality OFF disabled 1 powers off after 1 minute of no activity 5 powers off after 5 minute of activity 10 powers off after 10 minute of no activity End Readout End Advance to the next menu or return to the top of the current menu EN 18 Ranger Count 3000 Series 4 5 Unit Menu This sub menu activates units so they will be accessible with the Units button The units in the menu must be turned to be active Note Available units vary by model and local regulations Ranger Count 3000 Series EN 19 4 6 Print Menu Enter this menu t
12. the Accumulate ID number Display on displayed function of the weight or input for store an value to the the current l library COM port application record mode edit recall Primary Function Short Press Zero the display Off Units Turn scale Change the Switch Enter user off weighing between menu SE unit Check View the Audit Function Weigh Trail event Long Press Check counters Count and extended Off Yes No Back Acceptthe Advanceto Gobackto Exit the menu current the next the previous Abort the menu or menu or menu or calibration in setting setting setting progress Increment Decrement the the displayed displayed value value Yes No Back Accept the Advanceto Gobackto Exit the current the next the previous library Library setting library or library or Function setting setting Short Press Increment Decrement the the displayed displayed value value Menu Function Short Press Ranger Count 3000 Series EN 5 TABLE 3 1 Cont 0 9 Primary Enter numeric Enter dot on Clear the last Perform a Display Function values on the the display character tare otart Short Press display from the operation new APW display 12422 Switch owitch between between oecondary scale 1 and positive and Function scale 2 negative Long Press only available value second platform is connected Notes X
13. warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period During the warranty period OHAUS will repair or at its option replace any component s that proves to be defective at no charge provided that the product is returned freight prepaid to OHAUS This warranty does not apply if the product has been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than OHAUS In lieu of a properly returned warranty registration card the warranty period shall begin on the date of shipment to the authorized dealer No other express or implied warranty is given by OHAUS Corporation OHAUS Corporation shall not be liable for any consequential damages As warranty legislation differs from state to state and country to country please contact OHAUS or your local OHAUS dealer for further details OHAUS Corporation 7 Campus Drive ouite 310 Parsippany NJ 07054 USA Tel 1973 377 9000 Fax 41 973 944 7177 With offices worldwide Con oficinas alrededor del mundo Avec des bureaux dans le monde entier Weltweite Geshaftsstellen Con uffici in tutto il mondo www ohaus com 3 0 0 3 7 4 5 4 P N 30037454 2012 OHAUS Corporation all rights reserved todos los derechos reservados tous droits r serv s Alle Rechte vorbehalten tutti i
14. x 0 02 oz 96 oz x 0 03 oz 240 oz x 0 06 oz 480 ox x 0 2 oz 960 oz x 0 4 oz Approved Resolution 1 3000 Repeatability 0 0001 kg 0 0002 kg 0 0005 kg 0 001 kg 0 002 kg 0 0001 kg 0 0002 kg 0 0005 kg 0 001 kg 0 002 kg Non Approved models g kg Ib oz Ib oz Weighing Units EC and OIML Approved models g kg Measurement Canada and NTEP Approved models g kg Ib oz are Range Stabilization Time 3 LCD with white LED backlight 6 digit 7 segment 20 5 mm 0 8 in in characters 20 mechanical buttons Battery Operating Time at 20 C 210 hours with backlight turned off Construction ABS plastic housing with 304 stainless steel SST platform Approval Class Pan Dimensions Gross Weight 5 45 kg 12 Ib Weight Display Ranger Count 3000 Series 7 2 Drawings 311 mm 12 2 in 327 mm 12 9 in Figure 7 1 Dimensions EN 31 EN 32 Ranger Count 3000 Series 7 3 Communication The scale is equipped with an RS232 interface COM1 located under the Options Cover Connecting the scale to a computer enables you to operate the scale from the computer as well as receive data such as displayed weight 7 3 1 Interface Commands Communicate to the scale using the command characters listed in Table 7 2 below TABLE 7 2 OS Turn off Stable Only menu item and allow unstable print 1S Turn on Stable Only menu item and only print stable pr
15. 0 Ranger Count 3000 Series EN 7 3 2 Turning Scale On Off To turn the scale on press and hold the On Zero Off button for 1 second The scale performs a display test momentarily displays the software version and then enters the active weighing mode To turn the scale off press and hold the On Zero Off button until OFF is displayed 3 3 Counting Ranger Count 3000 has three display areas Weight Piece Weight and Count information are displayed in these areas respectively 3 3 1 Preset tare Enter a value with the numeric keypad The value will be displayed in the Count window Pressing Tare key the value in current unit will then be saved as the preset tare To clear the tare value clear the pan and press Tare key 3 3 2 APW Establishment If there is no APW established both the piece weight display and the count display will show O Positive Sampling 1 Place the sample on the pan 2 Input the sample size with the numeric keypad The value is displayed in the count window Alternatively the Sample key can be used instead Pressing the Sample key and Been is displayed in the count window Then input the sample size value only integer 3 Press the Sample key will be displayed in both the piece window and the count window 4 he established APW and count will be displayed Negative sampling 1 Place container with the samples on the pan 2 Tare the scale a NET 0 will be displayed 3 Remove the
16. Set the baud rate 300 300 bps 600 600 bps 1200 1200 bps 2400 2400 bps 4800 4800 bps 9600 9600 bps 19200 19200 bps Parity PAr 9 set the data bits and parity 7 EVEN lt data bits even parity 7 ODD 7 data bits odd parity 7 NONE 7 data bits no parity 8 NONE 8 data bits no parity Stop bit SEGP set the number of stop bits 1 1 stop bits 2 2 stop bits Handshake H ShAFE Set the flow control method Hardware handshaking is only available for COM1 menu NONE no handshaking ON OFF XON XOFF software handshaking Ranger Count 3000 Series EN 23 Alternate command RLt CP Define command character for the Print Tare and Zero commands Print Set the alternate command character for Print A to Z Tare Set the alternate command character for Tare A to Z Zero Set the alternate command character for Zero A to Z End COM1 End COM2 End Advance to the next menu or return to the top of the current menu 4 8 Lock Menu The Lock Menu is a software controlled option which can lock Menu settings to prevent tampering Reset no yes Lock Cal off on Lock Setup off on Lock Read off on Lock Unit off on Lock Print off on Lock COM off on End Menu Lock Exit menu Reset ESEt Reset the Lock menu to factory defaults NO no reset YES reset Lock Cal L CAL Set the status OFF Calibration menu is not locked ON Calibration menu is locked Lock Setup Set the
17. actor that corresponds to your location End Cal End Advance to the next menu or return to the top of the current menu Ranger Count 3000 Series EN 15 4 3 Setup Menu Enter this menu to set scale parameters Reset no yes Dual Scale off on Cap2 1 9999 Grad2 0 00005 0 5 Power on unit auto kg g Ib oz Ib oz Zero Range 2 1096 Auto Opt off on Auto Tare on off on acc Beeper Opt off on Beeper Signal off accept under over under over Beeper Key off on Accumulation off auto manual End Setup Exit menu Reset ESEt Reset the Setup menu to factory defaults NO not reset YES reset Dual Scale d ScALE set the status of the second scale input Scale 2 OFF disabled ON enabled Cap2 CAP 2 Set the capacity of Scale 2 1 9999 Grad2 G Rde Set the readability of Scale 2 0 00005 0 5 Power on unit Pur Set the unit of measure displayed at startup AUTO last unit in use when turned off kg kilograms 0 grams pounds 02 ounces Ib oz pound ounces Zero Range 05 0 oet the percentage of scale capacity that may be zeroed 2 Zero up to 2 percent of capacity 1096 zero up to 10 percent of capacity Auto Opt R GPt oet the Automatic Optimization OFF disabled ON APW automatically optimized EN 16 Ranger Count 3000 Series Auto Tare R ER E oet the automatic tare functionality OFF Automatic Tare is disabled ON
18. cal data is valid under the following ambient conditions Operating temperature 10 C to 40 C Relative humidity 2096 to 85 relative humidity non condensing Height above sea level Up to 2000 m Power AC power 100 240V 50 60 Hz internal rechargeable sealed lead acid battery EMC See Declaration of Conformity Pollution degree 2 Installation category Il Main supply voltage fluctuations up to 10 of the nominal voltage 7 1 Specifications TABLE 7 1 MODEL RC31P1502 RC31P3 RC31P6 RC31P15 RC31P30 1 5 kg x 0 00005 kg 3 kg x 0 0001 kg 6 kg 0 0002 kg 15 kg x 0 0005 kg 30 kg x 0 001 kg Capacity x Readability 1500 g 0 6000 0 0 3000 0 0 15000 g 0 30000 09 0 Max x d non approved 3 Ib x 0 0001 Ib 6 0 0002 Ib 15 Ib x 0 0004 Ib 30 Ib 0 001 Ib 60 Ib x 0 002 Ib 48 oz x 0 002 oz 96 oz 0 003 oz 240 oz 0 006 02 4800xx0 020z2 960 02 oz Maximum Displayed 1 30000 Resolution Minimum Recommended Saniple Weight 1g 0 002Ib 2g 0 004Ib 4g 0 01Ib 10g 0 02Ib 20g 0 04lb e Recommended 9 9959 0 000011b 0 01g 0 00002Ib 0 02g 0 00005lb 0 05g 0 00011b 0 1 g 0 0002lb 1 5 kg x 0 0005 kg 3 kg x 0 001 kg 6 kg x 0 002 kg 15 kg x 0 005 kg 30 kg x 0 01 kg Capacity x Readability 1500 0 0 15000 0 0 6000 9 0 3000 09 09 30000 0 10g Max e approved 3 Ib 0 001 Ib 6 Ib x 0 002 Ib 15 0 004 Ib 30 Ib x 0 01 Ib 60 Ib x 0 02 Ib 48 oz
19. cation The local weights and measures official or authorized service agent must perform the verification procedure 5 3 Sealing The local weights and measures official or authorized service agent must apply a security seal to prevent tampering with the settings Refer to the illustration below for sealing methods Lock Switch Paper Seal Wire Seal Unlocked Locked with Paper Seal Locked with Wire Seal Figure 5 1 Sealing 5 4 Audit Trail Seal USA only For jurisdictions that use the audit trail method the local weights and measures official or authorized service agent must record the current configuration and calibration event counter values at the time of sealing These values will be compared to values found during a future inspection Note A change to an event counter value is equivalent to breaking a physical seal The audit trail uses two event counters to record changes to configuration and calibration settings e The configuration event counter CFG will index by 1 when exiting the menu if one or more of the following menu settings are changed Dual Scale Capacity 2 Graduation 2 Zero Range Stable Range Auto Zero Tracking AZT Units kg g oz lb or Ib oz or Stable Printing Dual Scale Capacity 2 Graduation 2 Zero Range setting is locked at 2 e The calibration event counter CAL will index by 1 when exiting the menu if a Span Calibration or GEO setting change is made Note that the counter only indexes once
20. e of the initial verification must be carried out by an authorized and certified service organization established within the European Community or by the National Notified Body The first stage of the initial verification has been carried out at the manufacturers work It comprises all tests according to the adopted European standard EN 45501 1992 paragraph 8 2 2 If national regulations limit the validity period of the verification the user of the weighing instrument must strictly observe the re verification period and inform the respective weights and measures authorities Ranger Count 3000 Series EN 35 Disposal In conformance with the European Directive 2002 96 EC on Waste Electrical and Electronic Equipment WEEE this device may not be disposed of in domestic waste This also applies to countries outside the EU per their specific requirements The Batteries Directive 2006 66 EC introduces new requirements from oeptember 2008 on removability of batteries from waste equipment in EU Member States To comply with this Directive this device has been designed for safe removal of the batteries at end of life by a waste treatment facility Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment If you have any questions please contact the responsible authority or the distributor from which you purchased this device Should this device be pa
21. e weight window 3 Press the ID or Yes key c ECRLL is displayed and the data is retrieved Editing a record 1 Enter the ID number using the keypad and press the ID key rECRLL and l b are displayed in the weight and piece weight windows respectively 2 Long press the ID key Ed k and L b are momentarily displayed in the weight and piece weight windows respectively 3 Part number is displayed in the weight and piece weight windows Default setting is 000000 00008000 Use the keypad to enter part number if required 4 Press the Yes key to save the part number and advance to the preset Tare setting Use the keypad to enter a preset Tare value if required 5 Press the Yes key to save the preset Tare value and advance to the APW setting Use the keypad to enter a new APW value if required 6 Press the Yes key to save the APW and advance to the Limits setting Use the keypad to enter new under and over limits if required 7 Press the Yes key to save the limits SRUE is displayed in the weight window Press the Yes key to save the data Clearing a record 1 Enter the ID number using the keypad and press the ID key rECRLL and l b are displayed in the weight and piece weight windows respectively 2 Press the C key dELEEE is displayed in the weight window Press the Yes key to delete the data Ranger Count3000Series EN f 4 MENU SETTINGS The User Menu allows the customizing of scale set
22. ed value is accumulated automatically The gt icon will keep flashing until the weight is removed 3 4 2 Viewing and Clearing Statistical Data When the pan is cleared press the M key to display the statistical information To clear the accumulation data press the Tare key while the statistical information is displayed The display shows Ctx RCC Press the Yes key to clear the stored data and return to current mode Notes The item must be removed from the pan before the next item can be accumulated Only stable weights are stored to accumulation data Changing the mode will clear the accumulation data 3 5 Check Weighing Press and hold the Target key ChEcF is displayed in the weight window Release the Target key when UJE bH is displayed in the piece weight window Use this mode to compare the weight of items to a target weight range 1 Press the Target key UndEr is displayed in the weight screen The previous under limit will is displayed in the piece weight screen 2 Input the target weight with the numeric keypad The value will be displayed in the count window 3 Press the Yes key to accept the under limit value 8UEr is displayed in the weight screen 4 Repeat step 2 and 3 above to set the over limit 5 If required place an empty container on the pan and press Tare 6 Place sample material on the pan or in the container If the sample weight is under the target weight range the yellow LED will light If the
23. gs gt open Control Panel 2 Double click Accessibility Options in Control Panel Select the General tab Check Use Serial Keys and click the Settings button Select the Serial Port set the Baud rate to 9600 and click OK Close the Control Panel Run Excel to open one blank sheet Focus the cursor on one item At this time if scale sends data to the PC through the RS232 port the data will be put into the cell and the cursor will automatically move to the next vertical cell NO O1 5 OW Notes For Windows 7 setup contact Ohaus If the weighing value is a negative number set the target cell in TEXT format Otherwise Excel will not distinguish it as a negative number Please do not use this function during continuous printing Print Library P L b Print the library data No disabled Yes All library records stored in memory are printed End Print End Advance to the next menu or return to the top of the current menu EN 22 Ranger Count 3000 Series 4 7 COM1 Menu Enter this menu to define communication parameters Reset no yes Baud Rate 300 9600 19200 Parity 7 even 7 odd 7 none 8 none Stop bit 1 2 Handshake none On Off Alternate command Print A P Z Tare A T Z Zero A Z End COM Exit menu Note The COM2 menu is only displayed if a second interface is installed Reset ESEt Heset the COM menu to factory defaults NO noreset YES reset Baud Rate bRud
24. ial removed from the scale is within the target range In this case the UNDER and OVER limits are both negative values The UNDER limit must be greater than the OVER limit Place the item to be weighed on the scale and press Tare Remove a portion of the item until it is within the ACCEPT range Zero Check Zero check is used when comparing subsequent samples to an initial reference sample In this case the UNDER limit must be a negative value and the OVER limit must be a positive value Place the reference item on the scale and press Tare Remove the reference sample and place the item to be compared on the scale to determine if it is within the ACCEPT range EN 10 Ranger Count 3000 Series 3 7 Library When an item is processed on a regular basis the item s data may be stored in memory for future use This memory is referred to as the scale s Library Storing a record 1 Press the ID key SEoc and t b are displayed in the weight and piece weight windows respectively The first vacant ID number will blink in the count window 2 Press ID key to save the data to the displayed ID number Alternately a different ID number may be selected using the keypad and then pressing the ID key Note Maximum 30 records from O to 29 Retrieving a record 1 Enter the ID number using the keypad and press the ID key rECRLL and L b are displayed in the weight and piece weight windows respectively 2 ID number is displayed in th
25. int Interval Print x Print Interval 1 3600 sec OP turns auto print OFF oame as pressing Zero Key oame as pressing Tare Key Download Tare value in grams positive values only sending OT clears tare if allowed Print current unit g kg oz Ib oz t c custom oet scale to unit x 129 2 kg 3 lb 4 02 5 0 02 6 1 Enter Header line where x line number 1 to 5 text header text up to 24 alphanumeric characters 4 4 Enter Footer line where line number 1 to 2 text rear text up to 24 F text alphanumeric characters Global reset to reset all menu settings to the original factory defaults EscR p Escape key R Key H x text 7 3 2 Connections for RS232 The 9 pin female subminiature D connector COM 1 is provided for interfacing to other devices The pin connections are as follows Active pins PIN 2 TXD PIN 3 RXD PIN 5 Ground 5 4 3 Q0 0 CN PAN Pg S 9 8 7 Figure 7 2 RS232 Pins Ranger Count 3000 Series EN 33 7 4 Printouts The following sample print outs are generated by the Print button P Command or alternate print command The content of the printout is defined in the Print Content menu item A maximum of 24 characters can be printed on each line Description Comment PN 123456789926 Part Number line If Printx Content P N is ON 926 PCS Result line If Printx Content gt Result is ON 12 33 kg G Gro
26. layed during tare setting usy ae zero setting printing DD Action not allowed Function not executed Calibration value outside m Calibration error allowed limits rR E Unstable environment Move the scale to suitable Incorrect calibration weight location Use correct calibration weight Lo REF Low reference weight warning Increase reference weight Reference weight too small Weight on the pan is too small Unacceptable reference weight ito define a valid reference weight Increase reference EE Err weight Battery fails to charge Have battery replaced by fully l 3 Battery is defective Ohaus authorized service dealer Ranger Count 3000 Series EN 29 6 3 Service Information If the troubleshooting section does not resolve or describe your problem contact your authorized OHAUS service agent For service assistance or technical support in the United otates call toll free 1 800 526 0659 between 8 00 AM and 5 00 PM EST An OHAUS product service specialist will be available to provide assistance Outside the USA please visit our web site www ohaus com to locate the OHAUS office nearest you 6 4 Accessories 2nd scale Interface Kit 30037446 Ethernet Interface Kit 30037447 2 RS232 Interface Kit 30037448 USB Interface Kit 30037449 In Use Cover 30037450 Weigh Below Hook 30025993 Printer Contact Ohaus Communication Cables Contact Ohaus EN 30 Ranger Count 3000 Series 7 TECHNICAL DATA The techni
27. o set printing parameters Default settings are bold Note The Print2 menu is only displayed if a second interface is installed Reset no yes Stable Only off on Auto Print off on stable interval continuous accept Content P N gt off on Result gt off on Gross gt off on Net gt off on Tare gt off on Header gt off on Footer gt off on Mode gt off on Unit gt off on Info gt off on APW gt off on Acc gt off result all Layout Format gt S M Feed gt Line 4 Lines form Data Transfer off on Print Lib yes no End Print Exit menu Reset ESEt Reset the Print menu to factory defaults NO no reset YES reset Stable Only SEAbLE Set the printing criteria OFF values are printed immediately ON values are only printed when the stability criteria are met Auto Print A Pr int Set the automatic printing functionality OFF disabled ON STAB printing occurs each time the stability criteria are met INTER printing occurs at the defined interval ACCEPT printing occurs continuously CONT printing occurs each time the display is within the Checkweigh accept range and stability criteria are met Content Canehe Define the content of the printed data Part Number Set the status OFF disabled ON enabled EN 20 Ranger Count 3000 Series Result Set the status OFF 0 ON the displayed reading is printed
28. sample is within the target weight range the green LED will light If the sample is over the target weight range the red LED will light Ranger Count 3000 Series EN 9 3 6 Check Counting Press and hold the Target key ChEcF is displayed in the weight window Release the Target key when Count is displayed in the piece weight window 1 Press the Target key UndEr is displayed in the weight screen The previous under limit will is displayed in the piece weight screen 2 Input the sample size with the numeric keypad The value will be displayed in the count window 3 Press the Yes key to accept the under limit value 8UEr is displayed in the weight screen 4 Repeat step 2 and 3 above to set the over limit 5 If required place an empty container on the pan and press Tare 6 Place sample material on the pan or in the container If the sample quantity is under the target pieces range the yellow LED will light If the sample is within the target pieces range the green LED will light If the sample is over the target pieces range the red LED will light Positive Check Positive check is used to determine when the material added to the scale is within the target range In this case the UNDER and OVER limits must be positive values The OVER limit must be greater than the UNDER limit Add material to the scale until it is within the ACCEPT green range Negative Check Negative check is used to determine when the mater
29. samples from the container weight window will now display a negative net reading 4 Input the sample size with the numeric keypad The value will be displayed in the count window Alternatively the Sample key can be used instead Pressing the Sample key and is displayed in the count window Then input the sample size value only integer 5 Press the Sample key will be displayed in both the piece window and the count window 6 The established APW and count will be displayed EN 8 Ranger Count 3000 Series Entering a known APW 1 Press the APW key will be displayed in the piece window 2 Input the APW value 3 Press the APW key the new APW value will be displayed in piece weight window 3 3 3 Simple Counting 1 After a valid APW is established if required place a container on the pan and press Tare 2 Place the sample on the pan 3 he quantity will be displayed in the count window 3 4 Accumulation and Statistics The Accumulation feature enables manual or automatic totalizing of displayed values Statistical data is stored in memory for review and printing Accumulation works together with Check mode 3 4 1 Accumulating Displayed Values With ACCUMULATE set MANUAL place the item on the scale and press the M key to add the weight to accumulation data The gt icon will keep flashing until the weight is removed With ACCUMULATE set to AUTO place the item on the scale The display
30. ss value line If Printx gt Content gt Gross is ON and a tare value is entered 11 11 kg NET Net value line If Printx Content gt Net is ON and a tare value is entered 1 22 kg T Tare value line If Printx Content gt Tare is ON and a tare value is entered APW 1 2 g Information line If Printx Content Info is ON N 3 Accumulation line If Printx Content gt Accu is All ACC 2875 Pcs Accumulation line If Printx Content gt Accu is All BL o WM i Meu oy AB E EN 34 Ranger Count 3000 Series 7 5 Compliance Compliance to the following standards is indicated by the corresponding mark on the product Mark Standard This product conforms to the EMC Directive 2004 108 EC the Low Voltage Directive 2006 95 EC and the Non Automatic Weighing Instrument Directive 2009 23 EC The Declaration of Conformity is available online at europe ohaus com europe en home support compliance aspx AS NZS CISPR 11 CAN CSA C22 2 No 61010 1 04 UL Std No 61010 1 MC173467 Important notice for verified weighing instruments M Weighing Instruments verified at the place of manufacture bear one of the preceding marks on the packing label and the green M metrology sticker on the descriptive data plate They may be put into service immediately Weighing Instruments to be verified in two stages have no green M metrology on the descriptive data plate and bear one of the preceding identification marks on the packing label The second stag
31. ssed on to other parties for private or professional use the content of this regulation must also be related Disposal instructions in Europe are available online at europe ohaus com europe en home support weee aspx Thank you for your contribution to environmental protection FCC Note This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense Industry Canada Note This Class B digital apparatus complies with Canadian ICES 003 ISO 9001 Registration In 1994 OHAUS Corporation USA was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International BVQI confirming that the OHAUS quality management System is compliant with the ISO 9001 standard s requirements On June 21 2012 OHAUS Corporation USA was re registered to the ISO 9001 2008 standard EN 36 Ranger Count 3000 Series Limited Warranty OHAUS products are
32. status OFF Setup menu is not locked ON Setup menu is locked Lock Read Set the status OFF Readout menu is not locked ON Readout menu is locked Lock Unit L Un E Set the status OFF Unit menu is not locked ON Unit menu is locked EN 24 Ranger Count 3000 Series Lock Print Set the status OFF Print menu is not locked ON Print menu is locked Lock COM t C amp P7 Set the status OFF COM menu is not locked ON COM menu is locked End Lock End Advance to the next menu or return to the top of the current menu 4 9 End Menu Press Yes to advance to the Calibration menu Press No to exit the menu and return to the current application mode 4 10 Additional Features To use this feature remove AC power from the scale and remove the protective cover for the weigh below opening Install a suitable hook into the access hole at the bottom of the scale as shown Do not over tighten tighten finger tight Mount the scale onto an appropriate assembly that allows free working space below the hook See figure 4 1 Refer to section 6 4 for optional hook Note Never allow the scale to rest directly on the hook Figure 4 1 Setup for the Weigh Below Hook Ranger Count 3000 Series EN 25 TABLE 4 2 GEO CODES Elevation in meters O0 325 650 975 1300 1625
33. tings Note Additional Sub Menus may be available if Interface Options are installed See Interface User Manual for the additional setting information 4 1 Menu Navigation User Menu Mew GAL SELUP rEAd Unit co Printi 6001 LOCk End Notes Only visible when dual scale is set to On Some units may not be available in all models When LEGAL FOR TRADE is set to ON LFT Switch in locked position the menu settings are affected as follows Calibration R L menu is not accessible Dual Scale Capacity 2 Graduation 2 and Units are locked at their current settings Zero Range setting is locked at 296 Stable Range setting is locked at 1d Auto Zero Tracking setting is locked at 0 50 Stable Only is locked On Auto Print Continuous is disabled Lb oz is locked Off EN 12 Ranger Count 3000 Series Summary of button navigation functions in menu mode Yes Allows entry into the displayed menu Accepts the displayed setting and advances to the next item No okips by the displayed menu Rejects the displayed setting or menu item and advances to the next available item Back Moves backwards through the upper and middle level menus Backs out of a list of selectable items to the previous middle level menu Exit Exits from menu directly to the active weighing mode For menu items with numeric settings such as Capacity the current setting is displayed with all digits flashing Press the No bu
34. tton to begin editing The first digit is displayed flashing Press the No button to increment the digit or press the Yes button to accept the digit and move to the next digit Hepeat this process for all digits Press the Yes button when the last digit has been set The new setting is displayed with all digits flashing Press the Yes button to accept the setting or press the No button to resume editing This method also applies to setting Checkweigh under and over targets For End menu items pressing the Yes button advances to the next menu while pressing the No button returns to the top of the current menu Ranger Count 3000 Series EN 13 4 2 Cal Menu Enter this menu to perform calibrations Initial Calibration When the scale is operated for the first time a span calibration is recommended to ensure accurate weighing results Before performing the calibration be sure to have the appropriate calibration weights as listed in table 4 1 Ensure that the LFT switch calibration lock is set to unlocked position See figure 5 1 Or adjust the GEO setting according to your location see table 4 2 Procedure Press and hold Menu until PERU Menu is displayed When the button is released the display will show E A L Press Yes to accept SPAR will then be shown Press Yes to begin the span calibration D koal will be displayed Press Yes to accept 6 will be displayed while zero reading is stored Next the displa
35. unt 3000 Series 2 3 Selecting the Location Use the scale on a firm steady surface Avoid locations with excessive air current vibrations heat sources or rapid temperature changes Allow sufficient space around the scale 2 4 Leveling the Equipment The Ranger Count Series has a level indicator as a reminder that the scale should be leveled for accurate weighing To level the scale adjust the feet so the bubble is centered in the circle Be sure the equipment is level each time its location is changed evel Indicator YES NO Adjustable feet Figure 2 2 Level indicator 2 5 Connecting Power AC power is used to power the scale when battery power is not needed First connect the AC power cord supplied to the power input jack then connect the AC plug to an electrical outlet Figure 2 3A Connect the AC power Figure 2 3B Connect the AC power plug to the input jack on the back of plug to the proper AC supply scale Ranger Count 3000 Series EN 3 2 5 1 Battery Power The scale can be used on AC power immediately Allow the battery to charge for 12 hours before using the scale on battery power The scale will automatically switch to battery operation if there is a power failure or the power cord is removed With AC power the scale is constantly charging so the battery charge indicator see item 10 in figure 3 2 will remain lit The scale can be operated during charging and the battery is protected
36. y shows the calibration weight value Place the specified calibration mass on the pan Press Yes to accept the weight or No to select an alternate weight 6 will be displayed while the reading is stored The display will show donE if the calibration was successful The scale returns to the previous application mode and is ready for use TABLE 4 1 Required Span Calibration Mass sold separately 1500g 1 5kg 3lb 150000 15kg 30lb 30009 3kg 6lb 300000 30kg 60lb 6000g 6kg 15Ib o Note Pound masses are used when calibrating in the Ib unit EN 14 Ranger Count 3000 Series Zero2 Perform Span1 Perform Span2 Perform Linearity1 Perform Linearity2 Perform Geographic Adjustment Set 0 12 31 End Calibration Exit menu Bold always represents factory default value Zero2 2ErG Initiates a 2 scale zero calibration Span1 SPAR Initiates a span calibration procedure zero and span A span calibration is important when initially setting up the scale Span2 SPARC Initiates a 2 scale span calibration procedure Lint t Initiates a linearity calibration procedure zero mid point and span Lin2 t N2 Initiates a 2 scale linearity calibration procedure GEO GEG Geographical Adjustment Factor GEO is used to adjust the calibration based on the current location Settings from O to 31 are available with 12 being the default Refer to table 4 2 to determine the GEO f