Contents

1. 29 Metal Cover Replacement 30 Cord Sirap 101 e E EPI MAIER DUM PI Me ARUM UNITS 31 TOU BIC SC OMG A 32 General System Information Components Parts List HTAIR 7 33 Warranty Sitqiement nad SEO MET d PIU 34 35 and Repairs 36 AE HOVERTECH SE International 1 HSManual Symbol Reference o Attention Please read accompanying documents Type BF Applied Part Declaration of Conformity to Medical Device Directive Functional Earth Ground Alternating Current Underwriters Laboratory Agency Approval 120 V Medical Equipment with respect to electrical shock fire and mechanical hazards only in accordance with UL 60601 1 IEC EN 60601 1 CAN CSA 22 2 No 601 1 230 V Medical Equipment with respect to electrical shock fire and mechanical hazards only in accordance with UL 60601 1 IEC 60601 1 2 CAN CSA 22 2 No 601 1 HOVERSLING user MANUAL H T This End Up Temperature Humidity Date of Manufacture Keep Dry 2 s HOV E RTECH Z International HSManual Rev B HOVERSLING user MANUAL Intended Use and Precautions Indications Patients unable to assist in their own lateral or vertical transfer Patien
2. wear and tear Component parts particularly any optional equipment valve caps their attachments and cords will show wear with use over time and eventually may need to be refurbished or replaced This normal type of wear is not covered by our warranty but we will provide prompt high quality repair service and parts at a nominal cost HoverTech International s liability under this warranty and on any claim of any kind for any loss or damage arising out of connected with or resulting from the design manufacture sale delivery installation repair or operation of its products whether In contract or tort including negligence shall not exceed the purchase price paid for the product and upon expiration of the applicable warranty period all such liability terminates The remedies which this warranty provides are exclusive and HoverTech International shall not be liable for any incidental or consequential damages There are no warranties expressed or implied which extend beyond this warranty statement The provisions of these warranty clauses are in lieu of all other warranties expressed or implied and of all other obligations or liabilities on HoverTech International s part and neither assumes nor authorizes any other person to assume for HoverTech International any other liability in connection with Manufacturer sale or lease of said products HoverTech International makes no warranty of merchantability or fitness for a particular purpo
3. Button Control Panel Label Power Cord Jo US UK Euro Hose Nozzle Hose Air Filter Cover AER HOVERTEC SE Internationa 7 Rev B HSManua HOVERSLING user MANUAL Air Supply Keypad Functions Adjustable Speed Indication Adjustable Pressure Setting Q 8 HSManual Power Indication c HOVERMATT 28 34 Transfer Speed High Hm HOVERTECH nternationa Transfer Speed Low The Adjustable Keypad function has four different settings Pressing the button once will result in the lowest inflation setting available A second press of the button increases the air pressure and rate of inflation Pressing the button a third time will again increase the rate of inflation A fourth press of the button results in the highest inflation rate and air pressure available for HoverTech Accessories The STANDBY button may be pressed at any time to cease all air flow NOTE The LED will indicate the inflation speed by the number of flashes i e two flashes equals the second inflation speed All of the settings in the Adjustable range are substantially lower than the HoverMatt and HoverJack settings The Adjustable function is not to be used for transferring it is only for use with HoverTech Accessories which require a lower pressure for slow inflation Standby Used to stop inflation air flow HoverMatt 28 34 For use with 28 amp 34 HoverMatt Air Transfer Mattresses and 34 Hover
4. a Sling Placing HoverSling Patient in Chair 14 HSManual Position patient Release leg snaps Lean patient forward Place the apex of the sling s horseshoe behind the patient until you touch the seat ensuring the sling is centered on the patient s torso Have patient lean back against the sling to hold it in place Adjust leg sections Place the leg sections along the side of the hips and legs then place each leg section underneath each respective leg Cross inside leg support straps and attach all leg straps Cross inside leg support straps by taking one leg support strap and passing it through the other on the opposite side Attach inside and outside leg support straps to hanger bar Attach shoulder straps Attach shoulder straps to the hanger bar for the proper safety and comfort of the patient Color coded strap loops provide easy identification to match strap placement on the left and right side of the patient Shorter strap loops at shoulder blue beige create a more upright sitting position while longer shoulder and leg strap loops white increase recline angle and reduce hip flexion A HOVERTECH SE International Rev B HOVERSLING user MANUAL For HTAIR 2300 ONLY Guidance and Manufacturer s Declaration Electromagnetic Emissions The HoverTech International Air Supply is intended for use in the electromagnetic environment specified below The customer or the user of the HoverTech Internationa
5. a typical commercial or hospital environment NOTE U is the AC mains voltage prior to application of the test level HSManual HOVERTECH SE International Rev B HOVERSLING user MANUAL For HTAIR 2300 ONLY Guidance and Manufacturer s Declaration Electromagnetic Immunity The HoverTech International Air Supply is intended for use in the electromagnetic environment specified below The customer or the user of the HoverTech International Air Supply should ensure that it is used in such an environment Immunity test compliance Electromagnetic environment guidance test level level Portable and mobile RF communications equipment should be used no closer to any part of the HoverTech International Air Supply including cables Conducted RF IEC 3 Vrms than the recommended separation distance 61000 4 6 150 kHz to calculated from the equation applicable to the 80 MHz frequency of the transmitter Recommended separation distance Radiated RF IEC 3 V m d 1 2 61000 4 3 80 MHz to 2 5 GHz d 1 2 VP 80 to 800 MHz d 2 3 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occu
6. cover fold in the three flexible tabs Then insert the screws one at a time starting with the top middle followed by the sides To Remove Hose Attachment Snap 1 Disconnect hose from unit See page 27 2 Remove the phillips head screw and snap HOVERTECH SE International 29 Rev B HSManual HOVERSLING user MANUAL Metal Cover Replacement 1 Disconnect hose from unit See page 27 2 Remove the two phillips head screws on each side to detach the air filter covers Remove the air filters 3 Gently pull the damaged cover apart to allow clearance to remove it 4 Caretully re insert the new metal cover Reassemble the two filters and filter covers 5 Secure everything in place by reinserting four screws Metal Cover Air Filter Cover hy A HOVERTECH SE International HSManual Rev B HOVERSLING user MANUAL Cord Strap Replacement 1 Unsnap the strap and remove power cord 2 Detach the damaged cord strap by removing the screw as shown 3 Reattach strap by positioning in place and securing it with the screw provided Power Cord Strap HOVERTECH SE International 31 Rev B HSManual HOVERSLING user MANUAL Troubleshooting for HTAIR 1200 HTAIR 2300 Y AC from wall Check power cord connections at air supply and wall Check circuit breakers on rear of air supply unit HT Return unit for repair Return unit for repair IN Return unit for
7. preventive maintenance filter cleaning should be performed monthly Infection Control HoverTech International offers superior infection control with our single patient use HoverSling Transfer amp Lift Device Whatever the patient is lying on to keep the hospital bed clean may be placed on top of the HoverSling to help keep it clean If desired the protector sheet or disposable sheet may be used to cover the transfer mattress available for separate purchase This sheet may also be placed under the transfer mattress when log rolling the patient to prevent the bottom of the HoverSling from coming in contact with the hospital bed sheets If the HoverSling is used on an isolation patient the hospital should employ the same protocols procedures it utilizes for the bed mattress and or for the linen in that patient room 20 A SE International HSManual Rev B HOVERSLING user MANUAL Frequently Asked Questions 1 What is the weight limit of the HoverSling The patient weight limit is 700 lbs 318 kg for the 34 HoverSling and 1000 lbs for the 39 HoverSling 2 Is the HoverSling latex free Yes 3 Is the HoverSling FDA Approved HoverTech International is registered with the FDA The HoverSling is listed as a Class device and is exempt from pre market notification and approval gt HO V ERTECH SE International 2 1 HSManual HOVERSLING user MANUAL AIR SUPPLY NO USER
8. repair Check hose connections at air supply and mattress Check hose for rips tears Check mattress for rips tears Verify that air filters are clean BES HOVERTECH 32 International HSManual Rev B HOVERSLING user MANUAL Component Parts List for HTAIR 1200 HTAIR 2300 HoverTech Part Number Description HTA IL Information Labels gt HO V ERTECH SE International 33 Rev B HSManual HOVERSLING user MANUAL Warranty Statement The HoverSling and the HoverTech International Air Supplies are warranted to be tree from defects in materials and workmanship Warranty begins on date of in service by a HoverTech International representative or shipment date HoverSling Transfer amp Lift Device In the unlikely event that a problem arises as a result of a detect in materials or workmanship we will promptly replace your item at our expense and discretion using current models or parts performing the equivalent function upon receipt of the original item to our repair department You must pre notify HoverTech International by phone 800 471 2776 is to be insured by you against loss during transportation and must be shipped with transportation and or broker charges prepaid HoverTech International Air Supplies In the unlikely event that a problem arises as a result of a defect in materials or workmanship we will promptly repair your HoverTech International Air Supply or replace it if we feel that it ca
9. uses the Air Supply can be cleaned by wiping down using a damp cloth with soap and water or mild neutral detergent Dry using a clean dry cloth or disposable paper towel Plastic fitted hose covers are available not spray cleaners or liquids directly on the Air Supply NOTE CHECK YOUR LOCAL STATE FEDERAL INTERNATIONAL GUIDELINES BEFORE DISPOSAL A HOVERTECH SE International 1 9 HSManual HOVERSLING user MANUAL Preventive Maintenance Prior to use a visual inspection should be performed on the HoverTech International Air Supply to insure the power cord 15 not frayed or nicked and that there is no visual damage that would render the Air Supply unusable The HoverSling should have all of its safety straps sling support straps and handles reference Page 6 of the manual for all appropriate parts There should be no tears or holes any damage is found that would cause the HoverSling not to function as intended the HoverSling should be removed from use If any damage is found that would cause the Air Supply not to function as intended the Air Supply should be removed from use and returned to HoverTech International for repair see page 36 Returns and Repairs The Air Supply has air filters on either side of the motor These filters can be accessed by removing the small screws holding the filter cover in place Filters should be cleaned by holding under warm running water Allow to air dry As
10. HOVERTECH SE International HSManual Rev B HOVERSLING user MANUAL Introduction Using the HoverSling Transfer amp Lift Device The HoverSling is a product that combines HoverMatt technology and sling design into one single patient use device This combination allows seamless transition for lateral transfers and lift transfers utilizing one product that can stay with the patient It is radiolucent and artifact free so patients can remain on the device for all ancillary procedures including Diagnostic Radiology Nuclear Medicine MRI or Radiation Therapy The HoverSling is available in two sizes to accommodate patient girth The 34 HoverSling has a 700 Ibs 318 kg weight limit and the 39 HoverSling has a 1000 lbs 454 kg weight limit The Principle of the HoverSling Technology Lateral Transfer Application After the patient is placed on the HoverSling low pressure air from the Air Supply will inflate the mattress At the same time the air is supporting the patient the air is escaping from the perforations in the underside of the HoverSling The escaping air acts as a lubricant to reduce friction which facilitates effortless transfers With less force needed to transfer a patient there is less physical effort and strain expended by caregivers which results in a reduction of workers compensation injuries The HoverSling requires the caregiver to exert a force of approximately 1596 of the patient s body weight for the tra
11. HoverSling From HoverTech International User Manual HMS Meciical T 1300 106 003 F 07 3036 6625 E info hmsmedical com www hmsmedical com au ZF HOVERSLING www HoverMatt com HOVERSLING user MANUAL Table of Contents 02 Intended Use and Precautions 03 04 HOVERSLINC Introduction 05 Part Idenitification FloverS Ing 06 Part 0 07 Air oupply Keypad 08 Instructions for Use as a Transfer 09 Product Specifications Required Accessories 10 11 Use ds G uu 12 14 Electromagnetic Compatibility Chortl 15 18 Ee aa 7 19 Preventative Maintenance Infection eene 20 Freguently Asked QUESTIONS 21 AIR SUPPLY 22 Pan COMO E IM MEE UN 23 Power Cord Clarmp 24 Handle e SIBI 29 Feet or Bumper Rep GCS 26 POSE REMOVA uuu E E 27 Air Filter and Air Filter Cover Replacement 20001 28 Dust Cover Hose Attachment Snap
12. SERVICEABLE PARTS Only qualified service personnel shall perform repairs on the HoverTech International Air Supply 22 SS International HSManua Rev B HOVERSLING user MANUAL Part Identification Control Panel Dust Cover q9 Hose Attachment Snap Air Filter Cover Handle Attachment Hook Metal Cover Foot Grounding Pin Used for a Ground Continuity Bond Test Information Labels Circuit Breakers Power Cord Clamp 0 Power Cord Strap Bumpers Power gt US UK Euro ER HOVERTECH Ja Sz International EE Rev B HSManual HOVERSLING user MANUAL Power Cord Clamp Replacement HOVERSLING user MANUAL Handle Replacement 1 Remove the damaged handle by unscrewing the socket head screw from the barrel nut using two 5 32 allen wrenches as shown 2 Attach the new handle by reversing the process When tightening the screw be sure that the handle can rotate easily The screw is treated with thread lock to secure it in place 5 32 Allen rene LN Socket Head Screw 5 32 Allen Wrench HOVERTECH SE International 25 Rev B HSManual HOVERSLING user MANUAL Feet or Bumper Replacement 1 The feet and bumpers are held in place by a self adhesive backing Use a small flat bladed screwdriver to pry up an edge and gently remove the foot or bumper 2 Clean surface thoroughly to remove any excess ad
13. Sling used as a lateral transfer mattress HoverMatt 39 50 amp HoverJack For use with 39 amp 50 HoverMatt Air Transfer Mattresses 32 8 39 HoverJack Air Patient Lifts and 39 HoverSling AE HO V ERTECH SE International Rev B HOVERSLING user MANUAL Instructions for Use as a Transfer Mattress 1 Ensure sling support straps are stowed in sleeves and the four 4 snaps located at the center leg and foot sections are connected 2 Patient should be in a horizontal position for transfer repositioning Regardless of ease of patient movement for safety always use a minimum of two caregivers for the transfer 3 Place the HoverSling underneath patient using log rolling technique and attach patient safety straps Whatever the patient is lying on to keep the bed mattress clean can be placed on top of the HoverSling 4 Plug the Air Supply power cord into an electrical outlet 5 Insert the Air Supply hose nozzle into mattress in one of the two hose inlets located parallel to the foot end and snap in place Be sure transfer surfaces are as close as possible and brake the wheels of the bed stretcher 7 If possible transfer from a higher surface to a lower surface 8 Turn on Air Supply by choosing the appropriate speed HOVERMATT HOVERMATT Or Use for sizes Use for sizes 28 amp 34 39 amp 50 9 Grasp transter handles and pull patient on an angle either head first or feet first until patie
14. connected to the hanger bar before the patient is lifted Make sure patient is sitting securely in the HoverSling before transferring or lifting Never lift a patient higher than is needed to complete the lift transfer Never leave the patient unattended while using the HoverSling Use the HoverSling according to both the lift and HoverSling instructions f the weight limit differs among the patient lift hanger bar and HoverSling then the lowest weight limit applies AERE HOVERTECH SE International 3 Rev B HSManual HOVERSLING user MANUAL Intended Use and Precautions continued 4 Precautions HoverSling as an Air Transfer Mattress Never leave patient unattended on an inflated HoverSling Never attempt a lateral transfer on an uninflated HoverSling Ensure patient is centered on HoverSling before inflating When transferring to or from a low air loss bed set the bed mattress air flow to the highest level for a firm transfer surface Precautions HoverTech International Air Suppl Route the power cord in a manner to ensure freedom from hazard Avoid blocking the air intakes of the Air Supply Use this product only for its intended purpose as described in this manual WARNING For safety always use a minimum of two people during patient transfer CAUTION Avoid electric shock Do not open Air Supply WARNING Reference product specific manuals for additional operating instructions
15. hesive that may have been left behind Apply the new part by removing the backing material and position as shown Press firmly to ensure adhesion _ a gt EE HOVERTECH 6 SE International HSManual Rev B HOVERSLING user MANUAL Hose Removal 1 Remove the damaged hose by lifting the dust cover slightly and unsnapping the hose from the side of the unit as shown 2 Push the release button at the top of the unit to remove the hose 3 Attach the new hose by reversing the process Hose Release Button Dust Cover Hose Attachment Snap A HOVERTECH 27 Sz International HSManual HOVERSLING user MANUAL Air Filter and Air Filter Cover Replacement To Remove Air Filter Covers 1 Disconnect hose from unit See page 27 2 Remove the two phillips head screws on each side to detach the air filter covers 3 Re attach the new air filter covers and screws To Remove Air Filters 1 Follow the steps shown above to remove air filter covers 2 Remove air filters and replace 3 Re attach the air filter covers and screws For air tilter cleaning instructions please refer to Page 20 AE HOVE RTECH 28 SE International HSManual Rev B HOVERSLING user MANUAL Dust Cover Hose Attachment Snap Replacement To Remove Dust Cover 1 Disconnect hose from unit See page 27 2 Lift the cover flap to remove the phillips head screws that attach the dust cover 3 To replace dust
16. l Air Supply should ensure that it is used in such an environment Compliance Electromagnetic Environment Guidance The HoverTech International Air Supply uses RF energy only for its internal function Therefore RF emissions ne Group 1 its RF emissions are very low and are not likely to CISPR11 I cause interference in nearby electronic equipment CISPR11 The HoverTech International Air Supply is suitable Closs for use establishments other than domestic IEC 61000 3 2 and those directly connected to the public low Voltage fluctuations voltage power supply network that supplies Complies buildings used for domestic purposes flicker emissions 61000 3 3 HOVERTECH SE International 1 5 Rev B HSManual HOVERSLING user MANUAL For HTAIR 2300 ONLY Guidance and Manufacturer s Declaration Electromagnetic Immunity The HoverTech International Air Supply is intended for use in the electromagnetic environment specified below The customer or the user of the HoverTech International Air Supply should ensure that it is used in such an environment 60601 Electromagnetic Environment Immunity Test Compliance Level Level Guidance Electrostatic Discharge ESD IEC 61000 4 2 Electrical fast Transient burst IEC 61000 4 4 Surge IEC 61000 4 5 Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Power Freq
17. ng behind patient s back until the apex of the horseshoe reaches the coccyx of the patient Have patient lean back to hold sling in place Cross inside leg support straps and attach Release inside leg snaps Cross inside leg support straps by taking one leg support strap and passing it through the other on the opposite side Attach inside leg supports to hanger bar Attach outside leg support straps Attach outside leg support straps to hanger bar Attach shoulder straps Attach shoulder straps to the hanger bar for the proper safety and comfort of the patient Color coded strap loops provide easy identification to match strap placement on the left and right side of the patient Shorter strap loops at shoulder blue beige create a more upright sitting position while longer shoulder and leg strap loops white increase recline angle and reduce hip flexion HOVERTECH SE International Rev B HOVERSLING user MANUAL Instructions for Use as a Sling Disconnecting HoverSling Patient in Bed 1 Position patient and lower onto bed Position patient above center of bed Raise head of bed and lower patient on to bed 2 Detach strap loops Detach strap loops trom hanger bar 3 Place the straps back into the strap pockets Place the straps back into the strap pockets on the inside and outside edges of the HoverSling HOVERTECH SE International 13 Rev B HSManual HOVERSLING user MANUAL Instructions for Use as
18. nnot repaired at our expense and discretion using current models or parts performing the equivalent function upon receipt of the original item to our repair department You must pre notify HoverTech International by phone 800 471 2776 is to be insured by you against loss during transportation and must be shipped with transportation and or broker charges prepaid Should any HoverTech International product be returned that is not covered under warranty there will be a minimum 100 service charge plus shipping costs Lead time for repairs is approximately 2 weeks Please refer to the Return and Repairs section of this Manual for return instructions This warranty is not an unconditional guarantee for the life of the product Our warranty does not cover product damage that may result from use contrary to Manufacturer s instructions or specifications misuse abuse tampering or damage due to mishandling Warranty specifically does not cover product damage that may result from using an air supply that produces more than 3 5 psi to inflate the HoverSling Equipment that has been neglected improperly maintained repaired or altered by someone other than an authorized representative of Manufacturer or operated in anyway contrary to the operating instructions shall void this warranty AER HOVERTECH 34 SE International HSManual Rev B HOVERSLING user MANUAL Warranty Statement continued This warranty does not cover normal
19. nsfer Sling Application The sling application offers a general purpose design along with enhanced features to improve patient comfort It utilizes a traditional wing and loop system that can be used with most traditional spreader bars to integrate easily into the health care facility The Purpose of HoverSling Technology Consistent utilization of the HoverSling dramatically reduces back injuries to staff that are caused by lateral and vertical transfers and repositioning In addition fewer staff members are required to perform these tasks and a more comfortable transfer is provided for the patient HOVERTECH SE International 5 Rev B HSManual HOVERSLING user MANUAL Part Identification HoverSling Sling Support Straps 6 Patient Satety Strap 1 white beige and blue color coded straps for easy identification DO NOT USE TO TRANSFER Lateral Transfer Handles 4 N 2 NM T 4 Uy f MA m NY N Label Includes Model Air Supply Hose Inlets 2 Serial and Instructions for Use Before each use check to ensure Safety straps intact v Handles intact V No tears or holes v Do Not Launder label intact V No loose threads v Sling straps intact HOVERTECH SE International 6 Rev B HSManual HOVERSLING user MANUAL Part Identification Air Supply Carrying Handle Attachment Hook Power Cord Strap Hose Release
20. nt is in desired position 10 Ensure the patient is centered on the receiving equipment prior to deflation especially if the width of the equipment receiving the patient is less than the width of the HoverSling 11 Press Standby to deflate HoverSling and employ the bed stretcher rails NEVER LEAVE PATIENT UNATTENDED ON AN INFLATED HOVERSLING AER HOVERTECH SE International 9 Rev B HSManual HOVERSLING user MANUAL Product Specifications Required Accessories HoverTech International Air Supply Classification Not for use in the presence of flammable anesthetics or in a hyperbaric chamber or oxygen tent Type of protection against electric shock CLASS EQUIPMENT Degree of protection against electric shock TYPE BF APPLIED PART Protection against ingress of water Ordinary not protected Mode of operation CONTINUOUS OPERATION To remove supply mains unplug equipment from wall Use Temperature 50 to 104 F 10 to 40 C Use Humidity 1096 to 7096 Non Condensing Storage Shipping Temperature 40 to 176 F 40 to 80 C Storage Shipping Humidity 1096 to 7096 Non Condensing Power Input 120 V 60 Hz 10 A North American version 230 V 50 Hz 6 European Version Air Supply Dimensions 12 5x7 x7 inches 31 75 x 17 8 x 17 8 Air Supply Weight 11 Ibs 5 kg Air Supply Material Fire Retardant ABS Stainless Steel Power Cord Length 15 feet 457 cm Required Accessory for use as an air
21. r in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHZ and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters and electromagnetic site survey should be considered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above the device should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the device Over the frequency range 150 kHz to 80 MHz the field strengths should be less than 3 V m HOVERTECH SE International 1 7 HSManual HOVERSLING user MANUAL For HTAIR 2300 ONLY Recommended separation distances between portable and mobile RF communications equipment and the HoverTech International Air Supply The HoverTech International Air Supply is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the u
22. se There is no warranty that the goods will be fit for a particular purpose By accepting the goods the buyer acknowledges that buyer has determined the goods are suitable for the buyer s purposes MANUFACTURER S SPECIFICATIONS ARE SUBJECT TO CHANCE AE HOVERTECH SE International 35 Rev B HSManual HOVERSLING user MANUAL Returns and Repairs All products being returned to HoverTech International must have a Return Goods Authorization Number issued from the company Please call 800 471 2776 for an RGA Any products returned without the necessary RGA may cause a delay in the repair time If the product is not covered under warranty a minimum charge of 100 will be assessed for each repair Should a repair charge be assessed HoverTech International will notify the facility and a purchase order for the repair will need to be issued before the repair can be completed Lead time for repairs is approximately 2 weeks All products should be sent to HoverTech International 513 South Clewell St Bethlehem PA 18015 Attn Repair Dept RGA Z Phone 800 471 2776 Fax 610 694 9601 3 AE HOVERTECH 6 SE International HSManual Rev B HMS Meaclical Specialising in pieri Transfers T 1300 106 003 HoverTech International 513 South Clewell St Bethlehem PA 18015 Phone 800 471 2776 www HoverMatt com AE HOVERTECH SE International
23. ser of the HoverTech International Air Supply can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the HoverTech International Air Supply as recommended below according to the maximum output power of the communications equipment Rated maximum Separation distance according to frequency of transmitter output power of m transmitter 15 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 3 5 V vP d 3 5 V vP d 7 E vP For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 18 HOVERTECH SE International HSManual Rev B HOVERSLING user MANUAL Cleaning HoverSling Transfer amp Lift Device The HoverSling is a single patient use product and is not intended to be reprocessed IF the HoverSling is laundered the perimeter seam thread will dissolve Air Supply Cleaning and Maintenance In between patient
24. transfer mattress HoverTech International Air Supply Part HTAIR1200 North American Version Part HTAIR2300 European Version 10 HSManual gt HOVERTECH SE International Rev B HOVERSLING user MANUAL Product Specifications Required Accessories HoverSling Patient Weight Limit 34 86 cm 700 lbs 318 kg 39 99 cm 1000 lbs 454 kg HOVERSLING Transfer amp Lift Device Nylon Twill and Non Woven Polypropylene Fiber Width 34 86 cm Inflated or 39 99 cm Inflated 75 190 cm Inflated Required Accessories for use as a sling Any patient lift with 2 3 or 4 point hanger bars Intended for use with Mobile hoists with sling seats Hoist trolleys Stationary hoists fixed to the wall walls floor and or ceiling Stationary free standing hoists All HoverTech International Products are Latex Free For a full product listing go to www HoverMatt com HOVERTECH SE International 1 1 HSManual HOVERSLING user MANUAL Instructions for Use as a Sling Placing HoverSling Patient in Bed Note HoverSling has six 6 sling support straps that need to be attached 12 HSManual Position patient Option 1 Place the HoverSling underneath patient using log rolling technique with the apex of the sling s horseshoe at the coccyx of the patient Elevate the head of the bed Option 2 Elevate the head of the bed Lean patient forward and slide sli
25. ts whose weight or girth poses a potential health risk for the caregivers responsible for repositioning vertically transferring or laterally transferring said patients Contraindications Patients who are experiencing thoracic cervical or lumbar fractures that are deemed unstable unless using in conjunction with a spinal board on top of the HoverSling Intended Care Settings Hospitals long term or extended care facilities Precautions General Caregivers must verity that all caster brakes of bed and stretcher have been engaged prior to transfer or lift Always use a minimum of two caregivers when operating the HoverSling Product should only be used by trained personnel Only use attachments and or accessories that are authorized by HoverTech International Before use a risk assessment shall be carried out to ensure that the correct size of HoverSling is being used for the patient Precautions HoverSling as a Sling Ensure the HoverSling meets the patient s needs before using Aclinical assessment should be carried out by a qualified nurse and therapist before lifting patients using the HoverSling Do not use the HoverSling lateral transfer handles to lift Excessive force on the lateral transfer handles can cause the HoverSling to tear Make sure the HoverSling is not attached to any object other than the hanger bar Once the sling support straps are fully extended make sure that they are properly
26. uency 50 60 Hz magnetic field IEC 61000 4 8 6 kV contact 8 kV air 2 kV for power supply lines 1 kV for input output lines 1 kV line s to line s 2 kV line s to earth gt 5 7 gt 95 U 0 5 40 60 U For 5 cycles 70 30 dip in U For 25 cycles gt 5 gt 95 dip in U for 5 seconds kV contact 8 kV air 2 kV for supply mains kV for input output lines 1 kV line s to line s 2 kV line s to earth gt 5 gt 95 dip in U For 0 5 cycle 40 60 dip in U For 5 cycles 70 30 dip in U For 25 cycles gt 5 gt 95 dip in U for 5 seconds Floors should be wood concrete or ceramic tile If floors covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of the HoverTech International Air Supply requires continued operation during mains interruptions it is recommended that the HoverTech International Air Supply be powered from an uninterruptible power supply or a battery Power frequency magnetic fields should be at levels characteristic of