        TOS5200 User`s Manual - Kikusui Electronics Corp.
         Contents
1.      Basic item       Test voltage  Voltage     SET key  rotary knob       Limit voltage  LIMIT     LIMIT  SHIFT SET  key  rotary knob       Upper limit  UPPER     UPR  LWR key  rotary knob       Lower limit  LOWER     UPR  LWR key  rotary knob       on  off    ON  OFF  SHIFT UPR  LWR  key       Test time  TEST     TEST key  rotary knob       on  off    ON  OFF  SHIFT  TEST  key       Voltage rise time  RISE     TEST key  rotary knob       Additional items       Test voltage frequency  Frequency     MORE key  rotary knob       Start voltage  Start Voltage     MORE key  rotary knob       Voltage fall time  Fall Time     MORE key  rotary knob       Voltage current measurement mode   Measurement     MORE key  rotary knob             Test conditions  basic items        Test voltage  Voltage     For details on how to select settings and enter values  see    Panel Operation Basics        SET unn TEST MORE                         De ONOFF ONOFF KEY LOCK    LOCAL       For details on how to connect the TOS5200 to the DUT  see    Connecting to the Device under  Test  DUT         For details on the LOWER and UPPER settings  see also    About Judgment        Sets the test voltage  You cannot specify a voltage that is greater than or equal to the limit    voltage        Test voltage    0 00 kV to 5 50 kV       Limit voltage  LIMIT     Sets the limit voltage  This prevents an unnecessarily high voltage from being applied to the    DUT because of incorrect operations of the TOS5200        
2.      Internal construction            signal                The input signal circuit and the  output signal circuit share the     amp    Output signal 1 same common    Output   The input signals are pulled up to   controller Output signal 2   15 V  If the input terminals are  opened  the input signal circuit is          15 V put into the same state as when a  high level signal is applied   O Input signal 1       Input  signal  controller              15 V      O Input signal 2              P      gt  4 COM       Internal construction of the SIGNAL H  O connector    Input signal usage example    ee  S1  COM    C      COM          Using a make contact to control input    Use a make contact  such as a relay or switch  to set the input ter   minal to low level     Using a logic element to control input    Use a logic element  such as a transistor  in place of the switch in  the above example     Design the circuit so that a transistor collector current  ic  of 5 mA  or greater flows     Output signal usage example       A       m  400 mA or less    COM    Driving a relay          Relay    Use the output signal to drive a relay              To improve the safety of the circuit  we recom   L mend that you insert a protection fuse or connect  T30 Vor a diode    less          56    TOS5200    Starting a Test                          T30 V or    Low level during signal  less    COM generation          Obtaining a low level digital signal    R Use the output signal to obtain a low leve
3.    PART NO  18027963  9 KIKUSUI n  2 5                   User   s Manual    Withstanding Voltage Tester    TOS5200                                        FMA MAA                      DANGER    This product generates high voltage   Improper operation can lead to serious accidents   o To prevent accidents  be sure to read the section     Safety Precautions during Testing    in this manual   o Keep this manual close to the product so that the  operators can read the manual at any time     O    Search by Topic 8  Component Names 10    General Description 13  Product Overview 14  Options 16    Installation and Preparation 19  Connecting the Power Cord 20   Using the Protection Cover 21  Turning the Power On 22    Connecting to the Device under Test  DUT  24    Safety Precautions for Testing 27  Pre Test Inspection 28   Testing Precautions 28   Remote Control Precautions 29   Interrupting Testing or Operations 29  Emergency Measures 29   Forbidden Actions 30   About Malfunctions 30   To Use the Product for a Long Time Free of  Malfunctions 30       Panel Operation Basics 31  Parts of the Screen 32   Panel Operations 34   Panel Memory 37    Withstanding Voltage Test 39  About Judgment 40   Invalid Settings 40   Setting Test Conditions 41   CONFIG Settings 44   Starting a Test 48   Finishing a Test 51    External Control 53   SIGNAL I O Connector 54   Starting a Test 57   Recalling Panel Memory and Test Modes 57  Interlock Feature 58   STATUS OUT connector 59    Maintenance 61  Pre T
4.    Tests are only executed while the START switch is held down        Fail mode feature       This feature enables you to prevent remotely transmitted stop signals from clearing FAIL  judgments and PROTECTION modes        Timer feature    The test ends when the specified time elapses        Output voltage monitor feature   Volt Error     If output voltage exceeds    setting   350 V    or is lower than    setting   350 V     the TOS5200  switches to PROTECTION mode  output is turned off  and testing finishes        Memory    Up to three sets of test conditions can be saved to memory        Key lock    Protection functions    Locks panel key operations  settings and changes      Under any of the following conditions  the TOS5200 switches to the PROTECTION state   immediately turns output off  and stops testing  PROTECTION on the screen lights        Interlock Protection    An interlock signal has been detected        Power Supply Protection    An error was detected in the power supply        Volt Error Protection    While monitoring the output voltage  a voltage outside of the rated limits was detected   Withstanding voltage test   350 V       Over Load Protection    While monitoring the output power  power exceeding the output power limit was detected   Withstanding voltage test  550 VA       Over Heat Protection    The internal temperature of the TOS5200 became too high        Over Rating Protection    The output current was generated for a length of time that exceeds the reg
5.    VOT vd PROTECTION de  m    l Counts up the  l voltage rise time       48 TOS5200    Starting a Test       If the start voltage has been turned off and the voltage rise time  has not been set    After 0 1 seconds  the TEST LED lights in red  and the test starts     If the test time has been set or if the test time has been turned off    41 42 The TEST LED lights in red during testing  If the test time has been turned on  the display  ened Pel x Bs counts down from the set test time  If it is set to OFF  the display counts up the test time     The time display after  you start testing  differs depending on          I5 ni Nn 15   1 whether TIMER is set  L LI Tu LILI I mA   l to ON or OFF   TIMER ON The timer   s remaining test time is counted down        TIMER OFF The elapsed test time is counted up   When the test time exceeds 999 seconds     999    blinks        To change the voltage setting during a test    While the TEST LED is lit in red  press SET to select the voltage setting value  and then use  the rotary knob to change the value  The voltage is changed immediately     The changed voltage is immediately applied to the test  but the voltage value on the display  shows the measured value  After the test finishes  when the TOS5200 returns to the READY  state  the new voltage value is shown on the display     If you cannot start testing    In the following conditions  you will not be able to start testing  The READY LED will not light        e The memory number is not fi
6.   Input time width       5 ms minimum       1 INTERLOCK     If you open the positive and negative terminals  the output is turned off  and the  TOS5200 is switched to Protection mode    Open  If the resistance between the terminals is 1 2 KQ or greater    Short  If the resistance between the terminals is 1 KQ or less                                                                    2 PMO Panel memory selection signal   The selection signal is latched PMO  PM1  Recalled panel memory number  on the rising edge of the input H H Memory 1  Sep strobe signal to recall panel L H Memory 2  memory  H L Memory 3    The selection of memory is pri  L L    oritized over TEST SEL and  AUTO SEL   4 NC  5 NC  6 NC  7 NC  8 NC  9 STB Panel memory strobe signal input terminal   10 TEST SEL NA  11 AUTO SEL NA  12 COM Circuit common terminal        13 INTERLOCK     If you open the positive and negative terminals  the output is turned off  and the  TOS5200 is switched to Protection mode    Open  If the resistance between the terminals is 1 2 KQ or greater    Short  If the resistance between the terminals is 1 KQ or less                                14 HV ON On during testing and when a voltage remains across  the output terminals    15 TEST On during testing  excluding when voltage is rising or falling     16 PASS On for approximately 0 2 seconds when a PASS judgment occurs   On continuously when the PASS HOLD time is set to HOLD    17 U FAIL On continuously when UPPER FAIL results from judgme
7.  2  EN 61000 3 3  Applicable under the following conditions  The maximum length of all cabling and wiring connected to the  TOS5200 is less than 2 5 m   Shielded cables are being used when using the SIGNAL I O   The high voltage test lead TL31 TOS is being used   Electrical discharges are not occurring outside the DUT                    Safety  Complies with the requirements of the following directive and stand   ards   Low Voltage Directive 2006 95 EC  EN 61010 1 ed3  Class H  Pollution degree 2    Dimensions See    Outline drawing       Weight Approx  14 kg  30 9 Ib     Accessories   Power cord 1 pc        High voltage test lead  TL31 TOS     1 set  1 red wire and 1 black wire  each with alligator clips   1 5 m       SIGNAL I O plug    1 set  assembly type                         High voltage warning sticker 1 pc   Setup Guide 1 pe   Quick Reference English  1 pc    Japanese  1 pc   Safety Information 1 pe   CD ROM 1 pe           1 Limited to products that have the CE mark on their panels  Does not apply to specially made or modified TOS5200s    2 This product confirms to Class A  This product is intended for use in an industrial environment  This product may cause interference  if used in residential areas  Such use must be avoided unless the user takes special measures to reduce electromagnetic emis   sions to prevent interference to the reception of radio and television broadcasts    3 This product confirms to Group 1  This product does not generate and or use intentiona
8.  FUNCTION  ACW                TOS5200    Panel Operations       Locking panel operations  key lock     The key lock feature can be used to prevent changes to the test conditions due to incorrect  operations     To lock the panel operations  press KEY LOCK  SHIFT MORE     In this state  only the START and STOP switches are enabled    When keys are locked  the key lock icon  amp  appears on the screen  If you try to use a locked  key  the key lock icon  4  blinks for approximately 2 seconds     To release the key lock  hold down KEY LOCK  SHIFT MORE  until the key lock icon  amp  dis   appears        If you will run tests using conditions that you have recalled from memory  we recommend that  you use the protection cover        Talk mode    This feature enables you to control the product in the same manner as remote control without  using commands from a PC  It can reduce processing on the PC     If you turn talk mode on  you cannot control the product from a PC  To control the product  from a PC  turn talk mode off     For details on talk mode  see the communication interface manual     TOS5200     See  p  73    3    MEMORY  1          2          3             RECALL       TOS5200    Panel Memory       You can save up to three sets of test conditions  the test conditions currently being used  to  internal memory     Test conditions and values that can be saved      PROTECTION                               Test voltage 0 00 kV to 5 50 kV  Output freguency 50 Hz  60 Hz   Upper l
9.  If you will not use the product for some time or if  the operator will be away from the product  be sure to turn the POWER switch off     STOP    e  e The TEST and DANGER LEDs are both off        Emergency Measures    There are two actions that you must carry out if  due to a malfunction in the product or the  DUT  there is a possibility of an emergency occurring such as electric shock or damage to the  DUT    e Turn the POWER switch off     e Remove the power cord plug from the outlet     Turn the POWER switch off  Remove the power plug  from the outlet  Ge    POWER       KE E    TOS5200    Safety Precautions for       29    30    Forbidden Actions       Turning the power on and off repeatedly    After you turn the POWER switch off  wait for at least 10 seconds before you turn it back on   It is dangerous to do otherwise  because the protective features of the product may not work  effectively  This may cause the product to malfunction  and it may reduce the life of the  POWER switch and internal parts such as the fuses     About Malfunctions    Z WARNING Risk of electric shock     e Until you get the product fixed  make sure that nobody can use it   e For repairs  contact your Kikusui agent or distributor     If the TOS5200 is in one of the states explained below  it may be malfunctioning in a very  dangerous manner   it may not be possible to turn off the high voltage that is being gener   ated  If the tester is not operating properly  it may be generating a high voltage 
10.  Location    in the Safety  information manual  The following precautions pertain  only to this product     e When installing this product  be sure to observe the temper   ature and humidity ranges indicated below   Operating temperature range  0   C to  40   C   32   F to 104   F   Operating humidity range  20  rh to 80  rh   no condensation     e When storing this product  be sure to observe the tempera   ture and humidity ranges indicated below   Storage temperature range   20   C to  70   C    4 F to 158   F   Storage humidity range  90  rh or less  no condensation     Donotuse the productin a poorly ventilated location   The product uses forced air cooling  It sucks air through the  inlet holes on its right  and left panels  and then expels air  through its rear panel  Secure adeguate space around the  product s inlet and outlet holes to prevent the possibility of  fire caused by accumulation of heat   Allow at least 20 cm of space between the inlet and outlet  holes on the side and rear panels and the walls  or obsta   cles    Hot air  approximately 20  C  68   F  hotter than the ambient  temperature  is expelled from the outlet holes  Do not place  objects that are affected by heat near the air outlet     Left panel inlet ae  gt     Right panel inlet holes            Do not use this product near highly sensitive measuring  instruments or receivers   Noise generated by this product may affect other devices   At a test voltage of 3 kV or greater  the product may produce 
11.  Operations                                 nmmenneeneannaeeaaneneennnnanananeeannnnananaeeaaneneanaaae 3  Safety Precautions                              ieenennnennanaaanenennnenaaanaaeennananaannananannnanaaaeannnnnanna 4  Precautions Concerning Installation Location                              emnannneennnenannnae 4  Notations Used in This Manual                                  nemenenenanneeneeananannenenannanneeanannnenee 4  SEACH 4 0  41 K9  0   35 AA AAA 8  Component NAMES iis    53370010 anise e M  LLU pa kak EM ALS ni ll ede V  hk ah aa m AAA os 10    General Description    Product OVO VI OW MA EEGENEN 14  FeaAlurSS issi ska   asusin EE Eege AER ENEE SEENEN 14       Options    Installation and Preparation    Connecting the Power Cord       USING The Protection  COVER eier   T  rningthe Power OM sicsscssssscticesassscossectosnstsvsssocesensanusstssonnasnnaiogsesdhaveuonuateiuisssbisvnavedg odaconenaassssenddseshnnsaststesbdtessstavere  Checking indicators and the status of the interlock feature            s ssssssesreeseeeeee 22  Turning the POWER switch on    23  Turning the POWER switch oft    23   Connecting to the Device under Test  DUT  24  Using test lO e EE 24  Using the optional high voltage test probe   model HOT A TOS HPOZA  TO     26  Disconnecting test leads from the DUT    26    Safety Precautions for Testing        Erectestlospeioft  tege 28  BK eine PRECAUTIONS MSS AEA 28  Remote Control Precautions EE 29  Interrupting Testing or Operations EE 29  Em
12.  TOS5200       When the product is shipped from the factory  a protection cover is attached to the front  panel  This cover prevents unintentional changes to the test conditions  Remove this cover  when you want to set the test conditions     Even when the cover is attached  you can still recall settings from memory  start and stop  tests  perform remote operations  and control the TOS5200 through its USB port  If the tester  will be used to repeatedly perform tests with fixed conditions  such as when being used as  part of a manufacturing line  attach the protection cover to ensure safe operation of the tester   This is useful in preventing incorrect operation of the tester     If the cover is damaged or lost  contact your Kikusui agent or distributor   E Removing the protection cover    Loosen the screw  and then pull the  hook centered at the bottom of the  cover towards you to remove the pro   tection cover from the panel                    E Attaching the protection cover    Insert the tabs at the top of the cover into the slots in the panel  push the bottom part of the  cover until it is attached to the panel  and then use the screw to fix the cover in place     H Storing the protection cover     gt  When you want to use the   gt  TOS5200 without the protection  cover attached  such as when you     Use the screw to fasten i i    the cover in place  will repeatedly perform tests with  xs O frequently changed test condi      tions  you can store the protection  cover on t
13.  corona discharge between its test lead clips  This will gener   ate a significant amount of broadband RF emission  To min   imize this effect  keep the alligator clips away from each  other  Also  keep the alligator clips and test leads away from  conducting surfaces  especially sharp metal edges     Notations Used in This Manual    The TOS5200 Withstanding Voltage Tester is also referred  to as the TOS5200     Device under test is also referred to as DUT     The term    PC    is used to refer generally to both personal  computers and workstations     The following markings are used in the explanations in this  manual     Indicates a potentially hazardous situation which  if  ignored  could result in death or serious injury      N CAUTION    Indicates a potentially hazardous situation which  if  ignored  may result in damage to the product or other  property     NOTE    Indicates information that you should know     DESCRIPTION    Explanation of terminology or operation principle     See          Indicates a reference to detailed information   SHIFT key name  blue letters     Indicates an operation that requires you to press a key  indicated in blue letters while holding down SHIFT        Memo    Indicates useful information     TOS5200    Contents    TOS5200    About the Manuals sic  ssvcvvetestieeetesdeceeeetie NEEN EENS NEEN 2  Checking the Package Content             c ccccseeeeseseeceeeeeeeeeeeceeeeeesaaeeeeneteeseaeeeseneeetas 2  Notes to the supervisor ek 3  Dangerous
14.  if you let go of the probe   Thus  you can definitely check the status     Withstanding Voltage Test       off  0  The fail mode feature is turned off  factory default setting         on  1  The fail mode feature is turned on        Judgment result hold time  Pass Hold     This feature enables you to set the length of time that the TOS5200 maintains a PASS judg   ment state  If you set Pass Hold to HOLD  the measured results remain displayed on the  screen until you press STOP  When a FAIL judgment occurs  the FAIL judgment s measured  results remain displayed on the screen until you press STOP  regardless of the value of the  Pass Hold setting        50 ms  0  to 5 5 s  5  0  50 ms  factory default setting   1  100 ms  2  200 ms  3  15   4  25 5  55    HOLD  6  The results are maintained until you press STOP           Buzzer volume level for PASS judgment  PASS     You can setthe volume level of the buzzer that is sounded when a PASS judgment occurs        0to9 0  1  2  3  factory default setting   4  5  6  7  8 9       TOS5200    eoooe  A  QO    CONFIG Settings       Buzzer volume level for FAIL judgment  FAIL     You can set the volume level of the buzzer that is sounded when a FAIL judgment occurs        0to9 0  1  2  3  4  5  factory default setting   6  7  8  9       RS232C baudrate    You can setthe baud rate for the RS232C interface        9600  0  to 115200  4  0  9600  factory default setting   1  19200   2  38400  3  57600   4  115200       Talk mode    This featu
15.  power cord and all cables are correctly connected     Press the     side of the front panel POWER switch to turn the TOS5200  on    Check that all the front panel indicators light    The firmware version screen will be displayed for a few seconds  and then a message  indicating that the TOS5200 is in PROTECTION mode will be displayed    Check that the tester is in PROTECTION mode through the interlock feature     2 Check that nothing is connected to the SIGNAL I O connector     Firmware version screen       Interlock protection mode       4 Press the  O  side of the front panel POWER switch to turn the TOS5200  off     TOS5200    Turning the Power On       Turning the POWER switch on     See  p  72    1 Connect the included SIGNAL I O plug to the SIGNAL I O connector   Connecting the SIGNAL I O plug will release the interlock feature     2 Press the     side of the POWER switch to turn the TOS5200 on     3 Check the firmware version  Ver x xx  that is displayed on the screen   Check that the firmware version screen is displayed for a few seconds  that the setup  screen for setting the withstanding voltage test conditions is displayed thereafter  and  that the tester is then in READY mode  that the READY LED lights      _ PROTECTION      Firmware version 1 00 display example       The first time that the POWER switch is turned on  the firmware version is displayed  and  then the setup screen for setting the withstanding voltage test conditions is displayed  with  the factory 
16.  settings  If the remedy does not correct the problem  contact your Kikusui  agent or distributor     The power does not turn on          Symptom Items to check and possible remedy  See   The TOS5200 does not operate when Is the power cord disconnected  p  20  the POWER switch is turned on  Has the interlock feature been activated  p  22  Release the interlock feature  p  58       Unable to carry out panel operations    Symptom Items to check and possible remedy       Testing does not begin when the Is a probe connected to the REMOTE connector   START switch is pressed  When a probe is connected  only the probe   s START switch is enabled     Is a STOP signal being applied to the SIGNAL I O connector   When you are using the SIGNAL I O connector to control the TOS5200  the p  57  START switch on the panel is disabled     Is the TEST LED blinking   The TOS5200 is in the middle of the voltage rise time  The test will begin when p  48  the rise time elapses     p  26             If a PROTECTION message is displayed on the screen   The TOS5200 is in PROTECTION mode  See    Protection Functions     p     Is the double action feature on   Press STOP  and then press START within 0 5 seconds of pressing STOP  p  45  Alternatively  turn the double action feature off     74                You cannot start tests while panel memory is being accessed  p  37   Is a message displayed on the screen  40   A setting is invalid  Specify settings that are within the correct range  P   Panel oper
17. 0 5 seconds  If you do not press START within this time limit  the test does not start   the READY LED turns off   This feature ensures that tests are started safely because it  requires you to use both the STOP and START switches        off  0  The double action feature is turned off  factory default setting         on  1  The double action feature is turned on        Momentary feature    This feature enables you to perform tests only while you are pressing START  This ensures  safe testing because it means that your hand must be fixed to the panel or to an optional  RC01 TOS START switch  While this feature is on  testing stops as soon as you release  START  The stop operation is the same as if you had pressed STOP     Using this feature with the optional RC02 TOS  two hand type remote control box   See  p  16 CR      provides an even higher level of safety        off  0  The momentary feature is turned off  factory default setting         on  1  The momentary feature is turned on        Fail mode feature    This feature prevents FAIL judgments and protection modes from being cleared even when a  stop signal is sent with a command  To clear the FAIL judgment when the fail mode feature is  turned on  press STOP on the front panel     If you are using the optional high voltage test probe  HP01A TOS HP02A TOS  Seep  17    we recommend that you turn on the fail mode feature  When a test ends with a FAIL judgment  or ends in protection mode  the FAIL judgment is not cleared even
18. 1 TOS RC02 TOS remote control box    The remote control box can be used to start and stop withstanding voltage test  One model is  for use with one hand  and the other model is for use with two hands        RC01 TOS  one hand  RC02 TOS  two hands        Model DD 5P 9P remote control adapter cable    This is an adapter cable for connecting the product to a  remote control box     16 TOS5200    Options       Model HP01A TOS HP02A TOS high voltage test probe       This is a probe for generating the test voltage  To prevent the test  voltage from being generated unintentionally  this probe has been  designed so that the test voltage is only generated when the user  operates the probe with both hands    When using the test probe  make sure that the rated voltage is not                exceeded   Rated voltage  AC 4kV  DC 5 kV       Model PL02 TOS warning light unit    The warning light unit indicates that the TOS5200 is performing a  test  This enables you to see that a testis in progress from a dis   tance        TOS5200    General Description E    17    18    This page is intentionally blank     TOS5200       Installation and Preparation    This chapter describes how to prepare this  product for use     20    Connecting the Power Cord      Z WARNING e This product conforms to IEC Safety Class    equipment that has a protective con   ductor terminal   Be sure to earth ground the product to prevent electric shock     e The product is grounded through the power cord ground wire  Con
19. 18  Start voltage on and off OFF  Voltage rise time Dis  Voltage fall time OFF  Test voltage frequency 50 Hz  Interface setting USB  CONFIG settings Test Mode Double Action OFF  Pass Hold 50 ms  Momentary OFF  Fail Mode OFF  Buzzer Volume Level Pass 3  Fail 5  Baudrate 9600  Talk Mode OFF  Status Signal Output H V ON OFF  Test OFF  Pass OFF  Upper Fail OFF  Lower Fail OFF  Ready OFF  Protection OFF  Power ON OFF       TOS5200    List of Default Settings       Default panel memory values    TOS5200    There are three panel memory entries  Initially  these entries all contain AC withstanding volt   age test conditions that comply with safety standards     If you initialize the TOS5200  the panel memory entries are also returned to their default val                          ues    Memory number Setting Setting after initialization   1to3 Test mode ACW  Output frequency 50 Hz  Test voltage 0 00 kV  Upper limit 0 02 mA  Lower limit OFF  Test time 0 1s  Start voltage OFF  Voltage rise time Dis  Voltage fall time OFF       73    74    B Protection Functions       If one or more of the causes of the following problems occur  the protective circuits will acti   vate  and you will no longer be able to use the TOS5200  This is referred to as PROTECTION  mode    When the TOS5200 switches to PROTECTION mode     PROTECTION    on the screen lights   Follow the instructions below to remove the cause of the problem and clear the PROTEC   TION mode    If the TOS5200 has switched to PROTECTION 
20. 47  USB  Protection MESA MAA 74    79    80    vV    VEISION  EE 2   Volt Error Protection                         mmmmnnenennnneneenna 74  W   W COMP  2iiiiissta dadada v  ik ke a 32  40   Veure RE 64    TOS5200    KIKUSUI ELECTRONICS CORP     1 1 3 Higashiyamata  Tsuzuki ku  Yokohama  224 0023  Japan  Tel   81 45 593 7570 Fax   81 45 593 7571    Website http   www kikusui co jp       
21. Limit voltage    0 00 kV to 5 50 kV       TOS5200    Withstanding Voltage Test E    41    Setting Test Conditions       Upper limit  UPPER     Sets the upper current limit that is used in judgments  If a current that is greater than or equal  to the upper limit is measured  a U FAIL judgment occurs        Upper limit 0 01 mA to 110 mA       Lower limit  LOWER     Sets the lower current limit that is used in judgments  While LOWER is on  if a current that is  less than or equal to the lower limit is measured  an L FAIL judgment occurs     You can turn on and off the lower limit function using the ON OFF  SHIFT UPR LWR  key        Lower limit 0 01 mA to 110 mA       Test time  TEST     Set the test time  The test time begins when the voltage rise time elapses and ends when the  specified time elapses     You can turn on and off the test time using the ON OFF  SHIFT TEST  key     If the test time is set to off  PASS judgment is not performed   Press STOP to stop testing        Test time 0 1 s to 999s       Voltage rise time  RISE     Separate from the test time  you can set the time that the TOS5200 takes to raise the voltage  to the test voltage        Voltage rise time 0 1 s to 10 0 s       Test conditions  additional items     MORE Press MORE to display the items that are not displayed                                  Start voltage setting  0  off  1  on   Frequency setting  0  50 Hz  1  60 Hz     42 TOS5200    Setting Test Conditions       A 1 Press MORE to display the More 
22. PDF files are included in the accompanying CD ROM   You can view the PDF files using Adobe Reader 9 2 or later     Every effort has been made to ensure the accuracy of this  manual  However  if you have any questions or find any errors  or omissions  please contact your Kikusui agent or distributor     If you find any misplaced or missing pages in the manuals   they will be replaced  If the manual gets lost or soiled  a new  copy can be provided for a fee  In either case  please contact  your Kikusui agent or distributor  At that time  inform your  agent or distributor of the    Part No     written on the front cover  of this manual     After you have finished reading this manual  store it so that  you can use it for reference at any time     Firmware versions that this manual covers    This manual covers firmware versions 1 0X When contacting  us about the product  please provide us with   Model  marked in the top section of the front panel   The firmware version  see page 23   The serial number  marked in the bottom section of the rear  panel     Trademarks    Microsoft  Windows  and Visual Basic are registered trade   marks of Microsoft Corporation in the United States and or  other countries     All company names and product names used in this manual  are trademarks or registered trademarks of their respective  companies     Copyrights  The contents of this manual may not be reproduced  in whole  or in part  without the prior consent of the copyright holder     The specif
23. PROTECTION FMS                                  To save  press  KG Recall while    m holding down    SHIFT     34 TOS5200    Panel Operations       Selecting items    To select an item  use the SET  UPR  LWR  or TEST key     PROTECTION   MS       SET UPR LWR TEST MORE       LIMIT ON OFF ON OFF KEY LOCK    LOCAL          On the basic setup screen  press SET to select the voltage   Example  Selecting the voltage  blinking     If the item that you want to select is not displayed  press LIMIT  SHIFT SET   ON OFF   SHIFT UPR LWR   ON OFF  SHIFT TEST   or press MORE key     On the basic setup screen  hold down SHIFT and press SET to select the limit voltage   Example  Selecting the limit voltage  blinking  j  se nae             1  1  TEST   LIMIT 1 wm ON OFF ON OFF KEYLOCK  EI    LOCAL  1 m l i  FAIL Je LU  i  Ll I    EC    Panel Operation Basics D             SHIFT K    When you enter the setup screen for additional test conditions  MORE  or a CONFIG setup  screen  the MEMORY 2 and MEMORY 3 keys become left and right movement keys     MORE             ES MNA                      MANN  LEY  LILI          Entering values    You can set the value of the blinking item by turning the rotary knob        Turn the rotary knob to set the value  Rotary knob  Kai    1  TEST       aen r 1 TEST    1T a 1  e he MA s     i Lii td    If you hold down SHIFT and turn the rotary knob  you can change the setting resolution   While you hold down SHIFT  the values change quickly     Rotary knob        
24. Test Conditions setup  More  screen   PH  Memo   a ad 2 Press MEMORY 2 or MEMORY 3 to move to the item that you want to set   MEMORY 3 keys to move and then use the rotary knob to set the value    between settings  Each   time you press one of A    ee   these kts  the cursor Press STOP  or press a key other than MORE to exit the More Test Conditions setup screen  moves to the next setting  and return to the basic setup screen    After selecting a setting    use the rotary knob to set   its value     Frequency    Sets the test voltage frequency        50 Hz  0  The test voltage frequency is set to 50 Hz  factory default setting    60 Hz  1  The test voltage frequency is set to 60 Hz           Start voltage  Start Voltage     Separate from the test voltage  you can set the start voltage of withstanding voltage tests   You can set the start voltage to 50   of the test voltage  The output voltage reaches this  value approximately 0 1 seconds after the start of testing        off  0  The start voltage is not set  factory default setting    on  1  The start voltage is set to 50   of the test voltage           SH TEST LED    READY LED lit 1 Off   blinking   TEST LED lit    Output on       0 1s Voltage rise time    Voltage fall time  Fall Time     Separate from the test time  you can set the time that the TOS5200 takes to lower the voltage  after a withstanding voltage test ends  This setting is used only when a PASS judgment  occurs        off  0  The output voltage is shut off imme
25. Voltage rise time 0 1 sto 10 05  Resolution 0 15  Voltage fall time 0 1 s  OFF  only enabled when a PASS judgment occurs   Test time 0 1 sto 999 s  can be turned off  TIMER OFF   Resolution 0 1 s to 99 9 s  0 1 s  100 s to 999 s  1 s   Accuracy    100 ppm   20 ms  2   Q  1 Regarding the output time limits  Ka  Taking size  weight  and cost into consideration  the heat dissipation capability of the voltage generator that is used for withstanding 2  voltage tests has been designed to be one half that of the rated output  Use the TOS5200 within the following limits  If you use the    Oo  product in a manner that exceeds these limits  the output section may overheat  and the internal protection circuits may be activat  2  ed  If this happens  stop testing  and wait until the TOS5200 returns to its normal temperature   dp   Ambient temperature Upper limit Pause time Output time  t lt 40   C 50 mA  lt i  lt  110 mA Greater than or equal to the output time 30 min  max   i  lt  50 mA Not necessary Continuous output possible   Output time   voltage rise time   test time   voltage fall time   2 Regarding the test voltage waveform   Waveform distortions may occur if an DUT whose capacitance is dependent on voltage  for example  an DUT that consists of ce   ramic capacitors  is connected as the load  However  if the test voltage is 1 5 kV  the effect of a capacitance of 1000 pF or less can  be ignored   Because the product   s high voltage power supply uses the PWM switching method  if t
26. aks         Pre Test Inspection          e What do I need to keep in mind to use  the TOS5200 for a long time without  malfunctions      gt   To Use the Product for a Long Time  Free of Malfunctions             Component Names    Front Panel               IKIKUSUI TOS5200 3          OUTPUT    RMS AVE CONFIG MORE CAL RMT ei       LOWER W COMP          FUNCTION       VOLTAGE   CURRENT   TIMER                                  HIGH VOLTAGE    CONFIG                DANGER  A    Es CA    5  e EEN pee  e E MENON  E          SET UPR LWR TEST RISE   MORE    ON PFF   KEYLOCN  LOCAL                      7 8 Testconditions 9 10 11 12 13 1415161718 19    operating unit       The protection cover is  attached to the TOS5200  when the product is  shipped from the factory            REMOTE USB    g Oo  T                            20 Loosen the screw and pull this part towards you  to remove the cover   No  Name Function  1 DANGER LED Lights in red when a test is being performed  p 48  READY  Lights in white when the TOS5200 is ready to perform a test  p 48  TEST  Lights in red when a test is being performed  p 48  2   Status indicators      PASS  Lights in green when a test passes  p 51  FAIL  Lights in yellow when a test fails  p 51  3 Display Displays the settings  measured values  and other information  p 32  LU  e  10   TOS5200  o                                                                      No  Name Function   See    Press these keys to display the settings that are saved to mem
27. anenennnenneananaa 70  Over Heat Protection                           immnmnennni 74  Over Load Protection                                mumereanaa 74  Over Rating Protection      74  P  panel MOMONSY 1 isii tiimmvevatteraabkvdva aada t  nava uknatv atu ava 73  panel memory  saving                            nnmnenennneee 37  PASS  Sue 40  47  PASS Hold WE 45  power cord  ET 20  KONEN E E T 47  Power Supply Protection                                mea 74  power  turning ON                   mminerennaenenananenenennannanaa 22  PIOLEGUONM  eege ebe Eege eege 47  PROTECTION mode      74  protection mode parni 74  R  rack mount adapter                             mmnenennnneenannee 16  Ready eege daki E kat Maa TAA ja G Eege 47  Remote Protection                           mmmnerenenenenananenaa 74  RISE eege ees 32  RISE TIME desen 41  rise time control                              ssrnennennnaenennaa 32  41  S  SIGNAL I O connector                             mumeneneeanaa 54  SIGNAL I O Protection      74  Start Voltage                              imeeneneaneeannenaneeanna 43  T  talk MOU E 36  TEST eieiei eit Wie ee 47  kost PIODG E Senge ieed deeded apa aie 17  26  tast TPIS erreia tive Miedesatu ves ain chbeccebiesdncnsdouanseindestends 51  tast  SEAMING creo ccestsecenicdacetesdacsastcvescneasetneceensaceetsoeneaia  48  troubleshooting                           mremnnsenanme 76  U  EEA EE E E 40  Upper Fall ticceccsetcnscutctedscceadec cdddvate var ahav  eka vana taadu 
28. angerous  Never come   close to these parts during testing  LS   If connections are incomplete  the entire DUT may be   charged to a high voltage  This is dangerous  so be sure   to connect the DUT correctly     Be sure to connect the low voltage test lead  black  first     1 Check that the POWER switch is off and that the DANGER LED is off           KKIKUsUI TOS5200                      POWER             A       OUTPUT     Off Off    A DANGER Ae           d Check that there are no tears or breaks in the test lead insulation          p  62    Raise the front panel LOW VOLTAGE terminal   s cable lock  and then  connect the low voltage test lead  black      E    Raise the cable lock     74  AA  VA Ray Low voltage test lead  black         4 Lower the cable lock to secure the lead in place        24 TOS5200    Connecting to the Device under Test  DUT        TOS5200    Connect the low voltage test lead  black  to the DUT     Connect the high voltage test lead  red  to the DUT     a 63 071    Connect the high voltage test lead  red  to the front panel HIGH VOLT   AGE terminal        8 Check that the DANGER LED is turned off             KIKUSUI TOS5200                      OUTPUT     Off    A DANGER Ae    Reducing the effect of noise    Noise may be generated if the outputs are shorted or if the DUT insulation is damaged  Elec   tronic devices in the surrounding area may malfunction due to the effect of this noise  To  reduce the effect of noise  connect a toroidal core or a resisto
29. apses  when TIMER is set to ON    b  If a current greater than or equal to the upper limit  U FAIL  or less than or equal to the    lower limit  L FAIL  is measured   c  If you press STOP     After a test finishes  the DANGER LED turns off  and the high voltage output is turned off   If a test finishes under condition a or b given above  the judgment result is displayed on the    screen           PROTECTION       e zm TEST  KS  DU fw ELL A Sma ds    Measured voltage        Measured current Test time    If the test time  Timer  has been set    Example  Display of a PASS judgment  after testing finished    After the test time elapses  the PASS LED lights in green  and the test finishes  The PASS  LED remains lit in green for the length of time specified by Pass Hold  The READY LED then  lights in light white  and the TOS5200 switches to the READY state     READY PROTECTION   P    m 50Hz       For details on how measured values are judged  see    About Judgment      The operations that the TOS5200 performs after testing finishes for each judgment condition    are shown below              Operation PASS U FAIL L FAIL  Display PASS LED lights in green    FAIL LED  yellow  and FAIL LED  yellow  and  Measured value remains    UPPER    lights     LOWER    lights   displayed for the length of  time specified by Pass  Hold   Buzzer Sounds for 50 ms   Sounds until STOP is pressed   SIGNAL UO The PASS signal is gener  The U FAIL signal is gener  The L FAIL signal is gener   connector ate
30. ations are not being Is    KEY LOCK Er displayed  or blinking  in the upper right of the screen  p  36    registered  Release the key lock        Is a    RMT    displayed on the screen   The product is being controlled remotely  To control the TOS5200 from the      panel  press the LOCAL key to switch to local mode        The TOS5200 does not switch to Was a local lockout  LLO  command sent through the communication inter   local mode even when I press the face       LOCAL key  Use a communication command to clear LLO command        Unable to perform measurements             Symptom Items to check and possible remedy  The measured values are abnormal  Are the test leads connected to the correct terminals  24  Connect the test leads properly  p   e  o  76   TOS5200  o    Troubleshooting       PROTECTION mode cannot be cleared                         Symptom Items to check and possible remedy  PROTECTION mode cannot be Is an interlock signal being applied  58  cleared even if   press STOP and Stop the interlock signal input  E  EES Is the internal temperature too high   Confirm that the internal temperature of the TOS5200 has decreased  and then      clear the mode   Is the REMOTE connector connected properly  11  Check the REMOTE connector  P   During remote control operation  did the USB or RS232C connector become  disconnected or are there problems with the connector   Check the USB or RS232C connector   Unable to use remote control  Symptom Items to check and possible remed
31. below     Status indicators       switch on     Status indicators    These display the status     T  UPPER    Test condition display Icon area          pus AVE       LOWER W COMP         Data input                            Display Description  READY Lights when the TOS5200 is ready to perform a test  TEST Lights when a test is being performed  PASS Lights when a test result is success  Lights when a test result is fail          Lights when a protection function is activated       Test condition display    This displays test conditions                                               Display Description   RMS Lights when the measurement mode is true rms response  AVE Lights when the measurement mode is mean value response  CONFIG Lights when CONFIG items are displayed   MORE Lights when items other than CONFIG items are displayed  CAL Lights during calibration   RMT Lights when the TOS5200 is being remotely controlled  LIMIT Lights when a limit voltage is set   50 Hz Lights when the frequency is 50 Hz   60 Hz Lights when the frequency is 60 Hz   UPPER Lights when the ammeter is displaying the upper limit   LOWER Lights when the ammeter is displaying the lower limit    W COMP Lights during lower limit judgment   RISE Lights when the voltage rise time is displayed   TEST Lights when the test time is displayed       TOS5200    Parts of the Screen       TOS5200    Icon area    The TOS5200 status is displayed using icons or characters     Display icon Status             See   ep Mod
32. c settings that are  shown on the panel  what test  conditions can be set      gt     Test conditions  additional items         gt     CONFIG Settings             How do   use each of the two test leads      gt     Using test leads    p 24    How are measured values judged      gt     About Judgment               want to check that measurements are  being performed correctly before   start  testing      gt     Pre Test Inspection    p 62    How does the TOS5200 display the  judgment results      gt     When a test finishes             I want to know more about the interlock  feature      gt     Interlock Feature       How can   hold the PASS judgment  result        Judgment result hold time  Pass Hold     p 45             TOS5200    Operation    Maintenance          How do I save the present test conditions  and use them later      gt     Panel Memory    p 37       How do   start tests with a greater level of  safety      gt     Double action feature    p 45       How do   prevent unintentional key  operations from changing the test  conditions      gt     Using the Protection Cover        gt     Locking panel operations  key lock           How do   control the TOS5200 using  external signal input      gt     SIGNAL I O Connector          How do   release the interlock feature      gt     Turing the Power On          Troubleshooting    See    Troubleshooting    on page 76     TOS5200    e How do I check that the test leads are  not damaged  for example  that they  have no bre
33. d for the length of time ated until the FAIL judgment ated until the FAIL judgment    specified by the Pass Hold is cleared     setting     is cleared        1 The length of time that the buzzer sounds when a PASS judgment occurs is fixed to 50 ms  The    specified Pass Hold time has no effect     Withstanding Voltage Test E    51    52    Finishing a Test       To clear judgment results    Press STOP to switch the TOS5200 to the READY state  the READY LED lights      TOS5200       External Control    This chapter explains how to use the SIG   NAL I O connector to externally start tests  and recall panel memory entries and  sequence programs     SIGNAL I O Connector       uN Risk of electric shock  Ensure that all devices are off before you connect or discon   WARNING  nect cables between them     AC AUTION Risk of damage to the internal circuitry   e Keep the signal wire at least 500 mm away from the high voltage test lead and the DUT     Do not short the output voltage circuit     e If you use the TOS5200 with incomplete connections  burn outs caused by heating may  occur when the output is turned on     e Do not touch the contacts or attach insulators to them  Doing so may reduce the quality of  the contacts or cause other problems     e Do not short the  24 V of pin 24 to the chassis or the circuit common     The SIGNAL I O connector is the D sub 25 pin connector on the rear panel     This connector is used to start and stop tests  recall panel memory entries and test mod
34. d voltage rise time   15 p 35  p 41  ON  OFF key  SHIFT key   Turns the test time  Test Time  on and off   Communica   16 USB port USB interface connector  tion interface  manual  i  LOCAL key Switches between local mode and remote mode       SHIFT key Press to access the features that are written in blue       Press to select additional test condition settings   18 MORE Key Frequency  start voltage  voltage fall time  and measurement mode SES  KEY LOCK key  SHIFT key   Locks panel key operations  settings and changes   p 36  19 POWER switch Turns the power on      and off   O    p 22  20 Protection cover Cover designed to prevent incorrect operation of the TOS5200  p 21  1 This indicates an operation that requires you to press a key while holding down SHIFT   e  e  TOS5200 e 11  LI    12    Rear Panel         AWARNING S    DO NOT REMOVE COVERS  NO OPERATOR SERVICEABLE  PARTS INSIDE   REFER SERVICING TO QUALIFIED SERVICE PERSONNEL   UNPLUG THE MAINS SUPPLY CORD BEFORE HANDLING  AC INPUT THE CORD OR LOAD WIRES    IROD   SCY  400 240V KURDA IE  Wl DUSCHEN   50 60Hz 800VA MAX        ARO AAR IS SAL ANE te     EAT DIT EHL TCHS   ZER t   UEL  BTEBRISTEROT  DDTOTHERSL     Ko ce OO        STATUS OUT    conf    STRIP GAUGE       AWG 10  His ey OO         RS232C SIGNAL I O                                                                      KIKUSUI ELECTRONICS CORP  yo  ES  A 5  No  Name Function  1 Chassis terminal Functional ground terminal     2 AC LINE connector AC inlet p 20  3 Ai
35. default settings      The product stores the settings that are in use before it is turned off  so the next time that the  POWER switch is turned on  the TOS5200 starts with these settings     Turning the POWER switch off      N WARNING    TOS5200    Press the  O  side of the POWER switch to turn the TOS5200 off     The panel settings that were in use immediately before the POWER switch was turned off are  saved  If the POWER switch is turned off immediately after the settings have been changed   the last settings may not be stored     Risk of electric shock     e After you turn the POWER switch off  wait at least 10 seconds before you turn the  POWER switch back on  It is dangerous to do otherwise  because the protective fea   tures of the product may not work effectively    This may cause the product to malfunction  and it may reduce the life of the POWER  switch and internal parts such as the fuses     e Except in an emergency  do not turn the POWER switch off while output is being  generated     Installation and Preparation    KIE    Connecting to the Device under Test  DUT     AA WARNING Risk of electric shock  During testing  while the TEST LED or DANGER LED is lit    never touch the HIGH VOLTAGE terminal  test leads  and DUT     Using test leads    ZN WARNING Risk ofelectric shock  Never come close to this area  Parts of the included test leads near the alligator clips during testing    protrude from the vinyl insulation when the wires are   connected  These parts are d
36. diately after a test ends with a PASS  judgment  factory default setting     on  1  The output voltage falls over approximately 0 1 seconds after a test ends  with a PASS judgment           Measurement mode  Measurement     Selects the voltage current measurement mode        RMS  0  Measurement is performed using true rms response  factory default setting    AVE  1  Measurement is performed using mean value response              The measurement mode on Kikusui TOS9200 series  TOS8870A  TOS8850  TOS8850A   TOS5000 series  and TOS5000A series is mean value response     NOTE       TOS5200    Withstanding Voltage Test Q    44    CONFIG Settings    FUNCTION       CONFIG settings are provided on the CONFIG   screen and CONFIG   screen                             CONFIG 1 CONFIG 2  Double Action Double action feature H V ON Output when there is residual volt   age or during testing  Momentary Momentary feature Test Output during testing when the  test voltage is at the set value  Fail Mode Fail mode feature Pass PASS judgment output  Pass Hold PASS judgment result hold Upper Fail Output when upper limit judgment  time is FAIL  Pass Buzzer volume level for PASS Lower Fail Output when lower limit judgment  judgment is FAIL  Fail Buzzer volume level for FAIL Ready Output when READY is displayed  judgment  Baudrate RS232C baud rate Protection Output during PROTECTION  mode  Talk Mode Talk mode Power ON Output when the POWER switch    is on       1 Press CONFIG  SHIFT FUNCTION  to displa
37. displayed  and the icon corresponding to the memory num   ber is displayed in the icon area  The TOS5200 is now set to the recalled test condi   tions     Example  The test conditions  that were saved in MEMORY  2 have been recalled        If you change the recalled test conditions  the memory number disappears  Even if you set  the test conditions back to the values that are stored in memory  the memory number will not  reappear     TOS5200       Withstanding Voltage Test    This chapter explains withstanding voltage  test  from how to set the test conditions to  how to save test results     40    About Judgment       The TOS5200 judges whether a test results in PASS  L FAIL  or U FAIL on the basis of the  limits that are set in advance        Judgment result Description Display Buzzer   PASS When the test time elapses  TIMER is 0 sec  PASS LED lights in Sounds for  onds   if the condition    lower limit  lt  meas  green  50 ms    ured value  lt  upper limit    is true  the Measured value    judgment result is PASS  and the test ends    remains displayed for  the length of time speci   fied by Pass Hold           L FAIL If the condition    lower limit 2 measured FAIL LED  yellow  and Sounds  value    is true  the judgment result is    LOWER    lights  until STOP  LOWER  and the test is immediately is pressed   stopped    U FAIL If the condition    upper limit  lt  measured FAIL LED  yellow  and    value    is true  the judgment result is UPPER     UPPER    lights   and the t
38. e power cord is connected to a properly grounded outlet    e There is no damage such as tears or breaks in the test lead insulation     e When the POWER switch is turned on  the DANGER LED and the status indicators  light     e During testing  do not touch the items that are charged to a high voltage  the DUT   the test leads  and the areas near the output terminals     e During testing  do not turn the POWER switch off except in an emergency     Testing Precautions    During testing  the TEST and DANGER LEDs light  When these LEDs are lit  the TOS5200 is  generating a high voltage  During testing  be sure to wear rubber gloves for electrical work  If  obtaining these gloves is difficult  contact your Kikusui agent or distributor                                                     TT Late   e Lit   The TEST and DANGER LEDs light     A AS The TEST LED lights in red     TOS5200    Remote Control Precautions       When you are controlling the TOS5200 remotely  external signals are used to turn the high  voltage on and off  To prevent accidents  follow the safety measures given below     e Make sure that high voltages are not generated unintentionally     e Make it impossible to touch the DUT  test leads  test probes  and the areas near the  output terminals when high voltages are being generated     Interrupting Testing or Operations    zer Before you change test conditions or other settings  press the STOP switch  and then be sure  to check the following items to ensure safety 
39. el RC01 TOS RC02 TOS remote control box connected p  16  A Keys locked p  36  0068 Recalled memory number or CONFIG setup screen number p  37       Data input area    This area is used to enter the test voltage  limit voltage  upper and lower current limits  test  time  CONFIG values  and additional test items     The selected item blinks     READY PROTECTION P    TEST    TEST    6 55    The selected item  blinking        Panel Operation Basics D    33    ee    Panel Operations       Switching screens    On the TOS5200  you need to change the setup screen depending on the item you want to    d  Memo set   erer From the basic setup screen  you can switch to the additional test conditions setup screen or  POWER switch is turned CONFIG setup screen     on is referred to as the  basic setup screen     To display the setup screen for additional test conditions  press MORE     To display the CONFIG setup screen  press CONFIG  SHIFT FUNCTION    Press once to show the CONFIG   screen and twice to show the CONFIG   screen        To return to the basic setup screen  press STOP     Additional test conditions setup screen    Basic setup screen   MORE    RMS  READY   PROreG TION key        GaSe 7  CONFIG  SHIFT FUNCTION  key   CONFIG setup screen          CONFIG  SHIFT FUNCTION  key   CONFIG setup screen         To return to the basic setup  screen  press STOP           To recall a memory entry  press MEMORY 1  MEMORY 2  or MEMORY 3  To save  press  Recall while holding down SHIFT        
40. ergency ET 29  Sales Eu 30   Turning the power on and off repeatedly    AAA 30  About Malfunctions ENEE    30  To Use the Product for a Long Time Free of Malfunctions EE 30    Panel Operation Basics    Parts Of the Green  32  Panel Operations ESSEN E ud SSS iaaea 34  Switching Screens                              mernrennnenanenennnennnnnnunanaaennnnannaennneananannenannnannae 34  Selecting JAIL    21 PS AT A 35  Entering VAIWCS ai 35  Locking panel operations  key Joch   36  TAK TMOG E 36  Panel MEMORY sssisesscsscsnssscassonaseasioressaaneconsasiszvqsaveoconsusiusbunnasecs senaisspsnshnsa seconsssaistanssdaecdouipeastovnsvasdoredsgnnesnssaaislvgvivescs  staies 37    saving ere e E 37  Recalling test CONGITIONS EENEG 38    5 Withstanding Voltage Test    lee ee e Tu E 40  Effectiveness of the lower mt    40  Invalid Seting E 40  Setting Test COMAIUOMS  saccscascsssnsssssasacancsinnsasbessvosecsnnnansessonteacohnssasestcnduovsnusscegsdotsbaussoveascoduonsansagvencandieisissagneaedoseesisdine 41  Test conditions  basic Hemes   41  Test conditions  additional eme   42  CONFIG Se tei 1 4 tiiniktvsiv aired E naha aaasta njaa seas ETA i    CONFIG 1 settings  CONFIG 2 settings       Starting CR  Tostat Atesten e eea aaa  aieastetsasescceuvesdeasteetiny  When  artest starts sereni a a Main hota i eaea  To change the voltage setting during a Test    49  If you cannot Start testingan irinenn 49   Finishing  a TSS AAA EMASE ASUTADA 51  eeler 51  Whena test Un EE 51  To clear judgment re
41. es   and monitor the status of the TOS5200       Connector on the rear panel    XM3B 2522 D sub 25 pin female connector  socket   M2 6  x 0 45 screws  manufactured by OMRON Corporation      Compatible connector  plug   D sub 25 pin male  with fix screw M2 6     DO NOT REMOVE COVERS  NO OPERATOR SERVICEABLE  PARTS INSIDE   REFER SERVICING TO QUALIFIED SERVICE PERSONNEL     SIGNAL I O connector  OMRON corporation   XM3B 2522    RS2320 SIGNAL Si                      To avoid malfunctions caused by noise  use shielded D sub  E  25 pin connectors and a cable that is 2 5 m or less in length     For information about how to obtain replacement parts  contact your Kikusui agent or distrib   utor     For information about how to use these components  see the OMRON Corporation catalogs     Wire and tool that are necessary to make the connection       Wire Single wire  0 32 mm  AWG28  to 0 65 mm  AWG22  in diameter       Twisted wire  0 32 mm   AWG22  to 0 08 mm   AWG28   Wire stripper A wire stripper that matches the wires listed above          SIGNAL I O specifications                   Input signal Output signal  Low active control input Open collector output  High level input voltage 11Vto15V Output withstanding voltage 30 Vdc  Low level input voltage OVto4V Output saturation voltage Approx  1 1 V  at 25   C  Low level input current 5 mA max   Maximum output current 400 mA total  Input time width 5 ms minimum       1 Excluding the interlock signal    54 TOS5200    SIGNAL I O Connecto
42. ess STOP   888 SIGNAL I O Protection The SIGNAL I O connector s ENABLE Press STOP       signal has changed   gE USB Protection During remote control operation  the Check the USB connector  and then  D    USB connector was disconnected  or press STOP   a defect was detected        If you cannot clear the PROTECTION mode even after you have corrected all the causes  the  TOS5200 may be malfunctioning     Stop using the product  and contact your Kikusui agent or distributor      TOS5200    C Test Start Operation and Display    TOS5200       The following table shows the control sources that are enabled depending on the combination  of remote control sources                                               Combination Enabled remote Indicator on the  control source screen  Front panel Rear panel SIG    USB or RS232C    RMT    ED  REMOTE NAL I O connec   connector tor  the RR  ENABLE signal  is at low level   a             Panel Off Off  REMOTE connec   Off On  b Yes          tor  SIGNAL I O con  Off On  c     Yes      nector  d _ _ Yes USB connector or   On Off  RS232C connector  SIGNAL I O con  Off On  e Yes Yes      nector                75    D Troubleshooting       This section introduces troubleshooting measures  Typical symptoms are listed  Check whether any  of the items listed below apply to your case  In some cases  the problem can be solved guite easily     p  72 If none of the items apply to your case  we recommend that you initialize the TOS5200 to its    factory default
43. est Inspection 62    Specifications 63    Appendix 71   List of Default Settings 72  Protection Functions 74   Test Start Operation and Display 75  Troubleshooting 76    About the Manuals    These manuals are intended for users of the Withstanding  Voltage Tester and their instructors  Explanations are given  under the presumption that the reader has knowledge related  to electricity     Manual construction    H Setup Guide    This manual is intended for first time users of the product  It  gives an overview of the product  connecting procedures   safety precautions  etc  Please read this manual before you  operate the product     H Quick Reference    The quick reference briefly explains the control panel and  the basic operation of it     H Safety Information    This document contains general safety precautions for this  product  Keep them in mind and make sure to observe  them     H User   s Manual  this manual  PDF     This manual is intended for first time users of this product  It  provides an overview of the product  notes on usage  and  specifications  It also explains how to connect the product   configure the product  operate the product  perform mainte   nance on the product  and so on     H Communication Interface Manual  PDF     This manual explains how to control the product remotely  using commands  It is included on the CD ROM    The interface manual is written for readers with sufficient  basic knowledge of how to control measuring instruments  using a PC     
44. est is immediately stopped        1 The length of time that the buzzer sounds when a PASS judgment occurs is fixed to 50 ms  The  specified Pass Hold time has no effect     Effectiveness of the lower limit    Normally  even a good DUT will have a certain degree of leakage current  Setting the upper  limits just within the boundaries of the characteristic range of the DUT is useful in detecting  breaks in the test leads and poor connections  enabling you to perform highly reliable testing     In withstanding voltage tests  you can perform tests effectively by enabling the lower limit with  W COMP     Invalid Settings    If you specify an invalid value for a setting  a message blinks on the screen  While a message  is blinking  the READY LED turns off  and you cannot start a test  You also cannot select  other items    Specify a valid value     READY LED turns off _ LIMIT blinks    ER a Ri dl L g  LIMIT      VI TEST         Example of LIMIT indicator  blinking    Message Description       mA indicator blinking When the lower or upper limit is turned on  and the upper limit has been set  lower than the lower limit        LIMIT indicator blinking When you try to setthe test voltage higher than the set limit voltage        kV and mA indicators When the test voltage and the upper limit are set in the withstanding volt   blinking simultaneously age test settings such that their product exceeds 550 VA        TOS5200    Setting Test Conditions    Test condition       Panel operation  
45. h that has a voltage         13  rating of 30 Vdc or greater and a current rating of 10 mA or  INTERLOCK greater     Using a transistor or optical device    dI    Photocoupler    a 11  such as a transistor When the transistor collector current is 5 mA or less  the  i interlock feature is activated  To release the interlock fea   ture  allow an ic of 6 mA or greater to flow  and then press  STOP or apply a STOP signal        STATUS OUT connector       TOS5200    This is the output connector for connecting the optional warning light unit  PLO2 TOS     In CONFIG setting 2  select the status that you want to output  Select H V ON  Test  Pass   Upper Fail  Lower Fail  Ready  Protection  or Power ON  If you select multiple items  the sta   tus that is generated will be the logical sum of the items  When the selected status is true  the  TOS5200 generates a  24 Vdc signal     For details  see the    WARNING LIGHT PL02 TOS OPERATION MANUAL        External Control Q    59    60    This page is intentionally blank     TOS5200       Maintenance    This chapter explains daily inspections  such as measurement checks     62    Pre Test Inspection       Inspection of test leads and the judgment feature           p  41  p  42    Calibration    Allow a current to flow between the HIGH VOLTAGE and LOW VOLTAGE terminals to check  the judgment feature  During this inspection  the two test leads are shorted  so you can check  for breaks in the test leads at the same time    We recommend that you 
46. hat is recalled is shown in the table below     PMO PM1 TEST SEL AUTO SEL Recalled test mode  L L L    ACW          Interlock Feature       This feature links the TOS5200 to an external device to stop output appropriately  This  ensures the safety of the operator    While the interlock feature is active  even if you press the START switch or apply a start sig   nal from an external controller  the TOS5200 will not perform testing  While the interlock sig   nal is being applied  you cannot release PROTECTION mode by pressing the STOP switch  or applying a stop signal    By using the interlock feature  you can control the TOS5200 output from an external source   This ensures safe operation of the tester    The first time that you turn the POWER switch on after you purchase the TOS5200  the tester  will be in PROTECTION mode through the interlock feature  You can use the included SIG   NAL I O plug to easily release the PROTECTION mode    Attaching the included SIGNAL I O plug connects pin numbers 1 and 13  the INTERLOCK   and INTERLOCK  pins  Only use this plug to easily release the PROTECTION mode    When you are actually performing tests  use the interlock feature to ensure safety    During withstanding voltage test in which you are using tools   1  placing a cover over the  DUT so that output is turned off whenever the cover is removed to prevent electric shock and   2  placing a safety fence around the work area where withstanding voltage test is being per   formed so that 
47. he products bottom  panel  This is useful in preventing  the cover from being lost        Use the screw to fix the cover to  the product   s bottom panel     Installation and gel      21    22    Turning the Power On       Checking indicators and the status of the interlock feature          The first time that you turn the POWER switch on after you purchase the TOS5200  the tester  will be in PROTECTION mode through the interlock feature  Connect the included SIGNAL I   O plug to the SIGNAL I O connector to release the interlock feature    Only use the included SIGNAL I O plug to easily release the PROTECTION mode    When you are actually performing tests  use the interlock feature to ensure safety    During withstanding voltage test in which you are using tools   1  placing a cover over the  DUT so that output is turned off whenever the cover is removed to prevent electric shock and   2  placing a safety fence around the work area where withstanding voltage test is being per   formed so that output is turned off whenever the fence is opened are both examples of effec   tive safety measures        NOTE    When the TOS5200 is turned on  a self test is run  and all the indicators on the front panel  light  To ensure safety  check that all the indicators light before you use the TOS5200  It is  especially dangerous to use the tester if its DANGER LED is broken     When the power is turned on  the DANGER LED lights  but no voltage is generated         See  p  20    Check that the
48. he test voltage is 500 V or less  the switching  and spike noise proportions are large  The lower the test voltage  the greater the waveform is distorted   3 In either case  true rms or mean value  a response time of at least 50 ms is required to meet the measurement accuracy   4 Regarding ammeter and judgment accuracy   During withstanding voltage tests  current also flows in the stray capacitance of items such as the measurement leads and jigs   This current that flows in the stray capacitances is added to the current that flows in the DUT  and the sum of these currents is  measured  Especially if you want to perform judgments with high sensitivity and accuracy  it is necessary to consider methods to  limit the current that flows in these stray capacitances  such as by adding upper and lower limits   Output voltage 1 kV 2 kV 5 kV  When using 350 mm long test leads that are suspended in air  TYP  2 UA 4 pA 10 pA  When using the accessory  high voltage test lead TL31 TOS  TYP  16 pA 32 HA 80 HA  When the humidity is 70   or higher  add 50 pA   o      TOS5200 2 65  o    66    Specifications       Other features       TOS5200       Test mode    Double action feature    Tests can only be started by pressing and releasing STOP and then pressing START  within 0 5 seconds of releasing the STOP switch        Length of time to hold a PASS  judgment result    You can set the length of time to hold a PASS judgment  50 ms  100 ms  200 ms  1 s  2s   5 s  or HOLD        Momentary feature 
49. ications of this product and the contents of this man   ual are subject to change without prior notice        2014 Kikusui Electronics Corporation    Checking the Package Contents    When you receive the product  check that all accessories are  included and that the accessories have not been damaged  during transportation    If any of the accessories are damaged or missing  contact  your Kikusui agent or distributor     We recommend that you save all packing materials  in case  the product needs to be transported at a later date     ACN    Rating  125 Vac  10 A Rating  250 Vac  10 A Rating  250 Vac  10 A  Plug  NEMA5 15 Plug  CEE7 7 Plug  GB1002   85 AA 0003   85 AA 0005   85 10 0790                 Power cord  1 pc     The power cord that is provided varies depending on the             destination for the product at the factory shipment                                               High voltage test lead High voltage warning SIGNAL I O   1 set  sticker  1 pc   plug  1 set    TL31 TOS   A8 210 203  Assembly type     D sub plug unit      lt  gt              i Japanese  1 sheet    Z1 006 150   English  1 sheet    Z1 006 152   CD ROM  1 pc   Quick Reference                   CP  lt  gt                 Setup Guide  1 pc            Safety Information  1 pc             Z1 006 140   Z1 005 040     TOS5200    Notes to the supervisor    If the operators cannot understand the language used in this manual  translate the  manuals into the appropriate language     Make sure that the opera
50. imit 0 01 mA to 110 MA  Lower limit OFF  0 01 mA to 110 mA  Test time OFF  0 1 s to 999 s  Voltage rise time 0 1 s to 10 0 s   Voltage fall time OFF  0 1 s   Start voltage OFF  50         Initial panel memory settings    Initially  memory numbers 1 to 3 contain withstanding voltage test conditions  Feel free to use  these test conditions if they match the tests that you want to perform    For details on the test conditions that are initially contained in memory  see    Default panel  memory values        Saving test conditions    Display the basic setup screen  and then set the test conditions     Press MEMORY 1  MEMORY 2  or MEMORY 3 to select the memory num   ber  1 to 3  in which you want to save the current test conditions   The selected memory number blinks     Press SHIFT RECALL to save the conditions in the selected memory  number     READY   PROTECTION    Example of saving to memory  number 1       After the test conditions have been saved  the basic setup screen is displayed in READY  mode  the READY LED lights      Panel Operation a gt     38    Panel Memory       Recalling test conditions    1 Press MEMORY 1  MEMORY 2  or MEMORY 3 to specify the memory  number  1 to 3  from which you want to recall test conditions     MEMORY    i The test conditions that are saved in the memory number  1 to 3  that you specified  C  are displayed        2 6  2 Press RECALL to recall the test conditions that are saved in the speci     fied memory number    The basic setup screen is 
51. irrespective of  the settings made by the operator     Immediately turn the POWER switch off  and disconnect the power cord from the outlet  Stop  using the product immediately  and contact your Kikusui agent or distributor     e Even when you press the STOP switch  the DANGER LED remains lit     e During tests  the DANGER LED does not light     To Use the Product for a Long Time Free of Malfunctions    Taking size  weight  and cost into consideration  the heat dissipation capability of the voltage  generator that is used for withstanding voltage tests has been designed to be one half that of  the rated output  Use the TOS5200 within the following limits  If you use the product in a man   ner that exceeds these limits  the output section may overheat  and the internal protection cir   cuits may be activated  If this happens  stop testing  and wait until the TOS5200 returns to its  normal temperature     H Output limits during withstanding voltage testing          Ambient temperature Upper limit Pause time Output time  t lt 40   C SOMA  e up m   Greater than or equal to zo min  max   the output time  i  lt  50 mA Not necessary Continuous output possible        Output time   voltage rise time   test time   voltage fall time     TOS5200       Panel Operation Basics    This chapter describes how to select items  on the screen and how to enter values     32    Parts of the Screen          The screen that is used to set the basic test conditions is made up of the four parts shown  
52. l digital  signal     Starting a Test    ENABLE  IN   a p     o  READY  OUT     5 ms or more    1  I     n h F  START  IN  1      I  La      1 1      To use the SIGNAL I O connector to start a  test  set the ENABLE signal to low level  When  at least 1 ms has elapsed after the READY sig   nal was set to low level  set the START signal  to low level for at least 5 ms  After a valid  START signal has been detected  the READY  signal is set to high level     When the ENABLE signal is low  the START    1 ms or more    signal of the SIGNAL I O connector is enabled   and the panel   s START switch and the START  input of the REMOTE terminal are disabled     S  Recalling Panel Memory and Test Modes    Selecting the panel memory    The PM and STB signals are processed with the timing shown below  Check that the READY  signal is low level     The relationship between the PMO and PM1 signals and the panel memory number that is  recalled is shown in the table below             An interval of 1 ms or more  is required before the next  input signal is received   Apply the START signal for  at least 50 ms     n    STB  IN  1 ms      r     110 ms or more  ore          TOS5200    External Control Q    57    58    Interlock Feature                   PMO PM1 TEST SEL AUTOSEL Recalled panel memory number  H H L L Memory 1   L H L L Memory 2   H L L L Memory 3   L L L L           Selecting the test mode    The relationship between the PMO  PM1  TEST SEL  and AUTO SEL signals and the test  mode t
53. lly radio frequency energy  in the form of  electromagnetic radiation  inductive and or capacitive coupling  for the treatment of material or inspection analysis purpose    4 This product confirms to Class    Be sure to ground the protective conductor terminal of this product  If not grounded properly  safe   ty is not guaranteed     TOS5200    Specifications Q    70    Specifications       Outline drawing                                                                            A  a a    A d ZER 5  Sc   Se  65 Sa    250  9 84   F  wn         SIS A  o     9   q pS     265  10 43   r    TN  TN  e  TT Hes    gt i    Le Dal       SS  y  Ol  NO  aS MAX330  12 99        320  12 60  ji  ZS  el s  DI o  x lt  mM  GK                                     Unit  mm  inch     TOS5200    Appendix       A List of Default Settings   B Protection Functions   C Test Start Operation and Display  D Troubleshooting    72    A List of Default Settings    Initializing the TOS5200    PROTECTION   PR        Default values    Item       While holding down SHIFT  turn the POWER  switch on     When you initialize the TOS5200  all settings   such as test conditions  and saved data are  reset to the default values shown below     Setting after initialization                                                                                     Withstanding voltage Measurement mode RMS  EREEREHERG Test voltage 0 00 kV  Test voltage   s limit voltage 5 50 kV  Upper limit 0 02 mA  Lower limit OFF  Test time 0 
54. measured current     measured values test time       2    The input signals are all low active control  The input terminal is pulled up to  12 V by a resistor  Leaving the input terminal open  is equivalent to applying a high level signal     TOS5200       Specifications       General       TOS5200       Display    LCD  LED backlight custom design       Environment    Installation location    Indoors  at a height of up to 2000 m                   Spec guaranteed   Temperature 5   Cto 35   C  41   F to 95   F    range Humidity 20   rh to 80   rh  no condensation    Operating range Temperature 0   Cto 40   C  32 F to 104   F   Humidity 20   rh to 80   rh  no condensation    Storage range Temperature  20   C to 70   C   4   F to 158   F   Humidity 90   rh or less  no condensation        Power supply       Nominal voltage ran  range    Power  consumption    ge  allowable voltage    When no load is  connected  READY     100 Vac to 240 Vac  90 Vac to 250 Vac     100 VA or less       Rated load       Allowable frequency range    800 VA max   47 Hz to 63 Hz       Insulation resistance  between AC LINE and the chassis     30 MO or more  500 Vdc        Withstanding voltage  between AC LINE and the chassis     1500 Vac  1 minute       Earth continuity    25 Aac  0 1 Q or less       Electromagnetic compatibility  EMC  1    Complies with the requirements of the following directive and stand   ards   EMC Directive 2004 108 EC  EN 61326 1  Class A    EN 55011  Class A  Group 19   EN 61000 3
55. mode because of several causes  the cause  with the highest priority is displayed  After you eliminate the cause  the cause with the next  highest priority is displayed to indicate that there are more causes     If you cannot clear the mode even by pressing STOP  turn the TOS5200 off and then back on     Display example when an  interlock signal is detected                                  Panel display Protection circuit Description Remedy  E Interlock Protection An interlock signal was detected  Remove the interlock signal input       and then press STOP   pr Power Supply Protection An error was detected in the power You can clear the PROTECTION  e supply  mode by pressing STOP  but if the  protection occurs frequently  repairs  are necessary   8886 Volt Error Protection An output voltage outside of the rated Press STOP   LOL limits was detected   Withstanding voltage test   350 V  1p Over Load Protection An output voltage causing the output   Press STOP   om power limit to be executed was  detected   Withstanding voltage test  550 VA  Hp Over Heat Protection The internal temperature of the Confirm that the internal temperature  e TOS5200 became too high  of the TOS5200 has decreased  and  then press STOP   Ap Over Rating Protection The output current was generated for Press STOP   em a length of time that exceeds the regu   lated time   B8 Remote Protection A connection to or disconnection from Check the REMOTE connector  and      the REMOTE connector was detected  then pr
56. nect the protec   tive conductor terminal to earth ground        NOTE e Use the supplied power cord to connect to the AC line   If the supplied power cord cannot be used because the rated voltage or the plug shape is  incompatible  have a qualified engineer replace it with an appropriate power cord that is 3  m or less in length  If obtaining a power cord is difficult  contact your Kikusui agent or dis   tributor     e The power cord with a plug can be used to disconnect the product from the AC power line  in an emergency  Connect the plug to an easily accessible power outlet so that the plug can  be removed from the outlet at any time  Be sure to provide adeguate clearance around the  power outlet       Do not use the supplied power cord with other instruments        This product conforms to IEC Overvoltage Category II  energy consuming equipment that is sup   plied from a fixed installation      In addition to the supplied power cord  Kikusui also provides other 200 V power cords with  plugs  sold separately      1 Check that the POWER switch is turned off     2 Check that the AC power line meets the nominal input rating of the prod   uct     The product can receive a nominal power supply voltage in the range of 100 Vac to  240 Vac  90 Vac to 250 Vac  that has a freguency in the range of 47 Hz to 63 Hz     3 Connect the power cord to the rear panel AC inlet  and then connect the  power plug to an outlet that has a ground terminal     TOS5200    Using the Protection Cover   
57. nges  stable tests can  be performed even in locations where the power supply is unstable    The AC inlet is designed for worldwide use  The TOS5200 can be used without modifica   tion provided the nominal power supply voltage is within the range of 100 Vac to 240 Vac   90 Vac to 250 Vac  and the frequency is within the range of 47 Hz to 63 Hz     e Rise time control feature that gradually increases the test voltage    Instead of immediately applying the specified test voltage to the DUT after the test begins   this makes it possible to perform tests in which the voltage is raised gradually to the test  voltage  As required by withstanding voltage tests defined by standards such as IEC and  UL  this makes it possible to perform tests in which no more than half of the test voltage is  applied at the start of the test  and the test voltage is gradually reached over the specified  time     e Fall time control feature that gradually decreases the test voltage    The test voltage can be gradually decreased after a PASS judgment occurs during an AC  withstanding voltage test     e Window comparator feature for setting upper and lower judgment limits    You can set not only the upper limit  but the lower limit as well  This is useful in determin   ing whether there are breaks in test leads or whether there was a mistake during opera   tions  This leads to highly reliable tests     e Ability to save three sets of test conditions  You can save three sets of test conditions     e Impro
58. nt because a value greater  than or egual to the upper limit is detected    TOS5200    Specifications E    67    68    Specifications                                                             TOS5200  SIGNAL UO  continued   18 L FAIL o On continuously when FAIL results from judgment because a value less than or  equal to the lower limit is detected   19 READY O On when the TOS5200 is waiting  when it is in the READY state    20 PROTECTION o On when protective features have activated  the TOS5200 is in the Protection  state    21 START   Start signal input terminal   22 STOP   Stop signal input terminal   23 ENABLE l Start signal enable signal input terminal   24  24V      24 V internal power supply output terminal  maximum output current 100 mA   25 COM     Circuit common terminal   STATUS SIGNAL OUTPUT Output terminal for a warning light     Terminal A  24 V signal is generated from this terminal when output has been turned on   COM  24 V circuit common terminal       1    Talk mode can be set when RS232C is in use     Talk mode Description       0                Responds only to commands from a PC  factory default setting    Automatically responds at the start and end of a test  The TOS5200 status  settings  and measured values are  returned   Response at the  lt START gt   start of a test  Status  lt PASS gt    lt U_FAIL gt    lt L_FAIL gt    lt PROT gt   or  lt ABOUT gt   Response at the  end of a test Settings  Test number  program number  test mode  measured voltage  
59. or the following settings and conditions     The product is warmed up for at least 30 minutes   e Values indicated by    TYP    are typical values  They are not guaranteed performance val     ues       Values indicated by    rdng    are readout values   e Values indicated by    set    are settings     Withstanding voltage tester                            form     TOS5200  AC output   Output range 0 05 kV to 5 00 kV  section Setting    2   of set   20 V   at no load   accuracy  Setting range   0 00 kV to 5 50 kV  Resolution 10 V STEP  Max  rated output  500 VA  5 kV 100 mA   Max  rated voltage 5 kV  Max  rated current 100 mA  output voltage 0 5 kV or higher   Transformer rating 500 VA  Output voltage wave  Sine       Distortion    If the output voltage is 0 5 kV or more  3   or less  when no load or a pure resistive load  is connected         Crest factor    Within   2 3    output voltage 800 V or higher  at no load        Frequency    50 Hz or 60 Hz          Accuracy     0 5    excluding during voltage rise time        Voltage regulation    10   or less  when changing from maximum rated load to no load        Input line regulation     0 3    5 kV at no load  power supply voltage  90 V to 250 V        Short circuit current    200 mA or more  output voltage 1 0 kV or higher           Output method    PWM switching       Start voltage    The voltage at the start of withstanding voltage tests can be set to 50   of the test volt   age        Limit voltage    The test voltage uppe
60. ory   When test conditions or configuration items are being set  these keys corre   spond to the menus displayed on the screen   A MEMORY keys MEMORY 1 key  Displays the settings saved to MEMORY 1  p 37  p 44  MEMORY 2 key  Displays the settings saved to MEMORY 2   MEMORY 3 key  Displays the settings saved to MEMORY 3   RECALL key  Recalls settings from panel memory    SHIFT key   Saves the current settings to panel memory   START switch Starts testing  p 48  6 STOP switch Stops testing and clears the current status  p 51  p 74  7 nap VOLIAGE This terminal is for the high voltage line of the tester output  p 24  terminal  8 Se This terminal is for the low voltage line of the tester output  with cable lock   p 24  FUNCTION key Does not use  WW  g  ACW key   CONFIG key  SHIFT key   Displays the CONFIG setup screen  p 44  10 Rotary knob Changes settings  p 33  p 41     p 44  11 Screw hole Fasten a screw to this hole to fix the protection cover in place       SET key Press to select the voltage setting   12 p 35  p 41  LIMIT key  SHIFT key   Selects the voltage limit setting   UPR  LWR key Press to alternately select the upper and lower current limits   13 p 35  p 41  ON  OFF key  SHIFT key   Turns the lower limit judgment feature on and off   Specialized connector for connecting the optional remote control box  RC01   14 REMOTE connector  7OS RC02 TOS  or the high voltage test probe  HPO1A TOS HPO2A TOS      P 10  P 26  TEST  RISE key Press to alternately select the test time an
61. output is turned off whenever the fence is opened are both examples of effec   tive safety measures     E Interlock signal input conditions    Open across terminals  e When the resistance across the positive and negative terminals is 1 2 KQ or greater     s If you are using transistors or an optical device  when the current across the positive  and negative terminals is 5 mA or less     E Interlock signal release conditions    Short across terminals  e When the resistance across the positive and negative terminals is 1 KQ or less     s If you are using transistors or an optical device  when the current across the positive  and negative terminals is 6 mA or greater     TOS5200    STATUS OUT connector       How to use the interlock feature    When SIGNAL UO connector pins 1 and 13 are opened  the interlock feature is enabled   When the pins are shorted  the interlock feature is released     When the interlock feature is active  the TOS5200 is in PROTECTION mode  To release the  interlock feature  connect the included SIGNAL I O plug to the rear panel SIGNAL I O con   nector  Then press STOP to release PROTECTION mode     Using an open close switch    Door switch or    EE EE other type of switch A   A   111 yP When the contact is open  the interlock signal also            Sa becomes open  and the interlock feature is activated  To     release the interlock feature  close the contact  and then       press STOP or apply a STOP signal        Use a door switch or other type of switc
62. perform this pre test inspection before you use the TOS5200     The test leads are consumable parts  Check them periodically for tears and breaks     Breaks or tears in the insulation may cause electric shock or fire  If a break or tear is  found  stop using the damaged test lead immediately    Short the low voltage test lead  black  and high voltage test lead  red    Connect the low voltage test lead  black  to the LOW VOLTAGE terminal   Connect the high voltage test lead  red  to the HIGH VOLTAGE terminal     Set the test conditions   Be sure to set the upper limit  UPPER      Press START to start a test     oni ROKN      If a test results in a U FAIL judgment  based on this simple check  both  the test leads and the TOS5200 judgment feature are working properly     If a U FAIL result does not occur even after you perform this check multiple times  there may  be breaks in the test leads  Double check whether this is the case  If there are no breaks in  the test leads  the TOS5200 needs to be repaired    To have your TOS5200 repaired  contact your Kikusui agent or distributor     The product is calibrated before shipment  To maintain long term reliable performance  we  recommend periodic calibration     To have your product calibrated  contact your Kikusui distributor or agent     TOS5200       Specifications    This chapter contains the specifications  and gives the dimensions of the TOS5200     64    Specifications       Unless specified otherwise  the specifications are f
63. r       H SIGNAL I O connector pinout       Pin no  Signal name VO   TOS5200  INTERLOCK  If you open the positive and negative terminals  the output is turned off  and the TOS5200 is  switched to Protection mode    Open  The resistance between the two terminals is 1 2 kQ or greater    Short  The resistance between the two terminals is 1 KQ or less                                                                          PMO Panel memory selection signal   2     The selection signal is latched on the ris   PMO_ PM1_  Recalled panel memory number  ing edge of the input strobe signal to H H Memory 1  PMI recall panel memory  L H Memory 2  3       ANJA H L Memory 3  The selection of memory is prioritized  over TEST SEL and AUTO SEL  E E ae  4 NC           5 NC           6 NC           7 NC           8 NC           9 STB l Panel memory strobe signal input terminal   10 TEST SEL l NA  11 AUTO SEL l NA  12 COM       Circuit common terminal   INTERLOCK  If you open the positive and negative terminals  the output is turned off  and the TOS5200 is    switched to Protection mode   Open  The resistance between the two terminals is 1 2 KQ or greater        13 l                                                                                                                                                                                                                                                                                                 Short  The resistance between the two terminal
64. r limit can be set  0 00 kV to 5 50 kV        Output voltage monitor feature    If output voltage exceeds the specified value   350 V or is lower than the specified value    350 V  output is turned off  and protective features are activated                                                              Voltmeter   Digital Measurement   0 000 kV to 6 500 kV  range  Display o 000 kV  Accuracy V  lt  500 V     1 5   of reading   20 V   V2 500 V   1 5   of reading  Response   True rms  Mean value response rms display Can be switched  Hold feature After a test is finished  the measured voltage is held until the PASS or FAIL judgment is  cleared   Ammeter   Digital Measurement   0 00 mA to 110 mA  range  Display i   measured current  i lt  1mA 1mAsi  lt 10mA 10 mAsi lt 100mA   100 mA lt  i  0 000 mA o ooo mA 00 00 MA 000 0 MA  Accuracy    1 00 mA  lt i     1 5   of reading   i  lt  1 00 mA     1 5   of reading   30 pA   Response True rms  Mean value response rms display Can be switched          Hold feature       After a test is finished  the measured current is held until the PASS judgment is cleared           TOS5200    Specifications                                                                                                                      TOS5200  Judgment   Judgment method and  feature judgment operation  Judgment   Judgment method Display Buzzer poe Y  If a current that is greater than or equal to   FAIL LED and Generates  UPPER    the upper limit is detected  the outp
65. r of approximately 470 Q  between the tips of the high  and low voltage test leads and the DUT  Connect the toroidal  core or resistor as close to the DUT as possible     If you are connecting a toroidal core  it is effective to wrap the test leads two to three times  around a type of core that can be snapped on and that is often used with power cables  This  type of core is usually approximately 20 mm in diameter    If you are connecting a resistor  pay close attention to the power rating of the resistor  When  the upper limit is 10 mA or less  connect a resistor of approximately 470 A  3 W  30 kV  impulse withstanding voltage   Because this resistor causes the voltage to fall  the voltage  that is actually applied to the DUT is slightly lower than the voltage that is generated from the  product   s output terminals  when a 10 mA current flows  the voltage falls approximately 10 V      These methods are extremely useful in reducing the effect of noise     Connect the toroidal core or resistor as close to  the DUT as possible                                                  Toroidal core 470 2 3 W  HIGH VOLTAGE HIGH VOLTAGE Oww  terminal terminal  DUT DUT  LOW terminal LOW terminal 5 ml  Toroidal core 4702  3W  TOS5200 TOS5200    Installation and Preparation      26    Connecting to the Device under Test  DUT        Using the optional high voltage test probe   model HP01A TOS HP02A TOS     If you use the optional test probe instead of the test leads  you can use hands on cont
66. r outlet Air outlet for cooling      4 STATUS OUT connector Connector for connecting the optional warning light unit  PLO2 TOS p 17  p 59  5 Serial number The product s serial number      Communication  6 RS232C port RS232C interface connector interface  manual  7 SIGNAL I O connector External control signal connector p 54  1 A multipurpose ground terminal provided for the stable operation of the TOS5200  Use it if necessary   TOS5200       General Description    This chapter gives an overview of the  TOS5200 and explains the options that are  available for it     14      Product Overview       The TOS5200 Withstanding Voltage Tester performs withstanding voltage tests  which is one  of the four types of tests  that are required for ensuring the safety of electrical products     This product can perform AC withstanding voltage tests on electrical products and electrical  components in accordance with the requirements of safety and electrical standards and ordi   nances such as IEC  EN  UL  VDE  and JIS     It is suited to  1  research and development installations   2  test facilities for quality assur   ance testing and standard certification  and  3  manufacturing lines     The withstanding voltage tester is easy to use  safe  and reliable     Features    e Newly developed constant voltage output for stable testing    The TOS5200 is not affected by AC line interference  Because the output voltage is main   tained at a fixed value even if the AC line voltage or frequency cha
67. re enables you to send test results and the like from the product through the  RS232C interface without using commands from a PC  It can reduce processing on the PC        If you turn talk mode on  you cannot control the product from a PC  To control the product  from a PC  turn talk mode off        off  0  Talk mode is not used  factory default setting         on  1  Talk mode is used        CONFIG 2 settings    On the CONFIG 0 setup screen  you can set the output conditions when using the optional  warning light unit  PLO2 TOS     Example  Selecting H V ON  blinking                                   Wd l EE    Power ON setting  0  off  1  on   Protection setting  0  off  1  on   Ready setting  0  off  1  on     Lower Fail setting  0  off  1  on              Upper Fail setting  0  off  1  on              Pass setting  0  off  1  on           v   Test setting  0  off  1  on   v   H V ON setting  0  off  1  on     H V ON    Sets whether the 24 Vdc output is generated while there is a residual voltage and during test   ing        off  0  The 24 Vdc output is turned off while there is a residual voltage and during testing   factory default setting         on  1  The 24 Vdc output is turned on while there is a residual voltage and during testing        46 TOS5200    CONFIG Settings       Test    Sets whether the 24 Vdc output is generated while tests are being performed with the test  voltage at the specified voltage value   The 24 Vdc output is not generated during voltage rise 
68. rol to  start tests  For details  see the    OPERATION MANUAL HIGH VOLTAGE TEST PROBE  HP01A TOS HP02A TOS           Disconnecting test leads from the DUT    Check that the POWER switch is off and that the DANGER LED is off     Disconnect the high voltage test lead  red  from the front panel HIGH  VOLTAGE terminal     Disconnect the low voltage test lead  black      You can disconnect the low voltage test lead  black  first from either the DUT or the  TOS5200     3 Disconnect the high voltage test lead  red  from the DUT     TOS5200       Safety Precautions for Testing    This chapter describes the precautions that  must be followed to perform tests safely     28    Pre Test Inspection       Z N WARNING Risk of electric shock     e During testing  this product supplies a voltage of at least 5 kVac to an external  device  Handling this tester improperly may lead to a fatal accident  To prevent acci   dents  strictly follow the precautions and always pay the utmost attention to safety  concerns when you operate the TOS5200     e This product conforms to IEC Safety Class    equipment that has a protective con   ductor terminal   Be sure to earth ground the product to prevent electric shock     e The product is grounded through the power cord ground wire  Connect the protec   tive conductor terminal to earth ground     e During testing  be sure to wear rubber gloves for electrical work    Check the following items before you start testing  and always follow the precautions   e Th
69. s is 1 KQ or less      14 HV ON O On during testing and when a voltage remains across the output terminals  E  15 TEST O   On during testing  excluding when voltage is rising or falling   O  16 PASS o On for at least 0 2 seconds  the PASS HOLD time  when a PASS judgment occurs   o  On continuously when the PASS HOLD time is set to HOLD  E  On continuously when an UPPER FAIL judgment occurs because a value greater than or 2  17 U FAIL O eech x lt   equal to the upper limit is detected  UU  On continuously when LOWER FAIL results from judgment because a value less than or  18 L FAIL O EE  equal to the lower limit is detected   19 READY O   On when the TOS5200 is waiting  when it is in the READY state    20 PROTECTION O On when protective features have been activated   21 START l Start signal input terminal   22 STOP l Stop signal input terminal   23 ENABLE   Input terminal for the start signal   s ENABLE signal  If the ENABLE signal changes  the  TOS5200 is switched to Protection mode   24  24 V        24 V internal power supply output terminal  maximum output current 100 mA   25 COM       Circuit common terminal   NC  NC NC  STB     m NC  TEST SEL             _ NCC  AITO SEL      po PM1  COM PMO  INTERLOCK  m INTERLOCK   I  ta  12   11  10   9   8 zels lalale   28  4l bal bol bi  20 19  hal 17 16  hs 14   COM        HV ON   24V           TEST  ENABLE          _    i PASS  STOP    _ i J FAIL  START     L L FAIL  PROTECTION READY  e      TOS5200 S 55  9    SIGNAL I O Connector  
70. sults    52    External Control       SIGNA VO eeler  SIGNAL I O specifications                                 inreenensnnenenenaenenannennnneannaeaneennnnat  Internal construction                       Input signal usage example  Output signal usage example        s sssesseeeeeeesekesitsiteekeetrsttstrtknstnntnntnnstnntnnenneennee 56  Starting ECK EE 57  Recalling Panel Memory and Test Modes EENEG 57  Interlock TE 58  How to use the interlock feature A 59  STATUS e Ria 59  Maintenance  Pre Test   1151 072707   9  6 SAHA 62  Inspection of test leads and the judgment feature    s es 62  GC IAT ON ai aa eege 62  Specifications  Withstanding voltage Tester    64  Other FEALUFSS sieniin a rni ai rA a EEE avai eiS nS   ul  ENEE       Outline drawing    Appendix    A Listof Default Settings sesiiesiscrsscceccrsssssessensiesasisdenastacsiseciaacnsscas 72  Initializing the TOS5200                              strnnensnnenenenennenannnnananannananneannananenananee 72   B Protection Functions eebe 74  TOS5200    TOS5200    C Test Start Operation and Display               ccccssssssseeeeesssseeeeees 75  D  TROUDIGSIOOUIING scciccsascsnadhncrescccaseanasanccistonssccccuaebenetsoommacananananins 76  1470  EN 79    Search by Topic    Preparation          What accessories are included in the  package      gt     Checking the Package Contents          Before   start testing    want to check  that the TOS5200 is operating safely      gt     Turning the Power On    p 22    Other than the basi
71. tection mode   factory default setting     on  1  The 24 Vdc output is turned on when the TOS5200 is in protection mode    Power ON  Sets whether the 24 Vdc output is generated while the POWER switch is on    off  0  The 24 Vdc output is turned off when the POWER switch is turned on  fac   tory default setting     on  1  The 24 Vdc output is turned on when the POWER switch is turned on        TOS5200    Withstanding Voltage Test Q    47    OW    Starting a Test       Risk of electric shock  During testing  do not touch the test leads and the DUT     N WARNING    To start a test    1 Check that the TOS5200 is correctly connected to the DUT     2 When the READY LED lights in white  press START     PROTECTION FMS       READY  E    If the test does not start  see    If you cannot start testing        If you want to start another test after testing finishes  set a wait time between tests as neces   sary        When atest starts    The TEST and DANGER LEDs light  The TEST LED lights in red                    a Lit  ad  m  lt     e  ZO                                O   ii oe Lit   The TEST and DANGER LEDs light       Dancer  id  A Sr AR The TEST LED lights in red     If the start voltage and the voltage rise time have been set    After the voltage rises to the start voltage or after the voltage rise time elapses  the TEST  LED lights in red  and the test starts  While the voltage is rising  the TEST LED blinks in red   and the display counts up to the set rise time        Blinking 
72. time  Rise Time  and voltage fall time              Fall Time    off  0  The 24 Vdc output is turned off during testing  factory default setting    on  1  The 24 Vdc output is turned on during testing   Pass    Sets whether the 24 Vdc output is generated after a PASS judgment has occurred           off  0  The 24 Vdc output is turned off during testing  factory default setting    on  1  The 24 Vdc output is turned on during testing   Upper Fail    Sets whether the 24 Vdc output is generated after an upper limit FAIL judgment has occurred        off  0  The 24 Vdc output is turned off after an upper limit FAIL judgment occurs   factory default setting         on  1  The 24 Vdc output is turned on after an upper limit FAIL judgment occurs     Lower Fail    Sets whether the 24 Vdc output is generated after a lower limit FAIL judgment has occurred        off  0  The 24 Vdc output is turned off after an lower limit FAIL judgment occurs   factory default setting         on  1  The 24 Vdc output is turned on after a lower limit FAIL judgment occurs                 Ready  Sets whether the 24 Vdc output is generated while the READY LED is lit   off  0  The 24 Vdc output is turned off while the READY LED is lit  factory default  setting    on  1  The 24 Vdc output is turned on while the READY LED is lit   Protection    Sets whether the 24 Vdc output is generated while the TOS5200 is in protection mode                    off  0  The 24 Vdc output is turned off when the TOS5200 is in pro
73. tors understand the information in this manual before they  operate this product     Keep this manual close to the product so that the operators can read the manual at  any time    If the tester will be used to repeatedly perform tests with fixed conditions  such as  when being used as part of a manufacturing line  attach the protection cover to  ensure safe operation of the tester  This is useful in preventing incorrect operation of  the tester     Dangerous Operations    You will receive a potentially fatal electric shock if   e You touch an output terminal while output is being generated     e You touch a test lead that is connected to an output terminal while output is being  generated     e You touch the device under test  DUT  while output is being generated     e You touch a location that is electrically connected to an output terminal while output  is being generated     You may receive a potentially fatal electric shock if   e You operate the tester without grounding it   e You operate the tester without using rubber gloves for electrical work     e You come close to a location that is electrically connected to an output terminal  while output is being generated        TOS5200    AN Safety Precautions       When using this product  be sure to observe the    Safety  Precautions    in the Safety information manual      N Precautions Concerning Installa     tion Location       When installing this product  be sure to observe the    Pre   cautions Concerning Installation
74. ulated time        Remote Protection    A connection to or disconnection from the front panel REMOTE connector was detected        SIGNAL I O Protection    The rear panel SIGNAL I O connector s ENABLE signal has changed        USB Protection             The USB connector has been disconnected  or a defect was detected during remote con   trol operation           TOS5200    Specifications                                                       Interfaces  0000000000000000000000000000000000000000000000000000000000000000000  TOS5200  USB USB Specification 2 0  Standard type B socket  RS232C  Hardware D SUB 9 pin connector  EIA 232 D compliant   All functions except the POWER switch and key lock  Baudrate  9600  19200  38400  57600  115200 bps  Transmission system  Start stop synchronization  Data length  8 bits  Stop bits  1 bit  Parity bit  None  Program message CR LF during transmission  CR  LF  or CR LF during reception  terminator  REMOTE Front panel 9 pin MINI DIN connector   By connecting an optional device to this connector  you can control the starting and  stopping of tests remotely   SIGNAL UO Rear panel D sub 25 pin connector  Output Output method Open collector output  4 5 Vdc to 30 Vdc   specifications   Output withstanding 30 Vdc  voltage  Output saturation volt  Approx  1 1 V  25   C   age  Maximum output current   400 MA  TOTAL   Input High level input voltage   11 Vto 15 V  specifications    Low level input voltage  0 Vto4 V  Low level input current 5 MA max      
75. ut is    UPPER    ON a U FAIL  FAIL turned off  and an UPPER FAIL judgment   lights i    signal  occurs   If a current that is less than or equal to the   FAIL LED and  lower limit is detected  the output is    LOWER     LOWER turned off  and a LOWER FAIL judgment   lights  Generates  occurs  This judgment is not performed ON an L FAIL  FAIL A SE i   3  during voltage rise time  Rise Time  of all signal  tests and during the voltage fall time  Fall  Time  of withstanding voltage tests   If the specified time elapses without any PASS LED Generates  PASS problems  the output is turned off  and a lights ON a PASS  PASS judgment occurs  gnis  signal     If PASS HOLD is enabled  the PASS signal is generated continuously until the TOS5200  receives a STOP signal     The UPPER FAIL and LOWER FAIL signals are generated continuously until the TOS5200  receives a STOP signal     The FAIL and PASS buzzer volume levels can be changed     For PASS judgments  the length of time that the buzzer sounds for is fixed to 0 2 seconds   Even if PASS HOLD is enabled  the buzzer turns off after 0 2 seconds   Upper limit setting 0 01 mA to 110 mA  Lower limit setting 0 01 mA to 110 mA  OFF  Judgment accuracy  1 00 mA si     1 5   of set   i  lt  1 00 MA     1 5   of set   30 JA   Current detection method   Calculates the current s true rms value or mean value value and compares this value with  the reference value  Calibration Calibrated with the rms of a sine wave using a pure resistive load  Time 
76. ved safety    In addition to having features that enable you to view the output voltage  the TOS5200  also enables you to set the voltage limits  so you can prevent a voltage greater than what  is necessary from being generated unintentionally  This provides protection for the DUT     e Standard equipped USB port    The TOS5200 is standard equipped with a USB interface  You can use a PC or  sequencer to control test conditions and read measured values and test results        1 The four tests are the withstanding voltage  insulation resistance  earth continuity  and leakage   current tests     TOS5200    Product Overview       TOS5200    Light weight and easy to move    Even though the TOS5200 can generate 500 VA  which is sufficient for performing with   standing voltage tests  it only weighs 15 kg or less  so it can be moved by even a single  person     Protection against incorrect operations    In addition to the key lock feature  the TOS5200 has a protection cover for the part of its  panel that is used to change test conditions  This cover is useful in preventing incorrect  operations when you want to perform tests with fixed conditions     General Description E    15    Options       The following options are available for the TOS5200   For information about options  contact your Kikusui agent or distributor     Rack mount option    Name Model Note  KRA4 TOS For EIA inch racks  KRA200 TOS For JIS millimeter racks       Rack mount adapter             KRA4 TOS    Model RC0
77. xed while panel settings are being recalled from or    See  poe  saved to memory        See  p  51 54 e STOP is being pressed  including when a STOP signal is being applied to the SIG   bay Peoh es NAL UO connector         e    mA    is blinking                                      See  p  40  p The lower or upper limit is turned on  and the upper limit has been set lower than the  lower limit  Specify a valid value   e    kV    and    mA    are blinking simultaneously    When the test voltage and the upper limit are set in the withstanding voltage test settings   such that their product exceeds 550 VA  Specify a valid value   e      TOS5200 ku        Withstanding Voltage Test Q    50    Starting a Test                PASS    or    FAIL    is lit     You cannot start testing while a judgment result is displayed  Press STOP to switch the  TOS5200 to the READY state        PROTECTION    is lit     If PROTECTION is lit  the TOS5200 has switched to protection mode and will not allow  you to start testing  Resolve the issues causing the PROTECTION mode to be on        The double action feature has been turned on     When the double action feature is turned on  you cannot start tests just by pressing  START  Press STOP  and then press START within 0 5 seconds     TOS5200    Finishing a Test       To stop a test    STOP    K    Press STOP     When a test finishes     See  p  41   p  42       TOS5200    A test will stop under one of the following conditions     a  If the test time el
78. y  Remote control is not possible Is the baudrate set correctly  p  46    through the RS232C interface  Check the value        TOS5200       77    78    This page is intentionally blank     TOS5200    Index    A  accessories                              nnnennnnnnnnnnnnnneene ne enneenena 2  B  Buzzer Volume  Elle eebe ee 46  PASS ren ee 45  c  CONFIG setup screen   sssssesssnssssissrnsinsrrennernrerreenne 44  connection  DUT  E A E aie eee 24  D  default settings  list                           mrnnnene 72  Double Action    45  DUT  E t E EEE 24  F  Fail 4 0 e  MESA 45  IC UNN Le 43  fall time control                              rerreneenneeeeeeneeeenee 43  firmware version                            mmrennnnennnnnnnannnannanne 2  ee Ee 43  H  HPOTA TOS ou  ccccccccccccccceceeeeeeeeeeeessensssesseuseaeeeeneeeeeers 26  HPO2ART EG 26  I  initialization    72  10551210 17970  EPEE SEAD 62  judgment feature      62  test leads                             ennminenneeeeeea een 62  Interlock Protection    74  K  KEYLOCK orriei ta kii aint aeii 36  key ek  25454 1330 ama ee a 36  L  EA eieiei 40  Lower Fall EE 47  lower limit judgment 00    ee eee eeeteeeeeneeeteneeees 32  40  M  measurement EE 43  measurement check ou     ceeeeeecseceseceseseseeeesesentesseeaees 62  message  invalid setting                           ismenamv 40  TOS5200    Mesuremerit uk case siruina iiai 43  Momentary  geck magaja aema aa aada Desk 45  O  elei IC 16  outline drawing                miausereenansen
79. y the CONFIG setup screen   ACW coun Press once to show the CONFIG   screen and twice to show the CONFIG   screen     es    en    d Press MEMORY 2 or MEMORY 3 to move to the CONFIG item that you  want to set  and then use the rotary knob to set the value     To exit from the CONFIG setup screen and return to the basic setup screen  press STOP        Memo    The first screen that is dis     played when you perform viously were viewing will be displayed   configuration settings is   called CONFIG setup   screen     CONFIG 1 settings    If you enter CONFIG settings again without first turning the power off  the screen that you pre     On the CONFIG   setup screen  you can set the test mode and buzzer volume level     Example  Selecting the double action feature   blinking         0 to 9  10 setting levels                       v  Double action feature setting  0  off  1  on     A Fail mode function setting  0  off  1   Momentary function setting  0  off  1  on     eee ny    ID       Talk mode setting  0  off  1  on     RS232C baud rate setting   0  9 600  1  19 200  2  38 400  3  57 600  4  115 200     FAIL judgment buzzer volume level setting   0 to 9  10 setting levels     PASS judgment buzzer volume level setting    Judgment result hold time setting   0  50 ms  1  100 ms  2  200 ms  3  1 s  4  2 s  5 5 s  6  HOLD     on     TOS5200    CONFIG Settings       Double action feature    This feature enables you to start a test by first pressing STOP and then pressing START  within 
    