1. SAFETY ACTION NOTICE SHS SCOTLAND By arrangement with the Scottish Executive Health Department SAN SC 02 08 BAXTER COLLEAGUE 1 AND 3 VOLUMETRIC INFUSION PUMPS 09 APR 2002 MALFUNCTION OF ON OFF CIRCUITRY DUE TO FLUID INGRESS Medical Devices Page 1 of 6 Pages SUMMARY Baxter Colleague 1 and 3 volumetric infusion pumps may power on or power off without the user operating the controls and without any alert or alarm Advice is given on corrective measures and ongoing precautions BACKGROUND 1 Baxter Healthcare Ltd have established that fluid ingress into the main body of Colleague volumetric infusion pumps can cause a short circuit leading to a malfunction of the on off circuitry The problem affects Colleague 1 single channel pumps product codes FKM8151 and 2M8151K and Colleague 3 triple channel pumps product code 2M8153K The malfunction may result in the pump spontaneously turning itself on or off or rapidly cycling between on and off This may occur without any alerts or alarms if the Panel Lockout facility is not enabled 2 Fluid may enter the main pump body if the recommended cleaning instructions are not followed or if a significant quantity of IV solution should spill onto the pump Fluids penetrating the pump may settle on the ribbon cable behind the keypad Prolonged exposure may cause the ribbon cable to degrade and experience an intermittent short circuit causing the behaviour described above 3
2. power on or power off without a key press Most commonly when the circuit shorts a rapid power on off cycling occurs These events may occur with or without an alert or alarm and if they should occur during patient use there is a potential that delivery of therapy may be interrupted Fluid may enter the main pump body if the recommended cleaning instructions are not followed or if a significant IV solution spill into the pump occurs Fluids entering the pump may settle on the ribbon cable behind the keypad Over prolonged exposure the ribbon cable may degrade and experience an intermittent short circuit causing the behaviour described above During the next service activity on your pump an insulator should be placed in the area of the ribbon cable to prevent the effect of any fluid contact on this circuitry This activity should take no more than 5 minutes to complete In the interim any of the following actions will prevent this malfunction from occurring Colleague is a trademark of Baxter Healthcare Corporation SAN SC 02 08 09 APR 2002 Medical Devices Page 4 of 6 Pages Do not spray cleaning fluid onto the body of the pump or submerge the pump for cleaning Please remember that the Colleague Volumetric Infusion Pump is not watertight Under normal use and following the recommended cleaning instructions attached cleaning fluid will not enter the pump and the behaviour described above will not occur As recommended i
3. Cleaning the Exterior Clean the pump after each use with soapy water a solution of 10 bleach and water isopropyl alcohol up to 95 or a commercial cleaning solution that does not contain any of the ingredients listed above Follow the manufacturer s dilution instructions for concentrated cleaners Clean the exterior using a soft cloth sparingly dampened not soaked with cleaning solution To clean pumps that have been in an isolation area select a cleaner that cleans and disinfects Colleague is a trademark of Baxter Healthcare Corporation SAN SC 02 08 09 APR 2002 Medical Devices Page 6 of 6 Pages Colleague Infusion Pump 4 February 2002 Letter Reply Record Please complete and return this record to the FAX number listed below 01842 767134 Reply Record Completed By please print name Title please print Telephone Number Signature Date REQUIRED FIELD Facility Name and Address Colleague is a trademark of Baxter Healthcare Corporation
4. Baxter Healthcare issued an Urgent Device Correction letter on 4 February 2002 attached The letter informed customers of the problem and provided details of the corrective measures required fitting an insulator The letter also outlined interim measures to be taken until the insulator is fitted as well as Cleaning Instructions to be posted in all pump cleaning areas 4 Baxter service engineers and technical staff will fit the insulators to pumps on a next seen basis Baxter has also arranged to provide insulators and installation instructions Colleague Pumps Service Bulletin on a no cost basis where self fit by a suitably qualified person is preferred ACTION 5 This notice should be brought to the attention of all appropriate managers and staff 6 Staff should be made aware of the attached Urgent Device Correction letter from Baxter Healthcare Ltd in connection with Colleague 1 and 3 volumetric infusion pumps Suggested Distribution District Nursing Intensive Therapy Units O D A s Risk Management Social Services SCOTTISH HEALTHCARE SUPPLIES Gyle Square Edinburgh EH12 9EB A Division of the National Service Scotland for NHSScotland This notice was obtained from the INCIDENT REPORTING amp INVESTIGATION CENTRE public website P http www nhsscotland com shs hazards_safety adverse_p html FAX 0 1 3 1 3 14 0722 E MAIL iric shs csa scot nhs uk N WORD HAZ PUBLIC LIBRARY SAN 2002 SAN PSAN0208 doc SAFETY ACTIO
5. N NOTICE SHS By arrangement with the Scottish Executive Health Department SCCIFAND SAN SC 02 08 BAXTER COLLEAGUE 1 AND 3 VOLUMETRIC INFUSION PUMPS 09 APR 2002 MALFUNCTION OF ON OFF CIRCUITRY DUE TO FLUID INGRESS Medical Devices Page 2 of 6 Pages CORRECTIVE MEASURES 7 Baxter Colleague 1 and 3 volumetric infusion pumps with the following product code numbers should be identified FKM8151 2M8151K and 2M8153K 39 66 8 Pumps which show an unexpected power on power off or similar anomalies should be removed from service and Baxter Technical Service should be contacted for further advice 9 An insulator should be installed in the area of the ribbon cable during the next service activity on the pump as described Baxter s Colleague Pumps Service Bulletin Keypad Flex Connector Insulator Installation ONGOING PRECAUTIONS 10 Cleaning fluid should not be sprayed onto the body of the pump nor should the pump be submerged for cleaning 11 The pump cleaning instructions in the Baxter letter should be displayed in the pump cleaning area and should be followed 12 The panel lockout feature as described in the User Manual should be used during infusions ENQUIRIES Enquiries to the manufacturer should be addressed to General Enquiries Technical Enquiries Ms Fiona Taylor Reports of malfunction and requests for Baxter Infusion Systems Product Manager upgrading work or parts amp installation Baxter H
6. ealthcare Ltd instructions should be directed to Caxton Way Thetford Baxter Technical Service Norfolk IP24 3SE Tel 01604 704555 Fax 01842 704590 Tel 01842 767000 Fax 01842 767099 SCOTTISH HEALTHCARE SUPPLIES Gyle Square Edinburgh EH12 9EB A Division of the National Service Scotland for NHS Scotland This notice was obtained from the INCIDENT REPORTING amp INVESTIGATION CENTRE public website P http www nhsscotland com shs hazards_safety adverse_p html FAX 0 1 3 1 3 14 0722 E MAIL iric shs csa scot nhs uk N WORD HAZ PUBLIC LIBRARY SAN 2002 SAN PSAN0208 doc SAN SC 02 08 09 APR 2002 Medical Devices Page 3 of 6 Pages URGENT DEVICE 4 February 2002 CORRECTION Re Colleague Volumetric Infusion Pump Product Codes FKM8151 2M8151K HNM8151 DNM8151 GNM8151 CNM8151 WNM8151 PNM8151 2M8151 2M8153K HNM8153 DNM8153 GNM8153 CNM8153 WNM8153 PNM8153 2M8153 Dear Customer Baxter has recently completed an investigation into reports of an infrequently occurring malfunction of the on off circuit in the Colleague 1 amp 3 Volumetric Infusion Pump We are sending you this information so that you will understand the cause the possible consequences how to prevent it and what Baxter will be doing to correct it A malfunction of the on off circuitry can occur if fluid enters the main body of the pump and after long term exposure causes a short circuit A short circuit may then cause an unexpected
7. id receiving a repeat mailing of this letter please complete the attached reply record and return it by FAX to the numbers listed on the reply record We sincerely apologise for the inconvenience that this situation causes to you and your staff Thank you for your assistance Sincerely Patricia Dargan Regulatory QA d Associate Baxter Healthcare Colleague is a trademark of Baxter Healthcare Corporation SAN SC 02 08 09 APR 2002 Medical Devices Page 5 of 6 Pages Colleague Volumetric Infusion Pump Cleaning Instructions Please post in all pump cleaning areas Prior to cleaning the pump read this procedure carefully Follow this procedure to ensure that the pump is not damaged during cleaning and remains under Baxter warranty Contact your authorised Baxter Service Centre for further information CAUTIONS e Do not use hard instruments for cleaning e Do not spray liquids or foam directly on the pump e Do not submerge the pump into liquids of any kind e Do not clean disinfect or sterilise any part of the pump by autoclaving or with the use of ethylene oxide gas Doing so may damage the pump and void the warranty Only external parts of the pump should be disinfected e Do not use the following chemicals on the pump as they will damage the front panel acetaldehyde acetone ammonia benzene hydroxytoluene methylene chloride or ozone Do not use cleaners containing n alkyl dimethyl ethyl benzyl ammonium chloride
8. n the Colleague User Manual use the panel lockout feature as a standard procedure during infusions Pressing the black Panel Lockout button on the back of the pump will enable this feature When the Panel Lockout is enabled all inputs from the on off key are ignored and the behaviour described above should not occur or should it occur a failure alarm will sound Alternatively the Autolock feature can be enabled in the configuration for each personality This results in the automatic setting of the Panel Lockout 2 minutes after the start of an infusion Install a protective insulator at the location of the ribbon cable This insulator and the installation instructions will be provided to you at no charge by calling Baxter Technical Service on 01604 704555 If any of your pumps become exposed to excessive fluid from spills or inadvertent immersion please remove the pump from service and contact Baxter Technical Service on the above number for advice Remove from service any Colleague pumps in your inventory that are reported to show this on off or similar behaviour and contact Baxter Technical Service on the above number for advice Any pumps returned to Baxter or seen by a Baxter Engineer will have the insulator fitted on a next seen basis Please do not hesitate to contact your Baxter Infusion Systems Product Manager Fiona Taylor on 01635 206411 with any questions The Medical Devices Agency has been notified of this action In order to avo