Home
        MQSA Final Regulations Modifications and Additions to Policy
         Contents
1.  4  Emphasizing the need to minimize non  AEC component variability when conducting the  AEC performance test     FDA s guidance documents  including this guidance  do not establish legally enforceable  responsibilities  Instead  guidances describe the Agency s current thinking on a topic and should be  viewed only as recommendations  unless specific regulatory or statutory requirements are cited   The use of the word should in Agency guidances means that something is suggested or  recommended  but not required     Contains Nonbinding Recommendations    Background    This document is intended to provide guidance to mammography facilities and their personnel  It  represents the Food and Drug Administration   s  FDA  current thinking on the final regulations  implementing the Mammography Quality Standards Act  MQSA   Pub  L  102 539   The FDA  uses mandatory language  such as shall  must  and require  when referring to statutory or regulatory  requirements  The FDA uses non mandatory language  such as should  may  can  and recommend  when referring to guidance  It is the responsibility of the facility to read  understand  and follow the  final regulations     Under its own authority  a State may impose more stringent requirements beyond those specified  under MQSA and its implementing regulations  A facility may want to check with the State or local  authorities regarding their requirements     The Mammography Quality Standards Act was signed into law on October 27  1992  to es
2.  call the CDRH Facts On Demand system at 800 899 0381 or 301   827 0111 from a touch tone telephone  Press 1 to enter the system  At the second voice prompt   press 1 to order a document  Enter the document number 1435 followed by the pound sign       Follow the remaining voice prompts to complete your request     Contains Nonbinding Recommendations    Table of Contents   Guidance for Industry and FDA  Stati 1 20    sascessesestusscead soeqaleaadoacave aemedeatvsucaetsauaenpzateccasatede ine 1  InMtrOdUCHON 3220  ested EEE E ental ible Lge en Ieee 1  Background esnia aen Senate a uscd choca e ates E a a talons Socata hos a 2  Automatic Exposure Control  AEC  Performance Testing     Annual Physics Survey and  Mammography Equipment Evaluation         seseseeeeeeeseseeeeesessesressersreseesstesetsersstesseseresresseseresressesse 4  Air Kerma and AEC Reproducibility Annual Quality Control Test          sseeeeeeeeeeeeeeeeeeeeeseesseeeresee 14  Phantom Images Exposed in a Fully Automatic AEC Mode  if that is the Clinically used Technique  EE GRC Ulead Deh T EE cl IR olay yeh BR OI Rite bh Rie Be OR eo ee Reale tok 15  ally ASSUAGE RECOM  4 a esx geese a E a A s aa 16  Weekly Equipment Quality Control Tests        0 cccccessseceeseecessceceeaceceeneeceeneeceeneecseneeceeeeceeeeenaeees 17    Mammography Equipment Evaluations               cc cccssscecssnceceecceceeceeceeeeecseceeceeeeecsueeeceeeeesneeeenaeees 18    Contains Nonbinding Recommendations    Guidance for Industry and FDA 
3.  e  and  f  of this section until the next annual inspection has been  completed and FDA has determined that the facility is in compliance with the quality assurance requirements  or until the test has been performed two additional times at the required frequency  whichever is longer     Question 4  Is it acceptable for a technique chart to simply state that the facility  is using the unit in its fully automatic AEC mode  e g   BACE  OPDOSE  AUTO  FILTER  AOP  or similar modes  for all routine examinations     Yes  However  if such a technique chart is the only one available  the facility may not use the unit if  the automatic AEC mode listed on the technique chart fails during medical physicist testing  For this  reason  FDA strongly recommends that the facility develop  and have available  a technique chart  that includes other AEC modes  such as fixed kVp mode  and or manual techniques  The technique  chart could then be used as a guide for the radiologic technologist in the event that the fully  automatic AEC mode failed or was not suitable for use with a specific patient     16    Contains Nonbinding Recommendations    Weekly Equipment Quality Control Tests    Citation   900  12 e  2  i    ii    iii    iv    2  Weekly quality control tests  Facilities with screen film systems shall  perform an image quality evaluation test  using an FDA approved phantom  at least weekly      i  The optical density of the film at the center of an image of a standard FDA accepted phantom s
4.  even after October  28  2002     3  Confirming AEC performance by demonstrating that the AEC maintains the MOD within      0 15 OD      0 30 OD if done before October 28  2002  in all configurations used clinically by  the facility  The action limit applies only within each specific configuration tested and does not apply  to data collected across the different configurations     Because of conflicting recommendations that existed in the professional community regarding  measurement of AEC performance during mammography equipment evaluations  facilities that  measured AEC performance only in the contact configuration before October 28  2002 will not be  cited for failure to measure AEC performance for all clinically used configurations  However  those  that continue this practice after October 28  2002 will be subject to citation     Question 5  Must medical physicists test all AEC detectors for AEC  performance in mammography units with multiple AEC detectors and can the  testing procedures be modified if the detectors are in the same cassette holder   bucky      The general principle is that all AEC detectors must be tested  What is considered adequate testing  will depend on the arrangement of the AEC detectors in the mammography unit    1  Where a mammography unit has different AEC detectors in the different size  cassette holders  buckys   each detector must be tested separately as described  above in questions  3 or  4    2  Where a mammography unit has more than one A
5.  met  a technique chart shall be developed showing appropriate  techniques  kVp and density control settings  for different breast thicknesses and compositions that  must be used so that optical densities within  0 30 of the average under phototimed conditions can  be produced    B  After October 28  2002  the AEC shall be capable of maintaining film optical density  OD   within  0 15 of the mean optical density when thickness of a homogeneous material is varied over a  range of 2 to 6 cm and the kVp is varied appropriately for such thicknesses over the kVp range used  clinically in the facility    C  The optical density of the film in the center of the phantom image shall not be less than 1 20     900  12 b  10  i    ii   A  B    iii   Automatic Exposure Control     i  Each screen film system shall provide an AEC mode that is operable in all combinations of equipment  configuration provided  e g   grid  nongrid  magnification  nonmagnification  and various target filter  combinations      ii  The positioning or selection of the detector shall permit flexibility in the placement of the detector  under the target tissue    A  The size and available positions of the detector shall be clearly indicated at the X ray input  surface of the breast compression paddle    B  The selected position of the detector shall be clearly indicated    iii  The system shall provide means for the operator to vary the selected optical density from the normal     zero  setting     Question 1  What i
6. 4  How much variability from the  normal  optical density setting  must the system provide     11    Contains Nonbinding Recommendations    The regulations do not specify the range of variability that must be provided  only that some  variability is available     Question 15  Do all possible positions of the AEC detector have to be indicated  on the compression paddle     The intent of this regulation is to help the radiologic technologist optimally position the AEC  detector  Under some AEC detector designs it may be difficult to show all possible positions   Some detectors cover the entire area of the image receptor and once the exposure begins  they  automatically select the region of maximum density as the active area  For these systems  indication  of the entire potential active area  along with appropriate instructions  usually found in the user  manual   would satisfy this requirement  Since the area is automatically selected  the display of the  size of the detector is not required  Other designs may have an essentially infinite number of  locations under all or part of the image receptor  An indication of the complete range and detector  size  coupled with adequate instructions  would be sufficient  Still others may indicate the range of  multiple positions on the paddle  Again  this would satisfy the requirement  There may be other  methods employed that also satisfy the requirement  The key is that the operators know what areas  they may select and the size of the 
7. EC detector in a single cassette  holder  bucky   the physicist must test all the individually selectable AEC detectors  and may test the detectors using either of the following methods    i  All detectors as described above in questions  3 or  4  OR    i  One detector as described above in questions  3 or  4 AND comparing  the OD obtained at 4 cm from each of the other detectors to the MOD  obtained from the first detector  When results across different detectors are  compared  the medical physicist may use the action limit of     0 30 OD  even after October 28  2002    3  Where a mammography unit has multiple AEC detectors that are not individually  selectable by the operator  the AEC can be tested as if it was a single detector  An    Contains Nonbinding Recommendations    example of such a system is one with three fixed detectors in which the system  automatically chooses which detector will be active during the exposure  Similarly   a large field detector that automatically selects its active area needs to be tested only  as a single detector  However  a system with three fixed detectors  each of which  can be selected individually by the operator  needs to have all three detectors tested  as described in section  2 above  Please note that a detector that can be moved to  different positions by the operator is still considered a single detector and needs to  be tested at only one of those positions     Question 6  If the AEC performance test fails  does that automaticall
8. Guidance for Industry and FDA  Staff    The Mammography Quality  Standards Act Final Regulations  Modifications and Additions to  Policy Guidance Help System  6    Document issued on  August 19  2003    This document supersedes The Mammography Quality Standards Act  Final Regulations Modifications and Additions to Policy Guidance Help  System  6 Draft Guidance issued on February 19  2003    For questions regarding this document contact Charles Finder 301 827 0009     U S  Department of Health and Human Services    D    Food and Drug Administration  R H 3 Center for Devices and Radiological Health  p  Division of Mammography Quality   Yea  ie 6    and Radiation Programs  Office of Health and Industry Programs    Contains Nonbinding Recommendations    Preface    Public Comment    Written comments and suggestions may be submitted at any time for Agency consideration to  Dockets Management Branch  Division of Management Systems and Policy  Office of Human  Resources and Management Services  Food and Drug Administration  5630 Fishers Lane  Room  1061   HFA 305   Rockville  MD  20852     Alternatively  electronic comments may be submitted to http   www fda gov dockets ecomments   When submitting comments  please refer to Docket No  03D 0025  Comments may not be acted  upon by the Agency until the document is next revised or updated     Additional Copies    Additional copies are available from the Internet at  http   www fda gov cdrh mammography  or to  receive this document by fax 
9. Staff    The Mammography Quality Standards Act  Final Regulations Modifications and  Additions to Policy Guidance Help System   6    This guidance represents the Food and Drug Administration s  FDA s  current thinking  on this topic  It does not create or confer any rights for or on any person and does not  operate to bind FDA or the public  You can use an alternative approach if the approach    satisfies the requirements of the applicable statutes and regulations  If you want to  discuss an alternative approach  contact the FDA staff responsible for implementing this  guidance  If you cannot identify the appropriate FDA staff  call the appropriate number  listed on the title page of this guidance        Introduction    On February 19  2003  FDA published The Mammography Quality Standards Act Final  Regulations Modifications and Additions to Policy Guidance Help System  6 Draft Guidance for  public comment  During the public comment period  two respondents submitted a total of 14  comments  In addition  the National Mammography Quality Assurance Advisory Committee  reviewed the Draft Guidance during its April 28  2003 meeting and provided additional comments   In response to those comments  FDA has modified the guidance as follows by   1  Further clarifying the term    equipment configuration      2  Adding different image receptor sizes as separate equipment configurations   3  Not recommending that target filter combinations be tested as separate equipment  configurations  
10. d with the malfunctioning AEC  The facility can use this temporary  chart for up to 30 days  or until the problem has been corrected and the equipment passes the AEC  performance test  whichever comes first  If the AEC is completely non functioning  the medical  physicist may create a manual mode technique chart that includes all the appropriate manual  technique factors  Use of the manual mode would be acceptable under the complete failure  situation raised by the question  The facility can use manual techniques for up to 30 days while the  non functioning AEC is being repaired and can continue to use the unit on patients during this  period     The answer to the second question depends on the repair needed to fix the problem  If the repair is  classified as    major     see Table  Medical Physicist Involvement in Equipment Adjustments   Changes  or Repairs    then the medical physicist must be onsite to perform the post repair testing   If the repair is not classified as    major    then the post repair testing may be done under the medical  physicist   s oversight  In either event  the appropriate testing must be performed and passed within  the specified time frames     13    Contains Nonbinding Recommendations    Air Kerma and AEC Reproducibility Annual Quality  Control Test    Citation   900 12 e  5  v   Breast entrance air kerma and AEC reproducibility  The coefficient of variation for both air  kerma and mAs shall not exceed 0 05     Question 1  Must medical physicist
11. detector     Question 16  The position of the AEC detector is infinitely variable over the  entire area of the image receptor  How can the position of such a detector be  identified on the compression paddle     An indication of the range of coverage and the detector size  along with appropriate instructions   usually found in the user manual   would satisfy this requirement     Question 17  Do paddles designed to be smaller than the full size of the image  receptor have to display the AEC detector position and size     No  Paddles designed to be smaller than the full size of the image receptor do not have to display  AEC detector position and size  Paddles used only for invasive procedures do not have to display  AEC detector position and size because they are not covered by the regulations     Question 18  If the AEC performance is found to be outside the action limit  during physicist testing  can a facility adjust the density control settings or use  manual techniques until the unit is fixed  Would it require the physicist to come  and recheck it or if the repairman did so would that be satisfactory     12    Contains Nonbinding Recommendations    The answer to the first question is yes  According to 900 12 e  5     A   when the AEC  performance is found to be outside the action limit during physicist testing  the medical physicist may  create a temporary technique chart that includes the appropriate density settings  in addition to the  other technique factors  to be use
12. e complex in recent years  When units had only one AEC detector  a  single AEC mode  and a single target  filter combination  testing was relatively straightforward  That  is no longer the case for most units  The following guidance is designed to help medical physicists  adequately test a unit   s AEC performance without over testing the unit     During the annual physics survey  the physicist can limit testing of AEC performance to the contact  configuration  To fulfill MQSA requirements  all AEC detectors  that can be individually selected  by the operator  and all AEC modes used clinically over the 2 to 6 cm range in the contact  configuration must be tested  While there are several ways to do the test  medical physicists who  use the following guidance will have fulfilled this requirement  Note  Facilities that do not clinically  use their AEC in the 2 to 6 cm range  only use manual techniques  must still test the AEC to ensure  that at least one AEC mode for each available AEC detector meets the regulatory requirements     In order to minimize sources of variability  the physicist should use a single cassette  or same  cassette type   film from the same emulsion batch  and the same processing conditions throughout  Steps 1 and 2 below  see question   6 below      Step 1  Determine the Mean Optical Density  MOD    A  For an AEC detector used in the contact configuration  perform three exposures using 2  4   and 6 cm thicknesses of a homogeneous material  The exposures ar
13. e to be performed using  an AEC mode clinically used at each of the thicknesses  For example  if a facility typically  uses fixed kVp mode at 2 cm  fixed mA mode at 4 cm and OPDOSE mode at 6 cm  then  the medical physicist should use these same modes at those thicknesses when conducting  the AEC performance test  Note  Even if a facility clinically uses more than one AEC  mode at a particular thickness  no more than one of the AEC modes should be tested at  each thickness to establish the MOD  For example  if a facility clinically uses both the fixed  kVp and the AOP CONTRAST modes at 2 cm  the medical physicist should use the more  commonly used of these modes to determine the MOD    B  Measure the optical density of the images obtained at 2  4 and 6 cm  total of three images   and average them  This is your MOD     Step 2  Determine if the AEC detector used in Step 1 is within the regulatory action limit of     0 15  OD of the MOD      0 30 OD if done before 10 28 2002     Contains Nonbinding Recommendations    A  Check to see that all three of the optical densities obtained in Step 1B are within the action  limit when compared to the MOD  B  If ALL three ODs are within the action limit AND no other AEC modes are clinically used  in the 2 to 6 cm range  then this AEC detector has passed  The medical physicist then  needs to repeat Steps 1 and 2 for each additional AEC detector clinically used in the 2 to 6  cm range  See question  5 for additional guidance on testing multip
14. elected by the unit for the phantom exposures  Even small variations in kVp may lead to  significant variations in the mAs values obtained  While small variations in kVp are to be expected  when using the Full  Auto mode  large variations in kVp  greater than 1 kVp of the value usually  obtained  may indicate a problem and should be further evaluated  Facilities using the Full  Auto  mode that wish to monitor kVp and or mAs may want to establish baseline mAs values  corresponding to the specific kVp values usually encountered during phantom testing  In this way   they can account for the mAs variability that may be caused by small changes in kVp     Note about facility phantom QC  If the facility typically uses the Full  Auto AEC mode for its clinical  examinations  it must use this same AEC mode for its weekly phantom QC test     15    Contains Nonbinding Recommendations    Quality Assurance Records    Citation    900 12 d  2   Quality assurance records  The lead interpreting physician  quality control technologist  and  medical physicist shall ensure that records concerning mammography technique and procedures  quality  control  including monitoring data  problems detected by analysis of that data  corrective actions  and the  effectiveness of the corrective actions   safety  protection  and employee qualifications to meet assigned  quality assurance tasks  are properly maintained and updated  The quality control records shall be kept for  each test specified in paragraphs 
15. equirements     Question 11  When performing the weekly phantom image test must we  monitor kVp and or mAs     No  The only requirements on the weekly phantom image test are that the phantom image achieve  at least the minimum phantom scores established by the accreditation body and must be within the  action limits established for the three optical density requirements  FDA is aware that many facilities  are monitoring kVp and or mAs as part of their weekly phantom QC testing  This is not required   If a facility uses the Full  Auto mode and monitors kVp and or mAs  it will probably observe that   over time  the Full  Auto mode leads to small variations in the kVp selected by the unit for the  phantom exposures  Even small variations in kVp may lead to significant variations in the mAs  values obtained  While small variations in kVp are to be expected when using the Full  Auto mode   large variations in kVp  greater than 1 kVp of the value usually obtained  may indicate a problem  and should be further evaluated  Facilities using the Full  Auto mode that wish to monitor kVp  and or mAs may want to establish baseline mAs values corresponding to the specific kVp values    17    Contains Nonbinding Recommendations    usually encountered during phantom testing  In this way  they can account for the mAs variability  that may be caused by small changes in kVp     Mobile facilities should be aware of the following if they are monitoring mAs as part of their post   move pre exam tes
16. hall be  at least 1 20 when exposed under a typical clinical condition     ii  The optical density of the film at the center of the phantom image shall not change by more than   0 20 from the established operating level     iii  The phantom image shall achieve at least the minimum score established by the accreditation body  and accepted by FDA in accordance with 900 3 d  or 900 4 a  8      iv  The density difference between the background of the phantom and an added test object  used to    assess image contrast  shall he measured and shall not vary by more than  0 05 from the established operating  level     Question 10  We perform our weekly phantom images using an AEC mode  different from the Full Auto AEC mode that we typically use for patients  Is this  acceptable toward meeting the requirement     No  If the facility clinically uses the Full  Auto AEC mode for its standard breast patients  the  weekly phantom images must be obtained using that mode  FDA requires the weekly phantom  image be produced using the same clinical conditions that are used for its patients with the standard  breast  compressed breast thickness of 4 2 cm  with breast tissue consisting of approximately 50   adipose  fat  tissue and 50  glandular tissue in composition   Prior to performing mammography  on patients  the phantom image must achieve at least the minimum phantom score established by the  accreditation body and must be within the action limits established for the three optical density  r
17. ique factors  kVp  AEC mode  target filter  and density control setting  for the  different breast thicknesses and compositions so that optical densities  OD  within     0 15 OD       0 30 OD if done before October 28  2002  of the MOD under AEC testing conditions can be  produced     Question 9  A facility only performs screening mammography and never  performs any magnification studies  Must the medical physicist test the AEC in  the magnification configuration  even though the unit won   t be used in that  configuration by that facility     No  The intent of the regulation is to ensure that the AEC mode is operable in all equipment  configurations used clinically by the facility  The term  operable   means the AEC must meet the  performance requirements of 900 12 e  5 G  within the 2 to 6 cm range  One way is to have the  AEC tested in all the configurations provided by the system  An alternative method is to ensure that  the facility does not clinically use the AEC in those configurations not previously tested by the    10    Contains Nonbinding Recommendations    medical physicist  This can be accomplished by placing a label on the unit   s control panel listing the  configurations that cannot be used because they were not tested  These non operational  configurations must also be identified in the facility   s quality assurance records     Question 10  A facility   s x ray unit has a single AEC detector that may be  moved to any of three positions along the chest wall t
18. le AEC detectors    C  If ALL three ODs are within the action limit AND the facility clinically uses an additional  AEC mode s  in the 2 to 6 cm range  other than the ones used to originally establish the  MOD   the facility must test the additional AEC modes  The medical physicist needs to test  EACH additional AEC mode s  at any ONE clinically used thickness in the 2 to 6 cm  range  If the OD s  is within the action limit when compared to the MOD  then this AEC  detector has passed  The medical physicist then needs to repeat Steps 1 and 2 for each  additional AEC detector clinically used in the 2 to 6 cm range  See question  5 for  additional guidance on testing multiple AEC detectors    The medical physicist does not have to test the other clinically used equipment configurations during  the annual physics survey  but will have to test these configurations whenever a mammography  equipment evaluation involving the AEC is performed     Question 4  During the mammography equipment evaluation  must the medical  physicist test the AEC performance in all equipment configurations used  clinically by the facility or can it be limited to the contact configuration  What  action limits apply     During a mammography equipment evaluation  the AEC must be operable in all equipment  configurations  contact  magnification  and various image receptor sizes  used clinically by the  facility  The term  operable   means the AEC must meet the performance requirements of  900 12 e  5 G  withi
19. n the 2 to 6 cm range  Compliance with this requirement may be demonstrated  by any of the following three methods     l  Confirming AEC performance in the contact configuration  In the contact configuration  the  AEC must maintain the film optical density  OD  over the 2 to 6 cm range within the action limit of      0 15 OD      0 30 OD if done before October 28  2002  of the MOD  See question 3 for  additional guidance     AND   Confirming AEC performance in all other clinically used configurations  This can be done by  demonstrating that the AEC meets the density and reproducibility limits established by the  manufacturer for those other configurations    Note  Method  1 can be used only in those cases where the manufacturer has established AEC  performance standards for the non contact configurations provided     Contains Nonbinding Recommendations    2  Confirming AEC performance in the contact configuration  In the contact configuration  the  AEC must maintain the film optical density over the 2 to 6 cm range within the action limit of      0 15 OD  4   0 30 OD if done before October 28  2002  of the MOD    AND   Confirming AEC performance in all other clinically used configurations  This can be done by  comparing the contact configuration MOD with measurements obtained using the 4 cm thick  phantom in the other configurations used clinically at the facility  When results across different  configurations are compared  the facility may use the action limit of     0 30 OD
20. o nipple midline of the  breast  It cannot be placed under all areas of the breast  Would this meet the  intent of the regulation     Yes  It is not necessary that the AEC detector be mobile over the entire area of the breast     Question 11  On the facility   s x ray unit  the indication of the detector size and  position options is projected onto the input surface of the compression paddle   However  when the paddle is moved up and down  the indicated detector size  does not change with distance  Is this an acceptable indication under   900  12 b  10  ii  A      Yes  The size and positions indicated at the input surface should be indicative of the size and  positions of the detector in the plane of the detector  Compliance could be achieved by  representations permanently marked on the paddles or by a projected image that approximates the  size and position of the detector     Question 12  A facility   s unit indicates the selected position of the detector by  the relative position of the adjustment lever located on one side of the unit and  is only visible from that side of the unit  Does this meet the regulation     Yes  The relative position of the selector would be an adequate display of the detector position   and this display need be visible from only one location     Question 13  The position of the AEC detector is indicated by a knob under the  bucky that can be felt but not seen  Does this satisfy the requirement of being   Clearly indicated      Yes     Question 1
21. radiologic technologist can only adjust the exposure  by use of the density control  The actual names for the different AEC modes of operation will vary  with the different make and model of mammographic unit     For equipment testing involving the phantom  inspectors should use the same technique factors and  AEC mode of operation that the facility uses for its patients with the standard breast  compressed  breast thickness of 4 2 cm  with breast tissue consisting of approximately 50  adipose  fat  tissue  and 50  glandular tissue in composition   When a facility typically uses the Full  Auto AEC mode  for its clinical examinations  the inspector should make an exposure of the phantom using the Full   Auto AEC mode and record the kVp selected by the x ray system  This same kVp should be used  when the beam quality  HVL  testing is conducted in the manual mode of operation  In the event  that the displayed kVp after the exposure with the phantom has a three digit display  e g     25 7 kVp   but the manual mode only allows selection of two digits  e g   25 kVp   round up or  down based on the final digit  example  for 25 1 to 25 4  use 25 kVp  for 25 5 to 25 9  use   26 0 kVp      FDA is aware that many facilities are monitoring kVp and or mAs as part of their weekly phantom  QC testing  This is not required  If a facility uses the Full  Auto mode and monitors kVp and or  mAs  it will probably observe that  over time  the Full  Auto mode leads to small variations in the  kVp s
22. s meant by the terms    AEC        AEC mode        mean optical  density     and    configuration        In its Diagnostic X ray Performance Standard  FDA defines an automatic exposure control  AEC   as a device that automatically controls one or more technique factors in order to obtain a desired  quantity of radiation at a pre selected location  Such a device would automatically terminate the  exposure when the selected quantity of radiation had been delivered  The AEC may control the  selection of target material  focal spot  filter material  time  mA  mAs  kVp or a combination of any  or all of these factors     Contains Nonbinding Recommendations    AEC mode refers to the type of AEC being used  Typically available AEC modes can range from  fixed kVp and mA  where the kVp and mA are selected by the operator and the time is varied by  the AEC   to fixed kVp  where the kVp is selected by the operator and the mAs is varied by the  AEC   to various AEC modes in which all factors are varied by the AEC  Some of the more  automated AEC modes are known by brand names such as BACE  OPDOSE  AUTO FILTER  and AOP     Mean Optical Density  MOD  means the average of the optical densities measured on the images  produced with a given equipment configuration during the AEC performance test using 2  4  and 6  centimeter thicknesses of a homogeneous material     For AEC testing purposes  the only equipment configurations that need to be tested are the contact  configuration  the magnifica
23. s test all AEC detectors for AEC  reproducibility in mammography units with multiple AEC detectors in a single  cassette holder  bucky      No  Because the AEC detectors are also being evaluated as part of the AEC performance test   only a single detector per bucky needs to be tested by the medical physicist  Units that have  different AEC detectors in different buckys  e g   different AEC detectors for the different size  cassette holders  will need to have one detector in each bucky tested for reproducibility     Question 2  Must medical physicists test AEC reproducibility in all clinically  used AEC modes     No  The medical physicist can limit AEC reproducibility testing to the AEC mode used most  commonly for the standard breast     14    Contains Nonbinding Recommendations    Phantom Images Exposed in a Fully Automatic AEC  Mode  if that is the Clinically used Technique    Many mammographic x ray systems have more than one AEC mode of operation  A commonly  used AEC mode requires the radiologic technologist to set a specific kVp value with the AEC  automatically determining the mAs for the exposure  commonly called fixed kVp or Auto mAs  modes   The radiologic technologist can vary the exposure in this AEC mode by setting the kVp   adjusting the density control setting  or both  A more advanced AEC mode is where the system  automatically controls the kVp  mAs  and in some cases  target  filter combination  commonly called  the Full  Auto mode   In this latter mode  the 
24. so be tested at all clinically used thicknesses outside this range and  that the action limits specified in the regulations be applied to the extended test  If the unit cannot  meet these action limits outside the 2 to 6 cm range  FDA recommends that a technique chart be  developed showing appropriate technique factors  kVp  AEC mode  target filter  and density  control setting  for the different breast thicknesses and compositions so that optical densities  OD   within     0 15 OD  4   0 30 OD if done before October 28  2002  of the MOD under AEC  testing conditions can be produced     During the mammography equipment evaluation  as defined in 900 12 e  10    the medical physicist  must evaluate the AEC in all clinically used configurations  See Question 4   Section 900 12 e  10   requires that the AEC meet the requirements of 900 12 b  and  e   Under 900 12 b  10   the AEC  is required to be  operable  under  configurations provided   The term  operable   means the AEC  must meet the performance requirements of 900 12 e  5 G  within the 2 to 6 cm range  FDA also  recommends that in addition to the required testing in the 2 to 6 cm range  the unit also be tested in  all configurations at all clinically used thicknesses outside this range and that the action limits  specified in the regulations be applied to the extended test  If the unit cannot meet these action limits  outside the 2 to 6 cm range  FDA recommends that a technique chart be developed showing  appropriate techn
25. st Burdensome Approach    We believe we should consider the least burdensome approach in all areas of medical device  regulation  This guidance reflects our careful review of the relevant scientific and legal requirements  and what we believe is the least burdensome way for you to comply with those requirements   However  if you believe that an alternative approach would be less burdensome  please contact us so  we can consider your point of view  You may send your written comments to the contact person    Contains Nonbinding Recommendations    listed in the preface to this guidance or to the CDRH Ombudsman  Comprehensive information on  CDRH s Ombudsman  including ways to contact him  can be found on the Internet at    http   www  fda gov cdrh resolvingdisputes ombudsman html     Contains Nonbinding Recommendations    Automatic Exposure Control  AEC  Performance Testing      Annual Physics Survey and Mammography Equipment  Evaluation    Citations     900 12 e  5  i  A  B  C   Annual quality control tests  Facilities with screen film systems shall perform the   following quality control tests at least annually     i  Automatic exposure control performance    A  The AEC shall be capable of maintaining film optical density within  0 30 of the mean  optical density when thickness of a homogeneous material is varied over a range of 2 to 6 cm and the  kVp is varied appropriately for such thicknesses over the kVp range used clinically in the facility  If  this requirement cannot be
26. st can limit testing in the contact  configuration to one image receptor size  usually the small size   However  FDA does recommend  that in addition to this required testing  the medical physicist also measure the optical density  obtained using the large image receptor and a 4 cm thick homogeneous material and compare it to  the mean optical density  MOD  obtained for the small image receptor  When results across  different size image receptors  different equipment configurations  are compared  the physicist  should use the action limit of     0 30 OD     Question 8  The regulations in 900 12 e  5  i  require that an x ray unit pass an  annual test for AEC performance using 2  4  and 6 centimeter thicknesses of a    Contains Nonbinding Recommendations    homogeneous material  If a unit is used clinically at combinations of kVp and  filtration that include thicknesses outside the 2 to 6 cm range  must it meet the  AEC performance requirements at the thicknesses where it operates and must it  be tested at those technique factors under the annual quality control  requirements  What about AEC performance testing during a mammography  equipment evaluation     During the annual physics survey  the unit is not required to meet the AEC performance action limit  outside the 2 to 6 cm range and the medical physicist is not required to test the AEC using  thicknesses outside this range  However  we recommend that in addition to the required testing in  the 2 to 6 cm range  the unit al
27. tablish  national quality standards for mammography  The MQSA required that to provide mammography  services legally after October 1  1994  all facilities  except facilities of the Department of Veterans  Affairs  must be accredited by an approved accreditation body and certified by the Secretary of  Health and Human Services  the Secretary  or by an approved State certification body  The  authority to approve accreditation bodies  State certification bodies  and to certify facilities was  delegated by the Secretary to the FDA  On October 28  1997  the FDA published the MQSA  final regulations in the Federal Register  The final regulations  under which mammography facilities  are currently regulated  became effective April 28  1999  The FDA compiled all final guidance  related to MQSA into a computerized searchable Policy Guidance Help System in November  1998  The Policy Guidance Help System is available on the Internet at   www fda gov cdrh mammography robohelp start htm   This compliance guidance document contains guidance to update the Policy Guidance Help System   This document deals with new and previously issued guidance about the Automatic Exposure  Control  AEC  component of mammography units     Guidance information is periodically updated  Individuals wishing to get automatic notification of  such updates may subscribe to our E mail ListServ by visiting http   list nih gov cgi   bin wa SUBED1 mammography_cdrh 1 amp A 1 and following the directions there     The Lea
28. ting  Performing the post move pre exam test in the Full  Auto mode may be  problematic  due to the variability of kVp and mAs as previously mentioned   In these cases  the  facility may     de Use the AEC mode  with fixed kVp  to perform the post move pre exam test  even if they  use the Full  Auto mode for their patients with the standard breast  Note  The weekly phantom QC  test must be performed using the same clinical conditions that the facility uses for its patients with the  standard breast    OR  2  Use the Full  Auto mode and establish baseline mAs values corresponding to the specific  KVp values usually encountered during phantom testing  If the mAs value is within 10  of the  baseline value for the post exposure kVp value  the unit has passed that portion of the post move   pre examination test     Mammography Equipment Evaluations    Question 2  When performing an annual physics survey or mammography  equipment evaluation on a unit with multiple target filter combinations  what  tests or measurements must be performed for each combination     For a unit with multiple target filter combinations  the following tests must be performed for each  clinically used target filter combination     Focal spot condition  for different target materials  tracks  only       X ray field light field image receptor compression paddle alignment  for different  target materials  tracks  only     Beam quality and half value layer    System artifacts    18    
29. tion configuration  if used clinically   and the various image receptor  sizes  Due to advances in AEC design  the example of a target filter combination as an equipment  configuration given in the regulations is no longer applicable  Therefore  we will not enforce testing a  target  filter combination as a separate equipment configuration     Question 2  Can we continue to use technique charts after 10 28 2002     Yes  Facilities should continue to develop and use technique charts for their clinical exams   especially for AEC modes where kVp and other technique factors must be selected by the  technologist  The only place where the words    technique chart    appear in the regulations is in the  annual AEC performance test requirement  This regulation places a restriction on the use of a  specific factor of the technique chart  the density control setting  when the medical physicist is  performing the AEC test after October 28  2002  After that date  the medical physicist may not  adjust the density control setting while performing the AEC test in the 2 to 6 cm range  In other  words  the medical physicist may not use the density control setting to compensate for inadequate  performance of the AEC  When performing this test  the medical physicist may use a technique  chart to adjust other factors such as kVp  filter  anode track or AEC mode to the extent such  factors are used clinically  Ifthe AEC performance test fails  the medical physicist may create a  temporary techniq
30. ue chart that includes the appropriate density settings  in addition to the other  technique factors   This temporary technique chart may then be used by the facility for up to 30  days  or until the problem has been corrected and the equipment passes the AEC performance test   whichever comes first     When the AEC is functioning properly  the radiologic technologist shouldn   t need to adjust the  density control setting while imaging patients who are in the 2 to 6 cm range  If the radiologic  technologist needs to continually adjust the density control to achieve films of adequate density  the  AEC may need adjustment and the medical physicist should be consulted     The regulations do not restrict the use of technique charts by radiologic technologists  While a  properly functioning AEC should reduce the need to use the density setting component of a    Contains Nonbinding Recommendations    technique chart  radiologic technologists may use these charts to change the density control settings  whenever they believe it appropriate during the performance of clinical mammographic  examinations  In addition  the regulations do not preclude the use of the manual mode and under  that scenario  the use of technique charts is essential     Question 3  During the annual physics survey  how must the medical physicist  test AEC performance and what action limits apply     Due to the proliferation of mammography units with multiple AEC modes  testing of AEC  performance has become mor
31. y mean that  the AEC is the cause of the failure     No  Because the AEC performance test involves many parts of the imaging chain  the medical  physicist needs to make sure that the AEC is the part responsible for the failure  For example   problems with the processor  film emulsion or the use of different cassettes during the performance  of this test may lead to a failure that is not the fault of the AEC  Facilities are reminded  however   that whatever the cause of the failure it needs to be corrected within the appropriate time frame     Note  When conducting the AEC performance test  the physicist should try to minimize the  variation introduced by components such as film  cassette and processor  Different film emulsion  batches  cassettes or processors should not be used while conducting the AEC performance test   An exception would be where prior testing has shown that the different film emulsion batches   cassettes or processors would not introduce excessive variation in the test results  Introducing such  variability into the AEC performance test may lead to an inappropriate    failure    of this test     Question 7  For purposes of AEC testing  are different image receptor sizes  considered different    configurations        Different image receptor sizes are considered different configurations and have to be tested  separately during the mammography equipment evaluation  With respect to AEC performance  testing during the annual physics survey  the medical physici
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
PROFESSIONELLER CD-PLAYER  Panasonic AG-HMC70 Menu Information  FranceVFR`s French Riviera or Cote d`Azur    Copyright © All rights reserved. 
   Failed to retrieve file