1.         ready    indicator   e located on the    start    button  the indicator signals that the DiffSpin  is ready to process the next specimen  there is a brief delay  between cycles       See Figure 1 on the following page        acceleration     amp     duration    control dials   e adjust the acceleration and the duration of the cycle  They are  located on the side of the instrument and are usually set at their  mid point  the letter    C     for all blood samples  Setting    C     should provide adequate films for normal hematocrits  Turn  either control toward    E    to increase the spin duration or the  acceleration  These adjustments may be necessary for blood  samples from patients with disease states that result in major  changes in their blood flow characteristics      Duration  0 4   3 0 seconds    Acceleration  Final rotational speed is inconsequential   Acceleration of the platform  cradle  is  responsible for the quality of the blood film     3 1 Controls and Indicators  cont    NOTE  Some laboratories prefer  acceleration  setting of  D      To adjust the    acceleration    control to a slower setting  first  change the control to the desired setting  Then discharge the     memory    of the previous setting  Do this by placing a Slide  Holder  with slide  inside the DiffSpin  Close the cover and press  the start button  Spin only the slide and Slide Holder with no  specimen during this    reset    cycle  The next cycle will now  operate at the new speed and 
2.  THERE IS NO WARRANTY OF MER   CHANTABILITY OR OF FITNESS FOR ANY INTENDED PUPOSE  AND THAT THERE ARE NO OTHER REMEDIES OR WARRANTIES   EXPRESSED OR IMPLIED  WHICH EXTEND BEYOND THE DE   SCRIPTION ON THE FACE OF THIS AGREEMENT   NO AGENT OR  EMPLOYEE OF STATSPIN IS AUTHORIZED TO EXTEND ANY  OTHER WARRANTY OR TO ASSUME FOR STATSPIN ANY LIABIL   ITY EXCEPT AS SET FORTH ABOVE     19    Stat Spr       An IRIS Company  85 Morse Street telephone 800 782 8774  Norwood  Massachusetts 781 551 0100  02062 USA fax 781 551 0036    http   www statspin com    
3.  one hour of preparation with either a  Romanowsky stain  containing fixatives or fix within one hour with   water free   i e    lt 3  water  methanol for later staining     4 2    Operation    1        Insert a clean slide Product No  SL72   frosted edge out  into a  disposable Slide Holder  Product No  SC01     Place the Slide Holder  with slide  onto the cradle by  placing the holes in the tabs  located on the sides of the  Slide Holder  over the silver colored mounting posts and  press down gently until the Slide Holder comes to rest  flush on top of the cradle   See Figure 2      IMPORTANT  The labeled end of the slide should rest  between the slide retaining posts on the Cradle     With a StatSpin Transfer Pipette  Product No  PD50   place a  single drop of blood into each of the two holes located on the  top of the Slide Holder  See Section 3 2   The two drops must  pool together to ensure uniform coverrage  this takes a second  or two  Be careful not to leave any blood on the outside of the  disposable Slide Holder before spinning     Spin the slide immediately after adding the blood     Close cover firmly until it latches  Press    start    button  The  cycle is less than 3 seconds     When unit comes to a stop  squeeze the latch on the cover  lift  upward to open   See Figure 3 below     Figure 3       4 2    4 3    Operation  cont    7  Remove the disposable Slide Holder containing the slide     8  Remove the slide and air dry or oven dry as your lab procedures    
4.  products and believes that the quality  utility  and value we build into each instrument will provide complete customer  satisfaction  StatSpin warrants to the original purchaser that this  instrument will be free from defects in material or workmanship under  normal and proper use for a period of one year from the date of installa   tion  StatSpin will at its option  repair or replace any unit covered  under this warranty returned to StatSpin with shipping costs prepaid   Repaired or replaced instruments supplied under this warranty will  carry only the remaining portion of the original warranty  For warranty  terms and conditions outside the United States  contact your Autho   rized StatSpin Distributor     StatSpin shall not be responsible for damage to a StatSpin instrument  resulting from misuse  negligence  accident  or resulting from unautho   rized repairs  alterations  or improper installation  Additionally   StatSpin shall not be liable for any claims  losses or damages of any  third party or for lost profits or indirect or consequential damages of any  kind even if StatSpin has been advised of the possibility of such  damages  The purchaser shall be deemed liable for any and all claims   losses or damages incurred by the use or misuse of the StatSpin  instrument by the purchaser  its employees or others  following receipt  of the instrument     THIS WARRANTY IS GIVEN EXPRESSLY AND IN LIEU OF ALL  OTHER WARRANTIES  EXPRESSED OR IMPLIED  THE PUR   CHASER AGREES THAT
5.  when  assembled     10  REFERENCES    1   Blood Sample Preparation for Automated Differential  Systems   Rogers  Charles H   American Journal of  Medical Technology  Vol  39  No  11 435 442  November  1973     2   The Automated Differential  Pattern Recognition Sys   tems  Precision  and the Spun Smear   Kingsley   Thomas C   Blood Cells  483 487  1980     3   The Use of Wedge and Spun Blood for Reference WBC  Differential Counts   Abstract presented at International  Society of Laboratory Hematology  6th International  Symposium on Technological Innovations in Laboratory  Hematology  Berend Houwen  et al   Santa Fe  New  Mexico  March 1993     4     Causes for Poor Leukocyte Distribution in Manual  Spreader Slide Blood Films   Steine Martin  E A    American Journal of Medical Technology  Vol  46  No  9   May 1980     5   Automation and Laboratory Hematology  Impact on  Quality and Cost Effectiveness   B  Houwen  et al    American Society of Clinical Pathology   College of  American Pathologists  Hematology Seminar  Orlando   Florida  October 19  1993     6   Flow Cytometric White Blood Cell Differential  A Pro   posed Alternative Reference Method   L  Lebeck  et al    Sysmex Journal International  3  61 69  1993    7   A Critical Evaluation of the Manual   Visual Differential    Leukocyte Counting Method   Koepke  J A   et al    Blood Cells  11  173 186  1985     1 15 02 54 002663 001 Rev B    18    11 0    WARRANTY FOR THE DIFFSPIN CENTRIFUGE    StatSpin takes pride in its
6. DiffSpin     Operator s Manual    Stat Spin    An IRIS Company    Operator   s Manual    for    DiffSpins Slide Spinner    Model Number    M700 10  M700 12  M700 22    Covered under one or more of the following US patents     StatSpin and accessories   4 846 974   4 981 585  DiffSpin Slide Spinner   5 326 398    Other patents applied for     StatSpin is a registered trademark of StatSpin  Inc  an IRIS Com   pany  International Remote Imaging Systems      Copyright 2002 Printed in U S A     10     11     TABLE OF CONTENTS    INTRODUCTION AND THEORY OF OPERATION  1 1 Introduction  1 2 Clinical Utility    UNPACKING AND INSTALLATION   2 1 Check for damage sustained  during transport   2 2 Check for complete packaging   2 3 Installing the DiffSpin slide spinner   2 4 Power connection    WORKING WITH DIFFSPIN  3 1 Controls and indicators  3 2 Safety features and precautions    INSTRUCTIONS FOR USE   4 1 Collection and preparation of blood sample  4 2 Operation   4 3 Sample Volume   4 4 Analysis of the slide   4 5 Typical results   4 6 Slides    MAINTENANCE  5 1 Cleaning instructions    SERVICE   TECHNICAL DATA   NOTES   8 1 Safety Warning   8 2 General Note  TROUBLESHOOTING GUIDE  REFERENCES    WARRANTY FOR THE DIFFSPIN    Www w    OA    O ON    12    12    13    14    14    15    15    16    18    19    Warranty Card Inside back cover    UNIVERSAL PRECAUTIONS    Universal Precautions should be followed on all specimens   regardless of whether a specimen is known to contain an  inf
7. MPORTANT  The DiffSpin contains components  which have a potential SHOCK HAZARD  The unit  should only be opened by trained and qualified  personnel     7  TECHNICAL DATA    Power Requirements   Product   DS02 10 100 VAC  50 60 Hz  DS02 12 120 VAC  50 60 Hz  DS02 22 240 VAC  50 60 Hz    Running Time   0 4 to 3 0 seconds    Slide Holder   Disposable  accepts 1 x 3 inch microscope slides     Dimensions   6 5 inches  16 5 cm  diameter by 4 5 inches  11 4 cm   high     Weight   3 7 pounds  1 7 kg     14    8 1    8 2    8  NOTES    Safety Warning    This DiffSpin left the factory in perfect order from the point of  view of work safety  To keep it that way and to ensure safe  operation  the user must follow the instructions and warnings  given in this operating manual     General Note    The data and information provided in this manual correspond  to the state of knowledge existing at the time of printing this  manual  Any important changes will be taken into account  in the next edition of this manual or with the Package Insert  sheet included with the Slide Holders  In every case  the  respective package insert should be regarded as definitive     9  TROUBLESHOOTING GUIDE    DESCRIPTION POSSIBLE CAUSE OPERATOR RESPONSE  Power light not lit Power cord Verify that the power  unplugged from cord is plugged   DiffSpin or from securely into the   outlet  outlet and into the  DiffSpin    Non functioning Verify that outlet is   outlet  functioning properly    LED burned out Verify Diff
8. Spin is  still operational    Cradle does not Start button not Retry by pushing  rotate depressed fully  start button    Cover not closed Verify that cover is   properly  closed tightly    Unit not    ready     Verify    ready    light is  lit before initiating  cycle    Poor Quality Unclean slides  Verify that slides are  Blood Films without dust and or    fingerprints  See  Section 4 6  Slides      Inappropriate Verify sample volume   sample volume  refer to Section 4 3     16    DESCRIPTION POSSIBLE CAUSE    Poor Quality  Blood Films    Time and speed  settings not correct     Excessive amount  of blood     Blood escapes  Slide Holder    Blood on outside  of holder     Slide Holder does Slide holder not    not lay flat on Cradle positioned correctly     Slide and or  Holder dislodges  during spinning    Slide holder not    Smear area covers Slide placed  frosted edge of slide incorrectly into  Slide Holder     17    positioned correctly     OPERATOR RESPONSE    Verify position of  time and speed knob  at recommended    C     mark for average  hematocrit levels     Verify drops only large  enough to cover slide     Verify that no blood is  present on exterior of  slide holder after pipetting  blood onto slide     Check Figure 2  Section 3 2  and verify Slide Holder and  slide are oriented properly  on cradle     Check Figure 2  Section 3 2  and verify Slide Holder and  slide are oriented properly  on cradle     Verify that frosted edge  of slide protrudes from  Slide Holder
9. ectious agent  Universal Precautions have been stated by  Center for Disease Control in 1987 and updated in 1988 and    are reprinted in National Committee for Clinical Laboratory  Standards proposed guidelines     Protection of Laboratory  Workers From Occupationally Acquired Infections     Approved  Standard  NCCLS Document M29 A2  2001  NCCLS  National  Office  940 West Valley Road  Suite 1400  Wayne  PA 19087   1898  USA         1 1    1 2    1  INTRODUCTION AND THEORY OF OPERATION  Introduction    The DiffSpin Slide Spinner is a special purpose instrument for produc   ing uniform monolayer blood smears or films on microscope slides     DiffSpin is compact and easy to use  allowing the user to rapidly pre   pare reliable blood smears that are technique insensitive  The benefits  of DiffSpin are  1  it   s ability to produce consistent monolayer smears   and 2  the convenience and safety of disposable Slide Holders which  collect and contain the excess spun off blood and minimize aerosols     IMPORTANT  Insure the DiffSpin is on a level surface   Clinical Utility    The differential white blood cell count has long been recognized as  being useful in the diagnosis and prognosis of disease and the monitor   ing of therapy     There are five potential sources of error or methodological variables in  making the differential count     1  Errors in interpretation or classification of the cells    2  Statistical sampling errors    3  Non random or non uniform distribution of cell
10. ilable for count   ing  Depending on the aperture of the microscope  there will be  over 20 000 high power fields  100x  available to the user  This area  is always larger than the    feathered    area used on the wedge  smear     Red blood cell morphology varies with the method of preparation   2   Spun films may display slightly different morphology in some  areas of the slide than smears prepared by the wedge method   However  similar as with the wedge method the user will easily  identify those areas that enable accurate identification of red cell  morphology     Distribution of white blood cells is uniform through out a spun  blood smear  1   The uniformity of blood smears prepared by  DiffSpin depends on the quality of the microscope slides     Slides    The uniformity of blood smears prepared with either manual or  automated systems depends on the quality of the microscope  slides  For best results use StatSpin microscope slides  Prod   uct Number SL72  StatSpin slides are clean  dustfree and de   hydrated  which is a prerequisite for consistent results  Other  various commercially available pre cleaned slides may result in  inadequate blood films if the slide surface is covered with dust or  debris particles  or if the slides have been exposed to humid  storage conditions     Always use clean desiccated slides and handle them carefully     Avoid touching slide surface  Oil from fingers on the slide sur   face will result in a non uniform film     12    4 6    5 1    S
11. in Slide Preparation Kit  Product Number DSPK    Installing the DiffSpin Slide Spinner    Place the DiffSpin on any level surface acceptable for a laboratory  instrument  Position the DiffSpin away from direct sunlight and  sources of heat and cold     Power Connection    Before connecting the DiffSpin to the power  please check that the  nominal voltage  V   and frequency  Hz  specified at the rear of the  DiffSpin matches your outlet voltage and frequency  If it does not   please contact your supplier     WARNING  If the voltages do not match  do not connect  the DiffSpin to the power source     If the voltages match  connect the DiffSpin to a grounded outlet  using  the cable provided  This protective grounding must not be neutralized  by using an extension cord without a ground lead     Once the DiffSpin is connected to the power it is ready for use  The  DiffSpin does not have an ON OFF switch     3 1    3  WORKING WITH DIFFSPIN       Controls and Indicators       stop   open    button   e interrupts the cycle and stops the centrifugation  The    stop   open     button is rarely used as the cycles are short  It can be used to  release an accidentally latched cover        power    indicator  e green LED  illuminated when the DiffSpin is connected to a power  source     The DiffSpin is not provided with an    on off    switch  The unit is  normally left plugged in and    on     The cover is usually left down  but not latched        start    button  e initiates a spin cycle
12. indicate  When slide is dry it may be stained or processed in the  usual manner     Dispose of the contaminated Slide Holder as you would other  biohazardous material     IMPORTANT  Use Slide Holders only once  The Slide Holders are  designed for single use only     NOTE  There is a built in time delay between cycles  Wait until the     ready    indicator is lit before initiating the next cycle   See Figure 1  Section 3 1  A new cycle can not be initiated until the time delay  has elapsed     Sample Volume    For best results  use 50 ul of whole blood placed in each of the holes  in the slide holder  The drops should  pool  to create a bar formation   This bar formation will adequately cover the slide and provide good  viewing area     The density of the cellular layer depends on the sample s viscosity   cell content and surface tension of the glass slide  The volume  applied to the slide also affects the cell density on the blood film   Operator s preference of volume varies greatly     Sample Volume continued next page     4 3 Sample Volume  cont      The following are some suggestions for sample volumes    e For lower cell density  some laboratories report best results with  total volume of 40 ul  i e  0 040 ml  of blood pipetted into each  hole of the slide holder  With small quantities of blood  the slide  will not be uniformly coated with a film  however  the spacing  between the cells will be greater something that may be  preferred for faster cell recognition     e Fo
13. lides  cont      Slides must be fresh  Non uniform slide coverage  e g  streaking  or  fingering   is a result of poor quality slides that have degraded  resulting in increased surface tension  e g  glass slides stored for  months without protection from humidity and temperature   This  increased surface tension is found with poor quality slides  slides  which were exposed to humidity  or slides stored for excessive  lengths of time  If this condition persists try a new box of StatSpin  slides  Product Number SL72      5  MAINTENANCE  Cleaning Instructions    The DiffSpin requires no maintenance beyond routine cleaning   Prior to cleaning  the power cord should be disconnected   Clean the exterior surface and inner surface or bowl area with a  10  bleach solution on a damp cloth  Never have liquid inside  the bowl  Do not use abrasives or solvents  Clean the clear  cover with a glass plastic cleaner if required     Any machine or accessory containing accumulated blood and or  other chemical deposits must be cleaned prior to shipment to  the manufacturer distributor for service  This decontamination is  required by Federal Law  Title 48 and 49 of the Code of Federal  Regulations  and in accordance with the E P A s Regulations for  Biohazard Waste Management  This decontamination cannot  be performed by StatSpin personnel     13    6  SERVICE    The DiffSpin has no user serviceable parts  Should the  instrument require service please consult your Warranty for  instructions    I
14. n was developed to eliminate aerosol formation while pro   viding the higher quality of a spun slide  Using the DiffSpin  when the  slide rotates  the excess blood is captured by the slide holder  When  used according to the instructions  independent studies indicate there  is no detectable aerosol generation  5   In addition  because of the  eccentric spinning platform  slides prepared with the DiffSpin are not  completely covered with blood  Therefore  prepared specimen slides  are readily handled and easily labeled for identification     When the slide rotates  the excess blood is thrown off and captured  by the Slide Holder  The remaining blood wets the slide and the  thickness of the layer left behind depends on the sample   s viscosity   cell content and surface tension  Though the density of the cellular  layer may depend on the sample s viscosity  cell content and surface  tension of the glass slide the volume applied the the slide also affects  the cell density  See Section 4 3 Sample Volume     2 1    2 2    2 3    2 4    2  UNPACKING AND INSTALLATION    Check for Damage Sustained During Transport    The DiffSpin Slide Spinner and its accessories are delivered in one  package  Save the shipping carton in the event the instrument needs  to be returned for service  Notify the carrier and your DiffSpin distribu   tor if damage in shipment is evident     Check for Complete Packaging    e Centrifuge   e     Line Cord   e Operator   s Manual w Warranty Card   e     DiffSp
15. r higher cell density  and a larger area of the slide to be  covered   some laboratories apply 2 x 65 ul  i e   2 x 0 065 ml   drops of blood  Although this produces a more fully covered  slide  this much surface area is not really necessary  Do not  exceed 75 ul per drop as blood will exit the slide holder and  cause excessive aerosols  Examination Area shown below is  best for viewing red cell morphology     Figure 4       Examination Area    Representation of 2 x 50 ul drops    NOTE  The two drops of blood must merge or pool together to  ensure uniform coverage     Sample Volume continued next page    10    4 3 Sample Volume  cont      e The amount of blood added to create a monolayer film with the  DiffSpin is a matter of preference  The smaller volumes create a  more spread out cell monolayer under the microscope  but may not  have the same slide coverage as larger volumes     Also  other cell suspensions may also be used to create a mono   layer on a microscope slide     Pipettes    For best results use StatSpin pipettes  with a premeasured drop  size of 50 ul  Order StatSpin Product Number PD50     11    4 4    4 5    4 6    Analysis of the Slide    The peripheral blood smear made with your DiffSpin slide spinner  should be read according to your laboratory protocol     Typical Results    About 70  of the slide will be covered with a monolayer of blood  cells  The areas covered on typical DiffSpin slides are shown in  Figure 4  Nearly all of these covered areas are ava
16. s on the slide    4  Distortion of blood cell morphology by inappropriate  anticoagulant  blood storage and handling    5  Blood film preparation technique     The spinner technique addresses the third and fifth sources of these  potential errors  Amore complete explanation follows     1 2    Clinical Utility  cont      The traditional technique to prepare the microscope slide  the manual  wedge smear  introduces a source of error as it causes non random  distribution of cells which is dependent on cell size  1   This gener   ally results in an under reporting of monocytes  7  and a slight over   reporting of lymphocytes  This phenomenon is not surprising with  the visually apparent non uniform distribution gradient established using  the wedge technique     The slide spinner technique addresses both the non uniform distribu   tion of cells and preparation technique  By using a slide spinner  all  areas covered by cells are good  no searching for a  good  area   Total time to achieve accurate results is less  Indeed  the results  achieved with spun slides are more accurate than those prepared  using the wedge smear technique  6      Slide spinners were first developed in association with automated  differential counters to produce uniform monolayers independent of  operator skill  1   Using these early generation slide spinner instru   ments produced superior quality slides  however  the advantages were  offset by the hazards associated with aerosol generation     The DiffSpi
17. setting     Figure 1 DiffSpin  Front View       is   SN  S   Q       Ready Indicator        Power    STOP  OPEN START    FiK                a       Vo    Stop Open Start  Button Button    5    Control Dials    3 2    Safety features and precautions    e The DiffSpin slide spinner incorporates an interlock mechanism   This mechanism prevents the DiffSpin from being operated unless  the cover is closed     IMPORTANT  Do not attempt to defeat this safety  feature of your DiffSpin instrument     e The DiffSpin prepares blood smears for microscopic examination   Use of the DiffSpin places a responsibility upon administrative  personnel to insure adequate training of operators  Do not open  the DiffSpin until the slide has stopped rotating     WARNING  Any rotating device can generate a  mechanical safety hazard  For the user   s protection   the unit will NOT operate when the cover is open     e Donot leave any blood on the outside of the disposable Slide  Holder before spinning     e The DiffSpin is not explosion proof  Do not use in a potentially  explosive atmosphere     e StatSpin slides  1 x 3 inch  Product No  SL72     e Mount the Slide Holder so that it lies flat on the Cradle  Check  that the tabs are securely over the mounting pins  Failure to  ensure the Slide Holder is mounted properly may break the slide  and or result in damage to the instrument   See Figure 2  below     Holes for Blood Microscope Slide            Figure 2    Frosted or Label End    Rear Retaining Pos
18. ts       Slide Cradle  Mounting Posts for Slide Holder    4 1    4  INSTRUCTIONS FOR USE    Collection and Preparation of Blood Sample    The National Committee for Clinical Laboratory Standards  NCCLS   publication    Reference Leukocyte Differential Count  Proportional   and Evaluation of Instrumental Methods     Approved Standard   NCCLS Document H20 A  Vol  12 No  1  NCCLS  National Office   940 West Valley Road  Suite 1400  Wayne  PA 19087 1898  USA   indicates the following for visual differential count     Use whole venous blood collected in anticoagulanted tubes with  tripotassium ethylenediamine tetra acetate  K EDTA  1 5     0 15  mg  in liquid or powder form  per mL  see NCCLS document H3   A4      Note the specimen condition   Macroscopically visible clots are  cause for rejection of the specimen for analysis  Microscopically  visible platelet clumps consisting only of platelets are acceptable   provided their presence is noted     Slides are to be prepared within 4 hours of blood collection in  K EDTA     Blood samples are to be stored at room temperature until blood  smear has been prepared     Use StatSpin Slides  Product No  SL72     Make sure that the blood is well mixed  Use a blood tube rocker  or similar device if there is any delay in preparing the blood smear   Manual mixing is allowable as long as the blood tube is inverted  completely at least 20 times  This will mix the blood and  minimize artifacts such as rouleaux formation     Stain the film within
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
  HI 38061 Test Kit para Fosfato con Checker Disc  Composteur rotatif  RLComputing DNN User Manual v7  HIGH WALL SPLIT AIR CONDITIONER  Mode d emploi Prise en ma\n Paiement en ligne  Expert Conclusion + - TELE  Manual de usuario de la Yamaha MT01  SC-450 System Controller  Motorola FV500 Two-Way Radio User Manual    Copyright © All rights reserved. 