1. F BRAINLAB coat  Da KapellenstraBe 12   85622 Feldkirchen    Germany    phone   49 89 99 1568 O  fax   49 89 99 15 68 33    FIELD SAFETY NOTICE   PRODUCT NOTIFICATION    Subject  Cranial Navigation System     Standard Cranial Reference Array   For specific combinations an inaccuracy of more than 1mm might  be added to the registration result when exchanging Standard  Cranial Reference Arrays    www brainlab com       Product Reference  Standard Cranial Reference Array   Date of Notification  Aug 3  2012   Individual Notifying  Julia Mehltretter  MDR  amp  Vigilance Manager   Brainlab Identifier  11 12 22 TGE 2   Type of action  Device component exchange  advice regarding use of device    We are writing to advise you of the following effect Brainlab has internally determined for  specific combinations of Standard Cranial Reference Arrays when used for the exchange of  unsterile with sterile array in conjunction with the Brainlab Cranial Navigation System     This Notification letter is to provide you with corrective action information  and to inform you of  the actions Brainlab is taking to address this issue     Effect    Manufacturing tolerances of the Standard Cranial Reference Array influence the actual position  of its marker spheres  The differences between the individual arrays are supposed to be very  small  not significantly affecting navigation accuracy when the unsterile array is exchanged with  the sterile array during surgery     However  there are specific pairs 
2. and thank you in advance for your co operation   If you require further clarification  please feel free to contact your local Brainlab Customer  Support Representative     Customer Hotline   49 89 99 15 68 44 or  1 800 597 5911  for US customers  or by    E mail  support brainlab com  Fax Brainlab AG    49 89 99 15 68 33  Address  Brainlab AG  headquarters   Kapellenstrasse 12  85622 Feldkirchen  Germany     Aug 3  2012    Kind Regards     ne    Julia Mehltretter  MDR  amp  Vigilance Manager  brainlab vigilance brainlab com          Europe  The undersign confirms that this notice has been notified to the appropriate Regulatory Agency in  Europe     Attachments   Appendix A     List of affected serial numbers for article 41735  Appendix B     List of affected serial numbers for article 41730A    FORM 14 04  11 12 22 TGE 2 Rev 7 Page 3 of 5       BRAINLAB    Appendix A     List of affected serial numbers for article 41735    Article  41735 STANDARD CRANIAL REFERENCE ARRAY    O wo   O ero   Eo    www brainlab com       FORM 14 04  11 12 22 TGE 2 Rev 7 Page 4 of 5       BRAINLAB    Appendix B     List of affected serial numbers for article 41730A    Article  41730A REFERENCE CLAMP UNIVERSAL  WITH 2 ARRAYS     www brainlab com       FORM 14 04  11 12 22 TGE 2 Rev 7 Page 5 of 5    
3. ay     With this exchange an inaccuracy of more than 1 mm might be added to the registration result  if array pairs with specific combinations of tolerance limits are used     User Corrective Action   According to our records you are the owner of one or several of the potentially affected  Standard Cranial Reference Arrays     Therefore please check your Standard Cranial Reference Arrays and do not use  reference arrays with one of the serial numbers listed in Appendix A and B     Please note that for article 41730A REFERENCE CLAMP UNIVERSAL  WITH 2 ARRAYS   two reference arrays bear the same serial number  In this case both reference arrays    must no longer be used     Please remove the according reference arrays from clinical use  You will be provided with a  replacement accordingly     Where you can find the serial number        Serial number Article number    Figure 2     Where to find the Serial Number on the Standard Cranial Reference Array    FORM 14 04  11 12 22 TGE 2 Rev 7 Page 2 of 5       BRAINLAB    In any case  please continue to follow the instructions in the relevant user manuals     e The Brainlab Cranial ENT instruments are highly accurate and sensitive medical  devices and must be handled with extreme care  If you drop or otherwise damage an  instrument  return it immediately to Brainlab for testing  Failure to do so may lead to  serious injury to the patient     e Do not use damaged instruments     e Do not change the position of the headholder adapter whe
4. n attaching the sterile  reference array  This could result in inaccurate tracking and severe patient injury  www brainlab com    e Verify registration accuracy after exchanging the unsterile reference array for the  sterile reference array        e Do not move a reference array relative to the patient   s anatomy during the procedure   Changing the array   s position throws off the entire measurement coordinate system   endangering the patient     e To avoid unintentional movement of the reference array  do not place pressure on it    e lf the position of the reference array changes relative to the patient  a new registration  is required    e Anewregistration is required if the patient was repositioned after registration    e Check regularly that the marker spheres remain tightly screwed onto the pins    e Navigation accuracy critically depends on the condition of the marker spheres used     e Verify prior to use that the reflective surface of all marker spheres is in good condition   and not peeling     Brainlab Corrective Action    1  Existing potentially affected customers who own one or several of the potentially affected  Standard Cranial Reference Arrays receive this product notification letter    2  Brainlab will actively provide revised hardware to improve exchange compatibility to these  customers as of October 2012     Please advise the appropriate personnel working in your department of the content of  this letter     We sincerely apologize for any inconvenience 
5. of arrays that might add an inaccuracy of more than 1mm to  the registration result during the exchange due to the combination of their tolerance limits  The  Appendix lists the potentially affected arrays identified by Brainlab simulations     This effect could potentially cause an inaccurate display of instruments by the navigation  system in the region of interest  compared to the actual patient anatomy  If these inaccuracies  are not detected by user verification of navigation accuracy as described in the user manual   this could lead to serious injury or ineffective treatment of the patient     Figure1  Standard Cranial Reference Array    FORM 14 04  11 12 22 TGE 2 Rev 7 Page 1of5     BRAINLAB    Details   Reference arrays  e g  the Standard Cranial Reference Array  allow the navigation system to  track the location of the patient s anatomy throughout the procedure     Usually  two reference arrays are used     e One for preoperative procedures  patient registration   The reference array used for  registration may be unsterile as registration is usually performed in an unsterile envi     www brainlab com    ronment    e One for intraoperative navigation  The reference array used for navigation must be       sterilized before use and attached during draping    After the successful registration of the patient using the Brainlab navigation software it is  required to carefully remove the unsterile reference array  drape the patient and then attach the  sterile reference arr
    