        Electrical Transport Option (ETO) User`s Manual
         Contents
1.     I    reported   Rorive     R    T ingie   R    Drive Leads    where Rjeaas is the resistance due to leads and sample contacts  Table 3 1 lists the nominal drive  impedance for each current range when the drive feedback is disabled     3 2 PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011    Chapter 3 Section 3 2  Hardware ETO Module and Head    RLeads    V Drive    I sample I sample    Vsample Vsample  b  c   Figure 3 3  The equivalent circuit for ETO when drive feedback is disabled  The dashed line  represents the current source equivalent circuit        Table 3 1  ETO current drive output impedance with feedback disabled                                   Current Range Drive Output Impedance  100 mA 100 Q  10 mA 1 kQ  1mA 10kQ  100 uA 100 kQ  10uA 1MQ  1uA 10 MQ  100 nA 100 MQ          In most cases the electronics should be run with the drive feedback enabled  This will typically  yield the most accurate measurement  However  it may be useful to disabled the feedback if you  are measuring an IV curve on a sample with a non linear current voltage characteristic  such as a  diode  With feedback off the IV curve of a diode will contain more points at lower bias currents  when the device impedance is large     Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 3 3  September 2011    Section 3 2 Chapter 3  ETO Module and Head Hardware    3 2 2 Preamplifiers    The ETO head contains a programmable gain amplifier  PGA   a high gain amplifier  HGA  
2.    cooocccncccooncnoncnonononononoconacanccanccanccnnos 5 5  The immediate mode differential resistance measurement window               ceeeee 5 7  List of sequence commands available when ETO is active   ooooonoccnnccnocinoncnoncnonaconanos 5 8  The sequence command window for resistance measurements   oooccnocccoccnoncnancnnnannss 5 9  The sequence command window for differential resistance measurements             5 10  The sequence command window for IV measurementS   ocoonccnnncnnoccnoncnoncnonacnacana  ns 5 10  The sequence command window for creating a New Data File    eee 5 11   PPMS ETO User s Manual  1084 700  Rev  A1 V    September 2011    Contents  Table of Contents    Tables    Table 2 1     Table 3 1   Table 3 2     vi    A list of hardware components comprising the Electrical Transport Option for  a     PPMS and  b  VersaLab Systems isie nepaiso snieni nie as 2 1  ETO current drive output impedance with feedback disabled       o ooonconnncnnnnnncnoncn   3 3  Preamp ranges and their associated gains for both 4 wire and 2 wire measurement  MOSS it A ada 3 4  PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design    September 2011    P R E F A C E    Contents and Conventions    P 1 Overview    The preface contains the following info   e Section P 2 discusses the overall scope of the manual   e Section P 3 discusses the conventions of the manual     e Section P 4 discusses the safety guidelines for this option     P 2 Scope of the Manual    This manual discusses
3.   0 4 V 1X 10X 10X  130 mV 3X 10X 30X  40 mV 1X 100X 100X  13 mV 1X 300X 300X  4 mV 3X 300X 900X  1 3 mV 1X 3 000X 3 000X  0 4 mV 3X 3 000X 9 000X  130 uV 1X 30 000X 30 000X  40 uV 3X 30 000X 90 000X  2 Wire Mode  iene Module Gain Head Gain Total Gain  400 nA 1X 1X 1X  130 nA 3X 1X 3X  40 nA 1X 10X 10X  13 nA 3X 10X 30X  4nA 1X 100X 100X  1 3 nA 3X 100X 300X  3 4 PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design    September 2011    Chapter 3 Section 3 4  Hardware Sample Pucks    If you are measuring a sample with impedance greater than 10 MQ you will use the HZA and a  2 wire sample mounting configuration  In this configuration ETO applies a voltage and measures  the current through the sample  The preamp range is specified in Amps for these measurements   The measurement windows will automatically switch the drop down list to a selection of current  ranges when the 2 wire configuration is chosen     3 3 Cables and Jumpers    The ETO option includes one 25 pin cable to connect the module to the remotely mounted head   one 2 pin cable to power the cooling fans on the head  and one 8 pin cable to connect the head to  the sample chamber  The cable use and routing are shown in Figure 3 4  ETO kits intended for use  on PPMSs are shipped with long length module and fan cables  ETO kits intended for use on  VersaLab systems are shipped with short length cables  Both kits use the same sample cable to  connect the head to the sample chamber grey Lemo  Additional sample cables 
4.   4 2 2 Electrical Transport Option Event Log    The ETO Event Log window is displayed when ETO is activated  This window displays high   level hardware and software activity during ETO measurements  This log will also display  warnings  errors  and informational messages generated since option activation  If you have a  problem while a measurement is running  entries in the event log can help diagnose the problem     4 3 Electrical Transport Option Data Files    All of the data taken by the Electrical Transport Option is stored in a single data file  The file  name and location can be set on the Data File tab of the ETO Console or within a sequence  For  more information on changing the data file refer to Section 4 2 1 2  This section contains a list    and description of each of the data columns as well as the measurement types that use the  different data types                                      Data Type Units Description Resistance dV dI IV  Time Stamp   Time stamp associated with the x x x  row of data  Temperature K Sample Temperature X X X  Field Oe Magnetic Field  se Angle of the sample stage relative  paar e Posies deg to the magnetic field  Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 4 5    September 2011       Section 4 3    Electrical Transport Option Data Files                                                                            Chamber Pressure Torr Pressure inside the sample  chamber  ees Ohms Measure resistance  Resistance Std  Dev  Ohms 
5.   List of sequence commands available when ETO is active     5 5 1 ETO Resistance    Figure 5 6 shows the sequence command window for the resistance measurement  This window  contains the same controls as the immediate mode window  Follow the same steps outlined in  Section 5 4 1 to specify the measurement parameters  When you are finished press the OK  button and the measurement command will appear in the sequence window     Since this command allows simultaneous measurements on both channel 1 and 2 it is possible for  the two channels to have different measurement times  This can be caused by either very  different averaging times or by selecting different numbers of points to measure on each channel   In this case the sequence will wait for both channels to finish measuring before moving on to the  next sequence command     5 8 PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011    Chapter 5  ETO Measurements    AC Resistance Measurement  Channel 1  Enable Measurement  Excitation    Amplitude  10  m   Freq  15 vw  He    Preamp    Range  av Si     C  Autorange    Sample    Wiring LS eS    Measurement Configuration    Averaging    la Number of W     Time E Measurements          Data Logging       Section 5 5  Sequence Measurements    Channel 2  Enable Measurement    Excitation    Amplitude 110  m   Freq  1 Y Hz    Preamp  o    C  Autorange    Range  ay  Sample       Wiring 4 wire M  Measurement Configuration    Averagin Number of ol  Time ms  1  see Meas
6.   installation procedure  Follow the instructions on the screen to install or remove a new sample     The Sample Properties button will launch a separate window to allow you to enter properties of  the sample that you would like to store in the data file  This information is stored in the data file  header  For more information regarding the ETO data files refer to Section 4 3  Figure 4 3  shows the Sample Properties window containing the Wiring  Sample 1 Info  and Sample 2  Info tabs  The Wiring tab is used to specify the wiring configuration used on each sample  It  also lists the types of measurements that can be run with each configuration  The two Sample  Info tabs allow you to enter sample specific information such as name  material  and dimensions   For more information regarding the wiring configuration refer to Section 5 2     ETO Console       Setup   Data File Measurement       nstallation    lead Properties    Figure 4 1  Setup tab in the ETO Console window        EE  Sample Installation Wizard    System Status    296 77 K  Standby  Sealed    Instructions                Dpen Chamber    Cancel    Figure 4 2  The Sample Installation Wizard window        4 2 PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011    Chapter 4  Software    Section 4 2  Overview of Electrical Transport Option Software    Sample Configuration    Witing   Sample 1 Info   Sample 2 Info    Channel 1 Channel 2    Wing Scheme  4 wire Wiring Scheme  2 wie wv    Available Meas
7.  2 on the wiring tab  Enter any other sample specific information you desire on the  two sample tabs  This information will be entered into the header section of the data file     Switch to the Data File tab to select a new data file for your measurements  Press Browse to  enter a new file name and location  Checking the Create New File Version checkbox will add a  numeric suffix to the end of the file name if the file already exists  Press the View button to open  the data file in MultiVu  Note  The data file is not created until a measurement is initiated     5 4 Immediate Mode Measurements    After the sample puck has been loaded you are ready to take measurements in immediate mode   At this point it is a good idea to make a few test measurements in immediate mode to ensure your  sample is loaded properly  Open up a measurement window to perform the desired test  measurements  The immediate mode measurement windows are accessible through the  Measurement tab in the ETO Console  Select a measurement type and press the Launch  Measurement button  The following sections describe the process of setting up each type of  measurement     5 4 1 Taking Resistance Measurements in Immediate Mode    The AC Resistance Measurement window contains two identical sets of controls and indicators  corresponding to channel   and channel 2  Use these blocks to define the measurement  parameters and read the measurement results specific to each channel  A third section at the  bottom of the win
8.  and a  high impedance amplifier  HZA   The amplifiers used in a particular measurement are determined  both by the measurement type and the range specified in the measurement window  Four wire   measurements will use the PGA and the HGA when needed while two wire measurements will use    the PGA and the HZA at all times  The HGA has a typical noise specification of 1 nV     Hz  giving a noise floor of 10 nQ for a 100 mA excitation  The HZA  by contrast  is a sensitive amp  meter and is used to measure impedances above 10 MQ     The gain of the preamplifiers is determined in each measurement by selecting the range in the  preamp controls  The range is typically specified as a voltage  This can be interpreted as the  maximum sample voltage the system can measure at the chosen range  The lower the voltage range  the higher the total gain  In practice the system can measure up to 110  of the voltage specified by  the range  If you do not know the impedance of your sample ahead of time  you can elect to  autorange  In this case the system will automatically adjust the range based on the measured signal   Selecting autorange will delay the return of data while the system adjusts the gains to find the  appropriate range     Table 3 2  Preamp ranges and their associated gains for both 4 wire and 2 wire measurement                                                                                     modes   4 Wire Mode  TTAR Module Gain Head Gain Total Gain  4V 1X 1X 1X  1 3 V 3X 1X 3X
9.  resistivity pucks that mount into the sample  chamber  Each sample puck has a circuit board around the perimeter with 12 pads organized into  three channels  Each channel has 4 pads labeled I   I   V   V  in the standard 4 wire resistance  geometry  The circuit board has labels for each pad as well as the channel number corresponding  to each group to avoid confusion during sample mounting  ETO uses the channel 1 and channel  2 pads  There is a space in the center of the puck for physically mounting your sample substrate  onto the puck surface  For most measurements below room temperature Apeazon N grease is  adequate to secure your sample to the puck surface  However  more exotic mounting methods  may be necessary for larger samples     The wiring configuration used for each sample depends on the type of sample you are measuring   The ETO electronics has a 4 wire standard mode for samples whose resistance is below 10 MQ  and a 2 wire high impedance mode for samples whose resistance is 10 MQ up to 5 GQ  Both  wiring configurations are discussed below along with mounting and wiring examples  Since the  ETO has two independent sets of measurement electronics it is possible to wire up one sample in  standard mode and the other in high impedance mode     Note  How you wire up your sample determines what electronics you must use  You cannot  switch back and forth between standard and high impedance modes without physically changing  the sample wiring           WARNING     Do not 
10.  the operation of the Electrical Transport Option  ETO   option hardware   option software  and general use of the option     This manual does not provide detailed information about the Versalab system or the MultiVu  software application used to operate Versalab  This information can be found in the VersaLab  User   s Manual     P 3 Conventions in the Manual    File menu Bold text is used to distinguish the names of menus  options  buttons  and  panels appearing on the computer screen     File  gt  gt  Open The  gt  gt  symbol indicates that you select multiple  nested software options    lt Enter gt  Angle brackets distinguish the names of keys located on the PC keyboard     Note Text is set off in this manner to signal supplementary information about the  current task  the information may primarily apply in special circumstances     Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 vii  September 2011    Section P 4 Preface  Safety Guidelines and Regulatory Information Contents and Conventions    This symbol signals specific caution or conditions that could result in system  damage  bodily harm  or loss of life     This symbol signals electrical hazards that could result in bodily harm  or loss  of life  Used at all accessible 200 240 V power outlets        This symbol signals information on fusing     P 4 Safety Guidelines and Regulatory Information    Before using this product  please read the entire content of this manual and observe all  instructions  warnings a
11.  to configure the leads correctly in order to take full advantage of the  instrument   s sensitivity  The current leads are typically located at the edge of the sample  while the voltage leads lie between  Figure 1 1 a    When current is passed through the  sample an electric field is created  The voltage leads should be arranged such that they  measure a potential drop across a region where the electric field lines are relatively straight   This is achieved by placing the voltage leads in line with the current leads or by separating  the voltage leads by a distance that is small compared to their distance from the current leads   It is also important for a true four terminal measurement that the voltage and current leads do  not contact the sample at the same point  Otherwise  the reported resistance will include the  sample contact resistance     The lead geometry shown in Figure 1 1 a  works well for thin film sample with well defined  dimensions  Other lead geometries may be used depending on the nature of the sample  For  instance  when measuring a bar shaped sample you may attach the current leads to conductive  pads that contact the entire end of the bar and then make voltage leads that    1 2 PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011    Chapter 1 Section 1 3  Theory of Operation ETO Measure Types    contact the sample in a line parallel to those ends  You are responsible for determining the best  lead arrangement for your needs and for i
12.  wire     6  Use the Averaging Time text box to specify the measurement time in seconds     7  Use the Number of Measurements text box to specify the number of resistance points you  would like the measurement to include     8  Press the Measure button to start the measurement     AC Resistance Measurement    Channel 1 Channel 2  Enable Measurement Enable Measurement    Excitation Excitation  Amplitude   10 m   Freq  103   Hz Amplitude  10 m   Freq 200 Hz       Preamp Preamp    Range  1 3   x  C  Autorange Range  40m v     Autorange  Sample    Sample  y  Wiring 4 wire  x  Wiring   Measurement Configuration Measurement Configuration    Averaging   Number of   Averaging   Number of  Time 1 9  Measurements  io Time  1  see Measurements 10       Resistance     2nd Harmonic     dB Resistance       2nd Harmonic     dB    Phase Angle       31d Harmonic     dB Phase   ngle       31d Harmonic     dB    Temperature     Magnetic Field       Position     deg    Save Results to File    Figure 5 2  The immediate mode AC Resistance Measurement Window        During the measurement the Measure button will read Stop  Press this button at any time to end  the measurement  The measurement results will be displayed at the bottom of the window  The  Resistance displays the resistance measured in Q  The Phase Angle displays the phase angle in  degrees between the excitation signal and the measured signal  The 2  Harmonic and 3   Harmonic display the amplitude of each harmonic in dB  At the very bot
13. 084 700  Rev  A1 iii    September 2011    Contents  Table of Contents    CHAPTER 4  SOMA AAA AAA 4 1  NA NO 4 1  4 2 Overview of Electrical Transport Option Software     oooncccnncnnonnnoccnonononcnnn nono nono nono ncnn corn nconnccnnacns 4 1  ADA    IN A O NO 4 1  421 1  Setup Tab  nico lit iii died tido dales load 4 1  421 2     Data File Tab A A A ee ees 4 3  4 2 1 3  Measurement Tab  nosene Sa tities lirio llas inca 4 4  4 2 2 Electrical Transport Option Event Log    ee cesceseceseceseceseceseeeseeeeeeeaeecaaecnaecaeenaeenaeens 4 5  4 3 Electrical Transport Option Data Files           cee ecesccseesesseceseeseeecaeeaeeeeseeesecseeeeeesecaeeaeeeeeeaeeas 4 5  CHAPTER 5  IRA O tobeaceudseesssyes 5 1  A AS E EAT E EE AEA 5 1  5 2 Sample Montin S secreet isesi ti g eea i aa a a 5 1  5 2 1 4 Wire Mounting Configuration          esseeesesssesessesresresreerestesstsrrssetstssrentestessestesseeresseeeeestt 5 2  5 2 1 1 Hall Coefficient Mounting Configuration          eeeseeeseeeeseseesrssresrrsresresrresrsrrsreesesres 5 2  5 2 2  2 Wire Mounting ConfiguratlON    ooooccnoccnonononononoconaconaconncnnncnnn S rui aunean a risanni ia Ee 5 2  3 3     Sample  Installation aa 5 2  5 4 Immediate Mode MeasuremMentS    cooooccnncccnoncnononannnonononenonnnacnnnnonnnconnnnn nan nc nnnn nc nnnn rra nn annn nc cnn ccnnnss 5 3  5 4 1 Taking Resistance Measurements in Immediate Mode    oooocnnccnncccnoccnonnnoncnoncnnonononnnacnonns 5 3  3 42    Current  Voltage Curves iii A tia 5 5  5 4 3 Diff
14. PMS DynaCool CAN rack located on the right side of the  cryostat  The power switch for the CAN rack is located in back    Open the CAN rack door and verify that none of the modules installed have power  All of  the module LEDs should be dark    Install the two pieces of Velcro provided with the system on the back panel of the CAN rack  between the CAN rack and the cryostat shield  These should be in the center of the back  panel high enough to ensure the ETO sample cable can reach the grey lemo connector   Remove the cover for the second or third module bay from the back and carefully insert the  ETO module  Finger tighten the two mounting screws to secure the module in place  The  module should now be flush with the other modules or cover plates    Mount the ETO head on the Velcro between the CAN rack and the cryostat with the D shell  connector facing the back of the PPMS DynaCool    Connect the ETO Module Cable  3101 455 01  and Fan Cable  3101 457 01  between the  ETO module and the ETO head  Attach the ETO Sample Cable  3101 456 01  between the  ETO head and the grey lemo connector    Turn on the power to the CAN rack  Verify that both the PWR and COP indicator LEDs on  the module are green  It may take up to 30 seconds for the module to boot and the LEDs to  turn green  The COP LED will blink green until the option is activated     PPMS ETO User s Manual  1084 700  Rev  A1 2 3  September 2011    Section 2 2 Chapter 2  Initial Installation of Hardware and Software Instal
15. QuantumDesign       el  a   dl              Electrical Transport Option  ETO  User   s Manual    Part Number 1084 700  A1    Quantum Design  6325 Lusk Boulevard  San Diego  CA 92121    USA  Technical support  858  481 4400   800  289 6996  Fax  858  481 7410    Second edition of manual completed September 2011     Trademarks  All product and company names appearing in this manual are trademarks or registered trademarks of their respective holders     U S  Patents    Foreign Patents    c ONTENTS    Table of Contents    PREFACE  Contents and Conventions          sesssesoosseesoesooesossoesoossosssessossocssessossocesessoesoossosssesoossossoessoseossse vii  PE A O IN vil  B 2    Scope ofthe Mantal mesita dla tie o sii  vii  P 3    Conventionsan  the Manta A eee EOE  vil  P 4 Safety Guidelines and Regulatory Informati0N     cooconcnnnnnnncnnncnnnncnnnconnconnconncnnncnnnnnnacrnn cra conan viii  CHAPTER 1  Theory OF OperatioN vise asscisessscesessenssssenvavenavevssesseuncoosesusosecnesyeceyusvecesouteycesbexovesesstenscosevecsenshens 1 1  EN o AAA A A E NOA 1 1  1 2 Overview of the Electrical Transport OptiON     ooocnoccnoconococoncconnconnconn nono nona nono nonn conocio nnncnnncnnccnnnos 1 1  1 3 Electrical Transport Option Measurement Types   oooconoconoccnonncocnconnconncnnncnnncnnncnnn crac cnn conan rancia 1 2  La O CS A OD ne 1 2  13 11   Harmonics Detection n oerien aee aa eee a a aasa a ckecosanstdanea tees TEENE 1 3  EALS MAUL oerien it A ta ae ei Ags 1 3  1 3 2  Curre
16. Select     Sample1    Sample 2    AC Excitation    N O S    Frequency Y Hz ON ov  AC Amplitude 10 m              C  Disable Drive Feedback IV Sweeps a      Preamp  Gain 4yV v ae       Figure 5 3  The immediate mode IV measurement window     While the measurement is being performed the Measure button will display Stop  Press this at  any time to abort the current measurement  The status bar at the bottom of the window will  display the measurement progress  When the status bar is full the IV curve will be available in  the data file     Quantum Design    PPMS ETO User s Manual  1084 700  Rev  A1 5 5  September 2011    Section 5 4    Chapter 5    Immediate Mode Measurements ETO Measurements    5 4 3    5 6    Differential Resistance Curves    Differential resistance curves  like IV curves  can only be performed on one channel at a time   This measurement applies a small AC excitation and measures the differential resistance at  various DC bias points  Take the following steps to measure a differential resistance curve in  immediate mode     1     2     10     11   12     13     Select the radio button for Sample 1 or Sample 2 to select a channel    Select either 4 wire or 2 wire configuration from the Sample Wiring drop down list   Select the AC frequency from the Frequency drop down list    Use the DC Amplitude text box to specify the maximum value of the DC sweep    Use the Steps Quadrant text box to specify the number of DC steps per quadrant of the DC    sweep     The Step 
17. Size will be filled in automatically using the previously entered information  If  you prefer  you can enter the step size and the steps per quadrant will be recalculated     Select the DC sweep quadrants by dragging the start and stop cursors around on the sweep  graph     If the sweep does not start at zero bias  select the Ramp to Starting Point check box to  ramp the DC bias to the starting value prior to starting the measurement     Use the AC Amplitude text box to specify the amplitude of the AC excitation used to  measure the differential resistance  This value should be less than the DC sweep amplitude     Select the Range from the drop down menu     Use the Averaging Time text box to specify how long to average each data point     Use the Settling Time text box to specify how long in milliseconds to wait after each DC  step before beginning each new resistance measurement  100 ms is usually adequate     Press the Measure button to begin measuring the curve     PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011    Chapter 5 Section 5 5  ETO Measurements Sequence Measurements    dV dl Dialog    Selec DC Sweep     Sample 1    Sample 2  l   0   q    AC Excitation  Freq  976  A Hz        a Time  AC Amplitude  2 m      Preamp    Range     Step Size 5    Sample ac  g Steps Quadrant ho  Wiring      C  Ramp to Starting Point    Settings  Averaging Time pl   Settling Time  100  ms    an     Figure 5 4  The immediate mode differential resistance measuremen
18. The Electrical Transport Option supports three types of measurements   e Resistance  e IV Curves    e Differential Resistance    1 3 1 Resistance    The ETO is capable of both four terminal and two terminal resistance measurements  In the  classic four terminal geometry  4 wire mode   two leads pass a current through the sample  and two separate leads measure the potential drop across a section of the sample  Ideally the  voltage leads draw very little or no current  The current through the sample and the potential  drop across the sample can therefore be known to a high degree of accuracy  Ohm   s law can  then be used to calculate the resistance of the sample for the region between the two voltage  leads  This geometry eliminates the effects of lead contact resistance from the measurement  results     The resistance is calculated using Ohm   s law    V  R     1  7  1     where V is the measured potential drop across the sample and   is the current through the  sample  Unlike the AC Transport measurement option available from Quantum Design  ETO  does not calculate the resistivity  Since resistivity relies on sample specific parameters that  may not be know  or not known accurately at measurement time  the option reports the  resistance in units of Ohms  Resistivity can be calculated as follows     oes 2  oar  2     where R is the reported resistance    is the voltage lead separation  and A is the cross   sectional area through which the current is passed     It is important
19. The standard deviation of the  Ch1 Ch2 resistance  Phase Angel  Ch1 Ch2 deg  ee mA Wave form of the current  eres V Wave form of the voltage  ae Hz Excitation Frequency  uo g s Actual averaging time  AC Current wA The peak amplitude of the AC  Ch1 Ch2 drive excitation  eres mA The DC drive excitation  nd di V Magnitude of the voltage  aaa  V The in phase voltage amplitude  Quadrature Voltage V Douro E ase voljage  amplitude  aaa V The peak voltage of the AC drive  DC Voltage y The maximum DC voltage in the  Ch1 Ch2 channel  e Wench mA Magnitude of the current  rena me V The in phase current amplitude  Quadrature Current A The out of phase current  Ch1 Ch2 AE amplitude    1  Total gain   Module gain X head  can 90000 gain  29 Harmonic yv The in phase voltage amplitude of  the 2    harmonic  23  Harmonic yv The in phase voltage amplitude of  the 3    harmonic  An encoded integer containing  ETO Status Code N information on the current status of  the ETO hardware  ETO Measurement yv Integer indicating the measurement  Mode type for the row of data                         PPMS ETO User s Manual  1084 700  Rev  A1  September 2011    Quantum Design    CHAPTER 5    ETO Measurements    5 1    5 2    Introduction    This chapter contains the following information   e Section 5 2 discusses how to mount samples on the sample pucks   e Section 5 3 discusses how to make measurements in immediate mode and in sequence  mode     Sample Mounting    The Electrical Transport Option includes two
20. are necessary when  connecting ETO to one of the temperature options available on the PPMS        b  OA    Figure 3 4  Pictures of the ETO option installed onto  a  a PPMS system  b  a VersaLab system and  c  Dynacool  system  On the PPMS the head is mounted on the dewar using grip tape  VersaLab requires the use of the  accessory mounting kit  4372 079      3 4 Sample Pucks    ETO includes two sample pucks for mounting samples that interface with the VersaLab sample  chamber wiring  The pucks have a mounted circuit board with labeled contact pads for convenient  sample wiring on 3 channels  When using ETO  only channels 1 and 2 are used  Channel 3 may be  used with other measurement options or other third party electronics that you hook up to the base  system  The gold plated contact pads on the puck are useful for either wire bonding or soldering   For more information regarding sample wiring refer to Section 5 2     Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 3 5  September 2011    CHAPTER 4    Software    4 1    4 2    4 2 1    Introduction    This chapter contains the following information   e Section 4 2 presents an overview of the ETO software including the ETO console   e Section 4 2 2 discusses the ETO status log   e Section 4 3 reviews how to create data files and contains descriptions of each data  column contained within the file     Overview of Electrical Transport Option Software    ETO Console    Whenever the Electrical Transport Option is active t
21. dow displays base system information relevant to both channels  Figure 5 2  shows the resistance measurement window     To make a measurement enable one or both channels  set all of the measurement parameters  then  press the Measure button at the bottom  Resistance measurements will run until both channels  are finished  Make sure to check the Save Results to File checkbox if you want to save the data   Measurement results will be displayed in the bottom of the window  Take the following steps to  run an immediate mode measurement     1  Check the Enable Measurement checkbox corresponding to the channel s  you would like  to measure     2  Select either 4 wire or 2 wire from the Sample Wiring drop down menu corresponding to  the wiring configuration used when the sample was mounted     Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 5 3  September 2011    Section 5 4 Chapter 5  Impedance Mode Measurements ETO Measurements    3  Use the Amplitude text box to specify the peak AC amplitude of the sinusoidal excitation  applied during the measurement  Make sure the excitation is not large enough to damage  your sample  The Amplitude will have units of mA for 4 wire configurations and in V for  2 wire     4  Select the excitation frequency from the Freq drop down menu     5  Select the Range from the drop down menu or check the Autorange check box if you want  the range to be determined for you  The Range will have units of volts for 4 wire  configurations and amps for 2
22. driver adjusts the applied potential to maintain a constant output  current regardless of the sample and lead resistance  Figure 3 1  b  shows the load lines for the  current driver with feedback on  As the graph shows  the current is independent of the measured  voltage across the sample  The driver behaves as a current source for potentials up to the 30 V  compliance of the source  In this mode the current driver has high output impedance  This is the  default operating mode and is desirable for most measurements and samples     When feedback is off  the drivers can be represented as a voltage source behind a source resistance   Figure 3 3  a  shows the equivalent circuit when the feedback is turned off  In this case  the current  passing through the sample is dependent on the sample and lead resistance  Figure 3 3  c  shows  the load lines for the current driver as a function of the measured sample voltage  In this  configuration the sample current decreases with increasing sample voltage  If the sample or lead  impedance is comparable or larger than the source impedance the actual current will be  significantly lower than the requested current  ETO will correct for this discrepancy as best as  possible and the corrected current will be reported in the data file  The corrected current will be a  good approximation of the actual sample current in the limit that Rpeags  lt  lt  Roprive  For systems with  large lead resistance the current can be manually corrected as follows 
23. e AC current  response is measured with a current amplifier     The DSP is used to filter the AC response and pick out the portion of the response at the same  frequency and phase as the drive signal  All other components of the signal are filtered out   eliminating frequency dependent noise  DC offsets  and instrument drift  The DSP can also  calculate harmonic components of the measured signal     The current source has a minimum precision of 1 nA and a maximum current of 100 mA  It  is capable of supplying both DC and AC current with frequencies from 0 1 Hz to 200 Hz   The preamplifiers consist of a high gain amplifier  a programmable gain amplifier  and a high  impedance  current  amplifier  These three preamps  in conjunction with the current source   give the ETO a noise floor of 10 nQ and allow measurement of resistances up to 5 GQ     The Electrical Transport option is available on both PPMS and VersaLab systems  The ETO  software is integrated into the MultiVu software application used to control and monitor the  system hardware  You may use any MultiVu commands while working with the ETO  This  software integration allows you to utilize MultiVu sequences to fully automate system  operations and run a wide range of measurements without being present in the laboratory     Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 1 1    September 2011    Section 1 3 Chapter 1  ETO Measure Types Theory of Operation    1 3 Electrical Transport Option Measurement Types    
24. ediate mode the sequence command for IV curves can only perform  measurements on one channel at a time  To measure both channels requires two sequence    commands     IV Sequence    Select     Sample1    Sample 2    AC Excitation    Frequency 0 3   Hz       AC Amplitude 10 m        C  Disable Drive Feedback    Preamp    Gain  EN y   Sample    Wiring    Data Logging       Waveform Select    oy oY    ON Ow       IV Sweeps 1    v    Figure 5 8  The sequence command window for IV measurements     5 10 PPMS ETO User s Manual  1084 700  Rev  A1  September 2011    Quantum Design    Chapter 5    Section 5 5  ETO Measurements    Sequence Measurements    5 5 4 New Data File    Figure 5 9 shows the sequence command window to select or create a new data file within a  sequence  Insert this command whenever you need to switch files  such as between  measurements of different types  Press the Browse button to open the file dialog and select the  file  Selecting the Create New File Version radio button will create a new file with the name  specified or add a numeric suffix if the file already exists  Selecting the Append to File radio  button will add the data to an existing data file     NewDatafileDlg         2default dat             Browse    default  dat          File Action      Create New File Version       Append to File    Figure 5 9  The sequence command window for creating a New Data File        Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 5 11  September 2011    
25. en the file specified in the File Name field provided the file has already been  created     Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 4 3    September 2011    Section 4 2 Chapter 4  Overview of Electrical Transport Option Software Software    ETO Console    Setup   Data File   Measurement     Path    CAQdY ersal ab Electrical Transports          File Name      eto_default dat       Create new file version    Sample    Figure 4 4  Data File tab in the ETO Console window        4 2 1 3 MEASUREMENT TAB    The Measurement tab is used to open the measurement window for the selected measurement   The radio buttons on the right indicate the measurement type selected and the display window  gives an example of the type of data that can be collected  After you have selected the  measurement type press the Launch Measurement button to bring up the immediate mode  measurement window  While an immediate mode measurement window is active this tab is not  active  The active measurement window must be closed prior to launching a different  measurement window     4 4 PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011       Chapter 4    Section 4 3  Software    Electrical Transport Option Data Files    ETO Console    Setup   Data File   Measurement    AC Resistance Mode  1 0       Resistance  O dvdl  OW    El  ES    a  goo  3  E    e    s       0 100 110  Temperature  K     Launch  Measurement    Figure 4 5  Measurement tab in the ETO Console window      
26. enting the charge of the carriers     To measure the Hall potential arrange the voltage leads perpendicular to both the current and  field directions  Figure 1 2 shows a thin film rectangular sample wired up to detect the Hall  potential  In this example the magnetic field would be applied normal to the sample surface   Using Eq   1  you can rewrite Eq   4  in terms of the resistance  R  reported by ETO     RA  R      5    4B  Accurately measuring the Hall potential can be difficult  To measure the potential difference due  only to the Hall potential the voltage leads must be perfectly perpendicular to the bias field  If  this is not the case the measured potential will include a component due to the longitudinal  potential resulting from the sample resistance  The magnitude of the longitudinal potential is  typically much larger than the Hall potential making even small errors in lead placement  problematic  Both the Hall potential and the longitudinal potential are proportional to the bias  current so the AC filtering technique employed by the ETO does not eliminate this effect     V        I  V  l     Figure 1 2  Typical lead placement on a rectangular thin film sample for measurement of the  Hall potential  The magnetic field would be applied normal to the sample surface     PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011    Chapter 1    Section 1 3    Theory of Operation ETO Measure Types    1 3 2    133    Current  mA        Current Voltage Curv
27. er   Verify that the power is off  All of the module LEDs should be dark     Remove the cover plate for one of the rear  high power  module bays and carefully insert the  module into the bay  Gently finger tighten the two mounting screws to secure the module in  place  The module should now be flush with the other modules or cover plates     The ETO head needs to be mounted close to the Grey Lemo connector at the back of the  sample chamber  Determine the optimal mounting location for the ETO head and place the  two pieces of self adhesive grip tape on the PPMS dewar at this location  Make sure the  head is close enough that the sample cable can easily reach the Grey Lemo without stressing  the cable     Press the ETO head onto the grip tape to secure it in place  Make sure the head is not  touching any other dewar components     Connect the ETO Module Cable  3101 455 02  and Fan Cable  3101 457 02  between the  ETO module and the ETO head     PPMS ETO User s Manual  1084 700  Rev  A1  September 2011    Quantum Design    Chapter 2  Installation    7     Section 2 2  Initial Installation of Hardware and Software    Turn on the power to the Model 6000 tower  Verify that both the PWR and COP indicator  LEDs on the module are green  It may take up to 30 seconds for the module to boot and the  LEDs to turn green  The COP LED will blink green until the option is activiated     2 2 1 2 VERSALAB HARDWARE INSTALLATION    Refer to the VersaLab User   s Manual for any issues regarding t
28. erential Resistance Curves           cccccscccesscecsseceesseceeaeeceeeeecsaeceeaaecseaeeceaeeseaaeceeaeeneneeenaees 5 6  A Sequence Measurement A A 5 7  Sl BETO RESISTAOCE il A ee eo 5 8  A NI NN 5 9  NXS AA OT 5 10  o    New  Data AA A RT 5 11  iv PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design    September 2011    Contents  Table of Figures    Figures    Figure 1 1   Figure 1 2     Figure 1 3     Figure 2 1   Figure 3 1     Figure 3 2     Figure 3 3     Figure 3 4     Figure 4 1   Figure 4 2   Figure 4 3   Figure 4 4   Figure 4 5     Figure 5 1     Figure 5 2   Figure 5 3   Figure 5 4   Figure 5 5   Figure 5 6   Figure 5 7   Figure 5 8   Figure 5 9     Quantum Design    Typical lead placement on a rectangular thin film sample for  a  four wire and  b   two wire lead  geometries versidad rd didas 1 3  Typical lead placement on a rectangular thin film sample for measurement of the  Hall potential  The magnetic field would be applied normal to the sample surface  1 4  Two graphs showing current and voltage waveforms  a  and the resulting I V   curve  b  for a diode measured using the ETO     oooocccnncccinocacanocononcconnanonnncnnnncananenonnns 1 5    Activate the Electrical Transport Option using the Option Manager dialog box       2 4  Pictures of  a  the ETO module  CM H  and  b  the remotely mounted ETO head  Picture of the ETO module front panel showing the CH1 1 measured signal BNC     JH 1   CH1 drive signal BNC  JH 2   CH2 measured signal BNC  JH 3   CH2  dr
29. es    The ETO is capable of measuring the current voltage characteristics of any sample or device  In  this mode the electronics applies a triangular excitation waveform and measures a response  waveform  These two waveforms are combined to form the current voltage curve for the device  under test  The triangular excitation waveform always starts and ends at zero bias  Figure 1 3 a   shows both an excitation waveform and the response waveform for a diode plotted as a function  of time  The nonlinear device characteristics are clearly visible in the response waveform   Figure 1 3 b  shows resulting I V curve when the current waveform is plotted as a function of  the voltage waveform  The ETO software allows you to select the number of quadrants to  measure and in what order  The resulting curve will contain 256 points per quadrant  Both four   wire and two wire IV curves can be measured  In the four wire case a current excitation  waveform is used and a voltage response waveform measured  In the two wire case a voltage  excitation waveform is used and a current response waveform measured     10     A  aBeyon  Current  mA   o       ua     10  Time  Arb   3  2  1 0 1     b  Voltage  V     Figure 1 3  Two graphs showing current and voltage waveforms  a  and the resulting I V curve  b   for a diode measured using the ETO     Differential Resistance Curves    The ETO is capable of measuring differential resistance as a function of bias current or voltage   This is achieved by apply
30. he ETO Console window will be present  within the MultiVu window  This window can be minimized but will only close when the option  is deactivated  From the console window you can perform all basic system operations  such as  sample installation  setting up data files  and running measurements manually in immediate  mode  The ETO Console contains a Setup tab  Data File tab  and a Measurement tab  The  functionality of each tab is described below     4 2 1 1 SETUP TAB    The Setup tab is the default tab for the ETO Console and is displayed when the option is  initially activated  Figure 4 1 shows the Setup tab within the ETO Console  This tab is used to  install or remove samples and enter information about the samples that will be entered into the  data file     The Sample Installation Wizard button will launch a separate window for the wizard when  pressed  This wizard will guide you through the steps necessary to open the sample chamber to  either install or remove a sample  Figure 4 2 shows the Sample Installation Wizard window   The System Status panel displays the current system status and the Instructions panel displays  information and guidance regarding each step of the wizard  The wizard does not execute any    Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 4 1    September 2011    Section 4 2 Chapter 4  Overview of Electrical Transport Option Software Software    commands when started and requires that the user press the Open Chamber button to initiate the
31. he VersaLab system     1   2     Turn off the power to the VersaLab system     Open the hood covering the CAN rack and verify that none of the modules already installed  have power  All of the module LEDs should be dark     Install the Accessory Mounting Bracket to the VersaLab frame using the two  6 32 screws   washers  and nuts provided in the Accessory Mounting Kit     Remove the cover plate for the left most module bay as viewed when facing VersaLab   Carefully insert the module into the bay  Gently finger tighten the two mounting screws to  secure the module in place  The module should now be flush with the other modules or  cover plates     Mount the ETO Head to the accessory mounting bracket using the hook and loop fastening  tape on both the head and the bracket  Be sure to mount the head in the correct orientation     Note  When installed properly  the ETO head should not touch any of the mounting screws  and should not interfere with the hood when closed  Adjust the location of the head if  necessary     Connect the ETO Module Cable  3101 455 01  and Fan Cable  3101 457 01  between the  ETO module and the ETO head     Turn on the power to the VersaLab system  Verify that both the PWR and COP indicator  LEDs on the module are green  It may take up to 30 seconds for the module to boot and the  LEDs to turn green  The COP LED will blink green until the option is activiated     2 2 1 3 DYNACOOL HARDWARE INSTALLATION    1     Quantum Design    Turn off the power to the P
32. ing a small AC excitation on top of a DC offset bias  The AC response  is measured and used to calculate the differential resistance in the same manner as the standard  resistance measurement when no DC bias signal is applied  The differential resistance is a direct  measure of the first derivative of the IV curve at a given DC bias  This type of measurement is  very useful for examining small nonlinearities in a given device     Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 1 5    September 2011    CHAPTER 2    Installation    2 1    2 2    Introduction    This chapter contains the following information     e Section 2 2 lists the components of the Electrical Transport Option and describes the  procedure for installation on PPMs and VersaLab systems     Initial Installation of Hardware and Software    This section describes the procedures for the initial installation of the Quantum Design Electrical  Transport Option  ETO  onto the PPMS and VersaLab systems  These procedures apply only to  the first time you install and use ETO  Normal activation and use of the option is covered in  Chapter 4     Note  If your system came from Quantum Design with ETO already installed you do not need to  follow the procedures in this chapter  Refer to Chapter 4 for normal operation     Table 2 1 lists the components of the Quantum Design Electrical Transport Option  Please verify  that you have received all the components before you begin the installation process     Table 2 1  A 
33. ive signal BNC  JH 4   the head cable connector  JH 5   and the fan cable    AMAT A iden svadeseavadanceardadecensadabedanvaies 3 2  The equivalent circuit for ETO when drive feedback is disabled  The dashed line  represents the current source equivalent circuit  2    ee ee eeeeeereeereeeeeeeeensecnaeenaeens 3 3    Pictures of the ETO option installed onto  a  a PPMS system  b  a VersaLab system  and  c  Dynacool system  On the PPMS the head is mounted on the dewar using  grip tape  VersaLab requires the use of the accessory mounting kit  4372 079       3 5    Setup tab in the ETO Console Window     ooooonocccoconocononoconoconcconaconccnnn conc nnnanonacrna conan 4 2  The Sample Installation Wizard Window    oocoonconnconoccnocnnoonononcnonononoconoconoconccnnncnnnannos 4 2  The Sample Configuration Window    oooooccconccooncnoncnonnnonnnnnncnnncnn nono nono noconacnnccnnnannnannos 4 3  Data File Tab in the ETO Console Window   oooconcccnocnnocnnocncncncnnonnncnononncconccnnannnannnos 4 4  Measurement Tab in the ETO Console Window    ocococcccnocccnonanononononononanacnnnnccnnnccnnnos 4 5    Sample wiring examples   a  Shows mounted both a 2 wire sample on channel 1  and a 4 wire sample on channel 1   b  Shows mounted a 4 wire sample on    channel 1 and a 4 wire Hall sample on channel Zo   ooooonconnccnicccicacocononaconaconaconncanncnnos 5 2  The immediate mode AC Resistance Measurement Wind W    oococonocococcconcconncnnnannss 5 4  The immediate mode IV measurement Window 
34. lation    2 2 2 Install the ETO Software    Use the following instructions to install the VersaLab Electrical Transport Option software on  your PC   See the MultiVu User   s Manual for MultiVu installation instructions   The  installation wizard will install all of the option files in the C  QDPPMS ETOption    C  QDVersaLab ETOption   or C  QDDynacool ETOption directory for PPMS and VersaLab  respectively   1  Close all other applications prior to installing the ETO software   2  Double click the ETO Installer msi file to start the installation wizard  Follow the  instructions on the screen to install the software   Note  Be sure to select the correct base system type  either PPMS or VersaLab  when  prompted     3  Verify that the ETO software is installed correctly by activating it within the MultiVu  software     a  Open the MultiVu application software   b  Select Utilities gt  gt A ctivate Option     c  In the Option Manager dialog box select Electrical Transport from the Available  Options list and press the Activate button     d  Electrical Transport will move to the Active Options list inside the Option Manager     e  The Electrical Transport software will open the ETO Console window and the ETO  Event Log window  confirming that the software was install correctly     Option Manager    Available Options   Active Options            Electrical Transport      Activate    gt  gt     Connection  Diagrams          Close    Figure 2 1  Activate the Electrical Transport Optio
35. list of hardware components comprising the Electrical Transport Option for   a  PPMS   b  VersaLab  and  c  Dynacool systems                                       a  PPMS COMPONENT PART NUMBER   ETO Module CM H 4101 451   ETO Head EM QN 4101 455   ETO Module Cable 3101 455 02   ETO Sample Cable 3101 456 01  ETO Fan Cable 3101 457 02   ETO Accessory Kit 4084 710   Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 2 1    September 2011    Section 2 2    Initial Installation of Hardware and Software    2 2 1    2 2                                                     b  VERSALAB COMPONENT PART NUMBER  ETO Module CM H 4101 451  ETO Head EM QN 4101 455  Accessory Mounting Kit 4372 079  ETO Module Cable 3101 455 01  ETO Sample Cable 3101 456 01  ETO Fan Cable 3101 457 01  ETO Accessory Kit 4084 710   c  DYNACOOL COMPONENT PART NUMBER  ETO Module CM H 4101 451  ETO Head EM QN 4101 455  ETO Module Cable 3101 455 01  ETO Sample Cable 3101 456 01  ETO Fan Cable 3101 457 01  ETO Accessory Kit 4084 710    Chapter 2  Installation                Install the ETO Module  4101 451  and ETO Head  4101 455     Use this section only if the ETO electronics have not been installed on your PPMS or VersaLab  system     2 2 1 1 PPMS HARDWARE INSTALLATION    Refer to the PPMS User s Manual for any issues regarding the PPMS  These installation  instructions assume the Model 1000 tower and CAN bus have already been installed and tested     1   2   3     Turn off the power to the Model 1000 CAN tow
36. n using the Option Manager dialog box                    2 2 3 Configuring the Electronics    The ETO electronics are ready to use out of the box and do not require any calibration files on  the PC  All of the calibration and configuration information is stored inside the module SROM   Contact your customer service representative with any problems associated with the calibration  or SROM     2 4 PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011    C H APTER 3    Hardware    3 1 Introduction    This chapter contains the following information   e Section 3 2 discusses the operation of the ETO electronics   e Section 3 3 discusses the cables and jumpers    e Section 3 4 discusses the sample pucks    3 2 ETO Module and Head    The Electrical Transport Option electronics consists of the module  CM H  and the remotely  mounted head  EM QN   The module contains a Digital Signal Processor  DSP  that performs all  AC stimulus response calculations and a secondary processor for post processing and  communications  The remotely mounted head contains two sets of current drivers and  preamplifiers to allow for simultaneous  independent resistance measurements on channel 1 and  channel 2  Figure 3 1 shows the module and head enclosures         b     Figure 3 1  Pictures of  a  the ETO module  CM H  and  b  the remotely mounted ETO head  EM QN        To perform a basic transport measurement an AC drive signal is generated at the requested  frequency by the DSP  The cu
37. nd cautions  These are provided to help you understand how to safely and  properly use the Electrical Transport Option  ETO      Quantum Design Inc  disclaims any liability for damage to the system or injury resulting from  misuse or improper operation of the system  Please contact your Quantum Design representative  for any service issues     This product is NOT user serviceable by the user     viii PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011    CHAPTER 1    Theory of Operation    1 1    1 2    Introduction    This chapter contains the following information   e Section 1 2 gives an overview of the Electrical Transport Option     e Section 1 3 discusses the different measurements that are possible with ETO     Overview of the Electrical Transport Option    The Quantum Design Electrical Transport Option  ETO  enables users to make several  different types of transport measurements over a wide range of resistance values and sample  types  The option contains two separate channels each with their own dedicated electronics  to allow simultaneous  continuous resistance measurement of two different samples  Each  channel   s electronics consists of a precision current source and voltage preamplifiers coupled  to a Digital Signal Processor  DSP   Measurements are usually made by applying a  sinusoidal AC drive current and measuring the AC voltage response  However  a special 2   wire high resistance mode is available where an AC voltage is applied and th
38. nt Voltage Curves coincida aia 1 5  t33  Differential R  sistance CUYOS nad RA 1 5  CHAPTER 2  Iastalla on ala ace dla il iaa 2 1  21  ES A NN 2 1  2 2 Initial Installation of Hardware  amp  SO  tWAT8     ccooooocononoooccnonononononnnnconnnnnononnnnnononnnnnn nn nnconnnnnnnnnnnss 2 1  2 2 1 Install the ETO Module  4101 451   amp  ETO Head  4101 455   0oooooocnnnccccocncooonnnonanoconanonns 2 2  2 2 1 1 PPMS Hardware Installation  nen neike apiaster inie ia arean ttiie 2 2  2 2 1 2 VersaLab Hardware Installati0N            ooooonnnococcnonocnncnononnnonononononononononononnnnonnnnncnnnos 2 3  2 2 1 3 Dynacool Hardware Installati0N       ooonnonnnnnnnnnnnonoccnonncnoncnnnccnnocnoconccnnccnnccnnccnnncnnnos 2 3  2 2 2  smstallithe E TO S oft wares  cits site ia ii WA de es ES 2 4  22 3     Configuring the  Electronics  usisne essenin aee a ea thes nEaN RESTS 2 4  CHAPTER 3  PRAT WAL E E icbaectaicedectdaacastiaaeeneuidinsauchatesSeeusisoasscomaipecktinesscoanessuccauiniten 3 1  A A A ER 3 1  3 2     ETO Module  Head  naniesione a a iets aisle ats eesti 3 1  3 2 0    Current Driver  BIECtrOmiCs   csds30 cs  sesceccaceesesa ibe ck a a stead ehcsecceseaae netos casdanedeebesetdeosaace 3 2  S222  A clesees eote E Liki pees ier bettas hens ele neta RO RE hae rites 3 4  323      Cables  Se JUMpers   ttre  lox e ne ee A ecto ee Unga E E A lot nese owes 3 5  3 4     Sample  Pucks s  3s c shiteccieed os hie beeen eee RI ea See ede Blo ete eee 3 5  Quantum Design PPMS ETO User s Manual  1
39. nterpreting the resulting data     The Electrical Transport option is also capable of measuring large resistances using a two   terminal lead geometry  2 wire mode   This mode is typically used for resistances above 10 MQ   In 2 wire mode the system applies a voltage across the sample and uses the high impedance  amplifier to measure the current passing through the sample  The resistance is again calculated  using Ohms law  Figure 1 1 b  shows a sample wired up in the two terminal configuration  To  achieve the best results the sample should be rectangular or bar shaped with the leads far apart  yielding a well defined measurement region where the electric field line density is uniform  Lead  and contact resistance can usually be ignored in this case as long as they are much less than the  sample resistance     It V  V  l  I  V    a   b     Figure 1 1  Typical lead placement on a rectangular thin film sample for  a  four wire and   b  two wire lead geometries     1 3 1 1 HARMONIC DETECTION    The second and third harmonics of the measured response are reported in units of dB during a  resistance measurement  The units of dB are referenced to the voltage of the fandamental  response  This information can be related to nonlinearities in the sample but is often an indicator  of the amount of noise present during a measurement  The second and third harmonic  contribution is typically less than    50 dB for a good measurement     1 3 1 2 HALL COEFFICIENT    While it is not explici
40. rrent driver board in the head receives the drive signal and converts it  to a current that is applied to the sample  The preamplifier board in the head detects and amplifies  the sample response signal and sends it back to the DSP for processing  The DSP calculates the in   phase and out of phase components of the response signal as well as several harmonic amplitudes     Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 3 1  September 2011    Section 3 2 Chapter 3  ETO Module and Head Hardware    The module  CM H  contains four BNC outputs for use in monitoring the drive  JH 2 and JH 4   and response signals  JH 1 and JH 3  for both channel 1 and 2  The BNC labels are located on the  module front panel as shown in Figure 3 2  The drive signal full scale is 10 V  For example  on  the 100 mA range  a 50 mA peak excitation current would generate a 5 V peak signal at the BNC   The response signal is monitored behind the preamplifiers  The response signal measured at the  BNC needs to be divided by the total gain to obtain the voltage at the sample        Figure 3 2  Picture of the ETO module front panel showing the CH1 1 measured signal BNC  JH 1   CH1 drive  signal BNC  JH 2   CH2 measured signal BNC  JH 3   CH2 drive signal BNC  JH 4   the head cable connector   JH 5   and the fan cable connector  JH 6      3 2 1 Current Driver Electronics    The current drivers in ETO employ active feedback to approximate an ideal current source  When  the feedback is on  the current 
41. t window        While the measurement is being performed the Measure button will display Stop  Press this at  any time to abort the current differential resistance measurement  The status bar at the bottom of  the window will display the measurement progress  When the status bar is full the dV dI curve  will be available in the data file     5 5 Sequence Measurements    A sequence can be used to run a series of measurements sequentially or to change other system  properties while measuring  such as temperature or magnetic field  Any number of measurement  commands or system commands can be run from within a sequence  Figure 5 5 shows the list of  sequence commands available in MultiVu when ETO is active  All ETO specific commands are  listed under Measurement Commands  gt  gt  Electrical Transport     Note  For detailed information about creating and editing sequence files and for a discussion  about all standard sequence commands  refer to the PPMS or VersaLab User   s Manual     Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 5 7  September 2011    Section 5 5 Chapter 5  Sequence Measurements ETO Measurements    Sequence Commands        oe System Commands  Beep  Call Sequence or Script  Chamber Operations  Remark  Scan Field  Scan Temperature  Scan Time  Sequence Message  Set Field  Set Temperature  Shutdown  Wait  E Measurement Commands  E Electrical Transport  ETO dvdl  ETO Iv  ETO Resistance  New Datafile  Log Data  Sigma Log Data     lt     gt        Figure 5 5
42. ted to the I  pad and the V  pad as shown in Figure 5 1 a   channel 1  It is important that you use the I  and V  pad or the measured resistance will have the  wrong sign or value  Hall measurements are not possible in the 2 wire configuration        Figure 5 1  Sample wiring examples   a  Shows mounted both a 2 wire sample on channel 1  and a 4 wire sample on channel 1   b  Shows mounted a 4 wire sample on  channel 1 and a 4 wire Hall sample on channel 2     53 Sample Installation    The sample installation wizard is accessible from the Setup tab on the ETO Console  You are  encouraged to use this wizard as 1t automates the installation process and leaves the sample  chamber in the proper state for temperature cycling  Pressing the Sample Installation Wizard    5 2 PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011    Chapter 5 Section 5 4  ETO Measurements Immediate Mode Measurements    button will launch the wizard window  shown in Figure 4 2  Follow the instructions on the  screen to install your sample  When prompted by the wizard install the sample puck using the  puck extraction tool   For more information on sample installation refer to either the PPMS or  Versalab manual   Pressing cancel at any time during the wizard will abort the wizard safely     Enter sample information after installing the sample puck into the sample chamber  Press the  Sample Properties button to bring up the dialog window  Enter the wiring configuration for  sample 1 and
43. tly supported it is possible  with proper lead placement  to perform   4 wire Hall measurements using the Electrical Transport Option  When charged particles move  perpendicular to a magnetic field a force is exerted on them perpendicular to both the field and  the direction of particle motion  The force can be expressed as    F qvxB  3     This transverse force can often cause charge carriers to build up on one edge of a sample leading  to a potential difference across the sample  This potential difference is called the Hall potential   Vy  The sign of the Hall potential generally indicates the sign of the charge carriers and the    Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 1 3  September 2011    Section 1 3    Chapter 1    ETO Measure Types Theory of Operation    1 4    magnitude of the potential is related to the density of charge carriers in the sample  The Hall  coefficient  Ry  describes these two properties and is defined as     _ Ey V  A        B UB        4     where Ey is the Hall field  j is the current density  and B is the magnitude of the magnetic field   For a well defined geometry the current density is equal to JA where Z is the current and A is the  sample cross sectional area  and the Hall field is equal to V      Here    is the distance between  a set of transverse voltage leads used to measure the Hall potential  It can be shown that the  Ry  nq    with n representing the number of charge carriers per unit volume in the sample and q  repres
44. tom of the window are  displayed the Temperature  Magnetic Field  and Position of the samples inside the sample  chamber     5 4 PPMS ETO User s Manual  1084 700  Rev  A1 Quantum Design  September 2011    Chapter 5    ETO Measurements    Section 5 4  Immediate Mode Measurements    5 4 2 Current Voltage Curves    Current voltage curves can be taken on only one channel at a time  Take the following steps to  measure a current voltage curve in immediate mode     1     2     Select the radio button for Sample 1 or Sample 2 to select a channel   Select either 4 wire or 2 wire configuration from the Sample Wiring drop down list   Select the sweep frequency from the Frequency drop down list     Use the AC Amplitude text box to specify the peak amplitude of the sweep  The AC  Amplitude has the units mA with 4 wire configuration and V in 2 wire     Make sure the Disable Drive Feedback box is un checked  This feedback should only be  disabled if you are measuring a sample with a highly non linear I V characteristic  See  section XXXX for more details regarding the drive feedback     Select the type of excitation waveform desired  You have the option of performing a four   quadrant sweep or a two quadrant sweep in the positive or negative direction     Use the IV Sweeps text box to specify the number of sweeps to average together to create  the IV curve     Select the Range from the drop down menu   Press the Measure button to begin the measurement     IV Measurement    Select Waveform 
45. urements  0 u    Figure 5 6  The sequence command window for resistance measurements     5 5 2 ETO dV dI    Figure 5 7 shows the sequence command window for differential resistance measurements   Follow the same steps outlined in Section 5 4 3 to specify the measurement parameters  When  you are finished  press the OK button and the measurement command will appear in the sequence  window  As with the immediate mode  the sequence command for differential resistance can  only perform measurements on one channel at a time  To measure both channels requires two    sequence commands     Quantum Design    PPMS ETO User s Manual  1084 700  Rev  A1 5 9    September 2011    Section 5 5  Sequence Measurements    dV dl Sequence Dialog    Select       Sample 1 O Sample 2    AC Excitation    Frequency 15  w Hz    AC Amplitude  2 m       Preamp  Range    Sample  Wiring       Settings    Averaging Time  1   sec  Data Logging       Chapter 5  ETO Measurements    DC Sweep   l  0     l  Time    DC Amplitude  50       Step Size  5          Steps Quadrant  10       Ramp to Starting Point    Settling Time 100 ms    Figure 5 7  The sequence command window for differential resistance measurements     5 5 3 ETO IV    Figure 5 8 shows the sequence command window for the IV curve measurements  Follow the  same steps outlined in Section 5 4 2 to specify the measurement parameters  When you are  finished press the OK button and the measurement command will appear in the sequence  window  As with the imm
46. urements Available Measurements                      Figure 4 3  The Sample Configuration window     4 2 1 2 DATA FILE TAB    The Data File tab is used to specify the data file into which the ETO data is written  Pressing the  Browse button will bring up a standard Windows file interface that allows you to specify a file  name and location on the PC for the data file  You can use this window to either select an  existing file or create a new file name  If a new data file name is selected the data file is not  created immediately  Instead the file is created when the first data point is written to the file    The currently selected target file and its location are displayed in the Path and File Name fields   The Create new file version checkbox  when checked  will add a numeric suffix to the file name  displayed if the file already exists  If this box is unchecked the option will append to an existing  data file     The Sample Properties button is also contained in the Data File tab in the event the Setup tab is  not used  The functionality for this button is identical to that on the Setup tab and you can enter  information from either tab  The information contained in the Sample Properties window is  written to the data file when it is first created  New sample information can be entered at any  time before a new measurement is taken and the new file is created     The View button opens the selected data file and displays the data as a new graph in MultiVu  It  will always op
47. wire up both channels to the same sample as this may lead to errors  in the reported resistance due to stray capacitance                    Quantum Design PPMS ETO User s Manual  1084 700  Rev  A1 5 1    September 2011       Section 5 3 Chapter 5  Sample Installation ETO Measurements    5 2 1 4 Wire Mounting Configuration    To make a standard resistance measurement in 4 wire configuration you should mount your  sample as shown in Figure 5 1 a  for the sampled wired to channel 2  The two current leads  should be at the ends of your sample and the two voltage leads should be in line between the  current leads  Ensure that the polarity of the voltage leads matches that of the current leads or the  reported resistance will be negative     5 2 1 1 HALL COEFFICIENT MOUNTING CONFIGURATION    ETO can also be used to make Hall measurements  Figure 5 1 b  shows channel 1 wired for a  standard resistance measurement and channel 2 wired for a Hall measurement  The wiring  arrangement shown yields the proper sign for the Hall coefficient  It is important when wiring a  Hall sample that the voltage contacts are at the same zero field potential to avoid measuring  longitudinal signal  This is usually achieved by making sure the voltage contacts are  perpendicular to the current contacts  ETO does not support 5 wire Hall measurements     5 2 2 2 Wire Mounting Configuration    For samples whose resistance exceeds 10 MQ the 2 wire configuration is necessary  In this case  the sample is connec
    