Home
User Manual - BC Group International Inc.
Contents
1. connection to DUT for 1 Amp Ground Resistance Test PUSH TO TEST light touch key to initiate a test Chassis Plug for connection to Safety Analyzer Chassis 25 Port Current Sour CS 2000 U Layout Green LED indicator for visual feedback if a test is valid On Valid Test PUSH TO TEST light touch key to initiate a test Off No Functional Test Flashing Resistance is greater than 3 0 ohms i Chassis Plug for Male Cord for A connection to Safety connection to Safety Analyzer Chassis Analyzer Receptacle 65 2099 N Port Current Source Female Receptacle for connection to DUT Kelvin Cable for connection to DUT for 1 Amp Ground Resistance Test CS 2000 E Layout OPERATION The CS 2000 1 Amp current source is not a stand alone device It is designed to work in conjunction with a BC Biomedical SA 2000 Series Safety Analyzer or a DALE 544D 544D Lite 601 and 601E Safety Analyzer Push To Test Key Pressing and holding this key initiates a test This applies the 1 Amp current source to the Device Under Test DUT LED Indicator This LED gives the operator visual feedback on the test status The following table explains the meaning of each possible LED status indication LED Status Meaning OFF Test is not functional ON Test is Valid BLINKING Resistance is greater than 3 ohms Connections 1 The Line plug
2. on the CS 2000 plugs into the Safety Analyzer DUT Receptacle 2 The Chassis cable located at the top right corner of the CS 2000 plugs into the Chassis Connection port on the Safety Analyzer 3 The DUT plugs into the CS 2000 Receptacle cable 4 The detachable cable located at the bottom right corner of the CS 2000 is used to connect to the chassis of the DUT Below is a typical example for wiring configuration of the Safety Analyzer BC International SA 2005 shown the CS 2000 CS 2000 U model shown Current Source and the DUT in this case a Patient Monitor TESTING WARNING All connections to patients must be removed before connecting the DUT to the CS 2000 or Analyzer A serious hazard may occur if the patient is connected when testing with the Analyzer Do not connect any leads from the patient directly to the CS 2000 or Analyzer from the DUT 1 Connect the CS 2000 to the Safety Analyzer and the DUT as indicated in the Connections portion of the Overview Section 2 Set the Safety Analyzer for Forward Polarity with Hot and Neutral Closed 3 Set the Safety Analyzer to read earth resistance The safety analyzer display will read approximately 0 67 ohms NOTE When the CS 2000 is connected to the Safety Analyzer and the Push to Test button is not held bias voltages cause a reading to be displayed on the Safety Analyzer This reading is approximately 0 67 ohms This is not a true reading of the DUT Valid tests
3. 601 requirements where a 1 A source current is required This device interfaces directly with BC Biomedical SA 2000 series of Safety Analyzers as well as the Dale 544D amp 544D Lite and Dale 601 amp 601E Safety Analyzers CS 2000 SERIES e FACILITATES TESTING IEC601 REQUIREMENTS INTERFACES DIRECTLY TO BC BIOMEDICAL SA 2000 SERIES OF SAFETY ANALYZERS e INTERFACES TO DALE 5440 544D LITE 601 AND 601E SAFETY ANALYZERS 90 264 VAC OPERATION 50 60 Hz 1 AMP CURRENT SOURCE FOR RESISTANCE MEASUREMENTS NO PROTRUDING PUSH BUTTONS LIGHT TOUCH KEY USED TO OPERATE SMALL SIZE AVAILABLE MODELS e CS 2000 U STANDARD MODEL WITH HOSPITAL GRADE NEMA 5 15P LINE PLUG AND HOSPITAL GRADE NEMA 5 15R DUT TEST RECEPTACLE FOR USE IN THE USA AND OTHER COMPATIBLE COUNTRIES e CS 2000 E EUROPEAN MODEL WITH SCHUKO LINE PLUG AND SCHUKO DUT RECEPTACLE FOR USE IN APPLICABLE EUROPEAN COUNTRIES ACCESSORIES BC20 20221 SCHUKO GROUNDING ADAPTER e BC20 20114 REPLACEMENT KELVIN LEAD FOR CS 2000 e BC20 30106 SOFT SIDED CARRYING CASE LAYOUT This section looks at the layout of the CS 2000 and gives descriptions of the elements that are present Green LED indicator for visual feedback if a test is valid On Valid Test Off No Functional Test Flashing Resistance is greater than 3 0 ohms Male Cord for connection to Safety Analyzer Receptacle Female Receptacle for connection to DUT Kelvin Cable for
4. Biomedic Nm A UNIT OF BC GROUP INTERNATIONAL INC 1 AMP CURRENT SOURCE CS 2000 U CS 2000 E USER MANUAL BC BIOMEDICAL CS 2000 SERIES TABLE OF CONTENTS WARNINGS CAUTIONS NOTIGES aan een ii DESGRIP TON 1 LAYOUT ET 2 OPERATION E e 4 TESTING see aa AA AA 6 MANUAL 7 LIMITED WARRANTY nano 7 SPECIFICATIONS tc 8 NOTES atiende uoi b ento a 9 This User Manual covers the following units e CS 2000 U e CS 2000 E WARNING USERS The CS 2000 is for use by skilled technical personnel only WARNING USE The CS 2000 is intended for testing only and should never be used in diagnostics treatment or any other capacity where they would come in contact with a patient WARNING CONNECTIONS All connections to patients must be removed before connecting the DUT to the CS 2000 A serious hazard may occur if the patient is connected when testing with the unit Do not connect any leads from the patient directly to the unit or DUT while it is powered by the Analyzer WARNING CLEANING Disconnect Line Power to the CS 2000 before attempting to clean it Do not immerse The unit should be cleaned by wiping gently with a damp lint free cloth A mild detergent can be used if desired WARNING LIQUIDS Do not submerge or spill liquids on the CS 2000 Do not operate the CS 2000 if internal components not intended for use with fluids may have been exposed to fluid as t
5. are indicated only when the CS 2000 Status LED is continuously lit readings on the Safety Analyzer in all other situations should be ignored 4 On CS 2000 press and hold the Push to Test key 5 Verify the validity of the test by checking the LED on the CS 2000 NOTE A continuously lit LED indicates a valid test 6 Read the resistance on the Safety Analyzer MANUAL REVISIONS Revision Revisions Made Rev 01 Origination Rev 02 Updated Pictures Rev 03 Specifications Updated Misc Edits Rev 04 Format Updated Pictures Updated Misc Edits Rev 05 Misc Edits LIMITED WARRANTY WARRANTY BC GROUP INTERNATIONAL INC WARRANTS ITS NEW PRODUCTS TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE INTENDED THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF SHIPMENT EXCLUSIONS THIS WARRANTY 1 IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE BC GROUP INTERNATIONAL INC IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR ASSUME ANY LIABILITY REMEDIES THE PURCHASER S SOLE AND EXCLUSIVE REMEDY SHALL BE 1 THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS WITHOUT CHARGE 2 AT THE OPTION OF BC GROUP INTERNATIONAL INC THE REFUND OF THE PURCHASE PRICE P MANUALS BCGr
6. he internal leakage may have caused corrosion and be a potential hazard WARNING MODIFICATIONS The CS 2000 is intended for use within the published specifications Any application beyond these specifications or any unauthorized user modifications may result in hazards or improper operation CAUTION USAGE The CS 2000 is not a continuous duty device it is intended for short duration testing within the current limits specified Do not leave the DUT connected to this unit for extended time periods CAUTION ENVIRONMENT Exposure to environmental conditions outside the specifications can adversely affect the performance of the CS 2000 If the unit is outside of Operating specifications allow it to acclimate to specified conditions for at least 30 minutes before attempting to operate it CAUTION INSPECTION The CS 2000 should be inspected before each use for obvious signs of abuse or wear The unit should not be used and should be serviced if any parts are in question CAUTION SERVICE The CS 2000 is intended to be serviced only by authorized service personnel Troubleshooting and service procedures should only be performed by qualified technical personnel NOTICE SYMBOLS Symbol Description Caution Consult Manual for Further Information Electrical Caution Consult Manual for further information NOTICE ABBREVIATIONS AorAmp Association for the Advancement of Medical Instrumentati
7. on Celsius Degree s Device Under Test European Hertz International Electrotechnical Commission kilogram s millimeter s ohm s pounds Relative Humidity Root Mean Square second s United States of America Volt s Volt Ampere s Volt s Alternating Current Watt s NOTICE DISCLAIMER BC GROUP INTERNATIONAL INC WILL NOT BE RESPONSIBLE FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS NOTICE DISCLAIMER BC GROUP INTERNATIONAL INC RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME WITHOUT NOTICE IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE HOWEVER NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES NOTICE CONTACT INFORMATION BC BIOMEDICAL BC GROUP INTERNATIONAL INC 3081 ELM POINT INDUSTRIAL DRIVE ST CHARLES MO 63301 USA 1 800 242 8428 1 314 638 3800 www bcgroupintl com sales bcgroupintl com CS 2000 Series User Manual Copyright 2012 www bcgroupintl com Made in the USA 06 12 Rev 05 This Page Intentionally Left Blank vi BC BIOMEDICAL CS 2000 SERIES 1 AMP CURRENT SOURCE The CS 2000 Series is a 1 Amp Current Source for testing the ground continuity in equipment This device facilitates testing to IEC
8. oup CS2000_UM_Rev05 doc SPECIFICATIONS Electrical CURRENT SOURCE 1 00 to 1 02 ADC OPERATING LINE VOLTAGE 90 to 264 VAC 50 60 Hz POWER CONSUMPTION 12 VA 12 W This device is an accessory to a Safety LOAD CURRENT Analyzer refer to applicable Safety Analyzer User Manual for this Rating CS 2000 U NEMA 5 15P LINE PLUG CS 2000 E European CEE 7 7 Schuko DUT CS 2000 U Hospital Grade NEMA 5 15R RECEPTACLE CS 2000 E European CEE 7 7 Schuko 3 Pin Locking Plug for Safety Analyzer Chassis Cable Connection See Description section for CONNECTIONS Compatible Safety Analyzers Detachable Kelvin Cable BC20 201 14 for DUT Ground Connection Physical ENCLOSURE ABS Plastic CONSTRUCTION OVERLAY Back printed Lexan SIZE 7 38 3 94 x 1 64 Inches 187 5 x 100 1 x 41 7 mm WEIGHT lt 1 Lbs 0 45 kg 0 to 50 32 to 122 F OPERATING RANGE 10 to 90 RH Non Condensing STORAGE RANGE 40 to 60 C 40 to 140 F NOTES lt Group 5 YOUR ONE STOP BIOMED SHOP BC GROUP INTERNATIONAL INC 3081 ELM POINT INDUSTRIAL DRIVE ST CHARLES MO 63301 USA 1 800 242 8428 1 314 638 3800 www bcgroupintl com sales bcgroupintl com CS 2000 Series User Manual 06 12 Rev 05 Copyright 2012 Made in the USA