1.       OFITE    Dependable Products From People You Trust       Production Screen Tester    Part No  810 00 1    Instruction Manual    Updated 3 10 2014  Ver  1 0    OFI Testing Equipment  Inc   11302 Steeplecrest Dr    Houston  Texas   77065   U S A   Tele  832 320 7300   Fax  713 880 9886   www ofite com      Copyright OFITE 2015    Table of M Osea ee ccsecctacnsncaaveaosnark E a 2    DESCHID ON cin EA 2  Con ten ts SHeECINCAUONS varoen a 2  GCOMPONENTS ozin a aa 3  Sale orp a a a aa bes 4  eE ST a de uacatvoces aavadeeaesenddeestevinacs eersmnteeas 5  Operat O Maiin a a a 6  DiG AIM sheices de csndtendda rcs icutvencndsaicuczonsddasiwestaudtaulodsiessaestiniwbobeenstsvens  11  GEIFASSOIMNDIY  reaa E R 11  Advanced Screen Holder ASS CMDI          ccccccccseccseeccseeeseecneeeeaes 12  PAD DG IAIN onn a 13  Advanced Screen Holder ASSCMDI          ccccccccseccsesccnseeseeseseeaes 13  Cap Screen ASSCMDIV gecesaisceiseriiadic ti culesd detasehaenlevicedigaziesebuclds 14  Warranty and Return Policy ivccicccssecievccsctssiessideccescsesvetiennsasvesdves 15    OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA   Tel  832 320 7300   Fax  713 880 9886   www ofite com    Intro    Description    Specifications    The flow back characteristics of drill in fluids are critical to the productivity  of a well  If the fluid plugs the production screen  it will not only slow down  production  but it could also lead to screen erosion and costly remediation     The PST is designed to test flow
2.   2  Assemble the test cell    3  Pour the fluid into the test cell    4  Place the top cap on the cell body    5  Place the cell into the frame and secure it with the T screw   6  Apply pressure and start the timer    7  Record temperature  time  and fluid volume    8  Release the pressure     9  Disassemble the cell and clean all components     OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA   Tel  832 320 7300   Fax  713 880 9886   www ofite com    Opera tion 1  Before beginning a test  make sure each part of the cell is clean and dry   particularly the screen  Examine the gaskets for distortion and wear  Make  sure the screen is free of sharp edges  burrs  or tears     2  Measure the initial temperature of the fluid sample and record it for later  analysis     3  Assemble the test cell  to assemble the cap screens see 13 and 14    a  Assemble the Bottom Cap     i  Place a gasket in the inside groove in the bottom cap        Bottom Cap with Gasket    li  Apply lubrication to the inner bottom cap   810 00 007  and place  an o ring   170 13  in the designated seat               Bottom Cup Insert   Bottom Cup Insert with O ring  iii  Place the bottom cup insert into the bottom cap  Screw the fitting    130 78 045  into the bottom of the bottom cup insert     Ya  NPT Fitting       130 78 045   Bottom Cup  Insert     810 00 007     Bottom Cap     810 00 002     Bottom Cup  Insert     810 00 007     OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA  Tel  832 320 7300   F
3.   8 10 00 010A   8 10 00 010B   8 10 00 010C   8 10 00 010D   8 10 00 020 1    LP SS Tubing  2  x 1   4   Neoprene Gasket  Qty  2   Cell Pin  Qty  2   Threaded Insert W Set Screw  T screw   Felt Filter   CO  Pressuring Assembly With Top Cap  Ya   Polyurethane  Flex  Tubing  O ring Nitrile   Bottom Cap Assembly   Cell Body   O ring Nitrile   Cap Screen   Bottom Cup Insert   Plastic Main Insert   Thick Spacer Ring   Thin Spacer Ring  Qty  2    Advanced Screen Holder Assembly  Advanced Screen Holder   Thick Spacer   Thin Spacer  Qty  2   Locking Ring   Screen Tester Main Insert  SS 303     Tools    120 50 076  6  Combination wrench  Qty  2   120 58 07  a  Allen wrench for Threaded Insert for T Screw   170 27  2  Allen wrench for Base Cap Set Screws    OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA   Tel  832 320 7300   Fax  713 880 9886   www ofite com    Safety Carbon Dioxide gas is normally supplied in small bulbs or cartridges  which  contain approximately 900 PSI  6 206 kPa  pressure when new  Because  they are highly portable  they are usually used in field operations  These  bulbs should not be exposed to high heat  50  C 120  F  as they can explode if  over heated     Never transport CO  bulbs or cartridges by airplane without proper packaging   Cabin depressurization could cause them to explode     OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA   Tel  832 320 7300   Fax  713 880 9886   www ofite com 4    Quick Start 1  Measure the initial fluid temperature   
4.   OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA  Tel  832 320 7300   Fax  713 880 9886   www ofite com    5  Place a rubber gasket inside the top cap  Make sure it is seated all the  way around the cap  Then place the top cap onto the top of the cell body     6  Place the entire cell into the frame and secure the cell with the T screw     PRODUCTION SCREEN  TESTER    FI Testing Equipment  inc     ORME o  o 810 00 1  Oi  oe  lt        e i    4E   i y N 7    y       eor aa  as      ap with Gasket         Top C   7  Place a clean  dry 1 000 mL Beaker under the filtrate tube    8  With the regulator closed and the relief valve pushed in  place a CO   bulb into the sleeve and screw it onto the pressure assembly  This will    puncture the bulb and release the pressure     9  Turn the regulator T screw clockwise to adjust the pressure  Start the  timer     10  At the end of the test  back off the regulator and record the following     a  Ifthe fluid plugs the screen  record the time and the volume of fluid  collected     b  Ifthe fluid does not plug the screen  record the time required for all  the fluid to flow through the screen     11  Pull the relief valve out to release any remaining pressure     12  Remove the cell from the frame and disassemble it  Discard any  remaining fluid     13  After each test  disassemble the test cell and thoroughly clean all    surfaces with soap and water  Make sure all parts are clean and dry  before storing the unit     OFITE  11302 Steepl
5.  back of completion fluids on the rig site  It   is no longer necessary to ship fluid samples back to the lab and delay the  completion operation for days or weeks  Field fluids can be tested in real time  with samples of the actual production screen being used down hole     The PST now makes it possible to determine if the fluid remaining in the  annulus will flow back through the production screen     The Production Screen Tester is similar to the standard API filter press both   in appearance and operation  The user assembles a test cell with a sample of  the production screen at the bottom  After filling the cell with fluid  he applies  pressure and opens a valve at the bottom  The test will very quickly show  whether the screen and fluid are compatible  or if plugging is a possibility  And  the apparatus is simple enough that new screen samples can be installed   and tested quickly and easily     Pressure Source  CO    Cell Size  3   7 8  ID x 10 9   27 6 cm  Long   Sample Volume  1 000 mL   Screen Size  1 9   50 mm  Diameter and 2 5   63 5 mm  Diameter  Size  11 5  x 8  x 27   29 x 20 x 69 cm    Weight  24 Ib  10 9 kg     OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA  Tel  832 320 7300   Fax  713 880 9886   www ofite com 2     1 20 60 30   141 05   141 06   141 09   141 10   141 22   142 00   153 78   170 13    8 10 00 002   8 10 00 003   8 10 00 005   8 10 00 006   8 10 00 007   8 10 00 020   8 10 00 021   8 10 00 022    Components    Optional      8 10 00 010  
6. ax  713 880 9886   www ofite com 6       mbly Installation    Ber    Shut Off Valve Asse    iv  Use two     wrenches to install the Shut Off Valve Assembly    120 910 070  to the NPT connector on the inner cap as shown  above    b  Assemble the Main Insert  Plastic    810 00 020      i  Insert an o ring   170 13  into the internal portion of the insert and  lubricate the o ring as shown below  left      li  Install the external o ring   810 00 005  and lubricate the external  portion of the insert as shown below  right      Internal O ring    External O ring       OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA   Tel  832 320 7300   Fax  713 880 9886   www ofite com T    iii  Place the Insert into the bottom side of the cell and press down to  sit flush inside the cell        c  Assemble the cap screen   i  Insert spacers and screens as needed   li  Place an o ring   810 00 004  into the plastic insert     iii  Insert the cap screen   810 00 006  as shown below        i  Spacer Insert       iii  Cap Screen Insert Cap Screen Installed    OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA  Tel  832 320 7300   Fax  713 880 9886   www ofite com 8    d  Place the assembled bottom cap onto the cell and turn it to lock it into  place  Use a  2  allen wrench to tighten the set screws which secure  the cell cap to the cell     e  Turn the cell assembly right side up     4  Pour the freshly stirred sample fluid into the cell  leaving 0 5   13 mm  of  empty space at the top      
7. ecrest Dr   Houston  TX 77065 USA  Tel  832 320 7300   Fax  713 880 9886   www ofite com 10    Diagram CO  Pressuring  Assembly    Cell Assembly With Top Cap    142 00     Cell Body    810 00 003     Cell Pin  Qty  2     141 06     O ring  Nitrile    810 00 004     Cap Screen    810 00 006   O ring  Nitrile    810 00 005   Main Insert  Plastic     810 00 020        O ring  Nitrile    170 13       Thin Spacer  Pe phan   810 00 022     O ring  Nitrile    170 13     SS  O ring  Nitrile  Bottom Cup Insert SA   170 13       810 00 007   CY  Gasket  Neoprene  ea   141 05     Bottom Cap Assembly    810 00 002           w Fitting  Male Connector   g 14    130 78 045    LP SS Tubing  2  x 14     120 60 30     Shut Off Valve 1   4     120 910 070        i 4  Polyurethane Tubing     4153 78     OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA   Tel  832 320 7300   Fax  713 880 9886   www ofite com 11    Diagram  Advanced Screen    Holder Assembly   Optional     Locking Ring     810 00 010D     O ring    810 00 004   Thick Spacer    810 00 010B     Thin Spacer    810 00 010C     Screen Holder   Advanced    810 00 010A        OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA  Tel  832 320 7300   Fax  713 880 9886   www ofite com    12    Appendix The Advanced Screen Holder   810 00 010  is an optional screen assembly  which includes spacers to accommodate multiple screens of varying    Advanced Screen thickness up to    and 50 mm in diameter  Not all screens will require the  Holde
8. o guarantees to the buyer that any products will be provided     OFITE makes no other warranties or guarantees to the buyer  either express or implied   and the warranties provided in this clause shall be exclusive of any other warranties  including ANY IMPLIED OR STATUTORY WARRANTIES OF FITNESS FOR PURPOSE   MERCHANTABILITY  AND OTHER STATUTORY REMEDIES WHICH ARE WAIVED     This limited warranty does not cover any losses or damages that occur as a result of      Improper installation or maintenance of the products  e Misuse     Neglect  e Adjustment by non authorized sources  e Improper environment       Excessive or inadequate heating or air conditioning or electrical  power failures  surges  or other irregularities    e Equipment  products  or material not manufactured by OFITE     Firmware or hardware that have been modified or altered by a third party    e Consumable parts  bearings  accessories  etc      Returns and Repairs     Items being returned must be carefully packaged to prevent damage in  shipment and insured against possible damage or loss  OFITE will not be  responsible for equipment damaged due to insufficient packaging     Any non defective items returned to OFITE within ninety  90  days of invoice are subject to a  15  restocking fee  Items returned must be received by OFITE in original condition for it to  be accepted  Reagents and special order items will not be accepted for return or refund     OFITE employs experienced personnel to service and repair eq
9. ofite com 14    Warranty and  Return Policy    Warranty     OFI Testing Equipment  Inc   OFITE  warrants that the products shall be free from liens and defects in  title  and shall conform in all respects to the terms of the sales order and the specifications applicable  to the products  All products shall be furnished subject to OFITE   s standard manufacturing variations  and practices  Unless the warranty period is otherwise extended in writing  the following warranty shall  apply  if  at any time prior to twelve  12  months from the date of invoice  the products  or any part  thereof  do not conform to these warranties or to the specifications applicable thereto  and OFITE is  so notified in writing upon discovery  OFITE shall promptly repair or replace the defective products   Notwithstanding the foregoing  OFITE   s warranty obligations shall not extend to any use by the buyer  of the products in conditions more severe than OFITE   s recommendations  nor to any defects which  were visually observable by the buyer but which are not promptly brought to OFITE   s attention     In the event that the buyer has purchased installation and commissioning services on  applicable products  the above warranty shall extend for an additional period of twelve   12  months from the date of the original warranty expiration for such products     In the event that OFITE is requested to provide customized research and development for the buyer   OFITE shall use its best efforts but makes n
10. r Assembly use of every spacer  The thickness of screens can be different which may   Optional  eliminate the use of other spacers     Screen Holder   Advanced    810 00 010A     Thick Spacer    810 00 010B     Thin Spacer    810 00 010C   Locking Ring     810 00 010D        Advanced Screen Holder Assembly   810 00 010        1  Place spacers as needed into holder assembly    2  Place screen into screen holder and more spacers if needed   3  Place an o ring   810 00 004  into screen holder    4  Secure screen assembly with Locking Ring   810 00 010D     5  Place assembled screen holder into insert     OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA   Tel  832 320 7300   Fax  713 880 9886   www ofite com 13    Appendix    Cap Screen Assembly    The Cap Screen Assembly is an optional screen assembly which includes  spacers to accommodate multiple screens of varying thickness up to    and  2 5  in diameter  Not all screens will require the use of every spacer  The  thickness of screens can be different which may eliminate the use of other  spacers     Cap Screen    810 00 006        Thin Spacer  Qty  2     810 00 022     Thick Spacer    810 00 021     Cap Screen Assembly    Place spacers as needed into holder assembly   Place screen into screen holder and more spacers if needed   Place an o ring   810 00 004  into screen holder     Place assembled screen holder into insert     OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA  Tel  832 320 7300   Fax  713 880 9886   www 
11. uipment manufactured by us  as   well as other companies  To help expedite the repair process  please include a repair form with all  equipment sent to OFITE for repair  Be sure to include your name  company name  phone number   email address  detailed description of work to be done  purchase order number  and a shipping address  for returning the equipment  All repairs performed as    repair as needed    are subject to the ninety  90   day limited warranty  All    Certified Repairs    are subject to the twelve  12  month limited warranty     Returns and potential warranty repairs require a Return Material Authorization  RMA   number  An RMA form is available from your sales or service representative     Please ship all equipment  with the RMA number for returns or warranty repairs  to the following address     OFI Testing Equipment  Inc   Attn  Repair Department  11302 Steeplecrest Dr   Houston  TX 77065   USA    OFITE also offers competitive service contracts for repairing and or maintaining your lab  equipment  including equipment from other manufacturers  For more information about  our technical support and repair services  please contact techservice ofite com     OFITE  11302 Steeplecrest Dr   Houston  TX 77065 USA   Tel  832 320 7300   Fax  713 880 9886   www ofite com 15    
    