1. nstruction Manua Flat Bed Electrophoresis Chamber MULTI for lsoElectric Focussing Applications using IEF Gels or IPG Strips and for Flat Bed Gel Electrophoresis in general CP51 1 Warning Like all apparatus run by electricity these units are capable of delivering potentially lethal voltage when connected to a power supply They should be operated only by qualified technically trained personnel The horizontal electrophoresis units from ROTH are designed for long term laboratory use and to obtain reproducible results Please spend a few moments reading the instruction manual thoroughly DO NOT attempt to remove the outer casing or make repairs to our electrical range of products should any unit fail These units comply with the statutory CE safety rules 73 23 EEC Low voltage directive IEC 1010 1 1990 plus amendment 1 1992 EN 61010 1 1993 BS EN 61010 1 1993 Please verify that you received the unit completely and without any damage Any faults or losses have to be reported to ROTH immediately ROTH can not accept responsibility for goods that were sent back without informing them Please take a look at the packing list and check whether all components and accessories are present Please retain all packaging material until the warranty period has expired SPECIFICATIONS Technical specifications Unit dimensions WxDxH inc
2. Triton X100 non ionic detergent Art No 3051 2 to the surface of the cooling plate This will act as a heat transfer agent between the cooling plate and the gel during the electrophoresis run Place one end of the gel on its support onto the Cooling plate in the position required making contact with the applied Triton X100 on the lower edge Slowly lower the gel so that the Triton 5 spreads under the gel support plate expelling any air and ensuring a good contact to the cooling plate surface If on lowering air is trapped raise the gel plate so that it is released then lower the gel again Remove any excess Triton from around the gel with a tissue Cut filter paper wicks long enough to reach from the lower buffer chambers to the edge of the gels Dip the anode wick in the anode buffer allow to wet Form a bridge between the lower anode buffer and the anode edge of the gel Apply the filter paper to the gel surface so that it overlaps by 5 10 mm and makes good contact 9 Dip the cathode wick in the cathode buffer allow to wet Form a bridge between the lower cathode buffer and the cathode edge of the gel Apply the filter paper to the gel surface so that it overlaps by 5 10 mm and makes good contact 10 Apply the samples to the gel 11 Replace the safety lid 12 Connect to a suitable power supply 13 The unit is now ready for Electrophoresis OND l At the End of the Run Turn the power supply off Remove the safety
3. be run on the electrophoresis unit for many different applications The manufacturer s instructions should be followed and the unit used in the configuration most appropriate for the method recommended When using filter paper strips or buffer strips which are applied to the surface of the gel the connection should be using the suspended electrodes electrode plate see section F When using liquid buffer the connections should be made between the buffer in the lower chamber and the gel by use of filter paper wicks or equivalent see section H H Electrophoresis Using the Lower Buffer Chambers In addition to the use of the unit with the electrodes applied directly to the gel surface by means of the suspended electrodes electrode plate see section F the tank can be used with buffer in the lower chambers In this format the electrode plate and attached electrodes are not required The following instructions are general instructions for Electrophoresis where the gels are hand poured or precast gels in a horizontal format on a gel support Please consult a suitable reference text for gel compositions and exact running parameters Recommendations are given above 1 Remove the cooling plate and fill the lower chambers of the base unit with the required electrode solutions Do not overfill the chambers Ensure that the buffer level is below the level where the cooling plate locates Replace the cooling plate in the base unit Apply 1 5 ml of
4. flowing water from a tap or water bath flow cooler Static regulation is unlikely to sufficient for Isoelectric focusing and for this application a flow cooler is recommended e g Roth Art No X618 1 Static Temperature Regulation 1 Attach a short length of rubber hose to each connector 2 Incline the unit at an angle of approximately 45 degrees with the ports uppermost 3 Use a funnel to fill the Cooling Plate with de ionised water containing 0 02 w v sodium azide preservative to prevent bacterial and algal growth 4 When filled keep the unit inclined and attach clamps to each piece of rubber hose 5 The unit can be cooled before an electrophoresis run if required DO NOT FREEZE Active Temperature Regulation 1 2 Attach a short length of rubber hose to each connector Attach one end of the rubber hose to the outlet port of a circulating water bath flow cooler and the other end to the inlet port Alternatively attach one end of a rubber hose to a water supply and allow the other rubber hose drain to waste The maximum recommended water flow rate is 1 Litre min DO NOT exceed this figure If you are using a circulating water bath flow cooler which exceeds this flow rate you can attach a T connector in line One branch of the connector can return water to the bath and the other can flow to the Cooling Plate and incorporate a flow regulator such as an adjustable tubing clamp Measure and adjust the flow rate before attachi
5. lid and electrode holder if applicable Carefully remove the electrode wicks if used Proceed with processing the gel After Electro focusing remove the electrodes from the electrode holder and rinse carefully with distilled water to prevent corrosion by the strong acidic and basic solutions DO NOT submerge the socket connector 6 After Electrophoresis rinse the buffer chambers with distilled water again taking care not to submerge the socket connector Dry the chambers by using a vacuum line taking care not to damage the platinum electrodes 7 Rinse the apparatus with distilled water after Ensure that the connectors are clean and dry before usage or storage Oie OS E J Troubleshooting Guide Equipment Problem Comments No Current Reading Check the safety plug insertion Check the pin and socket connections Check the lid is placed on firmly When using the suspended electrodes on the electrode holder ensure that they are making good contact with the electrode strips on the gel If using the lower buffer chamber electrodes do not overfill chambers but ensure sufficient buffer is used so that good electrical contact is made between the buffer and gel Ensure that the filter wicks are completely wetted in contact with the buffer and attached to the gel surface Check the connection to mains fuses and power cables Current increases with time The Cathode and Anode polarities may be reversed Check
6. the connections the gel orientation and the pH of the applied electrode strips Condensation over the The Gel may be being run at too high a voltage Reduce the power settings entire surface of the Electrode holder The cooling may be insufficient for the application Check that the cooler is switched on Carl Roth GmbH Co KG SchoemperlenstraBe 3 5 76185 Karlsruhe Postfach 100121 76231 Karlsruhe Telefon 49 721 5606 0 Telefax 49 721 5606 149 E Mail info carlroth de 6 Internet www carlroth de s s 11 2014
7. Glass electrode frame optional Cabels use Electrodes lsoElectric Focusing using the suspended platinum wire electrodes The composition and running of IEF gels or IPG strips varies considerably depending on the range of pH required the application native or denaturing and the format of the gels The following instructions are general instructions for IsoElectric focusing applications where the gels are hand poured or precast in a horizontal format on a gel support Please consult a suitable reference text for gel compositions and exact running parameters Recommendations are given above 10 11 12 13 Apply 1 5 ml of Triton X100 non ionic detergent Art No 3051 2 to the surface of the cooling plate This will act as a heat transfer agent between the cooling plate and the gel during the electrophoresis run Place one end of the gel or IPS strip on its support onto the Cooling plate in the position required making contact with the applied Triton X100 on the lower edge Slowly lower the gel so that the Triton spreads under the gel support plate expelling any air and ensuring a good contact to the cooling plate surface If on lowering air is trapped raise the gel plate so that it is released then lower the gel IPS strip again Align the gel with the relevant electrode etchings if applicable Remove all excess Triton from the cooling plate Prepare two electrode strips one cathode one anode by cutting strips of filt
8. bles rated safe up to 10 000 volts e Gold plated electrical connectors corrosion free and rated safe up to 10 000 volts e Recessed power connectors integral with the safety lid e 0 2 mm diameter platinum electrodes 99 99 pure e Glass electrode weight for optimal contact situation e Silicone rubber dovetail seal provides leak free sealing and are easy to clean and or replace e User friendly clamping system e Wide range of accessories Environmental Conditions e This apparatus is intended for indoor use only e This apparatus can be operated safely at an altitude of 2 000 m e The normal operating temperature range is between 4 C and 65 C e Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C e The apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664 POLLUTION DEGREE 2 states that Normally only non conductive pollution occurs Occasionally however a temporary conductivity caused by condensation must be expected Packing List 1 Main Running Tank and lid Art Nr CP51 1 Cooling Plate Art Nr CP54 1 Electrode holder acrylic 1 Glass Electrode Frame Art Nr CP55 1 Cathode Electrode Art Nr CP53 1 Anode Electrode Art Nr CP52 2 Silicone tubing for cooling plate connectors 1 Glass electrode weight glass plate USING THE FLAT BED ELECTROPHORESIS CHAMBER A Safety Precautions READ the instructions before using th
9. e apparatus Always isolate electrophoresis units from their power supply before removing the safety cover Isolate the power supply from the mains FIRST then disconnect the leads DO NOT exceed the maximum operating voltage 5 000 V or current 150 mA DO NOT operate the electrophoresis units in metal trays 2 Acrylamide is a volatile cumulative neurotoxin and suspected carcinogen Wear effective protective clothing and follow recommended handling and disposal procedures Polymerised gels contain some non polymerised monomer Handle with gloves only Following the replacement of a platinum electrode have the unit inspected and approved by your safety officer prior to use DO NOT fill the unit with running buffer above the maximum fill lines 275 ml per lower chamber 1 200 ml total when using chamber reservoir electrodes DO NOT move the unit when it is running CAUTION During electrophoresis very low quantities of various gases are produced at the electrodes The type of gas produced depends on the composition of the buffer employed To disperse these gases make sure that the apparatus is run in a well ventilated area B General Care and Maintenance To remove the safety lid push thumbs down on the plastic lugs and lift the lid vertically with your fingers Clean and dry the apparatus with DISTILLED WATER ONLY IMPORTANT Acrylic plastic is NOT resistant to aromatic or halogenated hydrocarbons ketones esters alcohol
10. er paper 5 mm wide and slightly shorter than the gel width Moisten the Electrode strips with the relevant electrode solutions Drain if required Place the electrode strips 2 mm from the cathode and anode edges of the gel IPS strip The etched positions will help in positioning Ensure the electrode strips are slightly shorter then the gel on which they are applied This will prevent electrical contact along the edges of the gel Apply the samples and proceed with steps 9 to 13 if pre focusing is not required In case prefocusing is required then proceed with steps 9 to 13 before loading samples and repeating these steps Pre focusing is generally performed at lower voltage than the following separation step Using the markings on the electrode holder to align the electrodes in the correct position on the electrode holder This is achieved by loosening the screws on the electrodes sides opposite of the cables sliding them into the correct channel on the plate holder and retightening the location screws Place the electrode holder into the unit and lower carefully so that the platinum wire of the electrodes makes contact with the filter paper wicks on the gel Connect the electrodes to their respective sockets in the base unit Replace the safety lid and connect to a suitable power supply The unit is now ready to be switched on for the run G Electrophoresis Electrofocusing Using Precast Gels 1 Horizontal precast gels IPS strips can
11. l cooling coil 46 x 41 x 11 5 cm Inner tank dimensions WxDxH 37 x 31 x 8 5 cm Dimensions of electrode frame cooling plate and glas plate WxL 27 x 27 cm Maximum gel size 250 x 250 mm or 30 3mm IPG strips Recommended temperature for gel IEF 20 C Recommended temperature for IEF using IPG strips 4 C Recommended temperature for flat bed gel electrophoresis 4 C Power output connectors 2 mm shrouded Snap lock connectors for cooling coil Inner 10 mm Outer 12 mm Quick fit tubing Inner 10 mm Outer 12 mm Recommended running conditions for IEF Running conditions for Step 1 2 3 4 5 6 end IEF gels with length Voltage V 300 600 1 000 1 200 1 500 2 000 10 15 cm Time mins 20 20 30 30 30 60 Volt hour 100 200 500 600 750 2 000 Running conditions for Step 1 2 3 4 5 6 end IEF with IPG strips Voltage V 150 300 600 1 500 3 000 330 of 5 10 cm length Time mins 0 5 0 5 0 5 0 5 2 5 lt 20 Volt hour 75 150 300 750 7 500 Running conditions for Step 1 2 3 4 5 end IEF with IPG strips Voltage V 300 600 1 500 3 000 330 of 15 20 cm length Time mins 1 1 1 12 5 lt 20 Volt hour 300 600 1 500 37 500 Technical features e Rugged acrylic construction e Ceramic cooling plate connectable to chiller or water tap e Glass electrode frame for optimal cooling efficiency e All acrylic joints chemically bonded e Doubly insulated ca
12. ng the line to the gel unit E Installation and Preparation _ F Place the unit on the bench where it is to be used Place the prepared cooling plate see above in the base unit Ensure it locates properly with the pipes protruding from the cut outs in the side wall The unit should now be checked to make sure it is perfectly level with a spirit level Mount the electrofocusing electrodes on the electrode holder by removing the clamping screws opposite of the cables and placing the electrode assembly under the electrode holder Ensure that the electrodes are colour co ordinated with the appropriate coloured edge of the plate holder Replace the clamping screw Place the electrode plate onto the unit The electrode holder should locate snugly onto the cooling plate with the connectors in the correct position to plug into the correct anode cathode socket of the base unit Set the desired temperature of the cooler circulator if to be used and switch on at least 10 minutes prior to electrophoresis to allow the cooling plate to reach the desired temperature The normal range for IEF operations is 4 20 C The unit is now ready for electrofocusing or electrophoresis For instruction on these techniques see sections F G and H overleaf Elektrophoresis tank base unit Glass plate for weighing of electrodes Tubings for connection of V4 cooling plate P P Cooling plate d 9 RAD Electrode holder acrylic and
13. s over 25 and acids over 25 they will cause crazing especially of the UV transparent plastic and should NOT be used for cleaning DO NOT use abrasive creams or scourers The cooling plate upper surface cannot be cleaned with acetone Methanol and ethanol can be used Dry components with clean tissues prior to use e g ROTH tissues ref 0087 1 Before use and then on a monthly basis check the unit for any leaks at the bonded joints Place the unit on a sheet of dry tissue and then fill with DISTILLED WATER ONLY to the maximum fill line Any leakage will be seen on the tissue paper If any leakage is seen DO NOT ATTEMPT TO REPAIR OR USE THE APPARATUS but notify Carl Roth GmbH Co KG immediately Ensure that the connectors are clean and dry before usage or storage C Storage of Water Cooled Units Water cooled units can be stored with water in the cooling plate but 0 02 sodium azide should be added to prevent algal growth Store in a dark cupboard or cold room Alternatively drain the unit A small quantity of water will remain in the base core If algal growth does build up over a period of time fill the base core with neutral Decon overnight and then flush through with clean water D Using the Cooling Plate The Cooling Plate will already contain a small quantity of water from control tests The Cooling Plate can be used in two ways Static water can be used as a simple heat sink or the tank can be actively regulated using

Download Pdf Manuals

image

Related Search

Related Contents

user manual Bedienungsanleitung manuel d`utilisation manuale d  Ergotron StyleView Sit-Stand Combo Arm with Worksurface  Nextar T30 User's Manual  Belt Heat Flux Sensor USER MANUAL    Kingston Technology ValueRAM 4GB 400MHz DDR2 ECC Registered CL3 DIMM (Kit of 2) Dual Rank, x4  ASUS AI Suite II  CHANGER8 取扱説明書  