         Contents
1.     If the meter cannot be calibrated  or if its readings are erratic  first make sure that the  probe head contains enough fresh electrolyte  To fill the head with electrolyte  follow the  instructions in the    Calibrating the oxygen probe    section  If that doesn   t solve the  problem  replace the probe head diaphragm set  In either event  the next step should be  to recalibrate the probe     When the icon cx  appears in the left corner of the display  it   s time to replace the six     AA    batteries that power it  although measurements will remain valid for several hours  after the low battery indicator first appears   Replacing the batteries requires removing  the screws that hold the battery compartment cover in place  After inserting fresh  batteries in the correct orientation  tighten the screws to secure the cover     If the meter    freezes     like a computer  and buttons become unresponsive  try resetting  the instrument by pushing the RESET button on its right side  callout 1 15 of Fig  1  with  the end of a paper clip     Remove the battery when storing this product for an extended period of time   Do not drop or disassemble the instrument or immerse it in water     14    REPLACEMENT PARTS  amp  OPTIONAL ACCESSORIES  e Spare oxygen probe  e Spare probe head and diaphragm set  e Electrolyte for probe  e 9 VDC adapter for 110VAC power supply    WARRANTY INFORMATION    General Tools  amp  Instruments     General   s  DOM5512SD Dissolved Oxygen Meter with  Datal
2.   80 White Street   New York  NY 10013  212 431 6100    Remember to include a copy of your proof of purchase  your return address  and your  phone number and or e mail address     15    Test Equipment Depot   800 517 8431   99 Washington Street Melrose  MA 02176    TestEquipmentDepot com    
3.  To power off the DOM5512SD  press the POWER button and hold it until the  instrument responds with a long beep      When the meter powers on  a series of transient startup screens will briefly appear  Once  the display has stabilized  perform the following ten setup steps in the order presented     3  Format the SD card  Press the LOGGER SET button  callout 1 8  and hold it for at least  five seconds  until    Sd F    appears in the lower half of the display  Quickly  within three  seconds  press the REC ENTER button      If you press buttons too slowly in setup mode  the screen will revert to the normal  display  To return to the setup sequence  press and hold the LOGGER SET button again   To move ahead to the next field in the sequence or to the next parameter  press the  HOLD ESC button      Pressing the REC ENTER button causes the word    no    or    yES    to appear over the term     Sd F     Press the W button to make a selection  Choose    yES    whenever a new SD card  is being used  or when a used card is being repurposed  from use with another SD card  instrument or a camera  for example  and all data on it is to be erased  Choose    no    to  preserve any data on a card previously used with this instrument     If you choose    yES     after you press the REC ENTER button the instrument will prompt  you to confirm that decision by displaying the term    Ent    below    yES    and sounding three  beeps  To confirm that you want to begin the erasure formatting proced
4.  button  which has the words    Sampling check    stenciled below it  once   To change the sampling time  perform Step 5 of the setup procedure     1  To start automatic datalogging  press the REC ENTER button once  The top line of the  display will then show the term    REC     Pressing the LOGGER SET button at this point  will make REC flash and add the flashing term    LOGGER    at the top right of the display   This indicates that the instrument is currently storing measured values and their time  stamps in memory     2  To pause automatic datalogging  press the LOGGER SET button once  this action  makes the flashing term LOGGER disappear from the top right of the display and  changes the term    REC    from flashing to constant  Pressing the LOGGER SET button  again resumes automatic datalogging     3  To end automatic datalogging  press the REC ENTER button and hold it for at least  two seconds  This action causes the    REC    message to disappear     4  To log data manually  set the sampling time to zero using Step 5 of the setup  procedure  Then press the REC ENTER button once  The display will show the term     REC    on the top line  a value in the middle  and below it the letter    P    on the left and a  number from 1 to 99 on the same line to the right  The number indicates the position  on the SD card that will be used to store manually logged data     Now press the LOGGER SET button  This will cause the beeper to sound and the term     LOGGER    to briefly app
5.  the W or A button to decrease or increase the value and  then press the REC ENTER button to store that setting  Finally  press the Y or A  button to decrease or increase the flashing seconds value and then press the  REC ENTER button to store that setting     6  Enable or disable auto power off  Once the sampling time has been set and stored   the display will show the word    yES    over the term    PoFF     Press the W button until  the desired automatic power off management scheme     yES    for enable     no    for  disable  is displayed  Press the REC ENTER button to store the selection  If enabled   the power off function shuts off the DOM5512SD after a period of inactivity of ten  minutes     7  Enable or disable the beeper  Once the auto power off function has been enabled or  disabled  the display will show the word    yES    over the word    bEEP     Press the W  button until the desired setting     yES    or    no     is displayed  and then press the  REC ENTER button to store the selection     8  Choose a decimal point or comma to represent the decimal division between  integers and fractions  for example  American style 20 88 vs  European style 20 88    Once the beeper has been enabled or disabled  the display will show the word     bASIC    above the phrase    dEC     Press the W button to make    bASIC     American  style  or    Euro    appear in the upper display  as desired  Press the REC ENTER button  to store the selection     9  Select the temperature uni
6. DO  level is an excellent indicator of water  quality  For example  if the DO level of a body of water remains too high or too low for too  long  fish and other aquatic organisms will die or become more susceptible to disease   For quality control purposes  many industries other than fish farming and water pollution  control routinely measure and keep track of the DO levels of their process streams in real  time  Among them are food processing  photography  pulp and paper manufacturing   power generation  electroplating and facilities management     General Tools  amp  Instruments    DOM5512SD has all of the features needed by an  instrument suitable for precision measurement of DO levels in a wide range of  applications  Its 7 5 in  long  polarographic  electrolyte filled oxygen probe measures not  only the DO concentration of a liquid  but the liquid   s water temperature as well  The  meter uses a single custom microcontroller to display readings from 0 to 20  milligrams liter  mg L  with a resolution of 0 1 mg L and an accuracy of   0 4 mg L     Because it is microprocessor based  the DOM5512SD can make full use of the portability   reliability and large storage capacities that SD memory cards offer  Measurements can be  made automatically at any sampling rate between one second and nine hours  After the  instrument time stamps and stores the measurements on an SD card plugged into it  a  process called datalogging   the user can remove the card and plug it into a laptop 
7. Test Equipment 99 Washington Street  Depot Melrose  MA 02176  Phone 781 665 1400  1 800 517 8431 Toll Free 1 800 517 8431          DOM5512SD  USER   S MANUAL    hgw shack  DISSOLVED OLYGEN METER      GENEOAL MMO       DISSOLVED OXYGEN METER  WITH DATALOGGING SD CARD    Please read this manual carefully and thoroughly before using this product     TABLE OF CONTENTS    Introduction 32  soss0creeor ees snstereererers 3  Key Features ecco cenhchabalanasneesnsasicnue  amp   Operating Instructions            00ceeeeeee 4 13  What   s in the case             00eeees 4 5  Setup 2  deciedodedede ded eteesd lt e 6 8  Calibrating the Oxygen Probe             8 9  Normal Operation               0000 9 10  Holding and Storing Measurements          10  Automatic vs  Manual Datalogging           11  Transferring Data from the SD Card  to a Computer          0c cece eee 12 13  Specifications           0 0eceeeeeeeeaee 13 14  Maintenance  amp  Troubleshooting Tips             14  Replacement Parts  amp  Optional Accessories        15  Warranty Information              0000ceeeaee 15    Return for Repair Policy              0000e00es 15    INTRODUCTION    Thank you for purchasing General Tools  amp  Instruments    DOM5512SD Dissolved Oxygen  Meter with Datalogging SD Card  Please read this user   s manual carefully and thoroughly  before using the instrument     Many commercial and industrial processes carefully monitor the concentration of oxygen  contained in water because dissolved oxygen  
8. a screen that mixes both types of  content         y    E 3 G   oston Tate Tine Cal Valve Chl Unis Ch2 Valte Ch   arit  1 MBE 13 3829 78 eet 29 Dega  2 AWN BEW I8 n  N9 Deps      3 WWB 133882 I8 ug   29 Dogal  4 W BES I6 ugo   Z9 Degrae     5  2009 8 12  13 3236 16 mg   X9 Dege C  6 Oberle 133638 gh 79 Dega                      7 008l 13380 75 mg   29 Deget  8 202  1332 75 mge   29 Deg   C  9 OBZ BRL I5 u  ZBO Degat  10 WWB 13386 F5 mg  29 Deg C  2  1  MORN  133848 75 mge  ZO Dega    Fig  4  Typical Excel data only screen    79  78  7 7  7 6  7 5  74  7 3  7 2       12 e48  amp  4 Sw 1 Ww ds 4  Fig  5  Typical Excel graphics only screen    12                                 A B Z   E F G H  1_ Position Date Time Chi_Valne Chi_Unit Ch2_Valne Ch2_nnit   2 1 2009 8 12  13 38 29 7 8 mg L 27 9 Deyiee_C  3 2 2009 12 13 38 30 7 8 mg L 27 9 Degree _C  4 3  209 8 12  13 38 32 V8 mg L 2 9 Degree_C  5 4 DMOZ 13 38 34 7 6 mg L 27 9 Degree C  6 5 2009 8 12 13 38 36 7 6 mg L 27 9 Degree_C  7 6 2009 8 12  13 38 38 7 6 mg L 27 9 Degree C  g 7 2009 8 12 13 38 40 7 5 mg L 27 9 Degree_C  g g 2009   12 13 38 42 tf  mg L 27 4 Degres_C  10 9 2009 812 13 38 44 7 5 mg  27 9 Degrer_C  aij 10 2009 6812 13 38 46 7 5 mg L 27 9 Dewe C  12 11 2009 7 12 13 38 48 7 5 mg L 27 9 Degree_C  13 A  14  15  16    1     1k  19  20  21  22    23  24 9 19 11 12 13 14  2A  Fig  6  Typical Excel mixed graphics data screen  SPECIFICATIONS  Embedded microcontroller Custom single chip LSI device  Display type Liquid cryst
9. al with green backlight  Display size 2 0 x 1 5 in   52 x 38mm   DO measurement range 0 to 20 mg L  DO measurement resolution 0 1 mg L  DO measurement accuracy   0 4 mg L  Oxygen in air measurement range 0 to 100   Oxygen in air measurement resolution 0 1  Oo  Oxygen in air measurement accuracy  0 7  0   Temperature measurement range 32   to 122  F  0   to 50  C   Temperature measurement resolution 0 1  F or   C  Temperature measurement accuracy  1 5  F   0 8  C   Compensation for temperature Automatic    Compensation for altitude    0 to 29 200 ft   0 to 8 900 meters        13    Compensation for liquid salt content    0 to 50        Datalogging sampling time range    1 second to 8 hours   59 minutes  59 seconds             DO probe type Polarographic   SD memory card capacity 1 GB to 16 GB   Power source 6 Alkaline    AA    batteries or  9 VDC adapter   Power consumption 14 mADC  normal operation  with backlight off  and SD card not saving data   37 mADC with  backlight on and card not saving data  49 mADC  with backlight on and card saving data   Operating temperature 32   to 122  F  0   to 50  C    Operating relative humidity 0 to 85        Dimensions of meter dimensions    7 0 x 2 7 x 1 9 in   177 x 68 x 45mm        Weight of meter 1 08 Ib   489g   Dimensions of probe 7 5  long  x 1 1  across  in   190 x 28mm   Weight of probe 0 74 Ib   335g     MAINTENANCE  amp  TROUBLESHOOTING TIPS  Keep the probe clean and its head inside the included protective cover when not in use 
10. alue as the temporary default     Once the altitude compensation in feet has been set and stored  the display will then  either return to the first of the ten steps in the setup sequence  choosing whether or not  to format the SD memory card   or show the term    ESC     If    ESC    appears  press the  HOLD ESC button to exit the setup procedure and enter normal operating mode  In this  mode  the display will show    mg L    units by default    CALIBRATING THE OXYGEN PROBE    The DOM5512SD   s oxygen probe must be calibrated before you use it for the first time   and again before each use after a long period  weeks or months  of sitting idle  General  also recommends calibrating the probe before making any measurements that must be  extremely precise     1  Before calibrating the probe  make sure that it is full of fresh electrolyte  Use the  following procedure   while referring to Fig  2   either to fill the probe with electrolyte  or to replace old electrolyte with fresh electrolyte     1  Unscrew the probe head from the probe handle    2  Empty and dispose of any residual electrolyte from the probe head   s reservoir   3  Pour fresh electrolyte into the reservoir    4  Screw the probe head back into the probe body     PROBE PLUG          _______      f     HANDLE  2  Fill the probe head with electrolyte  A  To calibrate the DOM5512SD  use re    i        the following procedure while    referring to Figures 1 and 3        As    7      val caine  ELECTROLYTE    Fig  2  Fil
11. ear at the upper right of the display  As in automatic datalogging  mode  in this mode the instrument is storing measurements and their time stamps on the  SD card     In manual datalogging mode  however  measurements are being stored continuously  with  a sampling time of zero   and their locations on the card can be changed     5  To change the storage location of manually logged data  press the W button once   this causes the    P    to disappear from the left side of the display and the value on its  line to begin flashing  Once the flashing begins  you can use the W and A buttons to  change the flashing value to any number between 1 and 99  Once you have chosen the  storage location  press the REC ENTER button to save the setting  This causes the value  to stop flashing and the    P    to return     6  To end manual datalogging  press the REC ENTER button and hold it for at least three  seconds  This action causes the    REC    message to disappear from the top line of the  display     11    TRANSFERRING DATA FROM THE SD CARD TO A COMPUTER    After automatic or manual datalogging of measurements  remove the SD card from the  DOM5512SD and plug it into your computer either directly  if it has an SD card slot  or  through an SD card reader     Because the files containing time stamped data logs have the extension  xls  they open in  Microsoft   s Excel application  Figures 4  5 and 6 show three kinds of Excel presentations   a data only screen  a graphics only screen  and 
12. ess the button again     Whenever the instrument is in measurement mode  you also can check the current date  and time by briefly pressing the TIME button  Doing so causes both values to appear  briefly at the lower left of the display     HOLDING AND STORING MEASUREMENTS    1  To hold a measured value  press the HOLD ESC button during the measurement   Doing so will cause the word    HOLD    to appear at the top of the display  Pressing the  HOLD ESC button again releases the hold     2  To record and recall readings  press the REC ENTER button while making  measurements  This will make the term    REC    appear at the top of the display   Pressing the REC ENTER button again  briefly  will make the term    MAX    appear to the  right of    REC    and switch the display to the maximum value stored in memory during  the last recording session  Pressing the REC ENTER button again  briefly  will make the  term    MIN    appear to the right of    REC    and switch the display to the minimum value  stored during the last session     3  To exit recording mode  press and hold the REC ENTER button for at least three  seconds  until the term    REC    disappears from the top line of the display  The display  will then revert to showing the current reading     10    AUTOMATIC VS  MANUAL DATALOGGING    The DOM5512SD can automatically log data at a user selected sampling period from 1  second to 9 hours  To view the sampling time for which the instrument was set up  press  the LOGGER SET
13. g the POWER button on the front panel once   e Insert the probe at least 4 in   10cm  into the liquid to be measured     e Wait a few minutes for the instrument to automatically compensate for any  difference in temperature between the probe and the liquid     e Read the DO content of the liquid in mg L on the upper display and its temperature  on the lower display     e Rinse the probe head with tap water after each measurement   e Replace the probe head protective cover        A DO measurement tip    Agitating the liquid reduces the effect and measurement errors caused  by oxygen diffusing into the liquid from ambient air  At a minimum  this  can be done by shaking the probe while it is immersed  To minimize  measurement errors of this type  General recommends using a  magnetic agitator  especially in laboratory settings           2  To measure the oxygen content of a volume of air  first make sure that the oxygen  probe is full of fresh electrolyte  calibrated and correctly inserted  and that the meter is  powered on  Then press the W button  which has the word    Function    stenciled above  it  to make the display show     02    instead of    mg L     Wait at least five minutes until  the readings stabilize  To return to    mg L    readings  press the W button again     Whenever the DOM5512SD is in measurement mode  you can turn off the backlight   which is on by default  by briefly pressing  but not holding  the POWER ESC button   To reactivate the backlight  briefly pr
14. ling the probe head with electrolyte    C       3 1  Fig  3  The oxygen probe   s main components i   3 2  3 1     Probe plug  3 2     Probe handle  3 3     Temperature sensing metal  3 4     Probe head    3 5     Probe head protective cover    3 3           7          3  Plug the probe plug into the DO probe socket at the top of the DOM5512SD     4  Power on the instrument by pressing the POWER button on the front panel once    Powering the unit through a 9V AC DC adapter   an optional accessory   disables the  POWER button      5  Press the V button  which has the word    Function    stenciled above it  once  the  display will then show     02    instead of    mg L        6  Wait at least five minutes until the display readings stabilize     7  Use two fingers to press the REC ENTER and HOLD ESC buttons at the same time   Release your fingers once the display shows the following  indicating that air typically  has an oxygen content of 20 9      20 9    CAL  8  Press the REC ENTER button  The display will then count down from 30 seconds to 0   When the countdown is complete  press the W button once to return to the normal  display of    mg L      NORMAL OPERATION  1  To measure the percentage of oxygen dissolved in a liquid   e Make sure that the probe is full fresh electrolyte and has been calibrated recently   e Remove the protective cover from the probe head   e Plug the probe plug into the DO probe socket at the top of the instrument   e Power on the DOM5512SD by pressin
15. ogging SD Card is warranted to the original purchaser to be free from defects in  material and workmanship  Subject to certain restrictions  General will repair or replace  this instrument if  after examination  the company determines it to be defective in  material or workmanship for a period of one year     This warranty does not apply to damages that General determines to be from an  attempted repair by non authorized personnel or misuse  alterations  normal wear and  tear  or accidental damage  The defective unit must be returned to General Tools  amp   Instruments or to a General authorized service center  freight prepaid and insured     Acceptance of the exclusive repair and replacement remedies described herein is a  condition of the contract for purchase of this product  In no event shall General be liable  for any incidental  special  consequential or punitive damages  or for any cost  attorneys     fees  expenses  or losses alleged to be a consequence of any damage due to failure of  or  defect in any product including  but not limited to  any claims for loss of profits     RETURN FOR REPAIR POLICY    Every effort has been made to provide you with a reliable product of superior quality   However  in the event your instrument requires repair  please contact our Customer  Service to obtain an RGA  Return Goods Authorization  number before forwarding the unit  via prepaid freight to the attention of our Service Center at this address     General Tools  amp  Instruments
16. or  desktop computer either directly or via a USB card reader  The data logs are stored on the  card as files with the  xls extension that can be opened by Microsoft   s Excel application     The DOM5512SD has a backlit 1 5 x 2 0 in  display and is powered by six    AA    Alkaline  batteries or an optional 9 VDC AC adapter     KEY FEATURES  e Uses heavy duty polargraphic oxygen probe  e Precisely measures DO from 0 to 20 mg L  as well as liquid temperature  e Easy to calibrate  e Supports automatic and manual datalogging  e Time stamps and stores measurements on SD memory cards  e Sampling time settable from 1 second to 9 hours  e Automatically compensates for altitude and salt content of liquids  e 1 5 x 2 0 in  backlit liquid crystal display  e Powered by six    AA    batteries or optional 9 VDC adapter  e Readings switchable between   F and   C    OPERATING INSTRUCTIONS  WHAT   S IN THE CASE    The DOM5512SD comes fully assembled in a hard carrying case along with the following  accessories     e An oxygen probe   e Electrolyte to fill the probe   e Two spare probe heads with diaphragm set  e A 2 GB SD memory card   e This user   s manual    Figure 1 shows all of the controls and indicators on the front  back  top  bottom and right  side of the DOM5512SD  Familiarize yourself with their positions and functions before  moving on to the setup and calibration procedures     1 13    1 11    1 12  1 10   1 9  1 16  1 15  1 14    1 3  14    1 6  1 8       1 17    Fig  1  The con
17. t  Once the format of decimal divisions has been set and  stored  the display will show the term    t CF     Press the A or W button until your  preferred unit   degrees Fahrenheit    F  or degrees Celsius    F     is displayed  Then  press the REC ENTER button to store the setting as the default     10  Compensate DO measurements for the presence of salt in the liquid  Once the  temperature unit has been selected and stored  the term    SaLt    will appear in the  lower portion of the display  Press the A or W button repeatedly until the reading in  the upper portion of the display matches the known percentage of salt in a liquid  sample  Then press the REC ENTER button to store this value as the temporary default   until the instrument is powered off      11  Compensate DO measurements for altitude in meters  Once DO measurements  have been compensated for salt  the term    High     will appear in the lower portion of  the display  Press the A or W button repeatedly until the reading on the upper portion  of the display matches your altitude in meters  Then press the REC ENTER button to  store this value as the temporary default     12  Compensate DO measurements for altitude in feet  Once DO measurements have  been compensated for salt  the term    Highf    will appear in the lower portion of the  display  Press the A or W button repeatedly until the reading on the upper portion of  the display matches your altitude in feet  Then press the REC ENTER button to store  this v
18. trols and indicators and other physical features of the DOM5512SD    1 1 Liquid crystal display 1 10 Battery compartment cover   1 2 POWER button 1 11 Battery compartment cover   also controls backlight  screws   1 3 HOLD ESC button 1 12 Tripod attachment nut   1 4 REC ENTER button 1 13 Socket for DO probe   1 5 A button 1 14 Socket for 9V AC DC adapter   1 6 V button 1 15 RESET button   1 7 TIME button 1 16 RS 232 output jack   1 8 LOGGER SET button 1 17 SD card socket    1 9 Kickstand    SETUP    1  Choose the power source  Before using the DOM5512SD  be sure it is powered by  fresh batteries or an optional 9VDC AC adapter plugged into the bottom jack on its right  side  callout 1 14      To remove the battery compartment cover  remove the two screws holding it in place   callout 1 11   Then install six    AA    batteries in the correct orientation  using the polarity  marks on the inside of the compartment as a guide  Replace the cover by replacing the  two screws     2  Install an SD card  To prepare for setup  also install the supplied 2 GB SD memory  card or another card with a capacity from 1 GB to 16 GB in the socket on the bottom of  the DOM5512SD  callout 1 17   When installing the card  make sure its gold contacts  are facing front and push the card into the socket until you hear a click  To remove the  card  push it in until you hear a click and the card pops out     Now power on the instrument by pressing the POWER button  callout 1 2  to generate a  short beep  
19. ure  press the  REC ENTER     Ent    will then flash several times and the instrument will sound another  three beeps to confirm that the SD card has been erased and formatted     4  Set the date and time  Once the SD card has been set up  the word    dAtE    will appear  in the center of the display  along with the flashing value    00 00 00    at the lower left   above    yy mm dd     Set the current year by pressing the W or A button repeatedly until  the correct value appears above    yy     Press the REC ENTER button to store the setting   The next screen that appears will have the value above    mm    flashing  Use the W or    A button to navigate to the current month and press the REC ENTER button to store  the setting  When the next screen flashes the value above    dd     again use the Y or A  button to navigate to the current day and press the REC ENTER button to store the  setting   Once you have set the date  the display will prompt you to set the hour  minute and  second of the current time  Again use the W or A buttons to navigate to the correct  values  and the REC ENTER button to store the settings     5  Set the datalogging sampling time  Once the date and time have been set and  stored  the display will show a value below the letters    SP t     Press the Y or A  button to decrease or increase the flashing hour value until the desired number  appears below    SP t     Press the REC ENTER button to store that setting  The minutes  value will then flash  Press
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Octobre 2014  Mode d`emploi    ASPIRATEURS INDUSTRIELS TYPE KF  Enter o to this page the details for the document  Arrow Fastener T50 Use and Care Manual  Bouillie bordelaise - Documents pour le développement durable de l  prefeitura da estância hidromineral de poá  FICHA DATOS TÉCNICOS DESENGRASANTE DE MOTOR DS-600  取扱説明書 太陽光発電用パワーコンディショナ    Copyright © All rights reserved. 