1. Blank values are saved automatically 6 3 3 Negative controls You must set one negative control in each program unless the negative control factor is 0 6 3 4 Positive controls Displays the negative control setting Section 6 3 5 Standard This option is valid only when a program needs a standard If the program already has a standard value it can be used or a new value can be entered Note If a program sets a standard and the test results are good new standard values will overwrite the old values 6 3 6 Clear If marked wells are to be deleted press the clear button then click on the wells to be deleted 16 30 HumaReader HS User Manual 6 4 Select All Sample wells can be selected using this button Press the select all button to display this window Input start No amp well Start No 100 Start well A To limit the selection input the start number 1 999 and start well column number and line number and press the Ok button All of the wells in this range will be marked as samples With this option all selected programs and wells will be overwritten eg T mM g nm P H 09510661087100918091090109110921099109410951096 Program ABS Calculation Absorbance Select allf Clear all 6 5 Clear All Pressing the Clear all button 6 6 Test To begin testing the plate press the start button Press the stop button to stop testing Human Os Diagnost
2. Storage environment Lamp Standard wavelengths Abs range Measurement range Accuracy Precision Linearity Reading speed Warm up time CPU Memory Interface Display Input 10kg 450mm L x 330mm W x140mm H 110 250VAC 50 60Hz 250V 3 15A temperature 10 C 40 C humidity 20 85 20 C 50 C OSRAM64607 8V SOW 405 450 492 630nm substitute filters from 340 700nm available 0 3 500A 0 2 000A 0 2 0A 1 0 or 0 007A 0 2 0A 0 5 or 0 005A 2 0 or 0 007A Continuous mode lt 5s step by step mode lt 15s 1 minute Embedded RISC CPU 200 tests 1 000 patient data records and 10 000 sample records RS 232C serial interface parallel printer interface 5 7 LCD display 320x240 discernibility 256 gray scale Touch panel and pen external mouse special order Human EP Diagnostics Worldwide 5 30 3 OPERATION INSTRUCTIONS 3 1 Touch Panel amp Pen The HumaReader HS features a touch screen For operation a touch pen is included with the instrument accessories Alternatively an RS 232 mouse can be connected 3 2 Software Number Keypad Input patient namei lelle ctear ox Cancel In order to input numbers such as concentrations names etc a software keypad is provided 3 3 Software Keyboard Input Operator s Name caplalslalrlaln ilkl lil shit 2 x cv pn m 1 17 celati TAL Pe te In order to
3. may may not be mounted correctly Check it Driver motor problem Open the cover and check the driver motor Maybe use serial mouse Turn the HumaReader HS off Disconnect the mouse from instrument and turn it on again Problem with printer power supply Check connections Make sure the printer is turned on Check that the correct printer type is selected in the printer settings Turn the HumaReader HS on first then turn the printer on Make sure the printer is correctly connected to the HumaReader HS Clean the printer head and or change the ink cartridge See the printer user manual See the printer user manual See the printer user manual HUMAN Gesellschaft fur Biochemica und Diagnostica mbH Max Planck Ring 21 65205 Wiesbaden Germany Tel 49 61 22 99 88 0 Fax 49 61 22 99 88 100 e Mail human human de www human de 03 2008 04
4. pipetted into two consecutive wells the mean absorbance reading of both wells will be used to calculate a single concentration Human I Diagnostics Worldwide 11 30 3 Positive and negative settings window Qualitative Positive threshold value Negatives 1 000 sic Positive lt 000 SJEO Normal 0 000 u i 000 Qualitative mode With this option regular COV or reverse COV or none can be selected Regular COV Samples with values higher than the positive cutoff are labeled as positive Samples with values lower than the negative cutoff are labeled as negative Any samples with values falling between the negative cutoff and the positive cutoff are labeled as equivocal Reverse COV Samples with values lower than the positive cutoff are labeled as positive Sample with values higher than the negative cutoff are labeled as negative None if selected negative positive will not be determined Normal range values from the package insert can be input Pressing the Finished button returns to the program list window 5 3 Create program Pressing the new button opens a new program window into which the name parameters etc can be input see Edit Program If the program name exists the system will warn program exists 5 4 Delete program After selecting a program from the program list pressing the delete button will display the following window 7 Are you sure delete ABS Yes Mo P
5. printer off connect the parallel cable to the rear of the instrument Plug the other end of the parallel cable into the printer Human I Diagnostics Worldwide 3 30 2 FUNCTIONAL DESCRIPTION 2 1 Introduction 1 The HumaReader HS is a microprocessor controlled general purpose photometer system designed to read and calculate the results of ELISA assays in microtiter plate 2 All HUMAN Methods are preprogrammed 3 Multiple calculations are available Absorbance mode ABS Cut off mode Single standard mode Point to point mode absorbance multi point mode Linear regression mode Exponent regression mode Logarithm regression mode Cubic Spline Power regression mode 4 The instrument can evaluate up to 12 different tests in one microplate 5 Test reading time is 5s plate 6 Memory 200 programs 1 000 patient data records and 10 000 sample records 7 Features a department database operator database and system log database 2 1 1 HumaReader HS Front view 1 Power pilotlamp Lit when the instrument is powered on 2 Touch panel Display program 3 Plastic cover Plate carrier Microplate inside 4 30 HumaReader HS User Manual 2 1 2 HumaReader HS rear view 1 Power switch 2 Power socket 3 Hinge 4 Fuses 5 Hinge 6 RS 232 interface 7 PRINTER interface 8 PSTN connector 2 2 Technical Specifications Weight Overall dimensions Power Fuses Work environment
6. AN representative shall be relieved of any liability under this warranty if the product is not used in accordance with the manufacturer s instructions altered in any way not specified by HUMAN not regularly maintained used with equipment not approved by HUMAN or used for purposes for which it was not designed HUMAN shall be relieved of any obligation under this warranty unless a completed installation warranty registration form is received by HUMAN within 15 days of installation of this product This warranty does not apply to damages incurred in shipment of goods Any damage so incurred shall be re ported to the freight carrier for settlement or claim 3 INTENDED USE OF THE INSTRUMENT The instrument has to be used for the expected purposes and in perfect technical conditions by qualified personnel in working conditions and maintenance operations as described in this manual according to the GENERAL SAFETY WARNINGS This manual contains instructions for professional qualified operators 4 GENERAL SAFETY WARNINGS Use only chemical reagents and accessories specified and supplied by HUMAN and or mentioned in this manual Place the product so that it has proper ventilation The instrument should be installed on a stationary flat working surface free from vibrations Do not operate in area with excessive dust Work at room temperature and humidity according to the specifications listed in this manual Do not operate this instrument with cov
7. HUMAREADER HS User Manual C Cat No 16670 1 No DATE Rev 01 2007 02 02 2007 03 03 2008 04 REVISION LIST OF THE MANUAL REVISION DESCRIPTION First edition Correction of typing errors Addition of Calculation Mode Chapter 5 Human Diagnostics Worldwide 1 INTRODUCTION This manual is considered as a part of the instrument it has to be at the operator s hand as well as at the maintenance operator s availability For accurate installation use and maintenance please read the following instructions carefully In order to avoid instrument or personal damages carefully read the GENERAL SAFETY WARNINGS describing the suitable operating procedures In case of breakdowns or any troubles with the instrument apply to the local Technical Service 2 USER WARRANTY HUMAN warrants that instruments sold by one of its authorised representatives shall be free of any defect in material or workmanship provided that this warranty shall apply only to defects which become apparent within one year from the date of delivery of the new instrument to the purchaser The HUMAN representative shall replace or repair any defective item at no charge except for transportation expenses to the point of repair This warranty excludes the HUMAN representative from liability to replace any item considered as expendable in the course of normal usage e g lamps valves syringes glassware fuses diskettes tubing etc The HUM
8. aving 7 2 4 Print report Press the print button to print all marked patient reports Human I Diagnostics Worldwide 21 30 7 3 Print By Program list Select program and press Ok to display the program list window no Program Test date Press Print to print all of the results for the selected program 7 3 1 Preview report Select a program and press the preview button to display the following window m 7 3 2 Delete program Press the delete button and confirm with yes All marked programs and test data will be deleted 7 3 3 Print report Press the print button to print the standard curves and test data for all marked tests 22 30 HumaReader HS User Manual 7 3 4 Printing multiple records The HumaReader HS supports the printing of multiple records you can quickly print all negative or positive patient data for any given day Press the gather button to open this window Input condition _ Date 001 ipe spe Program ET ok 00009 Input the date select a program and results from the list and press OK All corresponding patient records will be displayed in the window 16 Name Sex Age Sample Press the print button to print all selected data Human I Diagnostics Worldwide 23 30 8 Communication 8 1 Communication with PC The HumaReader HS can be connected to a PC to send data to the PC With the HumaRead
9. be switched off safely Note Leave the HumaReader HS off for at least 30 seconds before turning it on again 28 30 HumaReader HS User Manual 13 INSTRUMENT SERVICE 13 1 Maintenance The HumaReader HS is designed to be a maintenance free instrument To insure optimum trouble free performance however the instrument should be kept dry and operated in an area free from excessive dust 13 2 Cleaning The Instrument Keep the working environment clean Use a slightly damp soft cloth to remove dirt or spills For decontamination 70 isopropanol is recommended Use a soft cloth to remove dirt or spills from the LCD screen Note The use of chemicals or abrasive cleaners is not recommended 13 3 Replacing parts 13 3 1 Changing fuses 1 2 3 Disconnect the power cord from the mains supply before replacing fuses Open the fuse box to the left of the power socket to change a fuse Fuse specification 250V 3 15A Reconnect the power cord and turn the instrument on Notice Use only a fuse with the same rating for replacement 13 3 2 Changing the lamp Turn the instrument off and open the instrument cover by unscrewing the screw on the front of the instrument Remove the four screws fastening the optical assembly cover Remove the cover Lift up the lamp with the terminal socket Pull the terminal socket from the lamp contacts Refit the terminal socket to the contacts of the new lamp OSRAM64607 8V 50W Replace the te
10. cel button in the absorbance window The instrument will automatically store the test results in the history database Human I Diagnostics Worldwide 19 30 7 REPORT 7 1 Report form Press report in the main menu to display this window Canon BIC 250 seri _ Report by patient Program Date Today AIl Co Je Select the format of the report Reportform Report by patient list or program list Daterange Select today or all 7 2 Report By Patient list Select patient and press the Ok button to open the following window no oste sample no_ name aj 1 Select one patient 2 Select all patients Clicking the diamond in the first line of the title bar will select all patients 20 30 HumaReader HS User Manual 7 2 1 Edit patient data Select a patient and press Edit The patient s information will be shown thus Input patient information Date 2003 06 09 Sample NO 001 Name Age Sex Male g Female MN Unknow Dept Operator Bed ID Sender 0 Make any desired changes Press Ok to save the changes Press cancel to exit without saving 7 2 2 Preview report Select a patient and press the preview button to display the patient s data Program name Cancel 7 2 3 Delete patient data Pressing the delete button Pressing yes will delete all marked data Press no to exit without s
11. e Press Ok to add the operator s name to the list If the operator s name already exists the system will display this warning Select a different operator name If database is full the system will report operator s records is over 2 Delete record Select an operator in the operator list and press the Delete button Press Yes to delete press No to cancel Human OP Diagnostics Worldwide 27 30 11 SYSTEM LOG Press the event log button in the main menu to view the history log The event log contains the date time type and status of each event The HumaReader HS can store a maximum of 2500 events The most recent even will be listed first Pressing Clear will delete all event records Press Cancel to return to the main menu 12 POWER OFF i i xP Power off Are you sure Press the Power off button in main menu to shut down the instrument Press Yes confirm or no to exit to the main menu Note If the instrument is not shut down normally through the software results and parameters edited during the session willl not be saved During shutdown the system will display this window Warning The power should not be switched off while the HumaReader HS is shutting down A power interruption during shutdown may damage the instrument When the data has been saved the following message will appear 5 Please power off The power can now
12. e Human m Diagnostics Worldwide 9 30 5 2 Edit Program Select an existing program Press the Edit button to display the following window Program 10599 Full name PO Reagent n First wavelength a50 al Second wavelength none Calibaration mode absorbance Double sample Blanks 0 050 Program If the name exists it should not be replaced or changed Full name For inputting the full name of the program optional Reagent For inputting the reagent name optional Wavelength and second wavelength Select the appropriate options Calibration mode Select from the list Double sample If checked samples will be pipetted into two consecutive wells and the mean absorbance reading of the two wells will be used to calculate a single concentration Blank This field is used to enter the out of range value for the blank absorbance If the blank absorbance is out of range system will display blank abs is out of range Press the button to continue to the next windows or press the K lt button to return to the previous window Pressing the cancel button will exit the program configuration window 1 Standard mode settings window Unit ba Double standard Standard Number Upto 8 can be selected Unit Units can be selected from the list If the required units are not included additional units can be entered in the Information menu Double standard If selected standards w
13. enter names and alphanumeric information a software keyboard is provided It can be used like a PC keyboard and can be moved on the screen using the drag and drop function 6 30 HumaReader HS User Manual 4 OPERATION 4 1 Power on After being switched on the instrument performs a self check and loads the stored data into the working memory An error report will be displayed if the instrument does not pass the self check 4 2 Main menu After the self check the main menu is displayed Human oF G Diagnostics Worldwide 7 30 5 Program By pressing the program icon a list of the preprogrammed HUMAN tests will be displayed No Program Full name Calculate 5 1 Mode setting The HumaReader HS supports 10 calculation modes 5 1 1 ABS mode Only the absorbance will be calculated and printed 5 1 2 Cut off mode Cut off formula Cov X x NC Y PC Fac NC is the mean of the negative controls PC is the mean of the positive controls Cov is the cutoff absorbance X Y Fac are coefficients that have any positive and negative numerical value including 0 and 1 Example Cut Off NC 0 1 x PC X 1 Y 0 1 Fac 0 In regular cutoff mode the HumaReader HS indicates that the sample is positive if the sample OD Cov gt 1 5 1 3 Calculation mode 1 Single standard mode A single calibrator material of known concentration is used so that concentrations of unknown samples may be Abs Conc calcu
14. er HS software you can store patient lists monitor curves print reports and collect QC data 1 Start the HumaReader HS and PC and connect the serial serial cable to the HumaReader HS serial port Plug the other end into the COM 1 port of the PC 2 Press the Pc button in in communication window to open the following window Waiting connection Send 3 Run the PC software and open the data send menu to receive the data 4 After data transmission is complete press the Cancel button to return to the main menu 24 30 HumaReader HS User Manual 9 SYSTEM SETTINGS Press the System button in the main menu to open the settings window Serial No 123456788 Hospital Cold harbor hospital pooo Zhz ps Qi Sound onO Off Cancel 1 Serial No Displays the instrument serial number 2 Hospital Enter the hospital name in this field 3 Date and time Date format is year month day time in 24 hour format 4 Printer The HumaReader HS supports 5 printer series the Canon BJC 250 series HP DJ 340 400 series HP LJ 5L 6L series Epson LQ series Epson Stylus series Select the correct series from the list Serial No 123456788 Hospital Cold harbor hospital 2 Date foo f 12 y Time pa im 108 Printer Epson stylus series p Contrast Canon BJC250 series HP DJ 340 400 series Sound HP LJSL 6L series Epson LQ series Note If the printer is not on the co
15. ers and panels removed Only use the power cord specified for this product with the grounding conductor of the power cord connected to earth ground Use only the fuse type and rating specified by the manufacturer for this instrument use of fuses with improper ratings may pose electrical and fire hazards To avoid fire or shock hazard observe all ratings and markings on the instrument Do not power the instrument in potentially explosive environment or at risk of fire Prior to cleaning and or maintaining the instrument switch off the instrument and remove the power cord For cleaning use only materials specified in this manual otherwise parts may become damaged It is recommended always to wear protective apparel and eye protection while using this instrument Respective warning symbols if appearing in this manual should be carefully considered Human a Diagnostics Worldwide 5 DISPOSAL MANAGEMENT CONCEPT The currently valid local regulations governing disposal must be observed It is in the responsibility of the user to arrange proper disposal of the individual components All parts which may comprise potentially infectious materials have to be disinfected by suitable validated procedures autoclaving chemical treatment prior to disposal Applicable local regulations for disposal have to be carefully observed The Instruments and electronic accessories without batteries power packs etc must be disposed of according to the re
16. gulations for the disposal of electronic components Batteries power packs and similar power source have to be dismounted from electric electronic parts and disposed off in accordance with applicable local regulations 6 INSTRUMENT DISINFECTION Analytical instruments for in vitro diagnostic involve the handling of human samples and controls which should be considered at least potentially infectious Therefore every part and accessory of the respective instrument which may have come into contact with such samples must equally be considered as potentially infectious Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated parts Before the instrument is removed from the laboratory for disposal or servicing it must be decontaminated disinfected Decontamination disinfection should be performed by a authorised well trained personnel observing all necessary safety precautions Instruments to be returned have to be accompanied by a disinfection certificate completed by the responsible laboratory manager If a disinfection certificate is not supplied the returning laboratory will be responsible for charges resulting from non acceptance of the instrument by the servicing centre or from authority s interventions 7 NOTICE Every effort has been made to avoid errors in text and diagrams however HUMAN GmbH assumes no responsibility for any errors which may appear in this publication It is the
17. he next window Blank mc pc sto ac ar 9 10 11 12 Program Calculation Select allf Clear all 14 30 HumaReader HS User Manual 6 2 Select Test Program The HumaReader HS supports 12 programs for one plate Press the New button to display the following window lan E E G a R Troi nmo nm P OO OT II AM Program ABS Calculation 4bsorbance Select allf Clear all Select the desired program and press OK NOTE If multiple programs are tested on one plate the programs must be set one by one STD standard CLR clear 6 3 Mark well Each well s function can be selected Select the function of the well first then click the well to be marked fat B _ D E F E H Program ABS Calculation Absorbance Select allf Clear all mark label sample 1 999 blank B negative controls NC positive controls PC standard S1 58 Quality Control QC Human AI Diagnostics Worldwide 15 30 6 3 1 Sample Click the well to be tested to mark it If the sample no is changed that well must be clicked again The following window will be displayed Input No 1 999 HEBBEBE BHRBROE ee E i m Input the sample no again If the no already exists the system will display the message sample no already exists input again 6 3 2 Blank The blank well option or no in different program A blank well can be set for only one program at a time
18. ics Worldwide 17 30 6 7 Results The test results will be displayed as below The absorbance results will be displayed in columns 1 6 pressing the button will display the absorbance results for columns 7 12 In field Al in the picture above 001 represents the sample number and 0 040 is the absorbance result Note If absorbance results are gt 3 500A the instrument will display 3 500 and calculate with 3 500 if the absorbance is lt 0 000A the instrument will display 0 000 and calculate with 0 000 The well result can be modified by clicking on the well which opens the window below Change Abs 0 3 000 POOooe BAHEN 000 0 Jonee E Input the absorbance value and press Ok Note Changing the standard or control absorbance will affect all of the results in the selected program 18 30 HumaReader HS User Manual 6 8 Calculation Results Press the Result button in the absorbance results window to display the calibration results tTo m mono gt 200 _ han UE TTE UE LO The values and negative positive symbol will be displayed for each well If the values in any wells is lt 0 0 the well ux will be flagged with 6 9 Printing Note OTA quantitative analysis OLA qualitative analysis Press the Print button in the testing window the HumaReader HS will print all test results 6 10 Storing Test Results Press the can
19. ill be pipetted into two consecutive wells the mean absor bance reading of the two wells will be used to calculate a single concentration Standard conc Standard concentration can be entered into the list using the software keypad 10 30 HumaReader HS User Manual Input standard conc eet te 2 Cut off mode settings window FORMULA Math List y Control v Logic List v cut off v cut off panco el Cut off formula parameter Cov X N Y P Fac Math list Control NC NC NC1 NC2 NC3 NC4 NC5 NCx PC PC1 PC2 PC3 PC4 PC5 PCx CR CR1 CR2 CR3 CR4 CR5 CRx BLK Logic list gt lt lt gt AND OR Type Cut off Val1 Val2 Val3 4314 Val5 Val6 Val7 4318 and Val9 Note NC PC CR Represents NCi and PCi when it acts as the range other represents mean value NCx PCx and CRx NCx PCx and CRx mean value of Nci PCi and CRi can not be omitted such as 1 2 NCx The can not include the AND OR Invalid input 0 12 lt NC lt 0 4 Valid input NC gt 0 12 NC lt 0 4 NC and PC should not be used with OR and AND NC1 lt PC1 AND PC2 lt NCx is permitted Range of negative and positive if absorbance of the negative control or positive control is out of range the HumaReader HS calculates using min or max absorbance Double controls if selected controls will be
20. k Ring 21 65205 Wiesbaden Germany Tel 49 61 22 99 88 0 Fax 49 61 22 99 88 100 e Mail human human de www human de Human oO Diagnostics Worldwide CONTENTS INSTALLATION 1 1 Unpacking 1 2 Environmental Requirements 1 3 Electrical Setup 1 4 External Printer Connection FUNCTIONAL DESCRIPTION 2 1 Introduction 241 HumaReader HS Front view 2 1 2 HumaReader HS rear view 2 2 Technical Specifications OPERATION INSTRUCTIONS 3 1 Touch Panel amp Pen 3 2 Software Number Keypad 3 3 Software Keyboard OPERATION 4 1 Poweron 4 2 Main menu Program 5 1 Mode setting 51 ABS mode 5 1 2 Cut off mode 5 1 3 Calculation mode 5 2 Edit Program 5 3 Create program 5 4 Delete program 5 5 Standard RUNNING THE TESTS 6 1 Plate Parameter Settings 6 2 Select Test Program 6 3 Mark well 6 3 1 Sample 6 3 2 Blank 6 3 3 Negative controls 6 3 4 Positive controls 6 3 5 Standard 6 3 6 Clear 6 4 Select All 6 5 Clear All 6 6 Test 6 7 Results 6 8 Calculation Results oO oo CO CO N N N A A A A VU Wp A A U W WD WW WwW PP e BE Pe B H uU eA BR WN N O PrRPrRrPrP FPF DAAAA OO VI PP RP B RB O oN N N 6 9 Printing 19 6 10 Storing Test Results 19 7 REPORT 20 7 1 Report form 20 7 2 Report By Patient list 20 7 2 1 Edit patient data 21 7 2 2 Preview report 21 7 2 3 Delete patient data 21 7 2 4 Print report 21 7 3 Print By Program list 22 13 1 Preview report 22 73 2 Delete program 22 7 3 3 Print rep
21. lated according to Beer s Law 8 30 HumaReader HS User Manual 2 Point to point mode 2 8 standards can be set The resulting calibrator curve is a series of lines connecting the calibrator points which may be entered in ascending or descending order of absorbance Abs Abs Conc Conc 3 ABS 2 8 standards can be set The highest absorbance standard will be assigned a value of 100 and each sample and standard is reported as a percent of the absorbance of the highest calibrator The resulting calibrator curve is a series of lines connecting the calibrator points which may be entered in ascending or descending order of absorbance 4 Linear regression mode 2 8 standards can be set This is a multi point standard mode that calculates a best fit linear equation based upon the standard points Equation Y kX b It is used when the absorbance and the concentration are linear 5 Exponent regression mode 2 8 standards can be set Equation Y ke This is used when the natural log of the absorbance is plotted against the concentration 6 Logarithm regression mode 2 8 standards can be set Equation Equation Y kLnX b This is used when the natural log of the concentration is plotted against the absorbance 7 Power regression mode 2 8 standards can be set Equation Y kX This is used when the natural log of the absorbance is plotted against the natural log of the concentration 8 Cubic Splin
22. mpatible list or is incorrectly selected it may not function properly 5 Contrast adjust press the button to increase the contrast and the button to decrease The new settings will be saved automatically 6 Sound Turn the speaker on or off 7 Touch panel calibration To calibrate the touch panel press the touch panel button to open the calibration window Human I Diagnostics Worldwide 25 30 Tap the cross center with the touchscreen pen Wait one second and the cross moves to next position auto matically Tap the cross with pen Repeat this proceedure 5 times When the process is complete the system will display the following window Note If the cross doesn t move after tapping it try again Tap anywhere on the screen within 10 seconds to save the new settings The system will return to the settings window 26 30 HumaReader HS User Manual 10 INFORMATION MANAGEMENT Press the information button in the main menu to open this window is Operator i Department i Sender Delete Cancel The HumaReader HS has an operator database and a department database Operators and department can be added to or deleted from either database The HumaReader HS can store 100 records in each of these two databases Select operator or department In this example the operator database has been selected 1 Add record Press the add button to input the operator name into the field Input operator nam
23. ort 22 7 3 4 Printing multiple records 23 8 Communication 24 8 1 Communication with PC 24 9 SYSTEM SETTINGS 25 10 INFORMATION MANAGEMENT 27 11 SYSTEM LOG 28 12 POWER OFF 28 13 INSTRUMENT SERVICE 29 13 1 Maintenance 29 13 2 Cleaning The Instrument 29 13 3 Replacing parts 29 13 3 1 Changing fuses 29 13 3 2 Changing the lamp 29 13 4 Troubleshooting 30 2 30 HumaReader HS User Manual 1 INSTALLATION 1 1 Unpacking Unpack the HumaReader HS carefully and check for any damage that may have been caused during transportation The package should contain the following HumaReader HS instrument Operator s manual Panel pen Power cable Printer cable RS 232 serial cable Lamp Fuses Keep the original package for future transportation 1 2 Environmental Requirements Locate the HumaReader HS to avoid exposure to excessive dust vibrations strong magnetic fields direct sunlight drafts excessive moisture or large temperature fluctuations Leave sufficient clearance 10 cm at both sides of the unit for adequate air circulation Note The instrument can be operated within an ambient temperature range of 10 40 C and humidity of 20 85 1 3 Electrical Setup Power requirements 110 250 VAC 50 60 Hz 80W Connect the supplied power cable to the rear of the instrument as shown Plug the other end of the power cable into an AC outlet 1 4 External Printer Connection With both the instrument and the external
24. policy of HUMAN GmbH to improve products as new techniques and components become available HJMAN GmbH therefore has to reserve the right to change specifications if necessary in the course of such improvements NOTICE Analytical instruments for in vitro diagnostic application involve the handling of human samples and controls which should be considered at least potentially infectious Therefore every part and accessory of the respective instrument which may have come into contact with such samples must equally be considered as potentially infectious BIOHAZARD The BIOHAZARD warning label must be affixed to instrument prior to first use with biological material Servicing Note Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated parts Before the instrument is removed from the laboratory for disposal or servicing it must be decontaminated Decontamination should be performed by authorised well trained personnel only observing all necessary safety precautions Instruments to be returned have to be accompanied by a decontamination certificate completed by the responsible laboratory manager If a decontamination certificate is not supplied the returning laboratory will be responsible for charges resulting from non acceptance of the instrument by the servicing centre or from authority s interventions HUMAN Gesellschaft f r Biochemica und Diagnostica mbH Max Planc
25. ress the Yes button to delete program Press No to cancel 12 30 HumaReader HS User Manual 5 5 Standard The standard test results and curve are stored by the system The operator can review and print the results After selecting one program from program list Press the standard button and the system will display the standard concentration and absorbance Calibrator conc ail Abs A 1 0 000 0 000 2 1 000 0 500 3 2 000 0 800 4 3 000 1 000 Program Demo Calculation mode Reg Pressing the curve button will display the standard curve 0 000 0 000 1 500 dgl 3 000 Program Demo Calculation mode Reg Press the print button to print the standard curve Note Old standards will be deleted if the program settings are modified Wavelength calculation mode standard number standard concentration etc Human Os Diagnostics Worldwide 13 30 6 RUNNING THE TESTS 6 1 Plate Parameter Settings Press the test icon in the main menu to open this window Plate direction 1 12 lt AH Test mode continue _ Step mixer setting mixer speed None mixer time S Plate direction Select the direction A H or 1 12 Test mode Continuous mode 5 seconds or step mode 15 seconds can be selected Mixer settings Mixer speed Select fast normal slow or none Mixer time Select 1 60 seconds Press the Ok button to continue to t
26. rminal socket with the new lamp Close the optical assembly cover Replace the screws Close the instrument cover replacing the retaining screw Switch the HumaReader HS on Human I Diagnostics Worldwide 29 30 13 4 Troubleshooting Symptom HumaReader HS cannot be switched on Lamp does not light RAM check error Lamp intensity is too high Lamp intensity is too low Motor does not work Chopper wheel moves too fast Filter does not return to zero position Plate does not return to rest position Plate does not move Open COM port error Printer does not work Print is faint or incomplete Paper jam Other printer error 30 30 HumaReader HS User Manual Causes and Remedies Check power supply cord Check fuses Check power voltage Turn HumaReader HS off for 30 seconds and turn it back on Check lamp voltage Check lamp replace if necessary Turn the HumaReader HS off for 30 seconds and turn it back on First turn on the HumaReader HS second turn the printer on If there is a cable connect error turn HumaReader HS off open instrument cover and reconnect cable between the two circuit boards Filters improperly installed Lamp is bad replace Motor is bad Open the cover of the instrument check motor Chopper driver motor error Open the cover of the instrument and check the chopper driver motor Filter wheel error Open the cover of the instrument and check the sensor The plate