Home
        Tenma 72-7210 Manual
         Contents
1.                                         Output Range  1 Display 2 Accuracy  2  MIN  MAX  Resolution  7mHz 500mHz 1mHz No  Specified  0 5Hz 5 Hz 0 01Hz   3  2   5 Hz 50 Hz 0 1Hz  50 Hz 500 Hz 1Hz   0 3  1   0 5k Hz 5k Hz 0 001kHz   0 03  1   5 kHz 50 kHz 0 01kHz  50kHz 500 kHz 0 1kHz  500kHz 5 MHz 0 001MHz  Notes     1     2     The specification is specified within 10  to 100  of maximum output range  and the    waveform characteristics are limited to the specified range also     The display resolution and accuracy is relative to the span of internal counter     Waveform Characteristics    Triangle Wave Linearity                Frequency Linearity  1Hz  100KHz  gt  98    gt 100 KHz No  Specified             Sine Wave Distortion                         Distortion Harmonic ratio  10Hz   100kHz   100kHz  500kHz 0 5MHz  5MHz   lt 2   lt  30dB  lt  20dB  Notes     1     The distortion isn t specified for less than 10 Hz        Square Wave       Transition time    Aberration           lt  50ns          5  of P P am    plitude     50 mV          Output Characteristics                   Amplitude  Load   MIN  CCW  MAX  CW  Attenuator Flatness  Attenuator ON   Attenuator OFF   1kHz  lt 500 kHz 0 5  5 MHz  No  0 2Vp p 2 20Vp p  20dB 1dB  1dB  3dB  Load  50 Q   0 1Vp p 2 10Vp p                      Notes     1  Output impedance is 5004 10    100kHz   2  The amplitude should consider additional tolerance of flatness    3  FCCW  Full counterclockwise  FCW  Full clockwise   4  Output prote
2.     Press EXT COUNTER switch on front panel  Connect external source  sine waveform   10MHz  amplitude of 10 mV rms   Connect a probe to TP600 and to the scope  Adjust  R610 for a square wave display on the scope       Remove the probe from TP600  then adjust C104 for counter reading to    10 0000 10 0002 MHz                            Function adjustment of Triangle wave    Set this FG controls as follows   FUNCTION TRIANGLE  AMPLITUDE MAX  DC OFFSET OFF  FREQUENCY 5 KHz  SYMMETRY OFF  SWEEP OFF       Set the FG controls as follows                    FUNCTION SQUARE  AMPLITUDE MAX  DC OFFSET OFF  FREQUENCY 5 KHz  SYMMETRY OFF                     Adjust R340 for a symmetrical square wave around zero       Frequency Adjustment      Connect the OUTPUT to an external counter     Set the FG controls as follows              FUNCTION TRIANGLE  AMPLITUDE Middle  DC OFFSET OFF                Turn the FREQUENCY control knobs to full CW and select the proper RANGE for a  reading of  gt 5 2 KHz on the display    Adjust R259 for a 50  symmetry waveform    Turn the FREQUENCY control knob CCW and set the frequency to 100 Hz on the    17    OO JO    display counter  It is important to stay in the same range       Adjust R249 for a 50  of symmetry waveform      Repeat steps 5 and 6 for best results on both ends      Select the Frequency range from 5Hz to 50 OHZ      Connect the output to an external counter  Tum the frequency knob to full CW position     Adjust R211 for a frequency reading t
3.    s l ik man sal bek ak ii k   4   O  tp  t CharacteriStiCS ssenari n kk ab e ka kk ak kib da kb kn 5   Other Inputs and Outputs io awa io ok a io a on kk l a ok oak pa kk kb ni pa 6   ee Eer 7  Section 4 GETTING STARTE E 9  Front AVG ie aaa e aj e e an ak ke a pa e n e ae a a kn a e on pen A ERE look balan 9    PLE ANE EN E 11  Preparing For 8 5 38 et a a a ste 12  Reference Operation DEE 13  Section 5 MAINTENAN e 15  ST NEE 15  Section 6 ADJUSTMENT PROCEDURES      u u u u u vdcsc senere reen 16  Required teste quipm ent E 16  lge WE ME 17    Section 1 SAFETY    Review the following safety precautions to avoid injury and prevent damage to this  instrument or any instruments connected to it     Read this manual carefully to ensure your personal safety and to prevent damage  either to the instruments or to equipment connected to them    Before applying power  ensure that the line selector is in the proper position for  power source being used    Make sure that the power cord and the fuse type   rating is specified for this  product before use    This product is grounding through the grounding conductor of the power cord  To  avoid electric shock  the grounding conductor must be connected to earth ground   Before making any connections to the input terminals of the products  ensure that  the properly grounded    To avoid personal injury  never operate the instrument without covers or panels  removed    To prevent product overheating  provide proper ventilation    Do not operat
4.  The main output will sweep the desired range of frequencies according to the input  level of SWEEP in out BNC     Observe the results on the oscilloscope     14    Section 5 MAINTENANCE    A  WARNING     To avoid electrical shock or damaging this instrument  never get water into the  meter     Cleaning    Before cleaning this instrument  make sure the power is switched in OFF position and the  power cord is disconnected from the AC outlet  To clean the meter  wipe the dirty parts  with gauze or soft cloth soak with diluted neutral detergent  Avoid getting the instrument  too wet to prevent the detergent from penetrating into inside parts and causing damages   After cleaning  leave the instrument until it dries completely  Don   t use chemicals  containing benzine  benzene  toluene  xylene  acetone or similar solvents     A  WARNING       The following instructions are for qualified personnel only  To avoid electrical  shock  do not perform any servicing other than that contained in the operating  instructions unless you are qualified to do so       When serving  use only specified replacement parts as described hereinafter       To avoid electric shock  disconnect the power cord from the unit before    changing fuse or input voltage selector     Selecting different input voltage  amp  protection fuse replacement    1  To select different input voltage level of AC 100V  120V 220V or 240V  simply switch to  the selected voltage  so the desired voltage is displayed in the windo
5.  for 100k Hz output       Adjust R309 for minimum distortion and  lt 1 2      To check frequency range for 50 Hz  If the distortion  gt  1 2   to adjust steps 3  7 again     Adjust R314 for a symmetrical waveform around zero     18           FTENMA rue    405 Pioneer Blvd  Springboro  Ohio  45066  http   www tenma com    19    
6.  under test     e Do not exceed the voltage rating of the circuit being powered     12    Reference Operation    This section describes several advanced functions of the Tenma 72 7210 function  generator  The variety of swept available from the function generator make it especially  useful for such applications as test servo system or amplifier response  distortion and  stability  It can use for FM generator  frequency multiplication  or as a variable   beat frequency oscillator  The synchronous and sweep outputs can be used as a source  for TTL logic or to synchronize as external device  such as an oscilloscope     Main output    Connect the Main output connector to the input BNC connector on an oscilloscope   Switch off the external counter button    Select the waveform by pushing function selector    Push UP DOWN to select a range of frequency output    Turn the frequency knob to adjust the frequency output  The output value will be  indicated on the display   Turn the amplitude knob to adjust the output level  The attenuator switch can be  pushed to attenuate the output level for    20 dB if necessary    7  Push DC offset switch to the ON position  an additional DC level will be used to shift  output level  The knob of DC offset is used to adjust DC level     aon      D    Symmetry    Follow the operation of Main output  but with the symmetry button switched ON   The frequency will be divided ten from the original frequency output around   Turn the knob of symmetry to adj
7. FSET KNOB SWITCH  The switch is set at OFF position for normal operation   The knob sets the DC level and therefore the polarity of the OUTPUT signal after the  switch ON       SWEEP RATE  Rotate the Knob to adjust the sweep rate  Adjusts how often the  frequency sweep reiterates     7  SWEEP SWITCHES  The ON switch is used to enable sweep function  The STOP  switch is used to see the stop frequency     8  STOP FREQ  KNOB  This knob is used to adjust the sweep width  Adjusts the range of  frequency that is traversed by each sweep     9  EXT COUNTER INPUT  External signal input BNC  The input frequency is from 5Hz to  50MHz     10 EXT COUNTER SWITCH  Push the switch to ON for external counter     11 FREQUENCY KNOB  Rotate this knob to set the desired frequency output of  generator     12  SYNC OUT BNC  This connector sends out synchronous signal of TTL level  the output  frequency is same as MAIN output     13  SWEEP IN OUT BNC  Sweep signal output and input connector  It will be an output  connector when you push the SWEEP switch ON  otherwise it is an input connector   which is voltage control frequency  VCF      14 WAVE SELECTOR  Push the switch to select the output waveforms     15 AMPLITUDE KNOB  Adjust the voltage within the presently selected range  Rotate this  knob to adjust the voltage level of main output signal  OUTPUT BNC      16 ATTENUATOR  Push the switch to turn attenuator ON     17  OUTPUT BNC  Main output of function generator     10    Rear Panel    In addit
8. TENMA          ET    User s Manual                5MHz SWEEP FUNCTION GENERATOR SYMMETRY          CO MHz  CD KHz         Ca Hz  OFF POWER CO mHz                            OFFSET          ON    EXT COUNTER PE SWEEP  IN IN OUT                                  TIL  SY          SWEEP                  STOP    un   MAX  STOP FREQ                                 MAX  AMPLITUDE    OUTPUT     amp     500    20Vp p                                        5MHz SWEEP FUNCTION GENERATOR SYMMETRY       TO MHz    _      ABABAB Ef we  OFF POWER    Lome                ol   FREQUENCY  EXT COUNTER                         TTL lt 5V    OFFSET             STOP    A  MIN MAX  STOP FREQ       MIN MAX  AMPLITUDE    OUTPUT     amp     son    20Vp p                               http   www tenma com                                                             TENMA     72 7210 5 MHZ FUNCTION GENERATOR    INSTRUCTION MANUAL       REVISION HI STORY       2004 05 10  Ee          Copyright Statement    Copyright   Tenma  All rights reserved  No part of this publication may be reproduced in  any form or by any manner or used to make any derivative such as translation   transformation  or adaptation without our authorization and permission  All the Trademarks  belong to their respective proprietors        CONTENTS       Section 1 SAFETY E 1  Section 2 INTRODUCTION ME 2  S6  ction 3    SPECIFICATION E 3  General Specifications E 3   El etricalS le EE 4  E ak TE A  Waveform Characteristics ki aien si kk se is n
9. ction  The generator main output is non destructively protected against  short circuit to ground or to any voltage practically available in electronic laboratories                                   DC Offset  Load DC Offset Knob 1 Ripple 2  FCCW FCW  No Load  lt  10V  gt   10V  lt  10mVp p  50Q  lt   5V  gt   5V  Notes     1  Set the switch of DC OFFSET to ON position  and set the function switches of Sine   Triangle and Square to OFF position  Then test the DC offset     2  Before testing ripple  set the frequency range to lowest range and turn frequency knob    to MIN position  Otherwise it will induce noise around 50mVp p based on different    frequency        Other Inputs and Outputs    Synchronous Output                         Output Output level   Transition time   Fan out    2  Impedance  50 Q typical    1   TTL level   gt  3V  lt  30 ns Typical 20 TTL    open circuit loads  Notes     1  The output Impedance is measured at 1kQ load or above   2  The capability of fan out depends on different TTL logics and the level is greater than  2V                            Symmetry  Range Symmetry Knob  FCCW FCW  Up to 500kHz  lt 10  2 90   Notes     1  Symmetry from 10  to 90  at full output amplitude terminated into 50Q    External Sweep IN                      Impedance Sweep Range Sweep Ratio  1  14K Q typical  0 2Hz 100Hz 1 100  Notes     1  The frequency output will be proportional to the level of SWEEP IN connector  0   5V  input for up to 100 1 frequency change approxima
10. e this product in wet  damp or explosive atmosphere    Marking on the instruments        Caution  referring to manual for  detailed instructions        Protective conductor terminated     It B gt              Section 2 INTRODUCTION    This Function Generator  FG  produces low distortion combined with a 6 digit   high resolution 50MHz counter  It is a rugged  easy to operate  excellent heat dissipation  and high stability instrument  The FG range is from 0 01 Hz to 5 MHz with 8 ranges  You  can also directly control amplitude and DC offset    The sweep function makes the output signal traverse a range of frequency  The START  and STOP frequency control is setting for the sweep rate and sweep width internally  The  sweep function can be input from an external DC signal     This FG has a frequency counter to count the signal frequency from 5 Hz to 50 MHz   The frequency counter features one six digit display with automatic decimal point  placement and LED indicator that show the measuring units  It provides high input  impedance and good sensitivity for all specified frequency     The symmetry function provides the user to vary the duty of frequency  When you  push the switch of symmetry ON  the output waveform symmetry is adjustable  The  sweep IN function provides external voltage source to control the output frequency  The  output frequency can be swept above or below the selected frequency to a maximum of  100 1  depending on the amplitude of input and the selected output fr
11. equency     The exceptional accuracy  its portability and stackable design makes the Tenma  72 7210 Function Generator an ideal instrument for using by the electronic engineer  the  communication technician  schools and laboratories     Section 3 SPECIFICATION    General Specifications             Display Six digits of seven segment LED   Output Sine  Square  Triangle  DC  Sync and Sweep  Input Sweep IN  External frequency counter       Warm up time    At least two hours       Operating Temperature    Storage Temperature    0  C 40  C  32  F to 104  F    10  C to 60  C       Altitude    Up to 2000 M       Relative Humidity    Up to 80  for 0  C to 28  C  Up to 70  for 28  C to 40  C             Dimension 261 x 71x 211 mm  WxH xD   Weight 1 8Kgs approx   Line Voltage AC 100V 120V    10   50 60Hz  FUSE 0 4A 250V    AC 220V 240    10   50 60Hz  FUSE  0 2A 250V       Power consumption    40VA maximum       Ventilation    Always keep the back of instrument far away the wall for  better ventilation  and nothing at among        Standard Accessories    User Manual in CD ROM  Power cord             Safety Designed in compliance with EN61010 1  IEC1010 1   Installation Category II  Pollution Degree 2 Environment   EMC Designed in compliance with EN61326              Electrical Specification    The accuracy is given as      of span   No  of least significant counts  at 23  C   5  C with    relative humidity less than 80  R H     Main Output    Range  Sine Triangle Square Waves          
12. ion to line voltage selector  fuse and power cord receptacle                                                  Figure 2  Rear Panel      VENTILATION FAN  Always keep nothing on the front of fan for better ventilation   2  POWER CORD SOCKET    LINE VOLTAGE SELECTOR AND FUSE HOLDER  To select line voltage and fuse  replacement     11    Preparing for Use    PRECAUTIONS    In order to avoid damaging this instrument  make sure that the unit is set to the correct line  voltage for your area  Also make sure that the correct fuse is used for the line voltage   These line voltages are 100V  120V  220V and 240V at 50 60HZ     A  WARNING    To avoid damage the equipment use only specified fuse when change the power line  voltage  Please refer to following table                SELECTOR LINE VOLTAGE FUSE  100V 92 110V 50 60Hz 0 4A  120V 108 132V 50 60Hz 0 4A  220V 198 242V 50 60Hz 0 2A  240V 216 264V 50 60Hz 0 2A                     Avoid touching the rear panel of the instrument during operation  The heat sink    is nearby the rear panel  contacting it could result in skin burns     e Avoid using the function generator in ambient temperatures above specified   Always keep sufficient air space between the rear of instrument and wall to    prevent internal heat build up       Although the function generator is protected against reverse polarity damage  the circuit being powered may not include such protection  Always carefully    observe polarity incorrect polarity may damage the equipment
13. o 52 0Hz on the external counter    10  Turn the frequency knob for an internal counter reading to 50 0Hz  Adjust R637 for a    display of 50 0  0 1Hz     11  Turn the frequency knob to 10Hz reading on the counter  Adjust R638 for a display    reading of 10 0Hz  0 1Hz     12 Repeat steps 10 and 11 for best result on both ends   13  Select the Frequency Range from 500KHz 5MHzZ  Rotate the frequency knob to full    CW  Adjust C217 for a reading  gt 5 3MHz     14  Turn the frequency knob to full CCW  adjust R262 for a reading of 30 5 KHz     ENDOU      Output amplitude Adjustment    Terminate the output in 50 ohms and connect an oscilloscope and a multi meter to the    output       With all function pushbuttons out and the knob of AMPLITUDE control at full CW    position  adjust R401 for output of OV  10 mV       Set the amplitude control to full CW position and select triangle waveform  Check for a    peak amplitude of  gt 10 2V       Set frequency to 5KHz and adjust R225 for a symmetrical around zero waveform       Function adjustment of Sine wave     Connect a distortion analyzer through a 50 ohms termination to the OUTPUT and    monitor with an oscilloscope also       Set the FG controls as follows                    FUNCTION SINE  AMPLITUDE MAX  DC OFFSET OFF  FREQUENCY 50Hz  SYMMETRY OFF                  Set frequency range to 50 Hz and FREQ  Knob for 50 Hz output     Adjust R320 and R309 for minimum distortion and  lt 1 2      Set frequency range to 500kHz  and FREQ  Knob
14. pedance  1MQ typical   The counter display will be 0 Hz with 500 load on no signal input   The maximum input voltage   lt  42Vpk       Section 4 GETTING STARTED    Front Panel    Figure 1 shows the front panel controls  connectors and indicators of FG     1 2 3 4 5 6 7 8    Li             SYMMETRY OFFSET                SMHz SWEEP FUNCTION GENERATOR   SWEEP  LI       A   few VO YA  hh PENN N   ZATNNAN                              f       AA    ow 1 II Ce   JA j   Jij N a           H H H H H g  e RANGE WEY AREY 7 VEY r VOI d      OFF PO    R JE ne N   a        MIN MAX MIN AX l    LI  gt  ON ON   RATE STOP STOP FRE Ji  Ni FREQUENCY   da  HI so TR      LI   A NA   SYNC SWEEP Ss Nh  ie  ai   2008 OUTPUT              WI       WY   N A Vs      LI MIN   J     18 A MIN MAX          de NE V a AMP  ITUDE II  Ge S         Figure 1  Front Panel       POWER SWITCH  To toggle instrument power ON and OFF       Counter READOUT  A six digit display that shows frequency generator or external  counter  The LED indicates the frequency units of MHz  KHz  Hz or mHz for the  counter readout  The decimal point and unit is automatically placed in the appropriate  position       RANGE BUTTONS  Push buttons of UP DOWN to set the frequency range of the  output signal  The frequency range will be 10 times difference on each step       SYMMETRY KNOB SWITCH  The knob is used to vary either positive pulse ramp or  negative pulse ramp  Push the symmetry switch to    ON    before rotating this knob       DC OF
15. tely  To avoid damage to this  instrument  ensure that the maximum voltage into this BNC is no more than   10 Vp    2  Turn off the Sweep ON switch     Internal Sweep Sweep OUT                                  Impedance Sweep Rate Output Level  MIN MAX MIN MAX  600 Q  gt 5s  lt 10 ms  gt 0 3V   4 5  typical  Notes     1  The output level of SWEEP OUT BNC is proportional to the knob of STOP FREQ    2  The frequency knob is set to full CCW for start frequency  and Set the knob of stop  frequency to full CW  The STOP switch is used to indicate stop sweep frequency  The  sweep width is variable up to 100 1 frequency change approximately     6       Counter Characteristics    Internal Counter                            Range Display Resolution    1 Accuracy  7mHz   500 mHz 1mHz No  Specified  0 50Hz  5 00Hz 0 01Hz    3Yo 2   5 0Hz  50 0 Hz 0 1Hz   1  999 Hz 1 Hz    0 3   1   1 000 9 999 kHz 0 001 kHz    0 03   1   10 00 99 99 kHz 0 01kHz  100 0 999 9 kHz 0 1 kHz  1 000 5 000 MHz 0 001MHz                Notes     1  The display resolution of internal counter is according to the ranges of main output  setting     External Counter                               Range Resolution Accuracy  2 Input Sensitivity  5  5   999 Hz 1 Hz   Time base error 20mV RMS  1   999 999 kHz   0 001 kHz se OU  1  20 0000 MHz  0 0001 MHz     Time base error   1 count    gt 20   50 MHz   0 0001 MHz 60 mV RMS             Notes   1  Time base  10 MHz    Ol i    Time base error  10 PPM at  23   5   C   Input im
16. ust duty cycle from 10  to 90      Observe the results on the oscilloscope     eoa    SYNC  OUTPUT    1  Connect the SYNC output connector to the input BNC connector on an oscilloscope    2  The level and symmetry will be a constant value  but the frequency will follow the Main  output  Please refer to main output for frequency change    3  The knobs for DC offset and amplitude do not affect the output    4  Observe the results on the oscilloscope     Internal Sweep  Sweep OUT    1  Connect the Main output connector to the CH1 BNC connector on the oscilloscope   2  Connect the SWEEP in out BNC to the CH2 BNC connector on the oscilloscope    3  Turn the frequency knob to set desired sweep start frequency    4  Turn the stop frequency knob to set stop frequency  Push the STOP switch to the ON  position to see the stop sweep frequency on the display  Turn off the STOP switch once  you finished the setting    Push the sweep switch to the ON position    The main output will sweep the desired range of frequencies  and the BNC sweep is  output a linear level    7  Turn the sweep rate knob to set how often the frequency sweep reiterates    8  Observe the results on the oscilloscope     oo    External Sweep IN    13    Aa aN  gt     D    Push the sweep switch to the OFF position    Connect the Main output connector to the CH1 BNC connector on the oscilloscope   Connect the SWEEP in out BNC to another function generator    Turn the frequency knob to set desired sweep start frequency   
17. w  Please make  sure that the protection fuse rating matches     2  When replacing fuse or selecting the input voltage  please use fuse ratings as  specified        For Service and repair and calibration please contact your local Tenma  distributor or go to http  Awww tenma com    15    Section 6 ADJUSTMENT PROCEDURES     CAUTION    The Following instructions are for qualified service personnel only  To avoid electrical  shock  do not perform any servicing other than contained in the operation instructions  unless you are qualified to do so     Before calibrating and adjusting  make sure that the instrument has warmed up for at  least 2 hours     Required test equipment    The following Table lists the test equipment necessary to perform the adjustments and the  maintenance of this instrument  Any equipment that satisfies the critical specifications  given in the table may be substituted for the recommended model     Instrument Type Required Specifications       Oscilloscope 100 MHz bandwidth  dual channel       DVM 4 5 digits  true RMS  Counter 100 MHz  2 CH  time interval capability  Distortion Analyzer 10 Hz     100 KHz    0 1     Function Generator 10 MHz frequency  Feed through termination 50 ohm    Cable Assembly BNC BNC  50 ohms  RG 58  BNC Adapters BNC to banana  BNC tees       16    Start to Adjust    A  Frequency Counter  10 MHz Oscillator Adjustments    1             2  Adjust R225 for a symmetrical around zero waveform     C  Function adjustment of Square wave  1 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Manual de Usuario Motorola SBG900 - Multitel  O texto para as questões de 01 a 05 encontra    Rabbit ® 6000 Microprocessor User`s Manual  Troubleshooting for 10-20kUE Three-Phase Inverter    Targus Blacktop Messenger Computer Case  Samsung Samsung E2152 Užívateľská príručka    Copyright © All rights reserved. 
   Failed to retrieve file