User manual - PONSEL Instrumentation for water quality

image

Contents

1. 2 2 Connecting the ACTEON 2020 transmitter and pH and Temperature sensors e Case the distance between the sensors and the transmitter is less than 10 meters Power supply cable 230V 110 V or 24VDC Line out cable Relay cable 2 channels 4 20mA cable 2 channels Redox ORP sensor PONCPC EHAN 10 Temperature sensor PONCPC T 10 Figure 2 Installing a measuring system PONSEL Ref NOTICE_ACTEON_2020_v003 V 003 10 58 User manual 2O00 e Case 2 the distance between the transmitter and the sensors is longer than 10 metres A watertight IP 68 connection box with integrated pre amplifier must be used REF PON BJAI E PH EH C Power supply cable 230 110V or 24VDC Line out cable Relay cable 2 channels 4 20mA cable 2 channels Redox ORP sensor PONCPC EHAN 10 Temperature sensor PONCPC T 10 Connection box PON BJAI E PH EH T Figure 3 Installation with a connection box with integrated pre amplifier PONSEL Ref NOTICE_ACTEON_2020_v003 V 003 11 58 PX User manual 20000 2 2 1 ACTEON 2020 wiring See appendix 15 at the end of the document 2 3 Tank mounting 2 3 1 Using the stand and protective hood A protective hood PON PDPVC 1 is available for mounting the ACTEON 2020 The hood is essential in the case of direct exposure to adverse weather or sunshine 250 276 Y 74 Figure 4 Mount and hood
2. R2 PONSEL 46 58 gs PONSEL User manual 2000 9 4 2 Configuring relays in mode 2 CONFIG MODE 2 ADJUSTMENT OF THE THRESHOLDS AND HYSTERESI Rel 1 connected to redox measurement Alarm threshold 150mV Hyst 10mv ESTEE ceshold R1 is on Rel 2 connected to redox measurement Alarm threshold 50mV Hyst 10mv Above the threshold R2 is on Use the Aand V keys to Enter Validate the choice increase the values Use the X and gt keys to navigate the selections Esc Cancel the procedure eo Configure the two relays and press ENTER WHAT DO YOU WANT TO DO 4 20mA outputs adjustment Relay adjustment Measure units Language Factory default value reminder In the above example the relays are configured as indicated below Alarm threshold Hysteresis R2 Alarm i threshold Hysteresis R1 Inactive Active oo cadi I Active R1 j i Inactive i z i Inactive Acti i i Active js 259 ek D Ee i 5 1 PONSEL 47 58 gs PONSEL User manual 2000 9 5 Adjusting measurement units ACTEON 2020 can be configured in C and F for temperature measurement CONFIG WHAT DO YOU WANT TO DO PAveraging sensor response gt Graphic screen gt 4 20mA outputs adjustment gt Relay adjustment gt Measure units gt Language gt Factory default value reminder Use the Aand V keys
3. User manual BOOO ACTEON 2020 Redox Temperature User manual ACTEON 2828 Redox 57 mv CAL THEOR 30 090 PONSEL Ref NOTICE ACTEON 2020 v003 V 003 1 58 quio User manual 2 000 IMPORTANT Read the manual carefully before switching the device on In order to maintain and ensure the good working order of the device users must comply with the safety precautions and warnings featured in this manual Assembly and activation Assembly electrical connection activation operation and maintenance of the measuring system must only be carried out by trained personnel duly authorised by the end user Trained personnel must be familiar with and comply with the instructions in this activation manual Make sure the power supply complies with the specifications on the nameplate before connecting the device A clearly labelled power switch must be positioned near the device Check all connections before turning the power on Do not use damaged equipment it may represent a hazard and should be labelled as faulty Repairs must only be carried out by the manufacturer or by a Ponsel after sales service PONSEL Ref NOTI CTEON 2020 v003 03 2 58 gs PONSEL User manual 2000 CONTENTS 1 THE MEASURING SYSTEM iinicassasa cH susa na EKERESUE CHAKFR ELM aSE ERR CDD RUM IE Sa E aUC ELI RR EE 6 1 1 PRE DASIC
4. 6 1 2 Redox sensor slope adjustment You can only adjust the slope of your sensor by carrying out the following procedure CALIBRATION SENSOR CALIBRATION ORP SENSOR gt Gain adjust Complete calibration gt Theoretical calibration TEMPERATURE SENSOR gt Gain adjust gt Complete calibration gt Theoretical calibration Use the Aand V keys to select ENTER Validate Che choice the type of calibration and the probe to be calibrated Select the menu and press ENTER Immerse the Redox sensor in the 240mV or 470mV ORP standard solution ORP SENSOR CALIBRATIO N Immerse the sensor in an ORP standard solution Measure 234 mV WAIT Nonstable measurement Use the Aand V keys to ENTER Validate the new slope adjust the buffer value OmV to 999mV ESC Leave the current slope x AWAIT MEASUREMENT STABILISATION Press the ENTER key to confirm gain adjust PONSEL Ref NOTICE_ACTEON_2020_v003 V 003 24 58 PUT User manual BDOOO Calibration Correct No If the second calibration point is incorrect an error message window appears see Error message information chapter 6 2 2 Yes ACTEON 2020 ORP 290 mv Comment The measurement window is 20 00 c displayed again once gain a adjust has been correctly Model R1 qM carried out 200 PONSEL 25 58 gs PONSEL User manual 2000 6 1 3 Returning to
5. 20 00 c mV Model RI R2 N ee MEAS car Loc Peony Use the D and keys to navigate the ACTEON menu 200 E Select the menu and press ENTER CALIBRATION WHAT DO YOU WANT TO DO PAveraging sensor response Graphic screen 4 20mA outputs adjustment Relay adjustment Measure units Language Factory default value reminder ESC Escape menu PONSEL 37 58 gs PONSEL User manual 2000 9 1 Configuring sensor response averaging This configuration makes the sensor response more stable and responsive By default the averaging procedure involves 10 measurements This means that the displayed value is the average measurement over the last 10 seconds CONFIG WHAT DO YOU WANT TO DO PAveraging sensor response Graphic screen 4 20mA outputs adjustment AVERAGING SENSOR RESPONSE CONFIG Relay adjustment Measure units Language Factory default value reminder Use the Aand V keys to select the configuration required ENTER Average on EJ measures Instantaneous answer Answer very strongly averaged Esc Cancel the procedure Enter Validate the averaging Use the Aand V keys to increase and decrease the averaging qe Select the averaging desired and press ENTER CONFIG WHAT DO YOU WANT TO DO PAveraging sensor response Comment Graphic screen 4 20mA outputs adjustment Relay adjustment Measure units La
6. 2000 9 7 Resetting factory default values ACTEON 2020 can be reinitialised with the default parameters CONFIG WHAT DO YOU WANT TO DO PAveraging sensor response Graphic screen 4 20mA outputs adjustment Relay adjustment Measure units Language Factory default value Use the Aand V keys to select the configuration required Select the menu and press ENTER DEFAULT VALUE REMINDER CONFIG ATTENTION you will loose all your personal adjustments gt No Esc Cancel the procedure Enter Validate the choice Use the Aand V keys to select the procedure confirmation oe Select and press ENTER CONFIG WHAT DO YOU WANT TO DO gt Graphic screen gt 4 20mA outputs adjustment gt Relay adjustment gt Measure units gt Language gt Factory default value reminder PONSEL 3 3 50 58 Ze PONSEL User manual BOOO After having reset the factory default parameters ACTEON 2020 will be configured with the following parameters Sensor response averaging 10 measurements Time base for trend line display 1 minute 4 20mA output adjustment Redox output The 4mA stop is set to 500mV and the 20mA stop is set to 500mV Temp C output The 4mA stop is set to 5 C and the 20mA stop is set to 35 C Relay output adjustment The relays are configured in mode 1 RI threshold OmV hysteresis 5mV abo
7. 68 connection box with integrated pre amplifier supplied with 10 metres of cable for pH Eh and temperature sensors connections at distances over 10 metres see Figure 3 Installation with a connection box with integrated pre amplifier PONCBMC 9 9 conductor coated cable for the Connection box Acteon connection PON ACT 24V Optional 24VDC power supply PON PDPCV 1 PVC stand and protective hood for one ACTEON transmitter PON PDPCV 2 PVC stand and protective hood for two ACTEON transmitters PON CASQ 1 PVC protective hood for one ACTEON transmitter PON CASQ 2 PVC protective hood for two ACTEON transmitters PONPPCC PH EH T Elbowed sensor holder perch ESHP for pH Eh or Temperature sensors Supplied with nozzle and glued connector PONPPCD PH EH T Straight sensor holder perch for pH Eh and Temperature sensors supplied with nozzle and glued connector SPECIFIC APPLICATION PONBUSE PH EH T Nozzle for pH Eh without cleaning system and Temperature sensors Complete with 50 diameter BP3P glued connector PONCOUDE 90 elbow for sensor holder perch closure PONSPFR2C Stainless steel ESHP or SSHP type mount for 2 perches 1 arm 2 sliders PONSPFRIC Stainless steel ESHP or SSHP type mount for 1 perch 1 arm 1 slider PONSPFR COUL Additional slider for QRPM Quick release perch mount systems 1 2 4 Accessories for Redox filling sensors PONCPC PH RV 10 PONRAC EH RV Connector f
8. Configure the 4 20mA output for temperature and press ENTER CONFIG WHAT DO YOU WANT TO DO Comment Both 4 20mA Graphic screen 4 20mA outputs adjustment outputs have been ini fei sU Ed configured correctly Measure units Language Factory default value reminder PONSEL 42 58 PUT User manual 20000 9 3 2 Calibrating 4 20mA outputs Calibrating the 4 20mA outputs improves their level of accuracy by taking into account the device load CONFIG 4 20mA OUTPUTS gt 4 20mA output adjustment gt 4 20mA output calibration Use the Aand V keys to select the configuration required press ENTER Select the LEE EIU Y ESS SEERE SEERE menu and Procedure CONFIG Insert an ammeter on 4 20mA outputs ACTEON 2000 AMMETER ENTER To continue ESC Escape menu qe Connect an ammeter to the 4 20mA outputs and press ENTER 4 20mA OUTPUTS CONFIG CURRENT GENERATION OF 4mA Enter the value Enter the value read read on the ORP on the temperature 4 20mA output 4 20mA output mA ESC Escape menu ENTER To continue Use the Aand V keys to increase and decrease the4mA value Use the X and gt keys to select the ORP 4 20mA output or the temperature 4 20mA output PONSEL 43 58 gs PONSEL User manual 2000 oe Adjust the 4mA outputs and press ENTER 4 20mA OUTPUTS CONFIG CURRENT GENERATION OF 20mA Enter the value Enter the value read read on the
9. ORP on the temperature 4 20mA output 4 20mA output PARMA Use the Aand V keys to increase and decrease the 20mA value Use the X and gt keys to select the ORP 4 20mA output or the temperature 4 20mA output s Adjust the 20mA outputs and press ENTER CONFIG WHAT DO YOU WANT TO DO PAveraging sensor response Graphic screen 4 20mA outputs adjustment Relay adjustment Measure units Language Factory default value reminder PONSEL 3 3 44 58 gs PONSEL User manual 2000 9 4 Adjusting relay outputs The 2 relays can be configured in 2 different modes Mode 1 is used to configure relay R1 on a Redox measurement value and relay R2 on a temperature measurement value Mode 2 is used to configure relays R1 and R2 on two Redox measurement values The 2 relays can be configured with the following attributes Alarm threshold Threshold trigger value Hysteresis Hysteresis value for relay switching prevents relay hunting Triggering direction Above Means the relay contact is closed above the alarm threshold Below Means the relay contact is open below the alarm threshold CONFIG WHAT DO YOU WANT TO DO PAveraging sensor response gt Graphic screen gt 4 20mA outputs adjustment gt Measure units gt Language gt Factory default value reminder Use the Aand V keys to select Esc Escape menu the configuration required ee S
10. for the transmitter PONSEL Ref NOTICE ACTEON 2020 v003 V 003 12 58 User manual BOOO 2 3 2 Installing the sensor with a sensor holder perch elbowed or straight and nozzle PONPPCC PH EH or PONPPCD PH EH It is best to use the elbowed sensor holder perch when immersing the sensor in a tank 2 20m Sensor holder perch amp Glued connector 9 Sensor holder nozzle W Lock nut Liquid level Flow direction of liquid 50cm minimum wN Sensor in the nozzle Figure 5 PVC Elbowed Sensor Holder Perch Use elbowed sensor holder perches in heavily soiled tanks to prevent fibre build up on the perch Installing the pH sensor in the nozzle 1 Dismantle the sensor holder perch then remove the nozzle from the bottom of the perch by removing its tightening screw 2 Introduce a solid wire into the perch then a string 3 Remove the sensor s protective casing and two O ring seals PONSEL Ref NOTICE_ACTEON_2020_v003 V 003 13 58 go OS User manual DOOO 4 Attach the string to the end of the sensor and introduce it into the top of the perch Use the string to pull the sensor to the bottom of the perch 5 Remove the string and place the sensor in the nozzle The sensor should protrude about twenty centimetres 6 Replace only the O ring seal nearest the glass section of the sensor 7 Adjust the tip of the sensor exactly flush with the end of the
11. u sssssessereereere renere renerne nere ener ene 15 Figure 8 Installing a temperature sensor inside a Pipe eee eeeceeeceeeceeceseceaeceeeeeeeeeeeaeeeaeeeeeeeeeeeeeseseaeenaees 16 Figure 9 Installing a temperature sensor in a pipe u u sssserserersere renere ener nennen nennen nennen eren en rennen ens 16 Figure 10 Redox sensor diagram eese ner e rin co ined aua Slo ea epa ce eee by epe eue ERE EEE RELEE EET SERER 53 Figure 11 Resetting ACTEON 2000 eese onenen sven EE NEEE E EEE EEEE VE E EEr ONET SS 56 Figure 12 Transmitter wiring terminal layout sess nennen eren eren rennen ens 57 PONSEL Ref NOTICE ACTEON 2020 v003 V 003 5 58 Ge ONSE User manual 2000 1 The measuring system 1 1 The basic system A measuring system requires the following basic elements 1 1 1 A Redox Temperature transmitter PONACTEON2020 EH T ACTEON 2020 Redox meter and thermometer transmitter The Redox Temperature transmitter comes with a 240mV buffer solution and a user manual 1 1 2 A Redox sensor Three sensors are available for different applications PONCPC EH 10 Plastogel electrolyte Eh sensor platinum tip with 10m cable connection perch mounting or in pipe installation PONCPC EHAN 10 Plastogel electrolyte Eh sensor platinum ring with 10m cable connection perch mounting or in pipe installation PONCPC EH RV 10 Glass bod
12. EEa KEDE KE veto S E E Ra 54 EMG noD r E E E E A EEE ESEE E RE 54 13 2 Temperature sensor ee 54 TS 251 Maintenance 35 soe rs rere eoim i KEE EF ER EEG HEE ee t REDER ENGE UD FUR Q eder tuer ens que eae UE eu SR E SRM FE dE 54 13 22 Mechanical diagram a ee IECUR C UI REDE URBE TERCER 55 14 QUESTIONS 8 ANSWENRS 2iii tr atacian nin ka cuo rEMFIIMI E MERE CREME EA ME FEM EM REM EP E EMEP LANA 56 14 1 Display screen troubleshooting eese esee ee eere enne setenta sins tn sts tu stesso ses Ness sesta sets suse ss Tages 56 15 APPENDIR eq 57 PONSEL 4 58 gs PONSEL User manual 2000 Figures Figure 1 Transmitter mounting diagram oo cesses eesecesecesecsecseecaeecsecseecaeeeeeeaecesecsecsaecsaecaaecaeeeaeseaeeensents 9 Figure 2 Installing a measuring system ssseeeeeeeeeeneeneen eene eene eene nerne tene trennen trennen renes 10 Figure 3 Installation with a connection box with integrated pre amplifier sse 11 Figure 4 Mount and hood for the transmitter enne ener nre nest nennen 12 Figure 5 PVC Elbowed Sensor Holder Perch essere eene trennen nennen 13 Figure 6 Cross section diagram of the sensor nozzle without cleaning system seseeeee 14 Figure 7 Installing a perch on a QRPM with one or two sliders
13. ON ORP SENSOR gt Gain adjust gt Complete calibration gt Theoretical calibration TEMPERATURE SENSOR gt Gain adjust gt Complete calibration gt Theoretical calibration Use the Aand V keys to select the type of calibration and the probe to be calibrated ENTER Validate the choice Select the 2 Complete calibratior menu of the temperature sensor and press ENTER CALIBRATION TEMPERATURE SENSOR Enter your name iurant ESC Cancel the procedure ENTER Start calibration Use the Dand lt q keys to move the cursor in the name section Use the Aand V keys to change the letters The scrolling order of the letters is A Z 0 9 gt space qe Enter your name and reference and press ENTER Immerse the sensor in a water and crushed ice mixture at 0 00 C PONSEL 27 58 Ze PONSEL User manual 2000 TEMPERATURE SENSOR CALIBRATIO N Immerse the sensor in water and ice Measure 0 15 C WAIT Nonstable measurement ESC Cancel the procedure ENTER Validate the zero Calibration Correct Yes Use the Aand V keys to adjust the water temperature value x AWAIT MEASUREMENT STABILISATION Press the ENTER key to confirm the first calibration point No If the first calibration point is incorrect an error message window appears see Error message information chapter 6 3 1 Immerse the sensor in water at ambient temperatu
14. R CALIBRATION IMPOSSIBLE Offset NON CORRECT Esc Escape menu PONSEL 33 58 gs PONSEL User manual 2000 If an error occurs during calibration in water at 0 C 1 Check the water temperature with a precision thermometer 2 Check the connection between the temperature sensor and ACTEON 2020 6 3 2 Ambient water calibration error CALIBRATION TEMPERATURE SENSOR CALIBRATION IMPOSSIBLE Slope NON CORRECT Esc Escape menu If an error occurs during calibration in water at ambient temperature 1 Check the water temperature with a precision thermometer 2 Check the connection between the temperature sensor and ACTEON 2020 PONSEL 34 58 PUT User manual 2000 7 VIEWING MEASUREMENT HISTORY The history menu can be used to consult the last 100 data items recorded and displayed on the trend line To view this data follow the instructions below 20 00 c Model RIX R2 N Use the gt and lt keys to navigate the ACTEON menu Select the menu and press ENTER HISTORY 20 00 c 4 hrs 24 hrs Use the gt andd keys to move the ESC Escape menu cursor along the trend line PONSEL Ref NOTK CTEON_2020_v003 03 35 58 Ze PONSEL User manual 2000 8 VIEWING THE SENSOR CALIBRATION REPORT The calibration logger is used to check the reports of the most recent calibrations Follow the ins
15. REDOX measurement theoretical calibration Theoretical calibration is carried out using the theoretical coefficients Follow the instructions below to carry out theoretical calibration CALIBRATION SENSORS CALIBRATION ORP SENSOR gt Gain adjust gt Complete calibration gt Theoretical calibration TEMPERATURE SENSOR gt Gain adjust gt Complete calibration gt Theoretical calibration Use the Aand V keys to select ENTER i date the choice the type of calibration and the probe to be calibrated press ENTER CALIBRATION THEORETICAL CALIBRATION ATTENTION you will erase your calibration coefficients to return to the theoretical coefficients gt Yes gt No Esc Cancel the procedure Enter Validate the choice Use the Aand V keys to select the procedure confirmation Select and press ENTER to return to the theoretical coefficients ACTEON 2020 ORP Indicates that the device has MARUR 2 9 0 mV ae been calibrated with the amp theoretical coefficients Trigo CAL 300 20 00 c mv Model RI R2 N a rsr i can LOG E CONF INFO 200 PONSEL 3 3 26 58 gs PONSEL User manual 2000 6 1 4 Two point temperature sensor calibration complete calibration Use the following procedure to completely calibrate your temperature sensor You will need a precision thermometer and water at 0 C for this procedure CALIBRATION SENSORS CALIBRATI
16. SEL User manual 2000 6 Calibrating the ACTEON 2020 6 1 Calibrating the sensors Select the calibration menu in the measurement window 20 00 c mV Model R1 200 R2 x Use the gt and lt keys to BES ee navigate the ACTEON menu oe Select the menu then press ENTER CALIBRATION SENSORS CALIBRATION ORP SENSOR gt Gain adjust gt Theoretical calibration TEMPERATURE SENSOR gt Gain adjust gt Complete calibration gt Theoretical calibration ENTER Validate the choice PONSEL 20 58 gs PONSEL User manual 2000 6 1 1 Two point Redox ORP sensor calibration complete calibration Use the following procedure to calibrate your Redox sensor You will need the 240mV or 470mV Redox buffer solution for this procedure CALIBRATION SENSORS CALIBRATION ORP SENSOR gt Gain adjust gt Complete calibration gt Theoretical calibration TEMPERATURE SENSOR gt Gain adjust gt Complete calibration gt Theoretical calibration Use the A and V keys to ENTER Validate the choice select the type of calibration and the probe to be calibrated Select the EENIGE menu and press ENTER CALIBRATION ORP SENSOR Enter your name iurant ESC Cancel the procedure ENTER Start calibration Use the gt and lt keys to move the cursor in the name section Use the Aand V keys to change the letters The scrolling
17. SYSTE sssecscccsanicerssiccessssosasesassessonesusecesactcctesaussusssasseuteeasesuessacdenasessssssesassecsousSonsdecscuasseascvsacbeasves 6 1 1 1 A Redox Temperature transmitter sessssessssseseeseeer ene nne ener eene trennen nenne ener enne 6 1 1 2 A Redox SENSOR ieissar menda redet tU iu sseagnteenedesicesesveeean nat sdnccetsedesnedscteenedseeesseetocess 6 1 1 3 A temperature Sensor iip step rU E oe E Hee 6 1 2 YO a SENSE SDS ES SEES 7 1 2 1 Seuil 7 1 2 2 sensor connector adapter 2 5 a Ope eroi RERO PEOR Gee eee E 7 1 2 3 Accessories for a tank mounted installation without cleaning system eene 7 1 2 4 Accessories for Redox filling sensors PONCPC PH RV 10 sess 7 1 2 5 Accessories for Redox PONCPC EH 10 and Temperature PONCPC T 10 sensor in pipe SLAM ATION E M SES 8 2 INSTALLATION Liouscicickcxa d pdcidadida aia eC uP PRIMERA CU EMPVRU OH Vend SAP ERE CERIS URDU bd 9 2 1 Mounting the ACTEON 2020 transmitter box eese esee e essen eese testen testen sns sneen vene ennen renees 9 2 2 Connecting the ACTEON 2020 transmitter and pH and Temperature sensors 10 22 1 ACTEON 2020 wiring pi ne HOHER P ei dies eid i iam 12 2 3 Tank Mounting 12 2 3 1 U
18. ater at ambient temperature TEMPERATURE SENSOR Immerse the sensor in water at ambient temperature Standard c Measure 20 20 c WAIT Nonstable measurement ESC Leave the current slope Use the Aand V keys to adjust the ENTER Validate the new slope ambient water temperature value x AWAIT MEASUREMENT STABILISATION Press the ENTER key to confirm slope adjustment PONSEL Ref NOTK CTEON_2020_v003 03 30 58 PUT User manual BDOOO Calibration Correct No If the second calibration point is incorrect an error message window appears see Error message information chapter 6 3 2 Yes ACTEON 2020 ORP MEASURE 2 9 0 my Comment The measurement window is 20 00 c displayed again once gain a adjust has been correctly Model R1 qM carried out 200 PONSEL 31 58 gs PONSEL User manual DOOO 6 1 6 Returning to temperature measurement theoretical calibration Theoretical calibration is carried out using the theoretical coefficients PT100 theoretical slope and offset zeroing Follow the instructions below to carry out theoretical calibration GALTERATTO SENSORS CALIBRATION N ORP SENSOR gt Gain adjust gt Complete calibration gt Theoretical calibration TEMPERATURE SENSOR Gain adjust Complete calibration Use the Aand V keys to select ENTER Validate the choice the type of calibration and the
19. e sensor is connected the shunt between terminal points 28 and 29 must be removed Figure 12 Transmitter wiring terminal layout Before shutting the ACTEON cover check that the cover sealing joint is correctly positioned in its groove PONSEL Ref NOTICE_ACTEON_2020_v003 V 003 57 58 PX User manual 2000 After sales service PONSEL MESURE AFTER SALES SERVICE 35 rue Michel MARION 56850 CAUDAN FRANCE sav O ponsel fr TEL 33 0 2 97 89 25 30 FAX 33 0 2 97 76 55 72 Serial number PONSEL 58 58
20. elect the menu and press ENTER CONFIG Configuration of the relays mode Mode 1 Alarms adjustment Mode 2 Regulation 2 relays Esc Escape menu Enter Validate the choice Use the Aand V keys to select the configuration required PONSEL Ref NOTI CTEON 2020 v003 03 45 58 gs PONSEL User manual 2000 9 4 1 Configuring relays in mode 1 ore ie MODE 1 ADJUSTMENT OF THE THRESHOLDS AND HYSTERESIS Rel 1 connected to redox measurement Alarm threshold 50mV Hyst 10mv Below the threshold R1 is on Rel 2 connected to temperature measurement Alarm threshold 20 01 C Hyst 0 06 C Above the threshold R2 is on Use the Aand V keys to Enter Validate the choice increase the values Use the lt and keys to navigate the selections Esc Cancel the procedure eo Configure the two relays and press ENTER CONFIG WHAT DO YOU WANT TO DO gt Graphic screen gt 4 20mA outputs adjustment gt Relay adjustment gt Measure units gt Language gt Factory default value reminder In the above example the relays are configured as indicated below Redox Alarm 50 TN threshold 40 ee oc Hysteresis Inactive Inactive Active i Active R1 NL DRM se Nes Temperature Alarm 20 01 i threshold 19 956 semen M osos Lie Hysteresis Inactive A ee Inactive EM e Active Active
21. er 4 20mA output precision it is advised to calibrate the outputs during activation CONFIG WHAT DO YOU WANT TO DO PAveraging sensor response gt Graphic screen gt 4 20mA outputs adjustment gt Relay adjustment gt Measure units gt Language gt Factory default value reminder Use the Aand V keys to select the configuration required Select the menu and press ENTER CONF 4 20mA OUTPUTS gt 4 20mA output adjustment gt 4 20mA output calibration PONSEL 40 58 gs PONSEL User manual DOOO 9 3 1 Adjusting 4 20mA output stop thresholds CONFIG 4 20mA OUTPUTS gt 4 20mA output adjustment gt 4 20mA output calibration Use the Aand V keys to select the configuration required Select the EPI T MIT MET REIS ERIS menu and press ENTER 4 20mA OUTPUTS CONFIG NUMBER 1 FOR ORP 4mA adjustment 20mA adjustment ERT mv 12 00mv E um choice Use the Aand V keys to increase and decrease the 4 and 20mA stop values Use the lt and gt keys to select 4mA or 20mA d Configure the ORP 4 20mA output and press ENTER 4 20mA OUTPUTS CONFIG No 2 FOR TEMPERATURE AmA adjustment 20mA adjustment Use the Aand V keys to increase and decrease the 4 and 20mA stop Esc Escape menu Enter Validate the choice values Use the X and gt keys to select 4mA or 20mA PONSEL R OTK CTEO 003 41 58 gs PONSEL User manual 2000
22. ied Eh filling sensor with 10 metre cable connection For cold liquids with very low conductivity less than 50uS cm and contaminating media it is recommended to use a filling sensor 1 1 3 A temperature sensor Two sensors are available for different types of installation PONCPC T 10 Temperature sensor with 10m cable connection perch mounting or in pipe installation PONSEL 6 58 Ze PONSEL User manual 2000 1 2 Accessories 1 2 1 Consumables PONMANU 2020 EH T Additional ACTEON 2020 user manual PONKCL 1M KCI 1M for Eh filling sensors PONKCL 2M KCI 2M for Eh filling sensors PONKCL 3M KCI 3M for Eh filling sensors PONPEPNET 125ml bottle of pepsin cleaning solution for Eh and pH sensors PONSOLCSV 125ml bottle of storage solution for pH Eh sensors PONEHSOL240 Buffer solution for Redox meters 240mV at 25 C 125ml bottle PONEHSOL470 Buffer solution for Redox meters 470mV at 25 C 125ml bottle PONFILP EHAN ACTEON 2020 earth wire 1 2 2 Sensor connector adapter These adapters are used to connect ACTEON 2020 units to PONSEL sensors with connectors PONADAP EHAN ACTEON 2020 adapter for PONCPC EHAN S10 and PONCPC EH S10 sensors PONADAP T Adapter for PONCPC T S10 sensor 1 2 3 Accessories for a tank mounted installation without cleaning system PON BJAI E PH EH T Watertight IP
23. n 6 1 6 Adjusting the graphic screen 9 2 CONFIGURATION 4 20mA output MEASURE adjustment Config of 4mA and 20mA stop switches 89 3 1 4 20mA output calibration Threshold adjustment in mode 1 0 Threshold adjustment in mode 2 89 4 2 con 59 2 N Measure unit configuration 9 5 Language configuration 9 6 GENERAL Factory default configuration reminder 9 7 INFORMATION PONSEL 18 58 gs PONSEL User manual 2000 5 The measurement window In measure mode the measurement screen displays various information ACTEON 2020 ORP 4 W M 1 R1 100 ode NEA zb 0 R2 N ursr can LOG J CONF INFo o Redox measurement Measurement state indicator THEOR CAL Use of the theoretical coefficients as calibration coefficients N CAL Incorrect calibration point on the probe No icon displayed The last calibration is correct o Temperature transmitted by the temperature sensor In the event of faulty wiring or uninstalled sensor the temperature will not be displayed Comment The following icon is displayed if the temperature is higher than the operating temperature 55 C HOT 0 R1 and R2 relays state and operating mode D Contact is inactive o ACTEON menu Trend line can be configured from 1 min to 24 hours with automatic scaling Contact is active PONSEL 19 58 gs PON
24. nguage Factorv default value reminder Increase the averaging if your measurement is unstable PONSEL 38 58 Ze PONSEL User manual 2000 Decrease the averaging if the measurement process is too slow 9 2 Configuring the trend line The trend line can be used to check regulation cycles and detect anomalies Follow the instructions below to adapt the time base to the trend line CONFIG WHAT DO YOU WANT TO DO PAveraging sensor response gt Graphic screen gt 4 20mA outputs adjustment gt Relay adjustment gt Measure units gt Language gt Factory default value reminder Use the Aand V keys to select the configuration required Select the DRE menu and press ENTER CONFIG GRAPHIC SCREEN Confiquration of the trend line Time base inutes Showing maximum range 11 hours Esc Cancel the procedure Enter Validate the choice oe Use the Aand V keys to increase the time base in order to adjust the duration of the trend curve 1 sec to 15 min Select the time base chosen and press ENTER CONFIG WHAT DO YOU WANT TO DO PAveraging sensor response PONSEL Graphic screen 4 20mA outputs adjustment Relay adjustment Measure units Language Factory default value reminder 39 58 gs PONSEL User manual 2000 9 3 Configuring the two 4 20mA outputs Comment For great
25. nozzle Secure the sensor in place with the tightening screw 8 Replace the nozzle on the perch and secure it with the tightening screw 9 Use the stand with a slider or another sort of system to set the perch in a vertical position Adjust the immersion parameters so that the nozzle always remains submerged and the submerged section is always at least 0 50m below the surface 10 To prevent fibre build up position the sensor so that the nozzle is facing in the same direction as the flow Diagram of the sensor in the nozzle without cleaning system Water earth cable Nozzle PONBUSE PH EH T EH SENSOR PONCPC EHAN 10 Figure 6 Cross section diagram of the sensor nozzle without cleaning system Comment The line out cable is to be connected to the ACTEON 2000 earth terminal PONSEL Ref NOTICE ACTEON 2020 v003 V 003 14 58 User manual ZBOOO 2 3 3 Installing an Elbowed Sensor Holder Perch ref PONPPCC CIR or Straight Sensor Holder Perch ref PONPPCD CIR on QRPM ref PONSPFR and PONSPFR2 1 Fix the stainless steel QRPM to the infrastructure 2 Next fix a sensor holder perch to the stainless steel QRPM as shown in the diagram below Figure 7 Installing a perch on a QRPM with one or two sliders A second slider may be added to install a second sensor holder perch for the temperature sensor see diagram above PONSEL Ref NOTICE ACTEON 2020 v003 V 003 15 58 PUT User ma
26. nual 2000 2 4 In pipe installation For in pipe installations use the tee assembly PON 4020 and the adapter sleeve PONAP1 Redox sensor PONCPC EH 10 or PONCPC EHAN 10 Sleeve PONAP1 Tee PON4020 een dk 20 27 20x 27 Figure 8 Installing a temperature sensor inside a pipe Temperature sensor PONCPC T 10 Sleeve PONAP1 ue Tee PON4020 T 10 2 20x27 connectors Figure 9 Installing a temperature sensor in a pipe PONSEL R NOTICI ACTEON 02 003 003 16 58 User manual BZBOOO 3 ACTEON 2020 transmitter 3 1 Control console ACTEON 2828 Redox MESURE S73 268 CAL THEOR 30 09 C 200 Modet Be imin e ENTER key for accessing menus or confirming actions ESC key for exiting menus or cancelling actions lt for moving left in menus gt for moving right in menus A for increasing a value or selecting the menu above V for decreasing a value or selecting the menu below Control screen Quarter turn screw to seal cover PONSEL Ref NOTICE_ACTEON_2020_v003 V 003 17 58 PUT User manual BOOO 4 Block diagram of ACTEON 2020 menus Redox sensor HISTORY Gain adjust 6 1 2 Complete calibration 6 1 1 CALIBRATION Theoretical calibration 86 1 3 Temperature Sensor CALIBRATION LOGGER Gain adjust 6 1 5 Complete calibration 6 1 4 Theoretical calibratio
27. or filling system PONRES EH RV 5 litre capacity electrolyte tank for filling system PONROB EH RV Electrolyte tank tap for filling systems PONROB EH RV 10 metre tube for connecting the tank to the strainer and for filling systems PONSEL 7 58 flg PONSEL User manual ROOO PONCREP EH RV Filling system strainer 1 2 5 Accessories for Redox PONCPC EH 10 and Temperature PONCPC T 10 sensor in pipe installation PON4020 Tee assembly fitted with 20 x 27 male female sockets for AP1 To be fitted on PVC piping PON40 40 diameter glued tee assembly without connectors for API PONAPI Adapter sleeve for pH and temperature sensor in pipe installation PONVCPO 63 Stainless steel 63mm clamp type assembly for in pipe installation 316 L for SS OXY Eh pH Temp C2E and C4E sensors To be fitted with the appropriate nipple To be fitted on stainless steel piping PONNIP PH EH pH Eh sensor nipple PONNIP PH EH T Nipple for 2 pH T or Eh T or pH Eh sensors PONNIP T Temperature sensor nipple PONSEL CTEO 3 3 8 58 User manual ZBOOO 2 Installation 2 1 Mounting the ACTEON 2020 transmitter box ACTEON 2020 mounting diagram Dimensions mm 156 5 181 195 3 Figure Transmitter mounting diagram PONSEL Ref NOTICE_ACTEON_2020_v003 V 003 9 58 User manual BOOO
28. order of the letters is A Z 0 9 gt space oe Enter your name or reference then press ENTER PONSEL Ref NOTICE ACTEON 2020 v003 V 003 21 58 gs PONSEL User manual 2000 ORP SENSOR CALIBRATION Automatic zero Standard Oo mV 0 Measure mV ESC Cancel the procedure ENTER Validate the pH7 e a No Calibration If the automatic zero is incorrect an error message Correct window appears see Error message information chapter 6 2 1 Yes Immerse the Redox sensor in the 240mV or 470mV ORP standard solution ORP SENSOR CALIBRATION Immerse the sensor in an ORP standard solution Standard 240 mV Measure 234 mV WAIT Nonstable measurement ESC Leave the current slope Use the Wand V keys to adjust ENTER Validate the new slope the buffer value from OmV to 999mV x AWAIT MEASUREMENT STABILISATION Press the ENTER key to confirm the second calibration point PONSEL Ref NOTICE_ACTEON_2020_v003 V 003 22 58 PUT User manual BDOOO Calibration No ee Correct If the second calibration point is incorrect an error message window appears see Error message information chapter 6 2 2 Yes ACTEON 2020 ORP 290 mv Comment The measurement window is 20 00 c displayed again once the ud calibration has been P rie correctly carried out 200 PONSEL 23 58 gs PONSEL User manual DOOO
29. probe to be calibrated Select the SERERE ren and press ENTER CALIBRATION THEORETICAL CALIBRATION ATTENTION you will erase your calibration coefficients to return to the theoretical coefficients gt No Esc Cancel the procedure Enter Validate the choice Use the Aand V keys to select the procedure confirmation er Select I and press ENTER to return to the theoretical coefficients ACTEON 2020 ORP 290 nv 20 00 c mv Model RIX R2 X a rsr i can LOG If CONF INFO 200 PONSEL 3 3 32 58 PONSEL User manual BOOO 6 2 Redox sensor calibration error message 6 2 1 Automatic zero measurement error iMi i Nar M ORP SENSOR CALIBRATION IMPOSSIBLE Offset NON CORRECT Esc Escape menu If you observe an error during automatic zero measurement Contact the after sales service see last page 6 2 2 240mV or 470mV buffer solution calibration error Wil ORP SENSOR CALIBRTION IMPOSSIBLE Slope NON CORRECT Esc Escape menu If an error occurs during calibration with the 240mV or 470mV buffer 1 Check if the buffer is fouled or out of date 2 Check the connection between the sensor and ACTEON 2020 3 Clean the platinum electrode the sensor s response will be weak if the platinum electrode is soiled 6 3 Temperature sensor calibration error message information 6 3 1 OC calibration error CALIBRATION TEMPERATURE SENSO
30. re SENSOR TEMPERATURE CALIBRATION Immerse the sensor in water at ambient temperature Standard Bom ec Measure 20 20 ec WAIT Nonstable measurement ESC Leave the current slope ENTER Validate the new slope Use the Aand V keys to adjust the water temperature value x AWAIT MEASUREMENT STABILISATION Press the ENTER key to confirm the second calibration point PONSEL Ref NOTICE_ACTEON 28 58 PUT User manual 2000 Calibration Correct No If the second calibration point is incorrect an error message window appears see Error message information chapter 6 2 2 Yes Comment The 20 00 c measurement window is displayed again once the Model R1 R2 N calibration has been sri eL Proc GONE INFO correctly carried out PONSEL 29 58 gs PONSEL User manual 2000 6 1 5 Temperature sensor slope adjustment If you observe a small measurement error you may only adjust the slope of your sensor by carrying out the following procedure CALIBRATION SENSORS CALIBRATION OPR SENSOR gt Gain adjust gt Complete calibration gt Theoretical calibration TEMPERATURE SENSOR gt Complete calibration gt Theoretical calibration Use the Aand V keys to select ENTER date the choice the type of calibration and the probe to be calibrated Select the menu and press ENTER Immerse the sensor in w
31. s renere rren enten eene ene eene retener ren 30 6 1 6 Returning to temperature measurement theoretical calibration 32 6 2 Redox sensor calibration error message ceeeeee eee eee esee seen setas tanta sitas ta sss ea setas eaae eaae tas e ease enes es 33 6 2 1 Automatic zero measurement CTTOL 00 eeeeceeeceesceceeecessceceeeeeeneeceaeeesaeecsueeeeneeceaeeeeaeecseeeeneeceeeeenaeenees 33 6 2 2 240mV or 470mV buffer solution calibration error eeeeeeeeseeeeeeenren nennen nnne 33 6 3 Temperature sensor calibration error message information eere eerte eren neta nnun 33 6 3 1 a emicuit 33 6 3 2 Ambient water calibration error eene essere renten renere ennen nene en nest tenter nes nene nnns 34 PONSEL 3 58 ee User manual 2000 7 VIEWING MEASUREMENT HISTORY ccccceeeeseeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeees 35 8 VIEWING THE SENSOR CALIBRATION REPORT eerte 36 9 CONFIGURING ACTEON 2020 vsivsiessivsisissinetsucnduumiicsiniansansododianinnieanindoduawsinin 37 9 1 Configuring sensor response averaging scccecscccessssscssesscescssseesssesesscsssscsssssssssesessessessesssssesseees 38 9 2 Configuring the trend line eee ee eee esten eese teens ense neta sons tn stes tn saneta sens enses sens enses ense tato neta an 39 9 3 Configuring the two 4 20mA outputs eese esee esee eese tn sensn atn ense sense sns ta
32. sensor must be rinsed in deionised water between measurements Failing this use a demineralised water 13 1 4 Maintenance A After use rinse the electrode in demineralised water Return the sensor to the protective casing after having checked that the casing contains several drops of storage solution or if not available KCL saturated solution water for long periods of time Do not store the REDOX sensor in the protective casing filled with water especially very clean Although the reference solution may have a considerable amount of KCI salt content it may be depleted and used up prematurely if the salt is distributed in water with a low ionic strength If you have no suitable storage solution or saturated KC solution a few drops is sufficient in both cases you can simply moisten the inside of the protective casing before storage so as to maintain an atmosphere conducive to hydration of the glass pH bulb 13 1 5 Cleaning Rinsing in demineralised water is normally quite adequate If the sensor is dirty soak it in an Eh cleaning solution pepsin HCl for one hour Inthe case of extreme poisoning sulphur etc gently rub the platinum ring with a scotch brite light duty scrub type sponge for example From time to time grease the O Ring seals on the delrin shaft with vaseline The casing will slide easier and storage water tightness is improved 13 2 Temperature sensor Stainless steel sen
33. sesto sees sosta sone seer 40 9 3 1 Adjusting 4 20mA output stop thresholds esseeeeseeeeeeeenerenenneen eene nennen 41 9 3 2 Calibrating 4 20MA outputs RR meeen eE aE o E ESE A TEE E EER ARNE 43 9 4 Adjusting relay outputs R 45 9 4 1 Configuring relays inmode Utes oo eR EUER IE Rea 46 9 4 2 Configuring relaysn mode 2 egi e OE e ERE EUR Et die reU 47 9 5 Adjusting measurement units Lees eee esee ee eese teen testes tassa sets osse doki sssese sssi sess snae tasse EEN sasse snae 48 9 6 Setting the langQuage ccecccccccrcccressrsessacsvssesaseessesaceasersssecssassonsonsscansenseesesesscescasdsssessasusasssscassesevsatoessoanson 49 9 7 Resetting factory default Values cscsccccscscsccsssccsscssscsssssssssssccsssssecsseesscssscesscsscesscssscsssssssessssesesees 50 10 INFORMATION MENU ias sidbtivs aas bn fat isa a Ma MEGA UN UI EE VPN E 51 11 ADJUSTING THE ACTEON 2020 DISPLAY CONTRAST 52 12 TECHNICAL SPECIFICATIONSS 5552252222262 444 42122522ZcE vC F2F FM xcd da aS ECC UV D rd x RR ida aa 52 EX e lf c 53 13 1 REDOX ORP SEMSOM sisisccidscccccestsconcescesencossossdsbavevdenssdessscsasextecesass socessces donde o CSE ESET SSRS TSTS rE esas asaos 53 13 1 Specifications E er EE E E E EE E E TER EE R 53 13 12 DES CAP OD erari M 53 BA MES CERERI E E E EE EE 53 es MEE e E Re E A E
34. sing the stand and protective hood ssseeeseeeseeeeee nennen nennen neret 12 2 3 2 Installing the sensor with a sensor holder perch elbowed or straight and nozzle PONPPCC PH EH or PONPPCD PH EH 5 5n ks Ge eons a A east PEE ERR GRAAE ort Py Rr ES 13 2 3 3 Installing an Elbowed Sensor Holder Perch ref PONPPCC CIR or Straight Sensor Holder Perch ref PONPPCD CIR on QRPM ref PONSPFR and PONSPFRO2 eene 15 2 4 Ti pipe installation RRR 16 3 ACTEON 2020 TRANSMITTER ins cinis Bw Deer in law RD n res Cw ard 17 3 1 Control feni e EKES 17 4 BLOCK DIAGRAM OF ACTEON 2020 MENUS eese 18 5 THE MEASUREMENT WINDIQVN nnn 22zu su cuba anu ELLE EDEL venu nunca auia E Cru u dan 19 6 CALIBRATING THE ACTEON 20920 EEEL ELLE nena nnns 20 6 1 Calibrating th S NSOrs cc icccscsessesoceossssscssssoonscassescsesscsssesacesscssscesdeccesassnsssvaccessousssussonsenacdeccetassoassessseseese 20 6 1 1 Two point Redox ORP sensor calibration complete calibration eese 21 6 1 2 Redox sensor slope adjustment 2 602 nennen nennen rennen rennen tenerent 24 6 1 3 Returning to REDOX measurement theoretical calibration sese 26 6 1 4 Two point temperature sensor calibration complete calibration esee 27 6 1 5 Temperature sensor slope adjustment sec
35. sor with DELRIN shaft Working temperature 20 to 55 C Pressure 5 bars max Cable stranded Measurement principle PT100 3 wire assembly 13 2 1 Maintenance Rinse the temperature sensor with a water jet every 3 months PONSEL 54 58 PUT User manual BOOO 13 2 2 Mechanical diagram 020 5 85 2 O ring seals 16 8 316 stainless steel tube 03 40 PONSEL Ref NOTICE ACTEON 2020 v003 V 003 55 58 User manual ZBOOO 14 Questions amp Answers 14 1 Display screen troubleshooting The LCD screen is too dark Press ESC then press the V key several times to reduce the contrast Strange characters are displayed on the LCD screen RESET the microcontroller and follow the instructions below to reinitialise the LCD screen 1 Open the ACTEON 2000 transmitter box 2 Insert a LEAD or PLASTIC PENCIL in the hole located on the upper part of the box DO NOT USE SCREWDRIVERS OR METAL INSTRUMENTS nn Figure 11 Resetting ACTEON 2000 3 Check that the display screen restarts Comment Your settings and calibrations are not lost but saved on the EEPROM PONSEL Ref NOTICE ACTEON 2020 v003 V 003 56 58 User manual BOOO 15 Appendix Wiring diagram for the ACTEON 2000 terminal Ej i If the temperature sensor is not connected terminal points 28 and 29 must be connected to the shunt supplied with the terminal If the temperatur
36. to select the configuration required oe Select the IETTPPPEPEEREER menu and press ENTER CONFIG eee Temperature gt C degree Celsius gt F degree Fahrenheit Esc Cancel the procedure Enter Validate the choice Use the Aand V keys to select the unit desired oe Select the required measurement unit and press ENTER CONFIG WHAT DO YOU WANT TO DO gt Graphic screen gt 4 20mA outputs adjustment gt Relay adjustment gt Measure units gt Language gt Factory default value reminder PONSEL CTEO 3 3 48 58 gs PONSEL User manual 2000 9 6 Setting the language ACTEON 2020 can be configured in French or English CONFIG WHAT DO YOU WANT TO DO PAveraging sensor response gt Graphic screen gt 4 20mA outputs adjustment gt Relay adjustment gt Measure units gt Factory default value reminder Use the Aand V keys to select the configuration required oe Select the ERREA menu and press ENTER CONFIG Select the language gt French gt English Esc Cancel the procedure Enter Validate the choice Use the Aand V keys to select the language required eo Configure the language and press ENTER CONFIG WHAT DO YOU WANT TO DO gt Graphic screen 4 20mA outputs adjustment Relay adjustment Measure units Language Factory default value reminder PONSEL CTEO 3 3 49 58 gs PONSEL User manual
37. tructions below to access this information 20 00 c Model RIX Use the Pand lt keys to navigate the ACTEON menu Select the menu and press ENTER CALIBRATION LOGGER CALIB LOG Calibration of redox sensor CORRECT Standard 1 OmV Standard 2 240mv Offset 1mv Slope 110 Name Ponsel Calibration Temperature c Stand Standard 2 20 00 C 0 01 C Offset 1 65 C Slope 110 Name Ponsel Esc Escape menu List of diagnostic messages from the last calibration Message Explanation CORRECT ACTEON 2020 has been calibrated correctly THEORETICAL coef ACTEON 2020 is using the theoretical coefficients slope and offset this message is displayed after a theoretical calibration THEORETICAL slope ACTEON 2020 is using the theoretical slope THEORETICAL offset ACTEON 2020 is using the theoretical offset ACTEON 2020 is incorrectly calibrated because the slope calculated during INCORRECT slope calibration is incorrect in this case ACTEON 2000 is using the last correctly calculated slope ACTEON 2020 is incorrectly calibrated because the offset calculated during INCORRECT offset calibration is incorrect in this case ACTEON 2020 is using the last correctly calculated offset PONSEL 36 58 gs PONSEL User manual DOOO 9 CONFIGURING ACTEON 2020 Select the configuration menu in the measurement window ACTEON 2020 ORP 290 mv
38. ve the threshold the R1 relay is active R2 threshold 50 00 C hysteresis 0 10 above the threshold the R2 relay is active Temperature measurement unit C Language English 10 Information menu The INFO window provides information on the type and version of the device The software and hardware versions will be requested when you call PONSEL Technical support or After sales service INFO ACTEON 2020 ORP Software version Hardware version Esc Escape menu PONSEL Temperature 51 58 gs PONSEL User manual 2000 11 ADJUSTING THE ACTEON 2020 DISPLAY CONTRAST The LCD screen contrast can be adjusted to modify the display This adjustment is only possible from the measurement window Model RI R2 N Press the ESC and V keys simultaneously to reduce the contrast Press the ESC and A keys simultaneously to increase the contrast 200 12 Technical specifications Power supply 230 115V AC 60Hz optional 12 24VDC 4 20mA outputs 2 galvanic isolation outputs max resistive load 700ohms Adjustable from 1 000mV to 1 000mV Adjustable from 10 00 C to 50 00 C Relay outputs 2 relays that can be configured in 2 different modes Adjustment in alarm mode 1 Redox and 1 temperature C threshold Adjustment in adjustment mode 2 Redox thresholds Interrupting capacity 3A under 230VAC or 50VDC Standard Temperature measurement accurac
39. y C 25 C PONSEL 52 58 Ge ONSE User manual 2000 13 Sensors 13 1 REDOX ORP sensor 13 1 1 Specifications Length of glass section 65mm Working range 0 2 000mV Stability lt 3mV shift in 24 hours Temperature of use 5 to 50 C Pressure Max 5 bars Cable anti static coaxial cable max length 10m Response time lt 20 seconds for 95 of the response Minimum measurement volume 0 5ml Reference solution 4M KCL polysaccharide gel 13 1 2 Description Eh sensor platinum ring The 4M KCL polysaccharide gel that the Ag AgCl bar is steeped in communicates electrically with the Redox medium through reverse junction burn in 1 The ring shaped electrode is suitable for continuous measuring The KCL permeates outwards through this reverse junction burn in The high osmotic pressure generated increases the protection against contamination The sensor can operate for up to 4 years 20 50mm 84mm O Ring seal Platinum ring 12mm ZA 5 4 70mm Figure 10 Redox sensor diagram 13 1 3 Activation Remove the protective casing by pulling and twisting the entire assembly Connect the electrode to the measurement box Dip the electrode in the water to be analysed PONSEL Ref NOTICE ACTEON 2020 v003 003 53 58 gs PONSEL User manual 2000 Comment The