DermaFloat LAL User Manual
Contents
1. User Service Manual Joerns Support Surface DermaFloat LAL Model To avoid injury read user s manual before using transforming wound management a a a eee ee Joerns Support Surfaces DermaFloat LAL Important Precautions Important Notice The equipment must be installed and operated in the manner for which it was intended Facility staff user is responsible for reading and understanding the product user manual and contacting Joerns Healthcare if anything in this manual is unclear Joerns will not be held responsible for any injuries resulting from failure to comply with the instructions and precautions in this manual AWarning Joerns specialty support surfaces are designed as mattress replacement systems The risk of entrapment may occur when the equipment is placed on bed frames that leave gaps of even a few inches between the mattress and the head panel foot panel and bed or side rails The equipment is NOT to be used when such gaps are present Facility staff user is responsible for ensuring that all mattresses properly fit the bed frames Joerns is not responsible for the placement of its equipment on bed frames that leave gaps between the mattress and the head panel foot panel or bed or side rails which present a risk of harm to residents AWarning An optimal bed system assessment should be conducted on each patient by a qualified clinician or medical provider to ensure maximum safety of the pati
2. Position the power cord to keep personnel from tripping over it Remove the existing mattress from the bed Unroll the mattress with the hose connection at the foot end of the bed and the therapy cells facing up Secure the ten 10 straps on the mattress securely to the movable part of the bed frame If the therapy pad is not already on the mattress place it on the mattress Attach the elastic straps to the mattress buckles around each corner of the mattress Attach the six 6 additional straps to the movable part of the bed frame Hang the control unit on the foot of the bed facing away from the bed Attach the hose connector marked CPR to the control unit Joerns Support Surfaces DermaFloat LAL e Plug in the control unit and the yellow Standby light will illuminate Press the Power button The control unit will start and the green light will illuminate Keep the control unit ON while the resident is on the mattress Fully inflate the mattress by selecting Autofirm When the mattress is fully inflated select the Therapy mode and place the resident on the mattress Select the appropriate Comfort Adjust level to prevent bottoming out i e providing greater than one inch of air between the resident s sacral area buttocks and the lower safety mattress as outlined below 1 Begin by placing the head of the bed in the appropriate position based on the resident s clinical condition 2 Select the highest or
3. e Integrated CairRails risk management side air bolsters two inch side bolsters that inflate on both sides of the resident along the mattress edge to provide additional support and to provide a gentle reminder to the resident that they are near the edge of the mattress Constructed with lower ingress egress area Therapy Pads Many healthcare facilities are facing the challenge of infection control Joerns quilted therapy pads are treated with an antimicrobial to protect the therapy pad itself from the growth of mold mildew and odor causing bacteria Key features and benefits Treated with a highly effective bacteriostat agent to inhibit the growth of bacteria and fungi Constructed from a very smooth nylon fabric with low friction and low shear properties to protect the resident s skin from damaging friction shearing forces Breathable moisture vapor permeable fabric allows air to circulate beneath the pad and wicks away excess moisture This keeps your resident dry and helps to prevent skin maceration Two way stretch therapy pad is designed for optimal comfort moisture vapor transfer stain resistance and ease of laundering Grounding Instructions AWarning Use a properly grounded three prong 120V AC outlet for this product Failure to use a grounded outlet could result in personal injury or damage to equipment or house wiring including risk of fire A qualified electrician should be contacted to correct the
4. s liy Guel kesk r as ca ken ea k ka xu c kan K k ku l ks ku K k S k KAW AK RW k k RA R W u AK A n KAWAR KEK d HEAR RN E WA dR Ru w KE K KARE R N 10 Resident Migration Traction Skin Care Bed Height ei ld E E E E E E E E T 10 Control Unit General Cleaning Disinfecting Therapy Pad Steam Cleaning Filter Cleaning Maint 1 n six seca k sa rayana nali kanan aya a n aa asa a a naa ceeds rra a xa ada wara kina kaka ARAA kara aa ka ba a aa a sala k naa nala ku a yad ba kara ala ala kaka kk kaka a RR k 11 Storage AN Came E k ye dad kinn a niban kk RE WE DM s RA K RE a KE KAR ka RE ER RAM aS a Ku Wa k WE RR WER kk KERARA SR SR RAMA WAR RRA RD 11 Control Unit Support Surface DYSLOM SPSCITICALI ONS sas ss noc nis ak a ka kk k sa keke ku dek HERA n kk KEKE KK CAK A cies Kik KRA MEKR KAKE KA Hu A k kK k N RR GK HA KAKE KA HEKA AER e k EAR A 12 Weight Dimensions Electrical Specifications Environmental Conditions Agency Approvals Call for Assistance Warranty Introduction The DermaFloate LAL provided by Joerns Healthcare is a unique therapy system that provides pressure redistribution by combining low air loss with pulsation Low air loss therapy has been demonstrated to reduce the risk of pressure ulcers as well as being a valuable aid in the treatment of pressure ulcers The pulsation feature may provide additional reside
5. wiring and ensure a properly grounded outlet Before installing this product have the electrical system checked to make sure the electrical circuits and the electrical service are properly grounded Having a three prong outlet does not necessarily mean it is grounded Sometimes two prong outlets are replaced with a three prong type even though there is no ground wire There is always a chance of a loose connection or poor installation of a ground wire that causes the loss of proper ground at the outlet Inadequate grounding at electrical outlets can occur even if there is a ground wire Wires can become loose over time at the connection to the outlet Note To install new wires on a circuit requires a qualified electrician How to Determine if Your Outlet has the Proper Grounding Most hardware stores sell circuit testers Figure 1 that can be used to test an outlet for proper grounding The tester plugs into an outlet and by observing the indicator lights you can determine if the outlet is properly grounded For a higher level of assurance an electrician should be requested to thoroughly test the electrical j pa TE i jii system with more reliable i le equipment lwara l If repair or replacement of the cord or plug is necessary Figure 1 please contact Joerns Healthcare for assistance Setup Warning For important precautions see page two Caution Do not place the control unit on the floor
6. Power Failure After a power failure the unit will start up in Standby mode Reset the preferred mode and comfort level as described in the Operation section In the event of a power failure the lower support cells should remain inflated for up to 12 hours In case of an extended power failure transfer the resident to a hospital mattress or other surface CairRails Integrated CairRails risk management air bolsters offer a bilateral side bolster solution designed to address healthcare s growing concerns of liability in relation to resident falls and entrapment CairRails are being recognized by some of the nation s leading healthcare systems for improving their resident safety and risk management programs CairRails are recommended for residents requiring additional support during resident care and transfer CairRails can help reduce costs while ensuring optimal clinical outcomes and increasing resident safety Features and Benefits e Abilateral side air bolster solution which can enhance your facilities entrapment risk management program Easy to engage Ready Valve for instant inflation and deflation Transfer friendly deflate for ease of assisted transfer or when bolsters are not required Unique contoured design allows ease of ingress egress while providing additional protection comfort and supports resident compliance Designed to fit on most key Joerns therapeutic support surfaces Promotes maximum independe
7. most firm Comfort Adjust setting 3 Hand Check Place a hand with three 3 fingers if head of bed at 30 or higher or four 4 fingers if head of bed lower than 30 stacked vertically beneath the cells of the mattress and above the safety mattress directly between the lowest point of the resident s sacral area buttocks The smallest finger should be resting on the safety mattress 4 Sequentially reduce the Comfort Adjust setting to the firmness level where the height of the three 3 or four 4 fingers can slide with minimal resistance between the resident s sacral area buttocks and the lower safety mattress This is the proper Comfort Adjust setting for the resident to assure proper inflation of the air cells and prevent bottoming out of the mattress 5 Document the resident s Comfort Adjust setting for future reference and re evaluate with the Hand Check as the resident s condition warrants The CairRails risk management side air bolsters can be inflated or deflated as required Locate the turn valve on the hose assembly between the mattress and the control unit Next inflate deflate the CairRails by moving the turn valve to the up inflate or down deflate position Note When inflating or deflating CairRails it is recommended that the support surface be in Autofirm mode Joerns Support Surfaces DermaFloat LAL Operation amp Awarning For important precautions see page two ACaution The resid
8. this critical issue Joerns Healthcare can offer you the expertise assistance and products to bring your facility into compliance Joerns Compliance Solutions Matching the right bed components in order to meet regulatory guidelines can be complex That s why Joerns offers a wide array of compliance options We assist customers in selecting compliant accessories recommended for their specific bed model Creating a Safer Care Environment While the guidelines apply to all healthcare settings hospitals nursing homes and home care long term care facilities have particular exposure since serious entrapment events typically involve frail elderly or dementia residents Joerns Support Surfaces DermaFloat LAL For More Information To learn more about compliance options with Joerns products visit our website at www joerns com or contact our Customer Care representatives at 800 826 0270 and ask for free informational publications To learn more about entrapment zones assessment methods and guidelines concerning entrapment contact Joerns Healthcare at 800 826 0270 or consult the FDA website http www fda gov MedicalDevices ProductsandMedicalProcedures GeneralHospitalDevicesandSupplies HospitalBeds default htm Joerns Support Surfaces DermaFloat LAL Table of Contents Important Precautions WOVE OC UCT lel j PAR N en na ney en a E Ka Si EKE LE EINEN A BEE SA EREY Ae DEN M via ya yer KEN Moisture C
9. Joerns Healthcare should perform preventative maintenance Preventative maintenance performed by unauthorized personnel could result in personal injury or equipment damage Any maintenance done without Joerns s authorization will invalidate any warranties on this product Storage and Care When the product is not in use store the power cord properly Failure to do so could result in personal injury Note Clean the DermaFloat LAL as described in the previous section prior to storage Control Unit The power cord may be wrapped around the unit for convenience Wrap the unit in a plastic bag for dust resistance then store the unit in an area appropriate for an electronic medical device Support Surface Gently roll up the support surface expelling any residual air for temporary storage The mattress should be wrapped in plastic and or a clean bag for storage 1 1 Joerns Support Surfaces DermaFloat LAL System Specifications Weight Control Units eee sarda 10 Ibs 4 5 Kg Van Spi 22 Ibs 10 Kg Maximum weight capacity 500 Ibs 227 Kg Dimensions Control unit 7 5 19 cm W x 12 25 31 cm H x 5 5 14 cm D Mattress 36 89 cm W x 80 203 cm L x 10 25 cm D 42 106 cm W x 80 203 cm L x 10 25 cm D Electrical Specifications USA 120V AC 60 Hz 0 6A Environmental Conditions Operating Conditions Ambient Temperature 10 C to 400C Relative Humidity 30 to 75 Non Conde
10. an directed angle of articulation and guard against risks of resident migration or inadvertent deflation of resident surface Skin Care Monitor skin conditions regularly particularly in areas where incontinence and drainage occur or collect and consider adjunct or alternative therapies for high acuity residents Early intervention may be essential to preventing serious skin breakdown Bed Height To minimize the risks of falls or injury the resident surface should always be in the lowest practical position when the resident is unattended Make sure areas under and around the frame are clear of objects persons and parts of body before adjusting height Cleaning AWarning Unplug the control unit from its power source Failure to do so could result in personal injury or equipment damage Warning Do not expose the unit to excessive moisture that would allow for liquid pooling Personal injury or equipment damage could occur Caution Do not use harsh cleansers detergents such as scouring pads and heavy duty grease removers or solvents such as acetone Equipment damage could occur Control Unit Wipe off dust If necessary clean the housing exterior with a disinfectant solution or a mild detergent and a damp cloth Then wipe dry General Cleaning If there is no visible soilage with possible body fluids we recommend that you clean the mattress system with a mild detergent and warm water If disinfection is d
11. ath the resident This may be due to perspiration incontinence or wound drainage On the DermaFloat LAL moisture is controlled via the specially treated breathable fluid proof urethane coated nylon therapy pad The moisture vapor permeable fabric of the therapy pad allows a sufficient amount of air to circulate beneath the pad and wicks away excess moisture Joerns Support Surfaces DermaFloat LAL Shear and Friction Reduction Shearing occurs when the skin is stationary in relation to the support surface while the underlying tissues and vessels are stretched and damaged When a resident s skin rubs against another surface the result is friction The top surface of the DermaFloat LAL therapy pad is constructed from a very smooth nylon fabric with low friction and low shear properties to protect the resident s skin from these damaging forces Indications for Use Note The selection of a pressure redistributing surface needs to be based on each individual resident s clinical condition diagnosis and or co morbidities The choice and use of a support surface is one factor in a holistic program of wound care and treatment Spinal Cord Injury The DermaFloat LAL is not recommended for use by residents with unstable spinal fractures Advice should be obtained from the appropriate physician before using the DermaFloat LAL system for these residents Pressure Redistribution Pressure Ulcers Rehabilitation Neurology Derma
12. d Suite 6 London ON Canada N5V 3A9 P 866 546 1151 F 519 451 8662 2011 Joerns Healthcare 6110055 RevD 11 4326 19748 Dearborn Street Chatsworth CA 91311 P 800 966 6662 F 800 232 9796
13. ent s head should be positioned in the center of the top section of the mattress When using the mattress system always ensure that the resident is positioned properly within the confines of the bed Do not let any extremities protrude over the side or between the bed rails when the mattress is being used Power Switch The Power switch is used to turn the power on and off Standby The unit starts up in Standby Press the Power button to inflate the mattress When the Standby light is on it may also indicate that there has been a power interruption and the therapy control unit is ready to be turned back on Press the Power button and reset the preferred mode of therapy and comfort level Modes of Operation Autofirm Autofirm mode provides maximum air inflation designed to assist both residents and caregivers during resident transfer and treatment The unit will automatically return to the mode it was in prior to Autofirm either Therapy or Pulsate in approximately 12 minutes Therapy Modes Therapy and Pulsate Therapy The unit starts in the Therapy mode which is the standard low air loss therapy Pulsate Pulsate alternates between a decrease and increase of the pressure in all cells every 30 seconds Comfort Adjust The Comfort Adjust function is located in the center of the control panel The DermaFloat LAL can be customized to meet individual resident needs within a therapeutic window See Setup section f
14. ent The assessment should be conducted within the context of and in compliance with the state and federal guidelines related to the use of restraints and bed system entrapment guidance including the Clinical Guidance for the Assessment and Implementation of Side Rails published by the Hospital Bed Safety Workgroup of the U S Food and Drug Administration Further information can be obtained at the following web address http www fda gov MedicalDevices ProductsandMedicalProcedures GeneralHospitalDevicesandSupplies HospitalBeds default htm When using the mattress system always ensure that the resident is positioned properly within the confines of the bed Do not let any extremities protrude over the side or between the bed rails when the mattress is being used ADanger Explosion Hazard Do not use in the presence of flammable anesthetics Do not use in the presence of smoking materials or open flame Air flowing through the air mattress will support combustion ADanger To reduce the risk of shock adhere to the following instructions Failure to do so could result in personal injury or equipment damage Immediately after using the DermaFloat LAL unplug it from its power source Do not place or store the product where it can fall or be pulled into a tub or sink Do not place or drop the product into water or other liquid Do not remove the back of the control unit Refer servicing to Joerns Healthcare AWarning To
15. ervice the control unit Please call Joerns Healthcare for any service requests 11 The therapy pad top cover of this product is not air permeable and may present a suffocation risk It is the responsibility of the caregiver to ensure that the resident can use this product safely Save These Instructions for Future Reference Bed System Entrapment Information In April 1999 the U S Food and Drug Administration FDA in partnership with representatives from the hospital and post acute bed industry including Joerns Healthcare national healthcare organizations resident advocacy groups and other federal agencies formed the Hospital Bed Safety Workgroup HBSW The workgroup s goal is to improve the safety of bed frames for patients and residents in all health care settings who are most vulnerable to the risk of entrapment The efforts of the FDA and the HBSW culminated in the FDA s release of recommended guidelines intended to reduce the risk of entrapment including dimensional limits for critical gaps and spaces between bed system components and clinical guidance for assessment and implementation of bed side rails in various health care settings Entrapment zones involve the relationship of bed components often directly assembled by the healthcare facility rather than the manufacturer Therefore compliance is the responsibility of the facility As the leading manufacturer of long term care beds and a frontrunner in addressing
16. esired you may use a combination cleanser disinfectant as explained in Disinfecting area Resident care equipment that does not come in contact with mucous membranes or non contact skin requires low level disinfection Wiping surfaces with a properly prepared detergent or disinfectant carries out low level disinfecting Processing of dirty resident care equipment should take place in a designated area away from clean or sterile supplies and food preparation areas Detergent disinfectants should not be mixed with other germicides or detergents Using the proper dilution ensures the most effective killing power of the disinfectant Wash hands often and well including after removal of gloves Resident care equipment that is used in isolation areas should be disinfected in accordance with all internal policies and procedures regarding such equipment Disinfecting When there is visible soilage and between residents we recommend that you disinfect the unit and mattress with a tuberculocidal disinfectant Disinfectant should be registered with the Environmental Protection Agency EPA Use rubber gloves and eye protection Prepare detergent disinfectant registered by EPA as hospital disinfectant solution according to instructions on label for correct use dilution With support surface deflated thoroughly wipe down entire mattress as air cells will lie flat Be sure to reach all areas underneath and in between air cells All
17. is disabled To disable press the up and down Comfort Adjust arrows simultaneously Nursing Procedures Recommended Linen Special linens are not necessary for the DermaFloat LAL There is no need for a bottom sheet as the therapy pad should be covering the therapy cells at all times The resident should never be lying directly on the therapy cells Based upon the resident specific needs the following linens may be utilized Draw or slide sheet to aid in positioning and to further minimize friction and shearing Incontinence barrier pad for residents incontinent of urine and or stool and residents with heavily draining wounds Add top sheet blanket and or bedspread as needed for resident comfort Keep the amount of padding between the resident and bed to a minimum for optimum performance Changing the Therapy Pad Place the therapy pad over the therapy cells fitting the corner of the cushions into the corner of the therapy pad similar to a fitted sheet Zip the therapy pad along each side of the mattress tub Resident Positioning and Comfort General Repositioning Residents should be turned and repositioned per individual turning schedule or per facility policy It may be helpful to activate the Autofirm mode to achieve a firm surface for repositioning purposes The unit will automatically return to the mode it was in prior Joerns Support Surfaces DermaFloat LAL to Autofirm in approximately 12 minutes or
18. nce by allowing caregiver to decide when added protection is required Inflation Deflation Set the control unit to Autofirm mode or the highest comfort control setting to achieve quick inflation Locate the Ready Valve by unsnapping the flap on the hose set Turn the Ready Valve to inflate or deflate as desired Return the control unit to the desired settings and snap the open flap closed Note CairRails are meant to provide a documentable and functional intervention for the risk management issues of falls and entrapment but in no way guarantee the prevention of falls or entrapment occurrences When using the mattress system always ensure that the resident is positioned properly within the confines of the bed Do not let any extremities protrude over the side or between the bed rails when the mattress is being used Optional Accessories Additional therapy pads available for purchase e Bariatric size available Joerns Support Surfaces DermaFloat LAL Troubleshooting Therapy Surface is Not Inflating Ensure the hose connection from the therapy mattress system mattress to the control unit is securely connected Ensure that the control unit is plugged into an AC outlet Ensure that the power is not on Standby If on Standby press the Power button Ensure that all air cells are connected to the internal mattress manifold Unable to Change Therapy Mode or Adjust Comfort Control Make sure the Lockout function
19. nsing Storage and Shipping Conditions Ambient Temperature 10 C to 40 C Relative Humidity 10 to 100 Control Unit Agency Approvals UL Classified Medical Equipment UL 60601 1 Can CSA C22 2 No 601 1 Call for Assistance If you have any questions or require service on a Joerns product please call Joerns Healthcare at 800 826 0270 Mattress weight capacity only total weight must not exceed 42 bed frame manufacturers specified load capacity Notes Joerns Support Surfaces DermaFloat LAL Joerns Support Surfaces DermaFloat LAL Joerns Healthcare Warranty Program for Joerns DermaFloat LAL Support Surfaces Joerns Healthcare warrants the DermaFloat LAL mattress to be sold free from defects in workmanship and materials under normal and proper use for a period of two 2 years on the mattress and one 1 year on the cover and electromechanical mattress components compressors valves printed circuit boards hoses and couplers Damages arising from improper use will not be covered by this warranty Improper use is defined as but not limited to those caused by Burns Use of improper chemical agents Needle punctures cuts or abrasions Excessive loads Staining Negligent or excessive usage Improper maintenance handling and or cleaning Failure to use in the manner indicated in the DermaFloat LAL user manual Any modification repair or alteration done to the DermaFloat LAL that was not a
20. nt comfort and enhance the lymphatic system The DermaFloat LAL is designed as a mattress replacement system for residents weighing up to 500 Ibs AWarning The risk of entrapment can arise when equipment is placed on bed frames that leave gaps of even a few inches between the mattress and the head panel foot panel and bed or side rails The equipment is NOT to be used when such gaps are present See Important Precautions section of this manual The DermaFloat LAL mattress replacement system is suitable for both the treatment of existing pressure ulcers Stage through Stage IV as well as for those who have been assessed at risk from the complications of immobility The DermaFloat LAL is quiet comfortable and simple enough for single caregiver installation featuring rapid inflation in just five minutes or less The user friendly controls allow for easy adjustment of resident comfort Additionally low friction and low shear materials together with average interface pressures well below capillary closure levels means that the DermaFloat LAL meets the comfort and clinical requirements of your residents up to 500 Ibs We have ensured that the DermaFloat LAL addresses the four key areas in the treatment of compromised skin pressure redistribution moisture control and reduction in both friction and shearing forces Moisture Control Residents are at risk for skin maceration if excess moisture is permitted to accumulate bene
21. ontrol Shear and Friction Reduction indications for USC visits cinveasenctacanotesnnscasnssncxennttanneiontnttntsannanandnnsastuhshaasasdnansinnsencnancanisannncasasnaanadnessthniivanstacsnanaseenexniants 5 Spinal Cord Injury Pressure Redistribution Pain Management Fe l UI Su xasi sara sa danka k xakan sa kasa ka kaka akil kaka xala E kalas aa a Kak kala secu a Ra a aya la r a d ala a kd a a aa adi aa ia a i s a ka la ala k E a k 6 Therapy Control Unit Features Therapy Mattress System Features Therapy Pads Operation Modes of Operation Additional Feature iiiciecc25c oe cece ceceSucsciek enian Senne cues dace sawa wa k k kul as bu Ea kan kan cuca a u a W KE Ka kw kk ka les xi kn j al de xw l n du awa e xu E kan 8 CPR Transport Power Failure CairRails Optional Accessories TI GUD lO SI TO O lak EPA nnn kk neka na enan r kk n an R ARAS E T E K k R k ERA KANA A K RRA k KER K EA KUN RA GERA Re A AE R K EKA ARA R KA AGR 9 Therapy Surface Not Inflating Unable to Change Therapy Mode or Adjust Comfort Control Nursing PO C U S i sis kis ark ka ue kaka An hesan ek ke kak k GA ku UW uc Ka RR Kub N A s n H KU K Ku Hek AT A Ru Ke Cu A G KRG K kul kK KAD A ku k a kawa kk A KERR ASO REKAR 9 Recommended Linen Changing the Therapy Pad Resident Positioning and Comfort Incontinence Saloy li GELIYA
22. or more information This function will not work in Autofirm mode Lockout This feature is to prevent any unauthorized changes to the resident settings To unlock and make adjustments to the settings press both up and down comfort arrows at the same time to disengage the Lockout function The Lockout function will return in approximately five minutes Comfort Adjust Autofirm Therapy Pulsate Figure 2 Note The unit is designed to lock out all the adjustment controls after the resident has been positioned correctly In approximately five minutes after the last button push the power on light begins to flash indicating Lockout is enabled Additional Features AWarning A possible fire hazard exists This product is suitable for use with oxygen administering equipment of the nasal mask or half bed length tent type only To prevent personal injury or equipment damage ensure that the oxygen tent does not extend below the mattress CPR The hose connection at the control unit is marked CPR Disconnect the hose from the control unit Deflation times will vary based on resident weight and profile To resume therapy reattach the hose to the control unit Transport To transport the resident in bed turn the control unit off Unplug the power cord from the outlet Do not disconnect the hose connection at the control unit The lower support cell will prevent the resident from bottoming out for up to 12 hours
23. ow to air dry If dust or other soiling has accumulated along air hoses remove using swabs moistened with detergent disinfectant as necessary Allow all components to air dry Wrap mattress in plastic and return to storage area Thoroughly wipe down outside of control unit and allow to air dry Cover with plastic and return to storage area Remove gloves and dispose wash hands Therapy Pad The therapy pad can be wiped down with a disinfectant solution or a mild detergent with a damp cloth If heavily soiled the therapy pad can be laundered in a washer and dryer with warm water no more than 120 Fahrenheit A non bleach detergent should be used sparingly Wipe dry or allow to air dry Steam Cleaning Do not use any steam cleaning device on the unit Excessive moisture can damage mechanisms in this unit Filter Cleaning Check the air filter on the rear of the unit regularly for buildup of dust dirt If buildup is visible turn off the control unit and disconnect the power cord from the wall outlet Remove the filter by grasping the filter and pulling outward Replace with the second supplied filter Ensure the replaced filter covers the entire filter region Hand wash the removed filter in warm soapy water and allow to air dry When dry store the filter in a safe place for the next filter maintenance Joerns Support Surfaces DermaFloat LAL Maintenance AWarning Only facility authorized personnel trained by
24. powertul with a reusable air filter Crisp easy to read graphics for intuitive set up and therapy control Automatic panel lock out to avoid unwanted or accidental adjustments Eight therapeutic comfort control settings to maximize resident compliance and promote healing Closed loop pressure sensor control system eliminates concerns of changes in mattress interface pressure due to ambient temperature and pressure changes Integrated swing out hanging brackets for fixing to most bed types Quick disconnect hose feature allows for rapid attach and CPR deflate at the control unit Therapy Mattress System Features Twenty individual therapy cells help to evenly distribute the resident s weight and maximize pressure redistribution Modular cell design mattress for ease of cleaning re assembly and cost effective service Lower two inch safety cell remains inflated without any power up to 12 hours Eight inch deep therapy cells are constructed of highly durable polyurethane coated nylon to provide adequate support and prevent bottoming out for most residents within weight limit Integrated low pressure heel section provides lower interface pressures for high risk residents Durable base tub is constructed from 100 heavy weight 1680 Denier nylon with 1 5 oz urethane coating and incorporates bed attachment loops for stability Maximum weight capacity of 500 Ibs Anti kink easy clean air supply hose set e Quick CPR deflation
25. reduce the risk of burns shock fire or personal injury adhere to the following instructions Failure to do so could result in personal injury or equipment damage 1 Use this product only for its intended purpose as described in this manual Only use attachments and or accessories that are recommended by the manufacturer 2 If this product has a damaged power cord or plug is not working properly has been dropped or damaged or has been dropped into water do not operate it For examination and repair return the product to Joerns 3 Keep the control unit and power cord away from heated surfaces e g space heaters 4 Never block the air openings of the product Do not place the control unit on a surface such as a bed or couch where the air opening and or filter compartment located on the back of the control unit may be blocked Keep the air openings free of lint and hair 5 Never drop or insert any object into any opening or hose 6 Do not spill food or liquids onto the control unit If a spillage does occur turn off the unit disconnect it from its power supply and allow at least 24 hours for drying 7 Do not use the product outdoors or where aerosol spray products are used 8 Plug this product only into a properly grounded outlet Refer to Grounding Instructions 9 Ensure nothing is placed on the power cord and ensure it is not located where it can be stepped on or tripped over 10 Do not attempt to s
26. tology Burns Amputations Pain Management AIDS Arthritis Oncology The DermaFloat LAL provides distribution of weight over a wide surface area which redistributes pressure against bony prominences and provides a soft gentle therapy surface to lie on For residents experiencing severe pain and discomfort due to pressure and or positioning limitations consider the DermaFloat LAL as an adjunct to pain management interventions Note Pressure redistribution and pain management are conditions and diagnoses for which the DermaFloat LAL may be indicated Occasionally there are orthopedic and neurological residents that require body positioning to be maintained in specific alignment The use of the DermaFloat LAL for these residents should be considered on an individual basis and discussed with the attending physician Joerns Support Surfaces DermaFloat LAL Features The DermaFloat LAL is comprised of two components Therapy control unit Therapy mattress system Therapy Control Unit Features True low air loss with greater than 100 liters of airflow per minute Three modes of operation Autofirm Therapy and Pulsate Pulsate mode oscillates the air throughout the mattress every 30 seconds providing increased resident comfort Autofirm mode provides maximum air inflation designed to assist both residents and caregivers during resident transfer and treatment Compact lightweight control unit is quiet robust and
27. uthorized in writing by Joerns Healthcare will void this warranty Damage caused by use in unsuitable environmental conditions abuse or failure to maintain the product in accordance with user and service instructions is not covered This warranty is extended to the original purchaser of the equipment Joerns Healthcare www joerns com email info joerns com Post Acute Acute HomeCare VA Government 5001 Joerns Drive Stevens Point WI 54481 P 800 826 0270 F 800 457 8827 Parts Joerns DermaFloat LAL contains various parts that wear from normal use Joerns Healthcare s obligation under this warranty is limited to supplying replacement parts servicing or replacing at its option any product which is found by Joerns to be defective When requested by Joerns parts must be returned for inspection at the customer s expense Credit will be issued only after inspection Service Most service requests can be handled by the facility Maintenance Department with assistance from the Joerns Healthcare Product Service Department Most parts requested can be shipped next day air at the customer s expense Should a technician be required one will be provided by Joerns Healthcare at our discretion Only the Joerns Healthcare Product Service Department can dispatch authorized technicians Manufactured by Joerns Healthcare LLC 1032 North 4th Street Baldwyn MS 38824 Canadian Office 1000 Clarke Roa
28. you can manually return to therapy mode once resident has been repositioned Unless counter indicated it is desirable to keep the back section of the bed in the flat position to provide optimal pressure redistribution and minimize the risk of shearing injuries Elevating Resident into Sitting Position The special properties of the DermaFloat LAL therapy pad reduce the opportunity for shear and friction that may occur when raising the back section of other bed systems As with any surface sliding can be expected therefore residents should be repositioned after elevation The knee gatch or knee section of the bed may be elevated first to help prevent the resident from sliding when the back section is elevated Incontinence Moisture against the skin surface leads to maceration or softening of the tissues To prevent maceration we recommend you use an incontinence barrier pad to absorb the excess moisture In the event of incontinence or excess drainage on the therapy pad you should wipe off the excess fluid from the bed surface Safety Information Resident Migration Specialty bed products are designed to reduce redistribute pressure and the shearing friction forces on the resident s skin The risk of gradual movement and or sinking into hazardous positions of entrapment and or inadvertent bed exit may be increased due to the nature of these products Traction With any traction or unstable fractures maintain physici