Home
        Poseidon MkVI User Manual
         Contents
1.      Figure 1 22   Closeup of hose connection     Figure 1 23    Push hose connector into  receiver until outer flange Is  flush with start of threads     Figure 1 24    Hand tighten the swivel  capture nut  Do NOT use  tools or over tighten     First  select two hoses to be used as the rear left and right breathing hoses  Figure 1 21   Insert  the end of one of the hoses into the left  inhalation  threaded breathing manifold  Figures 1 22  and 1 23   Pay particular attention to the condition of the radial o ring on the end of the hose  and to the radial sealing surface inside the lip of the threaded breathing manifold  The o ring   O ring groove  and sealing surface at the hose junction must all be clean and free of all debris   free of scratches and gouges  and properly lubricated before assembly  Be sure when inserting  the hose end into the threaded breathing manifold that the radial o ring does not extrude from  its circular groove  The radial o ring in the hose end should insert smoothly into the radial  sealing surface  Figure 1 22  until it is no longer visible and the top rim of the hose end is flush  with the ledge just above the threads on the manifold  See Figure 1 23      Once the hose fitting is properly inserted  Figure 1 23   slide down and hand tighten the hose  swivel nut  Figure 1 24   Make sure that the threads are properly engaged and are not cross   threaded  The fittings are designed for ease of assembly  and the capture nut should freely  rotate until 
2.     Figure 1 34 show how to connect the oxygen regulator to the oxygen cylinder valve  Both the  cylinder fitting and the regulator fitting and o ring should be carefully inspected for signs of  organic debris  grease  oil  and hydrocarbons  If the regulator o ring is damaged  gouged   scratched or otherwise worn or cut  replace it with an appropriate Poseidon oxygen cleaned  o ring     The oxygen cylinder valve  for European users  is an M26x2 female DIN thread  This is larger  than the G 5 8 DIN thread that is commonly used in both the United States and in Europe for  compressed air Scuba  The purpose of using different threads is to make it absolutely clear  that the oxygen delivery system is different from the diluent delivery system  Accidental con   nection and use of diluent through the oxygen regulator can be considered a compromise of  the oxygen cleaning requirements  If contamination occurs  the compromised components  should be taken to an authorized Poseidon Tech Center representative or a qualified tech dive  shop technician for oxygen cleaning        POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE 24    DANGER     All components exposed to pressurized oxygen   including  the oxygen cylinder  valve  and regulator   must remain free  of hydrocarbons  grease  oil  gasoline etc  and other organic  compounds  Never expose these components to compressed air   which can contaminate components with oil  Always have these  components cleaned and serviced by a Poseidon Tech Cen
3.     of the MkVI is the location  furthest behind the diver   s back when wearing the rig normally  Following is a  brief description of each of the major components     Primary display    The Poseidon MkVI is equipped with a custom glass liquid crystal display   LCD  with large  bright  crisp letters for easy reading at a glance underwater   It is designed specifically for recreational divers and presents only the informa   tion needed for safe operation  It includes an automatic high efficiency back   light that illuminates the panel when ambient light levels are low  It also  includes an infrared data port  which allows communication with a PC for dive  log downloading  parameter setting  and software upgrades  Two wet switch  contacts on the back of the display activate the MkVI electronics     Figure 1 1  Front view of assembled Poseidon MkVI        Right rear  breathing hose    Right side   exhalation   counterlung    Right side water    diversion  manifold    Right front  breathing       hose  Counterlung  Over  Pressurization  Valve  OPV     dr      Jetstream Octopus       Closed circuit open circuit  convertible mouthpiece                      CHAPTER 1 PAGE 1   Fe j D  rie   T   v i               Ie  ER      Left rear    breathing hose    Left side   inhalation   counterlung    Left side water  diversion manifold    Left front  breathing hose    Auto diluent  addition valve  ADV     Primary display    Head up display  HUD     POSEIDON MKVI USER MANUAL    Open circuit
4.    One of the important new features in the Poseidon MkVI is the Hyperoxic Linearity test  When  the oxygen sensors are calibrated during the pre dive routine  Chapter 2   the linearity of the  oxygen sensor response is only validated up to a PO  value of 1 0 bar   atm  i e   100  oxygen  at sea level   Most rebreathers assume that the sensor response remains linear at higher values   operational PO  setpoint values often exceed 1 0 bar   atm   However  in certain situations the  sensors may not be linear above 1 0 bar   atm  which can lead to a very dangerous situation   For example  if the sensor is not capable of responding to PO  values greater than 1 2 bar    atm  and the PO  setpoint is 1 2 bar   atm  the control system may flood the breathing loop  with dangerously high levels of oxygen while attempting to achieve a PO  value that the sensors  are not capable of registering     To overcome this problem  the Poseidon MkVI performs a test on the primary oxygen sensor  the first time a depth of 6 m   20 ft is achieved  The test injects a short burst of oxygen directly  onto the primary sensor to ensure the sensor response is linear up to a PO  value of 1 6 bar    atm  If the test passes  then the dynamic setpoint performs as described previously e  using  up to the    deep    PO  setpoint value when the depth exceeds 15 m   50 ft    However  if the  Hyperoxic Linearity test fails  then the maximum allowable setpoint is set at 1 0 bar   atm  The  reason for this is that th
5.   2  If test failure persists  contact an authorized  Poseidon Service Center for repair     1  Standard Response  2  If test continues to fail  attempt to reinstall  Firmware  may cause unrecoverable failure   3  If test failure  persists  contact an authorized Poseidon Service Center for repair     1  Standard Response  2  If test continues to fail  reset system parameters   3  If test failure persists  contact an authorized Poseidon  Service Center for repair     1  Standard Response  2  If test failure persists  contact an authorized  Poseidon Service Center for repair     1  Standard Response  2  If test continues to fail  attempt to reinstall  Firmware  may cause unrecoverable failure   3  If test failure persists  contact  an authorized Poseidon Service Center for repair     1  Standard Response  2  If test continues to fail  try a different battery   3  If test failure persists  contact an authorized Poseidon Service Center  for repair     1  Standard Response  2  If test failure persists  or if backlight does  not turn on during this test  contact an authorized Poseidon Service  Center for repair     POSEIDON MKVI    TH TIME  SEC     18    20    22    24    Zo    26    2     4 5    4 5    4 5    USER MANUAL    DESCRIPTION    HUD LED  This test measures the amount of electrical current  consumed by the red LED in the Head Up Display  HUD    when activated     Buddy Light LED  This test measures the amount of electrical  current consumed by the red LED in the battery  
6.   3  If test failure  persists  contact an authorized Poseidon Service Center for repair     Breathing hoses can only be re attached to the end fi ttings by a qualified Poseidon Service  Center  Do NOT attempt to re attach the fi tting without understanding the correct procedure  for doing so  An incorrectly attached fitting may appear to be conneted properly  can easily  and suddenly detach underwater  leading to a fl ooded breathing loop     1 Do NOT attempt to dive with a cracked absorbent cartridge  If gas leaks through the cartridge  wall  CO  may bypass the absorbent material  and enter the inhalation side of the breathing  loop  which can cause CO2 poisining  Always replace a cracked cartridge with a new one     Use the various adjustable straps on the counterlungs to secure them in the proper position   see Chapter 3 of this Manual   It may take several attempts in a pool or other confi ned body  of water to get them just right  but the effort is well worth it  Well positioned counterlungs  reduce the work of breathing     Loosen the cylinder cam straps on one of the two cylinders  and carfeuly adjust its height such  that both cylinders are the same height  When the cylinders are mounted properly  the unit  should stand upright without wobbling     1  Ensure all fi ttings are snuggly attached  2  Remove the hose from the leaking fitting to  inspect the o ring and sealing surfaces for signs of damage  and clean or replace o rings as  needed  3  If leaking persists  co
7.   Pressure  Dry in accordance with EN14143      With the nuts loose  pull the webbing handle on the bottom of the housing in a fashion  so that  it moves in the opposite direction from the electronics housing  Because o ring seals can    set     with time it may be necessary to restrain the tube with one hand while pulling with the other   see Figure 1 49   Make sure that the nuts are completely free from the heli coils before pulling  out the bottom lid     Once the dual radial seals on the end cap have cleared the end of the housing tube  the CO   absorbent cartridge will freely slide out  as shown in Figure 1 50     SofnoDive   797 Pre pack  CO  absorbent Cartridge       Figure 1 47   The CO  absorbent cartridge with the SofnoDive   797 Pre pack CO  absorbent Cartridge inside     POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE 31       WARNING     The CO  absorbent cartridge is intended for use with a single fill  of the standard 3 liter oxygen cylinder provided with the Poseidon  MkVI  The cartridge MUST be replaced whenever the oxygen  cylinder is re filled  When in doubt  discard the cartridge and  replace it with a fresh one        Base plate nut  x 4     Cartridge end cap       Figure 1 51  Figure 1 52  Figure 1 53   Remove the cartridge Remove the end plate  Lift and remove the lower sponge   top plate  inspect for collected water     Rinse  dry  and store the lower  sponge  disinfect if desired     The sponge should be dried after every dive  even if a repetitive dive is p
8.   The test number is displayed on the left side  of the screen  preceeded by a lower case    t     as shown in Figure 2 5   While the test is running   a    spinning wheel    is displayed on the right side of the display  where the remaining dive time  is normally displayed  This    spinning wheel    is represented by a    O    character in the right most    POSEIDON MKVI USER MANUAL    position  missing one of the segments  The missing segment rotates positions around the    0     in a clockwise direction  Different tests require different amounts of time to complete  some  require less than one second  others require 4 12 seconds to complete  Certain tests that  involve some action by the diver allow for up to 2 minutes to complete  if necessary     ka    4  Ki  mg               Figure 2 6  Test 17 failure  with flashing test  number and error code     DA       Figure 2 5  Jest 17  Backlight Power  consumption   displaying test number on the  left and the    spinning wheel    on the right     Two things are worth noting in Figure 2 5  First  the bar graph along the top of the LCD screen  is a progress bar  indicating how much time remains before the PST is complete  or how much  time is left for the diver to complete some required action  Second  the circle with the diagonal  line through it in the upper right corner of the screen is the    Do Not Dive    symbol  which is  displayed throughout the entire pre dive process  As long as this symbol is displayed  the dive  shoul
9.   contact an authorized  Poseidon Service Center for repair     1  Ensure the oxygen cylinder is connected to the oxygen regulator  and that  the valve is turned on  2  Ensure that the oxygen cylinder contains sufficient  oressure  3  If test failure persists  contact an authorized Poseidon Service  Center for repair     1  Ensure the diluent cylinder is connected to the diluent regulator  and that  the valve is turned on  2  Ensure that the diluent cylinder contains sufficient  oressure  3  If test failure persists  contact an authorized Poseidon Service  Center for repair     1  Standard Response  2  If test continues to fail  subject battery to a Learn  Cycle on the charger  see Chapter 1   3  If test continues to fail  try a different  battery  subject to decompression data issues   4  If test failure persists   contact an authorized Poseidon Service Center for repair     POSEIDON MKVI    TH Time  SEC     49    5O    53    120    120    120    USER MANUAL    DESCRIPTION    Positive Pressure Loop Test  Besides checking for leaks in the  breathing loop  this test checks several additional things   including leaks in any of the four solenoid valves  that gas  actually passes through each of the two metabolic solenoid  valves  that the dump valve on the right counterlung is secured   and that the depth sensor Is sensitive to small pressue changes   Because there are several things being checked during this  test  there are also several different kinds of failures  with differ
10.   go through the Pre Dive checklist located on the  sticker placed on the back side of the primary display  The Pre Dive checklist is also located  in chapter 2 on page 46     POSEIDON MKVI USER MANUAL    Servicing    Poseidon MkVI cylinders should be hydro tested once every five years and visually inspected  yearly  Regulators should be rebuilt every two years  Oxygen regulators  cylinders  and cylinder  valves should be oxygen cleaned every two years  All of these services form a part of the  Poseidon Tech Center biannual service  Contact your Poseidon representative for details     We recommend that you pay close attention to the primary display  at test 55  during the pre  dive check  At this test  the remaining number of weeks  until the next scheduled service date   is displayed     Once the service date is due  the diver is prompted to acknowledge  similar to    power up proce   dure     that he she has understood that service is needed  A grace period of 4 weeks is added     WARNING  It is recommended practice to don the rebreather  unit only after completion of the Pre Dive Procedures described  in the next Chapter  Not only is it easier to diagnose and correct  many test fails when the rig is not on the diver   s back  but also  the risk of oxygen fire is greatest when cylinders are pressurized   which is usually done during the Pre Dive Procedures        CHAPTER 1 PAGE 36         Figure 1 63   Ready for Pre Dive tests  As configured in this photo  the Poseidon MkVI
11.   graph will flash  The full scale  100   value of each of these bar graphs is established using the  PC Software     Do NOT ignore the remaining battery life indicator  If the battery    fails  the entire life support system  including alarms  may cease  to function  Failure to abort to open circuit and terminate the dive    could lead to serious injury or death        Figure 3 31   Diluent cylindepressure indicator     Figure 3 32   Oxygen cylinder pressure indicator        Temperature    Immediately to the left of the Battery Life Indicator  is the Temperature reading  This value is displayed  in units of centigrade when in metric mode  and  units of fahrenheit when in imperial mode   NOTE  These numbers will show the last two  decimals of the PO  value  if it has been set to     High Res    in the PC Config tool        Figure 3 30  Temperature     Ascent rate indicator    The bar graph along the very top of the Primary  Display indicates the divers current ascent  rate  It soans from left to right  and is not dis   played when the diver is not ascending  If the  bar is half way across the width of the screen   the diver is ascending at a rate of 9 m   29 5 ft  per minute  If the status bar is shown across  the entire width of the screen  the diver is  ascending at arate of 18 m  59 0 ft per minute     The segments in this graph will flash if the safe    Figure 3 33  Ascent rate indicator  ascent rate  10 m   33 ft per minute  is exceeded     POSEIDON MKVI USER MANUAL    S
12.  2  If test continues to fail  reset system parameters   HUD  Head Up Display   which contains user selectable con  EEPROM 3  If test failure persists  contact an authorized Poseidon Service Center for  figuration information  for internal errors or data corruption  repair    6 5 Backpack ROM   RAM   Fuses  This tests the RAM  ROM and 4 Bad RAM 1  Standard Response  2  If test failure persists  contact an authorized  fuse settings of the electronics in the backpack processor  5 Bad Fuse Poseidon Service Center for repair    The RAM is tested only when the battery is inserted  and the  results used for all subsequent power up routines  Other tests 3 Bad ROM 1  Standard Response  2  If test continues to fail  attempt to reinstall    are conducted on each power up routine     Firmware  may cause unrecoverable failure   3  If test failure persists  contact  an authorized Poseidon Service Center for repair     POSEIDON MKVI    TH Time  SEC     7    14    15    16    17    1    USER MANUAL    DESCRIPTION    Backpack EEPROM  This tests the EEPROM  static memory   in the backpack processor  which contains user selectable  configuration information  for internal errors or data corruption     Battery ROM   RAM   Fuses  This tests the RAM  ROM and  fuse settings of the electronics in the battery processor  The  RAM is tested only by the factory or when new firmware is  installed  and the results used for all subsequent power up  routines  Other tests are conducted on each power up routi
13.  Before the dive  make sure there is enough diluent  air  and oxygen to carry out the dive you  plan to do  The EU version of the Poseidon MkVI comes with a 3 liter   183 cubic inch  aluminum diluent  air  cylinder  with black valve knob  with a rated fill pressure of 204 bar    2958 psi  Filled to its maximum allowed working pressure  it holds 612 liter   21 6 cubic feet of  air  Because this cylinder is your open circuit  OC  bailout gas in the event of an emergency     Poseidon strongly recommends that this cylinder be full at the start of each dive  The included  oxygen cylinder  white valve knob  has the same capacity and pressure rating as the diluent  cylinder  but the recommended maximum filling pressure of oxygen is limited to 135 bar   2000 psi  for reasons of fire safety     Attach both the diluent and oxygen cylinders using the procedures described in Chapter 1  Do  not turn the cylinders valves on initially  as this will result in wasted gas during certain portions  of the pre dive tests  As described below  the cylinders should be turned on when the pre dive  checks reach Test number 44 and 45  The pre dive tests will fail if the pressure in the diluent  cylinder is less than 51 bar   739psi  or the oxygen cylinder is less than 34 bar   493 psi  Similarly   if starting a dive with only a marginal amount of gas above these minimum safety limits  these  gas pressure limits will be reached soon after the start of the dive  leading to an unsatisfying  diving experie
14.  Breathing hoses properly connected to the mouthpiece        CHAPTER 1 PAGE 28    The final step in assembling the breathing loop is the convertible open circuit   closed circuit  mouthpiece  Figure 1 45 shows the bottom half of the mouthpiece and the two closed circuit   CC  hose connections  Each CC connection contains within it a removable flow checkvalve  and o ring sealed carrier plate  These checkvalves constrain the direction of the breathing gas  from left to right  The left CC entry port into the mouthpiece is the    upstream    or    inhalation     port  and the right CC entry port is the    downstream    or    exhalation    port  Figure 1 45 shows  a close up view of the downstream CC checkvalve  It should be free of debris and should lie  smooth and uniformly flat against its removable carrier plate  If there is any curling  cuts   gouges or other damage to the checkvalve  remove the carrier plate and replace the check   valve only with a Poseidon original manufacture mouthpiece checkvalve  Attach the right hand  front breathing hose to the downstream CC port as shown in Figure 1 45  following the proce   dures previously described             Inhalation  left   CC port    Inhalation close  circurit checkvalve    Exhalation  right  CC port                  Exhalation close ee di checkvalve         Figure 1 45  msbdki the inhalation CC and  exhalation CC checkvalve on the mouthpiece   Attach the front breathing hoses to the inhalation VS  and exhalation OC ports on
15.  Module around the braided 90 cm hose and attach the  HUD to the moutpiece and make sure it is properly secured  See Figure 1 38 and 1 39        Diluent 1st stage IP port  E      Gong braided 90cm  Low Pressure hose        Long braided  90 cm Low Pressure hose    9 16    connection       Figure 1 39   Figure 1 37  Connect the long braided 90 cm Low Pressure hose to the  Connect the long braided 90 cm Low Pressure hose to the diluent 1st stage IP port  mouthpiece and attach the HUD to the mouthpiece     POSEIDON MKVI USER MANUAL       10  Jetstream Octopus    The Poseidon MkVI is delivered with an alternative air source  i e  Octopus  that should be  connected to the diluent 1st stage     To fully understand the funktions ofthe Jetstream NS please read the Jetstream Octopus  manual  WG    CH       A    Figure 1 40  i  The alternative air source  i e  Octopus  should be connected to the diluent 1st stage and the    LP hose should be routed in such way that it allows easy access to the alternative air source  in case of an emergency situation     CHAPTER 1 PAGE 26       Figure 1 41  Connect the alternative air source low pressure hose  to a low pressure port  marked    IP     on the diluent  black connection  wheel  1st stage     POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE 27       11  Front CC hoses to counter lung T section    Connect the left front CC loop hose to the inhalation counter lugn T connection  Lay out the two remaining breathing hoses  Connect the right front breathi
16.  O      When the displayed value is the PO   Setpoint  the first digit  to the left of the decimal  is shown using only the upper half of the  numeric value        Figure 3 14  PO  Set point value Figure 3 15  Real PO  value     The Poseidon MkVI incorporates a dynamic PO  setpoint value  which means the setpoint  changes depending on depth and decompression status  Two setpoint settings control what  the range of setpoint values will be during the dive  A    surface    setpoint value  default 0 5 bar    atm  establishes the PO  setpoint when at the surface  and a    deep    setpoint  default 1 2 bar    atm  establishes the PO  setpoint when at a depth greater than of 15 m   50 feet  Between  these two depths  the setpoint changes in small increments between these two values  Thus   when the depth is less than 15 m  50 ft  the setpoint will be some value between the    surface     setpoint and the    deep    setpoint  proportional  but not linearly so  to current depth  This  dynamic setpoint method helps prevent excessive PO     spikes    during descent  and excessive  oxygen wastage during ascents from no decompression dives     There are two exceptions to the dynamic setpoint method described above  The first is that  whenever a decompression ceiling exists  the setpoint will not drop below 0 9 bar   atm during  ascent  The second involves the Hyperoxic Linearity test on the primary oxygen sensor  as  described below           CHAPTER 3 PAGE 52    Hyperoxic linearity test 
17.  adequate substitute for  training from a qualified Poseidon MkVI instructor  Failure to  obtain proper training prior to using the Poseidon dall could lead    to serious injury or death        Monitoring alarms    The most important responsibility of anyone diving the Poseidon MkVI is to monitor the alarm  systems  There are three separate alarm systems  the Head Up Display  HUD  located on the  mouthpiece   the battery module  located on the main electronics module  behind the diver   s  head   and the Primary Display  Each of these systems is intended to get the attention of the  diver or the diver   s companions through visual  audio  and tactile signals  and convey clear  information to the diver concerning the status of the MkVI     DANGER     NEVER ignore or otherwise discount any of the alarm signals on    the Poseidon MkVI  Failure to respond appropriately to any of the  alarm signals could lead to serious injury or death        HUD vibrator    Perhaps the most important alarm signal on the Poseidon MkVI is a customized version of the  patented Juergensen Marine DIVATM vibrator system  located in the HUD mounted on the  mouthpiece  There are two ways this tactile alarm may be triggered  The first  and by far the  most important  alarm is a continuous pulsing vibration signal On Off On Off   etc  This signal  has one and only one meaning     Change the Mouthpiece Valve position NOW       In most cases  this signal will be triggered in association with an open circuit 
18.  breathing gas and to direct the flow to the gas processing unit in the backpack   One way checkvalves in the mouthpiece direct the exhaled gas such that it  moves from the mouthpiece to the right front  exhalation  breathing hose  into  the right water diversion manifold and into the right counterlung     During normal use  water will sometimes collect in both of the front breathing  hoses  but predominantly it will collect in the right front  exhalation  breathing  hose  The right side water diversion manifold directs the water into the right  counterlung  while the breathing gas continues through the loop towards the  CO  absorbent cartridge  At the bottom of the right counterlung is a variable   tension dump valve that can be used to vent the water periodically during the  course of a dive     The counterlungs  left and right  are each sized to be about half the volume of  a full breath for an average individual  This design     known as a    dual over   the shoulder    counterlungs     optimizes the ease of breathing underwater   Those familiar with open circuit diving will notice an immediate improvement  in diving comfort when using the Poseidon MkVI because of this design     POSEIDON MKVI USER MANUAL    Carbon dioxide absorbent cartridge    At the heart of all reoreathers is the requirement to remove the metabolically generated  carbon dioxide  CO   from the breathing loop and replace the oxygen consumed through  metabolism  The Poseidon MkVI is designed around a mod
19.  closed circuit mouthpiece    One of the most amazing of several technology breakthroughs in the Poseidon MkVI is its  switch able mouthpiece  It incorporates a high performance lightweight open circuit regulator   and you can breathe it just that way  just like standard scuba  With a simple quarter turn of an  easy to operate switch  the system is ready for fully closed circuit  bubble free  silent  depth   independent diving     Automatic diluent addition valve  ADV     The mouthpiece also contains a patented system that integrates an automatic diluent addition  valve  ADV   which compensates for depth related compression of the counterlung breathing  volume during descents  This ensures a full breath automatically  allowing for hands free  descents  The Poseidon MkVI incorporates this into the mouthpiece with a special mechanism  that adjusts the trigger tension on the open circuit second stage when diving in closed circuit  mode  so that gas is only added when the counterlung volume is insufficient to provide a full  breath on inhalation     Figure 1 2   HUD with red LED on        Heads up display  HUD     The mouthpiece also includes a snap in place head up display  HUD   The HUD contains its  own computer processor that communicates with other system processors via the network   and includes both a high intensity red LED to alert the diver of a potential problem  Figure 1 2    and a patented Juergensen Marine vibration system that provides a tactile alarm system to  adv
20.  flash for five seconds  unless the wet switch is  activated   It is important to note BOTH the test number  left side of the display   and the error  code number  right side of the display   because both of these values are needed to identify  the likely cause of the problem and  in some cases  determine the best course of action to  correct the problem        Figure A1 1  Jest 17  Backlight Power  consumption   displaying test number on  the left and the    spinning wheel    on the right     Figure A1 2  Test 17 failure  with  flashing test number and error code     POSEIDON MKVI USER MANUAL    Technically  an error code of    1    means that the test passed successfully  However  this should  never be displayed  because as soon as a test passes  the routine continues on to the next  test  An error code of    O    means that the test did not complete within the alloted time  For  tests requiring action by the user  t43 t45  t50   this usually happens when the action was not  performed within the time allowed  For the other tests  error code    O    is the result of a failure  of a processor to respond in time  such as from a network failure   which in many cases can be  solved with the standard response  as described below     Standard response to test failure    The first thing to do when any of the automatic pre dive tests fail  is make sure that the battery  is adequately charged and that it does not require a learn cycle  A low battery may cause one  or more of the test
21.  followed by a stream of gas bubbles  After  the water has been flushed  the loop vent valve can be tightened by rotating clockwise  and  the breathing loop volume and PO  can be restored to normal    A small amount of condensation may also collect in the inhalation portion of the breathing loop   between the CO  absorbent cartridge and the mouthpiece  Normally  this will only be a small  volume of water  and most will be absorbed by sponge trap     POSEIDON MKVI USER MANUAL CHAPTER 3 PAGE 60       Managing ascents    During an ascent from a rebreather dive  the oxygen partial pressure in the loop will begin to  drop  due to the dropping ambient pressure   The oxygen control system will likely begin to  compensate for this by injecting oxygen  however  during somewhat faster ascents  the  solenoid valve may not be able to keep up with the drop in loop PO  caused by the drop in  ambient pressure  This is not of great concern  unless the PO  gets so low that it triggers alarm  conditions  but it represents one more reason why it   s always good practice to ascend at a  slow and controlled rate     During the ascent  loop gas will be vented from the breathing loop due to expansion  For this  reason  dives involving many ascents and descents  up and down  can lead to excessive loss  of both diluent  during descents  in re filling the breathing loop  and oxygen  during ascents   while trying to maintain the set point      Ending the dive    After surfacing and exiting the water  t
22.  for a dive to be  started  at least 25  of maximum capacity      Sufficient Battery Capacity  This test ensures that the battery  has a sufficient charge to start a dive  The minimum necessary  charge depends on how long it has been  both in terms of time  and number of charge cycles  since the last Learn Cycle  see  Chapter 1   If 160 days have elapsed since the last Learn Cycle   this test will always fail        ERROR GODE    35 Bad Batt   Deco data   36 Bad Rig  Deco data   37 Ser  Num   Mismatch  38 Time  Mismatch    39 No Deco  data    O Timeout    O Timeout    O Timeout    5  Charge  Insufficient    APPENDIX PAGE 75    SOLUTION    The most common cause of failure of this test is inserting a battery from one  user into the rig of another user  In such cases  the decompression data will  not match  The first automatic pre dive routine to encounter this condition will  fail  alerting the diver to the mis matched data  The next time the electronics  are booted  this test should pass  and the system will assume the worst case  set of decompression data among the two disparate sets     1  Standard Response  2  If test failure persists  contact an authorized  Poseidon Service Center for repair     1  Ensure that the moutpiece is fully in the OC position  sometimes requires  firm pressing on the mouthpiece lever   2  Ensure that the HUD is properly  positioned on the top of the mouthpiece  and that the exhaust cover is  holding it tightly in place  3  If test failure persists
23.  if the sponge isn   t completely dry  to minimize the  chance of replacing the wrong CO  absorbent cartridge     After each day of diving    Open the breathing loop    At the end of each diving day  it is important to open up the breathing loop to allow the hoses  and other components to dry overnight  This is  by far  the best procedure for keeping the  inside of the breathing loop clean     All four breathing hoses should be removed from their attachment points  mouthpiece  shoul   der ports  and main housing   and placed such that water inside will drain out  and somewhere  with relatively dry  well circulated air     Remove the shoulder ports from the counterlungs and store them where they will dry  and will  be protected from accidental damage  Remove the counterlungs from the harness and  if  possible  hang them such that water will drain from them through the Shoulder Port connec   tion sockets     Remove the CO  absorbent cartridge and the two sponge water traps  Discard the absorbent  cartridge properly  and squeeze out the sponges and place them where they will be allowed to dry     Store the electronics    After removing the regulators from the oxygen and diluent cylinders  remove the electronics  module and place the entire electronics oneumatics assembly where it will be able to dry  Do  not attempt to disconnect the regulators from the electronics module  or disconnect the  mouthpiece from the supply hose  It is best to keep the entire electronics oneumatics as
24.  obligation  i e   the Decompression  Ceiling Alert is displayed   and the diver then ascends above the depth at which the Decom   pression Stop Alert is displayed  the down arrow will flash  In this situation  simply descend  gradually until the down arrow no longer flashes  and remain at that depth until the Decom   pression Stop Alert no longer displays     POSEIDON MKVI USER MANUAL    Battery life indicator    Near the bottom of the Primary Display  just to  the left of the Elapsed Dive Time value  is the  Battery Life Indicator  This indicator serves as a     fuel gauge    for remaining battery life  If the  remaining battery life is less than 20   this indi   cator will flash  and the screen will indicate that  the dive should be terminated  The more time  that has elapsed since the last battery Learn  Cycle  the greater percentage of battery charge  is needed to ensure 20  remaining power     DANGER        Figure 3 29  Battery life indicator        CHAPTER 3 PAGE 56    Cylinder pressure indicators    Along either side of the Primary Display are the two cylinder pressure indicators  represented  as bar graphs  The graph on the left side of the screen is for the diluent supply  and the graph  on the right side of the screen is for the oxygen supply  Each segment in the bars represents  approximately 10  of the total gas supply for each cylinder  When the pressure in either cylinder  drops below the minimum acceptable value  the remaining segments of the corresponding bar
25.  oxygen clean systems  gas banks  and compressors are already handled by those centers  and all you need to do is present your cylinders for filling     For those with less convenient access to such facilities  there are several practical consider   ations  It may be worthwhile to acquire several spare Poseidon MkVI oxygen cylinders and  have those pre filled for future dives when needed  If you will be visiting a dive resort or live  aboard boat  inquire ahead as to the availability of oxygen and or the availability of pre charged  cylinders on site that are compatible with the MkVI  Note that the MkVI oxygen cylinders sold  in Europe contain a DIN M26x2 thread  This is larger than the G 5 8 DIN thread that is more  standard around the world  Poseidon sells oxygen cleaned converters that will allow filling of  the M26x2 oxygen cylinder from a standard G 5 8 male DIN fitting        POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE 12       Part 2     Assembly    Before you start assembling your Poseidon MkVI eCCR  you should perform the following  preparations   e Make sure you have a scrubber that will last your planned dive s      e Fill your diluent tank with the correct gas   e Fill your oxygen tank with the correct gas     e Make sure your battery is charged and that a learn cycle has been performed  within the last 30 days     e Make sure you have all parts and that they are without damage     e Lubricate all o rings that you can access       Please note that the Besea harness a
26.  rer ee EE 71    POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE IV       Conventions used in this guide    This User   s Guide is NOT intended for use as a training manual  or in any way as a Text  photographs and figures copyright    2008 2011  substitute for proper training through a legitimate training agency approved by  Poseidon AB  It is only intended to provide basic information concerning the Poseidon  pace A ALL RIGHTS RESERVED    Throughout this Guide  special alert boxes have been inserted to draw attention to  critical information  Three levels of alerts are used in conjunction with color coded  triangle symbols  as follows     by Poseidon Diving Systems AB    Manual Version 2 2   December 2011    No part of this book may be reproduced or transmitted in any form or by any  means  electronic or mechanical  including photocopying  recording or by  any information storage retrieval system  without permission in writing from  an authorized representative of Poseidon Diving Systems AB     WARNING    Alert boxes that are YELLOW contain vital information that may  impact a diver   s safety and or proper function of the Poseidon  MkVI  Though generally not life threatening  information contained  in these boxes should not be ignored        POSEIDON MKVI USER MANUAL    Preface    Congratulations on your decision to purchase the Poseidon MkVI Closed Circuit  Rebreather  CCR   The Poseidon MkVI design includes several fundamental improve   ments over previous closed circuit rebrea
27.  routine that automatically  follows power up     Negative Pressure Loop Test    Before powering up the electronics  it is important to check the integrity of the breathing loop   A positive pressure loop test is conducted later on  as part of the normal power up and auto   mated pre dive test procedure  However  it is possible that some leaks in the breathing loop  will fail only when the external ambient pressure exceeds the pressure inside the breathing  loop  and thus will not be detected during a positive pressure loop test   For this reason  it is  important to conduct a manually negative pressure loop test before starting a dive     To conduct a manual negative pressure loop test  first secure the exhalation counterlung over   pressure checkvalve by tightening it inward to its full extent using a clockwise rotation  when  viewed standing in front of the valve and looking at the valve   Place the mouthpiece switch  lever to closed circuit  CC  position and inhale any residual gas within the breathing loop   exhaling it through the nose to remove it from the breathing loop  Repeat this procedure several  times until you have pulled as strong a vacuum on the breathing loop as you can  and then  quickly switch the mouthpiece to OC position to hold the vacuum inside the breathing loop   The breathing hoses will contract until no more breathing gas can be pulled from the loop        CHAPTER 2 PAGE 38    With the mouthpiece in the OC position  observe over a period of a minut
28.  screen to see which value s  is are also  flashing  As long as one of the other displayed  values on the Primary Display is flashing  the  General Alert symbol will also flash     Electronics alert    Located between the DO NOT DIVE symbol and  the General Alert symbol is a small icon that  resembles a personal computer with a lightning  bolt on the screen  This symbol indicates that a  problem has been detected with the electron   ics  such as a network failure  an unexpected  system re boot  or other detected errors  The  specific cause is recorded in the logged data  If  the Electronics alert symbol is shown during a  dive of after the completion of a Pre Dive test   ABORT the dive or DO NOT DIVE     Decompression ceiling alert    In the lower center of the Alarm Signal Area is  the Decompression Ceiling Alert  This symbol  will flash when the diver has incurred a decom   pression obligation  The Poseidon MkVI is not  intended for decompression diving  so the dive  should be terminated whenever this icon is dis   played  The diver should ascend towards the  surface at a slow and controlled rate  watching  the Primary Display for the Stop Alert and addi   tional decompression information  see below      USER MANUAL       Figure 3 9  General alert        Figure 3 10  Electronics alert        Figure 3 11  Decompression ceiling alert     Figure 3 12  Stop alert     CHAPTER 3 PAGE 51    Stop alert    The octagonal shape with a flat palm in the  center  located between the ABOR
29.  shows  the optional Platform backpack installed  but not the optional buoyancy compensator  wing  which is mounted behind the counterlungs     POSEIDON MKVI USER MANUAL       CHAPTER 2 PAGE 37    Chapter 2   Pre dive procedures    The Poseidon MkVI is a compact and very powerful life support system that offers an  unprecedented new experience in recreational diving  But it is also a complex assembly of  high technology that includes sensors  actuators  computers  and software that need to  operate reliably in an underwater environment  for the important purpose of keeping a diver  alive and healthy  For the same reasons that good pilots use pre flight checklists to ensure  their flying machine has a high probability of successful take off  flight  and landing  so does  the rebreather diver need to formalize the process leading up to a dive  The MkVI design team  has gone to extraordinary measures to automate the pre dive procedure and the operation of  the rig during a dive  This chapter explains the pre dive test procedures  including manual  actions that are required by the user  and how to interpret the results of the automated tests   should any of them fail to complete successfully     DANGER     Failure to properly and completely conduct the Pre Dive tests    and to ensure that the rig is operating properly could lead to  permanent injury or death  Do NOT skip the Pre Dive Procedure   Your life depends on it        Initial Pre dive procedures    Gas supply cylinders   
30.  t      see Figure A1 1   When each test is active  a    spinning wheel    is displayed on the right side  of the display  where the remaining dive time is normally displayed  This    spinning wheel    is  represented by a    O    character in the right most position  missing one of the segments  The  missing segment rotates positions around the    0    in a clockwise direction  The purpose of this  symbol is to assure the diver that the test is still actively being conducted  and the system has  not locked up     When a test completes successfully  the next test begins automatically  as represented by the  increasing    t    number on the left side of the Prmary Display  The bar graph located along the  top of the display  normally used as an ascent rate indicator  serves as a progress bar for the  test routine  starting with all segments lit  then eliminating segments from right to left as the  test or set of tests progresses  If a test fails  the test number flashes  and the    spinning wheel     indicator on the right side of the display is replaced by a flashing error code  indicating what  aspect of the test failed  Figure A1 2   This continues for approximately five seconds  after  which the electronics power down  if the wet switch is not activated and the system has not  entered Dive Mode due to exposure to depth      It is important to carefully monitor the automatic pre dive test routine  in case a test fails  Upon  test failure  the test number and error code only
31.  tanks should be mounted    on your left hand side     Make sure that tanks are filled with the correct gases     Left side   Diluent Right side   Oxygen       Figure 1 27  The Diluent  black knob  tank is mounted on the unit   s left hand side and the  oxygen  white knob  is mounted on the unit   s right hand side        CHAPTER 1 PAGE 20    The Poseidon MkVI is equipped with one mounting bracket holding two nylon cylinder straps  with cam buckles on each side  Attachment of the diluent cylinder  black valve handle  is illus   trated in Figures 1 27 and 1 28  After positioning the cylinder with the valve in the correct ori   entation  the strap should be pulled through the cam buckle to snug the cylinder so that it  cannot rotate  Figure 1 28   Maintain the tension on the strap while threading it through the end  slot of the cam buckle  Maintain tension while closing the cam  so as not to allow any slack to  pass back through the buckle  then swing the cam buckle closed  The cylinder should be  rigidly mounted to the gas processor housing once both upper and lower straps have been  oroperly attached and secured     Then repeat these steps on the oxygen cylinder  white valve handle      Cam buckle Strap       Figure 1 28   The tank bands should be tightned hard enough to prevent the tanks from rotating     POSEIDON MKVI USER MANUAL    7  E module    Check and make sure that the two o rigs around the e module are in place and without  damage     Align the e module so that the 
32.  the mouthpiece       ON  Tighten the hose nut oh the downstream    CE pore Sw  threads  DO NOT use tools or over tighten  mE    POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE 29       Inspect the upstream CC checkvalve  It should be free of debris and should lie smooth and  uniformly flat against its removable carrier plate  If there is any curling  cuts  gouges or other  damage to the checkvalve  remove the carrier plate and replace the checkvalve only with a  Poseidon original manufacture mouthpiece checkvalve  Attach the left front breathing hose to  the upstream CC port as shown in Figure 1 45  following the standard hose connection pro   cedures previously described  The breathing loop assembly is now complete     The result should now appear as shown in Figure 1 44     WARNING    Carefully inspect all breathing hose o rings when assembling the  breathing loop  and ensure that the fittings themselves are properly  attached to the hoses  Sometimes the fittings can become loose  and form an imperfect seal  Careless assembly can lead to poor  seals  and increase the risk of water entering the breathing loop  during a dive        DANGER     Ensure that the checkvalves are inserted properly and in the  correct orientation        WARNING    The hoses are designed to function properly in all normal diving  environments  However  if the hoses are exposed to temperatures  in excess of 70  C   158   F  they may become permanently  deformed  and would need to be replaced     POSEIDON MK
33. 1 16  attach the counterlungs to the shoulder straps of the  backpack  The MkVI is provided with a lower D ring and a crotch strap that connect to the  bottoms of each counterlung  Using this system the user can fix the counterlungs as high or  low on the harness straps as desired to reduce the work of breathing        Figure 1 16   Fasten the upper counterlung position with the adjustable length quick connect buckle   Attach the three velcro retainer flaps to the Platform harness straps     POSEIDON MKVI USER MANUAL    4  Rear CC hoses to counter lung    T connections   Connect the T connection to the top port in each counter lung                     Thread connection post       Iw      Figure 1 17   nsert the right hand water diversion  manifold into the port on the right counterlung     Locate the two counterlungs and their respective water diversion manifolds  also known as     shoulder ports    or    T ports     and lay those out for assembly   The purpose of these manifolds  is to prevent water that collects in the counterlungs from reaching the gas processor  The  water diversion manifold at the top of each counterlung diverts water coming from either of the  front breathing hoses into their respective counterlungs  Because of the direction of breathing  flow and the one way check valves in the mouthpiece  almost all water that leaks into the  mouthbit collects in the right counterlung  where it can be dumped via the checkvalve port at  the bottom of the counterlung  F
34. Buddy Light    when activated     HUD Vibrator  This test measures the amount of electrical  Current consumed by the vibrator motor in the Head Up  Display  HUD   when activated     Metabolic O2 Solenoid  1  This test measures the amount of  electrical current consumed by the first metabolic solenoid  valve  when activated  It does not check that the solenoid  actually opens and closes  verified during the Positive Pressure  Loop Test  t32      Metabolic O  Solenoid  2  This test measures the amount of  electrical current consumed by the second metabolic solenoid  valve  when activated  It does not check that the solenoid  actually opens and closes  verified during the Positive Pressure  Loop Test  t32      Oxygen Calibration Solenoid  This test measures the amount  of electrical current consumed by the oxygen calibration  solenoid valve  when activated  It does not check that the  solenoid actually opens and closes  verified during the Positive  Pressure Loop Test  t32      Diluent Calibration Solenoid  This test measures the amount of  electrical current consumed by the diluent calibration solenoid  valve  when activated  It does not check that the solenoid  actually opens and closes  verified during the Positive Pressure  Loop Test  t32         ERROR GODE    11 Current  too low   12 Current  too high    11 Current  too low   12 Current  too high    11 Current  too low   12 Current  too high    11 Current  too low   12 Current  too high    11 Current  too low   12 Current  
35. CHAPTER 3 PAGE 54    Maximum depth   Ceiling    In most circumstances the maximum depth  achieved during the dive is displayed below the  current depth  in the lower left corner of the  Primary Display  to the left of the word    Max      However  in the event that a diver has inadver   tently incurred a decompression obligation  this  value changes to represent the current decom   pression    ceiling     shallowest depth to which it  is safe to ascend   When displaying the ceiling  value  the value briefly changes to    cL     to  indicate    ceiling     every few seconds  as shown  below      POSEIDON MKVI USER MANUAL    Remaining dive tims  RDT     The Remaining Dive Time  RDT  value  shown as the large number on the right side of the  Primary Display  is based on various factors  including the remaining no decompression time  at the current depth  oxygen supply  remaining battery life  and oxygen toxicity units  OTUs    It represents the number of minutes remaining at the current depth before one of these para   meters is exceeded     199    is displayed if more than 199 minutes remain   When the value falls  below 5 minutes  it will flash  If a decompression ceiling is incurred  this value changes to  represent the total decompression time     ascent time plus decompression stop s      When displaying this value  the value briefly changes to    td        total decompression     every few  seconds  as shown on the previous page            CH  bd    Figure 3 25  Remaining 
36. Diluent Addition Valve     ADV  that compensates for breathing  volume in closed circuit mode  so that you will never run short of a full breath  during a dive     Easy maintenance     the entire rebreather breaks down rapidly for washing   drying  and storage  No tools needed     New software releases can be downloaded and installed over the internet  and  various system parameters can be customized to meet your particular dive  operating paradigm        r  SEM o te a        J e F P w J  d i p  H T    8   a  bg        Fa rii  E  A Wee 7  r i ca  i i      L L td  T J e     E    i        K       DU  Ch  Ke       POSEIDON MKVI USER MANUAL    Conformance with CE requirements    The information listed in this chapter are the CE requirements that has to be met to  get a CE approval and they are NOT the exact specifications of the Poseidon MkVI   The exact specifications and values of the MkVI can be found in the later chapters of this  user guide  The below text shows that the MkVI operates within these CE  requirements     In conformance with the European Standard EN 14143  section 8  the following  information is provided herein     8 1    This manual contains information that will enable trained and qualified persons to  assemble and use the Poseidon MkVI in a safe manner     8 2    This manual is written in the English language     8 3    The application of the Poseidon MkVI is as a diving apparatus  to be used for  recreational  no decompression diving with mixtures of air an
37. E 57  Monitoring gas Tee E or  Mo  nitormg remaining Welle EE St   Breathing underwater              sssssesrsrsrsrerrrrrrsrsreraranrsrrrsrarrriersrrrnrirnsrsrtreranrrrsrsrernrrni 58  COU NURO P ICONEN sea carat aa E E 58  SEELEN srap AG IU SU GIS Eeer 58  WOORD E 59  Me keane to BGC ONO P E E 59  Venting water from the Joop  59  EE 60  EnGING NO CUS sc satyhesecer ey acge eet gta aee aier bance anacepneeieces 60  odie OMNO WINE OSS MKV RE 61    Chapter 4   Post dive care and maintenance    AOF O OA EE 62  FOW le 62  Replacing the Oxygen  and CO  absorbent Ga  eesiriisier siiin 62  Removing the electronics ane el 62    POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE Ill       Replacnd the water Tal SOONG CS anien ena E EAA AO 63  Altereach day of a le e E 63  DS NS ON eT NOs ee 63  OO NS SN COGS eae SE 63  Long term storage and TE 63  EE eebe 63  Replacing Oxygen EE 64  Traveling with the Poseidon MV     66  Beidel reel 66    Appendix    roublesnooting guide    ENEE EE A th ti i ts sn ne ee 67  Standard response tO test FANUC 0    ececcecsecee eee ee eee eeeeeeeeeaeeeeaee sense eeeeeeteeeenes 68  WONG STN OTIC ee 68   FCI DO UOS eree ier EE E ERE 68  Ie ebie EE EE RE E A E E E E EE 69  EO O E a E E nimaneecoageetdant sectors 69  Depth reading difference           s sssssrsrsrsrrrrrrersrsrnrrrrsrsrsrerarirrsrsrrrrrnersrsrsrernrrne  69  CE en Re A E A E E E E 69  Hyperoxic linearity Test    70  FOW EPO alarm WOK E 70  What to do if   can   t resolve the problem    70    lable oi the E
38. ER MANUAL    INGO anh e Fressie LOOD TOS EE 38  Bled teller ee UD EEN 39  Power up Self test  test 1   D      40  Tesue tension 16S eer Eege 41  Open circuit mouthpiece position  test A7   42  Oxygen and diluent cylinder supplies  test 44  amp  Af    42  Battery power verification  test A9     43  Positive pressure loop test  test AO    43  Closed circuit mouthpiece position  test DI    44  Oxygen sensor calibration  test 3  45  Open Circuit regulator function test E 45  Service interval check  test D      46  NM TO EE 46    Chapter 3   Dive procedures    MONORI UA EE 47  EIER Ste EE A7  RIIT E 48  A Seea e A e E E E E A E E vn aiaaitde tam pacanei assur et iseinmen 48  Ruddy alert IOM EE 48   POPP OTIC breede et Ee 48  Blat ene LE 50  eea E i E A en ATTE N E E T ETE ATE 50  Abon  and ODEN crout alor GE Sab   NT OVE AG GE 50  ER EE 51  EICH ONCS EE 51  Decompression er iere rai iasnatienceuemotctaeanandieaiesaueeeuanamed ourencaeeoeeennseieeyets 51  BO Eeer 51  OMe E ee ee ere ree 51    BOR e te ee eg       CHAPTER 1 PAGE Il   EE EH icc  eet 52  OXYGEN Sensor EENEG  e genee 53  MOUNDIGCS DOSI E TE  EEEE E EE 54  Maximum depth   Geill EE 54   Pe ReMi tC  GVE UMAR D T orren E eeanauaiyeaen 55  EPP OS ln EE 99  Ascend descenqd amow uu    cccecceceececcece uence eee eee eee eeeee eee eseeeeee ese eeeeeeeaeeaeeesnentegeas 50  Batory iNo NN CLO UC erisia EE AEE EREEREER 56   emperature geed EEEE rra   56  Cynder pressure e ege E 56  Ascent rate nelle er EE 56  SSN EE en  MORRO TME 
39. I will successfully complete all Pre Dive checks  in about three minutes  and a screen will appear on the display that looks like that shown in  Figure 2 13  with a PO  value between 0 3 and 0 9  usually the number is towards the higher  end of this range because the loop is partially filled with oxygen during Tests 49 and 53   a  depth of O  a dive time of O  and a remaining dive time clock showing 199 minutes     At this point the pre dive tests are  completed  and the system is ready  to dive        Figure 2 13  Clear to dive        CHAPTER 2 PAGE 46             Test  1 38 55  AO  43 50 54    44 45  48  49  53    2  S  A    Pre Dive Checklist    Cylinders OFF  CLOSE OPV   OPEN CIRCUIT mode     Touch wet switch  keep dry for 5 sec   onds  then HOLD wet switch     Test 44  turn cylinders ON   Test 50  CLOSED CIRCUIT     Test 54  OPEN CIRCUIT  test bailout  regulator function     Confirm and Restart   Needs Servicing  if persistent   Use correct battery   Verify mouthpiece position   Turn cylinders on   Refill cylinders  Recharge Battery   Replace battery  Counterlungs half full or less   Verify Oxygen Sensors       Post Dive Checklist  Mouthpiece in OPEN CIRCUIT     Both cylinders OFF   Wet switch DRY   Purge ADV        Figure 2 14  General Pre Dive and Post Dive Checklists           POSEIDON MKVI USER MANUAL       CHAPTER 3 PAGE 47    Chapter 3     Dive procedures    DANGER    Do NOT attempt to use the Poseidon MkVI rebreather without  proper training  This Manual is NOT an
40. POSEIDON MKVI USER MANUAL       POSEIDON MKVI USER MANUAL       Table of Contents    TONS Oi ER     CONV ens U Sea in TSUN Eege IV   Bier V   Conformance With CE Heourements     vi   Chapter 1   Preperation  amp  Assembly   Part 1   Preperation   An Overview of the Poseidon MV  1  PET D EE    Open Circuit   Closed CirCuit Mouthpiece          cc ccecececcecneeete teeta eeeeeeaeeeeaeeseaee ees 2  Automatic diluent adaton valve  AD EE 2  Heads up dopa  FUD sie tt ht tt ee a ee 2  Breathing IO OO ee 3  Carbon dioxide absorbent cartridge           cccecceceeceeceeeeeeeceeeeseeeeeeeeaeeeeeeeaeeanenes 4  Gas injection nee UE 4  FS IS MOTU E 4  EE EE 4   UE EE      ENEE E E E E 5  EE eene 6  Cone teure Le e A E E O A A A A A TE 7  Decompression Cale ernir aE a aaa D  Beie Kn E EE 8   OF lees eessen ere 8    CHAPTER 1 PAGE     KEE anad WOON eegener 10   mlllele breede e EE 11  Part 2   Assembly  Eer ie 7 BOD 7 ee ME 13  Ze  MLPA ee Le EE 14  EE ngs KEE Tales  Mee 15  a  Rear CO hoses to elle ET e e DEE 16  D Rea OC TOOD  ere 18  6  Attaching th   Cylinders  EE 20   Gre Gg elei EEN 21  CO WS Le ET 23  Del OSS Oe PUD TO MOUNDO CE  E 20  1O dJe ream EE 26  11  Front CC hoses to counter lung T section            s srsrsirirrrrerrrrrrrrrrsrerersrsrrrersrne SE  12 PION OO ee e TEE 28  EE tee 20  SEENEN 36  Chapter 2   Pre dive procedures  Initial Pre dive procedures   EIERE  erter geegent 37   AA ABSO DERT cl  nn er er i ee ee of    intact Breathing LOOD Ver MCaI OM E 38    POSEIDON MKVI US
41. PTER 3 PAGE 57    Monitoring remaining dive time    As mentioned previously  the Remaining Dive Time  RDT  value is based on several different  factors  The value displayed represents the amount of remaining time  in minutes  for the most  limiting factor  If the limiting factor is remaining battery life  the value will count down consis   tently  regardless of depth  However  if the limiting factor is remaining oxygen supply  the value  could increase or decrease depending on the rate at which the diver is consuming oxygen   The value can change even more dramatically  and suddenly  when the limit is based on  remaining no decompression time  This is because a diver with only a few minutes remaining  at a depth of 30 meters  for example  may well have many more minutes remaining at a shal   lower depth  Conversely  the remaining minutes may suddenly decrease sharply when depth  increases  Thus  it   s extremely important to monitor this value throughout the dive  particularly  after increases in depth     Note that the RDT value is NOT an exact value  and it should be regarded as a    recom   mended    remaining dive time  rather then an absolute remaining dive time  In the event that a  diver inadvertently exceeds the no decompression limits and the dive requires decompression  stop s   the RDT value changes to display the remaining total decompression time  as  described previously     POSEIDON MKVI USER MANUAL    Breathing underwater    Counterlung placement    When prope
42. Poseidon MkVI is capable of displaying parameter values in either metric or imperial units   Both screens at the top of the next page show the same information  except that the left  screen shows the depth and temperature values in imperial units  and the right screen shows  the values in metric units  Depth units are indicated by an    FT    or    M     and temperature units  are indicated by a   F or   C  Also  the Display can be configured to represent decimal points as    HD HD    a period            or as a comma            depending on individual user preference     Figure 3 4   mperial units  Figure 3 5  Metric units        What follows is amore detailed description of each of the LCD screen elements  and what they  mean  It is important that all Poseidon MkVI divers become familiar with these symbols and  values  what they mean  and how to respond when they are not displaying appropriate values   or are flashing      Alarm signal area       CHAPTER 3 PAGE 50    Abort  and open circuit alerts    The most important alert symbols on the screen are also the largest  The ABORT  and Open   Circuit symbols  The ABORT  symbol is a large word ABORTI in inverted font color  Whenever  this is displayed  the dive should be immediately terminated  If accompanied by the Open   Circuit Alert icon  image of a diver   s mask  second stage regulator  a series of bubbles on  either side of the diver   s face  and a small up arrow above the diver   s mask   the diver must  immediately te
43. T symbol  and the DO NOT DIVE symbol in the center of  the upper half of the Alarm Signal Area  is dis   played in one of two circumstances  either the  diver is ascending too rapidly  or the diver has  reached the decompression stop depth     ceiling       In either case  the appropriate response is to  immediately stop ascending  and the diver should  maintain the current depth until the symbol  disappears           Figure 3 13  PO  value     PO  value    The oxygen partial pressure  PO   in the breath   ing loop is displayed prominently in the upper  left corner of the Primary Display  This is perhaps  the most important number on the entire screen   as maintaining an appropriate oxygen partial  pressure in the breathing gas is critical to ensure  safe diving  If the value departs substantially from  the current PO  setpoint  the value will flash  If the  value becomes dangeously high or dangerously  low  the diver will be prompted to switch to open   circuit mode and terminate the dive     POSEIDON MKVI USER MANUAL    PO  setpoint    Every few seconds  the PO  value will briefly  less than one second  switch to show the current  PO  Setpoint  Normally  this value will be the same as the current PO   because the system will  normally maintain the correct PO   e  Setpoint PO    In some cases  however  the value may  be slightly different  In any case  the PO  Setpoint value can be distinguished from the current  PO  value by the size of the first digit  either    1    or   
44. VI USER MANUAL    13  Scrubber    The largest physical element of the rebreather is the CO  absorbent system  also referred to as  the    Gas Processor    unit  it physically includes the gas sensing and electronics control modules    The outer shell of this system  shown in Figure 1 46  comprises the structural backbone of the  Poseidon MkVI and both the diluent and oxygen cylinders attach to the side of this extruded  aluminium tube  The top section of the housing is the mounting structure for the electronics  module  The main tube is the housing for the CO  absorbent cartridge  The base plate seals  the housing and also allows for modular loading of the CO  absorbent cartridge     Because CO  production rate is closely correlated with oxygen consumption  the Poseidon  MkVI was designed such that the CO  absorbent cartridge duration corresponds to the oxygen  cylinder capacity  Thus  the absorbent cartridge MUST be replaced whenever the oxygen  cylinder is refilled  To remove the spent cartridge  loosen all four  4  large hand nuts at the         Figure 1 46   Overview of the cartridge house and the CO  absorbent cartridge        Outer shell   Cartridge housing    CHAPTER 1 PAGE 30    bottom of the cartridge housing to the point where they spin freely  no tools are required  see  Figure 1 48      The canister has been tested for a duration of 180 minutes at 40m  water temp 4  C  anda  breathing rate of 40lom producing 1 6   CO  per minute at STPD  Standard Temperature  amp 
45. advise an immediate abort to  the surface in open circuit mode  Always pay particular attention  to the instructions regarding o ring installation and o ring surface  inspection        POSEIDON MKVI USER MANUAL    8  1st stages    Connect the 16 cm Low Pressure hose to a IP port on the diluent 1ststage and to the 9 16     connection on the pneumatics block marked DIL     Make sure the connection o ring is in place and without damages  See Figure 1 32     Figure 1 38 illustrate the assembly of the diluent cylinder with the gas processor stack  analo   gous to that of attaching the oxygen cylinder  Always inspect cylinder and regulator threads  and the regulator o ring before assembly  If the regulator o ring is damaged  gouged  scratched  or otherwise worn or cut  replace it with an appropriate Poseidon oxygen cleaned o ring        Figure 1 32   16 cm Low Pressure hose   s  to the pneumatics block        PAGE 23    CHAPTER 1       N we  Figure 1 33  Align the diluent regulator with  the diluent cylinder female DIN thread  Push  it in and hand tighten the diluent regulator  swivel nut     Figure 1 34  Align the oxygen regulator with  the oxygen cylinder female DIN thread  Push  it in and hand tighten the oxygen swivel nut     Connect the oxygen clean  white marking  16 cm Low Pressure hose to a IP port on the  oxygen clean 1st stage and to the 9 16    connection on the pneumatics block marked O   See  Figure 1 32  Make sure the connection o ring is in place and without damages 
46. aircraft  Before removing the valves from the cylinders   it is necessary to completely drain the cylinders of any gas pressure  If the cylinders are full  or  the valves opened such that the cylinders drain quickly  the metal of the cylinders and valves  will become cold  and produce beads of moisture  condensation   It   s important that this  moisture not be allowed to get inside the cylinder  so always allow the cylinders to warm back  up to room temperature  and wipe off any remaining moisture before attempting to remove the  valves from the cylinders     Removing valves from cylinders can sometimes be tricky  DO NOT attempt to use tools such  as hammers  mallets  wrenches  pliers  or other such devices to remove the valves  unless you  know exactly what you are doing  It is highly recommended that the cylinders be taken to a  qualified dive shop or service center to have the valves removed  The same is true when  replacing the cylinders at the dive destination  or upon returning from a trip  AS soon as the  valves are removed  be sure to insert an appropriate plastic plug into the threaded opening of  the cylinder  to prevent dirt  moisture  and other contaminants from entering the cylinders     DANGER     The cylinders have been specially cleaned for use with high   pressure oxygen  Allowing contaminants to enter the cylinders  risks fire and explosion  and could lead to serious injury or death        POSEIDON MKVI USER MANUAL       APPENDIX PAGE 67    Appendix   Troub
47. al rule of thumb  1 minute on the  battery charger in standard charge cycle mode will load 10 minutes of charge into the battery   Thus  if you charge while taking a 30 minute break between dives  you will have added 5 hours  of dive time to the battery     Leaving the Smart Battery in the Charger  Although it is acceptable to leave the smart  battery in the charger when not in use  it is recommended that the battery be left attached to  the Poseidon MkVI after a successful charge for the following reasons    e  f power to the charger fails at any time  then having the battery in the charger will actually  deplete the cell     approximately as fast as if the battery was installed in the rig and the rig  powered up    e Storing the battery in the Poseidon MkVI enables the depth sensor and the wet switch on  the back of the primary display  If someone wearing the MkVI accidentally falls into the  water  the system will automatically power up the rig  enhancing the probability for survival  of the user  This is only possible if the battery is charged and stored in the rig    e Storing the battery in the Poseidon MkVI reduces the probability of debris entry and impact  damage to the battery contacts in the electronic module        CHAPTER 1 PAGE 7       Long term storage    Allowing the smart battery to sit for long periods of time on shelf without recharge will lead to  oremature failure of the battery  The best storage method  if the battery will not be used for a  substantial 
48. ars from  the Display  the letters    CC    flash in the upper left corner of the LCD  the HUD vibrates  the  HUD and battery LEDs flash  and the audio speaker sounds  All of this indicates that the  mouthpiece should be placed in the Closed Circuit  CC  position  Ae soon as the mouthpiece  is in the Closed Circuit position and is detected by the HUD  the test will pass  The system  allows 2 minutes for this test to be completed  before timing out     POSEIDON MKVI USER MANUAL    As with the Open Circuit mouthpiece position  Test 43   if the mouthpiece appears to be in the  closed circuit position  but Test 50 does not pass  then make sure that the mouthpiece switch  is fully in the CC position  If the test still will not pass  then make sure the HUD is positioned  correctly in the mouthpiece  and is not twisted or otherwise adjar from its proper position  If no  amount of repositioning of the mouthpiece lever or HUD allows the system to pass Test 50   then contact an authorized Poseidon Service Center     Figure 2 10   Test 50     Open Circuit mouthpiece position           Oxygen sensor calibration  test 53     Test 53  Figure 2 11  calibrates the oxygen sensors  Part of this test is to ensure that the  oxygen supply is really oxygen  and that the diluent supply is really air  The system will start by  injecting pure oxygen directly on the primary oxygen sensor for 20 continuous seconds   thereby flooding the entire oxygen sensor chamber with enough oxygen to also calibrate 
49. as supply pressures     DANGER     Always open the oxygen cylinder valve slowly  Rapid pressurization  increases risk of fire  Opening the valve slowly reduces this risk   Carefully maintaining clean oxygen regulators  cylinders  and valves  before  during  and after your dives will further reduce this risk           CHAPTER 2 PAGE 43    Battery power verification  test 48     Immediately after passing the two gas pressure tests  the pre dive routine tests whether there  is Sufficient battery power to begin a dive  Test 48   The amount of power required depends on  how recently the battery was subjected to a Learn Cycle during charging  see Chapter 1   If  the Learn Cycle occurred recently  then the system is able to predict the remaining battery life  relatively accurately  and Test 48 will pass if the battery has at least 20  charge remaining   approximately 5 6 hours of typical dive time  or 4 hours of night diving time   The amount of  charge required to pass this test increases by 0 5  per day since the last Learn Cycle  such  that after 160 days with no Learn Cycle  Test 48 will not pass     This test will pass or fail immediately  If it fails  the only remedies are to re charge the battery   and or subject it to a Learn Cycle   or replace the battery with another one with greater charge   subject to decompression data discrepancies  as discussed previously for Test 40      Positive pressure loop test  test 49     One of the most basic pre dive tests for any rebreather 
50. ated   There should not be much resistance when inserting the sensor base into the electronics  module  If excessive resistance is evident  inspect the o rings to make sure they are seating  oroperly  and make sure the electrical wires are not pinched between the sensor and the sur   rounding walls     Flat edge of the sensor base faces     i  towards the top of the electronics   module    Figure 4 6     Insert the new oxygen sensor mio  fe   electronics module  with the flat edge     of the sensor base facing upwards        DANGER     The oxygen sensors are the most important components of any    rebreather  Handle them with care  and make sure the electrical  connections are clean and properly attached           CHAPTER 4 PAGE 66    Traveling with the Poseidon MkVI    Many people conduct most of their diving activities at destinations far from home  As such  it   s  entirely likely that Poseidon MkVI owners will want to travel with their rebreathers to far off  destinations  Indeed  a great deal of effort went into the design and development of the MkVI  to ensure that it was lightweight and easy to travel with     Preparing the cylinders    There are strict laws concerning the transport of pressurized gas cylinders on aircraft  and different  airlines will have different policies to assure compliance with these laws  At a minimum  most  airlines require that cylinder valves be removed  and that the cylinders themselves be available  for inspection prior to loading onto an 
51. bailout situation   thereby instructing the diver to switch the mouthpiece from closed circuit mode to open circuit  mode  Once the mouthpiece has been properly switched  the vibrator signal will stop   Occasionally  this signal will be triggered when the system is unable to detect the position of  the mouthpiece  perhaps because it is not completely set in one position or the other  open or  closed   If the HUD vibrator signal continues even after switching the mouthpiece  first make  sure the mouthpiece is completely switched to the new position  If the vibration continues   then switch the mouthpiece back to its original position  again making sure it is completely  rotated  If the HUD vibrator signal persists  then terminate the dive immediately in open circuit  mode    In rare cases  the vibrator is intended to prompt the diver to switch from OC position back to  CC position  This will only occur when the diluent supply is low  and the loop PO  is known to  be safe  The important thing is to adjust the mouthpiece position whenever it vibrates    The other HUD vibrator signal consists of a short  1 2 second     blip    that is triggered every 2  minutes whenever the RED HUD LED is flashing  see below   as an alert to view the Primary  Display  Do NOT change the mouthpiece position in response to a short  singular    blip    of the  HUD vibrator     POSEIDON MKVI USER MANUAL       HUD Light    The HUD incorporates a RED LED light  designed to signal that there is a possibl
52. be filled with oxygen  containing less than 0 4   impurities     The Poseidon MkVI may only be used with specially designed Sofnodive   797 pre   packed cartridges manufactured by Molecular Products     Only accessories and or other personal protective equipment specifically authorized  by Poseidon Diving Systems may be used with the Poseidon MkVI  All other third   party additions or modifications are not covered within the intended usage of this  equipment     8 5    The Poseidon MkVI is designed to extend the duration of recreational dives     POSEIDON MKVI USER MANUAL    Chapter 1     Preparation and Assembly    Part 1     Preparation    This chapter describes the steps to assemble and prepare the Poseidon MkVI  for diving  The MkVI is a modular device with several key systems  Each of  these systems is described in a sequence that naturally follows the way one  would service the rig     An overview of the Poseidon MV    Throughout this manual  the terms    left        right        front     and    back    refer to  specific areas of the Poseidon MkVI  Figures 1 1 and 1 4 illustrate these loca   tions and the main systems of the Poseidon MkVI  The    left    side of the rig  corresponds to a diver   s left side when wearing the rig normally  the    right     side of the rig corresponds to the right side of the diver when wearing the rig  normally  The    front    of the MkVI is the location furthest in front of a diver   s  chest when wearing the rig normally  the    back
53. cable junkction box point in the direction of the two canister  housing top breathing loop interface     Gently push the e module in place and tighten the two screws  securing the e module           Diluent cylinder high    CHAPTER 1 PAGE 21    At the heart of the Poseidon MkVI is an electronic  pneumatic  control  and user feedback  system  The electronics module  shown at the center of Figure 1 29  contains the primary  backpack computer system  the smart battery  and the pneumatics control block assembled  in one compact plug and play block     The electronics module contains its own processor  connected via a network to the processors  in the primary display  battery module  and the HUD  head up display  in the mouthpiece   Pneumatic connections to the oxygen and diluent regulators allow for PO  control and oxygen  sensor calibration and validation  This entire electronics system comes pre assembled from  Poseidon when you purchase the Poseidon MkVI  Several of these sub systems will be dis   cussed in detail later     To begin assembly of the electronics module into the gas processor backpack  set the car   tridge housing upright and on a solid flat base as shown in Figure 1 29        Primary display    Diluent regulator  Oxygen regulator    Mouthpiece    pressure Sensor       Electronics module SS CAM  Me      Oxygen cylinder high    ee A pressure sensor    Figure 1 29   nspect the Electronics Module and Related Components     POSEIDON MKVI USER MANUAL    Inspect the in
54. ch new cartridge is good for at least 3 hours of diving   some users may obtain greater range depending on metabolic oxygen consumption rates    Once you open the air tight shipping envelope the SofnoDive   797 cartridge is activated        Figure 1 54   Unpacking a fresh SofnoDive 797 CO   absorbent cartridge     Opening the sealed SofnoDive 797  air tight shipping envelope    POSEIDON MKVI USER MANUAL    Radial o ring seat to bottom y  of SofnoDive   797 Cartridge          Top    end of cartridge  has projecting ribs     Twin radial o ring  seals for cartridge  housing end cap    Figure 1 56    Loading the Cartridge End Cap into the  new replacement SofnoDive   797  cartridge     Figure 1 57    Loading the Poseidon MkVI Cartridge  Top Cap into the replacement Sofno   Dive   797 cartridge    Inspect the two larger radial o rings on the cartridge end cap  Figure 1 56   and if either is  damaged  cut  or gouged  replace it  the third o ring that mates with the cartridge does not  form a seal   Ensure that each o ring is lubricated and that there is no debris  dust  sand  etc   on any of the o rings  Insert a dry  clean  disinfected annular sponge into the receiver pocket  for the cartridge end cap water trap  this is the reverse of the procedure shown in Figures 1 52  and 1 53   Be sure the inner diameter of the sponge is locked under the retainer plate  see  Figure 1 52      Insert the end cap into the bottom of the SofnoDive   797 cartridge  The    bottom    end is the one  
55. counterlungs should  be no more than half inflated     IMPORTANT     Make sure the over pressure relief valve on the bottom of the right   exhale  counterlung is adjusted to the maximum cracking pressure   turned all the way in the clockwise direction   Also ensure that  the mouthpiece is in the OC position  that the oxygen cylinder  valve is turned on  and the counterlungs are not already inflated   Otherwise  Test 49 will fail  Make sure that nothing pushes against  the OPV  that will cause the T49 to fail     The sequence of events for Test 49  and the various corresponding tests that are conducted   are as follows  First  the system injects oxygen into the breathing loop until the counterlungs  are both full  but not tight  detected by the depth sensor as a slight pressure increase   This is  why it   s important that the counterlungs not already be fully inflated prior to starting Test 49   which can occur if the cylinders are turned on during tests 24 27   This initial inflation is done  via one of the two metabolic oxygen solenoid valves  thereby ensuring that this solenoid valve  is actually injecting gas when it is Supposed to     Figure 2 7    Test 43  Mouthpiece in OC position    displaying Open Circuit icon to indicate  that the mouthpiece must be placed in the  Open Circuit position           CHAPTER 2 PAGE 44    Once the counterlungs are fully inflated and the depth sensor detects a slight increase in  oressure  the system pauses and monitors internal loop press
56. d not be started     If a test fails  the routine is halted at the failed test  and the value displayed on the right side of  the screen changes from a countdown timer to an error code  see Figure 2 6   Both values  the test  number and the error code  flash for five seconds  before the rig powers down automatically  If  the wet switch is connected  e  wet   then the rig will not power down  and the system will  inject excessive oxygen intermittently  in case a diver is breathing on the loop    When one of the PSTs fails  consult Appendix 1 to understand further what the failed test  means  In most cases  the first thing to try is to run the automatic pre dive routine again by  activating the wet switch on the back of the Primary Display  If the same test fails again with the       CHAPTER 2 PAGE 41    same error code  wait for the electronics to power down  then eject and re seat the battery   see Chapter 1   If the automatic pre dive routine persistently fails any of these tests  contact a  Poseidon Tech Center for assistance  In general  repeated failure of any one of these automated  tests indicates a problem with the Poseidon MkVI that will not be serviceable by the user     DANGER     Do not attempt to conduct a dive until the system has successfully  completed all of the pre dive tests  Diving in spite of a failed  pre dive test is extremely dangerous  and could lead to serious  injury or death        Tissue tension test  test 40     As discussed in Chapter 1  the Pose
57. d on a full inhalation  then there is not enough gas in the breathing loop  This condition  should be corrected automatically by the ADV     Tips on buoyancy control    Controlling buoyancy while diving with a rebreather is considerably different from buoyancy  control with conventional open circuit scuba  To begin with  whereas a scuba diver needs to  manage buoyancy characteristics of two separate factors  the Buoyancy Control Device   BCD   and the exposure suit  i e   a wetsuit or a dry suit   A rebreather diver must manage  both of these  as well as the breathing loop of the rebreather  A complete discussion of  buoyancy control with closed circuit rebreathers is beyond the scope of this Manual  However   the following tips might be useful     Although most divers probably do not realize it  fine trim for diving with conventional scuba  gear is achieved through breathing  On each inhalation  the diver   s lungs expand and buoyancy  is increased  The opposite occurs on exhalation  However  this does not occur with a rebreather   the Poseidon MkVI included   because the buoyancy increase caused by expanding the lungs  on an inhaled breath is offset by the decreasing volume of the counterlungs  and vice versa    This may at first be disconcerting for an experienced scuba diver trying a rebreather for the  first time  because an inhalation done subconsciously to slightly increase buoyancy has no  effect  However  with practice  it becomes advantageous to be able to hover in 
58. d oxygen     The Poseidon MkVI is certified to a maximum operating depth of 40 meters  130 feet      Two gas supply mixtures are used with the MkVI  air and oxygen   gt 92  pure   and  the maximum depth for the resultant breathing mixture blended by the Poseidon  MkVI is 40 meters  130 feet      Use of the MkVI is limited to diving underwater  solely by persons who have received  proper training  and for use on no decompression dives in environments without  obstruction between the diver and the surface     Detailed instructions on assembly of the Poseidon MkVI  including descriptions of the  individual components  the specific connections between components  and the  various safety devices  are included within Chapters 1 and 2 of this manual     The user shall be able to understand the risk and make an assesment regarding the  risk of using the Poseidon MkVI  with input from the manual before a dive if the diver  believes it is needed        CHAPTER 1 PAGE VI    The operating temperature for the Poseidon MkVI is between a minimum of 4   Celsius   89   Farenheit   and a maximum of 35   Celsius  95   Farenheit   Operation at temperatures  outside of this range may lead to unreliable function     The Poseidon MkVI is intended for use on dives involving low to moderate work rates   typical of normal recreational diving activities  Although it is capable of sustaining  divers operating with high work rates  this is not its intended purpose     The Poseidon MkVI is intended to mai
59. d the CO  absorbent cartridge  and will  generally collect in the exhalation  right hand  counter lung  Sometimes  water will collect in the  exhalation hose  immediately downstream of the mouthpiece  If this water is sufficient to cause  gurgling noises with each breath  it can be poured into the exhalation counter lung by looking  upward and holding the hose in such a way so as to dump the water towards the right hand  shoulder port  In most cases  the water that collects inside the exhalation counter lung will not  disrupt the function of the Poseidon MkVI in any way  so it can be safely ignored  However   sufficient quantities of water could be returned to the breathing loop if the diver becomes  inverted  so it may be desirable to vent this water from the breathing loop altogether     To do this  the diver should first become negatively buoyant  or attach to a secure object on  the bottom  The breathing loop volume should be increased to at least 75  of maximum  capacity by manually adding diluent via the ADV  The loop vent valve at the bottom of the  exhalation counter lung should be rotated counterclockwise maximally to minimize the cracking  oressure  While in an upright orientation  the diver should then compress both counter lungs  by squeezing them against the chest with the elbows and upper arms  while simultaneously  exhaling through the mouth and depressing the loop vent valve to open it  If done correctly   water will be expelled from the loop vent valve first 
60. decompression  information is corrupted  e g  from inadvertant electrostatic  discharge  there is a possibility that the computer system may  only recognize the rebreather system   s stored decompression  data  In that event  and if the previous diver of that rebreather  did not incur as serious a decompression debt  then swapping  batteries could lead to serious injury or death from incorrect  decompression on subsequent dives           CHAPTER 1 PAGE 8    WARNING     If a user changes batteries with a Poseidon MkVI unit other  than the one they were most recently diving and then turns  on the power of the new Poseidon MkVI with their original  battery installed  the Pre Dive test routine will FAIL Test 40   decompression comparison between battery and backpack  computers   This is a warning that there is a difference in the  decompression data stored in the rebreather   s computer and the  battery just inserted  Once the display times out and goes blank   the system can be re started  and the Pre Dive test will pass  on this second attempt  The user assumes all responsibility for  their own decompression safety in this event  The rebreather will  calculate decompression based on the most conservative values  from each set of decompression data     Dive log data    The Poseidon MkVI automatically creates an extensive dive log every time the system is pow   ered up  The information stored in this log will be of significant interest in reconstructing dives  and learning about 
61. dive care and maintenance    Proper post dive procedures are important for any rebreather  and the Poseidon MkVI is no  exception  Such procedures not only ensure that the system will work correctly on the next  dive  but will also extend the functional life of the unit  This chapter is divided into four main  sections  including information on care and maintenance that should be done following every  dive  steps that should be done at the end of each diving day  long term care and storage  and  information concerning travelling with the rebreather        After each dive    The extended dive durations possible with the Poseidon MkVI will likely exceed the amount of  time most divers will want to spend on any one dive  As a consequence  in many cases it is  likely that divers will conduct more than one dive in a single day     Power down    After each dive  if the next dive will not occur within a few minutes  it   s important to follow the  steps listed at the end of Chapter 3 to shut the power down on the electronics system  Failure  to do so will not cause any risk to the diver or the Poseidon MkVI itself  but it will lead to  unnecessary battery consumption  thereby requiring re charging sooner than would otherwise  be the case     Replacing the oxygen and CO  absorbent cartridge    If the remaining oxygen supply is insufficient for a second dive and the cylinder needs to be  refilled  then it is imperative that the CO2 absorbent canister be replaced at the same time   Thi
62. dive time  RDT     Go d af e a Cn f Je oO                             CHAPTER 3 PAGE 55    Elapsed dive time    The number of minutes that have elapsed  during the dive  i e   the total dive time  is dis   played in the lower right corner of the Primary  Display  next to the small clock symbol printed  on the LCD glass  This value represents the  total elapsed time since the start of the dive  It  begins incrementing only when a dive has  started  and stops incrementing when the dive  ends  If a subsequent dive is conducted without  allowing the unit to power down  then the  elapsed dive time resets         D  00  Ae O     Figure 3 26  Elapsed dive time           Ascend descend arrow    Located in the center of the Primary Display  between the Current Depth value and the Remaining  Dive Time value  is a symbol that can display an up arrow  or a down arrow  When the up arrow  is displayed  the diver should immediately begin a safe  controlled ascent  The up arrow does  not necessarily mean that the dive must be terminated     it may only indicate that the diver is  approaching the no decompression limit at the current depth  in which case ascending a  certain amount may cause the up arrow to stop flashing  e  when the depth is shallow  enough that the diver has ample remaining no decompression time at the current depth          it bei  LEO    Figure 3 27  Ascend arrow        J   emm Lion    Figure 3 28  Descend arrow    In the unlikely event that a diver incurs a decompression
63. don Service Center for repair    RAM is tested only when the battery is inserted  and the   results used for all subsequent power up routines  Other tests 3 Bad ROM 1  Standard Response  2  If test continues to fail  attempt to reinstall   are conducted on each power up routine  Firmware  may cause unrecoverable failure   3  If test failure persists  contact  an authorized Poseidon Service Center for repair    3 1 Display EEPROM  This tests the EEPROM  static memory  in the 6 Bad 1  Standard Response  2  If test continues to fail  reset system parameters   Primary Display  which contains user selectable configuration EEPROM 3  If test failure persists  contact an authorized Poseidon Service Center for  information  for internal errors or data corruption  repair    4 1 HUD ROM   RAM   Fuses  This tests the RAM  ROM and fuse 4 Bad RAM 1  Standard Response  2  If test failure persists  contact an authorized  settings of the electronics in the HUD  Head Up Display   The 5 Bad Fuse Poseidon Service Center for repair    RAM is tested only when the battery is inserted  and the   results used for all subsequent power up routines  Other tests 3 Bad ROM 1  Standard Response  2  If test continues to fail  attempt to reinstallFirm    are conducted on each power up routine  ware  may cause unrecoverable failure   3  If test failure persists  contact an  authorized Poseidon Service Center for repair    5 1 HUD EEPROM  This tests the EEPROM  static memory  in the 6 Bad 1  Standard Response 
64. e 2 3   This flashing signals the  diver to again connect the two wet switch contacts  either with fingers  water  or a metal  object  for a period of about eight seconds  as indicated by the bar graph on the right side of  the screen increasing from top to bottom   The normal power up procedure consists of the    following three steps     e Connect the wet switch contacts momentarily until the initial screen appears  e Release the contacts for 5 seconds  until the two segments flash on the right side    of the screen  e Connect the wet switch contacts a second time and hold continuously for 8 seconds     The ascent rate bar graph across the top of the screen serves as a progress bar  indicating the  amount of time remaining for the diver to complete the power up confirmation routine  about  30 seconds   After completing the power up confirmation procedure  all of the segments in the  Primary Display will display momentarily  Figure 2 4   It   s important for the user to note whether  any segments are missing  and if so  not to attempt to dive the unit until it has been repaired    POSEIDON MKVI USER MANUAL    at an authorized service center  After displaying all segments  the system proceeds to the initial  start up tests     If the system is powered up by inserting the battery  the system continues directly to the    all   segments    screen  Figure 2 4   without the need to complete the power up confirmation routine     The Poseidon MkVI electronics system conducts 55 auto
65. e a problem  that a simple reboot might not  Be sure to allow the system to power down before attempting  to reset the battery        APPENDIX PAGE 68       Troubleshooting table    lf  after attempting the Standard Response to a test failure  the automatic pre dive routine  consistently fails on the same test  note the test number and error code for the failed test  and  consult the table on the following pages  Some of the solutions in this table suggest that the  system parameters be reset  or the firmware be re installed  as follows     e Reset System Parameters  In some cases  a test may fail because some of the user   selectable parameters have become corrupted  Thus  for certain tests  the Configuration  PC software can be used to reset system parameters    e Re install Firmware  In a few  rare  cases  a failed test may be due to corrupt or inconsistent  firmware  In such cases  the Bootstrap load PC software can be used to re install the  firmware     Hardware issues    Following the table for the automatic pre dive tests is a similar table for troubleshooting various  issues related to the Hardware other than the automatic pre dive routine     POSEIDON MKVI USER MANUAL    If you get an error on test 49    Test 49 is the Positive Pressure Loop test  PPLT  where the unit checks for leaks in the  breathing loop and if the solenoids are opening and closing they way they  should     If your unit fails on this test  check the following       Is the OPV valve on the exhalat
66. e or two whether the  breathing hoses expand from their contracted state  and the counterlungs show signs of  relaxing or inflating slightly  If they do  then there is a leak somewhere in the breathing loop   This could be caused by any number of reasons including but not limited to any of the  following     e  mproper hose connection  hose not connected or incompletely connected   e Missing or failed o ring in a hose connection or a Shoulder Port connection  e Tear in a counterlung or hose   e Failed overpressure checkvalve   e CO  cartridge lid not in place  or o rings damaged or missing   e Mouthpiece o rings damaged or missing       POSEIDON MKVI USER MANUAL    Electronics Power Up    Insert the battery following the procedures described in Chapter 1  which will automatically  power up the electronics  If the battery is already inserted  the electronics can be powered up  by connecting the wet switch on the back of the display with a pair of wet fingers  as shown in    Figure 2 1     What happens next depends on how the system is initially powered up  If it is powered up by  touching the wet switch  Figure 2 1   then the first screen displayed shows the installed  firmware version number and the Poseidon MkVI serial number  Figure 2 2   The firmware  version number is shown in large digits on the left side of the display screen  Because the MkVI  is designed to accept firmware updates  knowing the specific version number of the firmware  is extremely important when diagno
67. e oxygen  sensors  the main computer system and the junction for the cables leading to the display   cylinder pressure gages  and HUD  Two thumb wheel screws allow for easy removal of the  electronics module from the gas processor housing after diving     Smart battery    The smart battery  Figure 1 6  is another patent pending design of the Poseidon MkVI   It is a snap in power supply that allows operation of the rebreather for up to 30 hours when  fully charged  It also contains its own onboard computer  and stores not only your dive log  data  but also your decompression status  tissue tensions   Keeping track of repetitive dive  status  The smart battery communicates with the other system computers via the network   and contains two user feedback systems  The first system consists of two extremely bright  red LEDs  one facing up  the other facing backwards  that provide a wide viewing angle  the  second is a 2 frequency acoustic speaker that broadcasts a very audible tone through the  water  Both systems are primarily designed to convey the safety status of your diving rig to       CHAPTER 1 PAGE 4    your partner from a distance  Once the rig is properly turned off following a dive  the smart  battery can be removed and taken to a desk top charging station  Use and maintenance of the  smart battery are discussed later in this chapter        Figure 1 5  E module with battery inserted     POSEIDON MKVI USER MANUAL       Figure 1 6   Smart battery module    Smart battery ca
68. e oxygen sensors can read PO2 values  above 1 0     If the unit  for any reason  fails the hyperoxic Linearity test  the maximum setpoint used during  the dive is 1 0     There are a few things you can do  to increase the chance of a successful Hyperoxic Linearity test     When descending  give the unit time to perform the test  i e  don   t descend too fast between  6 m  20 swf  to 10 m  83 swf      Avoid repeated ascend descend depth changes between 6 m 20 swf  to 10 m  33 swf  until  the Hyperoxic Linearity test is completed        APPENDIX PAGE  ZU    How the PO  alarm works    The PO  status is processed in the following order    If the PO  is  lt  0 25 then a hypoxic alarm will occur immediately    If the PO  is  gt  1 8 then a hyperoxic alarm will occur immediately    If the PO  is  gt  1 6 and has been so for more then 1 minute  a hyperoxic alarm will occur     If abs PO    SP   gt  SP 4 and has been so for more then 2 minutes  a deviation alarm will occur    Note  SP   setpoint     In all other cases  no alarms will occur     What to do if   can   t resolve the problem    If you can   t resolve the problem you are experiencing  do the following     e Connect your Poseidon MkVI to a computer laptop using the PC Configuration program   available as a download from www poseidon com    e Download the Red Box Log file from the two last failed pre dice checks or from the last dive  where you experienced the problem    e lf you experienced problems during a dive  also downl
69. e primary oxygen sensor is Known to be linear to at least 1 0 bar   atm   based on the successful completion of the pre dive calibration process  Thus  as long as the  PO  does not exceed 1 0 bar   atm  the response value is known with confidence     Using the default    surface    and    deep    PO  setpoint values  a setpoint of 1 0 is not achieved  until the depth exceeds 6 m   20 ft  so there is no consequence on dives shallower than this  depth  even if the Hyperoxic Linearity test is never performed  Until the Hyperoxic Linearity test  passes successfully  the PO  setpoint value will be limited to 1 0 bar   atm     POSEIDON MKVI USER MANUAL    Oxygen sensor confidence    One of the most sophisticated features of the  Poseidon MkVI is the automatic oxygen sensor  validation system  which monitors the reliability  of the oxygen sensors throughout the dive   Through a series of algorithms  the system  assigns a confidence rating to current oxygen  sensor readings  based on several factors includ   ing primary sensor validation  dynamic response  of sensors  and a comparison between primary  and secondary sensor values  If  for some  reason  the system loses confidence in the  oxygen sensors  then every few seconds an error  will be displayed momentarily on the Primary  Display where the PO  value is normally dis   played     in a manner similar to how the PO   Setpoint is displayed  If there is no confidence in  the oxygen sensors  then    COQ    is displayed   Other level
70. e problem   Red   Under normal diving conditions  the RED light will periodically blink to serve as a reminder  to the diver to monitor the Primary Display  Whenever a problem has been detected by the  system or when any of the dive parameters are not within safe limits  the HUD light will flash  continuously RED  and the vibrator will    blib    every 60 seconds   In either case  the purpose of  the HUD light is to alert the diver to look at the primary display for further information     Figure 3 1  HUD light          CHAPTER 3 PAGE 48    Audio alarm    One of the two alarm systems located in the battery module is the audio alarm  It emits a loud  staccato tone that alternates between two frequencies as a signal to abort the dive  Whenever  the audio alarm is triggered  the diver should immediately terminate the dive and commence a  safe ascent to the surface  while monitoring the Primary Display  The audio alarm will continue  to sound whenever the mouthpiece is not in the correct position  or when the diver fails to  ascend in an abort situation     Buddy alert light    Also contained in the battery module is the buddy alert light  This consists of two separate  high intensity red LED lights that flash whenever the HUD Light is flashing  The purpose of this  alarm is to alert other nearby divers of a potential problem     Monitoring the primary display    Most of the information concerning the status of the dive and the various system parameters  is Communicated to the di
71. ectronics alert    Diluent gas pressure    General alert    Current depth    Remaining dive time    i  E  IO     OC OD emt Ll e     Temperature OR High res PO     Di       Maxumum depth  Flapsed dive time    Battery indicator  Ascend   descend indicator y       Figure 3 2  LCD display layout  Figure 3 3  Fields on the primary display     The next most important region is the upper left part of the screen  where the current PO   value is displayed  2 in figure 3 2   The lower half of the screen includes basic information  about depth  on the left side  3   and time  on the right side  4   The left and right edges of the DANGER    screen  5  include bar graphs that represent the current capacity of the diluent  left side  and If the Primary Display screen is ever blank while diving the  oxygen  right side  cylinders  as a percentage of total cylinder capacity  Finally  the top edge of  the screen  6  includes a bar graph that represents the current ascent rate of the diver     Poseidon MkVI  immediately commence an abort to the surface  in Open Circuit mode  regardless of whether the HUD Vibrator is  When the Poseidon MkVI electronics are started  via the wet switch  or when the battery is activated   Failure to do so could lead to serious injury or death   inserted into the unit   the LCD screen momentarily shows all elements of the display  as illus   trated below  Each of these elements is described in detail below        POSEIDON MKVI USER MANUAL    Units of measure    The 
72. ensors needs to be replaced  The troubleshooting guide table in  Appendix 1 lists all of the error codes for Test 53  If the test fails consistently with Error Code  67  68  72  73  or 76  the Primary oxygen sensor needs to be replaced  If the test fails consis   tently with Error Code 69  70  74  or 75  the Secondary oxygen sensor needs to be replaced    Note  Error Codes 66 and 71 of Test 53 are likely due to incorrect diluent or oxygen mixtures   but may suggest the need to replace both oxygen sensors      Included with the Poseidon MkVI is the Oxygen Sensor Removal tool  Figure 4 1   This tool is  specially designed to remove oxygen sensors from the electronics module  As shown in Figure  4 1  the tool is held with the forefinger and middle finger through two large holes on either side  of the plunger  with the thumb on the plunger button  similar to holding a syringe      With the splayed flange of the tool lined up with the hole of the oxygen sensor base  snap the  tool into the sensor base as shown in Figure 4 2 It   s important to note that the tool locks into  the oxygen sensor base when plunger button is pressed  Therefore  do NOT attempt to insert  or remove the tool from the oxygen sensor base while the button is pressed        CHAPTER 4 PAGE 64         Primary oxygen sensor    Secondary oxygen sensor    e    Oxygen sensor removal tool    Figure 4 1   Oxygen sensor removal tool     Do NOT press butten when  snapped into base    Figure 4 2   Removal tool snapped in
73. ent  solutions     Mouthpiece Closed Circuit Position  This test requires that the  mouthpiece be in the Closed Circuit  CC  position in order to  pass  During this test  if the system does not detect the CC  position  the red LED and the vibrator on the Head Up Display   HUD  will pulse continuously to signal the diver to adjust the  mouthpiece position     Oxygen Sensor Calibration  Like the Positive Pressure Loop  Test  t32   this test check several things besides performing an  oxygen sensor calibration  including the composition of the  diluent and oxygen supplies  the proper funcioning of both the  oxygen and diluent calibration solenoid valves  and other  parameters associated with the oxygen sensor behavior  The  mouthpiece must remain in the CC position for the duration of  this test        ERROR CODE    46 Failed to  Fill Loop  47 Solenoid 1  Failure    49 Loop  Leaking  48 Did not fill the loop  in time    50 Valve  Leaking    O Timeout    66 Diluent  FO2 Bad    71 Oxygen  FO2 Bad    APPENDIX PAGE     76    SOLUTION    1  Ensure mouthpiece is in OC mode  2  Ensure oxygen cylinder valve is  connected and turned on  with sufficient pressure  3  Ensure all connections   seals  and o rings for the breathing hose connections  Water Diversion  Manifolds  Electronics Module  and bottom cover of the gas processing unit  are attached and seated correctly  4  Inspect for tears  cuts or punctures in  the counterlungs and breathing hoses  5  If test failure persists in spite o
74. f a  sealed loop  contact an authorized Poseidon Service Center for repair     1  Ensure dump valve on right counterlung is turned all the way clockwise   2  Ensure oxygen cylinder valve is connected and turned on  with sufficient  pressure  3  Inspect for tears  cuts or punctures in the counterlungs and  breathing hoses  4  If test failure persists in spite of a sealed loop  contact  an authorized Poseidon Service Center for repair     1  Ensure breathing loop is stable during test  2  If test failure persists in spite  of a sealed and stable loop  contact an authorized Poseidon Service Center  for repair     1  Ensure that the mouthpiece is fully in the CC position  sometimes requires  firm pressing on the mouthpiece lever   2  Ensure that the HUD is properly  positioned on the top of the mouthpiece  and that the exhaust cover is  holding it tightly in place  3  If test failure persists  contact an authorized  Poseidon Service Center for repair     1  Ensure Diluent has correct oxygen percentage  2  If test failure persists   contact an authorized Poseidon Service Center for repair     1  Ensure Diluent has correct oxygen percentage  2  If test failure persists   contact an authorized Poseidon Service Center for repair     POSEIDON MKVI USER MANUAL       APPENDIX PAGE 77    Service Interval  This test ensures that the rebreather has been  properly serviced within the past two years  104 weeks   The  number of weeks remaining until Servicing is required is shown  in the lowe
75. fications on the cylinders if necessary     Breathing hoses should be stored in a clean  dry location  where the insides of the hoses are  exposed to open air  and in a way that allows them to be laid straight  It is important not to  bend them sharply or store them in a way that causes deformation of the circular cross section  of the hoses  lest such distortions become permanent     POSEIDON MKVI USER MANUAL    Electronics should be stored in a clean  dry environment  with the battery and oxygen sensors  removed and stored separately  The battery should be recharged periodically  as described in  Chapter 1  Keep in mind that oxygen sensors may need to be replaced if the rebreather is  stored for extended periods     First stage regulators should receive annual servicing  as needed  The open circuit regulator  built into the mouthpiece of the Poseidon MkVI should be serviced by a qualified Poseidon  Service Center prior to diving after an extended period of storage     Before storing the rebreather for extended periods  it   s good practice to lubricate the user   accessible o rings  to minimize the affects of aging and extended drying     If long term storage is expected to extend for several months or more  it is good practice to  break down the cylinders and rebreather for storage in the provided case  as described below     Replacing oxygen sensors    If the automated Pre Dive routine consistently fails on test 53  oxygen sensor calibration   one  or both of the oxygen s
76. g any position information at all     unknown       In the former case  the problem is likely due to the mouthpiece switch being in the wrong  position  one or both of the magnets inside the mouthpiece being damaged or corrupted  or a  problem with the magnet sensors in the HUD  The latter case would arise if the HUD was  unable to communicate reliably with the Display  In any case  if the displayed value of the  mouthpiece position is not what it should be  first check the actual position of the mouthpiece   make sure it is firmly and completely in one position or the other  and attempt to wiggle the  HUD slightly        Figure 3 20  Mouthpiece position Unknown     Figure 3 19  No Circuit mode     POSEIDON MKVI USER MANUAL    Current depth    Immediately beneath the PO  value  on the left  side of the screen  is the current depth reading   This value is shown in either metric or imperial  units  depending on which mode is selected  as  indicated by the    FT    or    M    symbol to the right  of the current depth value   In metric mode  the  value is show to the nearest tenth  0 1  of a  meter  when in imperial mode  the value is shown  to the nearest foot  This value will flash whenever  the maximum rated depth  40 m  is exceeded        Figure 3 21  Current depth        Figure 3 22  Vaximum depth ceiling        Figure 3 23  Display indicating    ceiling     amp   Total decompression        Figure 3 24  Depth to which it is safe to  ascend  amp  total decompression time     
77. he Poseidon MkVI electronics will continue to function  indefinitely  ensuring a life sustaining gas mixture is maintained in the breathing loop  until the  following four conditions have all been met  the depth is    0     the back of the Primary Display   where the wet switch contacts are located  has been dried  the pressure in the diluent regulator  and hoses has been vented  and the mouthpiece valve has been placed in the Open Circuit  position  Once these four conditions are met  the system will vent the oxygen gas supply  system  and power down the electronics        The recommended sequence of steps for the proper post dive shut down procedure is as follows     e Ensure mouthpiece is the Open Circuit position  as it should always be when not in use    e Turn off BOTH gas supply cylinders   e Thoroughly dry the back face of the Primary Display  in the vicinity of the wet switch contacts     e Vent the diluent gas from the system by pressing the manual purge button on the ADV     POSEIDON MKVI USER MANUAL       Safe diving with the Poseidon MkVI    NEVER hold your breath when breathing underwater     ALWAYS change the CO  absorbent cartridge whenever the oxygen cylinder is refilled or  replaced     ALWAYS remove the sponge from the top of the CO  absorbent cartridge after every dive  and squeeze as much moisture out of the sponge as possible  It is extremely important to  allow this sponge to dry as much as possible before starting a new dive     If the mouthpiece vibra
78. he reason the breathing hoses use radial seals  for example      so that you can adjust their positions and that of the mouthpiece without having to make and Figure 1 11   ypical     Radial   type o ring seal   break the connections        Radial o ring seals still require a retainer to prevent them from accidental disassembly during  diving  For hose connections we use rotating shells whose threads engage a capture thread  on the mating part  see Figure 1 11 for example      For face and radial o rings to properly work  the diver is responsible for ensuring the  following    e The o ring is clean and free of debris and scratches  no cuts  gouges  dust  dirt  sand  hair  etc    e The o ring is lubricated with an approved o ring grease    e The sealing surfaces are clean and free of debris  scratches and gouges    e The sealing surfaces are lubricated with an approved o ring grease    Figure 1 10  e The retainer mechanism  e g  hand screws  hand nuts  or threaded shells  is securely  Typical     Face     type o ring seal  in place        Face    o ring seal       POSEIDON MKVI USER MANUAL    Cylinders and regulators    The Poseidon MkVI  EU version only  is factory issued with two 3 liter aluminum cylinders with  Poseidon post style valves  see Figure 1 12   The oxygen cylinder has a white valve knob and  the diluent cylinder has a black valve knob  Both cylinders are pressure rated to 204 bar    3000 psi service pressure  However  the specified maximum safe FILL pressure for 
79. how the rig and you behave during a dive  A Windows based MkVI dive log  reviewer are available as a download from Poseidon  www poseidon com   In general  the unit  will store approximately 20 hours of dive time  more if the dives were of a simple nature with  uncomplicated profiles  Examples of the common types of data you can review are battery  information  dive time and depth  However  the dive log contains much  much more information   http   www poseidon com support discovery    O ring care and maintenance    The Poseidon MkVI is a computer controlled precision underwater instrument  Its Successful  continued operation depends upon preventing water from entering the breathing loop  gas  processor  and electronics systems  To do this  and to keep the rig modular and easy to use  and maintain  there are dozens of o ring seals  These fall into two design classes     axial    o ring  seals and    radial    o ring seals  Figure 1 10 shows a typical use of a axial o ring seal  as used    POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE 9       in the CO  absorbent cartridge lid  Axial o rings are laid into an annular groove in the body of  the object to be sealed     The object is then pressed perpendicularly agains a flat  clean sealing surface  The axial o ring  is then compressed along its top side by the flat mating surface and compressed into the  groove  This compression of the o ring causes it to seal against the sides of the groove and to  the flat mating surface  Because a 
80. hypoxic  a failure  of this test more likely indicates a failed oxygen sensor and   or  a broken wire  This test does not ensure proper function of the  sensor  verified during the calibration routine  t34      Depth Temperature Sensor Validation  This test ensures that  the temperature sensor embedded in the depth sensor is  working correctly        ERROR CODE    11 Current  too low   12 Current  too high    14 Locked on  15 Locked off  16 Defective    1 7 Locked on  18 Locked off  19 Defective    26 Voltage  low  2  Voltage  very low    26 Voltage  low  2  Voltage  very low    31 Sensor  Suspect    APPENDIX PAGE 74    SOLUTION    1  Standard Response  2  If test failure persists  or the Audio Alarm speaker  does not sound during this test  contact an authorized Poseidon Service  Center for repair     1  Standard Response  2  If test failure persists  contact an authorized  Poseidon Service Center for repair     1  Standard Response  2  If test failure persists  contact an authorized  Poseidon Service Center for repair     1  Inspect the primary oxygen sensor  the wires leading from it  and the  electrical contacts at the back of the sensor cavity  and replace the sensor  and or wires if either are suspect  2  Standard Response  3  If test failure  persists  contact an authorized Poseidon Service Center for repair     1  Inspect the secondary oxygen sensor  the wires leading from it  and the  electrical contacts at the back of the sensor cavity  and replace the sensor  and o
81. idon MkVI stores decompression data in two places  the  battery  and the main backpack computer  This allows a diver to switch to a spare battery while  maintaining decompression in the rebreather unit  or switch rebreather units and transfer the  active decompression data with the battery     Test 40  Tissue Tension Test  compares the stored decompression information in both the battery  and in the main backpack computer  If the two decompression states do not exactly match on  a compartment for compartment level test 40 will fail  Failure of this test is a notification to the  diver that the system has detected this discrepancy between the two sets of decompression  data  Re start the pre dive test        POSEIDON MKVI USER MANUAL    The most common cause for this test failure is when a diver inadvertently inserts the wrong  battery into the Poseidon MkVI unit  As with any test failure  the LCD will flash the test number  and error code for five seconds  If the mis match was unintentional  the diver should replace  the battery with the correct one for the rebreather unit being used  If the mis match cannot be  corrected  then upon activating the wet switch the next time  the system will resolve the  decompression difference by selecting the most conservative value for each compartment of  the decompression algorithm     Open circuit mouthpiece position  test 43     Test 43  mouthpiece OC position test  is automatically passed provided the mouthpiece was  left in the OC posi
82. igure 1 18            CHAPTER 1    Checkvalve port    Figure 1 18   Clockwise tighten the water diversion manifold into the right and left  counterlung shoulder port        PAGE 16    POSEIDON MKVI USER MANUAL    Figure 1 19 shows a cross section of the water diversion manifold  A quick inspection will  show that on one side  the    front     you can insert a finger and feel an open vertical tube leading  down to the threaded counterlung connection post  see Figure 1 19   If you insert a finger  through the other side  the    back      you will feel an interior convex cylindrical surface blocking  entry  In order for the water diversion manifolds to work properly  we recommend that you align  the gas diversion mainifolds to follow the direction of the airflow in the loop  i e  that    the front     of the mainfold on the exhalation lung is pointing towards the mouthpiece and that    the front     of the manifold on the inhalation lung is pointing towards the canister housing     Insert one of the water diversion manifolds  they are identical  into the top port of the right  counterlung  Figure 1 18   Inspect the o ring and sealing surfaces and be sure both are clean  and lubricated  Carefully screw the manifold into the port  taking care not to cross thread the  connection post and port threads  Carefully observe the radial o ring as you make the connec   tion to make sure the o ring does not pop out of its groove  Screw the manifold all the way  down with a clockwise rotatio
83. ion lung closed      Are all loop hoses connected      Are all o rings in the loop undamaged         Are all O rings in the loop mounted correctly      Is there any visible damage to the unit      Did the loop pass a negative loop test      Are the lungs empty when the PPLT starts      Is the bottom plate of the canister housing properly assembled with all four screws     Is the mouthpiece leaking     The PPLI is a very sensitive test  where the unit looks for changes in pressure in the loop   A failed PPLT can be caused by one or both lungs being pressed by the mouthpiece or any  other part of the unit     Make sure that the lungs are kept free from any external pressure  during the PPLI   We know from experience that almost all failed PPLT are caused by improper assembly of the loop     Proper assembly of the loop  regularly lubricating o rings in the loop and making sure that the  OPV valve on the exhalation lung is closed  rinsed with fresh water and cleaned will minimize  the risk of failing on test 49     Error on test 53    Test 53 is the oxygen sensor calibration test and this is quite complicated and based on a lot  of factors such as       Temperature of the sensors     Percentage of oxygen in the gases used    Response time from a sensor     Milli voltage of a sensor       APPENDIX PAGE 69    The test will start by injecting pure oxygen directly on the primary oxygen sensor for 20 continuous  seconds  After the calibration constants for oxygen are established  the 
84. ip the threads  Great force is not required to seal the module     WARNING     Deep  permanent scratches  cuts  gouges or other damage to the  polished radial sealing surfaces of the electronics module  and or  failure to ensure that all electronics module o rings are in place   free of debris  and are lubricated  could lead to water entry into  the top of the cartridge housing during a dive  eventually causing  a lock up of the breathing loop and forcing an immediate open   circuit abort to the surface        PAGE 22    CHAPTER 1         Outer radial seal        em SEH    H    W     a       A    S       A Inner radial seal  re         j       Outer radial  sealing surface    Inner radial  sealing surface    Figure 1 31  Firmly screw down the  hand nut on the right and the left side  of the electronics housing  use NO  processor housing tools     Figure 1 30  Align the cable junction box with the  front of the gas processor  closest to backpack    then insert the electronics module innto the    DANGER     The electronics module contains the most important life critical  element of the Poseidon MkVI  the oxygen sensors  A leak into  the electronics housing may contaminate the oxygen sensors  and prevent them from providing correct readings  Dual radial  o rings help to guard against this  and the computer system is    programmed to discover anomalies in the oxygen sensors and  to attempt to repair the situation  Detection of an oxygen sensor  anomaly will lead the computer to 
85. ir     POSEIDON MKVI    TH Time  SEC     29    30    31    34    35    38    4 5    Go    1 9    USER MANUAL    DESCRIPTION    Audio Alarm Speaker  This test measures the amount of elec   trical current consumed by the speaker in the battery  Audio  Alarm   when activated     Oxygen Cylinder Pressure Sensor Validation  This test includes  a series of tests that confirm that power can be supplied to the  oxygen cylinder pressure sensor  and that the signal from the  sensor is within limits  regardless of whether the cylinder valve  is turned on      Diluent Cylinder Pressure Sensor Validation  This test includes  a series of tests that confirm that power can be supplied to the  diluent cylinder pressure sensor  and that the signal from the  sensor is within limits  regardless of whether the cylinder valve  is turned on      Primary Oxygen Sensor Validation  This test measures the  voltage output from the primary oxygen sensor  to ensure it  exceeds a minimum threshold value  Although it is possible  that the gas mixture in the breathing loop is hypoxic  a failure  of this test more likely indicates a failed oxygen sensor and   or  a broken wire  This test does not ensure proper function of the  sensor  verified during the calibration routine  t34      Secondary Oxygen Sensor Validation  This test measures the  voltage output from the secondary oxygen sensor  to ensure it  exceeds a minimum threshold value  Although it is possible  that the gas mixture in the breathing loop is 
86. is to make sure that the breathing loop  is intact and not leaking  Water entry into the breathing loop can cause serious problems if  mixed with the CO  absorbent material in the cartridge  As discussed earlier in this Chapter  a  manual negative pressure loop test can help detect leaks in the breathing loop     Another common test is the Positive Pressure Loop Test  PPLT   which is similar to the Negative  Pressure Loop Test  except the test is performed by pressurizing the breathing loop with  positive pressure  Like the Negative Pressure Loop test  this test could very easily be performed  manually  However  one of the features of the Poseidon MkVI     the placement of the depth  sensor within the breathing loop     allows this test to be performed automatically  And so it is      in Test 49  Figure 2 9   Test 49 actually performs four separate tests  only one of which is the  PPLT  The other three are     e Verifies depth sensor is sensitive to small pressure changes  e Verifies that both metabolic oxygen solenoid valves are injecting gas    e Tests for leaks in all four solenoid valves     POSEIDON MKVI USER MANUAL    Before reaching this test  indeed  before Powering up the electronics   it   s important to make  sure that the over pressure relief valve on the bottom of the right  exhale  counterlung is  adjusted to the full clockwise position  Ae mentioned previously  the mouthpiece should be in  the OC position  and the oxygen cylinder should be turned on  Also  the 
87. ise the diver to switch from closed circuit to open circuit mode  or vice versa  The HUD  also contains a sophisticated sensor to detect which position the mouthpiece is in  closed   circuit or open circuit         CHAPTER 1       Mouthpiece switch in closed circuit position       Figure 1 3  Mouthpiece open   amp  closed circuit switch and Heads up display  HUD     ADV    PAGE    2    POSEIDON MKVI USER MANUAL       Electronics module                    Diluent regulator Oxygen regulator    Diluent cylinder valve Oxygen cylinder valve    Oxygen high  pressure sensor  Diluent high    pressure sensor  Right rear    breathing hose    Left rear breathing S  connection    hose connection    Diluent cylinder  Oxygen cylinder    Smart battery    Tank connection   Gas injection block bracket  Carbon dioxide  CO   removal  cartridge housing    Cylinder connection  bands    Figure 1 4  Back view of Poseidon MkVI     CHAPTER 1 PAGE 3    Breathing loop overview    The most visible elements of the front portion of the Poseidon MkVI comprise  the breathing loop  breathing hoses  convertible open and closed circuit  mouthpiece with built in automatic diluent addition valve  ADV   water diver   sion manifolds  sometimes referred to as    T ports    or    Shoulder Ports      and  the left  inhalation  and right  exhalation  counterlungs     The breathing loop is a compliant system  its volume changes in response to  breathing   Its purpose is to provide an external reservoir for the exhaled 
88. lanned   to avoid  potential problems with the oxygen sensors  It is also good practice to rinse  disinfect  and dry  the absorbent sponge following a dive  Any disinfectant solution that doesn   t have a negative  effect on plastics and or rubber can be used for this purpose  Poseidon recommends using a  Figure 1 49  Remove the End Plate  Figure 1 50  Remove the Absorbent Cartridge  disinfectant called Gigasept    or a disinfectant called Virkon     POSEIDON MKVI USER MANUAL    DANGER  Dangerous levels of carbon dioxide  CO   can cause  symptoms that include  but are not limited to  rapid breathing   severe headache  tunnel vision and disorientation  High CO  levels    can also increase the potential for oxygen toxicity  Diving a closed   circuit rebreather with an expended CO  absorbent cartridge  could lead to serious injury or death When in doubt  switch the  mouthpiece to open circuit mode and end the dive immediately        Next  remove the black top of cartridge interface cap  Figure 1 51   and clean  disinfect  dry   and stow that component  Note that this cap is fitted with two o rings     a top mounted axial  o ring that seals to the interior top of the cartridge housing  and a radial o ring that seals to the  top receiver pocket on top of the CO  absorbent cartridge  Replace these o rings if cuts or  gouges are present     Remove the pre packed absorbent cartridge  Figure 1 52  from the base plate and properly  dispose of the cartridge  Replacements  discussed belo
89. leshooting guide    This Appendix provides detailed information on possible problems that may occur when pre   paring or using the Poseidon MkVI for diving  It is divided into two main sections  The Auto   matic Pre Dive Tests  and Hardware Issues  The Automatic Pre Dive Test section includes all  of the automated tests according to each test number  with a description of what is being  tested and the possible failure modes  as well as possible causes and solutions  The Hardware  Issues section discusses various problems that can occur with the mechanical aspects of the  MkVI  and how to correct them  Many of the problems in both sections can be easily solved by  the diver  but some require repair at an authorised Poseidon Service Center     Never tamper with bolts or screws this may permanently destroy the unit  This is due to that  the inernal cavities is filled     Do not try to alter the clock setttings to bypass service intervals etc  This may set the unit in  an un operable state     Automatic ore dive test    As described in Chapter 2 of the Manual  the Poseidon MkVI electronics automatically conduct  a series of tests whenever the system is powered up  i e   whenever a battery is inserted into  the electronics module  or the wet switch on the back of the Primary Display is activated    While these tests are running  the test number is shown on the left side of the Primary Display   where the depth is normally shown   and the test number is preceeded by a lower case   
90. lified service center  Failure to  comply with these instructions could lead to an oxygen fire and may cause serious  injury or death     The Poseidon MkVI requires proper pre dive setup  and several important verification  procedures that must be carried out by the diver  Details of these procedures are  included in Chapters 1 and 2 of this Manual  The MkVI also incorporates many auto   matic system tests as part of the power up procedure  Using the MkVI without com   pleting these automatic system tests poses significantly increased risk to the diver     POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE VII       Chapter 3 of this Manual describes procedures for donning and fitting of the MkVI  to  ensure proper positioning on the diver  as well as instructions for proper use while  conducting a dive     Chapter 4 of this Manual describes appropriate post dive procedures and long term  storage and maintenance requirements for the Poseidon MkVI  including conditions  for storage  shelf life of certain components  and appropriate precautions  as well as  a maintenence and inspection schedule  Failure to comply with these procedures  may result in deteriorated and or damaged components  and can lead to improper  functioning of the equipment  A separate set of instructions detailing maintenance  requirements is also provided for reference purposes     8 4    The diluent cylinder for the Poseidon MkVI should only be filled with Grade E  or  equivalent  air     The oxygen cylinder should 
91. lute the oxygen when Filling your own cylinders is dangerous business  You are  mixed in the breathing loop  This diluting characteristic is a requirement of a closed circuit personally responsible for your well being and those around you  rebreather because oxygen is toxic at partial pressures greater than 1 6 bar   2 ATM  If one when you fill your own cylinders  Before you consider this option   were only using pure oxygen in a rebreather the safe operating depth would be limited to 6 obtain formal training on the operation of the equipment and in  meters   20 feet underwater  Another characteristic of a suitable diluent gas is that it is normally the maintenance of oxygen clean systems  Never overfill any  selected so that it is directly breathable as an open circuit gas at the maximum operating Poseidon MkVI cylinder  oxygen or diluent   Service all equipment  depth of the rebreather  General examples of diluent gases that can be used in rebreathers per the manufacturer   s recommendations     include  air  nitrox  trimix  and heliox  The MkVI is limited to 40m   130 feet depth and uses only  air as an allowed diluent  following established compressed air scuba sport diving practices   The decompression algorithm requires that air be used as the diluent in the Poseidon MkVI     Poseidon MkVI cylinders  both oxygen and diluent  should be filled by a qualified facility that is  properly trained and equipped to fill such cylinders  The requirements concerning maintenance  of
92. mated and semi automated tests  during the pre dive routine  This procedure verifies a wide variety of parameters  and takes  about 3 minutes to complete  A full description of all of the tests is included in Appendix 1 of  this Manual  but a general description     including tests that require intervention     is included here     Although there are 55 tests  the display only shows 36 of them because several test numbers  are reserved for future tests  Note that if the depth is greater than zero  the system automatically  shifts into Dive Mode  and alerts the diver to abort the dive due to a failure to complete the  ore dive routine     Figure 2 4  The Primary Display showing  d all LCD elements        DANGER   Do not attempt to breathe on the Poseidon MkVI during the    automated pre dive routine  Oxygen control is disabled during    portions of this routine  so doing so involves a risk of hypoxia  Do  not attempt to conduct a dive until the system has successfully  completed the pre dive routine           CHAPTER 2 PAGE 40    Power up self test  test 1   55     The first series of tests  numbers 1 55  are referred to as Power Up Self Tests  or PSTs   They  are internal checks on the functionality of all of the various sensors  computers  actuators  and  alarm systems in the Poseidon MkVI  You will see and hear the rig as it tests the HUD light and  vibrator  and the battery lights and speaker systems  Similarly  you may also hear the rig  opening and closing some of the ga
93. n sensor to the oxygen sensor base by screwing it into place  Make sure    the o ring around the base of the threads on the oxygen sensor is clean and free of any  damage  and that it seals properly when the sensor is screwed down snugly       Oxygen sensor serial number    Oxygen sensor base          CHAPTER 4 PAGE 65    Once the sensor is properly attached to the sensor base  the electrical connector on the elec   tronics module should be attached to the sensor  The sensor has three electrical contact pins  in a straight row  parallel to a flat plastic guide tab  Hold the connector so that the three contact  holes line up with the three pins on the sensor  and the two plastic pins on the connector  straddle the flat guide tab  Carefully push the connector without bending any of the pins  until  it is completely seated                    Electrical connector    Oxygen sensor  Figure 4 5  The oxygen sensor is attached to the  sensor base by threads  sealing by anio ring     POSEIDON MKVI USER MANUAL    With the electrical connector properly attached to the sensor  inspect the two radial o rings on  the oxygen sensor base to make sure they are clean and free of any damage  Ensuring that the  electrical connector is still firmly attached  slide the sensor into the electronics module  with the  flat part of the outer edge of the sensor base facing towards the top of the electronics module   Figure 4 6   Carefully press the sensor into the electronics module until it is firmly se
94. n the Primary Diplay  see Chapter 3      Hose Detached from Fitting  One of the breathing hoses becomes  detached from the fi tting mounted at the end of the hose     Cracked Absorbent Cartridge  The plastic housing of the SofnoDive   797  CO  absorbent cartridge can sometimes develop a crack if it is dropped  or due to mishandling during shipment     Counterlungs Shift Position Underwater  The counterlungs may shift position  underwater  floating up above the diver   s shoulders  or squeezing the  diver   s neck     Misaligned Cylinders  If the cylinders are not attached to the backpack  at the same height  the rig will wobble back and forth and be unstable  when standing upright     Leaking Fittings  A small stream of bubbles may be seen eminating from  one or more of the fi ttings on the open circuit mouthpiece supply hose   the hoses that connect the fi rst stage regulators to the pneumatics  block on the electronics module  or from one of the high pressure sensors     PO2 Setpoint Limited to 1 0 bar   atm  The system is confi gured for a     deep    PO2 setpoint value greater than 1 0 bar   atm  but the setpoint  never increases above 1 0 bar   atm  even when the depth is greater than  15117 50 Ti        SOLUTION    1  Ensure that the mouthpiece is fully in the OC or CC position  Sometimes requires firm  oressing on the mouthpiece lever   2  Ensure that the HUD is properly positioned on the top  of the mouthpiece  and that the exhaust cover is holding it tightly in place
95. n until the o ring is fully engaged with the radial sealing surface  of the counterlung port  Verify that the    front    of the water diversion manifold is pointed forward   so that the hose connecting the mouthpiece attaches to the    front    side of the diversion  manifold  If the manifold is not oriented correctly  unscrew the manifold  counterclockwise  just  until the front aligns in the proper direction  always be less than one full rotation  When you  have completed installation of the two water diversion manifolds in to the top port of both  counter lungs it should look as shown in Figure 1 20     Airflow    Figure 1 19    Cross section of the water  diversion manifold    Right T Connection        Water flow  To counterlung    Airflow Water       CHAPTER 1 PAGE 17    OF of        Figure 1 20   The recommended alignment of the two water diversion manifolds when installed  on the counter lungs     POSEIDON MKVI USER MANUAL    5  Rear CC loop hoses    Figure 1 21   Insert the rear left and right hose  into the gas processing house        Breathing loop    All of the components of the breathing loop discussed in this section were first introduced in  Figures 1 1 and 1 4 above  It is useful to note that all breathing hoses and hose fittings are  identical  There are a total of eight  8  hose connections to be made in the assembly of the  Poseidon MkVI  Assembly of these hoses starts at the gas processor  and continues forward  to the mouthpiece        CHAPTER 1 PAGE 18  
96. nce     CO  Absorbent Cartridge    Follow the procedures described in Chapter 1 for installation of a new SofnoDive   797 CO   absorbent cartridge  When conducting a repetitive dive  It   s critical to keep track of the hours  of personal use for the cartridge once it is installed  The absorbent cartridge must be changed  whenever the oxygen cylinder is re filled  While many people experience a strong reaction to  CO  buildup  as would result from diving with a depleted or missing cartridge  in the form of  un naturally rapid breathing rate  disorientation and the onset of a strong headache  some  people do not experience them  Do not risk CO  poisoning  Change the cartridge every three  hours of use or whenever the oxygen cylinder is recharged  whichever comes first     POSEIDON MKVI USER MANUAL    WARNING     Always replace the CO  Absorbent cartridge with a new  un used    absorbent cartridge whenever the oxygen cylinder is re filled  This  will minimize the risk of CO  poisoning        Intact Breathing Loop Verification    Inspect all breathing hose connections to ensure that they are properly attached  The attachment  nuts should be hand tight and the nuts should be screwed down flush against the receiver  manifolds in all 8 locations  two at the top of the gas processor  four at the shoulder ports  and  two at the mouthpiece   Also at this time  make sure the right counterlung dump valve is fully  closed  turned all the way clockwise   This is important for the pre dive
97. nd the 11 inch adapter are both sold separately and    _are not included in the Poseidon MkVI  The pre packed scrubbers are consumables and are    _sold separately and not included in the MkVI        POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE 14       2  Tank band    Figure 1 14a    Tank band newer type    The tank band buckles should be aligned so  that they are folded backwards when the tank  is securly locked in place                  Thread the tank band through the tank band  bracket  from the back towards the front     Thread the tank band buckle  D ring on to the tank band         Figure 1 14b    Tank band older type    Thread the tank band through the  tank band bracket and in to the  buckle  see figure      Fold the tank band and thread it back  through the same opening on the tank band  bracket     Adjust the length of the tank band so it fits  on the tanks you intend to use with your  Poseidon MkVI    Properly adjusted  the tanks should be firmly  attached  not being able to move in the tank  band  when the buckle is closed and  secured around the tank     Adjust the length of the tank band so  it fits on the tanks you intend to use  with your Poseidon MkVI    Properly adjusted  the tanks should  be firmly attached  not being able to  move in the tank band  when the  buckle is closed and secured around  the tank     Thread the tank band through thin webbing  strap loop  counterlung strap         POSEIDON MKVI USER MANUAL    1  Stabjacket   BCD   Wing    Attach the ca
98. ne     Battery EEPROM  This tests the EEPROM  static memory   in the battery processor  which contains user selectable con   figuration information  for internal errors or data corruption     Battery Data logger  This test ensures that the data log  circuitry in the Battery is functional and accessible     Firmware Version Compatibility  This test compares the  versions of firmware installed on each of the system proces   sors  and ensures they are compatible with each other     Battery State of Charge  This tests the circuitry that calculates  the State of Charge  SoC  for the battery  by measuring the  base level electrical current consumed by the electronics   Many of the tests that follow this test rely on an accurate SoC  calculation     Primary Display Backlight  This test measures the amount of  electrical current consumed by the backlight of the Primary  Display  when the backlight is turned on with maximum bright   ness  After this test has completed  the backlight remains on  for the remainder of the tests        ERROR GODE    6 Bad  EEPROM    4 Bad RAM  5 Bad Fuse    3 Bad ROM    6 Bad  EEPROM    13 Bad Chip    7 Battery   Mismatch  8 Non Batt    Mismatch    9 Current  too low   10 Current  too high    11 Current  too low   12 Current  too high    APPENDIX PAGE 72    SOLUTION    1  Standard Response  2  If test continues to fail  reset system parameters   3  If test failure persists  contact an authorized Poseidon Service Center for  repair     1  Standard Response
99. ner and outer radial sealing surfaces in the electronics module receiver pocket at  the top of the gas processor  see Figure1 30   These surfaces should be free from scratches   gouges  and dings  Ensure that these surfaces have a smooth application of lubricant and that  no debris  dust or other foreign material is present     Inspect both inner and outer electronics module radial o rings  Figure 1 30   If any of the  O rings are damaged  cut  or gouged  replace them  Ensure that each o ring is lubricated and  that there is no debris  dust  sand etc  on any of the o rings     Orient the electronics module such that the cable entry port is aligned towards the front of the  gas processor  as shown in Figure 1 30  and carefully press the electronics module into the  receiver pocket in the top of the gas processor housing as shown in Figure 1 30     Be sure when inserting the electronics module into the open receiver cavity at the top of the  gas processor that the radial o rings do not extrude from their circular grooves  The radial  O rings in the electronics module should insert smoothly into the gas processor receiver cavity  until they are no longer visible and the under side lip of the electronics module fastener hand  nut flange is flush with the top ledge of the gas processor receiver cavity   see Figures 1 30 and  1 31 for proper seating of the electronics module      Tighten the left and right retainer hand nuts as shown in Figure 1 31  Do NOT use tools or you  may str
100. ng hose to the front port  of the right counterlung water diversion manifold  Figures 1 43   All of the previously mentioned  Connect the right front CC loop hose to the exhalation counter lugn T connection  practices concerning inspection and lubrication of o rings and o ring receiving surfaces apply     Hand tighten the hose retainer nut as shown in Figure 1 43  Use NO tools  Repeat this process  for the left front breathing hose attachment to the front port of the water diversion manifold for  the left counterlung  The result should now appear as shown in Figure 1 42     Retainer nut capture thread    Sealing surface for    radial    o ring seal       Radial    o ring seal       Figure 1 43  Hand tighten  clockwise  the swivel nuts for the front breathing  Figure 1 42  Breathing hoses properly connected onto the water diversion manifolds  hoses at the front shoulder ports  DO NOT use tools or over tighten     POSEIDON MKVI USER MANUAL    12  Front CC hoses to mouthpiece    Make sure that the in  and exhalation check valves are installed in the mouthpiece  in their  correct position     Make sure the check valves are undamaged  that they aren   t folded in any way and that they  are clean     Make sure you have the mouthipece in the correct position Connect the front left CC loop hose  to the inhalation side CC loop hose connection of the mouthpiece     Connect the front right CC loop hose to the exhalation side CC loop hose connection of the  mouthpiece        Figure 1 44 
101. ng strap on the other counterlung  These two keep the counterlungs held together    It is well worth the time spent in shallow water making adjustments to these various straps until  the counterlungs fit comfortably and closely to the upper chest and shoulders  The better the  counterlung adjustment  the easier the breathing will De when underwater        POSEIDON MKVI USER MANUAL    Tips on breathing    Breathing underwater on a closed circuit rebreather  such as the Poseidon MkVI  is somewhat  different from breathing on land  or breathing with conventional scuba gear  As the diver  exhales  the counterlungs both expand  As the diver inhales  the counterlungs contract  The  direction of gas flow through the breathing loop is governed by the two check valves in the  bottom portion of the mouthpiece  The incorporation of two separate  over the shoulder  counterlungs on the MkVI helps to minimize the effort required to breathe underwater  but  there are a few tips that make breathing easier     The most important thing is to maintain an optimum volume of gas in the breathing loop  If  there is too much back pressure when exhaling  often felt in the cheeks   or if the overpressure  relief valve on the exhale  left  counterlung    burbs    gas at the end of an exhaled breath  then  the loop has too much gas  and some should be vented  e g   by exhaling through the nose    If the counterlungs    bottom out    and or the Automatic Diluent Valve  ADV  in the mouthpiece  is triggere
102. nsiter housing to your stabjacket   BCD   Wing using either the tank bands or an  11    adapter     As previously mentioned  the Poseidon MkVI is sold with an optional backpack and buoyancy  compensator  This is to allow more experienced divers the choice of using an existing personal  backpack  harness  and buoyancy compensator  Poseidon supplies the tightly integrated  Besea  harness  and buoyancy wing system for use with the MkVI  The front extrusion rail on  the Poseidon MkVI accepts the 11 inch adapter  see Figure 1 13   Attachment of the Besea is  as quick as aligning the 11 inch adapter pin bolts to the upper holes in the Besea harness as  shown in Figure 1 13  Make sure that you have correctly adjusted the length size of the Besea  harness back plate before aligning the 11 inch adapter    Once the correct alignment of the 11 inch adapter pin bolts have been found  secure the  11 inch adapter position on the front rail of the Poseidon MkVI     The optional buoyancy compensator wing can be quickly attached or removed from the Besea  harness using the four guide pins attached to the eight attachement bolts on the back of the  Besea harness  The Besea comes with a convenient carrying handle that can be used to  transport the entire assembled Poseidon MkVI to and from the dive preparation area        You can attach a Poseidon MkVI to your  BCD Stabjacket using a tank band   BCD Stabjacket using an 11    adapter        If your BCD Stabjacket is equipped with a mounting poin
103. ntact an authorized Poseidon Service Center or dive shop for  repair     This situation occurs when the Hyperoxic Linearity Test fails  or has not been completed  This  test is performed the fi rst time the depth reaches 6 m  20 ft  and setpoint values greater than  1 0 bar   atm are not allowed until after this test has been performed and passes  See Chapter  3 of the Manual     APPENDIX PAGE 78    
104. ntain a breathing gas mixture representing an  inspired oxygen partial pressure of between 0 5 bar  0 35 bar minimum  and 1 2 bar   1 4 bar maximum   The oxygen fraction of the mixture depends upon depth and set  point  At the surface the oxygen fraction will vary from 35  to 100  and the nitrogen  fraction from 65  to 0   At the maximum operating depth of 40 msw  the oxygen  fraction will vary from 20   set point   1 0  to 28   set point   1 4  and the nitrogen  fraction will vary from 80   set point   1 0  to 72   set point 1 4   Users must monitor  displays and alarm systems and respond appropriately if oxygen concentrations  become unsafe     The Poseidon MkVI requires the monitoring of a backlit liquid crystal display  LCD   screen  and thus should only be used when water visibility exceeds approximately  30 centimeters  Using the MkVI in visibility conditions that prohibit viewing of the LCD  screen poses increased risks of operation     The Poseidon MkVI incorporates high pressure oxygen as one of its supply gas  mixtures  and uses equipment that has been cleaned and prepared specifically for  use with high pressure oxygen  Appropriate care must be taken when handling such  mixtures  especially when filling cylinders and providing proper maintainance and  oxygen compatible cleanliness for all components exposed to high pressure oxygen   Components exposed to high pressure oxygen  e g   the oxygen regulator and associated  pneumatic Components  must be serviced by a qua
105. nterface is a large format flat screen panel that displays all  you need to know about consumables management     cylinder pressures  dive  time  depth  oxygen level  and it incorporates a sophisticated resource algorithm  that monitors all systems for you and tells you when its time to head up  If things  are not going correctly for whatever reason  the MkVI has audible  tactile  and  visual warning systems to get your attention  and to advise your diving partner of  your status     1 Unlike open circuit scuba  the rate of gas consumption on the Poseidon MkVI does not  depend on the depth of the dive  Instead  the oxygen supply depends on how fast the diver  consumes oxygen through metabolism  Divers with large muscle mass  and or working  hard underwater  will consume oxygen faster  leading to shorter total dive times  Conversely   smaller divers with less muscle mass  or relaxed divers who are not working hard  will gain  significant improvements in dive duration         CHAPTER 1 PAGE V    Switchable mouthpiece     in a significant new patented design  the Poseidon  MkVI gives you the ability to switch from closed circuit to open circuit operation  in one easy motion  without the need to search for a spare mouthpiece in an  emergency  The ultra compact switchable mouthpiece is lightweight and easy to  breathe  giving you high performance in both open circuit and closed circuit  modes  The MkVI mouthpiece also combines in the same housing a pressure  balanced    Automatic 
106. oad the Dive log from the dive in question    e Contact the dive center   dealer where you purchased your Poseidon MkVI and e mail  them the Red Box Log file s  and dive log s  you have downloaded from your MkVI     If you are asked to send in your e module for repair   analysis  please send the following parts   e Electronics module with primary display  HUD and HP sensors    e Battery  e Oxygen sensors    POSEIDON MKVI USER MANUAL APPENDIX PAGE 71       Always ensure that battery is adequately charged  but not over charged  before attempting automatic pre dive routine  The standard response to any test failure should be an attempted reboot   Repeated failures of the same test  including Error Code 0  may sometimes be solved by removing the battery  placing it on the charger for a few minutes  then re inserting on the electronics   NEVER remove the battery until after the system has powered down  Time values are maximum seconds allowed for each test    TH TIME  SEC  SOLUTION    DESCRIPTION ERROR GODE    1 1 9 System Data Log Integrity Test  This test ensures that the data 2 Bad Chip 1  Standard Response  2  If test failure persists  contact an authorized  log circuitry in the Primary Display is functional and accessible  Poseidon Service Center for repair    2   Display ROM   RAM   Fuses  This tests the RAM  ROM and 4 Bad RAM 1  Standard Response  2  If test failure persists  contact an authorized  fuse settings of the electronics in the Primary Display  The 5 Bad Fuse Posei
107. p radial o ring does  form an important seal  so take extra care to make sure the seal is snug and reliable      Figure 1 58  this is the reverse of the procedure shown in Figure 1 51 above   Inspect the  radial sealing surface at the bottom of the cartridge housing  Figure 1 59   This surface should  be free from scratches  gouges  and dings  Ensure that this surface has a smooth application  of lubricant and that no debris  dust or other foreign material is present     POSEIDON MKVI USER MANUAL       The importance of ensuring that all exposed o rings and their respective sealing surfaces are  smooth and clean cannot be overstated  Careless treatment of these o rings could cause an  an entry path for water into the system  leading to an aborted dive  or worse   A slow leak may  represent a minor inconvenience  but may eventually lead to more severe problems later  A fast  leak could lead to an immediate requirement for an open circuit abort to the surface  O rings  and their sealing surfaces are at the heart of reliable operation of the Poseidon MkVI  Pay  attention to this detail when you assemble the rig           Je e  2  Es sl E  Se  E         w a e  be a   Zen     A             Figure 1 60   nsert the CO  absorbent stack        CHAPTER 1 PAGE 34    Insert the assembled CO  absorbent stack into the cartridge housing  see Figure 1 60   Take  care during the final stage of assembly to align the four hand nuts with their respective threaded  connectors on the extruded cart
108. period of time  would be to top off the charge once a month by running it on the  normal charge cycle in the desktop charger  If this is not possible  then the best long term  solution is to leave the battery on the charger  with the power to the charger turned on   The  method of monthly top off of charge  however  will maximize the battery life  Store the battery  in a cool  dry environment     Decompression data    In the Poseidon MkVI rebreather  individual user decompression data are stored in both the  backpack computer and the smart battery computer  Thus  every user carries their decom   oression information with them when they remove the battery  If the same user dives the same  rig  then the diver will receive repetitive dive surface credit for the time spent on the surface   even if the battery is removed from the rig between dives   The decompression algorithm is a  9 compartment real time implementation of the industry proven DCAP decompression engine     It is strongly recommended that you use the same battery in the same Poseidon MkVI for any  repetitive series of dives  Once sufficient surface credit has cleared the decompression model  entirely  generally 24 hours of no diving  then you can swap batteries between rebreathers  without risk     POSEIDON MKVI USER MANUAL    DANGER     Ifa user swaps batteries with another Poseidon MkVI unit than the  one they were diving  and if they incurred a decompression debt   and if the battery computer memory storing the 
109. post       Figure 1 7   Left  Align the retainer screw posts on the Smart Battery with the receiver slots  at the top of the Electronics module  note that the 4 contact pins in the housing must align  with those on the battery    Right   push the battery into the slot  engaging the pins and the  radial o ring seal  until an audible    snap    is heard as the safety latch closes        Figure 1 8  Removing the smart battery    POSEIDON MKVI USER MANUAL    Charging    Included with the Poseidon MkVI is a proprietary multi function desk top charger unit that  includes adapters for most international power outlets  The battery charger has three status  lights arranged in a circular pattern on the open section of the base  These are  in counter   clockwise order from the lower left in Figure 1 9  power     Learn Cycle    status  and charge  status     Power Indicator Light  When green  the power is    on    and the charger is ready to operate  If  it occasionally    blips     the charger does not have external power  and is actually draining the  battery     Learn Cycle Indicator Light  The middle light indicates the status of a    Learn Cycle     The  battery has its own onboard computer that monitors the state of charge  Over a period of  weeks to months  the computer   s estimate of the remaining power in the battery gradually  decreases in accuracy  The computer can    re learn    what a full  100   charged battery should  look like using the Learn Cycle of the charger  The ba
110. r right corner of the primary display  where the  elapsed dive time is normally shown     67 Primary  Dil  Low  68 Primary  Dil  High  72 Primary  O2 Low  73 Primary  O2 High  76 Bad Time  Constant    69 Secondary  Dil  Low  70 Secondary  Dil  High  74 Secondary  O2 Low  75 Secondary  O2 High    7  Not in CC  Mode    81 Servicing  Required      Replace Primary oxygen sensor with a known good sensor  2  Ensure  E absorbent cartridge is installed correctly  3  Ensure breathing loop tem   perature is within range limits  4  If test failure persists  contact an authorized  Poseidon Service Center for repair     1  Replace Secondary oxygen sensor with a known good sensor  2  Ensure  CO  absorbent cartridge is installed correctly  3  Ensure breathing loop  temperature Is within range limits  4  If test failure persists  contact an autho   rized Poseidon Service Center for repair     1  Ensure that the mouthpiece is FULLY in the CC position  sometimes  requires firm pressing on the mouthpiece lever   and that the HUD is properly  positioned on the top of the mouthpiece        Contact an authorized Poseidon Service Center for servicing     POSEIDON MKVI USER MANUAL    CATEGORY    Mouthpiece    Breathing  Hoses    Absorbent  Cartridge    Counterlungs    Cylinders    Pneumatics    Electronics    DESCRIPTION    HUD Seating  If the Head Up Display becomes unseated or out of  alignment  there may be frequent failures of Pre Dive Test 28 or 33  or  errors concerning    No Circuit     nc  o
111. r wires if either are suspect  2  Standard Response  3  If test failure  persists  contact an authorized Poseidon Service Center for repair     1  Standard Response  2  If test continues to fail  ensure loop temperature  is within limits  3  If test failure persists  contact an authorized Poseidon  Service Center for repair     POSEIDON MKVI    TH Time  SEC     40    43    44    45    48    2    120    120    120    USER MANUAL    DESCRIPTION    Decompression Status Verification  In this test  the two sets of  decompression data  one stored in the rig electronics  and  one in the battery   are validated and compared  see relevant  discussion in Chapters 1  amp  2   In addition to comparing and  validating the two sets of tissue tension data  this test also  compares the serial numbers of the battery and the main elec   tronics  as well as the time stamp on both     Mouthpiece Open Circuit Position  This test requires that the  mouthpiece be in the Open Circuit  OC  position in order to  pass  During this test  if the system does not detect the OC  position  the red LED and the vibrator on the Head Up Display   HUD  will pulse continuously to signal the diver to adjust the  mouthpiece position     Sufficient Oxygen Supply Pressure  This test requires that the  oxygen cylinder contains sufficient pressure for a dive to be  started  at least 25  of maximum capacity      Sufficient Diluent Supply Pressure  This test requires that the  diluent cylinder contains sufficient pressure
112. re    Figures 1 7 shows the installation procedure for the smart battery  The battery contains four  female quick connect contact pins mounted on an extended cylindrical  o ring sealed post  that projects from the end of the battery  This mates to a receptacle with four corresponding  male fixed pins in the electronics module inside a sealing cavity for the radial o ring seal  Be  careful not to short the contact pins when the battery is not in the rig  and inspect the battery  socket receptacle on the electronics module prior to inserting battery to ensure there is no  water  Once the battery is properly installed in its docking slots and pushed all the way forward   an audible    CLICK    will be heard as the safety latch closes  The battery is now ready for  diving     To remove the battery from the electronics module  press the safety latch and firmly push the  top part of the battery outward  as shown in Figure 1 8  It is usually best to remove the battery  when the system is dry  to avoid water ingress to the electrical contacts     Safety    The smart battery uses a high energy density lithium ion rechargeable battery  similar to bat   teries used in laptop computers  If any liquid or discoloration is observed inside the clear  plastic battery housing  dispose of the battery immediately  Disposal of an old or failed smart  battery should be in accordance with local laws regarding disposal of Li ion laptop computer  batteries        CHAPTER 1 PAGE 5    Retainer  scr  w 
113. release of the mating pressure would result in a leak in a  face type seal  these require a securing mechanism that not only prevents the part from lifting  off accidentally  but also actively compresses the axial o ring against the flat mating surface  In  the case of the CO  cartridge  the cartridge end plate is equipped with four thumb screws to  secure it in place and tighten it down     A second  and more commonly used seal is the     radial     o ring  Figure 1 11 shows a typical  implementation in the Poseidon MkVI breathing hose and hose connection ports  In contrast  to a axial o ring seal  a radial seal involves a circular groove that goes around a cylindrical or  semi cylindrical object  it can be a rectangular object with rounded corners provided the  corners have a sufficiently large radius     an example of this is the Electronics Module dual  radial seals   In a radial seal  the groove is designed such that the o ring snaps into the groove  with a certain pre tension  Once seated the o ring cannot leave the groove  To complete the  seal  the side of the connection containing the radial o ring and groove is inserted into a cylin   drical receiving surface  As the o ring is inserted  the cylindrical surface uniformly compresses  the radial o ring and creates the seal against all contacting surfaces  The important distinction  is that with a radial seal it is possible to rotate the objects relative to one another and still retain  a good waterproof seal  This is t
114. ridge housing  There is only one possible orientation in which  all four screws will align with those on the housing  Be sure when you insert the end cap into  the cartridge that the radial o rings do not extrude from their circular grooves  see Figure 1 61      Once the end cap is pushed in and both radial o rings are no longer visible  tighten the four  hand nuts alternately  one at a time  so they seat the cap on evenly  Only tighten the screws  until the end cap edge is approximately 1 to 2 mm away from the bottom tube edge of the  extruded cartridge housing  see Figure 1 62   This is anormal gap because the hand nuts are  pushing the entire assembly against the top face ring in order to secure the seal of the cartridge  stack to the interior top sealing surface of the cartridge housing     Compression gap is normally 1 to 2  mm wide after successfull assembly    Ensure that o rings do not  extrude from their grooves during  assembly       Figure 1 61  Final insertion of end cap into  cartridhe housing     Figure 1 62  Tighten the four screws     POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE 35       If there is no gap between the end plate and the bottom edge of the cartridge housing It likely  means that you have forgotten to install the top axial o ring on top of the CO  absorbent cartridge   If this is the case  return to Step 14  above  and install the axial o ring then resume        Make sure the OPV is in the maximum closed position     When your unit has been assembled
115. rly adjusted  the Poseidon MkVI should rest easily on the diver   s back  It should  not feel awkward or loose  but rather it should be reasonably snug and comfortable  Specific  strap adjustments will depend on what style of harness is used  but each counterlung comes  with a set of three straps that can be looped around the harness shoulder straps  securing  both counterlungs firmly to the diver   s upper chest and shoulders  When properly positioned   both counterlungs should curve over the tops of the shoulders  such that the top ends are in  line with the diver   s back  They should hug the diver   s body closely  and not float up or shift  position as the diver swims in different orientations     V    av       ah   4  fs a  F    did ASP  M atitid ITT       CHAPTER 3 PAGE 58    Counterlung strap adjustments    Besides the three large straps for attachment to the harness  each counterlung has several  additional straps used to adjust positioning  At the top of each counterlung is a single adjust   able strap that curves behind the diver   s back and attaches to the corresponding cylinder  strap  This counterlung strap is used to adjust the positioning of the top of each counterlung   At the bottom of each counterlung are two more adjustable straps  The longer of these angles  straight down for attachment to a crotchstrap or a waist strap  and is used to keep the bottom  of the counterlung securely down  The shorter strap angles laterally and attaches to the cor   respondi
116. rly inserted into the  charger  See the text for definitions of the status indicator lights     POSEIDON MKVI USER MANUAL    The Learn Cycle Activation Button  Just above the Learn Cycle indicator light is a button   Pressing the button will manually initiate a Learn Cycle  It may be pushed at any time during a  regular charge cycle to initiate a Learn Cycle     The System will require a Learn Cycle if the smart battery is fully depleted  if it has been more  than 90 days since the last Learn Cycle  or if the cell has had more than 20 charge cycles since  the last Learn Cycle  The system will recommend a Learn Cycle if it has been more than 45  days since the last Learn Cycle or if the cell has had 10 or more charge cycles since the last  Learn Cycle     Charge Cycle Indicator Light  The right most light on the charger is the charge cycle indicator   and it has the following meanings when plugged in with the smart battery inserted    e Off  The battery is being discharged as part of a Learn Cycle    e Alternating Red and Green flashing once per second  No battery detected    e Both Red and Green flashing  Battery is being charged  more green as battery is charged    e Red continuously on  Charging has failed  may require a Learn Cycle     e Green continuously on  Charge cycle completed successfully  battery is fully charged   While charging  the light will flash rapidly when the battery is discharged  and will flash more  slowly as the battery becomes more charged  As a gener
117. rminate the dive and commence a safe ascent to the surface in open circuit mode        Figure 3 7  Alarm signal area     DO NOT DIVE alert    In the upper right corner of the Alarm Signal Area is a circle with a diagonal slash through it   This symbol is the    DO NOT DIVE    Alert  and it indicates that the system is not currently ready  to be used for diving  This symbol will always be activated when the Poseidon MkVI electronics  are first turned on  while the pre dive routine is being conducted     As mentioned previously  the upper right corner  of the screen is the alarm signal area  and under  normal circumstances it should be completely  blank  It was designed this way so that a quick  glance at the screen would be all that is neces   sary to know whether any alarm conditions are  active  A blank field in the upper right corner of  the screen means all systems are functioning  properly  and all parameters are working cor   rectly  In most cases  the signals will flash when    activated  further drawing attention  Figure 3 6  Abort  and open circuit alerts     Figure 3 8  DO NOT DIVE alert     POSEIDON MKVI    General alert    The triangle symbol with an exclamation point   located in the bottom right corner of the Alarm  Signal Area will flash in synchrony with any other  parameter s  on the screen that is are inappro   priate or out of acceptable range  This signal is  intended to catch the diver   s attention  and  prompt the diver to scan the other elements on  the
118. rned  on and have sufficient gas to conduct a dive  Following proper procedure  both cylinders will have  been in the off position when Test 44 is reached  if not  gas will be wasted during Tests 24 27   which verify that the four solenoid valves draw the correct amount of power when held open      Each of these two tests will allow up to two minutes to turn on each cylinder  The bottom one   two  or three segments of the respective cylinder pressure bar graphs will flash until sufficient  pressure is detected  Figure 2 8   When the system detects sufficient oxygen pressure  it then  waits until it detects sufficient diluent pressure  Provided the oxygen cylinder pressure is  greater than 34 bar   493 psi  and the diluent pressure is greater than 51 bar   739 psi  the  automated pre dive check will pass  the pre dive test routine will continue  There is no upper  limit for cylinder pressures for these two tests  However  it should be noted that the high   oressure sensors themselves have an upper limit to the pressure they can correctly read  The  high pressure sensor for the oxygen cylinder is limited to 207 bar   3097 psi  and the sensor  for the diluent is limited to 300 bar   4410 psi  Exposing either sensor to a pressure in excess  of these limits may yield unpredictable results  Also  oxygen pressures in excess of about 135  bar   2000 psi pose a substantially increased risk of fire     POSEIDON MKVI USER MANUAL       Figure 2 8  Tests 44 and 45  confirming sufficient g
119. s  especially tests 16 31  to fail  It   s also important to make sure the battery  is not over charged  In rare circumstances  the battery might actually be charged beyond its  intended capacity  and this can also cause certain tests to fail  If there is reason to suspect that  the battery may be over charged  insert the battery and or power up the electronics  and  maintain contact across the two wet switch terminals on the back of the Primary Display   forcing the power to remain on in the event of a test failure   After several minutes with the  power on  the battery should no longer be over charged  and the automatic pre dive routine  can be re started     If the battery is properly charged  and not over charged   there are still several actions that may  correct a persistent failure of one of the PSTs  namely     e Reboot  Simply allowing the electronics to power down  after a test failure   then activating  the wet switch again to re start the automatic pre dive test routine  can often correct a  failure in one of the tests    e Reset Battery  After repeated failures of the same test  allow the system to power down  following a failed test  then remove the battery from the electronics and insert it into the  battery charger  with the charger plugged into an appropriate power supply   After leaving  the battery on the charger for a few minutes  re insert the battery into the electronics   which will re start the automatic pre dive test routine  Sometimes  this will solv
120. s control valves  A very brief summary of these PSTs is as  follows     e Test 1  Confirms the main data logger is functional   e Tests 2 9  verify the ROM  RAM and EEPROM function in all four processors   e Test 14  Confirms the battery data logger is functional   e Test 15  Confirms the firmware version is consistent across all four processors   e Test 16  Confirms the power consumption calculations are functioning properly   e Tests 17 20  Confirm the power draw of the backlight  solenoid valves  and alarms  are correct   e Test 22  Vibrator current test   e Tests 24 27  Solenoid current test   e Test 29  Speaker current test   e Tests 30 31  Cylinder HP sensor validation   e Tests 34 35  PO  sensor validation HW test   e Test 38  Depth temperature sensor validation   e Test 40  Decompression status verification   e Test 43  Mouthpiece OC test   e Tests 44 45  Sufficient oxygen and diluent to go diving test   e Test 48  Sufficient battery power to go diving test   e Test 49  Positive pressure loop test  PPLT    e Test 50  Mouthpiece CC test   e Test 53  O2 calibration test   e Test 54  Open Circuit regulator test   e Test 55  Service interval test    Test numbers 1 16 are conducted with the backlight off  allowing more precise measurement  of the power consumption of individual components  such as the various alarms and solenoid  valves  Test 17 checks the power consumption of the backlight  Figure 2 5   and the backlight  remains on for the remainder of the pre dive tests
121. s is because the absorbent duration is keyed to the amount of oxygen contained in the  oxygen supply cylinder  As long as the absorbent cartridge is replaced whenever the oxygen  cylinder is refilled  the absorbent will always out last the oxygen supply     DANGER   The CO   absorbent cartridge MUST be replaced whenever the    oxygen supply cylinder is replaced or re filled  Failure to change  the absorbent cartridge in this fashion could lead to serious injury  or death        Removing the electronics module    Unless a subsequent dive is planned soon after the previous dive  its generally good practice  to remove the Electronics Module from the breathing loop  to allow inspection of the oxygen  sensors  and also to allow moisture from condensation to dry out  Avoid removing the battery  unless the e module is dry  If the electronics are to be removed completely  both gas supply  cylinders must first be de pressurized so that the regulators can be removed  Follow the  instructions included in Chapter 3 for proper power down procedures  which include depres   surizing both gas supply cylinders     POSEIDON MKVI USER MANUAL    Replacing the water trap sponge    If a surface interval between dives is planned to extend for an hour or more  it is a good idea  to remove the sponge from the Poseidon MkVI backpack and squeeze as much water out of  it as possible  It is best to replace the sponge and cartridge  which must be removed to access  the sponge  immediately afterwards  even
122. s of confidence based on various M  factors include    C1        C2     and    C3     The last of Dil CH   these     C3     is normal  and means the system  has high confidence in the sensors  The other  levels     CO        C1     amp     C2     generate errors  and  will trigger appropriate alarms     E  Ge    GH   0  e    SR Bo     FS ca Yt Toke    Out Ge  Sal    Figure 3 16  Oxygen sensor confidence alerts  CO  C1  C2        Mouthpiece position    The area where the PO  is normally displayed serves one additional function  to communicate  the current position of the Mouthpiece  As with the PO  Setpoint and Oxygen Sensor Confi   dence warnings  this information is displayed briefly every few seconds  There are four possible  values  which are     e    cc    in the upper half of the PO  display area  mouthpiece is in the Closed Circuit position   e    oc    with    o    in the upper half          CHAPTER 3 PAGE 53    e    c    in the lower half  mouthpiece is in the Open Circuit position    e    nc    with    n    in the lower half  and    c    in the upper half  mouthpiece is not fully in either  position   or    un    with    u    in the upper half  and    n    in the lower half  mouthpiece position  is unknown      The difference between    nc        no circuit     and    un        unknown     depends on whether the  mouthpiece is reporting that neither closed circuit nor open circuit is currently established     no  circuit      or whether the mouthpiece is not reportin
123. sembly  together  The mouthpiece should be in the closed circuit position to allow the check valves to  dry on all sides        CHAPTER 4 PAGE 63    Remove the battery from the electronics Module and recharge  if necessary  Be careful not to  mix up different batteries with different electronics modules  as they are keyed to each other        Long term storage and care    Storage    If the rebreather is not going to be used for extended periods  e g   in excess of several weeks  or months   it   s important to break down and store the rebreather properly  The first step is to  follow the instructions above for procedures to follow at the end of each diving day  Once  opened  CO  absorbent cartridges cannot be stored safely for extended periods of time  so  any opened cartridges should be discarded  It is also important to ensure that all components  are clean and dry before long term storage  to avoid problems of corrosion as well as mold and  other biological cultures     The cylinders should be removed from the backpack unit and stored in a clean  dry location   This will prevent corrosion on the cylinders that may form from residual moisture or salt in the  webbing material of the cylinder straps  and will also prevent permanent deformation of the  straps and the rubber cylinder mounts located on the sides of the backpack unit  Cylinders  should be stored with valves installed and at least some pressure inside the cylinder  Be sure  to maintain proper inspections and certi
124. sing problems  The serial number of the rebrether unit is  displayed in the lower left corner of the screen  The serial number is represented in    hexadecimal      each digit can be either a numeral  0 9   or one of six letters  A F   Letters are represented as        Wet switch contacts          Figure 2 1   Connect the wet switch on the back of  the Primary Display for Power Up           CHAPTER 2 PAGE 39       ee a GE Se    Figure 2 3  When the top and bottom  segments on the display starts to blink   re connect the wet switches     Figure 2 2   nitial displays during the first  two seconds of Power up  showing  firmware version and serial number     The system waits for the diver to perform a power up confirmation procedure before proceeding   There are two steps to this procedure  First  the diver must ensure that the wet switch contacts  on the back of the Primary Display are not connected to each other  In most cases  this is  accomplished by simply releasing the fingers from the wet switch contacts  In very wet conditions   it may be necessary to blow off excess water from the back of the Primary Display  If they are  connected  the right side of the display will flash an alternating    checker board    pattern  every  other segment of the oxygen pressure bar graph  indicating that the contacts need to be dried   After the wet switch contacts have been dry for five consecutive seconds  the top and bottom  segments on the right side of the screen begin to flash  Figur
125. system then injects  diluent  air  via the diluent calibration solenoid valve  In doing so  this test calibrates the primary  sensor  and confirms that the correct gas mixtures are used in the respective cylinders     This means  that the milli volt reading from a sensor alone can   t be used to establish if an  oxygen sensor is working properly or not  The response time of an oxygen sensor differs based  on the temperature of the oxygen sensor  This means that the temperature of an oxygen  sensor can have a big effect on the success of a Pre Dive calibration     If your unit stops at test 53  try the following to resolve the problem       Verify that the gas cylinders are connected to the correct pneumatics block LP connections   DIL O2       Make sure the cylinders contain the correct gas mix     If winter diving  warm the sensors up  in your pocket    If the unit keeps failing test 58  you might have to replace one or both oxygen sensors     Once the service date is due  the diver is prompted to acknowledge  similar to    power up pro   cedure     that he she has understood that service is needed  A grace period of 4 weeks is added     Depth reading difference    When you compare the depth reading on the Poseidon MkVI display and your wrist mounted  dive computer  placing them side by side  you might see a difference in depth  This is due to  the fact that the depth sensor in the MkVI is located at the bottom of the e module  located  behind your neck and not in the displa
126. t  for an 11    adapter  the Poseidon MkVI can be bolted to the    CHAPTER 1 PAGE 13                        e     11 inch Adapter  See separate assembly manual     Figure 1 13    Align the pin bolts of the 11 inch adapter with the  upper hole and the most suitable hole of the plastic   metal  optional  back plate of the Besea harness   Fasten it using the wing nuts     POSEIDON MKVI USER MANUAL    3  Counter lungs to BCD   Harness    Attach the counter lungs to the shoulder straps of your BCD Harness using the velcro straps  on the back side of each counter lugn     Upper counter lung buckle to tank band connection     Connect the small male plastic clip on the upper part of the counter lung to the female plastic  clip attached to the tank strap on the same side as the counter lung     Adjust the position of the counter lung using the strap on each male plastic clip         Shoulder strap    Right counterlung port    Velcro counterlung connection straps   Three on each side    Over pressure checkvalve port    Figure 1 15   Lay out both left and right hand counterlungs and their water diversion manifolds        Shoulder strap    Left counterlung port    CHAPTER 1 PAGE 15    The counterlungs on the Poseidon MkVI are designed to attach to the backpack straps and to  be moveable along the straps  An upper adjustable length quick connect buckle anchors the  top end of the counterlungs to the gas processor  Figure 1 16   Three velcro straps on the  back of each counterlung  Figure 
127. ter  representative or other qualified personnel  Always use oxygen       l compatible lubricants when servicing o rings and seals  Open    a j   oxygen cylinder valves slowly  Keep the oxygen regulator and    g   cylinder valve protected from the environment when removing and  storing  Never over fill the oxygen cylinder  as higher pressures  enhance the risk of fire  Failure to follow these precautions could  result in fire  explosion  serious injury  and death     Figure 1 35  Align the oxygen clean Figure 1 36  Align the diluent    Connect the oxygen clean  white marking  High Pressure sensor located on the E Module to a  HP port on the oxygen clean 1st stage  See Figure 1 35    Connect the second High Pressure sensor located on the E Module to a HP port on the diluent  1st stage  See Figure 1 36     WARNING     Never attempt to adjust the interstage pressure or otherwise  tamper with either of the first stage regulators  Excessive pressure  can cause the system to fail     POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE 25       9  LP hose  amp  HUD to mouthpiece   g  DH IJ Figure 1 38      Twine the HUD cable around      the braided 90 cm hose    L Kaz    Connect the long braided 90 cm Low Pressure hose to a IP port on the diluent 1st stage   See Figure 1 37  Make sure the connection o ring is in place and without damages     Connect the long braided 90 cm Low Pressure hose to the 9 16    connection on the mouth   piece  See Figure 1 39     Twine the HUD cable located on the E
128. ters    OC    flash in the upper left corner of the screen along with  the Open Circuit Bailout icon  indicating the need change the mouthpiece back to OC mode   When the mouthpiece is switched  the word    bREAtHE     scrolls across the two PO  value  digits in the upper right corner of the screen  This prompts the diver to test the function of the  open circuit regulator  After inhaling several breaths from the regulator  the test passes     POSEIDON MKVI USER MANUAL    Service interval check  test 55     The final test  Test 55  Figure 2 12  is also the simplest  This test merely ensures that the  rebreather unit does not require servicing  Each rebreather unit must be brought to a qualified  Poseidon Service Center at least once every two years  to receive updates and make any  necessary repairs or adjustments  When Test 55 is displayed  the number in the lower right  corner of the screen  adjacent to the small clock icon  indicates the number of weeks remaining  before servicing will be required  When this value gets low  return the rebreather to an authorized  Poseidon service center for maintenance  Upon passing Test 55  the pre dive routine is complete     Once the service date is due  the diver is  prompted to acknowledge  similar to    power  up procedure     that he she has understood  that service is needed  A grace period of 4  weeks is added     Figure 2 12   Test 55     Service interval check        Cleared to dive    Under normal circumstances the Poseidon MkV
129. tes  then change the mouthpiece position   If you hear the audio alarm  then change the mouthpiece position NOW     If the Heads Up Display light on the mouthpiece is STEADY ON  then ASCEND at a safe  and controlled rate to the surface     If the Heads Up Display light on the mouthpiece FLASHES  then STOP  look at the LCD  screen  In general this is a reminder to you to keep track of your PO   which is shown in the  upper left field of the display  However  other data are included on the display  including  directional arrows that advise you to go up  ascend  or go down  descend   The latter  would be active and flashing  for example  if you have accumulated decompression debt  and have ascended through a decompression stop  More information on the functionality  of the display is presented in Chapter 3        CHAPTER 3 PAGE 61    When in doubt  bail out   switch to open circuit  OC  and ascend in a controlled manner to  the surface     The default setpoint control algorithm is designed to allow for hands off control of the  system PO  during all phases of a dive  The Poseidon MkVI uses a proprietary method that  begins with a default control setpoint on the surface of 0 5 bar and gradually increases PO   to a maximum automatic value of 1 2 bar at a depth of 15 m   50 ft  Beyond this depth the  system will automatically control to a setpoint of 1 2 bar to the maximum operating depth  of the rig at 40 m     POSEIDON MKVI USER MANUAL       CHAPTER 4 PAGE 62    Chapter 4   Post 
130. the  secondary sensor  The use of oxygen to perform Test 49  PPLT  helps this test complete  properly  because the breathing loop will have already been pre charged with oxygen  After  the calibration constants for oxygen are established  the system then injects diluent  air  via the  diluent calibration solenoid       CHAPTER 2 PAGE 45    valve  In doing so  this test both calibrates the sensors  and confirms that the correct gas  mixtures are in the respective cylinders     This test is by far the most important of all the pre dive tests  as it is determining whether the  oxygen sensors are providing true values for the partial pressure of oxygen  PO    Failure of this  test can occur for a number of reasons  all of which the user should be familiar with  Most  causes relate directly to the oxygen sensors themselves     either bad or aged sensors failing  the test  or the presence of condensate on the sensors from a prior dive  If Test 53 fails persis   tently  verify that the gas cylinders are connected to the correct regulators  and make sure they  contain the correct gas mixtures  If the test continues to fail  one or both of the oxygen sensors  may need replacing  When changing oxygen sensors  it   s very useful to keep track of which  oxygen sensor was placed in which position  by noting the individual oxygen sensor serial  numbers     Figure 2 11   Test 53   Oxygen Sensor calibration        Open circuit regulator function  test 54     After completing test 53  the let
131. the nut locks against the lower thread flange on the manifold  Do NOT over tighten  or use any tools  as this may result in stripping the threads and ruining both the connector and  the manifold port     Repeat these steps with the right rear breathing hose  so the rig is as shown in Figure 1 21     POSEIDON MKVI    USER MANUAL    breathing hose        The next step is to attach the rear right breathing hose to the    back     see Figure 1 25   of the  water diversion manifold  Insert the hose as shown in Figures 1 25 Follow the same procedure  for O ring and sealing surface inspection and lubrication as previously described for all hose  connections  Tighten  out do NOT over tighten  the right rear hose nut onto the right rear  manifold thread     Repeat these steps for the left water diversion manifold and counterlung  Figure 1 26             3 Figure 1 25  Attach the right counterlung  aie water diversion manifold to the right rear    Radial    o ring seal  Sealing surface for  radial    o ring seal    Retainer nut capture thread    CHAPTER 1    PAGE 19       Figure 1 26  Attach the left rear breathing  hose to the left counterlung shoulder port     SS       POSEIDON MKVI USER MANUAL    6  Attaching the cylinders    Attach the two tanks to the canister housing tank connections and secure them with the tank  straps     If you imagin wearing the unit on your back  the oxygen  white valve knob  tank should be  mounted on your right hand side and the diluent  black valve knob 
132. the oxygen  cylinder is 185 bar   2000 psi  The reason is twofold  first  the risk of oxygen fire increases  sharply at higher pressures  and second  using a greater supply of oxygen increases the risk  that the CO  absorbent cartridge may not be sufficient to last for the entire dive on a single fill           Diluent cylinder    Oxygen cylinder    Figure 1 12  How the rig should appear when properly assembled        CHAPTER 1 PAGE 10    It is also extremely important to NOT replace either of regulators that are provided with the  Poseidon MkVI  The oxygen cylinder  valve  and regulator have been cleaned for oxygen  service at the rated cylinder pressure  using regulators or valves not so cleaned dramatically  increases the risk of oxygen fire and or explosion  More importantly  the provided regulators  have been adjusted with a lower interstage pressure for use with the oxygen and diluent  solenoid valves  Using different regulators with higher interstage pressures will cause the  solenoid valves to fail  and may lead to permanent damage     POSEIDON MKVI USER MANUAL CHAPTER 1 PAGE 11       Filling the cylinders    From a diving logistics perspective  the Poseidon MkVI differs from normal Scuba in that it  uses two separate gas supplies  a pure oxygen supply and a    diluent    Supply  Pure oxygen is    necessary for the control system to exactly replace the oxygen consumed by metabolism  The DANGER    definition of a diluent gas in a rebreather is any gas that serves to di
133. the water with  perfect buoyancy  while breathing continuously     The quickest and easiest way to fine tune buoyancy with a rebreather is via addition and  removal of gas to or from the breathing loop  To increase buoyancy slightly  a small amount of  gas can be added to the breathing loop via the ADV  either by manually engaging the purge       CHAPTER 3 PAGE 59    button  or by making an especially deep inhaled breath   To decrease buoyancy slightly  one  need only exhale through the nose to vent gas out of the breathing loop  except when certain  kinds of full face masks are used      New rebreather divers often have the most difficulty in very shallow water  where a slight  change in depth yields a proportionally large change in displacement  and  hence  buoyancy    This is especially true when the diver begins to ascend  which causes the counterlungs to  expand  leading to increased buoyancy  leading to further ascents  and expanding loop  volume  This can lead to a    run away    ascent that can be difficult to control  For this reason   it   s useful practice for rebreather divers to be in the habit of venting gas through the nose  whenever ascending  particularly from very shallow depths     Venting water from the loop    Even if a diver is very careful to prevent water from entering the breathing loop  there will always  be some water collecting due to condensation  Most of this will form on the    exhalation    side  of the breathing loop  between the mouthpiece an
134. ther designs  Among the breakthroughs are     e Extreme compactness     at only 18 kg   40 Ibs  ready to dive    and 9 kg   20 Ibs     travel weight    the MkVI is one of the world   s smallest rebreathers  Take it as  carry on baggage when flying  Enjoy the freedom  Yet it is not    lightweight    in terms  of performance     you get three depth independent hours1 of near silent diving     e The first truly auto calibrating and auto validating rebreather  The MkVI uses a  patented automated method to verify that the oxygen sensors are working  properly at all times     both before and during a dive     e An intelligent battery     it   s your personal power and data storage system  Plug it  in and the system knows it   s you  as well as your dive history  including repetitive  dive information  Remove the modular battery  put it in your shirt pocket  and take  it home to recharge it  When you are ready to dive  take the battery from the  Charger and plug it into the rebreather     e True plug and play carbon dioxide absorbent cartridges  Simple and fast to use   Pre packed axial cartridges of SofnoDive   797 absorbent give you 3 hours  minimum diving range1 and can be changed out and replaced in seconds     e The world   s most advanced rebreather interface     The MkVI contains five separate  warning and advisory systems so that you get the information you need to manage  your dive without the task loading normally associated with rebreathers  The main  Poseidon MkVI data i
135. tion following the previous steps  If  for some reason  the mouthpiece is not  in the OC position when Test 43 appears on the screen  the mouthpiece vibrator will activate  along with the HUD and battery LEDs and audio alarm  The letters    OC    will appear in the  upper left corner of the LCD screen and  along with the    Open Circuit    icon  diver with bubbles    will flash until the mouthpiece switch is placed in the OC position  Figure 2 7   The system allows  the diver up to two minutes to make this switch  The    Open Circuit    icon will continue to display  on the LCD until Test 50  when the mouthpiece needs to be switched to the closed circuit  CC   position     If the mouthpiece appears to be in the open circuit position  but Test 43 does not pass  then make  sure that the mouthpiece switch is fully in the OC position  If the test still will not pass  then  make sure the HUD is positioned correctly on the mouthpiece  and is not twisted or otherwise  moved from its proper position  If no amount of repositioning of the mouthpiece lever or HUD  allows the system to pass Test 43  then contact an authorized Poseidon Service Center     Figure 2 7    Test 43  Mouthpiece in OC position    displaying Open Circuit icon to indicate  that the mouthpiece must be placed in  the Open Circuit position           CHAPTER 2 PAGE 42       Oxygen and diluent cylinder supplies  test 44  amp  45     Tests 44 and 45 determine whether the Oxygen and Diluent cylinders  respectively  are tu
136. to sensor base     Ce    Figure 4 3   Press button to lock tool onto sensor base            Hold the button and pull the oxygen sensor  base out of the electronics module      Figure 4 4   Pull the sensor out while pressing button     POSEIDON MKVI USER MANUAL    With the Oxygen Sensor removal tool snapped into the oxygen sensor base  press the plunger  button with the thumb  Figure 4 3  to lock it in  While continuing to press the button  pull the  tool away from the electronics module  and the oxygen sensor base  with oxygen sensor  attached  will slide out easily  Figure 4 4      Whenever an oxygen sensor is changed  the serial number of the new sensor  and its position   Primary or Secondary  should be logged  This should also be done whenever the positions of  the two oxygen sensors are reversed  Doing so allows the history of the sensor to be tracked  over time and correlated with the logged data associated with that sensor  Such information  can be extremely valuable for detecting when a sensor is nearing the end of its life  The serial  number of the sensor is printed on the sensor label  as shown in Figure 4 5     Once the oxygen sensor base and sensor are removed from the electronics module  the elec   trical connection can be unplugged from the back of the sensor  Detach the tool from the  sensor base by releasing the plunger button and pulling it off  The oxygen sensor can then be  removed from the sensor base by unscrewing it  Figure 4 5      Attach the new oxyge
137. too high    11 Current  too low   12 Current  too high    11 Current  too low   12 Current  too high    APPENDIX PAGE 73    SOLUTION    1  Standard Response  2  If test failure persists  or if no faint    click    sound  from the main electronics module can be heard at the start of this test  contact  an authorized Poseidon Service Center for repair     1  Standard Response  2  If test failure persists  or the Battery LED does not  turn on during this test  contact an authorized Poseidon Service Center for  repair     1  Standard Response  2  If test failure persists  or the HUD does not vibrate  during this test  contact an authorized Poseidon Service Center for repair     1  Standard Response  2  If test failure persists  or the HUD LED does not  turn on during this test  contact an authorized Poseidon Service Center for  repair     1  Standard Response  2  If test failure persists  or if no faint    click    sound  from the main electronics module can be heard at the start of this test  contact  an authorized Poseidon Service Center for repair     1  Standard Response  2  If test failure persists  or if no faint    click    sound  from the main electronics module can be heard at the start of this test  contact  an authorized Poseidon Service Center for repair     1  Standard Response  2  If test failure persists  or if no faint    click    sound  from the main electronics module can be heard at the start of this test   contact an authorized Poseidon Service Center for repa
138. ttery computer keeps track of how long  it has been since the last time it has undergone a full Learn Cycle  If that time exceeds a certain  value  the computer will advise the user to perform an optional Learn Cycle  If the time since  the last Learn Cycle is very long  the computer may automatically initiate a Learn Cycle  The  Learn Cycle takes approximately 8 hours to complete  Once a Learn Cycle has been initiated   it can only be stopped either by successful completion of the Learn Cycle or by physically  removing the battery from the charger  not recommended         The Learn Cycle indicator light has the following meanings when plugged in with the smart   battery inserted    e Off  Learn Cycle not needed or not in progress    e Alternating Red and Green flashing once per second  Learn Cycle is recommended    e Both Red and Green simultaneously flashing once per second  Learn Cycle is in progress    e Red continuously on  Learn Cycle has failed  often caused by power loss or user  intervention     e Green continuously on  Learn Cycle successfully completed        CHAPTER 1 PAGE 6       a    SS sir ie 3  p           Learn cycle activation button    Power status on off    Learn cycle status    Charge status    Figure 1 9   Left    nserting the smart battery into the desktop charging station  The battery  screw posts align with the vertical slots  the battery is pushed down until the contact post  mates with the offset cylindrical receiver hole   Right   The battery prope
139. ular plug and play carbon dioxide  filter system  It is equipped to handle Molecular Products SofnoDive   797 axial flow pre   packed cartridges  Procedures for changing out the cartridge are presented in detail below  in the Cartridge Housing discussion     Gas injection module    In a fully closed rebreather like the Poseidon MkVI  oxygen is consumed by the diver and a  mechanism must be provided for replacement of that used oxygen  Otherwise the mixture  will slowly be depleted to dangerously oxygen low levels  hypoxia   The MkVI is designed to  maintain the partial pressure of oxygen  PO   well above hypoxic levels  and to also prevent  it from becoming too high  hyperoxia   This is achieved by a control system that uses sensors  that are responsive to the partial pressure of oxygen and a mechanism for the addition of  pure oxygen to the system when the sensors indicate that the oxygen level is below the  target value  known as the PO     setpoint     The gas injection module in the Poseidon MkVI  does this and much more  In a patented design  this module provides the mechanisms for  not only adding pure oxygen to make up for metabolized gas  but also to automatically cali   brate the oxygen sensors prior to diving  as well as validate the sensors during the course of  each dive     Electronics module    The electronics module provides a single plug and play component that includes the previ   ously described gas injection module and the smart battery  It also includes th
140. ure for 20 seconds  If any of the  four solenoid valves are leaking  the pressure inside the breathing loop will gradually rise   Assuming no increase in loop pressure is detected during this 20 second period  the second  metabolic solenoid valve is used to inflate the breathing loop to a higher internal pressure   When this happens  the counterlungs will be tightly inflated  and the internal pressure should  be slightly less than the cracking pressure of the over pressure relief valve on the bottom of the  right counterlung  when that valve is adjusted to its maximum cracking pressure  The system  then monitors the loop pressure for the next 20 seconds to determine whether the pressure  decreases  as by a leak in the breathing loop     IMPORTANT     While Test 49 is being conducted  be careful not to manipulate  the counterlungs too much  or do anything that might affect  the internal loop pressue  independently of the gas injected by  the metabolic solenoid valves  The test can be performed while  wearing the rebreather  as long as there is not too much motion  or instability of the counterlungs  It   s recommended that you don   t  let the mouthpiece and loop rest against the counter lungs during  test 49  as this might cause the test to fail     Closed circuit mouthpiece position  test 50     The mouthpiece should have been left in OC position following Test 43  as indicated by the     Open Circuit    icon on the Primary Display  At Test 50  Figure 2 10   the icon disappe
141. ut also validate that the oxygen sensor readings are  correct and accurate  Although there are many alarm systems built into this system  it is always  good practice for divers to regularly monitor the PO  value on the Primary Display screen  to  ensure that it is within limits  and that the value itself is not flashing     Monitoring gas supplies    The next most important parameters to monitor are the gas supplies  represented as bar  graphs on the left and right sides of the Primary Display  In particular  it is important to make  sure that the Air     Dil     pressure graph is not flashing  The electronics system will constantly  calculate whether there is enough air supply remaining to allow a safe open circuit bailout to  the surface  If there is not enough air to allow a safe open circuit bailout to the surface  the    Up  Arrow    will be displayed on the LCD Display  indicating that the diver should ascend to a shal   lower depth     The oxygen supply pressure should also be monitored to ensure there is a sufficient quantity  of oxygen remaining in the oxygen cylinder to complete the remainder of the dive in closed   circuit mode  Because these values change very slowly throughout the course of a typical  rebreather dive  there is a tendency to ignore them  As with other important parameters  there  will be warnings issued in case the oxygen supply pressure gets too low  but nevertheless  the  diver should be in the habit of monitoring this value regularly        CHA
142. ver via the Primary Display  It consists of a backlit liquid crystal  display  LCD   with pre printed numerals and symbols  and provides the diver with important  information concerning sensor readings  system messages  decompression status  and other  data during the course of the dive  It is extremely important that all Poseidon MkVI divers  understand how to read the information contained in the Primary Display  particularly concerning  various alarm conditions    Before even turning the Poseidon MkVI electronics on  it is useful to understand the general  layout of the Primary Display  and the logic behind how the information is organized  The  display is arranged in six regions  each presenting different kinds of information  The most  important region is the upper right corner of the screen  1 in the illustration   which contains  icons for alarm conditions  Under normal circumstances  this region should be blank  The  alarm condition icons  described in more detail below  are designed to symbolically represent  the nature of the problem  and most of them will flash when activated  This should be the first  part of the screen that a diver should glance at when monitoring the Primary Display  as it will  be immediately obvious if there are any alarm conditions  and what they are     CHAPTER 3 PAGE 49    POSEIDON MKVI USER MANUAL       Stop ascending  Ascent indicator        Ceiling alert        ABORT alert  DO NOT DIVE    Oxygen partial pressure Oxygen gas pressure    El
143. w in detail  are the pre packed Sofno   dive   797 manufactured by Molecular Products and available through Poseidon authorized  dive centers and dealers     With the cartridge removed  inspect the cartridge housing end plate  Figure 1 53  for scratches  and gouges to any of the o rings or o ring sealing surfaces  Replace o rings if required and  re lubricate  Lift uo and remove  Figure 1 53  the bottom annular sponge water trap  This  sponge should be relatively dry when diving in relatively warm water  but may be damp or  soaked when diving in cold water  Improper use of the rebreather may also allow water to enter  the breathing loop  and possibly reach the cartridge housing  While the water diversion mani   folds will trap most of the water  acrobatic swimming  rolls  flipping from head down to head up  orientation etc  can defeat the system if the user is not paying attention  causing water to reach  the sponge     It is good practice to rinse  disinfect  and dry the lower absorbent sponge following a dive  Any  disinfectant solution that doesn   t have a negative effect on plastics and or rubber can be used  for this purpose  Poseidon recommends using a disinfectant called Gigasept    or a disinfec   tant called Virkon        CHAPTER 1 PAGE 32    At this point it is time to load a fresh SofnoDive   797 CO  absorbent cartridge  Cartridges are  available two to a pack  Figure 1 54  and are sealed in an air tight envelope for long term  storage  As previously described  ea
144. with the smooth cylindrical outer face  Be sure when you insert the end cap into the cartridge  that the radial o ring does not extrude from its circular groove  The top radial o ring in the end  cap should insert smoothly into the CO  cartridge until it is no longer visible and the bottom of  the cartridge is flush with the ledge just above the two main radial o rings     Inspect the cartridge top plate  Figure 1 57  and its radial and axial o ring seals  If any of the  O rings are damaged  cut  or gouged  replace them  Ensure that each o ring is lubricated and       CHAPTER 1 PAGE 33    Radial o ring seal to top of  SofnoDive   797 cartridge       Figure 1 58   Insert cartridge top plate and inspect  the face seal o ring     Figure 1 59   Inspect and lubricate cartridge housing  end cap radial sealing surface     that there is no debris  dust  sand etc  on any of the o rings  Insert the cartridge top plate into  the top of the SofnoDive   797 cartridge  The    top    end is the one with the projecting ribs along  the circumference of the end of the cylinder  Be sure when you insert the cartridge top plate  into the cartridge that the radial o ring does not extrude from its circular groove  The radial  o ring in the cartridge top plate should insert smoothly into the CO  cartridge until it is no  longer visible and the top rim of the cartridge is flush with the ledge just above the radial o ring   Unlike the radial o ring that seats in the bottom end of the cartridge  the to
145. y it self     C1 alarms on land    When your Poseidon MkVI has passed the Pre Dive check you should always set the mouth   piece DV switch to its Open Circuit  OC  position     If you set it in Closed Circuit  CC  position  you will most likely get a C1 alert  This is normal     POSEIDON MKVI USER MANUAL    The reason for this alert is  that when the unit is turned on  and the mouthpiece is in CC mode   it will perform sensor validations  checking the PO  value  even when it is on land  If the differ   ence in PO  value is too small  when the unit compares the PO  value from the last validation  with the PO  reading from the most recent validation  the unit assumes that the primary oxygen  sensor is  frozen    and giving a false reading     If you get a C1 alert  when your unit is on land  do the following to turn the alert of     1  Set the DV switch to CC mode    2  Breath on the loop  to change the PO  value    3  Continue breathing on the loop until the next successfull sensor validation is performed    about 2 min  maximum    4  When the C1 alert goes away  set the DV switch in OC mode     If the C1 alert persists  after you   ve done the 4 steps above  then there   s something else  causing the C1 alert     Note   The C1 alert will NOT go away by switching the mouthpiece DV switch to OC mode only     Hyperoxic linearity test    When you descend and reach 6 m  20 swf  depth  the Discovey will do a Hyperoxic Linearity  test  The reason for the test is to make sure that th
146. ystem monitoring    Merely understanding how to read and interpret the information presented on the Poseidon  MkVI Primary Display is only the first step  All divers must learn to monitor the Primary Display  and alarm systems regularly throughout the dive  In addition to the parameters monitored  during an open circuit scuba dive  e g   depth  cylinder pressure  decompression status   a  closed circuit rebreather diver must also monitor other variables  such as the PO  of the  breathing gas and the remaining battery life  The MkVI is designed to make the task of monitoring  these parameters as easy and straightforward as possible  and alarm systems have been  incorporated to alert the diver when these parameters drift out of safe range  Nevertheless  it  is vitally important to the safety of the diver that good system monitoring habits be developed     Monitoring the PO  value    The most critical parameter to monitor on any closed circuit reobreather is the oxygen partial  pressure in the breathing loop  The most dangerous aspect of closed circuit rebreathers is the  fact that the oxygen concentration in the breathing gas is dynamic and can change  Consider   ing the lack of reliable physiological warning symptoms for impending hypoxia or CNS oxygen  toxicity  and the severity of these maladies while underwater  the importance of frequent PO   monitoring should be obvious  Fortunately  the Poseidon MkVI is designed to not only monitor  the PO  value in the breathing loop  b
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Manual de usuario - Sistema de Solicitudes de Habilitaciones  Manual - Penguin Computing  PDF : LQPB12 User Manual  DeLOCK USB 2.0 Ethernet adapter    Copyright © All rights reserved. 
   Failed to retrieve file