1. REF IS 4000 IDS iSYS Intact PINP immunodic Instructions for Use IN VITRO DIAGNOSTIC CE Intended Use For In Vitro Diagnostic Use The IDS iSYS Intact amino terminal propeptide of type procollagen Intact PINP assay is intended for the quantitative determination of intact PINP in human serum or plasma on the IDS iSYS Multi Discipline Automated Analyser Summary and Explanation An important step in the bone formation process is synthesis of type collagen which is the major organic component in bone matrix During collagen synthesis propeptides are released from both the amino and the carboxyterminal parts of the procollagen molecule These propeptides are secreted into the blood circulation The amino terminal propeptide of type procollagen PINP is probably the most specific and sensitive marker of bone formation PINP is a particularly useful marker for monitoring the efficacy of osteoporosis therapy with anabolic agents but it is also one of the best bone turnover markers for monitoring the efficacy of antiresorptive therapy 45 Method Description This assay is based on chemiluminescence technology Samples and calibrators are diluted in a diluent and a portion of this is incubated with a biotinylated anti PINP monoclonal antibody an acridinium labelled monoclonal antibody streptavidin labelled magnetic particles and an assay buffer The magnetic particles are captured usi
2. lt 0 1 1 bottle 16 8 mL BUF Buffer containing mouse and bovine proteins with sodium azide as preservative lt 0 1 1 bottle 4 95 mL DIL Buffer containing sodium azide as preservative lt 0 1 1 bottle 23 7 mL Calibrators CAL A CAL B Lyophilised buffer containing BSA with lt 1 0 sodium azide as preservative lt 0 05 reconstituted 2 each of 2 concentration levels 1 0 mL per bottle Mini CD Contains IFU for IDS iSYS reagents control ranges and XML files Materials Required But Not Provided IDS iSYS Multi Discipline Automated Analyser IS 310400 IDS iSYS PINP Control Set IS 4030 2 x 1 0 mL each of levels 1 2 and 3 IDS iSYS Cuvettes Cube IS CC100 box of 960 cuvettes IDS iSYS System Liquid IS CS100 5 L container ready to use IDS iSYS Wash Solution S CW100 10 L container ready to use IDS iSYS Triggers Set A and B IS CT100 2 x 250 mL per bottle ready to use IDS iSYS Cartridge Check System IS 6010 ready to use IDS iSYS Sample Cups 500 pL IS SC105 Procedure Reagent Cartridge The reagents provided in the cartridge are ready to use The Analyser automatically performs the mixing of magnetic particles to maintain homogeneity Before a new cartridge is loaded on board the Analyser mix the magnetic particles container by brisk rotation motion Avoid foam formation The barcode is read when the cartridge is loaded on the reagent tray
3. If the label cannot be read by the Analyser barcode reader a manual procedure exists to enter the barcode data see the IDS iSYS User Manual Load the cartridge on the reagent tray and wait for at least 40 minutes before starting the assay If the cartridge is removed from the reagent tray store the cartridge upright at 2 8 C in the dark Calibrators The Intact PINP calibrators are lyophilised Reconstitute immediately before use Add 1 0mL of distilled or deionised water to each bottle Replace the stopper Leave for 10 minutes to reconstitute with occasional gentle mixing by hand Avoid formation of foam Pipette approximately 200 uL of calibrators and controls into sample cups and place on the machine Discard the material in the sample cups after use Proceed according to the instructions of the IDS iSYS User Manual DO NOT return material to the calibrator vial IS 4000PLV02 2009 12 22 English Page 2 4 Analyser Calibration The two Intact PINP calibrators are required to perform the adjustment of the master curve The calibrators are supplied with the kit and calibrators from another lot must not be used Note that to perform a master curve adjustment controls MUST be run at the same time as the calibrators All data required for the calibration of the cartridge batch can be found on the mini CD Use calibrator levels A and B to adjust the master curve to the reagents on board the Analyser Check for the presence of an I
4. be separated as soon as possible after collection Store samples at 20 C Avoid repeated freeze thaw of samples Samples containing particulate matter must be centrifuged before performing the assay Do not use heat inactivated samples To minimise possible evaporation effects samples calibrators and controls should be measured within 3 hours of being placed on the Analyser Before performing assays make sure that samples calibrators and controls are at room temperature 20 25 C Note Some sample collection tubes that are commercially available might affect the results of testing in particular cases It is recommended to follow the instructions of the tube manufacturer especially when processing samples in primary tubes Procedure Materials Provided Reagent Cartridge MPS3 Magnetic particles coated with streptavidin in phosphate BSA buffer with sodium azide as preservative lt 0 1 1 bottle 2 3 mL Anti PINP monoclonal antibody labelled with an acridinium ester derivative in buffer containing bovine and mouse proteins with sodium azide as preservative lt 0 1 1 bottle 4 95 mL Device code IS 4000 REF IS 4000 IDS iSYS Intact PINP immunodiagnosticsystems Instructions for Use IN VITRO DIAGNOSTIC CE Ab BIO Anti PINP monoclonal antibody labelled with biotin in buffer containing mouse and bovine proteins with sodium azide as preservative
5. de as a preservative which may react with lead copper or brass IS 4000PLV02 2009 12 22 English Page 1 4 plumbing to form highly explosive metal azides On disposal flush with large volumes of water to prevent azide build up Handling Precautions Apart from the calibrators which are lyophilised the reagents provided in the kit are ready to use Refer to the Calibrator section of the procedure for reconstitution methodology Before a new cartridge is loaded onto the Analyser the magnetic particle container requires mixing by the operator with a brisk rotation motion This will resuspend the magnetic particles that have settled during shipment It is very important to avoid any foam formation Shelf Life and Storage of Reagents Prior to first use store the cartridge and the calibrators in an upright position in the dark at 2 8 C Do not freeze the cartridge or the calibrators Reagent shelf life Cartridge Calibrators Before opening at 2 8 C To the expiry date Cartridge 28 Days N A after opening at 2 8 C Calibrators N A 4 weeks after reconstitution at 2 8 C On board the Analyser 7 Days 3 hours Continuous on board stability Sample Collection and Storage The assay should be performed using serum standard sampling tubes or tubes containing serum separating gel or plasma lithium heparin sodium heparin ammonium heparin or potassium EDTA samples Samples should
6. don Tyne amp Wear NE35 9PD England Tel 44 191 519 0660 Fax 44 191 519 0760 e mail info uk idsplc com www idsplc com USA Immunodiagnostic Systems Inc 8425 N 90th Street Suite 8 Scottsdale AZ 85258 Tel 1 480 278 8333 Fax 1 480 836 7437 IS 4000PLV02 2009 12 22 English Page 4 4 e mail info us idsplc com www idsplic com Germany Immunodiagnostic Systems GmbH IDS GmbH Mainzer Landstrasse 49 60329 Frankfurt am Main Tel 49 69 3085 5025 Fax 49 69 3085 5125 e mail info de idsplc com www idsplc com France Immunodiagnostic Systems IDS 153 Avenue D Italie 75013 Paris France Tel 0 1 40 77 04 50 Fax 0 1 40 77 04 55 e mail info fr idsplc com www idsplc com Scandinavia Immunodiagnostic Systems Nordic a s IDS Nordic a s Marielundvej 30 2 Sal 2730 Herlev Danmark Tel 45 44 84 0091 e mail info nordic idsplc com www idsplc com Belgium Immunodiagnostic Systems S A Rue E Solvay 101 4000 Li ge Belgium Tel 32 4 252 26 36 Fax 32 4 252 51 96 e mail info be idsplc com www idsplc com Device code IS 4000
7. ing levels of intact PINP should be measured Other suitable control material can be used in addition to the IDS iSYS Intact PINP Control Set Controls should be tested at or near the beginning of every run containing patient samples and also during calibrations or according to local regulations It is recommended that the controls be routinely run in duplicate Laboratories should test controls at least once per shift Refer to the IDS iSYS Intact PINP Control Set IFU for preparation and handling instructions Determination of Sample intact PINP levels Process samples according to the IDS iSYS User Manual Device code IS 4000 REF IS 4000 IDS iSYS Intact PINP immunodiagnosticsystems Instructions for Use IN VITRO DIAGNOSTIC CE Calculation of Results The intact PINP concentration of each sample is calculated automatically The display of the concentrations screen or printed is produced upon user request The IDS iSYS Intact PINP Assay uses a 4 parameter logistic curve fit 4PL to calculate the intact PINP concentrations Measurement Range Reportable Range The reportable range of the assay is 2 230 ng mL Any value that reads below 2 ng mL should be reported as lt 2 ng mL Dilution Samples with intact PINP concentrations above the reportable range should be diluted manually with a low concentration human serum sample in a ratio of 1 in 2 The results for diluted samples
8. ions Samples were prepared by diluting a high patient sample with a low patient sample prior to assay Predicted Measured E Concentration Concentration Variation ng mL ng mL ng mL 5 8 7 6 1 8 30 5 31 5 0 9 3 55 3 55 0 0 2 0 80 0 78 4 AG 2 104 8 101 9 2 9 3 129 6 127 2 2 4 2 154 3 158 0 3 7 2 179 1 179 0 0 0 0 203 8 204 6 0 7 0 Device code IS 4000 REF IS 4000 IDS iSYS Intact PINP immunodiagnosticsystems Instructions for Use IN VITRO DIAGNOSTIC CE Method Comparison The IDS iSYS Intact PINP Assay was compared against a recognised radioimmunoassay for the quantitative determination of intact PINP following CLSI EP 9A2 Method Comparison and Bias Estimation Using Patient Samples A total of 363 samples selected to represent a wide range of intact PINP concentrations 8 7 230 ng mL was assayed by each method Linear regression analysis was performed on the comparative data IDS iSYS 0 86 RIA 1 9 95 Cl of the slope and intercept were 0 84 to 0 88 and 3 4 to 0 3 respectively correlation coefficient r 0 98 Specificity The specificity was assessed with the following analytes Analyte Cross Reactivity Intact PINP 100 Rat mouse PINP lt 0 01 PIINP lt 0 03 N mid Osteocalcin lt 0 02 Bibliography 1 Risteli J Risteli L 2006 Products of bone collagen metabolism In Seibel MJ R
9. must be multiplied by the dilution factor 2 and corrected for the concentration of the low sample Limitations of Use 1 As in the case of any diagnostic procedure results must be interpreted in conjunction with the patient s clinical presentation and other information available to the physician 2 The performance characteristics of this assay have not been established in a paediatric population 3 The following substances do not interfere in the IDS iSYS Intact PINP Assay when the concentrations presented in the following table are below the stated threshold Potentially Interfering Threshold Agent Concentration Lipid 2803 mg dL Bilirubin 200 mg dL Haemoglobin 500 mg dL Biotin 300 nmol L 4 The hook effect was tested using concentrations of PINP up to 2350 ng mL No hook effect was observed Expected Values Each laboratory should determine ranges for their local population The following range was determined using the IDS iSYS Intact PINP Assay and is provided for information only The 95 reference interval for normal adults collected from 150 apparently healthy adults was calculated by a non parametric method following the NCCLS guideline C28 A2 How to Define and Determine Reference Intervals in the Clinical Laboratory Normal Adults 27 7 127 6 ng mL n 150 Performance Data Representative performance data are shown Results obtained at individual laboratories may vary Sensitivi
10. ng a magnet and a wash step performed Trigger reagents are added and the resulting light emitted by the acridinium label is directly proportional to the concentration of intact PINP in the original sample Warnings and Precautions The IDS iSYS Intact PINP is for in vitro diagnostic use only and is not for internal use in humans or animals This product must be used strictly in accordance with the instructions set out in these Instructions for Use IFU IDS Ltd will not be held responsible for any loss or damage except as required by statute howsoever caused arising out of non compliance with the instructions provided CAUTION This kit contains material of animal origin Handle kit reagents as if capable of transmitting an infectious agent Appropriate precautions and good laboratory practice must be used in the storage handling and disposal of the kit reagents Disposal of kit reagents should be in accordance with local regulations Sodium Azide Xn Harmful Calibrators CAU contain sodium azide NaNs gt 0 1 w w lt 1 R22 Harmful if swallowed R52 53 Harmful to aquatic organisms may cause long term adverse effects in the aquatic environment S46 If swallowed seek medical advice immediately and show this container or label 36 37 Wear suitable protective clothing and gloves S60 This material and or its container must be disposed of as hazardous waste Some reagents in this kit contain sodium azi
11. ntact PINP cartridge on the reagent tray and the availability of the cartridge master curve in the database If the data for the lot of calibrators is not available on board the Analyser load the data using the mini CD provided with the calibrator Start the immunoassay calibration on the IDS iSYS Analyser according to the IDS iSYS User Manual The calibration is carried out in triplicate RLU CVs of gt 5 will result in a failed calibration One replicate may be removed to meet the calibration requirements As stated above please note that controls must also be run Verify and approve the calibration according to the calibration status displayed in the calibration windows and discard the calibrator from the sample tray after use Calibration The calibrators of this kit are traceable to the manufacturers working calibrators which are prepared from highly purified PINP whose value is assigned using amino acid composition analysis Calibration Frequency A new calibration is required e Each time a new lot of cartridges is loaded on board e Each time a new lot of trigger or cuvettes is used e When the control values do not fall within the defined ranges e When the calibration has expired e After Analyser service Verification of the calibration is automatic and managed by the Analyser Quality Control Use the IDS iSYS Intact PINP Control Set for quality control To ensure validity of results at least three controls with vary
12. obins SP Bilezikian JP editors Dynamics of bone and cartilage metabolism Principles and clinical applications second edition London Academic Press p391 405 2 Melkko J Kauppila S Niemi S Risteli L Haukipuro K Jukkola A Risteli J 1996 Immunoassay for intact amino terminal propeptide of human type procollagen Clin Chem 42 947 954 3 Chen P Satterwhite JH Licata AA Lewiecki EM Sipos AA Misurski DM Wagman RB 2005 Early changes in biochemical markers of bone formation predict BMD _ response to teriparatide in postmenopausal women with osteoporosis J Bone Miner Res 20 962 970 4 Finkelstein JS Leder BZ Burnett SA Wyland JJ Lee H De la Paz AV Gibson K Neer RM 2006 Effects of teriparatide alendronate or both on bone turnover in osteoporotic men J Clin Endocrinol Metab 91 2882 2887 5 Nenonen A Cheng S lIvaska KK Alatalo SL Lehtimaki T Schmidt Gayk H Uusi Rasi K Heinonen A Kannus P Sievanen H Vuori Vaananen HK Halleen JM 2005 Serum TRACP 5b is a useful marker for monitoring alendronate treatment comparison with other markers of bone turnover J Bone Miner Res 20 1804 1812 al Immunodiagnostic Systems Ltd IDS Ltd 10 Didcot Way Boldon Business Park Boldon Tyne amp Wear NE35 9PD England Tel 44 191 519 0660 Fax 44 191 519 0760 e mail info uk idsplc com www idsplc com Immunodiagnostic Systems UK Immunodiagnostic Systems Ltd IDS Ltd 10 Didcot Way Boldon Business Park Bol
13. ty The limit of blank LoB limit of detection LoD and limit of quantitation LoQ were determined with guidance from CLSI EP17 A Protocols for Determination of Limits of IS 4000PLV02 2009 12 22 English Page 3 4 1 feof 143 os 26 os 42 2 80 3 0 2 4 5 3 Detection and Limits of Quantitation using 60 blanks and 120 low level samples LoB 0 5 ng mL LoD lt 1 0 ng mL LoQ lt 1 0 ng mL Precision Precision was evaluated in accordance with a modified protocol based on CLS EP 5A2 Evaluation of Precision Performance of Quantitative Measurement Methods Three serum controls were assayed using three lots of reagents in quadruplicate once per day for 20 days on two analysers Within run Total ng mL SD CV SD CV 44 6 1 3 eo 1000 Recovery Recovery was assessed by adding intact PINP to samples rior to assay Sample PINP Measured Recovery Recovery Conc added ng mL ng mL ng mL ng mL 44 3 35 3 77 1 32 8 93 44 3 74 8 115 1 70 8 95 33 5 35 3 66 4 32 9 93 33 5 74 8 101 5 68 0 91 37 3 35 3 70 8 33 5 95 37 3 74 8 105 2 67 9 91 Mean 93 Linearity Linearity was evaluated based on CLSI EP 6A Evaluation of the Linearity of Quantitative Measurement Procedures A Statistical Approach Samples containing varying concentrations of intact PINP were assayed in duplicate The resulting mean concentrations were compared to predicted concentrat
Download Pdf Manuals
Related Search
Related Contents
取扱説明書 SD100R L Band Demodulator - downloads Untitled - Audioline Dicota D30657 Razor Wheelchair 13014961 User's Manual Massive Suspension light 36065/17/10 Huawei Ascend P6 8GB Black Bulletin d`inscription Séminaire GR 2002 EVGA 03G-P4-2781-KR NVIDIA GeForce GTX 780 3GB graphics card Copyright © All rights reserved.