CircuFlow™ 5208 Quick Setup Guide
Contents
1. DEVON MEDICAL PRODUCTS CircuFlow 5208 Quick Setup Guide www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 LB35 0001 Rev B 20140314 DEVON MEDICAL PRODUCTS Setup Cd TABLE OF CONTENTS Therapy Setup M Pressure Mode 1 Ob Gradient Mode n 1 111 09 Treatment 1 3 www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS SETUP 01 Open the shipping box and lift device up and out of box 02 Remove the protective foam covers on each side and the plastic bag 03 Check to make sure the box also con tains 1 power cord 1 air blocker and a user manual 4 Place the device on a flat and stable surface near the patient 0p Unwrap the power cord from the box and plug it into the back of the machine Plug the device into a safe and properly grounded outlet www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS TH ERA PY S ETU P Settings can only be modified or restored before or in between treatments There are three operating modes with this device Factory Default Mode Device comes with a factory default setting providing 50mmHg pressure on the distal chamber with a 496 of gradient pressure 60 minute treatment time 60 second cycle time and no chambers being skip
2. illuminate 00 Upon this display User can push the green Start button to start treatment Once the button is pressed the menu screen will ask for Unilat eral or Bilateral treatment options If only using one sleeve push down on the Shuttle Knob when unilateral is selected to confirm selec tion If using two sleeves turn the Shuttle Knob to the Bilateral option and push down on the Shuttle Knob to confirm selection Treat ment will start once a selection is made e f Pressure Mode is desired continue to steps 10 33 e f Gradient Mode is desired continue to steps 34 59 m www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS PRESSURE MODE 10 Upon this display User can access Pressure Mode by pushing the SHUTTLE KNOB that will take you to Therapy Mode Settings 1 1 This display shows Therapy Mode Settings 17 Turn the SHUTTLE KNOB to the right to select Pressure Mode Then push the SHUTTLE KNOB to access Pressure Mode it will allow you to set for the various parameters Pressure Cycle Time Treatment Time Skip Chamber Uncorking and Focus Mode Pressure Set Push the SHUTTLE KNOB to select Pressure Set than push the SHUT TLE KNOB on the individual pressures the pressure can be changed in all 8 chambers 13 This display shows the first parameter 14 Increase the Pres
3. 19406 866 446 0092 DEVON MEDICAL PRODUCTS PRESSURE MODE Continued 22 Push the SHUTTLE KNOB to confirm the new Treatment Time 23 Access the fourth parameter Skip Cham ber by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select Skip Chamber 24 Turn to the chamber you wish to skip by turning the SHUTTLE KNOB to the right or left Select one only No skip Chamber 1 Chamber 2 Chamber 3 Chamber 4 etc Skipping a chamber does not mean the chamber selected to skip will not fill with air 25 Push the SHUTTLE KNOB to confirm the Skipped Chamber 20 Access the fifth parameter Uncorking by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select Un corking 27 Select whether Uncorking should be on or off before the treatment begins by turning the SHUTTLE KNOB and pressing the knob down to confirm it 28 If Uncorking is on select what pressure to use for Uncorking selecting between 15mmHg or 20mmHg Push the SHUTTLE KNOB to confirm the pressure and scroll to the right to exit Uncorking settings www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS PRESSURE MODE Continued 20 Access the sixth parameter Focus Mode by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select Focus Mode 3l Choose whether Foc
4. Focus Mode by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select Focus Mode www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 Hid DEVON MEDICAL PRODUCTS GRADIENT MODE tmine hh Choose whether Focus Mode should be turned on or off If it should be on se lect which chambers to focus on and what pressure to use choosing between 40mmHg or 50mmHg b Press the SHUTTLE KNOB to confirm all selections T To exit and complete Gradient Mode setup turn the SHUTTLE KNOB to the right Push to select to Exit hg Turn the SHUTTLE KNOB to the right in Therapy Setting Mode until you see EXIT push SHUTTLE KNOB to select and you will be taken to the lt Start gt display Push the green Start button to start your cus tom treatment Please select Unilateral or Bilateral for one sleeve or two sleeve treatments by pressing down on the SHUTTLE KNOB to confirm proceed to step 60 www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 60 61 62 63 64 Do www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 DEVON MEDICAL PRODUCTS TREATMENT The pump will go through its initial prefill stage of reaching 10 mmHg in all 8 cham bers Each treatment will end after its set treat ment time has ela
5. OB to the right Push the shuttle knob to select Gradient 38 Increase the Gradient by turning the SHUTTLE KNOB to the right decrease by turning to the left gradient can range from 1 7 increments of 1 39 Push the SHUTTLE KNOB to confirm the new Gradient Af This display shows the second parameter Pressure turn the shuttle knob to the left Push the SHUTTLE KNOB to select Pressure User can access other parameters by turning the shuttle knob to the right left at any time www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 09 4 42 43 Ad 4b 4b 4 49 DEVON MEDICAL PRODUCTS GRADIENT MODE Cmi Increase the Pressure by turning the SHUTTLE KNOB to the right decrease the pressure by turning to the left pres sure can range from 40 100 mmHg increments of 1 mmHg Push the SHUTTLE KNOB to confirm the new Pressure Access the third parameter Cycle Time by turning SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select Cycle Time Increase the Cycle Time by turning the SHUTTLE KNOB to the right decrease time by turning to the left cycle time can range from 20 240 seconds in 1 second increments Push the SHUTTLE KNOB to confirm the new Cycle Time Access the fourth parameter Treatment Time by turning the SHUTTLE KNOB to the right Push
6. ped Follow steps 6 8 before pressing the green START button Pressure Mode This mode allows you to set the pressure in each of the eight chambers in the sleeve the duration of each cycle how long the treatment will last whether or not any chamber will be skipped if Uncorking is required and whether Focus Mode is necessary Gradient Mode This mode allows you to set the pressure gradient between chambers the pressure for the first chamber the duration of each cycle the duration of the treatment whether any chambers will be skipped if Uncorking is required and whether Focus Mode is necessary Attach top and or bottom snap connectors of 6 the garment to the snap connect ports which are located on the front panel of the device During a single garment session i e top only or bottom only insert the air blocker into the unused snap connect port 07 Putting the garment on a for leg garments unzip the garment all the way to the end Place the foot at the bottom of the garment and pull up the zipper while sup porting the garment to wrap around the leg b for arm garments slide the arm through the internal cavity of the garment c for arm and shoulder garments slide the arm through the internal cavity of the garment and snap together the belt connector around your torso 08 Press Main POWER SWITCH up to ON position which is located on the rear panel The green power indicator on the front panel will then
7. psed The user can also end the treatment at any time during a treatment session by pressing the STOP button on the front panel Note If the user presses STOP the treatment will not pick up from where it left off at After the treatment the device will vacuum air from the garment for 1 minute so that it will facilitate removal of the garment Not Pictured Note If using one of the larger size sleeves all the air may not vacuum out within 1 minute The vacuum setting can go through another 1 minute cycle by pressing STOP Once the sleeve is deflated press the pow er switch on the backboard to the OFF position and then unplug the power cord Once the power indicator light is off it is safe to remove the garment Not Pictured The garment should be loose enough now that you can unzip the garment and remove 066 446 0092 DEVON MEDICAL PRODUCTS
8. sure by turning the SHUTTLE KNOB to the right decrease the pressure by turning to the left pressure can range from 20 100 mmHg increments of 1 mmHg 15 Push the SHUTTLE KNOB to confirm the new Pressure This will take you to the next chamber follow step 14 to set the pressure in each chamber repeat until you have reached chamber 8 www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS PRESSURE MODE Continued 16 After chamber 8 the screen will read exit and push the SHUTTLE KNOB to con firm 17 Access the second parameter Cycle Time by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select Cycle Time 1 Q Increase the Cycle Time by turning the SHUTTLE KNOB to the right decrease time by turning to the left cycle time can range from 20 240 seconds increments of 1 second 19 Push the SHUTTLE KNOB to confirm the new Cycle Time 20 Access the third parameter Treatment Time by turning SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select Treatment Time 71 Increase the Treatment Time by turning the SHUTTLE KNOB to the right de crease time by turning to the left treat ment time can range from 15 180 min utes increments of 15 minutes 06 www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA
9. the shuttle knob to select Treatment Time Increase the Treatment Time by turning the SHUTTLE KNOB to the right de crease time by turning to the left treat ment time can range from 15 180 min utes increments of 15 minutes Push the SHUTTLE KNOB to confirm the new Treatment Time www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS GRADIENT MODE 40 Access the fifth parameter Skip Chamber by turning the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select Skip Chamber P Turn to the chamber you wish to skip by turning the SHUTTLE KNOB to the right or left Select one only No Skip Chamber 1 Chamber 2 Chamber 3 Chamber 4 etc Skipping a chamber does not mean the chamber selected will not fill with air h1 Push the SHUTTLE KNOB to confirm the Skipped Chamber 52 Access the sixth parameter Uncorking by turn ing the SHUTTLE KNOB to the right Push the SHUTTLE KNOB to select Uncorking 53 Select whether Uncorking should be on or off before the treatment begins by turning the SHUTTLE KNOB and pressing the knob down to confirm it HA If Uncorking is on select what pressure to use for Uncorking selecting between 15mmHg or 20mmHg Push the SHUTTLE KNOB to confirm the pressure and scroll to the right to exit Uncorking settings Hh Access the seventh parameter
10. us Mode should be turned on or off If it should be on se lect which chambers to focus on and what pressure to use choosing between 40mmHg or 50mmHg 31 Press the SHUTTLE KNOB to confirm all selections 32 To exit and complete Pressure Mode setup turn the SHUTTLE KNOB to the right Push to select to Exit 33 Turn the SHUTTLE KNOB to the right in Therapy Setting Mode until you see EXIT push SHUTTLE KNOB to select and you will be taken to the Start display Push the green Start button to start your cus tom treatment Please select Unilateral or Bilateral for one sleeve or two sleeve treatments by pressing down on the SHUTTLE KNOB to confirm Continue to step 60 08 www devonmedicalproducts com 1100 First Avenue Suite 202 King of Prussia PA 19406 866 446 0092 DEVON MEDICAL PRODUCTS GRADIENT MODE 34 Upon this display User can access Gradi ent Mode by pushing the SHUTTLE KNOB circular button that will take you to Thera py Mode Settings the indicator light will be yellow setting 3h This display shows Therapy Mode Settings 3b Turn the SHUTTLE KNOB to the right to select Gradient Mode Then push the SHUTTLE KNOB to access Gradient Mode it will allow you to set for the various parameters Gradient Pressure Cycle Time Treatment Time Skip Chamber Uncorking and Focus Mode 3 Access the first parameter Gradient by turning the SHUTTLE KN
Download Pdf Manuals
Related Search
Related Contents
Descargar Manual del Usuario Philips HR1324 GW QSG#25213 21in 13A Mower front side User`s Manual Front and Rear Parking Sensor System Tripp Lite PowerVerter Plus DC-to-AC Inverter User's Manual 生物由来製品、処方せん医薬品注) 血液凝固阻止剤 - DI 2014-10-20 Bedienungsanleitung HST-Tool DYNAMIC CHARACTERIZATION OF RIGID POLYURETHANE DELL 3Y PS 4H MC, Ext, PowerEdge R320 Copyright © All rights reserved.