Home
        Multifunction Process Calibrator
         Contents
1.       Pressure  Module                                        V mA FREQ  L 2  SOURCE                                gt  To Unit Under Test          TC RTD PULSE                EXIT CONFIG          CONFIG  SELECTION  kend                RECALL  dh 2  ENTER  maas I     AME  JJ                                                 bec47f eps  Figure 18  Connections for Sourcing Pressure    40    Setting 0   and 100   Output  Parameters    For current output  the Calibrator assumes that 0    corresponds to 4 mA and 100   corresponds to 20 mA   For other output parameters  0   and 100   points must  be set before using the step and ramp functions  Proceed  as follows     1  If necessary  press for SOURCE mode     2  Select the desired source function and use the arrow  keys to enter the value  This example is temperature  source using 100  C and 300  C values for source     3  Enter 100   C then press and hold Lee  to store the       value   4  Enter 300   C then press and hold  19   to store the  value     This setting can now be used for the following     e Manually stepping an output in 25    increments     e Switch between the 0 and 100   span points by  momentarily pushing or  10          Multifunction Process Calibrator  Setting 0   and 100   Output Parameters      Error Functionality    Percentage Error is available for every range on the  lower display  The calculation is based on a mA  percentage deviation from the value measured on the  upper display to the value sourced on th
2.      di B B B B Pene   E Ce  WEE  Adjust    js OPEN CLOSE     SWITCH  TEST                                        Q FREQ    RTD    PULSE    EXIT CONFIG                                     CONFIG  SELECTION          RECALL   ENTER       RIGGER STOI  ANT                                                    bec06f eps    Figure 5  Adjusting the Contrast    Multifunction Process Calibrator  Configuration Menus       Shut Down Mode Frequency Pulse Output Voltage   The Calibrator comes with a shut down mode set for 30 1  Press L  n until FREQ OUTPUT V Adjust appears  minutes  displayed for about 1 second when the on the display    Calibrator is initially turned on   When shut down mode is 2  Use         EI and Q to adjust the frequency pulse    enabled  the Calibrator automatically shuts down after      output voltage from 1 to 20 V   the elapsed time from when the last key was pressed              3  Press  er  to save the setting   1  Press  sitio  until SHUT DOWN appears on the EE  9  display  Pulse Output Frequency  2  Use  amp  and S to increase or decrease the time  1  Press  s  rt  until PULSE OUTPUT Hz FREQ Adjust  3  Press to save the setting  appears on the display     2  Use        and    to adjust the pulse output  CJC frequency from 2 CPM to 15 kHz    Cold Junction Compensation  CJC  is a value for the cold 3  Press to save the setting    end of a thermocouple at the Meters end     1  Press    t  until SELECT CJC appears on the  display     2  Use   and Q to select ON or
3.     Black       Red                            Se               FLUKE 726 PRECISION CALIBRATOR       MEASURE    0000    10000               ZERO  Error     s5econds     OPEN CLOSE    V mA SWITCH  Loop      TEST                               V mA 2 FREG                               TC RTD PULSE          EXIT CONFIG  100           CONFIG  SELECTION    eme  LE  RECALL  da 25   ENTER    TRIGGER STO                                                                             Figure 23  Calibrating a Chart Recorder    52    bec25f eps    Remote Control Commands    The Calibrator may be remotely controlled from a PC  running a terminal emulator program  The remote control  commands give access to all capabilities of the Calibrator  with the exception of pressure measurement     See the Fluke Web Site for the 726 Remote  Programming application note at  www fluke com processtools    HART  Functionality    The Calibrator has a user selectable 250 Q HART to  facilitate use with HART communication devices  The  resistor can be switched in or out using config selection  menus  Use a HART communicator when measuring mA  with loop power or sourcing mA        Maintenance  Replacing the Batteries    AA Warning    To avoid false readings  which could lead to  possible electric shock or personal injury   replace the batteries as soon as the battery  indicator  EM  appears     Figure 24 shows you how to replace the batteries        Multifunction Process Calibrator  Maintenance    Battery and  C
4.     Exits Configuration Menu     Increments output by 25 96 of span   Decrements output by 25   of span     Recalls from memory a source value corresponding to 0   of span and sets it as the source value   D  Press and hold to store the source value as the 0   value   Press and hold when powering up to identify the firmware version  The firmware version is shown in the       upper display for about 1 second after initialization        11    726  Users Manual    Table 4  Key Functions  cont      we  mme   ven O    Cycles through     N Slow repeating 0     100     0   ramp    TRIGGER STOP M Fast repeating 0     100     0   ramp    d Repeating 0     100     0   ramp in 25   steps  Used for the pulse train and totalizer functions   SO Increases or decreases the source level   d H Cycles through the 2   3   and 4 wire selections   Hem Moves through the memory locations of Calibrator setups   Moves through the configuration menus     Saves and recalls setups  amp  data   ENTER is used in the configuration menus     Recall    Used to enter and navigate the configuration menus     Cycles the Calibrator through MEASURE and SOURCE modes in the lower display     Selects TC  thermocouple  measurement and sourcing function in the lower display  Repeated  pushes cycle through the thermocouple types     Toggles between voltage  mA sourcing  or mA simulate functions in the lower display   Inserts a 250 Q resistor when in mA     Selects RTD  resistance temperature detector  measurement and so
5.     Figure 1  Standard Equipment    Multifunction Process Calibrator  Safety Information    Symbols  Symbols used on the Calibrator and in this manual are explained in Table 2     Table 2  International Symbols    AC   Alternating current Double insulated  ne  DC   Direct current Battery    L Risk of danger  Important information  See  Fann ground A Manual  Precedes Warning     Conforms to European Union directives Hazardous Voltage  Precedes Warning     Conforms to Canadian Standards Association directives        726  Users Manual    Getting Acquainted with the Calibrator       Input and Output Terminals    Figure 2 shows the Calibrator input and output terminals   Table 3 explains their use  manna    9008 H T    SOURCEMEASURE AUTOA Mr                                                                                                                                                 bec05f eps    Figure 2  Input Output Terminals and Connectors    Multifunction Process Calibrator  Getting Acquainted with the Calibrator    Table 3  Input Output Terminals and Connectors  SSES    1  Pressure module Connects the Calibrator to a pressure module or to a PC for a remote control  connector serial connector serial connection   2   O MEASURE V  mA terminals   Input terminals for measuring voltage  current  supplying loop power  HART  resistance  switch test options     Thermocouple  TC  Terminal for measuring or simulating thermocouples  This terminal accepts a    input output miniature pola
6.     Figure 20  Calibrating a Thermocouple Transmitter    46    Calibrating a Pressure Transmitter    The following steps explain calibrating a pressure  transmitter     1     Connect the Calibrator to the instrument under test  as shown in Figure 21    Press for current  upper display   If required   press again to activate loop power    Press S    lower display         Multifunction Process Calibrator    Calibrating a Pressure Transmitter    If necessary  press for SOURCE mode   Zero the pressure module     Perform checks at 0   and 100   of span and  adjust the transmitter as necessary     Note    When HART resistor mode is selected  the  250 Q resistor is turned on both mA channels     47    726  Users Manual                                  Hand Pump                                                                                    Measure mA   N Source Pressure  7   24 V Loop Power  Enabled                                  Pressure  Module    e  Error 35econde OPEN CLOSE                                         V mA  SWITCH  Loop  LN TEST  HART                                                                                                                      PI V mA Q FREQ  s TC RTD PULSE  ex coura   cn 100    RECALL  dh da 25    S ee  a e  Saab    red     ra   HART           Communicator                  Oy    e e e Black                      bec34f eps    Figure 21  Calibrating a Pressure to Current  P I  Transmitter    48    Calibrating an I P Device    The following 
7.     vi    List of Figures    Figure Title  1  Standard Equipment  ion dae  2  Input Output Terminals and Conpnechors sese eee eee eee eee eee  3  NOE EEE EE A oe ee eed  4  Elements of a Typical Display A  5  Adjusting the  Contrast username ed A o lab  6  Voltage to Voltage Test     7  Measuring Voltage and Current Output     8  Connections for Supplying Loop Power   9  Measuring Electrical Parameterg  AAA  10  Measuring Temperature with a Thermocouple A    11  Measuring Temperature with an RTD  Measuring 2   3   and 4 Wire Resistance  12  Gage and Differential Pressure Modules       13  Connections for Measuring Pressure    eee eee ee eee eee eee e  14  Connections for Simulating a 4  to 20 MA Transmitter     15  Electrical Sourcing Connections esse serere eee eee eee eee  16  Connections for Simulating a Thermocouple see eee eee ee eee ee ee eee eee ee eee  17  Connection for Simulating a 3  and 4 Wire DTD       vii    726  Users Manual    18   19   20   21   22   23   24   25     Connections for Sourcing Pressure see eee eee eee eee ee    SAVE DATA Menu    Showing Measurement Memory Location 3  1    Calibrating a Thermocouple Transmitter   Calibrating a Pressure to Current  P I  Transmitter     Calibrating a Current to Pressure  I P  Transmitter       Calibrating a Chart    Recorrido    Replacing the Batteries    eee eee    Replacement Parts       viii    Multifunction Process Calibrator    Introduction    The Fluke 726 Multifunction Process Calibrator  referred  to as   
8.    bec39f eps    Figure 6  Voltage to Voltage Test    17    726  Users Manual    Using Measure Mode  Measuring Electrical Parameters  Upper Display     To measure the current or voltage output of a transmitter   or to measure the output of a 700 Series pressure  module  use the upper display and proceed as follows     1  Press to select volts or current  LOOP should  not be on     2  Connect the leads as shown in Figure 7   Current Measurement with Loop Power    The loop power function activates a 24 V supply in series  with the current measuring circuit  allowing you to test a  transmitter when it is disconnected from plant wiring  To  measure current with loop power     1  Connect the Calibrator to the transmitter current loop  terminals as shown in Figure 8     2  Press while the Calibrator is in current  measurement mode  LOOP appears and an internal  24 V loop supply turns on     Note    When HART resistor mode is selected  the  250    resistor is turned on both mA channels     18             FLUKE 726 PRECISION CALIBRATOR    ZERO     Error Seconds     OPEN CLOSE                      LOOP    V mA SWITCH   2 TEST          HART        vma  To    FREQ                TC RTD          PULSE             TONFIG  SELECTION    SAVE E   RECALL  ENTER  p Ce     TRIGGER STOI          EXIT CONFIG       100           de 25   y 25                 AMT Retum to    Recall                Ka    Ee                            bec42f eps    Figure 7  Measuring Voltage and Current Output    Mult
9.  30V MAX ALL TERMINAL S   A  EASURE  MV A Process Temperature  H  H             Warning AN          30 V maximum to L  TC Miniplug              bec12f eps    Figure 10  Measuring Temperature with a Thermocouple    23    726  Users Manual    Using Resistance Temperature Detectors  RTDs     The Calibrator accepts RTD types shown in Table 6   RTDs are characterized by their resistance at 0  C    32   F   which is called the    ice point    or R   The most  common R  is 100 Q  The Calibrator accepts RTD  measurement inputs in two   three   or four wire  connections  with the three wire connection the most  common     four wire configuration provides the highest  measurement precision  and two wire provides the lowest  measurement precision     24    To measure temperature using an RTD input   1  If necessary  press for MEASURE mode     2  Press Lr   for the RTD display  Continue pressing  this key to select the desired RTD type     3  Press    or to select a 2   3   or 4  wire  connection                 4  Attach the RTD to input terminals as shown in  Figure 11     PRT Custom Curves    Up to three custom curves can be named and CVD  coefficients can be entered through the serial port  Names  can be up to six characters  For more information  see  the Application Note on the 725 726 CD     Multifunction Process Calibrator  Using Measure Mode    Table 6  RTD Types Accepted      RTDType   leePoint R    ween   oa   Ranee     1200 565     Pt500  385  500   0 00385  UC  200 to 630  P
10.  OFF   3  Press  ve  to save the setting   Celcius and Fahrenheit    C and   F     1  Press    t  until SELECT UNIT   C  or   F  appears  on the display     Use    and   to select   C or   F     Press  Aij to save the setting                          15    726    Users Manual    HART  Resistor ON OFF    1     16       Press  siten  until SELECT HART ON or OFF  appears on the display     Use to toggle ON or OFF   Press to save the setting   Note    When HART mode is selected  the 250 Q  resistor is turned on both mA channels              Getting Started    This section details some basic operations of the  Calibrator     Voltage to Voltage Test  To perform a voltage to voltage test     1     Connect the Calibrator s voltage output to its voltage  input as shown in Figure 6    Press  O  to turn on the Calibrator  Press to  select dc voltage  upper display     If necessary  press for SOURCE mode  lower  display   The Calibrator is still measuring dc voltage   the active measurements are visible in the upper  display    Press  Vma  to select de voltage sourcing    Press hl and    to select a digit to change  Press C  to select 1 V for the output value  Press and hold   to enter 1 V as the 0   value    Press to increase the output to 5 V  Press and  hold  100   to enter 5 V as the 100   value     Press  425   and to step between 0 and 100    in 25   step increments              Multifunction Process Calibrator  Getting Started       FLUKE  726 Precision  CALIBRATOR  MEASURE       
11.  avoid mechanically damaging the  pressure module     Never apply more than 10 ft  Ib   13 5 Nm   of torque between the pressure module  fittings  or between the fittings and the  body of the module  Always apply  appropriate torque between the pressure  module fitting and connecting fittings or  adapters     27    726  Users Manual    e Never apply pressure above the rated  maximum printed on the pressure  module     e Only use the pressure module with  specified materials  Refer to the printing  on the pressure module or the pressure  module instruction sheet for the  acceptable material compatibility     1  Connect a pressure module to the Calibrator as  shown in Figure 13  The threads on the pressure  modules accept standard 4 NPT pipe fittings  Use  the supplied 4 NPT to 4 ISO adapter if necessary     2  Press either  2   or 2    The Calibrator  automatically senses which pressure module is  attached and sets its range accordingly                 3  Zero the pressure module as described in the  module s Instruction Sheet  Modules vary in zeroing  procedures depending on module type  but all require  pressing for 3 seconds     Continue pressing to change pressure display  units to psi  mmHg  inHg  cmH 0 04   C  cmH 0 220    C  inH 0  4   C  inH 0220   C  inH  0 260   F  mbar   bar  kg cm     or kPa     28    Zeroing with Absolute Pressure Modules    To zero  adjust the Calibrator to read a known pressure   This can be barometric pressure  if it is accurately known   for a
12.  costs of repair replacement parts when product purchased in one country is submitted  for repair in another country     Fluke s warranty obligation is limited  at Fluke s option  to refund of the purchase price  free of charge repair  or replacement of a defective product which is  returned to a Fluke authorized service center within the warranty period    To obtain warranty service  contact your nearest Fluke authorized service center to obtain return authorization information  then send the product to that service  center  with a description of the difficulty  postage and insurance prepaid  FOB Destination   Fluke assumes no risk for damage in transit  Following warranty  repair  the product will be returned to Buyer  transportation prepaid  FOB Destination   If Fluke determines that failure was caused by neglect  misuse   contamination  alteration  accident  or abnormal condition of operation or handling  including overvoltage failures caused by use outside the product s specified  rating  or normal wear and tear of mechanical components  Fluke will provide an estimate of repair costs and obtain authorization before commencing the work   Following repair  the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges  FOB  Shipping Point      THIS WARRANTY IS BUYER S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES  EXPRESS OR IMPLIED  INCLUDING BUT  NOT LIMITED TO ANY IMPLIED 
13.  the Calibrator     is a handheld  battery operated  instrument that measures and sources electrical and  physical parameters  See Table 1     In addition to the functions in Table 1  the Calibrator also   has the following features and functions    e A split screen display  The upper display allows  users to measure volts  current  and pressure only   The lower display allows the user to measure and  source volts  current  pressure  resistance  temperature detectors  thermocouples  frequency   and ohms    s A thermocouple  TC  input output terminal and  internal isothermal block with automatic reference   junction temperature compensation    e Stores and recalls setups    e Manual and automatic stepping and ramping     e Stores and recalls calibration screens   e Control the Calibrator remotely from a PC running a  terminal emulator program     Contacting Fluke    To order accessories  receive operating assistance  or  locate the nearest Fluke distributor or Service Center   call     USA  1 888 44 FLUKE  1 888 443 5853   Canada  1 800 36 FLUKE  1 800 363 5853   Europe   31 402 675 200   Japan   81 3 3434 0181   Singapore   65 738 5655   Anywhere in the world   1 425 446 5500    USA Service  1 888 99 FLUKE  1 888 993 5853     Or  visit Fluke s Web site at www fluke com   To register your product  visit register fluke com    726  Users Manual    Table 1  Summary of Source and Measure Functions      Function   Measure    Source    Types E  J  K  T  B  R  S  L  U  N  C  XK  BP  
14.  to 20 mA transmitter  simulating  30        A     Accessories   56     Auto ramping output   42        B      Battery  replacing  53      C      Calibration   54     Celcius and Fahrenheit   Cleaning calibrator   54   Cold Junction Compensation  CJO    15   Commands  Remote Control   configuration menus   Connections  for pressure sourcing   39   Contact Information   1     Index       D       Data    Recall   44   display   contrast adjustment   Display   13      E       Electrical parameters  measurement   20     sourcing   32        F     frequency output voltage   15     67    726    Users Manual       G      Getting started   16      H      HART    configuration menu  16              UP device  calibrating   49    Input terminals    Input output terminals and connectors   table    9        K       Key functions  table   11   Keys   10        L     loop power   18     Loop power    simulating   30     68       M       Measure functions  summary  table    2   Measure mode   18   Measuring  pressure   27   temperature with RTDs  24   temperature with thermocouples   21   menus    configu ration   14      0     Output device  testing   51   Output terminals   8       p      Parts list    Percentage Error  41    Pressure  Measurement   27   Sourcing   38     pressure modules  compatibility   56    Pressure Modules  Compatibility   56    Pressure modules available   56     Pressure modules  zeroing   28   Pressure Switch Test   51    Pressure transmitter  calibrating   47   P
15.  to recall for later use     low battery condition or  a battery change does not jeopardize the stored settings     Store a Setup  To store a setup   1  Create the desired setup     2  Push  84   The right side of the display changes to  show SAVE SETUP and SAVE DATA     3  Push Rito select SAVE SETUP     Push    or    to select the desired memory location  at  the top of the LCD      5  Push to enter the setup     Recall a Setup  To recall a setup     1     Push  E  twice  The right side of the display    changes to show RECL SETUP and RECALL DATA     Push  88  again to RECL SETUP     Push    to select the desired memory location  at the  top of the LCD      Push  YE  to recall the setup from the proper  memory location           Multifunction Process Calibrator  Storing and Recalling Data    Storing and Recalling Data    Up to 40 data samples can be stored in a nonvolatile  memory to be recalled for later use  A low battery  condition or a battery change does not jeopardize the  stored settings     Storing Data    To store measurement data  use the following procedure   refer to Figure 19     1  Take the desired measurement     2  Push  8    The right side of the display changes to  show SAVE SETUP and SAVE DATA     3  Press to select SAVE DATA     Push  85  again  The open data point  bottom right  of the display  flashes     Use    and  9 to change the data point location  1 8         Push  85  to store the measurement and return the  unit to the measurement mode  Figure 
16. 19 shows a  reading stored in memory location 3  data point 1     43    726  Users Manual       Memory Locations  Save Data Mode Indicator       MEASURE 3   Data Point  SAVE  DATA    SOURCE      av em a                          bec46f eps  Figure 19  SAVE DATA Menu Showing Measurement  Memory Location 3  1    Recall Data    To recall data   1  Push twice  The right side of the display  changes to show RECL SETUP and RECALL DATA   2  Push to highlight RECL DATA  bottom right of the  display    Push FSE   Push h to choose the desired memory location  top  of the display      The data saved in that first memory location now  appears  There can be different measurements  stored  1 5  for each memory location  1 8      44    5  Push n  or    to select the correct data location   bottom right of display      6  Push to recall the data stored in that location     7  Push  to return to the same RECALL DATA  location to see the next saved measurement  2 of 5  for example     Pulse Train Source Read    Pulse Train Source Read counts input pulses or sources  output pulses  Use the configuration menus to set the  frequency and output voltage  Refer to    Configuration  Menus    earlier in this manual  The number of counts is  set through the main display and cannot be changed  while sourcing pulses  works as a trigger stop key  in this mode since ramping or stepping during a pulse  train is not relevant        Calibrating a Transmitter    Use the measurement  upper display  and source  l
17. FLUKE     726    Multifunction Process Calibrator    Users Manual    September 2005  O 2005 Fluke Corporation  All rights reserved   All product names are trademarks of their respective companies     LIMITED WARRANTY AND LIMITATION OF LIABILITY    Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service  The warranty period is three years and  begins on the date of shipment  Parts  product repairs  and services are warranted for 90 days  This warranty extends only to the original buyer or end user  customer of a Fluke authorized reseller  and does not apply to fuses  disposable batteries  or to any product which  in Fluke s opinion  has been misused  altered   neglected  contaminated  or damaged by accident or abnormal conditions of operation or handling  Fluke warrants that software will operate substantially in  accordance with its functional specifications for 90 days and that it has been properly recorded on non defective media  Fluke does not warrant that software will  be error free or operate without interruption     Fluke authorized resellers shall extend this warranty on new and unused products to end user customers only but have no authority to extend a greater or different  warranty on behalf of Fluke  Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable  international price  Fluke reserves the right to invoice Buyer for importation
18. RTD    Pt100 Q  385    Resistance  Pt100 Q  3926   Temperature Pt100 Q  3916     Detector  Pt200 Q  385   Pt500 Q  385     Pt1000 Q  385   Ni120  672   CU10    Pressure 29 modules ranging from 1 0 in  H O to 10 000 psi  Pulse 1 100 000 1 10 000  Freqency Max 10 kHz Frequency Range 2 CPM to 10 kHz  Other functions Loop supply  HART resistor  pressure switch test  save screen  step  ramp  memory  cold  junction compensation       Standard Equipment    If the Calibrator is damaged or something is missing   contact the place of purchase immediately  To order  replacement parts  see Table 8  The items listed below  and shown in Figure 1 are included with the Calibrator     TL75 test leads   AC72 alligator clips   Stackable alligator clip test leads   726 Product Overview  not shown in Figure 1     725 726 CD ROM  contains Users Manual  not  shown in Figure 1     4 AA Batteries  installed     Multifunction Process Calibrator  Standard Equipment    Safety Information    The Calibrator is designed in accordance with CAN CSA   C22 2 NO  61010 1 04  UL 61010 1  and ISA 82 02 01    AA Warning    To avoid possible electric shock or personal  injury  use the Calibrator only as specified in  this manual  otherwise the protection  provided by the Calibrator may be impaired     A Warning identifies conditions and actions that pose  hazard s  to the user  A Caution identifies conditions and  actions that may damage the Calibrator or the equipment  under test     726    Users Manual    AA Warn
19. V  lower display  0 000 20 000 0 010     2 mV    20 V  Source  0 000 20 000 0 010    2 mV  100 mV  Source  0 000 100 000 0 010     10 UV  90 mV  Read  0 000 90 000 0 010     10 UV    Maximum current output in voltage ranges is 1 mA with an output impedance of  lt  1 Q       DC mA Measurement and Source    Ranae Accuracy   g    of Reading   Floor   mA Read  Upper Display  0 000 24 000 0 010     2 pA  mA Read  Lower Display  0 000 24 000 0 010     2 uA    Maximum load on  mA source is 1 kN  With the HART resistor on  maximum load is 750 Q   Voltage input range on simulate mode is 5 to 30 V       59    726  Users Manual    Ohms Measurement    Ohms Read  low  Inn  400 00    Accuracy     of Reading   Floor     0 015     0 05 Q       Ohms Read  high  401 0 4000 0    Ohms Source    0 015   0 5 Q    Ohms Range Excitation Current from Accuracy  9 Measurement Device    of Reading   Floor     Ohms Source  low  400 0 0 1 to 0 5 mA    0 015     0 1 Q       0 015     0 05 Q       Ohms Source  high  1500 0 05 to 0 8 mA    0 015     0 5 Q       1500 4000 0 05 to 0 4 mA    Unit is compatible with smart transmitters and PLCs   Frequency response is  lt  5 mS       Frequency Measurement       0 015     0 5 Q                      Range Minimum Maximum    of See Floor   CPM Read 2 0 1000 0 0 05     0 1 CPM  Hz Read 1 0 1000 0 0 05     0 1 Hz  KHz Read 1 00 15 00 0 05     0 01 KHz          60       Multifunction Process Calibrator                Specifications  Frequency Source  Range Minimum Maxi
20. WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE  FLUKE SHALL NOT BE LIABLE FOR  ANY SPECIAL  INDIRECT  INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES  INCLUDING LOSS OF DATA  ARISING FROM ANY CAUSE OR  THEORY     Since some countries or states do not allow limitation of the term of an implied warranty  or exclusion or limitation of incidental or consequential damages  the  limitations and exclusions of this warranty may not apply to every buyer  If any provision of this Warranty is held invalid or unenforceable by a court or other  decision maker of competent jurisdiction  such holding will not affect the validity or enforceability of any other provision     Fluke Corporation Fluke Europe B V   P O  Box 9090 P O  Box 1186  Everett  WA 98206 9090 5602 BD Eindhoven  U S A  The Netherlands    11 99    Table of Contents    Title Page    Introduction  TTT  Contacting BI  Standard EoOuiIprment    sss sees eee eee  Safety Iss ens  TT   E Ile  Getting Acquainted with the Calibrator   Input and Output Terminal     Ral UT  Configuration Menus sese eee eee eee eee eee ee eee  Contrast Adjustment sese eee ee eee  Shut Down es  T    Celcius and Fahrenheit    C and   F   Frequency Pulse Output Voltage ss sees eee ee eee  Pulse Output Frequency asse eee eee eee eee ee eee       726  Users Manual    HART  Resistor ONIDER   eege ENEE  Getting Started ivccshenesiieee acs Held Adie eee neki ania  Voltage to Voltage Test     Using Measure Mode  diem ni ee sn andes  Measuring Elect
21. bec17f eps  Figure 14  Connections for Simulating a 4  to 20 mA Transmitter    31    726  Users Manual                   3  Press  Vma  for dc voltage  or  FRE  for frequency  and  for resistance   4  Enter the desired output value by pressing   and    O keys  Press hl and    to select a different digit to  To select an electrical sourcing function  proceed as change     follows     1  Connect the test leads as shown in Figure 15   depending on the source function     2  If necessary  press tor SOURCE mode     Sourcing Other Electrical Parameters    Volts  ohms  and frequency are also sourced and shown  in the lower display     32       Multifunction Process Calibrator  Using Source Mode                        KAY 30V MAX ALL TERMINALS ZN  RCE   MEASURE MEASURE  v v  3 D TC  ma  Hz  S RTD LOOP  m     Ca    Common    HART        44  Communicator           4                                      MEASURE    V v  Hz  el LOOP    Common                bec16f eps    Figure 15  Electrical Sourcing Connections    33    726    Users Manual    Simulating Thermocouples    Connect the Calibrator TC input output to the instrument  under test with the thermocouple wire and the appropriate  thermocouple mini connector  polarized thermocouple  plug with flat  in line blades spaced 7 9 mm  0 312 in     center to center   One pin is wider than the other     34    A Caution    Do not try to force a miniplug into the wrong  polarization     Figure 16 shows this connection  Proceed as follows to   
22. d first    Do not use the Calibrator if it operates abnormally  Protection may be impaired  When in doubt  have the  Calibrator serviced     Do not operate the Calibrator around explosive gas  vapor  or dust     Multifunction Process Calibrator  Safety Information    e When using a pressure module  make sure the process pressure line is shut off and depressurized  before connecting it or disconnecting it from the pressure module     e Use only 4 AA batteries  properly installed in the Calibrator case  to power the Calibrator   e Disconnect test leads before changing to another measure or source function   e When servicing the Calibrator  use only specified replacement parts     e To avoid false readings  which could lead to possible electric shock or personal injury  replace the  battery as soon as the battery indicator  I  appears     e Turn off circuit power before connecting the Calibrator mA and COM terminals in the circuit  Place  Calibrator in series with the circuit     e Do not allow water into the case   ACaution  To avoid possible damage to the Calibrator or to equipment under test   e Disconnect the power and discharge all high voltage capacitors before testing resistance or continuity   e Use the proper input jacks  function  and range for the measurement or sourcing application     726  Users Manual        AC72    Alligator Clips Sy  Y    TL75 e l      Test Lead Get x Y Oj  T 4    See                   ER  m  ho   Icom    Te  g                          bec01f eps
23. e  Compatibility is listed in Table 9  Pressure Modules and values that are too low to be read if inappropriate units  Fluke model numbers are listed in Table 10  Contact your are selected  This is prevented by displaying OL on the  Fluke representative about new pressure modules not display per the following table   listed here  l    e 7OOHTP 0 to 10 000 PSI Pump  e    700PTP  11 6 to 360 PSI Pump  e 700TC1 and 700TC2 Thermocouple Mini plug Kits    Table 9  Fluke Pressure Module Compatibility                                     Pressure Unit Module Compatibility   Psi Available on all pressure ranges   In  HO All ranges through 3000 psi   cm  HO All ranges through 1000 psi   Bar 15 psi and above   Mbar All ranges through 1000 psi   KPa Available on all pressure ranges   In Hg  Available on all pressure ranges   mm  Hg All ranges through 1000 psi   Kg cm  15 psi and above       56       Multifunction Process Calibrator  Accessories    Table 10  Pressure Modules    0 to HO    Fluke 700P05 0 to 30 psi  Fluke 700P09 0 to 1 500 psi       57    726  Users Manual    Table 10  Pressure Modules  cont         58    Multifunction Process Calibrator  Specifications    Specifications    Specifications are based on a one year calibration cycle and apply from  18  C to  28  C unless stated otherwise  All  specifications assume a five minute warmup period     DC Voltage Measurement and Source    Range Accuracy   9    of Reading   Floor   30 V  upper display  0 000 30 000 0 010     2 mV  20 
24. e  In SOURCE mode  the Calibrator     e generates calibrated signals for testing and  calibrating process instruments    e supplies voltages  currents  frequencies  and  resistances    e simulates the electrical output of RTD and  thermocouple temperature sensors    e measures gas pressure from an external source   creating a calibrated pressure source     Sourcing 4 to 20 mA  To select the current sourcing mode  proceed as follows     1  Connect the test leads in the mA terminals  left  column      If necessary  press for SOURCE mode     Press  Y ma  for current and enter the desired current  by pressing 5 2   and            30    Simulating a 4  to 20 mA Transmitter    Simulate is a special mode of operation in which the  Calibrator is connected into a loop in place of a  transmitter and supplies a known  settable test current   Proceed as follows     1  Connect the 24 V loop power source as shown in  Figure 14     If necessary  press for SOURCE mode   Press  Vma  until both mA and SIM display   Enter the desired current by pressing S      and    b        Multifunction Process Calibrator  Using Source Mode             MEASURE    0000       SOURCE    12000      ZERO  Error     358conde OPEN CLOSE    2  EF    2 FREQ    TC RTD    PULSE                            HART                                     EXIT CONFIG  CONFIG  SELECTION Land                            RECALL   ENTER          TRIGGER STOI                                        Readout or  Controller                
25. e lower display  0   mA and 100   mA are fixed to 4 and 20 mA  0    and 100   for the lower display are set in source using   9   and  100    refer to    Setting 0  and 100  Output  Parameters           Stepping and Ramping the Output  Two additional features are available for adjusting the  value of source functions     e Stepping the output manually with the  42   and  keys  or in automatic mode       e     Ramping the output    Stepping and ramping apply to all functions except  pressure  which requires use of an external pressure  source     41    726  Users Manual    Manually Stepping the mA Output  To manually step current output              e Use   25  or     25   to step the current up or down in  25   steps    e Touch either L  to go to 0    or  19   to go to  100      Auto Ramping the Output    Auto ramping can continuously apply a varying stimulus  from the Calibrator to a transmitter  while your hands  remain free to test the response of the transmitter     When  mr  is pressed  the Calibrator produces a   repeating 0     100     0   ramp in a choice of three   ramp waveforms      ZN 0     100     0   40 second smooth ramp   e NM 0   100    0   15 second smooth ramp   er 0  100  0  Stair step ramp in 25    steps  pausing 5 seconds at each step  Steps are  listed in Table 7              42    To exit ramping  press any button   Table 7  mA Step Values      Step leng O    Storing and Recalling Setups       Up to eight settings can be stored in a nonvolatile  memory
26. essing   and    Press     and    to select a different digit to edit     If the 726 display indicates Ex  HI  the excitation  current from your device under test exceeds the  limits of the 726     Multifunction Process Calibrator  Using Source Mode                       FLUKE  726 Precision CALIBRATOR   MEASURE                     SENSOR  TERMINALS          bec40f eps    Figure 17  Connections for Simulating 3  and 4 Wire RTD    37    726    Users Manual    Sourcing Pressure    The Calibrator sources pressure by measuring pressure  supplied by a pump or other sources  and displaying the  pressure in the SOURCE field  Figure 18 shows how to  connect a pump to a Fluke pressure module which makes    it a calibrated source     Many ranges and types of pressure modules are available  from Fluke  see  Accessories     Before using a pressure  module  read its instruction sheet  The modules vary in  use     Attach the appropriate pressure module for the process    media  and accuracy     pressure to be tested     Proceed as follows to source pressure     38    A Warning    To avoid a violent release of pressure in a  pressurized system  shut off the valve and    slowly bleed off the pressure before attaching    the pressure module to the pressure line     ACaution    To avoid mechanically damaging the  pressure module     Never apply more than 10 ft  Ib   13 5 Nm   of torque between the pressure module  fittings  or between the fittings and the  body of the module  Always apply  appr
27. hree RTD leads are matched   Source Accuracy is based on 0 5 to 3 0 mA excitation current  0 1 mA for pt1000 range           63    726  Users Manual    Loop Power Supply    Voltage  24 V  Maximum current  22 mA  Short circuit protected     Pulse Read and Pulse Source     gt  re  m          Source 10 000 1 Count 2 CPM to 10 kHz  Read   100 000   000  Pressure Measurement     Range   Resolution   Accuracy  Units  mode      Determined by 5 digits Determined by psi  inH20 4   C  inH20 20   C  kPa    Pushing  2  for 3 seconds                pressure module pressure module cm H20 4   C  cmH20 20   C  bar  stores present pressure value  mbar  kg cm2  mmHg  inHg as an offset and subtracts it    from the displayed value       64    General Specifications   Operating temperature   Storage temperature   Stability   Operating altitude   Relative Humidity    RH operating without condensation     Vibration  Safety    Power requirements  Protection Class  Size   Weight    Multifunction Process Calibrator  Specifications    90    10 to 30  C   75    30 to 40  C   45    40 to 50  C   35    50 to 55  C   uncontrolled  lt  10  C    Random  2 g  5 to 500 Hz    EN50082 1 1992 and EN55022  1994 Class B Criteria A or B CSA  C22 2 No 1010 1 1992  4 AA alkaline batteries    Pollution Degree II  96 x 200 x 47 mm   3 75 x 7 9 x 1 86 in     650 gm  1 Ib  7 oz        65    726  Users Manual    66       Q      0  output parameter  setting   41      1      100  output parameter  setting   41      4       4
28. ifunction Process Calibrator  Using Measure Mode                            OPEN CLOSE    y ZERO    s  a Seconds  V mA SWITCH  H  HART    MAME     Ez  V mA FREQ  SS 2  SOURCE   TC RTD PULSE  EXIT CONFIG   CONFIG  SELECTION    RECALL   a  b   m                                                                                                                   Figure 8  Connections for Supplying Loop Power    bec18f eps    19    726  Users Manual    Measuring Electrical Parameters  Lower Display     To measure electrical parameters using the lower display     proceed as follows     1  Connect the Calibrator as shown in Figure 9   2  If necessary  press for MEASURE mode  lower    display                     3  Press Vma  for de voltage or current   FREG  frequency  and  RTD Jfor resistance     20    for                               FLUKE 726 PRECISION CALIBRATOR       MEASURE    0000      MEASURE    0000      ZERO    t  seconds OPEN CLOSE  V mA SWITCH    gt   LOOP SN TEST   2     HART  V mA Q FREQ                                                          TC RTD PULSE       EXIT CONFIG  100           CONFIG  SELECTION             RECALL  ENTER       TRIGGER STO                                      Figure 9  Measuring Electrical Parameters       bec43f eps    Measuring Temperature  Using Thermocouples    The Calibrator supports 13 standard thermocouples   Table 5 summarizes the ranges and characteristics of  each     To measure temperature using a thermocouple     1  Select Celsius o
29. inals and  connect the pump to the pressure switch  The  polarity of the terminals does not matter     Make sure the vent on the pump is open and  if  necessary  zero the Calibrator  Close the vent after  zeroing the Calibrator              Press     to enter switch test mode  The upper  display indicates the applied pressure  CLOSE is  displayed to the right of the pressure reading to  indicate closed contacts        Slowly apply pressure with the pump until the switch  opens     Note    Pressure the device slowly to ensure accurate  readings  Run the test several times to confirm  repeatability     Multifunction Process Calibrator    Pressure Switch Test    OPEN displays once the switch is open  Slowly bleed the  pump until the pressure switch closes  RECALL appears  on the display     5              Press  118  to read the pressure values for when the  switch opened  for when it closed  and for the  deadband     Hold msn for three seconds to restart the test  Press  2  OR  tons  to exit the switch test                                      Testing an Output Device    Use the source functions to test and calibrate actuators   recording  and indicating devices  Proceed as follows     1     2     Connect the test leads to the instrument under test  as shown in Figure 23     Press  Vma  for current or de voltage  or  FREQ  for  frequency or resistance  lower display      If necessary  press for SOURCE mode                    51    726  Users Manual       Oto 1 V de  Input    wy
30. ing    To avoid possible electric shock or personal injury     Use the Calibrator only as described in the Users Manual or the protection provided by the Calibrator  may be impaired    Do not apply more than the rated voltage  as marked on the Calibrator  between the terminals  or  between any terminal and earth ground  30 V 24 mA max all terminals     Before each use  verify the Calibrator s operation by measuring a known voltage    Follow all equipment safety procedures    Use the proper terminals  mode  and range for the measuring or sourcing application    Never touch the probe to a voltage source when the test leads are plugged into the current terminals     Do not use the Calibrator if it is damaged  Before using the Calibrator  inspect the case  Look for cracks  or missing plastic  Pay particular attention to the insulation surrounding the connectors     Select the proper function and range for the measurement    Make sure the battery door is closed and latched before operating the Calibrator    Remove test leads from the Calibrator before opening the battery door    Inspect the test leads for damaged insulation or exposed metal  Check test lead continuity  Replace  damaged test leads before using the Calibrator    When using the probes  keep your fingers away from the probe contacts  Keep fingers behind the finger  guards on the probes    Connect the common test lead before connecting the live test lead  When disconnecting the test leads   disconnect the live test lea
31. ll but the 700PA3 module  The maximum range of  700PA3 is 5 psi  therefore the reference pressure must be  applied with a vacuum pump  An accurate pressure  standard can also apply a pressure within range for any  absolute pressure module  To adjust the Calibrator  reading  proceed as follows     1  Press  2    REF Adjust appears to the right of the  pressure reading     2  Use    to increase or to decrease the Calibrator  reading to equal the reference pressure     3  Press again to exit the zeroing procedure     The Calibrator stores and automatically reuses the zero  offset correction for one absolute pressure module so that  the module is not rezeroed every time you use it     Multifunction Process Calibrator  Using Measure Mode    Isolation  Valve        Gage  Module                                                                                                                                     FLUKE 726 PRECISION CALIBRATOR    MEASURE   iDO ie 7    0000      SOURCE                                                                            agyr  a Seconds muer   F                                                              1  too   2  bei   ee  Y ma FREQ    PULSE    FEN SCH   Differential MPAAAAALAALA A  N Module    RECALL   ENTER                      TRIGGER STOF  AME                                                    Tank                                           bec37f eps  Figure 13  Connections for Measuring Pressure    29    726  Users Manual    Using Source Mod
32. mum Accuracy  CPM Source 2 0 1000 0 05    Hz Source 1 0 1000 0 0 05    KHz Source 1 0 10 00 0 25 9  10 00 15 00 0 50                     Temperature  Thermocouples    me ae JL mn JL Zen A  Accuracy Accuracy    0  0 a    J 00 E 04  02     20 poa EC HE    00 ET    1000    EO e    CJC error outside of 23   5   C is 0 05   C    C       61       726  Users Manual    NN AS  Accuracy Accuracy   EEE ERE zzz eS    oo GE ts J 12      eo deo 1 at      oo  10001 08 06   f  k     10  C to 75  C 0 015     10 UV     of Reading   Floor     10   C to 75   C 0 015     10 pV     of Reading   Floor        Maximum current output in voltage ranges is 1 mA with an output impedance of  lt  1 Q    62    Multifunction Process Calibrator  Specifications    RTD Accuracy  Read and Source   ITS 90                                                                                   Range Minimum Maximum Accuracy  Ni120  672   80 00 260 00 0 15   200 00 100 00 0 15  Pt100  385  100 00 300 00 0 25  300 00 600 00 0 35  600 00 800 00 0 45   200 00 100 00 0 15  Pt100  3926  100 00 300 00 0 25  300 00 630 00 0 35   200 00 100 00 0 15  Pt100  3916  100 00 300 00 0 25  300 00 630 00 0 35   200 00 100 00 0 75  Pt200  385  100 00 300 00 0 85  300 00 630 00 0 95   200 00 100 00 0 35  Pt500  385  100 00 300 00 0 45  300 00 630 00 0 55   200 00 100 00 0 15  Pt1000  385  100 00 300 00 0 25  300 00 630 00 0 35  CU10  10 00 250 00 1 8  Notes  Read Accuracy is based on 4 wire input  For 3 wire input  add   0 05 assuming all t
33. ompartment    bec38f eps    Figure 24  Replacing the Batteries    53    726  Users Manual    Cleaning the Calibrator    ACaution    To avoid damaging the plastic lens and case   do not use solvents or abrasive cleansers     Clean the Calibrator and pressure modules with a soft  cloth dampened with water or water and mild soap     Service Center Calibration or Repair    Calibration  repairs  or servicing not covered in this  manual should be performed only by qualified service  personnel  If the Calibrator fails  check the batteries first   and replace them if needed     To locate an authorized service center  refer to   Contacting Fluke    at the beginning of the manual     54    Replacement Parts    Table 8 lists the part number of each replaceable part   Refer to Figure 25     Table 8  Replacement Parts    hem   Semer La fan    1   AAalkaline batteries   376756   4      2   Case screws   832246   4      Battery door 1 4 turn 948609 2  fasteners      s ei   Test lead  black 688066 1   Fi ET  Manual   AC72 alligator clip  red  AC72 alligator clip  black 1670652 1  contains User Manual       Maintenance    Multifunction Process Calibrator                   Figure 25  Replacement Parts    55    726  Users Manual    Accessories External Fluke Pressure Module Compatibility  For more information about these accessories  contact The output of Fluke 700P pressure modules can cause  your Fluke representative  Fluke Pressure Module the 726 s 5 digit display to overflow  or else produc
34. opriate torque between the pressure  module fitting and connecting fittings or  adapters     Never apply pressure above the rated  maximum printed on the pressure  module     Use the pressure module only with  specified materials  Refer to the printing  on the pressure module or the pressure  module instruction sheet for the  acceptable material compatibility     Multifunction Process Calibrator    Using Source Mode  Connect a pressure module to the Calibrator as 3  Zero the pressure module as described in the  shown in Figure 18  The threads on the pressure moduless Instruction Sheet  Modules vary in zeroing  modules accept standard 4 NPT pipe fittings  Use procedures depending on module type     the supplied   NPT to   ISO adapter if necessary  4  Pressurize the pressure line with the pressure    Press       lower display   The Calibrator source to the desired level as shown on the display   automatically senses which pressure module is  attached and sets its range accordingly           If desired  continue pressing S   to change  pressure display units to psi  mmHg  inHg   cmH 0 04   C  cmH 0220   C  inH 004   C   inH 0 020   C  inH O O 60   C  mbar  bar  kg cm     or  kPa     39    726  Users Manual                                          Hand Pump                                                                    Measure mA   N Source Pressure  1 24 V Loop Power  Enabled                               ZERO  Error Second     OPEN CLOSE    V mA SwrrcH   Loop     amp  TEST 
35. ower  display  modes to calibrate a transmitter  This section  applies to all but pressure transmitters  The following  example shows how to calibrate a temperature  transmitter  Use the following steps to calibrate a  transmitter     1     Connect the Calibrator to the instrument under test  as shown in Figure 20    Press for current  upper display   If required   press again to activate loop power    Press L Tre    lower display   If desired  continue  pressing this key to select the desired thermocouple  type        a    Multifunction Process Calibrator  Calibrating a Transmitter    If necessary  press for SOURCE mode     Set the zero and span parameters by pressing     and   Enter these parameters by pressing and  holding and 100    For more information on  setting parameters  see    Setting 0   and 100    Output Parameters        Perform test checks at 0 25 50 75 100   points by  pressing or  72    Adjust the transmitter as  necessary    Note    When HART resistor mode is selected  the  250 Q resistor is turned on both mA channels     45    726  Users Manual                            FLUKE 726 PRECISION CALIBRATOR       MEASURE LOOP                   Error ds OPEN CLOSE  V mA SWITCH   Loop    TEST    HART  V mA 2 FREG                                                                            CONFIG  SELECTION                ENTER                                                    4 Communicator                   l   l   ai HART  c   l                          bec44f eps
36. r Fahrenheit  depending on the  desired measurement  Refer to  Configuration  Menus    for more information     2  Attach the thermocouple leads to the appropriate TC  miniplug  then to the TC input output as shown in  Figure 10     Multifunction Process Calibrator  Using Measure Mode    A Caution  One pin is wider than the other  Do not try to force a  miniplug into the wrong polarization   Note  If the Calibrator and the thermocouple plug are at  different temperatures  wait one minute or more    for the connector temperature to stabilize after  plugging the miniplug into the TC input output     3  If necessary  press for MEASURE mode     Press L Te   for the thermocouple display  Continue  pressing this key to select the desired thermocouple    type        B    21    726  Users Manual    22       Table 5  Thermocouple Types Accepted    Positive Lead  H   Positive Lead Color Negative Lead Specified Range    Material   ANSI    lEC   Material    C   Orange    Platinum  10   Rhodium  Orange  20 to 1750    Tungsten 5  Rhenium None Tungsten 26  Rhenium 0 to 2316  GOST    90 5   Ni   9 5   Cr 56   Cu   44   Ni  200 to 800  Violet or Black  95  W  5   Re Red or Pink 80   W   20   Re 0 to 2500     American National Standards Institute  ANSI  device negative lead  L  is always red     International Electrotechnical Commission  IEC  device negative lead  L  is always white           Multifunction Process Calibrator  Using Measure Mode       SOURCE   MEASURE    iz  TD    com    ect   
37. rical Parameters  Upper Display  sese ee eee eee eee  Current Measurement with Loop Power   Measuring Electrical Parameters  Lower Display  see ee ee eee eee eee  Measuring Temperature sssrin a aa Eat  Using TherMocouples     ooooccconnnnccnnnnccccnnonnncnnnnnnncnnnrncrnnnr creian  Using Resistance Temperature Detectors  RTDs    PRT Custom Curve sees sese eee  Measuring GE  Zeroing with Absolute Pressure Modules AAA  Using Source Mode et penis eat  Sourcing 4 tO 20 mA     Simulating a 4  to 20 MA Transmitter  Sourcing Other Electrical Parameters see eee eee  Simulating Ria EE  Simulating RTDS      sreervrnnnnrnnrvrnnnnrnnnvrnnnnrnenvrnnenrnnsnrnnenvnsnnrnsennnsnnrnnsnvnsenresenvenenensenresenen  Sourcing Pressure  iii dr desk  Setting 0   and 100   Output Parameters eee eee ee eee eee    Error Functionality EE  Stepping and Ramping the Output     Manually Stepping the mA Output     Auto Ramping the Output     Storing and Recalling Setups  cnoic renie arua E 2925 EEN  Store TTT  STT arica       ii    Contents  continued     Storing and Recalling Data   Storing Data EE  Recall Data  EEE EE SEEEEEn   Pulse Train Soutceibeagd  ege Cvt dE EEN paste teres   Calibrating a Transmitter      Calibrating a Pressure Transmitter    Calibrating an UP Device A   Pressure Switch TOS tirita negra   Testing an Output Device sse sese eee ee eee eee   Remote Control Commande AAA   HARTA Functionality cocinada pt linde   Maintenance ccoo cada  Replacing the Batteres AAA  Cleaning the Calibrator e
38. rized thermocouple plug with flat  in line blades spaced 7 9 mm   0 312 in  center to center     5    6  SOURCE  MEASURE V  Terminals for sourcing or measuring voltage  resistance  pulse  frequency  and  RTD  Pulse  Hz  Q terminals   RTDs   7  SOURCE  MEASURE mA Terminals for sourcing and measuring current and performing 3 W and 4 W RTD  k terminals  3W  4W measurements  HART resistor option in mA mode           726  Users Manual    Keys  Figure 3 shows the Calibrator keys and Table 4 explains their use                                                                                                                                   bec41f eps    Figure 3  Keys    10    Multifunction Process Calibrator  Getting Acquainted with the Calibrator    Table 4  Key Functions    mpe men O O  1a        Turns the power on or off      Error  KA Toggles voltage  mA  or Loop Power and   Error measurement functions in the upper display   LOOP    SS Selects the pressure measurement function in the upper display  Repeated pushes cycle through the     2  different pressure units  Zeros pressure when held for 3 seconds     OPEN CLOSE   E  bes Activates the switch test           Turns backlight on or off      FREQ    Selects frequency sourcing or measurement   PULSE    Selects pulse sourcing or measurement     Recalls a source value from memory corresponding to 100   of span and sets it as the source value           EXIT CONFIG  30054 Press and hold to store the source value as the 100   value   o
39. simulate a thermocouple    1  Attach the thermocouple leads to the appropriate TC  miniplug  then to the TC input output as shown in  Figure 16    If necessary  press for SOURCE mode    Press  79   for the TC display  If desired  continue  pressing this key to select the desired thermocouple  type    4  Enter the temperature you want by pressing    and  O keys  Press hl and    to select a different digit to  edit        Multifunction Process Calibrator  Using Source Mode                               OTESTDCPWR O    Color depends      galle    E S on type of TC NN    ma   2   Fa    RTD PULSE                                              EXIT CONFIG    CONFIG  SELECTION 00                    A 25        RECALL   ENTER          y 25        TRIGGER STO   AMS Retum t ge    tetum to      Recall E                                                    oJ TC Miniplug          bec20f eps    Figure 16  Connections for Simulating a Thermocouple    35    726    Users Manual    Simulating RTDs    Connect the Calibrator to the instrument under test as  shown in Figure 17  Proceed as follows to simulate an    RTD     1   2     36    If necessary  press for SOURCE mode   Press LPT   for the RTD display     Note    Use the 3W and 4W terminals for measurement  only  not for simulation  The Calibrator simulates  a 2 wire RTD at its front panel  To connect to a  3 wire or 4 wire transmitter  use the stacking  cables to provide the extra wires  See Figure 17        Enter the desired temperature by pr
40. steps explain how to calibrate a device that   controls pressure  Proceed as follows    1  Connect the test leads to the instrument under test as  shown in Figure 22  The connections simulate a  current to pressure transmitter and measures the  corresponding output pressure     2  Press  upper display       gt     Multifunction Process Calibrator  Calibrating an I P Device       Press  Vma  for sourcing current  lower display    If necessary  press for SOURCE mode     Enter the desired current by pressing    and  amp    Press hl and Q to select different digits     49    726  Users Manual                                                                   aSeconds OPEN CLOSE                SWITCH  TEST                      FREQ                            PULSE       Come    SELECTION          RECALL   ENTER    TRIGGER STOF  AMF  CJ                   EXIT CONFIG  100        a 25           w 25                       50       Measure  Pressure  Source mA                Pressure       Module                                              Red I        HART          Communicator                               Figure 22  Calibrating a Current to Pressure  I P  Transmitter             bec28f eps    Pressure Switch Test    Note  This example uses a normally closed switch     The procedure is the same for an open switch  but the display reads OPEN instead of CLOSE     To perform a switch test     1     Connect the Calibrator mA and COM terminals to the  switch using the pressure switch term
41. t1000  385  1000 Q 0 00385 Q   C  200 to 630  Pt100  3916  100 Q 0 003916 Q   C  200 to 630    The IEC standard RTD commonly used in U S  industrial applications is the Pt100  385   a   0 00385 Q   C   Pt100  3916   a   0 003916 Q  C is also designated as JIS curve   Custom RTDs may also be added  see PRT Custom Curves       25    726  Users Manual    MEASURE    v v     a a     RTD LOOP  5 P    COM          30V MAYA  SOURCE   MEASUg  ma                    bec15f eps  Figure 11  Measuring Temperature with an RTD  Measuring 2   3   and 4 Wire Resistance    26    Measuring Pressure    Many ranges and types of pressure modules are available  from Fluke  see  Accessories     Before using a pressure  module  read its instruction sheet  The modules vary in  use  media  and accuracy     Figure 12 shows the gage and differential modules   Differential modules also work in gage mode by leaving  the low fitting open to atmosphere     To measure pressure  attach the appropriate pressure  module for the process pressure to be tested  and  proceed as follows     A Warning    To avoid a violent release of pressure in a  pressurized system  shut off the valve and  slowly bleed off the pressure before attaching  the pressure module to the pressure line                 Multifunction Process Calibrator  Using Measure Mode    Gage Differential             High                                                          gj11f eps    Figure 12  Gage and Differential Pressure Modules    ACaution    To
42. ulse Train Source Read  44        R     ramping   41     Recalling setups  42   43    Remote control commands   52    Repair   54    Resistance Temperature Detectors  accepted types  table    25    RTD    simulating   36   RTD  measuring  24   types   24      S     Safety information   Saving  Measurements   43   Servicing  54   Setup  Recall     recalling     storing   42   43   shut down mode   Thermocouple   Simulating  loop power   30   RTD   36   thermocouples   34   Source functions  summary  table      Source mode   30    Sourcing  4 to 20 mA   30   electrical parameters   32   sucia    thermocouples   Specifications   Standard eq uipment   3   stepping TZ  Stepping output   Storing setups   symbols   7           Temperature  measuring with RTD   24   measuring with thermocouple   21   Terminals    input   8  output   8    Thermocouple  measuring  21     measuring temperature   21   sourcing     types   thermocouples    accepted types   22   Transmitter    4 to 20 mA  simulating   Transmitter  calibrating      V     voltage to voltage test        Z     Zeroing pressure modules   28     Index  continued     69    726  Users Manual    70    
43. urcing function in lower  display  Repeated pushes cycle through the RTD types  Selects resistance mode     Selects the pressure measurement and sourcing function  Repeated pushes cycle through the  different pressure units                          12    Display    Multifunction Process Calibrator  Getting Acquainted with the Calibrator    Figure 4 shows the elements of a typical display     Low Battery    Mode       er A MEASURE LOOP 12345678 HART      desto  lt   B0565 ABE 7      EBBBB8B    Loop Memory Locations HART Mode  Indicator for Calibrator Setups Indicator       SOURCEMEASURE AUTOA MF                Ramp    Figure 4  Elements of a Typical Display    Units  Display       bec07f eps    13    726  Users Manual    Configuration Menus    Use the configuration menus to set or change these  parameters of the Calibrator     e Contrast Adjustment   e Shut Down Mode   e CJC on off   e  CPF   e Frequency Pulse output voltage  e Pulse output frequency   e HART resistor on off    To enter the configuration menus  press    m   Press  ENE   to save new configuration  Press  100   EXIT  CONFIG to exit configuration     Configuration menus are explained below   Contrast Adjustment  To adjust the contrast  see Figure 5      1  Press h  m  until Contst Adjust appears on the  display        2  Use    and to adjust the contrast up and down     3  Press to save the setting     14                   FLUKE 726 PRECISION CALIBRATOR  MEASURE LOOP 12345678     88888    SOURCEMEASURE AUTOA Mr  
44. x muiden atea  Service Center Calibration or Repair    eee ee eee eee  Replacement Parts ciocie ich ee el daw ae eee he   PA CCOSSOMES drake a dende de LOSH AE ESA  External Fluke Pressure Module Compatibility   Specification eee  DC Voltage Measurement and Source eee eee eee eee eee  DC mA Measurement and Source sse eee ee eee ee eee  Ohms Measurement cummins iii  Ohms ele  Frequency Measurement essere sseee esec essere esec e esec essere eser enen  Frequency Source sse ee eee eee eee eee  Temperature  Thermocouples    ssssssee essere esec e esec esse cesse essere ener  RTD Accuracy  Read and Source   ITS 90        rrnrrnnnvvnenvrnnnvenenvrnnnnrnrrvrnnnnnnrrvrnnrnreren       iii    726  Users Manual    Loop Power Supply    Pulse Read and Pulse Gource mesnsrrrnine n e a A aE E Ea E TaS  Pressure Measurement eee e eee  General Specifications iii navnene det etek eb   kes dannede    List of Tables    Table Title  1      Summary of Source and Measure Functions sees eee eee eee  2  International Symbole    AA  3  Input Output Terminals and Conpnechors sese eee eee eee ee eee ee cnn aran rana  4  Key Tele  TT  5  Thermocouple Types Accepted      ssssseessee esse esse e sees neee  D  RTD Types ee ee T  7  MA Step Values  iuris heade Rares  8  Replacement Parts          rrnnnnnvnnnnnnvnnnnnnvnrnnnnnnnnnnnvnnnnnnnvnennvnnnnnnvnnenennnnnnensnnnnnnnvnvnnenennnnensvnvennnnn  9  Fluke Pressure Module Compatibility sese eee eee ee eee ee eee ee  10  Pressure le TEE       726  Users Manual
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Untitled    Siemens Spectrodensitometer FD-5 User's Manual  Q5K-YSK-055-E11  Bedienungsanleitung Instructions de service Operating Instructions  取扱説明書 - Maxon  Sencor SHF 1010 humidifier  LG Electronics 50PY2DR Flat Panel Television User Manual  user manual Safety precautions  Sommaire    Copyright © All rights reserved. 
   Failed to retrieve file