Home
        OC-725C User`s Manual (Rev. 050816.2)
         Contents
1.   The resulting voltage  in mV  will be displayed on the METER in the  BATH ELECTRODE section  From that value  the resistance of the pipette can be calculated       exactly as described above  1 e   divide the readout by 10 to get the resistance in MQ    Since the current electrode has a larger diameter  its resistance should be less than  that of the voltage electrode  about 1 0 MQ or less      If no voltage display is present during the electrode test procedure and you   re sure that  the electrode is contacting the bath  perform the following checks     a  Make sure that all cables are connected properly    b  Check to see that the aluminum shield around the current electrode pipette  if  used  is not touching the bath solution or the electrode wire    c  Check the current electrode to see if there is a bubble in the pipette causing an open    circuit     Impaling the cell    10     11     12     13     Recheck the VOLTAGE ELECTRODE METER to verify that the potential is correctly offset to  read 0 0 mV and readjust the Vm OFFSET control if needed        Advance the VOLTAGE ELECTRODE until its tip is slightly depressing the plasma       membrane of the cell and depress the Vm BUZZ pushbutton  This will produce a 1 V  10  kHz oscillation at the voltage electrode  disrupting the membrane and causing the tip  of the electrode to impale the cell with no further movement of the micro manipulator   this technique will work best with  fresh  oocytes  i e   1 or 2 days post excisi
2.   the following measurements were made with a shorted input  with the  standard 0 5 uF capacitance model cell  and with a 0 22 uF capacitance modified model cell   All readings were recorded at 1 kHz  8 pole Bessel  and are RMS                             Noise  Feedback  Resistor Shorted Input Standard Model Cell   Modified Model Cell   0 5 uF   0 22 uF   10 kQ 75 pA 6 0 nA 4 4nA  100 kQ 28 pA 5 5 nA 4 4nA  1 MQ 22 pA 5 0 nA 4 0 nA             A lower value feedback resistor increases the speed of the bath clamp and can handle larger  currents without saturating  important when recording from high expression oocytes     Warner instruments  A Harvard Apparatus Company    A PROCEDURE FOR RECORDING FROM OOCYTES    Initial electrode placement    1     2     Make sure that the bath electrodes are submerged in the chamber  or in the agar  bridge wells with the agar bridges completing the circuit to the bath  and the oocyte is  stable on the chamber floor     Install the voltage and current pipettes onto their respective holders but do not yet  place them in the chamber bath solution     Voltage electrode placement    3     4     5     6     Advance the voltage recording electrode into the bath  The VOLTAGE ELECTRODE METER       will indicate  in mV  the potential between the electrode and the bath     If there is no voltage reading and you are sure that the pipette tip is in the bath  solution  perform the following checks     a  make sure that all cables are connected properly   
3.  7  switches are off     4  To convert to high side current measuring mode  turn S4 off and turn 82  S5  and S7 on        All other switches should remain in the off position     Optional voltage headstage    The optional DIFFERENTIAL INPUT VOLTAGE RECORDING HEADSTAGE  Model 7255DI  is used in       applications where the bath clamp headstage cannot be effectively employed  Two examples  are  1  situations where the solution path from oocyte to ground is very long and  2  when  recording from two oocytes in a common bath     Two 1 mm input jacks are located on the side of the headstage  V DIFF and GND  A  shorting jumper is supplied and is used for normal single ended recording  For double     Warner Instruments  A Harvard Apparatus Company    ended recording  the jumper is removed and a V DIFF ELECTRODE is connected to the V DIFF       jack as shown below     If a shield between the voltage and clamp current electrodes is used  it should be  connected to the headstage ground     Warner Instruments  A Harvard Apparatus Company    COMMENTS AND RECOMMENDATIONS    Membrane damage    Recording from the same cell at a later time requires that the cell remain healthy  during the interim incubation  The less damage done to the membrane during handling and  impaling the cell  the happier it will be  Use of the Buzz function should help minimize the  trauma from electrode penetration  Membrane damage can be further reduced by properly  isolating the experimental platform from vibratio
4.  CLAMP MODE switch is set to off     Return all controls to their initial settings when done with this test     Voltage electrode test    In actual practice  the voltage electrode test is used prior to entering the cell and indicates  the resistance of the electrode  When used in conjunction with the model cell  it measures  both the electrode and membrane resistance  2 MQ   This test is performed using controls  in the VOLTAGE ELECTRODE section        With the model cell in place  depress the Vm ELECRODE TEST pushbutton and observe a  reading of 20 mV on the VOLTAGE ELECTRODE METER  This corresponds to a 2 MQ reading  at a calibrated response of 10 mV MQ        On the oscilloscope  the Vm x10 output will read 200 mV which is x10 the applied test  voltage  Since the test current is being passed through 1 MQ  Rn   the I MONITOR output  will indicate 1 mV  which corresponds to 10 nA of current     With the CLAMP MODE switch set to off  the BATH ELECTRODE METER monitors Ve  voltage at the       current electrode   In this case  Ve will be a measure of the voltage across Rm and the  meter will indicate 10 mV  1 MQ   The Ve x10 output BNC on the rear panel can also be  checked to see that it reads 100 mV  meter reading x10      Buzz    This test is performed using controls in the VOLTAGE ELECTRODE section        Set the oscilloscope sensitivity to 5 V div and depress the BUzz pushbutton while monitoring  the Vm x10 output  A 10 kHz square wave of approximately 24 V p p will be gen
5.  off                    AC clamp test 100     Adjust the Vm OFFSET control to 0 0 V  ei  Apply a 0 8 V  100 Hz square wave to 40    the COMMAND IN  10 BNC  Monitor the   odi  Vm x10 and I MONITOR outputs on the F Pi    oscilloscope   40 4  Switch the CLAMP MODE switch to fast and       increase the GAIN until Vm reads 80 FO 2 4 6 8 10 12 14  mV  Verify that the Vm x10 BNC ms  reports 0 8 V  As you further increase 10  the cam control  you will see the rise    time of the oscilloscope trace become 4  faster since the speed of the clamp is 2  limited by the resistance of the    gt   current electrode and the capacitance  4  of the oocyte  If ringing  oscillation  is u  observed  decrease the cam setting to    obtain the fastest clean waveform as 0 2 4 6 8 10 12 14  ms    Warner Instruments  A Harvard Apparatus Company    shown on the previous page   The current signal  I MONITOR  shown in the figure  displays the high current spikes required to charge the oocyte capacitance      Set the HOLD control to a reading of 50 mV and switch the POLARITY toggle to pos  The square  wave will be displaced 50 mV in the positive direction  Switching the polarity to neg  will produce a  50 mV offset     Warner Instruments  A Harvard Apparatus Company    SETUP    The following instructions are designed to guide the user  step by step  through a  typical recording session involving a Xenopus oocyte  It is assumed that the user is already  familiar with the techniques of Xenopus oocyte excision 
6.  rather than complexity  we welcome any feedback  you may wish to provide      gt  Warner Instrument product numbers are presented using a bold type      gt  References to instrument panel control blocks are specified using UNDERLINED SMALL CAPS    e g   COMMANDS  CLAMP            gt  References to specific controls within a block are specified using NON UNDERLINED SMALL  CAPS   e g   MODE SWITCH  DC GAIN        Finally  references to individual control settings are specified in italic type   e g   slow   fast  100 mV      gt  Special comments and warnings are presented in highlighted text     Any other formatting should be apparent from context        Warner Instruments  A Harvard Apparatus Company    CONTROL DESCRIPTION    The OC 725C is comprised of three functional channels  A high impedance voltage  sensing channel with capacity compensation and input offset to measure membrane  potential  a current sensing channel with bath clamp to clamp the bath and measure the  membrane current  and a high voltage amplifier to deliver the clamping current     The complete voltage clamp system consists of the OC 725C  the voltage recording  probe with electrode holder  the current sensing bath probe with silver wire electrodes  and  the current cable with electrode holder     Front panel    The instrument front panel is divided into six control blocks titled VOLTAGE ELECTRODE   Vm   BATH ELECTRODES  Im   CLAMP  COMMANDS  and CURRENT ELECTRODE  The instrument rear  panel has BNC connec
7.  the OC 725C can be used for two purposes  First  novice  users will find it a convenient tool for gaining experience in the operation of the instrument   Additionally  it is a convenient tool for trouble shooting since the function of the  instrument can be quickly checked     A schematic of the  model cell is shown to the  right  The oocyte is  represented by a 1 MO  resistor shunted by a 0 47  uF capacitor  The voltage 1M t   1M  and current electrodes are i  each represented by 1 MO VA n  resistances and the bath to voltage to current   probe cable    probes are represented by  the 10 kO resistors        Warner Instruments  A Harvard Apparatus Company    Initial instrument settings    Connect the model cell to the OC 725C as shown on its cover  Be sure to connect the  ground wire to the ground mini jack on the side of the bath probe  Connect the Vm x10 BNC  and the I MONITOR BNC on the OC 725C to an oscilloscope     Set the instrument panel controls as follows and turn the POWER SWITCH on                                                     Control Control block Setting  POWER SWITCH off  Vm OFFSET VOLTAGE ELECTRODE Center of rotation   approximately 5 turns   OUTPUT GAIN BATH ELECTRODES 0 1 V uA  GAIN SELECT BATH ELECTRODES x1 0  CLAMP MODE switch CLAMP off  DC GAIN toggle CLAMP  out  GAIN CLAMP CCW to detent off  HOLD POTENTIAL COMMANDS 00 mV  HOLD POTENTIAL MULTIPLIER COMMANDS x1 0  HOLD POLARITY COMMANDS off  Ve OFFSET CURRENT ELECTRODE Center of rotation   approximate
8. 2  EN61000 4 2  EN61000 4 3  ENV50204  EN610000 4 4  EN610000 4 8  EN610000 4 11    Manufacturer s Name  Warner Instruments  LLC    Manufacturer s Address  1125 Dixwell Avenue  Hamden  CT 06514  Tel   203  776 0664    Equipment Description  Instrument Amplifier    Equipment Class  ITE Class A    Model Numbers  OC 725C    I the undersigned  hereby declare that the equipment specified  above  conforms to the above Directive s  and Standard s      Place  Hamden  Connecticut USA    Pene    Full Name  Burton J  Warner    Signature     Position  President       Warner Instruments  A Harvard Apparatus Company    Declaration of Conformity  CE MARKING  LVD     Application of Council Directive  73 23 EEC    Standards To Which Conformity Is EN61010 1 1993  Declared     Manufacturer   s Name  Warner Instruments  LLC    Manufacturer   s Address  1125 Dixwell Avenue  Hamden  CT 06514  Tel   203  776 0664    Equipment Description  Instrument Amplifier  Safety requirements for electrical  equipment for measurement and  laboratory use   Equipment Class  Class I    Model Numbers  OC 725C    I the undersigned  hereby declare that the equipment specified  above  conforms to the above Directive s  and Standard s      Place  Hamden  Connecticut USA  Signature     Gab pr    Full Name  Burton J  Warner    Position  President       Warner Instruments  A Harvard Apparatus Company    OC 725C Manual  Rev  050816 2 33    WEEE RoHS Compliance Statement    EU Directives WEEE and RoHS    To Our Valued Custo
9. OC 725C Manual  Rev  050816 2    Warner Instruments    Oocyte Clamp Amplifier  Model OC 725C       Warner Instruments  1125 Dixwell Avenue  Hamden  CT 06514   800  599 4203    203  776 0664   203  776 1278   fax    CE    OC 725C Manual  Rev  050816 2    Table of Contents    NOMENCEATURE DC 4  Text conventions EA AN A                             4  CONTROL DESCRIPTION 2 005058                   t 5  Front panel    NN RN 5  Voltage electrode  ete ee ee tete e ete Dee Ne ER ka Pu n cu a eee Lene eee Eee ere Rana Po heed a 5  Bath electrodes utis A HH RS e etta eis 6  Clamp sectioniz a siio e tor RE ee os ie eee e heben ert eee eade Saat ERE AEDs dn 6  C  mman  sa uincelenaichesneikunnnesheininsiehnhsstsehnninsersnhshtaesnnineiehehentnernshneissnhnhtehsnnineiehnhnsl  enenkl  kreshesitrahsthe 7  Current  electrode    rte ee e etna Forsitan ei a vedi a aia 7  horiWrud e                                      8  EDDIE FED HIDE ETE EL 9  Voltage recording headstages                eese esee eene enne ettet entente setis entente enne 9  Bath headstage AERA S Re dee ex 9  Currentelectrode c  ble    ii nennen ana decanos lees addin rE Lada PATER RAE i P S 9  Model Cell    ee iei ettet ottenere tti o Ea a Eee do LEER b E rne ode bord ea ed doceo PU LER De Leod 10  LICET E 10  Connecting to line power    eee eei eet n e D E Ree Se ee e eg Bae ibe een 10  High voltage ouputs  us etui tette IR c A be Pa eee e e daa Pe leas e ele lern ru eee 10  USING THE MODEL MEMBRANE                    ee e
10. and microinjection  for a review of  those techniques  see Colman  1984   It is also assumed that the user has some familiarity  with the basic circuitry of a two electrode voltage clamp  for review  see Hille  1984      Pipettes    Microelectrodes can be made using the same glass  tubing and dimensions  as those  used for a typical patch pipette and are usually filled with 3 M sterile filtered KCl  Unlike  the pipettes used as patching electrodes  microelectrode pipettes do not require fire  polishing nor coating with Sylgard  They will need to be broken off  however  to a relatively  large diameter to insure a fast response time by the clamp     For the voltage electrode  the pipette tip should be broken back to an O D  of 3 5 um   The current electrode pipette should be broken back to an O D  of 7 9 um  The resistances of  these pipettes should be about 2 MQ and 1 MQ  or less   respectively     1 5 to 2mm    re 5mm          BREAK TIP TO DIAMETER  OF 3 5 MICRON FOR V ELECTRODE  25mm 0 3mm OR 7 9 MICRON FOR   ELECTRODE    0 6mm    When installed  the current electrode pipette should be shielded from the voltage  electrode and that shield should be grounded to the circuit ground  This can be  accomplished by wrapping the current pipette with aluminum foil or by mounting a metal  screen or plate between the two pipettes  In either configuration  the shield can be  grounded by connecting it to the  ground  mini jack on the side of the bath probe  When  using the aluminum foil m
11. atus Company    
12. b  inspect the voltage electrode to see if there is a bubble in the pipette which will  cause an open circuit     Using the Vm OFFSET control  adjust the VOLTAGE ELECTRODE potential to read 0 mV  If the  junction potential of the voltage electrode can not be adjusted to 0 mV  the electrode  holder may be at fault   See Electrode Holders  page 17      To test the resistance of the voltage electrode pipette  depress the ELECTRODE TEST  button  This passes a 10 nA current across the voltage electrode  The VOLTAGE  ELECTRODE METER will display the resulting potential in mV  The resistance of the  electrode can be easily calculated by dividing the current into the potential  The  resulting answer will be expressed in Q  For example  if the electrode test indicates  that a potential of 25 0 mV is produced by the 10 nA test current  then    pa  ZMV 5 5x1050  25M0       I 10nA       Warner Instruments  A Harvard Apparatus Company    Current electrode placement    7     8     9     Advance the current electrode until the tip is in the chamber bath solution  Adjust Ve  OFFSET for a zero reading on the CURRENT ELECTRODE METER  This will establish a null       reference allowing the resting potential to be directly read     With the CLAMP SELECTOR switch in the off position  the resistance of the CURRENT  ELECTRODE pipette is tested in the same manner as the VOLTAGE ELECTRODE  Pressing the  Ve ELECTRODE TEST pushbutton will cause a 10 nA current to be passed across the  CURRENT ELECTRODE
13. d Apparatus Company    NEGATIVE CAPACITY COMPENSATION   C  has been added to the OC 725C allowing for its use  as an electrometer in intracellular measurements  Input capacitance up to 45 pF  can be neutralized using the two associated controls  A lit LED indicates the active  status of this circuit     In general  negative capacity compensation is not useful for oocyte clamp  applications since clamp speed is a function of  1  the current electrode resistance    2  the RC time constant of the oocyte  typically 1 MO in parallel with 0 5 uF   and   3  the compliance voltage of the clamp current     Bath electrodes    The BATH ELECTRODES control block contains  the BATH PROBE connector  the CURRENT METER   the GAIN SELECT switches  and the I MONITOR  output BNC   s        The CURRENT METER reads the voltage  Ve  of   d ev ion  the current electrode when the CLAMP   A BATH PROBE  MODE SELECTOR switch is off  see CLAMP    section   A lit LED indicates voltage       readings in mV     When in clamp mode  CLAMP MODE SELECTOR switch set to slow or fast   the CURRENT  METER displays the current  Im  sensed by the bath electrode  A lit LED indicates  current readings in A     Instrument gain is set by the two GAIN SELECT controls  Gain is selected by a 7 position  GAIN SELECTOR switch  which ranges from 0 1 to 10 in 1 2 5 steps  and a 3 position  toggle switch which selects the gain multiplier  x0 1  x1 0 and x10   LED s indicate  the gain multiplier selection  The combination of thes
14. d on the BATH  ELECTRODES METER in units of uA  Fast clamp speeds are  350 us when  measured with the model cell as described above     The GAIN control is a single turn potentiometer which varies the full bandwidth open   loop gain from 0 to 2000  A high Dc GAIN  109  can be switched in with the DC GAIN  toggle switch to provide a hard clamp when passing large currents from high  expression oocytes     Commands    The COMMANDS control block contains the HOLD controls and COMMAND COMMANDS  IN 10 input BNC     HOLD controls   HOLD potential is set with the DIGITAL POTENTIOMETER       thumbwheel and RANGE toggle switch  Ranges are  99 mV and        198 mV depending on the scale multiplier selected  x1 0 or  x2 0   Signal polarity  or off is selected with the associated toggle  switch     COMMAND IN  10 input BNC  Command signals from an external  generator or computer connected to this input are attenuated by       10  Maximum input is  10 V     Current electrode    The CURRENT ELECTRODE block includes the Ve OFFSET  ELECTRODE TEST  and Ve BUZZ  controls  This section also contains the OVER VOLTAGE indicator and a DIN connector for  the current electrode     Warner Instruments  A Harvard Apparatus Company       Ve OFFSET   With a range of  200 mV  center zero   this CURRENTE  control is used to adjust the offset voltage of the Ve OFFSET  current electrode  Use this control to establish a zero  reference before impaling the oocyte  Once the oocyte  has been pierced  the resting 
15. dditional  DC gain of 106 may be employed for high  conductance cells or leaky oocytes        Two clamp speeds are available  The Slow  mode is used for screening oocytes or for  applications not requiring rapid response  times  The Fast mode is used for accurate  voltage clamping of fast whole cell  currents  Response time in the Fast mode  is 350 us  10 90  rise time  when  applying a 100 mV step to a model cell     Bath Clamp Headstage  The current  measuring range of the OC 725C bath  clamp headstage is extended at both ends  by a 3 position range multiplier  This  allows smaller currents to be amplified to  usable levels while larger currents  up to  1 mA  can be recorded without saturation     The unique design of the bath clamp  eliminates the need for series resistance  compensation  It provides an accurate  measurement of bath current by creating  a virtual ground in the bath while  simultaneously clamping the bath  potential at zero     Buzz controls for each electrode aid in  penetration of cell membranes with a  minimum of leakage     Electrode Test for voltage and current  electrodes     Capacity Compensation for the Vm  voltage input     Dual Oocyte Studies  Studies involving  two oocytes in a common bath requires  two clamp amplifiers  Traditional bath  clamp headstages cannot provide effective  clamping because they cannot separate  the individual currents from the combined  currents appearing in the bath  The  OC 725C solves this problem by the  application o
16. e     Vm   Membrane potential may be recorded from the Vm x10 connector  VOLTAGE  ELECTRODE section      Ve   The voltage of the current electrode can be monitored from the Ve connector on the  REAR PANEL     The output will be the same as that reported on the CURRENT ELECTRODE METER        showing the voltage across the current electrode when the CLAMP SELECTOR switch is  in the off position   Recall that the meter indicates the current  Im  when the camp  SELECTOR switch is in either the slow or fast position      Warner Instruments  A Harvard Apparatus Company    Im   The current signal is available from the I MONITOR and I MONITOR FILTERED outputs   BATH ELECTRODES section         Gain telegraph     Automatic monitoring of the Im gain can be achieved by connecting the  rear panel GAIN TELE BNC to the appropriate input on your analog to digital converter     Power  The power cord should be connected to a properly grounded AC receptacle with the  line voltage specified on the instrument nameplate  REAR PANEL      Resting position of the controls    To begin  set the instrument controls to the following positions                                         Control Section Setting  POWER off   I MONITOR output BATH ELECTRODE 1 V uA  GAIN SELECT BATH ELECTRODE xl   HOLD COMMANDS 0 mV  POLARITY toggle COMMANDS off  MODE SELECT CLAMP off  CLAMP DC GAIN toggle CLAMP out  CLAMP GAIN CLAMP off  fully CCW   ALARM REAR PANEL off or on  as desired                    Turn POWER on     Usin
17. e  DC and may need to be IR and UV filtered if you plan to use it during recording     Minimally  the recording chamber can be a stable surface on which the oocyte will not  roll around  A disposable petri dish with a piece of nylon mesh on the bottom has been  successfully used for this purpose  The dish can be stabilized by a holder or by some wax  placed around its perimeter  The diameter of the dish needs to be large enough to  accommodate the oocyte and the two bath electrodes  The walls of the recording chamber  should be low enough to not interfere with electrode placement  Perfusion of the chamber  can be accomplished using a gravity fed system  Perfusate from the dish can be evacuated  using gentle vacuum  Use as small an aperture as possible to avoid disturbing the surface    of the perfusate in the dish        Although it is not required  a computer can be employed to control the command  voltage  Acquisition and display of data is also usually handled by computer  The OC 725C  is fully compatible with all commercially available software packages designed for  electrophysiological research     Finally  a microelectrode puller is necessary for making appropriately sized voltage and  current electrodes  Usually  the microinjection pipette puller can also be used to make  microelectrodes  You will need use of a microscope to break off the pipette tips     Warner Instruments  A Harvard Apparatus Company    APPENDIX    Specifications   Equipment is intended to be operat
18. e controls allows gain to be  set from 0 01 to 100     Current outputs are available from the 1 MONITOR BNC at full bandwidth   10 kHz  and  from the I MONITOR FILTERED BNC which is filtered at 1 kHz by an integral 4 pole  Bessel filter     Clamp section    The CLAMP control block contains the MODE SELECTOR switch as well as  the GAIN and DC GAIN controls     The CLAMP MODE SELECTOR switch selects for slow and fast clamp  speeds  or for off  These choices are described below     off   In the off position  the clamp amplifier is disconnected from       the current electrode  The voltage difference between the    Warner Instruments  A Harvard Apparatus Company    current electrode and the bath electrode  Ve  mV  is read on the METER in the  BATH ELECTRODES control block  This information is also available at the Ve x10       output BNC on the rear of the instrument        slow   The slow clamp mode is useful for screening of oocytes or where high clamp  speeds are not required  The slow clamp speed is approximately 0 5 ms when  measured with the model membrane  1 MQ shunted with 0 47 uF   In this  mode  measured currents are displayed on the BATH ELECTRODES METER in units  of pA        fast   Most oocyte clamping is performed in the fast mode  The clamp speed is  limited by the resistance of the current electrode and the oocyte membrane  capacitance  Therefore  the current electrode resistance must be kept as low as  possible to obtain the fastest clamp speeds  Currents are rea
19. ed in a controlled laboratory environment    Voltage recording channel  Vm   Input Impedance  Output Resistance  Vm OFFSET  Noise   Electrode Test  Negative Capacity  Vm Meter Range  full scale  Bath electrode channel  Im   Ve OFFSET  Noise   Im Clamp  clamp on   Open loop  clamp off     MONITOR    MONITOR FILTERED  4 pole Bessel    Gain Telegraph  Meter Ranges  full scale   Ve  clamp off   Im  x0 1 range   Im  x1 O range   Im  x10 range   Current electrode channel  V    Compliance Voltage  Alarm  Gain  Variable  DC  Electrode Test  Commands  Hold  internal   External input  attenuated by 10   Maximum external input  Power requirements  100  130 or 220 240 VAC  50 60 Hz  Dimensions  Enclosure  Voltage Headstage  Mounting Handle  Bath Headstage      All noise measurements made with an 8 pole Bessel filter     5 x10   Q  shunted by 3 pF  1000     200 mV at V probe input  50 uV RMS at 1 kHz   10 mV MQ   0   45 pF     199 9 mV      200 mV    5 5 nA RMS at 1 kHz  x7 range    28 pA RMS at 1 kHz  x7 range    0 01   100 V uA in 3 ranges  7 steps per range  Same as above  filtered at 1 kHz   0 2   2 6 VDC in 0 2 V steps      199 9 mV    199 9 pA    19 99 pA    1 999 pA      180 V    160 V    0   2000 AC DC  1x10   DC  switch selected  10 mV MQ      198 mV in 2 ranges  1 V in   0 1 V command  10 V    9x42x25 cm  HxW x D    1 25 x 5 cm  dia  x length   with 1 8 m cable   4 8 mm x 6 3 cm  dia  x length    2 8 x 3 5 x 4 2 cm  H x W x D   with 1 8 m cable    Warner Instruments  A Harvard A
20. ed when the membrane resistance between the current electrode  and the bath virtual ground goes to zero   This will damage the oocyte  For this reason       we recommend that the user enable the audible overload alarm to provide a warning  when the potential for such damage exists     SPECIAL CIRCUMSTANCES    High side current measuring    In studies of single oocytes  current is monitored by the bath clamp headstage   Experiments involving two oocytes in a common bath  such as gap junction studies   requires monitoring currents from each oocyte  This is done in the current output leg  in  series with and ahead of the current electrode      Two disadvantages of monitoring the current in this manner exist     a  The noise level of this signal is higher  However  this is usually not a serious  problem since currents are typically in the pA range    b  The voltage drop across the solution resistance  from oocyte to bath ground   becomes an error voltage since it is not subtracted out as when the bath clamp  headstage is used  This problem is overcome by using the optional DIFFERENTIAL  VOLTAGE HEADSTAGE     Configuration    The OC 725C current measuring circuit can be changed to the high side current       measuring mode by setting a dip switch on the main circuit board   1  First disconnect the power cord from the wall   2  Remove the two screws at the rear of the top cover and it off     3  Locate dip switch S10 on the circuit board  For normal operation S4 is on and all other
21. ee eese een setae tna tn aetas etos e esee esee se sete sete esse sas e tas ta setas sena 10  Initial  instrument ECOUTER E                         P                11  Test DEOCeGUTFes       eiecit A O NN 11  Offset Controls    tussi RR ueniet uti iei i RE 11  Voltage electrode est    nere lic aa RE 12    jn RR 12  Current  electrode testa rte En vei t eres Wai E aN evan Pe RU PHAR Te Ces 12   DC UE RT                                           13   AC clamp iesti  nicae AA RE E epe T e PR po e te ee 13   S1 DA B NJ             M                           HE 15  Pipettes pM                                       H    15  luongnocH                                     15  Bath probe anran                                              EAA 16  Electrode placement and grounding          sssesocecoesooessoecooecsocesocesocesocesoccssecsoesssessoecesecssecssecesessocesocssooses 16  Bath clamp electrode placement                 eese esee nan nn anno na anna na entente anna nc entes 16    Single oocyte setup with indirect ground                  esee eese eren ennt enne enne 17    OC 725C Manual  Rev  050816 2    Single oocyte setup with direct ground             esses eene eene tenente eene nne ener ens 17  D  al ooctyesel Up  iic eietbc iter ts eek bet Haee eoe EENE EL FR Ee ERR ERE HA Re EO Lea be HARI EAEE 17  Cable Ie                                            18  Resting position of the controls                        4  ccce ee ee eee eese seen esta sette s eb aes ea eee pese co
22. erated  as long as the button is depressed     Current electrode test    This test is performed using controls in the CURRENT ELECTRODE section        Warner instruments  A Harvard Apparatus Company    With the model cell in place  depress the Ve ELECRODE TEST pushbutton and observe a  reading of 20 mV on the CURRENT ELECTRODE METER  This corresponds to a 2 MQ reading  at a calibrated response of 10 mV MQ        With the CLAMP MODE switch set to off  the BATH ELECTRODE METER monitors Ve  voltage at the  current electrode   In this case  Ve will be a measure of the voltage across Rm and the  meter will indicate 10 mV  1 MQ         DC clamp test    This test is performed using controls in the CLAMP ELECTRODE section        Place the CLAMP MODE switch in the fast position and adjust the Vm OFFSET control for a  reading of 100 mV on the VOLTAGE ELECTRODE METER        Turn the GAIN control on and slowly turn the control clockwise until the meter reading  Vm   decreases to zero  The CURRENT ELECTRODE METER should read  0 10 uA        Set the HOLD POTENTIAL  COMMANDS section  to 100 mV  50 mV on thumbwheel and MULTIPLIER  toggle at x2   Select positive  pos  polarity  The VOLTAGE ELECTRODE METER should read  100 mV and the CURRENT ELECTRODE METER should read 0 00 uA           Switch to negative  neg  polarity  The VOLTAGE ELECTRODE METER should read 100 mV and the  CURRENT ELECTRODE METER should read 0 2 uA          Return the GAIN control fully CCW and turn the CLAMP MODE switch to
23. ethod  care must be taken to prevent the foil from touching the  surface of the bath solution at the bottom end of the pipette or the silver electrode wire at  the top end     Electrode holders    Voltage Electrode   The voltage recording electrode holder uses a silver wire for the  electrical coupling between the pipette and holder connector  Any silver wire contacting the  KCI solution in the pipette must be chlorided to reduce junction potentials  see Chloriding    Warner Instruments  A Harvard Apparatus Company    Procedure in Appendix   The pipette should contain just enough KCl so that approximately  1 2 inch of the chlorided wire is submerged  The pipette holder assembly is attached  directly to the voltage headstage prior to mounting in a micropositioner     Current Electrode   The current recording electrode also uses a silver wire for coupling   In an manner analogous to the voltage electrode  the current electrode wire must be  chlorided prior to assembly and use  The pipette holder assembly is mounted in a  micropositioner with the mounting rod supplied     Bath probe    The bath clamp is designed to maintain a virtual ground in the oocyte perfusate  The  bath probe should be positioned so that the silver electrode wires can be inserted into the  recording chamber or into the agar bridge wells  Sticky wax or tape is usually sufficient to  secure the unit when positioned on a flat surface or  alternatively  the unit can be held in  place on a separate stand  The ba
24. f an internal switch  permitting measurement of the current in  series with the current electrode instead  of in the bath  Additionally  an optional  differential voltage headstage is available  which subtracts the voltage drop across  the series resistance in the bath     Voltage Headstage Probe  The voltage  measuring headstage is a single ended   high impedance probe  Its small size   convenient mounting rod  and two meter  cable make for easy attachment to a  micropositioner  Warner electrode holders  having a 2 mm jack mount directly onto  the headstage     Voltage and Current Meters  Independent  meters provide simultaneous displays of  membrane voltage Vm and membrane  current Im  To assure proper impalement  of the current electrode  the current meter  displays membrane potential  Ve  from the  current electrode before the clamp circuit  is activated     Overload Alarm serves as a reminder  when the feedback amplifier reaches its  maximum output voltage  a condition  which could result in damage to the  oocyte     DC Offsets for voltage and current  electrodes     Warner Instruments  A Harvard Apparatus Company    NOMENCLATURE    Text conventions    This manual refers to amplifier controls at four functional levels  operational sections   control blocks  specific controls within a block  and settings of specific controls  To minimize  the potential for confusion  we have employed several text conventions which are specified  below  Since our goal is to provide clarity
25. g the gain select    A wider range of bath current  Im  measurements is now possible with the addition of  the GAIN SELECT toggle switch located above the GAIN control  The switch has 3 positions   x0 1  x1 and x10  Resistance values shown below each LED indicate the bath clamp  feedback resistor used for the current measurement     The chart on the next page shows the effect of the range selection on the Im output   Note also that range selection changes the sensitivity of the current meter  Typically   measured currents will fall in the x7 range  The lower and higher ranges are intended to  cover those applications where currents are beyond the x  range  Currents below 1 uA  should be monitored in the x10 range  Large currents above 100 pA require the x0 1 range   Since there is overlap in the ranges  the current being measured may be monitored in one of  two ranges in which case the choice may be made on the basis of noise or clamp speed     Warner Instruments  A Harvard Apparatus Company                         Gain select Headstage Im output range Im max output   Maximum meter  resistor  V uA   uA  reading  x0 1 10 kQ 0 01   1 0 10   1000 199 9 uA  x1 0 100 ko 0 1   10 1 100 19 99 uA  x10 1 MQ 1 0   100 0 1  10 1 999 pA                Other gain range selection considerations  The intrinsic noise of the current measuring circuit is a function of the bath clamp    feedback resistor with a larger resistor offering lower noise and greater signal resolution   For comparisons
26. in Figure A  This method  uses the bath clamp headstage to establish the bath ground and is preferred for two  reasons   1  current readings with the bath clamp will have the lowest noise level  and  2   properly placed bath clamp electrodes will negate the need for series resistance    compensation                              CURRENT ELECTRODE 7 CURRENT ELECTRODE 4    HOLDER AND CABLE   e    HOLDERANDCABLE     Vm HEADSTAGE L vor Vm HEADSTAGE  Ya SHIELD              CURRENT ELECTRODE J    CURRENT ELECTRODE    CLAMP CURRENT PATH CLAMP CURRENT PATH    E BATH RESISTANCE    BATH REFERENCE ELECTRODE           BATH    Figure A CLAMP  HEADSTAGE        Figure B    Single oocyte setup with direct ground   Applications where use of the bath clamp is not suitable  such as those with a very long  solution path to ground  can be configured using the alternate method of directly grounding  the bath as shown in Figure B  In this configuration  current is read from the  high side  of  the current output leg     This method also requires the use of the optional DIFFERENTIAL VOLTAGE HEADSTAGE  Two  disadvantages are   1  The noise levels of the current signal measured in the  high side  is  approximately double that obtained with a bath clamp  and  2  high levels of clamp current  could produce a substantial voltage drop across the solution series resistance     Dual ooctye set up   Dual oocyte setup is accomplished using two clamps as illustrated in Figure C  Both  clamps must be configured 
27. ly 5 turns                    Test procedures    In the following testing procedures  allow a tolerance of  1  on the readings taken  For  example  if the test response is indicated as 100 mV  a reading from 99 0 to 101 0 mV would  be within tolerance     Offset controls    Vm OFFSET  VOLTAGE ELECTRODE section   The full range of this control is  200 mV  This can be  verified by rotating the control first fully clockwise and then fully counterclockwise  while observing the VOLTAGE ELECTRODE METER  The displayed readings will indicate off          scale at the extremes of the control   s manipulation since the meter is only capable of  displaying  199 9 mV     Vm x10 output BNC  VOLTAGE ELECTRODE section   This output can be monitored using an  oscilloscope  The reported voltage will swing between  2 V as the Vm OFFSET control is       Warner Instruments  A Harvard Apparatus Company    manipulated throughout its full range  Set the Vm OFFSET to 0 0 reading on the meter  and verify that the Vm x10 reading on the scope also reads 0 V     Ve OFFSET control  CURRENT ELECTRODE section   The Ve OFFSET control is tested in the same  manner as the Vm OFFSET control  This control is adjustable when the CLAMP MODE switch       is set to off  The voltage of the Ve OFFSET is read from the BATH ELECTRODE METER when  the CLAMP MODE switch is set to off        Ve x10 output BNC  This BNC is located on the rear panel of the instrument and reports  the setting of the Ve OFFSET control when the
28. mers     Harvard Apparatus is committed to being a good corporate citizen  As  part of that commitment  we strive to maintain an environmentally conscious  manufacturing operation  The European Union  EU  has enacted two  Directives  the first on product recycling  Waste Electrical and Electronic  Equipment  WEEE  and the second limiting the use of certain substances   Restriction on the use of Hazardous Substances  RoHS   Over time  these  Directives will be implemented in the national laws of each EU Member State     Once the final national regulations have been put into place  recycling will  be offered for those Harvard Apparatus products which are within the  scope of the WEEE Directive  Products falling under the scope of the  WEEE Directive available for sale after August 13  2005 will be identified  with a    wheelie bin    symbol     Two Categories of products covered by the WEEE Directive are currently  exempt from the RoHS Directive   Category 8  medical devices  with the  exception of implanted or infected products  and Category 9  monitoring    and control instruments  Most of Harvard Apparatus    products fall into  either Category 8 or 9 and are currently exempt from the RoHS Directive   Harvard Apparatus will continue to monitor the application of the RoHS  Directive to its products and will comply with any changes as they apply    Special Collection Disposal Required       Do Not Dispose Product with  Municipal Waste        Warner Instruments  A Harvard Appar
29. n  Finally  hydraulically driven micro   manipulators will also reduce membrane damage while the electrodes are in the cell     Repeated recordings    Most recording sessions will involve repeating the above steps several times with many  different cells  Unless there is a concern about contamination of the bath solution by  something carried over from previous experiments  the pipettes can also be used repeatedly   They should be free of debris and should have approximately the same resistance as they  had in the previous recording  A significantly higher resistance could indicate that the  pipette is partially plugged with cellular debris     Make the following control settings before the next recording is carried out                 Control Section Setting  POLARITY COMMAND off  MODE SELECT CLAMP off  GAIN CLAMP 0   fully CCW                    Electrophysiology    If you are well versed in setting up electrophysiological equipment  you can safely skip  over the rest of this section  If  however  this is your initiation into electrophysiology  as it  may well be for some of you molecular biologists   then you may find the following  recommendations helpful     While the whole cell electrophysiological configuration is more forgiving than the patch  clamp  it is still important to minimize mechanical motion  The platform for your  experimental setup  therefore  should be mechanically well isolated  This will reduce  leakage around the electrodes  making the clamp more effec
30. nclosure and connected to earth through the power line  cord  A shorting link allows for interconnection of the two grounds  In most  experimental setups  separating the grounds will result in minimizing 50 60 Hz signal  interference from ground loops  However  trial and error will determine the best  results     Warner Instruments  A Harvard Apparatus Company    Additional components    Voltage recording headstages  7250V PROBE  Standard Version     The voltage probe is an active headstage housed  ina 1 25 x 5 cm cylinder  dia  x length   The  probe body is nickel plated and epoxy sealed for       corrosion resistance     Warner microelectrode holders with 2 mm jacks mate directly to the input pin on the  probe body  A mounting block and handle are supplied and facilitate attachment of  the probe to a micromanipulator  The handle can be mounted either axially or  perpendicular to the probe body     7255DI DIFFERENTIAL PROBE  Optional     This voltage probe is designed for applications where two oocytes share a common  bath or where the voltage drop across the solution resistance is to be measured and  subtracted from Vm  The headstage housing is approximately 2 cm longer than that  of the 7250V PROBE and has two additional inputs  CIRCUIT GROUND and V  DIFFERENTIAL  When the two inputs are shorted  the probe functions exactly the same  as the standard single ended 7250 PROBE     Bath headstage  The BATH PROBE is housed in a 2 8 x 3 5 x 4 2 cm aluminum  enclosure  Inputs are 
31. on   If the  buzz technique fails to cause electrode penetration  further advance the voltage  electrode until it  pops  through the membrane     The potential across the membrane will now be displayed on the VOLTAGE ELECTRODE  METER        Now  advance the current electrode until its tip is slightly depressing the plasma       membrane of the cell and depress the Ve BUZZ pushbutton  Similar to the voltage  electrode BUZZ  the current electrode BUZZ produces a 1 V  10 kHz oscillation across the  current electrode  This disrupts the cell membrane and causes the tip of the electrode  to impale the cell with no further movement of the micro manipulator  Once again  if    Warner Instruments  A Harvard Apparatus Company    the BUZZ technique fails to cause penetration  further advance the current electrode  until it  pops  through the membrane     Clamping the cell    14  Activate the clamp by switching the CLAMP MODE switch to either the slow or fast mode     15  Adjust CLAMP GAIN control clockwise as far as possible without illuminating the OVER  VOLTAGE LED  located in the CURRENT ELECTRODE section        tm    16  The clamped membrane potential can now be observed over time or it can be       manipulated by selecting the desired polarity and amplitude with the controls located  in the COMMANDS section  Alternatively  you can control the COMMAND voltage externally  from a computer by leaving the POLARITY toggle switch in the off position and connecting  the appropriate analog o
32. potential can be read  from Ve x10 output BNC or on the current meter  BATH  ELECTRODES section      ELECTRODE TEST   A voltage proportional to the resistance of  the current electrode  10 mV MQ  will be displayed on  the meter by depressing the ELECTRODE TEST push  button when the CLAMP MODE SELECTOR switch is in the off position        Ve BUZZ   The Vm BUZZ pushbutton facilitates penetration of the voltage electrode by  producing a 10 kHz square wave at the pipette tip     OVER VOLTAGE LED   If the voltage at the current electrode exceeds  160 V  the OVER  VOLTAGE lamp will light  An alarm will also sound when the rear panel ALARM  switch is in the on position     Rear panel    The line power connector and fuse are located on the rear panel  Operating voltage is  specified on the MODEL SERIAL NUMBER sticker applied to the rear of the instrument  The rear  panel also contains Ve x10 and ca TELEGRAPH output BNCs  the ALARM switch and instrument  GROUNDS     Ow    The Ve x10 output BNC monitors the voltage of the current electrode  x10  when the CLAMP  MODE SELECTOR switch is off    The GAIN TELEGRAPH output BNC provides a DC voltage indicating the gain setting of the  instrument  The output varies from 0 2 to 2 6 volts in 200 mV steps as shown in the  appendix     ALARM switch   Activates or deactivates the over VOLTAGE current electrode audible alarm     GROUNDS   Both CIRCUIT and CHASSIS grounding posts are located on the rear panel  CHASSIS is  common with the instrument e
33. pparatus Company    Gain telegraph outputs       Im Output  V uA     Gain Telegraph                         0 01 0 2V   0 02 0 4 V   0 05 0 6 V   0 1 0 8 V   0 2 1 0V   0 5 1 2V   1 1 4V   2 1 6V   5 1 8V   10 2 0 V   20 2 2V   50 24V   100 2 6 V   Gain select settings  Gain select Headstage Im output range Im max output Maximum meter   resistor  V uA   uA  reading  x0 1 10kQ 0 01   1 0 10   1000 199 9 uA  x1 0 100 kQ 0 1   10 1 100 19 99 uA  x10 1MQ 1 0   100 0 1  10 1 999 uA                      Noise from bath clamp feedback resistor       Noise                   Feedback Resistor Shorted Input Standard Model Cell Modified Model Cell   0 5 pF   0 22 uF   10 kQ 75 pA 6 0 nA 44 nA  100 KQ 28 pA 5 5 nA 44 nA  1MO 22 pA 5 0nA 4 0 nA             Warner Instruments  A Harvard Apparatus Company       References   Colman  A   1984   Translation of eukaryotic messenger RNA in Xenopus oocytes   Transcription and Translation  eds  B D  Hames and S J  Higgins  IRL Press  Oxford    Ch  10    Hille  B   1984   Ionic Channels of Excitable Membranes  Sinauer  Sunderland  MA   Ch  2     Zhou  J   Potts  J F   Trimmer  J S   Agnew  W S  and Sigworth  F J   1991   Multiple gating  modes of the ul sodium channel  Neuron 7  775 785    Warner Instruments  A Harvard Apparatus Company    Certifications    Declaration of Conformity  CE MARKING  EMC     Application of Council Directive  89 336 EEC    Standards To Which Conformity EN55022 Class A  Is Declared  EN61000 3 2  EN61000 3 3  EN50082 1 199
34. sets eR a seen 00000000 ease etas 19  Using the DII              V                   iSo csatos 19  Other gain range selection considerations                          eee esee eere e eee ee ee enne seta sette seta aset ease tates etas esten aee 20   A PROCEDURE FOR RECORDING FROM OOCYTES                eerie sees eese teet ee eo setae ttes setas e testae en aes 21  Initial electrode placement          0  2000 000022000000000000000200082000000000000000000050100000000000000800008000000000 000000000 21  Voltage electrode placement                 2  aaa Era casae va Te sa Va Te sa soe Eua b eR Te a eo There ee 21  Current electrode placement           0 02000s000ss00000000000000s200000200000000000000s200000000000000000008000080000000000 0000000000000 22  Impaling the cell e          M        22   ei noPAltacjee E                                  eesse 23  Clamping high conductance cells                       4   eee eee eee eee e eene eene eene etate tense ta seta cono conocconaconaconncconeconos 23  Uniclamiping  the Cell nC 23  O a AAA ts tris co sesvecnooi esov to cece ti vess sri nos svesno irosit Seske  23  SPECIAL CIRCUMSTANCES asisisscsdsccseccavsssccccscsseccsascosceecconesctestssetsecesssescsesbeustecsoscveccoseteecesvectsadsseteceses 24  High side current measuring    eue cose toco oae to eroe nino east en tese nt Vosa voca Uoc ee real osa eo oerip rares e 24  ConfigurattQms z odii e ede Hle Rago AA etd t oe Ra oe 24  Optional voltage headstage         corssonssonssons
35. ssnsesnssnnnennussnnssnnssnnnsnnssnnnsnnnssonssonsssnnssnsesnnssnunsnnssnnssnnssnnssnnne 24  COMMENTS AND RECOMMENDATIONS                 eee eee esee eene en ette sone en otto ste ense tust te tee ease tastes eese taste soa 26  MENTES O 26  Repeated recordings sesssisccccsssziscensccsosseessescensccseseesessecenscessscoocedvesscessesosgsctenssessoonascedsescedesenbssestersnessens 26  El  ctr  ph  siology        26  APPENDIX iiss E NO 28  Specifications   Equipment is intended to be operated in a controlled laboratory environment             28  Gain telegraph Outputs    eereo eee ie tno ron voro so tos ono essnee orase eae on eV unen tea nsee ee eE eue a apu ece apte eher 29  Gain select Settlligs      esee roe NN 29  Noise from bath clamp feedback resistor                              1  eee ee ee ee eee eere eee ee eee tosta esten aset tn sette setas seen 29  References poo                30  Certifications    EC                31    SE TERRIER 31    Unique Features    Additional Features    The model OC 725C Oocyte clamp is designed for two electrode  whole cell voltage clamping  of Xenopus oocytes  as well as for other large cells and cell structures such as squid axons   The instrument has several features making it ideal for these purposes     High Voltage Compliance  The OC 725C  combines high AC and DC gains and a  voltage compliance of  180 V to insure  fast  nonsaturating clamp performance  under nearly any condition  The AC clamp  gain is variable up to 2000  An a
36. th probe electrodes should also be chlorided before use  as  described above     Electrode placement and grounding    Three drawings  shown on pages 19 and 20  have been included to illustrate the  various ways a bath circuit can be configured  Most applications involve only a single oocyte  and Figures A and B illustrate these setups  Figure C shows a setup for recording from 2    oocytes in a common bath with the use of dual clamps       Bath clamp electrode placement   Proper placement of the bath electrodes  Iout and Isense  is important for obtaining optimum  performance  The Isense electrode  or the agar bridge associated with it  should be placed as  close to the oocyte as possible since this point is the virtual ground node  and on the same  side as the voltage recording electrode  The Iout electrode  or the agar bridge associated with  it   on the other hand  can be placed at a greater distance from the oocyte and should be on  the same side as the current electrode    It is recommended that the user not directly expose the electrode wires to the perfusate if  the recording session is to last for more than a few minutes  Instead  agar bridges should be  employed to provide a circuit between these electrodes and the bath  This protects the cell  membrane from the potential adverse effects of the silver wire     Warner instruments  A Harvard Apparatus Company    Single oocyte setup with indirect ground   Single oocyte studies are best accomplished with the setup shown 
37. tive and reducing noise in your  recordings  The latter is especially important when recording responses of certain ligand   gated channels where membrane potential changes may only be a few mV     In addition to mechanical isolation  the setup must also be isolated from external  electrical noise sources  These include motors  lamps  and wiring  The platform should be    Warner Instruments  A Harvard Apparatus Company    shielded from these sources of electrostatic radiation with a Faraday cage  All equipment  within the Faraday cage should be grounded to the rear panel instrument circuit ground  and is best achieved by connecting everything  including the cage  to a ground bus within  the cage  Then only one wire is run from the setup to the instrument ground     You will need to mount the voltage recording electrode headstage and the current  injecting electrode on micro manipulators  They need not be hydraulically driven but such  drives will minimize the damage to the cell during and after penetration  and will make for  better seals around the electrodes  Another advantage gained by reducing membrane  damage by electrodes is to enhance the possibility of making subsequent recordings from  the same cell  These suggestions are also important for minimizing mechanical noise in the  recorded data    You will need a dissecting scope for viewing the placement of the electrodes  Anything  more powerful than 40x will just get in the way  The light source for your scope should b
38. to read current from the  high side  as described in the section  titled HIGH SIDE CURRENT MEASURING  see page 26  and each clamp must be equipped with  the optional DIFFERENTIAL VOLTAGE HEADSTAGE     Warner Instruments  A Harvard Apparatus Company    pe CLAMP 1 poe e  A CLAMP 2 A       CURRENT ELECTRODE  HOLDER AND CABLE    CURRENT ELECTRODE  HOLDER AND CABLE                  Vm HEADSTAGE          CURRENT    Vm ELECTRODE 1 N j  q ELECTRODE 1 ELECTRODE 2    CLAMP CURRENT PATH    BATH RESISTANCE NI A    NN V DIFF ELECTRODE 2    A a         V DIFF ELECTRODE 1    BATH REFERENCE  ELECTRODE Figure C    Cable connections    Bath clamp headstage  After positioning the probe as described above  connect it to the  BATH PROBE socket  BATH ELECTRODES section         Voltage electrode headstage  The high impedance probe for recording membrane potential  should be mounted on a micro manipulator and connected to the VOLTAGE PROBE socket   VOLTAGE ELECTRODE section         High voltage current electrode  The holder should be mounted on a micro manipulator and  the cable connected to the I ELECTRODE socket  CURRENT ELECTRODE section         Command potential  If a computer or external generator is used for controlling the clamp  command potential  its signal should be connected to the front panel COMMAND IN  10  input  COMMANDS section      External monitoring  To monitor the microelectrodes potentials on an oscilloscope   computer or a chart recorder  the following connections should be mad
39. tors for GAIN TELEGRAPH OUTPUT and V  x10 output  an ALARM on off  switch  and binding posts for CIRCUIT and CHASSIS GROUND                 Voltage electrode       The VOLTAGE ELECTRODE control block contains VOLTAGE ELECTRODE  Vm         Vm  mv     the VOLTAGE PROBE CONNECTOR  the VOLTAGE HO  METER  the Vm OFFSET control  the ELECTRODE E G x  TEST and BUZZ pushbuttons  and the Vm OUTPUT E  c  BNC  Controls for NEGATIVE CAPACITY p       vm  COMPENSATION are also located in this block     rose VMOFFSET gc  Vm X10  The VOLTAGE METER reports the membrane AS    e  voltage  Vm  with a full scale range of    gt       199 9 mV        The VOLTAGE PROBE CONNECTOR is a 7 pin DIN connector for attachment of the voltage       probe to the instrument     The Vm OFFSET control is a 10 turn potentiometer providing up to  200 mV at the  VOLTAGE PROBE input for offset of membrane junction potentials     The Vm ELECTRODE TEST control is used to determine the internal resistance of the  voltage probe  When the pushbutton is depressed  a constant 10 nA current is  passed through the voltage electrode producing a voltage drop of 10 mV MQ of  probe resistance  The measured potential is displayed by the meter or reported at  the Vm x10 output BNC  at 100 mV MQ      The Vm BUZZ push button facilitates penetration of the voltage electrode by producing a  10 kHz square wave at the pipette tip     The Vm x10 ourPUT BNC reports the membrane voltage in mV multiplied by 10     Warner Instruments  A Harvar
40. two 1 mm pin jacks labeled I SENSE and  I our  The case is electrically grounded and a pin jack is  located on the side for connecting to shields     A plastic plate with two screw mounting slots is attached to       the probe base  The BATH PROBE connects to the control unit  with a 6 pin connector    Current electrode cable  A two meter shielded cable is supplied with a 2 mm  pin jack on one end to mate with the an electrode  holder  and a 3 pin connector on the other end to  mate with the instrument  The electrode holder       should have a handle for mounting to a  micromanipulator  An example is shown to the right     Warner Instruments  A Harvard Apparatus Company    Model cell    The model cell supplied with the OC 725C is useful as a training aid and as a  calibration and test device  It has connections for the voltage and current probes and to  the bath clamp allowing all aspects of the amplifier   s function to be tested     Comments    Connecting to line power  The model OC 725C is supplied with a 3 conductor power cord  One conductor provides  a connection between the instrument housing and the earth ground  Safe operation of  the instrument will be assured provided that the ground circuit in the power outlet is  wired correctly and is connected to earth     High voltage outputs    When handling the current electrode cable  be sure to set the GAIN CONTROL fully  clockwise and the CLAMP MODE switch to off     USING THE MODEL MEMBRANE    The model cell supplied with
41. utput from your computer DAC to the COMMAND IN  10 BNC  connector   See Cable Connections  page 20      17  Depending on the amplitude of the response you wish to record  you may adjust the  instrument GAIN to a higher or lower position  The CURRENT ELECTRODE METER should now  be displaying the current  in uA  that is delivered to hold the cell at the designated       command potential     Clamping high conductance cells   18  DC GAIN mode  CLAMP section  may be required to clamp high conductance  low  resistance  cells  This condition will be evidenced by the inability of the instrument to  maintain a DC holding potential  to within 1  or better of the set value  and the  maximum instrument gain is not sufficient to provide a hard clamp  DC GAIN mode  provides an additional DC gain greater than 10  while the AC gain remains at 2000  maximum for stability     Unclamping the cell    19  To unclamp the cell  turn the GAIN control  CLAMP section  fully counter clockwise to the  detent off position  This will also disengage the DC GAIN     O    Removing the electrodes    20  It is very important that the CLAMP GAIN be returned to the off position  fully counter   clockwise to click off  as described above and the CLAMP MODE selector switch be placed  in the off position before removing the current electrode from the cell     Warner Instruments  A Harvard Apparatus Company    Failure to perform the above steps will overload the feedback amplifier  due to the  large current generat
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Manuel d`installation et d`utilisation  ~~ 厚生労働省医薬食品局安全対策課長 (公 印 省 略)  Acer Aspire Ethos AS5942G-334G50Mn  Archos 35 Home Connect user manual  Certificat de garantie limitée  ASUS PU551JH User's Manual  Mr. Heater MH12 User's Manual  información pdf  COMMANDE DE MATERIEL - mcl      Copyright © All rights reserved. 
   Failed to retrieve file