2. Pre-installation requirements
Contents
1. 1 3 Warnings and precautions During installation When installing the unit always observe local and national safety regulation concerning the installation and use of dental x ray equipment Failure to install the unit in an approved loca tion according to these installation instructions may cause the device to be dangerous to both patient and operator The aperture plate in the collimator is made of lead Pb which is a toxic material Do not touch it with your bare hands Be careful when working on the unit when the covers have been removed Some mechanical parts have sharp edges that can cause injuries if mishandled Some internal components become hot during operation Take suitable precautions to avoid burning yourself Some electric components are high voltage and can be lethal if touched Whenever you are making adjustments to the unit disconnect it from the main power supply before making the adjustment If the unit is to be used with a 3rd party im aging application software not produced or supplied by Gendex Imaging Systems the 3rd party imaging software must comply with all applicable local laws and regulations on patient information software This includes for example the Medical Device Directive 93 42 EEC and or the FDA During unit operation Read and familiarize yourself with the warn ings and precautions in the unit User s Manual When checking the alignment of the radiation beam protect
2. 3 Installing the unit 3 2 Attaching the unit to the wall 1 CAUTION HEAVY OBJECT A minimum of two persons are required for the following task Transport the unit to the place where it is to be installed NOTE The transportation wheels optional are at tached to the base of the column with two M10 screws The transportation wheels can be used to wheel the unit to the place where it is to be used Mark the position of the two fixing holes on the wall according to the dimensions given in the section 2 6 Fixing hole dimensions Alternatively lift the unit into the position where it is to be used and then using the wall bar as a template mark the positions where the holes must be drilled Hold a spirit level on the side of the column and then on the front to help to position the unit so that it is as vertical as pos sible Installation and set up manual 209480 17 3 Installing the unit GXDP 300 3 Drill two holes with diameters suitable for the 3 8 10 mm fixing hardware being used and then attach the unit to the wall with the appro priate fixing hardware Do not fully tighten the SCrews yet 4 Remove the packaging material that protects the rotating unit and vertical carriage and remove the cover from the upper shelf 4 x M4 screws Place the spirit level widthwise on the upper shelf and level the unit in the left right direc tion Level the unit by moving the base of the col umn t
3. 1 Position the chin rest to the O position 2 Place the laser alignment tool supplied in the chin rest 4 Check that the lasers hit the corresponding grooves on the laser alignment tool Frankfort laser Cuspid laser If any of the lasers are not positioned correctly their positions must be adjusted oee 5 3 Aligning the lasers Installation and set up manual 209480 43 4 Checking the alignment GXDP 300 44 4 4 Completing the Installation Check that the unit functions correctly Take a test exposure to verify that the unit operates correctly and the image is transferred successfully to the PC Protect yourself from radiation when taking the exposure Check the up down keys drive the unit up and down Check that the light key switches the position ing lights on and off Check that when the emergency switch is pressed the unit stops operating Finally replace all the covers that were removed Installation and set up manual 209480 GXDP 300 5 Aligning the unit 5 Aligning the unit 5 1 Aligning the CCD Sensor If the sensor was not correctly aligned after car rying out the alignment check 4 1 Checking the alignment of the CCD Sensor the sensor must be repositioned and new beam images taken to confirm that it is correctly positioned 1 Remove the inner sensor cover from the rotat ing unit It is held in place with four 4 screws 2 To adjust the CCD sensor loosen the four nuts
4. A that hold the CCD sensor in place and then adjust its position with the two screws B on the left hand side Installation and set up manual 209480 45 GXDP 300 5 Aligning the unit Adjust and recheck the position of the CCD sensor until the beam is vertical and centered in the middle of the CCD aperture o Tum CCD f f Tum CCD i counter clockwise l clockwise Move Move cco CCD to right ta left i l I I gt 3 Touch the Arrow button to exit the Still Mode and return to the main menu Installation and set up manual 209480 46 GXDP 300 5 Aligning the unit 5 2 Adjusting the image geometry IMPORTANT NOTE In some cases you may have to adjust the position of both the rotating unit AND the chin rest Center ball not round If the center ball is NOT ROUND then the position of the rotating unit RU must be must be adjusted 1 Loosen the two nuts that hold the guide pin as sembly in place Turn the adjusting nut counter clockwise to move the rotating unit towards the column and clockwise to move it away from the column A move RU SAL towards column ACA E lt A A tes A 4 un i R M d TYLLLALEBELEESLELET MINA LLP B move RU away from column E Guide pin assembly amp A Move rotating unit towards column B Move rotating unit away from column Insta
5. AC CESSORY has been performed in accordance to the appropriate IEC 601 1 and or IEC 601 1 1 harmonized national standard 2 3 The dental imaging software The dental imaging software for example VixWin that is installed in the PC used with the unit must be MDD compliant 2 4 Space requirements When installing the unit make sure that there is enough space at the front and sides of the unit to allow patients to enter and exit the unit easily Patients in wheelchairs will require more space than standing patients the unit is positioned at least 73 1 85 m away from any non medical electrical or elec tromechanical equipment the unit is positioned so that the operator when protected from radiation can see and hear the patient during an exposure Installation and set up manual 209480 9 2 Pre installation requirements GXDP 300 2 5 Unit dimensions Standard column leg 2340 mm 92 1 in Short column leg 2250 mm 88 6 in Standard column leg 965 1740 mm 38 in 68 5 in Short column leg 875 1650 mm 34 5 65 in 175 215 mm 790 mm 6 9 8 5 in 31 in 980 mm 38 6 in 830 mm 32 7 in 10 Installation and set up manual 209480 GXDP 300 2 Pre installation requirements 2 6 Fixing hole dimensions aq 200 10 mm 7 7 8 3 8 in a 180410 mm 7 1 8 43 8 in 46010 mm 6 1 4 3 8 in i 140 10 mm 5 1 2 3 8 in ig OTO eee Standard column leg 22
6. ray equipment to a power supply Installation and set up manual 209480 2 3 Installing the unit 3 6 Preparing the PC GXDP 300 NOTE The PC and any other external device s con nected to the system must meet the IEC 60950 standard minimum requirements Devices that do not meet the IEC 60950 standard must not be con nected to the system as they may pose a threat to operational safety 1 28 Position the PC to be used with the unit at least 1 85 m 73 in away from the unit NOTE If the unit and PC are to be part of a dental system make sure that all the other system components and devices are installed con nected and configured correctly Refer to the documentation supplied with the other compo nents and devices for information on how to do this Switch the PC on and install GxPicture sup plied with the device then install VixWin or other dental imaging software that will be used with the device For information on how to do these refer to the GxPicture manual and the documentation supplied with the dental imaging software you are installing IMPORTANT NOTE During the installation when the Select Fea tures window appears make sure that you select the GXDP 300 option Installation and set up manual 209480 GXDP 300 3 Installing the unit 3 7 Connecting the unit to the PC The unit supports 10Base T 100Base TX Ethernet connections The PC must be connected to the unit in accor dance with I
7. to the documentation supplied with GxPicture Software Driver for information on how to do this In the Device Connection section click the IP based radio button to select it Then click the Enable changing IP address check box to activate the IP address field GXDP 300Setup m So Device settings Driver Info Device Connection is P based Device IP address Enable changing IP address 1 94 9 227 250 Send to Device 6 PC Key in the new IP address of the unit into the IP address field 7 PC Click the OK button to close the settings window 36 Installation and set up manual 209480 GXDP 300 3 Installing the unit Multi room Connection Multi room Connection allows the unit to be used by several PCs within a network A PC must re serve the unit before it can be used 1 PC In the GXDP 300 Setup window select the Settings page 2 PC Click the Use Multi room Connection check box to select it and then from the drop down list select a unique Workstation iden tifier number between 1 and 8 for the PC being configured Additional workstation information for exam ple user name location etc can be entered into the Friendly Name field Multi Hoom Connection Use Mult Room Connection Workstation identifier Friendly Name Device Autorelease timeout seconds a00 The Device Autorelease timeout is the length of time that the unit will remain reserved and usable b
8. yourself from radiation After the unit has been installed and set up advise the people who will operate the unit to familiarize themselves with the warnings and precautions in the User s Manual Installation and set up manual 209480 GXDP 300 1 Introduction 1 4 Unit description Main parts Column Motorized carriage Rotating unit Exposure warning lights Emergency stop button Press to stop rotate to release On off switch rear of column PC NOT SUPPLIED with MDD approved dental imaging software ZAECGCIBEN Installation and set up manual 209480 3 O OGCIBEN 1 Introduction GXDP 300 Head support Midsagittal laser Mirror Frankfort laser and laser positioning knob Cuspid positioning knob Chin rest Cuspid laser Patient handgrips Installation and set up manual 209480 GXDP 300 1 Introduction Unit controls N O O1 AUN A Side control panel Lasers key switches the patient positioning lasers on and off Up key drives the unit up Down key drives the unit down Return key drives the unit to the patient entry position B Main control panel Program selection keys panoramic bitewing temporomandibular joint Patient size selection keys extra large large medium and small Ready indicator light green unit ready for an exposure red unit is cooling down after an exposure No x rays key the unit can be operated without generating x rays Exposure va
9. 5 15mm 89 5 8 5 8 in Standard column leg 2290 15 mm 90 1 4 5 8 in Short Short column leg column leg 2200 15 mm 2185 15 mm 86 5 8 5 8 in 86 5 8 in Installation and set up manual 209480 11 2 Pre installation requirements GXDP 300 2 7 Fixing hardware and installation and setup tools Fixing hardware 12 The following tools and hardware are required to install and set up the unit NOTE THESE ARE NOT INCLUDED IN THE DELIVERY OF THE UNIT UNLESS OTHERWISE STATED WALL attachment requires 2x 3 8 10 mm diameter sets of fixing hard ware WALL and FLOOR attachment requires 2x 3 8 10 mm diameter sets of fixing hard ware for the wall and 2x 5 16 8 mm diameter sets of fixing hard ware for the floor CAUTION DO NOT use smaller diameter fixing hardware The type and length of hardware to be used will depend on the wall material and floor material to which the unit is to be attached Different wall and floor materials may require dif ferent fixing hardware Make sure that you use the correct fixing hardware for the wall material and the floor material IMPORTANT NOTE Fixing hardware is not required if the unit is to be attached to and used with the show stand Refer to the separate instructions supplied with the show stand Installation and set up manual 209480 GXDP 300 Installation and set up manual 209480 Installation tools 2 Pre insta
10. 8 From the upper shelf unscrew the guide pin and two holding screws from the base plate DO NOT remove the base plate or loosen the nuts that attach it to the unit i d a Rotate the guide pin and insert it into the hole in the base plate Note that the plain section of the guide pin must be downwards and the nut on top Plain section x downwards 20 Installation and set up manual 209480 GXDP 300 3 Installing the unit IMPORTANT NOTE If the guide pin does not fully enter the hole in the base plate the top of bearing is NOT level with the top of the base plate the rotating unit must be moved slightly To do this remove the transportation screws and packing pieces see step 11 and then slide the rotating unit gently backwards or for wards until the guide pin drops into place f ww A PE UC LON B w e gt A m EF E a 10 Secure the guide pin in position with the two holding screws Installation and set up manual 209480 21 3 Installing the unit GXDP 300 11 Remove the four 4 transportation screws painted RED and four 4 packing pieces that hold the rotating unit to the upper shelf Now rotate the rotating unit to the patient en try position Refer to the user manual Packing pieces x 4 screws x 4 IMPORTANT NOTE If the unit ever has to be relocated replace the packing pieces transportation scre
11. EC 60601 1 1 Connection options Standalone With the standalone configuration the unit is con nected to a PC and the unit then reserves the whole network interface card The unit and PC then function as a single standalone system with no access to other networks If itis necessary for the PC to be connected to another network the PC will require two network interface cards Network With the network configuration the unit is con nected to the PC via a network device With this configuration the PC is connected to the unit and to other networks with a single network card The unit is also accessible from other PC s in the same network same subnet For the network configuration you may need a new IP address for the unit from your local system administrator Network device L 00000011 Installation and set up manual 209480 29 3 Installing the unit GXDP 300 Connection method The unit is connected to the PC using an IP based connection method This connection method allows the unit to be used in either a Standalone or a Network Configuration IP based connection uses IP Protocol to estab lish a connection between the unit and the PC It requires the IP addresses of both the unit and the PC to be configured correctly IMPORTANT NOTE The default IP address of the unit is 194 9 227 251 and the subnet mask is fixed to 255 255 255 0 The unit offers two alternatives for confi
12. IP address 34 Installation and set up manual 209480 GXDP 300 3 Installing the unit Manually changing the IP address The IP address of the unit can be changed manu ally NOTE Always set the IP address of the PC before chang ing the IP address of the unit 1 Touch the service button on the main control P cox panel of the unit The configuration screen will appear on the display It shows the unit serial number and the GXDP 300 GO1300001 firmware version numbers Name GO1300001 Also there are two function buttons IP 194 9 227 251 IP configuration and INSTALLATION CMDS Installation com GRAPHICS 1 mands IP 194 9 227 251 Press IP 194 9 227 251 button The current IP address of the unit is shown INSTALLATION CMDS NOTE See section 4 for information about Installa tion Commands 3 The IP configuration modes will appear on the display Remote or Manual Current IP IP 194 9 227 251 Choose IP config mode MANUAL Touch the Manual button REMOTE Installation and set up manual 209480 35 3 Installing the unit GXDP 300 The Manual IP address configuration screen will appear Key in the new IP address When the new IP address has been keyed in the OK button will become active Touch the OK button to accept the change and return to the main display Touch Arrow button to discard the change and return to the main display 5 PC Open the GXDP 300 setup window Refer
13. Installation Manual EXN DP 300 Digital Panoramic X ray Unit REF pP3ootMIl 209480 rev 2 Documentation for Gendex GXDP 300 M and the operating software are copyrighted with all rights reserved Under the copyright laws this documentation may not be copied photocopied reproduced translated or reduced to any electronic medium or machine readable form in whole or part without the prior written permission of Gendex Dental Systems Gendex and Gendex GXDP 300 are registered trademarks of Gendex Language This is the original English language version of the manual 2013 02 This installation manual is for units with serial number S No 601300001 and above Manufactured for Gendex Imaging Systems 1910 North Penn Road Hatfield Pa 19440 Manufacturer PaloDEx Group Oy Nahkelantie 160 FI 04300 Tuusula Finland Tel 358 10 270 2000 Fax 358 9 851 4048 CE 0537 Contents GXDP 300 V Installation and set up manual 209480 GXDP 300 Contents Contents te Ey ge eg MMA 1 TL AOE TINS gg PA Me 1 1 2 Associated documentation eus noni EM A A e 1 1 3 Warnings and TOM e ei ame ei ay en UT 2 PUA ae eg EON 2 DUNO Ek 0 ig NT E 2 TA ADT e ge ERREUR 3 MA OA a E E A E E EA ME 3 ee gg eaaa E E E E 5 1 5Disposalofthetransportationpackaging 6 2 Pre installation requirements eese T a gg MN T 22 MO a a ene ee me e ee 9 2 9 The dental im
14. aging software kame ye idu anl le ale dede li nere li 9 2 Space requirements ea al ml aa ya ll ll m 9 2 9 tI p e A A A MY MM EMME 10 20 FIXNO Ee e pg EE YA YY OT 11 2 Fixing hardware and installation and setup tools 12 FING Burzi iiS NN E P YE 12 mk Ef pe ME EPUM 19 Calibration and setup tools sena puni amm aracda kam ad a UU US NEED 13 3 Installing the unit eeee eese eee 14 3 1 Preparing the unit for installation seeeeeeeeeeeeeee m 14 3 2 Attaching the unit to the wall eeeeeeeeeeeeeeee eee 17 5 Bo m e gg uot DIU DENN Rd MM PO RUNI eee 24 3 4 Installing the remote exposure switch sese 25 3 5 Connecting the unit to the power supply esee 2f SES cce neis da m mo 28 3 7 Connecting the unit to the PC RN 29 e e e eg T T 29 IA 1 gt NY RR YP 29 eg A A 29 One Ol ME O L e am e yea RET 30 Connection using thelPbasedmethod 31 ManuallychangingthelP address 30 Multi room e pe A PN 37 Installation and set up manual 209480 V Contents GXDP 300 4 Checking the alignment and finishing the installation 39 4 1 Checking the alignment of the CCD Sensor eeeeeeeeeeee 39 4 2 Checking the image geometry ball
15. ar of the unit and then secure the cable in position under one of the cable clamps before connecting the cable to the unit This will ensure that the cable cannot be acci dentally disconnected Installation and set up manual 209480 25 3 Installing the unit Exposure switch cable 26 GXDP 300 Attach the exposure switch hook to any flat smooth vertical surface The exposure switch hook has double backed tape Make sure that exposure switch is positioned so that the user can protect them self from radiation and can see the patient during a exposure Connect the exposure switch to the remote exposure switch OPTIONAL Exposures can be taken either with the expo sure switch or the button on the remote expo sure switch Installation and set up manual 209480 GXDP 300 3 Installing the unit 3 5 Connecting the unit to the power supply The unit must be connected to its own dedicated power supply e 100 120 VAC or e 220 240 VAC No other equipment should be connected to the same power supply Extension cables MUST NOT be used to connect the unit to the power supply 1 The power supply must be equipped with an approved type circuit breaker e 100 120 VAC min 16 A 250 VAC e 220 240 VAC min 10 A 250 VAC 2 Connect one end of the power lead to the back of the unit and the other end to the power sup ply Observe any national and local requirements with regards the connection of dental x
16. ccessories GXDP 300 4 Connect cable that comes from the remote ex posure light to connector J 09 on L1300 J709 L200 5 Replace the board mounting plate assembly and secure the cable that goes to the remote expo sure warning light in position under one of the cable clamps 56 Installation and set up manual 209480
17. er PIN screen will appear STILL MODE Key in 1204 and touch the OK key Then select IMAGE CALIBRATION IMAGE CALIBRATION Installation and set up manual 209480 99 5 Aligning the unit o4 GXDP 300 Remove the chin rest and make sure that there is nothing between the sensor and tube head Protect yourself from radiation and take an exposure PC Use the dental imaging software to exam ine the image It should be evenly grey all over with no granularity nor horizontal lines stripes If the image is not evenly gray repeat the calibration until it is If artefacts appear on the image make sure that there is nothing between the sensor and the tubehead Touch the Arrow button to exit the CALIBRA TION MODE Installation and set up manual 209480 GXDP 300 6 Installing accessories 6 Installing accessories 6 1 Remote exposure warning light optional CAUTION The warning light assembly must be positioned outside the room in which the unit is used in the location where it can be clearly seen by the person using the unit 1 Make sure that the unit is switched off and dis connected from the main power supply 2 Remove the board mounting plate assembly from the back of the column 4 screws 3 Route the cable that comes from the remote exposure light through one of the holes in the board mounting plate assembly to the rear of the assembly Installation and set up manual 209480 55 6 Installing a
18. guring its IP address Remote and Manual With the Re mote alternative the unit receives its IP address from the PC With the Manual alternative the user sets the IP address of the unit using the Graphical User Interface GUI of the unit It is recommended that Remote alternative is used when configuring the IP address of the unit How ever if problems occur due to firewalls etc use Manual alternative instead The Remote alternative is explained as part of the section Connection using IP based connection method The Manual alternative is explained in section Manually changing the IP address IMPORTANT NOTE The unit can have any IP address but the unit and PC must always be in the same subnet For example IP for PC 194 9 227 250 and IP for the unit 194 9 227 251 and the subnet mask is fixed to 255 255 255 0 30 Installation and set up manual 209480 GXDP 300 3 Installing the unit Connection using the IP based method 1 Standalone configuration Connect one end of the Ethernet cable to the unit the connector Ss de ad of the column and the other end to Network configuration Connect one end of the Ethernet cable to the unit the connector at the rear of the column and the other end to the network 2 Switch the Unit on 3 PC Start the PC and then open GxPicture Software Driver Open the GXDP 300 setup window Refer to the documentation supplied with GxPicture Software Driver for information on how to do t
19. he chin rest PAN 3 Setup the unit to take an panoramic expo sure at 66 kV Panoramic small NY NY NY Refer to the unit user s manual for information on how to do this 66 kV 4 Setthe chinrest to the zero 0 position Installation and set up manual 209480 41 4 Checking the alignment GXDP 300 Protect yourself from radiation Press and hold down the exposure switch to take an exposure PC Use the measuring tools in the dental imaging software to examine the dimensions of the pins and ball that appear on the image Magnify the image so that you can accurately measure the dimensions If the image geometry is correct the center ball must be round with a difference of no more than 0 2 mm between the horizontal diameter and the vertical diameter Also the distances from the side of the center ball to the inner edges of pins must be equal with a difference of no more than 0 5 mm between the two distances Horizontal diameter mmm oun N Vertical diameter 42 If the center ball is NOT ROUND and or the distances between the pins and the center ball are NOT WITHIN THE TOLERANCE then the rotating unit RU or the chin rest must be adjusted See 5 2 Adjusting the image geometry IMPORTANT NOTE In some cases you may have to adjust the position of both the rotating unit AND the chin rest Installation and set up manual 209480 GXDP 300 4 Checking the alignment 4 3 Checking the positioning lasers
20. his 4 PC In the Device Connection section click the IP based radio button to select it and then click the Enable changing IP address check box to activate the IP address field GXDP 300 Setup m Sa Device settings Driver Info Device Connection im P based Device IP address i Enable changing IP address Eo o Send to Device Installation and set up manual 209480 31 3 Installing the unit IP based 32 GXDP 300 o Stand alone configuration PC Key in the desired IP address of the unit for example 194 9 227 250 into the IP ad dress field Network configuration PC Key in the new IP address obtained from the network administrator into the IP address field Device IP address i Enable changing IP address i 94 9 227 250 Send to Device IMPORTANT NOTE The unit can have any IP address but the unit and PC must always be in the same subnet The subnet mask of the unit is fixed to 299 2909 259 0 NOTE With network configuration the default IP address of the unit must be changed to the IP address obtained from your local IP adminis trator With standalone configuration you can avoid changing the IP address of the unit by setting the IP address of the PC to the same sub net as the default value of the unit After changing the PC IP address in this way you can proceed directly to step 12 to finish the configuration Touch the service button on the main contr
21. ion win dow select INSTALLATION CMDS The PIN number screen will appear This is a security feature to prevent the user from ac cidentally accessing the installation service commands Key in 1204 and touch the OK key Installation and set up manual 209480 39 4 Checking the alignment SMMOPE MODE IMAGE CALIBRATION O CALIBRATION Correct 40 GXDP 300 The service function will appear on the screen Select STILL MODE Protect yourself from radiation Press and hold down the exposure switch to take an exposure PC Use the dental imaging software to exam ine the image that appears If the CCD sensor is correctly positioned the image will show a vertical black line the beam in the middle of a vertical light gray area the CCD aperture Both the beam black line and the CCD aper ture light gray area must be vertical and the bottom edge of the beam black line must be between 0 1 and 2 mm from the bottom edge of the image If the beam black line is not in the middle of the CCD aperture the gray area or the CCD aperture gray area is not vertical the CCD sensor must be adjusted See 5 1 Aligning the CCD sensor Touch the Arrow button to exit the Still Mode and return to the main menu Installation and set up manual 209480 GXDP 300 4 Checking the alignment 4 2 Checking the image geometry ball pin phantom test 1 PC Prepare the unit for an exposure 2 Replace t
22. llation and set up manual 209480 4T 5 Aligning the unit GXDP 300 2 Re tighten the guide pin assembly nuts after the adjustment 3 Take another ball pin exposure to confirm the center ball is round Also verify that the dis tances from center ball to left and right pins are equal if not proceed to the next section Distances between left and right pins not equal Distances between center ball and left and right pins not equal If the center ball is round but the distances be tween the pins and the ball are DIFFERENT then the chin rest must be must be adjusted 1 Loosen the two screws that hold the chin rest in place on the underside of the patient sup port 48 Installation and set up manual 209480 GXDP 300 5 Aligning the unit 2 If the right hand distance in picture C is greater that the left hand distance move the chin rest to the left If the left hand distance in picture D is greater that the right hand distance move the chin rest to the right Move chin rest to left Move chin rest to right 3 Take another ball pin exposure to confirm the unit is now correctly aligned Readjust if re quired 4 After the adjustment has been done re tighten the two screws that hold the chin rest in place Installation and set up manual 209480 49 5 Aligning the unit GXDP 300 5 3 Aligning the lasers Midsagittal and frankfort 1 Loosen the screw that holds the frankfort laser positio
23. llation requirements Electric drill opanners wrenches 7 10 17 two mm AF Hexagon socket wrenches Allen keys 1 5 8 mm Slot blade screwdrivers 0 5 x 3mm 0 6 x 3 5 Torx screwdrivers T20 T25 T30 opirit level Pliers and wire cutters ocissors Knife Transportation wheels Optional not supplied with the unit Calibration and setup tools Ball pin tool Laser alignment tool Line pair phantom digital test tool for countries where this calibration tool is required 13 3 Installing the unit GXDP 300 3 Installing the unit 3 1 Preparing the unit for installation 1 Transport the boxed to the location where the unit is to be installed 2 Remowve the straps that hold the box to the pal let 4 3 Lift off the top of the box and then the sides 14 Installation and set up manual 209480 GXDP 300 3 Installing the unit Remove all the accessories the accessory box and as much packaging material as pos sible NOTE DO NOT remove the packaging material that protects the upper shelf and the rotating unit Fold down the end of the box that is nearest the hinged section part of the column and then slide the unit over the edge of the pallet so that you can access the underside of the hinged section of the column Support the hinged column section before loosening the screws that hold it to prevent it falling and causing damage or injury Loosen the screws the securing nuts are inside the c
24. lues Service key Installation and set up manual 209480 5 1 Introduction GXDP 300 1 5 Disposal of the transportation packaging The unit is transported in a corrugated cardboard box that is attached to a wooden pallet Inside the box preformed expanded polystyrene packaging material protects the unit from shock and damage Packaging material Weight Ib kg Cardboard box and lid 46 3 21 0 Wooden pallet 16 3 7 4 Expanded polystyrene 16 8 7 6 Total weigh 79 4 36 0 When disposing of the transportation packaging materials make sure that you follow all local na tional and international regulations regarding the correct and safe disposal and or recycling of these materials Some transportation packaging materials may be non environmentally friendly These transportation packaging materials must be disposed of in ac cordance with all local national and international regulations regarding the disposal of non environ mentally friendly materials NOTE It is advisable to keep some transportation pack aging so that units can be repacked and easily and safely moved to new locations if required 6 Installation and set up manual 209480 GXDP 300 2 Pre installation reguirements 2 Pre installation reguirements 2 1 The Unit The unit is supplied in one box on a pallet The dimensions of the box with pallet are Length Width Height 76 34 41 191 cm 85 cm 103 cm The assembled unit weighs 265 Ib 120 kg Make
25. ning knob in position and then remove the knob 2 Remove the corner cover from the unit It is held in position with magnets 50 Installation and set up manual 209480 GXDP 300 5 Aligning the unit 3 Remove the mirror cover from the unit t is held in position with magnets and must be removed by pulling from the top 4 Loosen the clamping screws on the laser mod ule and then rotate the laser to position beam vertically Installation and set up manual 209480 51 5 Aligning the unit GXDP 300 Cuspid 1 Remove the bottom cover from the lower shelf 6 screws 2 To move the laser backwards or forwards loosen the clamp screw A and then adjust the position of the laser with adjusting screw B 3 Check the positions of the lasers again and then replace all the covers and adjusting knob 52 Installation and set up manual 209480 GXDP 300 5 Aligning the unit 5 4 Calibrating the CCD sensor NOTE The CCD sensor is calibrated at the factory and thus does not normally need to be calibrated again However if the dental imaging software prompts an error message Please calibrate the device on startup perform CCD re calibration as instructed below 1 PC Prepare the unit for an exposure amp 2 Touch the service button on the main control P ceres panel of the unit to open the configuration win dow 3 Select INSTALLATION CMDS INSTALLATION CMDS 4 The personal identification numb
26. o the left or to the right When the unit is level in the left right direction place the spirit level lengthwise on the upper shelf and level the unit in the forwards back wards direction Make large adjustments by moving the base of the column backwards or forwards 18 Installation and set up manual 209480 GXDP 300 3 Installing the unit Make small adjustments by loosening the two screws that hold the bracket to the support bar and pulling or pushing the top of the column forwards or backwards Bracket N Support bar NOTE If the wall bracket alone does not hold the unit in position rigidly enough or the unit is to be used in an area prone to earthquakes attach the base of the column to the floor To do this make sure that the unit is level and then mark the positions of the two fixing holes on the floor Then loosen the two screws that hold the bracket to the support bar and rotate the unit to one side so the you can access the mark for the floor hole Drill a hole suitable for the 5 16 8 mm fixing hardware being used 7 Rotate the unit to the other side and drill the second hole Turn the unit back to the original position and then attach the unit to the floor with the fixing hardware Installation and set up manual 209480 19 3 Installing the unit GXDP 300 Check once more that the unit is vertical and then tighten all the fixing screws and wall bracket fastenings
27. ol panel of the unit Installation and set up manual 209480 GXDP 300 GXDP 300 GO1300001 Name GO1300001 FW 1 00 CORE 36 GRAPHICS 1 IP 194 9 227 251 INSTALLATION CMDS Current IP IP 194 9 227 251 Choose IP config mode MANUAL REMOTE 3 Installing the unit The configuration screen will appear on the display It shows the unit serial number and firmware version numbers Also there are two function buttons IP 194 9 227 251 IP configuration and INSTALLATION CMDS Installation com mands Press IP 194 9 227 251 button The current IP address of the unit is shown NOTE See section 4 for information about Installa tion Commands The IP configuration modes will appear on the display Remote or Manual Select Remote Installation and set up manual 209480 393 3 Installing the unit GXDP 300 9 The IP address window will appear which shows the current IP address of the unit Current IP 10 PC Press the Send to Device button im P based Device IP address Enable changing IP address 94 9 227 250 Send to Device 11 The new IP address will appear on the display New IP Touch the OK button to accept the IP address and then the Arrow button to return to the main display 12 PC Click the OK button to close the settings window NOTE If the remote IP address setting fails use the MANUAL method See section Manually changing the
28. olumn on the sides of the column that hold the hinged column section in the ver tical position Turn the hinged column section down so that it is horizontal Secure it in posi tion with the two screws supplied M10 x 16 the fixing holes are on the underside rear of the column Tighten the two screws M10 x 16 on the sides of the column Installation and set up manual 209480 15 3 Installing the unit Top of column Wall mountin assembiy 16 GXDP 300 IMPORTANT NOTE If the unit is to be attached to the show stand refer to the separate installation in structions supplied with the show stand Lift the top of the column and slide the packing piece that supported it underneath so that you can access the underside of the column Attach the wall mounting assembly to the back of the column with the screws nuts and wash ers supplied M10 NOTE The 4 x M10 washers must be positioned BE TWEEN the wall mounting assembly and the rear of the column The wall mounting assembly is adjustable so that the unit can adjusted in the backwards forwards direction Make sure that the wall mounting assembly is in the mid position of its adjustment To do this loosen the two screws that hold the bracket part of the wall mounting assembly at tached to the column to the support bar to be attached to the wall and position it to the mid position Retighten the screws Installation and set up manual 209480 GXDP 300
29. pin phantom test 41 4 3 Checking the positioning lasers eee dinsel asidin aksi 43 4 4 Completing TC ITS til ER am ll mma sakin 44 5 Aligning pg 45 A le pipe 2 CCD pg A TT 45 5 2 Adjusting the image SONG n ala la ca atl al lem 47 cener DAMN TIO TOUNO SR A MA NM P 47 Distances between center ball and left and right pins not equal 48 SS eh ale ME e e E A E A DOTT 50 Midsagittal and frankfort EE YA 50 Slo A A ye 52 5 4 Calibrating the CCD sensor KM cdseceas tadsepiwncnadeest extechocouesetendestend 53 vi Installation and set up manual 209480 GXDP 300 1 Introduction 1 Introduction 1 1 About this manual This manual explains how to install and set up the GXDP 300 digital panoramic dental x ray unit the unit You will need to operate the unit during the align ment checking procedure Make sure that you know how to operate the unit before starting to install and set up the unit It is also essential that person installing and set ting up the unit should have attended the manu facturer s service training course 1 2 Associated documentation The GXDP 300 User s manual The GXDP 300 Service manual not supplied in the delivery GxPicture Software Driver documentation The documentation supplied with VixWin or any other dental imaging software that will be used with the device Installation and set up manual 209480 1 1 Introduction GXDP 300
30. sure that the floor where the unit is to be installed can support this weight The unit must be permanently attached to the wall and if required to the floor If this is not possible use the show stand If the area where the unit is to be installed is subject to earthquakes the unit must be per manently attached to the wall and the base of the column must be permanently attached to the floor If this is not possible use the show stand The fixing hardware NOT SUPPLIED to be used to permanently attach the unit to the wall floor must be of the correct type for the wall floor materials WALL hardware MUST be 3 8 10 mm FLOOR hardware MUST be 5 16 8 mm DO NOT use smaller diameter fixing hardware Refer also to section 2 7 for more information Make sure that the fixing hardware to be used to attach the unit to the wall and the wall itself can withstand a continuous pull out force of at least 660 Ib 300 kg Installation and set up manual 209480 2 Pre installation requirements GXDP 300 If the wall is made of thin material you may have to use a reinforcing plate on the rear side of the wall to hold the fixing hardware If the wall is not strong enough to support the unit or if you are not sure if the wall will support the unit use the show stand The place where the unit is to be installed and the position from where the user will take exposures must be correctly shielded from the radiation that is genera
31. ted when the unit is operated Since radiation safety requirements vary from country to country and state to state it is the responsibility of the installer to ensure that all safety requirements are met Do not install the unit in environments where there are corrosive or explosive vapours opecial steps regarding EMC need to be taken when installing the unit For more informa tion refer to the EMC declaration in the user s manual The unit must be installed at least 73 1 85 m away from any non medical electrical or elec tromechanical equipment The unit must be connected to its own dedi cated power supply e 100 120 VAC or e 220 240 VAC Extension cables MUST NOT be used with the unit Installation and set up manual 209480 GXDP 300 2 Pre installation requirements 2 2 The PC e he PC to be used with the unit must be in stalled in a location that meets all local and national safety requirements with regards the connection of a PC to an x ray device e The connection of the unit to the PC must meet EN60601 1 requirements The PC must be positioned at least 73 1 85 m away from the unit e The use of ACCESSORY equipment not com plying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system Consideration relating to the choice shall include use of the accessory in the PATIENT VICINITY evidence that the safety certification of the
32. the unit it is recommended that it is placed on the left hand side of the column Make sure that when the user is using the exposure switch they can protect them self from radiation and can see the patient during an exposure If the user does not want the exposure switch positioned on the unit make sure that it is positioned so that they can protect them self from radiation and can see the patient during an exposure Also make sure that the expo sure switch is positioned so that it is within the extension range of the spiral cable Make sure that the surface to which the switch hook is attached is flat smooth and free from dirt and grease 2 Connect the exposure switch cable to the con nector at the bottom rear of the column and then place the exposure switch on the expo sure switch hook 24 Installation and set up manual 209480 GXDP 300 3 Installing the unit 3 4 Installing the remote exposure switch 1 Connect one end of the remote exposure switch extension cable to connector J2 inside the remote exposure switch Attach the remote exposure switch to the wall in the required position 2 screws Make sure that the cable from J2 is positioned in the small cut out in the bottom of the remote exposure switch housing Also make sure that the cable does not press against the circuit board when the housing is attached to the wall Cable to unit Route the remote exposure switch extension cable to the re
33. ws and the guide pin assembly step 8 BEFORE remov ing the unit from the wall or exhibition stand 22 Installation and set up manual 209480 GXDP 300 3 Installing the unit 12 Attach the covers to the bottom of the column Attach the Front Cover A to the front of the column with the special long screws 4 they are inserted from the rear Do not tighten them yet A Special long screws x4 a Attach the Side Covers B to the sides of the column Note that the front edge of the Side Covers is positioned between the column and the Front Cover The rear edge is attached to the back of the column with two screws Tighten all the screws 12 Check that the bite rod can be inserted into the chin rest easily and that it stays in position Bite rod when released If it is too tight or too loose ad just the tightness with the set screw slot blade screwdriver 0 5 x 3mm Set screw Installation and set up manual 209480 23 3 Installing the unit GXDP 300 3 3 Exposure switch NOTE If the exposure switch is to be used with the re mote option refer to the section 3 4 Installing the remote exposure switch 1 The exposure switch hook allows the exposure switch to be attached to any flat smooth verti cal surface The exposure switch hook has double backed tape so that it can be posi tioned wherever the user prefers on the unit or somewhere else If the user wants the exposure switch posi tioned on
34. y a PC After the set time elapses the PC automatically releases the unit so that it can be used by another PC in the network the unit can be reserved in advance from another PC The default setting is 300 seconds This can be changed by keying in a new value Installation and set up manual 209480 3 3 Installing the unit 38 GXDP 300 3 PC Click OK to connect the PC to the unit NOTE An automatic technique will automatically lo cate the unit within the local area network and connect the PC PC Repeat the above process for all the other PCs in the network Make sure that you define the IP address of the unit in each PC and that the workstation identifier selected is not used in more than one PC Check the installation by starting image cap ture using the imaging software If the Use Multi room Connection was se lected the Workstation identifier of the PC 1 8 being used will appear on the unit display Installation and set up manual 209480 GXDP 300 4 Checking the alignment 4 Checking the alignment and finishing the installation 4 1 Checking the alignment of the CCD Sensor 1 PC Prepare the unit for an exposure NOTE If the dental imaging software prompts an er ror message Please calibrate the device on startup perform CCD re calibration as de scribed in section 5 4 Remove the chin rest Touch the service button on the main control panel of the unit to open the configurat