         Contents
1.  If this light fails to glow when the black button is pressed  replace  both batteries     BATTERY REPLACEMENT    Detach the battery connectors  Pull on the plastic straps to  remove the batteries  Replace with fresh zinc carbon or alkaline 9  volt batteries  Reinsert the plastic straps to secure batteries   Replace the bottom cover     CELL CUP    Self conditioning of the built in electrodes occurs each time the  button is pressed with a sample in the cell cup  This ensures  consistent results each time  With some samples a small downward  swing of the pointer is a result of this conditioning action  This  action is powerful and removes normal films of oil and dirt  However   if samples are allowed to dry in the cell cup  a film will build up   This film reduces accuracy  When there are visible films of oil  dirt   or scale in the cell cup or on the electrode  scrub them lightly with  a small brush and household cleanser  Thoroughly rinse out the  cleanser  and the meter is ready for accurate measurements     SOLUTIONS INSIDE THE METER   Your Dialysate Meter is a rugged instrument and will withstand  dialysate exposure around its cell  meter movement  and switches   However  care should be taken to keep solutions from leaking in  around the bottom cover  It is not sealed  to prevent condensation  from forming     Large amounts of calibration solution or dialysate will almost  certainly damage the meter movement or electronics  Such meters  should be returned to the Myron L C
2.  Myron L Company    Faulty instruments may be returned to us without prior  permission     METERS WITHIN TWO YEAR WARRANTY PERIOD  Failures    due to materials or workmanship will be repaired or replaced  our  option  without charge if returned freight prepaid  If failure is  deemed by the factory to have been caused by abuse or  tampering  the following procedure will apply     INSTRUMENTS CONTROLS OUT OF WARRANTY    Diagnosis will be made and repairs completed  providing the repair  charges are  70 00 or less    NOTE  Actual repair charges may be less than this amount     We will diagnose  but not repair  a returned meter and mail   an estimate of charges if ANY of the following apply    1  Repair charges will be more than  70 00   2  You specifically request an estimate of required repairs and  charges    3  The cost of required repairs exceeds one half the list price of  a new instrument    4  The instrument is over ten years old  Because of component  changes and improvements  such instruments can no longer  be repaired     NOTE  Unrepaired meters are discarded unless you want them  returned to you  If so  there is currently a  25 00 charge per  unrepaired instrument to cover diagnosis and handling    This warranty is limited to the repair or replacement of the  Myron L Dialysate Meter only  The Myron L Company assumes  no other responsibility or liability        2450 Impala Drive  Carlsbad  CA 92010 7226 USA    MYRON L  COMPANY    Printed in U S A     pH Conductivity In
3. SINGLE RANGE DIALYSATE METERS  Operation Instructions    User Manual for Models D 1       2450 Impala Drive    Carlsbad  CA 92010 7226 USA  Tel 760 438 2021 MYRON L    Fax 800 869 7668   760 931 9189 COMPANY    pH Conductivity Instrumentation    WWW  myro nl  com Accuracy    Reliability    Simplicity    DESCRIPTION    This manual describes your Myron L D 1 meter  tells you how to  use it  and how to keep it working accurately for many years    Your Model D 1 single range Dialysate Meter is a compact  instrument which operates on the principle of electrical conductivity   It will quickly determine if dialysate concentrations are correct  Its  dial provides an expanded 11 15 millimhos  millisiemens  scale  for accurate readings    A unique feature of Model D 1 is the green color band above the  dial  This provides an immediate visual confirmation of whether or  not the dialysate concentration is within the patient s  safe zone    The color band is adjustable and can be pre set by the doctor or  technician for each patient  Adjustment is made by turning the screw  on the face of the meter  This screw may be covered with tape or  a label to avoid tampering    The D 1 is 3 4  x 4 5  x 4 0   85 x 129 x 126mm  and weighs  less than one pound  0 45kg   D 1 Meters are completely self   contained  The built in cell is automatically temperature  compensated from 60   to 105  F  16   to 40   C   They are powered  by two 9 volt transistor batteries good for at least 2000 tests  or  one ye
4. ar shelf life  A light bulb  which glows when the black button  is pushed  serves as a battery reserve indicator     The pictures in this manual show the major operating parts of  your Myron L D 1 Dialysate Meter  Handle your D 1 and identify  these parts to become familiar with it              Plunger  Electrode    Cell Cup             Indicator    Pointer    Button    USING YOUR METER    NOTE  Before measuring dialysate samples  check the calibration  of your meter using Myron L Company 14 0 millimhos Standard  Solution  see CALIBRATION     1  Fill the cell cup to just above the top electrode with a sample of  the dialysate you want to test  The dipper end of the cell plunger  will deliver just the right amount  Be sure hands are clean and  dry to avoid contamination of the sample     NEVER FILL THE CELL BY DIPPING THE METER INTO  SOLUTIONS     2  Drop the plunger into the cell cup  Lift it about half way out of  the cell and drop it back in three or four times  Discard the  sample    NOTE  This procedure must be conscientiously performed for   samples at working  body  temperature to avoid artificially low   readings    3  Repeat steps 1 and 2 at least three times    4  Take another sample  drop in the plunger  and press the black  button  When the pointer stops  IMMEDIATELY read and  release the button    5  Repeat step 4 twice and record the last reading    NOTE  If using the  safe zone  color band and the reading is not   within the band  immediately confirm the calibra
5. e CELL CUP opposite  Rinse  the cell thoroughly and test the Standard Solution again  If the  D 1 Meter still does not indicate the correct value  recalibrate it  as described below     TO RECALIBRATE THE METER   1  Remove the bottom cover using fingernails or a small  screwdriver to loosen the front or rear edge  Identify the  Calibration Control  see photo below  so you can find it by touch  while calibrating    2  Test another sample of the Standard Solution  be careful to not  splash solution inside the meter     3  Adjust the Calibration Control until the meter pointer indicates  14 0     AOR LAIR EERIE SIE SN           Battery Strap    Lower Electrode     inside cell     Calibration    Control       IMPORTANT    The plunger enclosed with your Dialysate Meter has been  calibrated specifically for this instrument  If a different plunger is  used  the Dialysate Meter MUST BE RECALIBRATED with  Standard Solution  Make sure the plunger is completely seated in  the cell cup     ACCESSORIES    STANDARD SOLUTION   Your D 1 Meter has been factory calibrated with  Myron L type 14 0 millimhos Standard Solution Myron  L Company Dialysis Standard Solution is made with  deionized water and reagent grade Sodium Chloride   and is within 1 0  of Sodium Chloride reference  solutions  Quart and gallon sizes are available        MAINTENANCE  BATTERY CHECK    Model D 1 has a battery indicator glow light visible through the  small hole on the lower right hand corner of the meter face plate  
6. ompany for repair    If there are only a few drops inside the meter  dry it as follows   1  Shake excess solution out of the inside of the meter    2  Dab the exposed surfaces dry with an absorbent cloth or tissue   Avoid pushing any liquid into the Calibration Control or push  button switch    3  Air dry the meter in a warm area with the bottom cover off   Allow several hours for thorough drying    If the dialysate entered through a leak in the case or cell or if the  instrument shows erratic readings or other unusual behavior  return  it to the Myron L Company for servicing     SERVICE    lf your D 1 meter fails to operate properly    1  Check the glow light to make sure the batteries are good    2  Check the cell cup and electrodes to ensure they are clean   Any contamination could alter the reading  See  MAINTENANCE    Any service required other than battery replacement  cleaning   or calibration must be referred to the Myron L Company  When  returning an instrument  please pack it securely and enclose a  note specifying the problem  The plunger should be included  but  DO NOT ship with plunger in cell  damage could result      ORDERING    To order accessories or instruments  contact your nearest stocking  distributor  or the Myron L Company     WARRANT Y SERVICE    Myron L Dialysate Meters have a limited two year warranty  If  your instrument fails to operate properly  check the batteries and  calibration  If it still fails to function properly  return it prepaid to  the
7. strumentation    Accuracy     Reliability    Simplicity I age    Fax 800 869 7668   760 931 9189  www myronl com    Specifications subject to change without notice    Myron L Company 2001  W     
8. tion of your D 1   with 14 0 millimhos Standard Solution  see CALIBRATION     6  When you are finished with the meter  RINSE THE CELL CUP  with clean water  preferably R O   distilled  or deionized     CAUTION    DO NOT use with samples hotter than 105  F  40  C   The readings  WILL NOT be accurate    DO NOT splash solvents such as lacquer thinner  acetone   benzene  or chlorinated solvents on the plastic case    DO NOT fix or modify the meter  That will void your warranty  See  SERVICE for details or consult Myron L Company    DO NOT DIP THE INSTRUMENT INTO SOLUTIONS  If liquid does  get inside the instrument  see MAINTENANCE for instructions on  drying it     olor Band            ROOT ans        117        2 23    wr    15    hg   MYRON L  COMPANY    DIALYSATE METER    CALIBRATION ACCESSORIES    STANDARD SOLUTION  Myron L Company Standard Solution  has a known conductivity and concentration of Sodium Chloride   NaCl   Your Dialysate meter was calibrated at the factory using  our 14 0 millimhos  7 86 grams per liter NaCl  Standard Solution   You can keep your meter accurate by using the same Standard  Solution     CHECKING CALIBRATION  1  Test a sample of Myron L Company 14 0 millimhos Standard  Solution     CAUTION  Throw the Standard Solution away as you use it  Don t  put the used samples back in the bottle     2  Ifthe D 1 Meter does not indicate the 14 0 millimhos shown on  the Standard Solution bottle s label  first clean the cell  For  directions on how to do this  se
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Manual Fluke 113  User Guide Bluetooth Intelligent Serial Module ATMP Command Set  Bowmount Foot Control Trolling Motors CE  1年・通年・3単位・必修 電子制御工学科  Casio P460 User's Manual  SSPC-01 SSI/パラレル変換器 取扱説明書  Service Manual  MANUAL - Tanita  iStarUSA EAGE4402TG-MS disk array  概要版(PDF形式:151KB)    Copyright © All rights reserved. 