        KAM® KF KARL FISCHER MOISTURE ANALYZER
         Contents
1.          This cell is equipped with an  electrolysis electrode which has  two platinum screens separated  by a ceramic frit  membrane    This is the cell that holds  Generator Solution C                       Drying Tube   Removes moisture from the  the atmosphere to the  Detector Electrode Generator Solution Cell     Detects the conductivity of the fluid    Septum    Sample Injection Port    Generator Solution Cell  with Glass Top    Holds the Generator Solution A    Stir Bar    KFMANUAL 0112 6 KAM CONTROLS  INC     COMPONENTS CONTINUED    SIDE and REAR PANELS    Jack for the Detector Electrode  Insert the male plug for the Detector  Electrode in the jack marked DETECTOR       Q    Power Plug       Stirring Speed  The stirring speed of the bar in  the cell is preset  If a slower or  faster stirring speed is  required  adjust the knob   Stirring speed is automatically  corrected for varying viscosity     Jack for the Cathode Solution Cell  Insert the male plug for the Cathode Solution  Cell in the jack marked TITRATION    KFMANUAL 0112 7 KAM CONTROLS  INC     All Models  Before connecting the power cord to an AC power source  ensure power switch  on the Analyzer base is in the OFF position  press switch to the left      Portable Models  The power cord is kept inside the top of the portable case  When the  Analyzer is operated on battery power  the power cord is not necessary  Plug the portable  unit in to recharge     ASSEMBLY    1  Loosen the Hold Down Clamp  A  by  unscrewi
2.     The Stir Bar is operating    2  Determine the sample size  The standard moisture determination range for this  instrument is from 10 micrograms to 10 milligrams of water  In order to determine the  moisture content of the sample accurately and effectively  control the sample size  corresponding to the estimated moisture content of the sample  Use the following chart  to determine sample size     ESTIMATED MOISTURE CONTENT SAMPLE SIZE    1 100 pg or 0 0001  0 0100   odm    100 1000 pg or 0 010076  0 1000 76 0 1 1 ml    1000 10000 yg or 0 1000   1 000  0 01 0 1 ml       KFMANUAL 0112 10 KAM CONTROLS  INC     OPERATION CONTINUED    3  To draw a sample  first clean the syringe by rinsing the inside of the syringe barrel and  needle with the sample to be measured  Draw this sample into the syringe and  discharge the sample into a waste container at least three times    4  Draw a sample in to the syringe  The sample volume should be slightly larger than the  desired amount    5  Discharge air bubbles and excess sample fluid from the inside of the syringe by slowly  pushing the plunger to the desired sample size    6  Press RESET key to clear display to    0       7  Insert syringe needle through the Sample Injection Port on the Generator Solution Cell   ensuring that the needle tip is submerged in the Reagent    8  Inject the sample into the Reagent  Do not allow the sample to splash up against the  wall of the Generator Solution Cell or the needle to touch the sides of the cel
3.  Cell  Detector Electrode  Drying Tube  Sample Injection Port  Stir Bar    P  09064 DETECTOR ELECTRODE  P  09063 CATHODE SOLUTION CELL    P  09062 GENERATOR SOLUTION CELL    KFMANUAL 0112 15 KAM CONTROLS  INC     ORDERING INFORMATION    SYRINGES AND NEEDLES   P  09010 GASTIGHT SYRINGE  10 pl  P  09023 GASTIGHT SYRINGE  50 pl  P  09021 GASTIGHT SYRINGE  250 pl  P  09022 GASTIGHT SYRINGE  500 pl  P  09008 GASTIGHT SYRINGE  1 ml  P  09001 GASTIGHT SYRINGE  2 5 ml  P  09002 NEEDLE  19 gauge  4      P  09003 NEEDLE  14 gauge  4     SPARE PARTS   P  09049 SAMPLE INJECTION PORT  P  09044 STIR BAR   P  09050 DRYING TUBE   P  09029 RUBBER SEPTUM  pack of 50    P  09015 SPECIAL GREASE  25 g tube    P  370420 REPLACEMENT RECHARGEABLE BATTERY  for model 09031     P  370400 POWER ADAPTER  110 120 or 220 240 VAC    ACCESSORIES    P  09014 HOMOGENIZER DRIVE UNIT  110 120 or 220 240 VAC with speed control    P  09019 DISPERSING TOOL FOR HOMOGENIZER  18 mm  for 50 2000 ml volume     P  09009 DISPERSING TOOL FOR HOMOGENIZER  25 mm  for 500 4000 ml volume     P  09018 HOMOGENIZER STAND    KFMANUAL 0112    
4.  Cell and pour entire  premeasured  contents of KAM Generator Solution C  P  09006 in a 5 ml ampoule  into the Cathode  Solution Cell using the 35 mm funnel  enclosed   After pouring the solution  replace the  stopper     NOTE  Wrap the Generator Solution C ampoule in a towel to break tip before pouring     KAM CONTROLS Reagents ship  already measured at the appropriate  volumes for Reagent replacement       BLANK ELIMINATION OPERATION    Each time new Reagents are added to the Cathode and Generator Solution Cells a blank  elimination must be performed to remove any residual moisture within the cells     2     Plug unit in and turn the Power Switch ON  When the Analyzer is energized  the digital  display reads    0        Push the Titration ON key  The Stir Bar will begin rotating  Approximately 5 seconds  later  electrolysis current will flow  As the residual moisture in the cell decreases  because of the reaction with iodine  the detector potential also decreases  In  association with the detector potential  the electrolysis current decreases  proportionally  When the detector potential approaches the zero point  the electrolysis  current decreases to a level lower than 5 mA  where it finally comes into equilibrium  with the blank current  When the blank current becomes less than 1 mA  the  measurement is complete  The Analyzer is now ready for measurement     KFMANUAL 0112 9 KAM CONTROLS  INC     OPERATION CONTINUED  IN CASE OF A LARGE BLANK CURRENT    If moisture has coll
5. H  inches   Lab Model Approx  10 pounds   Portable Model Approx  20 pounds    3 KAM CONTROLS  INC        DISPLAY PANEL    A  Microgram Mode   Press to initialize measurement of the  mass of water  in micrograms  in the  injected sample  The numerical display  will show the micrograms of water in  the sample from 0 to 99 999     B  Volume Percentage Mode   Press once to initialize measurement of  the volume percentage of water in the  injected sample  Press repeatedly to  cycle through sample sizes to be  injected  percent mode 0 1  0 25  0 5   or 1 0 ml  dilution mode 0 5 or 1 0 ml    Selected size is displayed in the upper  right hand corner of the LCD readout     For calculations of weight percent other  than 0 1  0 25  0 5  and 1 0 ml  please  refer to the API MPMS Chapter 10 9   or the ASTM D4928     A  Microgram mode    B  Volume percentage mode    C  On  titration  we    D  Off  titration        TITRATION    E  Reset  titration     KFMANUAL 0112       o       C  ON  Titration    The ON key has two functions  1  It initiates the  flow or titration current which corresponds to the  moisture content in the sample  The titration current  flows until all of the moisture is consumed  titrated    2  The ON key also functions as the magnetic stir   rer start switch     Note  Between titrations there will be some current  flow to balance accumulated moisture     D  OFF  Titration    Press to end titration and the operation of the   magnetic stirrer  The following operations requ
6. Kam   Controls  Incorporated   simple   precision     KAM   KF KARL FISCHER MOISTURE ANALYZER    USER MANUAL  PER ISO 12937  ASTM D 4928   API MPMS 10 9  ISO 12937  IP 386    REV  0112    KAM CONTROLS  INC   3939 Ann Arbor Drive  Houston  Texas USA 77063  T   1 713 784 0000  F   1 713 784 0001  www KAM com    KAM CONTROLS is an ISO 9001 certified company     KAM CONTROLS  INC  reserves the right to make changes to this document without notice     SECTION    KFMANUAL 0112    TITLE    Introduction    Available models    Theory of operation    Specifications    Components     Display panel     Titration Cell assembly    Side and Rear Panels    Operation     Assembly     Adding Reagent     Blank elimination operation     Moisture content determination    Sample dilution    Maintenance     Reagent replacement procedure    Storage     Special Grease     Electrode cleaning procedures    Ordering information    PAGE    NN ND    KAM CONTROLS  INC           AVAILABLE MODELS and ACCESSORIES    KAM   LKF Karl Fischer Moisture Analyzer Lab Model  KAM  PKF Karl Fischer Moisture Analyzer Portable Model  KAM  Homogenizer     For a full list of consumables  accessories  and parts  see page 15     THEORY OF OPERATION    KAM Moisture Analyzers incorporate the coulometry principle applied to Karl Fischer titration to totally eliminate  the troublesome procedures involved in conventional water determinations  Available in a lab model as well as a  portable  KAM Moisture Analyzers are easy to use 
7. a forced air dryer     14  Place the entire assembly in a laboratory oven and continue to dry it for several hours  at 120  F     15  Fill the cell with Generator Solution C and reinstall in the Generator Solution Cell     The Generator Solution Cell is cleaned in a similar manner minus the cleaning of the inside of  the probe     KFMANUAL 0112 14 KAM CONTROLS  INC        JR DENI NFOURMALION  KAM KF KARL FISCHER MOISTURE ANALYZERS    P  09031 KAM PKF PORTABLE KARL FISCHER MOISTURE ANALYZER  Battery powered  Analyzer includes Recharger  120 VAC   Titration Cell  Electrodes  Stir Bar  Drying Tube   and Carrying Case     P  09042 KAM LKF LAB KARL FISCHER MOISTURE ANALYZER  includes Titration Cell   Electrodes  Stir Bar  Drying Tube  and 120 VAC Power Adapter     P  09041 STARTER KIT  Contains all required items for analysis  including  Generator    Solution A  Generator Solution C  Calibration Verification Solution  1 ml and 10 pl Gastight  Syringes with Needles  package of 100 Septums  2 Funnels  Special Grease and Silica Gel     SUPPLIES AND CONSUMABLES    STANDARD GENERATOR SOLUTIONS  P  09004 GENERATOR SOLUTION A  Anode   pack of 6  100 ml bottles  P  09006 GENERATOR SOLUTION C  Cathode   pack of 6  5 ml ampoules    P  09032 CALIBRATION VERIFICATION CHECK SOLUTION  box of 10  4 ml ampoules  1   water content   certificate included    P  09013 SILICA GEL  100 ml bottle  GLASSWARE    P  09052 COMPLETE TITRATION CELL ASSEMBLY  includes Generator Solution Cell   Cathode Solution
8. and fully automatic  The Portable Model has a rugged case  for field water determinations and the capability to operate continuously for 10 hours before needing recharging   Incorporating a special electrolysis current control system  Moisture Analyzers assure fast  accurate measurement  of even trace content     KAM Moisture Analyzers can be used to rapidly and accurately determine the water content of liquid hydrocar   bons for all custody transfer operations  pipeline  marine  or truck movement  The Moisture Analyzer can also be  used to analyze crude oils  distillates  transformer oils  jet fuels  chemicals  and most other liquids  In the plant  it  can monitor the moisture content of streams during start ups  shutdowns  upsets  and normal operations  This can  be especially important on units where expensive catalysts can be damaged by excessive moisture  The   unit also can be used by machinists as part of a preventive maintenance program  By analyzing lube oil on  compressors  turbines  etc   the Moisture Analyzer can detect cooling water leaks before they become severe  enough to damage equipment     Karl Fischer titrimetry is an accurate moisture measurement method utilizing the quantitative reaction of water with  iodine  This method is the standard method of moisture measurement because of its speed and high sensitivity  In  coulometric Karl Fischer titration  the sample is added to the Karl Fischer Reagent  iodine ion and sulfur dioxide as  principal components   
9. e Solution Cell and discard     2  Wipe the grease off the tapered joint with tissue paper wetted with acetone     3  Remove any exterior contamination with tissue paper and acetone being careful not  to damage the platinum wires and screen on the outside of the cell     KFMANUAL 0112 13 KAM CONTROLS  INC     MAINTENANCE CONTINUED    4  Fill the Cathode Cell with acetone and stopper the inlet joint  Shake the unit until any  dirt is removed  This should be repeated several times  Pour out and discard acetone     5  Wash the outer surface with acetone    6  Pour approximately 5 ml of nitric acid into the Cathode Solution Cell    7  Place the cell in a 50 ml beaker and fill the beaker with nitric acid to the same level as  is in the cell  Secure the Cathode Solution Cell so that it will not fall over  Allow the cell    to remain in the solution overnight     8  After the cell has stood overnight  discard the nitric acid in the cell  and rinse the inside  and outside of the cell with running tap water until the yellow color disappears     9  Place distilled water in the cell and place it in a 400 ml beaker filed with distilled water   10  Place the beaker on a hot plate and bring the water to a slow boil for at least one hour   11  Repeat this operation until there is no coloration of the water     12  Remove the Cathode Solution Cell from the water and fill it with acetone to overflowing   flooding the outside of the cell as well  Repeat several times     13  Dry the cell with 
10. ected within the inner area and on the inner wall of the Generator  Solution Cell  the blank current becomes larger than approximately 1 mA     In this case  press the RESET button and the Titration OFF button  When the Stir Bar has  ceased rotation  remove the Generator Solution Cell and gently swirl the Generator  Solution around the sides of the Cell in order to allow any moisture inside the cell to be  absorbed by the Reagent  Remount the Generator Solution Cell on the cell holder and  restart electrolysis     Once you have conducted this procedure a couple of times  the blank current decreases  gradually to near the zero point in most cases  If the blank current cannot be reduced to a  low level  the cause is a leakage of moisture from the atmosphere  In this case  a close  inspection of all joints and Silica Gel is recommended  If this does not work  try cleaning  and drying the Generator Solution Cell     If the titration continues and never reaches an end point  the Generator Solution A and  the Generator Solution C are generally worn out   Follow the steps on page 12 to replace  Reagent   The Analyzer is now ready for moisture content determinations     MOISTURE CONTENT DETERMINATION    Sample preparation and methodology must conform to the standards dictated in API MPMS  Chapter 10 9     1  Prior to sample analysis ensure that       Titration ON key is pressed and the digital display for titration and detection  is operative     The blank current is not fluctuating 
11. ire   that the titration has ceased and the magnetic   stirrer has stopped      Removing the cathode solution cell     Removing the detector electrode     Removing or gently swirling the generator  solution cell     Adding or removing Generator Solution A or  Generator Solution C    E  RESET  titration     Press to reset count to O and begin titration    KAM KARL FISCHER  MOISTURE ANALYZER    CHARGING        PRESS TO CHANGE SAMPLE SIZE  1 0  0 5  0 25  0 1 ML    FULL BATTERY       LOW BATTERY      HIGH MOISTURE         lt  Controls  Incorporated       4 KAM CONTROLS  INC        COMPONENTS CONTINUED    KAM KARL FISCHER  MOISTURE ANALYZER    dT  Charging indicator   Portable model only     CHARGING  ug e       Battery full indicator    v PRESS TO CHANGE SAMPLE SIZE FULL BATTERY    Portable model only   Ae 1 0  0 5  0 25  0 1 ML    LOW BATTERY    lt q Battery low indicator  TITRATION  Portable model only     HIGH MOISTURE    a High moisture indicator          cater  Incorporated       LCD DISPLAY    Mode indicator    Sample size indication    Status indicator    reads    Processing     during analysis and     Analysis Complete  when titration has Pia    finished  miini imum     n BE  WHEEL GL Percentage of water in    sample       Detector current  bar graph display  Titration current bar graph display    Mode indicator    Micrograms of water  in sample       KFMANUAL 0112 5 KAM CONTROLS  INC     COMPONENTS CONTINUED    TITRATION CELL ASSEMBLY    Cathode Solution Cell           
12. l    9   Electrolysis begins automatically by the injection of the sample  When the analysis is  completed  the buzzer sounds and the result is displayed on the digital readout and  can be recorded    NOTE  Do not press RESET during titration or the instrument will fail to produce correct  data    SAMPLE DILUTION    In cases where the viscosity of the sample is such that it cannot be pulled into a syringe or  when the anticipated water content is higher than 5   users may dilute the sample  according to the following steps     2     Dilute  mix  the sample with a ratio of 1 part  1 ml  sample to 99 parts  99 ml  of  Reagent grade xylene     The Analyzer automatically adjusts for the dilution factor based on the sample size  selected     KFMANUAL 0112 n KAM CONTROLS  INC     iura LN    REAGENT REPLACEMENT PROCEDURE    API Chapter 10 9 Section 7 states that Reagents should be replaced after 7 days     2     12     13     Carefully remove the stoppers from the Generator and Cathode Solution Cells  Care  should be exercised not to damage the two electrode connections     With a suitable suction or aspirator device  remove the old solution from the Generator  and Cathode Solution compartments     Wipe all tapered glass joints  male and female  with clean tissue to remove the old  grease     Pour a 5 ml ampoule of Generator Solution C  P  09006  into the Cathode  inner   Solution Cell using the 35 mm funnel     Grease the glass stopper with Special Grease  P  09015  and insert it i
13. lodine  generated electrolytically at the anode  reacts with water in the sample as shown in  Formula 1  lodine is generated in direct proportion to the quantity of electricity according to Faraday s Law     Formula 1       SO    H O   2HI   SO    One mole of iodine reacts quantitatively to one mole of water  therefore  1mg of water is equivalent to 10 71  coulombs  Formula 2   Based on this principle  water content can be directly determined from the quantity of  electricity required for electrolysis  This eliminates the need for addition of Reagent and tedious standardization    procedures     Formula 2  2l  2e 1    KFMANUAL 0112 2 KAM CONTROLS  INC     Method   Detection   Control   Sample Size   Range   Sensitivity   Precision   Titration Speed     Power Requirements     Dimensions     Weight     KFMANUAL 0112    Coulometric Karl Fischer titration  ASTM D 4928  API MPMS 10 9   ISO 12937  IP 386   Polarization detection   Automatic electrolysis current control   0 1  0 25  0 5  1 0 ml for direct reading of volume   H O  or less than  two grams  ml    10 pg   100 000 pg H O  i e  for a 100 pl sample  the direct display is  0 0010  to 10  H O    1 pg H O    5 pg for 10 pg   1000 pg   0 5   C V   for over 1000 pg    meets or exceeds API MPMS 10 9  ASTM D 4928  and IP 386    1000 pg H O min   max   at high H O concentrations    Operates on either AC or DC  AC   110 120  220 240 V  50 60Hz  DC   Rechargeable battery for portable model  12   14 V    Approx  15  W  x 10  L  x 9  
14. ng the compression nut  and slide  it upwards until it rests at the bottom of the  Thumb Screws  B      2  Slide the Generator Solution Cell  C   through the Hold Down Clamp until it  rests on top of the Stirring Plate  D   Then  tighten the compression nut on the Hold  Down Clamp to secure the cell in place        3  Prior to installation  apply a thin  layer of Special Grease to the tapered  joint of each piece of glassware in order  to ensure an air tight  moisture tight  Generator Solution Cell     4  Insert the Cathode Solution Cell  E    Detector Electrode  F  and Drying Tube   G  into the appropriately sized holes in  the top of the Generator Solution Cell  C      5  Place a Rubber Septum  H  beneath the  Sample Injection Port  I      NOTE  Do NOT grease the Sample  Injection Port     6  Insert the plug of the Cathode Cell  E  in  the jack labeled TITRATION  J and the  plug of the Detector Electrode  F  into the  jack labeled DETECTOR  K      NOTE  Orientation of the compression  nut can be reversed to accommodate left or  righthanded operation        KFMANUAL 0112 8 KAM CONTROLS  INC        OPERATION CONTINUED    ADDING REAGENT    Remove the stopper connected to the drying tube from the Generator Solution Cell  and pour entire  premeasured contents of KAM Generator Solution A  P  09004 in  a 100 ml bottle  into the Generator Solution Cell using the 55 mm funnel  enclosed    After pouring the solution replace the stopper     Remove the stopper from the Cathode Solution
15. nto the seat on  the cell     Pour a 100 ml bottle of Generator Solution A  P  09004  into the Generator Solution  Cell using the 55 mm funnel     Grease the detector and insert it into the seat on the cell     Insert Sample Injection Port  without grease  into the seat on the Generator Solution  Cell     Change the rubber septum  P  09029  inside of the Sample Port     Press the Titration ON button  Allow the residual moisture in the titration vessel to be  titrated until the end point is reached     Do not proceed beyond this point until the titration end point is reached     If the titration continues at a slow rate for a prolonged period  press the Titration OFF  button  Loosen the hold down clamp on the Generator Solution Cell  Carefully remove  the cell and gently swirl it to remove any moisture from the sides of the vessel  Replace  the cell and tighten the hold down clamp before resuming the titration  The end point  should then be reached very quickly     Set the Analyzer to the g or   mode  The apparatus is now ready for sample analysis     KFMANUAL 0112 12 KAM CONTROLS  INC        MAINTENANCE CONTINUED    STORAGE    When the Analyzer is out of use for an extended period of time  do not disconnect any of  the attachments on the Generator Solution Cell  Keep all parts mounted to prevent moisture  absorption in the Reagents  If the cell is left standing with the drying tube removed the  ceramic frit on the Cathode Solution Cell will absorb moisture     For storage ove
16. r 2 weeks time  all glass needs to be washed cleaned and dried   SPECIAL GREASE    To keep the Generator Solution Cell performing accurately and effectively be sure that a  thin layer of Special Grease is applied to all glass tapered joints  male and female  each  time the Reagents are replaced  Frequent greasing will prevent the glassware from   seizing and causing damage to the cell  The Special Grease does not absorb moisture  so it  will not hinder the results of the analysis  No other grease should be used with a KAM KF  Karl Fischer Moisture Analyzer due to the potential for the grease to absorb moisture and  interfere with titration if it is silicon based     KAM CONTROLS is not responsible for problems resulting from the use of improper grease   ELECTRODE CLEANING PROCEDURES    Dirty electrodes usually result in the following problems      A decrease in titration efficiency resulting in a longer measuring time     An increase in the blank  residual  value of titration current due to absorption of moisture  in the dirty area     Unstable titration current and lack of reproducible results     Staining on the ceramic frit at the bottom of the Cathode Solution Cell   NOTE  This procedure uses Reagent grade nitric acid  Because nitric acid produces   corrosive fumes  be sure and conduct all parts of this procedure in an appropriately    ventilated environment     Nitric acid should NOT be reused  Follow appropriate disposal procedures   1  Pour the contents from the Cathod
    