1. Oximeter Plus Oxi Go QuickCheck Pro OPERATOR S MANUAL VER1 0025 Fingertip Pulse Oximeter C Eon General Description Thank you for purchasing the Oxi Go QuickCheck Pro pulse oximeter This oximeter can be used to measure your Blood Oxygen Saturation in SpO2 and your pulse rate It should be used for Spot Monitoring only and not for continuous monitoring The Oxi Go QuickCheck Pro Finger Unit spot check oximeter is not a medical device and should only be used by healthy individuals who are performing non medical activities such as sports and aviation It is ideal for use during sports activities mountain climbing and piloting airplanes It is sold without a prescription only to consumers who do not require a medical prescription It is not to be used for wellness applications Package Contents One Oxi Go Pro Oximeter One lanyard Two AAA batteries One user manual Product Operation Scope Oximeters are sensitive to motion artifacts Keep hands still while taking a reading a aT N Oximeters require sufficient blood flow to obtain proper readings Poor blood circulation can result in inaccurate readings If your hands are cold or you have poor circulation Warm your hands by rubbing them together or use another method before attempting to obtain a reading A tourniquet blood pressure cuff or other blood flow hindrances may also result in inaccurate readings ga Fingernail polish or acrylic nail
2. determine the positioning of the sensor adequate circulation and skin sensitivity of the user 5 Do not stretch the rubber while applying the pulse oximeter sensor This may cause inaccurate readings or skin blisters Read the manual carefully before use The pulse oximeter has no SpOz alarms it is not for continuous monitoring as indicated by the symbol A p In addition to items described in the Product Operation Scope inaccurate measurements may be caused by FACTORS INCLUDING BUT NOT LIMITED TO 1 Autoclaving ethylene oxide sterilizing or immersing the sensors in liquid 2 Significant levels of dysfunctional hemoglobin such as carbonxy hemoglobin or methemoglobin 3 Intravascular dyes such as indocyanine green or methylene blue 4 SpO2 measurements may be adversely affected in the presence of high ambient light such as direct sunlight In bright light conditions cover the sensor area if necessary 5 Excessive user movement 6 High Frequency electrosurgical interference and defibrillators 7 Venous pulsations 8 The user has hypotension severe vasoconstriction severe anemia or hypothermia 9 Fingernail polish or false fingernails may cause inaccurate SpOz readings 10 Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components including batteries Technical Specifications 1 Display mode LED Measuring range for SPO2 35 99 Measuring
3. range for PR 30 235 BPM PR display mode bargraph Low power indication q oo pe ee ls Power consumption Less than 40mA Resolution 1 for SPO2 and 1BPM for Pulse Rate Measurement accuracy Battery standard Two AAA 1 5V Alkaline Battery SpO2 80 99 2 70 80 8 lt 70 no definition PR 30 235 BPM 2 or 2 BPM Ambient Humidity 15 80 in operation lt 93 in storage 6 Automatic power off The oximeter will power off automatically after 8 seconds when the finger is removed 7 It is best to store the product in a place where ambient temperatures range from 10 to 40C 14 104F 8 It is recommended that the product be kept in a dry place A damp ambient might affect its lifetime and even might damage the product EMC of this product comply with IEC60601 1 2 standard The materials which user can come into contact with are non toxici and comply with ISO10993 1 5 10 Guidance and manufacture s declaration electromagnetic emissions for all EQUIPMENT and SYSTEMS Guidance and manufacture s declaration electromagnetic emission The Pulse Oximeter is intended for use in the electromagnetic environment specified below Emission test RF emissions CISPR 11 RF emission CISPR 11 Compliance Group 1 Class B Electromagnetic environment guidance The Pulse Oximeter uses RF energy only for its internal function Therefore its RF emissions are very low and are not l
4. Mechanical Misplace for receive emission tube 4 Amp circuit malfunction 2 Replace the batteries 1 Change new battery 2 Please contact with local customer service center 3 Please contact with local customer service center 4 Please contact with local customer service center Symbol Definitions Symbol o SpO2 Warranty Definition The equipment type is BF Refer to user manual before application Hemoglobin saturation Heart rate BPM Low power indication No SpO2 Alarm Not for continuous monitoring Spot measurement only Power switch Serial No Limited 1 year warranty Manufactured for Oximeter Plus 18 Milburn Lane Roslyn Heights Phone Made in China NY 11577 516 626 6226
5. ikely to cause any interference in nearby electronic equipment The Pulse Oximeter is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Possible Problems and resolutions Problems Solution Possible reason SpO2 or PR is not be shown normally 1 Finger is not plugged correctly 2 User s Oxyhemoglobin value is too low to be measured Retry by plugging the finger Possible oximeter failure Seek medical attention 1 Finger might not be plugged SpO2 or PR is shown unstably The Oximeter can not be powered on deep enough 2 Finger is trembling or user s body is in movement status 1 Power of batteries might be inadequate or not be there at all 2 Batteries might be installed incorrectly 3 The Oximeter might be damaged Retry by plugging the finger Try not to move mM ON 1 Please replace batteries 2 Please reinstall the batteries 3 Please contact with local customer service center 1 Normal 1 The product is automatically Indication lamps are suddenly off Error or Error4 Displayed on screen powered off when no signal is detected longer than 8 seconds 2 Power quantity of the batteries is started being inadequate 1 Low power 2 Receiving tube being shielded or damaged together with broken connector 3
6. low for normal operation of the Oximeter Replace batteries when indicator symbol lights 4 Dispose of batteries as per local ordinances Using the Oximeter 1 Open the clamp as shown in the picture 2 Place one of your fingers into the rubber opening of the Oximeter your finger should touch the bottom portion before releasing the clamp When you put your finger into the Oximeter your nail must be facing upward towards the display portion of the oximeter 3 Press the switch button one time on front panel to turn the oximeter on 4 Keep your hands still for the reading Do not shake your finger during the test It is recommended that you do not move your body while taking a reading 5 Read data from display screen 6 There may be times when the signal strength is to low see heart rate bar graph indicator below for a proper reading Please switch to another finger or hand and follow instructions in this manual for low profusion blood flow 7 When no signal or low signal is detected the oximeter will power off automatically in 8 seconds 8 There are four display modes After turning on the oximeter each time you press the power switch the oximeter will switch to another display mode there are 4 display modes shown as follows SpOx PR tem wyd ods 198 77 ZL J6 Pulse Oximeter Pulse Oximeter ods Hd vt gir Pulse Oximeter Pulse Oximeter 9 When you press the power switch for more tha
7. n one second the brightness of the oximeter will be changed by degrees there are 10 levels on brightness the default level is level four Maintenance 1 Clean the oximeter and sensor with a soft cloth dampened with isopropyl alcohol Do not pour or spray any liquids onto the oximeter and do not allow any liquid to enter any openings in the device Allow the oximeter to dry thoroughly before reusing The rubber inside of the Oximeter is composed of medical grade rubber Itis non toxic and is not harmful to the skin 2 The pulse oximeter requires no routine calibration or maintenance other than replacement of batteries Brief Description of Front Panel Low power indicator JBJOWIXO S Nd ulse Oximeter SpO2 Waveform The height of the bar graph indicates the intensity of the pulse and signal strength The bar should be greater then 30 for a proper reading Warnings 1 Keep the oximeter away from young children Small parts such as the battery door batteries etc may be hazardous if swallowed 2 LD ats may cause strangulation in conditions that may cause it to twist around the neck Precautions for use 1 Do use the pulse oximeter in situations where alarms are required The device has no alarms 2 Explosion hazard Do not use the pulse oximeter in an explosive atmosphere 3 The pulse oximeter is intended only as a non medical device 4 Check the pulse oximeter sensor application site ey to
8. s obstruct the light transmission and may also result in inaccurate readings Your finger must be clean for proper reading The Oximeter must be clean for a proper reading If a reading is difficult to obtain switch to another finger or to the other hand There are a number of other conditions which may lead to an inaccurate reading including but not limited to recent medical tests that included an injection of dyes use of arterial catheters a weak pulse low levels of hemoglobin in the blood low perfusion the quality of your pulse elevated levels of dysfunctional hemoglobin the strength and type of light that you are in while using the Oximeter and the existence of cell phones radios and fixed transmitters within certain ranges of the oximeter during use The Oximeter will not alert you if your readings are out of normal range pe Ser gt g Operating Instructions Battery Installation 1 Install two AAA batteries into the battery compartment Match the plus and minus signs in the compartment If the polarities are not matched damaged may be caused to the oximeter 2 ass the battery door cover horizontally along the arrow shown as below Do not orce Install batteries in right polarity Incorrect placement may cause damage to the bracket Remove the batteries if the Oximeter will not be used for long periods of time 3 Battery indicator symbol on front panel display will light when battery voltage is too