1.    VOLCANO    Volcano Revo Option  For use with the Volcano s5    s5i   Series Systems    Operator   s Manual    Software Version Level 3 3 X    C     0086 Complies with the Council Directive 93 42 EEC     A    TUVRheinland           Volcano systems meet TUV   s safety requirements     A SPECIFIC      STANDARDS  TUVRheinland       Attention  Read Operator   s Manual and Instructions For  Use prior to using this device  Please contact your local  Volcano representative for translated versions     This device contains mercury  dispose according to  state local laws     Do not dispose of this device or its components   Improper disposal may be harmful to the environment  and human health  Please refer to  www volcanocorp com for instructions on disposal     Ia     Manufactured by     Volcano Corporation   2870 Kilgore Road   Rancho Cordova  CA 95670 USA   Phone  916 638 8008   Phone  800 228 4728  USA and Canada   Fax  916 638 8112  www volcanocorp com    Authorized European Representative     Volcano Corporation Europe BVBA SPRL  Excelsiorlaan 41   B 1930 Zaventem  Belgium   Phone   32 2 679 1076   Fax   32 2 679 1079       2012 Volcano Corporation  All rights  reserved  No part of this manual may be  reproduced in any form without the  written permission of Volcano  Corporation     Revision Date  November 2012    Page 2 of 22 505 0100 14 002    Table of Contents    MERTEN 4  PATENTS AND TRADEMARKG           cccsscsssssscssecssssesssnecsesssnssessecssnssessecsscssossscssesssses
2.  entire range of measurement achievable with the Revolution  45MHz Rotational IVUS Imaging Catheter  These are worst case measurement uncertainties  and represent situations where the ultrasound signals are entirely within tissue  such as  muscle  which has a very different speed of sound than blood  For measurement of luminal  boundaries  where the ultrasound only traverses through blood  the inaccuracy is    Distance Measurement    1  of the measurement    0 10mm     Area Measurement    5  of the measurement    0 10mm7     NOTE  A slight difference in total pullback length displayed on SpinVision  PIMr  and  ILD may exist due to the latent response in initiation termination of the pullback  relative to recording  This discrepancy only exists in the most proximal and distal  1mm of the video loop     In Line Digital measurement accuracy   e Two distance measurements on the longitudinal display  ILD    e Maximum Loop Distance  Z axis length    0 5 mm sec pull back  amp  30fps  limited by  memory of 5400 frames  90 mm   e Maximum Loop Distance  Z axis length    1 0 mm sec pull back  limited by device  dimension  150 mm   e Accuracy  including software and SpinVision  PIMr  accuracy    4  of the  measurement       NOTE  In rare situations  the actual pullback speed may differ from the expected value   This could result in erroneous length information derived from the ILD  This can be  caused by restricted motion of the PIMr or simultaneous activation of the speed  controls on t
3.  hand corner of the Home screen image     Perform    1     2     ing an Automatic Pullback    Using the  014    guide wire  advance the catheter to the most distal portion of the target  lesion according to standard intervention procedures    Select the pullback speed by pressing the AUTO button on SpinVision  PIMr   which  toggles between 0 5 mm and 1 0 mm per second  To zero the distance counter  press  RESET    Select Image on SpinVision  PIMr  or the Image On  Live  button on the system  screen to begin imaging  The SpinVision  PIMr  motor will begin spinning  and the  Revolution image will appear on the system screen  The absolute movement of the  transducer will be displayed on the digital readout on SpinVision  PIMr   The Image Off   Freeze   Save Frame  Record  Pullback   and Adjust Image buttons become visible  on the system screen    To stop imaging at any time  simply select the IMAGE button on SpinVision  PIMr  or  the Image Off  Freeze  button on the system  The image currently displayed on the  screen remains there  Note  Starting or stopping imaging also starts or stops the  rotation of the catheter        NOTE  If an image artifact is observed  it may be necessary to flush the imaging core with a 3    cc syr    inge of heparinized saline attached to the Revolution catheter port SpinVision  PIMr        5     Press PULLBACK on SpinVision  PIMr  or Record  Pullback  on the system  Pullback  begins at the selected speed  and the Volcano system begins to record the
4.  kind   expressed or implied  including any warranty of merchantability  suitability or fitness for a  particular purpose  respecting the misuse of the system  software or catheter  VOLCANO  Corporation assumes no responsibility or liability for incidental or consequential damages  which may result from re use or misuse of the system  software or catheter        WARNING  Always ensure that SpinVision  PIMr  is placed in a stable location and is  clear from items that may interfere or prevent smooth movement during use  The  SpinVision weight will transfer as the device performs a pullback           WARNING  SpinVision   s  PIMr  cable may be damaged if equipment is rolled on it  Do  not pull on the cable  place the cable in a high traffic area  or use excessive force as  the strain relief ends may be damaged  Do not use the system if the outer case or  wires appear damaged           WARNING  Do NOT advance SpinVision  PIMr  while the catheter is imaging           NOTE  SpinVision  PIMr  complies with the fluid ingress requirements of IEC 60529 IPX4  when configured for normal use  SpinVision  PIMr  should be located under a sterile drape  when configured for normal use        Page 6 of 22 505 0100 14 002       NOTE  Please refer to the Volcano system operator   s manuals for detailed information about  the operation of the Volcano system  Pay careful attention to all Cautions and Warnings           CAUTION  Do not use a damaged catheter  Do not re sterilize the catheter  Cath
5.  video loop        NOTE  SpinVision  PIMr  can accommodate a maximum pullback distance of 15 cm  after    which    it stops the pullback        Page 16 of 22    505 0100 14 002       NOTE  The pullback distance may be limited to 9 cm if the acquisition rate is set at 30  frames sec and a pullback rate of 0 5 mm sec is used  Adjust the acquisition rate to 15 f s or  pullback rate to 1 0 mm sec to obtain the entire 15 cm        6  Once the desired pullback distance is reached  you may stop the pullback by pressing  the PULLBACK button on SpinVision  PIMr   or by pressing the Stop button on the  system or software     Performing a Manual Pullback    1  Select MANUAL on SpinVision  PIMr   SpinVision  PIMr  unlocks from the sled so that  you can move it by hand     2  Select IMAGE on SpinVision  PIMr  or the Image On  Live  button on the Home  screen     3  Select PULLBACK on SpinVision  PIMr  or Record  Pullback  on the Home screen     4  Gently pull SpinVision  PIMr  back at the desired speed  Utilize the rubber grips on the  side of SpinVision  PIMr  to enhance control     5  Press STOP on the Home screen to stop recording the video loop        NOTE 1  The digital display can be reset at any time to give the absolute position           NOTE 2  If the SpinVision  PIMr  motor drive is left stationary for more than 30 seconds while  in Manual mode  the system will switch to Auto mode and lock the sled into place        WARNING  Do NOT advance SpinVision  PIMr  while the catheter i
6. ACTIR cece crceceeesecnaecaeeescsaesaeseeessecaeeseseaesaseatenee 19  CHAPTER 5  TECHNICAL SPECIFICATIONG           csccssssssssssssssssecsssssssecssesseseesessssssessoes 21  SPINVISION  PIMR  DIMENSIONS AND WEIGHTS 000    ccceeeecseeee caste ceeaeteeeaeeee 21  CATHE TER SPECIBICA TIONS vise sccsisesssectentovsesetessseupensogetesnoccountenssoaggssbscetus ovtedutssapuueed AS 21  MEASUREMENT  ACCURACY aroni oa a a A beer 22  PACKAGING EE 22    505 0100 14 002 Page 3 of 22    Warranty    NOTICE  Manufacturer   s Specifications and Policies Subject to Change  Volcano Corporation  reserves the right to make changes in the products described in this manual in order to improve  design or performance  Reproduction or distribution of any portion of this manual without the prior  written consent of Volcano Corporation is prohibited     LIMITED WARRANTY    Subject to the conditions and limitations on liability stated herein  Volcano Corporation     VOLCANO      warrants that the Volcano PIMr  SpinVision     the    System     as so delivered  shall materially conform  to Volcano   s then current specifications for the System  for a period of one year from the date of  delivery  ANY LIABILITY OF VOLCANO WITH RESPECT TO THE SYSTEM OR THE  PERFORMANCE THEREOF UNDER ANY WARRANTY  NEGLIGENCE  STRICT LIABILITY OR  OTHER THEORY WILL BE LIMITED EXCLUSIVELY TO SYSTEM REPAIR  REPLACEMENT OR  IF  REPLACEMENT IS INADEQUATE AS A REMEDY OR  IN THE OPINION OF VOLCANO   IMPRACTICAL  TO REFUND OF THE PR
7. ICE PAID FOR THE SYSTEM  EXCEPT FOR THE  FOREGOING  THE SYSTEM IS PROVIDED    AS IS    WITHOUT WARRANTY OF ANY KIND   EXPRESS OR IMPLIED  INCLUDING WITHOUT LIMITATION  ANY WARRANTY OF FITNESS   MERCHANTABILITY  FITNESS FOR A PARTICULAR PURPOSE OR NON INFRINGEMENT   FURTHER  VOLCANO DOES NOT WARRANT  GUARANTEE  OR MAKE ANY  REPRESENTATIONS REGARDING THE USE  OR THE RESULTS OF THE USE  OF THE SYSTEM  OR WRITTEN MATERIALS IN TERMS OF CORRECTNESS  ACCURACY  RELIABILITY  OR  OTHERWISE  Buyer understands that VOLCANO is not responsible for and will have no liability for  any items or any services provided by any persons other than VOLCANO  VOLCANO shall have no  liability for delays or failures beyond its reasonable control    Additionally  this warranty does not apply if    1  The System is operated in other than a manner prescribed by Volcano Corporation in  the Operator   s Manual  and or supplements    2  The System is operated in a manner that is not in conformance with purchase  specifications and specifications contained in the Operator   s Manual  and or  supplements    3  The System is not maintained in accordance with procedures in the Operator   s Manual   and or supplements    4  The System is repaired  altered  or modified in any way by other than Volcano  Corporation authorized personnel  or without Volcano Corporation authorization    Contact Volcano Corporation Field Service for instructions and issuance of a Return Material  Authorization if claims under this warra
8. ce  Try another catheter  If these  do not resolve the problem  call Volcano Technical  Support     The unit has failed  Call Volcano Technical  Support     The unit has failed  Call Volcano Technical  Support     The unit has failed  Call Volcano Technical  Support      Catheter rotation has stalled  Ensure the catheter  has been prepared and flushed per the catheter  IFU  If this does not correct the problem  call  Volcano Technical Support     Ensure the pullback has not come to the end of its  travel  Check the cable for binding and the  catheter for excessive drag in the body  If these  do not correct the problem  call Volcano Technical  Support     The unit has failed  Call Volcano Technical  Support        sheath    Non Uniform Rotational  Distortion  NURD     catheter    Artifact Cause Resolution  Degraded image due to Air surrounding the IVUS Flush with heparinized saline per  shadowing transducer within the catheter IFU    Excessive tightening of the  hemostasis valve    Excessive bending of the    Loosen the hemostasis valve    Make sure catheter is not bent  between SpinVision  PIMr  and  hemostasis valve  Flushing with  heparinized saline may also help        Ring artifact in image    sheath          Air surrounding the IVUS  transducer within the catheter IFU    Flush with heparinized saline per             505 0100 14 002    Page 19 of 22    This page is intentionally blank for pagination purposes     Page 20 of 22 505 0100 14 002    Chapter 5  Technical Specif
9. de  LIVE Gummi OFF Ring Down  OFF WH  OFF  Catheter  Revolution    Figure 4    Time Gain Compensation Dialog Box    To set the TGC     1     2   3     Page 14 of 22    Select the TGC button in the Adjust Image screen  Note that you must be imaging  with a Revolution catheter for the TGC feature to be enabled     Select from one of the preset profiles using the drop down box or arrow keys     Move the sliders or use the arrows to set the gain for each desired region  as indicated  by the number of mm shown to the left of each slider      Click Close to save your settings and close the dialog box  If you do not want to save  any changed settings  click Reset  The settings will be restored to the default settings     505 0100 14 002    Using Hi Q    Feature    The Hi Q feature allows selection from multiple Revolution imaging options to provide an  option that best allows the user to differentiate between blood speckle and tissue along  lumen border  It must be enabled in the Adjust Image screen before recording images     BETTOL TINS  storsice  stories e ereo oro V    06 04 2012 11 31 04       Time Gain Compensation    Radius   mm        E e      men nd  ET   a l    mme  sl  _ E      E TI  N FOON           Display Mode  LIVE    ChromaFio  OFF Ring Down  OFF Hit OFF Catheter  Revolution    Figure 5  Hi Q Settings    Hi Q has three Revolution imaging mode settings   e Off  Imaging equivalent to systems with software versions prior to 3 3 X  e Level 1  Imaging with a darker and fi
10. ecsesssnsesse 5  PATENTS sees dee Ee 5  TRADEMARKS  eegener eene eege eege EE Eet dees vena EEE E E soeiseas 5  WARNINGS AND PRECAUTIONS           cscsssssssssssssesscssssesssssnessesssnssessecsscssesssscsessssssecsesssnseesee 6  FOR USE ONLY BY TRAINED MEDICAL PERSONNEL A 6  READ AND REVIEW MANUAL BEFORE OPER ATION  A 6  CONTACT INFORMATION           csccssssssssssssssscsssssssscssessecsscsssssscssesssssnecsesssnsscssocsscsssssnecsesssssoeees 7  ENTENTE 9  DEVICE  DESCRIPTION  ss Eet totaa ahea A ta aaae aa test aA a ARa ted eg EE ae Eas eet EEeEE EE 9  INTENDED USE EE 10  INSTALLATION TE 10  CATHETERS E 10  CHAPTER 2  PREPARING FOR A CASE  eeseseesosseseecesoeseseeeosoeseeoceossesecocsosseseeeesoeseeoceoseeseesesoe 11  CONNECTING SPINVISION  PIMR  TO THE SYSTEM    11  PREPARING SPINVISION  PIMR  FOR USB teste resis reegen SOARES 11  CHAPTER 3  ELE LE 13  SETTING PULCEBACK SPEED psccsssiaieesioveseas ating an a a a A EES 14  USING TIME GAIN COMPENSATION  TGC  scisssicdsccisasasssaseccdnsansondnipancduavsonrassuesuadavsoevensves 14  USING HOM FEATURE nire r AON A AEA cat alee a catenin ca 15  PERFORMING AN AUTOMATIC PULLBACK     sesssesssssssssssssseserstssrsreresrssententsesseseenesersenee 16  PERFORMING A MANUAL PULLBACK 17  POST IMAGING PROCEDURE si ccccctcssccsssssesseistcsessasssnssbegdessosevsdensebevastoseossesbeavapentesebiees 17  CHAPTER 4  TROUBLESHOOTING            cscsssssssssssssscessesessecsscssessecssesssssnecsesssseseesesssrssessees 19  POTENTIAL IMAGING ARTI 
11. eters  are shipped sterile for single use  In the event of damage  contact Volcano Corporation  personnel        CAUTION  The top surface of SpinVision  PIMr  Models 808884001 and 808884001 R  can become hot to the touch during prolonged periods of use     Contact Information    United States   Volcano Corporation  2870 Kilgore Road  Rancho Cordova  CA 95670 USA  Phone  916 638 8008  Phone  800 228 4728  USA and Canada   Fax  916 638 8112  www volcanocorp com    Authorized European Representative   Volcano Corporation Europe BVBA SPRL  Excelsiorlaan 41  B 1930 Zaventem  Belgium  Phone   32 2 679 1076  Fax   32 2 679 1079    505 0100 14 002 Page 7 of 22    This page is intentionally blank for pagination purposes     Page 8 of 22 505 0100 14 002    Chapter 1  Overview    This manual is intended to provide information specific to the Volcano Revo option for use on  the Volcano system  Please refer to the Volcano system operator   s manuals for information  regarding the system and its functionality     Device Description    The Volcano Revo option kit is a required accessory to the Volcano system when using the  Revolution   45MHz Rotational Imaging Catheter for intravascular ultrasound imaging  This  option consists of a Patient Interface Module  PIM  called SpinVision   PIMr  which can be   simply connected to the existing Volcano system PIM cable        Figure 1  Volcano Revo Option Patient Interface Module  SpinVision PIMr     The ability to perform an automatic or manual pu
12. he PIMr and system        Packaging    Catheters are shipped sterile for single use  In the event of damage  contact Volcano  Corporation personnel        CAUTION  Do not use a damaged catheter  Do not re sterilize the catheter        Page 22 of 22 505 0100 14 002    
13. ications       NOTE  Refer to Volcano system operator   s manuals for system specifications        SpinVision  PIMr  Dimensions and Weights                            a Height Width Depth Weight  3 9 in 4 0 in 14 4 in  3 2 Ib  3 9 in 4 0 in 14 4 in 2 9 Ib       Catheter Specifications  Imaging catheter   Revolution 45MHz Rotational IVUS Imaging Catheter       Transducer type Piezoelectric       Frequency Revolution 45MHz Rotational IVUS Imaging Catheter 360   view   single element rotating transducer       Lumen characteristics   coaxial  guide wire lumen                                  Catheter Crossing Profile at Maximum guide  Transducer wire  Revolution 3 2F  1 07 mm  0 014     0 36 mm   Catheter Minimum ID guiding Length  cm  catheter clearance  Revolution 6F  1 63 mm  2  064    135                   NOTE  Refer to    Instructions for Use     IFU  that accompanies each catheter or go to  www volcanocorp com to view detail catheter use  specifications  and acoustic output        505 0100 14 002 Page 21 of 22    Measurement Accuracy    Measurements obtained with the Revolution 45MHz Rotational IVUS Imaging Catheter are  subject to the following measurement inaccuracies due to variations in tissue speed of sound  and display limitations  The measurement precision is limited in both relative and absolute    ranges   Distance Measurement  4 5    7 0  of measured value    0 10mm   Area Measurement   9    14  of measured value    0 10mm7        These uncertainties apply to the
14. llback is integrated into the SpinVision   PIMr  pullback device  The controls available on SpinVision  PIMr  include the following  buttons    e Reset  Resets the LED distance display to zero    e Image  Press once to begin imaging  Press a second time to stop imaging  The  Revolution catheter will rotate while in imaging mode     e Pullback  Press to begin an automatic pullback  Press again to stop the pullback   e Auto  Offers two automatic pullback rates  1 0 and 0 5 mm sec   e Manual  Offers manual control of pullbacks    505 0100 14 002 Page 9 of 22    Intended Use    The Volcano Revo option is used for qualitative and quantitative evaluation of vascular  morphology in the coronary arteries and vessels of the peripheral vasculature  It is used as  an adjunct to conventional angiographic procedures to provide an image of the vessel lumen  and wall structures  The pullback feature of the PIMr withdraws the Revolution catheter   s  imaging core within the protective sheath for a maximum of 15cm     Installation    To ensure proper operation and warranty coverage  the Volcano system must be installed  and tested by following the Device Installation and Operational Qualification that  accompanies each device     Catheters    Revolution 45MHz Rotational Imaging Catheter 89000       Note  The Revolution   catheter is intended for the intravascular ultrasound  examination of coronary arteries  Intravascular ultrasound imaging is indicated in  patients who are candidates for tra
15. m screen     505 0100 14 002 Page 11 of 22       WARNING  Always ensure that SpinVision  PIMr  is placed in a stable location and is  clear from items that may interfere or prevent smooth movement during use  The  SpinVision  PIMr  weight will transfer as the device performs a pullback           CAUTION  The IVUS PIM  SpinVision  PIMr  and the FFR Pimmette are strongly  attracted to the magnetic field of Stereotaxis units  The user should observe caution  in handling these devices when less than 110cm from the center of these fields in  order to prevent damage to the device or personal injury should these devices become  free inside these very strong magnetic fields     For reliable operation  SpinVision  PIMr  must be further than 110cm from the center of  the magnetic field of a Stereotaxis unit  Inside of this limit the image may become  frozen on the monitor until the unit is moved to outside this limit  Placement of the  SpinVision  PIMr  inside this limit may damage the SpinVision  PIMr         Page 12 of 22 505 0100 14 002    Chapter 3  Imaging    Once the Revolution 45MHz catheter and SpinVision  PIMr  device are connected  the  system will display the following on the Home screen         i VOLCANO  TNE EES     IMAGING OFF 09 29 2011 13 55 48    SS    Save Frame    Record  Pullback     Adjust Image        Patient   Home    VA   EndCase   Archive   Retrieve    Press Image to Begin Imaging            Display Mode  DISABLED    Chramaflo  OFF Ring Down  OFF WH  OFF  Cathe
16. ner blood speckle appearance than in Off  mode  default setting   e Level 2  Imaging with a darker and finer blood speckle appearance than in Level 1  Mode    To enable Hi Q  select Level 1 or Level 2 from the Hi Q drop down menu        NOTE 1  The selection of Hi Q settings should occur before recording video loops           NOTE 2  In Level 1 or Level 2 mode  diameter choices are limited to 8mm and 10mm  The  Off mode allows the user to use the full range of catheter Diameters  8mm  10mm  12mm   and 14mm  Diameter controls the depth of field view in which ultrasound data is displayed   Selecting Level 1 and Level 2 using unsupported diameters results in the message displays  shown below        For example  if Level 1 or Level 2 is selected and then Diameter is set to12mm or 14mm  the  following message displays     505 0100 14 002 Page 15 of 22       Hi Q does not support the 12mm or 14mm diameters  Select OK  to change the diameter and set Hi Q Off        Cancel    Figure 6  Disable Hi Q Message             Conversely  if Diameter is set to 12mm or 14mm and then Level 1 or Level 2 is selected  the    followi    ng message displays        Hi Q does not support the 12mm or 14mm diameters  Select OK  to set the Diameter to 10mm and enable Hi Q     Figure 7  Enable Hi Q Message             Disable Hi Q by selecting Off from the Hi Q drop down menu     When    the Hi Q feature is set to Off  a visual indicator  Hi Q Off  appears under the L VE text     located on the upper left
17. ng SpinVision  PIMr  for Use    1   2   3     Remove the sterile equipment drape from the Revolution box    While maintaining sterility  remove the sterile drape from its protective packaging    This step requires two people  The sterile person opens the sterile drape on the side  with the green arrows  placing his or her hands in the back of the fold s  to keep it  open  The second person places SpinVision  PIMr  into the sterile drape as far as it will  go in the direction of the green arrow on the drape    Unfold and pull the drape all the way out  This will cover SpinVision   s  PIMr  cable far  enough to extend out of the sterile field    Once the catheter has been prepped per the Instructions for Use  IFU   insert the  sterile drape catheter port into the SpinVision  PIMr  catheter port    Remove the sterile barrier sticker from the drape so you can insert the catheter    Insert the catheter into the SpinVision  PIMr  port with the black marks on the catheter  and SpinVision  PIMr  port aligned  then lock it in place by rotating it 30   clockwise  A  message on the Volcano system screen reads    Catheter detected     The system is  ready for imaging once the    Press Image to Begin Imaging    message displays    With the catheter telescope in the most distal position  clip the telescoping hub into the  SpinVision   s  PIMr  catheter arm    Press the Image button on SpinVision  PIMr  or the Image On  Live  button on the  system Home screen  The image appears on the syste
18. nsluminal interventional procedures        Page 10 of 22 505 0100 14 002    Chapter 2  Preparing for a Case    Refer to the Volcano system operator   s manuals for the general instructions on preparing for  an IVUS case     Connecting SpinVision  PIMr  to the System    SpinVision  PIMr  excites the catheter   s transducer element to transmit ultrasonic energy to  the surrounding tissue  It then amplifies and processes the resulting echo signals from the  transducer and sends them to the system through the connection on the system   s rear panel     1     Connect the PIM cable to the rear of SpinVision  PIMr   Orient the marker on the cable  connector to align with the marker on the PIM connector  then slide the connector in  If  you need to remove the cable from SpinVision  PIMr   rotate the collar on the  SpinVision  PIMr  cable connector counterclockwise a quarter turn and then pull back   SpinVision  PIMr  performs a diagnostic self test and displays any errors on the system  screen    The SpinVision  PIMr  connection will be confirmed in the lower right hand section of  the screen  A green light with the words    Rotational    will display once SpinVision   PIMr  is connected        WARNING  SpinVision   s  PIMr  cable may be damaged if equipment is rolled on it  Do  not pull on the cable  place the cable in a high traffic area  or use excessive force as  the strain relief ends may be damaged  Do not use the system if the outer case or  wires appear damaged        Prepari
19. nty become necessary and if the System or components of the  System are to be returned  The System or components will not be accepted for warranty purposes  unless the return has been authorized by Volcano Corporation     System parts or components repaired or replaced under warranty bear the same warranty expiration  date as the original equipment  Consumable parts  data disks  batteries  among others  are  warranted only against defects in materials and workmanship  System parts purchased outside the  original warranty period are warranted for a period of 90 days  subject to all of the restrictions  contained in this Limited Warranty  Use of unauthorized replacement parts may void the warranty  In  all cases  Volcano Corporation will be the sole judge as to what constitutes warrantable damage     Page 4 of 22 505 0100 14 002    Patents and Trademarks  Patents    This product  and the use thereof  may be covered by one or more of the following U S  and  international patents     7074188 8104479 JP 2005537052  7215802 8105237 JP 2007105450  7359554 EP 1534137 JP 2007518488  7463759 EP 1534139 JP 2007526083  7627156 EP 1536727 JP 2008526387  7789834 EP 1703842 JP 2008543511  7874990 EP 1732461 JP 2009530068  7899224 EP 1903944 JP 2011036680  7927275 EP 1933711 JP 2011245326  7930014 EP 1996080 JP 2011521730  7940969 EP2291121 JP4700622   7978916 JP 2005536265 JP4773092   7988633 JP 2005536289 JP4481824    Other U S  and international patents are pending     Trademarks    The f
20. ollowing are trademarks of Volcano Corporation     Volcano and the Volcano logo are trademarks of Volcano Corporation and are registered in  the United States and other countries     Revolution is a registered trademark of Volcano Corporation   Hi Q    is a trademark of Volcano Corporation     s5    is a trademark of Volcano Corporation  S5i   is a registered trademark of Volcano  Corporation     SpinVision is registered in the United States and is a trademark in the European Community  and Japan     In Line Digital is a trademark of Volcano Corporation and is registered in the European  Community     505 0100 14 002 Page 5 of 22    Warnings and Precautions       WARNINGS are used to indicate the possibility of severe personal injury  Follow the  instruction or procedure correctly to avoid injury to the operator  patient  or other  personnel     The warnings are identified by the exclamation symbol  Ab    CAUTIONS are used to indicate the possibility of damage to the equipment  Follow the  instruction or procedure correctly to avoid damage to the equipment        For use only by trained medical personnel       CAUTION  FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER  OF A PHYSICIAN  OR PROPERLY LICENSED PRACTITIONER        Read and review manual before operation    Carefully read and review the entire Volcano system operator s manual  Refer to the section     Warnings and Precautions        VOLCANO Corporation makes no warranty  representation or condition of any
21. s imaging        Post Imaging Procedure    1  Remove the Revolution catheter using standard interventional procedures     2  Detach the catheter from SpinVision  PIMr  by rotating the port counterclockwise 30   to  unlock it and removing the telescoping hub from the SpinVision  PIMr  arm     3  Dispose of the Revolution catheter     4  Remove SpinVision  PIMr  from the sterile drape ensuring that no blood or fluid comes  into contact with it     5  Dispose of the sterile drape     NOTE  After the SpinVision  PIMr  is removed  the screen area will continue to display the  appropriate information for the last type of PIM used     505 0100 14 002 Page 17 of 22    This page is intentionally blank for pagination purposes     Page 18 of 22 505 0100 14 002    Chapter 4  Troubleshooting    These error messages are displayed on the right hand side of the SpinVision  PIMr  message  display  Use the following table to read the description and recommended action     Meaning    SpinVision  PIMr  passed self test  Unable to read RFID    TGC memory failure    Analog high voltage failure    One of the pulse width control pins is  stuck low    Catheter motor failure    Pullback motor driver failed    EEPROM data error       Potential Imaging Artifacts    Suggested User Manual Text    This code displays briefly after the unit is powered  on and indicates that the unit is ready     The catheter may be faulty or may not be fully  rotated into place  Ensure the catheter is fully  rotated into pla
22. ter  Revolution    Figure 2  Main System Screen     Image Off    Select Image On  Live  button on the system screen or the IMAGE button on SpinVision   PIMr  to begin imaging  The SpinVision  PIMr  motor will begin spinning  and the Revolution  image will appear on the system screen     VOLCANO  i ME   Geen frames  ee A  mm    LIVE 0410212012 08 12 46          Pullback Rate      0 5 mm sec  amp  1 0 mm sec    Manual      Adjust Image        Patient     End Case a Archive F Retrieve    Press    Save Frame or Record          Display Mode  LIVE Ring Down  OFF WH  OFF Catheter  Revolution Image Export  ON    Figure 3 Image On    505 0100 14 002 Page 13 of 22    Setting Pullback Speed    The current pullback speed is indicated in the bottom of the ILD section  You can change the  pullback speed by selecting the radio button at the bottom of the ILD section on the system or  by selecting the SpinVision  PIMr  button for the desired speed  0 5mm sec  1 0mm sec or  Manual      Using Time Gain Compensation  TGC     After selecting the Adjust Image button in the Home screen shown above  the Time Gain  Compensation  TGC  feature shown below increases or decreases the gain of the input  signal with respect to distance from the transducer  This feature is only available with the  Revolution catheter  Six TGC profiles are available  Each can be adjusted to a user   s need         VOLCANO ESE cc tone como Y    06 04 2012 11 33 55       Time Gain Compensation    Radius   mm        Display Mo
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Coal Handling & Storage Exhibitor Kit    Mode d`emploi pour la pré-inscription par les étudiants aux  Agression. L`imam d`Empalot tabassé devant la  Accueil Délégation Paris Michel-Ange Budget, finances et  Belinea-1905S1    À nous les nombres  Dirty Shirley Manual - Friedman Amplification  Mesko MS 2236 hair dryer    Copyright © All rights reserved. 