Contents

1. 29 Spare Parts PrepStar SD 2 30 Common Spare Parts 30 Spare Parts Kit 30 Figures Figure 1 PrepStar Solvent Delivery Module 1 Figure 2 Top of pump with oil input 6 Figure 3 Rear Panel Solvent Delivery Module 7 Figure 4 Pumpheads 8 Figure 5 The Menu Display 9 Figure 6 Design of the Solvent Delivery Module 12 Figure 7 Schematic of Pump Head 21 Figure 8 Inlet Check Valve 24 Figure 9 Outlet Check Valve 24 PrepStar SD 2 1
2. 4 Preparing the Pump for Operation 5 Loading the Pump Oil 5 Electrical Connections 7 Fuse Replacement 7 Fluidics 7 Preliminary Test 9 Final Testing 10 System Description 11 Overview 11 The Solvent Delivery System 11 Materials Used in the Solvent Delivery Module 13 Characteristics of Titanium and Stainless Steel 13 Solvent Considerations 13
3. Be sure to support the pump head while unscrewing the bolts 4 Carefully remove the low pressure segment of the pump head from the system by sliding it off the plunger NOTE Pull the pump head directly toward you If significant pressure is placed on the plunger while the pump head is being removed the plunger could be broken 5 Remove and replace the seals from inside the pump head both seals should be replaced NOTE When you are removing the seals take care not to scratch other components of the pump 6 Use a spare plunger to ensure proper alignment of the components after they are put into the head NOTE When installing the seals ensure that the springs are facing the pump head Take care to ensure that the seals are not distorted bent or dented 7 Place the plunger into the pump head and press on the cap until it is flush with pump head 8 Remove the spare plunger and carefully slide the refurbished pump head on to the plunger on the pump MAINTENANCE AND TROUBLESHOOTING PrepStar SD 2 23 NOTE Inspect the plunger before reinstalling the pump head to ensure that it is not damaged 9 Replace the two screws 10 Replace all tubing 11 Pump acetone through the pump for 10 15 minutes at a rate of 100 mL min Acetone serves to swell the pump seals and provide a tighter fit The acetone can be recycled for this operation Cleaning and Changing the Check Valves Each pump head is equipped with
4. Operation 15 Overview 15 The Keypad and Display 15 The Display 15 The Keypad 17 Generating a Solvent Flow 18 Maintenance and Troubleshooting 19 Maintenance Overview 19 Solvent Considerations 19 Cleaning the System and Shutdown Procedures 20 ii 03 914903 00 1 Cleaning the System 20 Shutdown Procedures 20 Routine Maintenance
5. Introduction Overview The PrepStar SD 2 Solvent Delivery Module is designed to meet the needs of scientists and engineers who require the ability to deliver solvents at high flow rates up to 800 mL min at elevated pressures The module is depicted in Figure 1 Figure 1 PrepStar Solvent Delivery Module 2 03 914903 00 1 Manual Organization The user manual includes the following sections Introduction page 1 describes the overall function of the solvent delivery system and provides an outline of the manual Installation page 3 describes the required facilities and the installation procedure System Description Page 11 describes the characteristics of the various components which are in contact with the solvent Operation page 15 describes the application program that is used to set up and operate the system Maintenance and Troubleshooting page 19 describes a series of preventive maintenance operations and discusses how you change components that need to be replaced on a periodic basis It also discusses a variety of problems which you might encounter during operation of the instrument It includes a description of activities which should be performed to determine the cause of problems and explains how to repair the unit if necessary Appendix page 29 contains the specifications and a list of spare parts PrepStar SD 2 3 Installation Overview The PrepStar SD 2 So
6. 21 The Pump Head 21 Replacing the Pump Seals 22 Cleaning and Changing the Check Valves 23 Replacing the Pump Oil 25 Troubleshooting Overview 25 Troubleshooting Procedures 26 Appendix 29 Specifications 29 General Specifications 29 Solvent Delivery Module Pump 29 Complete Spare Parts Kit 29 Extended 1 year Limited Warranty 29 Installation Requirements 29 Physical Specifications
7. Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek CA 94598 1675 USA Varian Inc 2001 03 914903 00 Rev 1 PrepStar SD 2 Solvent Delivery Module Operation Manual PrepStar SD 2 i Table of Contents Introduction 1 Overview 1 Manual Organization 2 Installation 3 Overview 3 Site Requirements 3 Physical Location of the System 3 Space Weight Requirements 3 Environmental Requirements 4 Power Requirements 4 Unpacking and Inspection
8. for an abnormal response from the system Troubleshooting can be simplified and the amount of downtime can be reduced by a consideration of the following guidelines In almost all situations there is one proximate cause for the problem A fundamental knowledge of the role of each component in the system is extremely useful in diagnosis of the problem The availability of critical spare parts is extremely useful If it is suspected that a part is faulty replacing that part may rapidly solve the problem If any aspect of the operating conditions are changed run a before operation and an after operation to ensure that the effect of the change is well understood Do not consider any change as trivial 26 03 914903 00 1 If you change the supplier of organic solvents buffers salts etc ensure that the new and old reagents are equivalent by running the same operation with each In normal operation the Solvent Delivery Module is quiet If a change in the volume or the tone from the system is observed determine the cause of the noise Although this is an unscientific approach a change in the volume and or tone is frequently a good indication that something has changed A service department which is designed to assist in the diagnosing and repair of any problems that may occur is available If you need assistance please contact your nearest sales or service representative Additional trou
9. rate the final flow rate and the ramp time As an example if you select a ramp of 200 minutes and a starting flow rate of 200 mL min and a final flow rate of 100 mL min the flow will be 150 mL min after 100 minutes At the end of the ramp period the flow rate will remain at the final flow value setting until the STOP button is pressed RAMP is used to indicate the desired time for the change in the flow indicated above in min When the RUN button is pressed the flow rate will increase from the start flow rate to the final flow rate in a linear manner at the indicated rate until the flow rate is equal to the upper flow rate indicated above At that point the mobile phase will continue to be delivered at the flow rate indicated in the second FLO field until the STOP button is pressed TIME T indicates the period of time that has elapsed in the current operation i e since the RUN button or the RESET button was pressed This is not editable by the user PL indicates the user selected pressure limit for the system The maximum values are indicated in Table 1 The right field is the actual pressure during a run Table 1 Maximum Pressure Psi 2500 Bar 170 OPERATION PrepStar SD 2 17 The Keypad The keypad includes the following keys FIELD Moves the cursor to the next position Increases the value of the active character This is a two speed control if you click for a short pe
10. an inlet and an outlet check valve Both valves should be cleaned changed at the same time The check valve has a ball and seat design and includes a ruby ball with sapphire seat in a titanium housing The best way to determine if the check valves have deteriorated is to monitor the flow rate at a set of specific conditions For systems that are dedicated to a specific application the flow rate should be checked on a routine basis e g on a monthly basis If the system is used under varying conditions a standard set of conditions should be developed for this test When a defective check valve is found i e the observed flow rate is incorrect it should be cleaned or replaced as described below To clean the check valves 1 Remove appropriate plumbing from the check valve connector take care not to put undue stress on the manifolds 2 Remove the check valve connector with a 13 mm or wrench 3 Rinse the interior of the check valve assembly using an appropriate solvent an organic solvent such as methanol or acetonitrile will remove most materials 4 When the rinse solvent is clear replace the check valve assembly 24 03 914903 00 1 To replace the check valves 1 Remove appropriate plumbing from the check valve connector take care not to put undue stress on the manifolds 2 Remove the check valve connector with a 25 mm or 1 wrench 3 Carefully remove the check valve housing The entire housing sh
11. bleshooting information is contained in manuals provided with components such as the variable wavelength detector the recorder etc Troubleshooting Procedures A Component does not Operate If a component does not start the initial response should be to ensure that the component is connected to the power line and the fuse is good It may be worthwhile to turn the component off and on again Problem Probable Cause Solution Pump will not start Pump is not plugged in Plug pump in Circuit breaker off Turn circuit breaker on Fuse is blown in pump Check replace fuse Pump will not start Drive not reset on power up Reset drive by turning power off wait 30 seconds and turn power on MAINTENANCE AND TROUBLESHOOTING PrepStar SD 2 27 Fluidics Problems Problem Probable Cause Solution Abnormally High Pressure Tubing clogged Isolate clogged line Flush out at high flow rate Low flow rate No Pressure Pumps not primed Disconnect application and run at high flow rate Loose inlet lines Loose fittings between pump and system Tighten fittings on inlet pump lines Check valves Clean or replace valves Page 23 Low flow rate Under Pressure High pressure seal deteriorated Replace seal Page 22 Check valves not operating efficiently Clean or replace valves Page 23 Liquid leaking from back of pump heads Back seal deteriorated Replace Seal Page 22 Significant pressure variance
12. e used Cleaning the System and Shutdown Procedures Cleaning the System On a periodic basis flush the system with a volatile organic solvent such as methanol using a rapid flow rate The frequency of flushing is dependent on the use of the system and the nature of the solvent If the system is used on a daily basis the system should be flushed on a weekly basis If the instrument is used on a less than daily basis it should be flushed after each use Shutdown Procedures Short Term Shutdown Procedures If the system will not be used for a short period of time e g a few hours the solvent can be recycled to maintain a flow through the system A lower flow rate can be used to minimize the wear and tear on the pumps The upper pressure limit should be set approximately 500 psi 35 bar above the pressure required for the flow rate to protect the system in case of a blockage If the solvent is recycled for a significant period of time do not use this solvent for normal operation because some of the more volatile components may evaporate over time For short term shutdowns it is recommended that the power switch on the front panel is left on Long Term Shutdown Procedures The conditions below are to be used when the Solvent Delivery Module will not be used for a lengthy period or if you cannot maintain the flow of solvent during the shutdown period MAINTENANCE AND TROUBLESHOOTING PrepStar SD 2 21 To prepare the sys
13. eel PCG362720400V 4 Fuse Kit 220V PCG50220STKIT 1 Ferrule 1 8 PEEK PCG3690P0200 4 Spare Parts Kit The PrepStar SD 2 Spare Parts Kit 02 KIT ReS800 includes the above items Please call Varian for price and delivery information
14. lean it up and determine the cause of the leak before continuing PrepStar SD 2 15 Operation Overview The PrepStar SD 2 Solvent Delivery Module is controlled via the controller on the front panel In this section we describe the general mode of operation of the system via the controller on the Solvent Delivery Module and provide a definition of each control The Keypad and Display The keypad and the display provide control of the instrument when it is in manual mode All user entry is performed via the keypad as described below The Display The display has two lines which present the system status and allows for entry of the desired operating information as follows FLO is used to indicate the desired flow rate s The maximum value is 800 0 mL min The flow rate that will be delivered when the RUN button is pressed is dependent on the RAMP entry If a constant flow rate is desired enter the flow in the first flow field and a RAMP of zero When the RUN button is pressed the pump will deliver the indicated flow rate until the STOP button is pressed If the flow rate should change during the run enter the initial flow rate in the first flow field FL1 and the final 16 03 914903 00 1 flow rate in the second flow field FL2 When the RUN button is pressed the pump will deliver the indicated flow rate until the STOP button is pressed The flow rate at a given instant is determined by the starting flow
15. lvent Delivery Module Pump seals should be replaced on a periodic basis typically after 6 months to a year of operation To replace pump seals refer to Replacing the Pump Seals page 22 Each check valve is fitted with 1 sapphire seat and 1 ruby ball These materials have been chosen for their hardness and ensure reliable use of the pump for an extended period of time with a minimum of maintenance The wear and tear on the seats and ball depend to a great extent on the amount of particulate matter in the solvent thus the solvent should be well filtered If the mobile phase is well filtered e g no particulate matter the balls and seats should last for an extended period of time For information about changing seats and balls refer to Cleaning and Changing the Check Valves page 23 SYSTEM DESCRIPTION PrepStar SD 2 13 Materials Used in the Solvent Delivery Module The following materials are used in the Solvent Delivery Module titanium zirconia sapphire ruby and other ceramic materials PEEK and Teflon Characteristics of Titanium and Stainless Steel Solvent Delivery Modules include fittings tubing and other components fabricated from titanium This material is inert to most solvents and mobile phases Type 316 Stainless Steel is inert to most solvents and mobile phases however the following materials should be avoided Salts containing halogens halogenated acids at any pH acidic solutions wi
16. lvent Delivery Module is designed to be installed by the user This Section describes the site requirements preparing the system for use Site Requirements Physical Location of the System The system should be used in a facility free from extremes of temperature humidity and vibrations The distance between the system and the walls of the facility must be at least 8 20 cm This will provide suitable ventilation and will allow ready access in the event that any adjustments are needed If an island type laboratory table is available it should be used as it provides ready access to all parts of the system Space Weight Requirements 29 2 cm W x 19 7cm H x 57 5 cm D 11 5 x 7 75 x 20 25 25 kg 60 lb WARNING Avoid back strain and other injuries by following all safety precautions when lifting heavy objects 4 03 914903 00 1 Environmental Requirements Temperature Range 59 to 86 F 15 30 C Humidity Range 20 to 80 Relative Humidity Allowable temperature change 5 F hr 2 8 C hr Do not install near a window a ventilation duct or any other device which can cause a significant change in the temperature Use a ventilation exhaust line to remove solvent vapors from the immediate vicinity of the instrument Power Requirements 230 Vac 50 60 Hz 1000 VA A dedicated power line should be used for the system This line should be Electromagnetic Interference free and m
17. ould be removed 4 If necessary force the check valve out of its housing by using a rod inserted from the tube end 5 Take the check valve apart The sapphire seats and ruby ball should be discarded 6 Install new seats and ruby ball and reassemble the check valve When reassembling the check valve ensure that the ruby ball is seated against the shiny part of the seat NOTE The inlet check valve and outlet check valve are put together in a different order Refer to Figure 8 inlet check valve and Figure 9 outlet check valve for assistance A B E F C D Figure 8 Inlet Check Valve B E F A D C Figure 9 Outlet Check Valve Legend A Housing D Ruby Ball B PEEK Seal E Cage C Sapphire Seat F PEEK Spacer MAINTENANCE AND TROUBLESHOOTING PrepStar SD 2 25 1 Place the check valve into its housing and then into the pump head Tighten the connector The check valves should be snug finger tight plus turn with a wrench DO NOT OVERTIGHTEN the fittings as this may cause damage to the internal components 2 Replace the inlet tubing 3 Prime the pump It may be necessary to force solvent into the pump head using a squeeze bottle 4 Replace outlet tubing Replacing the Pump Oil The pump oil should be replaced every year or after 2000 hours of operation whichever comes first Meropa 150 Texaco oil should be used Troubleshooting Overview Troubleshooting is defined as the determination of the reason
18. over a short time period spikes Degas solvent if a persistent problem 28 03 914903 00 1 PrepStar SD 2 29 Appendix Specifications General Specifications Compact modular benchtop design Maximum operating pressure 800 2500 psi 165 bar Solvent Delivery Module Pump Includes all general specifications listed above Duplex zirconia plunger pump Titanium pump heads Flow Rate SD 2 80 to 800 mL min upper flow rate limit is dependent on solvent viscosity Pressure monitor displayed on control panel Complete Spare Parts Kit Extended 1 year Limited Warranty Installation Requirements Power 230 Vac 50 60 Hz 1000 VA Physical Specifications Solvent Delivery Module 11 5 W x 7 75 H x 20 25 D 29 2 cm x 19 7 cm x 57 5 cm Weight approximately 60 lb 27 kg 30 03 914903 00 1 Spare Parts PrepStar SD 2 Common Spare Parts Item Part Number Quantity in spare parts kit PE Zinc Chromate White Seals PCG601660010 16 Viton O Rings PCG60130088100 16 Check Valve Cartridge PCG549A00005T 6 Nut 1 8 Stainless Steel PCG369220200 8 Ferrule 1 8 LITE TOUCH PCG369000200LT 8 Luer Lock Adapter PCG369520100 5 Zirconia Plunger 10 mm PCG6030F3026771 1 Ferrule 1 4 Titanium PCG369000400Ti 8 Nut 1 4 Stainless Steel PCG369220400 8 Tee 1 8 Titanium PCG3632T0400V 1 Inverted Nut 1 4 Stainless St
19. pump heads Preliminary Test At this juncture it is suggested that a preliminary test is performed to ensure that the pump s deliver solvent and the fittings do not leak To perform the Preliminary Test 1 Power up the pump s After a few seconds the display will present the Menu display Figure 5 FL1 XXXmL PLXXXX XXXXpsi bar FL2 XXXmL RAMP XXX TXXXmn Figure 5 The Menu Display NOTE A detailed discussion of the operating program is presented in the Operation section beginning on Page 15 The instructions listed below are provided as a step by step protocol simply to test the system 2 Click the Field button until the cursor is in the first character in the FLO field Click on the key until the first character in the FLO field is 1 e g 100 mL min For two pump systems make certain that the flow is set for both pumps 3 Click on the Field button until the cursor is on the first character of the Press field If the system indicates the pressure limit in psi click on the key until the first character in the PRESS field is 2 e g 2000 psi mL min If the system indicates the pressure limit in bar move the cursor to the second character then click on the key until the first character in the PRESS field is 1 e g 100 bar 10 03 914903 00 1 4 Press the RUN button The pump will deliver solvent at the indicated flow rate Allow the system to deliver solvent for approximately five minutes and check fo
20. r leaks 5 Press the STOP button and continue the installation Final Testing The final testing is done with the pressure load e g HPLC column in place and is a high pressure test To perform the final test 1 Raise the flow rate to 150 mL min for 200 mL min systems or 600 mL min for 800 mL min systems 2 Press RUN The system should deliver solvent at the indicated flow rate If the system stops due to a pressure overload raise the pressure limit by 500 psi 33 bar NOTE Make certain that the pressure limit is not set above the maximum pressure indicated by manufacturer of any device that will be attached to the solvent delivery module If the pressure overload is reached again do not raise the pressure again Instead the cause of the blockage should be determined and corrected 3 Allow the pump to deliver solvent at the indicated flow rate for approximately 1 2 hr and check for leaks PrepStar SD 2 11 System Description Overview The PrepStar SD 2 Solvent Delivery Module system provides a convenient method to deliver solvent at a high pressure Two different pumps are available 200 mL min and 800 mL min The Solvent Delivery System The PrepStar SD 2 Solvent Delivery Module System is specifically designed for preparative chromatography The pump includes two pistons to provide ripple free delivery of the mobile phase The pump is shown in Figure 6 12 03 914903 00 1 Figure 6 Design of the So
21. rating and type 5 x 20 mm 250 Vac 5 Amp Type F WARNING FIRE HAZARD To avoid fire hazard replace fuses ONLY with fuse of the same rating and type Always switch off the AC power and uplug the power cord before replacing a fuse Fluidics The inlet port for the Solvent Delivery Module is the manifold that delivers solvent to the bottom of the pump The outlet port Fuse Compartment 8 03 914903 00 1 for the Solvent Delivery Module is the manifold connected to the upper check valves Figure 4 Pumpheads NOTE When connecting fittings to the system tighten all fittings fingertight plus a turn Overtightening may permanently damage a fitting NOTE All inlet lines should be filtered At the rear of each pumphead at the top and the bottom is a washing section see Figure 4 above These connections can be used to flush the rear of the piston seals This is especially useful when buffers are used as part of the mobile phase The washing port can be used two ways Washing sections connections Pump Outlet Pump Inlet INSTALLATION PrepStar SD 2 9 1 A wash bottle or syringe can be used to squirt water or methanol directly through the back of the pumphead This can be done individually through the two heads or by connecting the two heads together as shown in Figure 4 A small pump or a reservoir positioned above the pump could also be used to produce a continuous flow of wash liquid behind the
22. riod the value increases by one character if you hold the button down the value will increase on a continuous basis Decreases the value of the active character This is a two speed control if you click for a short period the value decreases by one character if you hold the button down the value will decrease on a continuous basis RUN Initiates the delivery of the mobile phase at the indicated flow rate If a RAMP has been selected the flow rate will start at the value indicated in the first flow field and increase decrease until the flow rate indicated by the second flow rate field The time for the change in set in the RAMP field If a flow rate ramp has been halted due to STOP command the flow rate ramp will continue when the RUN button is pressed STOP Terminates the delivery of the solvent and the RUN Time clock If a flow rate ramp is active the ramp will be held at the present point and will resume when the RUN button is pressed RESET If a flow ramp has been selected this button is used to set the flow rate that will be delivered when the RUN button is pressed to the initial flow rate This button is only active after the STOP button has been pressed if the button is not pressed the ramp will continue 18 03 914903 00 1 Generating a Solvent Flow To generate a solvent flow 1 Enter the desired flow rate 2 Enter the maximum pressure 3 Enter the ramp if desired 4 Pre
23. ss RUN PrepStar SD 2 19 Maintenance and Troubleshooting Maintenance Overview On a routine basis the following activities are recommended to maintain the system in maximum operating condition always use high purity solvent and ensure that particulate matter does not enter the system keep the system clean and follow a shutdown protocol replace consumable items on a periodic basis When a system is first installed it may be wise to obtain performance data from an application that can be readily obtained on a consistent basis This application can be used as a benchmark to monitor the performance of the system Lists of recommended spare parts are presented in the Appendix Solvent Considerations The use of high purity solvents will maximize performance and minimize problems with the system Water should be purified using a reverse osmosis system and high quality organic solvents should be used It is important to keep particulate matter out of the system The following steps are recommended Solvents should be filtered through a 0 22 m membrane filter before use to remove any particulate matter Filters should be checked to ensure that extractable materials are not present 20 03 914903 00 1 If the solvent includes salts and or buffers the flush ports on the pump can be used to rinse the system The rinse solution should be free of particulate matter e g de ionized water should b
24. t your sales representative immediately NOTE If there is any apparent damage to the unit or any component do not plug the unit into the power line Preparing the Pump for Operation Loading the Pump Oil It is necessary to pour the oil into the gear box of the pump before operation This should be done as follows in Figure 2 1 Turn the main power switch off and unplug power cord WARNING SHOCK HAZARD Dangerous voltages exposed when cover is removed This procedure must be performed by qualified personnel 6 03 914903 00 1 Figure 2 Top of pump with oil input 2 Remove the screws that attach the top cover to the pump 3 Remove the oil fill plug on the gear box 4 Pour the contents of the bottle containing the pump oil into the gear box 600 mL 5 Place the VENTED cap plug into the pump DO NOT REPLACE THE PLUG THAT YOU REMOVED IN STEP 2 6 Replace the top cover on the unit and replace the three screws Oil fill plug INSTALLATION PrepStar SD 2 7 Electrical Connections The power cord is should be inserted in the power socket in the lower left corner of the rear panel Figure 3 Figure 3 Rear Panel Solvent Delivery Module WARNING Do not plug the power cords into the line voltage until all connections have been made to all components in the system All power cords should be connected to a common power line that is properly grounded Fuse Replacement Always use fuses of the correct
25. tem for shutdown 1 Flush with water until all salts and buffers are removed 2 Flush with methanol so that the water is removed Routine Maintenance The Pump Head Under normal operating conditions it will be necessary to replace some components e g the pump seals and check valves The frequency of replacement is dependent on a variety of factors including the nature of the solvent the number of hours day that the system is used the efficacy of removal of particulate matter from the solvent and the pressure that the pump is required to deliver As a rough rule of thumb pump seals should be replaced every 6 12 months and check valves should be replaced after a few years or sooner depending on the usage and the solvent Figure 7 Schematic of Pump Head Legend A Outlet H Pumphead low pressure segment B Pumphead high pressure segment I O Ring C Inlet J Wash Spring D Spring Seal K Spring Seal E O Ring L Rulon Guide F PEEK Back up Ring M Retainer Plate G Rulon Guide N Screws O Plunger A B C D E F G H I J K L M N O 22 03 914903 00 1 Replacing the Pump Seals The pump seals should be replaced when leaks are observed from the pump Both seals should be replaced at the same time To replace the pump seals 1 Remove all inlet and outlet tubing 2 Remove all tubing for the wash ports if present 3 Unscrew the two screws which hold the pump head to the system CAUTION
26. th a pH below 2 Solvent Considerations Hazardous chemicals under pressure Use proper eye and skin protection The composition of the solvent used with the Solvent Delivery Module depends on the application In many cases the solvent contains an organic solvent such as methanol or acetonitrile and a buffer e g 0 05 M KH2PO4 H3PO4 pH 4 0 To maximize the efficiency and reliability of the instrument the following general guidelines for the solvent should be heeded Water should be purified using a reverse osmosis system High quality organic solvents should be used Mobile phases should be filtered through a 0 22 membrane filter before use to remove any particulate 14 03 914903 00 1 matter Filters should be checked to ensure that extractable materials are not present If you use a buffer solution ensure that fresh solutions are used Microbiological growth can occur in a few days this may change the composition of the buffer and or lead to particulate matter in the system NOTE Do not leave water buffers salts or other potentially corrosive materials in the system when it is not in use Failure to flush these material from the liquid path may result in damage to the system and could possibly void the warranty It is very important to be aware of materials which are potentially corrosive to the system NOTE If caked salt is observed on any part of the system e g by a fitting c
27. ust be connected to a true ground The system should not be connected to a power line that is susceptible to sudden changes in power requirements e g a power line that also serves an oven or compressor If there are significant variations in the power line a constant voltage buck booster transformer may be required Unpacking and Inspection Place the PrepStar SD 2 on a flat surface and inspect for damage Damage in transit is the responsibility of the carrier If any damage is found contact the carrier and immediately Many carriers must receive a damage claim within 7 days of delivery and will need to see the shipping container Inspect the shipping container for indications of damage In North America damage occurring during transit is the responsibility of the carrier and should be reported immediately to the carrier and to INSTALLATION PrepStar SD 2 5 Varian Analytical Instruments 2700 Mitchell Drive Walnut Creek California 94598 1675 Attention Customer Service 1 800 FOR HPLC Outside North America please contact the carrier and your local Varian office For contents of shipping container see the Standard Accessory Kit list in the shipping box Carefully unpack the PrepStar SD 2 Solvent Delivery System It is a good idea to save the shipping container and packing material They provide excellent protection for the instrument in case of future transit or storage If any parts are missing please contac

Download Pdf Manuals

image

Related Search

Related Contents

オイルクーラーユニット 取扱説明書  Cortador de cama plana digital Serie G3 Instrucciones de servicio  Guia de Referência para Implementação da NR20  dossier complet  Edital Pregão 036/2015    Omega OMNCPIR notebook cooling pad  Trans-ABySS 1.4.4 User Manual  M175 Sakamoto Suction Simulator  Philips Sports in ear headset SHQ1007  